Stryker FL28EX Service manual

5 (1)

MAINTENANCE MANUAL

ELECTRIC MED/SURG BED

Model FL28EX

TECHNICAL ASSISTANCE AND PARTS

Canada: 1 888 233-6888

United States: 1 800 327 0770

Outside Canada and the United States: Contact your local representative

Manufactured by Stryker

72-0905E REV A MM FL28EX

December 2005

F15-44-B

Printed in Canada

 

TABLE OF CONTENTS

 

1. INTRODUCTION...................................................................................................................

5

1.1

Specifications .................................................................................................................

5

1.2

Technical Support...........................................................................................................

6

1.3

Warning, Caution, Note Definition...................................................................................

6

1.4

Static Discharge Precautions..........................................................................................

6

1.5

Warranty.........................................................................................................................

7

 

Limited Warranty ..........................................................................................................

7

 

To Obtain Service and/or Parts ....................................................................................

7

 

Return Authorization.....................................................................................................

8

 

Damaged Merchandise ................................................................................................

8

1.6

Symbols..........................................................................................................................

8

2. PREVENTATIVE MAINTENANCE ........................................................................................

9

2.1

Bed Cleaning and Mattress Care ....................................................................................

9

2.2

Lubrication Requirements .............................................................................................

10

 

Actuator Screw Lubrication.........................................................................................

10

2.3

Preventative Maintenance Program..............................................................................

12

 

Annual Checklist.........................................................................................................

12

 

Recommended Spare Parts .......................................................................................

13

3. TROUBLESHOOTING ........................................................................................................

15

3.1

Troubleshooting Guide..................................................................................................

15

4. MAINTENANCE PROCEDURES ........................................................................................

18

4.1

Siderail Maintenance Procedures .................................................................................

18

 

Head Siderail Assembly Replacement........................................................................

18

 

Head Rail Replacement..............................................................................................

19

 

Head Siderail Mechanism Replacement.....................................................................

19

 

Foot Siderail Assembly Replacement .........................................................................

19

 

Foot Rail Replacement ...............................................................................................

20

 

Foot Siderail Mechanism Replacement ......................................................................

20

 

Siderail Control Fascia Replacement..........................................................................

21

 

Siderail Outer Control Board Replacement.................................................................

22

 

Siderail Inner Control Board Replacement..................................................................

23

 

Speakerphone Replacement (Optional)......................................................................

24

4.2

Foot Board Maintenance Procedures............................................................................

25

 

Foot End Control Fascia Replacement .......................................................................

25

 

Weigh System LCD Display Replacement..................................................................

25

 

Foot End Control Board Replacement ........................................................................

26

4.3

Mattress Support Maintenance Procedures ..................................................................

27

 

Foot Section Replacement .........................................................................................

27

 

Thigh Section Replacement........................................................................................

28

 

Seat Section Replacement .........................................................................................

28

 

Head Section Replacement........................................................................................

29

 

Fowler Actuator Replacement ....................................................................................

30

 

Knee Gatch Actuator Replacement ............................................................................

32

Hi-Lo Actuator Replacement .......................................................................................

33

Motor Control Board Replacement..............................................................................

35

Nurse Call (Optional)/GEN lll Pendant Control (Optional) Control Board Replacement

...................................................................................................................................

36

Power Connector Replacement ..................................................................................

37

Power Connector Fuse Replacement..........................................................................

37

Scale Control Board/Bed Exit Buzzer (Optional) Replacement....................................

38

Toroidal Transformer Replacement (100/200/220/240V International Series Beds -) ..

39

Load Cell Replacement...............................................................................................

40

Scale Calibration.........................................................................................................

41

Verifying the Scale Accuracy ....................................................................................

41

Scale Calibration ......................................................................................................

42

Cardiac Chair and Auto Contour Micro Switch Replacement ......................................

44

Cardiac Chair Micro Switch.......................................................................................

44

Micro Switch Common to the Cardiac Chair and Auto Contour Mechanisms ............

45

Auto Contour Micro Switch .......................................................................................

45

CPR Micro Switch Replacement .................................................................................

46

Micro Switch Attached to the Head Section Lever ....................................................

46

Micro Switch Attached to the Frame. ........................................................................

47

CPR Activation Cable Replacement............................................................................

