When the patient monitor and other devices are to be used together, their equipotential
grounding terminals should be connected together, eliminating the potential difference
between them.
2. AC Power Input
3. Multifunctional Connector
It outputs defibrillator synchronization signals, nurse call signals and analogy output
signals. For iMEC15S/iMEC15, it also can connect the AG module.
4. VGA Connector
It connects a secondary display, which extends the display capability of your monitor.
The contents displayed on the secondary display screen accords with those displayed on
the monitor screen.
5. USB Connector
It connects the USB devices, such as a barcode scanner.
6. Network Connector
1
5
6
2
2-6
Page 35
V
U
F
D
U
a
a
U
Z
F
H
S
V
D
g
Z
W
W
S
V
D
U
\
H
D
V
/
W
R
\
D
Q
H
V
D
Q
S
G
V
&
P
D
H
P
K
G
P
H
W
,WLVD
SDWLHQW
&,6WK
,WDOVR
'LVSO
7KLVSDWLHQW
ZDYHIRUPV
WDQGDUG5-
PRQLWRUIRU
RXJKH*DWH
RQQHFWVWKH
\6FUH
PRQLWRUDGR
$W\SLFDOGL
FRQQHFWRU
HPRWHYLHZ
D\
SDWLHQWPRQL
Q
WVDKLJKUH
SOD\VFUHHQL
KLFKFRQQHF
WKURXJKWKH
RUWR3&IRU
ROXWLRQ7)7
VVKRZQEHO
WVWKHSDWLHQW
$1RU+/
V\VWHPXSJU
/&'WRGLVS
Z
PRQLWRUWR
EDVHGV\VWH
GH
OD\SDWLHQWS
06RURWKHU
VVXFK+,6
UDPHWHUVDQ
DQG
3DWLH QW
7KLVD
SDWLHQW
SDWLHQW
WKH>P
WKH>P
$ODUP
,QIRUPDWLRQ
HDVKRZVWKH
QDPHDQGS
LQIRUPDWLRQ
tient Setup@
tient Demo
6\PEROV
LQGL FDWHV
LQGL FDWHV
$UH
SDWLHQWLQIR
WLHQWFDWHJRU
LVLQFRPSOHW
PHQX,IDS
raphics@PH
DODUPVDUHS
DODUPDUHUH
PDWLRQVXFK
LQGL F
,IQRSDWLH
DWLHQWKDVEH
QX
DXVHG
HW
DVGHSDUWPH
WHVWKDWQR
WLVDGPLWWH
QDGPLWWHG
WEHGQXPE
DWLHQWLVDG
VHOHFWLQJW
HOHFWLQJWKLV
URRPQX
LWWHGRUWKH
LVDUHDZLOO
DUHDZLOOHQ
EHU
QWHU
HU
Page 36
indicates alarm sounds are turned off.
indicates the system is in alarm off status.
3. Technical Alarm Area
This area shows technical alarm messages and prompt messages. When multiple
messages come, they will be displayed circularly. Select this area and the technical
alarm list will be displayed.
4. Physiological Alarm Area
This area shows physiological alarm messages. When multiple alarms occur, they will
be displayed circularly. Select this area and the physiological alarm list will be
displayed.
5. Waveform Area
This area shows measurement waveforms. The waveform name is displayed at the left
upper corner of the waveform. Select this area and the corresponding measurement
setup menu will be displayed.
6. Parameter Area A
This area shows measurement parameters. Each monitored parameter has a parameter
window and the parameter name is displayed at the upper left corner. The corresponding
waveform of each parameter is displayed in the same row in the waveform area. Select
this area and the corresponding measurement setup menu will be displayed.
7. Parameter Area B
For the parameters displayed in this area, their corresponding waveforms are not
displayed.
8. Prompt Message Area
This area shows the current configuration name, prompt messages, network status icons,
battery status icons, etc. For details about battery status symbols, refer to the 25
Batteries.
indicates patient monitor is connected to a wired network successfully.
indicates the patient monitor has failed to connect a wired network.
indicates the wireless function is working.
indicates the wireless function is not working.
indicates a USB disk is inserted.
2-8
Page 37
Scroll left to display more QuickKeys.
Scroll right to display more QuickKeys.
Enter the main menu
Enter standby mode
C
Review the patient’s data
Enter the
Stop all NIBP measurement
Zero IBP
Reset the alarm system
P
Change screen
Enter the patient setup menu
Trigger a manual
Have a split
Enter the
9. QuickKeys Area
This area contains QuickKeys that give you fast access to functions.
2.6 QuickKeys
A QuickKey is a configurable graphical key, located at the bottom of the main screen. They
give you fast access to functions. Their availability and the order in which they appear on
your screen, depend on how your patient monitor is configured.
The following QuickKeys can be displayed on the screen:
hange alarm settings
NIBP measurement menu
ause or restore alarms
event
-screen view of minitrends
volume setup menu
2-9
Page 38
Default configurations
Start cardiac output procedure
Perform calculations
Have a split
Have a split
Enter the full
Enter the
Start NIBP STAT measurement
Enter the [
Enter the PAWP measurement screen
Enter the privacy mode
Enter the
Enter the intubation mode
-screen view of another patient’s conditions
-screen view of OxyCRG trends
-screen 7-lead ECG screen
[Parameters] menu
Unit Setup] menu
night mode
You can also select your desired QuickKeys to display on the screen.
1. Select [Main Menu]→[Maintenance >>]→[Manage Configuration >>]→enter the
required password→[Ok].
