General information Sirona Dental Systems GmbH

Operating Instructions SIROTorque L
General information
+
Dear customer, Thank you for purchasing a SIROTorque L
Sirona.
The technical documentation supplied is also part of the product. You should always keep this documentation within reach.
Please read these operating instructions carefully to familiarize yourself with your SIROTorque L
Please observe also the Operating Instructions of the corresponding electric motor and of the straight and contra-angle handpieces.
If, having carefully studied the operating instructions, you are nevertheless unable to proceed, contact your local dental depot.
+
Your SIROTorque L
and Instruments Team
+
unit from
+
.
2 D 3489.201.01.03.02
60 52 208 D 3489
Sirona Dental Systems GmbH Contents Operating Instructions SIROTorque L
+

Contents

1 Warning and safety information ................................................................................ 5
2 Technical description ................................................................................................. 7
3 Controls and functional elements ............................................................................. 9
4 Installing and connecting the unit............................................................................. 10
5 Operation ..................................................................................................................... 13
6 Error messages ........................................................................................................... 18
7 System information..................................................................................................... 19
8 Cleaning and care ....................................................................................................... 20
9 Key assignment overview .......................................................................................... 22
10 Troubleshooting .......................................................................................................... 23
bеЦдблЬ
60 52 208 D 3489 D 3489.201.01.03.02
3
Contents Sirona Dental Systems GmbH
Operating Instructions SIROTorque L
+
4 D 3489.201.01.03.02
60 52 208 D 3489
Sirona Dental Systems GmbH 1 Warning and safety information Operating Instructions SIROTorque L
+

1 Warning and safety information

Intended use The SIROTorque L
converting the compressed air emitted by a dental treat­ment center into electrical energy. A dental handpiece can be powered by this energy. The electric motor of the SIROTorque L replaces the air drive port (for highspeed handpieces or air-driven motors) previously used on the dental treat­ment center.
Implantology is not permisible!
Endodontics are permissible only in connection with a SIRONiTi contra-angle handpiece!
SIROTorque L+ complies with the latest regulations reflecting the current state of the art. According to the regulations, this product may only be used for the described application by trained personnel, in compli­ance with the applicable occupational safety regulations and accident prevention measures as well as the present operating instructions.
According to these regulations, users are obliged to use only faultless materials, to ensure correct application and to protect themselves, the patient and other persons against hazards.
This unit is not intended for operation in areas subject to explosion hazards.
+
(motor BL (4) or motor BL ISO (5))
+
was developed for the purpose of
bеЦдблЬ
General safety information As manufacturers of dental medical equipment and in
the interest of the operational safety of your system, we stress the importance of having maintenance and repair of this product performed only by Sirona Dental Systems or by agencies expressly authorized by Sirona Dental Systems. Furthermore components must always be replaced with original spare parts upon failure.
When having such work done, we suggest that you
a certificate stating the type and extent of work
request performed, including tions of the rated parameters or of the operating ranges (if applicable), as well as the date, name of organization and signature.
Modifications to this system which might affect the safety of the system owner, patients or other persons are pro­hibited by law!
For reasons of product safety, this product may be oper­ated only with original Sirona accessories or third-party accessories expressly approved by Sirona. The user is responsible for any damage resulting from the use of non-approved accessories.
information about any modifica-
60 52 208 D 3489 D 3489.201.01.03.02
5
1 Warning and safety information Sirona Dental Systems GmbH
Operating Instructions SIROTorque L
It is not permitted to modify the design or construction of the unit. The system owner is obliged to use only technically faultless products. Prior to engaging the system into ser­vice, its proper functioning must be checked. In case of unusual noises, please check the unit, electric motor and straight/contra-angle handpiece.
The unit is not sterile.
Please observe the operating instructions for the rele­vant electric motors.
i
NOTE
When the supply voltage drops for a moment, you must reckon with a brief speed reduction.
+
Wireless phone interference with dental electrical equipment
To ensure safe operation of medical electrical equip­ment, the use of mobile wireless phones in practice or hospital environments must be prohibited.
Highlighting of warning and safety information To avoid personal injury and material damage, you must
also observe the warning and safety information pro­vided in the present operating instructions. All such infor­mation is highlighted by the signal words NOTE,
CAUTION or WARNING.
i
Symbols used Observe accompanying documents!
Disposal It generally applies that any disposal of this product must
comply with the relevant national regulations.
Please observe the regulations applicable in your coun­try.
Within the European Economic Community, Council Directive 2002/96/EU (WEEE) requires environmentally sound recycling/disposal of electrical and electronic devices.
Your product is marked with the adjacent symbol. Disposal of your product with domestic refuse is not compatible with the objectives of environmentally sound recycling/disposal. The black bar underneath the "garbage can" symbol means that it was put into circulation after Aug. 13, 2005 (see EN 50419:2005).
Please note that this product is subject to Council Directive 2002/96/EU (WEEE) and the applicable national law of your country and must be recycled or dis­posed of in an environmentally sound manner.
Please contact your dealer if final disposal of your prod­uct is required.
The unit must be disinfected prior to disposal!
CAUTION According to US Federal Law, this product may be sold
only to or by instruction of dentists, general medical practitioners and medical specialists.
60 52 208 D 3489
6 D 3489.201.01.03.02
Sirona Dental Systems GmbH 2 Technical description Operating Instructions SIROTorque L
+

