KROHNE MFC-300 User Manual

Page 1
Handbook
Handbook
MFC 300
MFC 300
MFC 300MFC 300
HandbookHandbook
Signal converter for mass flowmeters
Electronic revision: ER 3.3.xx (SW.REV. 3.4x)
The documentation is only complete when used in combination with the relevant documentation for the measuring sensor.
Page 2
: IMPRINT :::::::::::::::::::::::::::::::::::::::
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH.
Subject to change without notice.
Copyright 2012 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany)
2
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 3
MFC 300
CONTENTS
1 Safety instructions 7
1.1 Software History ............................................................................................................... 7
1.2 Intended Use..................................................................................................................... 9
1.3 Certifications .................................................................................................................... 9
1.4 Safety instructions from the manufacturer ................................................................... 10
1.4.1 Copyright and data protection .............................................................................................. 10
1.4.2 Disclaimer ............................................................................................................................. 10
1.4.3 Product liability and warranty .............................................................................................. 11
1.4.4 Information concerning the documentation......................................................................... 11
1.4.5 Warnings and symbols used................................................................................................. 12
1.5 Safety instructions for the operator............................................................................... 12
2 Device description 13
2.1 Scope of delivery............................................................................................................. 13
2.2 Device description .......................................................................................................... 14
2.2.1 Field housing......................................................................................................................... 15
2.2.2 Wall-mounted housing ......................................................................................................... 16
2.3 Nameplates .................................................................................................................... 17
2.3.1 Compact version (example) .................................................................................................. 17
2.3.2 Remote version (example) .................................................................................................... 18
2.3.3 Electrical connection data of inputs/outputs (example of basic version)............................ 19
3 Installation 20
3.1 Notes on installation ......................................................................................................20
3.2 Storage ........................................................................................................................... 20
3.3 Transport ........................................................................................................................ 20
3.4 Installation specifications .............................................................................................. 20
3.5 Mounting of the compact version................................................................................... 21
3.6 Mounting the field housing, remote version .................................................................. 21
3.6.1 Pipe mounting ....................................................................................................................... 21
3.6.2 Wall mounting....................................................................................................................... 22
3.6.3 Turning the display of the field housing version .................................................................. 23
3.7 Mounting the wall-mounted housing, remote version .................................................. 24
3.7.1 Pipe mounting ....................................................................................................................... 24
3.7.2 Wall mounting....................................................................................................................... 25
4 Electrical connections 26
4.1 Safety instructions.......................................................................................................... 26
4.2 Important notes on electrical connection...................................................................... 26
4.3 Requirements for signal cables provided by the customer........................................... 27
4.4 Connecting the signal cables ......................................................................................... 28
4.4.1 Connection of signal cable, field housing............................................................................. 29
4.4.2 Connection of signal cable, wall-mounted housing............................................................. 30
4.4.3 Connection of signal cable, 19" rack-mounted housing ...................................................... 31
4.4.4 Connection box of measuring sensor................................................................................... 32
4.4.5 Connection diagram.............................................................................................................. 33
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
3
Page 4
CONTENTS
4.5 Grounding the measuring sensor .................................................................................. 34
4.6 Connecting power, all housing variants......................................................................... 35
4.7 Inputs and outputs, overview ......................................................................................... 37
4.7.1 Combinations of the inputs/outputs (I/Os) ........................................................................... 37
4.7.2 Description of the CG number .............................................................................................. 38
4.7.3 Fixed, non-alterable input/output versions.......................................................................... 39
4.7.4 Alterable input/output versions............................................................................................ 41
4.8 Description of the inputs and outputs............................................................................ 42
4.8.1 Current output ...................................................................................................................... 42
4.8.2 Pulse and frequency output.................................................................................................. 43
4.8.3 Status output and limit switch .............................................................................................. 44
4.8.4 Control input ......................................................................................................................... 45
4.9 Electrical connection of the inputs and outputs ............................................................ 46
4.9.1 Field housing, electrical connection of the inputs and outputs........................................... 46
4.9.2 Wall-mounted housing, electrical connection of the inputs and outputs............................ 47
4.9.3 19" rack-mounted housing (28 TE), electrical connection of the inputs and outputs ......... 48
4.9.4 Laying electrical cables correctly......................................................................................... 48
4.10 Connection diagrams of inputs and outputs................................................................ 49
4.10.1 Important notes................................................................................................................... 49
4.10.2 Description of the electrical symbols................................................................................. 50
4.10.3 Basic inputs/outputs ........................................................................................................... 51
4.10.4 Modular inputs/outputs and bus systems.......................................................................... 54
4.10.5 Ex i inputs/outputs .............................................................................................................. 62
4.10.6 HART
MFC 300
®
connection .............................................................................................................. 66
5 Start-up 68
5.1 Switching on the power .................................................................................................. 68
5.2 Starting the signal converter ......................................................................................... 68
6 Operation 69
6.1 Display and operating elements .................................................................................... 69
6.1.1 Display in measuring mode with 2 or 3 measured values ................................................... 71
6.1.2 Display for selection of sub-menu and functions, 3 lines.................................................... 71
6.1.3 Display when setting parameters, 4 lines ............................................................................ 72
6.1.4 Display when changing parameters, 4 lines ........................................................................ 72
6.1.5 Using an IR interface (option) ............................................................................................... 73
6.2 Zero calibration (menu C1.1.1)....................................................................................... 74
6.3 Menu structure............................................................................................................... 76
6.4 Function tables ............................................................................................................... 79
6.4.1 Menu A, Quick Setup............................................................................................................. 79
6.4.2 Menu B, Test ......................................................................................................................... 81
6.4.3 Menu C, Setup ....................................................................................................................... 82
6.4.4 Set free units......................................................................................................................... 96
6.5 Description of functions ................................................................................................. 97
6.5.1 Reset counter in the menu "quick setup" ............................................................................ 97
6.5.2 Deleting error messages in the menu "quick setup"........................................................... 97
6.5.3 Mode (menu A8) .................................................................................................................... 98
6.5.4 Density calibration (menu C1.2.1)......................................................................................... 99
6.5.5 Temperature/density tables ............................................................................................... 102
6.5.6 Density mode (menu C1.2.2)............................................................................................... 105
4
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 5
MFC 300
CONTENTS
6.5.7 Pipe diameter (menu C1.1.3) .............................................................................................. 106
6.5.8 Concentration measurement (menu C2)............................................................................ 106
6.5.9 Flow direction (menu C1.3.1).............................................................................................. 106
6.5.10 Pressure suppression....................................................................................................... 106
6.5.11 Process control ................................................................................................................. 108
6.5.12 2 phase threshold (Menu C1.5.3) ...................................................................................... 109
6.5.13 Diagnosis values (menu C1.5.4...C1.5.6)........................................................................... 110
6.5.14 Graphic page (menu C6.5)................................................................................................. 110
6.5.15 Save settings (menu C6.6.2).............................................................................................. 110
6.5.16 Load settings (menu C6.6.3) ............................................................................................. 110
6.5.17 Passwords (Menu 6.6.4 Quick Set; Menu 6.6.5 Setup) ..................................................... 111
6.5.18 Low flow cutoff .................................................................................................................. 111
6.5.19 Time constant.................................................................................................................... 112
6.5.20 Dual phase pulse output ................................................................................................... 112
6.5.21 Timeouts in programming mode ...................................................................................... 112
6.5.22 Output hardware ............................................................................................................... 113
6.6 Status messages and diagnostic information.............................................................. 113
6.7 Function tests and troubleshooting ............................................................................. 118
6.8 Diagnostic functions..................................................................................................... 120
6.8.1 Temperature (menu B2.6) .................................................................................................. 120
6.8.2 Strain (menu B2.7 strain MT / B2.8 strain IC) .................................................................... 120
6.8.3 Frequency (menu B2.9)....................................................................................................... 120
6.8.4 Drive level (menu B2.10)..................................................................................................... 120
6.8.5 Sensor levels A and B (menu B2.11, B2.12) ....................................................................... 121
6.8.6 2 phase signal (menu B2.13)............................................................................................... 121
6.8.7 SE board or BE board temperature (menu B2.14 or B2.15) .............................................. 121
7 Service 122
7.1 Replacing the sensor or converter electronics ........................................................... 122
7.1.1 Replacing the sensor electronics (SE) ............................................................................... 122
7.1.2 Replacing the signal converter electronics (BE)................................................................ 123
7.2 Driver or sensor coil fault ............................................................................................ 125
7.2.1 OPTIMASS 1000................................................................................................................... 125
7.2.2 OPTIMASS 2000................................................................................................................... 126
7.2.3 OPTIMASS 3000................................................................................................................... 127
7.2.4 OPTIMASS 7000................................................................................................................... 128
7.2.5 OPTIMASS 8000k................................................................................................................. 129
7.3 Spare parts availability................................................................................................. 130
7.4 Availability of services .................................................................................................. 130
7.5 Returning the device to the manufacturer................................................................... 130
7.5.1 General information............................................................................................................ 130
7.5.2 Form (for copying) to accompany a returned device.......................................................... 131
7.6 Disposal ........................................................................................................................ 131
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
5
Page 6
CONTENTS
MFC 300
8 Technical data 132
8.1 Measuring principle (single tube) ................................................................................ 132
8.2 Technical data............................................................................................................... 134
8.3 Dimensions and weights .............................................................................................. 145
8.3.1 Housing ............................................................................................................................... 145
8.3.2 Mounting plate, field housing ............................................................................................. 146
8.3.3 Mounting plate, wall-mounted housing ............................................................................. 146
9 Description of HART interface 147
9.1 General description ...................................................................................................... 147
9.2 Software history ...........................................................................................................147
9.3 Connection variants...................................................................................................... 148
9.3.1 Point-to-Point connection - analogue / digital mode......................................................... 149
9.3.2 Multi-Drop connection (2-wire connection) ....................................................................... 150
9.3.3 Multi-Drop connection (3-wire connection) ....................................................................... 151
9.4 Inputs/outputs and HART® dynamic variables and device variables.......................... 152
9.5 Parameter for the basic configuration......................................................................... 153
9.6 Field Communicator 375/475 (FC 375/475) ................................................................. 154
9.6.1 Installation .......................................................................................................................... 154
9.6.2 Operation............................................................................................................................. 154
9.6.3 Parameter for the basic configuration ............................................................................... 154
9.7 Asset Management Solutions (AMS)............................................................................ 155
9.7.1 Installation .......................................................................................................................... 155
9.7.2 Operation............................................................................................................................. 155
9.7.3 Parameter for the basic configuration ............................................................................... 155
9.8 Field Device Manager (FDM) ........................................................................................ 156
9.8.1 Installation .......................................................................................................................... 156
9.8.2 Operation............................................................................................................................. 156
9.9 Process Device Manager (PDM)................................................................................... 156
9.9.1 Installation .......................................................................................................................... 156
9.9.2 Operation............................................................................................................................. 157
9.9.3 Parameter for the basic configuration ............................................................................... 157
9.10 Field Device Tool / Device Type Manager (FDT / DTM) .............................................. 158
9.10.1 Installation ........................................................................................................................ 158
9.10.2 Operation........................................................................................................................... 158
9.11 Appendix A: HART® menu tree for Basic-DD ............................................................ 158
9.11.1 Overview Basic-DD menu tree (positions in menu tree).................................................. 159
9.11.2 Basic-DD menu tree (details for settings)........................................................................ 160
9.12 Appendix B: HART® menu tree for AMS .................................................................... 164
9.12.1 Overview AMS menu tree (positions in menu tree) .......................................................... 165
9.12.2 AMS menu tree (details for settings)................................................................................ 166
9.13 Appendix C: HART® menu tree for PDM.................................................................... 171
9.13.1 Overview PDM menu tree (positions in menu tree).......................................................... 171
9.13.2 PDM menu tree (details for settings) ............................................................................... 174
6
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 7
MFC 300
1.1 Software History
The "Electronic Revision" (ER) is consulted to document the revision status of electronic equipment according to NE 53 for all GDC devices. It is easy to see from the ER whether troubleshooting or larger changes in the electronic equipment have taken place and how that has affected the compatibility.
Changes and effect on compatibility
1 Downwards compatible changes and fault repair with no effect on operation (e.g. spelling
mistakes on display)
2-_ Downwards compatible hardware and/or software change of interfaces:
H
HART
P PROFIBUS
F Foundation Fieldbus
M Modbus
X all interfaces
3-_ Downwards compatible hardware and/or software change of inputs and outputs:
I Current output
F, P Frequency / pulse output
S Status output
C control input
CI Current input
X all inputs and outputs
4 Downwards compatible changes with new functions
5 Incompatible changes, i.e. electronic equipment must be changed.
SAFETY INSTRUCTIONS 1
®
INFORMATION!
In the table below, "x" is a placeholder for possible multi-digit alphanumeric combinations, depending on the available version.
Release date Electronic Revision Changes and
Documentation
compatibility
2006-11-06 ER 3.1.0x
(SW.REV. 3.10 (2.21))
2006-12-12 ER 3.1.1x
(SW.REV. 3.11 (2.21))
2007-02-07 ER 3.1.2x
(SW.REV. 3.11 (2.21))
2007-03-12 ER 3.1.3x
(SW.REV. 3.11 (2.21))
2007-06-27 ER 3.1.4x
(SW.REV. 3.11 (2.22))
2007-04-02 ER 3.2.0x
(SW.REV. 3.20 (2.22))
2007-05-04 ER 3.2.1x
(SW.REV. 3.20 (2.22))
2007-05-25 ER 3.2.2x
(SW.REV. 3.20 (2.22))
- -
1; 2-P; 2-M MA MFC 300 R02
1; 2-M MA MFC 300 R02
1; 2-H MA MFC 300 R02
1 MA MFC 300 R02
1; 2-X; 2-P; 2-F MA MFC 300 R02
1 MA MFC 300 R02
1; 3-I MA MFC 300 R02
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
7
Page 8
1 SAFETY INSTRUCTIONS
MFC 300
Release date Electronic Revision Changes and
compatibility
2007-06-27 ER 3.2.3x
(SW.REV. 3.20 (2.22))
2007-07-16 ER 3.2.4x
(SW.REV. 3.20 (2.22))
2008-08-01 ER 3.3.0x
(SW.REV. 3.30 (3.02))
2008-08-25 ER 3.3.1x
(SW.REV. 3.30 (3.03))
2008-10-23 ER 3.3.2x
(SW.REV. 3.30 (3.03))
2009-05-13 ER 3.3.3x
(SW.REV. 3.30 (3.03))
2009-10-29 ER 3.3.4x
(SW.REV. 3.30 (3.03))
2009-12-07 ER 3.3.5x
(SW.REV. 3.30 (3.03))
2011-03 ER 3.3.6x
(SW.REV. 3.40 (3.04))
2011-06 ER 3.3.7x
(SW.REV. 3.40 (3.04))
1 MA MFC 300 R02
1; 2-F MA MFC 300 R02
1; 2-X; 4 MA MFC 300 R02
1 MA MFC 300 R02
2-M MA MFC 300 R02
2-F MA MFC 300 R02
1 MA MFC 300 R02
2-F; 2-X MA MFC 300 R02
1; 2-F MA MFC 300 R02
1 MA MFC 300 R03
Documentation
8
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 9
MFC 300
1.2 Intended Use
The mass flowmeters are designed exclusively to directly measure mass flow rates, product density and temperature as well to indirectly measure parameters such as the total volume and concentration of dissolved substances as well as the volume flow rate.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
WARNING!
