Olympus CLV-180 User manual

ISSUE 1
MAINTENANCE MANUAL
CLV-180
FR-1700
CLV-180

INTRODUCTION

Introduction
This manual is intended for Olympus-certified technicians. Use of this manual by other individuals is prohibited.
Precautions
Before Repair
(1) Notify the site manager of the intent to repair the unit and obtain his/her approval before
commencing.
(2) During repair, there is a risk of injury from the various tools and parts in the vicinity, as well as
from fluid leakage on the floor. Inform the relevant people to restrict access to the repair area.
(3) To prevent potentially dangerous health risks, avoid working in a closed room, i.e., select a
well-ventilated location when using organic solvents.
(4) In general, it is advisable to record the function and operation settings before repair, to enable
restoration of the settings after service.
(5) If the original settings cannot be known due to mechanical problems present at the time the unit
was accepted for repair, apply the factory-set values, or the safest settings (such as the lowest output levels). In such a case, inform the user that the settings have been changed.
(6) Guard against static electricity.
Use a conductive mat or wristband to discharge static electricity to prevent damaging the boards or other electrical components if it is necessary to touch them.
During Repair
(1) To prevent potentially dangerous health risks, thoroughly rinse any bodily areas that have come
into contact with organic solvents as soon as possible.
(2) When using organic solvents, handle flames such as those in alcohol lamps with caution,
because these solvents may ignite if exposed to flame.
In addition, always replace the lids back onto organic solvent containers before leaving the work-bench.
(3) Beware of electric shock.
Turn off the power and unplug the power cord before removing the cover of the unit.
(4) Beware of residual voltages.
The unit may contain residual charges in capacitor components. Take care to avoid electric shock when opening the top cover.
(5) To avoid personal injury and damage to the unit, heavy units should always be assembled or
disassembled by at least two people.
(6) Repair with extreme caution to avoid injury.
Use extra caution around metal parts because the edges may be sharp.
(7) Reassemble parts according to their original configuration. This regards the following items
especially: a) Insulators, such as insulating tubes and mylar sheets. b) Cable rerouting, clamps, and cores.
ISSUE1 INTRODUCTION
CLV-180
c) Shield parts and cover screws with toothed washers.
Failure to attach parts in their original configurations, even if it does not impair product function, poses the risk of noise radiation and reduced electrical safety.
(8) Use specified parts.
The parts and components of this product are designed to operate under certain anticipated vibration, heat, chemical exposure, and voltage conditions. Always replace parts with those specified in the parts list.
(9) Always tighten nuts/screws to the specified torque and only use components of the specified
dimensions.
(10) Do not reuse O-rings, E-rings, or packing. Always use new ones. (11) When reusing components, remove sealing tape or sealing compound, and clean the
components before use.
(12) Check that no screws or nuts are loose.
After Repair
(1) After repair, inform the site manager of the nature of the problems, the cause, the
countermeasures taken, and the parts replaced.
(2) Always notify the site manager if, during repair, liquid disinfectants, cleaners, or alcohol were
used. Inform the site manager that before using the unit, it is necessary to verify the
concentration levels of disinfectants to ensure that they have not been diluted.
(3) Restore the original unit settings as they were recorded before repair. Request the presence of
site manager to verify the settings with you.
Copyright
© 2005 Olympus Medical Systems Corp. All rights reserved. Unauthorized reproduction or distribution in part or in whole is prohibited.
Trademarks
OLYMPUS is a registered trademark of the Olympus Corporation. The company names, product names, and proprietary technical terms in this document are the trade-marks or registered trademarks of their respective owners.
