KROHNE OPTIMASS-8000K User Manual

Handbook
Handbook
OPTIMASS 8000K
OPTIMASS 8000K
OPTIMASS 8000KOPTIMASS 8000K
HandbookHandbook
Sensor for mass flow
Software revision: V2.2.xx
© KROHNE 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
: IMPRINT ::::::::::::::::::::::::::::::::::
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH.
Subject to change without notice.
Copyright 2010 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany)
2
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
CONTENTS
1 Safety instructions 5
1.1 Software History ............................................................................................................... 5
1.2 Intended use ..................................................................................................................... 5
1.3 CE certification ................................................................................................................. 5
1.4 Associated documents ..................................................................................................... 6
1.5 Pressure Equipment Directive (PED)............................................................................... 7
1.6 Safety instructions from the manufacturer ..................................................................... 7
1.6.1 Copyright and data protection ................................................................................................ 7
1.6.2 Disclaimer ............................................................................................................................... 8
1.6.3 Product liability and warranty ................................................................................................ 9
1.6.4 Information concerning the documentation........................................................................... 9
1.6.5 Warnings and symbols used................................................................................................. 10
1.7 Safety instructions for the operator............................................................................... 10
2 Device description 11
2.1 Scope of delivery............................................................................................................. 11
2.1.1 Meters with hygienic connections ........................................................................................ 12
2.2 Nameplates .................................................................................................................... 12
2.3 Temperature differential and thermal shock ................................................................ 12
3 Installation 13
3.1 Notes on installation ......................................................................................................13
3.2 Storage ........................................................................................................................... 13
3.3 Handling.......................................................................................................................... 14
3.4 Installation conditions ....................................................................................................15
3.4.1 Supporting the meter............................................................................................................ 15
3.4.2 Mounting the meter .............................................................................................................. 16
3.4.3 Gas / liquid build up .............................................................................................................. 17
3.4.4 Side mounting ....................................................................................................................... 17
3.4.5 Cross talk .............................................................................................................................. 18
3.4.6 Flange connections............................................................................................................... 18
3.4.7 Maximum pipework forces (end loadings) ........................................................................... 19
3.4.8 Pipework reducers................................................................................................................ 19
3.4.9 Flexible connections ............................................................................................................. 20
3.4.10 Hygienic installations.......................................................................................................... 20
3.4.11 Heating and insulation ........................................................................................................ 21
3.4.12 Purge ports ......................................................................................................................... 22
3.4.13 Burst dics ............................................................................................................................ 22
3.4.14 Zero calibration................................................................................................................... 23
3.4.15 Sunshades........................................................................................................................... 23
4 Electrical connections 24
4.1 Safety instructions.......................................................................................................... 24
4.2 Electrical and I/O connections ....................................................................................... 24
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
3
CONTENTS
OPTIMASS 8000K
5 Service 25
5.1 Spare parts availability...................................................................................................25
5.2 Availability of services .................................................................................................... 25
5.3 Returning the device to the manufacturer..................................................................... 25
5.3.1 General information.............................................................................................................. 25
5.3.2 Form (for copying) to accompany a returned device............................................................ 26
5.4 Disposal .......................................................................................................................... 26
6 Technical data 27
6.1 Measuring principle (twin tube) ..................................................................................... 27
6.2 Technical data................................................................................................................. 29
6.3 Measuring accuracy ....................................................................................................... 33
6.4 Guidelines for maximum operating pressure................................................................ 34
6.5 Dimensions and weights ................................................................................................ 36
6.5.1 Flanged versions................................................................................................................... 36
6.5.2 NAMUR dimensions .............................................................................................................. 40
6.5.3 Hygienic versions .................................................................................................................. 41
6.5.4 Heating jacket version .......................................................................................................... 45
6.5.5 Purge port option .................................................................................................................. 46
7 Notes 47
4
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
1.1 Software History
Release date Software version Documentation
Aug 2008 V2.2.xx MA MFC 300 R02
1.2 Intended use
This mass flowmeter is designed for the direct measurement of mass flow rate, product density and product temperature. Indirectly, it also enables the measurement of parameters like total mass, concentration of dissolved substances and the volume flow. For use in hazardous areas, special codes and regulations are also applicable and these are specified in a separate documentation.
1.3 CE certification
CE marking
SAFETY INSTRUCTIONS 1
MA OPTIMASS 8000k R01
This device conforms with the following EC directives:
EMC Directive 2004/108/EC
ATEX Directive 94/9/EC
Low Voltage Directive 2006/95/EC
Pressure Equipment Directive 97/23/EC
The manufacturer declares conformity and the device carries the CE mark.
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
5
1 SAFETY INSTRUCTIONS
1.4 Associated documents
This handbook should be read in conjunction with relevant documents in relation to:
hazardous areas
communications
concentration
corrosion
OPTIMASS 8000K
6
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
1.5 Pressure Equipment Directive (PED)
LEGAL NOTICE!
The Pressure Equipment Directive places legal requirements on both the manufacturer and the end user. Please read this section carefully!
Visual check
XXXXX
To ensure the PED integrity of the meter, you MUST check that the serial numbers on the converter nameplate and the sensor nameplate are the same.
To comply with the requirements of the Pressure Equipment Directive (PED) the manufacturer provides all the relevant technical data in the technical data section of this handbook. Secondary pressure containment is NOT supplied on this meter.
SAFETY INSTRUCTIONS 1
Tube failure
Where the meter is being used to measure high pressure gasses and / or gasses kept as liquids by high pressure and / or where there is a risk of tube failure because of the use of corrosive or erosive fluids, frequent pressure and / or thermal cycling, seismic or other shock loading, the burst disc option MUST be purchased. For more information, please contact your nearest representative.
DANGER!