If the device is not used according to the operating conditions (refer to chapter "Technical data"), the intended protection could be affected.
1.3 Certifications
CE marking
SAFETY INSTRUCTIONS 1
The device fulfils the statutory requirements of the following EC directives:
Low Voltage Directive 2006/95/EC
EMC Directive 2004/108/EC
as well as
EN 61010
EMC specification acc. to EN 61326/A1
NAMUR recommendations NE 21 and NE 43
The manufacturer certifies successful testing of the product by applying the CE marking.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
9
Page 10
1 SAFETY INSTRUCTIONS
1.4 Safety instructions from the manufacturer
1.4.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.
MFC 300
We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.
1.4.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.
10
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 11
MFC 300
1.4.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.
1.4.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of underneath icons.
SAFETY INSTRUCTIONS 1
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
11
Page 12
1 SAFETY INSTRUCTIONS
1.4.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.
MFC 300
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems. There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the specified sequence.
i RESULT
RESULT
RESULTRESULT This symbol refers to all important consequences of the previous actions.
1.5 Safety instructions for the operator
12
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device.
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 13
MFC 300
2.1 Scope of delivery
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
DEVICE DESCRIPTION 2
Figure 2-1: Scope of delivery
1 Device in the version as ordered 2 Documentation (calibration report, factory and material certification if ordered, CD-Rom with product documentation
for measuring sensor and signal converter)
3 Signal cable (only for remote version)
Signal converter / measuring sensor combination possibilities
Measuring sensor Signal converter MFC 300
Compact Remote field
housing
OPTIMASS 1000 OPTIMASS 1300 C OPTIMASS 1300 F OPTIMASS 1300 W OPTIMASS 1300 R
OPTIMASS 2000 OPTIMASS 2300 C OPTIMASS 2300 F OPTIMASS 2300 W OPTIMASS 2300 R
OPTIMASS 3000 OPTIMASS 3300 C OPTIMASS 3300 F OPTIMASS 3300 W OPTIMASS 3300 R
OPTIMASS 7000 OPTIMASS 7300 C OPTIMASS 7300 F OPTIMASS 7300 W OPTIMASS 7300 R
OPTIMASS 8000 OPTIMASS 8300 C OPTIMASS 8300 F OPTIMASS 8300 W OPTIMASS 8300 R
Remote wall­mounted housing
Remote rack­mounted housing
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
13
Page 14
2 DEVICE DESCRIPTION
2.2 Device description
The mass flowmeters are designed exclusively to directly measure mass flow rates, product density and temperature as well to indirectly measure parameters such as the total volume and concentration of dissolved substances as well as the volume flow rate.
Your measuring device is supplied ready for operation. The factory settings for the operating data have been made in accordance with your order specifications.
The following versions are available:
Compact version (the signal converter is mounted directly on the measuring sensor)
Remote version (electrical connection to the measuring sensor via field current and signal
cable)
MFC 300
Figure 2-2: Device versions
1 Compact version 2 Measuring sensor with connection box 3 Field housing 4 Wall-mounted housing 5 19" rack-mounted housing
14
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 15
MFC 300
2.2.1 Field housing
Figure 2-3: Construction of the field housing
1 Cover for electronics and display 2 Cover for power supply and inputs/outputs terminal compartment 3 Cover for measuring sensor terminal compartment with locking screw 4 Cable entry for measuring sensor signal cable 5 Cable entry for measuring sensor field current cable 6 Cable entry for power supply 7 Cable entry for inputs and outputs 8 Mounting plate for pipe and wall mounting
DEVICE DESCRIPTION 2
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin­free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
15
Page 16
2 DEVICE DESCRIPTION
2.2.2 Wall-mounted housing
Figure 2-4: Construction of wall-mounted housing
1 Cover for terminal compartments 2 Terminal compartment for measuring sensor 3 Terminal compartment for inputs and outputs 4 Terminal compartment for power supply with safety cover (shock-hazard protection) 5 Cable entry for measuring sensor cable 6 Cable entry for inputs and outputs 7 Cable entry for inputs and outputs 8 Cable entry for power supply
MFC 300
1 Turn lock to the right and open the cover.
16
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 17
MFC 300
2.3 Nameplates
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
2.3.1 Compact version (example)
DEVICE DESCRIPTION 2
Figure 2-5: Example of nameplate for the compact version
1 Approvals-related information: Ex approval, EC type test certificate, hygienic approvals, etc. 2 Approvals-related thresholds 3 Additional information on documentation, calibration and patents 4 Protection category 5 Approvals-related pressure and temperature thresholds 6 Electrical connection data 7 Software and hardware revision (Electronics Revision), CG number, order number for signal converter and measuring
sensor
8 Manufacturing date, serial number and TAG number 9 Product description
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
17
Page 18
2 DEVICE DESCRIPTION
2.3.2 Remote version (example)
MFC 300
Figure 2-6: Example of a nameplate for the remote version
1 Product designation, manufacturing date, serial number and TAG number 2 Software and hardware revision (Electronics Revision), CG number, order number for signal converter and measuring
sensor
3 Electrical connection data 4 Additional information on documentation, calibration and patents
18
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 19
MFC 300
DEVICE DESCRIPTION 2
2.3.3 Electrical connection data of inputs/outputs (example of basic version)
Figure 2-7: Example of a nameplate for electrical connection data of inputs and outputs
1 Power supply (AC: L and N; DC: L+ and L-; PE for 24 VAC; FE for 24 VAC and DC) 2 Connection data of connection terminal D/D- 3 Connection data of connection terminal C/C- 4 Connection data of connection terminal B/B- 5 Connection data of connection terminal A/A-; A+ only operable in the basic version
A = active mode; the signal converter supplies the power for connection of the subsequent devices
P = passive mode; external power supply required for operation of the subsequent devices
N/C = connection terminals not connected
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
19
Page 20
3 INSTALLATION
3.1 Notes on installation
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
3.2 Storage
Store the device in a dry, dust-free location.
Avoid continuous direct sunlight.
Store the device in its original packing.
Storage temperature: -50...+70°C / -58...+158°F
MFC 300
3.3 Transport
Signal converter
No special requirements.
Compact version
Do not lift the device by the signal converter housing.
Do not use lifting chains.
To transport flange devices, use lifting straps. Wrap these around both process connections.
3.4 Installation specifications
INFORMATION!
The following precautions must be taken to ensure reliable installation.
Make sure that there is adequate space to the sides.
Protect the signal converter from direct sunlight and install a sun shade if necessary.
Signal converters installed in control cabinets require adequate cooling, e.g. by fan or heat exchanger.
Do not expose the signal converter to intense vibration. The flowmeters are tested for a vibration level in accordance with IEC 68-2-3.
20
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 21
MFC 300
3.5 Mounting of the compact version
INFORMATION!
The signal converter is mounted directly on the measuring sensor. For installation of the flowmeter, please observe the instructions in the supplied product documentation for the measuring sensor.
3.6 Mounting the field housing, remote version
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools in compliance with the applicable occupational health and safety directives.
3.6.1 Pipe mounting
INSTALLATION 3
Figure 3-1: Pipe mounting of the field housing
1 Fix the signal converter to the pipe. 2 Fasten the signal converter using standard U-bolts and washers. 3 Tighten the nuts.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
21
Page 22
3 INSTALLATION
3.6.2 Wall mounting
Figure 3-2: Wall mounting of the field housing
MFC 300
1 Prepare the holes with the aid of the mounting plate. For further information refer to
plate, field housing
on page 146.
Mounting
2 Use the mounting material and tools in compliance with the applicable occupational health
and safety directives.
3 Fasten the housing securely to the wall.
Mounting multiple devices next to each other
a 600 mm / 23.6" b 250 mm / 9.8"
22
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 23
MFC 300
3.6.3 Turning the display of the field housing version
INSTALLATION 3
Figure 3-3: Turning the display of the field housing version
The display of the field housing version can be turned in 90° increments.
1 Unscrew the cover from the display and operation control unit. 2 Using a suitable tool, pull out the two metal puller devices to the left and right of the display. 3 Pull out the display between the two metal puller devices and rotate it to the required position. 4 Slide the display and then the metal puller devices back into the housing. 5 Re-fit the cover and tighten it by hand.
CAUTION!
The ribbon cable of the display must not be folded or twisted repeatedly.
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin­free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
23
Page 24
3 INSTALLATION
3.7 Mounting the wall-mounted housing, remote version
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools in compliance with the applicable occupational health and safety directives.
3.7.1 Pipe mounting
MFC 300
Figure 3-4: Pipe mounting of the wall-mounted housing
1 Fasten the mounting plate to the pipe with standard U-bolts, washers and fastening nuts. 2 Screw the signal converter to the mounting plate with the nuts and washers.
24
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 25
MFC 300
3.7.2 Wall mounting
Figure 3-5: Wall mounting of the wall-mounted housing
INSTALLATION 3
1 Prepare the holes with the aid of the mounting plate. For further information refer to
plate, wall-mounted housing
2 Fasten the mounting plate securely to the wall. 3 Screw the signal converter to the mounting plate with the nuts and washers.
on page 146.
Mounting multiple devices next to each other
a 240 mm / 9.4"
Mounting
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
25
Page 26
4 ELECTRICAL CONNECTIONS
4.1 Safety instructions
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
MFC 300
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
4.2 Important notes on electrical connection
DANGER!
Electrical connection is carried out in conformity with the VDE 0100 directive "Regulations for electrical power installations with line voltages up to 1000 V" or equivalent national regulations.
CAUTION!
Use suitable cable entries for the various electrical cables.
The measuring sensor and signal converter have been configured together at the factory. For this reason, please connect the devices in pairs.
26
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 27
MFC 300
ELECTRICAL CONNECTIONS 4
4.3 Requirements for signal cables provided by the customer
INFORMATION!
If the signal cable was not ordered, it is to be provided by the customer. The following requirements regarding the electrical values of the signal cable must be observed:
Specifications for standard signal cables
2 twisted double wire circuits
20 AWG twisted, tinned copper conductors (19 mm / 0.2")
Completely tinned copper shielding
Casing colour: grey
Colour of wires:
Pair 1: black / red Pair 2 : green / white
Test voltage: 500 VAC RMS (750 VDC)
Temperature range: -20...+105°C / -4...+221°F
Capacity: 200 pF/m / 61 pF/ft
Inductance: 0.7 µH/m / 0.2 µH/ft
Specifications for cables in hazardous areas
2 shielded twisted double wire circuits
20 AWG twisted, tinned copper conductors (19 mm / 0.2")
Casing colour: blue
Colour of wires:
Pair 1: black / red Pair 2 : green / white
Test voltage: 500 VAC RMS (750 VDC)
Temperature range: -20...+105°C / -4...+221°F
Capacity: 200 pF/m / 61 pF/ft
Inductance: 0.7 µH/m / 0.2 µH/ft
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
27
Page 28
4 ELECTRICAL CONNECTIONS
4.4 Connecting the signal cables
DANGER!
Cables may only be connected when the power is switched off.
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel against electric shocks.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
MFC 300
28
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 29
MFC 300
4.4.1 Connection of signal cable, field housing
Figure 4-1: Electrical connection of the signal cables, field housing
a = white b = green c = black d = red
ELECTRICAL CONNECTIONS 4
1 Remove the locking screw and open the housing cover. 2 Pass the prepared signal cable through the cable entry. 3 Secure the signal cable using the clip. 4 Connect the electrical conductors as shown. The shielding is connected to terminal S. 5 Close the housing cover and secure it with the locking screw.
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin­free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
29
Page 30
4 ELECTRICAL CONNECTIONS
4.4.2 Connection of signal cable, wall-mounted housing
Figure 4-2: Electrical connection of signal cable, wall-mounted housing
a = white b = green c = black d = red
MFC 300
1 Open the housing cover. 2 Open the cover and guide the prepared signal cable through the cable entry. 3 Connect the twisted shielding to terminal S. 4 Connect the electrical conductors to terminals +, -, A, B. 5 Press the plug into the connector. 6 Close the cover and the housing cover.
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin­free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged.
30
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 31
MFC 300
ELECTRICAL CONNECTIONS 4
4.4.3 Connection of signal cable, 19" rack-mounted housing
Figure 4-3: Electrical connection of signal cable, 19" rack-mounted housing
Connect the conductor to the multipolar plug according to the illustration.
The shielding of the signal cable can be connected to 22z, 22d, 24z or 24d.
Press the plug into the connector.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
31
Page 32
4 ELECTRICAL CONNECTIONS
4.4.4 Connection box of measuring sensor
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel against electric shocks.
MFC 300
Figure 4-4: Electrical connection to connection box of the sensor
a = white b = green c = black d = red
1 Remove the locking screw and open the housing cover. 2 Pass the prepared signal cable through the cable entry. 3 Secure the signal cable using the spring terminal. The shielding MUST
MUST also be connected to
MUSTMUST
the spring terminal.
4 Connect the electrical conductors as shown. 5 Close the housing cover and secure it with the locking screw.
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin­free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged.
32
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 33
MFC 300
4.4.5 Connection diagram
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel against electric shocks.
Wall housing
ELECTRICAL CONNECTIONS 4
Figure 4-5: Connection diagram for remote versions, wall housing
1 Terminal compartment for signal converter 2 Terminal compartment for measuring sensor 3 Connect shielding to spring terminal 4 Connect shielding to terminal S 5 Functional ground
Field housing
Figure 4-6: Connection diagram for remote versions, field housing
1 Terminal compartment for signal converter 2 Terminal compartment for measuring sensor 3 Connect shielding to spring terminal 4 Connect shielding to terminal S 5 Functional ground
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
33
Page 34
4 ELECTRICAL CONNECTIONS
19" rack-mounted housing
Figure 4-7: Connection diagram for remote versions, 19" rack-mounted housing
1 Terminal compartment for signal converter 2 Terminal compartment for measuring sensor 3 Connect shielding to spring terminal 4 Connect shielding to terminal S
(The shielding can be connected to 22z, 22d, 24z or 24d)
5 Functional ground
MFC 300
4.5 Grounding the measuring sensor
DANGER!
