Zeiss OPMI Pentero User Manual

OPMI® Pentero

Software Release 2.2

®
Brief instructions
G-30-1702-en
Issue2.0
Printed on 17. 03. 2008

Contents

Key to symbols Page 3
Powering up and positioning the system Page 4
Positioning the microscope Page 5
Attaching sterile drapes Page 6
Balancing the system Page 7
Central user interface (touchscreen) Page 8
Menu structure Page 9
Illumination Page 10
Lamp change Changing the lamp module
Adjusting the position of the handgrips Page 12
Configuring the handgrips Page 13
Configuring the foot control panel Page 14
Mounting the tube and eyepieces Page 15
Attaching documentation / coobservation equipment Page 16
Adjusting the tubes and eyepieces Page 17
Setting the eyepieces (compensation for ametropia) Page 18
Setting the adjustment speeds of focus and zoom Page 19
OR layout drawings Page 20
Exporting patient data to CD/DVD/USB/DICOM Page 22
Connecting an external monitor Page 24
What to do in an emergency - brief instructions Page 26
Lamp change during power failure Failure of the zoom function Failure of the focusing function Failure of the magnetic brakes Failure of the touchscreen Failure of the line voltage Error messages in the data injection system and on the touchscreen Failure of all control functions (Emergency mode) Blocking of individual magnetic brakes
Page 11 Page 11
Page 28 Page 30 Page 31 Page 32 Page 32 Page 32 Page 33 Page 34 Page 35
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008

Key to symbols

Caution:

These brief instructions are only an excerpt from the complete user manual. They cannot replace the detailed description provided in the user manual. Therefore, make yourself thoroughly familiar with the comments and warning notes included in the complete user manual G-30-1458.
Different symbols used in this manual draw your attention to safety as­pects and useful tips. These symbols are explained in the following.
Warning!
The warning triangle indicates potential sources of danger which may constitute a risk of injury for the user or a health hazard.
Caution:
The square indicates situations which may lead to malfunction, defects, collision or damage of the system.
Note: The hand indicates hints on the use of the system or other tips for the user.
Read the user manual!
®
OPMI
GmbH
and Pentero® are registered trademarks of Carl Zeiss Surgical
.
AutoDrape™, Superlux, FlexiTrack™ and MultiVision™ are trade­marks of
Carl Zeiss Surgical GmbH.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
O
P
M
I
P
e
n
t
e
r
o
On / Off
2
1
3

Switching on the system

Warning!
The correct use of the system is absolutely vital for safe operation.
Before starting, make sure to read the chapters "Safety" and "Prepa-
Therefore, please thoroughly familiarize yourself with the content of user manual G-30-1458 before starting up the system.
rations for use" of the user manual!
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008

Positioning the microscope

G-30-1458
AB
G-30-1458
Warning!
For safety reasons, the system must only be used when correctly bal­anced. Before starting, make sure to read the chapter "Preparations for use" of the user manual!
Press the brake release buttons (AB) to unlock all magnetic brakes. Keep the brake release buttons (AB) pressed and move the ceiling mount into the working position required using the handgrips. When you let go of the brake release buttons (AB), all magnetic brakes are locked.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
12

