FEATURES
Intended Use
Life Scope G5/G7 is intended for continuous
monitoring, recording, and alarming of multiple
physiological parameters of adults, pediatrics
and neonates in the OR, recovery room, ICU,
CCU, HCU, NICU, ER, ward and other areas.
Several types of input unit are available. See the
chart in the following page.
For CSM-1500, an input unit can be directly
connected to the main unit and can be used as
an integrated monitor.
For CSM-1700, an input unit can be connected to
the DAU (Data Acquisition Unit) and the input
part can be installed away from the main unit.
Transport function
Life Scope G5/G7 supports transport. Patient
information and review data can be transferred to
another bed by disconnecting the input unit from
the source monitor and connecting it to the
destination monitor. (Transport function requires
an optional QM-600P memory unit in the input
unit.)
A Life Scope PT BSM-1700 series transport
monitor can also be used as an input unit for the
Life Scope G5/G7 bedside monitor.
After connecting the input unit to the destination
bedside monitor in the network, the data in the
input unit is automatically transferred to the
bedside monitor and central monitor. All patient
information, including trend and waveforms, is
automatically transferred to maintain one
seamless patient record.
Nihon Kohden's networking technology enables
seamless monitoring across different central
monitors. You can see continuous past data even
from a different central monitor at a different site.
Up to 11 MULTI connectors are available.
Multilink connectors
A multilink connector receives data from an
external device such as an anesthesia machine
or ventilator and the monitor displays the
external device data with other monitoring data.
Number of Multilink connectors:
CU-151R/152R/171R/171R core unit: 2
JA-170P data aquisiton unit: 3
QI-151P rear option for CU-151R/152R: 2
QI-171P rear option for CU-171R/172R: 2
QI-152P side option for
CU-152R/171R/172R: 2
User Interface
Operation
Life Scope G5/G7 can be operated by touch
screen, mouse, remote control and bar code
scanner.
Touch screen operation
The touch screen provides intuitive operation.
Review while monitoring
The G-Scope function lets you review past data
without hiding the current vital signs and
waveforms. Just flick the side or bottom of the
screen to access three pre-assigned review
windows.
Wireless remote control
You can operate Life Scope G5/G7 by an
RY-910PA wireless remote control. The remote
control has six function keys to call up preset
windows. You can turn monitor on or off and
alarm on or off by a wireless remote control.
One remote control can operate up to 9 monitors
by changing the channel.
Smart cables and MULTI connectors
Smart cable technology lets you measure
different parameters without the need for
dedicated modules. When you plug a smart cable
into a MULTI connector, the monitor
automatically detects the type of parameter and
starts measuring.
Basic parameters of ECG, respiration, SpO2,
NIBP and temperature can be measured with
dedicated connectors. Other parameters can be
measured with MULTI connectors.
4
Input units
Model Description
Number
of MULTI
connector
Available parameters on
MULTI connector
AY-660P Input unit
1 IBP, CO2(mainstream)
AY-661P Input unit
1
IBP, CO2, TEMP, CO, BIS,
SpO2-2, TOF
AY-651P
(Nellcor
SpO2)
AY-631P
(Masimo
SpO2)
Input unit
1
IBP, CO2, TEMP, CO, BIS,
TOF
AY-663P Input unit
3
IBP, CO2, TEMP, CO, BIS,
SpO2-2, TOF
AY-653P
(Nellcor
SpO2)
AY-633P
(Masimo
SpO2)
Input unit
3
IBP, CO2, TEMP, CO, BIS,
TOF
AA-672P Smart unit
2
IBP, CO2, TEMP, CO, BIS,
SpO2-2, TOF
* SpO2-2 is only available in
the NK SpO2 configuration.
AA-674P Smart unit
4
AA-174P Multi amp unit
4
Note: AY-660P is not available in Europe.
5
Optimized for each site and
each specialist
OR:
Triple display
Life Scope G5/G7 allows three independent
displays. Different displays can provide different
information for the anesthesiologist, physician
and heart-lung machine operator or other
specialist. (CU-152R/171R/172R with QI-152P)
Multi-gas measurement
You can measure CO2, N2O, O2, and five
anesthetic agents (Halothane, Isoflurane,
Enflurane, Sevoflurane and Desflurane). Gases
can be measured quickly with small sampling
flow rate. (Requires an optional GF-210R
multi-gas unit or GF-220R multi-gas/flow unit.)
MAC value
LifeScope G5/G7 can calculate MAC (Minimum
Alveolar Concentration) which helps predict how
much anesthetic will be required during surgery.
(Requires an optional GF-210R multi-gas unit or
GF-220R multi-gas/flow unit.)
more intuitive approach to diagnostic and
therapeutic decision making in hemodynamic
management.
A target graph for PPV and esCCO provide
minimally invasive hemodynamic monitoring.
Life Scope G5/G7 also has the target graph
based on Forrester Classification to support the
intermittent invasive parameters such as cardiac
output by bolus thermodilution and pulmonary
wedge pressure.
Bypass function (All alarms off)
You can turn all alarms off indefinitely when a
heart-lung machine is connected to the patient.
Neuromuscular transmission
For customers who need an objective way to
monitor the dosage of relaxants for patients for a
safe anesthesia, surgery and extubation, the
NIHON KOHDEN NMT module AF-101P provides
patient focused handling with operation buttons
directly at the patient near module supporting all
common NMT testing methods.
Life Scope G5/G7 shows TOF ratio, TOF counts
and PTC on the display.
Respiratory loops
PV loop and FV loop can be displayed. This
helps respiration management of patients
connected to an anesthesia machine. (Requires
an optional GF-220R multi-gas/flow unit.)
BIS monitoring
BIS monitoring helps clinicians determine and
administer the precise amount of anesthesia
drug to meet the needs of each individual patient.
PPV/SPV
PPV (Pulse Pressure Variability) and SPV
(Systolic Pressure Variability) is an indicator of
intravascular volume. It is useful in guiding fluid
therapy for patients receiving mechanical
ventilation.
esCCO
The optional QL-005P esCCO software/ QL-006P
esCCO/Hemodynamics review program enable
esCCO measurement. esCCO is a new
technology to calculate cardiac output
non-invasively and continuously. esCCO uses
PWTT (Pulse Wave Transit Time) which is
obtained from the SpO2 and ECG-signals. It can
continuously display esCCO, esCCI, esSV,
esSVI, esSVR and es SVRI.
Hemodynamics Review Program
The optional QL-006P esCCO/Hemodynamics
Review Program provides trend display to
support advanced intensive therapy
management.
A hemodynamics graph provides a graphic view
of overall hemodynamic information. It allows a
ICU:
PPV/SPV
PPV (Pulse Pressure Variability) and SPV
(Systolic Pressure Variability) is an indicator of
intravascular volume. It is useful in guiding fluid
therapy for patients receiving mechanical
ventilation.
esCCO
The optional QL-005P esCCO software/ QL-006P
esCCO/Hemodynamics review program enable
esCCO measurement. esCCO is a new
technology to calculate cardiac output
non-invasively and continuously. esCCO uses
PWTT (Pulse Wave Transit Time) which is
obtained from the SpO2 and ECG-signals. It can
continuously display esCCO, esCCI, esSV,
esSVI, esSVR and es SVRI.
CVP_ET
Respiration affects the intrapleural pressure and
makes the CVP value unstable. Nihon Kohden’s
original technology of CVP_ET (end tidal CVP)
uses the mainstream CO2 signals to calculate the
CVP at the end tidal CO2 point. CVP_ET provides
a stable and physiologically correct CVP value.
Hemodynamics Review Program
The optional QL-006P esCCO/Hemodynamics
Review Program provides calculation and trend
display to support advanced intensive therapy
management.
