Zeiss N HBO 103, N XBO 75 Operating Manual

Page 1
N HBO 103 and N XBO 75
Microscope Illuminating
Systems
Operating Manual
Page 2
N HBO 103 and N XBO 75
Carl Zeiss Copyright Microscope Illuminating Systems
Knowledge of this manual is required for the operation of the instruments. Would you therefore please make yourself familiar with the contents of this manual and pay special attention to hints concerning the safe operation of the instrument.
The specifications are subject to change.
Unless expressly authorized, forwarding and duplication of this document, and the utilization and
communication of its contents are not permitted. Violations will entail an obligation to pay compensation. All rights reserved in the event of granting of patents or registration of a utility model.
Published by: Carl Zeiss
Microscopy
D-07740 Jena Phone: ++49-36 41 64-16 16 Telefax: ++49-36 41 64-31 44 E-mail: micro@zeiss.de Internet: www.zeiss.de/micro
Number of this manual: B 40-065 e Date of issue 08/99
2 B 40-065 e 08/99
Page 3
N HBO 103 and N XBO 75 Microscope Illuminating Systems Contents Carl Zeiss
Contents
1 Notes on Safety....................................................................................................4
1.1 General ..............................................................................................................................4
1.2 Regulations concerning Instrument Safety and Electromagnetic Compatibility (EMC) ..........4
1.3 Notes on Unpacking, Transport and Storage.......................................................................5
1.4 Hints on Use .......................................................................................................................5
1.5 Hints on Maintenance and Care ......................................................................................... 6
1.6 Hints on Disposal ................................................................................................................6
1.7 Hints on Warranty ..............................................................................................................6
2 Instrument Description .......................................................................................7
2.1 Components of the Microscope Illuminating Systems .........................................................7
2.2 Technical Data of the Microscope Illuminating Systems....................................................... 8
3 Start-Up ...............................................................................................................10
3.1 N HBO 103 Microscope Lamp - Lamp Insertion/Change .................................................... 10
3.2 N XBO 75 Microscope Lamp - Lamp Insertion/Change ......................................................12
3.3 Start-Up of N HBO 103 Microscope Lamp .........................................................................14
3.4 Start-Up of N XBO 75 Microscope Lamp ...........................................................................15
3.5 Focus and Center the HBO 103/XBO 75 Lamps in the Microscope Lamp Housing ......................16
3.6 Change the Collector........................................................................................................17
Illustrations
Fig. 1 Overall view of the N HBO 103 microscope illuminating system ..........................................7
Fig. 2 Overall view of the N XBO 75 microscope illuminating system ............................................7
Fig. 3 Line spectrum of the HBO 100/103 mercury short-arc lamps...............................................9
Fig. 4 Line spectrum of the XBO 75 xenon short-arc lamp ............................................................9
Fig. 5 Insert HBO 103 lamp ........................................................................................................10
Fig. 6 Insert XBO 75 lamp........................................................................................................... 12
Fig. 7 Rear of transformer ebq 100 dc........................................................................................14
Fig. 8 Front of transformer ebq 100 dc.......................................................................................14
Fig. 9 Rear of transformer ebq 75 isolated..................................................................................15
Fig. 10 Front of transformer ebq 75 isolated ................................................................................15
Fig. 11 Lamp arc and mirror image...............................................................................................16
Fig. 12 Adjusting components of microscope lamp housing .........................................................16
Fig. 13 Change the collector ........................................................................................................17
Certificate for quality management system in accordance with DIN EN ISO 9001/DIN EN 46001
EC Declaration of Conformity in accordance with EC directives 73/23/EEC and 89/336/EEC
B 40-065 e 08/99 3
Page 4
Safety N HBO 103 and N XBO 75
Carl Zeiss General, EMC Microscope Illuminating Systems
1 Notes on Safety
1.1 General
Please make yourself familiar with the contents of these operating instructions before attaching and starting up the N HBO 103 and N XBO 75 microscope illuminators. Additional information is available from our maintenance service or from authorized agencies.
