Omnicell External Return Bin User manual

External Return Bin Installation and Service Guide
67-2038 Rev D
This guide is CONFIDENTIAL and designed only for Omnicell Technical personnel and/or designated representatives.
This guide and accompanying software and/or hardware described in it are protected under copyright laws and may not be copied, wholly or in part, without the express written consent of Omnicell, Inc. The same proprietary and copyright notices must be attached to any permitted copies as were attached to the original documents.
Omnicell, Inc. 1201 Charleston Road Mountain View, CA 94043 (650) 251-6100 www.omnicell.com
Omnicell and the Omnicell design mark, OmniBuyer, OmniCenter, OmniRx, OmniSupplier, SafetyMed, SafetyPak, SafetyStock, and Sure-Med are registered trademarks. Anesthesia TT, Anesthesia Workstation, Anywhere RN, Executive Advisor, Flexbin, Medication Surveillance, OmniDispenser, OmniLinkRx, OmniScanner, OmniTrack, Omni TT, Open Touch, OptiFlex, OptiFlex MobileTrack, Point-to-Point Medication Safety, SecureVault, See & Touch, SinglePointe, TempCheck, Touch & Go, VSuite, and WorkflowRx are trademarks of Omnicell, Inc. in the United States and internationally. All other trademarks and trade names are the property of their respective owners.
Copyright 1999-2011 Omnicell, Inc. All rights reserved.
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.

Table of Contents

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Software Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Software Configuration Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
ERB Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Mounting Brackets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
OmniRx. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
OmniTT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Half-Cell Cabinet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
One-, Two-, and Three-Cell Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Anesthesia Workstation (AWS) G3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Anesthesia Workstation (AWS) G4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Wall and Table Templates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
iii
Installation Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Cable Assembly Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Cabinet Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Installation on G3 Cabinets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
OmniTT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Attaching the ERB to the OmniTT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Cabling for the OmniTT G3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
OmniRx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Attaching the ERB to the OmniRx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Cabling for the OmniRx G3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Half-Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Attaching the ERB to the Half-Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Cabling for the Half-Cell G3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
One-, Two-, and Three-Cell Cabinets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
Attaching the ERB to the One-, Two-, and Three-Cell Cabinets . . . . . . . . . . . . . 2-21
Cabling for One-, Two-, and Three-Cell G3 Cabinets . . . . . . . . . . . . . . . . . . . . . . 2-23
Anesthesia Workstation (AWS) G3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Attaching the ERB to the AWS G3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Cabling for the AWS G3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
Installation on G3 Cabinets with G4 Console Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . 2-40
Attaching the ERB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-40
Cabling for the OmniTT G3 With a G4 Console Upgrade. . . . . . . . . . . . . . . . . . . . . . 2-40
Cabling for the OmniRx G3 with a G4 Console Upgrade. . . . . . . . . . . . . . . . . . . . . . . 2-40
Cabling for the Half-Cell G3 with a G4 Console Upgrade . . . . . . . . . . . . . . . . . . . . . . 2-40
Cabling for the One-, Two-, Three Cell with a G4 Console Upgrade . . . . . . . . . . . . . 2-40
Cabling for the Anesthesia Workstation G3 with the G4 Upgrade . . . . . . . . . . . . . . . 2-41
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
iv Table of Contents
Installation on G4 Cabinets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-45
Wall or Table Mount. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-71
Daisy Chains. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-72
Implementation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-77
Attaching the ERB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-45
Cabling for the OmniTT G4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-46
Cabling for the ERB on the Left Side of the Cabinet . . . . . . . . . . . . . . . . . . . . . . . 2-46
Cabling for the ERB on the Right Side of the Cabinet . . . . . . . . . . . . . . . . . . . . . . 2-51
Cabling for the OmniRx G4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-54
Cabling for the Half-Cell G4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-54
Cabling for the One-, Two-, Three-Cell G4 Cabinets . . . . . . . . . . . . . . . . . . . . . . . . . . 2-55
Installing on the Anesthesia Workstation (AWS G4) . . . . . . . . . . . . . . . . . . . . . . . . . . 2-58
Attaching the ERB in the Upper Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-58
Attaching the ERB in the Lower Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-60
Cabling for the AWS G4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-61
Cabling for the AWS G4 with the ERB Mounted in the Upper Position . . . . . . 2-61
Cabling for the AWS G4 with the ERB Mounted in the Lower Position . . . . . . 2-66
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-77
Configuration and Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-77
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Return Door Solenoid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Return Door Handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Printed Circuit Board (PCB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Cam Locks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Access Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Cable Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Appendix A: Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Part Link Table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
ERB Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Documentation Feedback. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . FB-1
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
1-1

