This Quick Reference Guide is intended to provide easy access to
®
most common uses of the ZOLL
completely understand the intended use warnings, cautions, and
instructions in the R Series Operator’s Guide before using the
R Series unit.
2. Press down on RELEASE latches with thumbs and remove the paddles from their holders.
3. Apply gel to electrode surfaces or gel patches to patient. If using gel, rub electrodes
together to spread gel.
4. Place sternum paddle to the right of sternum, just below clavicle.
5. Place apex paddle left of nipple on anterior-axillary line.
6. Press paddles firmly against patient’s skin.
Cardioversion Electrode
Anterior / Posterior Placement
(Recommended)
BACKFRONT
R Series BLS/Plus
Defibrillation (AED)
1. Apply pads as shown.
2. Turn selector switch to ON.
3. System will begin to analyze or instruct
START CPR, depending upon configuration
chosen.
4. Once analysis begins, stand clear.
5. If rhythm is shockable, defibrillator
will charge and red SHOCK button
will illuminate.
6. If shock button illuminates, shout “clear” –
once all are clear, press SHOCK.
7. If no shock advised, check pulse – if no
pulse, start CPR immediately.
Defibrillation (Manual)
1. Apply pads.
2. Turn selector switch to ON.
3. Press MANUAL, then CONFIRM soft keys. System will switch to ALS mode
(depending upon configuration, your defibrillator may not require a second key
press to activate ALS mode).
4. Energy select will automatically display the correct adult energy per configuration.
5. If desired, energy may be adjusted up or down.
5. If rhythm appears shockable, press CHARGE button.
6. When red SHOCK button illuminates, shout “clear” and verify others are clear.
7. Press SHOCK button.
8. Check pulse – if no pulse, return to CPR immediately.
Defibrillation (Advisory)
1. Apply pads.
2. Turn selector switch to ON.
3. Press MANUAL then CONFIRM soft keys.
4. Press ANALYZE to start advisory function
and follow the instructions.
5. If rhythm is shockable, defibrillator
will charge and red SHOCK button
will illuminate.
6. If shock button illuminates, shout “clear” –
once all are clear, press SHOCK.
7. If no shock advised, check pulse – if no
pulse, start CPR immediately.
Page 3Page 2
Adult Defibrillation – Pediatric Defibrillation
R Series ALS
CPR
CPR Idle TimerCPR Index
Real CPR Help
Defibrillation (Advisory)
1. Apply pads.
2. Turn selector switch to DEFIB.
3. The energy will be displayed as configured
by your institution.
4. Shout “Clear, Analyzing”; stop CPR and
ensure patient is motionless.
5. Press ANALYZE button.
6. Read display message. If SHOCK ADVISED is
displayed, shout “Clear” and verify others
are clear.
7. Press SHOCK button.
8. If NO SHOCK ADVISED, check for a pulse
and if no pulse, resume CPR immediately.
Defibrillation (Manual)
1. Apply pads.
2. Turn selector switch to red area labeled
DEFIB.
3. Energy selected will be as configured by
institution protocol. You may change settings
if desired by using the ENERGY SELECT UP
and DOWN arrow keys.
4. If the rhythm appears to be shockable, press
the CHARGE button.
5. Shout “Clear” and verify others are clear.
6. Press SHOCK button.
Pediatric Defibrillation
Follow the instructions as outlined above for your unit with the following exceptions:
• OneStep Pediatric pads should be used on children up to eight years of age. For children over
eight years of age, use adult pads.
• When the OneStep Pediatric pad is in use, the R Series will automatically decrement the energy
delivered to 50, 70, and 85 J increments when these pads are connected. These levels are
appropriate for children from 30 kg or greater.
• When using OneStep Pediatric electrodes on children under 30 kg, an energy of 1 J per kilo is
recommended and will need to be set manually using the ENERGY SELECT UP and DOWN arrows.
• For all other electrodes, select the desired energy manually based upon 1 J per kilogram.
• Do not use AED or Advisory mode in children under eight years of age unless OneStep Pediatric
pads are attached or energy has been manually decremented in advisory mode.
1. Place ZOLL pads with CPR Sensor (OneStep Complete,
OneStep CPR, or CPR stat•padz
2. The CPR Sensor should be placed on the sternum where
your hands would normally be placed for CPR.
3. Place the heel of your hand on the sensor and begin
compressions normally.
4. If your compression rate and depth are appropriate to AHA/ERC Guidelines, the hexagon
on the screen of the R Series (CPR Index
5. If the depth of compression is insufficient, the message “PRESS HARDER” will be
heard and the “Depth” will be displayed next to the CPR Index. When compressions
are correct, the message “GOOD COMPRESSIONS” will be heard.
