Mindray iMEC 5,iMEC 6, iMEC 7, iMEC 10, iMEC 12, iMEC 15 User manual [gb]

Page 1
iMEC15S/iMEC15/iMEC12 iMEC10/iMEC8/iMEC7 iMEC6/iMEC5
Patient Monitor
Operator’s Manual
Page 2
Page 3
© Cop
yright 2014-2018 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights
reserved.
For this Operator’s Manual, the issue date is December 2018.
I
Page 4
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this manual.
This manual may refer to information protected by copyrights or patents and does not
convey any license under the patent rights of Mindray, nor the rights of others.
Mindray intends to maintain the contents of this manual as confidential information.
Disclosure of the information in this manual in any manner whatsoever without the written
permission of Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaption and translation of this
manual in any manner whatsoever without the written permission of Mindray is strictly
forbidden.
, , and are the registered trademarks or trademarks owned
by Mindray in China and other countries. All other trademarks that appear in this manual
are used only for editorial purposes without the intention of improperly using them. They
are the property of their respective owners.
II
Page 5
WARNING
organization that employs this equipment to carry
out a reasonable service/maintenance plan. Neglect of this may result in machine
Responsibility on the Manufacturer Party
Contents of this manual are subject to changes without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be
liable for errors contained herein nor for incidental or consequential damages in connection
with the furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
all installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national
and local requirements;
the product is used in accordance with the instructions for use.
z This equipment must be operated by skilled/trained clinical professionals.
z It is important for the hospital or
breakdown or personal injury.
III
Page 6
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting
from the improper use or application of the product or the use of parts or accessories not
approved by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to
Malfunction or damage caused by improper use or man-made failure.
Malfunction or damage caused by unstable or out-of-range power input.
Malfunction or damage caused by force majeure such as fire and earthquake.
Malfunction or damage caused by improper operation or repair by unqualified or
unauthorized service people.
Malfunction of the instrument or part whose serial number is not legible enough.
Others not caused by instrument or part itself.
IV
Page 7
Manufacturer:
Shenzhen Mindray Bio
Website:
www.mindray.com
E
service@mindray.com
Tel :
+86 755 81888998
Fax:
+86 755 26582680
EC-Representative:
Shanghai International Holding Corp. GmbH(Europe)
Address:
Eiffestraβe
Tel : 0049
Fax: 0049
Customer Service Department
Address:
-mail Address:
Mindray Building, Keji 12th Road South, High-tech Industrial
Park, Nanshan, Shenzhen 518057, P.R.China
80, 20537 Hamburg, Germany
-40-2513175
-40-255726
-Medical Electronics Co., Ltd.
V
Page 8
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely and in
accordance with its function and intended use. Observance of this manual is a prerequisite
for proper product performance and correct operation and ensures patient and operator
safety.
This manual is based on the maximum configuration and therefore some contents may not
apply to your product. If you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to the
equipment so that it can be obtained conveniently when needed.
Intended Audience
This manual is geared for clinical professionals who are expected to have a working
knowledge of medical procedures, practices and terminology as required for monitoring of
critically ill patients.
Illustrations
All illustrations in this manual serve as examples only. They may not necessarily reflect the
setup or data displayed on your patient monitor.
Conventions
Italic text is used in this manual to quote the referenced chapters or sections.
[ ] is used to enclose screen texts.
→ is used to indicate operational procedures.
VI
Page 9
Content
1 Safety ....................................................................................................................................... 1-1
1.1 Safety Information .......................................................................................................... 1-1
1.1.1 Warnings............................................................................................................. 1-2
1.1.2 Cautions ............................................................................................................. 1-3
1.1.3 Notes .................................................................................................................. 1-3
1.2 Equipment Symbols ........................................................................................................ 1-4
2 The Basics ............................................................................................................................... 2-1
2.1 Monitor Description ........................................................................................................ 2-1
2.1.1 Intended Use....................................................................................................... 2-1
2.1.2 Applied Parts ...................................................................................................... 2-1
2.2 Front View ....................................................................................................................... 2-2
2.3 Side View ........................................................................................................................ 2-4
2.3.1 iMEC15S/iMEC15/iMEC12/iMEC10/iMEC7/iMEC6 ...................................... 2-4
2.3.2 iMEC8/iMEC5 ................................................................................................... 2-5
2.4 Rear View ........................................................................................................................ 2-6
2.5 Display Screen ................................................................................................................ 2-7
2.6 QuickKeys ....................................................................................................................... 2-9
3 Basic Operations .................................................................................................................... 3-1
3.1 Installation ....................................................................................................................... 3-1
3.1.1 Unpacking and Checking ................................................................................... 3-1
3.1.2 Environmental Requirements ............................................................................. 3-2
3.2 Getting Started ................................................................................................................ 3-3
3.2.1 Turning Power On .............................................................................................. 3-3
3.2.2 Starting Monitoring ............................................................................................ 3-3
3.3 Disconnecting from Power .............................................................................................. 3-4
3.4 Using Keys ...................................................................................................................... 3-4
3.5 Using the Touchscreen .................................................................................................... 3-5
3.6 Setting the Screen ............................................................................................................ 3-5
3.7 Using the On-screen Keyboard ....................................................................................... 3-6
3.8 Using the Timer ............................................................................................................... 3-7
3.9 Using the Main Menu...................................................................................................... 3-8
3.10 Changing General Settings ............................................................................................ 3-8
3.10.1 Setting up a Monitor ......................................................................................... 3-9
3.10.2 Changing Language ......................................................................................... 3-9
3.10.3 Adjusting the Screen Brightness ...................................................................... 3-9
3.10.4 Showing/Hiding the Help ................................................................................. 3-9
3.10.5 Setting the Date and Time .............................................................................. 3-10
3.10.6 Adjusting Volume ........................................................................................... 3-10
3.11 Setting Parameters ........................................................................................................3-11
1
Page 10
3.11.1 Switching the Parameters On/Off .................................................................... 3-11
3.11.2 Accessing the Parameters Menu ..................................................................... 3-12
3.12 Operating Modes ......................................................................................................... 3-12
3.12.1 Monitoring Mode ........................................................................................... 3-12
3.12.2 Night Mode .................................................................................................... 3-12
3.12.3 Privacy Mode ................................................................................................. 3-13
3.12.4 Demo Mode .................................................................................................... 3-14
3.12.5 Standby Mode ................................................................................................ 3-14
4 Managing Patients ................................................................................................................................................. 
4.1 Admitting a Patient .................................................................................................................................. 4-1
4.2 Quick Admitting a Patient ..................................................................................................................... 4-2
4.3 Setting the Monitor Location (Not Available for iMEC15S/iMEC15)................................... 4-2
4.4 Querying and Obtaining Patient Information .................................................................................. 4-2
4.5 Querying from Local Facility (Not Available for iMEC15S/iMEC15) .................................. 4-3
4.6 Associating Patient Information .......................................................................................................... 4-3
4.7 Editing Patient Information................................................................................................................... 4-3
4.8 Discharging a Patient .............................................................................................................................. 4-4
4.9 Transferring a Patient .............................................................................................................................. 4-4
4.9.1 Transferring Data from the Monitor to a USB Drive ...................................................... 4-5
4.9.2 Transferring Data from the USB Drive to Monitor ......................................................... 4-5
4.10 Connecting to a Central Monitoring System ................................................................................ 4-6
5 Managing Configurations ...................................................................................................... 5-1
5.1 Introduction ..................................................................................................................... 5-1
5.2 Entering the [Manage Configuration] Menu ................................................................... 5-1
5.3 Setting Default Configuration ......................................................................................... 5-2
5.4 Saving Current Settings ................................................................................................... 5-2
5.5 Editing Configuration ...................................................................................................... 5-3
5.6 Deleting a Configuration ................................................................................................. 5-3
5.7 Transferring a Configuration ........................................................................................... 5-3
5.8 Loading a Configuration ................................................................................................. 5-4
5.9 Restoring the Latest Configuration Automatically .......................................................... 5-4
5.10 Modifying Password ..................................................................................................... 5-5
6 User Screens ............................................................................................................................ 6-1
6.1 Tailoring Your Screens .................................................................................................... 6-1
6.1.1 Changing the Wave Line Size ............................................................................ 6-1
6.1.2 Changing Measurement Colors .......................................................................... 6-1
6.1.3 Changing Screen Layout .................................................................................... 6-1
6.2 Viewing Minitrends ......................................................................................................... 6-2
6.2.1 Having a Split-Screen View of Minitrends ........................................................ 6-2
6.2.2 Setting Minitrends .............................................................................................. 6-3
6.3 Viewing OxyCRG ........................................................................................................... 6-3
6.4 Viewing Other Patients.................................................................................................... 6-4
6.4
.1 Care Group ......................................................................................................... 6-4
2
Page 11
6.4.2 Viewing the Care Group Overview Bar ............................................................. 6-5
6.4.3 Understanding the View Other Patient Window ................................................ 6-5
6.5 Understanding the Big Numerics Screen ........................................................................ 6-7
7 Alarms ..................................................................................................................................... 7-1
7.1 Alarm Categories ............................................................................................................. 7-1
7.2 Alarm Levels ................................................................................................................... 7-2
7.3 Alarm Indicators .............................................................................................................. 7-2
7.3.1 Alarm Lamp ....................................................................................................... 7-2
7.3.2 Alarm Message ................................................................................................... 7-3
7.3.3 Flashing Numeric ............................................................................................... 7-3
7.3.4 Audible Alarm Tones .......................................................................................... 7-3
7.3.5 Alarm Status Symbols ........................................................................................ 7-4
7.4 Alarm Tone Configuration .............................................................................................. 7-5
7.4.1 Setting the Minimum Alarm Volume ................................................................. 7-5
7.4.2 Changing the Alarm Volume .............................................................................. 7-5
7.4.3 Setting the Interval between Alarm Sounds ....................................................... 7-5
7.4.4 Changing the Alarm Tone Pattern ...................................................................... 7-6
7.4.5 Setting the Reminder Tones ............................................................................... 7-6
7.5 Understanding the Alarm Setup Menu ............................................................................ 7-7
7.5.1 Setting Alarm Properties for All Parameters ...................................................... 7-8
7.5.2 Adjusting Alarm Limits Automatically .............................................................. 7-8
7.5.3 Setting Alarm Delay Time ................................................................................ 7-12
7.5.4 Setting SpO
Technical Alarm Delay ............................................................... 7-13
2
7.5.5 Setting Recording Length ................................................................................. 7-13
7.5.6 Intubation Mode ............................................................................................... 7-13
7.6 Pausing Alarms ............................................................................................................. 7-14
7.7 Switching Off All Alarms .............................................................................................. 7-15
7.8 Resetting Alarms ........................................................................................................... 7-15
7.9 Latching Alarms ............................................................................................................ 7-16
7.10 Testing Alarms ............................................................................................................ 7-17
7.11 Using Care Group Alarms ........................................................................................... 7-17
7.11.1 Care Group Auto Alarms ................................................................................ 7-17
7.11.2 Resetting Care Group Alarms ......................................................................... 7-18
7.11.3 Switching Off the Remote Device Disconnection Alarm (not available for
iMEC15S/iMEC15)................................................................................................... 7-18
7.11.4 Setting Care Group Alert Tone (for iMEC12/iMEC10/iMEC8/iMEC7/iMEC6/
iMEC5)...................................................................................................................... 7-19
7.11.5 Setting Care Group Alert Tone (for iMEC15S/iMEC15) ............................... 7-20
7.12 When an Alarm Occurs ............................................................................................... 7-20
8 Monitoring ECG .................................................................................................................... 8-1
8.1 Introduction ..................................................................................................................... 8-1
8.2 Safety .............................................................................................................................. 8-1
3
Page 12
8.3 Preparing to Monitor ECG .............................................................................................. 8-2
8.3.1 Preparing the Patient and Placing the Electrodes ............................................... 8-2
8.3.2 Choosing AHA or IEC Lead Placement ............................................................. 8-2
8.3.3 ECG Lead Placements ........................................................................................ 8-3
8.3.4 Checking Paced Status ....................................................................................... 8-5
8.4 Understanding the ECG Display ..................................................................................... 8-6
8.5 Changing ECG Settings .................................................................................................. 8-7
8.5.1 Accessing ECG Menus ....................................................................................... 8-7
8.5.2 Choosing the Alarm Source ................................................................................ 8-7
8.5.3 Setting the ECG Lead Set................................................................................... 8-7
8.5.4 Choosing an ECG Display Screen ..................................................................... 8-7
8.5.5 Changing the ECG Filter Settings ...................................................................... 8-8
8.5.6 Setting the Notch Filter ...................................................................................... 8-8
8.5.7 Changing the Pacer Reject Settings ................................................................... 8-9
8.5.8 Adjusting the Minimum QRS Detection Threshold (For Mindray ECG Algorithm,
not available for iMEC15S/iMEC15) ......................................................................... 8-9
8.5.9 Changing ECG Wave Settings ......................................................................... 8-10
8.5.10 Enabling Smart Lead Off ................................................................................ 8-11
8.5.11 Setting the Alarm Level for ECG Lead Off Alarms ........................................ 8-11
8.5.12 Adjusting QRS Volume ................................................................................... 8-11
8.5.13 About the Defibrillator Synchronization ......................................................... 8-11
8.6 About ST Monitoring .................................................................................................... 8-12
8.6.1 Switching ST On and Off ................................................................................. 8-12
8.6.2 Changing ST Filter Settings ............................................................................. 8-12
8.6.3 Understanding the ST Display ......................................................................... 8-13
8.6.4 Saving the Current ST Segment as Reference .................................................. 8-14
8.6.5 Changing the Reference Segment .................................................................... 8-14
8.6.6 Deleting a Reference Segment ......................................................................... 8-14
8.6.7 Recording the ST Segment ............................................................................... 8-15
8.6.8 Changing the ST Alarm Limits ........................................................................ 8-15
8.6.9 Setting the ST Alarm Delay Time .................................................................... 8-15
8.6.10 Adjusting ST Measurement Points ................................................................. 8-15
8.7 QT/QTc Interval Monitoring (For Mindray ECG Algorithm, not available for
iMEC15S/iMEC15) ............................................................................................................ 8-16
8.7.1 QT/QTc Monitoring Limitations ...................................................................... 8-17
8.7.2 Enabling QT/QTc Monitoring .......................................................................... 8-18
8.7.3 Displaying QT/QTc Parameters and Waveform ............................................... 8-18
8.7.4 Entering the QT View ....................................................................................... 8-19
8.7.5 Saving the Current QTc as Reference .............................................................. 8-19
8.7.6 Changing QT Settings ...................................................................................... 8-20
8.8 About Arrhythmia Monitoring ...................................................................................... 8-21
8.8.1 Understanding the Arrhythmia Events ............................................................. 8-21
8.8.2 Changing Arrhythmia Alarm Settings .............................................................. 8-22
8.8.3 Changing Arrhythmia Threshold Settings ........................................................ 8-23
4
Page 13
8.8.4 Setting the Extended Arrhythmia ..................................................................... 8-24
8.8.5 Reviewing Arrhythmia Events ......................................................................... 8-24
8.9 Viewing the ECG Summary (not available for iMEC15S/iMEC15) ............................ 8-24
8.10 ECG Relearning .......................................................................................................... 8-25
8.10.1 Initiating an ECG Relearning Manually ......................................................... 8-25
8.10.2 Automatic ECG Relearning ............................................................................ 8-25
8.11 12-Lead ECG Monitoring (for iMEC15S/iMEC15 only) ........................................... 8-26
8.11.1 Entering the 12-lead ECG Monitoring Screen ............................................... 8-26
8.11.2 Setting ECG Waveform Sequence .................................................................. 8-27
8.11.3 Extending the rhythm lead waveform area ..................................................... 8-27
8.12 Troubleshooting .......................................................................................................... 8-27
9 Monitoring Respiration (Resp) ............................................................................................. 9-1
9.1 Introduction ..................................................................................................................... 9-1
9.2 Safety Information .......................................................................................................... 9-1
9.3 Understanding the Resp Display ..................................................................................... 9-2
9.4 Placing Resp Electrodes .................................................................................................. 9-2
9.4.1 Optimizing Lead Placement for Resp ................................................................ 9-3
9.4.2 Cardiac Overlay ................................................................................................. 9-3
9.4.3 Abdominal Breathing ......................................................................................... 9-3
9.4.4 Lateral Chest Expansion..................................................................................... 9-4
9.5 Choosing the Respiration Lead ....................................................................................... 9-4
9.6 Changing the Apnea Alarm Delay ................................................................................... 9-4
9.7 Changing Resp Detection Mode ..................................................................................... 9-4
9.8 Changing Resp Wave Settings ........................................................................................ 9-5
9.9 Setting RR Source ........................................................................................................... 9-5
9.10 Setting alarm properties ................................................................................................ 9-6
9.11 Switching Resp Measurement On/Off ........................................................................... 9-6
10 Monitoring PR .................................................................................................................... 10-1
10.1 Introduction ................................................................................................................. 10-1
10.2 Setting the PR Source .................................................................................................. 10-1
10.3 Selecting the Active Alarm Source .............................................................................. 10-2
10.4 QRS Tone .................................................................................................................... 10-2
11 Monitoring SpO
................................................................................................................. 11-1
2
11.1 Introduction .................................................................................................................. 11-1
11.2 Safety ............................................................................................................................ 11-2
11.3 Identifying SpO
Modules ............................................................................................ 11-2
2
11.4 Applying the Sensor ..................................................................................................... 11-3
11.5 Changing SpO
11.5.1 Accessing SpO
Settings ............................................................................................... 11-3
2
Menus ................................................................................... 11-3
2
11.5.2 Adjusting the Desat Alarm .............................................................................. 11-3
11.5.3 Setting SpO
Sensitivity .................................................................................. 11-3
2
5
Page 14
11.5.4 Changing Averaging Time ............................................................................... 11-4
11.5.5 Monitoring SpO
and NIBP Simultaneously ................................................... 11-4
2
