WS755 Water Sampler

Publication Number 39160712 01-957
- 1 -
Page 2
Global Water
800-876-1172 globalw.com

Table of Contents

I. WS755 Checklist Page 3 II. Inspection 3 III. Description 4 IV. Installing the Water Sampler 5 V. Installing the Pickup Hose 5 VI. The Control Panel 6 VII. Composite Sampling 8 VIII. Discrete Sampling 8 IX. Mixed Sampling 9 X. External Trigger mode 9 XI. Specifications 11 XII. Maintenance 12 XIII. Troubleshooting 12 XIV. Warranty 14 XV. Appendix A: Sample Size vs. Head Height 15 XVI. Appendix B: Battery Li fe 16 XVII. Appendix C: Input and Output Connections 16 XVIII. Appendix D: Accessories 17
* Copyright  Global Water Instrumentation, Inc. 2012
- 2 -
Page 3
Global Water
800-876-1172 globalw.com
Congratulations on your purchase of the Global Water WS755 Water Sampler. This instrument has been quality tested and approved for accurate and reliable operation. We are confident that you will find the WS755 to be a valuable asset for your appli cations. Should you require assistance, our technical st aff will be happy to help.

I. WS755 Checklist

1. WS755 Water Sampler with Transport Case

2. Two 20-Mesh Pickup Strainers w/ 15’ of Hose

3. 12V Battery in Battery Enclosure

4. 3-Stage Charge Controller

5. 18 Volt Regulated Battery Charger

6. Battery Cable

7. Two Water Sensor Shorting Plugs

8. Two Quick-Disconnect Fitting Caps and Plugs

9. WS755 Manual

II. Inspection

The WS755 unit was carefully inspected and cer tified by Global Water’s Quality Assurance Team before shipping. If any damage has occurred during shipping, please notify Global Water Ins trum entation, Inc. and file a claim with the carrier involved.

Use the checklist to ensure that ever y thing needed to operate the WS755 Water Sampler was received.
- 3 -
Page 4

III. Description

The Global Water WS755 Sampler is designed spec ifically to meet a wide variety of sampling requirements including industrial discharge, process cont r ol, water and wastewater treatment plants, sewers, rivers and streams.
The sampler consists of a rugged, rainproof and loc k able wheeled transport case with a comfort grip handle on top and retract able handle on the back. The enclosure houses a control panel with two peristaltic sampling pumps, tw o sample bottles, two pickup hoses and st r ainer s with quick-disconnect fittings, a rechargeable battery with 3-stage char ge c ontroller in a self contained battery box, and a energy efficient regulated batt er y c harger.
The WS755 Sampler can take two individual “time weighted” composite samples, or the sampler can be set to take full-bottle discrete or “Grab” samples. Each pump has its own sample bottle and control for setting the size of individual samples. This allows each pump to be set as either a compos ite or discrete sampler independently of the other or, in the case of two composite samplers, have different size settings for eac h pum p.
A delay timer can be set to start sampling after a preset time. This allows multiple samplers to be deployed in the field and have them turn on at the same time. The delay can also be used to hold off sampling to let the water source flush out debris or other contaminants .
After each sample, the pumps will reverse f or 15 seconds. This clears any debris from the strainer at the end of the pickup hose and also empties water from the hose so the next sample is not contaminat ed by the pr ev ious one.
Inputs are provided for triggering the sam pler based on water level, a rain gauge sensor or signals from an external process c ontroller or other monitoring device. Individual outputs are also available for monitoring when each of the pumps has taken a sample, using a data logger or similar recor ding dev ic e.
Global Water
800-876-1172 globalw.com
- 4 -
Page 5
Global Water
800-876-1172 globalw.com

