No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
E5250-90000
Edition
Edition 1, October 1995
Edition 2, January 1997
Edition 3, August 1997
Edition 4, January 2000
Edition 5, May 2000
Edition 6, January 2001
Edition 7, August 2003
Edition 8, July 2005
Edition 9, May 2008
Agilent Technologies, Inc.
5301 Stevens Creek Blvd
Santa Clara, CA 95051 USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Tec hn ol og y L i c e ns es
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or
contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights as
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name:
Agilent Technologies International sarl
Manufacturer’s Address:
Supplier’s Address:
Rue de la Gare 29
CH - 1110 Morges
Switzerland
Declares under sole responsibility that the product as originally delivered
Product Name:
Supplementary Information:
This DoC applies to above-listed products placed on the EU market after:
September 26, 2006
Date
Toshiyuki Kawaji
QA Manager
Agilent Technologies
For further information, please contact your local Agilent Technologies sales office, agent or distributor,
•Herstellerbescheinigung
GEÄUSCHEMISSION
Lpa < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 T. 19
•Manufacturer’s Declaration
ACOUSTIC NOISE EMISSION
Lpa < 70 dB
operator position
normal operation
per ISO 7779
NOTEThis ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme ?Hla norme NMB-001 du Canada.
This product complies with the WEEE Directive (2002/96/EC) marking
requirements. The affixed label indicates that you must not discard this electrical/
electronic product in domestic household waste.
Product Category: With reference to the equipment types in the WEEE Directive
Annex I, this product is classed as a “Monitoring and Control instrumentation”
product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Agilent office, or see
www.agilent.com/environment/product/ for more information.
Microsoft, Windows, Win dows NT, Visual C++, and Visual Basic ar e registered trademar ks of Microsoft
Corporation.
LabVIEW are registered trademarks of National Instruments Corporation.
Borland C++ Builder is registered trademark of International, Inc.LabWindows and
Safety Summary
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual may impair the
protections provided by the equipment. In addition, it violates safety standards of
design, manufacture, and intended us e of the instrument. Agilent Technologies, Inc.
assumes no liability for customer’s failure to comply with these requirements.
NOTEAgilent E5250A complies with INSTALLATION CATEGORY II for mains input
and INSTALLATION CATEGORY I for measurement input terminals, and
POLLUTION DEGREE 2 defined in IEC 1010-1.
Agilent E5250A is INDOOR USE products.
•GROUND THE INSTRUMENT
This is Safety Class I instrument. To minimize shock hazard, the instrument
chassis and cabinet must be connected to an electrical ground. The power
terminal and the power cable must meet International Electrotechnical
Commission (IEC) safety standards.
•DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
•KEEP AWAY FROM LIVE CIRCUITS
Operation personnel must not remove instrument covers. Component
replacement and internal adjustments must be made by qualified maintenance
personnel. Do not replace components with power cable connected. Under
certain conditions, dangerous voltages may exist even with the power cable
removed. To avoid injuries, always disconnect power and discharge circuits
before touching them.
•DO NOT SERVICE OR ADJUST ALONE
Do not attempt internal service or adjustment unless another person, capable of
rendering first aid and resuscitation, is present.
•DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not install substitute
parts or perform any unauthorized modification to the instrument. Return the
instrument to a Agilent Technologies Sales and Service Office for services and
repair to ensure that safety features are maintained.
•DANGEROUS PROCEDURE WARNINGS
Warnings, such as example below, precede potentially dangerous procedures
throughout this manual. Instructions contained in the warnings must be
followed.
WARNINGDangerous Voltage, capable of causing death, are present in this instrument.
Use extreme caution when handling, testing, and adju sti n g.
Safety Symbols
The general definitions of safety symbols used on equipment or in man uals are
listed below.
Instruction manual symbol: the product will be marked with this symbol when it is
necessary for the user to refer to the instruction manual in order to protect against
damage to the instrument.
Indicates dangerous voltage and potential for electrical shock. Do not touch
terminals that have this symbol when insrument is on.
Affixed to product containing static sensitive devices--use anti-static handling
procedures to prevent electrostatic discharge damage to compone nt.
Protective conductor terminal. For protection against electrical shock in case of a
fault. Used with field wiring terminals to indicate the terminal which must be
connected to ground before operating equipment.
Frame or chassis terminal. A connection to the frame (chassis) of the equipment
which normally includes all exposed metal structures.
Indicates earth (ground) terminal.
Alternating current.
Direct current.
ON (Supply).
OFF (Supply).
STANDBY (Supply).
CAT 1
WARNINGThe warning sign denotes a hazard. It calls attention to a procedure, practice,
CAUTIONThe caution sign denotes a hazard. It calls attention to an operating procedure,
Means INSTALLATION CATEGORY I. Measurement terminals on the rear panel
comply with INSTALLATION CATEGORY I.
condition or the like, which, if not correctly performed or adhered to, could result in
injury or death to personal.
practice, condition or the like, which, i f not correctly perfor med or adhered to , could
result in damage to or destruction of part or all of the product.
In This Manual
This manual is a user’s guide for Agilent E5250A, and consists of the following
chapters:
•Introduction
Provides an overview of the E5250A Low Leakage Switch Mainframe, E5252A
10×12 Matrix Switch, and E5255A 24 (8×3) Channel Multiplexer.
•Installation
Describes requirements to install the E5250A and tasks for installation.
•Executing Self-Test and Leak Test
Describes how to execute the three Self-Test items (Controller Test, Front Panel
Interface Test, and Relay Test) and the Leak Test.
•Setting up Measurement Environment
Explains how to connect your instruments to the E5250A input, and how to
connect the E5250A output to your wafer prober or test fixture.
•Controlling the E5250A
Introduces several methods for controlling the E5250A, gives basic information
for controlling the relay switc hes on the plug-in cards, and describes how to use
Virtual Front Panel (VFP) utility furnished with the E5250A.
•Programming the E5250A
Describes how to create programs that contain SCPI commands to control the
E5250A.
•Command Reference
Describes the SCPI commands available to operate the E5250A via GPIB
interface and the status reporting structure.
•VXIplug&play Driver
Introduces the VXIplug&play driver available for the E5250A.
•Executing Sample Programs
Explains how to execute a nd modify the sample programs stored in t h e pr ogram
disk that is furnished with the E5250A.
•Specifications
Lists specifications, typical data, supplemental data, and reference data for the
E5250A, E5252A, and E5255A.
•Error Messages
Lists and describes the error messages for the E5250A.
•SCPI Command Summary
This is a quick reference for the SCPI subsystem commands available for the
E5250A.
Text Co nventions
The following text conventions are used in this manual:
Screen Text Represents text that appears on screen of the controller.
Italic Refers to a related document, or is used for emphasis.
This chapter gives an overview of Agilent E5250A, E5252A, and E5255A.
•“Agilent E5250A Product Description”
•“Agilent E5252A Product Description”
•“Agilent E5255A Product Description”
•“Options and Accessories”
1-2Agilent E5250A User’s Guide, Edition 9
Agilent E5250A Product Description
Agilent E5250A Low Leakage Switch Mainframe is a computer-controlled
switching matrix mainframe designed for semiconductor dc characteristics
measurement applications. The E5250A has four slots for installing the modules
(plug-in cards) listed in Table 1-1.
These cards can be used in various switching matrix configurations in the E5250A.
For example, using four E5252As, you can configure a maximum 10-input
48-output switching matrix, or using four E5255As, you can configure a maximum
2-input × 96-output multiplexer.
The E5250A has 2 hard keys and 4 LED indicators on the front panel as shown in
Figure 1-1.
LINE key Used to turn the E5250A on or off.
Line LED Turns on when the E5250A is turned on.
Local/Self Test key If "Remote" LED is on, pressing this key releases the
E5250A from remote GPIB control by an external controller.
If "Remote" LED is off, pressing this key executes the Relay
Test. Before pressing this key, make sure the Relay Test
Adapter is connected. For details of Relay Test, refer to
Chapter 3. During Relay Test execution, the LED in this key
is ON.
Remote LED Turns on when the E5250A is in the GPIB remote status.
System Fail LED Turns on if system error occurs in the E5250A. Contact you r
nearest Agilent Technologies Service Center.
Fail LED Tu rns on if the E525 0A fails Self-test (in cluding Relay Test).
If this LED turns on after executing Relay Test, a plug-in
card may be defective. And if this LED turns on after turning
the E5250A on again, the E5250A is defective. If so, contact
your nearest Agilent Technologies Service Center.
Rear Panel Tour
Rear panel of the E5250A is shown in Figure 1-2. The E5250A has 4 slots for
installing plug-in cards, and 10 input connectors for connecting measur ement cables
from instruments.
SMU INPUT SMU INPUT connectors are for semiconductor dc
characteristics measurements.
The E5250A has six SMU INPUT ports , which are tri axial type
connectors. The connectors can also be used for a maximum
three Kelvin connection inputs by coupling two ports as
follows:
•SMU INPUT 1 and 2
•SMU INPUT 3 and 4
•SMU INPUT 5 and 6
1-4Agilent E5250A User’s Guide, Edition 9
Introduction
Agilent E5250A Product Description
AUX INPUT AUX INPUT connector s are for C-V measurements, pulse
input, and so on.
The E5250A has 4 AUX INPUT ports: HF1, HF2, CV1, CV2.
These are BNC type connectors.
The AUX INPUT connectors are used only with the E5252A
and are not used with the E5255A.
CAUTIONThe maximum measurement voltage that can be applied to any input terminal is
±200 Vdc. The maximum measurement current is 1 Adc at ±200 Vdc. The
maximum voltage that can be applied between input terminals is 300 Vdc. Do not
apply an input signal over these limits to the E5250A inputs. If you do, the E5250A
will be damaged.
If you use a bias source that has current limit capability, set the bias source current
limit to less than 1 Adc.
Figure 1-2E5250A Rear Panel
Agilent E5250A User’s Guide, Edition 91-5
Introduction
Agilent E5252A Product Description
Agilent E5252A Product Description
The E5252A 10×12 Matrix Switch is a 10-input to 12-output switching matrix card
for the E5250A. The E5252A is designed for semiconductor dc parametric
measurement applications that need to switch some instruments connected to
Device Under Test (DUT), or need to scan instrument input/output for many DUTs,
automatically.
The E5250A can be installed w ith maximum f our E525 2As. And the in put termin als
of the E5252As are connected together inside the E5250A. So, the E5250A can
configure a 10×24 matrix (two E5252As), 10×36 matrix (three E5252As), and
10×48 matrix (four E5252As). The E5252A block diagram and output connectors
are shown in Figure 1-3 and Figure 1-4.
Figure 1-3E5252A Block Diagram
1-6Agilent E5250A User’s Guide, Edition 9
Introduction
Agilent E5252A Product Description
Input Ports The E5252A has 10 input ports via the E5250A.
Input paths for SMU1 and SMU2 are designed for low current
measurement.
There are actually only six input paths, so two of the input paths
are shared by three ports each as follows:
•SMU5, HF1, and CV1 ports share same input path
•SMU6, HF2, and CV2 ports share same input path
So, for these two input paths, you can only use one input po rt at
a time. For example, you cannot use HF1 and CV1 input ports
at the same time.
