Carl ZeissCopyrightMicroscope Illuminating Systems
Knowledge of this manual is required for the operation of the instruments. Would you therefore
please make yourself familiar with the contents of this manual and pay special attention to hints
concerning the safe operation of the instrument.
The specifications are subject to change.
Unless expressly authorized, forwarding and duplication of this document, and the utilization and
communication of its contents are not permitted. Violations will entail an obligation to pay
compensation.
All rights reserved in the event of granting of patents or registration of a utility model.
Fig. 7 Rear of transformer ebq 100 dc........................................................................................14
Fig. 8 Front of transformer ebq 100 dc.......................................................................................14
Fig. 9 Rear of transformer ebq 75 isolated..................................................................................15
Fig. 10 Front of transformer ebq 75 isolated ................................................................................15
Fig. 11 Lamp arc and mirror image...............................................................................................16
Fig. 12 Adjusting components of microscope lamp housing .........................................................16
Fig. 13 Change the collector ........................................................................................................17
Certificate for quality management system in accordance with DIN EN ISO 9001/DIN EN
46001
EC Declaration of Conformity in accordance with EC directives 73/23/EEC and 89/336/EEC
B 40-065 e 08/993
Page 4
SafetyN HBO 103 and N XBO 75
Carl ZeissGeneral, EMCMicroscope Illuminating Systems
1Notes on Safety
1.1General
Please make yourself familiar with the contents
of these operating instructions before attaching
and starting up the N HBO 103 and N XBO 75
microscope illuminators. Additional information
is available from our maintenance service or
from authorized agencies.
The N HBO 103 and N XBO 75 microscope
illuminators should be used with Zeiss
microscopes exclusively.
The N HBO 103 and N XBO 75 microscope
illuminators may only be handled by trained
personnel.
To ensure that the illuminators function
properly with the microscope, the appropriate
microscope manual must be observed.
The manuals
"ebq 100 dc– Electronic transformer for the NHBO 103 microscope illuminator“ and
"ebx 75 isolated – Electronic transformer for
the N XBO 75 microscope illuminator“
must be observed.
Special regard must be paid to the notes on
instrument safety included in this manual.
The data sheets from the lamp manufacturers
must be observed when the HBO 103 and XBO
75 lamps are operated.
To guarantee the safe operation and function
of the N HBO 103 and N XBO 75 microscope
illuminators, it is necessary under any
circumstances to take the precaution measures
and observe the warnings contained in the
operating instructions.
The following symbols are used for the
warnings:
CAUTION!
Non-observance of the safety notes constitutes
a hazard for the user.
CAUTION!
Dangerous electrical voltage
CAUTION!
Disconnect the instrument from the line before
opening it!
CAUTION!
Hot surface
CAUTION!
Non-observance of the safety notes constitutes
the risk of damage to the instrument.
☞
Notes which must be observed when using the
N HBO 103 and N XBO 75 microscope
illuminators.
NOTE!
1.2Regulations concerning
Instrument Safety and
Electromagnetic
Compatibility (EMC)
The N HBO 103 and N XBO 75 microscope
illuminators have been designed and tested in
compliance with EN 61010, Part 1 (VDE 0411).
From the safety viewpoint, they left the factory
in a perfect state.
They meet the requirements of EC directives
73/23 and 89/336 and the EMC legislation,
version of Nov.9, 1992.
Radio-screened in accordance with EN 55011,
Class B, insensitivity to noise in accordance with
EN 50082-2.
Compliance with the EC directives mentioned is
documented by the CE-certificate.
4B 40-065 e 08/99
Page 5
N HBO 103 and N XBO 75Safety
Microscope Illuminating Systems Notes on unpacking, transport, storage and useCarl Zeiss
1.3Notes on Unpacking,
Transport and Storage
The lamps, especially the XBO 75, are under
high pressure even if they are not switched on.
They may therefore be removed from the
package only if the protection mask (Cat.No.
417009) and the safety gloves with artery
protection (Cat. No. 417008) are worn.
The XBO 75 has an additional plastic protection
sleeve which must not be removed when the
lamp is unpacked!