47

CPR Pneumatic Cylinder ............................................................................................

47

4.4 Base Maintenance Procedures......................................................................................

49

Brake/Steer Pedal Replacement .................................................................................

49

Steer Wheel Caster Replacement...............................................................................

50

Steer Wheel Assembly Replacement..........................................................................

50

Steer Wheel Swing Arm Assembly Replacement........................................................

51

Bed Caster Replacement ............................................................................................

52

Brake Rod Replacement.............................................................................................

53

Appendix A: Circuit Diagrams ...................................................................................................

54

Appendix B: Maintenance Menu: Signification of Error Codes and Digital Inputs ......................

59

Introduction Chapter 1

1. INTRODUCTION

This manual is designed to assist in the servicing of the Stryker's FL28EX Med/Surg Beds. Read it thoroughly before beginning any service on the bed. Qualified maintenance personnel should be able to refer to this manual at all time when servicing the bed.

This Maintenance Manual is an integral part of the bed and should be included if the bed is sold or transferred.

1.1 SPECIFICATIONS *

Safe Working Load**

500 lb (227 kg)

Scale System

 

- Capacity

Patients weighing up to 500 lb (227 kg)

- Accuracy

± 2 % for weight from 100 to 500 lb (45.3 to 227 kg)

 

± 2 lb for weight under 100 lb (45.3 kg)

- Operating Angular Range

-12° to +12°

 

 

Overall Length/Width

 

- Siderails Up

94 5/8 x 40" (240.34 cm x 101.6 cm)

- Siderails Down

94 5/8 x 39" (240.34 cm x 99.06 cm)

 

 

Weight w/Boards

472 lb (214.1 kg)

 

 

Patient Sleep Surface

35 x 80" (89 x 203 cm) adjustable to 82" (208 cm) and

 

84" (213 cm)

 

 

Recommended Mattress Size

35 x 80" (89 x 203 cm); 35 x 82" (89 x 208 cm); 35 x

Mattress Maximum Thickness

84" (89 x 213 cm)

6" (15.24 cm)

 

 

Min/Max Bed Height

14 1/2 to 29" (36.8 to 73.7 cm)

 

 

Fowler Angle

0 to 61°

 

 

Knee Gatch Angle

0 to 24°

- with Auto Contour

- without Auto Contour

0 to 32°

 

 

Trendelenburg/Reverse Trendelenburg

+14 to -14°

 

 

Environmental Conditions

 

- Transport and Storage

 

- Ambient Temperature

-40 to 70°C (-40 to 158°F)

- Relative Humidity

10 to 100%

- Atmospheric Pressure

500 to 1060 hPa

- Operating ***

 

- Ambient Temperature

18.3 to 26.7°C (65 to 80°F)

- Relative Humidity

20 to 80% without condensation

- Atmospheric Pressure

700 to 1060 hPa

 

 

Electrical Requirements**** - all electrical

100V , 50-60Hz, 7.5A - Two 250V, 10A Fuses

requirements meet CSA C22.2 No. 601.1,

120V , 50-60Hz, 4.0A (9.8A w/120V Optional Auxiliary

UL 60601-1 and IEC 60601-1, 60601-2-38

Outlet) - Two 250V, 10A Fuses

specifications.

200V , 50-60Hz, 3.2A -Two 250V, 6.3A Fuses

 

220V , 50-60Hz, 2.9A -Two 250V, 6.3A Fuses

 

240V , 50-60Hz, 2.7A -Two 250V, 6.3A Fuses

 

 

*Stryker pays special attention to product improvement and reserves the right to change specifications without notice.

**The Safe Working Load specified is the sum of the mattress and accessory weight (100 lb/45.4 kg) and the patient's weight.

***Operating environment recommended to ensure the scale system precision.

****The device has a 10% duty cycle.

5

BED
GROUNDING DIAGRAM

Maintenance Manual

1.2 TECHNICAL SUPPORT

For questions regarding this product, contact the following Technical Service department or your local representative:

Stryker Canada

Stryker Medical

1 888 233-6888

1 800 327-0770

45, Innovation Drive

3800, East Centre Avenue

Hamilton, Ontario, L9H 7L8

Portage, MI 49002

Canada

USA

1.3 WARNING, CAUTION, NOTE DEFINITION

The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.