Output Ports The E5252A has 12 output ports. Two ports can be coupled for
Kelvin connection out put fo r a maxi mum o f six Kelv in out puts:
•OUTPUT1 and OUTPUT2
•OUTPUT3 and OUTPUT4
•OUTPUT5 and OUTPUT6
•OUTPUT7 and OUTPUT8
•OUTPUT9 and OUTPUT10
•OUTPUT11 and OUTPU T12
The output ports are triaxial connectors.
Figure 1-4E5252A Output Ports
WARNINGDo not touch the force and guard terminals of the output connectors while the
E5250A is turned on. Dangerous voltages up to the maximum input voltage
may be present at the output connectors.
Agilent E5250A User’s Guide, Edition 91-7
Introduction
Agilent E5255A Product Description
Agilent E5255A Product Description
The E5255A 24 (8×3) Channel Multiplexer is a 2-input to 24-output multiplexer
card for the E5250A. The E5 255A cons ists of 3 blocks ( three 2×8 m ultiplexers). So,
one E5255A can be configured as a 2×8 multiplexer, 2×16 multiplexer, or 2×24
multiplexer. "2-input" means one BIAS input and one IV input.
The E5255A is designed for semiconductor device reliability testing which needs to
automatically scan instrument input/output for many DUTs.
A maximum of four E5255As can be installed in the E5250A, which can use the
multiplexers in several configurations. For example, the E5250A can configure
twelve 8-output multiplexers, six 16-output multiplexers, four 24-output
multiplexers, three 32-output multiplexers, two 48-output multiplexers, or one
96-output multiplexer.
The E5255A block diagram is shown in Figure 1-5.
BIAS Input The BIAS INPUT ports are used to apply ac or dc bias to DUT.
BIAS INPUT ports are BNC connectors. Each BIAS INPUT
port is assigned to a 2×8 multiplexer block as shown in Table
1-2. And you can make internal connections to connect a BIAS
INPUT port to multiple blocks. Refer to Chapter 2.
IV Input IV input is used for measuring/forcing dc current/voltage. You
can connect the IV input to desired SMU INPUT connector
(SMU1 to SMU6) of the E5250A by making an internal
connection. Refer to Chapter 2. The E5250A AUX INPUT
connectors are not used with the E5255A.
OUTPUT
Connectors The E5255A has three 8-channel coaxial output connectors as
shown in Figure 1-6. Each output con nector is assigned to a 2×8
multiplexer as shown in Table 1-2.
1-8Agilent E5250A User’s Guide, Edition 9
Agilent E5255A Product Description
Resistors Resistors connected between BIAS path and IV path are for
protecting DUT from electrical damage. When shipped from
factory, 0 Ω resistance is used in the E5255A. You can easily
replace with desired resistors. Refer to Chapter 2. The
following resistors are furnished with the E5255A:
•0 Ω resistors (3 sets with 10 resistors in each set)
•1.2 kΩ resistors (3 sets with 10 resistors in each set)
•22 kΩ resistors (3 sets with 10 resistors in each set)
If you replace initial 0 Ω resistors with other res istor s, you must
cut the initial resistors. Then, if you need 0 Ω resistors again,
use the ones listed above.
WARNINGBe careful about the following when selecting protective resistors:
•If only one E5255A is installed in the mainframe, total power consumption
by resistors must be less than 16W.
•If multiple E5255As are installed in the mainframe, total power
consumption by resistors must b e less than 32W.
Introduction
•Consider the current flows to re sistor when DUT shorts, and make sure the
power consumption by resistor will be within specif ication of resistor.
Do not use resisto rs th at hav e specifications inadequate for your
applications, which will result in resistors heating up, burning, or s moking.
Be careful about the heating of the rear panel of mainframe or plug-in card,
which may be heated by resistors, even if you follow this warning.
Agilent E5250A User’s Guide, Edition 91-9
BIAS Input 1
Output
Resistor
IV Input 1
BLOCK 1
1 2 34 56 7 8
BIAS Input 3
Output
Resistor
IV Input 3
BLOCK 3
17 18192021 222324
BIAS Input 2
Output
Resistor
IV Input 2
BLOCK 2
9 10111213141516
Introduction
Agilent E5255A Product Description
Figure 1-5E5255A Block Diagram
1-10Agilent E5250A User’s Guide, Edition 9
Table 1-2E5255A Input/Output Connectors
Introduction
Agilent E5255A Product Description
2×8 multiplexer
Block No.
Block1BIAS1 (51)Output 1 to 8Left
Block2BIAS2 (52)Output 9 to 16Center
Block3BIAS3 (53)Output 17 to 24Right
Figure 1-6E5255A BIAS INPUT Connectors and OUTPUT Connectors
WARNINGDo not touch the force and guard terminals of the OUTPUT connectors while
the E5250A is turned on. Danger ous vol tages up t o the maximum input vol tage
may be present at the OUTPUT connectors.
CAUTIONThe maximum voltage that can be applied to any input terminal is ±200 Vdc. The
maximum current is 1 Adc at ±200 Vdc. The maximum voltage that can be applied
between input terminals is 300 Vdc. Do not apply an input signal over these limits to
the E5255A inputs. If you do, the E5255A will be damaged.
BIAS INPUT
(Port No.)
Output Port No.
OUTPUT
Connector
(Location)
If you use a bias source that has current limit capability, set the bias source current
limit to less than 1 Adc.
Agilent E5250A User’s Guide, Edition 91-11
Introduction
Options and Accessories
Options and Accessories
This section lists the options and accessories available for the E5250A.
Table 1-3 lists the options available for the E5250A.
Table 1-3Options
Model
Number
E5250ALow Leakage Switch Mainframe
Option Ite mDescription
E5250A-301Relay Test Adapter
E5250A-A6JANSI Z540 compliant calibration
E5250A-UK6Commercial cal. certificate w/ test data
1-12Agilent E5250A User’s Guide, Edition 9
Table 1-4 lists accessories furnished with the E5250A, E5252A, and E5255A.
Table 1-4Furnished Accessories
Introduction
Options and Ac cessories
Model
Number
Part NumberDescription Qty.
E5250AE5250-90000Agilent E5250A User's Guide1
E5250-17001
E5250A VXI Plug&Play Driver Disk
a
E5250-17003Agilent E525 0A Progr am Disk. See next
page. (LIF formatted, 3.5in diskette)
04155-60007Module Extractor1
8710-2151Wrench (for removing plug-in card)1
−Power Cable1
E5252AE5252-90000Agilent E5252A Installat i on Gui de1
8710-2151Wrench (for removing plug-in card)1
E5255AE5255-90000Agilent E5255A Installat i on Gui de1
1250-0929BNC Short Cap3
E5255-65000BIAS Input Modification Kit
(BNC Open Cap × 2, wire × 2)
E5255-650010 Ω Resistor Set (10 resistors)3
1
1
1
E5255-65 0021.2 kΩ Resistor Set (10 re si stors)3
E5255-65 00322 kΩ Resistor Set (10 resistors)3
8710-2151Wrench (for removing plug-in card)1
a. The Agilent E5250A VXIplug&play driver supports Windows XP Pro-
fessional, Wi ndows 2000, Windows NT 4.0, Windows 98, and W i ndows
95.
Agilent E5250A User’s Guide, Edition 91-13
Introduction
Options and Accessories
Contents of the E5250A Program Disk:
Virtual Front
Panel Utility Interactive control program for the E5250A. Runs on Agilent
4155/4156's built-in IBASIC controller. Refer to Chapter 5.
VFP Data Upload
Library Subprogram library used for upl oading the E5250A con trol data
(made by Vi rtual Front Panel Utility) to your program. Runs on
HP BASIC. Refer to Chapter 6.
Capacitance
Compensation
Routine Subprogram to compensate for capacitance error caused by the
E5250A and E5252A. Ru ns on H P BAS IC . This subprogram is
used to compensate the capacitance data measured by Agilent
4284A Precision LCR Meter. Refer to Chapter 6.
Sample Programs Three sample programs: one for Vth and C measurement using
the E5252A, and two for HCI test and analysis using the
E5255A. Runs on HP BASIC. Refer to Chapter 9.
Self-test Utility Self-test and relay test program for the E5250A. Runs on the
4155/4156's built-in IBASIC controller. Refer to Chapter 3.
1-14Agilent E5250A User’s Guide, Edition 9
Table 1-5 lists accessories available for the E5250A.
16494BKelvin Triaxial Cable (4156 kelvin to E5250 kelvin)
16494CKelvin Triaxial Cable (4142 kelvin to E5250 kelvin)
16494D8 Channel Shielded Coaxial Cable
16494EWide Temperature 8 Channel Shielded Coaxial Cable
16495CConnector plate with 6 connectors for 16494D
16495DConnector plate with 12 connectors for 16494D
16495EBlank plat e
16495FConnector plate with 12×triaxial, intlk, GNDU
16495GConnector plate with 24×triaxial, intlk, GNDU
Option
Item
16494A-0011.5 m length
16494A-0023 m length
16494A-00380 cm length
16494B-0011.5 m length
16494B-0023 m length
16494B-00380 cm length
16494C-0011.5 m length
16494C-0023 m length
16494D-0011.5 m length
16494D-0023 m length
16494E-0013m length
16495F-001Bulkhead feedthrough connectors (female to female)
16495F-002Connectors to contacts for soldering
16495G-001Bulkhead feedthrough connectors (female to female)
16495G-002Connectors to contacts for soldering
Description
Agilent E5250A User’s Guide, Edition 91-15
Introduction
Options and Accessories
1-16Agilent E5250A User’s Guide, Edition 9
2Installation
Installation
This chapter describes requirements to install Agilent E5250A and the tasks for
installation, and is organized into the following three sections:
•“To Inspect the E5250A and Accessories” describes how to inspect the E5250A
when you receive it.
•“Requirements” explains the information required for operating, storing, and
shipping the E5250A.
•“Installing the E5250A” describes how to install the E5250A.
•“Maintenance” describes about performance verification and cleaning.
WARNINGThe maximum input voltage of the E5250A is ±200 V. And dangerous voltages
may be present at the E5250A output terminals. To prevent electric shock, you
must observe the following safety precautions when using the E5250A.
• Use a three-conductor ac power cable to connect cabinet (if used) and the
E5250A to an electric ground (safety ground).
• If you need to touch the force and guard terminals of the output connector,
turn off the E5250A and discharge any capacitors whenever possible.
• Warn workers around the E5250A about dangerous conditions.
2-2Agilent E5250A User’s Guide, Edition 9
Installation
To Inspect the E5250A and Accessories
To Inspect the E5250A and Accessories
When you receive and open the box that contains the E5250A, check the following:
•Before unpacking any components, inspect all boxes for any signs of damage
that might have occurred during shipment such as:
•Dents
•Scratches
•Cuts
•Water marks
•When you open the boxes that contain the E5250A, check the components
against the contents lists that are attached to the boxes.
•After installing the E5250A, check the E5250A operation by executing self-test
shown in Chapter 3.
If anything is wrong, notify your local Agilent Technologies sales office.
When shipping the E5250A
When you unpack the E5250A, retain all packing materials for future use. If it
becomes necessary to reship the E5250A, repack it in the original packing materials
and shipping carton.
Agilent E5250A User’s Guide, Edition 92-3
Installation
Requirements
Requirements
This section describes the following requirements for the E5250A.
•“Power Requirements”
•“Power Cable”
•“Operating Environment”
•“Storage and Shipping Environment”
Power Requirements
CAUTIONBefore applying ac line power to the E5250A, ensure that the correct power cable is
used.