A hemodynamics graph provides a graphic view
of overall hemodynamic information. It allows a
more intuitive approach to diagnostic and
6
therapeutic decision making in hemodynamic
management.
Mainstream EtCO2 for both intubated and
non-intubated patients
Life Scope G5/G7 can use the compact and
lightweight TG-980P & TG-920P cap-ONE
mainstream CO2 sensor kit to obtain smooth and
accurate EtCO2 and respiration rate within 5 or
10 seconds. Unique nasal adapter effectively
catches the nasal and oral expired CO2 without
dilution for non-intubated patient. cap-ONE mask
provides ETCO2 measurement and oxygen
supply for non-intubated patients.
8-channel EEG
EEG monitoring is available with the optional
AE-918P neuro unit. Up to 8 channels of EEG
can be monitored.
CSA/DSA
CSA and DSA show the power in each EEG
frequency band plotted over time. CSA and DSA
can help spot EEG trends and changes in brain
function which might not be obvious from the raw
waveforms.
CCU:
12 lead ECG simultaneous display
12-lead ECG waveform can be displayed on the
home screen. This makes it easier to recognize
multi-focus VPC, ST change and ventricular
hypertrophy.
12-lead ECG analysis
The 12-lead ECG analysis of Life Scope G5/G7
has the same accuracy and reliability as a
dedicated electrocardiograph. There is no need
to change electrodes to another instrument.
Arrhythmia detection
Upgraded arrhythmia detection lets you analyze
multi-channel ECG. The multi-template matching
method reduces false alarms and assures high
quality detection. 25 arrhythmia items can be
detected.
ST display and review
ST waveforms of all ECG leads can be displayed.
The reference ST waveforms are also displayed
so you can see changes in the ST waves. You
can adjust the ISO point, J point and ST point for
ST level measurement.
ST review shows the ST level on a multi-axis
chart. This helps the clinicians easily recognize
ST change and location in the heart.
NICU:
Neonate arrhythmia algorithm
Life Scope G5/G7 has software for monitoring
neonates. A dedicated neonate arrhythmia
recognition algorithm recognizes narrow QRS.
NIBP inflation pressure for neonate
Connection of a neonate NIBP air hose is
detected and the inflation pressure range is
automatically changed for patient safety.
OCRG
OCRG (oxycardiorespirogram) combines
compressed trends of beat-to-beat heart rate,
respiration, and oxygenation levels. OCRG can
help doctors detect the cause of apnea attack.
Dual SpO2
Monitoring the difference in SpO2 between right
upper limb (pre-ductal) and lower limb
(post-ductal) is useful to diagnose PPHN
(Persistent Pulmonary Hypertension of Newborn),
decide the timing to start and stop treatment as
well as the operation schedule.
aEEG (amplitude-integrated EEG)
Because the brain of a newborn baby is
immature, it is difficult to detect seizures by
observation. Life Scope G5/G7 provides aEEG
which makes it easier to detect brain seizure.
aEEG monitoring is also becoming important in
hypothermia therapy for neonatal
encephalopathy.
Mainstream CO2 sensors for neonate
Mainstream CO2 monitoring for neonate is
available with a cap-ONE TG-980P CO2 sensor
kit. The airway adapter YG-213T has a dead
space of only 0.5 ml.
PI (Pulse-amplitude Index)
The Pulse-amplitude Index indicates the
percentage of pulsatile signal in the entire
transmitted IR signal. PI supports to assess
peripheral perfusion.
(PI is displayed when using Nihon Kohden SpO2
sensors and Masimo sensors)
SQI bar graph for SpO2
The SQI (signal quality index) bar graph shows
the pulse waveform quality for SpO2
measurement. (SQI bar graph is available only
with Nihon Kohden SpO2)
Non-adhesive SpO2 probe
The TL-535U (P206) and TL-260T (P205A)
Multi-site Y probe are designed so that the
adhesive of the attachment tape does not stick to
the delicate skin of neonates or low birth weight
infants.
Network
LS-NET (Life Scope Networking)
You can connect Nihon Kohden central monitors
and bedside monitors by LS-NET (Life Scope
Networking with Ethernet LAN).
7
Interbed
When Life Scope bedside monitors are
connected in an LS-NET network, you can
exchange and view data of other bedside
monitors. Both individual display and 20 bed
display are available.
Connecting to a network printer without a central
monitor
You can connect a network printer to a Life
Scope G5/G7 monitor on the LS-NET through the
LAN port.
You can print real time and review data on A4
papers.
Other Features
Screen capture
You can save Life Scope G5/G7 screen capture
images and transfer PNG data by USB.
This is helpful for preparing clinical reports.
*QM-150P is required.
iNIBP
Nihon Kohden's unique iNIBP inflation algorithm
provides fast and painless measurement of NIBP.
iNIBP is available.
Nihon Kohden’s YAWARA CUFF 2 NIBP cuffs
prevent bruising, increase patient comfort and
reduce noise for more accurate NIBP
measurement.
PWTT-triggered NIBP measurement
PWTT (pulse wave transit time) is continuously
and non-invasively measured from the ECG and
SpO2. If a sudden, critical circulation change
happens between the periodic NIBP
measurements, PWTT may detect it and trigger
NIBP measurement to confirm it.
Respiration calculation
Life Scope G5/G7 can calculate respiration
dynamics, the flow rates and dosages for
medication titration. This is the ideal solution for
medical and paramedical personnel who need
rapid dose calculation. In addition to preset
drugs, you can add custom medications.
Graphical and tabular trends
Up to 72 hours of graphic and tabular trends of
all parameters can be saved and reviewed.
*When transport function is ON, up to 24 hours.
Full disclosure
Up to 72 hours of full disclosure waveforms for 5
parameters can be saved and reviewed.
*When transport function is ON, up to 24 hours.
Data time is synchronized between review windows
When one review window is changed to another
review window, the new review window open in
the same time as the first review window. For
example, to check the waveform information for
the time of an alarm file in the Alarm History
window, select the corresponding file and change
the full disclosure window. The full disclosure
waveform for the corresponding time is displayed.
This makes it easy to compare various vital sign
data for the patient at the selected time between
the review windows.
Alarm escalation
The alarm escalation function automatically
escalates the alarm priority to a selected level if
the alarm continues or SpO2 drops. Alarm
escalation applies to vital alarms (SpO2 limit,
apnea and high pressure system IBP) and
technical alarms (ECG and SpO2). Alarm
escalation can contribute to decreased medical
accidents and improved quality of care.
Sleep mode
Sleep mode prevents the monitor from disturbing
the patient during sleep or other times. The
screen is darkened and the sync sound and
alarm indicator are turned off. Sleep mode is
available when the Life Scope G5/G7 bedside
monitor is connected to the central monitor.
Drug calculation
Life Scope G5/G7 can calculate the flow rates
and dosages for medication titration. This is the
ideal solution for medical and paramedical
personnel who need rapid dose calculation. In
addition to preset drugs, you can add custom
medications.
Synthesized 18-lead ECG (synECi18)
Synthesized 18-lead ECG derives the waveforms of
the right chest leads (V3R, V4R, and V5R) and back
leads (V7, V8, and V9) from the standard 12-lead
ECG data.
The measurement procedure is the same as the
standard 12-lead ECG but more information can be
obtained.
18-lead synthesized ECG is expected to be useful in
detecting right side and posterior infarction.
Optional software QL-004P is required.