If it is suspected that the primary measuring tube has failed, de-pressurise the meter and remove it from service as soon as it is safe to do so.
1.6 Safety instructions from the manufacturer
1.6.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
7
1 SAFETY INSTRUCTIONS
We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.
1.6.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect, incidental, punitive and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.
OPTIMASS 8000K
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.
8
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
1.6.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.
1.6.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of underneath icons.
SAFETY INSTRUCTIONS 1
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
9
1 SAFETY INSTRUCTIONS
1.6.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.
OPTIMASS 8000K
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems. There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the specified sequence.
i RESULT
RESULT
RESULTRESULT This symbol refers to all important consequences of the previous actions.
1.7 Safety instructions for the operator
10
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device.
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
2.1 Scope of delivery
Compact version
1 Mass flowmeter. 2 Carton. 3 Documentation. 4 2.5 mm and 5 mm hex head tools. 5 CD-ROM and calibration certificate.
DEVICE DESCRIPTION 2
Remote version
1 Mass flowmeter. 2 Converter. This will be either: field (as shown), wall or rack. 3 Carton. 4 2.5 mm and 5 mm hex head tools. 5 CD-ROM and calibration certificate. 6 Documentation.
If any items are missing, please contact the manufacturer.
If your meter has flange connections, the flange specification is stamped on the outer edge of the flange. Check that the specification on the flange is the same as your order.
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
11
2 DEVICE DESCRIPTION
2.1.1 Meters with hygienic connections
1 Fully welded - the O-rings between the meter and the process pipework are not supplied as standard but can be or-
dered.
2 DIN 11864-2 Form A - the O-rings between the Form A and Form B parts of the connection are not supplied as standard
but can be ordered.
3 The 11864-2 Form B is not supplied as part of this connection but it can be ordered.
2.2 Nameplates
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
OPTIMASS 8000K
2.3 Temperature differential and thermal shock
Temperature differential
Temperature differential
Temperature differentialTemperature differential
The maximum difference between ambient temperature and process (operating) temperature is
210°C / 410°F.
Thermal shock
Thermal shock
Thermal shockThermal shock
Thermal shock occurs when there is a sudden and extreme change (shift) in process temperature. To prevent thermal shock on this meter, the manufacturer recommends that you
avoid a temperature shift greater than 120°C / 248°F
CAUTION!
Operation outside these limits may result in shifts in density and mass flow calibration. Repeated shocking may also lead to premature failure of the meter! However, higher thermal shocks are possible at lower working pressures. For more information, please contact your nearest representative.
12
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
3.1 Notes on installation
INFORMATION!
Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Check the packing list to check if you received completely all that you ordered.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
3.2 Storage
Store the device in a dry and dust-free location.
Avoid direct exposure to the sun.
Store the device in its original packing.
Do not allow the ambient temperature to fall below -50°C / -58°F or rise above +85°C /
+185°F.
INSTALLATION 3
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
13
3 INSTALLATION
3.3 Handling
OPTIMASS 8000K
14
1 Use a well maintained sling to lift the meter by the spigots. 2 DO NOT lift the meter by the converter housing or the electronics stem. 3 DO NOT lift the meter by the meter body. 4 DO NOT lift the meter using the flange bolt holes.
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
3.4 Installation conditions
3.4.1 Supporting the meter
Support for meters with flange connections
1 Support the meter as close as possible to the meter body 2 DO NOT leave a long pipe run between the meter and the support. This can cause damage to the meter, especially on
larger meter sizes.
INSTALLATION 3
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
15
3 INSTALLATION
3.4.2 Mounting the meter
Mounting positions
OPTIMASS 8000K
16
1 The meter can be mounted at an angle but it is recommended that the flow is uphill. 2 Avoid mounting the meter with the flow running downhill because it can cause siphoning. If the meter has to be mount-
ed with the flow running downhill, install an orifice plate or control valve downstream of the meter to maintain back­pressure.
3 Horizontal mounting with flow running left to right. 4 Avoid mounting meter with long vertical runs after the meter as it can cause cavitation. Where the installation includes
a vertical run after the meter, install an orifice plate or control valve downstream to maintain backpressure.
5 The meter can be mounted vertically but it is recommended that the flow is uphill. 6 Avoid mounting the meter vertically with the flow running downhill. This can cause siphoning. If the meter has to be
installed this way, install an orifice plate or control valve downstream to maintain backpressure.
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
3.4.3 Gas / liquid build up
In certain applications, the design of the meter can cause either gas or liquid to build up in the measuring tube.
1 Where liquids are being measured, mount the meter as shown. This will prevent gas building up in the measuring tube,
when there is no flow.
2 Where gases are being measured, mount the meter as shown. This will prevent liquids building up in the measuring
tube, when there is no flow.
INSTALLATION 3
3.4.4 Side mounting
The meter can be installed with the converter (or remote junction box) on the side of the meter so that the measuring tubes are sitting one above the other. Avoid this method of installation where there is a two phase process flow, or where the process fluid contains gas. If this situation cannot be avoided, please contact the manufacturer for advice.
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
17
3 INSTALLATION
3.4.5 Cross talk
Where more than one meter is being installed, a very high level of immunity to cross talk means that the meters can be mounted within close proximity to each other. The meters can be mounted either in series or parallel, as shown.
Meters in series
INFORMATION!
Where the meters are installed in series, it is strongly recommended that the process pipe diameter remains constant. For more information, please contact the manufacturer.
OPTIMASS 8000K
Meters in parallel
3.4.6 Flange connections
18
Tighten the flange bolts evenly and in turn.
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
Use a regular pattern to tighten the bolts evenly
3.4.7 Maximum pipework forces (end loadings)
INSTALLATION 3
Mass flowmeters have a maximum level of force (negative or positive) that can be applied to the ends of the meter. Refer to the table below for permitted forces.