WARNING

The personal safety of the patient or user may be involved. Disregarding this information could result in injury to the patient or user.

CAUTION

These instructions point out special procedures or precautions that must be followed to avoid damaging the equipment.

NOTE

Notes provide special information to make maintenance easier or important instruction clearer.

1.4 STATIC DISCHARGE PRECAUTIONS

The electronic circuits of the bed are protected from static electricity damage only while the bed is assembled. It is extremely important that all service personnel always use adequate static protection when servicing the electronic components of the bed.

Static Protection Equipment

The necessary equipment for a proper static protection is:

1 static wrist strap

1 grounding plug

1 test lead with a banana plug on one end and an alligator clip on the other.

Static Protection Procedure

1. Unplug the bed power cord from the wall outlet.

2. Insert the grounding plug into a

properly grounded hospital grade wall receptacle. Plug the banana plug of the test lead into the receptacle on the grounding plug. Connect the

alligator clip on the other end of the test lead to a ground point on the bed.

3.Place the static control wrist strap on your wrist. Connect the clip at the other end of the wrist strap cord to a ground point on the bed.

6

Introduction Chapter 1

1.5 WARRANTY

LIMITED WARRANTY

All Stryker products are guaranteed against material or manufacturing defects, improper operation of mechanisms, and premature wear of bed components under normal use conditions.

For questions regarding warranty, please contact the Technical Service department (see section 1.2) or your local representative.

TO OBTAIN SERVICE AND/OR PARTS

To Require Service

For an on-site diagnosis of a malfunction by one of our Field Service Representative, contact the Technical Service department (see section 1.2) or your local representative.

To Order Parts

Contact the Technical Service department (see section 1.2) or your local representative and provide the following information that can be found on the serial number plate and the manufacturer's nameplate affixed respectively to the right side of the frame at the foot end of the bed, and on the right side of the head end casing:

HEAD END

B

C

D

FOOT END

MANUFACTURER NAMEPLATE

SERIAL NUMBER PLATE

A

Figure 1.5

From the serial plate, write down the serial number (A).

From the manufacturer's nameplate, write down the bed model (B), the production number (C), ex. FL28-XXXX, and the Customer's Guide number (D).

Consult the parts lists and the drawings contained in the Customer Guide third section, “Parts Lists”, to identify the defective part. Write down the information.

Write down a description of the problem encountered while using the equipment.

NOTE

It is very important that you refer to the parts lists and drawings of the Parts Lists manual specific to the bed needing repair.

The Technical Service representative can help you identify the parts to be replaced. However, if an error occurs when ordering, the user remains responsible for the parts ordered.

Stryker will take back wrong parts ordered but will not assume shipping charges, and restocking fees will be charged to the user unless a Field Service Representative has been requested for an on-site diagnosis of the malfunction.

7

Maintenance Manual

RETURN AUTHORIZATION

Merchandise cannot be returned without approval from the Technical Service department. An authorization number will be provided, which must be clearly printed on the returned merchandise. Stryker reserves the right to charge shipping and restocking fees on returned items.

DAMAGED MERCHANDISE

Claims for damaged merchandise must be made with the carrier within fifteen (15) days of receipt of merchandise. DO NOT ACCEPT DAMAGED SHIPMENTS UNLESS SUCH DAMAGE IS NOTED ON THE DELIVERY RECEIPT AT THE TIME OF RECEIPT. Upon prompt notification, Stryker will file a freight claim with the appropriate carrier for damages incurred. Claims will be limited in amount to the actual replacement cost. In the event that this information is not received by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on the delivery notice at the time of receipt, the customer will be responsible for payment of the original invoice in full.

Claims for any short shipment must be made within five (5) days of invoice.

1.6 SYMBOLS

Warning, refer to accompanying documents

10A 250V

6.3A 250V

Fuse rating for 100V and 120V electric systems

Fuse rating for 200V , 220V and 240V electric systems

Protective Earth (ground)

Alternating Current

Type B Equipment

IPX4 Protection from liquid splash

8

Preventative Maintenance Chapter 2

2. PREVENTATIVE MAINTENANCE

2.1 BED CLEANING AND MATTRESS CARE

WARNING

When large fluid spills occur in the area of the circuit boards, cables and motors, immediately unplug the bed power cord from the wall outlet, remove the patient from the bed and clean up the fluid. Have maintenance completely check the bed. Fluids can have an adverse effect on operational capabilities of any electrical product. DO NOT put the bed back into service until it is completely dried and has been thoroughly tested for safe operation. Ensure, among other things, that the plastic components being used as covers for the siderail mechanism arms and the foot end casing are removed and that the parts they cover are thoroughly dried.