11.5.6 Sat-Seconds Alarm Management .................................................................... 11-4
11.5.7 Changing the Speed of the Pleth Wave............................................................ 11-5
11.5.8 Zooming PI Value ............................................................................................ 11-6
11.5.9 Setting the Alarm Level for SpO
11.5.10 Setting the SpO
Tone Mode ......................................................................... 11-6
2
Sensor Off Alarm ...................................... 11-6
2
11.6 Measurement Limitations ............................................................................................. 11-6
11.7 Masimo Information ..................................................................................................... 11-7
11.8 Nellcor Information ...................................................................................................... 11-7
11.9 Troubleshooting ............................................................................................................ 11-8
12 Monitoring NIBP ................................................................................................................ 12-1
12.1 Introduction ................................................................................................................. 12-1
12.2 Safety .......................................................................................................................... 12-2
12.3 Measurement Limitations ............................................................................................ 12-2
12.4 Measurement Methods ................................................................................................ 12-3
12.5 Setting Up the NIBP Measurement ............................................................................. 12-3
12.5.1 Preparing the Patient ...................................................................................... 12-3
12.5.2 Preparing to Measure NIBP ........................................................................... 12-3
12.5.3 Starting and Stopping Measurements ............................................................. 12-4
12.5.4 Correcting the Measurement if Limb is not at Heart Level ............................ 12-4
12.5.5 Enabling NIBP Auto Cycling and Setting the Interval ................................... 12-4
12.5.6 Starting a STAT Measurement ........................................................................ 12-5
12.5.7 Sequence Measurement˄not available for iMEC15S/iMEC15˅ ................ 12-6
12.6 Understanding the NIBP Numerics ............................................................................. 12-7
12.7 Changing NIBP Settings ............................................................................................. 12-7
12.7.1 Setting the Initial Cuff Inflation Pressure ....................................................... 12-7
12.7.2 Setting NIBP Alarm Properties ...................................................................... 12-8
12.7.3 Displaying NIBP List ..................................................................................... 12-8
12.7.4 Setting the Pressure Unit ................................................................................ 12-8
12.7.5 Switching On NIBP End Tone........................................................................ 12-8
12.8 Assisting Venous Puncture .......................................................................................... 12-9
13 Monitoring Temp ................................................................................................................ 13-1
13.1 Introduction ................................................................................................................. 13-1
13.2 Safety .......................................................................................................................... 13-1
13.3 Making a Temp Measurement ..................................................................................... 13-1
13.4 Understanding the Temp Display ................................................................................ 13-2
13.5 Changing Temperature Settings .................................................................................. 13-2
13.5.1 Setting the Temperature Unit ......................................................................... 13-2
13.5.2 Setting the Temperature Label (not available for iMEC15S
/iMEC15) .......... 13-2
14 Monitoring IBP .................................................................................................................. 14-1
6
Page 15
14.1 Introduction ................................................................................................................. 14-1
14.2 Safety .......................................................................................................................... 14-1
14.3 Measuring an Invasive Blood Pressure ....................................................................... 14-1
14.3.1 Setting Up the Pressure Measurement ............................................................ 14-1
14.3.2 Zeroing the Transducer .................................................................................. 14-2
14.4 Measuring ICP Using the Codman ICP Transducer (not available for iMEC15S/iMEC15)
............................................................................................................................................. 14-3
14.4.1 Zeroing the Codman ICP transducer .............................................................. 14-3
14.4.2 Measuring ICP ............................................................................................... 14-4
14.5 Understanding the IBP Display ................................................................................... 14-5
14.6 Changing IBP Settings ................................................................................................ 14-6
14.6.1 Changing a Pressure for Monitoring .............................................................. 14-6
14.6.2 Setting the Pressure Label Order .................................................................... 14-7
14.6.3 Setting Alarm Properties ................................................................................ 14-7
14.6.4 Changing Averaging Time .............................................................................. 14-7
14.6.5 Setting Up the IBP Wave ................................................................................ 14-7
14.6.6 Enabling PPV Measurement and Setting PPV Source ................................... 14-8
14.6.7 Setting the Pressure Unit ................................................................................ 14-8
14.7 Overlapping IBP Waveforms ...................................................................................... 14-8
14.8 Measuring PAWP ........................................................................................................ 14-9
14.8.1 Preparing to Measure PAWP ........................................................................ 14-10
14.8.2 Setting Up the PAWP Measurement .............................................................. 14-11
14.8.3 Understanding the PAWP Setup Menu ......................................................... 14-12
14.8.4 Performing Hemodynamic Calculation ........................................................ 14-12
14.9 Troubleshooting ........................................................................................................ 14-12
15 Monitoring Cardiac Output .............................................................................................. 15-1
15.1 Introduction ................................................................................................................. 15-1
15.2 Understanding the C.O. Display ................................................................................. 15-1
15.3 Influencing Factors...................................................................................................... 15-2
15.4 Setting Up the C.O. Measurement .............................................................................. 15-2
15.5 Measuring the Blood Temperature .............................................................................. 15-6
15.6 Changing C.O. Settings ............................................................................................... 15-6
15.6.1 Setting the Temperature Unit ......................................................................... 15-6
15.6.2 Setting Alarm Properties ................................................................................ 15-6
16 Monitoring Carbon Dioxide .............................................................................................. 16-1
16.1 Introduction ................................................................................................................. 16-1
16.2 Measuring CO
16.2.1 Making a Sidestream CO
16.2.2 Making a Microstream CO
16.2.3 Making a Mainstream CO
16.3 Changing CO
16.3.1 Accessing CO
............................................................................................................ 16-2
2
Settings ............................................................................................... 16-5
2
Menus .................................................................................... 16-5
2
Measurement ....................................................... 16-2
2
Measurement ..................................................... 16-3
2
Measurement ...................................................... 16-4
2
7
Page 16
16.3.2 Entering the Standby Mode ............................................................................ 16-5
16.3.3 Setting the CO
Unit ....................................................................................... 16-5
2
16.3.4 Setting up Gas Compensations ....................................................................... 16-6
16.3.5 Setting up Humidity Compensation ............................................................... 16-7
16.3.6 Setting the Apnea Alarm Delay ...................................................................... 16-7
16.3.7 Choosing a Time Interval for Peak-Picking ................................................... 16-8
16.3.8 Setting the Flow Rate ..................................................................................... 16-8
16.3.9 Setting up the CO
Wave ................................................................................ 16-8
2
16.4 Setting RR Source ....................................................................................................... 16-9
16.5 Setting Barometric Pressure Compensation ................................................................ 16-9
16.6 Measurement Limitations ............................................................................................ 16-9
16.7 Leakage test ............................................................................................................... 16-10
16.8 Troubleshooting the Sidestream CO
Sampling System ........................................... 16-10
2
16.9 Removing Exhaust Gases from the System .............................................................. 16-10
16.10 Zeroing the Sensor ................................................................................................... 16-11
16.10.1 For Sidestream and Microstream CO
16.10.2 For Mainstream CO
Modules .................................................................... 16-11
2
Modules .......................................... 16-11
2
16.11 Calibrating the Sensor ............................................................................................. 16-12
16.12 Oridion Information ................................................................................................ 16-12
17 Monitoring AG (for iMEC15S/iMEC15 only) ................................................................. 17-1
17.1 Introduction ................................................................................................................. 17-1
17.2 Identifying AG Module ............................................................................................... 17-2
17.2.1 Introduction of AG module ............................................................................ 17-2
17.2.2 Connection of AG module .............................................................................. 17-3
17.3 Understanding the AG Display ................................................................................... 17-3
17.4 MAC Values ................................................................................................................ 17-4
17.5 Preparing to Measure AG ............................................................................................ 17-5
17.6 Changing AG Settings ................................................................................................. 17-6
17.6.1 Setting Gas Unit ............................................................................................. 17-6
17.6.2 Setting the Apnea Alarm Delay ...................................................................... 17-7
17.6.3 Changing the Sample Flow Rate .................................................................... 17-7
17.6.4 Setting up the O
Compensation .................................................................... 17-7
2
17.6.5 Entering the Standby Mode ............................................................................ 17-8
17.6.6 Setting up the AG Wave ................................................................................. 17-8
17.6.7 Setting RR Source .......................................................................................... 17-8
17.7 Changing the Anesthetic Agent ................................................................................... 17-9
17.8 Measurement Limitations ............................................................................................ 17-9
17.9 Troubleshooting .......................................................................................................... 17-9
17.9.1 When the Gas Inlet is Blocked ....................................................................... 17-9
17.9.2 When an Internal Occlusion Occurs ............................................................. 17-10
17.10 Removing Exhaust Gases from the System ............................................................ 17-10
18 Clinical Score
(not available for iMEC15S/iMEC15)
8
...................................................... 18-1
Page 17
18.1 Overview ..................................................................................................................... 18-1
18.1.1 MEWS (Modified Early Warning Score) ....................................................... 18-1
18.1.2 NEWS (National Early Warning Score) ......................................................... 18-2
18.1.3 Customizable Score ........................................................................................ 18-2
18.2 Entering Score ............................................................................................................. 18-3
18.3 Calculating a Score ..................................................................................................... 18-3
18.4 Clinical Score Screen .................................................................................................. 18-4
18.4.1 Score Tile in the Main Screen ........................................................................ 18-4
18.4.2 Score Screen ................................................................................................... 18-5
18.5 Obtaining the Total Score ............................................................................................ 18-6
18.6 Setting the Interval of Calculating a Score .................................................................. 18-7
18.7 Managing Scores ......................................................................................................... 18-7
18.7.1 Importing the Score ........................................................................................ 18-7
18.7.2 Selecting Default Score .................................................................................. 18-7
18.7.3 Deleting the Score .......................................................................................... 18-8
18.7.4 Loading a Score .............................................................................................. 18-8
18.8 Setting Operator ID ..................................................................................................... 18-8
18.8.1 Selecting an Operator ID ................................................................................ 18-8
18.8.2 Adding an Operator ID ................................................................................... 18-9
18.8.3 Deleting Operator ID ...................................................................................... 18-9
18.8.4 Setting Operator ID Timeout .......................................................................... 18-9
18.9 Reviewing ................................................................................................................. 18-10
19 Freezing Waveforms .......................................................................................................... 19-1
19.1 Freezing Waveforms ................................................................................................... 19-1
19.2 Viewing Frozen Waveforms ........................................................................................ 19-1
19.3 Unfreezing Waveforms................................................................................................ 19-2
19.4 Recording Frozen Waveforms ..................................................................................... 19-2
20 Review ................................................................................................................................. 20-1
20.1 Accessing Respective Review Windows ..................................................................... 20-1
20.2 Reviewing Graphic Trends .......................................................................................... 20-1
20.3 Reviewing Tabular Trends .......................................................................................... 20-3
20.4 Events .......................................................................................................................... 20-5
20.4.1 Marking Events .............................................................................................. 20-5
20.4.2 Reviewing Events ........................................................................................... 20-5
20.5 Reviewing Waveforms ................................................................................................ 20-8
20.6 Reviewing OxyCRG ................................................................................................. 20-10
21 Calculations ........................................................................................................................ 21-1
21.1 Introduction ................................................................................................................. 21-1
21.2 Dose Calculations........................................................................................................ 21-2
21.2.1 Performing Calculations ................................................................................. 21-2
21.2.2 Selecting the Proper Drug Unit ...................................................................... 21-2
9
Page 18
21.2.3 Titration Table ................................................................................................ 21-3
21.2.4 Drug Calculation Formulas ............................................................................ 21-3
21.3 Oxygenation Calculations ........................................................................................... 21-4
21.3.1 Performing Calculations ................................................................................. 21-4
21.3.2 Entered Parameters ......................................................................................... 21-4
21.3.3 Calculated Parameters and Formulas ............................................................. 21-5
21.4 Ventilation Calculations .............................................................................................. 21-5
21.4.1 Performing Calculations ................................................................................. 21-5
21.4.2 Entered Parameters ......................................................................................... 21-6
21.4.3 Calculated Parameters and Formulas ............................................................. 21-6
21.5 Hemodynamic Calculations ........................................................................................ 21-7
21.5.1 Performing Calculations ................................................................................. 21-7
21.5.2 Entered Parameters ......................................................................................... 21-7
21.5.3 Calculated Parameters and Formulas ............................................................. 21-8
21.6 Renal Calculations ...................................................................................................... 21-9
21.6.1 Performing Calculations ................................................................................. 21-9
21.6.2 Entered Parameters ......................................................................................... 21-9
21.6.3 Calculated Parameters and Formulas ........................................................... 21-10
21.7 Understanding the Review Window .......................................................................... 21-10
22 Recording ............................................................................................................................ 22-1
22.1 Using a Recorder ......................................................................................................... 22-1
22.2 Overview of Recording Types ..................................................................................... 22-2
22.3 Starting and Stopping Recordings ............................................................................... 22-2
22.4 Setting up the Recorder ............................................................................................... 22-3
22.4.1 Accessing the Record Setup Menu ................................................................. 22-3
22.4.2 Selecting Waveforms for Recording .............................................................. 22-3
22.4.3 Setting the Realtime Recording Length ......................................................... 22-3
22.4.4 Setting the Interval between Timed Recordings ............................................. 22-3
22.4.5 Changing the Recording Speed ...................................................................... 22-3
22.4.6 Setting the IBP Wave Overlap Recordings ..................................................... 22-4
22.4.7 Clearing Recording Tasks .............................................................................. 22-4
22.5 Loading Paper ............................................................................................................. 22-4
22.6 Removing Paper Jam ................................................................................................... 22-5
22.7 Cleaning the Recorder Printhead ................................................................................. 22-5
23 Printing ............................................................................................................................... 23-1
23.1 Printer .......................................................................................................................... 23-1
23.2 Connecting a printer .................................................................................................... 23-1
23.3 Setting Up the Printer .................................................................................................. 23-1
23.4 Starting Report Printouts ............................................................................................. 23-2
23.5 Stopping Reports Printouts ......................................................................................... 23-2
23.6 Setting Up Reports ...................................................................................................... 23-3
23.6.1 Setting Up ECG Reports ................................................................................ 23-3
10
Page 19
23.6.2 Setting Up Tabular Trends Reports ................................................................ 23-3
23.6.3 Setting Up Graphic Trends Reports ............................................................... 23-4
23.6.4 Setting Up Realtime Reports .......................................................................... 23-4
23.7 End Case Reports ........................................................................................................ 23-4
23.8 Printer Statuses ............................................................................................................ 23-5
23.8.1 Printer Out of Paper ....................................................................................... 23-5
23.8.2 Printer Status Messages .................................................................................. 23-5
24 Other Functions .................................................................................................................. 24-1
24.1 Analog Output ............................................................................................................. 24-1
24.2 Exporting the Log ....................................................................................................... 24-1
24.3 Transferring Data ........................................................................................................ 24-1
24.3.1 Data Export System ........................................................................................ 24-1
24.3.2 Transferring Data by Different Means ........................................................... 24-2
24.4 Nurse Call ................................................................................................................... 24-2
24.5 Network Connection ................................................................................................... 24-4
24.5.1 Setting the Monitor Network .......................................................................... 24-4
24.5.2 Wireless Network ........................................................................................... 24-4
24.5.3 WLAN Test .................................................................................................... 24-4
24.5.4 WLAN Setup (Not Available for iMEC15S/iMEC15) ................................... 24-5
24.5.5 Viewing the MAC Address (Not Available for iMEC15S/iMEC15) .............. 24-5
24.5.6 Enabling the Data Encryption (Not Available for iMEC15S/iMEC15) ......... 24-5
24.5.7 Settting DNS (Not Available for iMEC15S/iMEC15) ................................... 24-6
24.5.8 Certificates Maintenance (Not Available for iMEC15S/iMEC15) ................. 24-6
24.5.9 Setting the Multicast Parameters .................................................................... 24-7
24.5.10 Connecting the monitor to the CMS ............................................................. 24-7
25 Batteries .............................................................................................................................. 25-1
25.1 Overview ..................................................................................................................... 25-1
25.2 Replacing a Battery ..................................................................................................... 25-2
25.3 Battery Guidelines ....................................................................................................... 25-2
25.4 Battery Maintenance ................................................................................................... 25-3
25.5 Battery Recycling ........................................................................................................ 25-4
26 Care and Cleaning .............................................................................................................. 26-1
26.1 General Points ............................................................................................................. 26-1
26.2 Cleaning ...................................................................................................................... 26-2
26.3 Disinfection ................................................................................................................. 26-3
26.4 Sterilization ................................................................................................................. 26-3
27 Maintenance ....................................................................................................................... 27-1
27.1 Regular Inspection ...................................................................................................... 27-1
27.2 Maintenance and Testing Schedule ............................................................................. 27-2
27.3 Checking Monitor and Module Information ............................................................... 27-3
27.4 Calibrating ECG .......................................................................................................... 27-3
27.5 NIBP Tests ................................................................................................................... 27-4
27.