ISSUE1 INTRODUCTION

Contents

Chapter 1: Product Specifications………………………………………...1-1
1 Product Outline…………………………………..…………………………………………..1-2 2 Features……… ……… ………….……… …………..…………………… …… ……… ……..1-2 3 Operational Conditions……………………………..……………………………………...1-3 4 Specifications……… ….… ……… … …….………… ..…………………… ……… …….1-4 5 Nomenclature and Functions…………………………………………..……………………..1-9
5-1 Front panel………………………………………………….…………………………….... 1-9
5-2 Control panel(buttons)……………………………………………………………………1-10 5-3 Control panel(indicators)………………………………………………………………...1-12 5-4 Rear and side panels……………………………………………………………………..1-14
Chapter 2: Troubleshooting………………………………………..…….2-1
1 Contents……… ……… ……… ……… ……… … ……… …………………………………… .2-2 2 Troubleshooting…………………………...…………………………………………………….2-2
2-1 Power input failure…………………………………………………………..………….2-3 2-2 Lamp failure…… ……… ……… ……… … ………… ……………………..…… …….2-4 2-3 Lamp malfunction…………….…… …… …………………….…………..… ……….2-4 2-4 Scope connection failure………………………………………………………………….2-5 2-5 High-brightness malfunction……………………………………………………………..2-6 2-6 Cooling Fan malfunction………………………………………………………………….2-7 2-7 Shield malfunction at Scope removal…………………………………………………..2-7 2-8 Pump malfunction………………………………………………………………………….2-8 2-9 Filter Change Malfunction………………………………………………………………..2-9 2-10 Outgoing light failure……………………………………………………………………2-11 2-11 Manual Brightness failure……………………………………………………………...2-11 2-12 Automatic Brightness failure………………………………………………………….2-12 2-13 Panel Malfunction……………………………………………………………………….2-13 2-14 Emergency Lamp malfunction………………………………………………………..2-15 2-15 Back-up malfunction…………………………………………………………………..2-16 2-16 Lamp Life Meter Reset malfunction…………………………………………………2-16 2-17 Transillumination malfunction……………………………………………………….2-16
Chapter 3: Precaution on disassemble and reassembly…….....3-1
1 Warning… … … … … … … … ..………………… … … … … … … … ..… … … ……………….3-2 2 Caution… ………… … … … … … ………… … … … … .…… … … … … … …… ………… … .3-2
Chapter 4: Assemble procedure…… ………… … ……….....4-1
1 Jigs and Tools………… ……… …………… …… ……………………………………… ….4-2 2 Areas…………… ………………………………………….…………………………………….4-2 3 Disassembling and Reassembling procedure……………………………………………….4-3
Chapter 5: Parts List… … ....… … … … … … … … … … .....5-1
1 Exploded Parts Diagram…………………………..…………………………………………….5-2
2 Parts List… … … … … … … … … … … … … … .… … … … … … … … ....… … … … ..…… … … .5-10
SpecificationCLV-180
Chapter 1: Products Specifications
ISSUE1
1-1
SpecificationCLV-180
PRODUCT SPECIFICATIONS
1 Product Outline
Intended use: When used in conjunction with an OLYMPUS endoscope, video system center, video monitor, accessories, and other related equipment, this product supports examination and endoscopic treatment by enabling observation through a fiberscope or through a video monitor.
Compatibility: This product may be used in conjunction with EVIS 100/130/140 series videoscopes, EVIS EXERA 145/160 series videoscopes, EVIS EXERA II 165/180 series videoscopes, V series videoscopes, VISERA series videoscopes, rigid light guides, fiberscopes, as well as EVIS EXERA II CV-180/165, and endoscopic ultrasound systems.
Durability: With adherance to the conditions listed below, this product has a usable life of 6 years from the manufactured/shipped date. Conditions: During this product’s usable life, the user must perform inspection before using, and
regular subsequent inspections as outlined in the attached materials and the operation manual. Should the inspections reveal that repairs are required, the user must have those repairs performed.
2 Features
(1) This is a light source unit for exclusive use with the EVIS EXERA II VIDEO SYSTEM CENTER
CV-180/165. However, this product is not to be used for surgery if used with the CV-165 (high-intensity function is not available).
(2) A mechanical detection unit is installed in the endoscope socket, enabling automatic switching to
the maximum amount of light appropriate for any endoscope connected. It is capable of distinguishing between GI-type scope connectors, LG connectors for use with the SP high-intensity function, and LG connectors for use with the non-SP, high-intensity function.
(3) When videoscopes or OES fiberscopes (including OES connectors used with BF and CHF type
endoscopes) are attached, the unit is capable of generating 1.3 times the output light intensity of the CLV-U4 EVIS 100 mode and the OES mode (OAI, OLA). It produces 1.6 times more maximum permissible illumination (white light) for the LG (non-OAI, OLA regions).
(4) When connected to a surgical fiberscope not compatible with the high-intensity RF connector, it
produces an output light intensity of the maximum permissible illumination (white light) for the LG equivalent to the CLV-U4 EVIS 100 mode and the OES mode.