Heart rate, VPC rate, QTc, QRSd, ST level,
respiration rate, NIBP (Sys, Dia, MAP), IBP (Sys,
Dia, Mean), PPV, SPV, SpO2, SpO2-2, delta SpO2,
pulse rate, PI, temperature, CO, CI, Ti (injectate
temperature), Tb (blood temperature), O2
concentration, EtCO2, BIS, inspired/ expired N2O
concentration, inspired/ expired CO2 partial pressure,
inspired/ expired CO2 concentration, inspired/ expired
O2 concentration, inspired/ expired anesthetic agent
concentration (Halothane, Isoflurane, Enflurane,
Sevoflurane, Desflurane), MAC (minimum
alveolar concentration), Ppeak (peak airway
pressure), PEEP (positive end expiratory
pressure), Pmean (mean airway pressure), MV
(minute volume), TVi (inspiratory tidal
volume),TVe (expiratory tidal volume), C
(compliance), R (airway resistance), Ri
(inspiratory airway resistance), Re (expiratory
airway resistance), I:E (inspiration expiration
ratio), SEF (90 or 95% spectral edge frequency),
MDF (median frequency), PPF(peak power
frequency), TP (total power), power of frequency
(Absδ, Absθ, Absα, Absβ, Absγ), power ratio of
frequency (%δ, %θ, %α, %β, %γ), CSA
(compressed spectral array), DSA (density
spectral array), TOF cnt, TOF rat, Twitch height
(Tw1 to Tw4), double burst stimulation (DBS 3.2,
DBS 3.3), tetanic stimulation (TET50, TET100),
post tetanic stimulation (PTC), temperature
(Temp, measured by the AF-101P NMT module),
With AP-170P hemodynamic unit;
CO, CVP, ScvO2, Hb, SaO2, PR, A Psys, A Pdia,
MAP, PCCO, PCCI, dPmax, SV, SVI, SVR, SVRI,
SVV, PPV, CPO, CPI, GEDV, GEDVI, ITBV,
ITBVI, EVLW, EVLWI, PVPI, GEF, CFI, CI, DO2,
DO2I, VO2, VO2I, O2ER
With INVOS monitor;
rSO2, SSI indicator, baseline (BL), blood volume
index (BVI), area under the curve (AUC), change
rate,
With CCO monitor;
CCO, CCI, SvO2, ScvO2, Tb, SV, SVI, SVV, HR,
SVR, SVRI, RVEF, EDV, EDVI, ESV, ESVI, DO2,
VO2, O2EI, SaO2, HRV, CF
With PiCCO monitor;
PCCO, PCCI, SV, SVI, SVR, SVRI, SVV, PPV,
dPmax, PR, AP, CO, CI, GEDV, GEDVI, ITBV,
ITBVI, GEF, CFI, EVLW, EVLWI, PVPI, TD Time,
CPO, CPI, ScvO2, SpO2, DO2, DO2I, VO2, VO2I,
9
O2ER, R15
With tcPO2/tcPCO2 monitor;
tcPO2, tcPCO2
With esCCO option;
esCCO, esCCI, esSV, esSVI, esSVR, esSVRI
Position of numeric data: Selectable (Left or right)
Number of screen layouts: 3 for each display (A layout
can be changed with one touch by function key).
Patient name display on the 2nd, 3rd display: Selectable
(On or OFF)
◆
Alarm
Alarm items: Upper/lower alarms, arrhythmia
alarms, interbed alarms, technical alarms
Alarm levels: Crisis (red, blinking), Warning
(yellow, blinking), Advisory (yellow or blue, lit)
Alarm indication: Message, highlighted numeric
value, blinking alarm indicator, alarm sound
Alarm suspend: Provided (for 1, 2, 3 min, Off)
When connected to a central monitor, this alarm
can be adjusted from the central monitor.
Alarm escalation: SpO2 lower, SpO2-2 lower,
press high pressure system upper/lower/mean,
technical alarm (check electrodes, cannot
analyze, check SpO2 probe, cannot detect pulse)
Alarm trigger delay: Available
Alarm configurable start-up values:
Master Alarm Neonatal, Master Alarm Pediatric,
Master Alarm Adult
Number of master setting:
Up to 4 for each of Adult, Pediatric and Neonatal
mode.
Alarm limit setting: Displayed for each
parameter on the monitoring screen.
Auto setting: Upper/Lower alarm, ST level
Interbed alarm setting: All, Crisis & Warning,
Crisis, Off
◆
Power requirement
AC: 100 to 240 V
DC (battery): 10.8 V
Line frequency: 50 or 60 Hz
◆
Battery
Battery pack SB-950P for CU-151R/152R
Application: For continuously monitoring a patient
when moving the patient to a different location or
when there is a power failure or sudden power
interruption.
Type of battery: Lithium ion
Number of batteries: 2
Battery operation time (with 2 battery packs):
CSM-1501: 5 hours
CSM-1502: 4 hours
The above information is a guideline when all the
following conditions are fulfilled and ECG and
respiration (impedance) are monitored.
- New and fully charged battery pack
- Used at room temperature (around 25 °C )
- Recorder module is not operating
- No alarm occurs
Battery pack SB-970P for CU-171R/172R
Application: For preventing data from losing when
there is a power failure or sudden power
interruption.
Type of battery: Lithium ion
Number of batteries: 1
Battery operation time: Three minutes
◆
Interface on core unit
CU-151R/152R/171R/172R core unit
Network: 1
USB socket: 3
HDMI: 1
Serial socket: 1
Multi-link socket: 2
CU-171R/172R core unit
Network: 1
USB socket: 3
HDMI: 1
Serial socket: 1
Multi-link socket: 2
Recorder socket: 1
DVI socket: 1
Display socket: 1
Multi-link socket: 2
USB socket: 4
Network socket: 1
◆
Recording
Recording mode:
- Manual: Real time waveform recording,
recording on the review, analysis and
calculation windows
- Automatic: Alarm recording, periodic
recording
Waveform recording:
- Recording width: ≥ 46mm
- Recording speed: 12.5, 25, 50 mm/s
- Number of traces: Maximum 3 traces
- Recorder items: Date and time, patient
information, measurement values, causes of
arrhythmia message recording, waveform
name, filter on/off, sensitivity, speed
◆
Network
Network communication (LS-NET):
Connects to the Nihon Kohden monitor network
and communicates with central monitor, other
bedside monitors and server.
10
Network communication (hospital network):
Connects to the hospital network (HL7 output).
◆
Review
Trend graph:
Storage capacity: 72 hours
Type of trend graph: 3 trend graphs, CSA, DSA
Number of parameters in each trend graph: Up
to 9
Number of channels for CSA/DSA: Up to 8
Vital sign list:
Storage capacity: 72 hours
Number of vital sign list: 3
Number of parameters displayed in each list: Up to
15
List interval: 1, 5, 10, 15, 30 or 60 min
NIBP list:
Storage capacity: 1,024 items
Number of parameters displayed in the NIBP list:
Up to 15
Hemodynamics list:
Storage capacity: 1,024 items
Lung function list:
Storage capacity: 128 items
Full disclosure:
Storage capacity: 72 hours (5 waveforms)
Display time for 1 trace: 60, 30, 20, 12 or 6 seconds
Zoom in window (displays a full disclosure a
waveform with 1 trace of 6 seconds, and additional
information): Available
ST recall:
Storage capacity: 4,320 items + 1 reference file
Storage interval: 1 minute
Item: ST level waveforms of all monitoring ECG
ST review (ST level on multi-axis chart): Available
Storage capacity: 16,384 items
Filters for displayed items: Alarm levels, Arrhythmia
types
Event graph window (displays all arrhythmia events
in the past 72 hours): Available
ECG analysis results:
Storage capacity: 18 items
Waveforms: standard 12 lead and 6 extra
leads*synthesized from standard 12-lead ECG
Analysis results as a graphic: Available
*Synthesized 6 extra leads requires QL-004P
Hemodynamic graph*:
Storage capacity: 72 hours (requires QL-006P)
Number of target graph (the blood circulation of the
patient is displayed visually using X-Y graph): 4
aEEG review:
Storage capacity: 72 hours
Display channel: 2ch, 4ch
Displayed traces: aEEG (1 to 8), DSA (1 to 8), HR,
SpO2
OCRG review:
Storage capacity: 72 hours
Short trend:
The latest 30 minute parameter data can be
displayed as a trend graph on the home screen.