For safety reasons, a safety holder is inserted in
the microscope illuminator instead of the bulb
during transport.
Both the N HBO 103 and N XBO 75 microscope
illuminators and the HBO 103 and XBO 75
lamps must be transported in their original
packaging.
This packaging and the plastic protection sleeve
of the XBO 75 must be kept for storage and
disposal.
The transport and storage temperatures must
be observed in the same way as those of the
microscope stands.
1.4Hints on Use
− The lamps, especially the XBO 75, are under
high pressure even if they are not switched
on.
They may only be inserted in the microscope
illuminator if a safety mask and safety gloves
(see chapter 1.3) are worn.
The safety regulations of the lamp
manufacturer must be observed when the
XBO 75 is inserted.
− The HBO 103 and XBO 75 lamps are under
high pressure during operation; they must
always be opereated in closed lamp
housings!
− The light of the HBO 103 and XBO 75 short-
arc lamps contains a high amount of UV
light:
Never look directly into the lamp or in
reflected light
Wear safety glasses (sunglasses) when
operating the lamps
Cover inspection openings with filters
− The UV radiation of the lamps produces
harmful ozone.
This gas must be carried off to the outside
via exhauster tubes and the working room
supplied with fresh air.
− Both the N HBO 103 and N XBO 75
microscope illuminators and the HBO 103
and XBO 75 lamps must be protected from
impact and humidity.
☞
Instead of the HBO 103, its predecessor HBO
100 can also be used.
B 40-065 e 08/995
Page 6
Instrument DescriptionN HBO 103 and N XBO 75
Carl ZeissTechnical DataMicroscope Illuminating Systems
1.5Hints on Maintenance and
Care
To enable the microscope illuminators to
function perfectly, it is necessary to use only
components and spares marketed or admitted
by us. When in doubt, please contact our
service staff.
With the exception of the work described in
this manual, the user must not perform any
repairs or changes himself.
We would like to emphasize that all other
changes and repairs must be performed by
specially authorized personnel.
Defective instruments or components may only
be repaired and maintained by our service
department.
− After expiry of the average operating time
(see data sheets from the lampmanufacturers), the lamps must be
exchanged; there is a risk of explosion!
The operating time of the HBO 103 and XBO
75 can be checked on the counter of the
relevant transformer.
(Fig. 5/Fig. 8)
The intensity of the lamps decreases when
the operating time increases, i.e.
homogeneous illumination of the object field
can then no longer be guaranteed.
− Always disconnect the instrument from the
line before each lamp exchange or any other
work inside the instrument.
− The safety regulations of the lamp
manufacturer must be observed when the
XBO 75 is changed.
− Check the cables of the N HBO 103 and N
XBO 75 when changing the lamp.
− Clean the housing of the microscope
illuminator when it is dirty.
Do not use any solvents.
Soak a cleaning cloth in a mixture of water
and washing-up liquid, remove the dirt and
dry the housing using a clean cloth.
1.6Hints on Disposal
Used HBO 103 and XBO 75 lamps and defective
microscope illuminators should not be treated
like domestic waste; they should be disposed of
in their original packaging in accordance with
the legal regulations.
When in doubt, please contact the Carl Zeiss
microscopy service.
− The lamps must be exchanged as described
in chapters 3.1/3.2.
− Before each lamp exchange, make sure to
allow the N HBO 103 or N XBO 75
microscope illuminator to cool down
sufficiently long in order to reduce the
pressure; wait at least for 15 minutes!
6B 40-065 e 08/99
1.7Hints on Warranty
The instrument manufacturer is not liable for
damage caused by unauthorized operation,
particularly the removal and exchange of
components, or the use of accessories from
other manufacturers. This forfeits all the claims
against warranty.
