Xylem WQ SC-FDO User Manual

Global Water
O-ring
800-876-1172 globalw.com
WQ SC-FDO
Handling of the sensor cap
z Dirt and moisture below the sensor cap can have
negative influences on the function of the sensor and decrease lifetime of the sensor cap. Please pay attention to clean and dry working environ­ment, when taking off the senor cap.
z Please do not touch the outer sensor membrane
with your fingers. Touch the sensor cap principally only at the sides (arrows in the figure on the left).
z Avoid strong mechanical forces on the sensor
membrane (pressure, scratches).
z The impact of light, particularly daylight, on the
inner side of the sensor cap will have, over time, an impact on the measuring properties and can shorten the lifetime of the sensor cap. Therefore, the inner side of the sensor cap should not be exposed to direct sunlight. Please only store sen­sor caps in light protected environments.
Installation
Replace the O-ring each time when replacing the sensor cap. Refer to the sensor operating manual
ba75909e01 07/2010
for detailed information.