VENTIlogicLS with leakage ventilation / met lekkagesysteem
VENTIlogicLS - single patient circuit with patient valve / systeem
met een slang met patiëntenventiel
VENTIlogicLS - double patient circuit with patient valve /
systeem met dubbele slang met patiëntenventiel
VENTIlogic plus with leakage ventilation / met lekkagesysteem
VENTIlogic plus - single patient circuit with patient valve /
systeem met een slang met patiëntenventiel
Instructions for use for devices up to serial number 9.999 and from
firmware version 2.07 upwards
Gebruiksaanwijzing voor apparaten tot serienummer 9.999 en vanaf
The respiratory air flows to the patient from here
via the patient circuit and the patient/ventilator
interface.
2 O2 coupling
Serves as an adapter for connecting the oxygen
source to the therapy device.
3 Control panel and displays
For controlling and monitoring the therapy device
and connected accessories.
4 Handle
For transporting the device.
5 Power cord
For connecting the therapy device to the power
supply.
6 VENTIpower connection
For connecting the VENTIpower mobile power
supply available as an accessory.
7 Card reader
Slot for a Weinmann memory card. Therapy data
are stored on the memory card which the doctor
can call up.
8 Serial ports
For connecting devices for displaying and
evaluating therapy data.
9 Filter compartment lid, air inlet
For covering and securely positioning the coarse
dust and fine filters.
10 Remote alarm connection
For connecting the hospital's internal nurse call
system or the VENTIremote alarm remote alarm
case for use outside the hospital.
11 Cable-securing clip
Prevents the device being disconnected from the
power supply inadvertently.
12 Power connection
This is where the power cord is connected to the
device.
13 Device ID plate (underneath)
Provides information about the device, such as
serial number and year of manufacture, for
example.
14 Oxygen connection
For connecting the oxygen supply if oxygen
supply has been prescribed.
15 Housing fan
Protects the device from overheating.
6Overview
Page 7
EN
21 Drying adapter
20 Sealing plug (2x)
18 Pressure
measuring tube
17 Adapter
22 Exhalation
system
23 Connection for
VENTIclick humidifier
19 Creased tube
16 Patient circuit
Therapy device with leakage ventilation
Key
16 Patient circuit
The air flows to the patient/ventilator interface
through the patient circuit. The patient circuit
consists of creased hose, pressure measuring tube
and adapter.
17 Adapter
For connecting the patient circuit to the device
outlet port.
18 Pressure measuring tube
For measuring therapy pressure.
19 Creased tube
Delivers respiratory air to the patient.
20 Sealing plug (2x)
For sealing off the pressure measuring tube
during cleaning (only with leakage ventilation).
21 Drying adapter
Required to dry the patient circuit with the aid of
the therapy device and for the function check.
22 Exhalation system
Carbon dioxide-enriched exhaled air escapes here
during therapy.
23 Connection for VENTIclick humidifier
For connecting the VENTIclick humidifier available
as an accessory (only with leakage ventilation).
Overview7
Page 8
EN
Therapy device, single patient circuit with patient valve
25 Connection for pressure
measuring tube (marked
blue)
28 Single patient circuit
29 Pressure-
measurement
tube (marked
blue)
30 Valve control
tube
31 Patient valve
27 Connection for
valve control tube
32 Connection for
patient
33 Connection for
device outlet port
26 Test adapter
24 Connection for O
2
sensor
Key
24 Connection for O2 sensor
For connecting an oxygen sensor which can be
used to measure oxygen concentration in
respiratory air (only with patient circuits with a
patient valve).
25 Connection for pressure measuring
tube (marked blue)
For connecting the pressure measuring tube to
the device.
8Overview
26 Test adapter
Required for the function check of the therapy
device.
27 Connection for valve control tube
For connecting the valve control tube to the
device.
28 Single patient circuit
Delivers respiratory air to the patient.
Page 9
EN
29 Pressure-measurement tube (marked
blue)
For measuring therapy pressure.
30 Valve control tube
For controlling (opening and closing) the patient
valve.
31 Patient valve
For routing the patient's exhaled air out of the
patient circuit.
32 Connection for patient
This is where the patient/ventilator interface is
connected.
33 Connection for device outlet port
This is where the patient circuit is connected to
the device outlet port of the therapy device.
Overview9
Page 10
EN
Therapy device, double patient circuit with patient valve (VENTIlogic LS
38 Device connection for
patient valve
35 Connection for pressure-
measurement tube
(marked blue)
40 Double patient circuit
42 Y-adapter
43 Patient valve
45 Valve control
tube
47 Pressure-
measurement tube
(marked blue)
37 Connection for
valve control tube
39 Opening for
exhaled air
41 Connection
for patient
44 Connection for
device inlet port
46 Connection for
device outlet port
36 Test adapter
34 Connection for O
2
sensor
only)
34 Connection for O2 sensor
For connecting an oxygen sensor which can be
used to measure oxygen concentration in
respiratory air (only with patient circuits with a
patient valve).
35 Connection for pressure-
measurement tube (marked blue)
For connecting the pressure measuring tube to
the device.
39 Opening for exhaled air
This is where the patient's exhaled air is routed
10Overview
36 Test adapter
Required for the function check of the therapy
device.
37 Connection for valve control tube
For connecting the valve control tube to the
device.
38 Device connection for patient valve
For connecting the patient valve to the device
inlet port of the therapy device.
out of the device.
Page 11
EN
40 Double patient circuit
Delivers respiratory air to the patient and from the
patient back to the device.
41 Connection for patient
This is where the patient/ventilator interface is
connected.
42 Y-adapter
When the double patient circuit is used, this
brings the inspiration and exhalation tubes
together and serves as an adapter for connection
to the patient/ventilator interface.
43 Patient valve
For routing the patient's exhaled air out of the
patient circuit.
44 Connection for device inlet port
This is where the patient circuit is connected to
the device outlet port of the therapy device.
45 Valve control tube
For controlling (opening and closing) the patient
valve.
46 Connection for device outlet port
This is where the patient circuit is connected to
the device inlet port for the patient valve (only
with double patient circuit with patient valve).
47 Pressure-measurement tube (marked
blue)
For measuring therapy pressure.
Overview11
Page 12
EN
Accessories
50 Carrying bag
WM 27706
49 VENTIclick
(only leakage
ventilation)
54 Protective bag
WM 27106
53 Set, O2 sensor
52 Bacteria filter
WM 24476 (only
valve ventilation)
51 Bacteria filter WM 24148
(only leakage ventilation)
48 VENTIpower
Key
48 VENTIpower
Available as an accessory, for mobile power
supply to the therapy device.
49 VENTIclick (only leakage ventilation)
Available as an accessory, for humidifying
respiratory air.
For protecting the device from contamination, in
particular when the device is used by several
patients (patient change).
53 Set, O2 sensor
For measuring oxygen concentration at the device
outlet port.
54 Protective bag WM 27106
For protecting the therapy device during
transport.
Page 13
EN
Special markings on the device
8
7
6
5
4
3
2
1
13
12
11
10
15
18
14
16 17
9
Left-hand side
1
Oxygen connection: maximum supply rate: 15 l/min at < 1000 hPa
VENTIlogic LS:
2
Opening for exhaled air when operated with double patient circuit with patient
valve; do not seal opening or block in any other way.
VENTIlogic plus:
Opening is not used with VENTIlogic plus.
Overview13
Page 14
EN
Front
P
~
VENTIlogic LS:
3
Connection for patient's exhaled air with double patient circuit with patient valve.
VENTIlogic plus:
Opening is not used with VENTIlogic plus.
4
Jack: electrical connection for VENTIclick respiratory air humidifier (only leakage
ventilation), max. current consumption at 40 V: 600 mA
5
6
7
8
9
10
11
12
13
Jack: electrical connection for oxygen sensor; max. 100 mV DC
Connection: pressure measuring tube (marked blue). Therapy pressure 0 - 50 hPa
(only for patient circuit with patient valve)
Connection: control tube for patient valve 0 - 50 hPa (only patient circuit with
patient valve)
Device outlet port: outlet for exhaled air at 0 - 45 hPa with patient circuits with
patient valve, 0 - 35 hPa with leakage ventilation
Device outlet port: only patient circuits with a diameter of Ø 15 mm - 22 mm are
permitted.
Right-hand side
Connection for optional attachments, e.g. Analog box D/A;
max. current delivery at 5 V: 50 mA
Connection for specialist staff to set therapy parameters using
WEINMANNsupport; max. current delivery at 12 V: 50 mA
Slot for MMC or SD card
Connection for power supply-independent operation with VENTIpower
14
15
14Overview
Rear
Connector for power supply input 115/230 V AC; 50/60 Hz
Connection for remote alarm: connection for nurse call system and VENTIremote alarm remote alarm case. Breaking capacity: 60 V DC/2 A; 42 V AC/2 A
Page 15
EN
16
Servicing label: indicates when the next service is due
17
18
SNSerial number
115 V/230 V ~,
50-60 Hz
Safety check label: (in Germany only) marks when the next safety check as per §6
STK
of the German law relating to users of medical devices is required
Device inlet port: inlet port for ambient air at room temperature
BF protection class
Protection class II, protective insulation
Year of manufacture
Do not dispose of device in domestic waste!
Electrical rating
Device operated with humidifier
Device operated without humidifier
Device ID plate (underneath)
Overview15
Page 16
EN
Control panel
8 Softstart key
1 LED for power supply
2 LIAM (insufflation)3 On/Off key
4 Alarm acknow-
ledgement key
with LEDs
5 Dial
6 Operating keys
7 Menu key
9 Humidifier key
with LED
Key
1 LED for power supply
The green LED comes on when there is a power
supply.
2 LIAM (insufflation)
For triggering a coughing episode or ventilating a
sigh.
3 On/Off key
For switching the therapy device on and off.
4 Alarm acknowledgement key with
LEDs
The alarm acknowledgement key is for the
temporary muting of alarms. The LED displays
alarms visually.
5 Dial
Central control of the therapy device, for
navigating in the menu.
6 Operating keys
For quick-setting by a doctor; deactivated in
patient mode.
7 Menu key
For switching from the default display to the
menu and vice versa.
8 Softstart key
For activating softstart and for setting softstart
time up to the maximum softstart time set in the
menu.
9 Humidifier key with LED
For setting humidifier stage. Six stages are
available. The LED indicates whether the
humidifier is activated.
16Overview
Page 17
EN
Default display during therapy
1 Active ventilation
mode
9 Bar chart for pressure display
8 Access to patient
menu
7 Respiratory phase
change display
11 Ventilation
parameters
12 Patient circuit
2 Status line
10 Oxygen
concentration
5 Battery charge status
4 Softstart display3 Alarm display
6 Memory card symbol
Key
1 Active ventilation mode
The active ventilation mode is displayed at this
point in the status line.
2 Status line
This is where device status information (such as
alarm state display, filter change or servicing due)
is displayed.
3 Alarm display
If an alarm has been muted, it is then shown in
the status line for 120 seconds.
4 Softstart display
Displayed when softstart has been activated.
Gives remaining softstart time in minutes.
5 Battery charge status
Displays the charge status of the internal battery.
When the battery is charging, the segments are
shown consecutively.
6 Memory card symbol
Appears if a memory card is present and there is
data saved on the memory card.
7 Respiratory phase change display
Indicates whether the current respiratory phase
change is spontaneous or mandatory
(spontaneous: S, mandatory: T); the display
changes from left (inspiration) to right
(exhalation) depending on respiratory phase;
mandatory exhalation is shown here.
Also indicates whether the trigger for inspiration
is blocked due to an activated trigger lockout time
at the start of expiration ( ).
8 Access to patient menu
Use the key adjacent to this menu item to switch
to the patient menu and back to the default
display.
9 Bar chart for pressure display
For the graphical display of therapy pressure.
Overview17
Page 18
EN
10 Oxygen concentration
Gives oxygen concentration in respiratory air in
percent.
11 Ventilation parameters
The relevant current ventilation parameters are
displayed depending on the active mode.
12 Patient circuit
The relevant symbol appears depending on the
patient circuit set.
18Overview
Page 19
EN
Symbols used in the display
SymbolSignificance
Status line:
Softstart active, remaining time faded in
Filter change required
Servicing required
At least one physiological alarm is activated
Acoustic signal of alarms muted for 120 seconds
All physiological alarms deactivated (exception for VENTIlogic LS: In VCV and
aVCV modes, the Pressure
deactivated)
Blower off (standby mode)
Device in battery mode, battery below 25 % capacity
Device in battery mode, battery below 50 % capacity
Device in battery mode, battery below 75 % capacity
Device in battery mode, battery above 75 % capacity
high
and Pressure
alarms cannot be
low
Segments displayed consecutively: device in power supply mode,
battery charging
Measured values are written to the MMC or the SD card.
Operation with the VENTIpower mobile power supply WM 27640
Device in TA mode, automatic analysis phase running
Device in TA mode, manually-triggered analysis phase running
LIAM enabled: displayed under the current mode, e.g.:
PCV
+LIAM
20Overview
Page 21
EN
SymbolSignificance
LIAM
LIAM (Lung Insufflation Assist Maneuver) active
Main window (Monitor):
IPAP
EPAP / PEEP
P
SIMV
hPa
f
B Trigger for inspiration blocked during expiration
Ti/T
V
T
V
Ti
VTe
Ti
Te
O2 (21%)
Inspiration pressure
Exhalation pressure
Specifies the inspiration pressure level of the back-up ventilation (SIMV mode
only)
Pressure given in hectopascals; 1.01973 hPa corresponds to 1 cm H2O.
