The issue date for the X Series Operator's Guide (REF 9650-001355-01 Rev. H) is May, 2014.
0123
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA USA
01824-4105
ZOLL International Holding B.V.
Newtonweg 18
6662 PV ELST
The Netherlands
If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to det ermine if
additional product information updates are available.
Canada, Industry Canada (IC) Notices .....................................................................A-36
Appendix BAccessories
xwww.zoll.com9650- 001355-01 Rev. H
Product Description
The ZOLL® X Series® unit is an easy-to-use portable defibrillator that combines defibrillati on
and external pacing with the following monitoring capabilities: ECG, CO-Oximeter, Noninvasive Blood Pressure, IBP, CO
resuscitation situations and its rugged , compact, lightweight design makes it ideal for transport
situations. It is powered by auxiliary power and an easily replaced battery pack that is quickly
recharged in the device when it is connected to auxiliary power. In addition, the unit’s battery
may be recharged and tested using a ZOLL SurePower™ Battery Charger Station.
Note:The X Series has defibrillation and pacing functionality, but some of the monitoring
functions are optional features. See the complete list of options in Fig. 1-1. All features
are included in this manual, but only purchased features will be available on your unit.
The product is designed for use in both the hospital and the rugged EMS environment. The
device is a versatile automated external defibrillator with manual capabilities and may be
configured to operate in Manual, Advisory or Semiautomatic modes. It can be configured to
start up in Semiautomatic (AED) mode or manual mode.
When operating in manual configuration, the device operates as a conventional defibrillator
where the device’ s char ging and disc harging i s fully controlled by the operator. In Advisory and
AED modes, some features of the device are automated and a sophisticated detection algorithm
is used to identify ventricular fibrillation and determine the appropriateness of defibrillator
shock delivery. Units may be configured to automatically charge, analyze, recharge, and
prompt the operator to “PRESS SHOCK”, depending on local protocols. The unit is switched
from AED mode to Manual mode for ACLS use by pressing the appropriate key on the front
panel.
Chapter 1
General Information
, Temperature, and Respiration. It has been designed for all
2
9650-001355-01 Rev. HX Series Operator’s Guide1-1
CHAPTER 1GENERAL INFORMATION
The X Series unit assists caregivers during cardiopulmonary resusc itatio n (CPR) by evaluat ing
the rate and depth of chest compressions and providing feedback to the rescuer.
Real CPR Help® requires the use of ZOLL OneStep™ CPR electrodes, OneStep™ Complete
electrodes, or CPR-D-padz
adaptively filtered, using the See-Thru C PR
compressions.
The unit has a large colorful LCD display of numerics and waveform data that provides easy
visibility from across the room and at any angle. ECG, plethysmograph, and respi ration
waveform traces can be displayed simultaneously, giving easy access to all patient monitoring
data at once. The display screen is configurable, so you can choose the best visual layout to fit
your monitoring needs. The X Series includes a transcutaneous pacemaker consisting of a pulse
generator and ECG sensing circuitry. Pacing supports both demand and fixed noninvasive
pacing for adult patients and adolescent, child, and infant pediatric patients.
The X Series has a patient data review and collection system that allow s you to view, store, and
transfer patient data. The X Series unit contains a printer and USB port, which you can use to
print the data and transfer it to a PC.
The X Series unit can send data through a wireless connection to remote locations. The
X Series unit can send 12-lead report snapshots (including trend data) to a recipient via a ZOLL
server. Ful l disclosure case s, which also contai n trend data, can be automatically retrieved from
the X Series unit using ZOLL RescueNet or ePCR software.
®
. When using these pads, the displayed ECG waveforms can be
®
feature, to reduce the artifact caused by chest
X Series Optional Features
The following features are optional in the X Series unit.
Note:All features are included in this manual, but only purchased features will be avai lable
on your unit.