CAUTION

Do not use harsh cleaners, solvents or detergents. Equipment damage could occur. Do not steam clean, hose off or ultrasonically clean the bed. Do not immerse any part of the bed. The bed electrical parts may be damaged by exposure to water.

Germicidal disinfectant, used as directed, and/or Chlorine Bleach products are not considered mild detergents. These products are corrosive in nature and may cause damage to your bed if used improperly. If these types of products are used, ensure the beds are wiped with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the beds will leave a corrosive residue on the surface of the bed, possibly causing premature corrosion of critical components. Failure to follow the above directions when using these types of cleaners may void this product warranty.

CLEANING BEDS

Hand wash all surfaces of the bed with a soft cloth moistened with a solution of lukewarm water and a mild detergent.

Wipe the bed clean and dry thoroughly to avoid build up of cleaning solution.

MATTRESS CARE

WARNING

Inspect the mattress after each use. Discontinue use if any cracks or rips, which may allow fluid to enter the mattress, are found in the mattress cover. Failure to properly clean the mattress, or dispose of it if defective, may increase the risk of exposure to pathogenic substances and may bring about diseases to the patient and user.

Inspection

Implement local policies to address regular care, maintenance, and cleaning of mattresses and covers. The cover cleaning procedure can be found below and on the bed label.

Inspect the mattress cover inner and outer surfaces and the zip fasteners regularly for signs of damage. If the mattress cover is heavily stained or soiled, or is torn, remove the mattress from service.

Cleaning

Stains: Wash with lukewarm water using a mild detergent. Rinse with water and let dry. For tough stains, use bleach diluted with ten parts of water.

9

Maintenance Manual

2.2 LUBRICATION REQUIREMENTS

The only components of the bed needing periodic verification and lubrication are the four actuator screws and the clevis pin holding the head actuator to the head section lever. They should be checked every year and lubricated every two years.

CAUTION

The bed uses oil-impregnated shoulder spacers at hinge points. Do not lubricate these shoulder spacers. When shoulder spacers are found worn, replace them.

ACTUATOR SCREW LUBRICATION

Required Tools:

No. 2 Phillips Screwdriver

1/4" Ratchet (w/6" extension) and 5/16" socket

OG2 Grease

Brush

Bungee Cord

Procedure:

A

B

B

Figure 2.2

Clevis Pin

1.Raise the bed fully up and apply the brakes.

2.Remove if applicable the head section plastic cover (optional).

3.Fully raise the head section.

4.Remove the head board.

5.Unplug the bed power cord from the wall outlet.

6.Apply grease all over the clevis pin.

10

Preventative Maintenance Chapter 2

7. Fully raise and lower the head section several times to spread the grease.

Head and Thigh Actuators

8.Raise the bed fully up and apply the brakes.

9.Remove if applicable the head and foot section plastic covers (optional).

10.Fully raise the head section and flatten the thigh section. Lower the four siderails.

11.Unplug the bed power cord from the wall outlet.

12.Lift and fold back the foot section toward the head end of the bed. Secure the foot section to the bed using a bungee cord.

13.Using a no. 2 Phillips screwdriver, remove the four screws (A) holding the cover plate to the frame and remove the plate. Remove the optional night light if present.

14.Using a 1/4" ratchet (w/6" extension) and a 5/16" socket, remove the two screws (B) holding the dust tube of each actuator. Push the tube back to uncover the screw threads.

15.Using a brush, apply grease on the screw threads. Make sure the grease reaches the bottom of the threads.

16.Replace the dust tubes.

17.Replace the cover plate.

18.Lower the foot section.

19.Plug in the bed power cord and, using the electric controls, raise and lower several times the Fowler and Knee Gatch to spread the grease evenly.

Hi-Lo Actuators

1.Lower the bed completely and apply the brakes. Flatten the mattress support.

2.Lower the four siderails.