(5) When used with an endoscope and light guide compatible with the high-intensity RF connector, the
high-intensity mode is available, creating a brighter illumination. Compared to the illumination of the CLV-U40 EVIS 100 mode or OES mode, approximately 2 times the maximum illumination (white light) is produced. In high-intensity mode, it is also possible to select equal brightness or double brightness of the CLV-U40 through the brightness mode switch.
(6) The optical filter used in NBI (narrowband imaging) observation has been loaded. NBI observations
are enabled when an NBI-compatible endoscope is connected to the unit.
(7) The optical filter used in PDD (photodynamic diagnosis) observation can be loaded (OE market).
One other optical filter for specialized observations may be loaded if required. (8) The following modes may be selected: a. Auto-ignition mode: the examination lamp automatically illuminates when the power is turned on. b. Manual ignition mode: the examination lamp will illuminate when the lamp switch is pressed once
power has been turned on. (9) The examination lamp can be manually turned off by continously pressing the LAMP switch for
approximately 1 second. (10) Heat exhaust is through the rear (rear heat exhaust). (11) The system automatically switches to the emergency lamp if the examination lamp does not come
on after pressing the lamp switch. As an emergency response, this will provide sufficient brightness
required for endoscope removal in situations when the examination lamp does not come on or if it
turns off. This will be indicated on the panel. (12) Brightness can be adjusted using the 17 steps of automatic and manual brightness available. (13) The air supply can be stopped completely as well as changed between high, medium or low supply. (14) Disconnection of the emergency lamp will be automatically indicated on the panel. (15) The transillumination switch enables the transmitted light illumination function (for GI endoscopes
ISSUE1
1-2
SpecificationCLV-180
only). (16) Panel settings are retained, and the light source unit will display at the previous settings the next
time the unit is turned on. However filter switch settings are not backed up, and will revert to the
regular filter. (17) The exterior and the operating panel may be disinfected with ethanol (70 ethyl or isopropyl
alcohol) or with a mild detergent. (18) This product has passed the OLYMPUS Eco-Product standard (Environmentally-friendly design
standard).
3 Operational Conditions
(1) Applicable video system centers: CV-180, CV-165 (2) Operational environment:
i) Use in a medical facility under the supervision of a medical doctor. ii) Do not apply this light source directly to the heart.However, it may be used with the heart when
used in combination with a TYPE CF applied part indicated by a mark on the instrument. iii) The outer casing of the unit must be grounded for safety. iv) Do not use the light source in a combustible atmosphere. v) Do not use the light source with the following electronic equipment or endoscopes:
a. Any devices designed to apply electronic treatment of a patient for which safe usage with
the CLV-180 has not yet been confirmed.
b. Any devices not designed to apply electronic treatment of a patient for which safety (e.g.,
leakage currents) has not yet be confirmed.
vi) Power supply:
Voltage: 100 – 120 V AC Frequency: 50/60Hz Input current: 500 VA Power fluctuation: Within ±10%
Frequency fluctuation: Within ±1 Hz
vii) Environment
a. Operating environment
Ambient temperature: 10 – 40°C Relative humidity: 30 – 85% Air pressure: 700 – 1060 hPa
b. Storage environment
Ambient temperature: 25 – 70°C Air pressure: 700 – 1060 hPa
ISSUE1
1-3
4 Specifications
Item Specification
SpecificationCLV-180
1. Videoscopes
2. EUS Videoscopes 1. EUS 130, 140, 160 series videoscopes
4. Light guide for use with rigid endoscopes
5. Rigid endoscopes
(Model names
omitted)
6. High-intensity
1. Type of endoscopes
1. Compatible endoscopes
compatible fiberscopes (for stiff eyepiece endoscopes)
7. High-intensity compatible videoscopes
Illumination light path Emergency lamp
1. Optics
2. Illumination function
1. EVIS 100, 130, 140 series videoscopes
2. EVIS EXERA 145, 160 series videoscopes
3. V series videoscopes
4. EVIS EXERA II 165, 180 series videoscopes
5. VISERA series videoscopes
1. OES 10, 20, 30, 40, 60 series fiberscopes 3. Fiberscopes
2. E, E2, E3 series fiberscopes High-intensity compatible light guides
Note: The high-intensity mode can be set by pressing the
high-intensity mode selection switch when connecting the
light guide. Not compatible with rigid products other than those with high-intensity compatible light guides.