Positioning the system at the operating table

The stand is provided with a handle which allows the reliable and easy guidance of the system during relocation. Please only use this handle for this purpose.
The stand base is equipped with the new FlexiTrak technology which makes it considerably easier for you to move and position the system in the OR. Two additional pedals on the base allow you to conveniently se­lect the required condition of the stand:
None of the pedals has been pressed:
Exact and sensitive positioning in the OR and at the operating table in all directions with a minimum of effort.
Right-hand pedal (2) is pressed down hard until it snaps in:
Reliable and precise straight-ahead movement, also over low thresh­olds (elevator). Press this pedal until it snaps in to set the two front casters for straight-ahead travel. The other casters remain steerable. When you slightly press pedal (1), all four casters will be steerable again.
Pressing left-hand pedal (1) to the stop will lock the stand base in po-
sition, preventing it from inadvertently rolling away. After the system has been finally positioned at the operating table, you should step on this pedal.
3 Possible positions for cranial procedures 4 Possible positions for face-to-face procedures (spine)
5 Surgeon 6 Assistant
Note: The system allows overhead positioning and can therefore be placed in any position behind the surgeon or assistant.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
3
4
5
6
5
6
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
DRAPE
Hier luftdicht abschließen
Make airtight here
Fermer hermétiquement ici
Cerrar hermélicamente aqui
Effective range of drape vacuum system
Sterile area
Non-sterile area
PHOTO
AUTOBALA NCE
PAT-FI LES
USER
REC START
VOICE CTRL
DICTATION
TELEPHONE
FREEZE
FULL SCREEN
FOCUS
359 mm
LIGHT
33%
ZOOM
5.9x
FOCUS SPEED
ZOOM SPEED
LIGHT
INTENSITY
DRAPE
CONFIG
MENU
2
1

Attaching sterile drapes

Caution:
When attaching the drape, please ensure that there is sufficient space for the tilt and rotation movements of the surgical microscope.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008

Balancing the system

SB
AB
Fine adjustment for mouth switch
G-30-1458
PHOTO
DRAPE
FREEZE
Autobalance Complete System
MENU
START
START
START
Autobalance Microscope
Drape Compensation
Autobalance
30. 01. 04
17:56
User:
XXXXXXXX
Patient:
XXXXXXX
Recorder Capacity:
Int:.......% Ext:.........%
USER
VOICE CTRL
DICTATION
TELEPHONE
FULL SCREEN
CLOSE
1
2
CONFIG
PAT-FILES
Fine adjustment mouth switch
-3
-
+
REC START
STOP
AUTOBALANCE
Warning! Make sure to observe the instructions in the chapter "Balancing the system" of user manual G-30-1458. For safety reasons, the system must only be used when correctly bal­anced. Despite the autobalance function, it may happen in exceptional cases that the surgical microscope is not correctly balanced. With an incorrectly balanced system, brake release may lead to uncon­trolled movements of the suspension system. For this reason, the bal­ancing procedure and the subsequent test must not be performed above the patient and only at a safe distance from other persons and devices.
Hold the microscope tightly at both handgrips.
Unlock the magnetic brakes (AB) and swing out the microscope until indicator (1) is centered in the blue area (2). Then you can perform the "Autobalance Complete System" procedure.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
10
MENU
USER CONFIG
Start
PAT-FILES

Central user interface (touchscreen)

Note: For detailed information on the submenus and their configuration, please see user manual G-30-1458.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
11
PHOTO
PAT-FILES
USER
CONFIG
VOICE CTRL
TELEPHONE
FREEZE
FULL SCREEN
FOCUS
359 mm
LIGHT
33%
ZOOM
5.9x
45
%
ZOOM
SPEED
LIGHT
INTENSITY
MENU
30. 01. 04
17:56
User:
XXXXXXXX
Patient:
XXXXXXX
Recorder Cap acity:
Int:.......% Ext:.... .....%
FOCUS
359 mm
DRAPE
AUTOBALANCE
REC START
DICTATION

Menu structure

The menu is structured as follows:
MENU The main menu is constantly displayed after the system has been started. You can use it for triggering still camera and video recording, for auto­matic balancing of the system, for activating the drape vacuum system and for switching the illumination on and off. The full-screen mode permits you to view images or the live video signal in full display size.
PAT-FILES Use the Patient Files menu to save, edit and manage patient data, videos and images.
USER The USER menu permits you to save user-specific settings for several dif­ferent users. In addition, you can select several different languages for user guidance here.
CONFIG The CONFIG menu permits you to enter the settings for the microscope and suspension system parameters.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
12