Please refer to the table in the technical data section of this Handbook for the maximum end loadings.
3.4.8 Pipework reducers
Always avoid extreme step changes in pipe size. Use pipework reducers, where there is a large difference between pipework size and meter flanges.
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
19
3 INSTALLATION
3.4.9 Flexible connections
Flexible connections can be used but because of the high flow rates associated with large diameter meters, it is recommended that flexible connections are not used on meters larger than size 80.
3.4.10 Hygienic installations
OPTIMASS 8000K
Mounting the meter for hygienic applications
1 Install the meter vertically to allow self draining. 2 DO NOT install the meter horizontally.
Where the meter has been approved by the sanitary requirements of the European Hygienic Engineering and Design Group, you MUST give consideration to:
Installation - install the meter at an angle to allow self-draining (see illustration).
Cleaning fluids - cleaning fluids should flow uphill with a velocity rate greater than
1.5 m/s / 5ft/s. If the process flow is downhill, install a flow restrictor downstream of the meter. This will make sure that the meter is completely filled with the cleaning fluid.
Process connections and seals MUST be in accordance with EHEDG documentation.
20
The manufacturer also recommends that you refer to EHEDG (www.ehedg.org) document number 8 "HYGIENIC EQUIPMENT DESIGN CRITERIA".
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
3.4.11 Heating and insulation
Insulation
Insulation
InsulationInsulation
The meter can be insulated to a maximum depth as shown (A). Do not insulate above this depth as this will cause the electronics to overheat.
Insulation
INSTALLATION 3
S15 S25 S40 S80 S100
Dimension A [mm] 75 75 75 75 75
Dimension A [inches] 2.9 2.9 2.9 2.9 2.9
Factory fitted heating jacket
Factory fitted heating jacket
Factory fitted heating jacketFactory fitted heating jacket
If the meter has been ordered with a heating jacket, it will be supplied with DN15 PN40 to DIN
2501 or 1/2¨ ASME150 flange connections.
Connecting / using the heating jacket
Use rigid or reinforced flexible hoses to connect the heating jacket to the heat source.
The heating jacket material is 304.
Suitable heating mediums are steam or hot oil. Avoid the use of heating mediums that can
cause crevice corrosion in Stainless Steel.
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
21
3 INSTALLATION
Where liquid is being used, set up the pipe configuration so that air can be vented from the system.
Where steam is being used, set up the pipe configuration so that condensation can be drained off.
Heat the jacket to working temperature before flowing the process fluid through the meter.
CAUTION!
The design of the heating jacket is such that NO end load pressure (negative or positive) can be applied. The manufacturer recommends that (where possible) flexible hoses are used to connect to the heat source.
CAUTION!
The maximum heating pressure and temperature for heating jackets is 10 barg at
°
C / 145 psig at 446°F
230
3.4.12 Purge ports
If the meter has been ordered with a purge port, it will be supplied with NPT female connections which are clearly marked. The connections are sealed with NPT plugs and PTFE tape.
OPTIMASS 8000K
CAUTION!
DO NOT remove these plugs.
The meter is factory sealed with a dry nitrogen gas fill and if moisture is allowed to enter the meter casing it will cause damage. The plugs should only be removed to purge the meter casing in the event that the primary measuring tube fails.
If it is suspected that the primary measuring tube has failed, de-pressurise the meter and remove it from service, as soon as it is safe to do so.
3.4.13 Burst dics
If the meter has been ordered with a burst disc, it will be supplied with the disc fitted. The failure
pressure of the disc is 4 barg at +20°C / 58 psig at +68°F.
CAUTION!
The fitted burst disc will be suitable for the flow rates and process conditions specified on the original order. If the process conditions alter in any way, it is recommended that you contact the manufacturer for advice regarding suitability.
If the process product is hazardous (in any way) it is recommended that an exhaust tube is connected to the NPT male thread of the burst disc and the pipe routed so that the process product can be discharged to a safe area. Use a tube with a diameter large enough AND routed in such a way, so that pressure cannot build up in the meter case.
22
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
3.4.14 Zero calibration
The procedure for zero calibration is contained in the converter handbook. However, the following information should be considered when installing the meter.
Zero calibration
INSTALLATION 3
1 Where the meter has been installed vertically, install shut-off valves either side of the meter to assist with zero cali-
bration.
2 If the process flow cannot be stopped, install a bypass section for zero calibration.
3.4.15 Sunshades
The meter MUST be protected from strong sunlight.
1 Horizontal installation 2 Vertical installation
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
23
4 ELECTRICAL CONNECTIONS
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
OPTIMASS 8000K
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
4.2 Electrical and I/O connections
For information regarding electrical and I/O connections, please refer to the handbook for the relevant signal converter.
24
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
5.1 Spare parts availability
The manufacturer adheres to the basic principle that functionally adequate spare parts for each device or each important accessory part will be kept available for a period of 3 years after delivery of the last production run for the device.
This regulation only applies to spare parts which are under normal operating conditions subjects to wear and tear.
5.2 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, technical support and training.
INFORMATION!
For more precise information, please contact your local representative.
5.3 Returning the device to the manufacturer
SERVICE 5
5.3.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:
Due to statutory regulations on environmental protection and safeguarding the health and safety of our personnel, manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.
This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle.
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products, you are kindly requested:
to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from such dangerous substances,
to enclose a certificate with the device confirming that is safe to handle and stating the product used.
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
25
5 SERVICE
5.3.2 Form (for copying) to accompany a returned device
Company: Address:
Department: Name:
Tel. no.: Fax no.:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
OPTIMASS 8000K
This medium is: water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such substances.
We have flushed out and neutralized all cavities in the device.
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned.