5.1 NIBP Leakage Test ......................................................................................... 27-4
11
Page 20
27.6 NIBP Accuracy Test .................................................................................................... 27-5
Tests ..................................................................................................................... 27-6
27.7 CO
2
27.7.1 CO
27.7.2 CO
27.7.3 Calibrating CO
Leakage Test ........................................................................................... 27-6
2
Accuracy Test ......................................................................................... 27-6
2
.............................................................................................. 27-7
2
27.8 AG Tests ...................................................................................................................... 27-9
27.8.1 AG Leakage Test ............................................................................................ 27-9
27.8.2 AG Accuracy Test ........................................................................................... 27-9
27.8.3 AG Calibration ............................................................................................. 27-10
27.9 Calibrating the Touchscreen ...................................................................................... 27-12
27.10 Electrical Safety Tests ............................................................................................. 27-13
28 Accessories .......................................................................................................................... 28-1
28.1 ECG Accessories ......................................................................................................... 28-1
28.2 SpO
Accessories ........................................................................................................ 28-4
2
28.3 NIBP Accessories ........................................................................................................ 28-6
28.4 Temp Accessories ........................................................................................................ 28-7
28.5 IBP/ICP Accessories ................................................................................................... 28-8
28.6 C.O. Accessories ......................................................................................................... 28-9
Accessories ........................................................................................................ 28-10
28.7 CO
2
28.8 AG Accessories(for iMEC15S/iMEC15 only) ........................................................... 28-11
28.9 Others ........................................................................................................................ 28-12
A Product Specifications .......................................................................................................... A-1
A.1 Monitor Safety Specifications ....................................................................................... A-1
A.2 Power Supply Specifications ......................................................................................... A-2
A.3 Physical Specifications .................................................................................................. A-3
A.4 Hardware Specifications ............................................................................................... A-3
A.5 Data Storage .................................................................................................................. A-6
A.6 Wireless Network .......................................................................................................... A-6
A.7 Measurement Specifications ......................................................................................... A-6
B EMC and Radio Regulatory Compliance ............................................................................ B-1
B.1 EMC ............................................................................................................................... B-1
B.2 Radio Regulatory Compliance........................................................................................ B-6
C Default Configurations ........................................................................................................ C-1
C.1 Parameters Configuration ............................................................................................... C-1
C.2 Routine Configuration .................................................................................................. C-14
C.3 User Maintenance Items ............................................................................................... C-18
D Alarm Messages .................................................................................................................... D-1
D.1 Physiological Alarm Messages ...................................................................................... D-1
D.2 Technical Alarm Messages ............................................................................................ D-3
12
Page 21
E Electrical Safety Inspection .................................................................................................. E-1
E.1 Power Cord Plug ............................................................................................................. E-1
E.2 Device Enclosure and Accessories ................................................................................. E-2
E.3 Device Labelling ............................................................................................................. E-2
E.4 Protective Earth Resistance ............................................................................................ E-3
E.5 Earth Leakage Test .......................................................................................................... E-3
E.6 Patient Leakage Current ................................................................................................. E-4
E.7 Mains on Applied Part Leakage ...................................................................................... E-4
E.8 Patient Auxiliary Current ................................................................................................ E-5
F Symbols and Abbreviations ................................................................................................... F-1
F.1 Symbols ........................................................................................................................... F-1
F.2 Abbreviations .................................................................................................................. F-3
G Declaration of Conformity .................................................................................................. G-1
13
Page 22
FOR YOUR NOTES
14
Page 23
WARNING
Indicates a potential hazard or unsafe practice that, if not avoided, could result in
CAUTION
potential hazard or unsafe practice that, if not avoided, could result in
Provides application tips or other useful information to ensure that you get the most
1 Safety
1.1 Safety Information
z
death or serious injury.
z Indicates a
minor personal injury or product/property damage.
NOTE
z
from your product.
1-1
Page 24
WARNING
ric shock, this equipment must only be connected to a supply
If the installation does not provide for a protective earth
rich
the performance specification due to aging,
patient. Remember that alarm settings should be customized according to different
To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling
Ensure that the patient monitor is supplied with continuous electric power during
s metal parts or connectors when in contact with the
1.1.1 Warnings
z This equipment is used to one patient at a time.
z Before putting the system into operation, the operator must verify that the
equipment, connecting cables and accessories are in correct working order and
operating condition.
z To avoid risk of elect
mains with protective earth.
conductor, disconnect it from the power line and operate it on battery power, if
possible.
z To avoid explosion hazard, do not use the equipment in the presence of oxygen-
atmospheres, flammable anesthetics, or other flammable agents (such as
gasoline).Use and store the equipment in specified environmental condition. The
monitor and accessesories may not meet
stored or used outside the specified temperature and humidity range.
z Do not open the equipment housings. All servicing and future upgrades must be
carried out by the personnel trained and authorized by our company only.
z Do not come into contact with patients during defibrillation. Otherwise serious
injury or death could result.
z Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level or off may result in a hazard to the
patient situations and always keeping the patient under close surveillance is the
most reliable way for safe patient monitoring.
z The physiological data and alarm messages displayed on the equipment are for
reference only and cannot be directly used for diagnostic interpretation.
z
hazard. Wrap and secure excess cabling to reduce risk of entanglement or
strangulation by patients or personnel.
z When disposing of the package material, be sure to observe the applicable waste
control regulations and keep it out of childrens reach.
z
work. Sudden power failure leads to the loss of patient data.
z Do not touch the equipment
patient; otherwise patient injury may result.
z Never mix patient electrode types or brands. Dissimilar metals or other
incompatibilities may cause considerable baseline drift and may increase trace
recovery time after defibrillation.
1-2
Page 25
CAUTION
, please
. For this reason make sure that all external devices
ray equipment or MRI devices are a possible source
Always install or carry the equipment properly to avoid damage caused by drop,
ns plug as isolation means to the mains power supply. Do
of
This manual describes all features and options. Your equipment may not have all of
1.1.2 Cautions
z To ensure patient safety, use only parts and accessories specified in this manual.
z At the end of its service life, the equipment, as well as its accessories, must be
disposed of in compliance with the guidelines regulating the disposal of such
products. If you have any questions concerning disposal of the equipment
contact us.
z Magnetic and electrical fields are capable of interfering with the proper
performance of the equipment
operated in the vicinity of the equipment comply with the relevant EMC
requirements. Mobile phone, X-
of interference as they may emit higher levels of electromagnetic radiation.
z Before connecting the equipment to the power line, check that the voltage and
frequency ratings of the power line are the same as those indicated on the
equipments label or in this manual.
z
impact, strong vibration or other mechanical force.
z Dry the equipment immediately in case of rain or water spray.
1.1.3 Notes
NOTE
z Put the equipment in a location where you can easily view and operate the
equipment.
z The equipment use a mai
not locate the equipment in a place difficult to operate the mains plug.
z In normal use, the operator shall stand in front of the equipment.
z Keep this manual in the vicinity of the equipment so that it can be obtained
conveniently when needed.
z The software was developed in compliance with IEC60601-1-4. The possibility
hazards arising from software errors is minimized.
z
them.
1-3
Page 26
General warning sign
Refer to instruction
manual/booklet
Power ON/OFF (for a part
of the equipment)
Battery
A
ALARM PAUSED
Alarm
Graphical recorder
Freeze/unfreeze waveforms
Main menu
NIBP start/stop key
Output
Equipotentiality
VGA output USB connector
Network connector
Gas outlet
Serial number
Inserted direction
Protection against vertically
falling water drops
Manufacturer
D
The
of
the Council Directive 93/42/EEC concerning medical devices and fulfils the
essential requirement
Note: The
AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY
DEFIBRILLATION
1.2 Equipment Symbols
NOTE
z Some symbols may not appear on your equipment.
lternating current
reset
product bears CE mark indicating its conformity with the provisions
s of Annex I of this directive.
product complies with the Council Directive 2011/65/EU.
indicator
ate of manufacture
Non-ionizing electromagnetic radiation
-PROOF TYPE CF APPLIED PART
1-4
Page 27
DEFIBRILLATION
The following definition of the WEEE label applies to EU member states
only.
This symbol indicates that this product should not be treated as household
waste. By ensuring that this product is disposed of correctly, you will help
prevent bringing potential negative conse
human health. For more detailed information with regard to returning and
recycling this product, please consult the distributor from whom you
purchased it.
* For system products, this label may be attached to the main unit only.
-PROOF TYPE BF APPLIED PART
quences to the environment and
1-5
Page 28
FOR YOUR NOTES
1-6
Page 29
WARNING
This patient monitor is intended for use only by clinical professionals or under their
guidance. It must only be used by persons who have received adequate training in its
2 The Basics
2.1 Monitor Description
2.1.1 Intended Use
The iMEC patient monitors are intended to be used for monitoring, displaying, reviewing,
storing and transferring of multiple physiological parameters including ECG, heart rate (HR),
respiration (Resp), temperature (Temp), pulse oxygen saturation (SpO
non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.),
carbon dioxide (CO
The monitors are to be used in healthcare facilities by clinical professionals or under their
direction.
z
use. Anyone unauthorized or untrained must not perform any operation on it.
) and anesthetic gas (AG) (for iMEC15S/iMEC15 only).
2
), pulse rate (PR),
2
2.1.2 Applied Parts
The applied parts of the iMEC series patient monitors are:
ECG electrodes and leadwires,
SpO
NIBP cuff
Temp probes
IBP/ICP transducer,
C.O. sensor
CO
AG sampling line, water trap, and airway adapter
sensor
2
sampling line/Nasal sampling cannula, water trap, airway adapter, mainstream
2
sensor, and mask
2-1
Page 30
2.2 Front View
6
1
7
8
2
3
1. Alarm lamp
When a physiological alarm or technical alarm occurs, this lamp will flash as defined
below.
High level alarms: the lamp quickly flashes red.
Medium level alarms: the lamp slowly flashes yellow.
Low level alarms: the lamp lights yellow without flashing.
2. Display Screen
3. Power On/Off Switch
Pressing this switch turns the patient monitor on.
When the monitor is on, pressing and holding this switch turns the monitor off.
An indicator is built in this switch. It turns on when the patient monitor is on and turns
off when the patient monitor is off.
9
10
5
4
11
12
2-2
Page 31
4. AC power LED
It turns on when AC power is connected.
5. Battery LED
On: when the battery is installed and the AC source is connected.
Off: when no battery is installed or the installed battery is malfunction, or no AC
source is connected when the patient monitor is power off.
Flash:when the patient monitor operates on battery power.
6.
Press to reset alarms.
7.
Press to pause or restore alarms.
8.
Press to freeze or unfreeze waveforms.
9. Knob
Rotate the Knob clockwise or anti-clockwise. With each click, the highlight jumps to the
neighboring item. When you reach your desired item, press the Knob to select it.
10.
If no menu is displayed on the screen, pressing it will enter the main menu. If there is a
menu displayed on the screen, pressing it will close that menu.
Press to start or stop recordings.
11.
12.
Press to start or stop NIBP measurements.
2-3
Page 32
1
2
3
4
5 6 7
8
9
10
11
2.3 Side View
2.3.1 iMEC15S/iMEC15/iMEC12/iMEC10/iMEC7/iMEC6
1. Recorder 2. Battery compartment
3. Connector for C.O. cable 4. CO
5. Slot for CO
7. Connector for SpO
watertrap (sidestream) 6. Connector for ECG cable
2
cable 8. Connector for NIBP cuff
2
gas outlet
2
9. Connector for Temp probe 1 10. Connector for Temp probe 2
11 Connector for IBP cable
2-4
Page 33
2.3.2 iMEC8/iMEC5
1
2 3 4
5
6
7
8
9
10
11
1. Recorder 2. Battery compartment
3. Connector for C.O. cable 4. CO
5. Slot for CO
7. Connector for SpO
watertrap (sidestream) 6. Connector for ECG cable
2
cable 8. Connector for NIBP cuff
2
gas outlet
2
9. Connector for Temp probe 1 10. Connector for Temp probe 2
11 Connector for IBP cable
2-5
Page 34
2.4 Rear View
iMEC15S/iMEC15/iMEC12/iMEC10/iMEC7/iMEC6 iMEC8/iMEC5
1 2
3
4
1. Equipotential Grounding Terminal
When the patient monitor and other devices are to be used together, their equipotential
grounding terminals should be connected together, eliminating the potential difference
between them.
2. AC Power Input
3. Multifunctional Connector
It outputs defibrillator synchronization signals, nurse call signals and analogy output
signals. For iMEC15S/iMEC15, it also can connect the AG module.
4. VGA Connector
It connects a secondary display, which extends the display capability of your monitor.
The contents displayed on the secondary display screen accords with those displayed on
the monitor screen.
5. USB Connector
It connects the USB devices, such as a barcode scanner.
6. Network Connector
1
5
6
2
2-6
Page 35

V
U
F
D
U
a
a
U
Z
F
H
S
V
D
g
Z
W
W
S
V
D
U
\
H
D
V
/
W
R
\
D
Q
H
V
D
Q
S
G
V
&
P
D
H
P
K
G
P
H
W
,WLVD
SDWLHQW
&,6WK
,WDOVR
'LVSO
7KLVSDWLHQW
ZDYHIRUPV
WDQGDUG5-
PRQLWRUIRU
RXJKH*DWH
RQQHFWVWKH
\6FUH
PRQLWRUDGR
$W\SLFDOGL

FRQQHFWRU
HPRWHYLHZ
D\
SDWLHQWPRQL
Q
WVDKLJKUH
SOD\VFUHHQL

KLFKFRQQHF
WKURXJKWKH
RUWR3&IRU
ROXWLRQ7)7
VVKRZQEHO
WVWKHSDWLHQW
$1RU+/
V\VWHPXSJU
/&'WRGLVS
Z
PRQLWRUWR
EDVHGV\VWH
GH
OD\SDWLHQWS
06RURWKHU
VVXFK+,6
UDPHWHUVDQ
DQG



 3DWLH QW
7KLV D
SDWLHQW
SDWLHQW
WKH>P
WKH>P
$ODUP 
,QIRUPDWLRQ
HDVKRZVWKH
QDPHDQGS
LQIRUPDWLRQ
tient Setup@
tient Demo
6\PEROV
LQGL FDWHV
LQGL FDWHV
$UH
SDWLHQWLQIR
WLHQWFDWHJRU
LVLQFRPSOHW
PHQX,IDS
raphics@PH
DODUPVDUHS
DODUPDUHUH
PDWLRQVXFK
 LQGL F
,IQRSDWLH
DWLHQWKDVEH
QX
DXVHG
HW
DVGHSDUWPH
WHVWKDWQR
WLVDGPLWWH
QDGPLWWHG
WEHGQXPE
DWLHQWLVDG
VHOHFWLQJW
HOHFWLQJWKLV

URRPQX
LWWHGRUWKH
LVDUHDZLOO
DUHDZLOOHQ
EHU
QWHU
HU

Page 36
indicates alarm sounds are turned off.
indicates the system is in alarm off status.
3. Technical Alarm Area
This area shows technical alarm messages and prompt messages. When multiple
messages come, they will be displayed circularly. Select this area and the technical
alarm list will be displayed.
4. Physiological Alarm Area
This area shows physiological alarm messages. When multiple alarms occur, they will
be displayed circularly. Select this area and the physiological alarm list will be
displayed.
5. Waveform Area
This area shows measurement waveforms. The waveform name is displayed at the left
upper corner of the waveform. Select this area and the corresponding measurement
setup menu will be displayed.
6. Parameter Area A
This area shows measurement parameters. Each monitored parameter has a parameter
window and the parameter name is displayed at the upper left corner. The corresponding
waveform of each parameter is displayed in the same row in the waveform area. Select
this area and the corresponding measurement setup menu will be displayed.
7. Parameter Area B
For the parameters displayed in this area, their corresponding waveforms are not
displayed.
8. Prompt Message Area
This area shows the current configuration name, prompt messages, network status icons,
battery status icons, etc. For details about battery status symbols, refer to the 25
Batteries.
indicates patient monitor is connected to a wired network successfully.
indicates the patient monitor has failed to connect a wired network.
indicates the wireless function is working.
indicates the wireless function is not working.
indicates a USB disk is inserted.