Overview

This document provides installation and service instructions for the External Return Bin (ERB) on the G3 cabinets, the G3 cabinets with the G4 electronics box, and the G4 cabinets.
The ERB is used to collect unused medication after it has been issued to a patient. Only one return bin is supported per cabinet.
The ERB can be mounted to:
OmniRx Color Touch cabinets
OmniTT Color Touch cabinets
Half-Cell Color Touch cabinets
One-, two-, and three-cell Color Touch cabinets
Anesthesia Workstations
Walls
Table tops
The ERB comes built with a standard 4-foot cable mounted on the right side of the bin.

Software Requirements

The cabinet to which the ERB is attached must be running Color Touch 5.3 or higher.

Software Configuration Options

Refer to the Color Touch user guide of the version on the cabinet for the options that support the External Return Bin.

Tools

The following tools are needed to install or mount the External Return Bin:
Nut driver (5/16", 11/32", and 1/4")
Allen wrench (3/32", 9/64" and 5/32")
Crescent wrench
T8 Torx driver
T10 Torx driver
T15 Torx driver
T20 Torx driver
Phillips screwdriver #2 for the ERB handle
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
1-2 Overview

ERB Components

ERB Components
Figure 1-1 shows the components from the External Return Bin manufactured assembly. For a
listing of kits, components or spares, see “Appendix A: Parts” on page A-1.
Return
Door
Figure 1-1. ERB Mfg. Assembly #14-7030

Mounting Brackets

Kit #20-6021 contains various mounting brackets. The mounting part numbers and type are stamped into the metal.
OmniRx
Access Door
Foam Pad
key #2204 key #2202
Metal
Retainer
Strain Relief
Molex
Connector
3-Pin
Header
Plug Cover
6-Pin
Header
AUX Connector
4 ft. Cable
Assy
4-Pin
Header
Figure 1-2. OmniRx brackets
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
OmniTT
Figure 1-3. OmniTT brackets
Half-Cell Cabinet
Overview 1-3
ERB Components
Figure 1-4. Half-cell cabinet brackets
One-, Two-, and Three-Cell Cabinet
Figure 1-5. One-, two-, and three-cell cabinet brackets
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
1-4 Overview
ERB Components
Anesthesia Workstation (AWS) G3
Figure 1-6. AWS G3 bracket
Anesthesia Workstation (AWS) G4
Figure 1-7. AWS G4 bracket
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
Wall and Table Templates
Drill templates are available for mounting an ERB to a wall or table.
Figure 1-8. Wall and table mount templates #60-0052
Overview 1-5
ERB Components
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
1-6 Overview
ERB Components
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.

Installation Instructions

Setup

Before installing the ERB, change the cabling configuration if necessary, and perform cabinet preparation.