Date: Signature:
Stamp:
5.4 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
26
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
6.1 Measuring principle (twin tube)
Meter from the side, showing tube layout
TECHNICAL DATA 6
1 Measuring tubes 2 Drive coil 3 Sensor 1 4 Sensor 2
Static meter not energised and with no flow
1 Measuring tubes 2 Drive coil 3 Sensor 1 4 Sensor 2
A Coriolis twin tube mass flowmeter consists of two measuring tubes 1 a drive coil 2 and two sensors (3 and 4) that are positioned either side of the drive coil.
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
27
6 TECHNICAL DATA
Energised meter
1 Measuring tubes 2 Direction of oscilation 3 Sine wave
When the meter is energised, the drive coil vibrates the measuring tubes causing them to oscillate and produce a sine wave 3. The sine wave is monitored by the two sensors.
Energised meter with process flow
OPTIMASS 8000K
1 Process flow 2 Sine wave 3 Phase shift
When a fluid or gas passes through the tubes, the coriolis effect causes a phase shift in the sine wave that is detected by the two sensors. This phase shift is directly proportional to the mass flow.
Density measurement is made by evaluation of the frequency of vibration and temperature measurement is made using a Pt500 sensor.
28
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
TECHNICAL DATA 6
6.2 Technical data
INFORMATION!
The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local representative.
Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle Coriolis mass flow
Application range Mass flow and density measurement of fluids, gases and solids
Measured values Mass, density, temperature
Calculated values Volume, referred density, concentration, velocity
Design
Basic System consists of a measuring sensor and a converter to process the
Features Fully welded maintenance free sensor with twin U-shaped measuring tube
Variants
Variants
VariantsVariants
Compact version Integral converter
Remote version Available with field, wall or 19" rack mount versions of the converter
Modbus version Sensor with integral electronics providing Modbus output for connection to a
output signal
PLC
Measuring accuracy
Mass
Mass
MassMass Liquid ±0.1% of actual measured flow rate + zero stability Gas ±0.5% of actual measured flow rate + zero stability
Repeatability Better than 0.05% plus zero stability (includes the combined effects of
Zero stability
Zero stability
Zero stabilityZero stability Stainless Steel ±0.004% of maximum flow rate with respective sensor size
Reference conditions
Reference conditions
Reference conditionsReference conditions
Product Water Temperature +20°C / +68°F
Operating pressure 1 barg / 14.5 psig
Effect on sensor zero point caused by a shift in process temperature
Effect on sensor zero point caused by a shift in process temperature
Effect on sensor zero point caused by a shift in process temperatureEffect on sensor zero point caused by a shift in process temperature Stainless Steel 0.001% of max flow per 1°C / 0.0005% per 1°F
Pressure effect on mass flow rate
Pressure effect on mass flow rate
Pressure effect on mass flow ratePressure effect on mass flow rate
Stainless Steel -0.015% per 1 barg / 0.001% per 1 psig
Density
Density
DensityDensity
Measuring range
Accuracy
On site calibration
Temperature
Temperature
TemperatureTemperature Accuracy ±1°C / ±1.8°F
repeatability, linearity and hysteresis)
400...3000 kg/m3 / 25...187 lbs/ft
±2 kg/m3 / ±0.13 lbs/ft ±0.5 kg/m3 / ±0.033 lbs/ft
3
3
3
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
29
6 TECHNICAL DATA
Operating conditions
Maximum flow rates
Maximum flow rates
Maximum flow ratesMaximum flow rates
S15 3510 kg/h / 128.7 lbs/min
S25 11700 kg/h / 429 lbs/min
S40 41600 kg/h / 1525.3 lbs/min
S80 110500 kg/h / 4051.6 lbs/min
S100 325000 kg/h / 11916.6 lbs/min
Assumes operating density 1000 kg/m3 / 62.4 lb/ft
Ambient temperature
Ambient temperature
Ambient temperatureAmbient temperature
Compact version with Aluminium converter
Compact version with Stainless Steel converter
Remote versions Standard temperature range: -40...+65°C / -40…+149°F
Hazardous Area versions Refer to temperature limits
Process temperature
Process temperature
Process temperatureProcess temperature
Standard temperature range
Standard temperature range
Standard temperature rangeStandard temperature range Safe area (compact and remote version) -70+230°C / -94+440°F
Hazardous area (standard temperature, compact version only)
Hazardous area (standard temperature, remote version only)
Hygienic connections -70+150°C / -94+302°F
Cryogenic temperature range
Cryogenic temperature range
Cryogenic temperature rangeCryogenic temperature range Safe area -195+40°C / -310+104°F Hazardous area -195+40°C / -310+104°F
Nominal pressure at 20
Nominal pressure at 20°C / 68
Nominal pressure at 20Nominal pressure at 20
Measuring tube
Measuring tube
Measuring tubeMeasuring tube PED 97/23/EC -1100 barg / -14.51450 psig
FM Pending
CRN / ASME B31.3 Pending
Fluid properties
Fluid properties
Fluid propertiesFluid properties
Permissible physical condition Liquids, gases, slurries
Permissible gas content (volume) Contact manufacturer for information.
Permissible solid content (volume) Contact manufacturer for information.