Figure 1-1 X Series Optional Features
Optional Feature
12-Lead ECG with Interpretation
(Masimo®) with SpCO® and SpMet
SpO
2
NIBP (with Smartcuf® and SureBPTM)
(Oridion® Microstream®)
EtCO
2
Temperature (2 Channels)
Invasive Pressures (3 Channels)
Advanced CPR Help
Pacing
Microphone
®
1-2www.zoll.com 9650-001355-01 Rev. H
How to Use This Manual
The X Series Operator's Guide provides information operators need for the safe and effective
use and care of the X Series product. It is important that all persons using this device read and
understand all the information contained within.
Please thoroughly read the safety considerations and warnings section.
Procedures for daily checkout and unit care are located in the Chapter 24, "Maintenance".
Operator ’s Guide Updates
An issue or revision date for this manual is shown on the front cover. If more than three years
have elapsed since this date, contact ZOLL Medical Corporation to determine if additional
product information updates are available.
All users should carefully review each manual update to understand its significance and then
file it in its appropriate section wi thin this manual for subsequent referenc e.
Product documentation is available through the ZOLL website at www.zoll.com. From the
Products menu, choose Product Manuals.
Unpacking
How to Use This Manual
Carefully inspect each container for damage. If the shipping container or cushion material is
damaged, keep it until the contents have been checked for completeness and the instrument has
been checked for mechanical and electrical integrity. If the contents are incomplete, if there is
mechanical damage, or if the defibrillator does n ot pass its electrical self-t est, U.S.A. customers
should call ZOLL Medical Corporation (1-800-348-9011). Customers outside of the U.S.A.
should contact the nearest ZOLL authorized representative. If the shipping container is
damaged, also notify the carrier.
9650-001355-01 Rev. HX Series Operator’s Guide1-3
CHAPTER 1GENERAL INFORMATION
Symbols Used on the Equipment
Any or all of the following symbols may be used in this manual or on this equipment:
SymbolDescription
Class II equipment.
Dangerous voltage.
General warning: Observe and follow all safety signs.
Hot surface.
Fragile, handle with care.
Keep dry.
This end up.
Temperature limitation.
Conformité Européenne Complies with medical device directive 93/42/EEC.
Type B patient connection.
Type BF patient connection.
Type CF patient connection.
Defibrillator-proof type BF patient connection.
Defibrillator-proof type CF patient connection.
1-4www.zoll.com 9650-001355-01 Rev. H
SymbolDescription
2%452.
,I)/.
RECYCLE
,I)/.
Fusible link.
Equipotentiality.
Alternating current (ac).
Direct current (dc).
Auxiliary power adapter operation.
Caution, high voltage.
Symbols Used on the Equipment
Earth (ground).
Negative input terminal.
Positive input terminal.
Power On/Off
Protective earth (ground).
Contains lithium. Recycle or dispose of properly.
Keep away from open flame and high heat.
Do not open, disassemble, or intentionally damage.
9650-001355-01 Rev. HX Series Operator’s Guide1-5
CHAPTER 1GENERAL INFORMATION
SymbolDescription
Do not crush.
Do not discard in trash. Recycle or dispose of properly.
Return to a collection site intended for waste electrical and electronic
equipment (WEEE). Do not dispose of in unsorted trash.
Date of manufacture.
Use by.
Latex-free.
Do not reuse.
Do not fold.
Not sterile.
Manufacturer.
Authorized representative in the European Community.
Serial Number.
Catalogue number.
Consult instructions for use.
Refer to instruction manual/booklet.
1-6www.zoll.com 9650-001355-01 Rev. H
SymbolDescription
MR
Conventions
This guide uses the following conventions:
Within text, the names and labels for physical buttons and softkeys appear in boldface type (for
example, “Press the Charge button or press the Pacer button”).
This guide uses uppercase italics for audible prompts and for text messages displayed on the
screen (for example, LEAD FAULT).
Conventions
Prescription only.
Battery charging status.
Do not use device, cables, or probes in an MRI environment.
War ning !Warning statements alert you to conditions or actions that can result in personal injury
or death.
CautionCaution statements alert you to conditions or actions that can result in damage to the unit.