3.Remove if applicable the head and foot section plastic covers (optional), to reach the Hi-Lo actuators. Otherwise, fully raise the head section, and lift and fold back the foot section toward the head end of the bed. Secure the foot section to the bed using a bungee cord.

4.Unplug the bed power cord from the wall outlet

5.Using a brush, apply grease on the threads through the dust tube side openings of both HiLo actuators. Make sure the grease reaches the bottom of the threads.

6.Lower the foot section.

7.Plug in the bed power cord and, using the electric controls, raise and lower the bed several times to spread the grease evenly.

11

Maintenance Manual

2.3 PREVENTATIVE MAINTENANCE PROGRAM

WARNING

When servicing use only identical replacement parts provided by Stryker.

ANNUAL CHECKLIST

All fasteners secure.

Inspect for excessive wear the oil-impregnated bronze shoulder spacers found at the bed hinge points. Do not lubricate these spacers. Replace as needed.

Check the grease present on the components detailed in section 2.2, lubricate if needed (see section 2.2). Lubricate them at least every two years.

On both sides of the bed, depress fully down the side of the pedal identified with a red sticker and ensure that the brakes are applied and the bed is immobilized. Toggle the pedal to neutral and ensure the brakes are released.

On both sides of the bed, depress fully down the side of the pedal identified with a green sticker and ensure that the steer wheel is engaged. Toggle the pedal to neutral and ensure that the steer wheel disengages.

Siderails move, latch and stow properly.

All controls of the foot end panel working properly, including LED's.

Calibrate the scale (see the scale calibration procedure on page 41).

All siderail controls working properly.

Ensure that the nurse call (optional) alarm sounds in the nurse station.

Ensure that the Communication Package (optional) controls operate properly.

Verify the CPR emergency release using both CPR release handles: raise the Fowler fully up and, using the CPR handle, lower the Fowler gradually to flat position by pulling, holding and releasing the handle several times. Ensure the Knee Gatch (if raised) also starts flattening when the Fowler is completely down. Following the complete lowering of the Fowler, wait approximately 30 seconds - the time for the Fowler actuator to reset - and verify that the actuator has indeed reset itself by raising the Fowler fully up.

Verify the Fowler, Knee Gatch and Hi-lo movements to ensure that the motion interrupt switch integrated to the four electric actuators is operating properly.

Auxiliary outlet (option available only with 120V beds) working properly.

Night light (optional) working properly.

No cracks in the boards, siderails, wheel covers, steer wheel hood (optional) and plastic covers (optional) of the head and foot sections.

Head end bumpers tightly secured to frame and working properly.

No rips or cracks in mattress cover. Replace if so.

Power cord intact.

No cables pinched or worn.

All electrical connections tight.

All grounds secure to the frame.

All casters roll properly. Check caster for cuts, wear, etc.

Measure current leakage and grounding continuity of the bed and the auxiliary outlet (optional). Check with our Technical Service (see section 1.2) for the acceptable values.

12

Preventative Maintenance Chapter 2

RECOMMENDED SPARE PARTS

The following is a list of recommended on hand spare parts for the FL28EX bed.

Description

Part Number

Parts List

Electric/Electronic Components

Motor Control Board

QDF25-0592

OL250001/010

Scale Control Board

QDF25-0593

L28-008

S.A. Scale Display

QDF25-0463

L28-008

Nurse Call (Optional)/ Gen lll Control Board (Optional)

QDF21-1163

OL280002

 

 

 

Foot Board Control Board

QDF21-1154

L28-014

 

 

 

Micro Switch

1325P003

L28-013

Push Button Micro Switch

QDF9159

L28-013

Power Connector

QDF2034

OL28001/010

 

 

 

Fuse - Fast Acting 10A, 250V for 100/120V Electric Systems

QDF8078

OL280001/011

 

 

 

Fuse - Slow Blow 6.3A, 250V for 200/220/240V Electric Systems

QDF8068

OL280012/013

 

 

 

/014

Power Cord with Straight N A Molded Plug

QDF8066

OL250053

Power Cord with 90° N A Molded Plug (Optional)

QDF8066-90D

OL250055

Foot Board/Control Board Cable

QDF28-0258

OL28001/010

 

 

 

Speakerphone w/Connectors (Optional)

QDF26-0111

OL280020

 

 

 