IV. Installing the Water Sampler

a. The sampler should be placed upright (it will not work if placed on its back or
side).
b. Open the sampler and remove the battery charger from the unit. Store the
charger for later use.
c. Remove the pickup hoses and connect to the sampler using the quick-
disconnect fittings on the side of t he enc los ur e. Place the protective caps over the protective quick-disconnect plugs and save for future use.
d. To secure the sample bottles:
1. Screw the bottle cap/float switch onto the sample bottle
2. Place the bottles into the sampler enclosure
3. Insert the end of the peristaltic pump’s hose in the hole at the top of the bottle cap
4. Plug the float switch lead into the jack on the front of the control panel
e. The sampling unit can be secured from vandalism and strong winds by one of
the following methods:
1. Mount unit on post and lock closed
2. Lock closed and chain handles to a solid structure (such as a tree, post, or building)
3. Enclose and lock unit in a steel electrical box
f. Avoid drilling holes in the enclosure if possible. If holes must be drilled in the
enclosure avoid drilling through t he c ontrol panel enclosure. Any holes drilled through the enclosure must be sealed with s ome t y pe of sealant to prevent water from entering the case and causing equipment failure.
g. The unit is water resistant, not water proof. The unit must be located well
above the expected water level to continue to provide reliable service.

V. Installing the Pickup Hose

Remove the pickup hoses from the enclosure and c onnec t to the sampler using the quick-disconnect fittings. Place the protective caps over the protective plugs and save for future use. When storing the hoses inside the enclos ur e, replace the protective caps and plugs to keep the fittings clean of dirt and contam ination. The pickup hoses should be installed in a wastewater discharge c hannel or stream. The pickup strainers should be submerged under water and should be situat ed to avoid contact with the channel bottom.
- 5 -
Page 6