X Series Indications for Use
The X Series is intended for use by trained medical personnel who are familiar with basic
monitoring, vital sign assessment, emergency cardiac care, and the use of the X Series.
The X Series is also intended for use by (or on the order of) physicians at the scene of an
emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other
similar areas of a hospital. The usage may be in an ambulance or at the scene of an emergency.
It is also intended to be used during the transport of patients. The X Series will be used
primarily on patients experiencing symptoms of cardiac arrest or in post trauma situation. It
may also be used whenever it is required to monitor any of th ose functions that are included (as
options) in the device. The X Series unit can be used on pediatric patients (as described in the
following table) and on adult patients (21 years of age or older) with and without heart
dysfunction.
9650-001355-01 Rev. HX Series Operator’s Guide1-7
CHAPTER 1GENERAL INFORMATION
Pediatric Patient SubpopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age.
Infant1 month to 2 years of age.
Child2 to 12 years of age.
Adolescent12 to 21 years of age.
When the pediatric patient is less than 8 years of age or weighs less than 55 lbs. (25 kg.), use
ZOLL Pedi-padz
®
pediatric defibrillation electrodes. Do not delay therapy to determine the
patient’s exact age or weight. AED mode and ALS rescue protocol are not intended for use on
neonatal patients.
Manual Defibrillation
Use of the X Series in the manual mode for external and internal defibrillation is indicated on
victims of cardiac arrest where there is apparent lack of circulation as indicated by:
• Unconsciousness.
• Absence of breathing.
• Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular
fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable
of producing hemodynamically significant heart beats.
The unit can also be used for synchronized cardioversion of certain atrial or ventricular
arrhythmias. Qualified medical personnel must decide when synchronized cardioversion is
appropriate.
The patient population will range from newborn (neonate) to adult.
Semiautomatic Operation (AED)
X Series products are designed for use by emergency care personnel who have completed
training and certification requirements applicable to the use of a defibrillator where the device
operator controls delivery of shocks to the patient.
They are specifically designed for use in early defibrillation programs where the delivery of a
defibrillator shock during resuscitation involving CPR, transportation, and definitive care are
incorporated into a medically-approved patient care protocol.
Use of the X Series in the Semiautomatic mode for defibrillation is indicated on victims of
cardiac arrest where there is apparent lack of circulation as indicated by:
• Unconsciousness.
• Absence of breathing.
• Absence of pulse.
1-8www.zoll.com 9650-001355-01 Rev. H
Specifications for the ECG rhythm analysis functi on are provided in the section “ECG Analysis
Algorithm Accuracy” on page A-35.
When the patient is less than 8 years of age or weighs less that 55 lbs. (25 Kg), you must use
ZOLL pediatric defibrillation electrodes. Do not de lay t herapy to det ermine pat ient’s exact age
or weight.
ECG Monitoring
The X Series is intended for use to monitor and/or record 3-, 5-, or 12-lead ECG waveform and
heart rate, and to alarm when heart rate is above or below limits set by the operator. The patient
population will range from newborn (neonate) to adult, with and without heart dysfunction.
CPR Monitoring
The CPR monitoring function provides visual and audio feedback designed to encourage
rescuers to perform chest compressions at the AHA/ERC recommended rate of 100
compressions per minute. Voice and visual prompts encourage a minimum compression depth
of at least 1.5 (3.8 cm) or 2.0 inches (5.0 cm), depending on the configuration, for adult
patients. The CPR monitoring function is not intended for use on patients under 8 years of age.
X Series Indications for Use
External Transcutaneous Pacing
This product can be used for temporary external cardiac pacing in conscious or unconscious
patients as an alternative to endocardial stimulation.
The purposes of pacing include:
• Resuscitation from standstill or bradycardia of any etiology:
• As a standby when standstill or bradycardia might be expected:
• Suppression of tachycardia.
• Pediatric pacing.