Setting the illumination in the "Light" configuration menu

30. 01. 04
17:56
User:
XXXXXXXX
Patient:
XXXXXXX
Recorder Capacity:
Int:.......% Ext:.... .....%
G-30-1458
STAND
MENU
Intensity Start Value
Light
On
Off
OPMI
LIGHT
Light Intensity Control
On
Off
Low
Medium
AUDIO/VIDEO
High
Remaining lamp lifetime
LIGHT
Brightness
Speed
30. 12. 04
17:56
User: XXXXXXXX Patient: XXXXXXX Recorder Cap acity:
Int:.......% Ext:........ .%
Threshold Value Light Warning
SYSTEM INFO
CONFIG
USER
PAT-F ILES
25%
24%
Lamp 2
in use
200h
Yes
MULTIVISION
TELEPHONE
FLUORESCENCE
PHOTO
AUTOBALANCE
PAT-F ILES
USER
REC START
VOICE CRTL
DICTATION
TELEPHONE
FREEZE
FULL SCREEN
FOCUS
359 mm
LIGHT
ON
ZOOM
5.9x
LIGHT ON/OFF
FOCUS SPEED
ZOOM SPEED
LIGHT
INTENSITY
DRAPE
CONFIG
MENU
33%
3
2
1

Illumination

Warning!
Before starting, make sure to read the notes on "Risk of phototoxic injuries caused by high illumination intensity" in chapter "Safety"!

Setting the illumination on the touchscreen

G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
13
4
3
5
2
1
6
G-30-1458
2
3
1
defective
lamp

Changing the lamp module

Lamp change

Open the cover. The lever for changing the lamp moves out. Pull out the lever as far as it will go without using force. If the lever has not been properly pulled, the lamp will not be changed. Pull the lever again to make the lamp snap into position. Close the cover again.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
14
1
1

Adjusting the position of the handgrips

You can adjust the position of the handgrips to meet your specific needs.
Open locking lever (1) and adjust the mounting bar or the handgrip only to the position required.
Choose a position of the handgrips which is most convenient for the surgical procedure to be performed. Make sure that there is enough room between the handgrips and the accessories mounted on the mi­croscope.
Firmly re-tighten locking lever (1).
Note: The handgrips including the locking levers can be rotated through approx. 180°.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008

Configuring the handgrips

1
Focus + Zoom +
Focus - Zoom -
C
USER
AUDIO/VIDEO
LIGHT
LEFT
MENU
PAT-FILES
FOOT
SWITCH
MOTORS XY
BRAKES
Photo
Autofocus
OPMI
CONFIG
Photo
Videorec
Light Intensit y
Fluorescence
MultiVision
Aux
RESET
Focus +
Focus -
Light +
Autofocus
Zoom +
Zoom -
Light -
STAND
A
2
HANDGRIP
No function
Factory set tings
RIGHT
31.12. 04
23:56 User: XXXXXXXX Patient:
: Ext:.........%
MULTIVISION
SYSTEM INFO
TELEPHONE
FLUORESCENCE
CONFIG
B
1
FOCUS
ZOOM
D E
G-30-1458
You can trigger the functions of the surgical microscope using the buttons on the handgrips or on a foot control panel. Programmable buttons can be configured for the requirements of each user. Both handgrips are identical. In the basic setting, joystick (1) can be used for motorized fine movement in the XY directions. You can configure both handgrips with the same functions (CONFIG), or you can configure different functions in the right and left handgrip. The CONFIG menu / STAND / HANDGRIPS permits you to select the functions to be assigned to programmable buttons A, B, D or E. The rocker switches for Zoom and Focus can be reconfigured by the user. Briefly press button (C) on the touchscreen. The display and functions will be swapped.
15
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Configuring the handgrip buttons Press CONFIG to configure both handgrips. For separate programming of the left or right handgrip, press the LEFT or RIGHT button respectively. The buttons are only displayed if the CONFIG button has been pressed. Press the button (A, B, D, E) to which you would like to assign a different function.Then press the relevant function in selection menu (2) to assign it to the selected button. The labeling of the button is changed and indi­cates the new function. Press the RESET button to restore the factory settings.
Note
: The joystick on the right handgrip is used for special functions in op-
tional applications.
Printed on 17. 03. 2008
16
Zoom - Fokus -
Zoom + Fokus +
Fokus + Fokus -
Zoom + Zoom -
E
USER
AUDIO/VIDEO
LIGHT
MENU
PAT-FILES
Autofocus
HANDGRIPS
OPMI
CONFIG
FOOT SWITCH
MOTORS XY
RESET
STAND
BRAKES
ROCKER SWITCH
A
C
B
D
E
Videorec
Photo
Aux
MultiVision
Fluorescen ce
Light Intensity
CONTROL
PANEL
1
Factory settings
No function
MULTIVISION
SYSTEM INFO
TELEPHONE
FLUORESCENCE
30. 12.
17:56 User: XXXXXXXX Patient: XXXXXXX
Focus
Focus -
Photo
Autofocus
Zoom +
Zoom -
Videorec
MultiVision
Light -
Light +
FOOT SWITCH

