Parr Instrument 1552 User Manual

Operating Instructions
No. 245M
PARR INSTRUMENT COMPANY
Rev. 0
4 / 2004
for the 1552 Water Cooler
The 1552 Water cooler is intended to provide a uniform supply of cold water for removing heat and facilitating temperature control in a Parr 1281 Isoperibol Calorimeter and the Parr 1563 Water Handling System. In these applications, the water cooler is normally set to deliver water at a nominal temperature of 10 °C (50 °F) plus or minus 2.8 °C (5 °F).
INSTALLATION
Remove the cooler from its shipping carton and inspect for any shipping damage. If damage is discovered, report it immediately to the delivering carrier and retain the shipping for inspection.
Connections and 3/8” tubing for connecting to the Model 1281 Parr Isoperibol Calorimeter will be found packaged separately in the Calorimeter container.
Install the water cooler at a sufficient distance from the calorimeter or water handling system so that heat from the compressor will not affect other operations. If operations are conducted in a room where the air temperature is not closely controlled, it may be advisable to install the cooler outside the room, or in a location where heat from the compressor will not influence the ambient temperature. In any installation, allow a minimum clearance of 6 inches in front and top of the unit, 3 inches in the rear, and at least 1 inch on the sides for circulation.
Use 1/4” nylon tubing (furnished with the cooler) only for water connections on the 1563 Water Handling System.
2 ea. 496VB Elbow, Union 1/4T Brass (for use w/115V) 2 ea. 232VB Conn.,Male 3/8NPTM X 1/4T (for use w/230V) 10 ft. HJ0025TB035 Nylon Tubing, ¼” Tube
Cut the tubing squarely. Connections are made by pushing the tubing into the elbow and forcing it down as far as it will go. Tighten the fitting swaging the ferrule onto the tube.
Connect the water supply to the inlet fitting nearest the outside of the top panel marked “Water In”. Connect the cold water discharge line from the fitting nearest the middle of the top panel and extend it to the designated cold water connection on the 1563 Water Handling System or to the cold water inlet on the 1281 Calorimeter.
211 Fifty-Third Street, Moline, Illinois 61265 USA
(309)762-7716 (800)872-7720 Fax (309)762-9453
http://www.parrinst.com – parr@parrinst.com
START UP
Fill the cooler with water and purge all air from the system. Check the power supply to be sure it matches the voltage and frequency stamped on the nameplate, then plug the cooler into an appropriate electric outlet. If the fan does not start, disconnect the unit immediately and remedy any electrical problem. A temperature adjustment screw is located in an opening on the front end of the cooler. To lower the water temperature, turn the control screw clockwise.
MAINTENANCE
Clean the cabinet louvers and condenser fins periodically with a brush, air hose, or vacuum cleaner. Excess dirt and poor ventilation can inhibit cooling and cause the compressor to trip the overload protector. The motors are lifetime lubricated and require no attention. If it becomes necessary to drain the system, there is a drain plug located near the base of the unit.
If repairs are necessary, the Parr Instrument Company will furnish the name of an authorized refrigeration repair facility in the user’s area.