Page 7
N HBO 103 and N XBO 75Instrument Description
Microscope Illuminating SystemsComponents, Catalogue NumbersCarl Zeiss
2Instrument
Description
2.1Components of the
Microscope Illuminating
Systems
1 Lamp housing
2 Connecting cable5 Lamp base
3 Collector6 Transformer ebq dc
4 Dovetail7 Line cable
Fig. 1 Overall view of the N HBO 103
Microscope Illuminating System
1 Lamp housing
2 Connecting cable5 Lamp base
3 Collector6 Transformer ebx 75 isolated
4 Dovetail7 Line cable
Fig. 2 Overall view of the N XBO 75
Microscope Illuminating System
The N HBO 103 and N XBO 75 microscope
illuminators include the following components
(Fig. 1 / Fig. 2):
− Illuminator with:
dovetail for attachment to the microscope,
reflector,
double-lens collector or quartz collector for
UV light,
heat-protection filter
− HBO 103 W/2 mercury vapor short-arc lamp
or
XBO 75 W/2 xenon short-arc lamp
− transformer ebq 100 dc for N HBO 103
or
transformer ebx 75 isolated for N XBO 75
− line cable with earth-contact plug
or
line cable with American flat plug
HBO 103XBO 75
Microscope
Illuminating Systems
Collector
Quartz collector
Lamp
Transformerebq 100 dc
Line cable
Table:Catalogue numbers
Lamp N HBO 103
000000-1007-980
Collector N HBO 103/XBO 75
Quartz collector N HBO 103/XBO 75
Mercury vapor-
short-arc lamp
HBO 103 W/2
380301-9350-000
000000-1003-928
Line cable with earth-contact plug
Line cable with American flat plug
000000-1007-976
000000-1007-977
380137-6750-000
380137-6740-000
Lamp N XBO 75
000000-1007-981
Xenon-
short-arc lamp
XBO 75 W/2
380079-9190-000
ebx 75 isolated
000000-1003-924
B 40-065 e 08/997
Page 8
Instrument DescriptionN HBO 103 and N XBO 75
Carl ZeissTechnical DataMicroscope Illuminating Systems
2.2Technical Data of the
Microscope Illuminating
Systems
Lamp housingLamp housing N HBO 103
Cat. No. 000000-1007-980
Length163 mm228 mm
Width141 mm141 mm
Height200 mm200 mm
Collector mount
Connection to standdovetaildovetail
Connection to transformerCable with special plugCable with special plug
Transformer
Cat.No. 000000-1003-928
Line connection90-250 V AC90-250 V AC
Line frequency50-60 Hz50-60 Hz
Power consumption265 VA130 VA
FuseF1/F2 T 3.15 AF1/F2 T 2 A
Lamp
∅ 40 mm∅ 40 mm
Transformer N for
HBO 103
Mercury vapor
short-arc lamp
Lamp housing N XBO 75
Cat.No. 000000-1007-981
Transformer N for
XBO 75
Cat. No. 000000-1003-924
Xenon-
short-arc lamp
For the technical data of the lamps
please see the information
provided by the manufacturer.
Recommended:
Average life of the lampsOSRAM
OSRAM
Type HBO 103 W/2
Cat. No. 380301-9350-000
Type HBO 100 W/2200 h
Type HBO 103 W/2300 h
OSRAM
Type XBO 75 W/2
Cat. No. 417030-9010-000
Type XBO 75 W/2
Cat. No. 417030-9011-000
Hamamatsu
Type Xenon lamp L217401 UV
Cat.No. 380079-9190-000
Type Xenon lamp L217401 OFR
Cat. No. 380053-9870-000
OSRAM
Type XBO 75 W/2 400 h
Hamamatsu
Type Xenon lamp 2000 h
8B 40-065 e 08/99
Page 9
N HBO 103 and N XBO 75Technical Data
Microscope Illuminating SystemsLine spectraCarl Zeiss
Fig. 3 Line spectrum of the HBO 100/103 mercury short-arc lamps
Fig. 4 Line spectrum of the XBO 75 xenon short-arc lamp
B 40-065 e 08/999
Page 10
Start-UpN HBO 103 and N XBO 75
Carl ZeissHBO 103 Lamp Insertion/ChangeMicroscope Illuminating Systems
are cooled down; allow the microscope
illuminator to cool down for 10...15 minutes
to avoid the risks of burns and explosion!