2. In the [Manage Configuration] menu, select [Edit Config.>>].
3. In the pop-up menu, select the desired configuration and then select [Edit].
4. In the pop-up menu, select [Screen Setup >>].
5. In the [Select QuickKeys] screen, select your desired QuickKeys and the order of them.
2-10
Page 39
WARNING
organization or
Devices connected to the equipment must meet the requirements of the applicable
1 safety standards for medical electrical equipment). The
rsonnel who connect devices to the equipment’s
1. If
leakage currents, please consult the manufacturers or else an expert in the field, to
material as they can be used if the equipment
3 Basic Operations
3.1 Installation
z The equipment shall be installed by personnel authorized by us.
z The software copyright of the equipment is solely owned by us. No
individual shall resort to juggling, copying, or exchanging it or to any other
infringement on it in any form or by any means without due permission.
z
IEC standards (e.g. IEC 60950 safety standards for information technology
equipment and IEC 60601-
system configuration must meet the requirements of the IEC 60601-1 medical
electrical systems standard. Any pe
signal input/output port is responsible for providing evidence that the safety
certification of the devices has been performed in accordance to the IEC 60601-
you have any question, please contact us.
z If it is not evident from the equipment specifications whether a particular
combination with other devices is hazardous, for example, due to summation of
ensure the necessary safety of patients and all devices concerned will not be
impaired by the proposed combination.
3.1.1 Unpacking and Checking
Before unpacking, examine the packing case carefully for signs of damage. If any damage is
detected, contact the carrier or us.
If the packing case is intact, open the package and remove the equipment and accessories
carefully. Check all materials against the packing list and check for any mechanical damage.
Contact us in case of any problem.
NOTE
z Save the packing case and packaging
must be reshipped.
3-1
Page 40
WARNING
When disposing of the packaging material, be sure to observe the applicable waste
during storage and transport. Before use,
please verify whether the packages are intact, especially the packages of single use
WARNING
requirements. Otherwise unexpected consequences, e.g. damage to the equipment,
The equipment use a mains plug as isolation means to the mains power supply. Do
z
control regulations and keep it out of children’s reach.
z The equipment might be contaminated
accessories. In case of any damage, do not apply it to patients.
3.1.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in this
manual.
The environment where the equipment is used shall be reasonably free from noises, vibration,
dust, corrosive, flammable and explosive substances. If the equipment is installed in a cabinet,
sufficient space in front and behind shall be left for convenient operation, maintenance and
repair. Moreover, to maintain good ventilation, the equipment shall be at least 2 inches (5cm)
away from around the cabinet.
When the equipment is moved from one place to another, condensation may occur as a result
of temperature or humidity difference. In this case, never start the system before the
condensation disappears.
z Make sure that the operating environment of the equipment meets the specific
could result.
z
not locate the equipment in a place difficult to operate the mains plug.
3-2
Page 41
WARNING
Contact your
3.2 Getting Started
3.2.1 Turning Power On
Once the patient monitor is installed, you can get ready for monitoring:
1. Before you start to make measurements, check the patient monitor for any mechanical
damage and make sure that all external cables, plug-ins and accessories are properly
connected.
2. Plug the power cord into the AC power source. If you run the patient monitor on battery
power, ensure that the battery is sufficiently charged.
3. Press the power on/off switch on the monitor’s front.
z Do not use the patient monitor for any monitoring procedure on a patient if you
suspect it is not working properly, or if it is mechanically damaged.
service personnel or us.
NOTE
z Check that visual and auditory alarm signals are presented correctly when the
equipment is powered on. Do not use the patient monitor for any monitoring
procedure on a patient if you suspect it is not working properly, or if it is
mechanically damaged. Contact your service personnel or us.
3.2.2 Starting Monitoring
1. Decide which measurements you want to make.
2. Connect the required patient cables and sensors.
3. Check that the patient cables and sensors are correctly connected.
4. Check that the patient settings such as [Patient Cat.], [Paced], etc, are appropriate for
5. Refer to the appropriate measurement section for details of how to perform the
your patient.
measurements you require.
3-3
Page 42
CAUTION
for 10
to forcibly shut down the monitor when it could not be shut down normally
of the patient monitor.
3.3 Disconnecting from Power
To disconnect the patient monitor from the AC power source, follow this procedure:
1. Confirm that the patient monitoring is finished.
2. Disconnect patient cables and sensors from the patient.
3. Make sure to save or clear the patient monitoring data as required.
4. Press and hold the power on/off switch. The patient monitor shuts down and you can
unplug the power cable.
z Although not recommended, you can press and hold the power on/off switch
seconds
or under some special situations. This may cause loss of data
NOTE
z To completely disconnect the power supply, unplug the power cord.
3.4 Using Keys
The monitor has three types of keys:
Softkey: A softkey is a graphic key on the screen, giving you fast access to certain
menus or functions. The monitor has two types of softkeys:
Parameter keys: Each parameter area or waveform area can be seen as a softkey.
You can enter a parameter setup menu by selecting its corresponding parameter
area or waveform area.
QuickKeys: QuickKeys are configurable graphical keys, located at the bottom of
the main screen. For details, refer to the section 2.6 .
Hardkeys: A hardkey is a physical key on a monitoring device, such as the main menu
hardkey on the monitor’s front.
Pop-Up Keys: Pop-up keys are task-related keys that appear automatically on the
monitor screen when required. For example, the confirm pop-up key appears only when
you need to confirm a change.