6. If compression rate is too slow, “Rate” will be displayed next to the index and
a 100 bpm metronome will be heard until rate is correct.
7. If your R Series does not display a CPR Index, your defibrillator has only basic CPR
feedback and only audible prompts as outlined above will be available.
®
) on patient.
Rate and Depth Indicators
™
) will begin to fill with color and remain filled.
See-Thru CPR
See-Thru CPR is a proprietary algorithm that filters the
CPR artifact from the ECG to allow users to see a close
approximation of a patient’s organized underlying rhythm
during CPR compressions, reducing interruptions in CPR.
Activating See-Thru CPR:
1. Press OPTIONS soft key.
2. Press TRACES soft key.
Using See-Thru CPR:
1. During CPR, view the filtered ECG
(second trace) on display to determine
if an organized rhythm appears to be
developing.
2. Pause compressions and visually
confirm if rhythm is shockable or not.
3. Press TRACE 2 soft key.
4. Press Filt ECG soft key.
3. If shockable, press CHARGE, shout
“Clear” and verify others are clear.
4. Press SHOCK button.
5. Resume CPR immediately for 2 minutes
before stopping CPR to check rhythm.
If non-shockable, resume compressions.
Unfiltered ECG signal during CPR
Signal filtered by See-Thru CPR
Page 4Page 5
Pacing / Cardioversion
Monitoring
Large T Wave
egative R Wave
N
acer Markers
P
Example of paced rhythm
Pacing
1. Place OneStep Pacing or One Step Complete Pads on patient.
To pace with other defibrillation pads, electrodes must be
placed on the patient and the three or five lead cable attached
to the defibrillator.
2. Turn Selector Switch to green area labeled PACER.
3. Turn the pacer RATE knob to desired rate.
4. Turn the pacer OUTPUT knob clockwise slowly until capture.
5. Slowly decrease output by turning OUTPUT knob counterclockwise until capture disappears, then increase by 10%.
6. Check pulse to verify mechanical capture.
7. To check patient’s underlying rhythm; press and hold 4:1 button.
Note: When using BLS or Plus models, manual mode must be activated to pace.
Dial settings for
R Series Pacing
ECG Monitoring
Using OneStep ECG Cable
1. Disconnect OneStep electrode from the OneStep Therapy cable.
2. Plug the OneStep ECG cable into the OneStep Therapy cable.
3. Abrade skin to remove outer layer of dead skin.
4. Snap leads on to each electrode.
5. Peel off protective backing from electrode.
6. Apply electrodes to patient’s skin.
Using Standard ECG cable
1. Unplug OneStep ECG cable from back of unit.
2. Plug ECG cable connector into back of unit.
3. Abrade skin to remove outer layer of dead skin.
4. Snap leads onto each electrode.
5. Peel off protective backing from electrode.
6. Apply electrodes to patient’s skin.
ECG Display
1. Turn selector switch to MONITOR on the R Series ALS unit.
If using BLS or Plus device, press MANUAL then CONFIRM
soft keys to activate ALS mode.
2. Press LEAD button to change leads.
3. Press SIZE button to change size of display.
Dial settings for
R Series Monitoring
End Tidal CO2(EtCO2) - Mainstream
Synchronized Cardioversion
1. Apply pads or paddles to patient.
2. Turn selector switch to DEFIB. If using BLS or Plus device,
press MANUAL then CONFIRM soft keys instead.
3. Press SYNC ON/OFF soft key.
4. Verify that arrows (i) display over R-waves and display reads SYNC.
Do not cardiovert if SYNC is OFF.
5. Follow Manual Defibrillation instructions and institution protocol
for energy settings.
Page 6Page 7
1. Plug EtCO2cable into the back of the unit in the yellow port marked CO2.
2. Attach sensor to patient’s ET tube.
3. Turn selector switch to grey area labeled MONITOR. If using BLS or Plus device,
turn ON, then press MANUAL then CONFIRM soft keys to use in ALS mode.
3. Press the PARAM soft key.
4. Press the CO2soft key.
5. Press the ENABLE EtCO2soft key.
6. If properly attached, the EtCO2values will display on the screen.
1. Connect the SpO2cable to the patient sensor and to
the blue SpO2port on the rear of the unit labeled SpO2.
2. Attach sensor to patient’s finger.
3. Turn selector switch to grey area labeled MONITOR on
the R Series ALS. If using BLS or Plus device, turn ON,
then press MANUAL then CONFIRM soft keys to use
in ALS mode.
4. If properly attached, the SpO2values will display
on the screen.