The Control Panel

1. Pump One.
2. Pump Two.
3. Start Delay: This setting delays sampling by a preset amount of time. If no delay is desired, set the control to zero. Only the first sample taken by each pump is delayed, except as described lat er in the section titled “External Trigger Mode”.
4. Sample Interval: Sets the time interval between composite samples . The “External Trigger” position is reserv ed for using the Trigger and Rain Sensor inputs to initiate sampling as described later in the section titled “External Trigger Mode”.
5. Pump 1 Sample Size: Sets the size for a single composite sample with pump one. The sample size settings are approximat e and apply to a head height of four feet. The switch position mark ed “Full ” causes the pump to work as a discrete sampler, filling the bot tle full in one sample. Note: In the “Full” position, the sampler will shut off the pump after one hour, even if the bottle is not full.
6. Pump 2 Sample Size: Sets the sample size for pump two and works the same as Pump 1 Sample Size described above.
7. Trigger In: These inputs work in “External Trigger” mode only, when selected by the Sample Interval cont r ol. A signal from an external device will cause a single sample to be taken by the corresponding pump according to its Size control. The Ext er nal Trigger m ode is des c r ibed in
Global Water
800-876-1172 globalw.com
- 6 -
Page 7
Global Water
800-876-1172 globalw.com
more detail later in the section titled “External Trigger Mode”. Refer to the Specifications section for details about the input pulse requirements.
8. Pulse Out: These outputs send a pulse to an optional recording or monitoring device each time t he c or r es ponding pump starts, allowing the time of the sample event to be recorded. R efer to the Specifications section for details about the output pulse.
9. Status: This red LED indicates the status of the sampler. While the sampler is waiting to be triggered, the light remains off. Once triggered, the LED will blink once every two seconds while t he Delay or Interval timers are running and while the pumps are operating. The LED remains lit constantly when both sample bottles ar e full or if the switches are not plugged in.
10. Power: This turns the power to the sampler on and off. Turning the switch off for one second resets all f unc tions and timers.
11. Battery: Connect the battery here using the supplied battery cable.
12. Water Sensor: When an optional wat er s ens or is plugged into this jack, sampling will not begin until the presence of water is detected (the Rain Sensor must also detect water). If this sens or input is not being used, plug one of the two shorting plugs supplied with the sampler into this jack. This input has no affect when triggering the sam pler with the Trigger Inputs in External Trigger mode.
13. Rain Sensor: When an optional water sensor is plugged into this jac k , sampling will not begin until the presence of water is detected, usually in a rain gauge (the Water Sensor must also detect water). If this sensor input is not being used, plug one of the two shorting plugs s upplied with the sampler into this jack. This input has no aff ect when tr igger ing the sampler with the Trigger Inputs in External Trigger mode. This input can also be used to take individual composite samples and is described in detail later in the section titled “External Trigger Mode”.
14. Bottle Switches: The sample bottle float s witches plug into the control panel here. For sampling to occur, a float s witch must be plugged in and the bottle must not be full. If the Status LED remains lit all the time, check these connections.
15. Test Pump: Pressing one of these buttons tests the operation of t he corresponding pump. These buttons will test the pumps operation regardless of whether a bottle swit c h is plugged in or not. The pump runs forward for as long as the button is held down, t hen r ev er s es for an equal amount of time, or 15 seconds, which ever is les s .
- 7 -
Page 8
Global Water
800-876-1172 globalw.com
VI. Composite Sampling
A composite sample is a series of smaller s amples put into the same bottle, thus showing an “average” sample over time. Set the Interval control to the desired t im e between samples (do not select the External T r igger pos ition). Set the Size control for the corresponding pump to the desired sample size. Note that the sample size settings are approximate and apply to a head height of four feet . For heights other than four feet, refer to the chart inside the front of the sampler. E ach pump c an hav e different size settings but will share the same Sample Interval s etting.
The sampler starts timing when both of the optional Rain and Water Sensor s detect moisture. If you do not have the optional water sensor s or do not need them for your application, defeat them by plugging one of the two shorting plugs supplied with t he s ampler into their jack. If both inputs are defeated, the sampler starts up triggered when the pow er switch is turned on. The Status LED starts blinking every 2 seconds once the sampler is triggered.
The Start Delay tim er begins running when the sampler is triggered by the Rain and Water Sensors and the first sample is taken when t he timer ex pir es . If the Delay control is set to zero, sampling begins immediately . The delay timer only applies to the first sam ple taken by each pump. At the end of each sample, the pumps rev er s e for 15 seconds.
Sampling continues as Trigger-Delay-Sample-Reverse-Interval-Sample-Reverse-Interval­Sample-Reverse … until the sample bot tles are full. For each pump, the same amount of Interval time is placed between samples, ev en if the size settings are different. When both bottles are full, the Status LED r em ains lit constantly.
VII. Discrete Sampling
A discrete sample is a single sample put into the s ame bottle, thus being a sample at one point in time. Set the Size control for the cor r espondin g pum p to “Full”. Once started, the pump will continue to run until the sample bot tle if full.
As with composite sampling, the Start Delay timer begins when the sampler is trigger by the Rain and Water Sensors and the first sample is tak en when the timer expires. If the Delay control is set to zero, sampling begins immediately. The Status LED blinks every 2 s ec onds while the delay timer or the pumps are running and remains lit constantly once both of t he sample bottles have been filled. At the end of sampling, the pumps reverse for 15 seconds.
- 8 -
Page 9
Global Water
800-876-1172 globalw.com
VIII. Mixed Sampling, Composite and Discrete Sampler

The WS755 Sampler can be configured as both a c omposite and discrete sampler at the same time since each pump has a separate Size control. Set one pump’s size setting to the desired composite sample size and the ot her to “Full”.

Again, the Start Delay timer begins when the sampler is trigger by t he Rain and Water Sensors and the first sample is taken when the timer ex pir es . If the Delay control is set to zero, sampling begins immediately . The Status LED blinks every 2 s econds while the delay timer and the pumps are running and remains lit c ons tantly once both of the sample bottles have been filled. At the end of each sample, t he pumps r ev er s e for 15 seconds.