There should be no difference in potential between the measuring sensor and the housing or protective earth of the signal converter!
The measuring sensor must be properly grounded.
The grounding cable should not transmit any interference voltages.
Do not use the grounding cable to connect more than one device to ground.
The measuring sensors are connected to ground by means of a functional grounding
conductor FE.
In hazardous areas, grounding is used at the same time for equipotential bonding. Additional grounding instructions are provided in the separate "Ex documentation", which are only supplied together with hazardous area equipment.
34
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 35
MFC 300
ELECTRICAL CONNECTIONS 4
4.6 Connecting power, all housing variants
DANGER!
The device must be grounded in accordance with regulations in order to protect personnel against electric shocks.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
The protection category depends on the housing versions (IP65...67 to IEC 529 / EN 60529 or NEMA4/4X/6).
The housings of the devices, which are designed to protect the electronic equipment from dust and moisture, should be kept well closed at all times. Creepage distances and clearances are dimensioned to VDE 0110 and IEC 664 for pollution severity 2. Supply circuits are designed for overvoltage category III and the output circuits for overvoltage category II.
Fuse protection (I breaker) to isolate the signal converter must be provided close to the device. The separator
must be marked as the separator for this device.
16 A) for the infeed power circuit, as well as a separator (switch, circuit
N
100...230 VAC (tolerance range: -15% / +10%)
Note the power supply voltage and frequency (50...60 Hz) on the nameplate.
The protective ground terminal PE
clamp terminal in the terminal compartment of the signal converter For the 19" rack-mounted housing please refer to the connection diagrams.
PE of the power supply must be connected to the separate U-
PEPE
INFORMATION!
240 VAC+5% is included in the tolerance range.
24 VDC (tolerance range: -55% / +30%) 24 VAC/DC (tolerance ranges: AC: -15% / +10%; DC: -25% / +30%)
Note the data on the nameplate!
For measurement process reasons, a functional ground FE
separate U-clamp terminal in the terminal compartment of the signal converter.
When connecting to functional extra-low voltages, provide a facility for protective separation (PELV) (acc. to VDE 0100 / VDE 0106 and/or IEC 364 / IEC 536 or relevant national regulations).
FE must be connected to the
FEFE
INFORMATION!
For 24 VDC, 12 VDC-10% is included in the tolerance range.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
35
Page 36
4 ELECTRICAL CONNECTIONS
Power supply connection (excluding 19" rack-mounted housing)
1 100...230 VAC (-15% / +10%), 22 VA 2 24 VDC (-55% / +30%), 12 W 3 24 VAC/DC (AC: -15% / +10%; DC: -25% / +30%), 22 VA or 12 W
Power supply connection for 19" rack-mounted housing (28 TE)
MFC 300
36
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 37
MFC 300
4.7 Inputs and outputs, overview
4.7.1 Combinations of the inputs/outputs (I/Os)
This signal converter is available with various input/output combinations.
Basic version
Has 1 current output, 1 pulse output and 2 status outputs / limit switches.
The pulse output can be set as status output/limit switch and one of the status outputs as a
control input.
Ex i version
Depending on the task, the device can be configured with various output modules.
Current outputs can be active or passive.
Optionally available also with Foundation Fieldbus and Profibus PA
Modular version
Depending on the task, the device can be configured with various output modules.
ELECTRICAL CONNECTIONS 4
Bus systems
The device allows intrinsically safe and non intrinsically safe bus interfaces in combination with additional modules.
For connection and operation of bus systems, please note the separate documentation.
Ex option
For hazardous areas, all of the input/output variants for the housing designs C and F with terminal compartment in the Ex d (pressure-resistant casing) or Ex e (increased safety) versions can be delivered.
Please refer to the separate instructions for connection and operation of the Ex-devices.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
37
Page 38
4 ELECTRICAL CONNECTIONS
4.7.2 Description of the CG number
Figure 4-8: Marking (CG number) of the electronics module and input/output variants
1 ID number: 2 2 ID number: 0 = standard; 9 = special 3 Power supply option 4 Display (language versions) 5 Input/output version (I/O) 6 1st optional module for connection terminal A 7 2nd optional module for connection terminal B
The last 3 digits of the CG number (5, 6 and 7) indicate the assignment of the terminal connections. Please refer to the following examples.
Examples for CG number
MFC 300
CG 320 11 100 100...230 VAC & standard display; basic I/O: Ia or Ip & Sp/Cp & Sp & Pp/S
p
CG 320 11 7FK 100...230 VAC & standard display; modular I/O: Ia & PN/SN and optional module PN/SN & C
CG 320 81 4EB 24 VDC & standard display; modular I/O: Ia & Pa/Sa and optional module Pp/Sp & I
p
Description of abbreviations and CG identifier for possible optional modules on terminals A and B
Abbreviation Identifier for CG No. Description
I
a
I
p
Pa / S
a
Pp / S
p
PN / S
N
C
a
C
p
C
N
IIn
a
IIn
p
- 8 No additional module installed
- 0 No further module possible
A Active current output
B Passive current output
C Active pulse output, frequency output, status output or limit switch
(changeable)
E Passive pulse output, frequency output, status output or limit switch
(changeable)
F Passive pulse output, frequency output, status output or limit switch acc.
to NAMUR (changeable)
G Active control input
K Passive control input
H Active control input to NAMUR
Signal converter monitors cable breaks and short circuits acc. to EN 60947-5-6. Errors indicated on LC display. Error messages possible via status output.
P Active current input
R Passive current input
N
38
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 39
MFC 300
4.7.3 Fixed, non-alterable input/output versions
This signal converter is available with various input/output combinations.
The grey boxes in the tables denote unassigned or unused connection terminals.
In the table, only the final digits of the CG no. are depicted.
Connection terminal A+ is only operable in the basic input/output version.
CG no. Connection terminals
A+ A A- B B- C C- D D-
Basic I/Os (standard)
1 0 0
Ia + HART® active 1
Ip + HART® passive 1
Ex i IOs (option)
2 0 0
3 0 0
2 1 0 Ia active PN / SNNAMUR
3 1 0 Ia active PN / SNNAMUR
2 2 0 Ip passive PN / SNNAMUR
3 2 0 Ip passive PN / SNNAMUR
Sp / Cp passive 2 Sp passive Pp / Sp passive 2
passive 2
C
p
Cp passive 2
passive 2
C
p
Cp passive 2
ELECTRICAL CONNECTIONS 4
Ia + HART® active
Ip + HART® passive
Ia + HART® active
Ip + HART® passive
Ia + HART® active
Ip + HART® passive
PN / SN NAMUR 2
PN / SN NAMUR 2
PN / SN NAMUR 2
PN / SN NAMUR 2
PN / SNNAMUR 2
PN / SNNAMUR 2
PROFIBUS PA (Ex i) (option)
D 0 0 PA+ PA- PA+ PA-
FISCO Device FISCO Device
D 1 0 Ia active PN / SNNAMUR
passive 2
C
p
D 2 0 Ip passive PN / SNNAMUR
passive 2
C
p
PA+ PA- PA+ PA-
FISCO Device FISCO Device
PA+ PA- PA+ PA-
FISCO Device FISCO Device
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
39
Page 40
4 ELECTRICAL CONNECTIONS
FOUNDATION Fieldbus (Ex i) (option)
E 0 0 V/D+ V/D- V/D+ V/D-
FISCO Device FISCO Device
E 1 0 Ia active PN / SNNAMUR
Cp passive 2
E 2 0 Ip passive PN / SNNAMUR
Cp passive 2
1 function changed by reconnecting 2 changeable
V/D+ V/D- V/D+ V/D-
FISCO Device FISCO Device
V/D+ V/D- V/D+ V/D-
FISCO Device FISCO Device
MFC 300
40
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 41
MFC 300
4.7.4 Alterable input/output versions
This signal converter is available with various input/output combinations.
The grey boxes in the tables denote unassigned or unused connection terminals.
In the table, only the final digits of the CG no. are depicted.
Term. = (connection) terminal
CG no. Connection terminals
A+ A A- B B- C C- D D-
Modular IOs (option)
4 _ _ max. 2 optional modules for term. A + B
8 _ _ max. 2 optional modules for term. A + B
6 _ _ max. 2 optional modules for term. A + B
B _ _ max. 2 optional modules for term. A + B
7 _ _ max. 2 optional modules for term. A + B
C _ _ max. 2 optional modules for term. A + B
ELECTRICAL CONNECTIONS 4
Ia + HART® active
Ip + HART® passive
Ia + HART® active
Ip + HART® passive
Ia + HART® active
Ip + HART® passive
Pa / Sa active 1
Pa / Sa active 1
Pp / Sp passive 1
Pp / Sp passive 1
PN / SN NAMUR 1
PN / SN NAMUR 1
PROFIBUS PA (option)
D _ _ max. 2 optional modules for term. A + B PA+ (2) PA- (2) PA+ (1) PA- (1)
FOUNDATION Fieldbus (option)
E _ _ max. 2 optional modules for term. A + B V/D+ (2) V/D- (2) V/D+ (1) V/D- (1)
PROFIBUS DP (option)
F _ 0 1 optional module for
term. A
Terminati on P
RxD/TxD­P(2)
RxD/TxD­N(2)
Terminati on N
RxD/TxD­P(1)
RxD/TxD­N(1)
Modbus (option)
G _ _ 2 max. 2 optional modules for term. A + B Common Sign. B
(D1)
H _ _ 3 max. 2 optional modules for term. A + B Common Sign. B
(D1)
1 changeable 2 not activated bus terminator 3 activated bus terminator
Sign. A (D0)
Sign. A (D0)
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
41
Page 42
4 ELECTRICAL CONNECTIONS
4.8 Description of the inputs and outputs
4.8.1 Current output
INFORMATION!
The current outputs must be connected depending on the version! Which I/O versions and inputs/outputs are installed in your signal converter are indicated on the sticker in the cover of the terminal compartment.
All outputs are electrically isolated from each other and from all other circuits.
All operating data and functions can be adjusted.
Passive mode: external power U
Active mode: load impedance R
R
450 Ω at I 22 mA for Ex i outputs
L
Self-monitoring: interruption or load impedance too high in the current output loop
Error message possible via status output, error indication on LC display.
Current value error detection can be adjusted.
Automatic range conversion via threshold or control input. The setting range for the
threshold is between 5 and 80% of Q smaller to larger range of 1:20 to 1:1.25).
Signaling of the active range possible via a status output (adjustable).
Forward / reverse flow measurement (F/R mode) is possible.
32 VDC at I 22 mA
ext
1kΩ at I 22 mA;
L
100%
MFC 300
, ± 0...5% hysteresis (corresponding ratio from
INFORMATION!
For further information refer to Connection diagrams of inputs and outputs on page 49
Technical data on page 134
to
.
and refer
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
42
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 43
MFC 300
4.8.2 Pulse and frequency output
INFORMATION!
Depending on the version, the pulse and frequency outputs must be connected passively or actively or according to NAMUR EN 60947-5-6! Which I/O version and inputs/outputs are installed in your signal converter are indicated on the sticker in the cover of the terminal compartment.
All outputs are electrically isolated from each other and from all other circuits.
All operating data and functions can be adjusted.
Passive mode:
External power supply required: U I 20 mA at f 10 kHz (override up to f I 100 mA at f 100 Hz
Active mode: Use of the internal power supply: U
I 20 mA at f 10 kHz (over range up to f I 20 mA at f 100 Hz
NAMUR mode: passive in accordance with EN 60947-5-6, f ≤ 10 kHz, over range up to f
Scaling: Frequency output: in pulses per time unit (e.g. 1000 pulses/s at Q
Pulse output: quantity per pulse.
Pulse width: symmetric (pulse duty factor 1:1, independent of output frequency) automatic (with fixed pulse width, duty factor approx. 1:1 at Q
fixed (pulse width adjustable as required from 0.05 ms...2 s)
Forward / reverse flow measurement (F/R mode) is possible.
All pulse and frequency outputs can also be used as a status output / limit switch.
max
12 kHz
32 VDC
ext
max
=24VDC
nom
ELECTRICAL CONNECTIONS 4
12 kHz)
12 kHz)
max
);
100%
) or
100%
CAUTION!
At frequencies above 100 Hz, shielded cables must be used to prevent radio interference.
INFORMATION!
For further information refer to Connection diagrams of inputs and outputs on page 49
Technical data on page 134
to
.
and refer
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
43
Page 44
4 ELECTRICAL CONNECTIONS
4.8.3 Status output and limit switch
INFORMATION!
Depending on the version, the status outputs and limit switches must be connected passively or actively or according to NAMUR EN 60947-5-6! Which I/O version and inputs/outputs are installed in your signal converter are indicated on the sticker in the cover of the terminal compartment.
The status outputs / limit switches are electrically isolated from each other and from all other circuits.
The output stages of the status outputs/limit switches during simple active or passive operation behave like relay contacts and can be connected with any polarity.
All operating data and functions can be adjusted.
Passive mode: external power supply required:
32 VDC; I 100 mA
U
ext
Active mode: use of the internal power supply:
=24VDC; I 20 mA
U
nom
NAMUR mode: passive in accordance with EN 60947-5-6
For information on the adjustable operating states refer to
Function tables
MFC 300
on page 79.
INFORMATION!
For further information refer to Connection diagrams of inputs and outputs on page 49
Technical data on page 134
to
.
and refer
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
44
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 45
MFC 300
4.8.4 Control input
INFORMATION!
Depending on the version, the control inputs must be connected passively or actively or according to NAMUR EN 60947-5-6! Which I/O version and inputs/outputs are installed in your signal converter are indicated on the sticker in the cover of the terminal compartment.
All control inputs are electrically isolated from each other and from all other circuits.
All operating data and functions can be adjusted.
Passive mode: external power supply required:
32 VDC
U
ext
Active mode: use of the internal power supply: U
nom
NAMUR mode: in accordance with EN 60947-5-6 (Active control input to NAMUR EN 60947-5-6: signal converter monitors cable breaks and short circuits acc. to EN 60947-5-6. Errors indicated on LC display. Error messages possible via status output.
For information on the adjustable operating states refer to
=24VDC
ELECTRICAL CONNECTIONS 4
Function tables
on page 79.
INFORMATION!
For further information refer to Connection diagrams of inputs and outputs on page 49
Technical data on page 134
to
.
and refer
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
45
Page 46
4 ELECTRICAL CONNECTIONS
4.9 Electrical connection of the inputs and outputs
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools in compliance with the applicable occupational health and safety directives.
4.9.1 Field housing, electrical connection of the inputs and outputs
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
INFORMATION!