The N HBO 103 and N XBO 75 microscope illuminators should be used with Zeiss microscopes exclusively.
The N HBO 103 and N XBO 75 microscope illuminators may only be handled by trained personnel.
To ensure that the illuminators function properly with the microscope, the appropriate microscope manual must be observed.
The manuals "ebq 100 dc – Electronic transformer for the N HBO 103 microscope illuminator“ and "ebx 75 isolated – Electronic transformer for
the N XBO 75 microscope illuminator“
must be observed. Special regard must be paid to the notes on instrument safety included in this manual.
The data sheets from the lamp manufacturers must be observed when the HBO 103 and XBO 75 lamps are operated.
To guarantee the safe operation and function of the N HBO 103 and N XBO 75 microscope illuminators, it is necessary under any circumstances to take the precaution measures and observe the warnings contained in the operating instructions.
The following symbols are used for the warnings:
CAUTION!
Non-observance of the safety notes constitutes a hazard for the user.
CAUTION!
Dangerous electrical voltage
CAUTION!
Disconnect the instrument from the line before opening it!
CAUTION!
Hot surface
CAUTION!
Non-observance of the safety notes constitutes the risk of damage to the instrument.
Notes which must be observed when using the N HBO 103 and N XBO 75 microscope illuminators.
NOTE!
1.2 Regulations concerning Instrument Safety and Electromagnetic Compatibility (EMC)
The N HBO 103 and N XBO 75 microscope illuminators have been designed and tested in compliance with EN 61010, Part 1 (VDE 0411). From the safety viewpoint, they left the factory in a perfect state.
They meet the requirements of EC directives 73/23 and 89/336 and the EMC legislation, version of Nov.9, 1992. Radio-screened in accordance with EN 55011, Class B, insensitivity to noise in accordance with EN 50082-2.
Compliance with the EC directives mentioned is documented by the CE-certificate.
4 B 40-065 e 08/99
Page 5
N HBO 103 and N XBO 75 Safety Microscope Illuminating Systems Notes on unpacking, transport, storage and use Carl Zeiss
1.3 Notes on Unpacking, Transport and Storage
The lamps, especially the XBO 75, are under high pressure even if they are not switched on. They may therefore be removed from the package only if the protection mask (Cat.No.
417009) and the safety gloves with artery
protection (Cat. No. 417008) are worn. The XBO 75 has an additional plastic protection sleeve which must not be removed when the lamp is unpacked!
For safety reasons, a safety holder is inserted in the microscope illuminator instead of the bulb during transport. Both the N HBO 103 and N XBO 75 microscope illuminators and the HBO 103 and XBO 75 lamps must be transported in their original packaging. This packaging and the plastic protection sleeve of the XBO 75 must be kept for storage and disposal.
The transport and storage temperatures must be observed in the same way as those of the microscope stands.
1.4 Hints on Use
The lamps, especially the XBO 75, are under
high pressure even if they are not switched on. They may only be inserted in the microscope illuminator if a safety mask and safety gloves (see chapter 1.3) are worn. The safety regulations of the lamp manufacturer must be observed when the XBO 75 is inserted.
The HBO 103 and XBO 75 lamps are under
high pressure during operation; they must always be opereated in closed lamp housings!
The light of the HBO 103 and XBO 75 short-
arc lamps contains a high amount of UV light:
Never look directly into the lamp or in
reflected light
Wear safety glasses (sunglasses) when
operating the lamps
Cover inspection openings with filters
The UV radiation of the lamps produces
harmful ozone. This gas must be carried off to the outside via exhauster tubes and the working room supplied with fresh air.
Both the N HBO 103 and N XBO 75
microscope illuminators and the HBO 103 and XBO 75 lamps must be protected from impact and humidity.
Instead of the HBO 103, its predecessor HBO 100 can also be used.