Configuring the foot control panel

You can trigger the functions of the surgical microscope using the but­tons on the handgrips or on a foot control panel. Programmable but­tons can be configured for the requirements of each user. The joystick on the foot control panel permits motorized fine move­ment of the microscope in the XY directions.
The rocker switches for Zoom and Focus can be reconfigured by the user. Briefly press button (E) on the touchscreen. The display and functions will be swapped.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Configuring the programmable buttons A, B, C, D
Press the button (A, B, C, D) to which you would like to assign a dif­ferent function.
Then press the relevant function in selection menu (1) to assign it to the selected button. The labeling of the button is changed and indi­cates the new function.
Press the RESET button to restore the factory settings.
Printed on 17. 03. 2008

Mounting the tube and eyepieces

G-30-1458
978
1
55 63224
10
Warning!
Change the modules and accessories before surgery and without the pa­tient!
Loosen securing screw (8) by a few turns.
Remove cover (4).
17
Attach spine adapter (3) to the microscope body and firmly
tighten se-
curing screw (8).
Attach binocular tube (2) to spine adapter (3) and firmly
tighten secur-
ing screw (7).
Insert widefield eyepieces (1) as far as they will go in the mounts pro­vided on the binocular tube.
Attach binocular coobservation tube (5) to the microscope body and
tighten securing screw (9).
firmly
Insert widefield eyepieces (6) as far as they will go in the mounts pro­vided on the binocular tube.
Before starting up the system, remove protective cap (10) from the ob­jective.
Warning! Check that all modules are securely seated. Firmly
tighten se-
curing screws (8) and (9)! Balance the surgical microscope, page 9.
Printed on 17. 03. 2008
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
18
G-30-1458
312
4
2 3

Attaching documentation / coobservation equipment

Loosen knurled ring (1).
Remove dust cover (2) and keep it in a safe place.
Insert accessory module (3) in the mount of the image exit port and carefully turn it until the guide projections fit into the grooves. Push the accessory module into the mount as far as it will go.
Screw knurled ring (1) onto accessory module (3) and firmly
tighten it.
The type of coobservation (lateral image exit ports: left/right, or opposite image exit ports: face to face) can be configured at the touchscreen (CONFIG/OPMI/TUBE).
The sliding mirror has two positions:
Left/Right: The light is directed to the lateral image exit ports. Face to Face: The light is directed to the tube mount at the back.
The position of the sliding mirror can also be switched manually using knob (4).
Warning! Check that all modules are securely seated. Firmly
tighten se-
curing screws (8) and (9)! Balance the surgical microscope, page 9.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008

