ZOLL X Series User Manual

X Series™ Service Manual
9650-001356-01 Rev. D
The issue date for the X Series Service Manual (REF 9650-001356-01 Rev. D) is November, 2013.
0123
ZOLL Medical Corporation
269 Mill Road Chelmsford, MA USA 01824-4105
ZOLL International Holding B.V.
Newtonweg 18 6662 PV ELST The Netherlands
If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to det ermine if additional product information updates are available.
Masimo and Rainbow are trademarks or registered trademarks of Masimo Corporation in the United States and/or other countries.

Table of Contents

Preface
Overview................................................................................................................................1
Safety Considerations............................................................................................................2
Additional Reference Material................................................................................................2
Conventions...........................................................................................................................3
Service Policy Warranty.........................................................................................................3
Technical Service...................................................................................................................3
Technical Service for International Customers ......................................................................4
Chapter 1 Maintenance Tests
X Series Overview.................................................................................................................5
Before You Begin the Maintenance Tests ..................................................................... 6
Equipment You Need to Perform the Maintenance Tests ............................................. 7
Accessories You Need for the X Series Options Maintenance Tests ............................ 7
Maintenance Tests.................................................................................................................8
Physical Inspection of the Unit ......................................................................................8
Operational Tests .......................................................................................................... 9
Leads Test ................................................................................................................... 15
Heart Rate Display Test ...............................................................................................16
Heart Rate Alarm Test .................................................................................................17
Power Supply Test .......................................................................................................19
Leakage Current Test .................................................................................................. 21
Shock Test ...................................................................................................................22
Advisory Test ............................................................................................................... 25
Synchronized Cardioversion Test ................................................................................26
Pacer Test ................................................................................................................... 27
Printer Test ..................................................................................................................28
Treatment Report Test .................................................................................................29
SpO2 Test (for SpO SpO2, SpCO, and SpMet Test (for SpO
EtCO2 Calibration (required for annual calibration) ..................................................... 33
EtCO2 Calibration Check .............................................................................................34
EtCO2 Test .................................................................................................................. 35
IBP Test .......................................................................................................................36
NIBP Calibration Check ...............................................................................................37
NIBP Leak Test ............................................................................................................40
Deflation Test ...............................................................................................................41
feature without SpCO and SpMet) ............................................ 30
2
feature with SpCO, or SpCO and SpMet) .. 31
2
9650-001356-01 Rev. D ZOLL X Series Service Manual i
TABLE OF CONTENTS
Inflation Test ................................................................................................................42
NIBP Test .................................................................................................................... 43
Temperature Test ........................................................................................................ 44
Wi-Fi Test (Optional) ....................................................................................................45
Audio Recording Test (Optional) ................................................................................. 47
Ch a p t e r 2 Troubleshooting
Overview..............................................................................................................................49
X Series Error Messages.....................................................................................................49
Chapter 3 Disassembly Procedures
Overview..............................................................................................................................61
Required Equipment............................................................................................................61
Safety Precautions...............................................................................................................62
Modules ............................................................................................................................... 62
Removing the Handle..........................................................................................................63
Removing the Printer Module..............................................................................................64
Disassembling the Main Unit...............................................................................................66
Removing the Front Enclosure Module ...............................................................................70
Removing the Display Module.............................................................................................72
Removing the EtCO2 and NIBP Modules............................................................................ 76
Removing the Monitor Board and SpO2 Modules ...............................................................80
Removing the CP Carrier Module........................................................................................ 86
Removing the Defibrillator Boards and Capacitor Bank Modules........................................88
Removing the AUX Power Connector ................................................................................. 94
Removing the Dock Connector............................................................................................96
Ch a p t e r 4 Replacement Parts
Overview..............................................................................................................................99
Replacement Parts List ................................................................................................99
Pictures and Diagrams .............................................................................................. 104
Ch a p t e r 5 Functional Description
Overview............................................................................................................................121
Circuit Distribution ......................................................................................................121
Monitor Board ............................................................................................................ 121
Defibrillator Pacer Module ......................................................................................... 124
Printer ........................................................................................................................125
Lithium Ion Battery .....................................................................................................126
ii 9650-001356-01 Rev. D
Appendix A
Overview............................................................................................................................127
X Series Electrical Hardware Interconnect Diagram.......................................................... 128
X Series Signal and Power Interconnect Schematic .........................................................129
Appendix B
Overview............................................................................................................................131
NIBP Simulator Performance .....................................................................................131
Fluke Biomedical BP Pump/Bio-Tek Instruments BP Pump ...................................... 132
Fluke Biomedical BP Pump/Bio-Tek Instruments BP Pump2 .................................... 132
Fluke Biomedical BP Pump/Dynatech Nevada Inc. CuffLink ..................................... 133
Clinical Dynamics Smart Arm ....................................................................................133
Maintenance Test Checklist
Overview............................................................................................................................135
9650-001356-01 Rev. D ZOLL X Series Service Manual iii
TABLE OF CONTENTS
iv 9650-001356-01 Rev. D