B 40-065 e 08/99 5
Page 6
Instrument Description N HBO 103 and N XBO 75
Carl Zeiss Technical Data Microscope Illuminating Systems
1.5 Hints on Maintenance and Care
To enable the microscope illuminators to function perfectly, it is necessary to use only components and spares marketed or admitted by us. When in doubt, please contact our service staff.
With the exception of the work described in this manual, the user must not perform any repairs or changes himself. We would like to emphasize that all other changes and repairs must be performed by specially authorized personnel.
Defective instruments or components may only be repaired and maintained by our service department.
After expiry of the average operating time
(see data sheets from the lamp manufacturers), the lamps must be exchanged; there is a risk of explosion! The operating time of the HBO 103 and XBO 75 can be checked on the counter of the relevant transformer. (Fig. 5/Fig. 8) The intensity of the lamps decreases when the operating time increases, i.e. homogeneous illumination of the object field can then no longer be guaranteed.
Always disconnect the instrument from the
line before each lamp exchange or any other work inside the instrument.
The safety regulations of the lamp
manufacturer must be observed when the XBO 75 is changed.
Check the cables of the N HBO 103 and N
XBO 75 when changing the lamp.
Clean the housing of the microscope
illuminator when it is dirty.
Do not use any solvents. Soak a cleaning cloth in a mixture of water and washing-up liquid, remove the dirt and dry the housing using a clean cloth.
1.6 Hints on Disposal
Used HBO 103 and XBO 75 lamps and defective microscope illuminators should not be treated like domestic waste; they should be disposed of in their original packaging in accordance with the legal regulations.
When in doubt, please contact the Carl Zeiss microscopy service.
The lamps must be exchanged as described
in chapters 3.1/3.2.
Before each lamp exchange, make sure to
allow the N HBO 103 or N XBO 75 microscope illuminator to cool down sufficiently long in order to reduce the pressure; wait at least for 15 minutes!
6 B 40-065 e 08/99
1.7 Hints on Warranty
The instrument manufacturer is not liable for damage caused by unauthorized operation, particularly the removal and exchange of components, or the use of accessories from other manufacturers. This forfeits all the claims against warranty.
Page 7
N HBO 103 and N XBO 75 Instrument Description Microscope Illuminating Systems Components, Catalogue Numbers Carl Zeiss
2 Instrument
Description
2.1 Components of the Microscope Illuminating Systems
1 Lamp housing 2 Connecting cable 5 Lamp base 3 Collector 6 Transformer ebq dc 4 Dovetail 7 Line cable
Fig. 1 Overall view of the N HBO 103
Microscope Illuminating System
1 Lamp housing 2 Connecting cable 5 Lamp base 3 Collector 6 Transformer ebx 75 isolated 4 Dovetail 7 Line cable
Fig. 2 Overall view of the N XBO 75
Microscope Illuminating System
The N HBO 103 and N XBO 75 microscope illuminators include the following components (Fig. 1 / Fig. 2):
Illuminator with:
dovetail for attachment to the microscope, reflector, double-lens collector or quartz collector for UV light, heat-protection filter
HBO 103 W/2 mercury vapor short-arc lamp
or XBO 75 W/2 xenon short-arc lamp
transformer ebq 100 dc for N HBO 103
or transformer ebx 75 isolated for N XBO 75
line cable with earth-contact plug
or line cable with American flat plug
HBO 103 XBO 75
Microscope Illuminating Systems
Collector
Quartz collector
Lamp
Transformer ebq 100 dc
Line cable
Table: Catalogue numbers
Lamp N HBO 103
000000-1007-980
Collector N HBO 103/XBO 75
Quartz collector N HBO 103/XBO 75
Mercury vapor-
