For this Operator’s Manual, the issue date is December, 2018.
I
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this manual. This
manual may refer to information protected by copyright or patents and does not convey any
license under the patent rights or copyright of Mindray, or of others.
Mindray intends to maintain the contents of this manual as confidential information.
Disclosure of the information in this manual in any manner whatsoever without the written
permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution,
rental, adaptation, translation or any other derivative work of this manual in any manner
whatsoever without the written permission of Mindray is strictly forbidden.
,andare the trademarks, registered or otherwise, of
Mindray in China and other countries. All other trademarks that appear in this manual are
used only for informational or editorial purposes. They are the property of their respective
owners.
II
Responsibility on the Manufacturer
Party
Contents of this manual are subject to change without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national and
local requirements; and
the product is used in accordance with the instructions for use.
WARNING
It is important for the hospital or organization that employs this equipment to
carry out a reasonable service/maintenance plan. Neglect of this may result in
machine breakdown or personal injury.
NOTE
This equipment must be operated by skilled/trained clinical professionals.
III
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
Malfunction or damage caused by improper use or man-made failure.
Malfunction or damage caused by unstable or out-of-range power input.
Malfunction or damage caused by force majeure such as fire and earthquake.
Malfunction or damage caused by improper operation or repair by unqualified or
unauthorized service people.
Malfunction of the instrument or part whose serial number is not legible enough.
Others not caused by instrument or part itself.
EC-Representative: Shanghai International Holding Corp. GmbH (Europe)
Address: Eiffestraβe 80, 20537 Hamburg, GERMANY
Tel: 0049-40-2513175
Fax: 0049-40-255726
V
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely and in
accordance with its function and intended use. Observance of this manual is a prerequisite for
proper product performance and correct operation and ensures patient and operator safety.
This manual is based on the maximum configuration and therefore some contents may not
apply to your product. If you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to the equipment
so that it can be obtained conveniently when needed.
Intended Audience
This manual is geared for clinical professionals who are expected to have a working
knowledge of medical procedures, practices and terminology as required for monitoring of
critically ill patients.
Illustrations
All illustrations in this manual serve as examples only. They may not necessarily reflect the
setup or data displayed on your ventilator.
Conventions
Italic text is used in this manual to quote the referenced chapters or sections.
[ ] is used to enclose screen texts.
→ is used to indicate operational procedures.
Password
A password is required to access different menus within the ventilator.
Indicates an imminent hazard that, if not avoided, will result in death or serious
injury.
WARNING
Indicates a potential hazard or unsafe practice that, if not avoided, could result in
death or serious injury.
CAUTION
Indicates a potential hazard or unsafe practice that, if not avoided, could result in
minor personal injury and/or product/property damage.
NOTE
Provides application tips or other useful information to ensure that you get the
most from your product.
1-1
1.1.1 DANGER
There are no dangers that refer to the product in general.
1.1.2 WARNING
WARNING
The ventilator must only be operated and used by authorized medical personnel
well trained in the use of this product. Any unauthorized or untrained personnel
should not perform any operations. It must be operated strictly following the
Operator’s Manual.
Before putting the system into operation, the operator must verify that the
equipment, connecting cables and accessories are in correct working order and
operating condition.
To avoid the risk of electric shock, this equipment must be connected to a properly
installed power outlet with protective earth contacts only. If the installation does
not provide for a protective earth conductor, disconnect it from the power line. In
this case, lithium ion batteries should be used temporarily to supply power to the
equipment.
Use external power source (AC power) before the batteries are depleted.
To avoid explosion hazard, do not use the equipment in the presence of flammable
anesthetic agent, vapors or liquids. When O
any fire sources.
Do not place the ventilator adjacent to any barrier, which can prevent cold air
from flowing, resulting in equipment overheat.
Do not open the case of the equipment, as you may suffer an electric shock. All
servicing and future upgrades must be carried out by the personnel trained and
authorized by us only.
Users should set alarm volume and alarm limits based on patients’ actual
condition. Do not rely exclusively on the audible alarm system for patient
monitoring. Adjustment of alarm volume to a low level may result in a hazard to
the patient. Always keep the patient under close surveillance.
The physiological parameters and alarm messages displayed on the screen of the
equipment are for doctor’s reference only and cannot be directly used as the basis
for clinical treatment.
To dispose of the package material, observe the applicable waste control
regulations. And keep the package material out of children’s reach.
is used, keep the ventilator away from
2
All staff should be aware that disassembling or cleaning some parts of the
ventilator can cause risk of infection.