Respiratory frequency
Respiratory phase switch triggered - spontaneous
Respiratory phase switch triggered - mandatory
Proportion of inspiration time in a respiratory cycle
Tidal volume
Tidal volume on inspiration
Tidal volume on exhalation
Inspiration time
Exhalation time
Mean oxygen concentration
Shown in brackets: measuring cell not calibrated, perform oxygen calibration
Overview21
Page 22
EN
Markings on the packaging
SymbolSignificance
Therapy device:
SNSerial number of device
Permitted temperature for storage: -40 °C to +70 °C
Permitted humidity for storage: max. 95 % relative humidity
Protect pack from wet
Do not tip over or drop pack
Safety information in the instructions
for use
Safety information in these instructions for use is marked as follows:
Warning!
Warns of risk of injury and potential material damage.
Caution!
Warns of material damage and potentially false therapy results.
Note:
Contains useful tips.
22Markings on the packaging
Page 23
EN
2. Description of device
2.1 Intended use
2.1.1VENTIlogic LS
VENTIlogic
with DIN EN ISO 10651-2.
The device can be used in both static or mobile operation, both at home and in appropriate
hospital departments.
The device can be used for weaning off invasive ventilation and converting to mask ventilation.
It is used on patients with medium to severe acute and chronic global respiratory
insufficiency with a tidal volume of at least 50 ml and a body weight of at least 5 kg.
2.1.2VENTIlogic plus
VENTIlogic plus is used for invasive and non-invasive life-support ventilation in accordance
with DIN EN ISO 10651-2.
The device can be used in both static or mobile operation, both at home and in appropriate
hospital departments.
LS is used for invasive and non-invasive life-support ventilation in accordance
Note
VENTIlogic LS is not a ventilator for intensive care purposes in accordance with DIN
EN ISO 80601-2-12.
Note
VENTIlogic plus is not a ventilator for intensive care purposes in accordance with
DIN EN ISO 80601-2-12.
The device can be used for weaning off invasive ventilation and converting to mask ventilation
It is used on patients with medium to severe acute and chronic global respiratory
insufficiency with a tidal volume of at least 50 ml and a body weight of at least 5 kg.
Indication
This corresponds to the following clinical pictures:
• obstructive respiratory disorders, such as COPD
• restrictive respiratory disorders such as scolioses, deformities of the thorax
Description of device23
.
Page 24
EN
• neurological, muscular and neuromuscular disorders, such as muscular
dystrophies, pareses of the diaphragm etc.
• central respiratory regulation disorders
• hypoventilation syndrome associated with obesity
2.2 Owner/operator and user qualification
As an owner/operator or user, you must be familiar with the operation of this medical
device. Observe the legal requirements for operation and use (in Germany, the regulations
governing owner/operators of medical devices apply in particular). Basic recommendation:
get a person authorized by Weinmann to provide you with proper instruction about the
handling, use and operation of this medical device.
2.3 Description of function
2.3.1Providing the therapy pressure
An electronically-controlled blower draws in ambient air through a filter and delivers it to
the device outlet port. From here, air flows through the patient circuit and the patient/
ventilator interface to the patient.
Sensors detect the pressure at the patient/ventilator interface and in the patient circuit, as
well as the respiratory phase change. The blower accordingly provides the respiratory
volume and the IPAP and EPAP / PEEP pressures prescribed by the doctor.
2.3.2Display and operation
The display shows the therapy mode and, as a function of the mode, the currently applied
values for CPAP, IPAP and EPAP / PEEP, respiratory frequency (f) and volume. Spontaneous
or mechanical respiratory phase switches and the pressure change are also shown in
graphical form. Ventilation parameters can be set in standby mode and in ventilation
mode. The device is operated by a number of keys that give direct access to the most
important parameters, such as IPAP, EPAP / PEEP, frequency, inspiration time and volume.
A dial is used to navigate through the menu. Parameters are shown in an LC display.
A key code is used to prevent therapy values from being adjusted inadvertently. Operation
is locked when a padlock symbol is shown on the display. In the Physician menu, it is
possible to show ventilation curves such as flow curves and pressure/volume loops
(VENTIlogic LS only) in addition to therapy values.
24Description of device
Page 25
EN
2.3.3Operating status
Three operating states are possible on the therapy device: on, off and standby.
If the device is switched on, therapy is in progress. In standby, the fan is switched off but
the device is ready for immediate operation by briefly pressing the on/off switch, provided
that the patient circuit is connected correctly. The settings on the device can be adjusted
in standby mode. If the device is switched off completely, the fan and display are also
switched off and no settings can be adjusted on the device.
2.3.4Leakage ventilation
When leakage ventilation is used, an exhalation system continuously flushes out the CO2containing exhaled air.
2.3.5Valve ventilation
In this case, exhalation is controlled by the patient valve.
When the single patient circuit with patient valve is used, the patient's exhaled air escapes
into the environment through the patient valve. The device controls the patient valve by
means of the valve control tube.
When the double patient circuit with patient valve is used (VENTIlogic LS only), an
exhalation tube also routes exhaled air into the ambient air through the device.
2.3.6Therapy modes
The therapy device can be operated in the following therapy modes:
• leakage ventilation:TA, S, T, ST, SX (VENTIlogic LS only), SXX (VENTIlogic LS
only), CPAP
• valve ventilation: PCV, aPCV, PSV, VCV (VENTIlogic LS only), aVCV
(VENTIlogic LS only), SIMV
The mode required for therapy is set on the device by the doctor supervising treatment.
The doctor can activate volume compensation in pressure-controlled modes S, T, ST, SX
(VENTIlogic LS only), SXX (VENTIlogic LS only), TA, PCV, PSV and aPCV. A minimum volume
and maximum pressure rise are set to achieve this. If the minimum volume is undershot,
the device automatically and continuously increases pressure up to the set maximum
pressure (therapy pressure + max. pressure rise).
In controlled modes T, PCV and VCV (VENTIlogic LS only) and in assisted-controlled modes
ST, PSV, aVCV (VENTIlogic LS only) and aPCV, the doctor can set respiratory frequency in
Description of device25
Page 26
EN
the range from 5 to 45 breaths per minute and inspiration time in the range from 15 % to
67 % of the respiratory period.
In S, ST, SX (VENTIlogic LS only), SXX (VENTIlogic LS only), PSV, aPCV, aVCV (VENTIlogic LS
only) and SIMV modes, the doctor can select one of 8 trigger stages for inspiration and one
of 14 trigger stages for exhalation (not with aPCV, aVCV).
In ST mode the expiratory trigger can be deactivated. The switch to exhalation is then on
a time-controlled basis.
In adaptive mode TA, the device automatically adapts to the individual respiratory rhythm
and the patient's individual breathing pattern and provides the therapy pressure at
precisely this rhythm and pressure curve.
If no breath into the device is taken in S mode, therapy pressure is automatically provided
at a minimum frequency of 5 breaths a minute.
CPAP mode does not provide any respiratory assistance. The therapy device provides a
constant positive therapy pressure in this mode.
The display shows therapy pressure and, as a function of mode, current values for IPAP and
EPAP / PEEP and respiratory frequency (f). Depending on the patient circuit used, tidal
volume (V
) is displayed in the case of a leakage system and tidal volume on inspiration (VTi)
T
in the case of valve ventilation. When the single patient circuit is used, only tidal volume on
inspiration can be measured, whilst with the double patient circuit (VENTIlogic LS only),
total tidal volume can be measured.
Spontaneous or mechanical respiratory phase switches and the pressure change are also
shown in graphical form.
2.3.7SIMV mode
SIMV mode (synchronized intermittent mandatory ventilation) is a mixture of mandatory
and assisted ventilation.
If there is no spontaneous respiration the device will mandatorily ventilate the patient once
the T
and an inspirational pressure level of P
In the case of spontaneous respiration the device switches to assisted ventilation using the
set IPAP value. The pressure level will then fluctuate cyclically at a frequency of f
ratio of Ti/T
case is dictated by the patient.
26Description of device
time has elapsed at a respiratory frequency of f
apnea
and an inspirational pressure level of P
ba
SIMV
.
, a ratio of Ti/Tba (backup)
backup
. The respiratory frequency in this
SIMV
SIMV
, a
Page 27
EN
2.3.8Softstart
The softstart function makes it easier to fall asleep or to become used to higher ventilation
pressures. The physician sets an initial pressure for inspiration and exhalation that gradually
rises to the therapy pressure during the softstart phase. This function must be activated by
the physician and is only available in CPAP mode.
2.3.9Auto switch-on (only leakage ventilation)
The device has an automatic switch-on function. If this is activated, the device can be
switched on by taking a breath into the breathing mask. The device is still switched off
using the On/Off key .
2.3.10 Uninterrupted power supply (UPS)
A built-in battery ensures an uninterrupted power supply in the event of a power outage.
The battery operating time will depend on the load and temperature range. Detailed
information on the different loads with the corresponding battery operating times are
provided in section 11. on page 101. The internal battery is automatically charged or
maintained in a charged state as long as the device is supplied with power.
2.3.11 Mobile power supply
An external mobile power supply via the VENTIpower battery is also possible; this is
available as an accessory.
2.3.12 LIAM (insufflation)
The Lung Insufflation Assist Maneuver function allows a higher volume to be administered
to the patient if the corresponding key is pressed; this supports coughing. This function has
to be enabled by the doctor supervising treatment.
2.3.13 Nurse call and remote alarm
The device has a remote alarm connection to support the monitoring of patient and device,
especially when VENTIlogic LS is used for life-support ventilation. All high and mediumpriority alarms, together with the No power supply alarm are passed to this connection.
All other alarms are displayed only on the device itself.
The remote alarm connection can be used to connect the device to the VENTIremote alarm
remote alarm case. In hospital, the device can be connected directly to the hospital's own
internal alarm system.
Description of device27
Page 28
EN
2.3.14 Recording therapy data
Therapy data are stored in the device on a removable MMC or SD card. The
WEINMANNsupport PC software can be used to enable a doctor to evaluate the therapy
data.
2.3.15 Analog output of therapy data
The device has an interface for connecting to analog box WM 27560. It is used for a timesynchronized display of therapy data such as pressure, flow, leakage and volumes e.g. on
a PSG. Mode-specific data such as trigger times in S mode or efforts and fightings in TA
mode can also be displayed.
28Description of device
Page 29
EN
3. Safety instructions
3.1 Safety information
Read these instructions for use carefully. They are a component of the device and must be
available at all times. Use the device exclusively for the intended purpose described (see
“2.1 Intended use” on page 23).
For your own safety and the safety of your patients and in accordance with the
requirements of Directive 93/42/EEC, please note the following.
3.1.1Life-support ventilation
Danger!
Increased resistance in the patient circuit can cause the alarm to fail!
Attaching an accessory can increase the resistance in the patient circuit. Depending
on the settings, this could prevent life-saving alarms from being triggered. For
example, if the Disconnection alarm fails then the patient may be put at risk.
– Make absolutely certain that the V
– Ensure that appropriate values are used for the V
– Check that the alarms are working.
– Carry out an alarm check every time an accessory is changed.
Warning!
Device failure if incorrect patient circuits used!
If patient circuits with a diameter smaller than Ø 15 mm are used, the device may
overheat.
– Use only patient circuits with a diameter of Ø 15 mm or more.
– Note that total permitted resistance may be exceeded even in patient circuits
with a diameter of Ø 15 mm when these are combined with bacteria filters.
T low
and V
alarms are active.
T high
and V
T low
Safety instructions29
T high
alarms.
Page 30
EN
Danger!
The alarm will not work if the wrong settings are used!
If the V
alarm has been deactivated, or incorrect settings have been used, then
T low
the alarm will not be triggered. If the patient is dependent on the ventilation device
then they are placed at great risk if the alarm fails.
– It is essential to ensure that alarm V
is activated for life-support ventilation
T low
(VENTIlogic LS only). Only if these conditions are met can a blockage (stenosis)
be detected.
– Set the V
alarm appropriately.
T low
Danger!
Failure of alarm function due to incorrect alarm settings in VCV and aVCV
modes (VENTIlogic LS only)!
If the Pressure
and Pressure
high
alarms have not been properly set in VCV
low
and aVCV ventilation modes, then these alarms will not be triggered. If these alarms
are not triggered the patient may be put at risk.
– Make absolutely certain that the Pressure
and Pressure
high
alarms are
low
active in the VCV and aVCV ventilation modes.
– Ensure that appropriate values are used for the Pressure
and Pressure
high
low
alarms.
Warning!
•An alternative ventilation option (e.g. a replacement device or a manual
ventilating bag) needs to be kept to hand for patients who are dependent on a
ventilation device, in case the device fails.
•It is critical that patients who are dependent on the ventilation device are
monitored by the person caring for the patient. Otherwise it is possible that
there will be no reaction to any alarms occurring on the device.
•Ensure that any alarms and malfunctions can be seen at all times and that the
person caring for the patient can take the necessary measures. Recourse can be
had to the VENTIremote alarm remote alarm case or the hospital's own internal
alarm system to support monitoring.
• With the single patient circuit and patient valve, the system only allows the
volume given off by the device to be displayed and monitored. With the valve
system, exhaled volume can only be displayed reliably with a double patient
circuit with patient valve (VENTIlogic LS only). For this reason, you should ensure
that patients dependent on the ventilation device are ventilated with a double
30Safety instructions
Page 31
EN
patient circuit or, if a single patient circuit is used, that exhaled volume is
monitored separately.
3.1.2Operating the device
Warning!
•Do not cover the device with blankets etc. The air inlet would be blocked and
the device could overheat. This may lead to inadequate therapy and to damage
to the device.
• All device openings must be freely accessible and may not be blocked by objects.