Non-Invasive Blood Pressure Monitoring
The X Series is intended for use to make non-invasive measurements of arterial pressure and
heart rate, and to alarm if either parameter is outside of the limits set by the user. Measurements
are made using an inflatable cuff on the patient's arm or leg. The patient population will range
from newborn (neonate) to adult.
Temperature Monitoring
The X Series is intended for use to make continuous temperature measurements of rectal,
esophageal, or surface temperatures, and to alarm if the temperature is outside of the limits set
by the user. The patient population will range from newborn (neonate) to adult.
9650-001355-01 Rev. HX Series Operator’s Guide1-9
CHAPTER 1GENERAL INFORMATION
SpO2 Monitoring
The X Series pulse CO-oximeter, with Masimo Rainbow® SET technology and the Rainbow
series of sensors, is intended for use for continuous noninvasive monitoring of functional
oxygen saturation of arterial hemoglobin (SpO
(SpCO), and/or methemoglobin saturation (SpMet). The pulse CO-oximeter and accessories
are indicated for use on adult, pediatric, and neonatal patients during both no motion and
motion conditions, and for patients who are well or poorly perfused, in hospitals, hospital-type
facilities, or in mobile environments.
Respiration Monitoring
The X Series is intended for use to continuously monitor respiration ra te and to alarm if the rate
falls outside of the range set by the operator. Because the measurement method actually
measures respiratory effort, apnea episodes with continued respiratory effort (such as
obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The
patient population will range from newborn (neonate) to adult.
CO2 Monitoring
), pulse rate, carboxyhemoglobin saturation
2
The X Series is intended for use to make cont inuo us nonin vasive measurem ent an d moni tori ng
of carbon dioxide concentration of the expired and inspired breath and breath rate. The patient
population will range from newborn (neonate) to adult.
Invasive Pressure Monitoring
The X Series is intended for use to display and make continuous invasive pressure
measurements from any compatible pressure transducer. The primary intended uses are arterial
blood pressure, central venous pressure and intracranial pressure monitoring. Any contraindications of the particular transducer selected by the user shall apply. The patient population
will range from newborn (neonate) to adult.
12-Lead Analysis
The 12-lead ECG Analysis is intended for use in acquiring, analyzing and reporting ECG data,
and to provide interpretation of the data for consideration by caregivers. The interpretations of
ECG data offered by the device are only significant when used in conjunction with caregiver
overread as well as consideration of all other relevant patient data. The 12-lead ECG Analysis
is intended for use on adults (> 18 years of age).
1-10www.zoll.com 9650-001355-01 Rev. H
X Series Product Functions
Defibrillator Function
The X Series contains a direct current (dc) defibrillator capabl e of delivering up to 200 jo ules. It
may be used in synchronized mode to perform synchronized cardioversion using the patient’s
R-wave as a timing reference. The unit uses paddles or disposable, pregelled electrodes for
defibrillation.
Defibrillator Output Energy
X Series defibrillators can deliver biphasic energy from 1 joule to 200 joules. The energy
delivered through the chest wall, however, is determined by the patient’s transthoracic
impedance. An adequate amount of e lectrolyte gel mu st be applied t o the paddles and a force of
10 to 12 kilograms (22 to 26.4 pounds) must be applied to each paddle in order to minimize this
impedance. If hands-free therapy electrodes are used, make sure that they are properly applied.
(Refer to the instructions on the electrode package).
X Series Product Functions
External Pacemaker
X Series defibrillators include a transcutaneous pacemaker consist in g of a pulse generator and
ECG-sensing circuitry. Noninvasive transcutaneous pacing (NTP) is an established and proven
technique. This therapy is easily and rapidly applied in both emergency and nonemergency
situations when temporary cardiac stimulation is indicated.
The output current of the pacemaker is continuously variable from 10 to 140 mA (the output
current is 0 mA when paused). The rate is continuously variable from 30 to 180 pulses per
minute (ppm), by increments of 5 ppm (10 bpm when greater than 100 ppm).
The pacing output pulse is delivered to the heart via ZOLL hands-free defibrillation/pacing
electrodes placed on the patient’s back and the precordium.
