SMARTSENS ORP 8150
SMARTSENS ORP 8150
SMARTSENS ORP 8150SMARTSENS ORP 8150
Digital ORP sensor
Category
II 1 G
EPL Ga
Supplementary instructions
Supplementary instructions
Supplementary instructionsSupplementary instructions
© KROHNE 07/2014 - 4003048502 - MA SMARTSENS ORP 8150 R02 en AD
CONTENTS
SMARTSENS ORP 8150
1 Safety instructions 3
1.1 General notes ................................................................................................................... 3
1.2 EC conformity ................................................................................................................... 3
1.3 Approval according to the IECEx scheme ........................................................................ 3
1.4 Safety instructions............................................................................................................ 3
2 Device description 4
2.1 Device description ............................................................................................................ 4
2.2 Marking............................................................................................................................. 5
2.3 Flammable products ........................................................................................................ 7
2.4 Equipment category .........................................................................................................7
2.5 Protection type for gas hazardous areas ......................................................................... 8
2.6 Temperature classes........................................................................................................8
2.7 Electrical data................................................................................................................... 9
3 Installation 10
3.1 Mounting......................................................................................................................... 10
4 Electrical connections 11
4.1 General notes ................................................................................................................. 11
4.2 Power supply .................................................................................................................. 11
4.3 Inputs/outputs ................................................................................................................ 11
4.4 Grounding and equipotential bonding............................................................................ 12
5 Operation 13
5.1 Starting up for the first time .......................................................................................... 13
5.2 Operation ........................................................................................................................ 13
5.3 Electrostatic charge .......................................................................................................13
6 Service 14
6.1 Maintenance ................................................................................................................... 14
6.2 Dismantling .................................................................................................................... 14
7 Notes 15
2
www.krohne.com 07/2014 - 4003048502 - MA SMARTSENS ORP 8150 R02 en
SMARTSENS ORP 8150
1.1 General notes
This additional Ex instruction applies to explosion-protected versions of the sensors with the
marking II 1 G. It completes the standard documentation for non-explosion protected versions.
The information given in this instruction contains only the data relevant to explosion protection.
The technical details given in the standard documentation for the non-explosion protected
versions remain unchanged unless they will be excluded or replaced by this instruction.
1.2 EC conformity
The manufacturer declares with the EC declaration of conformity on his own responsibility
conformity with the protection goals of directive 94/9/EC for use in hazardous areas with gas.
Conformity with harmonised standards was checked in accordance with EN 60079-0:2012 and
EN 60079-11:2012.
The EC declaration of conformity is based on the EC type examination certificate of the
Physikalisch Technische Bundesanstalt (PTB):
SAFETY INSTRUCTIONS 1
PTB 13 ATEX 2018 X
PTB 13 ATEX 2018 X
PTB 13 ATEX 2018 XPTB 13 ATEX 2018 X
You can download the EC Type Examination Certificate from the manufacturer's website.
1.3 Approval according to the IECEx scheme
Conformity with IECEx standards was tested in accordance with the IECEx Certification Scheme
for Explosive Atmospheres acc. to IEC 60079-0:2011 and IEC 60079-11:2012. The number of the
IEC certificate is:
IECEx PTB 13.0036 X
IECEx PTB 13.0036 X
IECEx PTB 13.0036 XIECEx PTB 13.0036 X
You can download the IEC certificate from the manufacturer's website.
1.4 Safety instructions
Assembly, installation, start-up and maintenance may only be performed by personnel trained in
explosion protection!
CAUTION!
Should operating conditions and locations require the observance of further standards,
guidelines and laws, this is the responsibility of the operator and/or those commissioned by him.
www.krohne.com07/2014 - 4003048502 - MA SMARTSENS ORP 8150 R02 en
3
2 DEVICE DESCRIPTION
2.1 Device description
The sensor is designed to measure the oxidation reduction potential (ORP) in liquids.
It has a connection for a 4..20 mA current output with HART
The connection is designed as a plug connector and at the same time supplies the sensor with
power.
SMARTSENS ORP 8150
®
communication.
1 Nickel plated brass body with VP 2.0 connector
2 PG 13.5 thread
3 Washer
4 O-ring
5 Glass shaft
6 Diaphragm
7 Platinum electrode
4
www.krohne.com 07/2014 - 4003048502 - MA SMARTSENS ORP 8150 R02 en
SMARTSENS ORP 8150
2.2 Marking
The marking of the sensor type SMARTSENS ORP 8150 is shown on the nameplates below. The
main plate is on the main part of the device and there is an additional marking on the glass.
DEVICE DESCRIPTION 2
Figure 2-1: Example nameplate ATEX
1 Manufacturer
2 Type
3 TAG number, order code, serial number D14
4 Ex-data acc. to PTB 13 ATEX 2018 X
5 Observe the installation and operating instructions
14xxxxxx = Year of manufacture 2014
1414
Figure 2-2: Example nameplate IECEx
1 Manufacturer
2 Type
3 TAG number, order code, serial number D14
4 Ex-data acc. to IECEx PTB 13.0036 X
5 Observe the installation and operating instructions
14xxxxxx = Year of manufacture 2014
1414
www.krohne.com07/2014 - 4003048502 - MA SMARTSENS ORP 8150 R02 en
5