• Maintain a safe distance between the therapy device and devices which emit HF
radiation (e.g. cell phones - see page 111), otherwise there may be
malfunctions.
• In order to prevent reinfection in the case of infectious diseases, we recommend
using a bacteria filter.
• The device is not suitable for use in an environment at risk of explosion.
• The device may not be operated with flammable anesthetics, nor may flammable
anesthetics be kept in the vicinity of the therapy device. Risk of fire/explosion!
• Ensure that ventilation tubes and cables are routed so that they cannot lead to
the patient being strangled.
• Ensure that there are no small parts close to the patient, otherwise they might
get into the patient's respiratory flow and put him/her at risk.
• Masks of third-party manufacture may only be used following authorization by
the manufacturer, Weinmann. The success of therapy is put in jeopardy by the
use of unauthorized masks or other types of patient/ventilator interface.
• If a pneumotachograph with a high flow resistance is used to determine flow at
the start of therapy or to check it, this may restrict trigger function. In the event
of queries, contact the manufacturer, Weinmann.
• No antistatic or electrically conductive tubes may be used.
• Please observe Section “6. Hygiene treatment” on page 65 to prevent infection
or bacterial contamination.
• With valve ventilation, masks with integrated or separate leakage ventilation are
prohibited.
• With leakage ventilation, only full-face masks with an integrated emergency
exhalation valve may be used.
Safety instructions31
Page 32
EN
• An exhalation system must always be used with leakage ventilation, otherwise
the CO
concentration in the breathing mask and tube would rise to critical
2
values and thus obstruct breathing.
• Always use a suitable water trap if the patient produces a great deal of secretion
during ventilation, otherwise fluid may get into the device. This can lead to
damage to the device and thus a risk to the patient.
• In ventilation modes with a trigger function on inspiration, hyperventilation may
result.
• Only converter cable USB-RS485 WM 93318 or converter box WM 93316 may
be connected to the RS485 serial port.
Caution!
• Check whether the power supply of the device matches that of your local power
supply. The device can operate with voltages of 115 V and 230 V. It
automatically adapts to one of these voltages.
• Ensure that the power cord is connected correctly. Always secure the power plug
with the cable-securing clip to prevent the plug being removed by mistake.
• The device must be connected to an easily accessible socket so that the plug can
be taken out quickly in the event of a fault.
• Do not use sockets with an On/Off switch or dimmable sockets.
• Do not set up the device close to a radiator and do not expose it to direct
sunlight, as this could overheat the device. Condensation could also form in the
humidifier used and condense in the patient circuit.
• Never push objects, cloths etc. into the openings of the device. This may block
inlets and outlets and lead to damage to the device.
• The device must be on standby or switched off for the MMC or SD card to be
inserted or removed, otherwise therapy data may be lost.
Note:
• The use of accessories in the respiratory flow, such as humidifiers and bacteria
filters, for example, may change the characteristics of the device. Subsequent
addition of these accessories may make it necessary to reset device parameters.
The total resistance of the ventilation system must not exceed 6 hPa at 60 l/min
for adults and 6 hPa at 30 l/min for children.
32Safety instructions
Page 33
EN
3.1.3Mobile operation
Warning!
•When operated on a wheelchair, this system requires a certificate of conformity.
To obtain this, consult a specialist dealer authorized by the manufacturer,
Weinmann; this dealer will also perform the attachment.
Caution!
• The internal battery is not intended for mobile operation. Always use
VENTIpower for a mobile power supply.
• Protect the therapy device in the carrying bag from direct sunlight and rain,
using a sunshade or rain cover for example. The carrying bag itself provides only
brief protection from sunlight and rain.
• During mobile operation there may be problems with the trigger. This can result
in inadequate ventilation. Have your physician adjust the trigger settings or
select a monitored ventilation mode.
Note:
• When planning your time, take account of the fact that battery running time is
considerably reduced at low outdoor temperatures.
3.1.4Oxygen supply
Warning!
•If oxygen is being supplied to the respiratory flow, smoking and naked flames
are forbidden. Risk of fire. The oxygen can accumulate in clothing, bed linen or
hair. It can only be removed by thorough ventilation.
• It is essential to follow the safety instructions in the instructions for your oxygen
system.
• Too high or too low an oxygen supply can be toxic and lead to severe
complications. We therefore recommend monitoring the oxygen supply with an
oxygen sensor. This oxygen sensor can replace neither blood gas analysis nor
direct FiO
• Risk of fire. Always shut off the oxygen supply first at the end of therapy. Then
leave the therapy device to run for a short time before turning it off, otherwise
some residual oxygen will remain in the device. In the event of a fault, this could
lead to a risk of fire.
measurement.
2
Safety instructions33
Page 34
EN
• To supply oxygen, use only the therapy device connection provided for this
purpose. Never supply oxygen via the patient/ventilator interface or the Tadapter, otherwise the oxygen supply cannot be switched off automatically if a
fault occurs.
Note:
• Supplying oxygen via a connection not intended for this purpose leads to an
erroneous volume display, as the oxygen flow cannot be included in the
measurement.
• Be aware that pressure at the patient connection opening may rise during
exhalation if you connect accessories (e.g. bacteria filter or respiratory air
humidifier).
• The UPS works only if the internal battery is present and charged. Recourse can
only be had to VENTIpower during an interruption to the power supply with the
device in the switched-on state and only in conjunction with the internal battery.
• If the therapy device has been stored or transported at operating temperatures
outside those quoted in the instructions for use, the therapy device may be
commissioned only once the temperature of the device is within the temperature
range permitted for operation.
• Do not transport the therapy device with the humidifier fitted, otherwise residual
water may run into the therapy device and damage it.
• The remote alarm connection is designed to switch a small protective voltage
(see “11. Technical data” on page 101). The device can be damaged by
excessively high voltages.
• If third-party items are used, functional failures may occur and fitness for use
may be restricted. Biocompatibility requirements may also not be met. Please
note that in such cases, any claim under warranty and liability will be voided if
neither the accessories nor genuine replacement parts recommended in the
instructions for use are used.
• Have servicing and maintenance work carried out only by the manufacturer,
Weinmann, or by specialist staff expressly authorized by the manufacturer.
• Have modifications to the device carried out only by the manufacturer,
Weinmann, or by specialist staff expressly authorized by the manufacturer.
34Safety instructions
Page 35
EN
Note:
• If the therapy device is stored or not used for a prolonged period, the battery will
discharge. This is a property of rechargeable batteries and is not a malfunction,
so we recommend checking charge status regularly and recharging the battery
with the aid of the therapy device if required.
• In the event of error messages, please see “8. Troubleshooting” on page 79.
3.2 Contraindications
The therapy device should not be used or should be used only with particular caution in
the case of the following diseases. In the individual case, the decision about therapy is the
responsibility of the doctor supervising treatment.
• Cardiac decompensation
• Severe cardiac arrhythmia
• Severe hypotension, particularly in combination with intravascular volume
depletion
• Severe epistaxis
• High risk of barotrauma
• Pneumothorax or pneumomediastinum
• Pneumoencephalus
• Skull trauma
• Status following brain surgery and following surgical intervention at the pituitary
gland or the middle/inner ear
• Acute sinusitis, otitis media or perforated eardrum
• Dehydration
Mask ventilation must particularly not be used in the case of severe swallowing disorders
(bulbar disorders) with the risk of aspirations.
Dangerous situations involving this therapy device have not yet been observed.
Safety instructions35
Page 36
EN
3.3 Side effects
When using the therapy device, the following undesired side effects may occur in shortterm or long-term use:
• pressure points on the face from the breathing mask and the forehead cushion
• reddening of facial skin
• blocked nose
•dry nose
• dry mouth in the morning
• feeling of pressure in the sinuses
• irritated mucous membrane in the eyes
• gastrointestinal insufflation of air (“bloating”)
• nosebleeds
• muscular atrophy in the case of long-term ventilation
36Safety instructions
Page 37
EN
4. Set up device
Warning!
Device failure if incorrect patient circuits used!
If patient circuits with a diameter smaller than Ø 15 mm are used, the device may
overheat.
– Use only patient circuits with a diameter of Ø 15 mm or more.
– Note that total permitted resistance may be exceeded even in patient circuits
with a diameter of Ø 15 mm when these are combined with bacteria filters.
®
– Note on using Intersurgical
When using Intersurgical
single-patient circuits in which Intersurgical valves from the patient circuits mentioned are used) the volume display on the device may deviate. The volume displayed
may deviate from that actually delivered by up to 105 ml.
– Use patient circuit WM 27181 from Weinmann to avoid this deviation in the
display.
The therapy device can be operated with both leakage and valve ventilation. The doctor
supervising treatment or the authorized specialist dealer converts the device to the variant
appropriate for the patient's form of treatment. The necessary settings on the device are
likewise made by the doctor supervising treatment.
You need only follow the relevant section depending on which patient circuit is being used.
Different functions and connection options for accessories are available depending on the
patient circuit. This is clearly indicated in the relevant sections.
patient circuits
®
patient circuits ref. 5183064 and ref. 5083 (and other
4.1 Set up and connect the device
Set up the device on a flat surface, e.g. on a bedside cabinet or on the floor next to the
bed. Maintain a distance of at least 5 cm between the wall and the rear of the device, as
the air inlet is at the rear of the device. A distance of at least 5 cm should also be
maintained from the left-hand side of the device to allow the heat produced by the device
to escape.
Set up device37
Page 38
EN
Caution!
Cable-securing clip
Do not cover the device with blankets etc. The air inlet would be blocked and the
device could overheat. This may lead to inadequate therapy and to damage to the
device.
1. Connect the power cord to the power connector of
the device.
To do this, lift the cable-securing clip, plug the plug
into the power connector and flip the cable-securing
clip over the plug. Always secure the power plug with
the cable-securing clip to prevent the plug being
removed by mistake.
2. Connect the power cord to a power supply socket.
The therapy device is designed for a supply voltage of
115 V ~ and 230 V ~.
The green LED for power supply comes on and the
start screen appears in the display.
Now leave the device connected to the power supply for at least 24 hours to charge the
internal battery.
The device is now operational.
4.2 Patient/ventilator interfaces
The therapy device is intended for operation with nasal masks, oronasal masks and full-face
masks as well as with endotracheal cannulas and endotracheal tubes. It is essential to
follow the instructions for use of the patient/ventilator interface in question.
4.3 Connect valve ventilation
When using valve ventilation, patient/ventilator interfaces with leakage openings may not
be used. Always use a patient valve. Inspiration and exhalation is controlled using the
patient valve.
4.3.1Single patient circuit
The single patient circuit consists of a ventilation tube, a pressure-measurement tube, a
valve control tube and a patient valve. The patient/ventilator interface must be connected
directly to the patient valve.
38Set up device
Page 39
EN
Caution!
1
3
2
4
P
1
3
2
4
1
2
3
4
5
P
The patient valve may not be covered when the device is in operation, otherwise
the exhaled air may not be routed away, obstructing breathing.
Proceed as follows to connect the single patient circuit to the therapy device.
1. Plug the free end of the ventilation tube (2) onto the
device outlet port.
2. Now connect the blue connector stub of the pressuremeasurement tube (1) to the connection of the device
which is likewise blue and marked .
3. Connect the valve control tube (3) to the connection
of the device marked .
4. Connect the patient valve (4) to the patient/ventilator
interface, e.g. a mask.
It is essential to follow the instructions for use of the
patient/ventilator interface and of the patient circuit.
4.3.2Double patient circuit (VENTIlogic LS only)
In addition to the ventilation tube which delivers air to the patient, the pressuremeasurement tube and the valve control tube, the double patient circuit also has an
exhalation tube which routes exhaled air back to the device and into the ambient air.
On the double patient circuit, the patient valve is located on the exhalation tube.
Proceed as follows to connect up the double patient circuit.
1. Plug the ventilation tube (1) onto the device outlet
port of the VENTIlogic LS.
2. Plug the exhalation tube (3), on the end of which is the
patient valve (4), onto the device inlet port for exhaled
air underneath the device outlet port.
3. Now connect the blue connector stub of the pressuremeasurement tube (2) to the connection of the device
which is likewise blue and marked .
Set up device39
Page 40
EN
The pressure-measurement tube is the same length as
1
2
3
5
4
P
the ventilation tube and leads to the Y-connecting
piece where the ventilation tube and the exhalation
tube are brought together.
4. Connect the valve control tube (5) to the connection
of the device marked .
The valve control tube leads straight from the patient
valve to the connection on the device and is therefore
shorter than the pressure measuring tube.
5. Connect the patient/ventilator interface, e.g. a mask.
It is essential to follow the instructions for use of the
patient/ventilator interface and of the patient circuit.
4.4 Connect leakage ventilation
Proceed as follows to connect up the leakage circuit.
1. Plug the click adapter of the patient circuit onto the
ventilation outlet on the device.
2. Connect the patient/ventilator interface, e.g. a mask.
It is essential to follow the instructions for use of the
patient/ventilator interface and of the exhalation
system.
4.4.1Connect leakage adapter using standard tapered connector
For leakage operation, an adapter is available as an option to allow tubes with sleeves with
an internal diameter of Ø 22 mm and a pressure-measuring tube with an internal diameter
≤ Ø 5 mm to be used. If the device is equipped with this adapter, proceed as follows to
connect the tube:
1. Push the ventilation tube onto the device outlet port
of the therapy device.
2. Now connect the pressure-measurement tube to the
connection on the device marked .
3. Connect the patient/ventilator interface, e.g. a mask.
40Set up device
4. It is essential to follow the instructions for use of the
patient/ventilator interface and of the exhalation
system.
