All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without
the prior written authorisation of KROHNE Messtechnik GmbH.
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
5
1 SAFETY INSTRUCTIONS
1.1 Intended use
DANGER!
For devices used in hazardous areas, additional safety instructions apply.
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and
corrosion resistance of the used materials against the measured fluid lies solely with the
operator.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
The OPTIBAR PC 5060 C process pressure transmitter is suitable for measuring the process
pressure and level of gases, vapours and liquids. The available measuring ranges and permitted
maximum working pressures for each are indicated on the nameplate and described in the refer
Technical data
to
points:
on page 59 section. To observe the intended use, adhere to the following
OPTIBAR PC 5060 C
• Observe the instructions in this document.
• Comply with the technical specifications (for further information refer to
page 59).
• Only suitably qualified personnel may install and operate the device.
• Observe the generally accepted standards of good practice.
Technical data
CAUTION!
•
Any modification to the device, including drilling, sawing, trimming, welding and soldering of
parts, or partially painting over or coating, is prohibited.
•
Neither is it permitted to use the device as a climbing aid e.g. for installation purposes, as a
holder for cables, pipes or other loads.
•
The mounting or installation of parts is only permitted as described in this document, or
insofar as it has been authorised by the manufacturer or a certified service partner.
on
6
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
1.2 Technical limits
The device was constructed solely for use within the technical limits indicated on the nameplate
and in the technical data. Applications outside of these limits are not permitted and could lead to
significant risk of accident. For this reason, observe the following limits:
• Do not exceed the maximum working pressure (MWP).
• Do not exceed the indicated permissible operating temperature range.
• The permissible ambient temperatures given may not be exceeded or undershot.
• Observe the ingress protection of the housing during use.
1.3 Permissible mediums
The device is designed to measure the pressure of vaporous, gaseous and liquid media. Prior to
using any corrosive or abrasive products, the operator must check the resistance of all materials
which are in contact with the product.
1.4 Certification
SAFETY INSTRUCTIONS 1
CE marking
The device fulfils the statutory requirements of the following EC directives:
• EMC Directive 2004/108/EC
• EMC specification acc. to EN 61326/A1
The manufacturer certifies successful testing of the product by applying the CE marking.
Devices with a permissible pressure PS ≤ 200 bar (20 MPa) comply with Article 3 Section (3) and
are not subject to a conformity assessment. These devices were designed and manufactured in
accordance with sound engineering practice (SEP).
The CE marking on the device does not apply to the pressure equipment directive.
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
7
1 SAFETY INSTRUCTIONS
1.5 Safety instructions from the manufacturer
1.5.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
OPTIBAR PC 5060 C
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.5.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
8
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
1.5.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation and operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.5.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of underneath icons.
SAFETY INSTRUCTIONS 1
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
9
1 SAFETY INSTRUCTIONS
1.5.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This warning refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
OPTIBAR PC 5060 C
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.6 Safety instructions for the operator
10
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
2.1 Scope of delivery
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
DEVICE DESCRIPTION 2
Figure 2-1: Scope of delivery
1 Device in the version as ordered
2 Mounting bracket
3 Documentation (test reports, factory and material certification (if ordered) and product documentation)
Optional accessories
• Gaskets
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
11
2 DEVICE DESCRIPTION
2.2 Device description
The OPTIBAR PC 5060 C is suited to applications in virtually all areas of industry. It is used to
measure the following types of pressure:
• Gauge pressure
• Absolute pressure
• Vacuum
The setup of the device is carried out via the display and adjustment unit. For further information
refer to
Keypad functions
A capacitive ceramic sensor element is used in the measuring cell.
The measuring device is supplied ready for operation. The factory settings for the process data
correspond to your order specifications.
For safety reasons do not exceed the measuring range or permissible process pressure. This
also applies when, based on the order, a measuring cell is installed with a higher measuring
range than the permissible pressure range of the process connection.
on page 41.
OPTIBAR PC 5060 C
2.2.1 Device design
The following drawing shows the basic components of the pressure transmitter.