For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from electrical interferences (EMC).
MFC 300
Figure 4-9: Terminal compartment for inputs and outputs in field housing
1 Open the housing cover. 2 Push the prepared cable through the cable entry and connect the necessary conductors. 3 Connect the shield if necessary. 4 Close the touch guard. 5 Close the housing cover.
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin­free and acid-free grease. Ensure that the housing gasket is properly fitted, clean and undamaged.
46
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 47
MFC 300
ELECTRICAL CONNECTIONS 4
4.9.2 Wall-mounted housing, electrical connection of the inputs and outputs
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
INFORMATION!
Use shielded cables for frequencies above 100 Hz. The electrical connection of the shielding must take place using 6.3 mm / 0.25" blade receptacles (insulation as per DIN 46 245) in the I/O terminal compartment.
Figure 4-10: Terminal compartment for inputs and outputs in wall-mounted housing
1 Open the housing cover. 2 Push the cables through the cable entry and connect them to the supplied connector plugs 3. 3 Connect the shield if necessary. 4 Route the connector plugs with the clamped conductors into the sockets provided for that pur-
pose.
5 Close the housing cover.
INFORMATION!
Ensure that the housing gasket is properly fitted, clean and undamaged.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
47
Page 48
4 ELECTRICAL CONNECTIONS
MFC 300
4.9.3 19" rack-mounted housing (28 TE), electrical connection of the inputs and outputs
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from electrical interferences (EMC).
Terminal A+ is only operable in the basic version.
Figure 4-11: Terminal compartment for inputs and outputs in rack-mounted housing
1 Shielding
Connect the conductor to the multipolar plug according to the illustration.
The signal cable shield is connected to the Pin S.
Press the plug into the connector.
4.9.4 Laying electrical cables correctly
Figure 4-12: Protect housing from dust and water
1 Lay the cable in a loop just before the housing. 2 Tighten the screw connection of the cable entry securely. 3 Never mount the housing with the cable entries facing upwards. 4 Seal cable entries that are not needed with a plug.
48
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 49
MFC 300
ELECTRICAL CONNECTIONS 4
4.10 Connection diagrams of inputs and outputs
4.10.1 Important notes
INFORMATION!
Depending on the version, the inputs/outputs must be connected passively or actively or acc. to NAMUR EN 60947-5-6! Which I/O version and inputs/outputs are installed in your signal converter are indicated on the sticker in the cover of the terminal compartment.
All groups are electrically isolated from each other and from all other input and output circuits.
Passive mode: An external power supply is necessary to operate (activation) the subsequent devices (U
Active mode: The signal converter supplies the power for operation (activation) of the subsequent devices, observe max. operating data.
Terminals that are not used should not have any conductive connection to other electrically conductive parts.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
ext
).
Description of the used abbreviations
I
I
a
P
a
P
N
S
a
S
N
C
a
C
N
IInaIInpCurrent input active or passive
Current output active or passive
p
P
Pulse/frequency output active or passive
p
Pulse/frequency output passive acc. to NAMUR EN 60947-5-6
S
Status output/limit switch active or passive
p
Status output/limit switch passive acc. to NAMUR EN 60947-5-6
C
Control input active or passive
p
Control input active acc. to NAMUR EN 60947-5-6: Signal converter monitors cable breaks and short circuits acc. to EN 60947-5-6. Errors indicated on LC display. Error messages possible via status output.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
49
Page 50
4 ELECTRICAL CONNECTIONS
4.10.2 Description of the electrical symbols
mA meter
0...20 mA or 4...20 mA and other RL is the internal resistance of the measuring point including the cable
resistance
DC voltage source (U
), external power supply, any connection polarity
ext
MFC 300
Table 4-1: Description of symbols
DC voltage source (U connection diagrams
Internal DC voltage source
Controlled internal power source in the device
Electronic or electromagnetic counter At frequencies above 100 Hz, shielded cables must be used to connect the counters. Ri Internal resistance of the counter
Button, NO contact or similar
), observe connection polarity according to
ext
50
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 51
MFC 300
4.10.3 Basic inputs/outputs
CAUTION!
Observe connection polarity.
Current output active (HART®), basic I/Os
U
I 22 mA
R
= 24 VDC nominal
int, nom
1kΩ
L
ELECTRICAL CONNECTIONS 4
Figure 4-13: Current output active I
a
Current output passive (HART®), basic I/Os
U
U
I 22 mA
U
R
Figure 4-14: Current output passive I
= 24 VDC nominal
int, nom
32 VDC
ext
1.8 V
0
(U
L
ext-U0
)/I
max
p
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
51
Page 52
4 ELECTRICAL CONNECTIONS
INFORMATION!
For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from electrical interferences (EMC).
Compact and field housing versions:
Compact and field housing versions: Shield connected via the cable terminals in the terminal
Compact and field housing versions:Compact and field housing versions: compartment.
Wall-mounted version:
Wall-mounted version: Shield connected using 6.3 mm / 0.25" push-on connectors
Wall-mounted version: Wall-mounted version: (insulation to DIN 46245) in the terminal compartment.
Any connection polarity.
Pulse/frequency output passive, basic I/Os
U
f
f
If the following maximum load resistance R
The minimum load resistance R
Can also be set as status output; for the electrical connection refer to status output
32 VDC
ext
in operating menu set to f
max
max
I 100 mA open: I 0.05 mA at U
= 32 VDC
ext
closed: U
U
= 0.2 V at I 10 mA
0, max
= 2 V at I 100 mA
0, max
in the operating menu set to 100 Hz < f
max
I 20 mA open: I 0.05 mA at U
= 32 VDC
ext
closed: U
U U
= 1.5 V at I 1 mA
0, max
= 2.5 V at I 10 mA
0, max
= 5.0 V at I 20 mA
0, max
reduced accordingly by parallel connection of R: f 100 Hz: R
f 1 kHz: R f 10 kHz: R
L, min
=(U
R
L, max
L, max
L, max
ext-U0
= 47 kΩ
= 10 kΩ
= 1 kΩ
)/I
max
L, min
connection diagram.
100 Hz:
10 kHz:
max
is exceeded, the load resistance RL must be
L, max
is calculated as follows:
MFC 300
52
Figure 4-15: Pulse/frequency output passive P
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
p
Page 53
MFC 300
ELECTRICAL CONNECTIONS 4
INFORMATION!
Any connection polarity.
Status output / limit switch passive, basic I/Os
U
I 100 mA
R
open:
The output is open when the device is de-energized.
X stands for the terminals B, C or D. The functions of the connection terminals depend on the
32 VDC
ext
= 47 kΩ
L, max
L, min
=(U
ext-U0
R
I 0.05 mA at U closed:
U U
= 0.2 V at I 10 mA
0, max
= 2 V at I 100 mA
0, max
settings refer to
)/I
max
= 32 VDC
ext
Function tables
on page 79.
Figure 4-16: Status output / limit switch passive S
p
Control input passive, basic I/Os
8VU
I
max
I
max
Switching point for identifying "contact open or closed": Contact open (off): U
Contact closed (on): U
Can also be set as a status output; for the electrical connection refer to status output connection diagram.
Figure 4-17: Control input passive C
1 Signal
32 VDC
ext
= 6.5 mA at U = 8.2 mA at U
24 VDC
ext
32 VDC
ext
2.5 V with I
0
8 V with I
0
= 0.4 mA
nom
= 2.8 mA
nom
p
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
53
Page 54
4 ELECTRICAL CONNECTIONS
4.10.4 Modular inputs/outputs and bus systems
CAUTION!
Observe connection polarity.
INFORMATION!
For further information on electrical connection refer to Description of the inputs and outputs on page 42
For the electrical connection of bus systems, please refer to the separate documentation for
the respective bus systems.
Current output active (only current output terminals C/C- have HART® capability), modular I/Os
U
int, nom
I 22 mA
1kΩ
R
L
X designates the connection terminals A, B or C, depending on the version of the signal converter.
.
= 24 VDC
MFC 300
Figure 4-18: Current output active I
a
Current output passive (only current output terminals C/C- have HART® capability), modular I/Os
U
I 22 mA
U
R
X designates the connection terminals A, B or C, depending on the version of the signal
Figure 4-19: Current output passive I
32 VDC
ext
1.8 V
0
= (U
L, max
converter.
- U0) / I
ext
max
p
54
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 55
MFC 300
ELECTRICAL CONNECTIONS 4
INFORMATION!
For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from electrical interferences (EMC).
Compact and field housing versions:
Compact and field housing versions: Shield connected via the cable terminals in the terminal
Compact and field housing versions:Compact and field housing versions: compartment.
Wall-mounted version:
Wall-mounted version: Shield connected using 6.3 mm / 0.25" push-on connectors
Wall-mounted version:Wall-mounted version: (insulation to DIN 46245) in the terminal compartment.
Any connection polarity.
Pulse/frequency output active, modular I/Os
U
f
f
If the following maximum load impedance R
The minimum load impedance R
X designates the connection terminals A, B or D, depending on the version of the signal
= 24 VDC
nom
in the operating menu set to f
max
100 Hz:
max
I 20 mA open: I 0.05 mA closed: U
= 24 V at I = 20 mA
0, nom
in operating menu set to 100 Hz < f
max
10 kHz:
max
I 20 mA open: I 0.05 mA closed: U
U U
= 22.5 V at I = 1 mA
0, nom
= 21.5 V at I = 10 mA
0, nom
=19V at I=20mA
0, nom
L, max
be reduced accordingly by parallel connection of R: f 100 Hz: R
f 1 kHz: R f 10 kHz: R
L, min
=(U
R
L, max
L, max
L, max
ext-U0
= 47 kΩ
= 10 kΩ
=1kΩ
)/I
max
is calculated as follows:
L, min
converter.
is exceeded, the load impedance RL must
Figure 4-20: Pulse / frequency output active P
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
a
55
Page 56
4 ELECTRICAL CONNECTIONS
INFORMATION!
For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from electrical interferences (EMC).
Pulse/frequency output passive, modular I/Os
U
f
f
If the following maximum load impedance R
The minimum load impedance R
Can also be set as status output; refer to status output connection diagram.
X designates the connection terminals A, B or D, depending on the version of the signal
32 VDC
ext
in the operating menu set to f
max
I 100 mA open: I 0.05 mA at U
= 32 VDC
ext
closed: U
U
=0.2V at I≤ 10 mA
0, max
=2V at I≤ 100 mA
0, max
in operating menu set to 100 Hz < f
max
open: I 0.05 mA at U
= 32 VDC
ext
closed: U
U U
=1.5V at I1mA
0, max
=2.5 V at I ≤ 10 mA
0, max
=5V at I≤ 20 mA
0, max
be reduced accordingly by parallel connection of R: f 100 Hz: R
f 1 kHz: R f 10 kHz: R
R
L, min
=(U
L, max
L, max
L, max
ext-U0
= 47 kΩ
= 10 kΩ
=1kΩ
)/I
max
L, min
converter.
100 Hz:
max
10 kHz:
max
is exceeded, the load impedance RL must
L, max
is calculated as follows:
MFC 300
56
Figure 4-21: Pulse frequency output passive P
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
p
Page 57
MFC 300
ELECTRICAL CONNECTIONS 4
INFORMATION!
For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from electrical interferences (EMC).
Compact and field housing versions:
Compact and field housing versions: Shield connected via the cable terminals in the terminal
Compact and field housing versions:Compact and field housing versions: compartment.
Wall-mounted version:
Wall-mounted version: Shield connected using 6.3 mm / 0.25" push-on connectors
Wall-mounted version:Wall-mounted version: (insulation to DIN 46245) in the terminal compartment.
Any connection polarity.
Pulse and frequency output passive PN NAMUR, modular I/O
Connection in conformity with EN 60947-5-6
open:
= 0.6 mA
I
nom
closed: I
= 3.8 mA
nom
X designates the connection terminals A, B or D, depending on the version of the signal converter.
Figure 4-22: Pulse and frequency output passive PN to NAMUR EN 60947-5-6
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
57
Page 58
4 ELECTRICAL CONNECTIONS
Status output / limit switch active, modular I/Os
Observe connection polarity.
U
I 20 mA
R
open:
X designates the connection terminals A, B or D, depending on the version of the signal
=24VDC
int
47 kΩ
L
I 0.05 mA closed: U
= 24 V at I = 20 mA
0, nom
converter.
MFC 300
Figure 4-23: Status output / limit switch active S
a
Status output / limit switch passive, modular I/Os
Any connection polarity.
U
I 100 mA
R
open:
The output is open when the device is de-energized.
X designates the connection terminals A, B or D, depending on the version of the signal
=32VDC
ext
= 47 kΩ
L, max
L, min
=(U
ext-U0
R
I 0.05 mA at U closed:
U U
=0.2V at I≤ 10 mA
0, max
=2V at I≤ 100 mA
0, max
converter.
)/I
= 32 VDC
ext
max
58
Figure 4-24: Status output / limit switch passive S
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
p
Page 59
MFC 300
ELECTRICAL CONNECTIONS 4
Status output / limit switch SN NAMUR, modular I/Os
Any connection polarity.
Connection in conformity with EN 60947-5-6
open:
= 0.6 mA
I
nom
closed:
= 3.8 mA
I
nom
The output is open when the device is de-energized.
X designates the connection terminals A, B or D, depending on the version of the signal
converter.
Figure 4-25: Status output / limit switch SN to NAMUR EN 60947-5-6
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
59
Page 60
4 ELECTRICAL CONNECTIONS
CAUTION!
Observe connection polarity.
Control input active, modular I/Os
U
External contact open:
Switching point for identifying "contact open or closed":
X designates the connection terminals A or B, depending on the version of the signal
Control input passive, modular I/Os
3VU
I
Switching point for identifying "contact open or closed":
X designates the connection terminals A or B, depending on the version of the signal
=24VDC
int
0, nom
=22V
U External contact closed:
I
= 4 mA
nom
Contact open (off): U Contact closed (on): U
0
0
converter.
32 VDC
ext
=9.5mA at U
max
=9.5mA at U
I
max
Contact open (off): U
ext
ext
24 V32 V
0
Contact closed (on): U
converter.
10 V with I
12 V with I
2.5 V with I
3V with I
0
nom
nom
nom
nom
MFC 300
= 1.9 mA
= 1.9 mA
=1.9mA
=1.9mA
60
Figure 4-26: Control input passive C
1 Signal
p
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 61
MFC 300
ELECTRICAL CONNECTIONS 4
CAUTION!
Observe connection polarity.