Page 41
EN
5. Note that maximum flow rate and the accuracy of
Exhalation system
Circuit
Sleeve
Patient/ventilator
interface
dynamic pressure may deviate if you are not using
Weinmann tubes.
4.4.2Information relating to a separate exhalation system
Caution!
Always use an exhalation system. Used air containing carbon dioxide escapes from
the patient/ventilator interface (e.g. a mask) through the exhalation system.
Without an exhalation system, the CO
concentration in the patient/ventilator
2
interface and the ventilation tube would rise to critical values and thus obstruct
breathing.
An exhalation system can either be integrated in the patient/ventilator interface, e.g. a
mask, or must be plugged in between the patient/ventilator interface and the patient
circuit in the form of an accessory.
If the patient/ventilator interface, e.g. the mask, does not have an exhalation system, a
separate exhalation system, e.g. Silentflow 2, must be used.
The exhalation system also allows a patient to breathe through his or her nose for a short
time, even if the device were to fail. In the case of full-face masks, breathing in the event
of a fault is through an emergency exhalation valve on the mask.
4.4.3Connect separate exhalation system (only with leakage
ventilation)
The sleeve on the end of the patient circuit has a diameter of 19.5 mm and fits over a
22 mm standard tapered connector.
To connect a separate exhalation system, proceed as
follows.
1. Plug the exhalation system into the sleeve of the circuit.
2. Connect the patient/ventilator interface to the
exhalation system.
Follow the instructions for use for the exhalation system and for the patient/ventilator
interface.
Set up device41
Page 42
EN
4.5 Connect humidifier
4.5.1Leakage ventilation
Warning!
The VENTIclick humidifier is not suitable for invasive ventilation.
The VENTIclick humidifier is connected between the
therapy device and the patient circuit. The inlet connector
stub and the electric connection of the heating element
must point towards the therapy device. Follow the
instructions for use for the VENTIclick.
Note!
An optional tube adapter is available as an accessory for the use of Fisher and Paykel
humidifiers. Be aware that technical data change when third-party humidifiers are used.
Deviations in the technical data are described in the assembly instructions for the tube
adapter.
4.5.2Valve ventilation
VENTIclick cannot be used in conjunction with valve ventilation. We recommend humidifiers
HC 850 or HC 550 from Fisher & Paykel with the corresponding patient circuit. Follow the
associated instructions for use.
4.6 Connect bacteria filter
Caution!
The bacteria filter represents an additional resistance in the air flow. This can cause
a change to the response characteristics of the trigger. If a bacteria filter is
connected subsequently, the doctor must therefore check the device parameters
and may need to reset them.
Note:
The bacteria filter may not be operated on the device for more than 24 hours.
Follow the instructions relating to period of use in “6. Hygiene treatment” on
page 65.
42Set up device
Page 43
EN
If the therapy device is intended for use by several patients (e.g. in a hospital), a bacteria
WM 24148
WM 24476
filter must be used to prevent infections.
4.6.1Leakage ventilation
In combination with leakage ventilation, use bacteria
filter WM 24148.
If the bacteria filter is used alone, it is connected directly
to the device outlet port and the ventilation tube is
plugged onto the bacteria filter.
If the VENTIclick humidifier is also to be connected,
proceed as follows.
1. Connect the VENTIclick humidifier directly to the therapy device.
2. Connect the bacteria filter to the outlet of the VENTIclick humidifier.
3. Connect the patient circuit to the bacteria filter.
4.6.2Valve ventilation
In combination with valve ventilation, use bacteria
filter WM 24476.
If the bacteria filter is used alone, it is connected directly
to the device outlet port and the ventilation tube is
plugged onto the bacteria filter.
If a humidifier and/or an oxygen sensor is to be connected as well, a different sequence
applies.
Combination with a humidifier
1. Connect the bacteria filter directly to the therapy device.
2. Connect the humidifier to the outlet of the bacteria filter.
3. Connect the patient circuit to the humidifier.
Combination with an oxygen sensor
1. Connect the oxygen sensor directly to the therapy device.
2. Connect the bacteria filter to the outlet of the oxygen sensor.
3. Connect the patient circuit to the bacteria filter.
Set up device43
Page 44
EN
Combination with an oxygen sensor and a humidifier
1. Connect the oxygen sensor directly to the therapy device.
2. Connect the bacteria filter to the outlet of the oxygen sensor.
3. Connect the humidifier to the outlet of the bacteria filter.
4. Connect the patient circuit to the humidifier.
4.7 Therapy with oxygen supply
4.7.1Supplying oxygen
Warning!
• If oxygen is being supplied to the respiratory flow, smoking and naked flames
are forbidden. Risk of fire. The oxygen can accumulate in clothing, bed linen
or hair. It can only be removed by thorough ventilation.
• To supply oxygen, use only the therapy device connection provided for this
purpose. The therapy device has a safety valve which shuts off the oxygen
supply in the event of a fault, otherwise the oxygen supply cannot be stopped
automatically if a fault occurs. Never supply oxygen via masks or the T-adapter.
Note:
Supplying oxygen via a connection not intended for this purpose leads to an
erroneous volume display, as the oxygen flow cannot be included in the
measurement.
A supply rate of max. 15 l/min at < 1000 hPa pressure at the inlet for the oxygen supply is
permitted.
Use an oxygen sensor to monitor oxygen supply.
The oxygen source must have an independent flow regulation device. Ensure that you only
set the oxygen flow rate prescribed by the doctor on your oxygen supply device. It is
essential in this case to follow the safety instructions for handling oxygen as well as the
instructions for use of the oxygen device used.
Caution!
If the oxygen is humidified before being supplied, a tank with an overpressure valve
must be used, otherwise an overpressure will result in the event of faulty operation
which could lead to the humidification tank bursting or to the oxygen supply tubes
slipping off.
44Set up device
Page 45
EN
Proceed as follows to supply oxygen:
3
2
1
1. Switch on the therapy device.
2. Connect the O
coupling supplied to the connector
2
stub provided on the therapy device.
3. Connect the oxygen source to the O
coupling.
2
4. Start the supply of oxygen. When doing so, it is
essential to follow the instructions for use of the
oxygen source in question. The device can now be
operated normally.
Proceed as follows to end supply of oxygen:
1. Shut off the oxygen supply.
2. Continue operating the device for a while without an oxygen supply to flush the
remaining oxygen out of the device. If this instruction is not followed, there is a risk of
fire in the event of a malfunction.
3. Remove the adapter for the oxygen supply from the device.
4. Switch off the device. The safety valve for the oxygen supply shuts off the oxygen
supply.
Oxygen can be supplied via an oxygen concentrator (e.g. Weinmann Oxymat 3), via the
central gas supply system (only with corresponding pressure reducer) of a hospital, in the
form of liquid oxygen with a continuous flow or of an oxygen cylinder with a corresponding
pressure reducer.
The oxygen sensor can only be used in conjunction with valve ventilation.
During measurement, oxygen concentration is averaged over several breaths and
displayed. Measured values depend on therapy pressure and on the temperature of
ambient and respiratory air. This is not a FiO
concentration on inspiration.
1. The adapter is delivered in three parts: the oxygen
measurement, but the mean value of oxygen
2
sensor (1), the T-adapter (3) and an air management
adapter (2). Screw the air management adapter onto
the oxygen sensor.
Set up device45
Page 46
EN
2. Plug the oxygen sensor (1) and air management
31
3
6
5
4
adapter into the T-adapter (3).
3. Plug the T-adapter (3) onto the device outlet port (5).
4. Connect the sensor (1) to the oxygen measuring jack
(4) with the aid of the cable.
5. Connect the patient circuit - with a bacteria filter if
required - as shown in the illustration.
6. Calibrate the oxygen sensor (see “7.3 Calibrate
oxygen sensor” on page 75).
4.8 Connect mobile power supply
VENTIpower is available for power supply-independent operation. VENTIpower facilitates a
mobile power supply.
Warning!
• VENTIpower must be switched on at the On/Off switch before the start of
therapy so that in the event of a power failure or the failure of the internal
battery, VENTIpower can assume supply to the therapy device.
• Do not use VENTIpower as an uninterrupted power supply. VENTIpower does
not switch on automatically if the power supply fails. This can lead to the
patient being put in danger, especially in the case of life-support ventilation.
46Set up device
If you want to use the therapy device with VENTIpower,
proceed as follows:
1. Connect VENTIpower to the therapy device.
2. Press the On/Off key to switch on VENTIpower.
Page 47
EN
VENTIpower now assumes the supply of power. The
symbol appears in the status line of the therapy
device. The power supply LED goes out.
Using VENTIpower, you can operate the therapy device
(independently of the power supply) in carrying bag
WM 27706. To do this, VENTIpower is accommodated in
the compartment provided in the bag and connected to
the therapy device.
The instructions for use of the VENTIpower should also be
followed, as should the safety instructions relating to
mobile operation and transport of the device in particular.
4.9 Operation in the event of a power failure
If the power supply should ever fail, the internal battery of the therapy device automatically
assumes supply of the device.
The message No power supply appears. The green power supply LED goes out. The
battery operating time will depend on the load and temperature range. Detailed
information on the different loads with the corresponding battery operating times are
provided in section 11. on page 101.
As soon as the power supply is restored, the device is automatically supplied from the
power supply again and the internal battery is charged. The green power supply LED comes
on and the consecutive segments in the battery symbol indicate the charging process in the
display.
Note
• If the alarm Battery capacity critical appears, action is required. In this case,
the battery has only about 25 % capacity left. This is enough for about 15
minutes. Keep an alternative ventilation option to hand.
• If the alarm Battery capacity highly critical appears, there is less than 10 %
capacity remaining. The device will switch itself off in a few minutes. Use the
alternative ventilation option at once.
Set up device47
Page 48
EN
5. Operation
2
1
4
53
5.1 Controls
5.1.1Function keys
The following functions can be called up directly in
ventilation mode by pressing the relevant key on the
device.
• Activate softstart and set softstart time (1)
Note:
Softstart can only be activated in CPAP mode.
• Activate humidifier and set humidifier stage (only
leakage ventilation) (2)
• Acknowledge alarms (3)
• LIAM (insufflation) (4)
After these keys are pressed, the corresponding menu appears in the display. You can
navigate within the menu using the dial (see “Navigating with the dial” on page 49).
The other functions (5) can only be operated by the doctor.
Menu key
Use the menu key to switch from Monitor to Menu.
Current values during therapy are displayed in Monitor. You can make settings to the
device in Menu.
The menu key has other functions (e.g. back) depending on context. The current function
is always displayed on the left of the display next to the menu key.
Of the functions in the patient menu, only the Calibrate O
ventilation mode. Drying process, Function check and Auto switch-on can only be
activated in standby mode. Alarm list and Filter change can be selected at any time.
Acknowledge alarm
Use the alarm acknowledgement key to acknowledge an acoustic alarm and mute it for
120 seconds.
sensor function is available in
2
48Operation
Page 49
EN
5.1.2Navigating with the dial
1
2
The dial (1) is the central control of the therapy device.
You can use the dial to select menu items, navigate
within the menu windows and set values for individual
menu items.
To familiarize yourself with navigation using the dial, we
recommend switching to Menu first. Press the menu key
(2) to do so. You can then try out the functions described
below.
Select menu items
• Move the dial clockwise to move the selection bar in the display downwards.
• Move the dial anticlockwise to move the selection bar in the display upwards.
• Press the dial to confirm selection of a menu item and to open the corresponding
submenu or to select a value you want to change.
Set values
• Move the dial clockwise to increase a value.
• Move the dial anticlockwise to decrease a value.
• Press the dial to save a value.
Exit menu item
Move the dial clockwise until the selection bar in the
display is on back, cancel or close depending on
context. Then press the dial. The display switches back to
the next menu up.
Alternatively, you can exit a menu item by pressing the
menu key (back, cancel or close will appear in the
display to the left of the menu key depending on
context).
Operation49
Page 50
EN
5.2 Start up the device
5.2.1Operating states
Three operating states are possible on the therapy device: on, off and standby.
If the device is switched on, therapy is in progress.
In standby, the blower is switched off, but the device is immediately operational with a
brief press of the On/Off key as long as the patient circuit is connected correctly. Settings
can be made on the device in standby mode.
If the device is switched off completely, the blower and the display are likewise switched
off and no settings can be made on the device.
5.2.2Starting up
1. Connect the device to the power supply with the aid
of the power cord. The standby screen appears on the
display.
2. First, perform a function check (see “7. Function
check” on page 71).
50Operation
3. Connect the patient circuit to the patient/ventilator
interface. It is essential to follow the relevant
instructions for use for the patient/ventilator interface,
the patient circuit and, if appropriate, the exhalation
system.
Caution!
Always use a separate exhalation system for
leakage ventilation (e.g. Silentflow leakage
ventilation), otherwise the CO
would rise to critical values in the patient/
ventilator interface and tube and thus obstruct
the patient's breathing.
concentration
2
Page 51
EN
4. To switch on the device, press the On/Off key
On/Off key
Default display
briefly. The device is now in ventilation mode. The
patient menu can be accessed via the menu key.
When Auto switch-on (only with leakage ventilation) is
activated, you can also put on the patient/ventilator
interface and switch on the therapy device by taking a
breath (see “5.3 Activate/deactivate Auto switch-on
(only leakage ventilation)” on page 52). The operating
hours and the Weinmann software version appear in the
display for about 3 seconds.
The device starts to pump air through the patient circuit. The display switches to the default
display.
5.2.3Displays on screen
Ventilation parameters such as set therapy mode,
therapy pressures (CPAP pressure only in CPAP mode)
in hPa, the selected patient circuit and current
respiratory frequency in 1/min are shown in the
display.