Protection category (acc. to EN 60529) IP 67, NEMA 4X
Installation conditions
Inlet runs None required
Outlet runs None required
C / 68°F
C / 68C / 68
F
FF
Standard temperature range: -40...+60°C / -40…+140°F Cryogenic temperature range: -25...+40°C / -13…+104°F Standard temperature range: -40...+55°C / -40…+131°F Cryogenic temperature range: -25...+40°C / -13…+104°F
Cryogenic temperature range: -20...+65°C / -4…+149°F
-40+190°C / -40+370°F
-40+230°C / -40+440°F
3
OPTIMASS 8000K
30
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
TECHNICAL DATA 6
Materials
Measuring tube Stainless Steel AISI 316 / 316L (1.4401 / 1.4404) dual certified
Spigot Stainless Steel AISI 316 / 316L (CF3M / 1.4409) dual certified
Flanges Stainless Steel AISI 316 / 316L (1.4401 / 1.4404) dual certified
Outer casing Stainless Steel AISI 304 (1.4301)
Heating jacket version
Heating jacket version
Heating jacket versionHeating jacket version
Heating tubes and insulation jacket Stainless Steel AISI 304 (1.4301)
All versions
All versions
All versionsAll versions
Sensor electronics housing Stainless Steel 316L (1.4409)
Junction box (remote version) Die cast Aluminium (polyurethane coating)
Optional Stainless Steel 316 (1.4401)
Process connections
Flange
Flange
FlangeFlange DIN DN15150 / PN40100 ASME ½...6" / ASME 150600
JIS 15A...100A / 10...20K
Hygienic (S100 only)
Hygienic (S100 only)
Hygienic (S100 only)Hygienic (S100 only)
Tri-clover 1...4"
Tri-clamp DIN 32676 DN25..100
Tri-clamp ISO 2852 1...4"
DIN 11864-2 Form A DN25...100
Male thread DIN 11851 DN25...100
Male thread SMS 1...4"
Male thread IDF / ISS 1...4"
Male thread RJT 1...4"
Electrical connections
Electrical connections For full details, including: power supply, power consumption etc., see
I/O For full details of I/O options, including data streams and protocols, see
technical data for the relevant converter.
technical data for the relevant converter.
Approvals
Mechanical
Mechanical
MechanicalMechanical
Electromagnetic compatibility (EMC) acc. to CE
European Pressure Equipment Directive PED 97-23 EC (acc. to EN13445-3)
ASME B31.3
ATEX (acc. 94/9/EC)
ATEX (acc. 94/9/EC)
ATEX (acc. 94/9/EC)ATEX (acc. 94/9/EC)
OPTIMASS 8300k C non Ex i Signal outputs
OPTIMASS 8300k C non Ex i Signal outputs
OPTIMASS 8300k C non Ex i Signal outputsOPTIMASS 8300k C non Ex i Signal outputs
Ex d connection compartment II 1/2 G - Ex d [ib] IIC T4....T1 Ga/Gb
Namur NE 21/5.95
2004/108/EC (EMC)
2006/95/EC (Low Voltage Directive)
II 2 D - Ex t IIIC Txx°C Db
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
31
6 TECHNICAL DATA
Ex e connection compartment II 1/2 G - Ex de [ib] IIC T4....T1 Ga/Gb
II 2 D - Ex t IIIC Txx°C Db
OPTIMASS 8300k C Ex i signal outputs
OPTIMASS 8300k C Ex i signal outputs
OPTIMASS 8300k C Ex i signal outputsOPTIMASS 8300k C Ex i signal outputs
Ex d connection compartment II 1/2 (1) G - Ex d [ia/ib] IIC T4....T1 Ga/Gb
II 2 (1) D - Ex t [ia Da] IIIC Txx°C Db
Ex e connection compartment II 1/2 (1) G - Ex de [ia/ib] IIC T4....T1 Ga/Gb
II 2 (1) D - Ex t [ia Da] IIIC Txx°C Db
OPTIMASS 8000k / 8010k C
OPTIMASS 8000k / 8010k C II 1/2 G - Ex ib IIC T4T1 Ga/Gb
OPTIMASS 8000k / 8010k COPTIMASS 8000k / 8010k C
II 2 D - Ex t IIIC Txx°C Db
ATEX (acc. 94/9/EC) temperature limits
OPTIMASS 8000K
OPTIMASS 8000k / 8010k C with or without heating jacket / insulation. Minimum process temperature T
1
m
OPTIMASS 8000k / 8010k C with or
-40ºC
Ambient temp. T
°C
amb
-40...+65 80 T4 T130
-20...+65 -195...80 T4-T1 T130
Max. medium temp. T
140 T3 T195
230 T2-T1 T280
without heating jacket / insulation, Cryogenic Applications 1
OPTIMASS 8300k C Aluminium converter housing - with or without heating jacket / insulation. Minimum process temperature T
-40ºC 1
m
-40...+40 60 T4 T125
120 T3 T190
190 T2-T1 T265
-40...+50 120 T3 T190
190 T2-T1 T260
-40...+55 55 T4-T1 T125
-40...+60 2 60 T4-T1 T130
OPTIMASS 8300k C Stainless Steel converter housing - with or without heating jacket / insulation. Minimum process temperature T
-40ºC 1
m
-40...+40 60 T4 T125
120 T3 T195
190 T2-T1 T265
-40...+45 55 T4 T125
190 T2-T1 T260
-40...+50 2 50 T4-T1 T120
OPTIMASS 8300k C Aluminium or
-25...+40 -195...60 T4-T1 T125 Stainless Steel converter housing with or without heating jacket / insulation, Cryogenic Applications 1
1 For sensors with additional paint coating, please refer to manufacturer 2 depending on I/O option. Please call for more information.
°C
m
temp. °C
Temp. class Max. surface
32
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
TECHNICAL DATA 6
Maximum end loadings
S15 S25 S40 S80 S100
Flanges
Flanges
FlangesFlanges 20°C 40 barg 25 kN 38 kN 48 kN 99 kN 150 kN
100 barg 17 kN 19 kN 15 kN 20 kN 100 kN
230°C 32 barg 12 kN 18 kN 23 kN 45 kN 100 kN
52 barg 10 kN 15 kN 60 kN
Hygienic (all connections)
Hygienic (all connections)
Hygienic (all connections)Hygienic (all connections) 130°C 10 barg 5 kN 9 kN 12 kN 12 kN 18 kN
These (axial) loads have been calculated, based on 316L schedule 40 process pipework, where un-radiographed butt welds have been used in pipe joints.