3-4
Page 43
Area C
Area B
Area A
3.5 Using the Touchscreen
Select screen items by pressing them directly on the patient monitor’s screen. You can enable
or disable touchscreen operation by pressing and holding the [Main Menu] QuickKey for 3
seconds. A padlock symbol
is displayed if touchscreen operation is disabled.
3.6 Setting the Screen
You can enter the [Screen Setup] window as shown below by selecting [Main Menu]ė
[Screen Setup]ė[Screen Layout >>]. In this window, you can allocate the positions of the
parameters and waveforms. The parameters or waveforms whose positions are not allocated
will not be displayed.
The ECG parameter and the first ECG waveform always display in the first row. The
configurable areas can be classified as Area A, Area B, and Area C.
In Area A, you can choose to display the parameters (having waveforms) and their
waveforms. Each parameter and the associated waveform are displayed in the same row.
In Area B, you can choose to display the parameters and their waveforms. When there is
no parameter displayed in area C, both the parameters and their waveforms will be
displayed in area B. Otherwise, only the parameters will be displayed.
In Area C, you can choose to display Timer and all the parameters whose associated
waveforms will not be displayed.
The screen can automatically adjust to ensure the best view based on your screen setup.
3-5
Page 44
WARNING
] window will
If no corresponding parameter or waveform is displayed on the monitor screen, you should
perform the following inspections:
Check the connection of the lead, cable, or sensor.
Enter the [Screen Setup] window for the desired display configuration.
Check that the parameter is switched on in [Parameters Switch] window.
z The parameters whose positions are not allocated in the [Screen Setup
not be displayed. However, the monitor can still give alarms of these parameters.
3.7 Using the On-screen Keyboard
The onscreen keyboard enables you to enter information.
Use the
Use the
Select the
Select the
Select the
key to delete the previously entered character.
key to toggle between uppercase and lowercase letters.
key to confirm what you have entered and close the onscreen keyboard.
to access the symbol keyboard.
to exit the symbol keyboard.
3-6
Page 45
Corresponding events are created when the timer is started, paused or stopped. You
3.8 Using the Timer
To display the timer in the main screen, follow this procedure:
1. Select [Main Menu]ė[Screen Setup>>]ė[Screen Layout >>] to access the [Screens]
window.
2. Select [Screen Setup] tab.
3. In the Area C, select [Timer] from the drop-down list of the desired parameter area.
Refer to 3.6 Setting the Screen.
4. Select
Select [Start] or [Pause] to start or pause the timing.
Select [Clear] to clear current timing result.
Select [Setup] to access the [Timer Setup] window, in which you can set the
to exit the window. The main screen will display the timer.
[Direction] to [Up] or [Down]. If you select [Down], you should set:
[Run Time(h:min:s)]: The available time range is 0 to 100 hours, and the default
time is 5 minutes.
[Reminder Vol]: During the last 10 seconds of the countdown, the system issues
reminder tone. The available volume range is 0 to 10. 0 means off, and 10 the
maximum volume.
Note
z
can review these events in the [Events] page of the [Review] window.
3-7
Page 46
1
2
3
4
3.9 Using the Main Menu
To enter the main menu, select [Main Menu] QuickKey or the hardkey on the monitor’s
front. Most of monitor operations and settings can be performed through the main menu.
Other menus are similar to the main menu and contain the following parts:
1. Heading: gives a sum-up for the current menu.
2. Main body: displays options, buttons, prompt messages, etc. The menu button with “>>’’
enlarges a secondary window to reveal more options or information.
3. Online help area: displays help information for the highlighted menu item.
: select to exit the current menu.
4.
3.10 Changing General Settings
This chapter covers only general settings such as language, brightness, date and time, etc.
Measurement settings and other settings can be referred to in respective sections.
3-8
Page 47
3.10.1 Setting up a Monitor
In situations where you install a patient monitor or change the patient monitor’s application
site, you need to setup the patient monitor as follows:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. In the [User Maintenance] menu, select, in turn, [Monitor Name], [Department] and
[Bed No.], and then change their settings.
3.10.2 Changing Language
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. In the [User Maintenance] menu, select [Language] and then select the desired
language.
3. Restart the patient monitor.
3.10.3 Adjusting the Screen Brightness
1. Select the [Main Menu]→[Screen Setup >>]→[Brightness].
2. Select the appropriate setting for the screen brightness. 10 is the brightest, and 1 is the
least bright.
If the patient monitor operates on battery power, you can set a less bright screen to prolong
the operating time of the battery. When the patient monitor enters standby mode, the screen
will change to the least brightness automatically.
3.10.4 Showing/Hiding the Help
The patient monitor provides online help information. The user can display or hide the help
as required.
1. Select [Main Menu]→[Screen Setup >>].
2. Select [Help] and toggle between [On] and [Off].
3-9
Page 48
CAUTION
Changing date and time will affect the storage of trends and events and may cause
3.10.5 Setting the Date and Time
1. Select [Main Menu] →[Maintenance >>]→[System Time >>].
2. Set the date and time.
3. Select [Date Format] and toggle between [yyyy-mm-dd], [mm-dd-yyyy] and
[dd-mm-yyyy].
4. Select [Time Format] and toggle between [24h] and [12h].
If your patient monitor is connected to a central monitoring system (CMS), the date and time
are automatically taken from that CMS. In that case, you cannot change the date and time
settings on your patient monitor.
z
data missing.
3.10.6 Adjusting Volume
Alarm Volume
1. Select the [Alarm Setup] QuickKey→[Others], or [Main Menu]→[Alarm
Setup >>]→[Others].