1-2
WARNING
Maintenance menu can only be accessed when the equipment is disconnected from
the patient.
Positive pressure ventilation may be accompanied by some side effects such as
barotrauma, hypoventilation, hyperventilation, etc.
Using high frequency electrosurgery equipment, defibrillators, or short-wave
treatment equipment in the vicinity of the ventilator may interfere with its
operation and pose a risk of patient injury.
Do not use antistatic or conductive masks or patient tubing. They can cause burns
if they are used near high frequency electrosurgery equipment.
Do not use the ventilator in a hyperbaric chamber.
If the equipment internal monitoring system malfunctions, an alternative plan
must be available to ensure adequate level of monitoring. The operator of the
ventilator must be responsible for patient’s proper ventilation and safety under all
circumstances.
As required by the relevant rules and regulations, oxygen concentration should be
monitored when the equipment is used on the patient. If your ventilator is not
configured with such monitoring function or this function is turned off, use a
monitor which complies with the requirements of ISO 80601-2-55 for oxygen
concentration monitoring.
All analog or digital products connected to this system must be certified to the
specified IEC standards (such as IEC 60950 for data processing equipment and
IEC 60601-1 for medical electrical equipment). All configurations shall comply
with the valid version of IEC 60601-1. The personnel who are responsible for
connecting the optional equipment to the I/O signal port shall be responsible for
medical system configuration and system compliance with IEC 60601-1 as well.
Do not touch the patient when connecting the peripheral equipment via the I/O
signal ports or replacing the O
cell, to prevent patient leakage current from
2
exceeding the requirements specified by the standard.
This equipment is not suitable for use in an MRI environment.
When the ventilator’s gas supply input system fails or has faults, please contact us
immediately for service by specified personnel.
The ventilator shall not be used with helium or mixtures with Helium.
Do not move the ventilator before removing the support arm from it, in order to
avoid the ventilator getting tilted during the movement.
The oxygen and air gas mixer of the ventilator is without grease and thus no
de-grease process is needed. Do not use lubricants that contain oil or grease, and
rubber hose assembly should not be contaminated with grease. Lubricants will
burn or explode when exposed to high O
concentrations.
2
1-3
WARNING
The maximum pressure of hose is 1.4MPa@21℃ and please check whether gas
supply pressure meets hose requirements before usage.
Hose connectors adopt standardized gas terminal connector with gas nature.
Different types of gas and gas with different pressures shall not be exchanged with
each other.
Hose may be aging quickly by long-term exposure to acidity, alkalinity or
ultraviolet rays.
Don’t cascade two or more hose assemblies together.
The ventilator arm could bear 1kg maximally and don’t hang over 1kg goods.
After the ventilator is installed or the main control board is replaced, the altitude
must be reset. After resetting the altitude value, please perform flow calibration
(factory).
When disconnecting fast connectors, please operate by two hands to prevent
potential injury caused by sudden pressure release.
Do not block the air intake on the side of the ventilator.
To prevent interrupted operation of the ventilator due to electromagnetic
interference, avoid using the ventilator adjacent to or stack with other device. If
adjacent or stacked use is necessary, verify the ventilator’s normal operation in the
configuration in which it will be used.
To prevent possible personal injury and equipment damage, ensure that the
ventilator is secured to the trolley or placed on the safe and smooth surface.
To prevent possible equipment damage, avoid tipping over the ventilator when
crossing thresholds.
To prevent possible equipment damage, push the brake down when parking the
ventilator.
Avoid the use of polluted air. When the equipment uses air as gas source for
ventilation, if the air is polluted, harmful substance may enter the patient tubing.
To prevent patient injury caused by equipment malfunction, when the alarm
[Technical Error**] occurs, remove the equipment immediately, record failure
code, and contact the Customer Service Department.
To prevent possible ventilator malfunction, do not spill liquid onto the ventilator.
Backup air supply could cause gas to be heated. To reduce the temperature of gas
inside the tubing and prevent patient injury accordingly, ensure that the length of
patient tubing from the humidifier to Y piece is greater than 1.2m.
The internal electrical power source is to be used temporarily if the integrity of the
protective earth conductor or the protective grounding system in the installation is
in doubt.
1-4
WARNING
Nebulization or humidification can increase the resistance of breathing system
filters, and that you need to monitor the filter frequently for increased resistance
and blockage.
The ventilation accuracy can be affected by the gas added by use of a nebulizer.
The ventilator shall not be used with nitric oxide.
Check if the alarm limit settings are appropriate before taking measurement.