Proper operation of the device, together with correct electrode placeme nt, is critical to
obtaining optimal results. Every operator must be thoroughly familiar with these operating
instructions.
ECG Monitoring
The patient’s ECG is monitored by connecting the patient to the unit via a 3-, 5-, or 12-lead
patient cable or hands-free therapy electrodes. The ECG waveform is presented on the display
along with the following information:
• averaged heart rate, derived by measuring R to R intervals
• lead selection - I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (with ECG cable),
9650-001355-01 Rev. HX Series Operator’s Guide1-11
CHAPTER 1GENERAL INFORMATION
Electrodes
The X Series units will defibrillate, cardiovert, and monitor ECG using hands-free therapy
electrodes. The X Series unit will pace using ZOLL hands-free therapy electrodes.
Energy Select, Charge and Shock controls are located on the paddles and front panel. When
using hands-free therapy electrodes, you must use the controls on the front panel of the unit. To
switch between paddles and hands-free therapy electrodes, remove the multifunction cable
(MFC or OneStep) from the apex paddle and connect the hands-free therapy electrodes to the
cable.
You should always check the expiration date on the electrode packaging. Do not use expired
electrodes, which might result in false patient impedance readings and affect the level of
delivered energy, or cause burns.
This symbol on the electrode package is accompanied by the expiration date.
For Stat-padz
lower right corner of the label, below the lot number.
Note:ZOLL electrodes contain no hazardous materials and may be disposed of in general
trash unless contaminated with pathogens. Use appropriate precautions when
disposing of contaminated electrodes.
When the patient is less than 8 years old or weighs less than 55 lb. (25 kg), use ZOLL OneStep
Pediatric defibrillation electrodes. Do not delay therapy to determine the patient’s exact age or
weight.
®
II, this symbol does not appear; the expiration date appears on the
Batteries
X Series models use an easily replaced rechargeable li thium-ion battery pack (the Su r ePower II
Battery Pack). A new, fully charged battery pack typically delivers more than 6 hours of ECG
monitoring. Use of other functions (such as the defibrillator, printer, or pacemaker) reduces this
time.
When a LOW BATTERY icon appears on the display and the unit emits three beeps in
conjunction with the displayed battery icon, the battery must be replaced and recharged.
You can charge the battery by either of the following methods:
• Internal charging — plug the X Series into an auxiliary power adapter to automatically
begin charging the installed battery pack. The front panel battery indicator operates as
follows:
When the indicator is:It means:
Steady yellowBattery is charging.
Steady greenBattery is charged.
Alternating yellow and
green
Not litNo battery in device.
The charge state cannot be
determined or a battery charging
fault has been detected.
Note:Upon power up, it takes approximately 45 seconds for the LEDs on the battery to
accurately display run time.
1-12www.zoll.com 9650-001355-01 Rev. H
• External charging — use the ZOLL SurePower Battery Charger with the X Series battery
adapter to charge the battery pack and test the battery’s capacity. For details, refer to the
SurePower II Battery Pack Guide.
The Recalibration LED icon () lights for approximately 10 seconds (after you press and
release the Display button) if the battery ne eds to be calibrated . If the Recalibration LED lights,
the runtime indicator will not display run time for that battery. For best performance of the
battery, you should recalibrate the battery as soon as possible.
To manually recalibrate the SurePower II Battery Pack, you can insert the battery into the
SurePower Charger Station and perform a Manual Test (for mo re information, see the ZOLL
SurePowerCharger Station Operator’s Guide).
After you recalibrate the battery, the Recalibration LED will only flash when you press the
Display button.
Ready For Use (RFU) Indicator
The X Series has an RFU indicator on the front panel that indicates if the device is ready for
use. The RFU indicator has three states which are described in the following table.
StateDescriptionAction
X Series Product Functions
Ready for UseThe device is ready for use. Patient
monitoring, defibrillation, and
pacing parameters are functional
and the battery is above the low
battery capacity.