The E5250A can operate from any single-phase ac power source supplying 100 to
240 V ±10 % in the frequency range from 47 to 63 Hz. The maximum power
consumption is 100 VA. For details, refer to Chapter 10, “Specifications.”
Power Cable
In accordance with international safety standards, this instrument is equipped with a
three-wire power cable. When connected to an appropriate ac power outlet, this
cable grounds the instrument frame. The type of power cable shipped with each
instrument depends on the country of destination. Refer to the following table for
the part numbers of the power cables available.
If the plug on the cable does not fit the power outlet, or the cable is to be attached to
a terminal block, cut the cable at the plug end and re-wire it. This work should be
performed by a qualified electrician all local electrical codes being strictly observ ed.
The color coding used in the cable will depend on the cable supplied. If a new plug
is to be connected, it must meet local safety requirements and include the following
features:
•Adequate load-carrying capacity. (refer to Chapter 10, “Specifications.”)
•Ground connection.
•Cable clamp.
2-4Agilent E5250A User’s Guide, Edition 9
Installation
Requirements
•Plug: BS 1363/A,
250 V, 10 A
•Cabl e : 8120-4420,
8120-1351
•Plug: NEMA 6-15P,
250 V, 10 A
•Cabl e : 8120-3996,
8120-0698
•Plug: JIS C 8303,
125 V, 12 A
•Cabl e : 8121-0743,
8120-4753
•Plug: AS 3112, 250 V ,
10 A
•Cable: 8120-4419,
8120-1369
•Plug: SEV 1011, 250
V, 1 0 A
•Cable: 8120-2104
•Plug: Israel SI 32, 250
V, 1 0 A
•Cable: 8120-5182
•Plug: CEE 7 sheet
VII, 250 V, 10 A
•Cable: 8120-4519,
8120-1689
•Plug: SR 107-2-D1,
250 V, 10 A
•Cable: 8120-2956
•Plug: IRAM 207310A, 250 V, 10 A
•Cable: 8120-6870
•Plug: NEMA 5-15P,
125 V, 10 A
•Cable: 8120-6825,
8120-1378,
8120-1395
•Plug: IS 1293 and IS
6538, 250 V, 10 A
•Cable: 8121-1582,
8120-4211
•Plug : CEI 23-16, 250
V, 10 A
•Cable: 8120-6978
•Plug: GB 1002 figure
3 , 250 V, 10 A
•Cabl e : 8120-8376
•Plug : CS 0017:2003,
250 V, 10 A
•Cable: 8120-8871,
8120-0674,
8121-1638
•Plug: SANS 164-1,
250 V, 10 A
•Cable: 8121-0564,
8120-4211
•Plug: CNS 10917-2,
125 V, 10 A
•Cable: 8121-1635,
8120-6825
Agilent E5250A User’s Guide, Edition 92-5
Installation
Requirements
WARNINGFor protection from electrical shock, the power cable ground must not be
defeated.
Operating Environment
The E5250A must be operated within the following environmental conditions:
•T emp e rature: 5 °C to 40 °C (41 °F to 104 °F)
•Humidity: 5 % to 80 % RH at 40 °C (104 °F), non-condensing
•Altitude: 0 m to 2000 m
Storage and Shipping Environment
The E5250A should be stored or shipped in environments within the following
limits:
•Temperature: −40 °C to 70 °C (−40 °F to 158 °F).
•Humidity: 5 % to 90 % RH at 65 °C (149 °F).
•Altitude: 0 m to 15,240 m
Protect the E5250A from temperature extremes to prevent condensation from
forming inside the E5250A.
2-6Agilent E5250A User’s Guide, Edition 9
Installation
Installing the E5250A
Installing the E5250A
This section describes how to install the E5250A.
•“To Install Plug-in Card”
•“To Install Blank Panel”
•“To Configure E5255A”
•“E5255A Configuration Examples”
•“To Set GPIB Ad dress”
•“To Connect GPIB Cable”
WARNINGTo prevent electrical shock, turn off the mainframe and remove the power
cable before starting the instruction.
CAUTIONBe careful about the module pins used for internal connection to th e mainframe. The
pins can be damaged easily.
Use clean handling and anti-static procedures when removing, configuring, and
installing the modules. The modules contain components that can be damaged by
static electricity.
NOTEPlug-in module is shipped from the factory after the specifications are confirmed.
Agilent can guarantee that the modules will function and that the module
performance is designed to meet its speci ficat ion s. However, if you have any
trouble, contact your nearest Agilent Technologies service center.
If you require that the modules be guaranteed to their specifications (for ISO
compliance, etc.), then you must perform a calibration on the instrument (modules
and mainframe together). For calibration, contact your nearest Agilent Technologies
service center.
Agilent E5250A User’s Guide, Edition 92-7
Installation
Installing the E5250A
To Install Plug-in Card
The E5252As and E5255As are plug-in cards (modules). You install these cards in
card slots 1 to 4 of the E5250A. The following procedure explains the card
installation and removal:
1. Turn off the E5250A, then wait at least 10 seconds before you remove or install
a card.
2. Remove a blank panel or a card attached to the slot you want to install a new
card, as follows.
To remove blank panel, do as follows:
a. Loosen the screws on both the left and right edges of the blank panel by
using the wrench furnished with the E5250A.
b. Remove the blank panel.
To remove a card, do as follows:
a. Screw the Module Extractor into the screw hole for the extractor. See Figure
2-1. Module Extractor is furnished with the E5250A.
b. Loosen the screws on left and right edges of the card by using the wrench
furnished with the E5250A.
c. Gently pull the Module Extractor to extract the card from the card slot.
3. Install the card as follows:
a. Align the card with the left and right slot guide rails. The component side
should be fac i ng up.
b. Push the card into the slot until you feel it seat firmly into the connector at
the back of the card slot.
c. Screw in the screws on the left and right edges of the card by using the
wrench furnished with the E5250A.
4. Execute the Relay Test and Leak Test:
After installing the cards, perform the Relay Test and Leak Test described in
Chapter 3 to check the operation and leak current of the E5250A with installed
cards.
CAUTIONIf you want to confirm that the E5250A with installed cards meets its specifications,
return the E5250A with installed cards to your nearest Agilent Technologies Service
Center to do the Performance Verification.
2-8Agilent E5250A User’s Guide, Edition 9
Figure 2-1Module Extractor and Location of the Screw Hole
Module Extractor
Screw Hole
for Module Extractor
Agilent E5255A
Agilent E5252A
Installation
Installing the E5250A
To Install Blank Panel
CAUTIONTo prevent thermal damage to the E5250A cards, be sure that blank panels (Agilent
part number E5250-60003) are installed in all unused slots.
If the blank panel is not installed to cover an unused slot, install the blank panel as
follows:
1. Align the blank panel over the unused slot.
2. Screw in the screws on the left and right edges of the blank panel by using the
wrench furnished with the E5250A.
Agilent E5250A User’s Guide, Edition 92-9
Installation
Installing the E5250A
To Configure E5255A
The Agilent E5255A has three 2-input 8-output multiplexers. You can configure
several multiplexers by modifying some internal connections. Component locations
on the E5255A are shown in Figure 2-2, and the relationship of the multiplexer
blocks and the component s is sho wn in Table 2-1. Table 2-2 shows the initial s ettin g
of the E5255A internal connections. The setting does not depend on the slot in
which the module is installed.
IV input connection is defined by the connection of IVin cable and the setting of
DIP switch shown in Figure 2-2. For actual examples of modifications, see
“E5255A Configuration Examples”.
DIP SW
Each DIP switch has four bit switches. Bits 1 to 3 are used to specify which SMU
input connector is connected to the IV input. Bit 4 of DIP switch is used to specify
whether the protective resistors are 0 Ω or not. The DIP switche s m ust be set
properly after you complete the internal connection. Make sure you set the DIP
switches correctly. If not, dangerous voltages may be applied to unexpected output
connectors.
Each multiplexer block has an independent BIAS input. You can change the BIAS
input internally using the BIASin/BIASout terminals shown in Figure 2-2, or
externally using the BIAS INPUT connectors. For actual examples of modifications,
see “E5255A Configuration Examples”.
To Use Multiplexer Blocks on another E5255A
When a module is installed in a mainframe slot, six module pins are connected to
the mainframe internal bus. So, if multiplexers on different E5255As use the same
module pin, the IV inputs are connected together via the internal bus.
Agilent E5250A User’s Guide, Edition 92-11
Installation
Installing the E5250A
For example, if a mainframe installs four E5255As which the IVin1 cable is
connected to the SMU1 input connector and the IVin2 and IVin3 cables are not
connected, the E5250A configures a 32-output multiplexer that uses the SMU
INPUT 1 connector as the IV input.
To Mount Protective Resistors
To protect the device (DUT) from electrical damage, you can mount resistors
between the IV input and BIAS input of multiplexer. The following procedure
shows how to mount the resistors. See Figure 2-2 for the component locations.
When shipped from the factory, 0 Ω resistors are mounted, and bit 4 of all DIP
switches is set to OFF (0).
1. Cut the original resistors (0 Ω) off.
2. Loosen the screws from both sides of the resistor holder.
3. Put resistors in holder, and tig hten screws. And confirm that the resistors are
held properly. Resistors should be mounted on each multiplexer block that you
will use (8 resistors/block).
4. Set bit 4 to ON (1) for the DIP switch of the related block. For example, if you
mount resistors for Block 1 and 2 m ultip lexers, s et bi t 4 to ON for DIP SW1 and
SW2.
To Select Protective Resistors
Be careful about the following when selecting protective resistors. Also, be careful
about the heating of the rear panel of mainframe or plug-in module, which may be
heated by resistors, even if you follow this warning.
•If only one E5255A is installed in the mainframe:
Total power consumption by resistors must be less than 16W.
•If multiple E5255As are installed in the mainframe:
Total power consumption by resistors must be less than 32W.
•Consider the current flows to resistor when DUT shorts, and make sure the
power consumption by resistor will be within specification of resistor. Do not
use resistors that have specifications inadequ ate for your applications , which
will result in resistors heating up, burning, or smoking.
2-12Agilent E5250A User’s Guide, Edition 9
Installing the E5250A
E5255A Configuration Examples
The following example configurations are described:
To make this multiplexer, you need a E5255A. See Figure 2-3 for example setup.
The example uses the SMU INPUT 1 connector as the IV input, and uses 0 Ω
resistors.
T o make this multiplexer, you need a E5255A, two wires (furnished), and two BNC
open caps (furnished). See Figure 2-4 for example setup. The example uses the
SMU INPUT 1 connector as the IV input, and uses 0 Ω resistors.
1. Connect the IVin1 cable to the SMU1 input connector on the E5255A.
2. Connect the IVin2 cable to the IVout1 connector.
3. Connect the IVin3 cable to the IVout2 connector.
4. Set bit 1 to ON (1) and bit 2 thru 4 to OFF (0) for DIP SW1 to SW3.
2-14Agilent E5250A User’s Guide, Edition 9
Installation
To Bias Source
BNC T-adapter
BNC cable
Agilent E5255A
Installing the E5250A
5. Connect wires between BIASout1 and BIASin2, also between BIASout 2 and
BIASin3.
6. Install the module into the slot 1 of the mainframe. For installing module, see
“To Install Plug-in Card” on page 2-8.