2 Technical description

Model designation SIROTorque L
Transformer primary voltage 120 V 50/60 Hz 120 V~ 60 Hz 230 V∼ 50 Hz
TIYM 24 - 02 M TDYN - 01 M TIYE 24 - 02 M 100 VA 75 VA 100 VA
Transformer secondary voltage 24 VAC
Cooling air pressure Nominal operating pressure: 44 psi (3 bar),
maximum operating pressure: 73 psi (5 bar) depending on control pressure ISO 4-hole connection
Cooling air consumption Depending on control pressure (min. 18 NL/min,
at nominal operating pressure of 44 psi (3 bar))
Spray air pressure 2.7 + 0.1 bar
Spray water 2 + 0.1 bar, min. 100 ml/min
Air requirements Dry, oil-free, dirt-free, not contaminated.
If necessary: Use compressor with dry air system, install air filter (50 µ), blow through hoses before connnecting them.
Water requirements The spray water must be dirt-free according to the appli-
cable regulations. It must not be contaminated.
+
bеЦдблЬ
Operating conditions Ambient temperature: 50 °F – 104 °F (10 °C – 40 °C)
Relative humidity: 30 % – 75 % Air pressure: 700 hPa – 1060 hPa
Transport and storage conditions Temperature: -40 °F – 158 °F (-40 °C – +70 °C)
Relative humidity: 10 % – 95 % Air pressure: 500 hPa – 1060 hPa
Protection class Protection class I
Degree of protection against electric shock Type B applied parts
Degree of protection against ingress of water Ordinary equipment (not protected)
Mode of operation Intermittent operation 60 s, 10 min pause
(“INT 1/10 MIN“)
Dimensions Width: 150 mm, depth: 130 mm, height: 70 mm
Weight (without electric motor and hose) approx. 560 g
Weight of transformer approx. 2300 g
Year of manufacture (on rear side of unit)
OMuu
60 52 208 D 3489 D 3489.201.01.03.02
7
2 Technical description Sirona Dental Systems GmbH
Operating Instructions SIROTorque L
Tests/approvals This unit is intended exclusively for operating electric
motors approved by Sirona. This applies to: BL motor and BL ISO motor
+
8 D 3489.201.01.03.02
60 52 208 D 3489

