Parr Instrument 1356 User Manual

No. 434M
1356 ISOPERIBOL BOMB CALORIMETER
Service Manual
1. Calorimeter Operation
Operation……………………………………………………………………5 Corrections Fixed………………………………………………………… 6 Corrections and Final Report Data Entry Report Option …………………………………………………………… 7 Generate Report……………………………………………………………. 7 Display Reports …………………………………………………………… 8 Clear Memory Edit Memory Calorimeter Standardization ………………………………………………9 Heat of Combustion Calculation ………………………………………… 9 Calculations ……………………………………………………………….. 9 Correction Thermochemical……………………………………………….. 9 Details Thermochemical Calculation ……………………………………. 10 Sample Spiking ……………………………………………………………. 10 Spiking Sample Calculations ……………………………………………….11 Sample handling ………………………………………………………….12
2. 1108 Bomb Maintenance and Safety Instructions Bomb Maintenance ……………………………………………………….. 11 500 Firings…………………………………………………………………. 11
3. Parts Replacement
Motor………………………………………………………………………. 14
Battery…………………………………………………………………….. 14 Cover adjustment………………………………………………………….. 14 Thermistor………………………………………………………………….. 15 Controller………………………………………………………………… 15
Controller Combined …………………………………………………….. 15 Keyboard Cover Replacement………………………………………………………. 16 Support Rod ………………………………………………………………. 17
4. Error Messages Preperiod/Postperiod
Misfire …………………………………………………………………… 19 Exceeded Calibration Limit………………………………………………. 19 Oxygen Charge Low……………………………………………………… 19
5. Miscellaneous Misfire-wire intact
1356 Precison Calculate Tests per Oxygen Cylinder
Volatile Spiking Sulfuric acid correction …………………………………………………… 21 Use of inert gas-heat capacity ISO/BSI corrections
…………………………………………………………………. 7
………………………………………………………………8
………………………………………………………………. 8
………………………………………………………………….. 16
……………………………………………………. 18
………………………………………………………… 19
……………………………………………………………… 20
…………………………………………………………… 21
……………………………………………………… 23
………………………………………………. 7
…………………………………….. 20
…………………………………………….. 22
MSDS sheet benzoic acid…………………………………………………. 41 Data Logger 1356 ………………………………………………………… 24 Screen Bubbles ……………………………………………………………. 24 Stirrer Harness A1573E2………………………………………………….. 24 Replacing LCD Display and Keypad……………………………………. 25 Replacement I/O Board ……………………………………………………. 25 Longer Post Periods Power Supply Board Replacement CPU Board Replacement………………………………………………….. 27
6. Parts 5 year………………………………………………………………………. 27 Parts Information 5000 tests Assembly Hinge ………………………………………………………….. 29 1563 Water Handling……………………………………………………… 30
7. Diagrams
Front View 1356 Calorimeter…………………………………………….. 32 Upper Link…………………………………………………………………. 33 Lower Link………………………………………………………………… 33 Left Linkage Bracket ……………………………………………………… 34 Right Mounting Plate ………………………………………………………35 Left Mounting Plate……………………………………………………….. 35 Right Linkage Bracket…………………………………………………….. 36 Cover Assembly 1356…………………………………………………. 37 Hinge Assembly…………………………………………………………… 37 1563 Flow Diagram……………………………………………………….. 38 1563 Electrical Diagrams………………………………………………….. 39
………………………………………………………………… 28
………………………………………………………. 25
……………………………………….. 26
………………………………………………………….. 27
4
1. Calorimeter Operation
A. Operating the Calorimeter
All operations required to standardize the Calorimeter, or test an unknown sample, should proceed step-wise in the following manner:
1. Turn on the calorimeter, go to menu page 1 and press YES key on line 3 to activate the pump and heater.. The bomb parts should be wetted and then dried in the manner used at the conclusion of a test. This serves to wet all sealing parts, as well as leaving the bomb with the same amount of residual water which will exist in all subsequent testing.
Prepare the sample and charge the oxygen bomb by attaching the hose to the
bomb and pressing the O2 Fill key.
The throughput of the 1356 Calorimeter can be increased by using multiple
bombs and water buckets. With this arrangement, the calorimeter can operate almost continuously since the operator will be able to empty a bomb and recharge it while a run is in progress. A bomb and bucket for the next run will be ready to go into the calorimeter as soon as it is opened. Each bomb and bucket combination will have to be standardized separately and the proper energy equivalent for each set must be used for the heat of combustion calculation.