7. Connect BNC open caps to BIAS INPUT 52 and 53 connectors. So, the BIAS
INPUT 51 connector will be the BIAS input.
This example internally connects all BIAS INPUT connectors together. So, when
bias is applied to a BIAS INPUT connector, the bias will also be applied to all the
other BIAS INPUT connectors.
Do not forget to connect the BNC open caps to the unused BIAS INPUT conn ectors.
Otherwise, you will be shocked if you touch the connectors while bias is applied.
DO NOT connect the BNC short cap to the connectors. If you do, the E5255A and
the Bias Source will be destroyed when the bias is applied.
To connect BIAS inputs externally.
You can use the BIAS INPUT connectors instead of the BIASin/out internal
connections shown above. See Figure 2-5. This example uses three BNC cables and
two BNC-T type (f-m-f) adapters. Recommended p arts fo r the BNC-T type adap ters
are listed below.
•Agilent part number 1250-0781: female(side)-male(center)-female(side)
•Agilent part number 1250-2405: male(side)-female(center)-female(side)
Figure 2-5Interconnection of BIAS INPUT Connectors
To make this multiplexer, you need four E5255As, eight wires (furnished), five
BNC open caps (furnished), and three BNC cables. The example uses the SMU
INPUT 1 connector as the IV input, and uses 0 Ω resisto rs.
1. See “24-output multiplexer (1-BIAS input, 1-IV input, 1 module)” and make
four 24-output multiplexers. Then the IVin1 cable on all modules must be
connected to the SMU1 input connector.
2. Install the modules into the mainframe. For installing module, see “To Install
Plug-in Card” on page 2-8.
3. Connect BNC open caps to the BIAS INPUT 51 connector of the module in the
slot 1 and all BIAS INPUT 52 connectors.
4. Connect BNC cables between the following connectors:
Do not forget to connect the BNC open caps to the unused BIAS INPUT conn ectors.
Otherwise, you will be shocked if you touch the connectors while bias is applied.
DO NOT connect the BNC short cap to the connectors. If you do, the E5255A and
the Bias Source will be destroyed when the bias is applied.
2-16Agilent E5250A User’s Guide, Edition 9
Installation
Installing the E5250A
To Set GPIB Address
Every device on the GPIB bus must have a unique address.
If you need to change the GPIB address of the E5250A, make sure the E5250A is
turned off. With a small flatblade screwdriver, set the GPIB ADDRESS switch on
the rear panel to the new address (0 to 30).
The new GPIB address is recognized only at power on. The E5250A leaves the
factory with the GPIB address set to 22.
To Connect GPIB Cable
To connect the E5250A with a computer or peripheral device via GPIB (IEEE Std.
488), connect an GPIB cable between the GPIB connector on the E5250A rear panel
and the GPIB connector on the peripheral device.
Including the controller , a total of 15 GPIB interfaces can be connected on the same
GPIB bus. The following are rules for connecting GPIB interfaces:
•If total number of interfaces ≤ 10:
Max. total cable length = number of interfaces × 2 m.
•If total number of interfaces > 10:
Max. total cable length = 20 m.
•Maximum cable length between interfaces is 4 m.
•Star connection and cascade connection are allowed. Loop connection is not
allowed.
Agilent E5250A User’s Guide, Edition 92-17
Installation
Maintenance
Maintenance
Maintenance should be performed periodically to keep the E5250A in good
condition.
Performance Verification
Performance verification must be performed periodically so that the instruments
satisfy the specifications, and keep a good condition. It is recommended to perform
the performance verification once a year at least. For the performance verification,
contact your nearest Agilent Technologies Service Center . T rained service personnel
will perform the performance verification.
Cleaning
Before performing cleaning, turn off the instrument, and disconnect power cable
from the rear panel. Use a dry cloth to clean the external case parts.
To prevent electrical shock, do not perform cleaning when the instrument is turned
on, and do not use a wet cloth.
2-18Agilent E5250A User’s Guide, Edition 9
3Executing Self-Test and Leak Test
Executing Self-Test and Leak Test
This chapter describes how to execute the Self-Test and the Leak Test.
The Self-Test consists of three tests that check the operation of Agilent E5250A
firmware, LEDs, and key that are on the E5250A front panel, and the relays on the
cards installed in the E5250A.
Self-Test should be executed every day as an operation check before using the
E5250A.
The Leak Test checks leakage current of cards installed in the E5250A.
Leak Test can be executed by using Agilent 4155/4156 Semiconductor Parameter
Analyzer with the Self-Test Utility progr am that is o n the E5250A Program Disk
furnished with the E5250A.
NOTEPerformance Verification
The Self-Test and the Leak Test cannot be used to verify the specifications of the
E5250A, E5252A, and E5255A.
To verify the specifications, return the E5250A with the cards installed to your
nearest Agilent Technologies Service Center. The performance verification should
be done by service personnel who are trained to service the E5250A, and should be
performed once a year.
•“Executing Self-Test” introduces the test items of the Self-Test and describes
how to execute the Self-Test.
•“Using the Self-Test Utility” describes how to execute the Self-Test and the
Leak Test interactively from Agilent 4155/4156 semiconductor parameter
analyzer.
3-2Agilent E5250A User’s Guide, Edition 9
Executing Self-Test
The Self-Test of the E5250A consists of the following three tests:
Table 3-1E5250A Self-Test Items
Test Item Description Executed by Refer to
Executing Self-Test and Leak Test
Executing Self-Test
Controller
Test
Front
Panel
Interface
Test
Relay TestChecks the operation of
Checks the operation of
the E5250 A firmware.
Checks the operation of
the LEDs and key that
are on the E5250A front
panel.
the relays on the cards
installed in the E5250A.
Turning on the E5250A.page 3-4
Entering SCPI Command.pa ge 3-6
Using Self-Test Utility.page 3-10
Entering SCPI Commandpage 3-6
Using Self-Test Utility.page 3-10
Pressing the Local/
Self Test key.
Entering SCPI Command.pa ge 3-6
Using Self-Test Utility.page 3-10
page 3-4
Agilent E5250A User’s Guide, Edition 93-3
Executing Self-Test and Leak Test
Executing Self-Test
To Execute Self-Test (Standalone)
Controller Test is executed automatically when you turn on the E5250A. You
execute the Relay Test by pressing the
panel.
Executing the Controller Test
1. Turn on the E5250A.
Local/Self Test key on the E5250A front
2. Wait until the LED in the
Local/Self Test key t urns off.
If System Fail LED or Fail LED stays on after executing the Controller Test, contact
your nearest Agilent Technologies Service Center.
Executing the Relay Test
Requirements:
•Relay Test Adapter (E5250A-301)
•BNC Short Caps (3 caps are furnished with each E5255A)
Procedure:
1. Disconnect all cables from the E5250A input connectors.
2. Open all outputs of the plug-in cards as follows: If cable is connected to output
connector of plug-in card, open output at DUT end of cable. If end of cable is
Kelvin connection, disconnect the cable or make non-Kelvin connection at end
of cable.
3. Attach the Relay Test Adapter to the E5250A input connectors as shown in
Figure 3-1.
4. Set switch on the Relay Test Adapter to Relay Test position.
5. For each E5255A card installed in the E5250A:
a. Disconnect any cables from the BIAS INPUT connectors of the E5255A.
b. Connect a BNC Short Cap to each E5255A BIAS INPUT connector.
6. (If Remote LED is on, press the
Local/Self Test key to return to Local control.
The Remote LED should turn off.)
7. Press the
8. Wait until the LED in
Local/Self Test key. The Relay Test executes.
Local/Self Test key turns off.
3-4Agilent E5250A User’s Guide, Edition 9
NOTEIf Fail LED is ON after Relay Test
Relay Test Adapter
After executing the relay test, if the Fail LED turns on, the cause is one of the
following:
• The Relay Test Adapter is not connected or a cable is still connected to input
connector.
• Output is not open. See step 2 of previous procedure.
• You did not properly install the E5255A. See “To Configure E5255A” on page
2-10.
• A plug-in card may be defective. Contact your nearest Agilent Technologies
Service Center.
Figure 3-1Relay Test Adapter Connection
Executing Self-Test and Leak Test
Executing Self-Test
NOTESettings after Relay Test
After Relay Test, the E5250A settings are change d to same as after :SYST:CPON
command execution.
Agilent E5250A User’s Guide, Edition 93-5
Executing Self-Test and Leak Test
Executing Self-Test
To Execute Self-Test using External Controller
You can execute Self-T est fr om an external controller by using SCPI commands in a
program.
This section describes the requirements, SCPI commands, and a s ample program for
Self-Test execution from an external controller. For details about the SCPI
commands, refer to Chapter 7.
Requirements
•Controller
•GPIB Cable
•Relay Test Adapter
•BNC Short Caps (3 caps are furnished with each E5255A)
SCPI Commands for Self-Test
Table 3-2 shows the SCPI commands used to execute Self-Test. Keyword of the
SCPI command first level is :DIAG.
The :DIAG commands ignore the E5250A channel configuration mode (NORMAL
or AUTO).
For :DIAG commands that require a card number, you specify 1, 2, 3, 4, or ALL.
If the Fail LED turns on after executing the Self-Test, the E5250A or modules may
be defective. Contact your nearest Agilent Technologies Service Center.
NOTEBefore executing Relay Test, be careful about following:
• Attach the Relay Test Adapter to the E5250A input connectors. Refer to Figure
3-1. Then set the switch on the adapter to the Relay Test position.
• Open all outputs of the plug-in cards as follows: If cable is connected to output
connector of plug-in card, open output at DUT end of cable. If end of cable is
Kelvin connection, disconnect the cable or make non-Kelvin connection at end
of cable.
• For each E5255A installed in the E5250A, connect a BNC short cap to each
E5255A BIAS INPUT connector.
3-6Agilent E5250A User’s Guide, Edition 9
Table 3-2SCPI Commands for Self-Test
Test Item Description SCPI Command
Executing Self-Test and Leak Test
Executing Self-Test
Controller
Test
Executes Controller Test,
then returns test result.
Returns test result.:DIAG:TEST:FRAM:STAT? CONT
Clears test result.:DIAG:TEST:FRAM:CLE CONT
Front Panel
Interface
Executes I/F Test, then
returns test result.
(I/F) Test
Returns test result.:DIAG:TEST:FRAM:STAT? FPAN
Clears test result.:DIAG:TEST:FRAM:CLE FPAN
Relay TestExecutes Relay T est, then
returns test result.
Returns test result.:DIAG:TEST:CARD:STAT? ALL
Clears test result.:DIAG:TEST:CARD:CLE ALL or
NOTEFront Panel Interface Test
After the :DIAG:TEST:FRAM:EXEC? FPAN command is executed, the Remote,
System Fail, Fail, and Local/Self Test LEDs on the E5250A front panel blink for
about 10 seconds. If an LED does not blink, the LED is defective.
:DIAG:TEST:FRAM:EXEC? CONT
:DIAG:TEST:FRAM:EXEC? FPAN
:DIAG:TEST:CARD:EXEC? ALL
or Card No. (1, 2, 3, or 4)
or Card No. (1, 2, 3, or 4)
Card No. (1, 2, 3, or 4)
Confirm that the four LEDs are blinking, then press the
Local/Self Test key on the
E5250A front panel within 10 seconds. The LEDs should stop blinking.
If you do not press the
Local/Self Test key within 10 seconds, the E5250A returns a
fail result for the Front Panel Interface Test, which turns on the Fail LED.