2. Select [Alm Volume] and then select the appropriate volume: X-10, in which X is the
minimum volume, depending on the set minimum alarm volume (refer to the chapter
Alarms), and 10 the maximum volume.
Key Volume
When you press the navigation knob or the touchscreen, or the hardkeys on the panel, the
patient monitor prompts you by making a sound of the key volume you have set.
1. Select the [Volume Setup] QuickKey, or [Main Menu]→[Screen Setup >>].
2. Select [Key Volume] and then select the appropriate volume. 0 means off, and 10 the
maximum volume.
3-10
Page 49
QRS Volume
The QRS tone is derived from either the HR or PR, depending on which is currently selected
as the alarm source in [ECG Setup] or [SpO2 Setup]. When monitoring SpO
variable pitch tone which changes as the patient’s saturation level changes. The pitch of the
tone rises as the saturation level increases and falls as the saturation level decreases. The
volume of this tone is user adjustable.
1. Select the [Volume Setup] QuickKey, or the ECG parameter window→[Others >>], or
parameter window.
the SpO
2
2. Select [QRS Volume] or [Beat Vol] and then select the appropriate volume. 0 means
off, and 10 the maximum volume.
, there is a
2
3.11 Setting Parameters
3.11.1 Switching the Parameters On/Off
To switch the parameters on or off,
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the
required passwordė[Others].
2. Configure the [Para Switch Authority] to [Unprotected] or [Protected].
If [Para Switch Authority] is configured to [Unprotected], select[Main Menu]ė
[Screen Setup>>]ė[Screen Layout >>]ė[Parameters Switch] to switch the
parameters on or off.
If [Para Switch Authority] is configured to [Protected], the parameter switch is
password protected. To switch the parameters on or off, select [Main Menu]ė
[Maintenance >>]ė[User Maintenance >>]ėenter the required passwordė
[Others >>]ė[Parameters Switch Setup >>].
When a parameter is switched off, its corresponding parameter module stops working, and its
parameter value and waveform are not shown on the monitor display.
NOTE
z ECG is always selected, and you cannot switch it off.
3-11
Page 50
WARNING
brightness, alarm volume, QRS
when the setting value is
3.11.2 Accessing the Parameters Menu
Select [Parameters >>] from the main menu or select corresponding parameter area or
waveform area to access a parameter setup menu.
3.12 Operating Modes
Your monitor has different operating modes. Some are password protected. This section lists
the major operating modes.
3.12.1 Monitoring Mode
This is the normal, everyday working mode that you use for monitoring patients. Your
monitor automatically enters the monitoring mode after being turned on.
3.12.2 Night Mode
To avoid disturbing the patient, night mode may be used.
If a technical alarm or physiological alarm occurs, the alarm lamp will flash. The flashing
color and frequency match the alarm level as follows:
High level alarms: the lamp quickly flashes red.
Medium level alarms: the lamp slowly flashes yellow.
Low level alarms: the lamp turns yellow without flashing.
7-2
Page 75
7.3.2 Alarm Message
When an alarm occurs, an alarm message will appear in the technical or physiological alarm
area. For physiological alarms, the asterisk symbols (*) before the alarm message match the
alarm level as follows:
High level alarms: ***
Medium level alarms: **
Low level alarms: *
Additionally, the alarm message uses different background color to match the alarm level:
High level alarms: red
Medium level alarms: yellow
Low level alarms: yellow
You can view the alarm messages by selecting the physiological or technical alarm area.
7.3.3 Flashing Numeric
If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in
alarm will flash every second, and the corresponding alarm limit will also flash at the same
frequency indicating the high or low alarm limit is violated.
7.3.4 Audible Alarm Tones
The alarm tone is distinct from heart beat tone, keystroke tone and pulse tone in frequency.
This monitor has three choices of alarm tones and patterns: ISO, Mode 1 and Mode 2. For
each pattern, the alarm tones identify the alarm levels as follows:
ISO pattern:
High level alarms: triple+double+triple+double beep.
Medium level alarms: triple beep.
Low level alarms: single beep.
Mode 1:
High level alarms: high-pitched single beep.
Medium level alarms: double beep.
Low level alarms: low-pitched single beep.
7-3
Page 76
When multiple alarms of different levels occur simultaneously, the patient monitor
lamp and give alarm sounds
Some physiological alarms, such as asystole, are exclusive. They have identical alarm
gical alarms, but their alarm
say, when an exclusive physiological
simultaneously,
Mode 2:
High level alarms: high-pitched triple beep.
Medium level alarms: double beep.
Low level alarms: low-pitched single beep.
NOTE
z
will select the alarm of the highest level, light the alarm
accordingly, while all the alarm messages are displayed circularly on the screen.
z
tones and alarm lights with normal high level physiolo
messages are displayed exclusively. That is to
alarm and a normal high level physiological alarm are triggered
only alarm message of the exclusive physiological alarm is displayed.
7.3.5 Alarm Status Symbols
Apart from the aforementioned alarm indicators, the patient monitor still uses the following
symbols telling the alarm status:
indicates alarms are paused.
indicates alarms are reset.
indicates the alarm sound is turned off.
indicates individual measurement alarms are turned off or the system is in alarm
off status.
7-4
Page 77
7.4 Alarm Tone Configuration
7.4.1 Setting the Minimum Alarm Volume
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Minimum Alarm Volume] and toggle between 0 and 10.
