ZOLL R Series User Manual

0 (0)

R Series® Configuration Manual

9650-0902-01 Rev. K

The issue date for the R Series Configuration Guide (REF 9650-0902-01 Rev. K) is May, 2012.

If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to determine if additional product information updates are available.

© 2012 ZOLL Medical Corporation. All rights reserved. R Series, pedi-padz, Real CPR Help, Code-Ready, and OneStep are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other countries.

Masimo is a registered trademark of Masimo Corporation in the United States and/or other countries. All other trademarks are the property of their respective owners.

ZOLL Medical Corporation

269 Mill Road Chelmsford, MA USA 01824-4105

ZOLL International Holding B.V.

Newtonweg 18

6662 PV ELST

The Netherlands

0123

 

Table of Contents

Chapter 1

Introduction

 

Entering System Configuration Mode ....................................................................................

2

Printing Configuration Settings and Activity Logs ..................................................................

3

Uploading Data from a Compact Flash Card .........................................................................

6

Saving Configuration to a Compact Flash Card.....................................................................

8

Restoring Default Values .......................................................................................................

9

Accessing and Changing Configuration Settings.................................................................

10

Leaving Configuration Mode.................................................................................................

11

Chapter 2

Configuration Settings

 

General Settings ..................................................................................................................

14

Device Identifier ...........................................................................................................

14

Configuration Mode Passcode .....................................................................................

16

Code Markers ..............................................................................................................

17

ECG Settings

.......................................................................................................................

19

Notch Filter ..................................................................................................................

19

Display Lead ..........................................................................................on Power-Up

19

Frequency ...................................................................................................Response

19

Gain on Power .......................................................................................................Up

20

Enable Leads .........................................................................................................Off

20

Enable Pacer ...............................................................................................Detection

20

Audio Settings......................................................................................................................

21

QRS Volume ..........................................................................................on Power-Up

21

Low Battery .........................................................................................................Tone

21

Enable Corrective ................................................................................Voice Prompts

21

Basic Defibrillation .................................................................................................Settings

22

Defib Default ...................................................................................................to Pads

22

Basic Auto ......................................................................................Energy Escalation

22

Retain Sync ...............................................................................................After Defib

22

Defib Ready ................................................................................................Hold Time

23

Enable Remote ...................................................................................................Sync

23

9650-0902-01 Rev. K

ZOLL R Series Configuration Manual

i

TABLE OF CONTENTS

 

Advisory Defib......................................................................................................................

24

Energy Level: Shock 1 .................................................................................................

24

Energy Level: Shock 2 .................................................................................................

25

Energy Level: Shock 3 .................................................................................................

25

Ped Energy Level Shock 1 ..........................................................................................

25

Ped Energy Level Shock 2 ..........................................................................................

26

Ped Energy Level Shock 3 ..........................................................................................

26

Auto Analyze ................................................................................................................

26

Enable Check Pulse Prompt ........................................................................................

27

Check Patient Prompt ..................................................................................................

27

Auto Charge in Advisory Mode ....................................................................................

27

Enable Voice in Advisory Mode ...................................................................................

28

Display DO CPR Message ..........................................................................................

28

CPR Message After No Shock Advised .......................................................................

29

CPR Message After Shock ..........................................................................................

29

Stacked Shocks ...........................................................................................................

30

Restart Analysis After CPR ..........................................................................................

30

CPR Interval ................................................................................................................

30

CPR Settings .......................................................................................................................

31

Enable CPR Voice Prompts .........................................................................................

31

Enable CPR Display ....................................................................................................

31

Enable CPR Metronome ..............................................................................................

31

Target CPR Depth .......................................................................................................

31

Display Fully Release Prompt ......................................................................................

32

Pace Settings.......................................................................................................................

33

Pace Rate Setting on Power-Up ..................................................................................

33

Async Softkey in Pace Mode .......................................................................................

33

Alarms Settings....................................................................................................................

34

Alarm Limits at Power Up ............................................................................................

35

Alarms Active at Power Up ..........................................................................................

39

Display Settings ...................................................................................................................

40

Display Time ................................................................................................................

40

Select Trace 2 At Power Up ........................................................................................

40

Select Trace 3 At Power Up ........................................................................................

41

ECG Color ...................................................................................................................

41

SpO2 Color ..................................................................................................................

