4000188 Rev. ACopyright 2002 Bio-Rad Laboratories Inc.
Warranty Statement
Bio-Rad Model GS-800 Calibrated Imaging Densitometer
This warranty may vary outside of the United States. Please contact your local Bio-Rad
office for the exact terms of your warranty.
Bio-Rad Laboratories warrants to the customer that the Model GS-800 Calibrated Imaging
Densitometer (Catalog Number 170-7980 or 170-7981) will be free from defects in material
and workmanship, and will meet all performance specifications for the period of one year
from the date of shipment. This warranty covers all parts and labor.
In the event that the instrument must be returned for repair under warranty, the
instrument must be packed for return in the original packaging.
Bio-Rad shall not be liable for any incidental, special or consequential loss, damage, or
expense directly or indirectly arising from the use of the Model GS-800 Calibrated Imaging
Densitometer. Bio-Rad makes no warranty whatsoever in regard to products or parts
furnished by third parties, such being subject to the warranty of their respective
manufacturers. Service under this warranty shall be requested by contacting the
customer’s nearest Bio-Rad office.
This warranty does not extend to any instruments or parts thereof that have been subject
to misuse, neglect or accident, or that have been modified by anyone other than Bio-Rad or
Bio-Rad's authorized agent, or that have been used in violation of Bio-Rad instructions.
For any inquiry or request for repair service, contact Bio-Rad Laboratories. Inform BioRad of the model and serial number of your instrument.
2
3
Table of Contents
Section 1 General Information .......................................................................................................6
The Model GS-800 Calibrated Imaging Densitometer is a high performance, calibrated
imaging densitometer that converts transparent and opaque electrophoretic samples into
digital data. These data are analyzed with Quantity One® software operating in either
Microsoft Windows® or Macintosh® computer systems. For a step-by-step guide to
scanning and analyzing images, please refer to your Quantity One software user manual.
The main features of the Model GS-800 Calibrated Imaging Densitometer include:
• Adjustable resolution from 36.3 microns to 127.0 microns;
• Reflectance and transmittance
• Scanning area of 29 cm by 40 cm
• Sampling rate of up to 700 dots per inch (dpi).
• Stationary platen for higher reliability
• Variable wavelength lamp/filter architecture for improved color discrimination
• Analog to digital conversion at full 12-bit accuracy
• Calibrated transmission tablet to ensure accurate OD readings
• Reflective gray scale tablet to ensure reproducible reflectance values
• Supported by Quantity One software, which provides versatile image display,
optimization, and quantitation features
1.2 Installation Qualification and Operational Qualification ( IQ/OQ )
The calibration of the Model GS-800 Calibrated Imaging Densitometer can be validated by
following the GS-800 Calibrated Densitometer Installation Qualification and Operational
Qualification protocols. These protocols are available as an accessory to the GS-800
Calibrated Imaging Densitometer ( Catalog number 170-7956 ). These straightforward and
simple to follow IQ/OQ protocols include instructions for the installation of the hardware
and software, and instructions for the verification of the reflectance and transmittance
calibration functions.
Following proper installation, a linear dynamic range of 3.0 OD for the
Transmissive scanning and 2.0 OD for Reflectance scanning can be confirmed with
greater than 95% confidence. For additional details, contact your local Bio-Rad
Representative.
6
1.3 Safety
The Model GS-800 Calibrated Imaging Densitometer system uses high voltage and high
current and should be operated with care at all times. To avoid shock, set up the GS-800
Densitometer in a dry area. Immediately wipe up any spilled solutions.
•Turn off the Densitometer, and then disconnect the power plug when you want to
clean the Densitometer case or glass plate, or when the Densitometer needs service or
repair.
•Place the Densitometer on a level surface.
•To ensure proper ventilation, allow a minimum of 15 cm free space around each side of
the Densitometer.
•Do not leave photographs, gels, or film on the glass plate for excessive periods of time.
The heat from the light source in the Densitometer may cause them to deteriorate.
•Do not operate the Densitometer when the environmental temperature falls below 5° C
or rises above 40° C.
•Do not operate the Densitometer when the environmental humidity falls below 25% or
rises above 85%.
1.4 Regulatory Compliance*
Declaration of Compliance
The party responsible for product compliance
Corporate Name:UMAX Technologies, Inc.
Address:3561 Gateway Blvd. Fremont, CA 94538, U.S.A.
Telephone No.:510-651-4000
*See UMAX PowerLook 2100XL Operation Manual
7
Section 2 Setting Up
2.1 Checking the Contents
Your Model GS-800 Calibrated Imaging Densitometer system should arrive complete with
one of each of the following:
•GS-800 Calibrated Imaging Densitometer
•GS-800 Calibrated Imaging Densitometer User Guide
•Quantity One™ Image Analysis Software
•UMAX PowerLook 2100XL Operation Manual
•25/50-pin SCSI Cable
•Power Cord
• Photoperfect and Adobe Photoshop LE
•MagicScan CD and manual
These components may arrive in several boxes. If your system is missing any of these
items, immediately contact your local Bio-Rad office for a prompt replacement. Pleaseretain all packaging material. Additional charges will be assessed if packaging is not
available for instrument warranty service shipping.
Caution: Use care when unpacking your equipment. The equipment is delicate and
damage can occur if it is dropped or otherwise mistreated.
2.2 Unpacking the Contents
Unpacking the Model GS-800 Calibrated Imaging Densitometer shipping container:
Warning: The Model GS-800 Calibrated Imaging Densitometer weighs approximately 45.8
lbs. (20.8 kg). To avoid personal injury, two people may be needed to lift the Densitometer
out of the shipping container. Lifting the unit just by the lid may cause damage to the
instrument
1. Carefully lift the Model GS-800 Calibrated Imaging Densitometer out of the shipping
carton and place it on a stable, dry, flat surface. The chassis of the Model GS-800
Calibrated Imaging Densitometer attempts to conform to any uneven surface. This
may result in a poor-quality scan.
2. Remove the plastic wrapping and the packing materials from the Densitometer.
3. The Model GS-800 Calibrated Imaging Densitometer should be located close to the
computer system to which it will be connected. Select a flat, dry surface that
provides adequate ventilation on all sides to prevent overheating.