Note: If the device is plugged into
the auxiliary power adapter, the
Ready for Use indicator may display
even if the battery is depleted.
Check the status of the battery
before removing the device from the
auxiliary power adapter.
None required.
9650-001355-01 Rev. HX Series Operator’s Guide1-13
CHAPTER 1GENERAL INFORMATION
StateDescriptionAction
FlashingOne or more of the following has
occurred:
• The battery is not properly
installed.
• A low battery is installed.
• A battery fault has occurred.
• There is no battery installed
while connected to auxiliary
power.
• One or more patient monitoring
parameters have failed self-test
(NIBP, SpO
Temp).
• The front panel button self-test
failed.
• The speech database self-test
failed.
Do Not UseOne or more of the following has
occurred:
• The battery is not properly
installed.
• No battery is installed and
auxiliary power is not present.
• A very low battery (below
software shutdown limit) was
installed.
• ECG, defibrillator, or pacer selftests have failed, or other critical
self-tests have failed.
, CO2, IBP, or
2
Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to flash, remove the unit
from service and contact the
appropriate technical personnel or
the ZOLL Technical Service
Department.
Install a fully charged battery in the
unit and check the RFU indicator
again. If the RFU indicator
continues to display the Do Not
Use symbol, remove the unit from
service and contact the appropriate
technical personnel or the ZOLL
Technical Service Department.
1-14www.zoll.com 9650-001355-01 Rev. H
Safety Considerations
All operators should review these safety considerations before using the X Series unit.
X Series units are high-energy defibrillators capable of delivering 200 joules. To completely
deactivate the unit, press the power switch to turn the unit off.
To manually disarm a charged (or charging) defibrillator, do one of the following:
• Press the Disarm quick access key .
• Change the selected energy.
• Press the power switch to turn the unit off.
For safety, the X Series automatically disarms if left charged for more than 60 seconds if the
shock button ( ) is not pressed.
Warnings
General
Federal (U.S.A.) law restricts this defibrillator to sale by or on the order of a physician.
Safety Considerations
Only appropriately trained, skilled personnel who are familiar with equipment operation should
perform emergency defibrillation. The prescribing physician should determine what training,
such as Advanced Cardiac Life Support (ACLS) or Basic Life Support (BLS) certification, is
appropriate.
Only skilled personnel trained in Advanced Cardiac Life Support (ACLS) and who are familiar
with equipment operation should perform synchronized cardioversion. The precise cardiac
arrhythmia must be determined before attempting defi bril lat io n.
These operating instructions describe the functions and proper operation of the X Series
products. They are not a substitute for a formal patient care training course. Operators should
obtain formal training from an appropriate authority before using this defibrillator for patient
care.
Proper operation of the unit and correct electrode placement is critical to obtaining optimal
results. Operators must be thoroughly familiar with proper device operation.
The use of external pacing/defibrillation electrodes, accessories, or adapter devices from
sources other than ZOLL is not recommended. ZOLL makes no representations or warranties
regarding the performance or effectiveness of its products when used with pacing/defibrillation
electrodes or adapter devices from other sources. Defibrillator failures attributable to the use of
pacing/defibrillation electrodes or adapters not manufactured by ZOLL might void ZOLL’s
warranty.
At receipt of shipment, check pacing/defibrillation electrodes to ensure compatibility.
Allow ample slack in cables to make sure that cables do not tug at electrodes.
Do not disassemble the unit. A shock hazard exists. Refer all problems to authorized service
personnel.
Follow all recommended maintenance instructions. If a problem occurs, obtain service
immediately. Do not use the defibrillator until it has been inspected by appropriate personnel.
9650-001355-01 Rev. HX Series Operator’s Guide1-15
CHAPTER 1GENERAL INFORMATION
The X Series unit might not perform to specifications when stored at the upper or lower
extreme limits of storage temperature and then immediately put into use. The X Series unit
should not be stored or used outside of the environmental limits p rovided in Appendix A of this
manual.
A void using the X Series adjacent to, or stacked on, other equipment. If unavoidable, verify that
the unit operates normally in this configuration before clinical use.