This short trend graph can be dragged left or
right by touch panel operation.
◆
ECG
Leads:
3-electrode cable: I, II, III
6-electrode cable: I, II, III, aVR, aVL, aVF, 2 from
V1 to V6
10-electrode cable: I, II, III, aVR, aVL, aVF, V1 to V6
Frequency response:
DIAG mode: 0.05 to 150 Hz (–3 dB)
ST mode: 0.05 to 18 Hz (–3 dB)
MONITOR mode: 0.3 to 40 Hz (–3 dB)
MAXIMUM mode: 1 to 18 Hz (–3 dB)
25 items: ASYSTOLE, VF, VT, EXT TACHY, EXT
BRADY, VPC RUN, V BRADY, SV TACHY,
TACHYCARDIA, BRADYCARDIA, PAUSE,
COUPLET, EARLY VPC, MULTIFORM, V
RHYTHM, BIGEMINY, TRIGEMINY, FREQ VPC,
VPC, A-Fib, End A-Fib, IRREGULAR RR,
PROLONGED RR, NO PACER PULSER, PACER
NON-CAPTURE
ST level measurement:
Number of measurement channels:
3-electrode: 1 ch
6-electrode: 8 ch
10-electrode: 12 ch
ST level measuring range: ±2.5 mV
ST wave display on the home screen: Selectable
ST value display on the home screen: Selectable
Graphical display of ST change (ST review):
Available
QTc/QRSd measurement:
QTc/QRSd lead: All, Trace 1, I, II, III, aVR, aVL ,
aVF, V1 to V6
QTc/QRSd value display on the home screen:
Selectable
Operation mode: Manual, STAT, periodic, SIM
(depends on the SITE setting)
Measurement mode: Adult, child or neonate is
recognized by connected air hose
mean (MAP)
Operation mode: Manual, STAT, periodic, SIM
(depends on the site setting)
Other display: Oscillation graph, PR, Cuff
pressure (displays cuff pressure during
measurement)
Venous puncture mode: Available*
*target pressure is settable
PWTT triggered NIBP measurement: Available
iNIBP (NIBP measurement in inflation method):
Available
Old data display: Dim or hide
Time setting until data is recognized as old: 5,
Display range: 30 to 300 beats/min
Declared range: 30 to 300 beats/min
Counting accuracy*1 (rms*2): ±3%±1 beat/min
PI(Pulse-amplitude Index): available
SQI bar graph: available
SpO2-2: available
Delta SpO2: available
SpO2-2, Delta SpO2 cannot be monitored by
using the MULTI socket on the AY-660P.
Nellcor SpO2 , AY-651/653P, BSM-1753
SpO2 measurement:
Display range: 1 to 100% SpO
Declared range: 70 to 100 % SpO2
Measuring accuracy *1 (rms*2):
±2%SpO2 (adult)
±3%SpO2 (neonate)
Pulse rate measurement:
Display range: 20 to 300 beats/min
Declared range: 20 to 250 beats/min
Counting accuracy*1 (rms*2): ±3 beat/min
SpO2-2: available when OxiMax N-600x pulse
oximeter is connected.
Delta SpO2: available when OxiMax N-600x
pulse oximeter is connected.
Masimo SpO2 , AY-631/633P, BSM-1733
SpO2 measurement:
Display range: 1 to 100% SpO
Declared range: 70 to 100 % SpO
Measuring accuracy1 (rms*2):
No motion:
±2%SpO2 (adult)
±3%SpO2 (neonate)
Motion:
±3%SpO2 (adult)
±3%SpO2 (neonate)
Pulse rate measurement:
Display range: 25 to 240 beats/min
Declared range: 25 to 240 beats/min
Counting accuracy*1 (rms*2):
±3 beat/min (No motion)
±5 beat/min (Motion)
SpO2-2: available when Masimo pulse oximeter
is connected.
Delta SpO2: available when Masimo pulse
oximeter is connected.
PI (Perfusion Index): available
Signal IQ : available
2
2
2
◆
Temperature
Thermistor probe: 400 series (YSI)
Number of channels: Up to 4
Delta TEMP: 2
Measuring range: 0 to 45°C, 32 to 113°F
Measuring accuracy*:
injectate temperature (Ti), blood temperature
(Tb), Thermodilution curve (delta Tb)
Measuring range:
Injectate temperature (Ti): 0°C to 27°C (32 to
81°F)
Blood temperature (Tb): 15°C to 45°C (59 to
113°F)
Thermodilution curve (delta Tb): 0°C to 2.5°C (32
to 37°F)
Cardiac output (CO): 0.5 to 20 L/min
Ti: ≤ 0.025°C RMS/0.15°Cp-p
Tb: ≤ 0.016°C RMS/0.096°Cp-p (correspond to
37°C)
Delta Tb: ≤ 0.005°C RMS/0.03°Cp-p
Temperature drift:
Ti: ±0.005°C /°C
Tb: ±0.005°C /°C
Frequency response (delta Tb): DC to 12 Hz (–
3 dB)
Injectate volume range: 3, 5, 10 mL
Display update cycle: Updated every
measurement
◆
Bispectral Index, BIS
For the QP-910P BISx/BIS processor
specifications, refer to the QP-910P BISx/BIS
processor manual.
Input channels: 1 or 2 (depends on the BIS
sensor type)
Measuring parameters: Bispectral Index (BIS),
95% Spectral Edge Frequency (SEF),
Suppression Ratio (SR), EMG, Signal Quality
Index (SQI)
◆
TOF
For the AF-101P NMT module specifications,
refer to the AF-101P NMT module manual.
13
Stimulation mode:
Following stimulation modes are available.
Stimulation automatically stops when the NMT
module detected the device malfunction.
• Single stimulation
• TOF
• DBS
• TET
• PTC
Operation keys on the AF-101P: TOF, PTC,
1Hz, CAL
Stimulation current, output pulse range,
temperature sensor:
Stimulation current: 0 to 60 mA ±30% (at peak)
Selectable range: 1 mA steps
Output pulse range: 200 μs ±30%/300 μs ±30%
Output voltage: 0 to 300 V
Skin impedance: 100 to 5,000 Ω
Temperature sensor: 20.0 to 42.5°C ±0.5°C
◆
EEG
EEG can be monitored with the AE-918P neuro
unit. For the AE-918P neuro unit specifications,
refer to the AE-918P neuro unit manual.
Number of channels: Up to 8
Numeric display: SEF, MDF, TP, Abs Delta, Abs
Maximum undistorted input: > ±2 mV
Polarization voltage: > ±700 mV
Noise: <3 µV p-p (0.53 to 30 Hz)
Input impedance: >15 MΩ at 10 Hz
CMRR: >110 dB (in isolation mode)
Electrode impedance check: >10 kΩ within ±
20%
EEG Sensitivity: 10 µV/1 mm within ±5%
Frequency characteristics:
High range: 70 Hz ±20% at 70% amplitude (−3
dB)
Low range: 0.08 Hz ±20% or time constant 2 s
±20% at 70% amplitude (−3 dB)
AC filter: rejection ratio > 26 dB
Sampling frequency: 200 Hz (A/D Conversion 4
kHz)
Data display update cycle: Every 3 s or when
alarm is generated
Trend item: CSA(8ch), DSA (8ch), aEEG (8ch),
HR, SpO2
Auto impedance check: Available
Montage: Up to 5
◆
Gas
Gas can be monitored with the GF-210R
multi-gas unit or GF-220R multi-gas/flow unit.