Overview

Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to review this section thoroughly before attempting to use or service the X Series unit.
Chapter 1—Maintenance Tests describes step-by-step procedures for various maintenance tests.
Chapter 2—Troubleshooting provides a listing of error messages to help the service technician detect faults and repair them.
Chapter 3—Disassembly Procedures describes step-by-step procedures for disassembling modules in the X Series unit.
Chapter 4—Replacement Parts List displays a complete list of ZOLL part numbers for field replaceable parts available for the X Series unit, allowing the service person to identify and order replacement parts from ZOLL.

Preface

The ZOLL® X Series™ Service Manual is intended for the trained biomedical technician whose responsibility is to identify malfunction s and/or make repairs at the subassembly level. The X Series Service Manual has five main sections and two appendixes.
Chapter 5—Functional Description provides technical descriptions for the X Series major subassembly modules.
Appendix A—X Series Electrical Hardware Interconnect Diagram and X Series Signal and Power Interconnect Schematic.
Appendix B—Contains simulators and settings that may be used to assess the performance of the NIBP module.
9650-001356-01 Rev. D ZOLL X Series Service Manual 1
PREFACE

Safety Considerations

The following section describes general warnings and safety considerations for operators and patients. Service technicians should review the safety considerations prior to servicing any equipment and read the manual carefully before attempting to disassemble the unit. Only qualified personnel should service the X Series unit.
Federal (U.S.A.) law restricts this unit for use by or on the order of a physician. Safety and effectiveness data submitted by ZOLL Medical Corporation to the Food and Drug
Administration (FDA) under section 510(K) of the Medical Device Act to obtain approval to market is based upon the use of ZOLL accessories such as disposable electrodes, patient cables and batteries. The use of external pacing/defibrillation electrodes and adapter units from sources other than ZOLL is not recommended. ZOLL makes no representations or warranties regarding the performance or effectiveness of its products when used in conjunction with pacing/defibrillation electrodes and adapter units from other sources. If unit failure is attributable to pacing/defibrill ation e lectrodes or adapter units not manufactured by ZOLL, this may void ZOLL's warranty.
Only qualified personnel should disassemble the X Series uni t.
WARNING! This unit can generate up to 2775 volts with sufficient current to cause lethal shocks.
All persons near the equipment must be warned to “STAND CLEAR” prior to dischargin g the defibrillator.
Do not discharge the unit’s internal defibrillator energy more than three times in one minute or damage to the unit may result.
Do not discharge a battery pack except in a ZOLL SurePower Do not use the X Series in the presence of flammable agents (such as gasoline), oxygen-rich
atmospheres, or flammable anesthetics. Using the unit near the site of a gasoline spill may cause an explosion.
Do not use the unit near or within puddles of water.

Additional Reference Material

In addition to this guide, there is a X Series Operator’s Guide which is a comprehensive reference work that describes all the user tasks needed to operate the X Series.
TM
Charging Station.
2 9650-001356-01 Rev. D

Conventions

Conventions
WARNING! Warning statements describe conditions or actions that can result in personal injury or death.
Caution Caution statements describe conditions or actions that can result in damage to the unit.
Note: Notes contain additional information on using the defibrillator.

Service Policy Warranty

In North America: Consult your purchasing agreement for terms and conditions associated with your warranty. Outside of North America, consult a ZOLL authorized representative.
In order to maintain this warranty, the instructions and procedures contained in this manual must be strictly followed. For additional information, please call the ZOLL Technical Service Department 1-800-348-9011 in North America.