The minimum alarm volume refers to the minimum value you can set for the alarm volume,
which is not affected by user or factory default configurations. The setting of minimum
alarm volume remains unchanged when the patient monitor shuts down and restarts.
7.4.2 Changing the Alarm Volume
1. Select the [Volume Setup] QuickKey or the [Alarm Setup] QuickKeyė[Others], or
[Main Menu]→[Alarm Setup >>]ė[Others].
2. Select the appropriate volume from [Alm Volume]: X-10, in which X is the minimum
volume, depending on the set minimum alarm volume, and 10 the maximum volume.
3. Select [High Alarm Volume] to set the volume of the high priority alarm as [Alm
Volume+0], [Alm Volume+1] or [Alm Volume+2].
4. Select [Reminder Vol] to set the volume of the reminder tone as [High], [Med] or
[Low].
When alarm volume is set to 0, the alarm sound is turned off and a
the screen.
symbol appears on
7.4.3 Setting the Interval between Alarm Sounds
You cannot change the interval between alarm tones if you choose mode 1 or 2 as your
desired alarm tone pattern. For these two patterns, the interval between alarm tones
identifies the alarm levels as follows:
Mode 1:
Interval between high level alarm tones: continuously.
Interval between medium level alarm tones: 5 s.
Interval between low level alarm tones: 20 s.
Mode 2:
7-5
Page 78
WARNING
When the alarm sound is switched off, the patient monitor will give no audible alarm
about
Interval between high level alarm tones: 1 s.
Interval between medium level alarm tones: 5 s.
Interval between low level alarm tones: 20 s.
If you choose the ISO pattern, you can change the interval between alarm tones. To change
the interval between alarm tones:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
Interval (s)] in turn and then select the appropriate settings.
z
tones even if a new alarm occurs. Therefore the user should be very careful
whether to switch off the alarm sound or not.
z Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level may result in a hazard to the patient.
Always keep the patient under close surveillance.
7.4.4 Changing the Alarm Tone Pattern
To change the alarm tone pattern:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Alarm Sound] and toggle between [ISO], [Mode 1]and [Mode 2].
User or factory default configurations exert no impact on the setup of alarm tone pattern.
The alarm tone pattern remains unchanged after the monitor restarts.
7.4.5 Setting the Reminder Tones
When the alarm volume is set to zero, or the alarm tone is reset or switched off, the patient
monitor issues a periodical reminder tone. To set the reminder tones:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
7-6
Page 79
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Set the [Reminder Tones] to [On], [Off] or [Re-alarm]. When [Re-alarm] is selected,
the current physiological alarms and technical alarms marked with “Ĝ” will be
re-generated after the [Reminder Interval] if the alarm condition persists.
To set the interval between reminder tones, select [Reminder Interval] and toggle between
[1min], [2min] and [3min].
In addition, you can set the volume of alarm reminder tones. To set the volume of alarm
reminder tones, select [Main Menu]→[Alarm Setup >>]→[Others] or the [Alarm Setup]
QuickKey→[Others]. Then, select [Reminder Vol] and toggle between [High], [Medium]
and [Low].
7.5 Understanding the Alarm Setup Menu
Select the [Alarm Setup] QuickKey or [Main Menu]→[Alarm Setup >>] to enter the
[Alarm Setup], where you can:
Set alarm properties for all parameters.
Change ST alarm settings.
Change arrhythmia alarm settings.
Set the threshold for some arrhythmia alarms.
Change other settings.
Please refer to the Monitoring ECG for how to change ST alarm settings, how to change
arrhythmia alarm settings and how to set the threshold for some arrhythmia alarms.
7-7
Page 80
WARNING
ineffective. For example, High oxygen levels may predispose a premature infant to
retrolental fibroplasia. If this is a consideration do NOT set the high alarm limit to
When monitoring patients that are not continuously attended by a clinical operator,
properly configure the alarm system and adjust alarm settings as per the patient's
7.5.1 Setting Alarm Properties for All Parameters
In the main menu, select [Alarm Setup >>]→[Parameters]. You can review and set alarm
limits, alarm switches, alarm level and alarm recordings for all parameters.
When a measurement alarm occurs, automatic recording of all the measurement numerics
and related waveforms is possible when the measurement’s [On/Off] and [Record] are set
on.
z Make sure that the alarm limits settings are appropriate for your patient before
monitoring.
z Setting alarm limits to extreme values may cause the alarm system to become
100%, which is equivalent to switching the alarm off.
z
condition.
7.5.2 Adjusting Alarm Limits Automatically
The monitor can automatically adjust alarm limits according to the measured vital signs,
using the auto limits function. When auto limits are selected, the monitor calculates safe
auto limits based on the latest measured values.
To get accurate auto alarm limits, you need to collect a set of measured vital signs as a
baseline. Then, in the main menu, select [Alarm Setup >>]→[Parameters]→[Auto Limits]
→[Ok]. The monitor will create new alarm limits based on the measured values.
Before applying these automatically created alarm limits, confirm if they are appropriate for
your patient in the mass alarm setup menu. If not, you can adjust them manually. These
alarm limits will remain unchanged until you select auto limits again or adjust them
manually.