Sirona Dental Systems GmbH 3 Controls and functional elements

1 2
Endo
3
Sirona
Dental
Systems
GmbH
Fabrikstraße
31
D-64625
Bensheim
Germany
TIYE24-02M
(230V)
TIYM24-02M
(120V)
Operating Instructions SIROTorque L
+
3 Controls and functional elements
3 64
1
7
1. SIROTorque L+ supply unit
2. Control panel
3. Supply hose for BL or BL ISO electric motor
4. BL electric motor
5. BL ISO electric motor
6. Supply hose of an air-driven instrument (highspeed handpiece, air motor) of the treatment center
7. Connecting cable to transformer
8. Connection piece for electric motor hose
9. Connection piece for highspeed handpiece or air motor hose from treatment center
10. ON/OFF switch (only for a part of the unit)
5
2
11. Connection piece for transformer cable
bеЦдблЬ
12. Transformer with connecting cable to
11
10
3
8
Fabrikstraße
SironaDental System
D-64625
Germ
Bensheim
TIYE24-02M(230V)
TIYM24-02M(120V)
0123
5
sGmbH
31
any
9
6
7
SIROTorque supply unit
13. Power cable for transformer
14. Keys for setting motor speed higher (+) or lower (–)
15. Burr speed digital display
16. Motor speed digital display
17. Gear ratio digital display for straight/ contra-angle handpiece being used
18. Counterclockwise motor rotation symbol (flashes)
13
60 52 208 D 3489 D 3489.201.01.03.02
12
SIROTorque L
Bur Speed
Motor Speed Gear Ratio
17
18 2319
Motor Speed Gear Ratio
14 15 16 20 21 22
19. Instrument light "ON" symbol
20. Keys for setting gear ratio of straight/ contra-angle handpiece higher (+) or lower (–)
21. Key for instrument light ON/OFF
22. "Reverse" key for counterclockwise rotation of motor
+
23. Four colored save keys for presettings of differently colored straight/contra-angle handpieces
9