Technical Service

If the ZOLL X Series unit requires service, contact the ZOLL Technical Service Department: Telephone: 1-978-421-9655; 1- 800-348-9011 Fax: 1-978-421-0010 Email: techsupport@zoll.com Have the following information available for the Te chnical Service representative:
Unit serial number.
Description of the problem.
Department where equipment is used.
Sample chart recorder strips or electronic log files documenting the problem, if applicable.
Purchase Order to allow tracking of loan equipment.
Purchase Order for a unit with an expired warranty.
If the unit needs to be sent to ZOLL Medical Corporation, obtain a service request number (SR#) from the T e chn ic al Service representative. Return the unit in its original container to:
ZOLL Medical Corporation 269 Mill Road Chelmsford, Massachusetts 01824-4105
Attn: Technical Service Department (SR#) Telephone: 1-800-348-9011; 1-978-421-9655 FAX: 978-421-0010
9650-001356-01 Rev. D ZOLL X Series Service Manual 3
PREFACE

Technical Service for International Customers

International customers should return the unit in its origin al co nt ain er to the neare s t aut hor ize d ZOLL Medical Corporation Service Center. To locate an authorized service center, contact the International Sales Department at ZOLL Medical at the above address.
4 9650-001356-01 Rev. D

X Series Overview

This chapter includes step-by-step procedures for various maintenance tests. At a minimum, a trained biomedical technician must perform a CO once per year. A CO calibration. It is necessary to perform maintenanc e tests after a module is repaired to ensure that the functions of the X Series unit work properly . Use the checklist at t he back of this document (ZOLL X Series Maintenance Tests Checklist) to record your results of the maintenance tests.
Because the X Series units must be maintaine d ready for immediate use, it is important for users to conduct the Operator’s Shift Checklist procedure at the beginning of every shift. This procedure can be completed in a few minutes and requires no additional test equipment. (See the ZOLL X Series Operator’s Guide for the Operator’s Shift Checklist.)
Chapter 1

Maintenance Test s

calibration and an NIBP calibration check
calibration check may be performed at any time to verify the CO2
2
2
9650-001356-01 Rev. D ZOLL X Series Service Manual 5
CHAPTER 1MAINTENANCE TESTS
This chapter describes the following maintenance tests:
1.0 Physical Inspection of the Unit
2.0 Operational TestsOperational Tests
3.0 Leads T e st
4.0 Heart Rate Display Test
5.0 Heart Rate Alarm Test
6.0 Power Supply Test
7.0 Leakage Current Test
8.0 Shock Test
9.0 Advisory Test
10.0 Synchronized Cardioversion Test
11.0 Pacer Test
12.0 Printer Test
13.0 Treatment Report Test
14.0 SpO2 Test (for SpO
15.0 SpO2, SpCO, and SpMet Test (for SpO
16.0 EtCO2 Calibration (required for annual calibration)
17.0 EtCO2 Calibration Check
18.0 EtCO2 Test
19.0 IBP Test
20.0 NIBP Calibration Check
21.0 NIBP Leak Test
22.0 Deflation Test
23.0 Inflation Test
24.0 NIBP Test
25.0 Temperature Test
26.0 Wi-Fi Test (Optional)
27.0 Audio Recording Test (Optional)
feature without SpCO and SpMet)
2
feature with SpCO, or SpCO and SpMet)
2

Before You Begin the Maintenance Tests

Assemble the tools listed in following section, “Equipment You Need to Perform the
Maintenance Tests.”
Keep an extra fully charged X Series battery available.
Schedule an hour to conduct the entire maintenance test.
Perform the tests in the order presented.
Perform all the steps of each test procedure.
Complete all the steps of the procedure before evaluating the test results.
6 9650-001356-01 Rev. D