Note: 1.01973 hPa correspond to 1 cm H
If softstart is activated, the softstart display
appears with the remaining softstart time, max. 30
minutes, or the max. softstart time (see “5.8 Set
softstart (in CPAP mode only)” on page 55) specified
by the doctor.
The bar chart shows the pressure curve for inspiration
and exhalation.
The respiratory phase change display shows whether
the current respiratory phase was triggered
spontaneously by the patient (
T). Depending on respiratory phase, the display
(
switches from left (inspiration) to right (exhalation).
The illustration shows an exhalation triggered by the
machine.
S) or by the machine
O.
2
Operation51
Page 52
EN
See the section entitled “ Symbols used in the display”
Softstart display
Bar chart
on page 19 for an explanation of the other symbols in
the display.
Note:
If a therapy in TA mode is intended, see also the
section entitled “5.12 Therapy in TA mode (only
leakage ventilation)” on page 60.
5.3 Activate/deactivate Auto switch-on (only
leakage ventilation)
When Auto switch-on is activated, the therapy device switches on automatically as soon as
you breathe into the patient circuit. The device does not switch off again automatically
when the patient takes off the patient/ventilator interface. You can only switch off the
therapy device using the On/Off key .
Note
Auto switch-on can only be activated or deactivated in standby mode.
1. Start up the therapy device (see “5.2 Start up the device” on page 50).
2. Press the menu key to do so. The patient menu appears in the display.
52Operation
Page 53
EN
3. Use the dial to select the menu item Auto switch-on
and confirm the selection by pressing the dial. Now
select on or off using the dial. Confirm the selection
by pressing the dial. The selection bar switches back to
Auto switch-on. The current setting (on/off) is now
shown again in the Auto switch-on menu line.
4. Exit the menu again by pressing the menu key, now
assigned the back function. Auto switch-on is now
activated or deactivated.
5.4 Alarm list
5.4.1Storage of alarms
All alarm types listed in the tables “Physiological alarms”and “Technical alarms” are
recorded in an alarm list with date, time and duration once the alarm threshold is reached.
Up to 200 alarms can be stored. After that, the oldest alarm in each case is overwritten.
To call up the alarm list, select the menu item Alarm list in the patient menu using the dial
and confirm your selection by pressing the dial.
The alarm list is retained even if the entire power supply (power supply and internal battery)
fails.
In this case, the data can be called up for up to two years. The alarm list is overwritten after
two years or when servicing is carried out.
Alarms are always retained in the event of a power failure.
Operation53
Page 54
EN
5.5 Adjust brightness
You can change the brightness of the display:
1. In the patient menu, select the Brightness menu item
using the dial.
2. Select the desired brightness level using the dial.
3. Confirm your selection by pressing the dial.
Tip
To switch off the display completely during therapy (at
night for example), press the dial during therapy. Switch
the display back on by pressing the dial again.
5.6 LIAM info
Note
A detailed explanation of the LIAM function can be found in the section entitled
“5.11 LIAM (insufflation)” on page 57.
1. In the patient menu, use the dial to select the LIAM info menu item.
2. Confirm your selection by pressing the dial.
You will find the following values and their residual
running times under LIAM info:
• Duration
• Interval
• Cycles
• Plateau signal (activated or deactivated)
54Operation
Page 55
EN
5.6.1Activate/deactivate plateau signal
In patient mode you can switch the plateau signal on or
off under LIAM info:
1. In the LIAM info menu, use the dial to select the
plateau signal.
2. Confirm your selection by pressing the dial.
3. Use the dial to select the status plateau signal on
or plateau signal off .
4. Confirm your selection by pressing the dial.
5.7 Overview
Under the menu item Displays > Summary you can
have the current Settings, Actual values and Alarms
displayed.
1. In the Patient menu, use the dial to select the
Summary menu item.
2. Confirm your selection by pressing the dial.
3. Use the dial to select the desired submenu.
4. Confirm your selection by pressing the dial.
5.8 Set softstart
(in CPAP mode only)
If your doctor has enabled the softstart function for you, softstart time can be selected in
5-minute increments up to a maximum time of 30 minutes. Your doctor may limit the
maximum time to less than 30 minutes.
Note:
The softstart function is only available in CPAP mode.
Operation55
Page 56
EN
The setting can be set in both standby and in ventilation mode. To do so, proceed as
follows.
1. Keep the softstart key depressed until the
Softstart window appears.
2. Change the softstart time using the dial (turn to the
right = increase time, turn to the left = reduce time).
Alternatively, you can also increase the softstart time
in 5-minute increments by pressing the softstart key
several times (once the maximum time has been
reached, the device starts back at the minimum time
of 5 minutes if you press again).
3. To save the softstart time, press the menu key or
the dial.
The softstart time displayed is saved and the softstart
window closes automatically. If no key is pressed for 4
seconds, the softstart time displayed is likewise saved.
The settings are retained after the device is switched off.
Note:
• The therapy device automatically starts in softstart mode if this was activated the
last time it was used.
• The softstart function can be switched on or off at any time by briefly pressing
the softstart key .
5.9 Set VENTIclick humidifier
(only leakage ventilation)
With the VENTIclick humidifier (only with leakage ventilation), you can humidify the
respiratory air provided by the therapy device. Heating output can be selected in 6 stages.
The setting can be set in both standby and in ventilation mode. To do so, proceed as follows.
1. Press the humidifier key for approx. 3 seconds until
the Humidifier level window appears. Change the
heating output (stages 1-6) of the humidifier using the
dial or repeatedly press the key.
2. Press the menu key or the dial to store the heating
output. The Humidifier level window closes
automatically. If no key is pressed for 4 seconds, the
heating output displayed is likewise saved. The
settings are retained after the device is switched off.
56Operation
Page 57
EN
Also follow the instructions for use of the VENTIclick or, if using valve ventilation, the
instructions for use of the corresponding humidifier. For valve ventilation we recommend
humidifiers HC850 or HC550 from Fisher & Paykel with the corresponding patient circuit.
Note:
• The therapy device automatically starts with VENTIclick activated if it was
activated the last time the device was used. VENTIclick can be switched on and
off at any time if you briefly press the humidifier key . If the humidifier is
activated, the green LED next to the humidifier key comes on.
• If you press the humidifier key without a humidifier being connected, the
therapy device does not switch on this function.
5.10 Humidifier for patient circuits with patient
valve
VENTIclick cannot be used in conjunction with valve ventilation. We recommend humidifiers
HC 850 or HC 550 from Fisher & Paykel with the corresponding patient circuit for this case.
Follow the associated instructions for use.
5.11 LIAM (insufflation)
5.11.1 Information about the function
LIAM stands for Lung Insufflation Assist Maneuver. LIAM is a pressure-controlled
hyperinsufflation maneuver with the aim of administering an increased tidal volume which
can be used in all ventilation modes except CPAP and SIMV. LIAM can be used to support
coughing or for alveolar recruitment (similar to ventilation on sighing). In the case of
neuromuscular diseases, in particular, LIAM can be useful in expanding both thorax and
lung. With regular use, there can accordingly be a positive impact on the course of vital
capacity.
Operation57
Page 58
EN
LIAM can only be enabled by the doctor and only triggered during ventilation. The
Duration in min/h
(e.g. 20 min)
Pressure
Time
Interval in sec/min/h
(e.g. every 5 min)
Cycles
(e.g. 3 x consecutively)
Ventilation alternating with LIAM
maneuver includes at least one LIAM stroke consisting of insufflation and subsequent
exhalation.
Your doctor uses the Duration parameter to specify the period for which LIAM is applied.
The Interval parameter states at what intervals LIAM will be repeated. Within one interval,
either one LIAM stroke (Cycles = 1) or up to 10 consecutive LIAM strokes are executed.
LIAM to support a coughing maneuver
LIAM can be used to support a coughing maneuver by initially expanding lung and thorax
adequately during hyperinsufflation. This makes more air available for the subsequent
cough. At the same time, optimized initial tensioning of the lung and thorax increases the
effectiveness of the cough on exhalation.
58Operation
Page 59
EN
Pressure
Time
Ti
LIAM
Te
LIAM
IPAP
max
ΔP
LIAM
Signal which can be
switched off
1 sec. plateau phase
Initially, the pressure curve within the individual LIAM stroke is comparable with a normal
ventilation stroke. When the IPAP pressure level is reached, however, pressure continues to
rise in linear form to maximum pressure IPAP
(IPAP + ΔP
max
) and is maintained for
LIAM
one second (plateau phase). The plateau phase at the end of insufflation is to facilitate
the coordination (closure of the glottis) of a coughing maneuver. The start of the plateau
phase is audibly emphasized by an optional plateau signal. This plateau signal can be
switched on and off in the menu under LIAM info (see “5.6.1 Activate/deactivate plateau
signal” on page 55). You can also have the following values displayed under LIAM info:
Duration, Interval and Cycles. LIAM ends automatically once the set duration has
elapsed or can be cancelled manually (see “5.11.3 Canceling LIAM” on page 60).
5.11.2 Method
You can trigger LIAM manually during ventilation. Press
the key to do so. The device switches to LIAM mode
and insufflation is started to synchronize with the next
inspiration.
You can perform the whole process yourself several
times. Press the key again to do so.
Operation59
Page 60
EN
5.11.3 Canceling LIAM
LIAM can be interrupted at any time. Press the key to do this. Then LIAM is canceled
and the device reverts to the preset ventilation mode. If LIAM is then to be carried out
again, begin the process by pressing the key again.
5.12 Therapy in TA mode (only leakage
ventilation)
In TA mode the therapy device adapts to the personal respiratory rhythm of the patient.
Proceed as follows to achieve optimum adaptation to your respiratory rhythm.
1. Put on the patient/ventilator interface shortly before you switch on the therapy device.
2. Switch on the therapy device and continue to breathe
normally. The symbol AA (automatic analysis) appears
in the display. The device initially sets a constant
pressure of 4 hPa and analyzes the respiratory rhythm
of the patient.
As soon as the device has detected the patient's
respiratory rhythm (after 10 - 20 breaths), it provides the
prescribed therapy pressure in precisely this rhythm.
If the respiratory rhythm should change during therapy,
the therapy device detects this automatically and
performs a new analysis. The symbol A
display again. The patient can also continue to breathe
normally during this phase. Pressure is reduced to 4 hPa
again for 10 - 20 breaths. As soon as the device has
detected the new respiratory rhythm, it provides the
therapy pressure in this new rhythm.
A
appears in the
60Operation
Page 61
EN
5.13 After use
1. Switch the device to standby by keeping the On/Off
key depressed for approx. 2 seconds until the blower
switches off. The duration of the previous therapy
appears in the display. The device then switches to
standby.
2. Disconnect the patient/ventilator interface from the
patient circuit and the patient circuit from the device.
3. Clean the patient/ventilator interface, patient circuit
and device in accordance with the instructions for use.
For information on this, see the section entitled
“Hygiene treatment” on page 65.
5.13.1 Switch device off completely
If the device is connected to a power supply, do not let it be turned off completely. To
completely disconnect the device from the power supply, turn the device to standby mode
and disconnect the device from the power supply. For optimal battery charging, however,
we recommend not to disconnect the device from the power supply.
To switch off the device completely in battery-operated mode, first switch to standby by
pressing the On/Off key for about 2 seconds until the blower switches off. Then press the
On/Off key again for at least 2 seconds until the device switches off completely and the
display goes out.
5.13.2 Mobile therapy data check
The therapy device has a memory card reader for MMC or SD cards which can be used to
save therapy data on a memory card. In discussion with the doctor supervising treatment,
this allows the patient's therapy data to be read out independently of the location of the
device, as the data can be transported on the memory card.
The following data are stored on the memory card:
– therapy pressure in hPa
– respiratory flow in l/min
– volume, leakage corrected, in ml
– current respiratory phase
– mean leakage flow in l/min
– current ratio Ti/T in %
– current respiratory frequency in 1/min
Operation61
Page 62
EN
– ratio of spontaneous inspiration to total number of inspirations in %
– ratio of spontaneous exhalation to total number of exhalations in %
– number of effort episodes, related to number of breaths (TA mode only)
– number of fighting episodes, related to number of breaths (TA mode only)
– tidal volume of last inspiration in ml
– mean respiratory minute volume in ml/min
– current physiological alarms
– current technical alarms
– current warnings
This data can be read out from the memory card and displayed with the aid of
WEINMANNsupport.
If a card is in the device and therapy data is being recorded, a symbol appears in the
status line. If the symbol does not appear, the memory card is defective, absent or not yet
recognized by the device.
Caution!
The device must be on standby or switched off for the MMC or SD card to be
inserted or removed, otherwise therapy data may be lost.
Note
The SD card can only be detected by the device when ventilation mode is running.
After inserting the memory card, run the device briefly until the SD card is detected
and the symbol is displayed in the status line.
Proceed as follows to remove the memory card.
62Operation
1. The slot for the memory card is located on the side of
the device under a rubber cover. Pull on the rubber
cover to get at the memory card.
2. To remove the memory card, press briefly on the
memory card in the device. A spring mechanism now
pushes the memory card out a little way.
3. Remove the memory card.
4. Cover the slot for the memory card again using the
rubber cover.
Page 63
EN
Proceed as follows to put the memory card back in.
1. Pull on the rubber cover to get at the slot for the
memory card.
2. Push the memory card into the slot with the cut-off
corner pointing upwards.
3. Briefly press on the card so that the card can engage
in the device with the aid of the spring mechanism.
4. Cover the slot for the memory card again using the
rubber cover.
Caution!