X. External Trigger Mode

When the Sample Interval control is set to the “External Trigger” position, the sampler can be triggered externally by inputs fr om either the Trigger In jacks or the Rain Sensor input. This mode is useful in applications where sampl ing oc c ur s bas ed on r eadings from a process controller, SCADA system, flow monitor, auto-dumping rain gauge or other remot e monitoring devices.
Using the Rain Sensor as a trigger source: This mode is normally used with an auto-dumping rain gauge that empties itself every 24 hours. Set the Sample Interval control to Ext ernal Tr igger . Set each pumps Size control based on desired composite sample size or s et t o Full f or a dis c r ete sample.
The sampler will be triggered when both the Rain and W ater Sensors detect moisture. If the Water Sensor is not needed, use one of the shorting plugs supplied with the sampler to defeat it by plugging it into the Water S ens or input jack . Place the Rain Sensor at the desired height in the rain gauge. When t he Rain Sensor detects moisture, the sampler will start.
Once triggered by the Rain Sensor, the Start Delay timer begins running and the Status LED begins blinking every 2 seconds. When the tim er ex pir es , the pumps take a sample based on their individual Size settings. When the samples are complete, the pumps reverse for 15 seconds, then stop. The sampler cannot be re-trigger ed until the rain gauge empties itself and the Rain Sensor goes dry. The next t ime t he s ens or detects moisture, the Start Delay timer will start over again and another sample will be taken when it expires.
- 9 -
Page 10
Global Water
800-876-1172 globalw.com
If a pump’s Size control is s et to take a full-bottle discrete sample, further triggers will have no affect on that pump. If a pump is set to take composite samples, each one will r epr es ent the conditions at the time of that “qualifi ed r ain ev ent”. This process will continue until both the sample bottles are full and the Status LED remains lit all the time.
Using the Trigger Inputs as a source: These inputs are generally used for connecting to a flow monitor, processor controller, or other external monitoring device.
When using the Trigger In input jacks, leave the Rain Sensor input jack disconnected. Eac h of the two Trigger Inputs control the c or r es ponding pump separately from the other.
Upon being triggered by one of the Trigger Inputs, the Start Delay timer begins running. When the delay timer has expired, the corr es ponding pump t ak es a s am ple bas ed on its Size setting. Holding the Trigger Input active past the end of the sample will cause the pump to continue until the Trigger Input is r eleas ed. When the sample is complete, the pump reverses, and then stops.
Flow Proportional Sampling
a. To set the WS755 up to take flow proportional samples , a flow monitor must
be attached to the control panel. In t his c onfiguration, the sampler is designed to take a sample every time a specific amount of flow has passed through the flow monitor.
b. The flow monitor must provide a momentary switch closure (250ms or longer)
or a pulse of 4-30VDC, whenever a sample is to be taken. The output of the flow monitor must be wired to a 2-wire RCA jack as show n in A ppendix C. The phone plug is then inserted into the Tr igger In jack on the control panel.
c. Set the Sample Interval knob to Ex ternal Trigger mode and the Sample Size
knob to the desired sample size. The sample will be taken, flow proportionally, as directed by the flow meter.
- 10 -
Page 11
XI. Specifications
Operating Temperature: 0° to +70°C Dimensions 22”H X 18"W X 10"D Weight: 30lb (Shipping Weight 32lbs) Materials:
Enclosure: Expanded UV protected PVC Bottles: Two 1 gallon Polyethylene Pickup Hoses: 15' nylon reinforced 1/4" ID polyethylene
6-Mesh strainers available as options. Pump Tubing: Neoprene 1/4" ID, 7/16” OD
Sample Pumps:
Flow Rate: 1000 ml per minute at 4 ft. head Type: Peristaltic
Maximum Lift: ~20 feet Battery: Rechargeable 5 AH Gel Cell Battery Life: One Pump running: ~1 hour
Start Delay: 16 time s ettings from 0 to 12 hours Composite Interval: 15 time settings from 5 min. to 12 hours
Sample Size: 15 composite sam ple s iz es from 50ml to
External Trigger Inputs: 250mS minimum pulse width Pulse Outputs: 5VDC one-second pulse
1000ohm output impedance Bottle Switch Inputs: Switch closure Input Floating read switch in bottle Rain and Water Sensors: Optional moisture sensors or switch
Internal Fuse: 10A Slow-Blow Charger 18VDC Universal Switching Supply,
2.1mm Connector, Center Positive
Global Water
800-876-1172 globalw.com
flexible tubing with 20-Mesh intak e strainers and quick-disconnect f ittings.
Two Pumps running: ~ ½ hour Standby: 3 months while retaining power to run both pumps to capacity
plus an External Trigger mode sett ing 2 liters plus a Full Bottle discrete setting
(Approximate sizes at 4 foot head) Switch closure or 4-30VDC
closure inputs
- 11 -
Page 12