− The lamp may only be removed from the
packaging and inserted in the microscope
illuminator if a protection mask and safety
gloves are worn (see chapter 1.3).
− All clamping adjustments must be performed
carefully; pronounced heat during operation
may result in loose contacts.
1 Fixation screw for lamp housing
2 Control wheel for collector
3 Cooling body
4 Fixation screw for lamp mount in cooling body
5 Spring lever
6 Lamp mount
7Lamp
Fig. 5Insert HBO 103 lamp
10B 40-065 e 08/99
Page 11
N HBO 103 and N XBO 75Start-Up
Microscope Illuminating SystemsHBO 103 Lamp Insertion/ChangeCarl Zeiss
Procedure:
• Switch off the transformer
• Disconnect the microscope lamp housing
from the transformer
• Allow the microscope lamp housing to cool
down for 10...15 minutes
• Wear protection mask and safety gloves.
• Remove microscope lamp housing with
cooled-down lamp from the microscope
• Move collector in front position by turning
the control wheel (5-2) in the direction of
the arrow
• Loosen fixation screw (5-1) for lamp
housing; remove the lamp housing
• Press spring lever (5-5) downwards to loosen
the lamp locking; remove defective lamp
(5-7) with cooling body (5-3) from the
mount (5-6) in the lamp base.
• Loosen fixation screw (5-4) on the cooling
body: remove defective lamp and dispose of
it in accordance with the appropriate
regulations (see section 1-6 - Notes ondisposal)
For initial start-up, remove the safety
appliance first.
• Tighten the fixation screw on the cooling
body.
• Insert lamp with cooling body in the relevant
mount in the lamp base and make sure that
the lateral lead to the cooling body remains
accessible; use the spring lever for fixation.
Do not exert any force when inserting the
lamp.
• Attach lamp housing to the lamp base; make
sure that the contact pin fits into the
provided opening;
tighten fixation screw for lamp housing.
• Insert cathode (smaller base) of the new
lamp into the cooling body until stop;
When inserting the lamp, make sure to use
the correct hole diameter in the cooling
body; the mount in the cooling body is
marked with an H.
B 40-065 e 08/9911
Page 12
Start-UpN HBO 103 and N XBO 75
Carl ZeissXBO 75 Lamp Insertion/ChangeMicroscope Illuminating Systems
3.2N XBO 75 Microscope Lamp
Lamp Insertion/Change
− Before inserting the lamp, make sure that
no connection to the line is available.
− The lamps may only be changed when they
are cooled down; allow the microscope
illuminator to cool down for 10...15 minutes
to avoid the risks of burns and explosion!
− The XBO 75 lamp is under high pressure
even if it is not switched on.
It may only be removed from the packaging
and inserted in the microscope illuminator if
a protection mask and safety gloves are
worn. (See section 1.3)
Fig. 6 Insert XBO 75 lamp
− The safety regulations of the lamp
manufacturer must be observed when the
lamp is inserted or changed.
(See section 1.3 - Notes on use)
− All clamping adjustments must be performed
carefully;
pronounced heat during operation may
result in loose contacts.
1 Fixation screw for lamp housing
2 Control wheel for collector
3 Cooling body
4 Fixation screw for lamp mount in cooling body
(screw is hidden in the illustration by the cooling body)
5 Spring lever
6 Lamp mount
7Lamp
12B 40-065 e 08/99
Page 13
N HBO 103 and N XBO 75Start-Up
Microscope Illuminating SystemsXBO 75 Lamp Insertion/ChangeCarl Zeiss
Procedure:
• Switch off the transformer
• Disconnect the microscope lamp housing
from the transformer
• Allow the microscope lamp housing to cool
down for 10...15 minutes
• Wear protection mask and safety gloves.
• Remove microscope lamp housing with
cooled-down lamp from the microscope
• Move collector in front position by turning
the control wheel (6-2) in the direction of
the arrow
• Loosen fixation screw (6-1) for lamp
housing;
remove the lamp housing
• Press spring lever (6-5) downwards to loosen
the lamp locking; remove defective lamp
(6-7) with cooling body (6-3) from the
mount (6-6) in the lamp base;
Make sure to observe the safety regulations
of the lamp manufacturer.