For the GF-210R multi-gas unit or GF-220R
multi-gas/flow unit specifications, refer to the
manual for each unit.
Measurement method: Sidestream gas
sampling
Numeric display on the home screen: Up to 10
Warm-up time:
About 1 minute to CO2 measurement
About 6 minutes to measurement with
guaranteed accuracy
Sampling rate:
200 mL/min ±20 mL/min
Total system response time:
≤ 5.0 seconds (when YG-610P sampling line is
connected)
CO2 measurement:
Measurement method: Non-dispersive infrared
ray absorption
Measuring range: 0 to 10 vol%
Measuring accuracy: ± (0.43 vol% + 8% rel.)*
* Essential performance in EMC standard
N2O measurement:
Measurement method: Non-dispersive infrared
ray absorption
Measuring range: 0 to 100 vol%
Measuring accuracy: ± (2 vol% + 8% rel.)*
* Essential performance in EMC standard
Alarm (N2O(I), N2O(E)):
Upper limit: 1 to 100% in 1% steps, OFF
Lower limit: OFF, 0 to 99% in 1% steps
O2 measurement:
Measurement method: Paramagnetic
Measuring range: 5 to 100 vol%
Measuring accuracy: ± (2.5 vol% + 2.5% rel.)*
* Essential performance in EMC standard
Anesthetic agent measurement:
Measurement method: Non-dispersive infrared
ray absorption
Measured items: HAL (Halothane), ISO
(Isoflurane), ENF (Enflurane), SEV
(Sevoflurane), DES (Desflurane)
Measuring range:
HAL, ISO 0 to 8.5 vol%
ENF, SEV 0 to 10 vol%
DES 0 to 20% vol%
Measuring accuracy: ± (0.2%vol% +15%rel)*
* Essential performance in EMC standard
MAC:
MAC calculating method: selectable, uncorrected
MAC, ambient pressure corrected MAC or
enhanced MAC correction
Respiration rate:
Measuring range: 0, 4 to 60 counts/min
◆
Flow/Paw
Flow/Paw can be monitored with the GF-220R
multi-gas/flow unit. For the GF-220R
multi-gas/flow unit specifications, refer to the
GF-220R multi-gas/flow unit manual.
Measuring accuracy: −30 to +300 mmHg Available
measuring range for parameter calculation is 0 to 300
mmHg.
Measuring accuracy: ±4 mmHg (when the specified
cable and pressure transducer are used)
Injectate temperature:
Measuring range: 0 to 30°C (32 to 86°F)
Measuring accuracy: ±0.1°C (when the specified
cable and catheter are used)
Blood temperature:
Measuring range: 25 to 45°C (77 to 113°F)
Measuring accuracy: ±0.1°C (when the specified
cable and catheter are used)
Available measuring range for parameter
calculation is 0 to 300 mmHg.
Measuring accuracy: ±4 mmHg (when the
specified cable and pressure transducer are
used)
◆
CeVOX
Measuring method: Spectrophotometry
Measuring parameters: Central Venous Oxygen
Saturation (ScvO2), Oxygen Extraction Ratio
(O2ER), (DO2)*, Oxygen Delivery Index (DO2I)*,
Oxygen Consumption (VO2), Oxygen
Consumption Index (VO2I)*
*Can be calculated when ProAQT and CeVOX
are both measured.
Central Venous Oxygen Saturation:
Measuring range: 1 to 99%
Measuring accuracy: Depends on the accuracy
of the optical module. ±3% when the PC3040
CeVOX module is used.
The accuracy is guaranteed when the central
venous oxygen saturation is between 50% and
80%.
esSVI and esSVRI, patient information (BSA) is
required. To calculate esSVR and esSVRI, ART
and CVP measurement is required.)
Measuring range:
esCCO: 0.5 to 20.00 L/min, 0.01 L/min steps
esCCI: 0.5 to 20.00 L/min/m2, 0.01 L/min/m2 steps
esSV: 0 to 300 mL, 1 mL steps
esSVI: 0 to 200 mL/m2, 1 mL/m2 steps
◆
ECG/BP Output
ECG waveform output: Output the ECG on the
first trace
BP waveform output: Output the BP waveform
measured with MULTI socket 1 or the highest
priority BP waveform
HT pulse output: Outputs pulse based on the
ECG on the first trace
Output impedance:
ECG: ≤100 Ω
BP: ≤100 Ω
Output-waveform:
ECG: ±5.0 V (at 1 mV/V ± 5% sensitivity)
BP: –0.5 to +3.0 V (at 100 mmHg/V ± 1%
sensitivity)
HT: 5.0 to 15.0 V (Open collector output: 0.5 to
50 mA)
Frequency response:
ECG: ≥ 0.5 to 100 Hz (≥ –3 dB) (No
reproducibility of pace maker pulse)
BP: DC to 20 Hz ±3 Hz (–3 dB)
HT pulse width: 15 ms
Gain:
ECG: 1000
Offset:
ECG: ≤ ±50 mV
BP: ≤ ±10 mV
Delay:
ECG: 20 ms max
BP: 40 ms max
HT: 20 ms max
AA-174 multi amp unit:
156 W x 63 H x 190 D mm (excluding cable)
0.82 kg (excluding cable)
JA-170P data Aquisition Unit:
182 W x 182 H x 136 D mm (excluding cable)
1.1 kg (excluding cable)
QF series interface and IF series
communication cable:
65 W x 23 H x 44 D mm (excluding cables)
0.13 kg
RY-910PA remote controller:
45 W x 35 H x 135 D mm
0.08 kg
WS-151P recorder unit:
77 W x 73 H x 120 D mm
0.35 kg
WS-171P recorder unit:
99 W x 82 H x 142 D mm
0.62 kg
◆
Dimensions and Weight (approximate)
CU-151R core unit:
341 W × 294 H × 204 D mm
6 kg
CU-152R core unit:
403 W × 302 H × 204 D mm
7 kg
CU-171R core unit:
403 W × 300 H × 107 D mm
6.3 kg
CU-172R core unit:
475 W × 326 H × 115 D mm
8.3 kg
AY-660/661/663/651/653/631/633P input unit:
83 W × 176 H × 145 D mm
1.3kg
AA-672/674P smart expansion unit:
38 W × 165 H × 145 D mm
0.5kg
BSM-1700 bedside monitor:
147 W x 194 H x 94 D mm
1.4 kg
16
Safety Standard
IEC 60601-1:2005+Amendment 1:2012
IEC 60601-1-2:2007
IEC 60601-1-6:2010+Amendment 1:2013
IEC 60601-1-8:2006+Amendment 1:2012 1, 2
IEC 60601-1-9:2007+Amendment 1:2013
IEC 60601-2-26:2012
IEC 60601-2-27:2011
IEC 80601-2-30:2009+Amendment 1:2013
IEC 60601-2-34:2011
IEC 60601-2-49:2011
IEC 62304:2006+Amendment 1:2015
IEC 62366:2007+Amendment 1:2014
EN ISO 14971:2012
ISO 10993-1:2009
ISO 80601-2-55:2011 3
ISO 80601-2-56:2009
ISO 80601-2-61:2011 3
CAN/CSA-C22.2 NO. 60601-1:14
CAN/CSA-C22.2 NO. 60601-1-6:11+AMD1
CAN/CSA-C22.2 NO. 60601-1-8-08+AMD1
CAN/CSA-C22.2 NO. 60601-2-26:14
CAN/CSA-C22.2 NO. 60601-2-27:11
CAN/CSA-C22.2 NO. 80601-2-30-10+AMD1
CAN/CSA-C22.2 NO. 60601-2-34:12
CAN/CSA-C22.2 NO. 60601-2-49:11
CAN/CSA-C22.2 NO. 62366:14
CAN/CSA-C22.2 NO. 80601-2-55:14
CAN/CSA-C22.2 NO. 80601-2-56:12
CAN/CSA-C22.2 NO. 80601-2-61:14
AAMI-EC13:2002
AAMI-EC57:2012
ANSI/AAMI ES60601-1:2005/(R)2012,
AND C1:2009 AND A2:2010(R)2012
ANSI/AAMI/IEC 60601-1-8:2006 & A1:2012
ANSI/AAMI/IEC 60601-2-27:2011
ANSI/AAMI/IEC 80601-2-30:2009 & A1:2013
EN 300 328 V2.1.1
EN 301 893 V2.1.1
EN 62311:2008
EN 62479:2010
1 Except for the interbed alarm sound
2 To comply with the requirements from international
standards for distributed alarm systems, use the
monitor with a central monitor that complies with these
standards. The system will not comply with some
of the requirements from international standards for
distributed alarm systems if used with a non-compliant
central monitor.