Figure 2-2: Basic components of single chamber pressure transmitter
1 Housing cover, optional with display and adjustment module below
2 Housing with electronics
3 Process assembly with measuring cell
12
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
Figure 2-3: Basic components of double chamber pressure transmitter
1 Housing cover
2 Housing with electronics
3 Process assembly with measuring cell
4 Housing cover, optional with display and adjustment module below
5 Operating and display module
DEVICE DESCRIPTION 2
2.3 Nameplates
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
Figure 2-4: Example for nameplate for a OPTIBAR * 5060
1 Observe the installation and operating instructions
2 CE marking and marking of notified body
3 Hardware and Software version
4 Product name and type code
5 Nominal range
Permissible process pressure
6 Permissible temperature range
7 Electronics power supply and signal output
8 Ingress protection and material of wetted parts
(Diaphragm, process connections, gasket and fill fluid)
9 Approvals and approval guidelines
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
13
2 DEVICE DESCRIPTION
2.4 Terms and abbreviations
The following terms and abbreviations are used in this document.
URL
URL
URLURL
Upper Range Limit
LRL
LRL
LRLLRL
Lower Range Limit
URV
URV
URVURV
Upper Range Value
LRV
LRV
LRVLRV
Lower Range Value
SPAN
SPAN
SPANSPAN
Span
CAL SPAN
CAL SPAN
CAL SPANCAL SPAN
calibrated Span
TD
TD
TDTD
Turn Down
The following applies: URV ≤ URL, CAL SPAN ≤ SPAN, TD ≥ 1
Example: URL
Upper measuring range limit. Also called nominal range. The highest value that
can be measured by a particular device.
Lower measuring range limit. The lowest value that can be measured by a
particular device.
The calibrated measuring range or the highest adjusted measured value. This
value corresponds to the 20 mA signal.
The calibrated measuring range or the lowest adjusted measured value. This
value corresponds to the 4 mA signal.
Span or measuring range. SPAN = URL – LRL
Calibrated or adjusted measuring span. CAL SPAN = URV – LRV. Also called
"cSPAN".
This is the span set to the 4…20 mA output.
The ratio from the measuring span to the adjusted measuring span.
TD= SPAN/(CAL SPAN)= (+URL)/(CAL SPAN)
URL= 3 bar
URLURL
LRL
LRL= 0 bar
LRLLRL
URV
URV= 2 bar
URVURV
LRV
LRV= 0.5 bar
LRVLRV
OPTIBAR PC 5060 C
SPAN
SPAN= 3 bar
SPANSPAN
CAL SPAN
CAL SPAN= 1.5 bar TD
CAL SPANCAL SPAN
TD= 2:1
TDTD
14
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
2.5 Sealing concept
2.5.1 Sealing concept for recessed installation
Figure 2-5: Recessed installation of the CERTEC® measuring cell
1 Measuring cell
2 Measuring cell seal
3 Process diaphragm
4 Process connection
5 Seal for the process connection (optional)
DEVICE DESCRIPTION 2
2.5.2 Sealing concept for flush mounted installation with single seal
Figure 2-6: Flush mounted installation of the CERTEC® measuring cell
1 Measuring cell
2 Measuring cell seal
3 Process diaphragm
4 Process connection
5 Seal for the process connection (optional)
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
15
2 DEVICE DESCRIPTION
2.5.3 Sealing concept for flush mounted installation with double seal
Figure 2-7: Flush mounted installation of the CERTEC® measuring cell with double seal
1 Measuring cell
2 Measuring cell seal
3 Process diaphragm
4 Process connection
5 Additional front measuring cell seal
2.5.4 Sealing concept for hygienic installation
OPTIBAR PC 5060 C
Figure 2-8: Hygienic installation of the CERTEC® measuring cell
1 Measuring cell
2 Measuring cell moulded seal
3 Process diaphragm
4 Process connection
5 Gap-free seal for process connection
16
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
3.1 General notes on installation
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Protection category of the housing
The housing of the differential pressure transmitter fulfills the requirements for ingress
protection in accordance with IEC 60529. Housing for protection category IP69K in accordance
with ISO 20653 is also available. For further information refer to
INSTALLATION 3
Technical data
on page 59.
CAUTION!