41

CPR Color ....................................................................................................................

41

EtCO2 Color .................................................................................................................

42

NIBP Color ...................................................................................................................

42

SPO2 at Power Up .......................................................................................................

42

ii

www.zoll.com

9650-0902-01 Rev. K

NIBP Settings (Optional)......................................................................................................

43

Blood Pressure Units ...................................................................................................

43

Trigger NIBP Meas on BP Alarm .................................................................................

43

Trigger NIBP Meas on HR Alarm .................................................................................

43

NIBP Auto Interval Default ...........................................................................................

44

Adult Initial Cuff Inflation ..............................................................................................

44

Pediatric Initial Cuff Inflation ........................................................................................

44

Neonatal Initial Cuff Inflation ........................................................................................

44

NIBP Patient Mode at Power Up .................................................................................

45

NIBP Button Press & Hold ...........................................................................................

45

Enable Motion Tolerance .............................................................................................

45

EtCO2 Settings (Optional)....................................................................................................

46

Default EtCO2 Units .....................................................................................................

46

Displayed Zoom Level .................................................................................................

46

Default EtCO2 Comp Setting .......................................................................................

46

Power Save Mode at Power Up ...................................................................................

47

EtCO2 Average On Power Up .....................................................................................

47

Printer & Report Settings .....................................................................................................

48

Set Report Restart Delays ...........................................................................................

48

Auto Generate Strips ...................................................................................................

48

Print 3 Leads When Leads Sel. ................................................................................... 48

Readiness Test Settings ......................................................................................................

49

Automatic Readiness Test ...........................................................................................

49

Readiness Test Time ...................................................................................................

49

Print Readiness Test Results ......................................................................................

49

Enable Readiness Test Audible Alert (Future Use) .....................................................

50

One Step Padz Installed ..............................................................................................

50

AED Mode ...........................................................................................................................

51

Display ECG in AED Mode ..........................................................................................

51

Display HR in AED Mode .............................................................................................

51

Manual Confirm Enable ...............................................................................................

51

Extra CPR Interval Before 1st Analysis .......................................................................

52

Duration of Extra CPR Interval ....................................................................................

52

Display Remaining CPR Timer ....................................................................................

52

9650-0902-01 Rev. K

ZOLL R Series Configuration Manual

iii

TABLE OF CONTENTS

Wi-Fi Settings (Optional)......................................................................................................

53

Full Disclosure Transfer Mode .....................................................................................

53

Defib History Transfer Mode ........................................................................................

53

Infrastructure Mode Options ........................................................................................

53

DHCP/Static IP ............................................................................................................

53

Local IP Address ..........................................................................................................

54

Gateway IP Address ....................................................................................................

55

Subnet Mask ................................................................................................................

56

Use DNS ......................................................................................................................

57

Primary/Secondary DNS IP Address ...........................................................................

57

SSID Names (up to 2 entries) ......................................................................................

58

Server Name (Full Disclosure) .....................................................................................

59

Server IP (Full Disclosure) ...........................................................................................

60

Server Port (Full Disclosure) ........................................................................................

61

Server Name (Defib History) ........................................................................................

62

Server IP (Defib History) ..............................................................................................

63

Server Port (Defib History) ...........................................................................................

65

Passphrase/Keys (up to 2 entries) ...............................................................................

66

Enable Report ID .........................................................................................................

67

Wi-Fi Region ................................................................................................................

67

Wi-Fi Radio ..................................................................................................................

67

iv

www.zoll.com

9650-0902-01 Rev. K

Chapter 1

Introduction

The R Series® is designed with several user selectable features that allow operators to configure the device according to their protocols and local requirements. This manual describes the configuration settings (options and defaults) and how to change them for ALS, Plus, and BLS models. References to AED settings are for Plus and BLS models only.

Note: Your unit may not contain all features listed. For BLS/Plus units, some features are available in manual mode operation only.

For more information, see the following sections:

“Entering System Configuration Mode” on page 2

“Printing Configuration Settings and Activity Logs” on page 3

“Uploading Data from a Compact Flash Card” on page 6

“Saving Configuration to a Compact Flash Card” on page 8

“Restoring Default Values” on page 9

“Accessing and Changing Configuration Settings” on page 10

“Leaving Configuration Mode” on page 11

9650-0902-01 Rev. K

1

Chapter 1 Introduction

Entering System Configuration Mode

In order to make any changes to unit configuration, you must first power on the unit in System Configuration mode.