When operating the unit with the power supply unit, always connect the unit to an
easily accessible outlet so that it can be unplugged quickly in the event of a
malfunction.
No modification of this equipment is allowed.
Stop using the ventilator and contact us immediately when the buzzer sounds.
Please place cables of neonatal flow sensor correctly, to avoid patients from
becoming entangled or unplanned extubation.
System leakage, such as leakage caused by an uncuffed endotracheal tube, may
influence airflow readings, including airflow parameters, pressure, dead space, and
CO
production.
2
When ventilator is connected to patient, do not remove or replace fuse, or perform
any other maintenance tasks. Such tasks must be performed when the patient is not
using the ventilator.
Please ensure that the AC power cord is disconnected before removing or replacing
the fuse.
HAZARD can exist if different ALARM PRESETS are used for the same or
similar equipment in any single area. Please read the manual and confirm the
correct alarm pre-settings for the ventilator before using it.
1-5
1.1.3 CAUTION
CAUTION
The ventilator must be inspected and serviced regularly by trained service
personnel.
To ensure patient safety, always prepare resuscitator for use.
Always have a special person attend and monitor the operation of the equipment
once the ventilator is connected to the patient.
During the operation of the ventilator, do not disassemble the inspiration safety
valve and expiration valve unless in standby status.
To ensure patient safety, use only parts and accessories specified in this manual.
At the end of its service life, the equipment, as well as its accessories, must be
disposed of in compliance with the guidelines regulating the disposal of such
products.
Magnetic and electrical fields are capable of interfering with the proper
performance of the equipment. For this reason, ensure that all external devices
operated in the vicinity of the equipment comply with the relevant EMC
requirements. Mobile phone, X-ray equipment or MRI devices are a possible
source of interference as they may emit higher levels of electromagnetic radiation.
This system operates correctly at the electrical interference levels identified in this
manual. Higher levels can cause nuisance alarms that may stop mechanical
ventilation. Pay attention to false alarms caused by high-intensity electrical fields.
Before connecting the equipment to the power line, check that the voltage and
frequency ratings of the power line are the same as those indicated on the
equipment’s label or specified in this manual.
Always install or carry the equipment properly to avoid damage caused by
dropping down, impact, strong vibration or other mechanical force.
Check whether the repetitive patient tubing is damaged or leaked before usage. If
so, don’t use such tubing.
To electrically isolate the ventilator circuits from all poles of the supply mains
simultaneously, disconnect the mains plug.
To minimize the risk of fire, do not use supply hose assembly that is worn or
contaminated with combustible materials like grease or oil.
It is the clinician’s responsibility to ensure that all ventilator settings are
appropriate.
To prevent possible patient injury, ensure the ventilator is set up for appropriate
patient type with the appropriate patient tubing. Ensure that the System Check or
1-6
CAUTION
tubing check is performed before each patient.
Perform Flow Sensor Calibration before the first use, or when the measured values
have deviations.
To prevent possible patient injury, ensure the ventilation parameters are set up
properly before ventilating the patient.
To ensure the accuracy of oxygen monitoring, replace an exhausted O
cell as soon
2
as possible or use an external monitor that complies with ISO 80601-2-55.
A fan failure could result in oxygen enrichment inside the ventilator and a
subsequent fire hazard.
To reduce the risk of explosion, do not force the chemical O
cell open or place it
2
close to a source of heat.
When ventilating with a mask, avoid high airway pressures. High pressures may
cause gastric distension.
Peak pressures exceeding 33 cmH
O may increase the risk of gastric insufflation
2
when the ventilation type is non-invasive. When ventilating with such pressures,
consider using an invasive mode.
To reduce the risk of fire, use only tube systems approved for medical purposes
and for use with oxygen between the oxygen source and ventilator.
To reduce the risk of fire, ensure adequate ventilation at the rear of the ventilator.
To reduce the risk of fire, switch off the oxygen source when the ventilator is not in
a ventilating mode.
Avoid putting the ventilator in the storage environment of more than 50℃ for a
long time. Such environment may damage or shorten the battery lives of internal
battery and O
cell.
2
Use the original packing materials to ship the ventilator.
To prevent fire hazard, use only specified fuses or fuses with the same type, rated
voltage, and rated current as the existing fuses. When it is necessary to replace
fuses, contact the Customer Service Department.
The ventilator is suitable for use within the PATIENT ENVIRONMENT.
Additional MULTIPLE SOCKET- OUTLET or extension cord shall not be
connected to the system.
Before moving the ventilator, ensure that the casters and brakes can work
properly, and the main unit is locked on the trolley.