3 The CSM-1701 and CSM-1702 bedside monitors do
not comply with clause 201.11.8.101.1 when the
SB-970P battery pack is not installed.
17
Environment
Transport and Storage Environment
Temperature: −20 to +65°C (−4 to +149°F)
−15 to +55°C (+5 to 131°F) (recording paper)
Humidity: 10 to 95% RH
Atmospheric pressure: 700 to 1060 hPa
Operating Environment and Power
Operating environment
Temperature: 5 to 40°C (41 to 104°F)
SpO2 accuracy is guaranteed at surrounding
temperature of 18 to 40°C (64 to 104°F).
Humidity: 30 to 85% RH (noncondensing)
NIBP accuracy is guaranteed at surrounding humidity of
15 to 85% RH
(noncondensing).
Atmospheric pressure: 700 to 1060 hPa
Power
Rated voltage: AC: 100 to 240 V
DC (battery): 10.8 V
Allowable power fluctuation range:
AC: 90 to 264 V
DC (battery): 9.0 to 12.3 V
Power input: AC: ≤ 120 VA
DC (battery): ≤ 100 W
Line frequency: 50 or 60 Hz
Allowable frequency deviation: ±2%
Battery charging current (SB-950P): during operation:
500 mA ± 50 mA
stand-by: 2000 ± 750 mA
Electrode lead, 3 electrode,
clip, AHA, K911A
(BR-903PA), 0.8 m
✓✓
ECG connection cord, IEC,
K922 (JC-906P),3 m
✓
✓
ECG connection cord, AHA,
K922A (JC-906PA), 3 m
✓
✓
SpO² connection cord, K931
(JL-900P), 2.5 m
✓ ✓ ✓
✓
Air hose for NIBP, S902
(YN-901P),3.5 m
✓ ✓ ✓
✓
NIBP cuff,13 cm width, 23-33
cm range, S951D (YP-713T)
✓
✓
NIBP cuff, 16 cm width, 33-45
cm range, S951E (YP-714T)
✓
✓
Order code
Y212C
Y212D
Model
YO-60IY3
YO-60IY4
Electrode lead, 3 electrode, clip, IEC,
K911 (BR-903P), 0.8 m
✓
✓
ECG connection cord, IEC, K922
(JC-906P),3 m
✓
✓
Air hose for NIBP, S902 (YN-901P),3.5
m
✓
✓
NIBP cuff,13 cm width, 23-33 cm
range, S951D (YP-713T)
✓
NIBP cuff, 16 cm width, 33-45 cm
range, S951E (YP-714T)
✓
When ordering Nellcor (AY-653P/651P)/Masimo
(AY-633P/631P) input unit, order accessory set
STANDARD ACCESSORIES
Bedside monitor core unit, CU-151/152R
Power cord (220-240V) 1
Bedside monitor core unit, CU-171/172R
Power cord (220-240V) 1
Recorder module, WS-151P/WS-171P
Recording paper, FQW50-2-100 1
Thermal head cleaner pen 1
below.
Input unit, AY-660P, Nihon Kohden SpO2
None
Input unit, AY-661P, Nihon Kohden SpO2
None
Input unit, AY-663P, Nihon Kohden SpO2
None
Input unit, AY-651P, Nellcor SpO2
JL-650P, Nellcor SpO2 connection cord 1
Input unit, AY-653P, Nellcor SpO
2
JL-650P, Nellcor SpO2 connection cord 1
OPTIONS
Input unit
Input unit, AY-660P, 1 MULTI connector
(IBP/CO2), Nihon Kohden SpO2, 1 x Temp, no
ECG/BP out
Input unit, AY-661P, 1 MULTI connector, Nihon
Kohden SpO2
Input unit, AY-663P, 3 MULTI connectors, Nihon
Kohden SpO2
Input unit, AY-651P, 1 MULTI connector, Nellcor
SpO2
Input unit, AY-653P, 3 MULTI connectors, Nellcor
SpO2
Input unit, AY-631P, Masimo SpO2
None
Input unit, AY-631P, 1 MULTI connector, Masimo
SpO2
Input unit, AY-633P, 3 MULTI connectors,
Input unit, AY-633P, Masimo SpO2
None
Masimo SpO2
Smart unit, AA-672P, 2 MULTI connectors
Smart unit, AA-674P, 4 MULTI connectors
Bedside monitor, BSM-1763, Nihon Kohden SpO2
Side Cover 1
Bedside monitor, BSM-1753, Nellcor SpO2
Side Cover 1
JL-650P, Nellcor SpO2 connection cord 1
Bedside monitor, BSM-1733, Masimo SpO2
Side Cover 1
Memory unit, QM-600P, for storing data in the
AY-600 series input unit
Handle for AY series input unit, DH-600P
Multi amp unit, AA-174P, 4 MULTI connectors
Bedside monitor, BSM-1763, 3 MULTI
connectors, Nihon Kohden SpO2
Bedside monitor, BSM-1753, 3 MULTI
connectors, Nellcor SpO2
Bedside monitor, BSM-1733, 3 MULTI
connectors, Masimo SpO
2
Data acquisition unit, JA-170P
OPTIONAL ACCESSORIES
Accessory set
Data acquisition unit (DAU)
Data acquisition unit, JA-170P
DAU connection cable, YS-118P0, 0.9m
DAU connection cable, YS-118P1, 2.8m
DAU connection cable, YS-118P2, 5m
DAU connection cable, YS-118P3, 10m
Modules & units
BIS processor, QE-910P
Connection cable for QE-910P, YJ-671P
Neuro unit, AE-918P
Extension cord for AE-918P, BM-918P, 1.5m
NMT module, AF-101P
Multi-gas Unit, GF-210R
Multi-gas/Flow unit, GF-220R
Connection cable for GF-210/220R, YJ-600P
(K977), 0.3 m
Connection cable for GF-210/220R, YJ-601P
19
(K978), 2.5 m
Hemodynamic unit, AP-170P
Recorder
Recorder module for CU-151R/152R, WS-151P
Recorder module for CU-171R/172R, WS-171P
Recorder cable for CU-171R/172R, YS-118P8,
Unit holder, DH-150P
Adapter, DH-151P
Channel adapter, DH-152P
Recorder holder, DH-153P
IF holder, DH-154P
Holder, DH-155P
Unit holder, DH-156P
Adapter, DH-157P
GF holder, DH-158P
Unit holder DH-159P
Adapter, DH-172P
Unit holder, DH-190P
Unit mount, DH-223P
Holder, DH-692P
Interface mount, DH-916P
Holder, DI-170P
Stand for CU-171R/172R, DM-150P
Handle for CU-151R/CU-152R, DZ-150P
Cart for CU-151R/152R, KC-600P
Cart for CU-171R/172R, KC-150P
Counter top mount, KG-600P
Wall mount kit for CU-151R/152R, KG-951P
Wall mount kit for CU-171R/172R, KG-150P
The list is continuously increasing. Please contact your Nihon Kohden representative for updated list.