2. Fill the calorimeter bucket by first tarring the dry bucket on a solution or trip balance; then add 2000 (+/- 0.5) grams of water. Distilled water is preferred, but de-mineralized or tap water containing less than 250 ppm of dissolved solids is satisfactory. The water temperature should be approximately 1 to 2ºC below the room temperature. It is not necessary to use exactly 2000 grams, but the amount selected must be duplicated within +/-0.5 gram for each run. Instead of weighing the bucket, it can be filled from an automatic pipette, or from any other volumetric device if the repeatability of the filling system is within +/-0.5 ml. Try to make the starting temperature as repeatable as possible.
3, To speed and simplify the bucket filling process, and to conserve water and
energy, Parr offers a closed-circuit Water Handling System (No. 1563). This provides a water supply, cooled to the starting temperature and held in an automatic pipette ready for delivery in the exact amount needed to fill the bucket. A 1552 Water Cooler is required when using the 1563 Water Handling System. Instructions for this automatic system are given in Operating Instruction No. 245 and 246M.
4. Set the bucket in the calorimeter. Attach the lifting handle to the two holes in the side of the screw cap and partially lower the bomb in the water. Handle the bomb carefully during this operation so that the sample will not be disturbed. Push the two ignition lead wires into the terminal sockets on the bomb head. Orient the wires away from the stirrer shaft so they do not become tangled in the stirring mechanism. Lower the bomb completely into water with its feet spanning the circular boss in the bottom of the bucket. Remove the lifting handle and shake any drops of water back into the bucket. Check for any bubbles coming from the bomb, if bubbles are seen, do not fire the bomb. Resolve the problem.
5. Close the calorimeter cover
5
Press the START key . The calorimeter will prompt for a Cal ID. The calorimeter will then prompt for sample identification number by displaying Sample ID on the display. Enter the correct sample ID by using any number, up to six digits, to identify the sample. The calorimeter will check its memory and will not accept duplicate sample ID numbers. Enter this value. The system will now prompt for a sample weight, enter the sample wieght.
6. The calorimeter will now take over and conduct the test. During the time it is establishing the initial equilibrium, it will display the PRE-PERIOD. Once the bomb has been fired, the POSTPERIOD will be displayed. The calorimeter will check to make certain that a temperature rise occurs and will then look for the final equilibrium conditions to be met. If it fails to meet either the initial or final equilibrium conditions, or if it fails to detect a temperature rise within the allotted time, the calorimeter will terminate the test and advise the user of the error.
7. At the conclusion of the test, the calorimeter will signal the user and print the results.
8. Open the cover, detach the lead wires from the bomb and remove the bucket with the bomb. Remove the bomb from the bucket and open the knurled valve knob on the bomb head to release the residual gas pressure before attempting to remove the cap. This release should proceed slowly over a period of not less than one minute to avoid entrainment losses. After all pressure has been released, unscrew the cap, lift the head out of the cylinder and place it on the support stand. Examine the interior of the bomb and fuel capsule for soot or other evidence of incomplete combustion. If such evidence is found, the test will have to be discarded.
9. Wash all interior surfaces of the bomb with a jet of distilled water and collect the washings in a beaker.
10. Use Fixed Corrections or remove all unburned pieces of fuse wire from the bomb electrodes; straighten them and measure their combined length in centimeters. Subtract this length from the initial length of 10 centimeters and multiply this burned length by 2.3 calories per cm (for Parr 45C10 Fuse Wire) to obtain the fuse correction. The scale on the fuse wire card can be used to obtain this value directly. If fixed corrections are used no entry is required.
11. Use fixed corrections or titrate the bomb washings with a standard sodium carbonate solution using methyl orange, red or purple indicator. A 0.0709N sodium carbonate solution is recommended for this titration to simplify the calculation. This is prepared by dissolving 3.76 grams of Na2CO3 in the water and diluting to one liter. NaOH or KOH solutions of the same normality may be used. Enter the titer value for the acid correction. If fixed acid corrections has been programmed, no entry is required.
12 Analyze the bomb washings to determine the sulfur content of the sample if it exceeds 0.1 percent. Methods for determining sulfur are discussed in Operating Instructions No. 207M.
13. Turn off the calorimeter at the power switch.
6
B. Entering Corrections and Obtaining the Final Report
Final reports for each test can be obtained via the REPORT key, whenever the user is prepared to enter the corrections for acid, sulfur and fuse. Refer to the Reporting Generation for the steps necessary to initiate a report from the calorimeter.