Adjusting the tubes and eyepieces

12
345 345
1 Adjusting the interpupillary distance
This is done by pressing the two eyepieces together or pulling them apart. The correct position has been reached when the two eyepiece images merge into one.
2 PD adjustment knob
The correct position has been reached when the two eyepiece images merge into one.
3 Eyecup
Adjust the eyecups in such a way that the entire field of view can be seen.
Viewing with eyeglasses: Screw in the eyecups all the way.
19
Viewing without eyeglass-
es:
4 Diopter setting ring
The eyepieces provide compensation for ametropia. Eyeglass wearers using their glasses during work should set the diopter setting ring to 0. Turn the ring until the optimum setting has been achieved (see following page).
5 Diopter scale
For reading off the prescription set.
Adapt the eyecups to the viewer's field of view by screwing them out­ward. The three white marking rings facilitate the adjustment.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
20
Bring the surgical microscope into its starting position within the focusing range. Adjust the minimum magnification on the surgical microscope. Bring the surgical microscope into the position required. Adjust your interpupillary distance on the binocular tube. Adjust your prescription on the eyepieces. Please note that instrument myopia can occur.
Emmetropes adjust the diopter scale to 0 diopter (D).
Eyeglass wearers (who perform surgery wearing their eyeglasses) adjust the diopter scale to 0 diopter (D).
Ametropes (who know their prescription and perform surgery without wearing their glasses) adjust the diopter sca­le to their prescription.
Ametropes (who do not know their prescription
and perform surgery without wea-
ring their glasses)
Adjust both eyepieces on the main microscope to +8 D. Adjust both eyepieces on the assistant's microscope to +8 D (if applicable).
Eyepieces without
reticle:
Eyepieces with
reticle:
Remove the binocular tube and the eyepieces from the microscope body and point them at a distant ob­ject
*)
, i.e. use them like a pair of bi­noculars. The image of the object is still unsharp.
Turn the diopter scale on the eyepiece slowly clockwise until the reticle is imaged sharply.
Turn the diopter scale of one eye­piece slowly
clockwise until the ob­ject is imaged sharply. If necessary, repeat this procedure three times and take the average of the rea­dings. Adjust the second eyepiece in the same manner.
Focus the surgical microscope on the object. The reticle and the object must be in focus at the same time.
Mount the tube and eyepieces on the microscope body and tighten the securing screw firmly
.
Adjust the second eyepiece in the same manner until the object is seen sharply.
Adjust the eyecups in such a way that you can see the full field of view. Adjust maximum magnification on the surgical microscope and focus on the object. Adjust the working magnification required. When the magnification is changed, the focal plane is retained, but the depth of field changes.
Note
: Enter the prescription values in the relevant user profile.
*) Warning! Never use the sun as the distant object!

Setting the eyepieces (compensation for ametropia)

G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008

Setting the adjustment speeds of focus and zoom

USER
STAND
AUDIO/VIDEO
MULTIVISION
LIGHT
SYSTEM INFO
TELEPHONE
FLUORESCENCE
Autofocus Automatic
Micromanipul ator
Focus Stop
Focus
Start Value
MENU
PAT-FILES
ZOOM
DIAPHRAGM
TUBE
On
Off
On
Off
On
Off
On
Off
50% 300 mm
Focus
Zoom Link
Focusing Aid
Laser Spots
FOCUS
OPMI
CONFIG
30. 12. 04
17:56 User: XXXXXXXX Patient: XXXXXXX
USER
STAND
AUDIO/VIDEO
LIGHT
MENU
PAT-F ILES
DIAPHRAGM TUBE
50%
Speed
FOCUS
OPMI
CONFIG
ZOOM
Gamma Start
MULTIVISION
SYSTEM INFO
TELEPHONE
FLUORESCENCE
30. 12. 04
17:56 User: XXXXXXXX Patient: XXXXXXX
1.2
Focus Speed
ZOOM
SPEED
FOCUS
SPEED
50%
ZOOM
5.9x
FOCUS 359 mm
MENU
ZOOM
SPEED
FOCUS SPEED
50%
ZOOM
5.9x
FOCUS 359 mm
MENU
1
323
4
1
2
3
4
5
In the main menu:
In the configuration menu:
Press the focus or zoom button in the main menu. Use the displayed slider to set the focus or zoom speed as required. These settings can also be performed in the configuration menu. Press CONFIG/OPMI and select the Focus or Zoom tab. Use the slider to set the focus or zoom speed as required.
21
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
22
USB
DVD