XII. Maintenance

Sampler
The Global Water WS755 Water Sampler requires minim al maintenance. The sampler enclosure is rainproof and rugged. Av oid ex pos ur e to extremely rough usage. Routinely wipe the carrying case and control panel face, rinse the pickup hose and debris strainer, and wash the sample bottles with mild soap and warm water. Additional plastic and glass sam ple bottles, neoprene tubing for the sampler pump, bottle caps/float s w itches, removable debris strainers, and pickup hose can be purchased from Global Water (see Appendix D section).
Battery
If the pumps run slowly, this is an indication t hat t he battery requires recharging. Global Water recommends fully rechar ging batteries approximately once per month. In addition, the battery should be rec har ged before any extended use. NOTE: The battery life will be maximized if rec har ged before it drains below 11 volts.
Global Water
800-876-1172 globalw.com
To recharge the battery, plug the included charger into the charging jack on the outside on the s ampler enclosure. The WS755 has a 3-Stage charge controller installed inside t he battery enclosure that will prevent over-charging. T o pr ev ent damage to the battery and controller, use only the supplied c harger. Additional batteries and batt er y c har ger s ar e availabl e
from Global Water (see Appendix D).

XIII. Troubleshooting

a. Check that the battery is firmly connected to the control panel and confirm
that it is fully charged.
b. Press the Pump Test buttons on the front panel. The test buttons should
always run the pumps, regardless of the s tate of any of the inputs.
c. Turn the power switch off for 1-2 s ec onds then back on to reset the sampler. d. Check the Status LED on the front panel.
i. The LED is off: If the water sensors ar e being us ed, check the Rain
and Water Sensor inputs. Sensors must be in contact with moisture
and unused inputs must have one of the shorting plugs supplied with
- 12 -
Page 13
Global Water
800-876-1172 globalw.com
the sampler plugged into the jack. If the Trigger Inputs are being used,
confirm that External Trigger mode is s elec ted on the Interval control.
Check the pulses coming from the tr igger ing dev ic e and c onfirm they
meet specifications. Test the Trigger Inputs by plugging one of the
shorting plugs supplied with the sampler into the jack.
ii. The LED is blinking: Check the Start D elay and Interval settings. The
sampler may be waiting for the Delay tim er to run out or it may be
timing the interval between composite samples.
iii. The LED is on constantly: Check the bot tle switches. The light
remains on when both sample bottles are f ull or when both float
switches are not plugged in or working properly. Plug one of the
shorting plugs supplied with the sampler into one of the bottle switch
jacks, the light should go out or st art blink ing.
Other issues e. Call us for tech support: 800-876-1172 or (979) 690-5560 (many problems
can be solved over the phone). Fax: (979) 690-0440 or Email:
globalw@globalw.com.
Be prepared to describe the problem being experienced inc luding specific details of the application and installation and any additional pertinent information.
f. In the event that the equipment needs to be returned to the factory f or any
reason, please call to obtain a RMA # (Return Material Authorization). Do not return items without a RMA # on the outside of the package. Decontaminate the WS705 prior to ret ur ning.
Include a written statement describing the problems. Send the package with shipping prepaid to G lobal Water’s factory address.
Insure the shipment, as the warranty does not cover damage incurred during transit.
g. When calling for tech support, please have the following information r eady ;
1. Model #.
2. Unit serial number.
3. P.O.# the equipment was purchased on.
4. Global Water’s sales number or the invoice number.
5. Repair instructions and/or specific problems relating t o the product.
- 13 -
Page 14