Night Light (Optional)

QDF9539

OL250018

Load Cell w/Long Cable

QDF14-1367

L28-028

Load Cell w/Short Cable

QDF25-0218

L28-028

 

 

 

Auxiliary Power Outlet (Optional)

QDF8024

OL250029

 

 

 

5A Circuit Breaker (for Optional Auxiliary Power Outlet)

QDF9025

OL250029

Hi-Lo Actuator

25-0567

OL280001/010

Head Section Actuator

28-0403L

OL280001/010

Thigh Section Actuator

28-0377

OL280001/010

 

 

 

Toroidal Transformer (International Series Bed)

14-1160

OL280001/010

 

 

 

Stand Off Pins: Motor Control Board

QP23-0258

OL280001/010

Stand Off Pins: Scale Control Board

QDF8011

L28-008

Strain Relief Bushing

QDF9541

OL280001/010

Female Screw Lock

QDF2047

 

 

 

 

 

Foot Board Components

 

 

 

 

 

Foot End Control Panel Self-Sticking Fascia w/Scale and Bed

QDF25-0461

OP250002

Exit 3 Zone (optional)

 

 

Foot End Control Panel Self-Sticking Fascia w/Scale and Bed

QDF25-0528

OP250003

Exit 1 Zone (optional)

 

 

Foot End Control Panel Self-Sticking Fascia w/Scale w/o Bed

QDF25-0529

OP250004

Exit (optional)

 

 

Round Self-Sticking Screw Cover

QDF28-0334

L28-014

Head/Foot Board Leg Caps

QPC21-3855

L28-011-014

 

 

 

 

Siderail Components

 

 

 

 

 

Siderail Right Patient Ctrl Self-Sticking Fascia w/o Nurse Call

QDF28-0248

OL280019

nor Communications Package (Optional)

 

 

13

Maintenance Manual

Siderail Left Patient Ctrl Self-Sticking Fascia w/o Nurse Call nor

QDF28-0249

OL280019

Communications Package (Optional)

 

 

 

 

 

Siderail Right Patient Ctrl Self-Sticking Fascia w/Nurse Call, w/o

QDF28-0252

OL280020

Communications Package (Optional)

 

 

Siderail Left Patient Ctrl Self-Sticking Fascia w/Nurse Call, w/o

QDF28-0253

OL280020

Communications Package (Optional)

 

 

Siderail Right Patient Ctrl Self-Sticking Fascia w/Nurse Call and

QDF28-0250

OL280021

Communications Package (Optional)

 

 

Siderail Left Patient Ctrl Self-Sticking Fascia w/Nurse Call and

QDF28-0251

OL280021

Communications Package (Optional)

 

 

Outer Right Ctrl Panel Self-Sticking Fascia

QDF28-0139

L28-001

Outer Left Ctrl Panel Self-Sticking Fascia

QDF28-0140

L28-001

Outer Ctrl Panel Control Board

QDF21-1151

L28-001

Inner Ctrl Panel Control Board

QDF21-1169

L28-001

"Lift To Release" Self-Sticking Sticker

QDF28-0146

L28-001

 

 

 

Oblong Self-Sticking Screw Cover

QDF28-0145

L28-001

 

 

 

 

Mattress Support Components

 

 

 

 

 

Head Section Plastic Cover (optional)

28-0101

OL280032

Foot Section Plastic Cover (optional)

28-0103

OL280032

 

 

 

Steel Head Section (Optional)

28-0197L

OL280007

 

 

 

Steel Foot Section (Optional)

28-0196L

OL280007

 

 

 

Seat Section

28-0345L

OL280032

Thigh section

28-0034L

OL280032

 

 

 

CPR Mechanism Long Cable

QDF19-0815

OL280034

 

 

 

CPR Mechanism Short Cable

QDF19-0354

OL280034

 

 

 

CPR Mechanism Pneumatic Cylinder

QDF5090

OL280034

Foot End Mattress Retainer

17-0211L

OL250022/023

 

 

 

Hi-Lo Mechanism Components

 

 

 

 

 

Right "Stub-Acme" Nut

QP13-0677-05

OL280001/010

 

 

 

Nut Support

QPA25-0380

OL280001/010

Base Components

 

 

Foot End Casing Plastic Cover

QP28-0111

L28-005

 