7-8
Page 81
Module
Parameter
Low alarm limit
Auto alarm limits
range
Adult
pediatric
Neonate
Adult
pediatric
Neonate
ECG
HR/PR
(HR
0.8)
or
bpm
(whichever
is greater)
(HR
30) or
90
(whichever
is greater)
(HR
1.25)
or
bpm
(whichever
is smalle)
(HR
200bpm (whichever
is smaller)
Adult/pediatric: 35
to
Neonate:
to 225
Resp
RR
(RR
0.5)
or
rpm
(whichever
is greater)
(RR
10) or
30
rpm
(whichever
is greater)
(RR
1.5)
or
rpm
(whichever
is smaller)
(RR
rpm
(whichever is
smaller)
Adult/pediatric: 6
to
Neonate:
to 90
SpO2
SpO
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range
NIBP
NIBP-S
(SYS × 0.68
+ 10
mmHg
(SYS – 15)
or 45mmHg
(whichever
is greater)
(SYS × 0.86
+ 38
mmHg
(SYS
105mmHg
(whichever is
smaller)
Adult:
270
Pediatric:
to 185
Neonate: 40 to 115
NIBP-D
(Dia
0.68
+ 6
mmHg
(Dia
15)
or 20mmHg
(whichever
is greater)
(Dia
0.86
+ 32
mmHg
(Dia
80mmHg (whichever
is smaller)
Adult:
to 210
Pediatric:
to 150
Neonate:
to 90
NIBP-M
(Mean ×
0.68
)
mmHg
(Mean – 15)
or 35mmHg
(whichever
is greater)
(Mean ×
0.86
35)
mmHg
(Mean
95
mmHg
(whichever is
smaller)
Adult:
to 230
Pediatric:
to 165
Neonate:
to 105
Tem p
T1 (T1 – 0.5)ć
(T1
0.5) ć
(T1 + 0.5)ć
(T1 + 0.5)
1 to 49
T2 (T2 – 0.5)ć
(T2
0.5) ć
(T2 + 0.5)ć
(T2 + 0.5)
1 to 49
TD
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range
The monitor calculates the auto limits based on the following rules.
High alarm limit
/
/
×
40
×
2
6
)
×
)
+ 8
–
bpm
–
–
–
240
30
×
+ 40) or
240
×
+ 25) or 85
55
55
10
+ 15) or
)
×
+ 15) or
)
+ 15) or
+
ć
45 to
45
25
25
20
30
30
25
ć
–
ć
ć
7-9
Page 82
Module
Parameter
Low alarm limit
Auto alarm limits
range
Adult
pediatric
Neonate
Adult
pediatric
Neonate
IBP˖
ART/
Ao/
UAP/
BAP/
FAP /
LV/
P1
P4
(Arterial
pressure)
IBP-S
(SYS × 0.68
ˇ
10)
mmHg
(SYS – 15)
or 45mmHg
(whichever
is greater)
(SYS × 0.86
ˇ
mmHg
(SYS
105mmHg
(whichever is
smaller
Adult: 45 to 270
Pediatric
45 to 185
Neonate
35 to 115
IBP-D
(Dia
0.68
+ 6
mmHg
(Dia
15)
or 20mmHg
(whichever
is greater)
(Dia
0.86
+ 32
mmHg
(Dia
80mmHg (whichever
is smaller)
Adult:
225
Pediatric:
to 150
Neonate:
to 90
IBP-M
(Mean ×
0.68
8)mmHg
(Mean – 15)
or 35mmHg
(whichever
is greater)
(Mean ×
0.86
35)mmHg
(Mean
95mmHg (whichever
is smaller)
Adult:
245
Pediatric:
to 180
Neonate:
to 105
IBP:PA
IBP-S
SYS
0.75
SYS
0.75
SYS
1.25
SYS
3 to
IBP-D
Dia
0.75
Dia
0.75
Dia
1.25
Dia
IBP-M
Mean × 0.75
Mean × 0.75
Mean × 1.25
Mean
IBP:
CVP/
ICP/
LAP/
RAP/
UVP/
P1
P4
(Venous
pressur
e
IBP-M
Mean × 0.75
Mean × 0.75
Mean × 1.25
Mean
3 to
IBP:CPP
CPP
CPP
0.68
+ 8mmHg
(CPP – 15)
or 35mmHg
(whichever
is greater)
CPP
0.86
ˇ35mmHg
(CPP
95mmHg (whichever
is
Adult:
235
mmHg
Pediatric:
to 175
mmHg
Neonate:
to 100
mmHg
)
High alarm limit
/
–
×
38)
/
×
)
+ 15) or
)
+ 15) or
:
:
25 to
25
20
-
+
×
×
-
)
×
×
×
+
×
×
×
7-10
+ 15) or
× 1.25
× 1.25
× 1.25
× 1.25
ˇ15) or
smaller)
30 to
30
25
120mmHg
40mmHg
20 to
25
25
Page 83
Module
Parameter
Low alarm limit
Auto alarm limits
range
Adult
pediatric
Neonate
Adult
pediatric
Neonate
CO2
EtCO2
0 to
32mmHg:
remains the
same
0 to
32mmHg:
remains the
same
0 to
32mmHg:
remains the
same
0 to
remains the same
Same as the
measurement range
32
35mmHg:
29mmHg
32
35mmHg:
29mmHg
32
35mmHg:
41mmHg
32
41mmHg
35
45mmHg:
(etCO
6)
mmHg
35
45mmHg:
(etCO
6)
mmHg
35
45mmHg:
(etCO
+6)
mmHg
35
(etCO
mmHg
45
48mmHg:39
mmHg
45
48mmHg:39
mmHg
45
48mmHg:51
mmHg
45
48mmHg:51
mmHg
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg: remains
the same
FiCO2
N/A N/A
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range
awRR
awRR × 0.5
or
(whichever
is greater)
(awRR –
10
30
rpm
(whichever
is greater)
awRR × 1.5
or
rpm
(whichever
is smaller)
(awRR+25
85
rpm
(whichever is
smaller)
Adult/pediatric: 6
to
Neonate:
to 90
AG
EtCO2
(AG)
Same as CO
FiCO2
(AG)
awRR
awRR × 0.5
or 6
(whichever
is greater)
awRR – 10
or 30
rpm
(whichever
is greater)
awRR × 1.5
or 30
rpm
(whichever
is smaller)
awRR+25 or 85 rpm
(whichever is
smaller)
Adult/pediatric: 6
to 55
Neonate: 10 to 90
FiAA/
EtAA
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range
FiO
Same as the
Same as the
Same as the
Same as the default
Same as the
High alarm limit
/
/
32mmHg:
to
to
to
to 35mmHg:
to
to
to
to 45mmHg:
-
2
to
6 rpm
to
) or
module
2
-
2
2
to
30
to
+6)
2
) or
55
10
rpm
/
2
7-11
Page 84
Module
Parameter
Low alarm limit
Auto alarm limits
range
Adult
pediatric
Neonate
Adult
pediatric
Neonate
EtCO2
default
alarm limit
default
alarm limit
default
alarm limit
alarm limit
measurement range
FiN2O/
EtN
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range
C.O.
