WS705 Water Sampler

Publication Number 39180712 01-956
- 1 -
Global Water
800-876-1172 globalw.com

Table of Contents

I. WS705 Checklist Page 3 II. Inspection 3 III. Description 4 IV. Installing the Water Sampler 5 V. Installing the Pickup Hose 5 VI. The Control Panel 6 VII. Composite Sampling 8 VIII. Discrete Sampling 8 IX. External Trigger mode 9 X. Specifications 11 XI. Maintenance 12 XII. Troubleshooting 12 XIII. Warranty 14 XIV. Appendix A: Sample Size vs. Head Height 15 XV. Appendix B: Battery Life 16 XVI. Appendix C: Input and Output Connections 16 XVII. Appendix D: Replacement Parts and Accessories 17
* Copyright  Global Water Instrumentation, Inc. 2012
- 2 -
Global Water
800-876-1172 globalw.com
Congratulations on your purchase of the Global Water WS705 Water Sampler. This instrument has been quality tested and approved for accurate and reliable operation. We are confident that you will find the WS705 to be a valuable asset for your appli cations. Should you require assistance, our technical st aff will be happy to help.

I. WS705 Checklist

1. WS705 Water Sampler with Transport Case

2. 20-Mesh Pickup Strainer w/ 15’ of Hose

3. 12V Battery in Battery Enclosure

4. 3-Stage Charge Controller

5. 18 Volt Regulated Battery Charger

6. Battery Cable

7. Two Water Sensor Shorting Plugs

8. Quick-Disconnect Fitting Cap and Plug

9. WS705 Manual

II. Inspection

The WS705 unit was carefully inspected and cer tified by Global Water’s Quality Assurance Team before shipping. If any damage has occurred during shipping, please notify Global Water Ins trum entation, Inc. and file a claim with the carrier involved.

Use the checklist to ensure that ever y thing needed to operate the WS705 Water Sampler was received.
- 3 -

III. Description

The Global Water WS705 Sampler is designed spec ifically to meet a wide variety of sampling requirements including industrial discharge, process cont r ol, water and wastewater treatment plants, sewers, rivers and streams.
The sampler consists of a rugged, rainproof and loc k able wheeled transport case with a comfort grip handle on top and retr ac table handle on the back. The enclosure houses a control panel with peris taltic sampling pump, sample bottle, pickup hose and strainer with quic k -dis c onnec t fittings, a rechargeable battery with 3-stage charge controller in a s elf contained battery box, and a energy efficient regulated battery c har ger .
The WS705 Sampler can take individual “t im e weighted” composite samples, or it can be set to take a full-bottle discrete or “Gr ab” sam ple. This is accomplished with a Sample Size control having one position being res er v ed for “Full” bottle”. A Sample Interval control sets t he tim e between individual composite samples and has a provision for allowing the sampler to be triggered from external sources.
A Start Delay timer can be set to start sampling after a preset time. This allows multiple samplers to be deployed in the field and have them turn on at the same time. The delay can also be used to hold off sampling to let the water source flush out debris or other contaminants .
After each sample, the pump will reverse for 15 s ec onds . This clears any debris from the strainer at the end of the pickup hose and also empties water from the hose so the next sample is not contaminat ed by the pr ev ious one.
Inputs are provided for triggering the sam pler based on water level, a rain gauge sensor or a signal from an external process controller or other monitoring device. An output is also available for monitoring when the pump is taking samples, using a data logger or similar recording device.
Global Water
800-876-1172 globalw.com
- 4 -
Global Water
800-876-1172 globalw.com

IV. Installing the Water Sampler

a. The sampler should be placed upright (it will not work if placed on its back or
side).
b. Open the sampler and remove the battery charger from the unit. Store the
charger for later use.
c. Remove the pickup hose and connect it to the sampler using the quick-
disconnect fitting on the side of the enc los ur e. Place the protective cap over the protective quick-disconnect plug and save for future use.
d. To secure the sample bottle:
1. Screw the bottle cap/float switch onto the sample bottle
2. Place the bottle into the sampler enclosure
3. Insert the end of the peristaltic pump’s hose in the hole at the top of the bottle cap
4. Plug the float switch lead into the jack on the front of the control panel
e. The sampling unit can be secured from vandalism and strong winds by one of
the following methods:
1. Mount unit on post and lock closed
2. Lock closed and chain handles to a solid structure (such as a tree, post, or building)
3. Enclose and lock unit in a steel electrical box
f. Avoid drilling holes in the enclosure if possible. If holes must be drilled in the
enclosure avoid drilling through t he c ontrol panel enclosure. Any holes drilled through the enclosure must be sealed with s ome t y pe of sealant to prevent water from entering the case and causing equipment failure.
g. The unit is water resistant, not water proof. The unit must be located well
above the expected water level to cont inue to provide reliable service.

V. Installing the Pickup Hose

Remove the pickup hose from the enclosure and connec t it to the sampler using the quick-disconnect fittings. Place the protective cap over the protective plug and save for future use. When storing the hose inside the enclosure, replace the prot ec tive cap and plug to keep the fittings clean of dirt and c ontamination. The pickup hose should be installed in a wastewater discharge c hannel or s tream. The pickup strainer should be submerged under water and should be situated to av oid c ontact with the channel bottom.
- 5 -

VI. The Control Panel

1. Pump.
2. Start Delay: This setting delays sampling by a preset amount of time. If no delay is desired, set the control to zero. Only the first sample taken by the pump is delayed, except as described later in the section titled “External Trigger M ode”.
3. Sample Interval: Sets the time interval between composite samples . The “External Trigger” position is res er v ed for using the Trigger and Rain Sensor inputs to initiate sampling as des c r ibed later in the section titled “External Trigger M ode”.
4. Sample Size: Sets the size for a single composite sample. The sample size settings are approximate and apply to a head height of four feet. The switch position marked “Full” caus es the pump to work as a discrete sampler, filling the bottle full in one sample. Note: In the Full position, the sampler will shut of f the pump after one hour, even if the bottle is not full.
5. Trigger In: This input works in “External Trigger” mode only, as selected by the Sample Interval contr ol. A signal from an external device will cause a single sample to be taken accor ding to the Size control. The External Trigger mode is descr ibed in more detail later in the section titled “External Trigger Mode”. Refer to the Specifications section for details about the input pulse requirements.