 

 

Plastic Base Tube Cover

QP25-0023

L28-020

 

 

 

6" Caster w/Locking Mechanism

R25-0388-13

OL280003

 

 

 

5th Steer Wheel

RL5

L28-006

Right Wheel Cover

QP28-0299-13

L28-020

Left Wheel Cover

QP28-0300-13

L28-020

 

 

 

Brake/Steer Pedal

QP28-0130

L28-006

 

 

 

 

Miscellaneous

 

 

 

 

 

OG2 Grease

M0027

 

Touch-Up Paint - Opal

PD018

 

 

 

 

Ty-rap

QDF9518

 

 

 

 

Nylon Cable Dia. 3/4"

QDF9532

 

Coupling for Ty-rap

QDF9533

 

14

Troubleshooting Chapter 3

3. TROUBLESHOOTING

Please consult the following troubleshooting guide before calling the Technical Service department (see section 1.2).

3.1 TROUBLESHOOTING GUIDE

PROBLEM/FAILURE

WHAT TO CHECK

No power to bed

• Is the power cord connected to the power

 

connector and plugged into the wall outlet?

 

• Is the bed power switch at the head end of

 

the bed turned on?

 

• Is the power cord severed? Replace if

 

needed.

 

• Are the two fuses inside the power

 

connector still operational (see page 37)?

 

• Verify power at wall outlet.

 

 

No bed up or down motion when:

• Is the total lockout activated (LED (padlock)

• the siderail command is used

on)? If so, deactivate it.

• the foot board command is used

• Is the cable of the siderail control panel

properly connected to the bed receptacle

 

 

under the mattress support?

 

• Check points of the “No power to bed”

 

problem described above.

 

 

No Fowler up or down motion when:

• Is the Fowler lockout activated (LED

• the siderail command is used

(padlock) on)? If so, deactivate it.

• the foot board command is used

• Is the total lockout activated (LED (padlock)

on)? If so, deactivate it.

 

 

• Is the cable of the siderail control panel

 

properly connected to the bed connector

 

under the mattress support?

 

• Check points of the “No power to bed”

 

problem described above.

 

 

The Fowler does not fully raise

This situation happens when the CPR handle

 

is used to partly lower the Fowler. The use of

 

the CPR mechanism for this purpose creates a

 

situation where the Fowler motor is out of sync

 

with the actual position of the Fowler. To

 

correct the situation:

 

• Completely lower the Fowler using the

 

CPR handle or the Fowler down control to

 

enable the Fowler motor to reset itself.

 

Refer to the "Emergency CPR Release"

 

section of the Operations Manual for more

 

information.

 

 

15

Maintenance Manual

No Knee Gatch up or down motion when:

• Is the Knee Gatch lockout activated (LED

• the siderail command is used

(padlock) on)? If so, deactivate it.

• the foot board command is used

• Is the total lockout activated (LED (padlock)

on)? If so, deactivate it.

 

 

• Is the cable of the siderail control panel

 

properly connected to the bed connector

 

under the mattress support?

 

• Check points of the “No power to bed”

 

problem described above.

 

 

No Auto Contour motion

• Is the Knee Gatch or total lockout activated

 

(LED (padlock) on)? If so, deactivate it.

 

• Check and replace if needed the two Auto

 

Contour limit switches (see page 45).

 

 

Improper operation of the cardiac chair: Fowler

• Check and replace if needed (see pages

and Knee Gatch raise but final step of the

44 and 45) the two Cardiac Chair limit

cardiac chair position not reached (reverse

switches.

Trend).

 

Improper operation of the CPR positioning:

• Check and replace if needed (see pages

Knee Gatch does not lower and/or the Fowler

46 and 47) the two CPR limit switches.

actuator does not reset.

 

The weight shown on the scale display is

• Zero the bed without the patient in it (see

incorrect.

the "Scale System Usage" section in the

 

Operations Manual).

 

• Calibrate the scale (see page 41).

 

 

Scale display reads OVERLOAD or shows

• The weight present on the bed is superior

incorrect weight.

to the scale capacity, which is 500 lb (227

 

kg). If the patient weight is superior to 500

 

lb (227 kg), do not use the scale.