BT
Adult˖
(BT
N/A
Adult˖
(BT
N/A
Same as the
measurement range
High alarm limit
/
/
O
2
– 1)ć
– 1)ć
7.5.3 Setting Alarm Delay Time
You can set the alarm delay time for over-limit alarms of continuously measured parameters.
If the alarm-triggered condition disappears within the delay time, the patient monitor will
not give the alarm. You can set the [Alarm Delay] in the [Others] window of [Alarm
Setup] menu.
To set the alarm delay time,
1. Select [Main Menu]→ [Maintenance >>]→ [User Maintenance >>]. Enter the
required password and then select [Ok].
2. Select [Alarm Setup >>]→ [Alarm Delay].
Alarm delay is not applied to the following physiological alarms:
Apnea
ST alarms
Arrhythmia alarms
ECG Weak Signal
Resp Artifact
No Pulse
Nellcor SpO
Measurements of noncontinuous parameters over alarm limits
HR over alarm limits
Anesthetic Mixture's MAC >3
You can set [Apnea Delay] and [ST Alarm Delay] separately.
over alarm limits
2
7-12
Page 85
7.5.4 Setting SpO2 Technical Alarm Delay
You can set [Tech. Alarm Delay] in the [Others] tab of the [Alarm Setup] menu. The
options are [Off], [5s], [10s] and [15s]. The delay is effective to the following technical
alarms: SpO
Sensor Off, SpO2 Too Much Light, SpO2 Low Signal and SpO2 Interference.
2
7.5.5 Setting Recording Length
You can change the length of the recorded waveforms. In the [Others] window of the
[Alarm Setup] menu, select [Recording Length] and toggle between [8 s], [16 s] and [32
s]:
[8 s]: 4 seconds respectively before and after the alarm or manual event trigger
moment.
[16 s]: 8 seconds respectively before and after the alarm or manual event trigger
moment.
[32 s]: 16 seconds respectively before and after the alarm or manual event trigger
moment.
7.5.6 Intubation Mode
When performing intubation during general anesthesia, you can set the patient monitor to
enter intubation mode in order to reduce unnecessary alarms. Intubation mode is available
for Resp, CO
[Intubation Mode] button to disable respective physiological alarms.
The default intubation time is 2 minutes. You can also change the time by following this
procedure:
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the
2. Select [Alarm Setup >>], and set the [Intubation Mode Period] to [1 min], [2 min],
and AG parameters. In the setup menu of these parameters, you can choose
2
required password.
[3 min], or [5 min].
7-13
Page 86
you
7.6 Pausing Alarms
If you want to temporarily prevent alarms from sounding, you can pause alarms by pressing
hardkey on the monitor’s front. When alarms are paused:
the
No alarm lamps flash and no alarms are sounded.
No numeric and alarm limit flash.
No alarm messages are shown.
The remaining pause time is displayed in the physiological alarm area.
The
The patient monitor enters into the alarm paused status as soon as it is turned on. The alarm
pause time is fixed to be 2 minutes.
When the alarm pause time expires, the alarm paused status is automatically cancelled and
the alarm tone will sound. You can also cancel the alarm paused status by pressing the
hardkey.
The alarm pause time can be set to [1 min], [2 min], [3 min], [5 min], [10 min], [15 min]
or [Permanent]. The default alarm pause time is 2 minutes.
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the
2. Select [Alarm Setup >>]ė[Alarm Pause Time] and then select the appropriate
You can also temporarily prolong the alarm pause time after the monitor entersthe alarm
paused status:
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the
2. In the [Alarm Setup] menu, set the [Max. Alarm Pause 15min] to [Enable].
3. In the physiological alarm area, select a proper time in the [Alarm Pause Time] menu.
alarms paused symbol is displayed in the alarm symbol area.
required password.
setting from the popup list.
required passwordė[Alarm Setup >>].
NOTE
z [Max. Alarm Pause 15min] is configured to [Disable] by default. In this case,
cannot prolong the pause time.The prolonged pause time is only effective to the
current paused alarms.
7-14
Page 87
WARNING
Please be very
7.7 Switching Off All Alarms
If [Alarm Pause Time] is set to [Permanent], the patient monitor enters into the alarm off
status after the
As for physiological alarms, no alarm lamps flash and no alarms are sounded.