Control input active CN NAMUR, modular I/Os
Connection acc. to EN 60947-5-6
Switching point for identifying "contact open or closed":
Contact open (off): U Contact closed (on): U
= 6.3 V with I
0, nom
0, nom
= 6.3 V with I
Detection of cable break:
8.1 V with I 0.1 mA
U
0
Detection of cable short circuit:
1.2 V with I 6.7 mA
U
0
X designates the connection terminals A or B, depending on the version of the signal converter.
< 1.9 mA
nom
nom
> 1.9 mA
Figure 4-27: Control input active CN to NAMUR EN 60947-5-6
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
61
Page 62
4 ELECTRICAL CONNECTIONS
4.10.5 Ex i inputs/outputs
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
INFORMATION!
For further information on electrical connection refer to Description of the inputs and outputs on page 42
Current output active (only current output terminals C/C- have HART® capability), Ex i I/Os
Observe connection polarity.
U
I 22 mA
R
X designates the connection terminals A or C, depending on the version of the signal
.
= 20 VDC
int, nom
450 Ω
L
converter.
MFC 300
Figure 4-28: Current output active Ia Exi
Current output passive (only current output terminals C/C- have HART® capability), Ex i I/Os
Any connection polarity.
U
I 22 mA
U
R
X designates the connection terminals A or C, depending on the version of the signal
Figure 4-29: Current output passive Ip Exi
32 VDC
ext
4 V
0
=(U
L, min
converter.
ext-U0
)/I
max
62
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 63
MFC 300
ELECTRICAL CONNECTIONS 4
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
INFORMATION!
For frequencies above 100 Hz, shielded cables are to be used in order to reduce effects from electrical interferences (EMC).
Compact and field housing versions:
Compact and field housing versions: Shield connected via the cable terminals in the terminal
Compact and field housing versions:Compact and field housing versions: compartment.
Wall-mounted version:
Wall-mounted version: Shield connected using 6.3 mm / 0.25" push-on connectors
Wall-mounted version:Wall-mounted version: (insulation to DIN 46245) in the terminal compartment.
Any connection polarity.
Pulse and frequency output passive PN NAMUR, Ex i I/Os
Connection acc. to EN 60947-5-6
open:
= 0.43 mA
I
nom
closed: I
= 4.5 mA
nom
X designates the connection terminals B or D, depending on the version of the signal converter.
Figure 4-30: Pulse and frequency output passive PN as per NAMUR EN 60947-5-6 Exi
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
63
Page 64
4 ELECTRICAL CONNECTIONS
INFORMATION!
Any connection polarity.
Status output/limit switch SN NAMUR, Ex i I/Os
Connection acc. to EN 60947-5-6
open:
= 0,43 mA
I
nom
closed: I
= 4,5 mA
nom
The output is closed when the device is de-energized.
X designates the connection terminals B or D, depending on the version of the signal
converter.
MFC 300
Figure 4-31: Status output/limit switch SN to NAMUR EN 60947-5-6 Exi
64
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 65
MFC 300
ELECTRICAL CONNECTIONS 4
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
INFORMATION!
Any connection polarity.
Control input passive, Ex i I/Os
5.5 V U
I
= 6 mA at U
max
I
= 6.5 mA at U
max
Switching point for identifying "contact open or closed": Contact open (off): U
Contact closed (on): U
X designates the connection terminals B, if available.
32 VDC
ext
24 V
ext
32 V
ext
3.5 V with I 0.5 mA
0
5.5 V with I 4 mA
0
Figure 4-32: Control input passive Cp Exi
1 Signal
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
65
Page 66
4 ELECTRICAL CONNECTIONS
4.10.6 HART® connection
INFORMATION!
In the basic I/O the current output at connection terminals A+/A-/A always has HART® capability.
For modular I/O and Ex i I/O, only the current output module for the connection terminals
®
C/C- has HART
HART® connection active (point-to-point)
capability.
MFC 300
Figure 4-33: HART® connection active (Ia)
1 Basic I/O: terminals A and A+ 2 Modular I/O: terminals C- and C
3 HART
®
communicator
The parallel resistance to the HART® communicator must be R 230 Ω.
66
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 67
MFC 300
ELECTRICAL CONNECTIONS 4
HART® connection passive (Multi-Drop mode)
I: I0% 4mA
Multi-Drop mode I: I
U
32 VDC
ext
R ≥ 230 Ω
4mA = I
fix
0%
Figure 4-34: HART® connection passive (Ip)
1 Basic I/O: terminals A- and A 2 Modular I/O: terminals C- and C
3 HART 4 Other devices with HART
®
communicator
®
capability
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
67
Page 68
5 START-UP
5.1 Switching on the power
Before connecting to power, please check that the system has been correctly installed. This includes:
The device must be mechanically safe and mounted in compliance with the regulations.
The power connections must have been made in compliance with the regulations.
The electrical terminal compartments must be secured and the covers have been screwed
on.
Check that the electrical operating data of the power supply are correct.
Switching on the power.
5.2 Starting the signal converter
The measuring device, consisting of the measuring sensor and the signal converter, is supplied ready for operation. All operating data have been set at the factory in accordance with your order specifications.
MFC 300
When the power is switched on, a self test is carried out. After that the device immediately begins measuring, and the current values are displayed.
Figure 5-1: Displays in measuring mode (examples for 2 or 3 measured values) x, y and z denote the units of the measured values displayed
It is possible to change between the two measured value windows, the trend display and the list with the status messages by pressing the keys and . For information about possible status messages, their meaning and cause refer to page 113.
Status messages and diagnostic information
on
68
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 69
MFC 300
6.1 Display and operating elements
Figure 6-1: Display and operating elements (Example: flow indication with 2 measuring values)
1 Indicates a possible status message in the status list 2 Tag number (is only indicated if this number was entered previously by the operator) 3 Indicates when a key has been pressed 4 1st measured variable in large representation 5 Bargraph indication 6 Keys (see table below for function and representation in text) 7 Interface to the GDC bus (not present in all signal converter versions) 8 Infrared sensor (not present in all signal converter versions)
OPERATION 6
CAUTION!
The use of a jumper is only permitted for custody transfer devices to lock the access to custody transfer relevant parameters. For non custody transfer devices (i.e. process instruments) this jumper must not be used!
INFORMATION!
The switching point for the 4 optical keys is located directly in front of the glass. It is recommended to activate the keys at right angles to the front. Touching them from the side can cause incorrect operation.
After 5 minutes of inactivity, there is an automatic return to measuring mode. Previously changed data is not saved.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
69
Page 70
6 OPERATION
MFC 300
Key Measuring mode Menu mode Submenu or function
mode
> Switch from measuring
mode to menu mode; press key for 2.5 s, "Quick Start" menu is then displayed
^ Reset of display Return to measuring
or Switch between display
pages: measured value 1 + 2, trend page and status page(s)
Esc (> + ↑) - - Return to menu mode
Table 6-1: Description of key functionality
Access to displayed menu, then 1st submenu is displayed
mode but prompt whether the data should be saved
Select menu Select submenu or
Access to displayed submenu or function
Press 1 to 3 times, return to menu mode, data saved
function
without acceptance of data
Parameter and data mode
For numerical values, move cursor (highlighted in blue) one position to the right
Return to submenu or function, data saved
Use cursor highlighted in blue to change number, unit, setting and to move the decimal point
Return to submenu or function without acceptance of data
70
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 71
MFC 300
6.1.1 Display in measuring mode with 2 or 3 measured values
Figure 6-2: Example for display in measuring mode with 2 or 3 measured values
1 Indicates a possible status message in the status list 2 Tag number (is only indicated if this number was entered previously by the operator) 3 1st measured variable in large depiction 4 Bargraph indication 5 Depiction with 3 measured values
6.1.2 Display for selection of sub-menu and functions, 3 lines
OPERATION 6
Figure 6-3: Display for selection of sub-menu and functions, 3 lines
1 Indicates a possible status message in the status list 2 Menu, sub-menu or function name 3 Number relating to 2 4 Indicates position within menu, sub-menu or function list 5 Next menu, sub-menu or function
(_ _ _ signalise in this line the end of the list)
6 Current menu, sub-menu or function 7 Previous menu, sub-menu or function
(_ _ _ signalise in this line the beginning of the list)
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
71
Page 72
6 OPERATION
6.1.3 Display when setting parameters, 4 lines
Figure 6-4: Display when setting parameters, 4 lines
1 Current menu, sub-menu or function 2 Number relating to 1 3 Denotes factory setting 4 Denotes permissible value range 5 Permissible value range for numeric values 6 Currently set value, unit or function (when selected, appears with white text, blue background)
This is where the data is changed.
7 Current parameter (open with >) 8 Factory setting of parameter (non-alterable)
MFC 300
6.1.4 Display when changing parameters, 4 lines
Figure 6-5: Display when changing parameters, 4 lines
1 Current menu, sub-menu or function 2 Number relating to 1 3 Denotes the change of a parameter (simple to check changed data when browsing through lists) 4 Next parameter 5 Currently set data from 6 6 Current parameter (for selection press key >; then see previous chapter) 7 Factory setting of parameter (non-alterable)
72
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 73
MFC 300
6.1.5 Using an IR interface (option)
The optical IR interface serves as an adapter for PC-based communication with the signal converter without opening the housing.
INFORMATION!
This device is not part of the scope of delivery.
For more information about activation with the functions A6 or C6.6.6 refer to Function tables on page 79
.
OPERATION 6
Figure 6-6: IR interface
1 Glass panel in front of the control and display panel 2 IR interface 3 LED lights up when IR interface is activated. 4 Suction cups
Timeout function
Following activation of the IR interface in Fct. A6 or C6.6.6 the interface must be properly positioned and attached to the housing with the suction cups within 60 seconds. If this does not happen within the specified time period, the device can be operated using the optical keys again. Upon activation, the LED 3 lights up and the optical keys no longer function.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
73
Page 74
6 OPERATION
6.2 Zero calibration (menu C1.1.1)
Following installation, perform zero calibration prior to commissioning the device. Finalise the installation before performing zero calibration. Changes (pipe system or calibration factor) made after the zero calibration may affect the accuracy, making it necessary to perform zero calibration again.
Observe the following for reliable zero calibration:
The measuring sensor should be completely filled with the product at the expected process pressure and temperature.
The product may contain no air or gas, especially when it comes to horizontal installations. Prior to zero calibration, it is recommended that the product be flushed at a high flow rate (>50%), for 2 minutes.
After flushing, re-establish zero flow by closing off the corresponding valves.
Set zero calibration automatically or manually using the operating controls. The converter cover must be installed on the display for automatic calibration.
A) Automatic calibration
MFC 300
Key Display Description and setting
> A Quick Setup Press and hold for 2.5 s, then release
2 x C Setup
3 x > C1.1.1 Zero Calibration
> Calibrate Zero?
Break
Calibrate Zero?
Automatic
^ Please wait countdown from 40 s
Zero Calibration +XX.XXX%
5 x ^ Save Configuration?
Yes
^ Display page
the key.
Display of measured zero calibration in %. (Warning, value can be changed!)
74
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 75
MFC 300
OPERATION 6
B) Manual calibration
Key Display Description and setting
> A Quick Setup Press and hold for 2.5 s, then release
2 x C Setup
3 x > C1.1.1 Zero Calibration
> Calibrate Zero?
Break
3 x Calibrate Zero?
Manual
Zero Calibration +XX.XXX%
5 x ^ Save Configuration?
Yes
^ Display page
the key.
Display currently stored zero calibration in %. (Warning, value can be changed!)
Possible manual entry of zero calibration.
Storing the displayed zero calibration.
Under certain conditions, zero calibration is not possible and will be cancelled:
The product is still flowing. The shut-off valves are not closed tightly enough.
There are still gas bubbles in the fluid.
Remedy: Flush sensor and repeat calibration
With some media, it may be difficult to perform zero calibration. In such cases, there are various methods to still achieve good zero calibration:
Medium Possible solutions
Media that tend to evaporate or outgas Increase pressure.
Two-phase media (slurries), that contain solids that may fall out.
Two-phase media in which the solids or gas­forming components cannot be separated.
Only fill the sensor with the carrier medium.
Fill sensor with another liquid, e.g. water.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
75
Page 76
6 OPERATION
6.3 Menu structure
INFORMATION!
Note the key function within and between the columns.
MFC 300
Measuring mode Select
menu↓ ↑
^ Press
>2.5s
A Quick Setup > ^A1 Language >
>
Select menu and/or submenu
A2 Tag
A3 Reset > ^3.1 Error Reset
3.2 Reset Totaliser 1
3.3 Reset Totaliser 2
3.4 Reset Totaliser 3
A4 Analogue Outputs 4.1 Measurement
4.2 Unit
4.3 Range
4.4 Low Flow Cutoff
4.5 Time Constant
A5 Digital Outputs 5.1 Measurement
5.2 Pulse Value Unit
5.3 Value p. Pulse
5.4 Low Flow Cutoff
A6 GDC IR interface
A7 Zero Calibration
A8 Operation Mode
Select function and set data >
^
76
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 77
MFC 300
OPERATION 6
Measuring Mode Select
menu↓ ↑
^ Press
>2.5s
B Test > ^B1 Simulation > ^1.1 Mass Flow >
>
Select menu and/or sub-menu
1.2 Density
1.3 Temperature
1. Current Output X
1. Status Output X
1. Status Output X
1. Pulse Output X
B2 Actual Values > ^2.1 Operating Hours
2.2 Mass Flow
2.3 Volume Flow
2.4 Velocity
2.5 Density
2.6 Temperature
2.7 Strain MT
2.8 Strain IC
2.9 Tube Frequency
2.10 Drive Level
2.11 Sensor A Level
2.12 Sensor B Level
2.13 2 Phase Signal
2.14 SE PCB Temperature
2.15 BE PCB Temperature
2.16 Act. Operat. Mode
B3 Information > ^3.1 C Number
3.2 Sensor Electronics
3.3 SW.Rev.MS
3.4 SW.Rev.UIS
3.5 Electronic Revision ER
Select function and set data >
^
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
77
Page 78
6 OPERATION
MFC 300
Measuring Mode Select
menu↓ ↑
^ Press
>2.5s
C Setup > ^C1 Process Input > ^1.1 Calibration >
^ > ^C3 I/O (Inputs/Outputs) > ^3.1 Hardware >
^ > ^C4 I/O Totalisers > ^4.1 Totaliser 1 >
^ > ^C5 I/O HART > ^5.1 PV is >
^ > ^C6 Device > ^6.1 Device Info >
↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ >
Select menu and/or sub-menu
1.2 Density
1.3 Filter
1.4 System Control
1.5 Self Test
1.6 Information
1.7 Factory Calib.
1.8 Simulation
> ^C2 Concentration >
3. Current Output X
3. Frequency Output X
3. Pulse Output X
3. Status Output X
3. Limit Switch X
3. Control Input X
4.2 Totaliser 2
4.3 Totaliser 3
5.2 SV is
5.3 TV is
5.4 4V is
5.5 HART Units
6.2 Display
6.3 1. Meas. Page
6.4 2nd Meas. Page
6.5 Graphic Page
6.6 Special Functions
6.7 Units
6.8 HART
6.9 Quick Setup
Select function and set data >
^
^
^
^
^
^
78
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 79
MFC 300
OPERATION 6
6.4 Function tables
INFORMATION!