2-8
Page 37
Scroll left to display more QuickKeys.
Scroll right to display more QuickKeys.
Enter the main menu
Enter standby mode
C
Review the patient’s data
Enter the Stop all NIBP measurement
Zero IBP
Reset the alarm system
P
Change screen
Enter the patient setup menu
Trigger a manual
Have a split
Enter the
9. QuickKeys Area
This area contains QuickKeys that give you fast access to functions.
2.6 QuickKeys
A QuickKey is a configurable graphical key, located at the bottom of the main screen. They
give you fast access to functions. Their availability and the order in which they appear on
your screen, depend on how your patient monitor is configured.
The following QuickKeys can be displayed on the screen:
hange alarm settings
NIBP measurement menu
ause or restore alarms
event
-screen view of minitrends
volume setup menu
2-9
Page 38
Default configurations
Start cardiac output procedure
Perform calculations
Have a split
Have a split
Enter the full
Enter the
Start NIBP STAT measurement
Enter the [
Enter the PAWP measurement screen
Enter the privacy mode
Enter the
Enter the intubation mode
-screen view of another patient’s conditions
-screen view of OxyCRG trends
-screen 7-lead ECG screen
[Parameters] menu
Unit Setup] menu
night mode
You can also select your desired QuickKeys to display on the screen.
1. Select [Main Menu]→[Maintenance >>]→[Manage Configuration >>]→enter the
required password→[Ok].
2. In the [Manage Configuration] menu, select [Edit Config.>>].
3. In the pop-up menu, select the desired configuration and then select [Edit].
4. In the pop-up menu, select [Screen Setup >>].
5. In the [Select QuickKeys] screen, select your desired QuickKeys and the order of them.
2-10
Page 39
WARNING
organization or
Devices connected to the equipment must meet the requirements of the applicable
1 safety standards for medical electrical equipment). The
rsonnel who connect devices to the equipment’s
1. If
leakage currents, please consult the manufacturers or else an expert in the field, to
material as they can be used if the equipment
3 Basic Operations
3.1 Installation
z The equipment shall be installed by personnel authorized by us.
z The software copyright of the equipment is solely owned by us. No
individual shall resort to juggling, copying, or exchanging it or to any other
infringement on it in any form or by any means without due permission.
z
IEC standards (e.g. IEC 60950 safety standards for information technology
equipment and IEC 60601-
system configuration must meet the requirements of the IEC 60601-1 medical
electrical systems standard. Any pe
signal input/output port is responsible for providing evidence that the safety
certification of the devices has been performed in accordance to the IEC 60601-
you have any question, please contact us.
z If it is not evident from the equipment specifications whether a particular
combination with other devices is hazardous, for example, due to summation of
ensure the necessary safety of patients and all devices concerned will not be
impaired by the proposed combination.
3.1.1 Unpacking and Checking
Before unpacking, examine the packing case carefully for signs of damage. If any damage is
detected, contact the carrier or us.
If the packing case is intact, open the package and remove the equipment and accessories
carefully. Check all materials against the packing list and check for any mechanical damage.
Contact us in case of any problem.
NOTE
z Save the packing case and packaging
must be reshipped.
3-1
Page 40
WARNING
When disposing of the packaging material, be sure to observe the applicable waste
during storage and transport. Before use,
please verify whether the packages are intact, especially the packages of single use
WARNING
requirements. Otherwise unexpected consequences, e.g. damage to the equipment,
The equipment use a mains plug as isolation means to the mains power supply. Do
z
control regulations and keep it out of children’s reach.
z The equipment might be contaminated
accessories. In case of any damage, do not apply it to patients.
3.1.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in this
manual.
The environment where the equipment is used shall be reasonably free from noises, vibration,
dust, corrosive, flammable and explosive substances. If the equipment is installed in a cabinet,
sufficient space in front and behind shall be left for convenient operation, maintenance and
repair. Moreover, to maintain good ventilation, the equipment shall be at least 2 inches (5cm)
away from around the cabinet.
When the equipment is moved from one place to another, condensation may occur as a result
of temperature or humidity difference. In this case, never start the system before the
condensation disappears.
z Make sure that the operating environment of the equipment meets the specific
could result.
z
not locate the equipment in a place difficult to operate the mains plug.
3-2
Page 41
WARNING
Contact your
3.2 Getting Started
3.2.1 Turning Power On
Once the patient monitor is installed, you can get ready for monitoring:
1. Before you start to make measurements, check the patient monitor for any mechanical
damage and make sure that all external cables, plug-ins and accessories are properly
connected.
2. Plug the power cord into the AC power source. If you run the patient monitor on battery
power, ensure that the battery is sufficiently charged.
3. Press the power on/off switch on the monitor’s front.
z Do not use the patient monitor for any monitoring procedure on a patient if you
suspect it is not working properly, or if it is mechanically damaged.
service personnel or us.
NOTE
z Check that visual and auditory alarm signals are presented correctly when the
equipment is powered on. Do not use the patient monitor for any monitoring
procedure on a patient if you suspect it is not working properly, or if it is
mechanically damaged. Contact your service personnel or us.
3.2.2 Starting Monitoring
1. Decide which measurements you want to make.
2. Connect the required patient cables and sensors.
3. Check that the patient cables and sensors are correctly connected.
4. Check that the patient settings such as [Patient Cat.], [Paced], etc, are appropriate for
5. Refer to the appropriate measurement section for details of how to perform the
your patient.
measurements you require.
3-3
Page 42
CAUTION
for 10
to forcibly shut down the monitor when it could not be shut down normally
of the patient monitor.
3.3 Disconnecting from Power
To disconnect the patient monitor from the AC power source, follow this procedure:
1. Confirm that the patient monitoring is finished.
2. Disconnect patient cables and sensors from the patient.
3. Make sure to save or clear the patient monitoring data as required.
4. Press and hold the power on/off switch. The patient monitor shuts down and you can
unplug the power cable.
z Although not recommended, you can press and hold the power on/off switch
seconds
or under some special situations. This may cause loss of data
NOTE
z To completely disconnect the power supply, unplug the power cord.
3.4 Using Keys
The monitor has three types of keys:
Softkey: A softkey is a graphic key on the screen, giving you fast access to certain
menus or functions. The monitor has two types of softkeys:
Parameter keys: Each parameter area or waveform area can be seen as a softkey.
You can enter a parameter setup menu by selecting its corresponding parameter
area or waveform area.
QuickKeys: QuickKeys are configurable graphical keys, located at the bottom of
the main screen. For details, refer to the section 2.6 .
Hardkeys: A hardkey is a physical key on a monitoring device, such as the main menu
hardkey on the monitor’s front.
Pop-Up Keys: Pop-up keys are task-related keys that appear automatically on the
monitor screen when required. For example, the confirm pop-up key appears only when
you need to confirm a change.
3-4
Page 43
Area C
Area B
Area A
3.5 Using the Touchscreen
Select screen items by pressing them directly on the patient monitor’s screen. You can enable
or disable touchscreen operation by pressing and holding the [Main Menu] QuickKey for 3
seconds. A padlock symbol
is displayed if touchscreen operation is disabled.
3.6 Setting the Screen
You can enter the [Screen Setup] window as shown below by selecting [Main Menu]ė [Screen Setup]ė[Screen Layout >>]. In this window, you can allocate the positions of the
parameters and waveforms. The parameters or waveforms whose positions are not allocated
will not be displayed.
The ECG parameter and the first ECG waveform always display in the first row. The
configurable areas can be classified as Area A, Area B, and Area C.
In Area A, you can choose to display the parameters (having waveforms) and their
waveforms. Each parameter and the associated waveform are displayed in the same row.
In Area B, you can choose to display the parameters and their waveforms. When there is
no parameter displayed in area C, both the parameters and their waveforms will be
displayed in area B. Otherwise, only the parameters will be displayed.
In Area C, you can choose to display Timer and all the parameters whose associated
waveforms will not be displayed.
The screen can automatically adjust to ensure the best view based on your screen setup.
3-5
Page 44
WARNING
] window will
If no corresponding parameter or waveform is displayed on the monitor screen, you should
perform the following inspections:
Check the connection of the lead, cable, or sensor.
Enter the [Screen Setup] window for the desired display configuration.
Check that the parameter is switched on in [Parameters Switch] window.
z The parameters whose positions are not allocated in the [Screen Setup
not be displayed. However, the monitor can still give alarms of these parameters.
3.7 Using the On-screen Keyboard
The onscreen keyboard enables you to enter information.
Use the
Use the
Select the
Select the
Select the
key to delete the previously entered character.
key to toggle between uppercase and lowercase letters.
key to confirm what you have entered and close the onscreen keyboard.
to access the symbol keyboard.
to exit the symbol keyboard.
3-6
Page 45
Corresponding events are created when the timer is started, paused or stopped. You
3.8 Using the Timer
To display the timer in the main screen, follow this procedure:
1. Select [Main Menu]ė[Screen Setup>>]ė[Screen Layout >>] to access the [Screens]
window.
2. Select [Screen Setup] tab.
3. In the Area C, select [Timer] from the drop-down list of the desired parameter area.
Refer to 3.6 Setting the Screen.
4. Select
Select [Start] or [Pause] to start or pause the timing.
Select [Clear] to clear current timing result.
Select [Setup] to access the [Timer Setup] window, in which you can set the
to exit the window. The main screen will display the timer.
[Direction] to [Up] or [Down]. If you select [Down], you should set:
[Run Time(h:min:s)]: The available time range is 0 to 100 hours, and the default
time is 5 minutes.
[Reminder Vol]: During the last 10 seconds of the countdown, the system issues
reminder tone. The available volume range is 0 to 10. 0 means off, and 10 the
maximum volume.
Note
z
can review these events in the [Events] page of the [Review] window.
3-7
Page 46
1
2
3
4
3.9 Using the Main Menu
To enter the main menu, select [Main Menu] QuickKey or the hardkey on the monitor’s
front. Most of monitor operations and settings can be performed through the main menu.
Other menus are similar to the main menu and contain the following parts:
1. Heading: gives a sum-up for the current menu.
2. Main body: displays options, buttons, prompt messages, etc. The menu button with “>>’’
enlarges a secondary window to reveal more options or information.
3. Online help area: displays help information for the highlighted menu item.
: select to exit the current menu.
4.
3.10 Changing General Settings
This chapter covers only general settings such as language, brightness, date and time, etc.
Measurement settings and other settings can be referred to in respective sections.
3-8
Page 47
3.10.1 Setting up a Monitor
In situations where you install a patient monitor or change the patient monitor’s application
site, you need to setup the patient monitor as follows:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. In the [User Maintenance] menu, select, in turn, [Monitor Name], [Department] and
[Bed No.], and then change their settings.
3.10.2 Changing Language
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. In the [User Maintenance] menu, select [Language] and then select the desired
language.
3. Restart the patient monitor.
3.10.3 Adjusting the Screen Brightness
1. Select the [Main Menu]→[Screen Setup >>]→[Brightness].
2. Select the appropriate setting for the screen brightness. 10 is the brightest, and 1 is the
least bright.
If the patient monitor operates on battery power, you can set a less bright screen to prolong
the operating time of the battery. When the patient monitor enters standby mode, the screen
will change to the least brightness automatically.
3.10.4 Showing/Hiding the Help
The patient monitor provides online help information. The user can display or hide the help
as required.
1. Select [Main Menu][Screen Setup >>].
2. Select [Help] and toggle between [On] and [Off].
3-9
Page 48
CAUTION
Changing date and time will affect the storage of trends and events and may cause
3.10.5 Setting the Date and Time
1. Select [Main Menu] →[Maintenance >>]→[System Time >>].
2. Set the date and time.
3. Select [Date Format] and toggle between [yyyy-mm-dd], [mm-dd-yyyy] and
[dd-mm-yyyy].
4. Select [Time Format] and toggle between [24h] and [12h].
If your patient monitor is connected to a central monitoring system (CMS), the date and time
are automatically taken from that CMS. In that case, you cannot change the date and time
settings on your patient monitor.
z
data missing.
3.10.6 Adjusting Volume
Alarm Volume
1. Select the [Alarm Setup] QuickKey[Others], or [Main Menu][Alarm
Setup >>][Others].
2. Select [Alm Volume] and then select the appropriate volume: X-10, in which X is the
minimum volume, depending on the set minimum alarm volume (refer to the chapter
Alarms), and 10 the maximum volume.
Key Volume
When you press the navigation knob or the touchscreen, or the hardkeys on the panel, the
patient monitor prompts you by making a sound of the key volume you have set.
1. Select the [Volume Setup] QuickKey, or [Main Menu]→[Screen Setup >>].
2. Select [Key Volume] and then select the appropriate volume. 0 means off, and 10 the
maximum volume.
3-10
Page 49
QRS Volume
The QRS tone is derived from either the HR or PR, depending on which is currently selected
as the alarm source in [ECG Setup] or [SpO2 Setup]. When monitoring SpO
variable pitch tone which changes as the patient’s saturation level changes. The pitch of the
tone rises as the saturation level increases and falls as the saturation level decreases. The
volume of this tone is user adjustable.
1. Select the [Volume Setup] QuickKey, or the ECG parameter window→[Others >>], or
parameter window.
the SpO
2
2. Select [QRS Volume] or [Beat Vol] and then select the appropriate volume. 0 means
off, and 10 the maximum volume.
, there is a
2
3.11 Setting Parameters
3.11.1 Switching the Parameters On/Off
To switch the parameters on or off,
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the required passwordė[Others].
2. Configure the [Para Switch Authority] to [Unprotected] or [Protected].
If [Para Switch Authority] is configured to [Unprotected], select[Main Menu]ė
[Screen Setup>>]ė[Screen Layout >>]ė[Parameters Switch] to switch the
parameters on or off.
If [Para Switch Authority] is configured to [Protected], the parameter switch is
password protected. To switch the parameters on or off, select [Main Menu]ė [Maintenance >>]ė[User Maintenance >>]ėenter the required passwordė [Others >>]ė[Parameters Switch Setup >>].
When a parameter is switched off, its corresponding parameter module stops working, and its
parameter value and waveform are not shown on the monitor display.
NOTE
z ECG is always selected, and you cannot switch it off.
3-11
Page 50
WARNING
brightness, alarm volume, QRS
when the setting value is
3.11.2 Accessing the Parameters Menu
Select [Parameters >>] from the main menu or select corresponding parameter area or
waveform area to access a parameter setup menu.
3.12 Operating Modes
Your monitor has different operating modes. Some are password protected. This section lists
the major operating modes.
3.12.1 Monitoring Mode
This is the normal, everyday working mode that you use for monitoring patients. Your
monitor automatically enters the monitoring mode after being turned on.
3.12.2 Night Mode
To avoid disturbing the patient, night mode may be used.
To activate the night mode:
1. Select [Main Menu]→[Screen Setup >>]→[Night Mode >>].
2. In the pop-up menu, set the desired brightness, alarm volume, QRS volume, key volume,
NIBP end tone, or whether to stop NIBP measurement or not. When [Stop NIBP] is
selected, all the NIBP measurements terminate after entering the night mode.
3. Select the [Enter Night Mode] button.
To cancel the night mode:
1. Select [Main Menu]→[Screen Setup >>]→[Night Mode >>].
2. Select [Ok] in the popup.
z Before entering night mode, confirm the settings of
volume, and key volume. Pay attention to the potential risk
a bit low.
3-12
Page 51
WARNING
3.12.3 Privacy Mode
Privacy mode is only available when a patient who is admitted at a patient monitor is also
monitored by the central station.
To activate the privacy mode, select [Main Menu]→[Screen Setup >>]→[Privacy Mode].
The patient monitor behaves as follows as soon as the privacy mode is activated:
The screen turns blank and [Under monitoring. Press any key to exit the privacy
mode.] is displayed.
Monitoring and data storing continue but patient data is only visible at the central
station.
Alarms can still be triggered. But all audible alarms are suppressed and the alarm light is
deactivated at the patient monitor.
All system sounds are suppressed, including heart beat tone, pulse tone, all prompt tones,
etc.
To cancel the privacy mode, press any key.
The patient monitor exits the privacy mode automatically in one of the following situations:
The patient monitor disconnects from central station.
The alarm [Battery Too Low] or [System will shut down soon. Please replace the
batteries or use the external power.] is presented.
The touchscreen is locked automatically in the privacy mode.
z During privacy mode, all audible alarms are suppressed and the alarm light is
deactivated at the patient monitor. Alarms sound only at the central station.
3-13
Page 52
WARNING
The Demo mode is for demonstration purpose only. To avoid that the simulated data
are mistaken for the monitored patient’s data, you must not change into Demo mode
Otherwise, improper patient monitoring and delayed treatment
3.12.4 Demo Mode
In Demo mode, the monitor can demonstrate its major functions when patient or patient
simulator is not connected. The Demo mode is password protected.