IXV. Warranty

a. Global Water Instrumentation, Inc. warrants that its produc ts are free from
defects in material and workmanship under normal use and service for a period of one year from date of shipment from factory. Global Water’s obligations under this warranty ar e lim ited to, at Global Water’s option: (I) replacing or (II) repairing; any produc ts determined to be defective. In no case shall Global Water’s liability exceed t he pr oduc ts original purchase price. This warranty does not apply to any equipment that has been repaired or altered, except by Global Water Inst r um entation, Inc., or which has been subject to misuse, negligence or accident. It is expressly agreed that this warranty will be in lieu of all warranties of fitness and in lieu of the warranty of merchantability.
b. The warranty begins on the date of the product’s invoice.
Global Water
800-876-1172 globalw.com
- 14 -
Page 15
Actual Sample Size vs. Sample S ize Control for Various He a d He ights
0
200
400
600
800
1000
1200
1400
1600
1800
2000
0 250 500 750 1000 1250 1500 1750 2000 2250
Actual Sample Size (ml)
Sample Size Control Setting (ml)
14 Feet
2 Feet
Global Water
800-876-1172 globalw.com

Appendix A: Sample Size vs. Head He ight

Use the following graph to see how sample s iz e changes with head height. The Sample Size controls on the control panel are calibrated at a height of 4 feet. Increasing the head height decreases the sample size by a predictable amount. To determine the appropriat e Sample Size setting for a desired volume; find the sample size on the horizontal axis, follow straight up to the line corresponding to your head height, then look across to the vertical axis and read the setting for the Size control on the control panel.
- 15 -
Page 16
Global Water
800-876-1172 globalw.com
Appendix B: Battery Life
While battery life depends on several f actors such as charge, condition and temperat ur e; it can generally be assumed that the s am pler c an deployed in the field for 3 months waiting to be triggered, while still retaining enough c har ge to run a single pump for one hour (about 10 gallons) or both pumps for ½ hour (5 gallons). To prevent large current surges which occur when the pumps first turn on, the WS755 is designed s o that only one pump will turn on or off at a time. A ½ second time delay is inserted between these events, extending the life of the charge and the life of the batt er y. The battery should generally be charged at least once per month and the voltage should not be allowed to fall below 11 volts. The internal charge controller will prevent the overcharging of the battery if the charger is left connected all the time.

Appendix C: Input and Output Connections

- 16 -
Page 17
Part Descrip tion
Part Number
Unit
12V 5Ah Battery
00-010
Each
Battery Charger
01-998
Each
1 Gallon Plastic Sample B ottle
00-418
Each
1 Gallon Plastic Bottle Cap/Float Switch
CB0200
Each
Replacement Pickup Hose
00-546
Feet
20-Mesh Pic kup Strainer Assembly
CA0600
Each
Replacement 20-Mesh Strainer
01-881
Each
Optional 6-M esh Strainer
01-945
Each
Quick-Disconnect Fitting Plug & Cap
CA0850
Each
Tubing Quick-Disconnect Fitting
CA0860
Each
Replacem ent Pump Tubing
00-744
Feet
Rain and Wa ter Sensors
CA0400
Each
Auto-Drain Rain Gauge
CH0000
Each
Complete Stormwater Kit
CTU000
Each

Appendix D: Accessories

Global Water
800-876-1172 globalw.com
- 17 -
Loading...