Depending on the device version, not all functions are available.
6.4.1 Menu A, Quick Setup
No. Function Settings / descriptions
A1 Language
A1 Language Language selection depends on the device version.
A2 Tag
A2 Tag
A3 Reset?
A3 Reset?
A3.1 Error Reset Reset Errors? Select: no/yes
A3.2 Reset Totaliser 1 Reset Totaliser? Select: no / yes (available if activated in C6.9.1)
A3.3 Reset Totaliser 2 Reset Totaliser? Select: no / yes (available if activated in C6.9.2)
A3.4 Reset Totaliser 3 Reset Totaliser? Select: no / yes (available if activated in C6.9.3)
Measuring point identifier (Tag no.) (also for HART® operation), appears in the LC display header (up to 8 digits).
A4 Analogue Outputs (only for HART®)
A4 Analogue Outputs Applicable to all current outputs (terminals A, B and C), frequency outputs
A4.1 Measurement Select measurement: volume flow / mass flow / temperature / density /
A4.2 Unit Selection of the unit from a list, depending on the measurement.
A4.3 Range 1) Setting for main current output (range: 0...100%)
A4.4 Low Flow Cutoff 1) Setting for main current output (sets output value to "0")
A4.5 Time Constant 1) Setting for main current output (applicable to all flow measurements)
(terminals A, B and D), limit switch (terminals A, B, C, and / or D) and the 1st display page / line 1.
velocity /diagnosis 1 / diagnosis 2 /
Depending on the settings for the concentration measurement, the following measurements are possible: diagnosis 3 / concentration 1 / concentration 2 / concentration flow 1 / concentration flow 2
2) Use for all outputs? (also use this setting for Fct. A4.2...A4.5!) Setting: no (applies only to the main current output ) / yes (applies to all analogue outputs)
Setting: 0x.xx (format and unit, depending on measurement, see A4.1 and A4.2 above)
2) Use for all outputs? Make setting, see Fct. A4.1 above!
Setting: x.xxx ± x.xxx% (range: 0.0…20%) (1st value = switching point / 2nd value = hysteresis), condition: 2nd value 1st value
2) Use for all outputs? Make setting, see Fct. A4.1 above!
Setting: xxx.x s (range: 000.1…100 s)
2) Use for all outputs? Make setting, see Fct. A4.1 above!
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
79
Page 80
6 OPERATION
No. Function Settings / descriptions
A4 Station Address
A4 Station Address For Profibus / FF / Modbus devices.
A5 Digital Outputs
A5 Digital Outputs Valid for all pulse outputs (terminals A, B and/or D) and totaliser 1.
A5.1 Measurement 1) Select measurement: volume flow / mass flow / concentration flow 1
2) Use for all outputs? (also use this setting for Fct. A5.2...A5.5!) Setting: no (only for pulse output D) / yes (for all digital outputs)
A5.2 Pulse Value Unit Selection of the unit from a list, depending on the measurement.
A5.3 Value p. Pulse 1) Setting for pulse output D (volume or mass value per pulse)
Setting: xxx.xxx in l/s or kg/s
2) Use for all outputs? Make setting, see Fct. A5.1 above!
A5.4 Low Flow Cutoff 1) Setting for pulse output D (sets output value to "0")
Setting: x.xxx ± x.xxx% (range: 0.0…20%) (1st value = switching point / 2nd value = hysteresis), condition: 2nd value 1st value
2) Use for all outputs? Make setting, see Fct. A5.1 above!
MFC 300
A6 GDC IR interface
A6 GDC IR interface After this function has been activated an optical GDC adapter can be
connected to the LC display. If approximately 60 seconds pass without a connection being established or after the adapter is removed, then the function is exited and the optical keys are active once again.
break (exit function without connection)
activate (the IR interface (adapter) and interrupt the optical keys)
A7 Zero Calibration
A7 Zero Calibration Zero calibration, calibration sequence as described in Fct. C1.1.1...1.1.4
A8 Operation Mode
A8 Operation Mode Set the operating mode.
Setting: measure / stop / standby
For detailed information refer to
Mode (menu A8)
on page 98.
80
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 81
MFC 300
6.4.2 Menu B, Test
No. Function Settings / descriptions
B1 Simulation
B1 Simulation Displayed values are simulated.
B1.1 Mass Flow Simulation of mass flow
Set value (range and units depend on measurement)
break (exit function without simulation)
Query: start simulation?
Settings: no (exit function without simulation) / yes (start simulation)
B1.2 Density Sequence and settings similar to B1.1, see above!
B1.3 Temperature
B1. Current Output X Simulation X
B1. Pulse Output X
B1. Frequency Output X
B1. Control Input X
B1. Limit Switch X
B1. Status Output X
X stands for one of the connection terminals A, B, C or D stands for Fct. B1.4...1.7
X stands for one of the connection terminals A, B, C or D Sequence and settings similar to B1.1, see above! For pulse output a set number of pulses are output in 1 s!
OPERATION 6
B2 Actual Values
B2 Actual Values Display of current values;
B2.1 Operating Hours Device operating hours
B2.2 Mass Flow Current unfiltered mass flow
B2.3 Volume flow Current unfiltered volume flow
B2.4 Velocity Current unfiltered flow velocity
B2.5 Density Current unfiltered density
B2.6 Temperature Current unfiltered temperature
B2.7 Strain MT Current value for the measuring tube strain
B2.8 Strain IC Current value for the inner cylinder strain
B2.9 Tube Frequency Current vibration frequency of the measuring tube
B2.10 Drive Level Current drive level to activate vibration
B2.11 Sensor A Level Current vibration amplitude
B2.12 Sensor B Level
B2.13 2 Phase Signal 2 phase indicator value
B2.14 SE PCB Temperature Temperature of sensor electronics
B2.15 BE PCB Temperature Temperature of signal converter
B2.16 Act. Operat. Mode Current operating mode
Exit the displayed function by pressing the ^ key.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
81
Page 82
6 OPERATION
B3 Information
B3 Information
B3.1 C Number CG number, cannot be changed (input/output version)
B3.2 Sensor Electronics
B3.3 SW.REV.MS LC display:
B3.4 SW.REV.UIS
B3.5 "Bus interface" Only appears with Profibus, Modbus and FF.
B3.6 Electronic Revision ER LC display see Fct. B3.3 and B3.4
6.4.3 Menu C, Setup
No. Function Settings / descriptions
C1 Process Input
1st line: ID No. of the circuit board 2nd line: software version 3rd line: production date
MFC 300
C1.1 Calibration
C1.1 Calibration
C1.1.1 Zero Calibration Display of current zero calibration value.
Query: calibrate zero?
Setting: break (return with ^) / standard (factory setting) / manual (display last value; set new value, range: -10…+10%) / automatic (shows the current value as the new zero calibration value)
C1.1.2 Zero Add. Offset Direct setting of zero offset
C1.1.3 Pipe Diameter Set the pipe diameter in mm to calculate flow velocity
C1.1.4 Flow Correction Defines additional correction for mass flow;
Range: -100...+100%
C1.2 Density
C1.2.1 Density Calib. Start density calibration
For detailed information refer to
99.
C1.2.2 Density Density mode selection: Actual (back key ^) / fixed (a fixed value is used for
the density (e.g. standard density)) / Reference (calculates process density based on a reference temperature)
C1.2.3 Fixed Density Value Set the fixed value (e.g. standard density) for the density.
Appears only if density mode "fixed" has been selected in Fct. C1.2.2.
C1.2.3 Density Ref. Temp. Set the reference temperature for the reference density option
Appears only if density mode "reference" has been selected in Fct. C1.2.2.
C1.2.4 Ref. Density Slope Set the slope for the reference density option
Appears only if density mode "reference" has been selected in Fct. C1.2.2.
Density calibration (menu C1.2.1)
on page
82
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 83
MFC 300
OPERATION 6
C1.3 Filter
C1.3 Filter
C1.3.1 Flow Direction Define polarity of flow direction.
Forwards (according to the arrow on the measuring sensor) or backwards (in the opposite direction to the arrow)
C1.3.2 Press. Supp. Time Set the pressure suppression time, range: 0.0...20.0 s
C1.3.3 Press. Supp. Cutoff Cutoff settings for the pressure suppression; range: 0.0...10.0%
C1.3.4 Density Averaging Set the time constant for the density measurements; range: 1.0...20.0 s
C1.3.5 Low Flow Cutoff Set the low flow cutoff; range: 00.0...10.0%
C1.4 System Control
C1.4 System Control
C1.4.1 Function Set the system control.
C1.4.2 Sys. Ctrl. Condition Set the condition for activating the system control.
C1.4.3 Sys. Ctrl. Max Limit Defines the upper limit for the condition selected in C1.4.2
C1.4.4 Sys. Ctrl. Min Limit Defines the lower limit for the condition selected in C1.4.2
Select: inactive (off) / flow = 0 (flow to zero)
Selection: density or temperature
C1.5 Self Test
C1.5 Self Test
C1.5.1 Max. Rec. Temp. Display of maximum recorded sensor temperature
C1.5.2 Min. Rec. Temp. Display of minimum recorded sensor temperature
C1.5.3 2 Ph. Threshold Defines the process-dependent sensitivity for 2 phase signal error message.
C1.5.4 Diagnosis 1 Defines the parameter for the respective diagnostic value.
C1.5.5 Diagnosis 2
C1.5.6 Diagnosis 3
Select: off (goes to zero) / sensor average (sensor amplitude A+B) / sensor deviation / drive level / MT frequency / strain MT / strain IC / 2 phase signal
C1.6 Information
C1.6 Information
C1.6.2 V No. Sensor Shows the order number of the measuring sensor.
C1.6.3 SE Serial No. Displays the serial number of the sensor electronics
C1.6.4 SE Version Displays the version of the sensor electronics
C1.6.5 SE Interface Displays the interface version of the sensor electronics
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
83
Page 84
6 OPERATION
C1.7 Factory Calib.
C1.7 Factory Calib.
C1.7.1 Sensor Type Displays the sensor type
C1.7.2 Sensor Size Displays the nominal size of the sensor
C1.7.3 Sensor Material Displays the material of the sensor
C1.7.4 Max. Allowed Temp. Displays the maximum permitted temperature for the sensor
C1.7.5 Min. Allowed Temp. Displays the minimum permitted temperature for the sensor
C1.7.6...
1.7.30
C1.8 Simulation
C1.8 Simulation
C1.8.1 Mass Flow As in B1.1
C1.8.2 Density As in B1.2
C1.8.3 Temperature As in B1.3
CF1...CF27 Displays the sensor calibration coefficients (not CF9 or CF10)
MFC 300
No. Function Settings / descriptions
C2 Concentration
C2 Concentration See supplementary concentration manual
C3 I/O (Inputs/Outputs)
C3.1 Hardware
C3.1 Hardware Assignment of connection terminals dependent on signal converter version:
C3.1.1 Terminal A Select: Off (switched off) / Current Output / Frequency Output /
C3.1.2 Terminal B Select: Off (switched off) / Current Output / Frequency Output /
C3.1.3 Terminal C Select: Off (switched off) / Current Output / Status Output / Limit Switch
C3.1.4 Terminal D Select: Off (switched off) / Frequency Output / Pulse Output / Status Output /
active / passive / NAMUR
Pulse Output / Status Output / Limit Switch / Control Input
Pulse Output / Status Output / Limit Switch / Control Input
Limit Switch
84
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 85
MFC 300
C3. Current Output X
C3. Current Output X X stands for one of the connection terminals A, B or C
C3..1 Range 0%100%
C3..2 Extended Range Defines the min. and max. limits.
C3..3 Error Current Specify error current.
C3..4 Error Condition The following error conditions can be selected.
C3..5 Measurement Measurements for activating the output.
C3..6 Range 0100% of the measurement set in Fct. C3..5
C3..7 Polarity Set polarity, please note Flow Direction in C1.3.1!
C3..8 Limitation Limitation before applying the time constant.
C3..9 Low Flow Cutoff Sets output value to "0"
C3..10 Time Constant Range: 000.1100 s
C3..11 Special Function Automatic Range; select:
stands for Fct. no. C3.2 (A) / C3.3 (B) / C3.4 (C)
®
HART
current output: 4…20 mA
Current range for the selected measurement, e.g. 4...20 mA, corresponds to 0…100%
Note: with a 0...20 mA current output, HART in Fct. C6.8.1 must be shut off! xx.x xx.x mA; range: 0.0020 mA
(condition: 0 mA 1st value 2nd value ≤ 20 mA)
xx.x xx.x mA; range: 03.521.5 mA (condition: 0 mA 1st value 2nd value 21.5 mA)
xx.x mA; range: 322 mA (condition: outside of extended range)
Select: error in device (error category [F]) / application error (error category [F]) / out of specification (error category [S])
Select measurement: Volume Flow / Mass Flow / Temperature / Density / Velocity / Diagnosis 1 / Diagnosis 2 / Depending on the settings for the concentration measurement, the following measurements are possible: Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 / Concentration Flow 2
0xx.xx _ _ _ (format and unit depend on the measurement, see above)
Select: both polarities (plus and minus values are displayed) / positive polarity (display for negative values = 0) / negative polarity (display for positive values = 0) / absolute value (use for the output)
±xxx …±xxx%; range: -150…+150%
x.xxx ± x.xxx%; range: 0.020%
(1st value = switching point / 2nd value = hysteresis), condition: 2nd value 1st value
Off (switched off)
Automatic range (range is changed automatically, extended lower range, only makes sense together with a status output)
External range (change by control input, extended lower range, control input must also be activated)
OPERATION 6
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
85
Page 86
6 OPERATION
MFC 300
C3..12 Threshold Appears only when Fct. C3..11 threshold is activated between extended
C3..13 Information Serial no. of the I/O board, software version no. and production date of the
C3..14 Simulation Sequence see B1. Current Output X
C3..15 4mA Trimming Trimming of the current at 4 mA
C3..16 20mA Trimming Trimming of the current at 20 mA
and normal range. The automatic range function always changes from the extended to the normal range when the 100% current is reached.
The upper 100% value of the hysteresis is then = 0. The threshold is then the hysteresis value, instead of "threshold ± hysteresis" as shown in the display.