The X Series unit should be installed and put in to service a ccordi ng to the EM C informati on in
Appendix A of this manual.
Do not use internal paddles while the X Series unit’s auxiliary power source is connected to an
aircraft AC power operating at a frequency of 400 Hz.
The use of accessories, transducers, and cables other than those specified in this manual and
related X Series option manual inserts may result in increased emissions or decreased immunity
of the X Series.
Perform functional test of internal paddles prior to use.
Do not use or place the unit in service if the Ready For Use indicator (at the upper right of the
front panel) displays a red circle with a line through it.
Carefully route patient cabl es to avoid tri pping over the m, or inadverte ntly pulli ng the unit onto
the patient.
Always inspect the unit for damage if it has been dropped.
Only authorized personnel should use the Supervisor menus.
If uncertain about the accuracy of any measurement, first check the patient’s vital signs by
alternate means, and then make sure the monitor is functioning correctly.
ECG Monitoring
Implanted pacemakers might cause the heart rate meter to count the pacemaker rate during
incidents of cardiac arrest or other arrhythmias. Dedicated pacemaker detection circuitry may
not detect all implanted pacemaker spikes. Check the patient's pulse; do not rely solely on heart
rate meters. Patient history and physical examination are important factors in determining the
presence of an implanted pacemaker. Pacemaker patients should be carefully observed. See
“Pacemaker Pulse Rejection:” on page A-15 of this manual for disclosure of the pacemaker
pulse rejection capability of this instrument.
Use only ECG electrodes that meet the AAMI standard for electrode performance
(AAMI EC-12). Use of electrodes not meeting this AAMI standard could cause the ECG tr ace
recovery after defibrillation to be significantly delayed.
Prior to attempting synchronized cardioversion, ensure the ECG signal quality is good and that
sync markers are displayed above each QRS complex.
Do not place electrodes directly over an implanted pacemaker.
The X Series unit detects ECG electrical signals only. It does not detect a pulse (effective
circulatory perfusion). Always verify pulse and heart rate by physical assessment of the patient.
Never assume that the display of a nonzero heart rate means that the pat ient has a pulse.
Excessive artifact can result due to improper skin preparation of th e electrode site s. Follow skin
preparation instructions in Chapter 6: “Monitoring ECG.”
1-16www.zoll.com 9650-001355-01 Rev. H
Equipment such as electrocautery or diathermy equipment, RFID readers, electronic art icle
surveillance (EAS) systems, or metal detectors that emit strong radio frequency signals can
cause electrical interference and distort the ECG signal displayed by the monitor, thereby
preventing accurate rhythm analysis. Ensure adequate separation between such emitters, the
device, and the patient when performing rhythm analysis.
Shock Hazard: Use of accessories, other than those specified in the operating instructions, may
adversely affect patient leakage currents.
Certain line-isolation monitors may cause interference on the ECG display and may inhibit
heart rate alarms.
Monitoring ECG through the paddles may result in inaccurate heart rate display due to artifact.
Defibrillation
The ZOLL X Series can deliver 200 joules o f electrical energy. If this electrical energy is not
discharged properly, as described in the this manual, the electrical energy could cause persona l
injury or death to the operator or bystander.
To avoid possible damage to the X Series unit, turn off pacing before defibrillating the patient
with a second defibrillator.
Warnings
After a synchronized cardioversion, the SYNC mode may be cleared after each shock or
disarm. The user may have to reselect (press) the SYNC button after each synchronized
cardioversion shock performed on a patient. In Defib/Pacer Default sett ings in the Supervisor
Setup menu, the X Series can be configured to remain in the SYNC mode after each
synchronized cardioversion.
Synchronized cardioversion can be performed in the paddle monitoring mode. However, it is
possible that artifact can be produced by the moving paddles, which could cause t he
defibrillator to trigger on the artifact. It is recommended that monitoring in leads I, II or III be
used during synchronized cardioversion. Paddle monitoring should not be used for elective
cardioversions procedures.