When covering the memory card with the rubber cover, take care not to push in
the card accidentally, as doing so will eject it from the device. Ejecting the card may
cause loss of therapy data.
5.14 Travel with the therapy device
5.14.1 Bags for the therapy device
The therapy device has two bags, a protective bag (WM 27106) and a carrying bag for
mobile use (WM 27706).
The protective bag WM 27106 is supplied and is for
protecting the device but not for mobile operation.
The carrying bag WM 27706 is available as an accessory
and allows the device to be operated on a mobile basis.
The therapy device can be operated together with
VENTIpower in this bag.
Operation63
Page 64
EN
5.14.2 Before starting mobile operation
You should only transport the therapy device any distance in the protective bag WM 27106
intended for it. If you want to use the therapy device on a mobile basis, you must use it in
carrying bag WM 27706.
Proceed as follows to operate the device on a mobile
basis.
1. First connect the VENTIpower mobile power supply to
the therapy device.
2. Fit the patient circuit and the patient/ventilator
interface.
3. Now put the therapy device and the VENTIpower in
the relevant compartments in the bag and switch on
both devices.
4. Secure the devices with the hook-and-loop closures in
the bag.
5. A fabric tube is attached to the front of the bag with
a hook-and-loop closure to protect the devices from
moisture and rain. Pass the ventilation tubes through
this fabric tube.
6. Close the bag and check whether both devices are
reliably secured in the bag and cannot wobble or fall
out.
7. Attach the tube of fabric and the ventilation tubes to the side of the therapy device
using the hook-and-loop attachment provided.
Follow the instructions for use for the VENTIpower for operation in the bag.
If you want to take the therapy device and accessories onto an aircraft as hand baggage,
find out before departure whether any formalities are involved. You can obtain a certificate
for transport in an aircraft from the manufacturer, Weinmann.
64Operation
Page 65
EN
6. Hygiene treatment
This product may contain disposable items. Disposable items are intended to be used
only once. So use these items only once and do not reprocess them. Reprocessing
disposable items may impair the functionality and safety of the product and lead to
unforeseeable reactions as a result of ageing, embrittlement, wear, thermal load, the
effects of chemical processes, etc.
6.1 Intervals
You should check the filters at regular intervals and wipe down the housing and the filter
compartment lid with a damp cloth. You should also observe the following intervals:
6.1.1Leakage ventilation
IntervalActivity
– Clean the patient/ventilator interface in accordance with the
relevant instructions for use.
– Clean the patient circuit.
– Clean bacteria filter WM 24148 in accordance with the
Daily
Every 24 operating
hours
Weekly
Every 1000 operating
hours
Every 6 months
instructions for use.
– Clean the humidifier in accordance with the relevant
instructions for use.
– Clean the exhalation system in accordance with the instructions
for use every time it is used.
– Change the particulate filter in bacteria filter WM 24148.
– Clean coarse dust filter.
– Clean fan filter.
– Change fine filter (filter change indicator ), earlier if dirty.
– Change coarse dust filter, earlier if dirty or worn.
– Change pressure measuring tube (see “9.4 Change pressure-
measurement tube (only leakage ventilation)” on page 94),
earlier if dirty.
– Change fan filter.
Hygiene treatment65
Page 66
EN
IntervalActivity
Annually– Change patient circuit.
Follow the relevant instructions for use on the hygiene treatment for the patient/ventilator
interface.
6.1.2Valve ventilation
IntervalActivity
– Clean the humidifier in accordance with the relevant
Daily
Weekly
Every 1000 operating
hours
Every 6 months
The patient circuits for single and double patient circuits with patient valve (double patient
circuit only on VENTIlogic LS) are disposables and cannot be subjected to a hygiene
treatment. Follow the instructions for use for the patient circuit in question.
Follow the relevant instructions for use for the hygiene treatment of the patient/ventilator
interface.
instructions for use.
– Change bacteria filter WM 24476.
– Clean coarse dust filter.
– Clean fan filter.
– Change fine filter (filter change indicator ), earlier if dirty.
– Change coarse dust filter, earlier if dirty or worn.
– Change fan filter.
6.2 Clean leakage ventilation
6.2.1Clean patient circuit
1. Pull the patient circuit off the device and the exhalation system.
2. Pull out the one end of the pressure measuring tube
(shake a little if necessary) and seal it with the sealing
plug supplied. At the other end, seal the small opening
of the adapter using the second sealing plug so that
no water can penetrate.
3. Clean the creased tube with a little detergent in hot
water and make sure no residues are left behind. Flush
the inside of the tube through thoroughly in the
process.
66Hygiene treatment
Page 67
EN
4. Rinse the creased tube thoroughly inside and out using clean hot water.
5. Thoroughly shake out the patient circuit.
6. Hang up the patient circuit and leave to drip-dry well to stop moisture getting into the
therapy device.
7. Remove the plugs from the pressure-measurement tube.
6.2.2Dry the patient circuit using the therapy device
If water ever gets into the pressure measuring tube by accident, the
leakage circuit must be dried with the aid of the therapy device.
This function can only be activated in standby mode. Likewise press the
On/Off key to switch the device to standby.
To start the drying process, proceed as follows.
1. Plug the red drying adapter supplied into the device
outlet port.
2. Plug the adapter for the patient circuit onto the red
drying adapter.
3. Press the menu key to do so. The patient menu
appears in the display.
4. Use the dial to select the menu item Drying process
and confirm this selection by pressing the dial.
The message Drying process active! 30 min appears.
This display remains active throughout the entire
drying process and indicates remaining drying time.
After the drying process is complete, the device
switches off.
5. If you want to interrupt the drying process, press the
menu key (cancel). The display switches back to
the default display, the device switches back to
standby.
If the patient circuit still has damp places after drying,
start the drying process again.
6. Remove the drying adapter from the device outlet port.
Hygiene treatment67
Page 68
EN
6.3 Clean the housing
Warning!
• Risk of electric shock. Switch the device off completely before cleaning (see
“Switch device off completely” on page 61).
• Ensure that no liquids get into the device. Never immerse the device in
disinfectants or other liquids, otherwise damage to the device and thus a hazard
to users and patients may result.
Proceed as follows to clean the housing.
1. Wipe down the device and the power cord with a soft damp cloth. The therapy device
must be completely dry before the device is started up.
2. Take off the filter compartment lid.
3. Remove the coarse dust filter as described in
“9.3 Change filter”.
4. Clean the filter compartment lid under running water
until there are no residues. Then dry it carefully.
5. Insert the coarse dust filter and the filter compartment
lid as described in “9.3 Change filter” on page 91.
6. Remove the fan filter on the front of the device and clean it as described in
“9.3 Change filter”.
7. Put the fan filter back in.
6.4 Clean coarse dust filter/change fine filter
1. Take off the filter compartment lid as described in “9.3 Change filter”.
2. Remove the coarse dust filter from out of the filter compartment lid and clean it with
clean running water until there are no residues.
3. Change the fine filter if required.
4. Allow the coarse dust filter to dry. The coarse dust filter must be completely dry before
the device is started up.
5. Put the coarse dust filter back in and close the filter compartment lid.
Note:
The fine filter cannot be cleaned. It is changed every 1000 operating hours.
68Hygiene treatment
Page 69
EN
6.5 Clean the fan filter
The fan filter protects the housing fan from dirt.
To clean the fan filter, proceed as follows:
1. Remove the fan filter according to the instructions in section “Change fan filter” on
page 93.
2. Clean the fan filter with fresh running water until it is free of residue.
3. Let the fan filter dry. The fan filter must be completely dry before starting the device.
4. Insert the fan filter again according to the instructions in section “Change fan filter”
on page 93.
6.6 Clean the accessories
To clean the accessories, see the section entitled "Hygiene treatment" in the
corresponding instructions for use.
6.7 Disinfect, sterilize
If required, e.g. following infectious diseases or unusual contamination, you can also
disinfect the housing, the power cord, the patient circuit (only leakage ventilation), the
bacteria filter housing, (only leakage ventilation), VENTIclick and VENTIpower. See the
instructions for use for the disinfectant used. We recommend wearing suitable gloves (e.g.
household or disposable gloves) for disinfecting.
6.7.1Device
The housing and the power cord of the therapy device are cleaned simply by wiping with
disinfectant. We recommend terralin
®
protect for this purpose.
6.7.2Patient circuit (leakage ventilation)
We recommend gigasept® FF as disinfectant. When using gigasept® FF, take the same
steps as described under “6.2 Clean leakage ventilation”.
Rinse all parts thoroughly in distilled water following disinfecting. Allow the parts to dry
completely.
Allow the patient circuit to drip dry. Dry the patient circuit with the therapy device as
described in section 6.2 on page 66.
Hygiene treatment69
Page 70
EN
• Creased hose WM 24130 (transparent) can be washed in water at temperatures
of up to 70 °C. It may not be sterilized.
• Creased tube WM 24120 (gray) can be steam-sterilized with devices to EN 285.
Temperature: 134 °C, minimum retention time 3 minutes. Follow EN 554/
ISO 11134 with regard to validation and monitoring.
6.7.3Patient circuit (valve ventilation)
Patient circuits with patient valve are not suitable for reuse.
Follow the enclosed instructions for use in this regard.
6.7.4Oxygen sensor
The housing of the oxygen sensor is cleaned by wiping with disinfectant. No further
cleaning or hygiene treatment is possible. If the oxygen sensor has been used without a
bacteria filter before a change of patient, it must be replaced.
6.7.5Accessories
To disinfect/sterilize the accessories, see the section entitled “Hygiene treatment” in the
corresponding instructions for use.
6.8 Change in patients
If the device is operated with a bacteria filter, observe the following.
• Change bacteria filter WM 24476
or:
• sterilize bacteria filter WM 24148 and change the particulate filter inside it.
If the device is to be used for another patient without a bacteria filter being used, it must
be subjected to a hygiene treatment beforehand. This must be performed by the
manufacturer, Weinmann, or by an authorized specialist dealer.
The procedure for hygiene treatment is described in the service sheet and in the servicing
and repair instructions for the therapy devices.
70Hygiene treatment
Page 71
EN
7. Function check
7.1 Intervals
Perform a function check of the device monthly. One exception to this is the oxygen sensor.
The oxygen sensor needs calibrating daily.
We recommend checking the battery capacity before each use.
If you discover faults during the function check, you may not use the therapy device again
until the faults have been rectified.
7.2 Method
1. Assemble the therapy device so that it is ready to function.
2. Seal the opening of the patient circuit, e.g. with a sealing plug. For hygiene reasons,
suitable disposable gloves should be worn if you seal the opening of the patient circuit
with your thumb or hand.
3. Switch on the device by pressing the On/Off key . If the device works perfectly, two
different acoustic signals should sound after switching on and both LEDs next to the
alarm acknowledgement key should come on.
4. If TA mode is active, wait approx. 4 minutes until the analysis phase is complete. Then
perform the tests described below.
5. If softstart is switched on, switch it off by pressing the softstart key .
Depending on the operating mode set, now test the following functions:
Mode
Function
Triggering–•–•••––••–••
Flow sensor/
pressure
sensor
Alarms•••••••••••••
Oxygen
supply
TASTST
•••• • • • • • • •••
•••• • • • • • • •••
SX*SXX
*
CPAP PCV aPCV PSV
Function check71
VCV*aVCV
*
SIMV
Page 72
EN
If the values/functions quoted below are not met, send the device to your specialist dealer or
to the manufacturer, Weinmann, for repair.
* These modes are only available with VENTIlogic LS.
7.2.1Triggering
1. Switch off the device by keeping the On/Off key depressed for 2 seconds.
2. Connect the bacteria filter (see “4.6 Connect bacteria filter” on page 42).
3. Connect the patient circuit to the patient/ventilator interface.
4. Switch on the device by pressing the On/Off key .
5. Put on the patient/ventilator interface, e.g. a mask, and breathe normally. Your
respiratory frequency must be above the frequency prescribed by the physician to
prevent the unit’s timer from activating.
The device must react to the change in respiratory phases by changing pressure level.
Note:
If the trigger for exhalation does not work, it may have been switched off. Ask your
doctor whether this is the case.
An exhalation trigger is not available in aPCV and aVCV modes (VENTIlogic LS only).
7.2.2Check flow measurement and flow sensors/pressure sensors
(leakage ventilation)
Note:
A function check of the flow sensors/pressure sensors can only be carried out in
standby mode.
1. Plug the red drying adapter supplied into the device outlet port.
2. For hospital staff only: Ensure that the patient circuit set on the device is the same as
the patient circuit actually in use.
3. Press the menu key to access patient mode.
4. Use the dial to move the black selection bar to
Function check and press ENTER.
5. Confirm the function check in the following window.
The message window Function check running!
opens. The remaining duration of the function check
is displayed.
72Function check
Page 73
EN
6. If the function check is completed successfully, the
message Function check ok! appears.
If the function check is not successful, the message
Device system implausible appears. In this case
refer to section “8.1 Faults” on page 79.
7. Press the menu key to return to the standard display.
7.2.3Check flow measurement and flow sensors/pressure sensors
(leakage ventilation with leakage adapter for standard
tapered connector)
1. Connect the patient circuit to the device.
2. Plug the test adapter supplied onto the patient
connection.
3. Perform the subsequent steps as descibed in
“7.2.2 Check flow measurement and flow sensors/
pressure sensors (leakage ventilation)” from step 2.
7.2.4Check flow measurement and flow sensors/pressure sensors
(valve ventilation)
Function check using a single patient circuit with patient valve
1. Connect the patient circuit to the device.
2. Plug the test adapter supplied onto the patient
connection.