Cable Assembly Configuration

Important: The ERB cable assembly is configured to be installed on the right side of cabinets with the
cables leading to the back. If the ERB is to be installed on the left side of the cabinet, the cable assembly must be switched to the left side. This avoids having to bend cables back across the ERB to reach the back of the cabinet.
If the ERB is to be installed on the right side of the cabinet, skip to “Cabinet Preparation” on page 2-10.
If the ERB is to be installed on the left side of the cabinet, continue in this section to change the cabling configuration.
2-1
1. Place the ERB on a flat clean surface.
2. Open the access door using both keys. Key #2204 unlocks the access door. Key #2202 overrides the solenoid.
Figure 2-1. Unlock the access door
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-2 Installation Instructions
Setup
3. Lay the ERB down as shown in Figure 2-2, and unscrew the two 4-40 flat head fasteners from under the ERB.
Figure 2-2. Unscrewing the fasteners
4. Turn the ERB upright and lift out both I/O cover plates.
Figure 2-3. Removing the I/O cover plates
5. Remove the PCB/solenoid metal mount.
a. Open the return door.
b. Push the pawl of the solenoid to the left.
Figure 2-4. Pushing the pawl
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
Installation Instructions 2-3
Setup
c. Loosen, but do not remove, the two nuts on either side of the PCB/solenoid plate with a
5/16" nut driver.
Figure 2-5. Loosening the nuts on both sides of the PCB/solenoid plate
d. Maneuver the PCB/solenoid mount around the sheet metal tabs and nuts. Remove the
lower rim first.
Figure 2-6. Removing the PCB/solenoid mount
Under the PCB/solenoid mount is an internal cover to reduce the radiated emissions level.
6. Remove the RFI cover that is under the PCB/solenoid mount.
a. Loosen the two hex nuts on either side of the internal cover using a 5/16” nut driver or
socket.
Figure 2-7. Loosening hex nuts
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-4 Installation Instructions
Setup
b. Disconnect the yellow solenoid cable for convenience.
Figure 2-8. Disconnecting solenoid cable
c. Lift the RFI cover off the two studs. Slide the wiring rubber grommets down the two
keyhole slots and remove the RFI cover completely.
Figure 2-9. PC board with RFI cover removed
7. Unplug the PowerCom cable from J1 on the PC board.
Figure 2-10. J1 location on the PC board
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
Installation Instructions 2-5
Setup
8. Remove the gray PowerCom cable from the right side of the ERB and set it aside.
Figure 2-11. Removing the PowerCom cable
9. Squeeze and pull the black grommet to remove it from the opening.
10. Remove the black plug cover by pushing against the groove from the inside with a screwdriver.
Figure 2-12. Removing the plug cover
11. Thread the PowerCom cable through the rear cutout on the opposite side of the ERB. The correct orientation of the cable assembly exiting from the left side of the ERB is shown in the figure below.
Figure 2-13. PowerCom cable routing
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-6 Installation Instructions
Setup
12. Cover the gray PowerCom cable with the black grommet taken off in step 9. Squeeze the grommet together and push through the double D cutout on the left side.
Figure 2-14. Securing the grommet around the PowerCom cable
13. Snap the plug cover into the left-side D cutout adjacent to the cable assembly. The plug covers the opening which will be used for future hardware developments.
Figure 2-15. Snapping in the plug
14. Rotate the lower plates. The open side of the left lower plate should face left. The open side of the right lower plate should face down. This allows the left cover plate to protect the cables being routed out of the ERB. The right cover plates will seal the cable openings on the right side of the ERB.
Figure 2-16. Orientation for left cover plates
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
Installation Instructions 2-7
Setup
15. Install the left I/O cover plates. Take care not to pinch any wires.
Note: The 6-pin white connector is not used.
6-pin
connector
Figure 2-17. Left I/O cover plate
16. Re-insert the PowerCom cable into J1 on the PC board. Make sure all 6 pins on the cable connector are inserted.
Figure 2-18. J1 location
17. Secure the left I/O cover with the 4-40 flat head fastener removed in step 3 on page 2-2.
Figure 2-19. Replacing the fasteners
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-8 Installation Instructions
Setup
18. Reinstall the RFI cover.
a. Slide the two grommets back into the keyhole openings on the top and bottom of the RFI
cover. Hold the RFI cover over the studs against the rear wall. Make sure wires are not pinched.
Figure 2-20. RFI rubber grommet
b. Reconnect the 2 position yellow wire solenoid cable, and slide the internal cover over the
studs.
Figure 2-21. Sliding RFI over studs
c. Retighten the two hex nuts. Tighten using the 5/16” nut driver or socket.
19. Reinstall the PCB/solenoid mount with the solenoid inserted first into the ERB housing, clearing the return door. Keep the external return door open while reinstalling the mount plate.
Figure 2-22. Installing the PCB/solenoid mount
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
Installation Instructions 2-9
Setup
20. Maneuver the plate onto the two studs behind the nuts and position the bottom rim behind the sheet metal tabs. Be careful not to pinch the wires behind the plate.
21. Tighten the two loose nuts with a 5/16" nut driver.
Figure 2-23. Tightening the nuts
22. Install the right I/O cover plates.
23. Secure the right I/O cover plates with the 4-40 flat head fastener removed in step 3 on page 2-2. The side that does not have the cable installed has the two D cutouts covered by the I/O cover.
Figure 2-24. Covered cutouts
24. Close the return door.
25. Check that the solenoid lock mechanism is latching correctly.
Figure 2-25. Complete assembly
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-10 Installation Instructions