C. Manual Entry
During the reporting process, the calorimeter will prompt the user to enter the following values: Fuse Correction: Key in the Fuse Wire Correction and press the ENTER key. The default setting for this value is to be entered in calories. Acid Correction: Key in the Acid Correction and press the ENTER key. The default setting for this value is to be entered in milliliters of standard alkali required to titrate total acid or calories. Sulfur Correction: Key in the Sulfur Correction and press the ENTER key. The default setting for this value is entered as percent sulfur in the sample. Enter these values when requested by the corresponding prompt. After the last entry has been made, the calorimeter will automatically produce a final report. If values for these corrections are not available, the user can use the SKIP key to pass over any of these corrections. However, a final report will not be printed until an entry is made for each of the three correction factors.
D. Fixed Corrections
In many cases, fixed values for fuse and acid can be used without introducing a significant error since the corrections are both relatively small and constant. Fixed sulfur corrections can also be used whenever a series of samples will be tested with a reasonably constant sulfur content. Details for applying fixed corrections are found in the
Details Thermochemical Calculation Any value setup as a fixed correction will be
automatically applied and the calorimeter will not prompt the user for this value.
E. Report Option Selection
Data can be transferred over the RJ45 port to a 40 or 80 column printer to provide a printed report. This port can also be used to transmit data to a host computer. In this case, the data will have to be received, stored and formatted by programs residing in the host computer. If you wish to transfer data to a computer ask for software from Parr, we can quote a price for the software. The default setting sends the calorimetric reports to the printer.
F. Report Generation
There are two kinds of calorimeter reports, which can be issued; preliminary and final. Preliminary reports are generated at the conclusion of a test run when one or more of the calorimeter corrections (FUSE, ACID, SULFUR, SPIKE) is not fixed. A final report contains all of the final or fixed calorimetric corrections needed in order to give either an energy equivalent or heat of combustion value.
7
Reports may be obtained by pressing the REPORT key. To obtain a block of reports between two specified sample numbers, press the REPORT key. Enter the first sample number to the display, and press the ENTER key. The system will then prompt for the last sample ID, if only one report is desired enter the same sample ID, followed by the ENTER key. If a block of reports is desired enter the last sample in the block followed by the ENTER key. During the reporting process, the printed reports will indicate whether the reports are final or preliminary. The preliminary reports will require corrections to be entered. Preliminary reports will remain preliminary and the energy equivalent or heat of combustion value, which is reported will reflect the fixed constants as set by the operator. The printing of large numbers of reports may be avoided at any time using the RESET key. The reset action will take effect after the current report has been completely transmitted.
G. Displayed Reports
Reports may also be obtained through the display on the 1356 Calorimeter. The procedure for obtaining reports on the display is the same as for obtaining printed reports. The calorimeter will hold data for 500 tests within its memory. These tests may be either preliminary, final, determination or calibration reports. Once the memory of the calorimeter is filled, any attempt to start a new analysis will cause the calorimeter to display MEMORY FULL, to avoid this error message go to Menu page 7 and line 6. Change the Overwrite from OFF to On, then the oldest test will be deleted and replaced by the most recent test. The alternative is to clear memory of tests.
H. Clearing Memory
This capability allows the user to delete Sample ID numbers and all related data and results for a single report, a sequence of reports or for all reports. To clear a single report, press the CLEAR MEM key. The calorimeter will prompt for the first sample number in the block and then prompt for the last sample number in the block, then press the ENTER key. To clear all reports, use the sequence procedure with 1 as the first sample number and 999999 as the last number of the sequence.
I. Editing Memory
The user is able to add data and information to the previously gathered information for a test by using the Memory Editing procedures described in the manual on page 8-1. Press the F3 key. The calorimeter will then prompt for the first sample number, which is keyed in and press the Enter key. The calorimeter will then prompt for the las t sample number in the block, enter the ID and press the Enter key. If one wishes to edit one sample, then enter the same number for the first and last sample ID. Highlight the data field to be edited by pressing the Up or Down Arrow key
8
Press the Clear key, enter the new value on the keyboard and press the Enter key. This sequence is canceled by pressing the RESET key. Press the Escape key, when one is done editing. If more than 1 sample is to be edited, the screen will display the next sample’s information.
J. Standardizing the Calorimeter
The calorimeter will calculate the average EE value from standardization tests that have been made.
K. Calculating the Heat of Combustion
While the Model 1356 Calorimeter will automatically make all of the calculations necessary to produce a gross heat of combustion for the sample. It is important that the user understand these calculations to ensure that the instrument is set up so that the calculations match the procedures used and that the units are consistent throughout the entire procedure and calculations.