short-arc lamp
HBO 103 W/2
380301-9350-000
000000-1003-928
Line cable with earth-contact plug
Line cable with American flat plug
000000-1007-976
000000-1007-977
380137-6750-000
380137-6740-000
Lamp N XBO 75
000000-1007-981
Xenon-
short-arc lamp
XBO 75 W/2
380079-9190-000
ebx 75 isolated
000000-1003-924
B 40-065 e 08/99 7
Page 8
Instrument Description N HBO 103 and N XBO 75
Carl Zeiss Technical Data Microscope Illuminating Systems
2.2 Technical Data of the Microscope Illuminating Systems
Lamp housing Lamp housing N HBO 103
Cat. No. 000000-1007-980
Length 163 mm 228 mm
Width 141 mm 141 mm
Height 200 mm 200 mm
Collector mount
Connection to stand dovetail dovetail
Connection to transformer Cable with special plug Cable with special plug
Transformer
Cat.No. 000000-1003-928
Line connection 90-250 V AC 90-250 V AC
Line frequency 50-60 Hz 50-60 Hz
Power consumption 265 VA 130 VA
Fuse F1/F2 T 3.15 A F1/F2 T 2 A
Lamp
40 mm 40 mm
Transformer N for
HBO 103
Mercury vapor short-arc lamp
Lamp housing N XBO 75
Cat.No. 000000-1007-981
Transformer N for
XBO 75
Cat. No. 000000-1003-924
Xenon-
short-arc lamp
For the technical data of the lamps please see the information provided by the manufacturer. Recommended:
Average life of the lamps OSRAM
OSRAM
Type HBO 103 W/2 Cat. No. 380301-9350-000
Type HBO 100 W/2 200 h Type HBO 103 W/2 300 h
OSRAM
Type XBO 75 W/2 Cat. No. 417030-9010-000 Type XBO 75 W/2 Cat. No. 417030-9011-000
Hamamatsu
Type Xenon lamp L217401 UV Cat.No. 380079-9190-000 Type Xenon lamp L217401 OFR Cat. No. 380053-9870-000
OSRAM
Type XBO 75 W/2 400 h
Hamamatsu
Type Xenon lamp 2000 h
8 B 40-065 e 08/99
Page 9
N HBO 103 and N XBO 75 Technical Data Microscope Illuminating Systems Line spectra Carl Zeiss
Fig. 3 Line spectrum of the HBO 100/103 mercury short-arc lamps
Fig. 4 Line spectrum of the XBO 75 xenon short-arc lamp
B 40-065 e 08/99 9
Page 10
Start-Up N HBO 103 and N XBO 75
Carl Zeiss HBO 103 Lamp Insertion/Change Microscope Illuminating Systems
3 Start-Up
3.1 N HBO 103 Microscope Lamp Lamp Insertion/Change
Before inserting the lamp, make sure that
no connection to the line is available.
The lamps may only be changed when they
are cooled down; allow the microscope illuminator to cool down for 10...15 minutes to avoid the risks of burns and explosion!
The lamp may only be removed from the
packaging and inserted in the microscope illuminator if a protection mask and safety gloves are worn (see chapter 1.3).
All clamping adjustments must be performed
carefully; pronounced heat during operation may result in loose contacts.
1 Fixation screw for lamp housing 2 Control wheel for collector 3 Cooling body 4 Fixation screw for lamp mount in cooling body 5 Spring lever 6 Lamp mount 7Lamp
Fig. 5 Insert HBO 103 lamp
10 B 40-065 e 08/99
Page 11
N HBO 103 and N XBO 75 Start-Up Microscope Illuminating Systems HBO 103 Lamp Insertion/Change Carl Zeiss
Procedure:
Switch off the transformer
Disconnect the microscope lamp housing
from the transformer
Allow the microscope lamp housing to cool
down for 10...15 minutes
Wear protection mask and safety gloves.
Remove microscope lamp housing with
cooled-down lamp from the microscope
Move collector in front position by turning
the control wheel (5-2) in the direction of the arrow
Loosen fixation screw (5-1) for lamp
housing; remove the lamp housing
Press spring lever (5-5) downwards to loosen
the lamp locking; remove defective lamp (5-7) with cooling body (5-3) from the mount (5-6) in the lamp base.
Loosen fixation screw (5-4) on the cooling
body: remove defective lamp and dispose of it in accordance with the appropriate regulations (see section 1-6 - Notes on disposal) For initial start-up, remove the safety appliance first.