The loads shown are the maximum permitted static load. If loads are cycling (between tension and compression) these loads should be reduced. For advice, consult the manufacturer.
6.3 Measuring accuracy
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
0
Measuring error
The measuring error is obtained from the combined effects of accuracy and zero stability.
Reference conditions
Product Water Temperature +20°C / +68°F
Operating pressure 1 barg / 14.5 psig
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
33
6 TECHNICAL DATA
6.4 Guidelines for maximum operating pressure
Notes:
Ensure that the meter is used within its operating limits
All hygienic process connections have a maximum operating rating of 10 barg
at 130°C / 145 psig at 266°F
Pressure / temperature de-rating, all meter sizes in metric (flanged connections as
Pressure / temperature de-rating, all meter sizes in metric (flanged connections as
Pressure / temperature de-rating, all meter sizes in metric (flanged connections as Pressure / temperature de-rating, all meter sizes in metric (flanged connections as per EN 1092-1)
per EN 1092-1)
per EN 1092-1)per EN 1092-1)
OPTIMASS 8000K
X temperature [°C] Y pressure [barg]
1 Measuring tube PED certification 2 Hygienic connection
34
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
Pressure / temperature de-rating, all meter sizes, in imperial (flanged connections as
Pressure / temperature de-rating, all meter sizes, in imperial (flanged connections as
Pressure / temperature de-rating, all meter sizes, in imperial (flanged connections as Pressure / temperature de-rating, all meter sizes, in imperial (flanged connections as per ASME B16.5)
per ASME B16.5)
per ASME B16.5)per ASME B16.5)
TECHNICAL DATA 6
X temperature [°F] Y pressure [psig]
1 Measuring Tube PED certification 2 Hygienic connection
Flanges
DIN flange ratings are based on EN 1092-1 2007 table G.4.1 material group 14EO
ASME flange ratings are based on ASME B16.5 2003 table 2 material group 2.2
JIS flange ratings are based on JIS 2220: 2001 table 1 division 1 material group 022a
Notes
The maximum operating pressure will be either the flange rating or the measuring tube rating, WHICHEVER IS THE LOWER!
The manufacturer recommends that the seals are replaced at regular intervals. This will maintain the hygienic integrity of the connection.
WHICHEVER IS THE LOWER!
WHICHEVER IS THE LOWER!WHICHEVER IS THE LOWER!
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
35
6 TECHNICAL DATA
6.5 Dimensions and weights
6.5.1 Flanged versions
OPTIMASS 8000K
1 Compact version 2 Remote version
Meter weights (all flanges)
Weight [kg]
S15 S25 S40 S80 S100
Aluminium (compact) 13.8 22.3 30.8 62.3 103.8
Stainless Steel (compact) 19.2 28.4 36.9 68.4 109.9
Aluminium (remote) 11 19.5 28 59.5 101
Stainless Steel (remote) 11.8 20.3 28.8 60.3 101.8
Heating jacket add 7.5 10 11.5 16 20
Weight [lb]
S15 S25 S40 S80 S100
Aluminium (compact) 30.4 49 67.8 137 228.4
Stainless Steel (compact) 42.2 62.5 81.2 150.5 241.8
Aluminium (remote) 24.2 42.9 61.6 130.9 222.2
Stainless Steel (remote) 26 44.7 63.4 132.7 224
Heating jacket add 16.5 22 25.3 35.2 44
36
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
TECHNICAL DATA 6
Measuring tube in Stainless Steel
Dimensions [mm]
S15 S25 S40 S80 S100
A 185 280 320 415 450
C1 (compact) 359 374 380 407 433
C2 (remote) 282 297 303 330 356
D 160
E 60
F 137
G 123.5
H 98.5
I 73 102 114 168 220
Dimensions [inches]
S15 S25 S40 S80 S100
A 7.3 11 12.6 16.3 17.7
C1 (compact) 14.1 14.7 15 16 17
C2 (remote) 11.1 11.7 11.9 13 14
D 6.3
E 2.4
F 5.4
G 4.9
H 3.9
I 2.9 4 4.5 6.6 8.7
Flange connections
Dimension B [mm]
S15 S25 S40 S80 S100
PN40
PN40
PN40PN40
DN15 510 - - - -
DN25 512 600 - - -
DN40 - 608 700 -
DN50 - - 715 893 -
DN80 - - - 915 984
DN100 - - - - 998
DN150 - - - - 1018
PN63
PN63
PN63PN63
DN50 - - 741 921 -
DN80 - - - 943 1012