L. General Calculations
Basically, the calculation of the gross heat of combustion is done using the following equation: Where: Hc =Gross heat of combustion. T =Observed temperature rise. EE =Energy equivalent of the calorimeter being used. e1 =Heat produced by burning the nitrogen entrapped in the bomb to form nitric acid. e2 =Extra heat produced due to burning sulfur to sulfur trioxide and forming sulfuric acid instead of sulfur dioxide. e3 =Heat produced by the burning fuse wire. m =Mass of the sample. For convenience and by tradition, these calculations are made in calories, grams, and degrees Celsius, and then converted to other units if required. The other units desired are set on Menu page 2, line 2. Temperature rise automatically. Corrections for heat leaks are applied automatically. Similarly, the method for extrapolating the end point of the test is discussed in the dynamic method description. Energy equivalent abbreviated as EE) is determined by standardizing the calorimeter as described in the Standardization Section of the calorimeter manual. It is an expression of the amount of energy required to raise the temperature of the calorimeter one degree. It is commonly expressed in calories per degree Celsius. Since it is directly related to the mass of the calorimeter, it will change whenever any of the components of the calorimeter (i.e. the bomb, bucket, or amount of water) is changed.
M. Thermochemical Corrections
Nitric acid correction
. The 1356 Calorimeter produces a corrected temperature rise reading
. The energy equivalent (represented by W in the above formula, or
. In the high pressure oxygen environment within the oxygen bomb,
9
nitrogen that was present as part of the air trapped in the bomb is burned to nitric oxide which combines with water vapor to form nitric acid. All of this heat is artificial since it is not a result of the sample burning. Sulfur correction. In the oxygen rich atmosphere within the bomb, sulfur in the sample is oxidized to sulfur trioxide, which combines with water vapor to form sulfuric acid. This liberates additional heat over the normal combustion process, which converts sulfur to sulfur dioxide. The sulfur correction removes this excess heat from the calculations. Fuse wire correction. The wire used for a fuse to ignite the sample is partially consumed in the combustion. Thus, the fuse generates heat both by the resistance it offers to the electric firing current and by the heat of combustion of the wire that is actually burned. It is normally assumed that the heat generated by the electrical resistance will be the same when standardizing the bomb and when testing an unknown sample, and can therefore be ignored. Significant variances can, however, occur in the amounts of fuse wire actually burned in each test. So this energy is subtracted to account for the heat of combustion of the metal.
N. Thermochemical Calculation Details
Traditionally, standard solutions and procedures have been established to simplify the calculations and thermochemical calculations.
ACID and SULFUR Corrections. In certain ASTM methods, the amount of sodium carbonate used to titrate the bomb washings is equated with e1. Users may find it convenient to enter a fixed value for the acid correction and avoid the need to determine this correction for each test. To enter Fixed Acid Corrections,go to Menu page 5 and turn on line 1 Fixed Fuse for Standardization or line 4 for Determination. Total errors of more than 3 calories will seldom occur when using Fixed Acid Corrections. Fixed Sulfur Corrections can be entered if a series of samples contain a constant amount of sulfur. For Standizations, line 3 is always ON. For determination runs to enter Fixed Sulfur Corrections, turn on Line 6 and enter the estimated value.
O. Spiking Samples
It is sometimes necessary to add a spiking material to samples, which are very small, have a low heat of combustion, or have a high moisture content to add sufficient heat to drive the combustion to completion. Benzoic acid is an excellent material for spiking for all of the same reasons that it is a good standard material. White oil is also an excellent material; particularly for liquid samples. The 1356 Calorimeter can automatically compensate for the addition of spiking materials to these samples. When Use Spiking is turned on, on Menu page 2.3 and line 1, the heat of combustion of the spiking material must be entered on Menu page 2.3, line 2. During data entry when using a spike, the calorimeter will prompt for both the sample weight and the spike weight.
10
Spiking Calculations
[{EE x T(temp rise)}-e1-e2-e3-[mass(oil) x Heat Combustion(oil)]/sample m grams
2. 1108 Oxygen Bomb Maintenance and Safety Instructions
Oxygen Bomb Maintenance
Under normal usage Parr oxygen bombs will give long service if handled with reasonable care. However, the user must remember that these bombs are continually subjected to high temperatures and pressures which apply heavy stresses to the sealing mechanism. The mechanical condition of the bomb must therefore be watched carefully and any parts that show signs of weakness or deterioration should be replaced before they fail. Otherwise, a serious accident may occur.