Equipment You Need to Perform the Maintenance Tests

The equipment listed below is used in the maintenance procedures in this chapter. You can substitute an equivalent device; however, not all simulators and analyzers will produce the same results. Be sure to follow the manufacturer’s recommendations for conducting the maintenance tests.
Fluke Impulse 7000DP Analyzer, with appropriate interface adapter (Fluke Biomedical
BP Pump/Dynatech Nevada Inc. part number 3010-0378)
ECG simulator
Fluke Index 2 SpO
IBP/Temp simulator
NIBP simulator
5% CO
CO
Stop watch
Battery Current Test Fixture 9100-3055-TF
Agilent E3633A Power Supply
Agilent 34401 Digital Multi-meter
calibration gas cylinder with CO2 sample line (kit available from Scott Medical
2
Products, PN 4653ORF-BD)
sampling line airway adapter
2
simulator
2
X Series Overview

Accessories You Need for the X Series Options Maintenance Tests

3-lead, 5-lead and 12-lead ECG cables. (12-lead cable needed if 12-lead option is
installed.)
Dual lumen NIBP hose
IBP cable
SpO
EtCO
Paddles
Printer paper
Battery
Auxiliary power source
USB thumb drive
PC w/ Windows media player and speakers
cable and sensor
2
cable and sensor
2
9650-001356-01 Rev. D ZOLL X Series Service Manual 7
CHAPTER 1MAINTENANCE TESTS

Maintenance Tests

1.0 Physical Inspection of the Unit

Observe this... Pass/Fail
Housing
1.1
Is the unit clean and undamaged?
oo
1.2 Does the unit show signs of excessive wear?
1.3 Does the handle work properly?
1.4 Does the recorder drawer open and close properly?
1.5 Are input connectors clean and undamaged?
1.6 Are there any cracks in the housing?
1.7 Does the front panel have any damage or cracks?
1.8 Are there any loose housing parts?
1.9 Paddles (if applicable)
Do the adult and pedi plates have major scratches or show signs of damage?
1.10 Do the adult shoes slide on and off easily to expose the covered pedi plates?
1.11 Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
1.12 Cables
Are all cables free of cracks, cuts, exposed or broken wires?
1.13 Are all bend/strain reliefs undamaged and free of excessive cable wear?
oo oo oo oo oo oo oo
oo oo oo
oo oo
1.14 Battery
Is the X Series battery fully charged?
1.15 Is the battery seated in the battery well correctly?
1.16 Is battery housing free of cracks and damage?
1.17 Are connectors free of damage and excessive wear?
8 9650-001356-01 Rev. D
oo oo oo oo

2.0 Operational Tests

The X Series unit comes equipped with routine operational tests that can be performed periodically. With the exception of the Physical Inspection of the Unit test, these tests contain instructions within them that provide guidance during the tests.
The operational tests on the unit can be accessed by doing the following:
1. Turn on the X Series unit. Note: If the unit is in AED mode, press the Manual quick access key to go into manual
mode. Use the navigation keys to sel ect th e fou r di gits in the Manual Mode pass code. Press SAVE when you are finished. If the unit has not been config ured to enter a pass code, the message Exit to Manual Mode is displayed. Use the na vigation ke ys to select Yes.
2. Press the More ( ) then the Setup ( ) quick access keys.
3. Use the navigation keys to select Supervisor>Service>Device Tests. The operational tests
are displayed. Use the navigation keys to select one of the tests.
Note: You need a supervisor passcode to enter the Supervisor menu .
Maintenance Tests
9650-001356-01 Rev. D ZOLL X Series Service Manual 9
CHAPTER 1MAINTENANCE TESTS
2.1 Keypad Test
This test checks all the front panel buttons on the unit to make sure they are working correctly. Follow the instructions and use the navigation keys to move around the display and make
selections.
Note: To exit screen after test is complete, press the “Enter” button 3 times to display arrow.
2.2 LED Test
This test checks all the Light Emitting Diodes in the unit to make sure they are working correctly.
Follow the instructions and use the navigation keys to move around the display and make selections.
Note: The speed of the flash varies by color.
10 9650-001356-01 Rev. D
Maintenance Tests
2.3 LCD Test
This test checks the colors of the Liquid Crystal Display on the unit. Follow the instructions and use the navigation keys to move around the display and make
selections.
2.4 RFU Indicator Test
This test checks the Ready For Use Indicator on the unit to make sure it is working correctly. Follow the instructions shown on the display.
9650-001356-01 Rev. D ZOLL X Series Service Manual 11
CHAPTER 1MAINTENANCE TESTS
2.5 Audio Test
This test checks the audio tones on the unit to make sure they are working correctly. Follow the instructions and use the navigation keys to move around the display and make
selections.
2.6 Printer Test
This test checks the printer to make sure it is working correctly. Follow the instructions shown on the display.
12 9650-001356-01 Rev. D
2.7 Battery Test
This test checks the battery to make sure it is working correctly. Follow the instructions on the display.
Maintenance Tests
2.8 Fan Test
This test checks the fan in the unit to make sure it is working correctly. Follow the instructions and use the navigation keys to move around the display and make
selections.
9650-001356-01 Rev. D ZOLL X Series Service Manual 13
CHAPTER 1MAINTENANCE TESTS
2.9 USB Test (Optional)
Note: Before starting this test, contact the ZOLL Technical Service Department to obtain a
test file, or create a file (see instructions below). Once you have obtained a copy, save
it on a USB thumb drive. This test allows you to perform the USB test to make sure it is working correctly. Follow the instructions and use the navigation keys to move around the display and make
selections.
When these tests are completed, use the navigation keys to exit out of the Service menu.
Creating a Test File
1. Create a Text file using Microsoft Notepad.
2. Open new file in Notebook.
3. Click in the first line of the file and type “Thunderbolt Testfile".
4. Save file as Tbolt_usb.txt.
14 9650-001356-01 Rev. D
Maintenance Tests