Installation on G3 Cabinets

Cabinet Preparation

1. Perform a graceful shutdown on the cabinet that is to have an ERB mounted.
2. Unplug the power cable.
3. Refer to the Omnicell Drug Handling Policy and Procedures for information regarding the removal of drugs from the internal return bin drawer.
Caution: The pharmacist must be present when servicing any drawers where medication is present.
4. Empty the contents of the internal return bin drawer if it exists.
Important: Once the ERB is enabled, the internal return bin is no longer accessible.
Installation on G3 Cabinets
This section contains ERB installation instructions for various G3 cabinet types. For installation on G3 cabinets that have the new G4 E-box, see “Installation on G3 Cabinets with G4 Console
Upgrade” on page 2-40. For installation on G4 cabinets, see “Installation on G4 Cabinets” on
page 2-45.
This section describes installation procedures for the following:
“OmniTT” on page 2-10
“OmniRx” on page 2-17
“Half-Cell” on page 2-19
“One-, Two-, and Three-Cell Cabinets” on page 2-21
“Anesthesia Workstation (AWS) G3” on page 2-26

OmniTT

Attaching the ERB to the OmniTT
1. Remove the two screws securing the transport handle on the right side of the OmniTT Color Touch cabinet, then remove the original transport handle.
Figure 2-26. Removing the original transport handle
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
Installation Instructions 2-11
Installation on G3 Cabinets
2. Insert the longer end of the new transport handle into the left slot followed through with the other end into the right slot.
Figure 2-27. Installing the new transport handle
3. Secure the handle with the two screws removed in step 1. Note the threaded hole orientation.
4. Place the External Return Bin on a flat clean surface, before opening the access door with both keys. Use key #2204 to unlock the access door and key #2202 to override the solenoid.
Figure 2-28. Opening the access door
5. Align the two drilled holes on the transport handle to the two drilled holes inside the ERB. Secure the handle with two #8 hex nuts.
Figure 2-29. Screw hole locations
6. Close and lock the access door
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-12 Installation Instructions
Installation on G3 Cabinets
Cabling for the OmniTT G3
1. Lift the console cover from the back, then prop it open with a switch panel housing or the prop arm on a newer electronics tray.
Figure 2-30. Propping the cover
2. Review the back panel layout in Figure 2-26.
prop arm
Figure 2-31. Back panel layout
3. Remove the AUX cable connector at the rear of the electronic sled by pushing down on the AUX connector and pulling it out through the back.
Figure 2-32. Removing the AUX cable connectors
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
Installation Instructions 2-13
Installation on G3 Cabinets
4. Remove the second AUX cable connector.
Figure 2-33. Removing the other Aux connector
5. Unplug the 4-pin connector from J24 (labeled AUX) on the PowerCom board, then remove the cable.
Figure 2-34. Unplugging the 4-pin connector
6. Insert the black AUX connector from the External Return Bin retrofit cable assembly (#42-7052) into the first opening.
Figure 2-35. Inserting the AUX connector from the ERB cable assembly
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-14 Installation Instructions
Installation on G3 Cabinets
7. Plug in the 4-pin header connector on the retrofit AUX cable assembly into J24 (labeled AUX) on the PowerCom board.
Figure 2-36. Plugging in the 4-pin connector
8. Plug in the 3-pin connector with the blue and yellow cables on the retrofit AUX cable assembly into J16 (labeled MAG) on the PowerCom board.
Figure 2-37. 3-pin connector location
9. Thread the ERB cable connector (white) through the second AUX opening.
Figure 2-38. Threading the ERB cable connector
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
Installation Instructions 2-15
Installation on G3 Cabinets
10. Slip the metal retaining bracket over the white connector and position it on the inside rear of the metal rail.
Figure 2-39. Slipping on the retaining bracket
11. Snap on the black grommet over the outside ERB cable.
12. Push the grommet half-way through the second AUX opening.
13. Snap the metal retainer into the groove of the black grommet. The grommet covers the cable so that the metal retainer snaps in the grommet groove and secures the cable to the inside back of the electronic sled.
Figure 2-40. Snapping the grommet and retainer in place
Figure 2-41. Inside view of retainer snapped on grommet to secure the cable
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-16 Installation Instructions
Installation on G3 Cabinets
14. Connect the ERB 6-pin female connector to the Molex mini-fit Jr. 6-pin connector on the retrofit AUX cable assembly.
Figure 2-42. Connecting the 6-pin connector
15. Tuck all the cables out of the way. Use cable ties as needed.
Figure 2-43. Tucked cables
16. Remove the prop, then close the console cover.
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.