Do not fire the bomb if gas bubbles are observed anywhere indicating a possible gas leak. Disassemble the parts and install new seals immediately. Keep the 397A compression nut on the valve needle tightened firmly at all times.
Frequent tightening is important. This nut, if slightly loose, may allow a leak to develop during the rapid pressure rise upon ignition. This type of leak may not be detectable before firing; but if it develops, the hot gases can ignite the 20VB valve seat and burn through the head. Do not use extreme force when closing the needle valve. A moderate but firm turn on the valve knob should be sufficient to stop all gas flow. Excessive needle pressure will deform and possibly close the gas passage. If this happens, unscrew the valve body and replace the 20VB valve seat. Accumulated salt deposits may also clog the gas passage, making it difficult to release pressure at the end of a run. To avoid this, clean the passage through the valve needle and deflector nut with a small drill.
Firings 500
All O-rings and 20VB valve seats should be replaced after 500 firings for the standard 1108 Oxygen Bomb and 250 firings for the 1108CL Oxygen Bomb for chlorine content of samples is 25% or greater. A Parr 475A service clamp offers a convenient means for clamping the bomb head firmly in a vise without damaging the head when replacing any of the bomb head parts. When replacing the 230A head gasket, stretch the new O-ring and let it snap into place to be sure that it moves freely in its groove and is not twisted. To replace the valve seat, unscrew the 397A compression nut; remove the valve stem and the old seat, and disassemble all of the parts. Drop a new 20VB valve seat into the body and push it down into place. Slide a 7VBCM Monel washer, two 238A O-rings and the 378A packing cup onto the valve needle with the needle pointed upward; then adjust the parts on the needle so that the tip of the needle is flush with or slightly recessed into the bottom of the packing cup. Insert this assembly into the 369A outlet valve body and press it firmly against the valve seat by tightening and 397A compression nut to 100 inch-pounds of torque. The 238A sealing ring in the insulated electrode should be replaced with the same frequency as the 20VB valve seat. Also keep the 411A terminal nut tight at all times. As
11
Sample preparation
12
the 238A sealing ring ages and hardens it becomes a partial electrical conductor, permitting misfires and producing unwanted heating effects. Periodic replacement will eliminate this potential problem. The threads on the screw cap should be checked routinely for any burrs or other deformity. After long use, the threads on the screw cap may become worn to the point where they will no longer provide a safe closure for the bomb, and the screw cap will have to be replaced. The following procedure can be used to check the extent to which the threads have become worn.
Assemble the bomb with the head in the cylinder and count the number of turns required to bring the screw cap down firmly against the head. Then open the bomb; remove the head and replace the screw cap, but turn it down to only one-half of the turns previously counted. This will usually be about four turns. With the screw cap in this position, use a dial gage to measure the vertical deflection when lifting the screw cap upward. If this measurement exceeds 1/32 inch (0.030”), the screw cap is unsafe and should be discarded. The cylinder can then be returned to the factory for inspection. If the threads on the cylinder are in good condition, a new screw cap can be custom-fitted to the cylinder.
Never under any circumstances use oil on valves or fittings which handle compressed oxygen. This precaution applies to all of the oxygen bomb parts as well as to
the oxygen filling connection.
The 1108CL Bombs will resist chlorine, fluorine or bromine in the presence of moisture. If samples yielding appreciable amounts of these elements are burned in a Parr bomb, the bomb should be emptied and washed as quickly as possible after each combustion. If the interior of the bomb cylinder should become etched, the resistance of the metal to further attack can be improved by restoring the surface to its original highly polished condition. Bombs needing polishing or other repair work can be returned to the factory. A periodic overhaul and test at the factory will help to keep any Parr oxygen bomb in first-class condition.
Parr oxygen bombs can be returned at any time for repair and testing. A factory test is recommended after every 5000 firings, or sooner if the bomb has been subject to any of the following conditions: (a) fired with an excessive charge, (b) ignition of any internal components, (c) machined by any source other than the factory, (d) damaged by corrosive vapors that might have exceeded 80% of the corrosion allowance, or (e) any changes in the threads on the bomb cylinder and/or screw cap. When returning a bomb to the factory, ship it to Parr Instrument Company, 211-53rd Street, Moline, Illinois 61265, with the package marked for the attention of the Repair Department. A purchase order covering the repair work should be included with the shipment or mailed to the same address, as no repairs will be started without specific instructions. Be sure to include a return shipping address and the name and telephone number of the individual to be contacted if questions arise concerning excessive repair costs or other problems. Individual repair parts can be ordered from any Parr dealer or direct from the factory.
13
Loading...
+ 28 hidden pages