If you pressed
it means the
Local/Self Test key within 10 seconds, and the Fail LED still turns on,
Local/Self Test key is defective.
If you find any problem, contact your nearest Agilent Technologies Service Center.
Agilent E5250A User’s Guide, Edition 93-7
Executing Self-Test and Leak Test
Executing Self-Test
Self-Test Programming Example
This sample program is effective for an E5250A that has four card s installed. If your
E5250A does not have four cards installed, delete the Relay Test program lines for
the empty card slots. The program runs on HP BASIC.
10 ! Self-Test by Ext er na l Co nt ro ll er
20 !
30 REAL A
40 ASSIGN @Hp5250 TO 722
50 OUTPUT @Hp5250;":DIAG:TEST:FRAM:CLE CONT"
60 OUTPUT @Hp5250;":DIAG:TEST:FRAM:CLE FPAN"
70 OUTPUT @Hp5250;":DIAG:TEST:CARD:CLE ALL"
80 DISP "Press Continue to go to Controller Test."
90 PAUSE
100 ! Controller Tes t
110 OUTPUT @Hp5250;":DIAG:TEST:FRAM:EXEC? CONT"
120 ENTER @Hp525 0; A
130 IF A=0 THEN GOTO 160
140 DISP "Failed . Pr es s Co nt in ue to go to I/ F Te st ."
150GOTO 170
160DI SP "Pass ed . Pr es s Co nt in ue to go to I/ F Te st ."
170 PAUSE
180 ! Front Panel I/ F Te st
190 OUTPUT @Hp5250;":DIAG:TEST:FRAM:EXEC? FPAN"
200 DISP "Press Lo ca l/ Se lf Tes t Ke y wi th in 10 se c. "
210 ENTER @Hp5250;A
220 IF A=0 THEN GOTO 250
230 DISP "Failed . Pr es s Co nt in ue to go to Re la y Te st ."
240 GOTO 260
250 DISP "Passed . Pr es s Co nt in ue to go to Re la y Te st ."
260 PAUSE
270 ! Relay Test
280 OUTPUT @Hp5250;":DIAG:TEST:CARD:EXEC? 1"
290 ENTER @Hp5250;A
300 IF A=0 THEN GOTO 330
310 DISP "Card 1 failed. Press Continue to test Card 2"
320 GOTO 340
330 DISP "Card 1 passed. Press Continue to test Card 2"
340 PAUSE
350 OUTPUT @Hp5250;":DIAG:TEST:CARD:EXEC? 2"
360 ENTER @Hp5250;A
370 IF A=0 THEN GOTO 400
380 DISP "Card 2 failed. Press Continue to test Card 3"
390 GOTO 410
400 DISP "Card 2 passed. Press Continue to test Card 3"
410 PAUSE
420 OUTPUT @Hp5250;":DIAG:TEST:CARD:EXEC? 3"
3-8Agilent E5250A User’s Guide, Edition 9
Executing Self-Test and Leak Test
Executing Self-Test
430 ENTER @Hp5250;A
440 IF A=0 THEN GOTO 470
450 DISP "Card 3 failed. Press Continue to test Card 4"
460 GOTO 480
470 DISP "Card 3 passed. Press Continue to test Card 4"
480 PAUSE
490 OUTPUT @Hp5250;":DIAG:TEST:CARD:EXEC? 4"
500 ENTER @Hp5250;A
510 IF A=0 THEN GOTO 540
520 DISP "Card 4 failed."
530 GOTO 550
540 DISP "Card 4 passed."
550 END
Line Number Description
30 to 90Clears all previous test results.
100 to 170Executes the Controller Test, then returns test result.
180 to 260Executes the Front Panel Interface Test, then returns the test
result.
270 to 340Executes the Relay Test for the card installed in slot 1 of the
E5250A, then returns test result.
350 to 410Executes the Relay Test for the card installed in slot 2 of the
E5250A, then returns test result.
420 to 480Executes the Relay Test for the card installed in slot 3 of the
E5250A, then returns test result.
490 to 550Executes the Relay Test for the card installed in slot 4 of the
E5250A, then returns test result.
Agilent E5250A User’s Guide, Edition 93-9
Executing Self-Test and Leak Test
Using the Self-Test Utility
Using the Self-Test Utility
The Self-T est Utility is a Self-Test program for the E5250A, and runs on the Agilent
4155/4156's built-in IBASIC controller.
Self-Test Utility can execute all Se lf-Test items listed in Table 3-1 and can also
execute the Leak Test, which checks leakage current of the E5250A with installed
cards. You execute the Self-Test items and Leak Test interactively (using softkeys)
from the 4155/4156.
This section explains the operation of the Self-Test Utility.
•“Requirements”
•“To Start the Self-Test Utility”
•“To Execute Self-Test”
•“To Execute Leak Test”
Requirements
The following equipment is required to execute the Self-Test Utility:
Table 3-3Required Equipment
Agilent
Model/Part No.
4155 or 4156Semiconductor
Parameter Analyzer
10833A/B/CGPIB Cable1
E5250-17003E5250A Program
Disk (furnished)
E5250A-301
(E5250-60002)
1250-0929BNC Short Cap3 for each E5255A
3-10Agilent E5250A User’s Guide, Edition 9
Relay Test Adapter1 Relay Test
Description Qty. Usage
1 Executing
Self-Test
Utility
1
(furnished with the
E5255A)
Executing Self-Test and Leak Test
Using the Self-Test Utility
To Start the Self-Test Utility
1. Before turning on the 4155/4156 and E5250 A , conn ect the GPIB cabl e bet ween
the E5250A and the 4155/4156.
2. Turn on the 4155/4156 and E5250A.
3. Set the 4155/4156 to "SYSTEM CONTROLLER" mode on SYSTEM:
MISCELLANEOUS page.
4. Insert the E5250A Program Disk into the 4155/4156 disk drive.
5. Press the
Display key twice to display the "All IBASIC" screen. This key is
located in the IBASIC area on the 4155/4156 front panel.
6. Select GET softkey on the 4155/4156 screen. The following is displayed:
GET "_"
7. Type SELFTEST between the double quotations of GET statement. Then press
Enter key. Wait until the 4155/4156 finishes getting the file.
8. Press
Run key, which is in the IBASIC area on the 4155/4156 front panel. The
following message is displayed:
Input Agilent E5 25 0A GPI B Ad dr es s (d ef au lt 722 ): ?
9. To use default GPIB address 722, just press the Enter key. If not, type the GPIB
address of your E5250A, then press
Enter key. MAIN MENU of the E5250A
Self-Test Utility and followin g softkeys are displayed on the 4155/4156:
Softkey Description
SELF TESTDisplays the SELF TEST MENU.
LEAK TESTDisplays the LEAK TEST MENU.
QUITQuits the Self-Test Utility.
Agilent E5250A User’s Guide, Edition 93-11
Executing Self-Test and Leak Test
Using the Self-Test Utility
To Execute Self-Test
On the MAIN MENU of the E5250A Self-Test Utility, select the SELF TEST
softkey to display the E5250A SELF TEST MENU and following softkeys:
Softkey Description
CARD1 RELAYFor executing Relay Test for Card1.
CARD2 RELAYFor executing Relay Test for Card2.
CARD3 RELAYFor executing Relay Test for Card3.
CARD4 RELAYFor executing Relay Test for Card4.
CONTROLLERExecutes Controller Test immediately.
FRONT PANELExecutes Front Panel I/F Test imm ediately.
MAIN MENUReturns to MAIN MENU of Self-Test Utility
immediately.
Also, the most recent Self-Test result (P/F/N) is displayed on the 4155/4156 screen.
Figure 3-2Example of SELF TEST MENU
Agilent E5250A Tes t Ut il it y
** SELF TEST MENU **
CARD1 RELAY ... St ar t Sl ot 1 E5 25 5A Rel ay Tes t [N]
CARD2 RELAY ... St ar t Sl ot 2 E5 25 5A Rel ay Tes t [N]
CARD3 RELAY ... St ar t Sl ot 3 E5 25 5A Rel ay Tes t [N]
CARD4 RELAY ... St ar t Sl ot 4 E5 25 5A Rel ay Tes t [N]
CONTROLLER .. . St ar t E5 25 0A Con tr ol le r Te st [P]
FRONT PANEL ... St ar t E5 25 0A Fro nt Pan el UIF Tes t [N ]
MAIN MENU... Return to MAIN MENU
[P] indicates te st was pas se d.
[F] indicates te st was fai le d.
[N] indicates test was not executed.
Prepare the rela y te st ada pt er s
before you select "CARDn RELAY" key.
Select desired softkey.
_
3-12Agilent E5250A User’s Guide, Edition 9
Executing Self-Test and Leak Test
Using the Self-Test Utility
Executing the Controller Test
To execute the controller test, press the CONTROLLER softkey.
If [F] result is displayed after the test, the E5250A is defective. Contact your
nearest Agilent Technologies Service Center.
Executing the Front Panel Interface Test
To execute the front panel interface test, do the following:
1. Press the FRONT PANEL softkey.
2. Confirm that the following four LEDs on the E5250A front panel are blinking.
And press the
•Remote LED
•System Fail LED
•Fail LED
•Local/Self Test LED
Local/Self Test key on the E5250A front panel within 10 seconds.
NOTEFail Condition for Front Panel Interface Test
If an LED does not blink after selecting the FRONT PANEL softkey, the LED is
defective.
If you do not press the
If you pressed
it means the
Local/Self Test key within 10 seconds, and the Fail LED still turns on,
Local/Self Test key is defective.
Local/Self Test key within 10 seconds, the Fail LED turns on.
Contact your nearest Agilent Technologies Service Center.
Agilent E5250A User’s Guide, Edition 93-13
Executing Self-Test and Leak Test
Using the Self-Test Utility
Executing the Relay Test
To execute the Relay Test, do the following:
1. Press the CARDn RELAY softkey (n=1 to 4). The following message is
displayed on the 4155/4156 screen depending on card:
•Message for the E5252A:
Attach the relay test adapter on input terminal and
open all output of slot n.
•Message for the E5255A:
Attach the relay tes t ad ap te r an d th e BN C sh or t ca p to
E5250A/E5255 A in pu t. And ope n al l ou tp ut of sl ot n.
Where, n=1 to 4.
And the following softkeys are displayed.
Softkey Description
CONTINUEExecutes the Relay Test for the card immediately.
CANCELCancels the Relay Test execution.
2. Disconnect all cables from the E5250A input terminals. And attach the Relay
Test Adapter to the E5250A input terminals. For each E5255A card, connect a
BNC short cap to each E5255A BIAS INPUT port.
3. Open all outputs of the plug-in cards as follows:
If cable is connected to output connector of plug-in card, open output at DUT
end of cable. If end of cable is Kelvin connection, disconnect the cable or make
non-Kelvin connection at end of cable.
4. Press the CONTINUE softkey to execute the Relay Test.
Or press the CANCEL softkey to cancel test execution.
5. Repeat steps 1 to 4 for the each card you want to test.
If [F] result is displayed after executing the Relay Test, the card is defective.
Contact your nearest Agilent Technologies Service Center.
3-14Agilent E5250A User’s Guide, Edition 9
Executing Self-Test and Leak Test
Using the Self-Test Utility
To Execute Leak Test
NOTELeak Test is for checking leakage current of the E5250A with plug-in cards
installed. If leak test result is too large for your application, contact nearest Agilent
T echnologies Service Center. Leak test does not verify specifications of the E5250A
and installed cards.