Procedure

Follow these steps to power on the unit in System Configuration mode:

Step Action

1.Turn the selector switch to OFF, and wait 10 seconds.

 

Simultaneously press and hold the right-most and

 

System Configuration

 

 

 

 

 

 

 

 

left-most softkeys (below the monitor) while turning

 

Set Configuration Passcode

 

the selector switch to MONITOR (for ALS models)

 

 

or Power Off to Exit

 

 

 

 

 

 

 

 

 

or ON (for Plus and BLS models), and hold the

0

0 0 0

? ?

?

?

 

softkeys for approximately 4 seconds.

 

 

 

 

 

 

 

The unit powers on in System Configuration mode.

 

0

1

2

3

Delete

 

 

 

 

 

 

 

 

 

 

2.

Enter the passcode using the appropriate softkeys.

 

 

 

 

 

 

 

Note: When you enter System Configuration for

 

 

 

 

 

 

 

the first time, enter the default pass code of

 

System Configuration

 

 

 

 

 

 

 

 

 

 

00000000.

Enter Configuration Passcode

 

To change a digit, use the Delete softkey to

 

or Power Off to Exit

 

 

 

 

 

 

 

 

 

 

highlight the entry and re-enter the digit.

0

0 0 ?

? ?

? ?

 

 

When all 8 digits have been entered, the unit

 

0

1

2

3

Delete

 

 

 

automatically advances to the Select Language

 

 

screen.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.

Use the Prev Value and Next Value softkeys to

 

 

 

 

 

 

 

 

toggle the highlighted area between the language

 

System Configuration

 

 

options.

 

 

 

 

 

Select Language

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ENGLISH

 

 

 

 

 

 

 

FRANCAIS

 

 

 

 

 

 

 

Prev

Next

 

Exit

 

 

 

 

 

 

 

 

 

 

 

Value

Value

Enter

Config

4.Press the Enter softkey to select the highlighted

language.

All subsequent configuration menus will be in the System Configuration selected language.

The unit automatically advances to the Main Configuration menu.

Read

Restore From Change Exit

Print Defaults Card Config Config

2

9650-0902-01 Rev. K

Printing Configuration Settings and Activity Logs

Printing Configuration Settings and Activity Logs

To print the current configuration or activity log of your unit while the unit is in the System Configuration mode, press the Print softkey from the Main System Configuration menu. For instructions on putting the unit in System Configuration mode, refer to the previous section, “Entering System Configuration Mode.”

Printing Configuration Settings

In System Configuration mode, press the Print softkey, then the Print Config softkey to print a stripchart of all the unit’s configuration settings. See a copy of a sample Configuration Report below.

Note: In normal operation, you can print the configuration by pressing the following softkeys: Options, More, then Print Config.

9650-0902-01 Rev. K

3

ZOLL R Series User Manual

Chapter 1 Introduction

4

9650-0902-01 Rev. K

Printing Configuration Settings and Activity Logs

Printing the Activity Log

In System Configuration mode, press the Print softkey, then the Print Activity Log softkey to print a stripchart of the unit’s activity log. The log consists of the last 1,024 user entries or system prompts. See a copy of a sample Activity Log report below.

9650-0902-01 Rev. K

5

Chapter 1 Introduction

Uploading Data from a Compact Flash Card

You can upload data from a compact flash card to perform the following tasks:

Copy the configuration of one unit to any number of units

Install new software to the unit

Copy a second language to the unit

Procedure

Follow these steps to load a configuration, another language, or software from a compact flash card:

Step Action

1.Install the compact flash card in the R Series unit,

and power on the unit in System Configuration mode

as described in “Entering System Configuration System Configuration Mode” on page 2.

 

Read

 

 

Restore

From

Change

Exit

Print Defaults

Card

Config

Config

2.

Press the Read From Card softkey to display the

 

 

 

 

 

Read Card screen.

System Configuration

 

Use the Prev Value and Next Value softkeys to

 

Read Which Type of Card?

 

toggle the highlighted area between the following

 

 

 

 

 

 

options:

 

 

 

 

 

LANGUAGE

 

CONFIGURATION

 

LANGUAGE - allows you to upload a second

 

 

PROGRAM

 

language to the unit. English is always on the

 

 

 

 

 

unit as default, and one other language can be

 

Prev

Next

 

 

 

added.

