Velleman DVM322DI User Manual

DVM322DI -- SOUND LEVEL METER
1. Introduction and Safety Information
Thank you for buying the DVM322DI ! We advise you to read this manual thoroughly. You should only use this device for the applications described in this manual. Unauthorised applications may cause damage to the device.
Environmental Conditions :
- Max. relative humidity : 90%
- Operation temperature : 0 - 40°C
Maintenance
- All repairs and maintenance activities not described in this manual should be handled by qualified personnel.
- Wipe the housing clean from time to time with a dry cloth. Do not use abrasives or solvents on this device.
Safety Symbols
This device is EMC-compatible.
2. Features
This device was designed according to IEC651 type 2, ANSI S1.4 type 2 for sound level meters. The DVM322DI meets the requirements of safety engineers and industrial safety offices for the monitoring of sound quality in various environments.
- RS-232 interface with PC.
- data logging capacity with memory for 32000 measurements.
3. Specifications
Standards : IEC type 2, ANSI S1.4 type 2 Accuracy : ± 1.5dB (under conditions @ 94dB, 1KHz)
Frequency Range : 31.5Hz-8KHz Measuring Level : 30-130dB Frequency Weighting : A-C Microphone : ½" electret condenser microphone Digital Display : 4-digit LCD, resolution 0.1dB, display period 0.5sec. Analogue Display : 50-segment bargraph, resolution 0.1dB, display period 50mS Time Weighting : FAST = 125mS or SLOW = 1sec. MAX/MIN Hold : To hold the highest and lowest values obtained during measurements Level Ranges : a total of 4 ranges at step 10dB : 30-80dB, 50-100dB, 80-130dB, 30-130dB Dynamic Range : 100dB Alarm Function : OVER indicator (value is higher than the upper limit of the measuring range) and
UNDER indicator (value is lower than the lower limit of the range).
AC Output : 1Vrms at full scale
DVM322DI GB
1
Output impedance : 600 FS (full scale= the upper limit of each range). DC Output : 10mV/dB
Output impedance : ± 1K Power Supply : 1 x 9V-battery
Autonomy : ± 50hrs of continuous operation with an alkaline battery DC adapter : Voltage 9VDC Supply Current 30mA DC Socket : pin = ground, housing = +, external Ø = 3.5mm, internal Ø = 1.35mm Electromagnetic Compatibility : RF field = 3V/m, total accuracy = specified accuracy + 0.5dB Operating Temperature : 0 to 40°C / 32 to 104°F Operating Humidity : 10 to 90% Storage Temperature : -10 to +60°C / 14 to 140°F Storage Humidity : 10 to 75% Memory Size : 32000 measurements Dimensions : 10.8 (L) x 2.5 (W) x 1.2 (H) inch // 275 (L) x 64 (W) x 30 (H) mm Weight : Accessories : 9V-battery, carrying case, screwdriver, user manual, windscreen, 3.5mm plug, RS-
285g (including battery)
±
232 cable, software for Windows.
4. Calibration
Use a standard sound level calibrator (OPTION DVM1356).
1. Use the following settings : Display : dBA Time Weighting : FAST Measurement Mode : MAX MIN function deactivated Level Range : 50 to 100dB
2. Connect the microphone with the calibrator through the microphone connection of the DVM322DI.
3. Activate the calibrator and adjust the CAL potentiometer. The display will indicate the desired level range. This device has been calibrated by the manufacturer. Maximum accuracy can be expected during a one-year period after calibration. Consequently, the device should be recalibrated every twelve months.
acoustic calibrator
screwdriver
5. Preparing for Measurements
a) Inserting the battery : Remove the battery cover at the back of the device and insert one 9V-battery.
Observe the polarity indications inside the battery compartment !
b) Replacing the battery : The c) Connecting an DC adapter : Connect the OPTIONAL adapter (PSU03R) with the DC 9V connector of the
DVM322DI's side panel.
DVM322DI GB
-symbol will flash on the display when the battery needs to be replaced.
2
6. Operating Precautions
a) Wind blowing around the microphone causes additional extraneous noise. Mount the windscreen when
performing measurements under windy circumstances.
b) Recalibrate the instrument prior to use if it has not been used for some time or if it was last used in a difficult
environment.
c) Do not store or operate the instrument under excessive temperatures or humidity for a prolonged period of time. d) Keep the microphone dry and avoid severe vibrations. e) Remove the battery if the device is not to be used for several days.
7. Description of the Front, Side and Back Panels
1. Windscreen
Use the windscreen when the wind-force exceeds
10m/sec.
2. Display
See 8. Description of the LCD Display” below.
3. Power & Backlight Button
Use the
or to turn the backlight ON or OFF. Press the button once to turn the sound level meter ON and press it again to turn on the backlight. Hold the button for three seconds to deactivate the device.
4. MAX/MIN Hold Button
Press the MAX/MIN button to activate the MAX. & MIN. recording mode. Select the proper level range
beforehand to ensure that the readings will not fall outside the measurement range. Press once to select MAX value. Press again to select MIN value. Press a third time to select the current value (MAX/MIN annunciator is blinking). Hold the MAX/MIN button for two seconds to exit the MAX/MIN mode. Note that the MAX/MIN mode will clear every time a different level range is selected or the operator alternates between A and C weighting.
5. REC Button
Press the button to start recording. “REC” appears on the display. Press the button again to stop recording
6. Frequency Weighting Selection Button
A : A-weighting, for general sound level measurements. C : C-weighting, to check the low-frequency content of the monitored noise. There is a large amount of low-
frequency noise if the value for the C-weighted level is much higher than that for the A-level.
7. Time Weighting Selection Button
FAST : for normal measurements SLOW : for checking the average level of fluctuating noise.
8. Level Range Control Button
Press this button to select the various levels in this order : Lo → Med → Hi → Auto → Lo → …
9. Microphone
½” electret condenser microphone
10. CAL potentiometer
Calibration potentiometer used to calibrate the device.
11. RS-232 Interface
The RS-232 signal output is a 9600bps N 8 1 serial interface.
12. Signal Output Terminal
AC : 1 Vrms per range step. Output impedance = 100. The output signal is transmitted via a standard
DC : Output : 10mV/dB. Output impedance = 1KΩ. The output signal is transmitted via a standard 3.5mm
-key to switch the device ON or OFF
3.5mm coaxial socket. The signal is located on the pin
coaxial socket. The signal is located in the middle.
DVM322DI GB
3
Loading...
+ 6 hidden pages