To enter the Demo mode:
1. Select [Main Menu][Maintenance >>].
2. Select [Demo >>]. Enter the required password and then select [Ok].
To exit the Demo mode, select [Main Menu][Maintenance >>][Exit Demo].
z
during monitoring.
could result.
3.12.5 Standby Mode
In standby mode, you can temperately stops patient monitoring without turning off the
monitor. To enter the standby mode, select the Standby QuickKey
3-14
.
Page 53
4
4.
1
n
o
o
a
t
i
h
r
n
t
[
h
u
c
Y
N
[
P
o
P
W
a
[
A
i
m
g
P
a
p
u
r
S
P
a
h
e
h
m
m
m
c
P
i
s
t
u
e
l
r
t
s
k
e
w
a
p
e
m
h
h
r
[
I
a
e
s
r
t
a
i
w
d
n
e
e
n
t
p
d
c
u
i
o
m
e
d
h
a
e
d
a
e
e
o
p
p
t
s
k
e
e
t
n
i
S
t
y
u
o
d
d
G
i
w
s
t
y
n
a
t
d
r
e
i
]
n
Admit
The
patient mon
ected. This
con
rec
mmended t
rec
rdings, repo
To
dmit a patie
1. Select
2. Select
3. If [Dis
4. In the
5. Select
Man
from t
previo
data.
select:
aging
ing a P
tor displays
allows yo
at you fully
ts and netwo
t:
he [Patient
Discharge
e previous p
s patient. T
harge Pati
[
es] to apply
[
o] to clear t
Patient De
[
atient Cat.]
s
me measure
[
aced] deter
hen the [Pa
w
veform.
Ok].
atients
tient
hysiologica
to monito
admit a pa
king device
etup] Quic
atient] to cl
tient, the ne
e monitor m
nt] button a
the data sav
e data saved
ographics]
determines t
ents, and w
ines whethe
ed] is set to
data and sto
a patient
ient so that
.
Key, or [M
ar any prev
patient’s
kes no disti
pears dimm
d in the pati
in the patie
enu, enter
e way your
at safety an
to show pa
No], pace p
es them in t
hat is not
you can cl
in Menu]ė
ous patient
ata will be s
ction betwe
d, directly s
nt monitor t
t monitor.
he demogra
atient moni
alarm limit
e pulse mar
lse marks ar
e trends as s
dmitted ye
arly identif
Patient Set
ta. If you d
ved into the
n the old an
lect [Admit
the new pat
hic details, o
or processes
are applied
s on the EC
not shown
oon as a pati
. However,
your patie
p >>].
not erase d
data of the
the new pa
Patient] an
ient, o
f which:
and calculat
for your pat
waveform.
n the ECG
nt is
it is
t, on
ta
ient
then
s
ent.
W
RNING
z
[Patient
Cat.] and [
patient
s fully adm
patient
might n
z
For pac
patient
ECG si
z
For non
onitor use
ot be correc
ed patients,
monitor co
nal is too w
-paced pati
aced] will a
tted or not.
the default
for your p
you must s
ld mistake
ak.
nts, you mu
lways conta
If you do n
settings fro
tient.
t [Paced] to
a pace puls
t set [Pace
4-1
n a value, r
t specify set
the curre
[Yes]. If it
for a QR
] to [No].
gardless of
ings for the
t configura
s incorrectl
and fail to
whether the
e fields, the
ion, which
set to [No
alarm whe
, the
the
Page 54
4.2 Quick Admitting a Patient
Use [Quick Admit] only if you do not have the time or information to fully admit a patient.
Complete the rest of the patient demographic details later. Otherwise, the
displayed in the patient information area.
1. Select the [Patient Setup] QuickKey, or [Main Menu]ė[Patient Setup >>].
2. Select [Quick Admit]. If a patient has been admitted at present, select [OK]to
discharge the current patient. If .no patient is admitted, you can choose either:
[Yes] to apply the data in your patient monitor to the new patient, or
[No] to clear any previous patient data.
3. Enter the patient category and paced status for the new patient, and then select [Ok].
symbol will always be
4.3 Setting the Monitor Location (Not Available for
iMEC15S/iMEC15)
To set the monitor location, follow this procedure:
1. Select [
2. Input the following location of the monitor:
Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the required
password.
[Facility]: your facility name.
[Department]: your department name.
[Room No.]: room number.
[Bed No.]: bed number.
4.4 Querying and Obtaining Patient Information
The monitor can obtain patient information from HIS through eGateway. To query or obtain patient
information from HIS,
1. Select [Main Menu]ėaMaintenance >>]aiUser Maintenance >>]aintenance >>obtain patient ė[Network Setup >>]ė[Gateway Comm Setting >>], and set[IP Address]
and [Port]. Set [ADT Query] to [On].
2. Click patient information area to enter the [Patient Demographics] menu.
3. Select [Obtain Patient Info. >>] to enter the [Obtain Patient Information] menu.
4. Input query condition and then select [Query]. The monitor will display the obtained
patient information.
5. Select a patient and then click [Import]. Then the monitor will update the information
of corresponding patient.
4-2
Page 55
6. Select to exit the [Obtain Patient Information] menu.
NOTE
z The option [Obtain Patient Information] is available in the [Patient Setup] menu
only when [ADT Query] is set to [On].
z When obtaining patient information from HIS, the monitor only update patient
inforamtion. The patient’s monitoring data is not changed and the patient is not
discharged.
4.5 Querying from Local Facility (Not Available for
iMEC15S/iMEC15)
You can query the patient information from either the local facility or all networked facilities. To set
where to query, follow this procedure:
1. Select [
2. Set [
Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the required
passwordė[
Select [Yes] to query only from local facility.
Select [No] to query from all networked facilities.
Network Setup >>]ė[Gateway Comm Setup >>].
Query From Local Facility].
4.6 Associating Patient Information
After associating patient information with HIS, the monitor will automatically update patient
information if corresponding information in HIS has been changed. The monitor can associate
patient’s MRN, visit number, first name, last name, date of birth, and gender with HIS.
NOTE
z A keyword takes effect only when being defined in eGateway. Refer to eGateway
Integration Manager Installation Guide for details.
z The monitor displays corresponding patient information only when all the keywords
have been inputted.
4.7 Editing Patient Information
To edit the patient information after a patient has been admitted, or when the patient information is
4-3
Page 56
incomplete, or when you want to change the patient information:
1. Select the [Patient Setup] QuickKey, or [Main Menu]ė[Patient Setup >>].
2. Select [Patient Demographics] and then make the required changes.
3. Select [Ok].
You can also input the patient’s visit number in the [
Number
] option needs to be enabled.
To d i spl ay t h e [
1. Select [
2. Set [
Visit Number] option in the [Patient Demographics] menu:
Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the required
password.
Visit Number] to [On >>].
Patient Demographics] menu, but the [Visit
4.8 Discharging a Patient
To discharge a patient:
1. Select the [Patient Setup] QuickKey, or [Main Menu]ė[Patient Setup >>].
2. Select [Discharge Patient]. In the popup menu, you can either:
Directly select [Ok] to discharge the current patient, or
Select [Standby] then [Ok]. The patient monitor enters the standby mode after
discharging the current patient, or
Select [Cancel] to exit without discharging the patient.
NOTE
z Discharging a patient clears all history data in the monitor.
4.9 Transferring a Patient
You can transfer patient data between monitors with a USB drive without re-entering the patient
demographic information. Transferring of patient data enables you to understand the patient’s
history condition. The patient data that can be transferred includes: patient demographics, trend data,
alarm events and parameters alarm limits.
Select [Others >>] from [User Maintenance] menu. In the popup menu, you can set [Transferred
Data Length]. The default is [4 h]. You can also set [Data Transfer Method]. The default is [Off].
4-4
Page 57
4.
9
9
A
d
r
t
a
f
[
[
n
v
f
c
p
e
e
e
r
y
t
s
m
i
a
p
u
a
t
h
gD
u
w
r
gD
r
f
l
d
r
o
e
o
r
t
h
a
e
.
i
s
e
o
]
h
r
m
w
o
o
i
p
t
R
ǂ
i
e
n
o
f
n
a
n
a
g
i
m
u
e
n
d
U
u
m
t
n
u
i
a
d
n
n
y
n
t
r
t
y
]
u
r
t
s
n
w
a
c
h
c
r
m
u
U
a
e
o
n
n
.1 Trans
W
z
Do not
z
After a
patient
approp
NOTE
z
The sys
after tr
RNING
ischarge a
patient is s
category, p
iate for this
em automa
nsferring t
errin
atient befo
ccessfully
ced status
patient.
ically switc
e patient d
ata from
e the patien
ransferred,
and alarm
es on the H
ta.
the Mon
is successf
check if th
limits setti
alarm an
tor to a
lly transfer
patient se
gs, etc) on
lethal arrh
SB Driv
ed.
tings (espe
the monito
thmia alar
e
ially
are
1. Select
2. Select
3. Wait u
4. Remo
4.
.2 Trans
1. Conne
2. In the
3. After
Please
conten
[Ok] i
mediu
Main Men
Transfer to
til the follo
remove the
e the USB d
errin
t the USB d
opup menu,
S
lect [Trans
S
lect [Cance
S
lect [Unloa
d
ive.
ou select [T
s that need t
selected, th
and monit
D
fferent Patie
S
me Patient:
]ė[Patient
Storage Me
ing messag
USB drive.]
ive from pat
ata from
ive to the de
you can:
er] to transf
Transfer] t
USB Drive
ansfer], in t
be transfer
monitor co
r and deals
The m
ts:
data fr
accord
In the
Sel
Setup >>].
dium]. In th
appears: [T
ent monitor.
the USB
stination mo
r the patient
cancel the
to not trans
e popup me
ed. [Patient
pares the p
ith the patie
nitor erases
m the stora
ng to the pat
opup dialog
ect [Ye s ] to
popup men
ansfer to st
Drive to
itor.
data to the
peration of
er the patie
u you can f
Demograph
tient inform
t data base
ll the curre
e medium, a
ent categor
box, you ca
erge the pa
, select [Ok
rage medi
Monitor
onitor, o
ansferring p
t data and to
rther select
cs] must be
tion stored i
on the follo
t patient dat
d loads the
:
ient data in t
.
m successf
atient data.
unload the
he patient d
elected. Aft
both the st
ing.
, transfers th
onfiguratio
e monitor a
l.
SB
ta
r
rage
e patient
d
4-5
Page 58
4.
1
o
d
n
s
A
B
e
d
r
t
a
n
o
i
r
i
p
i
c
w
u
f
p
e
u
a
t
h
o
n
t
M
t
t
o
s
o
n
e
n
r
d
t
h
a
t
e
.
a
i
m
e
a
t
g
R
n
n
n
u
r
e
a
m
.
d
u
f
e
n
d
r
(
n
g
n
n
s
r
t
y
m
d
a
u
d
e
u
k
c
r
m
o
a
c
r
If y
4. Wait u
succes
W
z
The US
sure th
z
Do not
z
Do not
files ma
z
After a
patient
approp
NOTE
z
The sys
after tr
0 Con
ur patient m
All pat
transfe
All pat
on the
admitt
bi-dire
til the follo
ful.].
RNING
drive you
USB drive
ischarge a
remove the
y be damag
patient is s
category, p
iate for this
em automa
nsferring t
ecting t
nitor is con
ent informa
red to the C
ent informa
atient moni
ng a patient,
tional contr
Sel
ing messag
se may hav
or data tra
atient befo
storage me
d.
ccessfully
ced status
patient.
ically switc
e patient d
a Cen
ected to a c
ion, measure
S.
ion, measure
or and CMS
discharging
l can be ach
st
rage mediu
ect [No] to
a
d to transfer
appears: [T
e write-prot
sfer is in re
e the patien
ium durin
ransferred,
and alarm
es on the H
ta.
ral Mo
ntral monito
ment data a
ment data a
For some f
patient, sta
eved betwe
.
rase all the
the patient d
ansfer fro
ect function
d/write mo
is successf
data trans
check if th
limits setti
alarm an
itoring
ing system
d settings on
d settings ca
nctions such
ting/stoppin
n your patie
urrent patie
ta from the
storage me
In this case
e.
lly transfer
er process.
patient se
gs, etc) on
lethal arrh
ystem
CMS):
the patient
be displaye
as editing p
NIBP meas
t monitor an
nt data in th
torage medi
dium
, please ma
ed.
Otherwise,
tings (espe
the monito
thmia alar
onitor can b
d simultane
tient inform
urements, et
d the CMS.
monito
m.
e
data
ially
are
usly
tion,
.,
For
etails, refer
to the CMS’
instructions
for use.
4-6
Page 59
5
15.2
u
r
m
a
e
o
A
n
e
e
i
n
i
s
m
i
.
o
x
n
h
e
g
s
o
m
m
n
a
g
h
t
u
t
M
k
c
g
r
o
g
f
s
m
n
t
s
I
o
a
i
e
i
r
p
a
f
n
f
.
u
g
c
n
e
a
r
e
o
s
a
y
a
m
.
n
t
o
r
h
a
f
n
q
d
e
y
o
r
a
m
w
Man
5.
Introd
When perfo
adjust the
settings is c
monitor off
change som
configurati
The system
These
These
These
For the imp
see appendi
W
The co
manag
Param
Conve
setting
User
format
aging
ction
ming contin
onitor’s setti
lled a confi
rs different
e settings fro
n as a user c
RNING
figuration
ment tasks
configuratio
ter configur
tems relates
tional confi
tems define
.
aintenance i
tems relates
rtant config
Configura
Confi
uous monito
ngs accordin
uration. All
ets of confi
m a certain s
nfiguration.
anagement
ust be per
items can b
tion items
to parameter
uration ite
ow the mo
ems
to user main
ration item
ion Default
uratio
ing on a pat
g to the pati
wing you to
uration to su
et of configu
function is
ormed by cl
e classified
, e.g., wave
s
itor works, e
enance setti
and their de
nformation
ns
ent, the clini
nt’s conditio
configure th
t different p
ation and th
assword pr
inical profe
s:
orm gain, al
.g., screen la
gs, e.g., unit
ault values
al professio
. The collec
monitor m
tient catego
n save the c
tected. The
sionals.
rm switch,
out, record,
setup, time
nd user mai
al often nee
ion of all th
re efficientl
ies. You can
anged
configurati
larm limits..
print and ala
ormat and d
tenance ite
s to
se
, the
n
m
ta
s,
Enteri
1. Press t
2. Select
and th
g the [
e hard
[Maintenan
n select [Ok
anage
ey on the m
e >>][M
].
Config
nitor’s front
nage Confi
5-1
ration]
to enter the
uration >>]
Menu
ain menu.
Enter the re
uired pass
ord
Page 60
5.3 Setting Default Configuration
The monitor will load the pre-set default configuration in the following cases.
The patient monitor restarts after quitting over 120 seconds.
A patient is admitted.
A patient is discharged.
Patient data is cleared.
Patient category is changed.
To set default configuration:
1. Select [Select Default Config. >>] in the [Manage Configuration] menu.
2. In the [Select Default Config.] menu, select [Load the Latest Config.] or [Load
Specified Config.].
When you select [Load Specified Config.], the configuration (adult, pediatric or neonate) to
be restored is subject to the patient category. This configuration can be either factory
configuration or saved user configuration. Take adult as an example, select [Default Adu
Config.] and toggle between [Defaults] or user configuration(s).
NOTE
To know what configuration is restored when the patient monitor starts, enter the
main screen to check the prompt information at the lower part of the screen
(displayed for about 10 seconds).
5.4 Saving Current Settings
Current settings can be saved as user configuration. Up to 3 user configurations can be saved.
To save current settings:
1. Select [Save Current Settings As >>] in the [Manage Configuration] menu.
2. In the popup dialog box, enter the configuration name and then select [Ok].
5-2
Page 61
5.5 Editing Configuration
1. Select [Edit Config. >>] in the [Manage Configuration] menu.
2. The popup menu shows the existing configurations on the monitor. Selecting [Config.
on USB drive >>] will show the existing configurations on the USB drive. Select the
desired configuration and then select the [Edit] button.
3. Select [Alarm Setup >>], [Screen Setup >>] or [Parameter >>] to enter the
corresponding menu in which settings can be changed. The changed items of alarm
setup will be marked in red.
4. You can select [Save] or [Save as] to save the changed configuration. Select [Save] to
overwrite the original configuration. Select [Save as] to save the changed configuration
in another name.
5.6 Deleting a Configuration
1. Select [Delete Config. >>] in the [Manage Configuration] menu.
2. The popup menu shows the existing user configurations on the monitor. Selecting
[Config. on USB drive >>] will show the existing user configurations on the USB drive.
Select the user configurations you want to delete and then select [Delete].
3. Select [Yes] in the popup.
5.7 Transferring a Configuration
When installing several monitors with identical user configuration it is not necessary to set
each unit separately. An USB drive may be used to transfer the configuration from monitor to
monitor.