22
CONSUMABLES
ECG and Respiration (Impedance)
Cable/cord
K911 Electrode lead, BR-903P, for 3 electrodes,
clip, IEC, 0.8 m
K911A Electrode lead, BR-903PA, for 3
electrodes, clip, AHA, 0.8 m
K917 Electrode lead, BR-923P, for 3
electrodes, clip, IEC, 1.5 m
K916 Electrode lead, BR-963P, for 3 electrodes
on one lead, clip, IEC, 0.8 m
K905 Electrode lead, BR-973P, for 3 electrodes,
clip, IEC, 1.25 m, for low impedance
neonate/child
K906 Electrode lead, BR-975P, for 5 electrodes,
clip, IEC, 1.25 m, for low impedance
neonate/child
K912 Electrode lead, BR-906P, for 6 electrodes,
clip, IEC, 0.8 m
K912A Electrode lead, BR-906PA, for 6
electrodes, clip, AHA, 0.8 m
K918 Electrode lead, BR-926P, for 6 electrodes,
clip, IEC, 1.5 m
K922 ECG connection cord, JC-906P, 3/6
electrodes, IEC, 3 m
K922A ECG connection cord, JC-906PA, 3/6
electrodes, AHA, 3 m
K925 ECG connection cord, JC-916P, 3/6
electrodes, IEC, 1.5 m
K921 ECG connection cord, JC-900P, 10
electrodes, IEC, 3 m, use with disposable
electrode with lead
K921A ECG connection cord, JC-900PA, 10
electrodes, AHA, 3 m, use with disposable
electrode with lead
K901 ECG patient cable, BJ-900P, 10 electrodes,
IEC, 3.8 m
K901A ECG patient cable, BJ-900PA, 10
electrodes, AHA, 3.8 m
L021 Adapter, JC-990P, for low impedance
neonate/child
Disposable electrodes (Vitrode series)
G207 L-150X, for general use, X-ray
(radiolucent), 30 electrodes × 5/set, 35 mm dia,
Vitrode L
G210D F-150M, for adult/child, 3 elec- trodes ×
50/set, 25 × 45 mm, Vitrode F
G210C F-150S, for neonate/child, 3 electrodes ×
50/set, 18 ×36 mm, Vitrode F
G221 G-600, for ICU/OR, 30 electrodes ×
20/set, 47 × 47 mm, Vitrode G
G272A V-090M3, for general use, 3 electrodes ×
30/ set, DIN, 25 × 45 mm, lead 1.0 m, Vitrode V
G278A V-09IO3, for general use, 3 electrodes ×
30/ set, DIN, 25 × 45 mm, lead 1.5 m, Vitrode V
G272B V-040M4, for general use, 4 electrodes ×
10/ set, 25 × 45 mm, lead 1.0 m, DIN, Vitrode V
G273A V-04IO4, for general use, 4 electrodes ×
10/ set, DIN, 25 × 45 mm, lead 1.5 m, Vitrode V
G272C V-060M6, for general use, 6 electrodes ×
10/ set, DIN, 25 × 45 mm, lead 1.0 m, Vitrode V
G274A V-06IO6, for general use, 6 electrodes ×
10/ set, DIN, 25 × 45 mm, lead 1.5 m, Vitrode V
G271A V-120S3, for NICU, 3 electrodes × 40/set,
DIN, 20 × 20 mm, Vitrode V
G300A N-03IS3, for NICU, 3 electrodes x
10/set, DIN, 14 x 25 mm, Vitrode N
G300D N-01IS3, for NICU, 3 electrodes x
10/set, DIN, 15 mm dia, Vitrode N
2.5 m
K935 Nellcor SpO2 connection cord, JL-650P, 3 m
K936 Masimo SpO2 connection cord, for LNOP
series, JL-630P, 3.6 m
K937 Masimo SpO2 connection cord, for LNCS
series, JL-631P, 3 m
K943A Masimo SpO2 connection cord, for RD
SET series, JL-632P, 3.6 m
Probes (BluPRO series)
P225F Finger probe, TL-201T, finger-clip type
P311C Finger probe, TL-631T3, regular size,
attached to finger or toe by tape
P310C Finger probe, TL-630T3, large size,
attached to finger or toe by tape
P225G Multi-site probe, TL-220T, attached to
adult/child finger/toe or neonate instep and sole
by tape
P259 Cottony tape, 340703, secures probe and
cord for stable monitoring, stretchable, 20
pcs/set
P267 Probe fastener, YS-093P2, 30 pcs/set, for
TL-631T3/TL-630T3
Disposable probes (BluPRO series)
P203A TL-271T, for adult finger or toe, 24
pcs/set, 0.8 m
P204A TL-271T, for adult finger or toe, 5
pcs/set, 0.8 m
P203E TL-271T3, for adult finger or toe, 24
pcs/set, 1.6 m
P204E TL-271T3, for adult finger or toe, 5
pcs/set, 1.6 m
P203B TL-272T, for child finger or toe, 24
pcs/set, 0.8 m
P204B TL-272T, for child finger or toe, 5
pcs/set, 0.8 m
P203F TL-272T3, for child finger or toe, 24
pcs/set, 1.6 m
P204F TL-272T3, for child finger or toe, 5
pcs/set, 1.6 m
P203C TL-273T, for neonate instep, 24 pcs/set,
0.8 m
P204C TL-273T, for neonate instep, 5 pcs/set,
0.8 m
P203G TL-273T3, for neonate instep, 24
pcs/set, 1.6 m
P204G TL-273T3, for neonate instep, 5
pcs/set, 1.6 m
P203D TL-274T, for infant finger or toe, 24
pcs/set, 0.8 m
P204D TL-274T, for infant finger or toe, 5
pcs/set, 0.8 m
P203H TL-274T3, for infant finger or toe, 24
pcs/set, 1.6 m
P204H TL-274T3, for infant finger or toe, 5
23
pcs/set, 1.6 m
Multi-site Y probe
P205A Multi-site Y probe, TL-260T (for adult,
child, neonate), 5 pcs/set
P260E Sponge attachment tape S for TL-260T
multi-site Y probe, 24 pcs/set
P260F Sponge attachment tape L for TL-260T
multi-site Y probe, 24 pcs/set
P256 Ear clip adapter for TL-260T, reusable
Disposable probes
P228A TL-051S, for adult/neonate, 5 pcs/set,
0.8 m cable
P228B TL-052S, for adult/neonate, 5 pcs/set,
1.6 m cable
P229A TL-061S, for child/infant, 5 pcs/set, 0.8
m cable
P229B TL-062S, for child/infant, 5 pcs/set, 1.6
m cable
P260C Foam tape, for
TL-051S/052S/061S/062S, 100 pcs
P206 TL-535U, for neonates and preterm
infant, 5 pcs/set
P264A Attachment tape, for TL-535U, S: 1kg or
less, 24 pcs
P264B Attachment tape, for TL-535U, L: 2.5kg
or less, 24 pcs
Nellcor SpO2 probes are available from
MEDTRONIC (www.medtronic.com) or their
suppliers
Masimo SpO2 probes are available from
MASIMO (www.masimo.com) or their suppliers
NIBP
Reusable Cuffs (Latex free)
S943A YP-960T: for infant, 5 cm cuff width
S943B YP -961T: for child, 7 cm cuff width
S943C YP-962T: for child, 10 cm cuff width
S944B YP -963T: for adult, 13 cm cuff width
S944C YP-964T: for adult, 15 cm cuff width
S944D YP-965T: for thigh, 19 cm cuff width