4 Installing and connecting the unit Sirona Dental Systems GmbH

Sirona
Dental
Systems
GmbH
Fabrikstraße
31
D-64625
Bensheim
Germany
TIYE24-02M
(230V)
TIYM24-02M
(120V)
Operating Instructions SIROTorque L
4 Installing and connecting the unit
The SIROTorque L+ was developed for the purpose of converting the compressed air emitted by a dental
1
5
6
4
3
treatment center into electrical energy. A straight/con­tra-angle dental handpiece can be powered by this energy. The electric motor of the SIROTorque L
+
(motor BL (4) or motor BL ISO (5)) replaces the air drive port (for highspeed handpieces or air-driven motors) previously used on the dental treatment center.
It is connected and controlled solely via the hose cou­pling (6a) of the air-driven instrument.
Control is effected by means of the pneumatic foot switch (A) which is available on the unit side. Power is supplied by an external transformer (12).
Site of installation
Mount the SIROTorque L+ (1) at an easily accessible location on top of or adjacent to the dentist element of the treatment center.
Be sure to choose the installation site of the SIROTorque L connected to the dentist element with the previously used highspeed handpiece or air motor hose (6). Ensure that the unit has a safe and stable stand.
+
so that the SIROTorque L+ can easily be
+
Connection
Detach the highspeed handpiece, quick coupling or air
A
motor used until now from the supply hose and plug con­nection piece (6a) into the four-hole connection (9) on the rear side of the unit.
ATTENTION
With a 5-hole hose, remove the lamp from the hose prior to operation.
Fit the electric motor hose (3) to the connection piece (8) on the rear side of the SIROTorque L+. Attach the corre­sponding electric motor to the motor hose (see Operat­ing Instructions of motor). Place the motor in the holder on the dentist element which was previously occupied by the air-driven instrument.
13
7
12
11
8
3
Connect the connector of the transformer connecting
(230V) (120V)
TIYE24-02M
TIYM24-02M
0123
5
H
b
m
G
s
m
ste
1
3
Sy
l
e
ta
im
e
n
e
sh
D
n
a
rikstraß
e
n
b
B
y
5
Fa
n
2
Siro
a
6
4
rm
-6
e
D
G
cable (7) to the connection piece (11) of the SIROTorque L
+
.
Plug the power cable (13) into the transformer.
All cables must be safely routed, avoid stumbling hazards!
Plug the transformer power plug into a grounded socket
9
6a
7
outlet. Do not use any transformer other than the one approved and supplied by Sirona, i.e. TIYM24-02M (120 V) or TDYN-01M (120V) or TIYE24-02M (230 V)!
60 52 208 D 3489
10 D 3489.201.01.03.02
Sirona Dental Systems GmbH 4 Installing and connecting the unit
Sirona
Dental
SystemsGmbH
Fabrikstraße
31
D-64625
Bensheim
Germany
TIYE24-02M
(230V)
TIYM24-02M
(120V)
Operating Instructions SIROTorque L
+
Supplementary notes for installing
+
C
D
120V/60Hz
7
13
1
SIROTorque L
The connecting cable (7) between the transformer (12) and the SIROTorque L+ (1) must be run in such a way that it can not be damaged by moving appliances such as chairs with wheels, carts, etc.:
Place the transformer (12) directly next to the supply unit (C).
Run the connecting cable (7) so that it does not touch the floor anywhere. To do this, use the adhesive cable clamps (D) supplied.
Run the connecting cable (7) in such a way that it cannot be damaged at the supply unit.
Install the power cable (13) of the transformer as required by the relevant national regulations.
12
Satellite version
Upon your request, the supplied SIROTorque L+ table­top unit may be converted to a "satellite version" by your service engineer. Conversion kit, Order No: 60 57 223
bеЦдблЬ
In this version, the control panel is separated from the SIROTorque L+ tabletop unit and is connected to the latter by a permanently wired cable having a length of approx. 3 m.
A
The control panel (A) can be fastened at a place that is easily accessible for the user, e.g. in the control panel front area, either with screws (from the top) or with the Velcro strap included in the scope of supply (from the front).
Fastening of the control panel is done at your own risk, make sure not to damage any cables etc. with the screws.
+
supply unit (B) can be installed
B
The SIROTorque L e.g. below the dentist element.
Establish all connections as described under "Connection".
ATTENTION
Repair and maintenance work on the SIROTorque L+ must be performed only by properly trained service engi­neers. Before opening the housing, pull the power plug or disconnect all poles from the power supply so as to deen­ergize the unit. Wait for at least 10 seconds. The compressed air delivered by the highspeed hand-
)
V
0
3
)
(2
V
0
M
2
2
(1
-0 4
M
2
2
E
-0
IY
4
T
2
M
IY
T
0123
5
bH
m
G
s
m
1
3
Syste
e
ental
sheim
D
rikstraß
na
Ben
5
Fab
ny
2
Siro
a
46
-6 erm
D
G
piece or air motor hose must be dry, oil-free and dirt-free. It must not be contaminated! If necessary, install a filter or water separator. The spray water must be dirt-free according to the appli­cable regulations. It must not be contaminated.
60 52 208 D 3489 D 3489.201.01.03.02
11
4 Installing and connecting the unit Sirona Dental Systems GmbH
max
Operating Instructions SIROTorque L
Connecting the BL or BL ISO electric motor
A
BL motor
12
A
Connection to the supply hose
Push back the cap nut (1) and insert the small tubes and contact pins of the motor in the supply hose coupling (2) as far as they will go. The arrow on the hose coupling must line up with the notch on the motor (A). Make sure that the pins and small tubes are inserted properly (i.e. are not bent).
First press the cap nut (1) gently onto the thread; then turn it counterclockwise until a faint click can be heard. Only then should you screw it onto the motor hand-tight.
BL ISO motor
+
BL motor
BL ISO motor
2
1
Adjusting the cooling spray
BL motor
3
4
The cooling spray is adjusted with the control ring (3).
The maximum water flow rate is set when the two marks (4) are lined up opposite each other.
BL ISO motor
The cooling spray is set with the rotating sleeve (5) at the rear of the hose coupling.
The threshold arrow (6) shows the direction of rotation for increasing or decreasing the cooling spray.
ATTENTION
6
During preparation work, a sufficient amount of cooling media (water component > 50 ml/min) must always be available. Insufficient cooling leads to overheating of the cavity and to damage to the tooth substance.
Please always make sure that the flow rate is sufficient!
5
12 D 3489.201.01.03.02
60 52 208 D 3489