Range: 5.080%
(1st value = switching point / 2nd value = hysteresis), condition: 2nd value 1st value
circuit board
Reset to 4 mA restores the factory calibration.
Used for HART® setting.
Reset to 20 mA restores the factory calibration.
Used for HART® setting.
C3. Frequency Output X
C3. Frequency Output X X stands for one of the connection terminals A, B or D
C3..1 Pulse Shape Specify the pulse shape.
C3..2 Pulse Width Only available if set to "fixed" in Fct. C3..1.
C3..3 100% Pulse Rate Pulse rate for 100% of the measuring range
C3..4 Measurement Measurements for activating the output.
C3..5 Range 0100% of the measurement set in Fct. C3..4
C3..6 Polarity Set measured value polarity, please note flow direction in C1.3.2!
C3..7 Limitation Limitation before applying the time constant.
stands for Fct. no. C3.2 (A) / C3.3 (B) / C3.5 (D)
Select: symmetric (about 50% on and 50% off) / automatic (constant pulse with about 50% on and 50% off at 100% pulse rate) / fixed (fixed pulse rate, setting see below Fct. C3..3 100% pulse rate)
Range: 0.05…2000 ms Note: max. setting value Tp [ms] 500 / max. pulse rate [1/s], gives the pulse
width = time where the output is activated
Range: 0.010000 1/s Limitation 100% pulse rate 100/s: I
Limitation 100% pulse rate > 100/s: I
Select measurement: Volume Flow / Mass Flow / Temperature / Density / Velocity / Diagnosis 1 / Diagnosis 2 / Depending on the settings for the concentration measurement, the following measurements are possible: Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 / Concentration Flow 2
0xx.xx _ _ _ (format and unit depend on the measurement, see above)
Select: both polarities (plus and minus values are displayed) / positive polarity (display for negative values = 0) / negative polarity (display for positive values = 0) / absolute value (use for the output)
±xxx …±xxx%; range: -150…+150%
max max
100 mA20 mA
86
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 87
MFC 300
C3..8 Low Flow Cutoff Sets output value to "0":
x.xxx ± x.xxx%; range: 0.020%
(1st value = switching point / 2nd value = hysteresis), condition: 2nd value 1st value
C3..9 Time Constant Range: 000.1100 s
C3..10 Invert Signal Select: Off (activated output generates a high current at the output, switch
closed) / On (activated output generates a low current at the output, switch open)
C3..11 Phase Shift w.r.t. B Only available when configuring the A or D terminal and only if output B is a
pulse or frequency output. If setting in Fct. C2.5.6 is "Both Polarities", the phase shift is prefixed by a symbol, e.g. -90° and +90°.
Select: off (no phase shift) / 0° phase shift (between outputs A or D and B, inversion possible) / 90° phase shift (between outputs A or D and B, inversion possible) / 180° phase shift (between outputs A or D and B, inversion possible)
C3.3.11 Special Functions This function is only available at the terminal B frequency output. At the
same time, 2 frequency outputs must be available: 1st output at terminal A or D / 2nd output at terminal B
The B output is operated as a slave output, controlled and set using master output A or D
Selection: off (no phase shift) / phase shift w.r.t. D or A (slave output is B and master output is D or A)
C3..12 Information Serial no. of the I/O board, software version no. and production date of the
circuit board
C3..13 Simulation Sequence see B1. Frequency Output X
OPERATION 6
C3. Pulse Output X
C3. Pulse Output X X stands for one of the connection terminals A, B or D
C3..1 Pulse Shape Specify the pulse shape.
C3..2 Pulse Width Only available if set to "fixed" in Fct. C3..1
C3..3 Max. Pulse Rate Pulse rate for 100% of the measuring range
C3..4 Measurement Measurements for activating the output.
C3..5 Pulse Value Unit Selection of the unit from a list, depending on the measurement.
C3..6 Value p. Pulse Set value for volume or mass per pulse.
C3..7 Polarity Set polarity, please note Flow Direction in C1.3.2!
stands for Fct. no. C3.2 (A) / C3.3 (B) / C3.5 (D)
Select: Symmetric (about 50% on and 50% off) / Automatic (constant pulse with about 50% on and 50% off at 100% pulse rate) / Fixed (fixed pulse rate, setting see below Fct. C3..3 100% pulse rate)
Range: 0.05…2000 ms Note: max. setting value Tp [ms] 500 / max. pulse rate [1/s], gives the pulse
width = time where the output is activated
Range: 0.0…10000 1/s Limitation 100% pulse rate 100/s: I
Limitation 100% pulse rate > 100/s: I
Select: Volume Flow / Mass Flow
xxx.xxx, range in [l] or [kg] (volume or mass for current output C3..6)
At max. pulse rate see above 3..3 Pulse Output.
Select: both polarities (plus and minus values are displayed) / positive polarity (display for negative values = 0) / negative polarity (display for positive values = 0) / absolute value (use for the output)
max max
100 mA20 mA
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
87
Page 88
6 OPERATION
C3..8 Low Flow Cutoff Sets output value to "0"
x.xxx ± x.xxx%; range: 0.020%
(1st value = switching point / 2nd value = hysteresis), condition: 2nd value 1st value
C3..9 Time Constant Range: 000.1100 s
C3..10 Invert Signal Select: Off (activated output generates a high current at the output, switch
closed) / On (activated output generates a low current at the output, switch open)
C3..11 Phase Shift w.r.t. B Only available when configuring the A or D terminal and only if output B is a
pulse or frequency output. If setting in Fct. C2.5.6 is "Both Polarities", the phase shift is prefixed by a symbol, e.g. -90° and +90°.
Select: off (no phase shift) / 0° phase shift (between outputs A or D and B, inversion possible) / 90° phase shift (between outputs A or D and B, inversion possible) / 180° phase shift (between outputs A or D and B, inversion possible)
C3.3.11 Special Functions This function is only available at the pulse output of terminal B. At the same
time, 2 pulse outputs must be available: 1st output at terminal A or D / 2nd output at terminal B
The B output is operated as a slave output, controlled and set using master output A or D
Selection: off (no phase shift) / phase shift w.r.t. D or A (slave output is B and master output is D or A)
C3..12 Information Serial no. of the I/O board, software version no. and production date of the
circuit board
C3..13 Simulation Sequence see B1. pulse output X
MFC 300
88
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 89
MFC 300
OPERATION 6
C3. Status Output X
C3. Status Output X X (Y) stands for one of the connection terminals A, B, C or D
C3..1 Mode The output shows the following measuring conditions:
C3..2 Current Output Y Only appears if output A...C is set under "mode" (see above), and this output
C3..2 Frequency Output Y and
Pulse Output Y
C3..2 Status Output Y Only appears if output A...D is set under "mode" (see above), and this output
C3..2 Limit Switch Y and Control
Input Y
C3..2 Off Only appears if output A...D is set under "mode" (see above) and this output
C3..3 Invert Signal Off (activated output supplies a high current, switch closed)
C3..4 Information Serial no. of the I/O board, software version no. and production date of the
C3..5 Simulation Sequence see B1. status output X
stands for Fct. no. C3.2 (A) / C3.3 (B) / C3.4 (C) / C3.5 (D)
out of specification (output activated, signals application error or error in device refer to application error (output activated, signals application error or error in device refer to polarity flow (polarity of the current flow) / over range flow (over range of the flow) / totaliser 1 preset (activates when totaliser X preset value is reached) / totaliser 2 preset (activates when totaliser X preset value is reached) / totaliser 3 preset (activates when totaliser X preset value is reached) / output A (activated by the status of output Y, additional output data see below) / output B (activated by the status of output Y, additional output data see below) / output C (activated by the status of output Y, additional output data see below) / output D (activated by the status of output Y, additional output data see below) / off (switched off) / empty pipe (when pipe empty, output activated) / Error in device (when error, output activated)
is a "current output".
Polarity (is signalled)
Over Range (is signalled)
Automatic Range signals lower range
Only appears if output A, B or D is set under "mode" (see above), and this output is a "frequency/pulse output".
Polarity (is signalled)
Over Range (is signalled)
is a "status output".
Same signal (like other connected status output, signal can be inverted, see below)
Only appears if output A...D / input A or B is set under "mode" (see above), and this output / input is a "limit switch / control input".
Status off (is always selected here if status output X is connected with a limit switch / control input Y.
is switched off.
On (activated output supplies a low current, switch open)
circuit board
Status messages and diagnostic information
Status messages and diagnostic information
on page 113 /
on page 113 /
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
89
Page 90
6 OPERATION
MFC 300
C3. Limit Switch X
C3. Limit Switch X X stands for one of the connection terminals A, B, C or D
C3..1 Measurement Select: Volume Flow / Mass Flow / Diagnosis 1...3 / Flow Velocity /
C3..2 Threshold Switching level, set threshold with hysteresis
C3..3 Polarity Set polarity, please note Flow Direction in C1.3.2!
C3..4 Time Constant Range: 000.1100 s
C3..5 Invert Signal Select:
C3..6 Information Serial no. of the I/O board, software version no. and production date of the
C3..7 Simulation Sequence see B1. limit switch X
stands for Fct. no. C3.2 (A) / C3.3 (B) / C3.4 (C) / C3.5 (D)
Temperature / Concentration Flow 1 / Density
xxx.x ±x.xxx (format and unit depend on the measurement, see above)
(1st value = threshold / 2nd value = hysteresis), condition: 2nd value 1st value
Select: both polarities (plus and minus values are displayed) / positive polarity (display for negative values = 0) / negative polarity (display for positive values = 0) / absolute value (use for the output)
Off (activated output generates a high current, switch closed)
On (activated output generates a low current, switch open)
circuit board
C3. Control Input X
C3. Control Input X
C3..1 Mode X stands for connection terminal A or B
C3..2 Invert Signal off (control input is activated when a current is applied at the input by voltage
C3..3 Information Serial no. of the I/O board, software version no. and production date of the
C3..4 Simulation Sequence see B 1. control input X
stands for Fct. no. C3.2 (A) / C3.3 (B)
Off (control input switched off) / hold all outputs (hold current values, not display and totalisers) / output Y (hold current values) / all outputs to zero (current values = 0%, not display and totalisers) / output Y to zero (current value = 0%) / all totalisers (reset all totalisers to "0") / totaliser "Z" reset (set totaliser 1, (2 or 3) to "0") / stop all totalisers / stop totaliser "Z" (stops totaliser 1, (2 or 3) / zero outp.+stop Tot. (all outputs 0%, stop all totalisers, not the display) / external range Y (control input for external range of current output Y) - also make this setting on current output Y (no check if current output Y is available) / Error reset (all resettable errors are deleted) Zero Calibration
to passive inputs or a low-value resistor to active inputs)
on (control input is activated when no current is applied at the input, low voltage to passive inputs or a high-value resistor to active inputs)
circuit board
90
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 91
MFC 300
OPERATION 6
No. Function Settings / descriptions
C4 I/O Totalisers
C4.1 Totaliser 1 Set function of totaliser
C4.2 Totaliser 2
C4.3 Totaliser 3
C4..1 Totaliser Function Select: Absolute Total (counts positive and negative values) / +totaliser
C4..2 Measurement Selection of the measurement for totaliser
C4..3 Low Flow Cutoff Sets output value to "0"
C4..4 Time Constant Range: 000.1100 s
C4..5 Preset Value If this value is reached, positive or negative, a signal is generated that can be
C4..6 Reset Totaliser Sequence see Fct. A3.2, A3.3 und A3.4
C4..7 Set Totaliser Set totaliser to the desired value.
C4..8 Stop Totaliser Totaliser stops and holds the current value.
C4..9 Start Totaliser Start totaliser after that totaliser is stopped.
C4..10 Information Serial number of the I/O circuit board, software version number and
stands for 1, 2, 3 (= totaliser 1, 2, 3) The basic version (standard) has only 2 totaliser!
(counts only the positive values) / -totaliser (counts only the negative values) / off (totaliser is switched off)
Select: volume flow / mass flow / conc. flow 1 (depends on the settings for the concentration measurement)
Range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis), condition: 2nd value 1st value
used for a status output at which "preset totaliser X" has to be set.
Preset value (max. 8 digits) x.xxxxx in selected unit, see C6.7.10 + 13
Select: break (exit function) / set value (opens the editor to make the entry)
Query: set totaliser?
Select: no (exit function without setting the value) / yes (sets the totaliser and exits the function)
Select: no (exits the function without stopping the totaliser) / yes (stops the totaliser and exits the function)
Select: no (exits the function without starting the totaliser) / yes (starts the totaliser and exits the function)
production date of the circuit board.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
91
Page 92
6 OPERATION
No. Function Settings / descriptions
C5 I/O HART
C5 I/O HART
C5.1 PV is Current output (primary variable)
C5.2 SV is (secondary variable)
C5.3 TV is (tertiary variable)
C5.4 4V is (4th variable)
C5.5 HART Units Changes units of DVs (dynamic variables) in the display
Selection / display of the 4 dynamic variables (DV) for HART®.
The HART® current output (terminal A basic I/Os or terminal C modular I/Os) always has a fixed link to the primary variables (PV). Fixed links of the other DVs (1-3) are only possible if additional analogue outputs (current and frequency) are available; if not, the measurement can be freely selected from the following list: in Fct. A4.1 "measurement".
stands for 1, 2, 3 or 4 X stands for connection terminals A...D
Break: return with ^ key
MFC 300
HART display®: copies the settings for the display units to the settings for DVs
Standard: sets factory defaults for DVs
C5..1 Current Output X Shows the current analogue measured value of the linked current output.
C5..1 Frequency Output X Shows the current analogue measured value of the linked frequency output,
C5..1 HART Dynamic Var.
The measurement cannot be changed!
if present. The measurement cannot be changed!
Measurements of the dynamic variables for HART®.
Linear measurements: Volume Flow / Mass Flow / Diagnosis Value / Fow Velocity
Digital measurements: Totaliser 1 / Totaliser 2 / Totaliser 3 / Operating hours
No. Function Settings / descriptions
C6 Device
C6.1 Device Info
C6.1 Device Info
C6.1.1 Tag Settable characters (max. 8 digits):
C6.1.2 C Number CG number, non-alterable (I/O version)
C6.1.3 Device Serial No. Serial no. of the system, cannot be changed
C6.1.4 BE Serial No. Serial no. of the electronic assembly, cannot be changed.
C6.1.5 SW.REV.MS Serial no. of the circuit board, version no. of the main software, production
C6.1.6 Electronic Revision ER Shows ID no., electronic revision no. and production date;
AZ; az; 09; / - , .
date of the circuit board
Contains all hardware and software changes.
92
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 93
MFC 300
OPERATION 6
C6.2 Display
C6.2 Display
C6.2.1 Language Language selection depends on the device version.
C6.2.2 Contrast Adjust display contrast for extreme temperatures.