3. The subsequent steps are described in “Check flow
measurement and flow sensors/pressure sensors
(leakage ventilation)” starting from step 2.
Function check73
Page 74
EN
Function check using a double patient circuit with patient valve (VENTIlogic LS
only)
1. Connect the patient circuit to the device.
2. Plug the test adapter supplied onto the patient
connection.
3. The subsequent steps are described in “Check flow
measurement and flow sensors/pressure sensors
(leakage ventilation)” starting from step 2.
7.2.5Alarms
The therapy device performs a self-test on the sensor system when the On/Off key is
pressed which also checks that the alarm system is working. If a fault occurs during the selftest, an error message appears in the main screen (see also “8. Troubleshooting” on
page 79).
1. Check buzzer and LEDs:
Ensure each time you switch on that two different acoustic signals sound one after the
other and that the yellow and red LEDs come on at the same time.
2. Check the No power supply alarm (power supply failure alarm):
Start up the therapy device. Now take the power cord out of the socket. The internal
battery assumes power supply, the two buzzers sound and the yellow LED comes on.
The low-priority alarm window No power supply appears. Plug the power cord into
the power socket again. The alarm should no longer be displayed.
7.2.6Oxygen supply
Caution!
If a compressed gas system is used instead of an oxygen concentrator, an
overpressure valve must be fitted.
Note:
A function check can only be performed on the oxygen valve if the oxygen supply
has been activated by the doctor beforehand.
Proceed as follows to perform a function check of the oxygen valve.
1. Start up the therapy device.
2. For hospital staff only: Ensure that the oxygen supply is activated.
3. Connect the oxygen source to the therapy device and start it up.
74Function check
Page 75
EN
4. At the oxygen source, set the oxygen flow to the value prescribed by the doctor.
5. Select one of the following procedures depending on whether or not you are using an
oxygen sensor.
With an oxygen sensor
Ensure that the oxygen sensor is connected correctly and has been calibrated
beforehand. The oxygen valve is working correctly if an oxygen concentration > 21 %
is shown in the display and the prescribed oxygen flow rate can be set on the oxygen
source.
Without oxygen sensor
The oxygen valve is working correctly if the prescribed oxygen flow rate can be set on
the oxygen source.
7.3 Calibrate oxygen sensor
7.3.1General
If oxygen is supplied during therapy, oxygen concentration is measured at the device outlet
port so as to ensure that the patient is always adequately supplied with oxygen.
To ensure the accuracy of the measurement, calibration should be performed daily.
Calibration is necessary in the case of
– unsettled weather (air pressure, temperature) or
– changes in therapy pressure.
Always perform calibration with the device warmed up (approx. 20 minutes after switching
on).
Perform calibration of the oxygen sensor once a day to avoid the weather affecting
measuring results. The device will issue the relevant message each day to remind you to
perform calibration. This message also appears if:
– the device has previously been disconnected from the power supply
– 24 hours have passed since the last calibration
– the oxygen sensor has been disconnected electrically from the device and then
reconnected.
The sensor is calibrated at a proportion of 21 % oxygen (ambient air). To ensure this, the
oxygen supply is deactivated by the safety valve during calibration.
Function check75
Page 76
EN
7.3.2Precautions
Take the following precautions when calibrating.
1. Shut off the oxygen source.
2. Operate the device for approx. two minutes without an oxygen supply to flush out the
remaining oxygen in the device.
7.3.3Perform calibration
Proceed as follows for calibration.
1. Ensure that the oxygen sensor is connected.
2. Operate the device.
3. Press the menu key . The selection bar is on
Calibrate O
4. Confirm your selection by pressing the dial.
The calibration process is started and remaining
calibration time is displayed.
The device then automatically switches back to the
patient menu.
5. Check the display: the oxygen sensor is working
perfectly if mean oxygen concentration is 21 %.
sensor.
2
7.3.4Shelf life of the oxygen sensor
The shelf life of the oxygen sensor depends on the concentration of the oxygen supplied,
on ambient temperature, on duration of use and on the device settings. Under normal
conditions (ambient temperature 21 °C, 40 % oxygen concentration) the sensor lasts 6
months.
Note:
The sensor becomes exhausted even if it is not connected.
7.3.5Change oxygen sensor
Once the oxygen sensor is exhausted, remove it from the device. Connect a new oxygen
sensor as described in “Measure oxygen concentration (only valve ventilation)” on
page 45.
76Function check
Page 77
EN
Note:
Depending on storage time and temperature, the sensor requires a little time for
measured values to stabilize. As a result, after unpacking from the original
packaging and connecting the sensor, you should wait about 30 minutes before
calibrating the new sensor.
7.4 Energy supply
7.4.1Power supply
Connect the device to the power supply.
The power supply is working perfectly if the green power supply LED comes on
permanently and the standby screen appears in the display.
7.4.2Internal battery and energy failure alarm
1. For the procedure for the function check of the power supply failure alarm, see the
chapter entitled “Alarms” on page 74.
If supply is not assumed by the battery without interruption, either this or the therapy
device is defective. In this case have the device including its internal battery checked by
an authorized specialist dealer or by Weinmann.
2. Check battery capacity (shown in the status line of the display).
If fewer than three segments appear in the battery display, the battery does not have
full capacity. In this case, charge the battery by connecting the device to the power
supply.
7.4.3Operation with VENTIpower
1. Connect VENTIpower to the therapy device. Ensure that VENTIpower has full capacity.
Charge VENTIpower if necessary before continuing the function check.
2. Switch on VENTIpower and connect the therapy device to the power supply.
3. Disconnect the therapy device from the power supply.
The system is working perfectly if the No power supply alarm is displayed, an acoustic
signal sounds and VENTIpower assumes supply. The symbol is displayed in the status
line.
If this is not the case, the following faults may be present:
Function check77
Page 78
EN
– VENTIpower empty: in this case, recharge the VENTIpower and repeat the
function check.
– VENTIpower not connected correctly: in this case, check the connecting cable
and repeat the function check.
– VENTIpower defective, therapy device defective: have the devices repaired by an
authorized specialist dealer or by Weinmann.
78Function check
Page 79
EN
8. Troubleshooting
8.1 Faults
Caution!
If faults occur which cannot be eliminated at once, contact the manufacturer,
Weinmann, or your specialist dealer immediately to have the device repaired. Do
not continue operating the device in order to prevent even greater damage.
Fault/fault messageCause of faultRemedy
Device cannot be
switched on by taking a
breath
Softstart cannot be
switched on
Softstart time cannot be
set to 30 min.
Change filter
Filter change indicator
appears
Battery discharged
Clock not setInternal clock not set
Arrange maintenance
Service indicator
appears
Auto switch-on not
activated
Valve ventilation
connected
The softstart function is
deactivated
Maximum softstart time
limited by doctor
Filter dirty
Internal battery of device
exhausted
Maintenance interval
expired
Activate Auto switch-on (only leakage
ventilation) (5.3, Page 52)
Clarify with your doctor whether the softstart
function can be enabled for your therapy
Softstart is only available in CPAP mode
Clarify with your doctor whether the maximum
softstart time for your therapy can be extended
to 30 minutes
Press alarm acknowledgement key,
clean/change filter as soon as possible (6.4,
Page 68)
Clean/change filter as soon as possible (6.4,
Page 68)
Press alarm acknowledgement key, have battery
replaced by a specialist dealer so that course of
therapy is recorded correctly
Press alarm acknowledgement key, have clock
set by a specialist dealer so that course of
therapy is recorded correctly
The device needs to be checked or serviced by
Weinmann or a specialist dealer as soon as
possible
Troubleshooting79
Page 80
EN
Fault/fault messageCause of faultRemedy
Initial contact with
oxygen sensor
Calibrate O2 sensor
Fan failureFan no longer workingHave device repaired
Humidifier failure
Internal battery not
detected
Internal batt. cannot be
charged
Device system
implausible
Housing fan implausible!
Sensor system
implausible!
Flow measurement
implausible!
Pressure measurement
implausible!
Last calibration
performed over 24
hours ago
Device was switched off
completely beforehand
Humidifier no longer
working
Battery defective
Non-approved battery in
use
Battery operated outside
the permitted
temperature
– Function check not
performed correctly
–Device defective
– Incorrect tube system
installed
Perform calibration of oxygen sensor
Have humidifier replaced or repaired
Have device repaired
Protect device from direct sunlight, do not
operate close to a radiator
Ensure that the device is operated within the
permitted temperature range
Acknowledge message, switch off device
completely (see “Switch device off completely”
on page 61). Perform another function check.
Have settings checked by attending doctor. If the
fault occurs again, have device repaired
8.2 Alarms
A distinction is made between two kinds of alarm.
• Physiological alarms are those alarms which affect the patient's ventilation
directly.
• Technical alarms are those alarms which affect the configuration of the device.
80Troubleshooting
Page 81
EN
The alarms are classified into three priorities:
• low-priority alarms, indicated by the symbol in the alarm window, a
continuously illuminated yellow LED and an acoustic alarm issued (buzzer)
• medium-priority alarms, indicated by the symbol in the alarm window,
a flashing yellow LED and an acoustic alarm issued (buzzer)
• high-priority alarms, indicated by the symbol in the alarm window,
a flashing red LED and an acoustic alarm issued (buzzer)
8.2.1Deactivate alarms
The doctor supervising treatment can decide which physiological alarms to activate or
deactivate. If the symbol appears in the status line, all the physiological alarms have
been deactivated by the supervising doctor (see “Physiological alarms” on page 82).
8.2.2Acknowledging alarms
If a fault triggers an alarm (in this case: disconnection
alarm), you can have the acoustic alarm paused for about
120 seconds by pressing the alarm acknowledgement
key .
The default display appears again after the acoustic
alarm has been acknowledged. The fault which has not
yet been rectified continues to be displayed in the status
line and the alarm LED flashes (or stays on) until the fault
is rectified.
If the fault is not rectified within 120 seconds of the
acknowledgement, the acoustic alarm (buzzer) sounds
again.
You can find troubleshooting assistance in the following tables.
Troubleshooting81
Page 82
EN
8.2.3Sequence of displays if alarms are triggered simultaneously
If several alarms are triggered simultaneously, they are displayed in accordance with the
hierarchy shown below.
1. High-priority alarms
2. Medium-priority alarms
3. Low-priority alarms
If a new alarm with a higher priority occurs after one alarm is displayed, the alarm display
switches to the higher priority. The low-priority alarm is retained and is displayed again
once the high-priority alarm has been eliminated if it still applies.
8.2.4Physiological alarms
DisplayAlarmCause of faultRemedy
Filter dirtyClean/change filter
Adjust headgear/headband so
IPAP
(Leakage)
low
(pressure-controlled
only)
V
T low
Minimum therapy
pressure undershot.
Medium priority
Minimum respiratory
volume undershot.
High priority
Patient/ventilator
interface leaking
Patient/ventilator
interface defective
Settings implausible
Filter dirtyClean/change filter
Patient/ventilator
interface leaking
Patient/ventilator
interface defective
Settings implausible
that the patient/ventilator
interface seals, possibly
replace it
Replace patient/ventilator
interface
Have the settings checked by
the doctor supervising
treatment
Adjust headgear/headband so
that the patient/ventilator
interface seals, possibly
replace it
Replace patient/ventilator
interface
Have the settings checked by
the doctor supervising
treatment
82Troubleshooting
Page 83
EN
DisplayAlarmCause of faultRemedy
VT
high
O
2 high
O
2 low
Frequency
Frequency
Pressure
(VENTIlogic LS only)
*Minimum respiratory
low
*Maximum respiratory
high
high
Maximum tidal
volume exceeded.
Medium priority
Maximum oxygen
concentration
exceeded at device
outlet port.
Medium priority
Minimum oxygen
concentration
undershot at device
outlet port.
Medium priority
frequency undershot.
Low priority
frequency exceeded.
Low priority
Maximum pressure
exceeded. Low
priority, after 10
breaths, rises to
medium priority
Leak in the single patient
circuit(only with patient
circuit with patient
valve)
Patient breathing as well
Oxygen supply too high
as a result of oxygen
flow rate being
incorrectly set
Oxygen sensor
incorrectly calibrated
Oxygen flow rate set too
low
LeakFind and eliminate leak
Oxygen supply
interrupted
Oxygen sensor
incorrectly calibrated
Apnea in spontaneous
breathing mode
Patient hyperventilating
Various possible causes,
e.g. reduction in lung
impedance
Find and eliminate leak,
replace patient circuit if
required
Have settings checked by the
doctor supervising treatment
Check whether the oxygen
flow rate prescribed by the
doctor is set correctly at the
oxygen source. Have settings
checked by the doctor
supervising treatment if
appropriate
Calibrate oxygen sensor
Check whether the oxygen
flow rate prescribed by the
doctor is set correctly at the
oxygen source. Have settings
checked by the doctor
supervising treatment if
appropriate
Check oxygen supply and
connections
Calibrate oxygen sensor
Have settings checked by the
doctor supervising treatment
Calm patient and move
towards a “normal”
respiratory frequency. Call a
doctor
Have settings checked by the
doctor supervising treatment
Troubleshooting83
Page 84
EN
DisplayAlarmCause of faultRemedy
Filter dirtyClean/change filter
Adjust headgear/headband so
Pressure
(VENTIlogic LS only)
Minute volume low*
Minute volume
high*
low
Minimum therapy
pressure undershot.