Reusable Cuffs (YAWARA-CUFF 2 series,
Latex free)
S951A YP-710T: for infant, 5 cm cuff width
S951B YP -711T: for child, 7 cm cuff width
S951C YP-712T: for adult small, 10 cm cuff
width
S951D YP-713T: for adult, 13 cm cuff width
S951E YP -714T: for adult large, 16 cm cuff
width
S951F YP -715T: for thigh, 19 cm cuff width
Disposable Cuffs
S945C YP-810P: for infant, 6 cm cuff width, 20
pcs
S945D YP-811P: for child, 8 cm cuff width, 20
pcs
S946E YP -812P: for adult, 10 cm cuff width, 20
pcs
S946F YP -813P: for adult, 14 cm cuff width, 20
pcs
S946G YP-814P: for adult, 15 cm cuff width, 20
pcs
S946H YP-815P: for adult, 17 cm cuff width, 20
pcs
S946I YP-816P: for adult, 18 cm cuff width, 20
pcs
S946J YP -817P: for thigh, 20 cm cuff width, 20
pcs
S948A YP-820P: for neonate, 2 cm cuff width,
10 pcs, Luer connector
S948B YP -821P: for neonate, 3 cm cuff width,
10 pcs, Luer connector
S948C YP-822P: for neonate, 4 cm cuff width,
10 pcs, Luer connector
S948D YP-823P: for neonate, 4.5 cm cuff
width, 10 pcs, Luer connector
S948E YP-824P: for neonate, 5 cm cuff width,
10 pcs, Luer connector
S948F YP -820S: for neonate, 2 cm cuff width,
10 pcs, SnapQuick connector
S948G YP -821S: for neonate, 3 cm cuff width,
10 pcs, SnapQuick connector
S948H YP-822S: for neonate, 4 cm cuff width,
10 pcs, SnapQuick connector
S948I YP-823S: for neonate, 4.5 cm cuff width,
10 pcs, SnapQuick connector
S948J YP-824S: for neonate, 5 cm cuff width,
10 pcs, SnapQuick connector
Disposable Cuffs (Nihon Kohden original
cuff)
S954A YP-840T: for infant, 5cm cuff width, 20
pcs
S954B YP-841T: for child, 7 cm cuff width, 20
pcs
S954C YP -842T: for adult, 10 cm cuff width, 20
pcs
S954D YP -843T: for adult, 13 cm cuff width, 20
pcs
S954E YP-844T: for adult, 16 cm cuff width, 20
pcs
S954F YP-845T: for thigh, 19 cm cuff width, 20
pcs
Air hoses
S901 YN-900P: 1.5 m
S902 YN-901P: 3.5 m
S903 YN-990P: extension hose, 1.5 m
S904 YN-920P: for neonate, 1.5 m, Luer
connector
S905 YN-921P: for neonate, 3.5 m, Luer
connector for IEC 2nd edition
S904A YN-924P: for neonate, 1.5 m, SnapQuik
connector
S905A YN-925P: for neonate, 3.5 m,
SnapQuick connector for IEC 3rd edition
24
IBP
Merit Medical (Argon):
K951 IBP connection cord, JP-900P, Merit
Medical blood pressure transducers are
available from Merit Medical (www.merit.com)
or their supplier.
Edwards Lifesciences:
L901 IBP connection cord, JP-920P, for
Edwards Lifesciences transducer, 3.5 m
Edwards Lifesciences blood pressure
transducers are available from Edwards
Lifesciences Corporation (www.edwards.com)
or their suppliers.
Biosensors International:
K957 IBP connection cord, JP-960P, for
Biosensor transducer, 3.5 m
Biosensors blood pressure transducers are
available from BIOSENSORS INTERNATIONAL
(www.biosensorsintl.com) or their suppliers.
Others:
K952 IBP connection cord, JP-910P
Temperature
K961 Temperature connection cord, JT-900P,
30 cm
Thermistor Probe
P240B For adult rectum/esophagus, 401J
P241B For child rectum/esophagus, 402J
P242D DisG2c type, 409J
P249A Probe cover, for 401J, 10 pcs/set
K962 CO connection cord, JT-950P, 2.0 m
Merit Medical:
Merit Medical injectate supply systems and
thermodilution catheters are available from
Merit Medical (www.merit.com) or their
supplier.
Edwards Lifesciences:
Edwards Lifesciences injectate supply systems
and thermodilution catheters are available from
Edwards Lifesciences (www.edwards.com).
Gas monitoring
P252 Temperature insulation pad, for 409J, 60
pcs/set
CO2 (Mainstream measurement)
Semi-quantitative method
P903 CO
JG-900P), 3 m
P922A CO2 sensor, TG-101T, 1 m
K981 CO2 adapter, JG-900P, 2 m
R801 Airway adapter, YG-101T, 50 pcs/set
P907 CO
JG-920P), 3.5 m
P923 CO2 sensor, TG-121T, 2 m
K984 CO2 adapter, JG-920P, 1.5 m
R804 Airway adapter, YG-111T, 30 pcs/set
V921 Nasal adapter, YG-121T, for nasal, 30
pcs/set
V922 Nasal adapter, YG-121T, for nasal/oral,
30 pcs/set
V923 Nasal adapter, YG-122T, for nasal/oral, 30
pcs/set, with O2 supply tube adapter
V927A Oxygen cannula, 25 pcs/set
L461 EEG connection cord, JE-905P, 2 ch, 3 m
L462 EEG connection cord, JE-906P, 8 ch, 3 m
H503A EEG disk electrode, NE-113A, 26
pcs/set, 1.5 m, DIN
H503 EEG disk electrode, NE-114A, 26
pcs/set, 0.8 m, DIN
H527 EEG collodion electrode, NE-136A, 12
pcs/set, 0.7 m, DIN
H537A EEG needle electrode, NE-224S, 20
25
pcs/set, 1.5m, DIN
H471 Disk electrode, BE-910P, 0.8 m
H472 Collodion electrode, BE-920P, 0.8 m
K461 EEG electrode lead, BE-930P, 0.8 m
F507 EEG paste, Z-181BE, 180 g tube x 2/box,
ELEFIX
H564B EEG Headband, BE-090AS, Size S
H564C EEG Headband, BE-090AM, Size M
K621B Extension cord, BM-09IS8
BIS
BIS sensors are available direct from
MEDTRONIC (www.medtronic.com) or their
suppliers
TOF (NMT module)
Main cable, JW-101P
Stimulation cable, BF-101P
Acceleration transducer, TA-101P
Temperature sensor, TT-101P
Hand adapter, YS-114P1
Transducer adapter for thumb, YS-114P2
Transducer adapter for orbicularis oculi, YS-114P3
Holder, DH-101P
Hemodynamic unit
IBP connection cord for PiCCO, JP-170P
CO connection cord for PiCCO, JT-170P
Other consumables are available direct from
PULSION Medical Systems
(http://www.pulsion.com/) or their suppliers
Recording
A721 Recording paper, FQW50-2-100, 10
pcs/set, 50 mm x 20 m
Y011 Thermal head cleaning pen, 5 pcs/set
ECG/BP out
K974 ECG/BP output cable, YJ-910P, 5 m
K975 ECG/BP output cable, YJ-920P, 0.3 m