As for physiological alarms, no numeric and alarm limit flash.
No physiological alarm messages are shown.
[Alarm Off] is displayed in the physiological alarm area with red background.
As for technical alarms, no alarms are sounded.
hardkey is pressed. During the alarm off status,
The
You can cancel the alarm off status by pressing the
z Pausing or switching off alarms may result in a hazard to the patient.
alarm off symbol is displayed in the alarm symbol area.
hardkey.
careful.
7.8 Resetting Alarms
By selecting the QuickKey, you can reset the alarm system to acknowledging the
on-going alarms and enable the alarm system to respond to a subsequent alarm condition.
For physiological alarms, except the NIBP-related alarms, when the alarm system is reset:
The alarm sound is reset.
A Ĝ appears before the alarm message, indicating that the alarm is acknowledged.
The icon
The parameter numeric and alarm limits still flash.
The indication of alarm lamp for the physiological alarm depends on the alarm light setting.
When [Alarm Light on Alarm Reset] is set to [On], the alarm lamp remains flashing.
When [Alarm Light on Alarm Reset] is set to [Off], the alarm lamp stops flashing.
Technical alarms give different alarm indicators when the alarm system is reset:
appears in the alarm symbol area.
7-15
Page 88
For some technical alarms, including the NIBP-related alarms, a Ĝ appears before
the alarm message and
is acknowledged.
Some technical alarms are changed to the prompt messages.
Some technical alarms are cleared. The monitor gives no alarm indications.
For details about the indications of technical alarms when the alarm system is reset, refer to
D.2 Technical Alarm Messages.
To set [Alarm Light on Alarm Reset]:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Alarm Light on Alarm Reset], and toggle between [On] and [Off].
The default setting for [Alarm Light on Alarm Reset] is [On].
appears in the alarm symbol area, indicating that the alarm
7.9 Latching Alarms
The alarm latching setting for your patient monitor defines how the indicators of the
physiological alarms behave when you do not acknowledge them:
If you do not “latch” the physiological alarms, their alarm indications disappear when
the alarm condition ends.
If you “latch” the physiological alarms, all visual and audible alarm indications last
until you acknowledge the alarms, except that the measurement numeric and violated
alarm limit stop flashing as soon as the initial alarm condition goes away.
You can separately latch the visual indications or simultaneously latch the visual and the
audible indication.
When the visual indications are latched, the visual indications, including alarm lamp,
alarm message and its background remains when the alarm condition ends.
When the audible indications are latched, the monitor issues alarm sounds when the
alarm condition ends.
To latch a physiological alarm:
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the
required password.
2. Select [Alarm Setup >>]ė[Latching Alarms>>].
3. In the [Latching Alarms] menu, select how you want to latch the alarms.
7-16
Page 89
Changing of alarm priority may affect the latching status of corresponding alarm.
Please determine if you need to reset the latching status for the specific alarm when
The rules for latching the alarms are:
You can separately select [Latching Visual Signal].
Selecting [Latching Audible Signal] simultaneously latches the visual signal.
Selecting alarms of lower priority simultaneously latches the alarms of higher priority.
NOTE
z
you have changed its alarm priority.
z When the alarm system is reset, the latched physiological alarms are cleared.
7.10 Testing Alarms
When the monitor starts up, a selftest is performed. In this case the alarm lamp is lit in
yellow and red respectively, and the system gives a beep. This indicates that the visible and
audible alarm indicators are functioning correctly.
For further testing of individual measurement alarms, perform the measurement on yourself
(for example SpO
appropriate alarm behaviour is observed.
or CO2) or use a simulator. Adjust alarm limits and check that
2
7.11 Using Care Group Alarms
7.11.1 Care Group Auto Alarms
When a Care Group is set up on your monitor, a flashing symbol will appear beside the
QuickKeys area if any monitor in your Care Group, which is not currently viewed by your
monitor, is alarming. The alarm symbol is shown as below.
The background colors of the alarm symbols indicate alarm levels, and are the same as
those of the corresponding alarm messages. If multiple alarms are active in the Care Group,
the background color is the same as that of the highest-level alarm message. For more
information about the alarm message and background color, see7.3.2 Alarm Message.
When a patient monitor in the Care Group is disconnected, the flashing symbol is shown as
below.
7-17
Page 90
The department and bed label of the alarming monitor appear on the symbol. You can enter
the view other patient window by pressing this symbol.
7.11.2 Resetting Care Group Alarms
You can reset the alarms presented on the viewed bed by pressing the [Alarm Reset] from
the current monitor’s [View Other Patient] window. To enable this function:
1. For iMEC12/iMEC10/iMEC8/iMEC7/iMEC6/iMEC5 monitor, select [Main Menu]→
[Maintenance>>]→[User Maintenance>>]→enter the required password→[Alarm
Setup>>]→[Other Bed Alarm Setup>>]. For iMEC15S/iMEC15 monitor, select
[Main Menu]→[Maintenance>>]→[User Maintenance>>]→enter the required
password→[Alarm Setup>>].
2. Set [Reset Other Bed’s Alarms] to [On].
The alarms presented on the current monitor can also be reset from another monitor viewing
this monitor. To do so, proceed as follows:
1. For iMEC12/iMEC10/iMEC8/iMEC7/iMEC6/iMEC5 monitor, select [Main Menu]→
[Maintenance>>]→[User Maintenance>>]→enter the required password→[Alarm
Setup>>]→[Other Bed Alarm Setup>>] in the current monitor. For