The first digit stands for the protection of the inner electronic components against the ingress of
foreign bodies including dust. The first digit "6" means that the housing is dust-proof. The
second digit designates the protection of the inner electronic components against the ingress of
water. The second digit "6" means that the housing is waterproof and also resistant against a
strong jet of water resists. The number "7" means that the case is waterproof even under water
for a given pressure and time. The number "8" means that the housing is permanently
waterproof even under water.
3.3 Packaging
CAUTION!
Devices for oxygen applications are sealed in PE foil and provided with a label "Oxygen! Use no
Oil". Remove this foil just before mounting the device! After removing the protection for the
process connection the label OOOO
grease and dirt should be avoided. Danger of explosion!
Your device was protected by packaging during transport. Its capacity to handle normal loads
during transport is assured by a test following ISO 22248. The packaging of standard devices
consists of environmentally friendly, recyclable cardboard and PE foil. For special versions, PE
foam or PE foil is also used. Dispose of the packaging material via specialised recycling
companies.
will be visible on the process connection. Penetration of oil,
2222
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
17
3 INSTALLATION
3.4 Storage
CAUTION!
Observe the storage information found on the packaging. Labels on the original packaging must
always remain legible and may not be damaged.
• Store the device in a dry and dust-free location.
• Avoid lasting direct exposure to the sun.
• Store the device in the original packaging supplied.
• Do not expose to aggressive media.
• Avoid mechanical shocks.
• Storage temperature of -40 to +80°C / -40 to +176°F.
• Relative air humidity of 20 to 85%.
3.5 Transport
• Use original packaging for transport and ensure that the packaging does not get crushed or
damaged by sharp objects or other boxes.
• When transporting by ship, use seaworthy outer packing.
OPTIBAR PC 5060 C
3.6 Installation specifications
INFORMATION!
Observe the relevant directives, ordinances, standards and accident prevention regulations (e.g.
VDE/VDI 3512, DIN 19210, VBG, Elex V, etc.).
Ensure that all of the parts in the process are suitable for the current process conditions. This
includes in particular:
• Parts active in the measurement
• Process connection
• Process seal
Process conditions include in particular:
• Process pressure
• Process temperature
• Chemical properties of the media
• Abrasion and mechanical impact
18
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
3.7 Mounting
CAUTION!
•
Prior to installing the pressure transmitter, it is essential to verify whether the version of the
device on hand completely fulfils the technical and safety requirements of the measuring
point. This applies in particular to the measuring range, overpressure resistance,
temperature, explosion protection and operating voltage.
•
Check the materials used for the wetted parts (e.g. gasket, process connection, separating
diaphragm etc.) for suitability as regards process compatibility.
3.7.1 Rotating the housing
The transmitter housing can be rotated 350° for better readability of the display or access to the
wiring. A stop prevents the housing from being rotated too far.
• On all 2 chamber housings, the locking screw must first be loosened at the neck of the
housing.
iThe housing can then be rotated to the desired position.
• Once the desired position is reached, the locking screw is tightened again.
INSTALLATION 3
3.7.2 Mounting the display and adjustment module
The optional display and adjustment module can be set in any one of four different positions at
90° intervals. The installation of the adjustment module is carried out as per the illustrations
below. To do so, unscrew the housing cover and insert the adjustment module clockwise. The
display can be installed rotated at 90°. It is not necessary to interrupt the power supply.
Figure 3-1: Installation in single chamber housing
1 Insert the display and adjustment module into the housing
2 Turn the display and adjustment module clockwise
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
19
3 INSTALLATION
Figure 3-2: Installation in double chamber housing
1 Insert the display and adjustment module into the housing
2 Turn the display and adjustment module clockwise
3 Mounting on top
4 Mounting at side
OPTIBAR PC 5060 C
3.7.3 Temperature limits
Higher process temperatures often mean also higher ambient temperatures for electronics and
connection cables. Make sure that the upper temperature limits for the environment of the
electronics housing and connection cable are not exceeded. For further information refer to
Technical data
on page 59.
3.8 Instructions for oxygen applications
Oxygen and other gases can be explosive when brought into contact with oils, grease and
plastics, so the following measures must also be taken:
• All components of the plant, such as e.g. measuring devices must be cleaned according to the
requirements of BAM (DIN 19247).