 

Value

Value Enter

 

 

 

CONFIGURATION -allows you to read the configuration data from a card, thereby configuring this unit.

PROGRAM - allows you to install new software in either the PD processor or the ECG processor.

Note: Do not install an older version of software over a newer version because it will cause software problems.

3.

Press the Enter softkey to initiate reading and

 

 

 

 

 

 

storing the information from the installed compact

System Configuration

 

 

flash card.

 

 

 

 

 

 

 

 

Once the card has been read and loaded, the unit

 

Card Read

 

 

 

displays the CARD READ message (for a language

Press Change Config to View

 

 

 

 

 

 

 

 

card) or the CARD READ and PRESS CHANGE

 

 

 

 

 

 

CONFIG TO VIEW messages (for a configuration

 

 

 

 

 

 

card).

 

Restore

From

Change

Exit

 

 

 

 

Read

 

 

 

 

Print

Defaults

Card

Config

Config

If you want to review configuration settings, press the Change Config softkey. See “Accessing and Changing Configuration Settings” on page 10 for more information. If you want to leave

6

9650-0902-01 Rev. K

Uploading Data from a Compact Flash Card

System Configuration mode, press the Exit Config softkey. See “Leaving Configuration Mode” on page 11 for more information.

If the unit encounters an error reading the memory card, one of the following error messages is displayed:

Error Message

Description

 

 

CARD READ FAILED

No response is detected from the card reader.

 

 

BAD CARD

The unit cannot read the card.

 

 

NO CARD INSERTED

The card is not installed in the data card slot.

 

 

9650-0902-01 Rev. K

7

Chapter 1 Introduction

Saving Configuration to a Compact Flash Card

Whenever you exit configuration mode, you can optionally save the configuration to a compact flash card. This allows you to copy the configuration from one unit to any number of units.

Procedure

Follow these steps to save configuration data to a compact flash card:

Step Action

1.Power on the unit in System Configuration mode as

described in “Entering System Configuration Mode”

on page 2. System Configuration

 

Read

 

 

Restore

From

Change

Exit

Print Defaults

Card

Config

Config

2.Press the Exit Config softkey to display the Exit

Configuration screen, and then install the flash card

System Configuration

in the R Series unit.

 

 

 

 

 

Save Changes and Exit

 

Discard Changes and Exit

 

Return to Main Menu

 

Prev

Next

 

Item

Item Enter

3.Press the Enter softkey to save the configuration changes and exit.

 

The unit displays the SAVING CONFIGURATION

System Configuration

 

Save Changes to Card?

 

and DO NOT POWER OFF messages and then the

 

 

 

 

 

 

Save Changes To Card screen appears.

 

 

 

 

 

 

 

NO

 

 

 

 

YES

 

 

Prev

Next

 

 

Item

Item Enter

 

 

 

 

 

 

4.

Press the Next Item softkey to highlight YES, then

 

 

 

 

 

press the Enter softkey.

 

 

 

 

 

The unit displays the SAVING TO CARD message

 

 

 

 

 

and then the POWER OFF TO EXIT

 

 

 

 

 

CONFIGURATION message.

 

 

 

 

8

9650-0902-01 Rev. K

Restoring Default Values

Restoring Default Values

You can restore all configuration settings to their factory default settings. Only the display language, if different from the default, does not change.

Procedure

Follow these steps to restore all configuration settings to default values:

Step Action

1.Power on the unit in System Configuration mode as

described in “Entering System Configuration Mode”

on page 2.

System Configuration

 

 

Read

 

 

Restore

From

Change

Exit

Print Defaults

Card

Config

Config

2.

Press the Restore Defaults softkey to reconfigure

 

 

 

 

 

 

settings to their default settings.

System Configuration

 

 

 

 

 

Once default settings are restored, the unit displays

 

Defaults Restored

 

 

the DEFAULTS RESTORED and PRESS CHANGE

Press Change Config to View

 

 

 

 

 

 

 

 

CONFIG TO VIEW messages.