CHF-CB30L, CHF-CB30S, URF-P3 (LF-TP, LF-DP, LF-GP) *1 (ENF-GP) *2 *1 Applicable when used in combination with the following
light guides: A3290, A3291, A3292, A3293, A3294, A3295,
A3296, A3297, A3298. *2 Applicable when using in combination with light guide
A3293. LTF-V, LTF-V2, LTF-V3
A4940A, A4941A, A4942A, A4943A, A4800A, A4801A,
A4802A, A4803A, A4804A, A4805A, A50000A, A50001A,
A50010A, A50011A, A50020A, A50021A Condenser lens
Optical filters Turret board (emergency lamp, filter, etc.) Diaphram Light-adjustment mesh turret
(1) Each of the optical filters on the turret plate are
automatically placed in the light path according to the
normal observation mode (GI/SP), special observation
mode (GI/SP), or high-intensity mode (SP) settings. (2) In accordance with the various observational modes
above, the mesh filter with desired aperture ratio mounted
on the mesh turret can be automatically placed in the light
path. (3) The emergency lamp is automatically placed in the light
path when the examination lamp malfunctions. (4) There are four special filter frames provided to enable the
use of special filters.
ISSUE1
1-4
SpecificationCLV-180
Examination lamp
(1) Type: Xenon short arc lamp with an eliptic mirror (2) Model: MD-631
(3) Life: 500 hours (when lit continuously) (4) The lamp will either illuminate when the power is switched
on (auto-ignition funtion), or it can be turned on by
pressing the LAMP switch on the front panel. The lamp
can be turned off by pressing the LAMP switch
2. Illumination light
Emergency lamp
(1) Type: Halogen 35 W (with reflector)
continuously for one second.
(2) Life: Average of at least 500 hours
Method (1) Manual adjusment only for fiberscopes and rigid-type
endoscopes (excluding when used with a CV-180 and
camera head). (2) Automatic and manual adjustment available for
videoscopes. (3) Automatic adjustment is available when the CV-180 and
camera head are used with rigid endoscopes and
light-source cables.
NBI special observation mode
(1) The filter mode switch enables selection of the NBI special
observation mode when a light adjustment cable and the
CV-180 are connected to an NBI-compatible endoscope. (2) When the NBI special observation mode is selected, the
NBI special filter is placed in the illumination light path. Note: The CV-180 power switch must be ON.
PDD normal observation mode (OE) PDD is option 1 for non-OE regions
(1) PDD normal observation mode will be activated when the
light guide is connected to a light adjustment cable
equipped with the PDD function, and a camera head for
exclusive use during PDD special observation are
connected to the CV-180. (2) During PDD normal observation mode, the PDD normal
filter is placed in the illumination light path.
2. Illumination function
Note: The CV-180 power switch must be ON.
The PDD normal filter and PDD special filter are
required for the PDD function.
ISSUE1
PDD special observation mode (OE) PDD is option 1 for non-OE regions
3. Brightness adjustment
PDD special observation mode (OE) (Fiberscope) PDD is option 1 for non-OE regions
Optional observation mode (OE) PDD is option 2 for non-OE regions
(1) From the PDD normal observation mode, the PDD special
observation mode may be selected by pressing the filter
mode switch. (2) During PDD special observation mode, the PDD special
filter is inserted in the illumination light path. (3) Press the switch again to insert the PDD normal filter. The
filters can be exchanged by pressing the switch. Note: The CV-180 power switch must be ON.
(1) With the CV turned OFF, attach the light-supply cable with
the PDD function. When the light guide is attached,
pressing the filter mode switch will select the PDD special
observation mode. (2) During PDD special observation mode, the PDD special
filter is inserted in the illumination light path.
Note: The normal filter is used for PDD observation during
observation through a fiberscope, rather than the PDD normal filter.
(1) This function is reserved for experimental use.
1-5
SpecificationCLV-180
available through the endoscope distal end when
High-intensity mode (1) When a high-intensity compatible endoscope is inserted, the
high-intensity mode may be selected by pressing the high-intensity switch. High-intensity mode may not be selected with endoscopes that are not high-intensity compatible.
(2) When a high-intensity compatible endoscope is inserted, the
high-intensity mode switch illuminates. This enables the selection of the high-intensity mode (LED above the switch illuminates) or the normal mode (LED above the switch goes out).