C6.2.3 Default Display Specification of the default display page that is returned to after a short
C6.2.4 Self Test Not available at this time.
C6.2.5 SW.REV.UIS Serial no. of the circuit board, version no. of user software, production date
Setting: -9…0…+9
This change takes place immediately, not just when setting mode is exited!
delay period.
Select: none (the current page is always active) / 1. meas. page (show this page) / 2. meas. page (show this page) / status page (show only status messages) / graphic page (trend display of the 1st measurement)
of the circuit board
C6.3 and C6.4 1st Meas. Page and 2nd Meas. Page
C6.3 1st Meas. Page stands for 3 = meas. page 1 and 4 = meas. page 2
C6.4 2nd Meas. Page
C6..1 Function Specify number of measured value lines (font size)
Select: one line / two lines / three lines
C6..2 1st Line Variable Specify 1st Line Variable.
Select measurement: Volume Flow / Mass Flow / Temperature / Density / Velocity / Diagnosis 1 / Diagnosis 2 / Depending on the settings for the concentration measurement, the following measurements are possible: Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 / Concentration Flow 2
C6..3 Range 0100% of the measurement set in Fct. C5..2
0xx.xx _ _ _ (format and unit depend on the measurement)
C6..4 Limitation Limitation before applying the time constant.
xxx%; Range: -150…+150%
C6..5 Low Flow Cutoff Sets output to "0": x.xxx ± x.xxx %; range: 0.020%
(1st value = switching point / 2nd value = hysteresis), condition: 2nd value 1st value
C6..6 Time Constant Range: 000.1100 s
C6..7 1st Line Format Specify decimal places.
Select: automatic (adaptation is automatic) / X(=none)X.XXXXXXXX (max. 8 digits) depends on size of font
C6..8 2nd Line Variable Specify 2nd Line Variable (only available if this 2nd line is activated)
Select: Bar Graph (for measurement selected in the 1st line ) / Volume Flow / Mass Flow / Temperature / Density / Velocity / Bar Graph / Totaliser 1 / Totaliser 2 / Totaliser 3 / Operating Hours / Diagnosis 1 / Diagnosis 2 Depending on the settings for the concentration measurement, the following measurements are possible: Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 / Concentration Flow 2
C6..9 2nd Line Format Specify decimal places.
Select: automatic (adaptation is automatic) / X(=none)X.XXXXXXXX (max. 8 digits) depends on size of font
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
93
Page 94
6 OPERATION
C6..10 3rd Line Variable Specify 3rd Line Variable (only available if this 3rd line is activated)
Select: Volume Flow / Mass Flow / Temperature / Density / Velocity / Totaliser 1 / Totaliser 2 / Totaliser 3 / Operating Hours / Diagnosis 1 / Diagnosis 2 Depending on the settings for the concentration measurement, the following measurements are possible: Diagnosis 3 / Concentration 1 / Concentration 2 / Concentration Flow 1 / Concentration Flow 2
C6..11 3rd Line Format Specify decimal places.
Select: automatic (adaptation is automatic) / X(=none)X.XXXXXXXX (max. 8 digits) depends on size of font
MFC 300
C6.5 Graphic Page
C6.5 Graphic Page
C6.5.1 Select Range Graphic page always shows trend curve of the measurement of the 1st page
C6.5.2 Range Set the scaling for the Y axis. Only available if "manual" is set in C6.5.1.
C6.5.3 Time Scale Set the time scaling for the X axis, trend curve
/ 1st line, see Fct. C6.3.2
Select: Manual (set range in Fct. C6.5.2) / Automatic (automatic depiction based on the measured values) Reset only after parameter change or after switching off and on.
+xxx ±xxx%; range: -100+100%
(1st value = lower limit / 2nd value = upper limit), condition: 1st value 2nd value
xxx min; range: 0…100 min
C6.6 Special Functions
C6.6 Special Functions
C6.6.1 Reset Errors Reset Errors?
Select: no/yes
C6.6.2 Save Settings Save current settings.
C6.6.3 Load Settings Load saved settings.
C6.6.4 Password Quick Set Password required to change data in the quick setup menu.
C6.6.5 Password Setup Password required to change data in the setup menu
Select: cancel (exit function without saving) / backup 1 (save in storage location 1) / backup 2 (save in storage location 2).
Query: continue copy? (cannot be done afterwards) Select: no (exit function without saving) / yes (copy current settings to storage backup 1 or backup 2)
Select: break (exit function without loading) / factory settings (load in state as delivered) / backup 1 (load data from storage location 1) / backup 2 (load data from storage location 2)
Query: continue copy? (cannot be done afterwards) Select: no (exit the function without saving) / yes (load data from the selected storage location)
0000 (= to Quick Setup menu without password) xxxx (password required); range 4 digits: 0001…9999
0000 (= to Quick Setup menu without password) xxxx (password required); range 4 digits: 0001…9999
94
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 95
MFC 300
OPERATION 6
C6.6.6 GDC IR Interface After this function has been activated an optical GDC adapter can be
connected to the LC display. If approximately 60 seconds pass without a connection being established or after the adapter is removed, then the function is exited and the optical keys are active once again.
Break (exit function without connection)
Activate (the IR interface adapter and interrupt the optical keys)
If approximately 60 seconds pass without a connection being established, then the function is exited and the optical keys are active once again.
C6.7 Units
C6.7 Units C6.7.1 Volume Flow m³/h; m³/min; m³/s; l/h; l/min; l/s (l = litres); IG/s; IG/min; IG/h
C6.7.2 Text free unit For text to be specified refer to C6.7.3 [m³/s]*factor Specification of the conversion factor, based on m³/s:
C6.7.4 Mass Flow kg/s; kg/min; kg/h; t/min; t/h; g/s; g/min; g/h;
C6.7.5 Text free unit For text to be specified refer to
C6.7.6 [kg/s]*factor Specification of the conversion factor, based on kg/s:
C6.7.7 Velocity m/s; ft/s C6.7.9 Temperature °C; K; °F C6.7.10 volume m³; l (litres); hl; ml; gal; IG; in³; ft³; yd³; barrel
C6.7.11 Text free unit For text to be specified refer to C6.7.12 [m³]*factor Specification of the conversion factor, based on m³:
C6.7.13 Mass kg; t; mg; g; lb; ST; LT; oz;
C6.7.14 Text free unit For text to be specified refer to
C6.7.15 [kg]*factor Specification of the conversion factor, based on kg:
C6.7.16 Density kg/l; kg/m³; lb/ft³; lb/gal; SG
C6.7.17 Text free unit For text to be specified refer to C6.7.18 [kg/m³]*factor Specification of the conversion factor, based on kg/m³:
C6.7.19 Pressure Pa; kPa; bar; mbar; psi (no free units possible);
ft³/h; ft³/min; ft³/s; gal/h; gal/min; gal/s; barrel/h; barrel/day free unit (set factor and text in the next two functions, sequence see below)
Set free units
xxx.xxx refer to
lb/s; lb/min; lb/h; ST/min; ST/h (ST = Short Ton); LT/h (LT = Long Ton); free unit (set factor and text in the next two functions, sequence see below)
Set free units
on page 96
Set free units
xxx.xxx refer to
free unit (set factor and text in the next two functions, sequence see below)
Set free units
on page 96
Set free units
xxx.xxx refer to
free unit (set factor and text in the next two functions, sequence see below)
Set free units
on page 96
Set free units
xxx.xxx refer to
free unit (set factor and text in the next two functions, sequence see below)
Set free units
on page 96
Set free units
xxx.xxx refer to
only if current input available.
Set free units
on page 96
on page 96:
on page 96:
on page 96:
on page 96:
on page 96:
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
95
Page 96
6 OPERATION
C6.8 HART
C6.8 HART
C6.8.1 HART
C6.8.2 Address
C6.8.3 Message Set required text:
C6.8.4 Description Set required text:
C6.9 Quick Setup
C6.9 Quick setup Activate quick access in Quick Setup menu:
C6.9.1 Reset Totaliser 1 Reset Totaliser 1 in Quick Setup menu?
C6.9.2 Reset Totaliser 2 Reset Totaliser 2 in Quick Setup menu?
C6.9.3 Reset Totaliser 3 Reset Totaliser 3 in Quick Setup menu?
Switch HART® communication on/off:
Select: HART on (HART not activated) current = 0…20 mA
Set address for HART® operation:
Select: 00 (point-to-point operation, current output has normal function, current = 420 mA) / 01...15 (multidrop operation, current output has a constant setting of 4 mA)
AZ ; az ; 09 ; / - + , . *
AZ ; az ; 09 ; / - + , . *
Select: yes (switched on) / no (switched off)
Select: yes (activated) / no (switched off)
Select: yes (activated) / no (switched off)
Select: yes (activated) / no (switched off)
®
activated) current = 420 mA / HART off (HART
MFC 300
®
6.4.4 Set free units
Free units Sequences to set texts and factors
Texts
Texts
TextsTexts
Volume flow, mass flow and density: 3 digits before and after the slash
Volume, mass: xxx (max. 3 digits) Permissible characters: AZ ; az ; 09 ; / - + , .*; @ $ % ~ () [] _
Conversion factors
Conversion factors
Conversion factorsConversion factors
Desired unit = [unit see above] * conversion factor
Conversion factor Max. 9 digits Shift decimal point: ↑ to the left and ↓ to the right
xxx/xxx (max. 3 digits before / after the slash)
96
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 97
MFC 300
6.5 Description of functions
6.5.1 Reset counter in the menu "quick setup"
INFORMATION!
It may be necessary to activate resetting of the counter in the menu "quick setup".
Key Display Description and setting
> quick setup Press and hold for 2.5 s, then release the key.
> language ­2 x reset -
> reset errors -
counter 1 Select desired counter. counter 2 counter 3
> reset counter
no
or reset counter
yes
^ counter 1,2 (or 3) Counter has been reset.
3 x ^ Measuring mode -
(Counter 3 is optional)
-
-
OPERATION 6
6.5.2 Deleting error messages in the menu "quick setup"
INFORMATION!
For the detailed list of the possible error messages refer to Status messages and diagnostic information on page 113
Key Display Description and setting
> quick setup Press and hold for 2.5 s, then
> language ­2 x reset -
> reset errors -
> reset?
no
or reset?
yes
^ reset errors Error has been reset.
3 x ^ Measuring mode -
.
release the key.
-
-
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
97
Page 98
6 OPERATION
6.5.3 Mode (menu A8)
The device can be switched to "standby". In this mode, all flow values are set to zero and the counter values are "frozen". The temperature and density values are displayed normally and output via the outputs. The "standby" indicator on the display shows either the "frozen" counter value or simply "standby". In this mode, the measuring tubes continue to vibrate and the device can immediately return to "measuring" mode if necessary.
There is also a "stop" mode. In this mode, the sensor is switched off and is no longer vibrating. To return to measuring mode from this mode, the measuring device must complete "start-up" phase again before measurement continues.
The measuring device can be switched into "standby" mode either using the optical keys on the display or using the control input. You can only switch to "stop" mode using the optical keys.
Setting the mode (from measuring mode):
Key Display Description and setting
MFC 300
must go through the
mustmust
> A Quick Setup Press and hold for 2.5 s, then release
> A8 Mode
Measuring
> Mode
Measuring
Mode
Standby
Mode
Stop
3 x ^ Save Configuration?
Yes
^ Display page
the key.
If "standby" or "stop" is selected, the device immediately switches to this operating state. To return to the measuring mode, go to menu A8 and select "measure".
INFORMATION!
When changing from the "stop" to "standby" mode, the device goes through the entire start-up phase.
In addition to the "standby" mode, the system control function also allows you to switch to a comparable "standby" status completely automatically depending on the current operating temperature or density.
98
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Page 99
MFC 300
6.5.4 Density calibration (menu C1.2.1)
Mass flowmeters are calibrated for density in the factory. Density calibration is based on 2 calibration points. In the factory, air and water are used under reference conditions. The result of this calibration is saved in the converter electronics and stored in the factory settings. Regardless, various applications require maximum accuracy which can only be achieved with on-site calibration.
Available options:
Option Explanation
OPERATION 6
1 Point Calibration One of the 2 saved calibration points is replaced by customer calibration.
2 Point Calibration The user recalibrates both calibration points.
Default The signal converter reactivates the factory settings for the density
Manual The user can read the current density calibration points and enter them if
The signal converter decides which of the 2 calibration points is modified.
calibration.
necessary (according to sensor calibration certificate).
Example of 1 point calibration with town water
Key Display Description and setting
> A Quick Setup Press and hold for 2.5 s, then release
2 x C Setup
2 x > C1.1 Calibration C1.2 Density
2 x > C1.2.1 Density Calib.
C1.2.1 Density Calibration?
C1.2.1 Density Calibration?
C1.2.1 Density Calibration?
C1.2.1 Density Calibration?
^ C1.2.1 DCF1
Press until
^ C1.2.1 Single Pt. Calib.
C1.2.1 Single Pt. Calib.
^ C1.2.1 Perform Calibration
5 x ^ Save Configuration?
^ Measuring Mode
C1.2.1 DCF1
Break
Default
Manual
2 Point Calibration
1 Point Calibration
XXXXXXXXX
Town Water
Break
OK
Passed
Yes
the key.
Press ^ to exit density calibration menu.
Press ^ for density calibration. Factory calibration on OK (6 x ^).
Press ^ to read or change the current density calibration value.
Press ^ to start 2 point calibration.
Press ^ to start 1 point calibration.
Press until you reach town water.
Press ^ to start town water calibration.
Press ^ to start single point calibration.
www.krohne.com02/2012 - 4000498602 - MA MFC 300 R03 en
99
Page 100
6 OPERATION
On-site density calibration:
Ensure that the device has been properly installed and functions smoothly.
If air (empty) is used as the medium, the measuring tube must be completely dry and free of
liquids and solids. If possible, blow dry air into the measuring tube to empty it.
If liquids are used, flush for a few minutes at a high flow rate to remove gas bubbles.
Set the flow rate to a typical value (50% of the nominal flow is ideal).
If the process temperature is higher than the ambient temperature, wait until the system has
stabilised.
When it comes to 1 point and 2 point calibration, you can choose from the options "empty", "pure water", "town water" and "other". The reference values for the specified products are stored in the converter.
Density calibration has failed if "calib. error." is displayed. There are several possible reasons for this:
The device is not in "measuring" mode.
The calibration points are too close together.
One or more calibration points did not pass the plausibility test.
Flow, pressure, temperature or system are not stable?
Please check your system and try again.
If this results in another failed calibration, please contact the manufacturer.
MFC 300
100
www.krohne.com 02/2012 - 4000498602 - MA MFC 300 R03 en
Loading...