3.0 Leads Test

Equipment 3-lead, 5-lead, and 12-lead cables, Fluke Impulse 7000DP Analyzer
Note: Test each cable separately.
Test Setup Connect the lead wires appropriate for each test to the Fluke Impulse 7000DP Analyzer or
equivalent, then connect the ECG cable to the X Series unit.
Do this... Observe this... Pass/Fail
3.1 Turn on the X Series unit.
Note: Make sure the lead is not displayed.
3.2 Disconnect one lead from the simulator.
3.3 Reconnect the lead. Repeat step 2 with the remaining leads.
3 .4 Repeat steps 2 and 3 for 5 lead and 12
lead cables.
No ECG LEAD FAULT message displayed.
The ECG LEAD F AULT message displays within 3 seconds (if configured).
Wa it for ECG LEAD FAULT message to clear from the display (if configured).
Note: If heart rate alarm sounds, on the X Series unit:
Press the More quick access key ( ) key to access the second set of quick access keys.
Press the Alarm quick access key (). Press the Alarm Cancel quick access key
().
oo
oo
oo
oo
9650-001356-01 Rev. D ZOLL X Series Service Manual 15
CHAPTER 1MAINTENANCE TESTS

4.0 Heart Rate Display Test

Equipment Fluke Impulse 7000DP Analyzer
ECG Cable (3-lead, 5-lead or 12-lead)
Test Setup Press Lead quick access key until II displays.
Connect the ECG leads to the ECG studs on Analyzer.
Do this... Observe this... Pass/Fail
4.1 Set the ECG Simulator to 120BPM. The Heart Rate displays as 120 +/- 4 bpm.
oo
16 9650-001356-01 Rev. D