(3) The current settings for the high-intensity mode switch are
retained when the power is turned OFF. When a high-intensity compatible endoscope is inserted, illumination will either be the high-intensity mode or normal mode, depending on the settings retained by the system. Initial factory setting: normal mode.
(4) When the high-intensity mode is terminated (normal mode),
3. Brightness adjustment
illumination is equivalent to the CLV-U40 EVIS 100 and OES mode.
(5) High-intensity mode provides approximately twice the amount
of illumination as the SP normal observation mode.
Manual adjustment (1) Mechanical diaphram (set on the panel)
(2) 17 steps
Automatic adjustment (1) Mechanical diaphram with constant illumination control
(image illumination)
(2) 17 steps
2. Illumination function
Front panel display method in special
Observation mode Front panel display method
observation mode
NBI special observation mode
The NBI LED lights up (white)
PDD normal observation mode
PDD special observation mode
The PDD LED lights up (green) (OE only)
The PDD LED lights up (white) (OE only)
Equipment (1) Forced air cooling with a fan (rear exhaust)
4. Cooling
Air supply pump (1) Diaphragm system Air supply pressure (1) Three air supply pressure levels (high, medium, low and
stop)
(2) Maximum air supply pressure: less than 53.9 kPa
Control (1) Through air supply switch settings
1. Air
Water supply (1) Supply is
a water supply container is attached.
3. Air/Water supply
function
2. Water
ISSUE1
1-6
SpecificationCLV-180
switch)
Endoscope and light guide One-touch connection Light-supply cable (MAJ-1411) Connector on the rear-panel
(Light-supply cable for CV connection)
Expansion cable (MAJ-202/972) System connector on the rear-panel
(compatible with Endo GATE/Endo ALPHA)
4. Connection
Foot switch (MAJ-1391) Foot switch connector on the rear-panel Emergency lamp indicator Indicates if the emergency lamp is disconnected or if the
emergency lamp is in use (when the main lamp will not illuminate).
Lit up: Emergency lamp in use (the main lamp will not illuminate).Blinking: Emergency lamp is disconnected or has been removed.
indication
5. Emergency
Setting retention (1) Settings prior to powering off the unit are retained.
Manual / automatic brightness settings Airflow settings Light intensity level High-intensity mode setting Special observation function display
6. Panel
Lamp life indicator (with the power ON, the indicator may
be reset when the examination lamp is extinguished by continuously pressing the counter reset switch.)
Switch illumination system (not including counter reset
On endoscope removal (1) When the endoscope is removed, the light emanating from
1. Dimming
Exterior
2. Disinfection
Temperature switch (1) To ensure safety, when the internal temperature rises
3. Alarm
7. Safety
Protection Class 1 unit (3P power supply)
(1) All switches illuminated (not including counter reset switch)
the scope connector is dimmed by the dimmer plate located in the endoscope socket.
(1) Disinfect with ethanol for disinfection (70 ethyl or
isopropyl alcohol) or with a mild detergent.
above the specified value, the temperature switch turns off and the electrical current supply to the lamp is terminated. The unit detects that the temperature switch has shut off, and sounds an alarm.