Tighten the fixation screw on the cooling
body.
Insert lamp with cooling body in the relevant
mount in the lamp base and make sure that the lateral lead to the cooling body remains accessible; use the spring lever for fixation.
Do not exert any force when inserting the lamp.
Attach lamp housing to the lamp base; make
sure that the contact pin fits into the provided opening; tighten fixation screw for lamp housing.
Insert cathode (smaller base) of the new
lamp into the cooling body until stop;
When inserting the lamp, make sure to use the correct hole diameter in the cooling body; the mount in the cooling body is marked with an H.
B 40-065 e 08/99 11
Page 12
Start-Up N HBO 103 and N XBO 75
Carl Zeiss XBO 75 Lamp Insertion/Change Microscope Illuminating Systems
3.2 N XBO 75 Microscope Lamp Lamp Insertion/Change
Before inserting the lamp, make sure that
no connection to the line is available.
The lamps may only be changed when they
are cooled down; allow the microscope illuminator to cool down for 10...15 minutes to avoid the risks of burns and explosion!
The XBO 75 lamp is under high pressure
even if it is not switched on. It may only be removed from the packaging and inserted in the microscope illuminator if a protection mask and safety gloves are worn. (See section 1.3)
Fig. 6 Insert XBO 75 lamp
The safety regulations of the lamp
manufacturer must be observed when the lamp is inserted or changed.
(See section 1.3 - Notes on use)
All clamping adjustments must be performed
carefully; pronounced heat during operation may result in loose contacts.
1 Fixation screw for lamp housing 2 Control wheel for collector 3 Cooling body 4 Fixation screw for lamp mount in cooling body
(screw is hidden in the illustration by the cooling body) 5 Spring lever 6 Lamp mount 7Lamp
12 B 40-065 e 08/99
Page 13
N HBO 103 and N XBO 75 Start-Up Microscope Illuminating Systems XBO 75 Lamp Insertion/Change Carl Zeiss
Procedure:
Switch off the transformer
Disconnect the microscope lamp housing
from the transformer
Allow the microscope lamp housing to cool
down for 10...15 minutes
Wear protection mask and safety gloves.
Remove microscope lamp housing with
cooled-down lamp from the microscope
Move collector in front position by turning
the control wheel (6-2) in the direction of the arrow
Loosen fixation screw (6-1) for lamp
housing; remove the lamp housing
Press spring lever (6-5) downwards to loosen
the lamp locking; remove defective lamp (6-7) with cooling body (6-3) from the mount (6-6) in the lamp base; Make sure to observe the safety regulations of the lamp manufacturer.
Tighten fixation screw of the cooling body.
Insert lamp with cooling body in the relevant
mount (6-6) in the lamp base; make sure that the lateral lead to the cooling body remains accessible and that the melting nipple does not lie in the optical axis; use the spring lever for fixation.
Do not exert any force when inserting the lamp.
Remove the plastic protection from the lamp
Attach lamp housing to the lamp base; make
sure that the contact pin fits into the provided opening; tighten fixation screw for lamp housing.
Loosen fixation screw (6-4) on the cooling
body; remove defective lamp and dispose of it in accordance with the appropriate regulations (see section 1-6 - Notes on disposal) For initial start-up, remove the safety appliance first.
Insert anode (larger base) of the lamp in the
cooling body until stop;
When inserting the lamp, make sure to use the correct hole diameter in the cooling body; the mount in the cooling body is marked with an X. The filling tip and, if available, the ignition wire must be positioned laterally.
B 40-065 e 08/99 13
Page 14
Start-Up N HBO 103 and N XBO 75
Carl Zeiss HBO 103 Lamp Microscope Illuminating Systems
3.3 Start-Up of N HBO 103 Microscope Lamp
Do not perform any work while the lamp is exposed, do not look directly into the lamp or its reflected light; wear safety glasses (sunglasses
Procedure:
Connect the microscope lamp housing to
the transformer as described in the supplied
Fig. 7 Rear of transformer ebq 100 dc
ebq 100 dc operating instructions:
Figure 7
Fig. 8 Front of transformer ebq 100 dc
Connect transformer to the line.