• Tighten fixation screw of the cooling body.
• Insert lamp with cooling body in the relevant
mount (6-6) in the lamp base;
make sure that the lateral lead to the cooling
body remains accessible and that the melting
nipple does not lie in the optical axis;
use the spring lever for fixation.
Do not exert any force when inserting the
lamp.
• Remove the plastic protection from the lamp
• Attach lamp housing to the lamp base; make
sure that the contact pin fits into the
provided opening;
tighten fixation screw for lamp housing.
• Loosen fixation screw (6-4) on the cooling
body; remove defective lamp and dispose of
it in accordance with the appropriate
regulations (see section 1-6 - Notes ondisposal)
For initial start-up, remove the safety
appliance first.
• Insert anode (larger base) of the lamp in the
cooling body until stop;
When inserting the lamp, make sure to use
the correct hole diameter in the cooling
body; the mount in the cooling body is
marked with an X.
The filling tip and, if available, the ignition
wire must be positioned laterally.
B 40-065 e 08/9913
Page 14
Start-UpN HBO 103 and N XBO 75
Carl ZeissHBO 103 Lamp Microscope Illuminating Systems
3.3Start-Up of N HBO 103
Microscope Lamp
Do not perform any work while the lamp is
exposed,
do not look directly into the lamp or its
reflected light;
wear safety glasses (sunglasses
Procedure:
• Connect the microscope lamp housing to
the transformer as described in the supplied
Fig. 7Rear of transformer ebq 100 dc
ebq 100 dc operating instructions:
Figure 7
Fig. 8 Front of transformer ebq 100 dc
• Connect transformer to the line.
• Switch on microscope lamp housing via line
switch at the front of the transformer; the
lamp will ignite automatically:
Figure 8
Meaning of the luminiscence diodes
"Power" Line control in the line switch
"Lamp"The green LED lights up when the
lamp is on
"Temp." The yellow LED indicates inadmissible
heating of the electronics unit; the
output voltage is switched off.
Make sure to provide sufficient,
unhindered ventilation of the
transformer; when the admissible
temperature has been reached, the
instrument automatically reverts to
lamp operation.
14B 40-065 e 08/99
Page 15
N HBO 103 and N XBO 75Start-Up
Microscope Illuminating SystemsXBO 75 LampCarl Zeiss
3.4Start-Up of N XBO 75
Microscope Lamp
Do not perform any work while the lamp is
exposed,
do not look directly into the lamp or its
reflected light;
wear safety glasses (sunglasses
Procedure:
• Connect the microscope lamp housing to
the transformer as described in the supplied
Fig. 9Rear of transformer ebq 75 isolated
ebx 75 isolated operating instructions: Figure
9
Fig. 10 Front of transformer ebq 75 isolated
• Connect transformer to the line.
• Switch on microscope lamp housing via line
switch at the front of the transformer; the
lamp will ignite automatically:
Figure 10
Meaning of the luminiscence diodes
"Power"Line control in the line switch
"SAFETY" The green LED indicates whether the
safety circuit is closed and
whether the instrument is ready
for operation.
"Lamp"The green LED lights up when the
lamp is on
The green LED blinks when the lamp
must be changed
"TEMP"The yellow LED indicates inadmissible
heating of the electronics unit; the
output voltage is switched off.
Make sure to provide sufficient,
unhindered ventilation of the
transformer; when the admissible
temperature has been reached, the
instrument automatically reverts to
lamp operation.
B 40-065 e 08/9915
Page 16
Start-UpN HBO 103 and N XBO 75
Carl ZeissFocusing and CenteringMicroscope Illuminating Systems
3.5Focus and Center the HBO
103/XBO 75 Lamps in the
Microscope Lamp Housing
Do not perform any work while the lamp is
exposed,
do not look directly into the lamp or its
reflected light; wear safety glasses (sunglasses