5.0 Heart Rate Alarm Test

I, II, III...
Equipment Fluke Impulse 7000DP Analyzer
Do this... Observe this... Pass/Fail
5.1
Press the Leads quick access key
() on the X Series unit and
select lead II.
5.2 On the X Series unit:
Press the More quick
access key ( ) key to access the second set of quick access keys.
Press the Alarm quick
access key ().
Press the Limits softkey.
5.3 On the X Series unit, select HR/PR
Alarm in the Alarm Settings menu and use the navigation keys to do the following:
Set the Upper Limit Enable
to ON.
Set the Lower Limit
Enable to ON.
Set the Lower limit to 35.
Set the Upper limit to 150.
Use the navigation keys to exit the Alarm Settings menu.
5.4 On the X Series unit, press the
Alarm Cancel quick access key ().
5.5
Remove a lead wire from the
simulator
.
ECG
Maintenance Tests
II displays next to the Leads quick access key.
The Alarm Settings menu displays.
The red alarm suspension timer displays at the top of the screen next to the message area.
T
he ECG LEAD FAULT message
displays.
oo
oo
oo
5.6
On the X Series unit, press .
5.7
Reattach ECG Lead wire to
ECG
simulator.
5.8 Set simulator to 30 BPM or lower.
5.9
On the X Series unit, press .
The Alarm Suspend key flashes The ECG LEAD FAULT message no
longer displays. Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR LOW ALARM message displays.
The Alarm Suspend key flashes.
.
The
oo
oo
red alarm suspension timer displays at the top of the screen next to the message area.
5.10 Set the simulator to 120 BPM.
5.11
On the X Series unit, press .
9650-001356-01 Rev. D ZOLL X Series Service Manual 17
Alarm is enabled. The alarm suspend key flashes. The alarm suspension timer disappears.
oo
CHAPTER 1MAINTENANCE TESTS
Do this... Observe this... Pass/Fail
5.12 Set simulator to 160 BPM or higher.
5.13
On the X Series unit, press
the
Alarm Silence/Reset button ()
.
Heart Rate Value is highlighted and the alarm tone sounds. The HR/PR HIGH ALARM message displays.
Alarm is suspended for 90 seconds. The alarm suspension timer appears in black.
oo
oo
18 9650-001356-01 Rev. D
Maintenance Tests

6.0 Power Supply Test

Equipment Battery Current Test Fixture (P/N 9100-3055-TF), Power Supply, Digital Multi-meter.
Test Setup
1. Set the Power Supply to 12V VDC 10 A.
2. Turn off the Power Supply output.
3. Connect the positive (PS+) post of the Power Supply to t he p ositiv e (PS+) post (PS Red) of
the Battery Current Test Fixture.
4. Connect the negative (PS-) post of the Power Supply to the negative (PS-) post (PS Black) of
the Battery Current Test Fixture.
5. With the Battery Current Test Fixture out of the battery well, turn on the Power Supply
output.
6. Set the Digital Multi-meter to measure DC I DC Amps.
7. Set the rear/front switch on the Digital Multi-meter to rear.
8. Connect the Digital Multi-meter (I+) terminal of the Battery Current Test Fixture to the red
terminal (I) on the back of the Digital Multi-meter.
9. Connect Digital Multi-meter (I-) terminal of the Battery Current Test Fixture to the black
terminal (LO) on the back of the Digital Multi-meter.
Do this... Observe this... Pass/Fail
6.1 Install the Battery Current Test
Fixture into battery well and wait about five seconds.
6.2 Make sure the X Series unit is
turned off.
6.3 Press the I Measure switch on the
Battery Current Test Fixture.
6.4 Turn on the X Series unit, and wait
about ten seconds until the power has stabilized.
Verify that the reading on the digital multi-meter is between
0.583mA-0.770mA.
oo
9650-001356-01 Rev. D ZOLL X Series Service Manual 19
CHAPTER 1MAINTENANCE TESTS
Do this... Observe this... Pass/Fail
6.5 On the X Series unit, do the
following:
Press the More ( ) then the
Setup ( ) quick access keys.
Use the navigation keys to
select Supervisor>Display/ Configuration>Device Info and select Restore All Factory Defaults.
Note: You need a supervisor
passcode to enter the Supervisor menu. Also the unit will reboot during the of restoring factory defaults.
6.6 Press the I Measure switch on the
Battery Current Test Fixture.
Observe the current measurement on the DMM, and verify that the reading is between 0.47A-1.25A.
oo
20 9650-001356-01 Rev. D
Maintenance Tests

7.0 Leakage Current Test

Equipment See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Setup See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Repeat leakage test with accessories: MFC, external paddles, internal paddles, and anterior/ posterior paddles. Perform these tests at the line-power voltage and frequency used in your installation.
Maximum Leakage Acceptance Limits
Patient Leakage Mains on Applied Part
Normal Single Fault Single Fault ECG 10 MFC 10
Earth Leakage 2500
50 100 50 100
Normal Single Fault
5000
9650-001356-01 Rev. D ZOLL X Series Service Manual 21
CHAPTER 1MAINTENANCE TESTS
I, II, III...