OmniRx

Installation Instructions 2-17
Installation on G3 Cabinets
Attaching the ERB to the OmniRx
1. Preassemble the two #8-32 x 3/8" socket head cap screws to the modified transport handle in the kit.
Figure 2-44. Assembling the transport handle
2. Remove the screws that secure the original transport handle on the right side of the OmniRx Color Touch cabinet.
Figure 2-45. Removing the original transport handle
3. Remove the original transport handle and replace it with the modified transport handle.
4. Secure the new handle with the screws removed in step 2.
5. Place the mounting plate over the modified transport handle. Align the drilled holes over the studs on the handle.
Figure 2-46. Mounting plate
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-18 Installation Instructions
Installation on G3 Cabinets
6. Align the two drilled holes on the ERB to the two studs on the OmniRx mounting bracket plate.
Figure 2-47. Aligned screw holes
7. Open the ERB access door with both keys. Use key #2204 to unlock the access door and key #2202 to override the solenoid.
8. Fasten the ERB to the mounting plate with the two #8 hex nuts.
Figure 2-48. Hex nut locations
9. Close and lock the access door.
Cabling for the OmniRx G3
Connect the ERB to the OmniRx using the same procedures as for the OmniTT and half-cell cabinets. See “Cabling for the OmniTT G3” on page 2-12.
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.

Half-Cell

Installation Instructions 2-19
Installation on G3 Cabinets
Attaching the ERB to the Half-Cell
1. Install the half-cell mounting bracket to the right side of the half-cell using (4) #8-32 x 1/4" flat
head cap screws.
Figure 2-49. Installing mounting bracket
2. Align the drilled holes circled on the ERB to the half-cell bracket plate.
Figure 2-50. Aligned drill holes
© 2011 Omnicell, Inc. External Return Bin Installation and Service Guide/67-2038 Rev D
2-20 Installation Instructions
Installation on G3 Cabinets
3. Open the ERB access door with both keys. Use key #2204 to unlock the access door and key #2202 to override the solenoid.
Figure 2-51. Opening the access door
4. Secure the ERB with four #8 hex nuts.
#8 hex nut locations
Figure 2-52. Hex nut locations
5. Close and lock the access door.
Cabling for the Half-Cell G3
Connect the ERB to the half-cell cabinet using the same wiring procedures as for the OmniTT and OmniRx. See “Cabling for the OmniTT G3” on page 2-12.
External Return Bin Installation and Service Guide/67-2038 Rev D © 2011 Omnicell, Inc.
Loading...
+ 82 hidden pages