Exporting patient data to CD/DVD/USB stick/DICOM

Open PAT-FILES menu (1).
Open "LIST" tab (2).
Activate the required patient in list (3).
Open the image folder of the patient (4).
Select the images you want to export and mark them by pressing Mark button (5).
Press "SAVE" button (6).
Select the storage medium to be used in the SAVE menu: CD/DVD, USB STICK or DICOM (7).
Note: The display shows the number of images selected for export.
Insert a new, blank data carrier in the CD/DVD drive or plug a USB stick into the port on the touchscreen (8).
9 Export in the DICOM format
See DICOM option
10 Save images anonymously
Enables anonymous saving of the selected images.
Press "COPY" (11) to export the images. The data export process is indicated by a progress bar.
Press the button on the CD/DVD drive to remove the CD/DVD.
In the local image folder, exported images are marked by the storage symbol .
Note: The system is not intended for the permanent saving of data. You can use CDs/DVDs, a USB stick, an external hard drive or a hospital server (DICOM option) for data backup. All users are responsible for saving their own data.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
1
!
Files are being transferred This process may take a few minutes.
Data export
CANCEL
2
3
8
11
5
6
4
7
5
9
10
30. 12. 04
17:56 User: XXXXXXXX Patient: Benninger Ra lph Recorder Capacity:
Int:.......% Ext:..... ....%
PAT-FILES
MENU
USER
CONFIG
PATIENT
IMAGES
VIDEOCLIPS
LOAD
SAVE
BACK
SORT
Benninger Ra lph
Bachmann E.
Busch W.
ADD
DELETE
EDIT
LIST
30. 12. 04
17:56 User: XXXXXXXX Patient: Benninger Ral ph Recorder Capacity:
Int:.......% Ext:......... %
PAT-FILES
MENU
USER
CONFIG
PATIENT
IMAGES
VIDEOCLIPS
LOAD
SAVE
BACK
NEXT
EDIT
SELECT ALL
FULL
Mark
Delete
INDEX
1 X
4 X
12 X
30. 12. 04
17:56
User: XXXXXXXX Patient: Benninger Ral ph Recorder Capacity:
Int:.......% Ext:.........%
PAT-FILES
MENU
USER
CONFIG
PATIENT
IMAGES
VIDEOCLIPS
LOAD
SAVE
3
COPY
Archive images anonymously
Export in DICOM format
0
0
CD/DVD
USB STICK
DICOM
On
On
Off
On
Off
23
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
24

Connecting an external monitor

Caution:
Only connect the monitor to a wall outlet which is provided with a properly connected protective ground conductor.
1 Video signal output port BNC (VBS)
for connecting an external monitor. Suitable for devices capable of processing VBS signals, or if video sig­nals have to be transmitted over relatively long distances.
2 Video signal output port for an external monitor (Y/C)
for connecting a further external monitor. Y/C provides higher video image quality than VBS.
The two ports on the control panel are independent of each other and can be used as required.
Note: Monitors connected to the video signal output ports (BNC or Y/C) display the live image of the camera.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
25
BNC1
Y/C2
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
26

What to do in an emergency

1Failure of illumination - changing the xenon lamp:
Open flap (1)
Change the xenon lamp by pulling grip (8) page 28
2 Failure of the zoom function:
Manually adjust the magnification using zoom knob (2). If the motorized zoom function becomes active of its own accord (e.g. travels to the stop), set emergency switch (7) to the emergency mode (position 2).
3 Failure of the focusing function:
Manually adjust the working distance using focusing knob (3). If the motorized focusing function becomes active of its own accord (e.g. travels to the stop), set the emergency switch to the emergency mode (posi­tion 2)
4 Some of the magnetic brakes are blocked:
page 30
page 31
Switch off power switch (4). As soon as the blue screen appears (approx. 10 sec), switch the system back on.
The OPMI functions (zoom, focus, light and magnetic brakes) are available again after approx. 15 seconds. The computer and touchscreen, however, are disa­bled.
If the magnetic brakes are still blocked:
Hold the microscope on its body (not on the handgrips) and position it manually by overcoming the braking ef­fect.
5 Failure of the touchscreen:
Do not under any circumstances touch the touch­screen, since this can result in changes to settings and parameters. Zoom, focus, illumination and brakes can still be operated.
6 Error messages in the data injection system:
System errors are displayed in the microscope's inte­grated data injection system. You can delete these messages by acknowledgement using the joystick of the right handgrip (pushbutton) or the touchscreen.
page 35
page 32
page 32
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
7 Failure of control functions:
5
1
4
7
2
3
8
6
Set emergency switch (7) to the emergency mode (po­sition 2). Zoom and focus must then be operated man­ually (2, 3).
Hold the microscope on its body (not on the handgrips) and position it manually by overcoming the braking ef­fect.
27
page 34
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
28