To export the current monitor’s configuration:
1. Connect the USB drive to the monitor’s USB port.
2. Select [Export Config. >>] in the [Manage Configuration] menu.
3. In the [Export Config.] menu, select the configurations and [User Maintenance
Settings] to export. Then select the [Export] button. A status message will report
completion of the transfer.
5-3
Page 62
To import the configuration on the USB drive to the monitor:
1. Connect the USB drive to the monitor’s USB port.
2. Select [Import Config. >>] in the [Manage Configuration] menu.
3. In the [Import Config.] menu, select the configurations and [User Maintenance
Settings] to import. Then select the [Import] button. A status message will report
completion of the transfer.
5.8 Loading a Configuration
You may make changes to some settings during operation. However, these changes or the
pre-selected configuration may not be appropriate for the newly admitted patient. Therefore,
the monitor allows you to load a desired configuration so as to ensure that all the settings are
appropriate for your patient.
To load a configuration,
1. Select [Load Configuration >>] from the main menu.
2. The popup menu shows the existing configurations on the monitor. Selecting [Config.
on USB drive >>] will show the existing configurations on the USB drive.
3. Select a desired configuration.
4. Select [View] to view the configuration details. In the popup menu, you can select
[Alarm Setup >>], [Screen Setup >>] or [Parameter >>] to view the corresponding
contents. The alarm setup items which are different than those currently used are marked
in red.
5. Select [Load] to load this configuration.
5.9 Restoring the Latest Configuration Automatically
During operation, you may make changes to some settings. However, these changes may not
be saved as user configuration. To prevent the changes from losing in case of a sudden power
failure, the patient monitor stores the configuration in real time. The saved configuration is
the latest configuration.
The monitor restore the latest configuration if restarts within 60 seconds after the power
failure. And it will restore the default configuration rather than the latest configuration if
restarts 120 seconds later after the power failure. The monitor may load either the latest
configuration or the default configuration if restarts from 60-120 seconds after the power
failure.
5-4
Page 63
5.10 Modifying Password
To modify the password for accessing the [Manage Configuration] menu,
1. Select [Modify Password >>] in the [Manage Configuration] menu.
2. Input a new password in the popup menu.
3. Select [Ok].
5-5
Page 64
FOR YOUR NOTES
5-6
Page 65
6 User Screens
6.1 Tailoring Your Screens
You can tailor your patient monitor’s screens by setting:
Wave line size
The color in which each measurement’s numerics and waveform are displayed
The parameter to be monitored.
Changing some settings may be hazardous. Therefore, those setting are password-protected
and can be modified by authorized personnel only. Once change is made, those who use the
patient monitor should be notified.
6.1.1 Changing the Wave Line Size
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the
required password.
2. Select [Others >>].
3. Select [Wave Line] and toggle between [Thick], [Mediate] and [Thin].
6.1.2 Changing Measurement Colors
1. Select [Main Menu][Screen Setup >>][Measurement Color Setup >>].
2. Select the color box next to your desired measurement and then select a color from the
popup menu.
6.1.3 Changing Screen Layout
Select the [Screens] QuickKey, or [Main Menu]→[Screen Setup >>]→[Screen Layout >>]
to enter the [Screens] menu.
You can choose the desired screen type in the [Choose Screen] window.
You can select the parameters and waveforms you want to view in the [Screen Setup]
window. For details, please refer to the section 3.6 Setting the Screen.
You can select the parameters you want to view on big numerics screen in the [Big
Numerics Screen Setup] window.
6-1
Page 66
Minitrend View
You can switch on or off the connected parameter modules in the [Parameters Switch]
window. If a parameter module is switched off, parameter values and waveforms will
not display on the screen.
6.2 Viewing Minitrends
6.2.1 Having a Split-Screen View of Minitrends
You can split the normal screen so that one part of the screen, on the left hand side,
continuously shows graphic minitrends beside waveforms as shown in the figure below.
To have a split-screen view of minitrends, you can:
Select [Minitrends] QuickKey, or
Select [Screens] QuickKey→[Choose Screen][Minitrends Screen]
Select [Main Menu]→[Screen Setup >>]→[Screen Layout >>]→[Choose
Screen][Minitrends Screen]
.
, or
The split-screen view provides minitrends for multiple parameters. In each field, the label and
scale are respectively displayed at the top and left. The time is displayed at the bottom of the
minitrends shown view.
6-2
Page 67
2
1
3
4
5
6
6.2.2 Setting Minitrends
Select the minitrends area. From the pop-up [Minitrend Setup] menu, you can:
Select the parameters to be displayed, or
Select [Minitrend Length] and then select the appropriate setting.
6.3 Viewing OxyCRG
To have a split screen view of OxyCRG, you can:
Select [OxyCRG] QuickKey, or
Select [Screens] QuickKey[Choose Screen][OxyCRG Screen]
Select [Main Menu]→[Screen Setup >>]→[Screen Layout >>]→[Choose
Screen][OxyCRG Screen]
.
, or.
The split-screen view covers the lower part of the waveform area and shows HR trend, SpO
trend, RR trend, and a compressed wave (Resp wave or CO
controls:
1. OxyCRG Event
You can enter the [Review] menu by selecting the [OxyCRG Event] button.
2. Trend length list box
In the trend length list box, you can select [1 min], [2 min], [4 min], or [8 min].
3. Setup
Select the [Setup] button to enter the [Setup] menu, in which you can select the
parameters for display, the time length to be saved before and after an event, and the
6-3
wave). At the bottom, there are
2
2
Page 68
scale of the graphic trends and waveform. The trend area can display two parameter
trends, e.g. HR trend and RR trend, simultaneously.
4. Auto Scale
Select the [Auto Scale] button, and the system automatically adjusts the scaling.
5. Print
Select [Print] to print out the realtime OxyCRG.
6. Record
Through this button, you can print out the currently displayed OxyCRG trends by the
recorder.
6.4 Viewing Other Patients
6.4.1 Care Group
You can select up to 10 patient monitors (including telemetry) connected to the same LAN
into a Care Group. This lets you:
View information on the monitor screen from another bed in the same Care Group.
Be notified of physiological and technical alarm conditions at the other beds in the same
Care Group.
To have a Care Group:
1. Open the [View Other Patient] window by:
Selecting [Others] QuickKey, or
 Selecting [Screens] QuickKey→[Choose Screen]→[View Others Screen]→ ,
or
Selecting [Main Menu]→[Screen Setup >>]→[Screen Layout >>]→[Choose
Screen >>]→[View Others Screen]→
2. Select [Setup] in the [View Other Patient] window.
3. Select the desired patient monitors from the [Connected Monitor List], and then select
button. The selected patient monitors constitute a Care Group.
the
.
6-4
Page 69
1
2
3
4
5
6.4.2 Viewing the Care Group Overview Bar
The Care Group overview bar locates at the bottom of the [View Other Patient] window. In
the overview bar, the department and bed label for any Care Group beds are displayed. For
telemetry, # is displayed before the department label. The color in which a Care Group bed
appears matches its status:
Red: indicates the bed is giving high-level physiological alarms or the telemetry is
giving alarm, such as nurse call or event.
Yellow: indicates the bed is giving medium- or low-level physiological alarms, or
medium-level technical alarms.
Blue: indicates the bed is giving low-level technical alarms.
Grey: indicates the bed fails to be networked or stays in the standby mode.
You c a n view a Care Group bed’s alarms by selecting it from the care group, and as well you
can select the [View This Patient] button to view this bed in the [View Other Patient]
window.
For more details about Care Group alarms, refer to the 7 Alarms chapter.
6.4.3 Understanding the View Other Patient Window
When you first open the [View Other Patient] window, the patient monitor automatically
selects a monitor from the network to display in the [View Other Patient] window.
6-5
Page 70
WARNING
have delay. Do not rely on
The [View Other Patient] window covers the lower part of the waveform area and consists
of:
1. Information Area: shows the patient information (including department, bed number,
patient name, etc.), network status symbol.
2. View Area: shows physiological waveforms and parameters. You can switch a
waveform area to a parameter area by selecting your desired waveform area and then
selecting [Switch to Parameter Area], or switch a parameter area to a waveform area
by selecting your desired parameter area and then selecting [Switch to Waveform
Area].
3. Care Group Overview Bar.
4. Message Area: shows physiological, technical and prompt messages from the currently
viewed patient monitor. It also shows the alarm given by the telemetry such as nurse call
or event. By selecting this area, you can enter the [Alarm Information List] to view all
physiological, technical and prompt messages coming from the currently viewed patient.
5. [Alarm Reset] button
When [Reset Other Bed’s Alarms] is set to [On] in [Maintenance]ė[User
Maintenance]ė[Alarm Setup], the [Alarm Reset] button appears on the [View
Other Patient] window. You can reset the alarm system for the selected monitor
by pressing the button. Refer to section 7.11.3 Resetting Care Group Alarms for
details.
When [Reset Other Beds Alarms] is set to [Off], there is no button appearing on
the [View Other Patient] window.
Additionally, you can change a waveform or parameter for viewing
To change a waveform for viewing, select the waveform segment where you want a new
waveform to appear and then select the waveform you want from the popup menu.
To change a parameter for viewing, select the parameter window where you want a new
parameter to appear and then select the parameter you want from the popup menu.
z The data presented in the [View Other Patient] window
this window for realtime data.
6-6
Page 71
6.5 Understanding the Big Numerics Screen
To enter the big numerics screen:
1. Select the [Screens] QuickKey, or [Main Menu]→[Screen Setup >>]→[Screen
Layout >>]→[Choose Screen].
2. Select [Big Numerics]→
.
You can select your desired parameters to display in this screen: select the [Screens]
QuickKey→[Big Numerics Screen Setup] and then select the parameters you want. For
parameters having a waveform, the waveform will also be displayed.
6-7
Page 72
FOR YOUR NOTES
6-8
Page 73
WARNING
similar equipment in any single area, e.g. an intensive care unit or cardiac operating
or other monitors
7 Alarms
Alarms, triggered by a vital sign that appears abnormal or by technical problems of the
patient monitor, are indicated to the user by visual and audible alarm indications.
z A potential hazard can exist if different alarm presets are used for the same or
room.
z If your patient monitor is connected to the central monitoring system (CMS) or
other monitors, alarms can be displayed and controlled remotely. Remote
suspension, inhibition, or reset of monitor alarms via the CMS
may cause a potential hazard. For details, refer to the operator’s manual of the
CMS and the other monitors.
7.1 Alarm Categories
By nature, the patient monitor’s alarms can be classified into three categories: physiological
alarms, technical alarms and prompt messages.
1. Physiological alarms
Physiological alarms, also called patient status alarms, are triggered by a monitored
parameter value that violates set alarm limits or an abnormal patient condition.
Physiological alarm messages are displayed in the physiological alarm area.
2. Technical alarms
Technical alarms, also called system status alarms, are triggered by a device
malfunction or a patient data distortion due to improper operation or mechanical
problems. Technical alarm messages are displayed in the technical alarm area.
Apart from the physiological and technical alarm messages, the patient monitor shows some
messages telling the system status or patient status. Messages of this kind are included into
the prompt message category and usually displayed in the prompt information area. Some
prompt messages that indicate the arrhythmia events are displayed in the physiological
alarm area. For some measurements, their related prompt messages are displayed in their
respective parameter windows.
7-1
Page 74
7.2 Alarm Levels
By severity, the patient monitor’s alarms can be classified into three categories: high level,
medium level and low level.
Physiological alarms Technical alarms
High
level
Medium
level
Low
level
Indicate that your patient is in a
life threatening situation, such
as Asystole, Vfib/Vtac and so
forth, and an emergency
treatment is demanded.
Indicate that your patient’s vital
signs appear abnormal and an
immediate treatment is
required.
Indicate that you patient’s vital
signs appear abnormal and an
immediate treatment may be
required.
Indicate a severe device malfunction or an
improper operation, which could make it possible
that the monitor cannot detect critical patient status
and thus threaten the patient’s life.
Indicate a device malfunction or an improper
operation, which may not threaten the patient’s life
but may compromise the monitoring of vital
physiological parameters.
Indicate a device malfunction or an improper
operation, which may compromise a certain
monitoring function but will not threaten the
patient’s life.
7.3 Alarm Indicators
When an alarm occurs, the patient monitor will indicate it to the user through visual or
audible alarm indications.
Alarm lamp Alarm message Flashing numeric Audible alarm tones
7.3.1 Alarm Lamp
If a technical alarm or physiological alarm occurs, the alarm lamp will flash. The flashing
color and frequency match the alarm level as follows:
High level alarms: the lamp quickly flashes red. Medium level alarms: the lamp slowly flashes yellow. Low level alarms: the lamp turns yellow without flashing.
7-2
Page 75
7.3.2 Alarm Message
When an alarm occurs, an alarm message will appear in the technical or physiological alarm
area. For physiological alarms, the asterisk symbols (*) before the alarm message match the
alarm level as follows:
High level alarms: *** Medium level alarms: ** Low level alarms: *
Additionally, the alarm message uses different background color to match the alarm level:
High level alarms: red Medium level alarms: yellow Low level alarms: yellow
You can view the alarm messages by selecting the physiological or technical alarm area.
7.3.3 Flashing Numeric
If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in
alarm will flash every second, and the corresponding alarm limit will also flash at the same
frequency indicating the high or low alarm limit is violated.
7.3.4 Audible Alarm Tones
The alarm tone is distinct from heart beat tone, keystroke tone and pulse tone in frequency.
This monitor has three choices of alarm tones and patterns: ISO, Mode 1 and Mode 2. For
each pattern, the alarm tones identify the alarm levels as follows:
ISO pattern:
High level alarms: triple+double+triple+double beep. Medium level alarms: triple beep. Low level alarms: single beep.
Mode 1:
High level alarms: high-pitched single beep. Medium level alarms: double beep. Low level alarms: low-pitched single beep.
7-3
Page 76
When multiple alarms of different levels occur simultaneously, the patient monitor
lamp and give alarm sounds
Some physiological alarms, such as asystole, are exclusive. They have identical alarm
gical alarms, but their alarm
say, when an exclusive physiological
simultaneously,
Mode 2:
High level alarms: high-pitched triple beep.
Medium level alarms: double beep.
Low level alarms: low-pitched single beep.
NOTE
z
will select the alarm of the highest level, light the alarm
accordingly, while all the alarm messages are displayed circularly on the screen.
z
tones and alarm lights with normal high level physiolo
messages are displayed exclusively. That is to
alarm and a normal high level physiological alarm are triggered
only alarm message of the exclusive physiological alarm is displayed.
7.3.5 Alarm Status Symbols
Apart from the aforementioned alarm indicators, the patient monitor still uses the following
symbols telling the alarm status:
indicates alarms are paused.
indicates alarms are reset.
indicates the alarm sound is turned off.
indicates individual measurement alarms are turned off or the system is in alarm
off status.
7-4
Page 77
7.4 Alarm Tone Configuration
7.4.1 Setting the Minimum Alarm Volume
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Minimum Alarm Volume] and toggle between 0 and 10.
The minimum alarm volume refers to the minimum value you can set for the alarm volume,
which is not affected by user or factory default configurations. The setting of minimum
alarm volume remains unchanged when the patient monitor shuts down and restarts.
7.4.2 Changing the Alarm Volume
1. Select the [Volume Setup] QuickKey or the [Alarm Setup] QuickKeyė[Others], or [Main Menu]→[Alarm Setup >>]ė[Others].
2. Select the appropriate volume from [Alm Volume]: X-10, in which X is the minimum
volume, depending on the set minimum alarm volume, and 10 the maximum volume.
3. Select [High Alarm Volume] to set the volume of the high priority alarm as [Alm
Volume+0], [Alm Volume+1] or [Alm Volume+2].
4. Select [Reminder Vol] to set the volume of the reminder tone as [High], [Med] or
[Low].
When alarm volume is set to 0, the alarm sound is turned off and a
the screen.
symbol appears on
7.4.3 Setting the Interval between Alarm Sounds
You cannot change the interval between alarm tones if you choose mode 1 or 2 as your
desired alarm tone pattern. For these two patterns, the interval between alarm tones
identifies the alarm levels as follows:
Mode 1:
Interval between high level alarm tones: continuously.
Interval between medium level alarm tones: 5 s.
Interval between low level alarm tones: 20 s.
Mode 2:
7-5
Page 78
WARNING
When the alarm sound is switched off, the patient monitor will give no audible alarm
about
Interval between high level alarm tones: 1 s.
Interval between medium level alarm tones: 5 s.
Interval between low level alarm tones: 20 s.
If you choose the ISO pattern, you can change the interval between alarm tones. To change
the interval between alarm tones:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [High Alarm Interval (s)], [Med Alarm Interval (s)] and [Low Alarm
Interval (s)] in turn and then select the appropriate settings.
z
tones even if a new alarm occurs. Therefore the user should be very careful
whether to switch off the alarm sound or not.
z Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level may result in a hazard to the patient.
Always keep the patient under close surveillance.
7.4.4 Changing the Alarm Tone Pattern
To change the alarm tone pattern:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Alarm Sound] and toggle between [ISO], [Mode 1]and [Mode 2].