 

 

 

 

 

 

 

 

 

Read

 

 

 

 

 

Restore

From

Change

Exit

 

 

Print

Defaults

Card

Config

Config

If you want to review configuration settings, press the Change Config softkey. See “Accessing and Changing Configuration Settings” on page 10 for more information. If you want to leave System Configuration mode, press the Exit Config softkey. See “Leaving Configuration Mode” on page 11 for more information.

9650-0902-01 Rev. K

9

Chapter 1 Introduction

Accessing and Changing Configuration Settings

You can change the current value for any configuration setting. You access most of the settings in the same way as described below. Additional steps are required to access and modify several settings. For those settings, the additional steps are provided with the setting description in the next chapter.

Procedure

Follow these steps to access the settings.

Step Action

1.Press the Change Config softkey from the Main System Configuration screen.

The unit advances to the Select Configuration

System Configuration

Select Configuration Category

Category screen.

 

 

CPR

 

General

 

 

ECG

 

PaceS t/Review

 

Audio

 

Alarms

 

Basic Defib

 

Display

 

Advisory Defib

 

Printer & Report

 

 

Page 1 of 2

 

 

 

Main

 

 

 

Enter Menu

System Configuration

Select Configuration Category

Readiness Test

AED Mode

NIBP

EtCO2

Wi-Fi

Page 2 of 2

Main

Enter Menu

2.Use the softkeys on the screen to move through the list and select a category, or to exit as follows:

— moves the highlight up, down, left, and right through the list of categories.

Enter — provides you access to the options of the highlighted category. Press this softkey once you have highlighted the appropriate category.

Main Menu — returns you to the Main System Configuration screen.

3.Once you select and access the options for a category, you can highlight the specific setting you want to change, then press the Change Value softkey until the appropriate value is displayed.

Note: Some options require and evoke an additional screen to set the value (for example, setting the Device Identifier). Specific instructions for those options are provided in the next chapter.

10

9650-0902-01 Rev. K

Leaving Configuration Mode

Step Action

4.When you are finished, you can either press the Category List softkey to return to the Category screen, or press the Prev Item or Next Item softkey to move to another option whose settings you want to modify.

Leaving Configuration Mode

Procedure

Follow the step below to leave System Configuration mode:

Step Action

1.Press the Exit Config softkey from the Main System

Configuration screen to display the Exit

System Configuration

Configuration screen.

 

 

Use the Prev Item and Next Item softkeys to toggle

 

 

the highlighted area between the following options:

Save Changes and Exit

 

SAVE CHANGES AND EXIT - saves all

Discard Changes and Exit

Return to Main Menu

configuration changes you made and prompts

 

 

you to Save to Card. If you specify Yes, the

Prev

Next

configuration information is saved to the installed

Item

Item Enter

 

Compact Flash card. Lastly, the message

POWER OFF TO EXIT CONFIG is displayed.

Power off the unit.

DISCARD CHANGES AND EXIT-disregards the changes you made and displays the message

POWER OFF TO EXIT CONFIG. Power off the unit.

RETURN TO MAIN MENUreturns you to the Main System Configuration screen, where you can make more changes by selecting the Change Config softkey. You do not have to reenter the passcode.

Note: No changes are made if you turn the unit off without saving the configuration.

9650-0902-01 Rev. K

11

Chapter 1 Introduction

12

9650-0902-01 Rev. K

Chapter 2

Configuration Settings

This chapter describes all of the configuration settings on the R Series and includes the following sections.

“General Settings” on page 14

“ECG Settings” on page 19

“Basic Defibrillation Settings” on page 22

“Advisory Defib” on page 24

“CPR Settings” on page 31

“Pace Settings” on page 33

“Alarms Settings” on page 34

“Display Settings” on page 40

“NIBP Settings (Optional)” on page 43

“EtCO2 Settings (Optional)” on page 46

“Printer & Report Settings” on page 48

“Readiness Test Settings” on page 49

“AED Mode” on page 51

“Wi-Fi Settings (Optional)” on page 53

Note: The instructions in this chapter assume that you powered on the unit in System Configuration mode as described in “Entering System Configuration Mode” on page 2, and are familiar with how to access the settings, see “Accessing and Changing Configuration Settings” on page 10.

9650-0902-01 Rev. K

13

Chapter 2 Configuration Settings

General Settings

General settings are attributes that affect the unit’s basic operation and accessibility and include:

Device Identifier

Configuration Mode Passcode

Code Markers

Device Identifier

The Device Identifier feature allows you to set an 11-digit device identifier code. This code is divided into two sub codes: the site ID (consisting of eight alphanumeric characters) and the device ID (consisting of three alphanumeric characters). The device identifier is printed on the summary report and fax printouts.