To avoid stress to the defibrillator or the tester, never attempt to repeatedly charge and
discharge the defibrillator in rapid succession. If a need for repetitive testing arises, allow a
waiting period of at least 2 minutes after every third discharge.
In the SYNC mode, the defibrillator does not discharge without a command signal (R-w ave
detection) from the ECG monitor indicate d by a SYNC marker on the trace and a flashing
SYNC indicator.
If conductive gel forms a continuous path between the defibrillator electrodes, delivered energy
may be dramatically reduced to zero. In this case, reposition the electrodes to eliminate the
shunting path before attempting additional shocks.
Improper defibrillation technique can cause skin burns. To limit possible skin burns, use only
ZOLL defibrillation gel on paddles, ensure the gel covers the entire paddle surface and press
firmly against patient’s chest.
If a new energy level is selected after the CHARGE button is pushed and while the defibrillator
is charging or charged, the defibrillator will disarm. The CHARGE button will need to be
pressed again to charge to the new energy level.
Prior to defibrillation, disconnect from the patient any medical electronic device that is not
labeled “defibrillation protected.”
9650-001355-01 Rev. HX Series Operator’s Guide1-17
CHAPTER 1GENERAL INFORMATION
Before charging the defibrillator, verify that the energy selected on the display is the desired
output.
Defibrillation takes priority over external pacing. Should the defibrillator be charged during the
administration of external pacing, the pacer turns off and the defibrillator charges to the
selected energy.
Pacing
Ventricular fibrillation does not respond to pacing and requires immediate defibrillation.
Therefore, the patient’s dysrhythmia must be determined immediately, so that you can employ
appropriate therapy . If t he patient is in ventricula r fibrillation and defibrilla tion is successful but
cardiac standstill (asystole) ensues, you should use the pacemaker.
Ventricular or supraventricular tachycardias can be interrupted with pacing, but in an
emergency or during circulatory collapse, synchronized cardioversion is faster and more
certain.
Pulseless electrical activity (PEA) can occur following prolonged cardiac arrest or in other
disease states with myocardial depression. Pacing might then produce ECG responses wi th out
effective mechanical contractions, making other effective treatment necessary.
Pacing can evoke undesirable repetitive resp onses, tachycardia, or fibrillation in the presence of
generalized hypoxia, myocardial ischemia, cardiac drug toxicity, electrolyte imbalance, or
other cardiac diseases.
Pacing by any method tends to inhibit intrinsic rhythmicity. Abrupt cessation of pacing,
particularly at rapid rates, can cause ventricular standstill and should be avoided.
Noninvasive temporary pacing can cause discomfort of varying intensity, which occasionally
can be severe and preclude its continued use in conscious patients.
Similarly, unavoidable skeletal muscle contraction might be troublesome in very sick patients
and might limit continuous use to a few hours. Erythema or hyperemia of the skin under the
hands-free therapy electrodes often occurs; this effect is usually enhanced along the perimeter
of the electrode. This reddening should lessen substantially withi n 72 hours.
There have been reports of burns under the anterior electrode when pacing adult patients with
severely restricted blood flow to the skin. Prolonged pacing should be avoided in these cases
and periodic inspection of the underlying skin is advised.
There are reports of transient inhibition of spontaneous respiration in unconscious patients with
previously available units when the anterior electrode was placed too low on the abdomen.
The pacing rate determination can be adversely affected by artifact. If the patient’s pulse and
the heart rate display are sig nificantly different, external pacing pulses may not be delivered
when required.
Artifact and ECG noise can make R-wave detection unreliable, affecting the HR meter and the
demand mode pacing rate. Always observe the patient closely during pacing operations.
Consider using asynchronous pacing mode if a reliable ECG trace is unob tai nab le .
Transcutaneous pacing should not be used to treat V FIB (ventricular fibrillation). In cases of
V FIB, immediate defibrillation is advised.
1-18www.zoll.com 9650-001355-01 Rev. H
Loading...
+ 298 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.