Switch on microscope lamp housing via line
switch at the front of the transformer; the
lamp will ignite automatically:
Figure 8
Meaning of the luminiscence diodes
"Power" Line control in the line switch
"Lamp" The green LED lights up when the
lamp is on
"Temp." The yellow LED indicates inadmissible
heating of the electronics unit; the output voltage is switched off. Make sure to provide sufficient, unhindered ventilation of the transformer; when the admissible temperature has been reached, the instrument automatically reverts to lamp operation.
14 B 40-065 e 08/99
Page 15
N HBO 103 and N XBO 75 Start-Up Microscope Illuminating Systems XBO 75 Lamp Carl Zeiss
3.4 Start-Up of N XBO 75 Microscope Lamp
Do not perform any work while the lamp is exposed, do not look directly into the lamp or its reflected light; wear safety glasses (sunglasses
Procedure:
Connect the microscope lamp housing to
the transformer as described in the supplied
Fig. 9 Rear of transformer ebq 75 isolated
ebx 75 isolated operating instructions: Figure
9
Fig. 10 Front of transformer ebq 75 isolated
Connect transformer to the line.
Switch on microscope lamp housing via line
switch at the front of the transformer; the lamp will ignite automatically:
Figure 10
Meaning of the luminiscence diodes
"Power" Line control in the line switch
"SAFETY" The green LED indicates whether the
safety circuit is closed and whether the instrument is ready for operation.
"Lamp" The green LED lights up when the
lamp is on The green LED blinks when the lamp must be changed
"TEMP" The yellow LED indicates inadmissible
heating of the electronics unit; the output voltage is switched off. Make sure to provide sufficient, unhindered ventilation of the transformer; when the admissible temperature has been reached, the instrument automatically reverts to lamp operation.
B 40-065 e 08/99 15
Page 16
Start-Up N HBO 103 and N XBO 75
Carl Zeiss Focusing and Centering Microscope Illuminating Systems
3.5 Focus and Center the HBO 103/XBO 75 Lamps in the Microscope Lamp Housing
Do not perform any work while the lamp is exposed, do not look directly into the lamp or its reflected light; wear safety glasses (sunglasses
Fig. 11 Lamp arc and mirror image
Direct microscope lamp housing against
wall (approx. 3m away)
1 Collector operating wheel 2 Lamp adjustment - lateral 3 Lamp adjustment - height 4 Mirror adjustment- focus 5 Mirror adjustment- height 6 Mirror adjustment - lateral
Fig. 12 Adjusting components of microscope lamp
housing
Switch on the lamp, starting time < 3
minutes
Adjust collector via operating wheel (12-1)
and produce a sharp image of the lamp arc on the wall; adjust the mirror until the mirror image is in focus: Fig. 11 / Fig. 12
both images should - be identical in size
- be identical in focus
- lie side to side
If this is not the case, use ball-headed
screwdriver to readjust the adjusting components for the lamp and the mirror:
Figure 12
Attach adjusted microscope lamp housing
to the microscope via the dovetail.
16 B 40-065 e 08/99
Page 17
N HBO 103 and N XBO 75 Start-Up Microscope Illuminating Systems Change the Collector Carl Zeiss
3.6 Change the Collector
Loosen clamping ring (13-1)
Remove heat protection filter (13-2)
Pull out collector operating wheel (13-5)
against resistance until the pin allows access to groove (13-4)
Remove collector from the illuminator
Push in new collector until pin of the
pushed-out operating wheel locks into the groove; the collector can be adjusted via the
1 Clamping ring 2 Heat protection filter 3 Collector 4 Groove 5 Collector operating wheel
Fig. 13 Change the collector
operating wheel.
Insert heat protection filter;
tighten clamping ring.
B 40-065 e 08/99 17