Values are as follows:

Possible Values:

0 through 9, A through Z, blank and underscore inclusive for each digit.

 

 

Default:

00000000 000

 

 

Procedure

Follow these steps to set up a device identifier.

Step

Action

 

 

 

 

 

 

 

 

 

 

 

 

1.

From the Select Configuration Categories screen,

 

 

 

 

 

 

select General, then press the Enter softkey.

 

 

 

 

 

 

The General Settings screen appears.

System Configuration

 

 

 

 

General Settings

 

 

 

DEVICE IDENTIFIER

ABCDEFGH 777

 

CONFIGURATION

Set/Review

 

 

00000000

 

 

MODE PASSCODE

 

 

 

 

CODE MARKERS

Config/Review

 

 

 

 

Page 1 of 1

 

 

 

Change

Prev

Next

Category

Main

Exit

 

Value

Item

Item

List

Menu

Config

2.Select DEVICE IDENTIFIER and press the

Change Value softkey.

The Device Identifier screen appears.

System Configuration

 

 

 

 

Set Site ID

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

0

0

0

0

0

0

0

 

 

 

 

 

 

 

Set Device ID

 

 

 

 

 

 

 

 

 

0

0

 

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A

B

C

D

E

F

 

G

H

I

Cancel

 

 

 

J

K

L

M N

O

 

P

Q

R

Del

 

 

 

S T U V W X Y, Z 0 < >

 

 

 

1

2

3

4

5

6

 

7

8

9

_

 

 

 

 

 

 

 

 

 

 

 

 

 

Enter

 

 

 

 

 

 

 

 

 

 

 

 

Char Return

14

9650-0902-01 Rev. K

General Settings

Step Action

3.In the character selection table:

lets you move to the next or previous characters.

Use the softkeys as follows:

— moves the highlight up, down, left, and right through the character selection table.

Enter Char — enters the highlighted character for 8 digits of the Site ID and the 3 digits of the Device ID.

Return — returns you to the General Settings screen. Press the Return softkey when you are finished entering the IDs. Note that selecting Enter also returns you to the General Settings screen.

9650-0902-01 Rev. K

15

Chapter 2 Configuration Settings

Configuration Mode Passcode

This feature allows you to set an 8-digit system configuration passcode. Once set, only those people who know the passcode can access the configuration mode and modify settings.

Values are as follows:

Possible Values: 0 through 3 inclusive for each digit

Default: 00000000

Procedure

Follow these steps to set up a configuration mode passcode:

Step

Action

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.

From the Select Configuration Categories

 

 

 

 

 

 

 

screen, select General, then press the Enter

 

System Configuration

 

 

softkey.

 

 

 

The General Settings screen appears.

 

 

General Settings

 

 

 

 

 

 

 

 

 

DEVICE IDENTIFIER

ABCDEFGH 777

 

 

 

 

 

Set/Review

 

 

CONFIGURATION

00000000

 

 

MODE PASSCODE

 

 

 

 

CODE MARKERS

Config/Review

 

 

 

 

 

Page 1 of 1

 

 

 

Change

 

Prev

Next

Category

Main

Exit

 

 

 

Value

 

Item

Item

List

Menu

Config

2.Select CONFIGURATION MODE PASSCODE and press the Change Value softkey.

 

The Set Configuration Passcode screen appears.

System Configuration

 

Set Configuration Passcode

 

 

 

 

 

 

0

0

0

0

0

0

0

0

 

 

Prev

 

Inc

 

Dec

Next

 

 

 

 

 

 

Digit

 

Digit

 

Digit

Digit Enter

 

 

 

 

 

 

 

 

 

 

3.

Use the softkeys as follows:

 

 

 

 

 

 

 

 

 

Prev Digit — moves the highlight left to the

 

 

 

 

 

 

 

 

 

previous digit.

 

 

 

 

 

 

 

 

 

Inc Digit or Dec Digit — increments or

 

 

 

 

 

 

 

 

 

decrements the value of the selected digit.

 

 

 

 

 

 

 

 

 

Next Digit — moves the highlight right to the

 

 

 

 

 

 

 

 

 

next digit to be entered.