Sirona Dental Systems GmbH 5 Operation

Sirona
Dental
Systems
GmbH
Fabrikstraße
31
D-64625
Bensheim
Germany
TIYE24-02M
(230V)
TIYM24-02M
(120V)
Operating Instructions SIROTorque L
+
5 Operation
ATTENTION
Make sure that the compressed air is always dry, clean and not contaminated. Dirty or humid compressed air leads to premature wear of the motor.
Permissible speed range: 90 – 40,000 rpm
The compressed air emitted at the highspeed handpiece hose connection piece should have a pressure of at least 44 psi (3 bar) to enable the electric motor to reach its maximum speed.
ATTENTION
The displayed speed is an approximate value. The actual speed may differ from the displayed speed. If you need an accurate speed, you should operate the motor at a speed close to 40,000 rpm and use the corre­sponding contra-angle handpieces. e.g. desired speed: 347 rpm. – Set motor to 40,000 rpm. – Use reduction gear instrument (115 : 1).
bеЦдблЬ
Switching the unit on
with the ON/OFF switch (10) on the rear side of the unit.
The display lights up and the settings of the last hand-
10
TIYE24-02M (230V)
TIYM24-02M (120V)
0123
5
bH
m
sG
31
ental System
D
Bensheim
Fabrikstraße
Sirona
any
erm
D-64625
G
piece used appear (even if they deviate from the stored values).
ATTENTION
This unit offers a higher power output than a highspeed handpiece. You should practice treatment on a phantom element before using it on a patient.
60 52 208 D 3489 D 3489.201.01.03.02
13
5 Operation Sirona Dental Systems GmbH
1 2
Endo
3
Operating Instructions SIROTorque L
+
Motor Speed Gear Ratio
Bur Speed
Motor Speed Gear Ratio
A
1:1 1:4.2 1:5
2.4 : 1 6:1 24 : 1 115 : 1
+
SIROTorque L
B
1:0.5 1:1 1:1.5 1:2 1:2.5 1:3
– 1:9 1:9.5
1.5 : 1
1.5 : 1.5
1.5 : 2
1.5 : 2.5
1.5 : 3
1.5 : 9.5 2:1 2:1.5 2:2
– 2:9.5
– 10 : 1 10 : 1.5 10 : 2 10 : 2.5
– 11 : 1 11 : 2 11 : 3 11 : 4
– 11 : 9 12 : 1 12 : 2 12 : 3 12 : 9
– 199 : 1 199 : 2 199 : 3
– 199 : 9
20
+
Setting the gear ratio
Attach the selected straight/contra-angle handpiece to the motor.
ATTENTION
Be sure to compare the gear ratio of the straight/ contra-angle handpiece you're using with the "Gear Ratio" display reading. If these two values are not identical, the burr speed display reading is not correct, since it is calculated by multiplying the motor speed by the gear ratio.
Set the gear ratio of the attached straight/contra-angle handpiece with the +/- keys (20). This is also possible when the motor is running.
Preselection of the gear ratio occurs in the following order. The most popular straight/contra-angle hand­pieces appear first straight/contra-angle handpieces are displayed from 1:0.5 to 199:9
(A). Then additional possible
(B).
14 D 3489.201.01.03.02