ISSUE1
4. Electrical
shock
Region
1. EU/ EFTA Y
2. Japan N
regulations
5. Applicable
3. USA Y
Sold
MDD Class:II a (rule: 11-1) CE marking: CEO197
Pharmaceutical Law (Application of the Pharmaceutical law to assist Asian market response)
FDC laws
1-7
SpecificationCLV-180
4. Canada Y
5. Singapore Y
6. Other Y Common to all
countries
Class: II Singaporean medical device regulations
K-FDA (Korea) SDA (China) IEC 60601-1 (Medical electrical equipment - Part 1: General requirements for safety): 1988 + A1, A2 IEC 60601-1-1 (Medical electrical equipment - Part 1-1: General requirements for safety): 2000 IEC 60601-1-2 (Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility
- requirements and tests): 2001 IEC 60601-2-18 (Medical electrical equipment - Part 2: Particular requirements for the safety of endoscopic equipment): 1996 + A1 ISO 14971(Medical devices -- Risk management): 2000 ISO 9000-3(Software engineering -- Guidelines for the application of ISO 9001:2000 to computer software): 1997 ISO 8600-1(Optics and photonics -- Medical endoscopes and endotherapy devices -- Part 1: General requirements): 1997 ISO 7000(Graphical symbols for use on equipment) : 2004 IEC 60417-1: 2002, -2: 1998(Graphical symbols for use on equipment)+ A1, A2
Canadian Medical Device and Equipment Regulation
EUEFTA
EN 980 (Graphical symbols for use on equipment): 2003
2. Japan JIS T1005 (Style manual for instruction manuals): 1983
JIS T0601-1 (Medical electrical equipment -- Part 1: General requirements for safety): 1999
6. Applicable external regulations
JIS T0601-1-1(Medical electrical equipment -- Part 1-1: General requirements for safety -- Collateral standard: Safety requirements for medical electrical systems): 1999 JIS T0601-1-2(Medical electrical equipment -- Part 1: General requirements for safety -- 2. Collateral standard: Electromagnetic compatibility -- Requirements and tests): 2002
3. USA UL 60601-1: 2003
4. Canada CAN/CSA-C22.2 No.601.1-M90: 1990
CAN/CSA-C22.2 No.601.1-S1: 1994
5. Other Olympus Eco-Product Standard Ambient temperature 10 – 40°C
Relative humidity: 30 – 85% Air pressure 700 to 1060hPa
Usage environment Do not use in a combustible atmosphere Type of protection against
electrical shocks
8. Other
However, this instrument may be applied to the heart if it is used with equipment classified as TYPE CF applied part. (indicated by the symbol)
Class 1 Degree of protection: Type BF applied part
Voltage 100 – 240 V AC
1. Usage restrictions
Frequency 50/60 Hz Power fluctuation
Frequency fluctuation
Within ±10% Within ±1 Hz
Power consumption 500VA
1-8
ISSUE1
SpecificationCLV-180
User servicing Examination lamp, fuses
OLYMPUS servicing Emergency lamp replacement, special filter attachment
2. Par need replacment
Weight Dimensions 383 (W) × 536 (D) × 162 (D) mm (maximum)
Panel Panel selection varies by region
3. Other
Power cord Cord set with a 3-prong hospital grade
Fuse Fuse capacity
Warranty period As specified in the applicable laws for each country
5 Nomenclature and Functions
5-1 Front panel
Approximately 15.4
370 () × 475 (D) × 150 (D) mm (standard)
AC 100 - 240 English (3E, 6E, 6LA) AC 100 - 240 Symbol (6S) AC 100 - 120 English (3OA) AC 100 English (1J)
With plug (1J, 3OA, 3E) Plug-less (6E, 6S,6LA)
Littel (manufacturer) 215008: 8A
ISSUE1
1-9
SpecificationCLV-180
5-2 Control panel (buttons)
ISSUE1
1-10
SpecificationCLV-180
(1) Transillumination button When pressing this button, light emitted from the endoscopes distal end becomes brighter for 7
seconds, then returns automatically to its original brightness level.
(2) Intensity mode selector Press to switch between the high-intensity mode and normal intensity mode when using an
endoscope compatible with the high-intensity mode.
(3) Filter mode button Pressing this button activates the NBI observation function.
(4) Air feed button Pressing this button starts or stops the air feed from the endoscopes distal end.
(5) Airflow regulator button This button is use to control the pressure of the air being fed from the endoscope.
(6) Brightness adjustment buttons These buttons are pressed to adjust the brightness level.
(7) Auto/manual brightness selector This selector is pressed to select automatic or manual brightness control.
(8) Lamp button This button is pressed to turn ON or OFF the examination (xenon) lamp.
(9) Counter reset button After replacing the examination (xenon) lamp, the lamp hour meter is reset by pressing this button
for 1.
ISSUE1
1-11
5-3 Control panel (Indicators)
SpecificationCLV-180
ISSUE1
1-12
SpecificationCLV-180
(1) Transillumination indicator This indicator lights when the transillumination function is activated.
(2) High intensity mode indicator The indicator lights when high-intensity mode is selected.
(3) Air feed ON/OFF indicator This indicator lights when the air feed from the endoscope’s distal end is activated.
(4) NBI observation mode indicator This indicator lights in green to indicate that the light source can perform the NBI observation. This indicator lights in green continuously when an endoscope with the NBI observation
compatibility is connected and the light source is ready for the NBI observation. Also, the indicator lights in white when NBI observation mode is active.