 

 

 

 

 

 

 

 

 

Enter — saves the values that you entered

 

 

 

 

 

 

 

 

 

and returns the unit to the General

 

 

 

 

 

 

 

 

 

Configuration menu.

 

 

 

 

 

 

 

 

16

9650-0902-01 Rev. K

General Settings

Code Markers

This setting allows you to review and change the preconfigured list of Code Markers for use in DEFIB, MONITOR, and PACE modes. Code Marker settings are language dependent. You can specify up to three groups (MONITOR, DEFIB, and PACE) of 20 entries each.

Procedure

Follow these steps to review, change, or reset code marker settings:

Step

Action

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.

From the Select Configuration Categories

 

 

 

 

 

 

 

screen, select General, then press the Enter

 

System Configuration

 

 

softkey.

 

 

 

The General Settings screen appears.

 

 

General Settings

 

 

 

 

 

 

 

 

 

DEVICE IDENTIFIER

ABCDEFGH 777

 

 

 

 

 

Set/Review

 

 

CONFIGURATION

00000000

 

 

MODE PASSCODE

 

 

 

 

CODE MARKERS

Config/Review

 

 

 

 

 

Page 1 of 1

 

 

 

Change

 

Prev

Next

Category

Main

Exit

 

 

 

Value

 

Item

Item

List

Menu

Config

2.

Select CODE MARKERS and press the Change

 

Value softkey.

 

The Select Code markers screen appears.

 

 

 

System Configuration

 

 

 

 

 

General Settings

 

 

 

 

 

 

 

 

 

 

 

 

DEFIB

 

 

Config/Review

 

 

 

CODE MARKERS

Set/Review

 

 

 

 

 

 

MONITOR

 

 

Config/Review

 

 

 

CODE MARKERS

 

 

 

 

 

 

 

PACE

 

 

Config/Review

 

 

 

CODE MARKERS

 

 

 

 

 

 

 

 

Page 1 of 1

 

 

 

 

Change

 

Prev

Next

Enter

 

 

 

 

 

 

 

Value

 

Item

Item

Return

 

 

 

 

 

 

 

 

 

 

3.

Select either DEFIB CODE MARKERS,

 

 

 

 

 

 

 

 

 

MONITOR CODE MARKERS, or PACE CODE

 

 

 

 

 

 

 

 

 

MARKERS, then press the Change Value

 

 

 

System Configuration

 

 

softkey to toggle between Change/Review or Set

 

Defib Markers

Markers List

 

to Defaults.

 

 

 

 

 

 

 

CPR

 

 

 

Item 1: CPR

 

 

 

 

If you select Set to Defaults, press the Enter

 

Item 4: LIDO

 

 

DOPA

 

 

 

 

Item 2: Intubate

 

 

EPI

 

 

softkey to set the chosen code markers to

 

Item 3: EPI

 

 

Atrop

 

 

 

Item 6: B-block

 

 

Bicarb

 

 

their default values.

 

Item 5: Narcan

 

 

PHEN

 

 

 

Item 8: Mag Sulf

 

 

Oxygen

 

 

 

 

Item 7: Atrovent

 

 

Aspirin

 

 

If you select Change/Review, press the Enter

Change

 

Prev

Next

Delete

Return

 

 

 

softkey to access the screen that allows you

Value

 

Item

Item

to End

 

 

 

 

 

 

 

 

 

 

to change the marker, as shown.

 

 

 

 

 

 

 

 

 

Note: This example shows the screen that

 

 

 

 

 

 

 

 

 

allows you to change DEFIB CODE

 

 

 

 

 

 

 

 

 

MARKERS.

 

 

 

 

 

 

 

 

 

Only 8 items are displayed on the

 

 

 

 

 

 

 

 

 

screen at one time. When you get to the

 

 

 

 

 

 

 

 

 

last item on the screen and press the

 

 

 

 

 

 

 

 

 

Next Item softkey, you see the next 8

 

 

 

 

 

 

 

 

 

items. When you get to the last item on

 

 

 

 

 

 

 

 

 

that screen and press the Next Item

 

 

 

 

 

 

 

 

 

softkey, you see the remaining 4 items.

 

 

 

 

 

 

 

 

9650-0902-01 Rev. K

17

Loading...
+ 51 hidden pages