DN100 - - - - 1024
DN150 - - - - 1058
PN100
PN100
PN100PN100
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
37
6 TECHNICAL DATA
OPTIMASS 8000K
Dimension B [mm]
S15 S25 S40 S80 S100
DN15 522 - - - -
DN25 548 634 - - -
DN40 - 642 741 - -
DN50 - - 753 933 -
DN80 - - - 953 1024
DN100 - - - - 1048
DN150 - - - - 1098
ASME 150
ASME 150
ASME 150ASME 150
½¨ 528 - - - - ¾¨ 538 - - - -
1¨ 544 518 - - ­1½¨ - 642 741 - ­2¨ - - 745 925 ­3¨ - - - 937 1008 4¨ - - - - 1022 6¨ - - - - 1046
ASME 300
ASME 300
ASME 300ASME 300
½¨ 538 - - - - ¾¨ 548 - - - -
1¨ 556 642 -- - ­1½¨ - 656 755 - ­2¨ - - 757 937 ­3¨ - - - 957 1028 4¨ - - - - 1040 6¨ - - - - 1066
ASME 600
ASME 600
ASME 600ASME 600
½¨ 550 ¾¨ 560 - - - -
1¨ 568 656 - - ­1½¨ - 670 771 - ­2¨ - - 775 957 ­3¨ - - - 975 1046 4¨ - - - - 1086 6¨ - - - - 1116
JIS 10K
JIS 10K
JIS 10KJIS 10K
50A - - 706 893 -
80A - - - 913 -
JIS 20K
JIS 20K
JIS 20KJIS 20K
15A 508 - - - -
25A 512 598 - - -
40A - 608 707 - -
38
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
TECHNICAL DATA 6
Dimension B [mm]
S15 S25 S40 S80 S100
50A - - 713 893 -
80A - - - 913 -
100A - - - - 1020
Dimension B [inches]
S15 S25 S40 S80 S100
PN40
PN40
PN40PN40
DN15 20 - - - -
DN25 20.2 23.5 - - -
DN40 - 24 27.8 -
DN50 - - 28 35.2 -
DN80 - - - 36 38.7
DN100 - - - - 39.3
DN150 - - - - 40.2
PN63
PN63
PN63PN63
DN50 - - 29 -
DN80 - - - 37 39.8
DN100 - - - - 40.3
DN150 - - - - 41.6
PN100
PN100
PN100PN100
DN15 20.5 - - - -
DN25 21.6 25 - - -
DN40 - 25.3 28.9 - -
DN50 - - 29.4 36.7 -
DN80 - - - 37.5 40.3
DN100 - - - - 41.3
DN150 - - - - 43.2
ASME 150
ASME 150
ASME 150ASME 150
½¨ 20.8 - - - - ¾¨ 21.2 - - - -
1¨ 21.4 20.4 - - ­1½¨ - 25.3 29.2 - ­2¨ - - 29.3 36.4 ­3¨ - - - 36.9 39.7 4¨ - - - - 40.2 6¨ - - - - 41.3
ASME 300
ASME 300
ASME 300ASME 300
½¨ 21.2 - - - - ¾¨ 21.6 - - - -
1¨ 21.9 25.3 -- - -
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
39
6 TECHNICAL DATA
OPTIMASS 8000K
Dimension B [inches]
S15 S25 S40 S80 S100
1½¨ - 25.8 29.7 - ­2¨ - - 29.8 36.9 ­3¨ - - - 37.7 40.5 4¨ - - - - 41 6¨ - - - - 42
ASME 600
ASME 600
ASME 600ASME 600
½ 21.6 - - - - ¾¨ 22 - - - -
1¨ 22.4 25.8 - - ­1½¨ - 26.4 30.4 - ­2¨ - - 30.5 37.7 ­3¨ - - - 38.4 41.2 4¨ - - - - 42.8 6¨ - - - - 44
JIS 10K
JIS 10K
JIS 10KJIS 10K
50A - - 28 35.2 -
80A - - - 35.9 -
JIS 20K
JIS 20K
JIS 20KJIS 20K
15A 20 - - - -
25A 20.2 23.5 - - -
40A - 23.9 27.8 - -
50A - - 28 35.1 -
80A - - - 35.9 -
100A - - - - 40.2
6.5.2 NAMUR dimensions
The following face to face dimensions comply with NAMUR NE132
PN40
PN40 Dimensions [mm +0.0 / -5.0]
PN40PN40
S15 S25 S40 S80 S100
DN 15 510 - - - -
DN 25 - 600 - - -
DN 50 - - 715 - -
DN 80 - - - 915 -
Dimensions [inches +0.0 / -0.12]
DN 15 20.1 - - - -
DN 25 - 23.6 - - -
DN 50 - - 28.1 - -
DN 80 - - - 36 -
40
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
6.5.3 Hygienic versions
Hygienic connections: all welded versions
TECHNICAL DATA 6
Dimension B [mm]
S15 S25 S40 S80 S100
Tri-clover
Tri-clover
Tri-cloverTri-clover 1¨ 485 - - - ­1½¨ - 580 - - ­2¨ - - 675 - ­3¨ - - - 850 ­4¨ - - - - 911
Tri-clamp DIN 32676
Tri-clamp DIN 32676
Tri-clamp DIN 32676Tri-clamp DIN 32676
DN25 475 - - - -
DN40 - 570 - - -
DN50 - - 668 - -
DN80 - - - 859 -
DN100 - - - - 924
Tri-clamp ISO 2852
Tri-clamp ISO 2852
Tri-clamp ISO 2852Tri-clamp ISO 2852 1¨ 481 - - - ­1½¨ - 586 - - ­2¨ - - 666 - ­3¨ - - - 846 ­4¨ - - - - 911
DIN 11864-2 Form A (Female)
DIN 11864-2 Form A (Female)
DIN 11864-2 Form A (Female)DIN 11864-2 Form A (Female)
DN25 512 - - - -
DN40 - 617 - - -
DN50 - - 715 - -
DN80 - - - 919 -
DN100 - - - - 984
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
41
6 TECHNICAL DATA
OPTIMASS 8000K
Dimension B [inches]
S15 S25 S40 S80 S100
Tri-clover
Tri-clover
Tri-cloverTri-clover 1¨ 19 - - - ­1½¨ - 23 - - ­2¨ - - 26.6 - ­3¨ - - - 33.5 ­4¨ - - - - 36
Tri-clamp DIN 32676
Tri-clamp DIN 32676
Tri-clamp DIN 32676Tri-clamp DIN 32676
DN25 19 - - - -