The "OP.1" and "OP.2" are prepared for future use. (5) Airflow regulator indicators
One of these indicators lights to indicate the current airflow pressure level setting ("L" (Low), "M"
(Medium) or "H" (High)).
(6) Emergency lamp indicator This indicator lights when the emergency lamp (halogen) is in use, and blinks when the emergency
lamp (halogen) is disconnected or not mounted.
(7) Brightness level indicators These indicators display the current brightness level.
(8) Auto/manual indicator This indicator displays the brightness selector setting ("auto" or "manual").
(9) Lamp ON/OFF indicator This indicator lights when the examination lamp (xenon lamp bulb) lights.
(10) Lamp hour meter This indicator displays the total working hours of the examination (xenon) lamp.
ISSUE1
1-13
5-4 Rear and side panels
SpecificationCLV-180
1-14
ISSUE1
SpecificationCLV-180
(1) Fuse box The fuses protect the light source from electrical surges.
(2) SYSTEM terminal The terminal accepts connection from an external unit.
(3) CV terminal This terminal is the receptacle for the light source cable to connect the light source to the EVIS
EXERA II video system center.
(4) Lamp ignition selector This selector is set to select automatic or manual ignition of the examination lamp. When the
automatic ignition is selected, turning ON the light source lights the examination lamp simultaneously. When the manual ignition is selected, pushing the lamp button on the control panel lights the examination lamp.
(5) Foot switch terminal This connector is the receptacle for the foot switch (MAJ-1391) for use in the naked-eye PDD
observation.
(6) Water container holder This holder is used for the installation of the water container.
(7) Potential equalization terminal For safety, this terminal is connected to a potential equalization terminal of the other equipment
connected with the light source, and the electric potential of their equipment are made the same.
(8) Lamp cover This cover has to be removed to replace the examination lamp.
ISSUE1
1-15
Chapter 2: Troubleshooting
2-1
ISSUE1

TroubleshootingCLV-180

TroubleshootingCLV-180
1.
Contents
Symptom Failure Mode 2-1 No power Power input failure
2-2 Xenon Lamp not lit Lamp failure 2-3 Lamp going out Lamp going out 2-4 Scope not mounting Scope connection failure 2-5 Incapability of discriminating GI, SP, or high-brightness
SP <GI Scope> AIR button or Transillumination button cannot be selected when GI Scope is connected. <SP>
AIR button, Transillumination button, and high-brightness button can be selected when SP
Scope is connected. <High-brightness SP> High-brightness button cannot be selected when high-brightness Scope and Light Guide are connected.
2-6 Cooling fan not working Cooling fan malfunction 2-7 Exit light from Light Source when Scope is removed Shield failure at scope removal 2-8 Pump not working when Pump ON switch is pushed, or
no change in air supply volume
2-9 Turret Board not rotating when Special Light
Observation Scope is connected and Mode switch is pushed
2-10 No light from Scope Exit light failure 2-11 View area is dark or excessively light Manual Brightness failure 2-12 View area is dark or excessively light Automatic Brightness failure 2-13 No change in LED display when switch is pushed Panel malfunction 2-14 Emergency Lamp is not automatically lit when Xenon
Lamp is put out
2-15 Values set before shutdown are not saved when power
shuts down
2-16 Time not changing to "0" when Lamp Life Meter Reset
switch is pushed
2-17 No change in light quantity when Trans Illumination
switch is pushed
Scope discrimination failure
Pump malfunction
Filter changing malfunction
Emergency Lamp malfunction Back up malfunction Lamp Life Meter Reset malfunction Transillumination malfunction
2-2
ISSUE1
2. Troubleshooting CLV-180
2-1 Power input failure
Incapable to input power 2-1-1 Block diagram
TroubleshootingCLV-180
2-1-2 Estimated location of failure No
Estimated failure location Inspection method
Input voltage 1 Verify the voltage of wall socket.
* Conformity with commercial power voltage standard.
Power Cable 1 Connect the Power Cable to wall socket and verify voltage at inlet side.
* Conformity with commercial power voltage standard.
Body Inlet Plug 1 Confirm that the inlet plug is not bend or distorted.
Visual inspection
Fuse 1 Confirm that the fuse was not burned out.
Power Switch
Interlock Switch
Flat Cable 1 Verify connection of the Flat Cable
1 Confirm that the Power Switch inside the equipment is pushed when
the Main Switch on the Front Panel is pushed.