The Leak Test applies 10 Vdc to the specified input port of the E5250A, and
measures the leakage current. The test limits are shown in Table 3-4.
You specify an input port of the E5250A, and specify the 4155/4156 SMU to use.
The Leak Test automatically connects the specified input port to each output port
consecutively , and uses the 4155/415 6 SMU to measure the leakage current for each
channel.
You need to execute this test for each input port.
Table 3-4Test Limits of Leak Test
Card Instrument Connection Input Port Test Limit
E5252A 4155 non-KelvinIV 1, IV2±4.12 pA
IV3, IV4, IV5, IV6±1.01 nA
4156 non-KelvinIV1, IV2±1.14 pA
IV3, IV4, IV5, IV6±1.01 nA
KelvinIV1, IV2±2.24 pA
IV3, IV4, IV5, IV6±2.02 nA
E5255A4155non-KelvinIV1 to IV6±4.12 pA
4156non-Kelvi nIV1 to IV6±1.14 pA
Agilent E5250A User’s Guide, Edition 93-15
Executing Self-Test and Leak Test
Using the Self-Test Utility
To execute the Leak Test, do the following:
1. On MAIN MENU of the E5250A Self-Test Utility, select LEAK TEST softkey.
The following message is displayed:
Input Agilent 41 55 /4 15 6 GP IB add re ss (de fa ul t 81 7) :?
2. To use default GPIB address 817 for the 4155/4156, just press Enter key.
If not, type GPIB address of the 4155/4156, then press
Enter key.
The LEAK TEST MENU with the following softkeys are displayed:
Softkey Description
START TESTFor defining the Leak Test connection. Need to define the
E5250A input port and 4155/4156 SMU port .
MAIN MENUReturns to MAIN MENU of the Self-Test Utility.
3. Select START TEST softkey. Following message/softkeys are displayed:
Select IV Port of Leak Test by softkey.
Softkey Description
IVn (n=1 to 6)For specifying the E5250A SMU INPUTn port for which
you want to perform the leak measurement. If only the
E5255As are installed, only IVn softkeys displayed are for
SMU INPUTn internally connected to the E5255A IV input.
CANCELCancels the Leak Test execution.
3-16Agilent E5250A User’s Guide, Edition 9
Executing Self-Test and Leak Test
Using the Self-Test Utility
4. Select desired IVn softkey (n=1 to 6) for input port you want to test.
If the E5252A is installed in your E5250A, and the 4156 is used for the Relay
Test, then when you press IV1, IV3, or IV5 softkey, the follo wing message and
softkeys are displayed:
Is IVn Port a Kelvin connection?
Where n=1, 3, or 5.
Softkey Description
YESDeclares that the port specified by the IVn softkey is used
for the Kelvin connection.
NODeclares that the port specified by the IVn softkey is used
for the non-Kelvin connection.
CANCELCancels the Leak Test execution.
Select the YES or NO softkey.
The following message and the softkeys are displayed:
Select SMU channel for IVn leak measurement by softkey.
Where n=1 to 6.
Softkey Description
SMUm
(m= 1 to 6)
Declares that you will connect the 4155/4156 SMUm to th e
E5250A port specified by the IVn softkey. SMU5 and
SMU6 are displayed only if 5th and 6th SMU are installed
in the 41501 expander of the 4155/4156.
CANCELCancels the Leak Test execution.
Agilent E5250A User’s Guide, Edition 93-17
Executing Self-Test and Leak Test
Using the Self-Test Utility
5. Select the desired SMUm softkey (m=1 to 6). Following messag e is displayed
depending on the connection type:
•Message for Kelvin connection:
Disconnect Kelvin cable from IVn terminal.
And disconnect all cables from output terminals.
•Message for non-Kelvin connection:
Disconnect cable from IVn terminal.
And disconnect all cables from output terminals.
Where n=1 to 6. And the following softkeys are displayed.
Softkey Description
CONTINUETo continue setting up the Leak Test.
CANCELCancels the Leak Test execution.
6. Disconnect all cables from the E5250A input/output terminals. Then connect
triaxial cable or Kelvin triaxial cable to the 4155/4156 SMU terminal you
specified in step 5 because the next step executes the zero offset cancel.
7. Select the CONTINUE softkey. The following message is displayed depending
on the connection type:
•Message for Kelvin connection:
Connect SMUm to IV n by Kel vi n ca bl e.
•Message for non-Kelvin connection:
Connect SMUm to IV n by cab le .
Where m=1 to 6 (specified by SMUm softkey) and n=1 to 6 (specified by IVn
softkey).
And the following softkeys are displayed.
Softkey Description
CONTINUEExecutes the Leak Test. Consecutively connects input port
specified by IVn softkey to each output port and makes
leak measurement.
CANCELCancels the Leak Test execution.
3-18Agilent E5250A User’s Guide, Edition 9
Executing Self-Test and Leak Test
Using the Self-Test Utility
8. Connect the cable from the 4155/4156 SMU port specified by the SMUm
softkey (m=1 to 6) in step 5 to the E5250A input port specified by the IVn
softkey (n=1 to 6) in step 4.
•Use Kelvin triaxial cable for Kelvin connection
•Use triaxial cable for non-Kelvin connection
9. Press the CONTINUE softkey to execute the Leak Test.
After the Leak Test is completed for the input port specified by the IVn softkey,
the test result (PASS or FAIL) is displayed on the 4155/4156 screen. And the
following softkeys are displayed:
Softkey Description
REPORT CRTDisplays the Leak Test results on the 4155/4156 screen.
LEAK TESTReturns to the LEAK TEST MENU.
T o display the leak current measurement d ata on the 4155/4 156 sc reen, pr ess the
REPORT CRT so ftkey.
10. To return to the LEAK TEST MENU, select the LEAK TEST softkey. And to
perform the Leak Test for another input port, repeat steps 3 to 9.
Agilent E5250A User’s Guide, Edition 93-19
Executing Self-Test and Leak Test
Using the Self-Test Utility
3-20Agilent E5250A User’s Guide, Edition 9
4Setting up Measurement Environment
Setting up Measurement Environment
This chapter explains how to connect your instruments to Agilent E5250A input,
and how to connect the E5250A output to your wafer prober or test fixture.
You can connect the E5250A output (E5252A or E5255A) to your prober or fixture
in the following three ways:
•Via the available connector plates.
•(For E5255A only) Directly using Agilent 16494E coaxial cable.
•Via connectors that you mount directly on your own connector plate or test
fixture.
This chapter explains how to do this in the following sections:
•“Connector Plates” introduces connector plates available for the plug-in card.
For installing the connector plates on your shielding box for prober or fixture,
refer to Agilent 16495 Installation Guide.
•“Connecting 8-Channel Shielded Coaxial Cable” explains how to use the
16494E coaxial cable, which is used to directly connect the E5255A output to
the DUT, DUT socket, or any fixture you want without using a connector plate.
•“Mounting Connectors Directly” describes connector hole dimensions for
mounting connectors directly on your own connector plate or test fixture, and
describes the information to connecting the connectors to prober, sock et or DUT.
•“Connecting the E5250A Input” explains how to connect instruments to the
E5250A input.
•“Connecting the E5250A Output” describes the information to connect the
plug-in card output to a connector plate, connectors on your own shielding box
and so on.
•“Measurement Cable Length” is a reference for deciding total cable length at
your site by calculating the total guard capacitance according to the cable length
from instrument to Device Under Test (DUT).
4-2Agilent E5250A User’s Guide, Edition 9
Setting up Measurement Environment
Connector Plates
Connector Plates
A connector plate is the interface to connect the E5250A to your prober or test
fixture. The following connector plates are available for the E5250A. To install the
connector plate, refer to Agilent 16495 Installation Guide.
Connector Plates for the E5252A
16495F Half size connector plate for E5252A
16495F-001 has 12 triaxial through connectors (female to
female), an Intlk connector, and a GNDU connector (triaxial
through, female to female). The back of the Intlk connector is
designed for soldering.
16495F-002 has 12 triaxial connectors, an Intlk connector, and
a GNDU connector. The back of each co nnector is designed for
soldering.
16495G Full size connector plate for E5252A
16495G-001 has 24 triaxial through connectors (female to
female), an Intlk connector, and a GNDU connector (triaxial
through, female to female). The back of the Intlk connector is
designed for soldering.
16495G-002 has 24 triaxial connectors, an Intlk connector, and
a GNDU connector. The back of each co nnector is designed for
soldering.
Connector Plates for E5255A
16495C Half size connector plate for E5255A
The 16495C has six 8-channel shielded coaxial connectors, an
Intlk connector, and a GNDU connector. The back of each
connector is designed for soldering.
16495D Full size connector plate for E5255A
The 16495D has twelve 8-channel shielded coaxial connectors,
an Intlk connector, and a GNDU connector. The back of each
connector is designed for soldering.
Agilent E5250A User’s Guide, Edition 94-3
Setting up Measurement Environment
Connector Plates
Blank Plate
16495E Half size blank plate.
This plate is used to cover openings when you made too big
openings for mounting the connector plate. You will use this
plate to cover openings if you mount the half size connector
plate in openings made for the full size connector plate.
4-4Agilent E5250A User’s Guide, Edition 9
Setting up Measurement Environment
Total 8 ea.
DUT
Shielding (Ground)
A
B
SMU
Input
BIAS
B
A
Force
Bias line
Guard
Bias line
Ground
Connecting 8-Channel Shielded Coaxial Cable
Connecting 8-Channel Shielded Coaxial Cable
This section introduces Agilent 16494E wide temperature 8-channel shielded
coaxial cable, and explains how to connect the 16494E to the Device Under Test
(DUT).
WARNINGDo not connect the 16494E cable to the E5255A or any instrument until
connecting to the DUT as described below. If you ignore this warning, you
maybe exposed to dangerous voltage.
The 16494E coaxial cable is used to connect the E5255A output directly to the DUT,
DUT socket, or any fixture yo u want without using the con nector plate. One 16 494E
is used for one 8-channel output connector of the E5255A. To use 24 output
channels (three 8-channel output connectors), you need three 16494Es.
Instrument SideThe instrument side of the 16494E cable can be connected directly to the E5255A
output.
DUT SideSolder the open side of th e 16494E directl y to DUT, DUT socket, or your t est fixtu re
as shown in Figure 4-1.
Figure 4-1Example of DUT Connection
NOTEFor highly accurate current forcing and measurements with minimized leakage,
extend the guard to surround the signal lines as close as possible to the DUT.
Agilent E5250A User’s Guide, Edition 94-5
Setting up Measurement Environment
Mounting Connectors Directly
Mounting Connectors Directly
Previous sections described the information to use the available accessories , also the
16495 Installation Guide provides the information how to install connector plates
on the shielding box. However, you may need to directly mount connectors on your
own connector plate or test fixture.
In this case, you need to make connector holes and directly mount the connectors
yourself. This section has dimension information about the hole s y ou n eed t o m ake.
Do the following:
1. Select the appropriate parts for your situation. Refer to Table 4-1.
2. Make holes and mount the connectors. Refer to Table 4-2.
For kelvin connection, use the 16494B Kelvin cable. Kelvin cable requires the
kelvin triaxial connector which has two connector holes and three screw holes.