Emergency instructions - overview

Illumination failure - changing the xenon lamp

Open flap (3) as far as it will go. Lever (1) for changing the lamp moves out.
Pull out lever (1) as far as it will go, without using too much force. If you do not properly pull the lever, the lamp is not changed. Pull the lever again to make the lamp snap in.
Close the flap again.
Order a new xenon lamp. Cat. No. (2) is shown on the lamp module.
Note: – Only change the used lamp if replacement is available on site.
Always operate the illumination system with both xenon lamps. If one of the lamp fails, switch to the second lamp. Leave the failed lamp in the lamp housing until it is replaced by a new lamp.
Every time you switch on the system, it indicates that you are using
the backup lamp, until the used lamp is replaced.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
29
1
2
3
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
30
1

Failure of the zoom function

Failure of the zoom function while all other functions of the suspen­sion system work correctly
Adjust the magnification manually using zoom knob (1).

Failure of the zoom drive

In the event of a failure of the zoom drive (e.g. zoom drive moves con­stantly into an end position):
Set the mode selector switch on the connection panel to emergency operation. This ensures that the light source continues to be operative without any restrictions.
Adjust the magnification manually using zoom knob (1).
Continue surgery by manually operating the suspension system and the surgical microscope by overcoming the locking effect of the brakes.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008

Failure of the focusing function

1
Failure of the focusing function while all other functions of the sus­pension system work correctly
Adjust the working distance manually using focusing knob (1).

Failure of the focusing drive

In the event of a malfunction of the focusing drive (e.g. focusing drive moves constantly into an end position):
Set the mode selector switch on the connection panel to emergency operation. This ensures that the light source continues to be operative without any restrictions.
Adjust the working distance manually using focusing knob (1).
31
Continue surgery by manually operating the suspension system and the surgical microscope by overcoming the locking effect of the brakes.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
32

Failure of the magnetic brakes

If the magnetic brakes fail (magnetic brakes are locked), you can manu­ally move the axes of the stand by overcoming the locking effect of the brakes.

Touchscreen failure

Warning!
Do not under any circumstances touch the touchscreen. There is a possi­bility that only the illumination of the display has failed. In this case, you would inadvertently call menus or change settings when touching the screen.
Note: If the touchscreen no longer responds to your entries, you can still con­tinue to operate the OPMI Pentero with the current settings.
To restore the touchscreen function, proceed as follows:
Switch the suspension system off and back on again after a short time
Check all functions of the surgical microscope and suspension sys-
You can continue surgery in the standard operating mode.

Failure of the line voltage

In the event of power failure, the uninterruptible power supply (UPS) au­tomatically starts to operate. It ensures for a short period of time that no data will be lost.
(approx. 2 minutes).
After power-up, the system automatically goes into the standard oper­ating mode after it has completed a self-test (approx. 30 seconds).
tem.
Note: The UPS does not power the light source and microscope functions.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
If no power is available over a prolonged period of time, the system is shut
1
down. As soon as line power is back, the user is informed accordingly (Power OK) and all subsystems are re-initialized.
In the event of power failure (magnetic brakes are locked), you can man­ually position the arm of the suspension system with the microscope by overcoming the locking effect of the brakes.