Visual Inspection
2 Confirm conductivity of the Power Switch when power is ON.
1 Confirm that the switch is pushed by the Lamp Door.
Visual Inspection
2 Confirm conductivity of the Interlock Switch when power is ON.
Visual Inspection
Main Board (J12) <--> Front Panel
Tester
Tester
Tester
Tester
Tester
Interlock Jig
2-3
ISSUE1
TroubleshootingCLV-180
Main Board
Converter
Power Switch LED 1
Front Panel Unit 1
2-2 Lamp failure
Xenon Lamp of light source not lit 2-2-1 Block diagram
Confirm that the input voltage to Main Board J16 is correct.
1
J16 (Main Board) <- -> Converter) 1pin : DC +15V±0.75V 2pin : DC +12V±0.6V 3pin : DC + 5V±0.25V 4pin : GND
5pin : DC -15V±0.75V YES Replace Main Board NO→ Replace Converter Confirm that the Power Switch LED is lit: NO Replace the Power Switch LED Confirm that the Front Panel of LED is lit: NO Replace the Front Panel
Tester
2-2-2 Estimated location of failure No Estimated failure location Inspection method
Power input failure Refer to 2-1
Emergency Lamp Malfunction Refer to 2-16
Panel malfunction Refer to 2-13
Thermo Switch /
Main Board
ISSUE1
Status of Thermo Switch when buzzer is sounding
1
Thermo Switch Less than 85 Conductive 85 or higher Non-conductive Within standard -> Replace Main Board Out of standard -> Replace Thermo Switch
2-4
Main Board 1
Surrounding the Lamp
Xenon Lamp /
Igniter
Converter
1 Verify Lamp attachment.
2 Check the area surrounding the Lamp House.
1
1 Xenon Lamp not lit even after procedures 1-7
Confirm that the signals in the Main Board are correct:
Main Board J6/2Pin HIGH : XENON Lamp lit*HIGH : approx. 5 V LOW : XENON Lamp not lit
H -> L when the Ignition Button is pushed ON to turn on Lamp
L -> H when the Ignition Button is pushed OFF to turn off Lamp (Long push)
NO →Replace Main Board
Confirm that Lamp is attached properly. Confirm that heat compound is applied properly. Confirm that the area surrounding the Xenon lamp contacts is clean.
* Cleaning is required if the discoloration is found.
No unnecessary objects (e.g. metal chip) No leakage of high voltage pulse Check connection with Terminal F and Terminal R
Confirm there is enough distance between Round Terminal and Lamp
House Clattering noise at ignition YES →Replace Xenon Lamp NO →Igniter
YES →Replace Converter
2-3 Lamp malfunction
Lamp going out 2-3-1 Estimated location of failure
No
③ ④ ⑤
Estimated location failure Inspection method
Log 1 Verify frequency of Lamp failure errors.
* Refer to EVIS communication Checker for verification method.
CLV-180 Communication Cable Power input failure Refer to 2-1 Emergency Lamp malfunction Refer to 2-16 Cooling Malfunction Refer to 2-6 Lamp failure Refer to 2-2
2-4 Scope connection failure
Scope not mounting 2-4-1 Estimated location of failure
No
① ②
ISSUE1
Estimated location failure Inspection Metod
Scope 1 Confirm that Scope is a compatible one. Scope Socket 1 Confirm that the Scope Socket is free of abnormalities.
* Insert Heat Cover Positioning Jig to Scope Socket and confirm that it's not caught inside.
2-5
TroubleshootingCLV-180
Evis communication checker
Heat Cover Positioning Jig
2-5 High-brightness malfunction
Not changing to high-brightness function mode 2-5-1 Block diagram
TroubleshootingCLV-180
2-5-2 Estimated location of malfunction
No
① ② ③
Estimated location failure Inspection Method
High-brightness scope 1 Confirm that scope is compatible for high-brightness. Panel malfunction Refer to 2-13 Scope Sensor /
Main Board
Confirm following status of J15 connector on Main Board when
1
high-brightness scope is connected: Main Board J15
When high-brightness scope is connected Pin 2: HIGH Pin 5: LOW Pin 8: LOW Pin 9: GND
* HIGH: approx. 5 V YES Replace Main Board NO Replace Scope Sensor
Tester
2-6
ISSUE1
Loading...
+ 56 hidden pages