Please use dry and clean medical compressed air and oxygen as gas supply. Water
in gas supply can cause equipment malfunction.
1-7
1.1.4 NOTE
NOTE
Put the ventilator and its accessories in a location where you can easily see the
screen and access the operating controls.
Keep this manual close to the equipment so that it can be obtained conveniently
when needed.
The software was developed in compliance with IEC 62304. The possibility of
hazards arising from software errors is minimized.
This manual describes all features and options. Your equipment may not have all
of them.
When the oxygen supply is insufficient, the ventilator will automatically switch to
air supply. When the air supply is insufficient, the ventilator will automatically
switch to oxygen supply.
The ventilator is equipped with barometric pressure sensors, and has the function
of barometric pressure compensation.
1.2 Equipment Symbols
AUDIO PAUSED
Recent Alarm
Alarm settings
Nebulizer
ALARM OFF
Clear alarm
O2↑key
Tools key
Setup
Preventive
maintenance
Standby key
Screen Capture
1-8
History
Freeze
Inspiratory trigger
icon
Adult (male)
Pediatric (male)
Backup Air Supply
Invasive Ventilation
Non-Invasive
Ventilation
Adult (female)
Pediatric (female)
Adjust screen
Neonate
brightness/volume to
night mode
Adjust screen
Start O2 therapy timer
brightness/volume to
day mode
Stop O2 therapy timer
Caution
Fuse
Battery LED
RS-232 connector
USB connector
Display connector
Oxygen sensor
connector
Reset O2 therapy
timer
Equipotentiality
Protective earth
ground
AC power
VGA output
connector
Network connector
Nurse call connector
Pneumatic nebulizer
connector
Lock
Unlock
1-9
Power switch
Expiration connector
Oxygen supply
connector
Date of manufacture
Serial number
Keep dry
Humidity limitation
Neonatal flow sensor
connector
Inspiration connector
Air supply connector
Manufacturer
European community
representative
Temperature
limitation
Atmospheric pressure
limitation
IP21
This way up
Recyclable
Degree of protection
against harmful
ingress of water
Fragile, handle with
care
Stacking limit by
number
Not suitable for use
in an MRI
environment
High Efficiency
Particle Air (HEPA)
Water trap indicator
installation instruction
Refer to the operator's
manual
Defibrillation-proof
BF application part
Ventilator gas outlet
DEFIBRILLATION-
PROOF TYPE CF
APPLIED PART
1-10
Reset the paramagnetic oxygen sensor (NOTE: This operation can be
performed only by the Customer Service Department or authorized
personnel.)
The following definition of the WEEE label applies to EU member
states only.
This symbol indicates that this product should not be treated as
household waste. By ensuring that this product is disposed of
correctly, you will help prevent bringing potential negative
consequences to the environment and human health. For more detailed
information with regard to returning and recycling this product, please
consult the distributor from whom you purchased it.
* For system products, this label may be attached to the main unit
only.
The product bears CE mark indicating its conformity with the
provisions of the Council Directive 93/42/EEC concerning medical
devices and fulfills the essential requirements of Annex I of this
directive.
Unified circulation mark indicates that products marked them passed
all specified in the technical regulations of the Customs Union of the
procedure for the assessment (confirmation) of conformity and
complies with the requirements applicable to all the products technical
regulations of the Customs Union.
1-11
FOR YOUR NOTES
1-12
2 The Basics
2.1 System Description
2.1.1 Intended Use
This product is intended to be used in intensive care situations within a professional
healthcare facility, or during transport within a professional healthcare facility. This product
is intended to provide ventilation assistance and breathing support for adult, paediatric and
neonate patients. The product should be operated by properly-trained and authorized medical
personnel. This equipment is not suitable for use in an MRI environment.
2.1.2 Contraindications
There are no absolute contraindications for this product. For some special diseases, however,
some necessary treatments shall be taken for ventilator mechanical ventilation, or special
ventilation modes shall be adopted to prevent possible patient injury.
2.1.3 Components
The ventilator consists of a main unit (including pneumatic circuit, electronic system,
mechanical structure, display, CO
(model: C3), trolley, and support arm.
The ventilator is suitable for use within the PATIENT ENVIRONMENT. Connect the patient
to the ventilator via the patient breathing circuit. The applied part of the ventilator is
breathing masks.
module, SpO2 module), backup air supply, air compressor
2
2-1
2.2 Equipment Appearance
2.2.1 Front View
8
7
6
5
4
3
2
11
12
9
13
10
1
2-2
Loading...
+ 250 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.