Error messages in the data injection system and on the touchscreen

Major system errors are displayed in the microscope's integrated data in­jection system. You can delete these messages by acknowledgement using joystick (1) of the right handgrip (pushbutton) or the touchscreen.
33
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
34
1
2
3

Failure of all control functions (Emergency mode)

Check whether power switch (3) has been switched on.
Remove magnetic plate (1) which covers emergency mode switch (2).
Set switch (2) from position 1 to position 2 (emergency mode).
The microscope and suspension system functions have now been com­pletely deactivated. The emergency mode ensures that the illumination continues to function with constant and sufficient intensity.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008

Individual magnetic brakes are blocked (OPMI can not be moved at all or only to a limited extent)

It is possible that the integrated safety switches are activated and block some magnetic brakes. This can considerably restrict maneuverability of the microscope. Proceed as follows to reset the microscope functions and to allow surgery to be finished:
Turn off the system at the power switch. It can be switched on again as soon as the blue screen appears (approx. 10 seconds).
The system is operative again after approx. 15 seconds.
Computer and touchscreen are inoperative.
The following microscope functions are still available:
35
Motorized focusing via handgrip or foot control panel
Motorized magnification setting via handgrip or foot control panel
Focus and zoom speeds
function
Brake functions AB (All Brakes) and SB (Selected Brakes) on the
handgrip
Brightness setting of xenon illumination via handgrip or foot control
panel. Perhaps only a constant illumination value of approx. 70% (depending on the respective cause of the defect).
Video image on an external monitor (external video cable of micro-
scope camera).
System balancing is retained with the same setting as before the mal-
function. Unfortunately, new autobalance is not possible. Therefore, please do not change the OPMI configuration.
After completion of surgery, we recommend you to contact the respon-
sible Carl Zeiss service team to ensure that no hardware defect is present.
correspond to the speeds set before the mal-
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
36
0297

CE label

Subject to change

The system meets the essential requirements stipulated in Annex I to the 93/42/EEC Directive governing medical devices. The system is labeled with:
Subject to changes in design and scope of delivery as a result of ongoing technical development.
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008

Index

Index 37
A
Adjustment speeds of focus and zoom, setting of Archiving, not permanent 22
B
Brief instructions, no replacement for user manual
C
CE label Changing the xenon lamp 28 Compensation for ametropia 20 Computer and touchscreen are inoperative. 35 CONFIG 11 Configuring the foot control panel 16 Connecting the stand 6 Control functions, failure of 34
36
21
3
D
Data export to CD/DVD/USB/DICOM Diopter scale 19 Diopter setting ring 19 Drapes, attaching 8
E
Emergency instructions, overview Emergency mode 34 Emergency, what to do 26 Error messages in the data injection system and on the touchscreen 33 Exporting patient data to CD/DVD/USB/DICOM 22 External monitor, connection of 24 Eyecup 19 Eyepieces, mounting of 17 Eyepieces, setting of 20
F
Failure of all control functions Failure of magnetic brakes 32 Failure of the focusing function 31 Failure of the graphic touchscreen 32 Failure of xenon lamp 28 Focusing function, failure 31
34
22
28
H
Handgrips, adjustment
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
14
Printed on 17. 03. 2008
38 Index
I
Individual magnetic brakes are locked Interpupillary distance 19
L
Lamp change
M
Magnetic brakes, failure Main menu 11 MENU 11
P
PAT-FILES PD adjustment knob 19 Positioning at operating table, system 6
R
Risk of phototoxic injuries
35
28
32
11
12
S
Subject to change Switching on the system 4 System, positioning at operating table 6
T
Touchscreen Tube, mounting of 17 Tubes and eyepieces, adjustment of 19
U
Uninterruptible power supply USER 11 User interface (touchscreen) 10
V
Video signal output port BNC (VBS) Video signal output port for an external monitor (Y/C) 24
X
Xenon lamp, failure
Z
Zoom function, failure
36
10
32
24
28
30
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
39
G-30-1702-en OPMI® Pentero® Software Release 2.2 Issue 2.0
Printed on 17. 03. 2008
Carl Zeiss Surgical GmbH
A Carl Zeiss Meditec Company
73446 Oberkochen Germany
Fax: ++49 (0) 7364-20 48 23 Email: surgical@meditec.zeiss.com Internet: www.meditec.zeiss.com
G-30-1702-en Printed in Germany CG-ST
Loading...