tekmar 083 User Manual

- Data Brochure
D 083
Duct Sensor 083
The tekmar Duct Sensor 083 is designed to measure the average temperature of the air stream in a heating, cooling, or ventilation duct. The 083 has a 6” (150 mm) long probe connected to a 2” x 4” electrical box for easy wiring.
The Duct Sensor 083 probe should be located in the middle of the air stream in order to measure the average temperature of the duct. Drill a 3/8” (10 mm) hole in the duct and place the probe through the hole.
drill 3/8” (10 mm) diameter hole in duct
05/07
18 AWG or similar wire
Ensure that the foam ring on the electrical box around the probe fits tight against the duct.
Foam ring
1 of 4 © 2007 D 083 - 05/07
A 4” (102 mm) flange mounted on the electrical box has two holes to allow sheet metal screws (not included) to fasten the sensor to the duct.
Wiring the Sensor
Open the electrical box by unscrewing the top screw. Connect 18 AWG or similar wire to the two terminals provided in the enclosure and run
the wires from the 083 to the control. Do not run the wires parallel to telephone or power cables. If the sensor wires are located in an area with strong sources of electromagnetic interference (EMI), shielded cable or twisted pair should be used or the wires can be run in a grounded metal conduit. If using shielded cable, the shield wire should be connected to the Com Sen terminal on the control and not to earth ground.
Follow the sensor testing instruction in this brochure and connect the wires to the control. Close the electrical box by tightening the top screw.
Testing the Sensor
A good quality test meter capable of measuring up to 5,000 kΩ (1 kΩ = 1000 Ω) is required to measure the sensor resistance. In addition to this, the actual temperature must be measured with either a good quality digital thermometer, or if a thermometer is not available, a second sensor can be placed alongside the one to be tested and the readings compared.
First measure the temperature using the thermometer and then measure the resistance of the sensor at the control. The wires from the sensor must not be connected to the control while the test is performed. Using the chart on the following page, estimate the temperature measured by the sensor. The sensor and thermometer readings should be close. If the test meter reads a very high resistance, there may be a broken wire, a poor wiring connection or a defective sensor. If the resistance is very low, the wiring may be shorted, there may be moisture in the sensor or the sensor may be defective. To test for a defective sensor, measure the resistance directly at the sensor location.
Do not apply voltage to a sensor at any time as damage to the sensor may result.
© 2007 D 083 - 05/07 2 of 4
Loading...
+ 2 hidden pages