• Depending on the seal material, certain temperatures and pressures must not be exceeded
in oxygen applications, refer to
CAUTION!
Devices for oxygen applications are sealed in PE foil and provided with a label "Oxygen! Use no
Oil". Remove this foil just before mounting the device! After removing the protection for the
process connection the label OOOO
grease and dirt should be avoided. Danger of explosion!
Technical data
will be visible on the process connection. Penetration of oil,
2222
on page 59.
20
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
3.9 Venting
CAUTION!
The filter element causes a delayed pressure equalisation when the housing cover is opened or
closed quickly. During the process the measurement can change by up to 15 mbar for up to 5
seconds.
CAUTION!
In order to ensure effective ventilation, the filter element must be always free of deposits.
CAUTION!
Do not use a high-pressure cleaner to clean the housing. The filter element may become
damaged and as a result moisture can penetrate into the housing. The exception to this is the
IP69K single chamber housing.
The ventilation for the electronics housing is assured via a filter element in the vicinity of the
cable glands, which is permeable to air but water-absorbent.
INSTALLATION 3
Figure 3-3: Venting in non-Ex, Ex ia and Ex d ia versions
1 Single chamber housing, plastic, stainless steel precision casting
2 Single chamber housing, aluminium
3 Single chamber housing, electro-polished stainless steel
4 Double chamber housing, plastic
5 Double chamber housing, aluminium
6 Filter element
The following devices feature a dummy plug instead of a filter element:
• Devices in the IP 66 / IP 68 (1bar) protection category - venting via capillary tube in
permanently connected cable
• Devices with absolute pressure
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
21
3 INSTALLATION
Devices in Ex d version
Figure 3-4: Ventilation Ex-housing
1 Single chamber housing, aluminium and stainless steel precision casting
2 Double chamber housing, aluminium and stainless steel precision casting
3 Rotating metal ring
4 Filter element
The filter element is integrated into the sensor assembly via a rotating metal ring. Align the
metal ring towards the bottom to better protect the filter element from deposits. Devices with
absolute pressure feature a dummy plug instead of a filter element.
OPTIBAR PC 5060 C
Devices with second process barrier
Figure 3-5: Ventilation gas-proof feedthrough
1 Filter element
For devices with a second process barrier, the sensor assembly is completely encased by an
additional, gas-proof feedthrough. Additional venting is not required for absolute pressure
sensors. With relative pressure sensors, the ambient pressure is measured and compensated by
an additional sensor in the electronics.
22
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
INSTALLATION 3
Devices in IP69K version
Figure 3-6: Ventilation IP69K
Filter element
Devices with absolute pressure feature a dummy plug instead of a filter element.
3.10 Measurement setup for process pressure measurement
The following points should be observed in this application:
• The signal converter must be mounted above the measuring point.
Figure 3-7: Measurement setup for measuring the process pressure of gases
1 Pressure transmitter
2 Shut-off valve
3 Tapping point
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
23
3 INSTALLATION
3.11 Measurement setup for steam measurement
The following points should be observed in this application:
• The pressure transmitter should be connected via a siphon to protect the measuring cell
from non-permitted high temperatures.
• Siphon to be kept free of insulation.
• When using superheated steam, the siphon must be filled with water prior to start-up.
OPTIBAR PC 5060 C
Figure 3-8: Measurement setup for steam measurement
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
OPTIBAR PC 5060 C
Figure 3-13: External housing (IP 68)
Electronic insert
Cable gland for the power supply
Cable gland for the sensor connection cable
INSTALLATION 3
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
27
4 ELECTRICAL CONNECTIONS
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
OPTIBAR PC 5060 C
4.2 Notes for electrical cables
DANGER!
The device must be grounded to a spot in accordance with regulations in order to protect
personnel against electric shocks.
DANGER!
Cables may only be connected when the power is switched off! Since the transmitter has no
switch-off elements, overcurrent protection devices, lightning protection and/or energy isolating
devices must be provided by the customer.
28
www.krohne.com04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en
Loading...
+ 64 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.