User or factory default configurations exert no impact on the setup of alarm tone pattern.
The alarm tone pattern remains unchanged after the monitor restarts.
7.4.5 Setting the Reminder Tones
When the alarm volume is set to zero, or the alarm tone is reset or switched off, the patient
monitor issues a periodical reminder tone. To set the reminder tones:
1. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password.
7-6
Page 79
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Set the [Reminder Tones] to [On], [Off] or [Re-alarm]. When [Re-alarm] is selected, the current physiological alarms and technical alarms marked with Ĝ will be
re-generated after the [Reminder Interval] if the alarm condition persists.
To set the interval between reminder tones, select [Reminder Interval] and toggle between
[1min], [2min] and [3min].
In addition, you can set the volume of alarm reminder tones. To set the volume of alarm
reminder tones, select [Main Menu]→[Alarm Setup >>]→[Others] or the [Alarm Setup]
QuickKey→[Others]. Then, select [Reminder Vol] and toggle between [High], [Medium]
and [Low].
7.5 Understanding the Alarm Setup Menu
Select the [Alarm Setup] QuickKey or [Main Menu]→[Alarm Setup >>] to enter the
[Alarm Setup], where you can:
Set alarm properties for all parameters.
Change ST alarm settings.
Change arrhythmia alarm settings.
Set the threshold for some arrhythmia alarms.
Change other settings.
Please refer to the Monitoring ECG for how to change ST alarm settings, how to change
arrhythmia alarm settings and how to set the threshold for some arrhythmia alarms.
7-7
Page 80
WARNING
ineffective. For example, High oxygen levels may predispose a premature infant to
retrolental fibroplasia. If this is a consideration do NOT set the high alarm limit to
When monitoring patients that are not continuously attended by a clinical operator,
properly configure the alarm system and adjust alarm settings as per the patient's
7.5.1 Setting Alarm Properties for All Parameters
In the main menu, select [Alarm Setup >>]→[Parameters]. You can review and set alarm
limits, alarm switches, alarm level and alarm recordings for all parameters.
When a measurement alarm occurs, automatic recording of all the measurement numerics
and related waveforms is possible when the measurement’s [On/Off] and [Record] are set
on.
z Make sure that the alarm limits settings are appropriate for your patient before
monitoring.
z Setting alarm limits to extreme values may cause the alarm system to become
100%, which is equivalent to switching the alarm off.
z
condition.
7.5.2 Adjusting Alarm Limits Automatically
The monitor can automatically adjust alarm limits according to the measured vital signs,
using the auto limits function. When auto limits are selected, the monitor calculates safe
auto limits based on the latest measured values.
To get accurate auto alarm limits, you need to collect a set of measured vital signs as a
baseline. Then, in the main menu, select [Alarm Setup >>]→[Parameters]→[Auto Limits]
[Ok]. The monitor will create new alarm limits based on the measured values.
Before applying these automatically created alarm limits, confirm if they are appropriate for
your patient in the mass alarm setup menu. If not, you can adjust them manually. These
alarm limits will remain unchanged until you select auto limits again or adjust them
manually.
7-8
Page 81
Module
Parameter
Low alarm limit
Auto alarm limits
range
Adult
pediatric
Neonate
Adult
pediatric
Neonate
ECG
HR/PR
(HR
0.8)
or
bpm
(whichever
is greater)
(HR
30) or
90
(whichever
is greater)
(HR
1.25)
or
bpm
(whichever
is smalle)
(HR
200bpm (whichever
is smaller)
Adult/pediatric: 35
to
Neonate:
to 225
Resp
RR
(RR
0.5)
or
rpm
(whichever
is greater)
(RR
10) or
30
rpm
(whichever
is greater)
(RR
1.5)
or
rpm
(whichever
is smaller)
(RR
rpm
(whichever is
smaller)
Adult/pediatric: 6
to
Neonate:
to 90
SpO2
SpO
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range
NIBP
NIBP-S
(SYS × 0.68
+ 10
mmHg
(SYS – 15)
or 45mmHg
(whichever
is greater)
(SYS × 0.86
+ 38
mmHg
(SYS
105mmHg
(whichever is
smaller)
Adult:
270
Pediatric:
to 185
Neonate: 40 to 115
NIBP-D
(Dia
0.68
+ 6
mmHg
(Dia
15)
or 20mmHg
(whichever
is greater)
(Dia
0.86
+ 32
mmHg
(Dia
80mmHg (whichever
is smaller)
Adult:
to 210
Pediatric:
to 150
Neonate:
to 90
NIBP-M
(Mean ×
0.68
)
mmHg
(Mean – 15)
or 35mmHg
(whichever
is greater)
(Mean ×
0.86
35)
mmHg
(Mean
95
mmHg
(whichever is
smaller)
Adult:
to 230
Pediatric:
to 165
Neonate:
to 105
Tem p
T1 (T1 – 0.5)ć
(T1
0.5) ć
(T1 + 0.5)ć
(T1 + 0.5)
1 to 49
T2 (T2 – 0.5)ć
(T2
0.5) ć
(T2 + 0.5)ć
(T2 + 0.5)
1 to 49
TD
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range
The monitor calculates the auto limits based on the following rules.
High alarm limit
/
/
×
40
×
2
6
)
×
)
+ 8
bpm
240
30
×
+ 40) or
240
×
+ 25) or 85
55
55
10
+ 15) or
)
×
+ 15) or
)
+ 15) or
+
ć
45 to
45
25
25
20
30
30
25
ć
ć
ć
7-9
Page 82
Module
Parameter
Low alarm limit
Auto alarm limits
range
Adult
pediatric
Neonate
Adult
pediatric
Neonate
IBP˖
ART/
Ao/
UAP/
BAP/
FAP /
LV/
P1
P4
(Arterial
pressure)
IBP-S
(SYS × 0.68
ˇ
10)
mmHg
(SYS – 15)
or 45mmHg
(whichever
is greater)
(SYS × 0.86
ˇ
mmHg
(SYS
105mmHg
(whichever is
smaller
Adult: 45 to 270
Pediatric
45 to 185
Neonate
35 to 115
IBP-D
(Dia
0.68
+ 6
mmHg
(Dia
15)
or 20mmHg
(whichever
is greater)
(Dia
0.86
+ 32
mmHg
(Dia
80mmHg (whichever
is smaller)
Adult:
225
Pediatric:
to 150
Neonate:
to 90
IBP-M
(Mean ×
0.68
8)mmHg
(Mean – 15)
or 35mmHg
(whichever
is greater)
(Mean ×
0.86
35)mmHg
(Mean
95mmHg (whichever
is smaller)
Adult:
245
Pediatric:
to 180
Neonate:
to 105
IBP:PA
IBP-S
SYS
0.75
SYS
0.75
SYS
1.25
SYS
3 to
IBP-D
Dia
0.75
Dia
0.75
Dia
1.25
Dia
IBP-M
Mean × 0.75
Mean × 0.75
Mean × 1.25
Mean
IBP:
CVP/
ICP/
LAP/
RAP/
UVP/
P1
P4
(Venous
pressur
e
IBP-M
Mean × 0.75
Mean × 0.75
Mean × 1.25
Mean
3 to
IBP:CPP
CPP
CPP
0.68
+ 8mmHg
(CPP – 15)
or 35mmHg
(whichever
is greater)
CPP
0.86
ˇ35mmHg
(CPP
95mmHg (whichever
is
Adult:
235
mmHg
Pediatric:
to 175
mmHg
Neonate:
to 100
mmHg
)
High alarm limit
/
×
38)
/
×
)
+ 15) or
)
+ 15) or
:
:
25 to
25
20
-
+
×
×
-
)
×
×
×
+
×
×
×
7-10
+ 15) or
× 1.25
× 1.25
× 1.25
× 1.25
ˇ15) or
smaller)
30 to
30
25
120mmHg
40mmHg
20 to
25
25
Page 83
Module
Parameter
Low alarm limit
Auto alarm limits
range
Adult
pediatric
Neonate
Adult
pediatric
Neonate
CO2
EtCO2
0 to
32mmHg:
remains the
same
0 to
32mmHg:
remains the
same
0 to
32mmHg:
remains the
same
0 to
remains the same
Same as the
measurement range
32
35mmHg:
29mmHg
32
35mmHg:
29mmHg
32
35mmHg:
41mmHg
32
41mmHg
35
45mmHg:
(etCO
6)
mmHg
35
45mmHg:
(etCO
6)
mmHg
35
45mmHg:
(etCO
+6)
mmHg
35
(etCO
mmHg
45
48mmHg:39
mmHg
45
48mmHg:39
mmHg
45
48mmHg:51
mmHg
45
48mmHg:51
mmHg
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg: remains
the same
FiCO2
N/A N/A
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range
awRR
awRR × 0.5
or
(whichever
is greater)
(awRR –
10
30
rpm
(whichever
is greater)
awRR × 1.5
or
rpm
(whichever
is smaller)
(awRR+25
85
rpm
(whichever is
smaller)
Adult/pediatric: 6
to
Neonate:
to 90
AG
EtCO2
(AG)
Same as CO FiCO2
(AG)
awRR
awRR × 0.5
or 6
(whichever
is greater)
awRR – 10
or 30
rpm
(whichever
is greater)
awRR × 1.5
or 30
rpm
(whichever
is smaller)
awRR+25 or 85 rpm
(whichever is
smaller)
Adult/pediatric: 6
to 55
Neonate: 10 to 90
FiAA/
EtAA
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range FiO
Same as the
Same as the
Same as the
Same as the default
Same as the
High alarm limit
/
/
32mmHg:
to
to
to
to 35mmHg:
to
to
to
to 45mmHg:
-
2
to
6 rpm
to
) or
module
2
-
2
2
to
30
to
+6)
2
) or
55
10
rpm
/
2
7-11
Page 84
Module
Parameter
Low alarm limit
Auto alarm limits
range
Adult
pediatric
Neonate
Adult
pediatric
Neonate
EtCO2
default
alarm limit
default
alarm limit
default
alarm limit
alarm limit
measurement range
FiN2O/
EtN
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the
default
alarm limit
Same as the default
alarm limit
Same as the
measurement range
C.O.
BT
Adult˖
(BT
N/A
Adult˖
(BT
N/A
Same as the
measurement range
High alarm limit
/
/
O
2
– 1)ć
– 1)ć
7.5.3 Setting Alarm Delay Time
You can set the alarm delay time for over-limit alarms of continuously measured parameters.
If the alarm-triggered condition disappears within the delay time, the patient monitor will
not give the alarm. You can set the [Alarm Delay] in the [Others] window of [Alarm
Setup] menu.
To set the alarm delay time,
1. Select [Main Menu] [Maintenance >>] [User Maintenance >>]. Enter the
required password and then select [Ok].
2. Select [Alarm Setup >>] [Alarm Delay].
Alarm delay is not applied to the following physiological alarms:
Apnea
ST alarms
Arrhythmia alarms
ECG Weak Signal
Resp Artifact
No Pulse
Nellcor SpO
Measurements of noncontinuous parameters over alarm limits
HR over alarm limits
Anesthetic Mixture's MAC >3
You can set [Apnea Delay] and [ST Alarm Delay] separately.
over alarm limits
2
7-12
Page 85
7.5.4 Setting SpO2 Technical Alarm Delay
You can set [Tech. Alarm Delay] in the [Others] tab of the [Alarm Setup] menu. The
options are [Off], [5s], [10s] and [15s]. The delay is effective to the following technical
alarms: SpO
Sensor Off, SpO2 Too Much Light, SpO2 Low Signal and SpO2 Interference.
2
7.5.5 Setting Recording Length
You can change the length of the recorded waveforms. In the [Others] window of the
[Alarm Setup] menu, select [Recording Length] and toggle between [8 s], [16 s] and [32
s]:
[8 s]: 4 seconds respectively before and after the alarm or manual event trigger
moment.
[16 s]: 8 seconds respectively before and after the alarm or manual event trigger
moment.
[32 s]: 16 seconds respectively before and after the alarm or manual event trigger
moment.
7.5.6 Intubation Mode
When performing intubation during general anesthesia, you can set the patient monitor to
enter intubation mode in order to reduce unnecessary alarms. Intubation mode is available
for Resp, CO
[Intubation Mode] button to disable respective physiological alarms.
The default intubation time is 2 minutes. You can also change the time by following this
procedure:
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the
2. Select [Alarm Setup >>], and set the [Intubation Mode Period] to [1 min], [2 min],
and AG parameters. In the setup menu of these parameters, you can choose
2
required password.
[3 min], or [5 min].
7-13
Page 86
you
7.6 Pausing Alarms
If you want to temporarily prevent alarms from sounding, you can pause alarms by pressing
hardkey on the monitor’s front. When alarms are paused:
the
No alarm lamps flash and no alarms are sounded.
No numeric and alarm limit flash.
No alarm messages are shown.
The remaining pause time is displayed in the physiological alarm area.
The
The patient monitor enters into the alarm paused status as soon as it is turned on. The alarm
pause time is fixed to be 2 minutes.
When the alarm pause time expires, the alarm paused status is automatically cancelled and
the alarm tone will sound. You can also cancel the alarm paused status by pressing the
hardkey.
The alarm pause time can be set to [1 min], [2 min], [3 min], [5 min], [10 min], [15 min]
or [Permanent]. The default alarm pause time is 2 minutes.
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the
2. Select [Alarm Setup >>]ė[Alarm Pause Time] and then select the appropriate
You can also temporarily prolong the alarm pause time after the monitor entersthe alarm
paused status:
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the
2. In the [Alarm Setup] menu, set the [Max. Alarm Pause 15min] to [Enable].
3. In the physiological alarm area, select a proper time in the [Alarm Pause Time] menu.
alarms paused symbol is displayed in the alarm symbol area.
required password.
setting from the popup list.
required passwordė[Alarm Setup >>].
NOTE
z [Max. Alarm Pause 15min] is configured to [Disable] by default. In this case,
cannot prolong the pause time.The prolonged pause time is only effective to the
current paused alarms.
7-14
Page 87
WARNING
Please be very
7.7 Switching Off All Alarms
If [Alarm Pause Time] is set to [Permanent], the patient monitor enters into the alarm off
status after the
As for physiological alarms, no alarm lamps flash and no alarms are sounded.
As for physiological alarms, no numeric and alarm limit flash.
No physiological alarm messages are shown.
[Alarm Off] is displayed in the physiological alarm area with red background.
As for technical alarms, no alarms are sounded.
hardkey is pressed. During the alarm off status,
The
You can cancel the alarm off status by pressing the
z Pausing or switching off alarms may result in a hazard to the patient.
alarm off symbol is displayed in the alarm symbol area.
hardkey.
careful.
7.8 Resetting Alarms
By selecting the QuickKey, you can reset the alarm system to acknowledging the
on-going alarms and enable the alarm system to respond to a subsequent alarm condition.
For physiological alarms, except the NIBP-related alarms, when the alarm system is reset:
The alarm sound is reset.
A Ĝ appears before the alarm message, indicating that the alarm is acknowledged.
The icon
The parameter numeric and alarm limits still flash.
The indication of alarm lamp for the physiological alarm depends on the alarm light setting.
When [Alarm Light on Alarm Reset] is set to [On], the alarm lamp remains flashing.
When [Alarm Light on Alarm Reset] is set to [Off], the alarm lamp stops flashing.
Technical alarms give different alarm indicators when the alarm system is reset:
appears in the alarm symbol area.
7-15
Page 88
For some technical alarms, including the NIBP-related alarms, a Ĝ appears before
the alarm message and
is acknowledged.
Some technical alarms are changed to the prompt messages.
Some technical alarms are cleared. The monitor gives no alarm indications.
For details about the indications of technical alarms when the alarm system is reset, refer to
D.2 Technical Alarm Messages.
To set [Alarm Light on Alarm Reset]:
1. Select [Main Menu][Maintenance >>][User Maintenance >>]→enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Alarm Light on Alarm Reset], and toggle between [On] and [Off].
The default setting for [Alarm Light on Alarm Reset] is [On].
appears in the alarm symbol area, indicating that the alarm
7.9 Latching Alarms
The alarm latching setting for your patient monitor defines how the indicators of the
physiological alarms behave when you do not acknowledge them:
If you do not “latch the physiological alarms, their alarm indications disappear when
the alarm condition ends.
If you latch the physiological alarms, all visual and audible alarm indications last
until you acknowledge the alarms, except that the measurement numeric and violated
alarm limit stop flashing as soon as the initial alarm condition goes away.
You can separately latch the visual indications or simultaneously latch the visual and the
audible indication.
When the visual indications are latched, the visual indications, including alarm lamp,
alarm message and its background remains when the alarm condition ends.
When the audible indications are latched, the monitor issues alarm sounds when the
alarm condition ends.
To latch a physiological alarm:
1. Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the
required password.
2. Select [Alarm Setup >>]ė[Latching Alarms>>].
3. In the [Latching Alarms] menu, select how you want to latch the alarms.
7-16
Page 89
Changing of alarm priority may affect the latching status of corresponding alarm.