3. Make interlock circuit as shown in “To Make Interlock Circuit” on page 4-8.
4. Make cable connection between connector and DUT (device under test). Refer
to “To Connect Connectors to DUT” on page 4-13.
Table 4-1 lists the parts required to mount connectors on your shielding box. From
this table, select the parts that you need to mount. Quantity of parts depends on your
site.
4-6Agilent E5250A User’s Guide, Edition 9
Table 4-1Recommended Parts
2 − ∅11.3∅3.22 − M3 x 0.5
14111114
2.8
10.3
∅11.3
10.3
63.5
59
12
6
4
−
R1.5
2
− ∅
3
10
o
1.8
8.2
5
.
1
Setting up Measurement Environment
Mounting Connectors Directly
Usage
Making interlock
circuit
Agilent Part
No.
1252-1419CInterlock Connector (6 pin, female)
3101-0302 or
Switch
3101-3241
1450-0641LED (V
F
8150-5680Wire
Connecting
E5252A output
Connecting
E5255A output
1250-2457Triaxial Connector (female)
8121-119 1Low Noise Coaxial Cable
1251-2367Connector Frame (female)
1251-0179Coaxial Connector (female) (8ea. required for one frame.)
0380-3070Screw (2ea. required for one frame.)
2260-0002Nut (2ea. required for one frame.)
2190-0913Washer (2ea. required for one frame.)
8121-119 1Low Noise Coaxial Cable
Table 4-2Dimen sions of Connector Holes
Kelvin Triaxial Connector (in mm)Triaxial Connector (in mm)
Description
≅ 2.1 V @ IF = 10 mA)
8 Channel Shielded Coaxial Connector (in mm)Interlock Connector (in mm)
Agilent E5250A User’s Guide, Edition 94-7
Setting up Measurement Environment
Mounting Connectors Directly
To Make Interlock Circuit
The interlock circuit is to prevent electric shock when touching measurement
terminals.
You must install an interlock circui t on sh iel d i ng bo x t o prev ent dangerous voltages
when door of the shielding box is open.
Figure 4-2 shows the pin assignments of the interlock connector mounted on your
own connector plate or test fixture.
Figure 4-2Interlock Connector Pin Assignments
WARNINGDangerous voltages of up to the maximum voltage of SMUs may be present at
force, guard, and sense terminals when the interlock terminals are shorted.
To Install Interlock Circuit
Install the interlock circuit as follows.
1. Mount two mechanical switches on your shielding box, so that the switches
close when the door of the shielding box is closed, and open when the door is
opened. For the dimensions of the switch, refer to Figure 4-3 and Figure 4-4.
2. Use wire to connect the two switches in series between pin number 1 and 2 (or
3) of the interlock co nnector. See Figure 4-2.
For example, Agilent 4155/4156 is connected to the interlock circuit, it cannot force
more than ±40 V when the doo r is open. When door is closed, it can force more than
±40 V.
4-8Agilent E5250A User’s Guide, Edition 9
Setting up Measurement Environment
Mounting Connectors Directly
Figure 4-3Dimensions of Interlock Switch (Agilent part number 3101-0302)
Figure 4-4Dimensions of Interlock Switch (Agilent part number 3101-3241)
Agilent E5250A User’s Guide, Edition 94-9
Setting up Measurement Environment
Mounting Connectors Directly
To Install LED Circuit
The 4155/4156 semiconductor parameter analyzer's Intlk connector provides the
interlock signal and a LED drive signal.
If a LED is connected between pin 4 and pin 5 (or 6) of the interlock connector, the
LED lights to indicate high voltage output when more than ±40 V is forced from an
SMU in the 4155/4156.
To install LED circuit on your shield in g box, do followin g:
1. Mount LED on your shielding box. For the dimensions of the LED, refer to
Figure 4-5.
2. Use wire to connect the LED between pin 4 and pin 5 (or 6) of the interlock
connector. Refer to Figure 4-2.
Figure 4-5Dimensions of LED (Agilent part number 1450-0641)
4-10Agilent E5250A User’s Guide, Edition 9
Setting up Measurement Environment
Mounting Connectors Directly
To Connect Interlock Circuit to Instrument
Before beginning the measurement, connect the interlock circuit to the interlock
connector of the instrument as follows.
•For Agilent 4155/4156 semiconductor parameter analyzer:
Connect Intlk connector of the 4155/4156 to the interlock connector by using
Agilent 16493J Interlock cable that is furnished with the 4155/4156. You can
connect it directly without using any adapter.
•For the instruments which has a BNC-Type interlock connector:
If you use an instrument that has a BNC-type interlock connector (such as
Agilent 4142B), connect interlock connectors of instrument and the interlock
circuit as follows:
Required Parts:
Agilent
Model No.
16435A1Interlock Cable Adapter
16493J1Interlock Cable
Procedure:
1. Connect interlock connector of instrument to the BNC connector of the
16435A (interlock cable adapter) by using the BNC cable that is furnished
with the adapter.
2. Connect the 16435A adapter to the interlock connector by using the 16493J
Interlock cable.
Quantity Description
Agilent E5250A User’s Guide, Edition 94-11
Setting up Measurement Environment
Mounting Connectors Directly
NOTETo Check Interlock Circuit
If you use the 4155/4156, you can easily check the interlock circuit as follows:
1. Connect the Intlk connector of the 4155/4156 to your interlock circuit.
2. Press
System front-panel key, then select CALIB/DIAG primary softkey to
display the SYSTEM: SELF-CALIBRATION/DIAGNOSTICS page.
3. In the CALIB/DIAG field, select DIAG secondary softkey.
4. In the CATEGORY field, select I/O PERIPH secondary softkey.
5. Move pointer to the 403 (INT.) Interlock LED field.
6. Select EXECUTE secondary softkey.
7. Confirm the following:
•LED turns on within 1 sec from when interlock circuit is shorted.
•LED turns off within 1 sec from when interlock circuit is open.
To stop the interlock test, select STOP secondary softkey.
4-12Agilent E5250A User’s Guide, Edition 9
Setting up Measurement Environment
to
DUT
Coaxial cable
Triaxial
connector
Insulator
Plate
Guard
Sense
Common
Guard
Force
Common
to
DUT
Coaxial cable
Triaxial
connector
Insulator
Plate
Guard
Sense
Common
Wi re
Guard
Force
Common
Mounting Connectors Directly
To Connect Connectors to DUT
Following figure shows a connection example from the connector to DUT. Use
low-noise coaxial cable (Agilent part number: 8121-1191).
Kelvin connectionsnon-Kelvin connections
This connection is available only for the
Kelvin connectors. This connection can cancel
effects of cable resistance by connect sense
line and force line as close as possible to DUT
terminal.
Following connection is for the Kelvin
connectors. For the triaxial connectors or the
8 ch shielded connectors, ignore SENSE
terminal, and make connection only for
FORCE terminal. Measurement results
include residual resistance from the
connection cable.
CAUTIONNever connect the guard terminal to any output, including circuit common, chassis
ground, or the guard terminal of any other unit. Doing so may result in an
emergency condition.
Agilent E5250A User’s Guide, Edition 94-13
Setting up Measurement Environment
Insulator (black)
Outer conductor
(for Guard signal)
Insulator (clear)
Conductive layer (black)
Center conductor
(for Force/Sense signal)
approx. 10 to 15 mm
min 2 mm
Cover here using sleeve
Mounting Connectors Directly
For highly accurate current forcing and measurements while minimizing leakage,
surround all force and sense lines from SMU by a guard as far as possi ble, and make
cables stable by taping.
For reducing capacitance measurement error, tape the cables to any grounded
materials such as sh ielding box . If you use probe card, ground ing the p robe card al so
reduces capacitance measurement error.
NOTELow-Noise Coaxial Cable
When you make a SMU output connection by using low-noise coaxial cable
(Agilent part number: 8121-1191), shave the Conductive layer and the Insulator
(clear) by knife so that the Center Conductor is insulated from the Conductive layer.
Refer to following figure.
4-14Agilent E5250A User’s Guide, Edition 9
Setting up Measurement Environment
Connecting the E5250A Input
Connecting the E5250A Input
This section explains how to connect instruments to the E5250A.
•“To Connect Instruments to Input Connectors”
•“To Connect Bias Source to the E5255A Bias Input”
•“To Connect GNDU to the E5250A Input”
WARNINGTurn off all instruments that will be connected to the E5250A. And do not turn
them on until the connection described in this section is completed. If you
ignore this warning, you may be exposed to dan gerous voltage.
To Connect Instruments to Input Connectors
The E5250A input connector has six SMU input connectors and four AUX input
connectors.
Table 4-3 shows cables required to connect the instrument for typical applications
using the E5250A.
You can make Kelvin connections when using the 4156 HRSMU, 41501 HPSMU,
4142B SMU, E5270 SMU. Use cables as follows:
To connect Kelvin triaxial cable to the E5250A input, use pairs (couple ports) of
SMU INPUT connectors (1&2, 3&4, or 5&6) of the E5250A.
The input signal for each Kelvin input is as follows:
•Odd input port number: FORCE
•Even input port number: SENSE
Agilent E5250A User’s Guide, Edition 94-15
Setting up Measurement Environment
Connecting the E5250A Input
Table 4-3E5250A Input Connections
Application
DC
Measurement
Instrument
Agilent
Model No.
4155MPSMUSMU1 to 616494A-003 Triaxial cable
4156HRSMUSMU1 to 616494B-003 Kelvin Triaxial cable or
41501HPSMUSMU1 to 616494B-003 Kelvin Triaxial cable or
4142BSMUSMU1 to 616494C-001 Kelvin Triaxial cable or
Output
Connector
VSUAUX1 to 4BNC cable
VMUAUX1 to 4BNC cable
Intlk−16493J Interlock cable
VSUAUX1 to 4BNC cable
VMUAUX1 to 4BNC cable
Intlk−16493J Interlock cable
MPSMUSMU1 to 616494A-003 Triaxial cable
PGUHF1 or 2BNC cable
GNDUSMU1 to 616493H GNDU cable
VSUAUX1 to 4BNC cable
VMUAUX1 to 4BNC cable
GNDUSMU1 to 616493H GNDU cable
Intlk−16435A Interlock cable adapter
E5250A
Input
Connector
(for direct connection to Connector Plate)
16494A-003 Triaxial cable
(for direct connection to Connector Plate)
16494A-003 Triaxial cable
and E5250-60004 Adapter
−16493H GNDU cable
(for direct connection to Connector Plate)
16494A-001 Triaxial cable
and E5250-60044 Adapter
−16493H GNDU cable
(for direct connection to Connector Plate)
and 16493J Interlock cable
(for direct connection to Connector Plate)
Cable Required
4-16Agilent E5250A User’s Guide, Edition 9
Setting up Measurement Environment
Agilent 4156
SMU
AUX
Connector Plate
Agilent E525 0 A
INTLK
GNDU
Kelvin Triaxial Cable
BNC Cable
Triaxial Cable
INTLK Cable
Connecting the E5250A Input
Application
C
Instrument
Agilent
Model No.
Output
Connector
C MeterHigh, LowCV1 and 2BNC cable
E5250A
Input
Connector
Measurement
Pulse InputPulse
OUTPUTHF 1 or 2BNC cable
Generator
Bias InputPower
OUTPUTAUX1 to 4BNC cable
Supply
In the table above,
•AUX means CV1, CV2, HF1, or HF2 connector.