8.0 Shock Test

WARNING! Take the necessary precautions to guard against shock or injury before you start
conducting the defibrillator tests.
Keep hands and all other objects clear of the multi-function cable connections and defibrillator analyzer when discharging the defibrillator.
Before you discharge the defibrillator, warn everyone near the equipment to STAND CLEAR.
Equipment Fluke Impulse 7000 DP Analyzer (or equivalent), stop watch
Do This... Observe This... Pass/Fail
8.1 Turn on the Analyzer.
8.2 On the Analyzer, select the DEFIB function.
8.3 On the Analyzer, select ENERGY.
8.4 Connect the X Series MFC connector to the pads connection.
8.5 On the X Series unit, apply power to the unit using a battery or an AC adapter.
8.6 Connect the MFC cable to the X Series unit.
8.7 Turn on the X Series unit, and press the Leads quick access key
PADS.
8.8 On the X Series unit, set energy level to 5J by pressing the Energy Select DOWN arrow .
8.9 On the X Series unit, press the CHARGE button.
8.10 On the X Series unit, press the Shock button.
8.11 On the Analyzer, observe and record the energy measurement.
() to select
The unit charges to 5J, the red Shock button illuminates and a ready tone sounds.
Energy is delivered at 5J. Verify that the delivered energy is within 15% based on the values in the table “Delivered Energy at 200J Defibrillator Setting into a Range of Loads” on page 24.
oo
oo
8.12 On the X Series unit, set the energy level to 50J by pressing the Energy Select UP arrow.
8.13 On the X Series unit, press the CHARGE button.
22 9650-001356-01 Rev. D
The unit charges to 50J, the red Shock button illuminates and a ready tone sounds.
oo
Maintenance Tests
Do This... Observe This... Pass/Fail
8.14 On the X Series unit, press the Shock button.
8.15 On the Analyzer, observe and record the energy measurement.
8.16 On the X Series unit, set energy level to 100J by pressing the Energy Select UP arrow.
8.17 On the X Series unit, press the CHARGE button.
8.18 On the X Series unit, press the Shock button.
8.19 On the Analyzer, observe and record the energy measurement.
8.20 On the X Series unit, set energy level to 200J by pressing the Energy Select UP arrow.
8.21 On the X Series unit, press the CHARGE button.
Energy is delivered at 50J. Verify that the delivered energy is within 15% based on the values in the table “Delivered Energy at 200J Defibrillator Setting into a Range of Loads” on page 24.
The unit charges to 100J, the Shock button illuminates and a ready tone sounds.
Energy is delivered at 100J. Verify that the delivered energy is within 15% based on the values in the table “Delivered Energy at 200J Defibrillator Setting into a Range of Loads” on page 24.
The unit charges to 200J, the red Shock button illuminates and a ready tone sounds.
oo
oo
oo
oo
8.22 On the X Series unit, press the Shock button.
8.23 On the Analyzer, observe and record the energy measurement.
8.24 On the X Series unit, press the CHARGE button.
8.25 When the X Series unit is charged, start the stop watch.
8.26 Turn off the X Series unit.
Energy is delivered at 200J. Verify that the delivered energy is within 15% based on the values in the table “Delivered Energy at 200J Defibrillator Setting into a Range of Loads” on page 24.
The unit charges to 200J, the red Shock button illuminates, and a ready tone sounds.
Verify that after 50 seconds the tone changes and within 10 seconds the unit does an internal discharge and displays DISARMED TIMEOUT.
oo
oo
oo
9650-001356-01 Rev. D ZOLL X Series Service Manual 23
CHAPTER 1MAINTENANCE TESTS
Delivered Energy at 200J Defibrillator Setting into a Range of Loads
Selected
Energy
5 J 3 J5 J6 J6 J6 J6 J615%
50 J 35 J 54 J 59 J 61 J 62 J 61 J 59 J ±15% 100 J 71 J 109 J 119 J 122 J 125 J 123 J 119 J ±15% 200 J 142 J 230 J 2 49 J 253 J 269 J 261 J 260 J ±15%
25 50 75 100 125 150 175 Accuracy*
Note: For a complete listing of X Series Delivered Energy at Every Defib rillator Setting into a Range
of Loads
, see Appendix A in the X Series Operator’s Guide.
Load
24 9650-001356-01 Rev. D
Loading...
+ 114 hidden pages