Low priority
High priority
Medium priority
Patient/ventilator
interface leaking
Patient/ventilator
interface defective
Settings implausible
Minimum minute
volume undershot
Maximum minute
volume exceeded
that the patient/ventilator
interface seals, possibly
replace it
Replace patient/ventilator
interface
Have settings checked by the
doctor supervising treatment
Have the settings checked by
the doctor supervising
treatment
Have the settings checked by
the doctor supervising
treatment
Apnea
Low priorityApnea
Have the settings checked by
the doctor supervising
treatment
* These alarms are only activated 2 minutes after
ventilation starts.
8.2.5Technical alarms
Caution!
If faults occur which cannot be eliminated at once, contact the manufacturer,
Weinmann, or your specialist dealer immediately to have the device repaired. Do
not continue operating the device in order to prevent even greater damage.
DisplayAlarmCause of faultRemedy
Restore power supply and
Battery capacity
critical
Restore power supply and
charge battery. In the event of
a long power failure, connect
the VENTIpower or use an
alternative ventilation option
Have device repaired
Internal battery not
detected
O
measurement
2
defective
O
valve failure
2
Excessive pressure
Low priority
Medium priority
Medium priority
High priority
Device switches off.
Disconnection
High priority
Excessive
temperatureHigh priority
Device switches off.
Excessive
temperature
Medium priority
Battery defective
An unapproved battery
Have device repaired.
is being used
Disconnected,
exhausted or defective
sensor
Check oxygen sensor and
replace if necessary
Safety valve defective,
no oxygen supply
Have device repaired
possible
Pressure sensor defective Have device repaired
Patient circuit is
incorrectly connected to
the device or not
Check tube system
connected at all
Device being operated
with patient/ventilator
interface open (not put
Put on patient/ventilator
interface or switch off device
on)
Device has overheated,
for example as a result
of direct sunlight or
other radiated heat
Device operated outside
permitted temperature
range
Allow device to cool down,
find a more suitable setup
location.
Use alternative ventilation
option
Allow device to cool down,
choose more suitable setup
location. Keep alternative
ventilation option to hand
Troubleshooting85
Page 86
EN
DisplayAlarmCause of faultRemedy
Check power cord is firmly
connected. If necessary, check
the function of the socket by
connecting a different device
(e.g. a lamp).
Connect device to a power
socket and charge internal
battery
Disconnect VENTIpower from
the therapy device and
recharge. Continue operating
therapy device from a power
socket.
Check tube system and
replace patient circuit if
required
Change patient circuit or have
settings corrected by
physician supervising
treatment
Change patient circuit or have
settings corrected by doctor
supervising treatment
Check tube system
Check valve control tube for
damage, replace patient
circuit if necessary
Connect valve control tube
Check tube system
Display gone out
Fault in tube system
Control pressure
high
Control pressure low
Acoustic signal for at
least 120 seconds, no
display.
High priority
Medium priority
High priority
High priority
No power supply and
internal battery is
discharged
The internal battery and
VENTIpower have
discharged
Device defectiveHave device repaired
Exhalation tube not
connected
Incorrect patient circuit
connected
The patient circuit
selected in the menu
does not correspond to
the patient circuit
connected
Device defectiveHave device repaired
The patient circuit
selected in the menu
does not correspond to
the patient circuit
connected
Valve control tube and
pressure measuring tube
the wrong way round
The valve control tube
between the device and
the patient valve is
incorrectly connected
Valve control tube and
pressure measuring tube
the wrong way round
86Troubleshooting
Page 87
EN
DisplayAlarmCause of faultRemedy
Check valve control tube for
damage, replace patient
circuit if necessary
Connect valve control tube
Have settings checked by the
doctor supervising treatment
Check patient circuit and
patient/ventilator interfaces
and change if necessary
Check patient circuit and
patient/ventilator interfaces
and change if necessary
Ensure that the exhaled air
can escape freely at the
patient valve
Check tube system
Disconnect device from power
supply and switch off. Check
connections for tubes. Restore
power connection and
perform function check. If the
fault continues to occur, have
device repaired
Only use converter cable USBRS485 WM 93318 or
converter box WM 93316
Have device repaired
Pressure
permanently low
V
permanently low
T
EPAP
high
Continuous pressure
Control voltage
failure
Device start error
The valve control tube
between the device and
the patient valve is
incorrectly connected
Device defectiveHave device repaired
High priority
Implausible ventilation
settings
Permanent large leak
Filter dirtyClean/change filter
High priority
Permanent large leak
Device defectiveHave device repaired
The pressure drop on
exhalation is inadequate
High priority
(possible cause: pressure
drop at patient valve too
slow)
High priorityDevice defective
RS485 incorrectly
High priority
connected
Internal faultHave device repaired
High priority
High priority
Device switches off.
Parameters could not be
loaded
Blower does not switch
back on after power
failure
Troubleshooting87
Page 88
EN
DisplayAlarmCause of faultRemedy
IPC failure
Change deviceHigh priority
Device switches off.
Sensor system failure
Change deviceHigh priority
System monitoring
failure
No power supply
Blower failure
Change deviceHigh priority
CPU failure
Change deviceHigh priority
Device switches off.
Low priority
Low priorityPower supply failed
Device switches off.
Device switches off.
Device defectiveHave device repaired
Sensor system defective Have device repaired
Voltage monitoring
failure
Device defectiveHave device repaired
Internal processing
faulty, device defective
Have device repaired
Keep alternative ventilation
option to hand. Check power
cord is firmly connected. If
necessary, check the function
of the socket by connecting a
different device (e.g. a lamp)
Have device repaired
I2C failure
VENTIpower failure
Warning: Device
switched off!
Unknown device
88Troubleshooting
Medium priorityDevice defectiveHave device repaired
Low priority
High priority
High priorityDevice defectiveHave device repaired
VENTIpower discharged
VENTIpower defectiveHave VENTIpower repaired
Device has been
switched off
Restore connection to power
supply. Charge VENTIpower
Turn the device on again
Acknowledge alarm
Page 89
EN
8.2.6Storage of alarms
Once the alarm threshold is reached, all the alarm types listed in tables “8.2.4Physiological
alarms” and “8.2.5Technical alarms” are recorded in an alarm list with date, time and
duration. Up to 200 alarms can be stored. After that, the oldest alarm in each case is
overwritten. The alarm list can be called up by the doctor supervising treatment or by you
(see “5.4 Alarm list” on page 53).
The alarm list is retained even if the entire power supply (power supply and internal battery)
fails. In this case, the data can be called up for up to two years. The alarm list is deleted
after two years or when maintenance is carried out.
In any event, alarms are retained in the event of a power failure.
8.2.7Passing on of alarms via the remote alarm connection
All high and medium-priority alarms are switched to a remote alarm connection in parallel.
The only low-priority alarm which is passed on to the connection is the No power supply
alarm.
If the device is operated in a hospital, the therapy device can be connected to the hospital’s
nurse call system via the remote alarm connection.
For home ventilation, the therapy device can also be connected to the VENTIremote alarm
box via the remote alarm connection. The remote alarm box remotely transmits and
amplifies the acoustic and visual alarm signals output by the therapy device. It is of
particular assistance to nursing staff and to relatives when looking after the patient in the
home environment.
Troubleshooting89
Page 90
EN
9. Maintenance and safety checks
9.1 Intervals
We recommend having servicing, safety checks and repair work carried out only by
the manufacturer, Weinmann, or by a specialized dealer expressly authorized by
the manufacturer.
Check both filters regularly for dirt.
• The coarse dust filter and the fan filer should be cleaned once a week and
changed no later than every 6 months.
• The fine filter must be changed after no more than 1000 operating hours (filter
change symbol appears in the display).
For hygiene reasons, we recommend replacing the following parts at the intervals
specified.
9.1.1Leakage ventilation
• Pressure measuring tube every 6 months - earlier if dirty
• Clean the patient/ventilator interface in accordance with the relevant
instructions for use
• Clean the exhalation system in accordance with the relevant instructions for use
• If necessary, other accessories in accordance with the associated operating
instructions
9.1.2Valve ventilation
• The patient circuits with patient valve should be replaced in accordance with the
relevant instructions for use
• Clean the patient/ventilator interface in accordance with the associated
instructions for use
• If necessary, other accessories in accordance with the associated instructions for
use
90Maintenance and safety checks
Page 91
EN
9.2 Battery
9.2.1Care of the internal battery
Fully charge the battery before carrying out the following steps. Leave the device
connected to the power supply for at least 15 hours.
We recommend discharging the battery once a month by operating the device until the
alarm Battery capacity critical appears (approx. 25 % capacity remaining). Then
recharge the battery until full by operating the device from the power supply. This boosts
the shelf life and the service life of the battery.
If the battery lasts less than 45 minutes if having previously been at full capacity, have the
battery replaced by the manufacturer, Weinmann, or by specialist staff expressly
authorized by the manufacturer.
We recommend that the device remain connected to the power supply at all times, except
for the monthly discharge procedure. This ensures that the internal battery is always fully
charged and ready to use.
9.2.2Storage
If the device is to be operational at all times, we recommend leaving the device connected
to the power supply even if it is not used for prolonged periods. This ensures that the
battery is always fully charged.
9.3 Change filter
9.3.1Change coarse dust filter
Use original filters from Weinmann only. Using third-party filters invalidates any claim
under warranty and may result in restricted function and bioincompatibility.
1. If the VENTIclick humidifier is connected, first disconnect it from the device. This will
avoid water running into the device when the filter is changed. Follow the instructions
for use for the VENTIclick.
2. Press on the latch of the filter compartment lid and
take it off.
Maintenance and safety checks91
Page 92
EN
3. Take the coarse dust filter out of the filter
Filter change indicator
compartment
lid and dispose of it in normal domestic waste.
4. Place the clean coarse dust filter in the filter
compartment lid.
5. Put the filter compartment lid into the cutout in the
housing bottom edge first. Then push the filter
compartment lid into the housing until the latch
engages.
9.3.2Change fine filter
The fine filter needs changing when it has gone dark, but in any event after no more than
1000 operating hours. In the latter case, the message Change filter will appear in the
display.
Acknowledge the message by pressing the alarm
acknowledgement key . The filter change symbol
then appears permanently in the status line.
Proceed as follows to change the fine filter:
1. Press on the latch of the filter compartment lid and
take it off.
2. Replace the fine filter with a new fine filter
WM 15026. Dispose of the old filter in normal
domestic waste.
3. Put the filter compartment lid into the cutout in the
housing bottom edge first. Then push the filter
compartment lid into the housing until the latch
engages.
9.3.3Reset filter change indicator
After you have changed the fine filter, you must reset the filter change indicator. This is
necessary even if the filter was changed before 1000 operating hours expired so the filter
change symbol was not in the display.
92Maintenance and safety checks
Page 93
EN
1. To reset the filter change indicator, press the menu
key and use the the dial to select Filter change
from the patient menu. Press the dial to call up the
Filterchange menu.
2. The question Reset filter change? appears. Select
YES with the dial and confirm the selection by pressing
the dial.
If you want to cancel the process, select NO with the
dial and press the dial. The process is cancelled.
If you select and confirm YES with the dial, the
message Filter change reset!appears for approx. 3
seconds.
9.3.4Change fan filter
Proceed as follows to change the fan filter.
1. Turn the device over so that the underneath points upwards.
2. Grasp the fan filter with your fingertips and pull it out of the opening.
3. Take a new clean filter and put it in. Ensure that the
fan filter is straight and that the corners are not tilted
by the opening.
4. Turn the device back over so that the top points
upwards.
5. Dispose of the used filter with domestic waste.
Maintenance and safety checks93
Page 94
EN
9.3.5Bacteria filter
Adapter
Pressuremeasurement
tube
Sleeve
When using bacteria filter WM 24148, change the particulate filter in the bacteria filter in
accordance with the associated instructions for use.
When using bacteria filter WM 24476, change the bacteria filter in accordance with the
associated instructions for use.
1. Release the sleeve of the creased tube from the
adapter.
2. Pull the pressure-measurement tube out of the
creased tube.
3. Pull the pressure-measurement tube off the adapter.
4. Push the new pressure-measurement tube onto the
adapter.
5. Hold up the creased tube and guide in the free end of
the new pressure measuring tube.
6. Push the sleeve of the creased tube onto the adapter.
7. Dispose of the old pressure-measurement tube.
9.5 Safety check
The legally-specified interval for performance of the safety check (Sicherheitstechnische
Kontrolle - STK) in accordance with §6 of the German law governing medical devices
[Medizinprodukte-Betreiberverordnung - Germany only] is 2 years.
In addition, servicing must be carried out as a preventive maintenance measure at the
following intervals:
• after every 8000 operating hours (servicing symbol
94Maintenance and safety checks
appears in the display)
Page 95
EN
or
• after 2 years (see service label on the rear of the device)
depending on which comes first.
The safety check and servicing include:
–check for completeness
–visual inspection for mechanical damage
–filter change
–cleaning the device
–replacement of any defective parts
–complete check of device functions and pressure displays
–battery change
–change of internal battery
–final check in accordance with test instruction WM 27104
–testing the power supply failure alarm
9.6 Disposal
9.6.1Device
Do not dispose of the device with domestic waste. To
dispose of the device properly, contact a licensed,
certified electronic scrap disposal merchant. This address
is available from your Environment Officer or from your
local authority.
9.6.2Internal battery
Do not dispose of the internal battery with domestic
waste. To dispose of the battery properly, contact a
licensed, certified electronic scrap disposal merchant.
This address is available from your Environment Officer or
from your local authority.
9.6.3Oxygen sensor
Do not dispose of the oxygen sensor with domestic
waste, as it contains electrolyte and lead. It should be
disposed of in accordance with applicable national or
regional regulations.
You can order accessories and spare parts separately if required. A current list of
accessories and spare parts can be ordered on the Internet at www.weinmann.de or via
your dealer.
100Scope of supply
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.