DN40 - 22.5 - - -
DN50 - - 26.3 - -
DN80 - - - 34 -
DN100 - - - - 36.4
Tri-clamp ISO 2852
Tri-clamp ISO 2852
Tri-clamp ISO 2852Tri-clamp ISO 2852 1¨ 19 - - - ­1½¨ - 23 - - ­2¨ - - 26.2 - ­3¨ - - - 33.3 ­4¨ 36
DIN 11864-2 Form A (Female)
DIN 11864-2 Form A (Female)
DIN 11864-2 Form A (Female)DIN 11864-2 Form A (Female)
DN25 20 - - - -
DN40 - 24.3 - - -
DN50 - - 28.2 - -
DN80 - - - 36 -
DN100 - - - - 38.7
42
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
Hygienic connections: adapter versions (male thread)
S15 S25 S40 S80 S100
TECHNICAL DATA 6
Dimension B [mm]
Male thread DIN 11851
Male thread DIN 11851
Male thread DIN 11851Male thread DIN 11851
DN25 490 - - - -
DN40 - 593 - - -
DN50 - - 695 - -
DN80 - - - 893 -
DN100 - - - - 976
Male thread SMS
Male thread SMS
Male thread SMSMale thread SMS 1¨ 472 - - - ­1½¨ - 583 - - ­2¨ - - 678 - ­3¨ - - - 855 ­4¨ - - - - 924
Male thread IDF/ISS
Male thread IDF/ISS
Male thread IDF/ISSMale thread IDF/ISS 1¨ 485 - - - ­1½¨ - 580 - - ­2¨ - - 675 - ­3¨ - - - 850 ­4¨ - - - - 911
Male thread RJT
Male thread RJT
Male thread RJTMale thread RJT 1¨ 496 - - - ­1½¨ - 591 - - ­2¨ - - 686 - ­3¨ - - - 861 ­4¨ - - - - 922
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
43
6 TECHNICAL DATA
OPTIMASS 8000K
Dimension B [inches]
S15 S25 S40 S80 S100
Male thread DIN 11851
Male thread DIN 11851
Male thread DIN 11851Male thread DIN 11851
DN25 19.3 - - - -
DN40 - 23.3 - - -
DN50 - - 27.4 - -
DN80 - - - 35 -
DN100 - - - - 38.4
Male thread SMS
Male thread SMS
Male thread SMSMale thread SMS 1¨ 18.6 - - - ­1½¨ - 23 - - ­2¨ - - 26.7 - ­3¨ - - - 33.7 ­4¨ - - - - 36.4
Male thread IDF/ISS
Male thread IDF/ISS
Male thread IDF/ISSMale thread IDF/ISS 1¨ 19 - - - ­1½¨ - 22.8 - - ­2¨ - - 26.6 - ­3¨ - - - 33.5 ­4¨ - - - - 35.9
Male thread RJT
Male thread RJT
Male thread RJTMale thread RJT 1¨ 19.5 - - - ­1½¨ - 23.3 - - ­2¨ - - 27 - ­3¨ - - - 33.4 ­4¨ - - - - 36.3
44
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
6.5.4 Heating jacket version
TECHNICAL DATA 6
Dimensions [mm]
S15 S25 S40 S80 S100
Heating connection size PN40 DN25 or ASME 150 ½¨
A 590 ±5.0 692 ±5.0 715 ±5.0 891 ±5.0 956 ±5.0 B 440 ±3.0 542 ±3.0 565 ±3.0 741 ±3.0 806 ±3.0 C 130 ±3.0 210 ±3.0 230 ±3.0 320 ±3.0 340 ±3.0 D 344 ±3.0 453 ±3.0 499 ±3.0 622 ±3.0 682 ±3.0 E 221 ±3.0 316 ±3.0 356 ±3.0 451 ±3.0 486 ±3.0 F 226 ±3.0 254 ±3.0 266 ±3.0 322 ±3.0 372 ±3.0
Dimensions [inches]
S15 S25 S40 S80 S100
Heating connection size PN40 DN25 or ASME 150 ½¨
A 23.2±0.2 27.2 ±0.2 28 ±0.2 891 ±0.2 37.6 ±0.2 B 17.3 ±0.12 21.3 ±0.12 22.2 ±0.12 29 ±0.12 31.7 ±.012 C 5 ±0.12 8.7 ±0.12 9 ±0.12 12.6 ±0.12 13.4 ±0.12 D 13.5 ±0.12 17.8 ±0.12 19.6 ±0.12 24.5 ±0.12 26.9 ±0.12 E 8.7 ±0.12 12.4 ±0.12 14 ±0.12 17.7 ±0.12 19.1 ±0.12 F 8.9 ±0.12 10 ±0.12 10.5 ±0.12 12.7 ±0.12 14.6 ±0.12
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
45
6 TECHNICAL DATA
6.5.5 Purge port option
OPTIMASS 8000K
Dimensions [mm]
S15 S25 S40 S80 S100
A 150 200 215 300 305
B 300 400 430 600 610
Dimensions [inches]
S15 S25 S40 S80 S100
A 5.9 7.9 8.5 11.8 12
B 11.8 15.7 17 23.6 24
46
www.krohne.com 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
OPTIMASS 8000K
NOTES 7
www.krohne.com08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en
47
KROHNE product overview
Electromagnetic flowmeters
Variable area flowmeters
Ultrasonic flowmeters
Mass flowmeters
Vortex flowmeters
Flow controllers
Level meters
Temperature meters
Pressure meters
Analysis products
Measuring systems for the oil and gas industry
Measuring systems for sea-going tankers
Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 D-47058 Duisburg (Germany) Tel.:+49 (0)203 301 0 Fax:+49 (0)203 301 10389 info@krohne.de
© KROHNE 08/2010 - 4000921002 - MA OPTIMASS 8000k R02 en - Subject to change without notice.
The current list of all KROHNE contacts and addresses can be found at: www.krohne.com
Loading...