Please determine if you need to reset the latching status for the specific alarm when
The rules for latching the alarms are:
You can separately select [Latching Visual Signal].
Selecting [Latching Audible Signal] simultaneously latches the visual signal.
Selecting alarms of lower priority simultaneously latches the alarms of higher priority.
NOTE
z
you have changed its alarm priority.
z When the alarm system is reset, the latched physiological alarms are cleared.
7.10 Testing Alarms
When the monitor starts up, a selftest is performed. In this case the alarm lamp is lit in
yellow and red respectively, and the system gives a beep. This indicates that the visible and
audible alarm indicators are functioning correctly.
For further testing of individual measurement alarms, perform the measurement on yourself
(for example SpO
appropriate alarm behaviour is observed.
or CO2) or use a simulator. Adjust alarm limits and check that
2
7.11 Using Care Group Alarms
7.11.1 Care Group Auto Alarms
When a Care Group is set up on your monitor, a flashing symbol will appear beside the
QuickKeys area if any monitor in your Care Group, which is not currently viewed by your
monitor, is alarming. The alarm symbol is shown as below.
The background colors of the alarm symbols indicate alarm levels, and are the same as
those of the corresponding alarm messages. If multiple alarms are active in the Care Group,
the background color is the same as that of the highest-level alarm message. For more
information about the alarm message and background color, see7.3.2 Alarm Message.
When a patient monitor in the Care Group is disconnected, the flashing symbol is shown as
below.
7-17
Page 90
The department and bed label of the alarming monitor appear on the symbol. You can enter
the view other patient window by pressing this symbol.
7.11.2 Resetting Care Group Alarms
You can reset the alarms presented on the viewed bed by pressing the [Alarm Reset] from
the current monitor’s [View Other Patient] window. To enable this function:
1. For iMEC12/iMEC10/iMEC8/iMEC7/iMEC6/iMEC5 monitor, select [Main Menu]
[Maintenance>>][User Maintenance>>]enter the required password[Alarm
Setup>>][Other Bed Alarm Setup>>]. For iMEC15S/iMEC15 monitor, select
[Main Menu][Maintenance>>]→[User Maintenance>>]→enter the required
password[Alarm Setup>>].
2. Set [Reset Other Bed’s Alarms] to [On].
The alarms presented on the current monitor can also be reset from another monitor viewing
this monitor. To do so, proceed as follows:
1. For iMEC12/iMEC10/iMEC8/iMEC7/iMEC6/iMEC5 monitor, select [Main Menu]
[Maintenance>>][User Maintenance>>]enter the required password[Alarm
Setup>>][Other Bed Alarm Setup>>] in the current monitor. For
iMEC15S/iMEC15 monitor, select [Main Menu][Maintenance>>][User
Maintenance>>]enter the required password[Alarm Setup>>] in the current
monitor.
2. Set [Alarm Reset By Other Bed] to [On].
3. In the other monitor, select the [Alarm Reset] button from the [View Other Patient]
window.
WARNING
Resetting care group alarms may cause a potential hazard. Please act with caution.
7.11.3 Switching Off the Remote Device Disconnection Alarm
(not available for iMEC15S/iMEC15)
The monitor can provide an alarm if a viewed bed device is disconnected. By default, the
function is enabled. To disable the alarm, follow this procedure:
1. In the current monitor, select [Main Menu][Maintenance>>][User
Maintenance>>]enter the required password[Alarm Setup>>][Other Bed
Alarm Setup>>].
2. Set [Other Bed Disconnection Alm] to [Off].
7-18
Page 91
7.11.4 Setting Care Group Alert Tone (for iMEC12/iMEC10/
iMEC8/iMEC7/iMEC6/iMEC5)
7.11.4.1 Setting the Alarm Reminder
Whena monitor in the Care Group issues an alarm, your patient monitor prompts you by
giving alert tone. To set the alert tone, follow this procedure:
1. In the current monitor, select [Main Menu]ė[Maintenance>>]ė[User
Maintenance>>]ėenter the required passwordė[Alarm Setup>>]ė[Other Bed
Alarm Setup>>].
2. Set the [Alarm Reminder].
[Repeat]: The monitor gives continuous alert tone when the alarm occurs at the
viewed bed is the same level as the setup level in the monitor. To set which alarm
level applies to continuous alert tone, see section 7.11.4.2Setting the Alarm
Level.
[Once]: The monitor gives a single alert tone when an alarm occurs at the viewed
bed.
[Off]: The monitor do not give any alert tone when an alarm occurs at the viewed
bed.
7.11.4.2 Setting the Alarm Level
When [Alarm Reminder] is set to [Repeat], you can set which alarm level of the viewed
bed alarm applies to the continuous alert tone. To set the alarm level of the viewed bed
alarm, follow this procedure:
1. In the current monitor, select [Main Menu]ė[Maintenance>>]ė[User
Maintenance>>]ėenter the required passwordė[Alarm Setup>>]ė[Other Bed
Alarm Setup>>].
2. Set the [Alarm Lev].
[All]: This monitor gives continous alert tone to all the alarms of the viewed bed
when [Alarm Reminder] is set to [Repeat].
[High Only]: This monitor gives continous alert tone only to high level alarms of
the viewed bed when [Alarm Reminder] is set to [Repeat].
[High&Med]: This monitor gives continous alert tone to high level and mediate
level alarms of the viewed bed when [Alarm Reminder] is set to [Repeat].
NOTE
z The setting of the [Alarm Lev] is valid only when [Alarm Reminder] is set to
[Repeat].
7-19
Page 92
7.11.5 Setting Care Group Alert Tone (for iMEC15S/iMEC15)
When a monitor in the Care Group issues an alarm, your patient monitor prompts you by
giving an alert tone. To set the alert tone:
1. In the main menu, select [Screen Setup >>]ė[Screen Layout >>]ė[Choose Screen].
2. In the [Choose Screen] window, select [View Others Screen].
3. In the view other patient window, select [Setup>>], and set [Alert Tone] to [Repeat],
[Once] or [Off].
7.12 When an Alarm Occurs
When an alarm occurs, observe the following steps and take proper actions:
1. Check the patient’s condition.
2. Confirm the alarming parameter or alarm category.
3. Identify the source of the alarm.
4. Take proper action to eliminate the alarm condition.
5. Make sure the alarm condition is corrected.
For troubleshooting specific alarms, see appendix D Alarm Messages.
7-20
Page 93
WARNING
Make sure the conductive parts of electrodes and associated connectors for applied
parts, including the neutral electrode, should not contact any other conductive parts
odically inspect the electrode application site to ensure skin quality. If the skin
frequency surgical procedure, ensure
surgery unit (ESU) shall properly contact the
CAUTION
grounded instrument near the patient and electrosurgery
8 Monitoring ECG
8.1 Introduction
The electrocardiogram (ECG) measures the electrical activity of the heart and displays it on
the patient monitor as a waveform and a numeric. For iMEC12, iMEC10, iMEC8, iMEC7,
iMEC6 and iMEC5, ECG monitoring provides 3-, 5-lead ECG monitoring, ST-segment
analysis and arrhythmia analysis. For iMEC15S and iMEC15, ECG monitoring provides 3-,
5- and 12-lead ECG monitoring, ST-segment analysis and arrhythmia analysis.
8.2 Safety
z Use only ECG electrodes and cables specified by the manufacturer.
z
including earth.
z Peri
quality changes, replace the electrodes or change the application site.
z Use defibrillation-proof ECG cables during defibrillation.
z Do not touch the patient, or table, or instruments during defibrillation.
z This equipment is not suitable for direct cardiac application.
z To reduce the hazard of burns during high-
that the monitor’s cables and transducers never come into contact with the
electro-surgery unit (ESU).
z The neutral electrode of the electro-
patient. Otherwise, burns may result.
z Interference from a non-
interference can cause problems with the waveform.
8-1
Page 94
electrodes are used and applied in accordance with the manufacturer’s instructions
NOTE
z After defibrillation, the screen display recovers within 10 seconds if the correct
for use.
8.3 Preparing to Monitor ECG
8.3.1 Preparing the Patient and Placing the Electrodes
1. Prepare the patient’s skin. Proper skin preparation is necessary for good signal quality at
the electrode, as the skin is a poor conductor of electricity. To properly prepare the skin,
choose flat areas and then follow this procedure:
Shave hair from skin at chosen sites.
Gently rub skin surface at sites to remove dead skin cells.
Thoroughly cleanse the site with a mild soap and water solution. We do not
recommend using ether or pure alcohol, because this dries the skin and increases
the resistance.
Dry the skin completely before applying the electrodes.
2. Attach the clips or snaps to the electrodes before placing them.
3. Place the electrodes on the patient.
4. Attach the electrode cable to the patient cable and then plug the patient cable into the
ECG connector.
8.3.2 Choosing AHA or IEC Lead Placement
1. Select the ECG parameter window or waveform area to enter the [ECG Setup] menu.
2. Select [Others]→[Lead Set] and then select [3-lead], [5-lead], [12-lead] (for
iMEC15S/iMEC15 only) or [Auto] according to the applied electrodes.
3. Select [Main Menu]→[Maintenance >>]→[User Maintenance >>]→enter the
required password
4. Select [Others >>]→[ECG Standard] and then select [AHA] or [IEC] according to
the standard that is applied for your hospital.
8-2
Page 95
8.3.3 ECG Lead Placements
The electrode placement illustrations in this chapter adopt the AHA standard.
3-Leadwire Electrode Placement
Following is an electrode configuration when using 3 leadwires:
RA placement: directly below the clavicle and near the right shoulder.
LA placement: directly below the clavicle and near the left shoulder.
LL placement: on the left lower abdomen.
5-Leadwire Electrode Placement
Following is an electrode configuration when using 5 leadwires:
RA placement: directly below the clavicle and near the right shoulder.
LA placement: directly below the clavicle and near the left shoulder.
RL placement: on the right lower abdomen.
LL placement: on the left lower abdomen.
V placement: on the chest.
8-3
Page 96
The chest (V) electrode can be placed on one of the following positions:
V1 placement: on the fourth intercostal space at the right sternal border.
V2 placement: on the fourth intercostal space at the left sternal border.
V3 placement: midway between the V2 and V4 electrode positions.
V4 placement: on the fifth intercostal space at the left midclavicular line.
V5 placement: on the left anterior axillary line, horizontal with the V4 electrode
position.
V6 placement: on the left midaxillary line, horizontal with the V4 electrode position.
V3R-V6R placement: on the right side of the chest in positions corresponding to those
on the left.
VE placement: over the xiphoid process.
V7 placement: on posterior chest at the left posterior axillary line in the fifth intercostal
space.
V7R placement: on posterior chest at the right posterior axillary line in the fifth
intercostal space.
12-Leadwire Electrode Placement (for iMC15S/iMC15 only)
12-lead ECG uses 10 electrodes, which are placed on the patient’s four limbs and chest. The
limb electrodes should be placed on the soft skin and the chest electrodes placed according to
the physician’s preference.
8-4
Page 97
WARNING
When using electrosurgical units (ESU), patient leads should be placed in a position
is equal distance from the Electrosurgery electrotome and the grounding plate
Lead Placement for Surgical Patients
The surgical site should be taken into consideration when placing electrodes on a surgical
patient. e.g. for open-chest surgery, the chest electrodes can be placed on the lateral chest or
back. To reduce artifacts and interference from electrosurgical units, you can place the limb
electrodes close to the shoulders and lower abdomen and the chest electrodes on the left side
of the mid-chest. Do not place the electrodes on the upper arm. Otherwise, the ECG
waveform will be very small.
z
that
to avoid burns to the patient. Never entangle the ESU cable and the ECG cable
together.
z When using electrosurgical units (ESU), never place ECG electrodes near to the
grounding plate of the ESU, as this can cause a lot of interference on the ECG
signal.
8.3.4 Checking Paced Status
It is important to set the paced status correctly when you start monitoring ECG. The paced
symbol
The pace pulse markers “|” are shown on the ECG wave when the patient has a paced signal.
If [Paced] is set to [No] or the patients paced status is not selected, the symbol
shown in the ECG waveform area.
To change the paced status, you can select either:
the patient information area, or
[Main Menu]→[Patient Setup]→[Patient Demographics], or,
the ECG parameter window or waveform area→[Others >>],
and then, select [Paced] from the popup menu and toggle between [Ye s] and [No].
If you do not set the paced status, the patient monitor issues a prompt tone when pace pulse is
detected. At the same time, the paced symbol flashes and the message [Please confirm the
pace of patient] appears in the ECG waveform area. Then, please check and set the paced
status of the patient.
is displayed in the ECG waveform area when the [Paced] status is set to [Ye s ].
8-5
will be
Page 98
WARNING
For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the
patient monitor could mistake a pace pulse for a QRS and fail to alarm when the
not rely entirely on rate meter alarms when monitoring
1
1
3
2
3
4
2
z
ECG signal is too weak. Do
patients with pacemakers. Always keep these patients under close surveillance.
z For non-paced patients, you must set [Paced] to [No].
z The auto pacer recognition function is not applicable to pediatric and neonatal
patients.
z False low heart rate indicators or false Asystole calls may result with certain
pacemakers because of pacemaker artifact such as electrical overshoot of the
pacemaker overlapping the true QRS complexes.
8.4 Understanding the ECG Display
Your display may be configured to look slightly different.
1. Lead label of the displayed wave
2. ECG gain
3. ECG filter label
4. Notch filter status
When a paced signal has been detected, the pace pulse marks “|” are shown on the ECG wave
if the [Paced] has been set to [Ye s ].
1. Current heart rate alarm limits
2. Current heart rate
3. Heart beat symbol
8-6
Page 99
is in use, a question mark (?) may display on the right
NOTE
z When an electro-surgery unit
of the HR value. This indicates that there is high frequency interference
For 12-lead ECG display screen, refer to the section 8.11 12-Lead ECG Monitoring (for
iMEC15S/iMEC15 only).
8.5 Changing ECG Settings
8.5.1 Accessing ECG Menus
By selecting the ECG parameter window or waveform area, you can access the [ECG Setup]
menu.
8.5.2 Choosing the Alarm Source
In most cases the HR and PR numerics are identical. In order to avoid simultaneous alarms
on HR and PR, the monitor uses either HR or PR as its active alarm source. To change the
alarm source, select [Alm Source] in the [ECG Setup] menu and then select either:
[HR]: if you want the HR to be the alarm source for HR/PR.
[PR]: if you want the PR to be the alarm source for HR/PR.
[Auto]: If the [Alm Source] is set to [Auto], the patient monitor will use the heart rate
from the ECG measurements as the alarm source whenever a valid heart rate is available.
If the heart rate becomes unavailable, for example the ECG module is turned off or
becomes disconnected, the patient monitor will automatically switch to PR as the alarm
source.
8.5.3 Setting the ECG Lead Set
You can set the [Lead Set] by selecting [ECG Setup][Others>>]. You can set the [Lead
Set] as [Auto] if the auto lead detection function is available.
8.5.4 Choosing an ECG Display Screen
When monitoring with a 5-lead set or 12-lead set, you can select the [Screens] Quickkey. In
the [Choose Screen] window, choose the screen type as:
8-7
Page 100
WARNING
[Normal Screen]: The ECG waveform area shows 2 ECG waveforms.
[ECG 7-Lead Full-Screen]: The whole waveform area shows 7 ECG waveforms only.
[ECG 7-Lead Half-Screen]: The upper half part of the whole waveform area displays 7
ECG waveforms.
When monitoring with a 12-lead set, you can also choose the screen type as [ECG 12-Lead
Full-Screen].
When the screen type is set to [Normal Screen], cascaded ECG waveforms can be displayed.
To cascade ECG waveforms:
1. Select the [Screens] Quickkey[Screen Setup].
2. Select [ECG1 Casc.] in the second row. A cascaded waveform is displayed in two
waveform positions.
8.5.5 Changing the ECG Filter Settings
The ECG filter setting defines how ECG waves are smoothed. To change the filter setting,
select [Filter] from [ECG Setup] and then select the appropriate setting.
[Monitor]: Use under normal measurement conditions.
[Diagnostic]: Use when diagnostic quality is required. The unfiltered ECG wave is
displayed so that changes such as R-wave notching or discrete elevation or depression of
the ST segment are visible.
[Surgery]: Use when the signal is distorted by high frequency or low frequency
interference. High frequency interference usually results in large amplitude spikes
making the ECG signal look irregular. Low frequency interference usually leads to
wandering or rough baseline. In the operating room, the surgery filter reduces artifacts
and interference from electrosurgical units. Under normal measurement conditions,
selecting [Surgery] may suppress the QRS complexes too much and then interfere with
ECG analysis.
[ST]: Use when ST monitoring is applied.
z The [Diagnostic] filter is recommended when monitoring a patient in an
environment with slight interference only.
8.5.6 Setting the Notch Filter
The notch filter removes the line frequency interference. Only when [Filter] is set to
[Diagnostic], the [Notch Filter] is adjustable.
8-8
Loading...