 

If not the case, zero the scale after having

 

removed the patient from the bed. This

 

situation often results from equipment

 

added to the bed without using the

 

add/remove equipment procedure (see the

 

Operations Manual). The weight of this

 

equipment is then added to the patient's

 

weight.

 

• One or more load cell cable not properly

 

connected or completely disconnected from

 

the scale control board (see figure 4.3E,

 

page 38 for the connecting positions of the

 

load cell cable).

 

 

Angle shown on the scale display is incorrect

• Calibrate the scale (see page 41).

 

 

16

 

 

Troubleshooting Chapter 3

 

 

 

 

 

 

Scale display does not turn on when the scale

• Press ENTER key twice.

 

 

is activated.

• Turn the main power switch off and turn it

 

 

 

on.

 

 

 

• Lift the foot board and gently replace it into

 

 

 

position taking care to completely push it

 

 

 

down so that the connectors match

 

 

 

correctly.

 

 

 

• Check points of the “No power to bed”

 

 

 

problem described above.

 

 

 

• Scale user interface cable not properly

 

 

 

connected or completely disconnected from

 

 

 

the scale control board (J6 connector, see

 

 

 

drawing L28-008 in the parts list manual).

 

 

 

 

 

 

Scale displays "No scale found".

• Press the ENTER key. If message

 

 

 

persists, contact our Technical Service

 

 

 

(see section 1.2).

 

 

 

If message disappears, data displayed is

 

 

 

correct and current procedure can be

 

 

 

resumed.

 

 

 

 

 

 

All lockout LEDs (padlock icons) flash or the

• The scale is unable to measure the

 

 

word Error appears at the bottom of the LCD

weight due to an electrical problem.

 

 

display.

Please contact our Technical Service

 

 

 

(see section 1.2).

 

 

 

 

 

 

Bed Exit system (optional) does not operate

• Verify that the load cell cables are properly

 

 

properly.

connected to the scale control board (see

 

 

 

figure 4.3E, page 38 for the illustration of

 

 

 

the load cell cable connecting positions).

 

 

 

 

 

 

Bed Exit system (optional) does not work at all

• Check points of the “No power to bed”

 

 

or alarm goes off when Bed Exit is activated.

problem described above.

 

 

 

• One or more load cell cable are not

 

 

 

properly connected or are completely

 

 

 

disconnected from the scale control board

 

 

 

(see figure 4.3E, page 38 for the load cell

 

 

 

cable connecting positions).

 

 

 

 

 

 

Nurse call (optional) or Bed Exit (optional)

• Check points of the “No power to bed”

 

 

signal does not reach the nurse desk.

problem described above.

 

 

 

• Check the connecting cable.

 

 

 

• Check the wall outlet.

 

 

 

• Contact the Technical Service (see 1.2)

 

 

 

 

 

17

Stryker FL28EX Service manual

Maintenance Manual

4. MAINTENANCE PROCEDURES

WARNING

Only field technicians from Stryker or service personnel trained by Stryker should perform the procedures detailed in this maintenance manual, especially those related to the Scale (option) and Bed Exit (option) systems. Failure to observe this restriction can result in serious damage to material and/or severe injury to people.

To prevent injury when working under the bed with the bed in the high position, always place blocks under the mattress support frame and apply the brakes.

Always unplug the bed power cord from the wall outlet when servicing or cleaning the bed.

NOTE

Throughout this maintenance manual, the words “right” and “left” refer to the right and left sides of a patient lying face up on the bed.

4.1 SIDERAIL MAINTENANCE PROCEDURES

HEAD SIDERAIL ASSEMBLY REPLACEMENT

Required Tools:

 

 

1/2" Wrench

3/16" Allen Key

Cutting Pliers

Procedure:

 

 

B

B

A

Figure 4.1A

1.Run the bed fully up and apply the brakes.

2.Remove the head section plastic cover (optional). Fully raise the head section and raise the siderail needing repair.

3.Unplug the power cord from the wall receptacle.

4.Loosen the lock ring (A) and unplug the siderail cable.

5.Using cutting pliers, remove the Ty-raps holding the siderail cable to the frame.

6.Using a 3/16" Allen key and a 1/2" wrench, remove the four locknuts/bolts (B) holding the siderail assembly to the head section and remove the assembly. Support the assembly when removing the last bolts.

18

Loading...
+ 42 hidden pages