•The E5250A does not have input connectors for the interlock. Connect directly
from instrument to the connector plate which has an interlock connector.
•Can't use Agilent 41422A/41423A (HCU/HVU) of the 4142B with the E5250A.
•T o connect instrument output that uses banana plug to AUX connector , use Dual
Banana plug to BNC Adapter (Agilent part number 1251-2277).
Figure 4-6 shows an example of connecting the E5250A to the 4156.
Figure 4-6Connection Example with Agilent 4156
Cable Required
Agilent E5250A User’s Guide, Edition 94-17
Setting up Measurement Environment
Connecting the E5250A Input
To Connect Bias Source to the E5255A Bias Input
The E5255A BIAS INPUT connector is a BNC-type connector.
Use BNC cable to connect the bias source to the BIAS INPUT of the E5255A.
If output terminal of the bias source is a banana plug, use a Dual Banana Plug to
BNC Adapter (Agilent part number 1251-2277).
To Connect GNDU to the E5250A Input
Ground Unit (GNDU) of Agilent 4142B Modular DC Source/Monitor or Agilent
41501 Expander (for Agilent 4155/4156) is usually connected to GNDU connector
on the connector plate by using Agilent 16493H GNDU cable. This connection
allows current over 1A.
If you want to connect the GNDU through the E5250A to connector plate, test
fixture or DUT, connect the GNDU to the E5250A input as shown below.
CAUTIONMake sure current to GNDU is less than 1 A if you connect GNDU to the E5250A.
The maximum input current of the E5250A is 1A.
Required Parts:
Agilent
Model/Part No.
16493H1GNDU Cable
E5250-600441Triaxial to Dual Triaxial Adapter
Procedure:
1. Connect the adapter to the E5250A SMU INPUT connector (1&2, 3&4 or 5& 6)
as shown in Figure 4-7.
2. Connect the GNDU cable between the instrument GNDU connector and the
triaxial connector on the adapter.
4-18Agilent E5250A User’s Guide, Edition 9
Quantity Description
Figure 4-7Connection of Triax-Dual Triax Adapter
SMU INPUTAUX
To
SMU
1
23456
Agilent E5250A
From
GND
TRIAX-DUAL TRIAX
Adapter
Agilent 16493H Cable
To
SMU INPUT
To
GNDU
By this connection, input signal for SMU INPUT connectors is as follows:
SMU INPUT Port NumberInput Signal
Setting up Measurement Environment
Connecting the E5250A Input
If you make non-Kelvin connection for the E5250A outputs, use the odd input port
only. Because non-Kelvin connection uses only Force signal for measurement.
Agilent E5250A User’s Guide, Edition 94-19
1, 3, or 5Force
2, 4, or 6Sense
Setting up Measurement Environment
B
A
SMU
Input
AUX
B
A
Force or Sense
Signal line
Guard
Ground
Ground
Connecting the E5250A Output
Connecting the E5250A Output
This section describes h ow to conn ect t he E525 2A an d E5255A output to connector
plate, your test fixture, and so on.
•“To Connect the E5252A Output”
•“To Connect the E5255A Output”
WARNINGTurn off the E5250A and all instruments connected to the E5250A. And do not
turn them on until the connection described in this section is completed. If you
ignore this warning, you may be exposed to dan gerous voltage.
To Connect the E5252A Output
The E5252A output connector is shown in Figure 4-8. The E5250A input signal
(SMU input or AUX input) appears at the E5252A output connector as shown in
Figure 4-8 when input-output switching path is made.
Figure 4-8E5252A Output Connector and Output Signal
You can easily connect the E5252A output connector to the connector plate or your
test fixture by using cables shown in Table 4-4.
For Kelvin connection, use Kelvin triaxial cable listed in Table 4-4. To make a
Kelvin output port (1, 3, 5, 7, 9 and 11), couple two E5252A output po rt s as fo l lo ws:
Kelvin Output PortOutput Port Number
11 (Force) and 2 (Sense)
33 (Force) and 4 (Sense)
55 (Force) and 6 (Sense)
77 (Force) and 8 (Sense)
99 (Force) and 10 (Sense)
1111 (Force) and 12 (Sense)
Figure 4-9 shows the 16494B Kelvin triaxial cable for the E5252A.
Figure 4-916494B Kelvin Triaxial Cable for E5252A
Connecting the E5250A Output
Agilent E5250A User’s Guide, Edition 94-21
Setting up Measurement Environment
B
A
SMU
Input
BIAS
B
A
Force
Bias line
Guard
Bias line
Ground
To
E5255A
To
Connector Plat e
FEMALE
MALE
Connecting the E5250A Output
To Connect the E5255A Output
The E5255A output connector is shown in Figure 4-10. The E5250A SMU INPUT
signal or the E5255A BIAS INPUT signal appears at the E5255A output connector
as shown in Figure 4-10 when input-output switching path is made.
Figure 4-10E5255A Output Connector and Output Signal
You can easily connect the E5255A output to the con nector plate or your tes t fixture
by using the cable shown in Table 4-5 and Figure 4-11. The cable can be connected
to the E5255A output ports labeled as 1−8, 1−8 (9−16), or 1−8 (17−24).
Figure 4-1116494D 8-Channel Shielded Coaxial Cable for E5255A
4-22Agilent E5250A User’s Guide, Edition 9
Setting up Measurement Environment
Measurement Cable Length
Measurement Cable Length
This section describes how to calculate the total guard capacitance when using a
Source Monitor Unit (SMU), such as in Agilent 4155/4156 Semiconductor
Parameter Analyzer, Agilent 4142B Modular DC Source/Monitor, Agilent E5270
Series of Parametric Measurement Solution, and so on.
When using an SMU, the length of measurement cables is limited by the guard
capacitance of the cables. Guard capacitance means the capacitance between the
signal line (FORCE or SENSE) and the GUARD line.
T ab le 4-6 lists the guard capacitance for each element of the E5250A measurement
environment.
When using the 4155/4156/4142B/E5270, the maximum limit of the guard
capacitance is approximately 900 pF. So, you add the following and total must be
less than 900 pF:
•Guard capacitance of cable from SMU to E5250A SMU INPUT.
•Guard capacitance of E5250A and installed plug-in cards.
•Guard capacitance of cable from plug-in card output to connector plate.
•Guard capacitance of cable from connector plate to DUT.
•Other capacitances, such as for probe card.
The following is an example to calculate the guard capacitance of a measurement
This chapter explains how to control Agilent E5250A.
•“Methods for Controlling the E5250A” introduces several methods for
controlling the E5250A.
•“Switch Control Basics” introdu ces bas ic infor mati on for contro l lin g the relay
switches on the E5252A and E5255A plug-in cards.
•“Using Virtual Front Panel Utility” describes how to use Virtual Fron t Panel
(VFP) utility furnished with the E5250A. The VFP is for Agilent 4155/4156
users.
5-2Agilent E5250A User’s Guide, Edition 9
Methods for Controlling the E5250A
This section introduces several methods for controlling the E5250A. Table 5-1 lists
the E5250A control methods, and the operating systems on which each control
method can be used.
Table 5-1Summary of Control Methods
Operating System (Controller)
Controlling the E5250A
Methods for Controlling the E5250A
Control
Method
Creating Your
Program
Modifying
Sample
Program
Using VFPNoNoNoYespage 5-25
HP BASIC
(workstation)
YesYesYesYesChapter 6
Y esNoNoNoChapter 6,
RMB/UX
(workstation)
BASIC for
Windows
(PC)
IBASIC
(4155/4156)
Refer to
and 7
7, and 8
Creating Your Own Program
You can create a control program using the SCPI Commands describ ed in Chapt er 7
on all the operating systems listed in T able 5-1. Refer to Chapter 6 for how to create
the program.
In an control progr am ( which runs on HP BASIC opera ting s ystem), you can us e the
Capacitance Compensation Routine and the VFP Data Upload Library furnished
with the E5250A.
Capacitance Compensation Routine
This routine is an subprogram runs on HP BASIC, and is used to compensate f or the
capacitance measurement error caused by the E5250A when the E5252A is
installed. This subprogram is used with Agilent 4284A Precision LCR Meter. You
enter the capacitance data measured by the 4284A, and this subprogram returns the
compensated value.
To use this subprogram, link it to your program. For details, refer to Chapter 6.
Agilent E5250A User’s Guide, Edition 95-3
Controlling the E5250A
Methods for Controlling the E5250A
VFP Data Upload Library
This subprogram library is used to convert the control data created by the Virtual
Front Panel (VFP) Utility to data that can be used with your program runs on HP
BASIC. To use this library, link it to your program. For details, refer to Chapter 6.
CAUTIONOpen or close the relay without allowing the output voltage to affect the relay by
using the following procedure.
1. Disable t he output voltage of bia s source
2. Connect the relay
3. Enable the output voltage of bias source
4. Make measurements
5. Disable t he output voltage of bia s source
6. Disconnect the relay
If the relay is opened and output voltage reaches the relay, the relay may be
damaged or its service life reduced.
Modifying a Sample Program
The sample programs furnished with the E5250A are useful for creating your
program on HP BASI C. You can create your desired program easily by modi f ying a
sample program. To understand the sample programs and how to modify, refer to
Chapter 9.
Also, you can use the Capacitance Compensation Routine and the VFP Data Upload
Library introduced in “Creating Your Own Program” on page 5-3.
Using the Virtual Front Panel (VFP) Utility
The Virtual Front Pa nel (VFP ) Utility is used to control the E5250A by using
Agilent 4155/4156 Semi conductor Paramet er Analyzer' s bu ilt- in Inst rument BAS IC
(IBASIC) controller. You can control the E5250A interactively without the control
program.
Refer to “Using Virtual Front Panel Utility” on page 5-25.
5-4Agilent E5250A User’s Guide, Edition 9
Controlling the E5250A
Switch Control Basics
Switch Control Basics
This section introduces basic information for controlling the relay switches on the
E5252A and the E5255A.
•“Channel List and Channel Configuration”
•“Connection Rule”
•“Connection Sequence”
•“Bias Mode”
•“Couple Port”
•“E5252A Channel List”
•“E5255A Channel List”
Agilent E5250A User’s Guide, Edition 95-5
Controlling the E5250A
Channel No.(4 digits)
Card No.(1digit)
@
Channel No.(4 digits)
Card No.(1digit)
(
:
,
)
Channel No.
(4 digits)
Card No.
(1digit)
@
Channel No.
(4 digits)
Card No.
(1digit)
(
:
,
)
Switch Control Basics
Channel List and Channel Configuration
Some SCPI commands require the channel_list parameter, which determines the
relays to close/open in the E5252A or the E5255A to connect/disconnect the d esired
channels. The following figure shows the basic syntax of th e channel_list parameter.
Figure 5-1Syntax of channel_list Parameter
The channel_list depends on Channel Configuration mode as described later. In the
channel_list, you can specify one or more channels.
Each specified channel consists of five digits as follows:
Card Number: One digit decimal number: 0 to 4.
Specifies the slot number of the card that you want to control.
Depends on Channel Configuration mode. Refer to Table 5-2.
Channel Number: Four digit decimal number:
First two digits: specifies input port.
Last two digits: specifies output port.
Depends on type of card and Channel Configuration mode.
Refer to “E5252A Channel List” on page 5-15 and “E5255A
Channel List” on page 5- 19.
5-6Agilent E5250A User’s Guide, Edition 9
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.