ALTOFLUX electromagnetic flowmeters are precision measuring instruments designed for the
linear flow measurement of process liquids.
The process liquids must be electrically conductive: ≥ 20 µS/cm
The full-scale rangeQ
ALTOFLUX M900:
DN 10 - 300 /
ALTOFLUX 3080 K:
DN 10 - 300 /
This is equivalent to a flow velocity of 0.3 - 12 m/s, or 1 – 40 ft/s.
3
/8” - 12” Q
3
/8” - 12” Q
can be set as a function of the meter size:
100%
ALTOFLUX electromagnetic flowmeters are designed solely for measuring the volumetric
flowrate of electrically conductive, liquid process products.
Flowmeters with ALTOFLUX primary heads are not certified for use in hazardous locations.
Other flowmeters series are available for such applications.
Responsibility as to suitability and intended use of these electromagnetic flowmeters rests solely
with the operator.
Improper installation and operation of the flowmeters (systems) may lead to loss of warranty.
In addition, the ”General conditions of sale” forming the basis of the purchase contract are
applicable.
If ALTOFLUX flowmeters need to be returned to Krohne, please note the information given on
the last-but-one page of this manual. Krohne regret that they cannot repair or check your
flowmeter(s) unless accompanied by the completed form sheet.
System description
= 0.1 - 3050 m3/hr = 0.37 – 13 440 US GPM
100%
= 0.1 - 3050 m3/hr = 0.37 – 13 440 US GPM
100%
Product liability and warranty
2 Installation and operating instruction ALTOFLUX M900 and 3080 K
Standards and approvals
Please refer to the installation and operating instructions for the signal converter.
Items included with supply
ALTOFLUX M900
primary heads
• Primary head in the size as ordered • Compact flowmeter in the size as ordered
• Connecting wires for grounding, refer to
Section 7 “Grounding”
• Certificate of calibration data • Certificate of calibration data
• Grounding rings (optional), if ordered • Grounding rings (optional), if ordered
Fitting accessories (stud bolts, nuts, gaskets, etc.) are not supplied with the flowmeter,
to be provided by customer!
ALTOFLUX 3080 K
compact flowmeters
• Connecting wires for grounding, see
Section 7 “Grounding”
• Installation and operating instructions for the
signal converter
Handling
Do not lift flowmeter by the signal converter
housing or the terminal box.
Do not set flowmeter down on signal converter
housing or terminal box.
PLEASE NOTE
the temperature limits for storage and transport, see Page 4.
Installation and operating instruction ALTOFLUX M900 and 3080 K 3
1 Important information for installation: PLEASE NOTE !
• Temperatures
Refer to Section 11 “Limits” for operating pressure and vacuum load based on flange
standards and type of tube liner.
Compact system 3080 K
ALTOFLUX M900 -25 to +60 °C (-13 to +140 °F)
In storage -20 to +60 °C (- 4 to +140 °F), (Neoprene liner)
avoid moisture and sunlight.
Transport - 5 to +50 °C (- 4 to +140 °F), (Neoprene liner)
avoid moisture and sunlight.
• Location and position as required,
but electrode axis
X – • – • – • – X
must be approximately horizontal in a
horizontal pipe run.
Y terminal box or converter housing
• Measuring tube must be completetly filled at all times.
• Direction of flow is arbitrary. Arrow on flowmeter can normally be ignored. For exceptions,
refer to Section “Factory settings” in the installation and operating instructions for the signal
converter.
• Stud bolts and nuts: to fit, make sure there is sufficient room next to the pipe flanges.
• Vibration: support the pipeline on both sides of the compact flowmeter. Level of vibration in
conformity with IEC 068-2-34: below 2.2g for compact flowmeters in the frequency range of
20-150 Hz with the IFC 090 K.
• Do not expose to direct sunlight, fit a sunshade if necessary, not included with flowmeter,
to be provided by customer.
• Large meter sizes (≥ DN 200 / ≥ 8”), use adapter pipes to allow axial shifting of the
counterflanges and to facilitate installation.
• Strong electromagnetic fields, avoid in vicinity of flowmeter
• Straight inlet run minimum of 5 × DN and outlet run minimum of 2 × DN,
(DN = meter size), measured from the electrode axis.
Ambient temperature Process temperature
-25 to +60 °C (-13 to +140 °F)
-25 to +40 °C (-13 to +104 °F)
-25 to ≤ +60 °C (-13 to ≤ +140 °F)
-25 to > +60 °C (-13 to ≤ +194 °F)
-25 to > +60 °C (-13 to > +140 °F)
-25 bis +60 °C (all other liners)
-25 bis +60 °C (all other liners)
4 Installation and operating instruction ALTOFLUX M900 and 3080 K
• Vortex and corkscrew flow: increase length of inlet and outlet runs or install flow
conditioners.
• Mixing different process liquids: install flowmeter upstream of mixing point or at an
adequate distance downstream (minimum of 30 × DN), otherwise display may be unsteady.
• Plastic pipes and internally coated metal pipelines: grounding rings required, see Section 7
“Grounding”.
• Insulated pipeline: do not insulate flowmeter.
• Zero setting not necessary. To check, it should be possible to set “zero” flow velocity in the
completely filled measuring tube. Shutoff valves should therefore be provided either
downstream of the flowmeter or upstream and downstream of the flowmeter.
2 Suggestions for installation
To avoid measuring errors due to gas/air inclusion or to pipe running empty, please observe the
following:
Highest point of pipe run Downpipe over 5 m (16 ft) length
(Air bubbles collect in measuring
tube - faulty measurements!) flowmeter.
Install air valve ✇ downstream of
Preferred
locations
Downpipe
“Zero” flow
velocity.
Line drained.
Faulty
measurements!
Long pipeline
Always install control and shutoff valves
downstream of flowmeter.
Horizontal pipe run
Install in slightly ascending pipe section. If
not possible, assure adequate velocity to
prevent air, gas or vapor from collecting in
uppper part of flow tube.
Pumps
Open feed or discharge Never install flowmeter on pump suction side.
Install meter in low section of pipe.
open discharge
Installation and operating instruction ALTOFLUX M900 and 3080 K 5
3 Instrument nameplate
ALTOFLUX M900
separate primary head
Series No.
Primary head
constant
Meter size
DN in mm and
inch equivalent
Liner
Hard rubber
Type designation
Electrode material
Hastelloy C4
Magnetic field frequency
(here 1/6 of power frequency)
ALTOFLUX M900
Flange pressure rating
or flange class
Tag No.
Protection category
to IEC 529/EN 60529
Insulation class
of field coils
Liner materials
Electrode materials
H Hard rubber HB Hastelloy B2
NE Neoprene HC Hastelloy C4
PUI Irathane PT Platinum
T Teflon®-PTFE TA Tantalum
W Soft rubber TI Titanium
V4A Stainless steel 1.4571 / SS 316-Ti
®
Teflon
is a registered trademark of Du Pont
Instrument nameplate for compact flowmeters
see installation and operating instructions for the signal converter.
4 Flowmeter versions
ALTOFLUX M900 Separate primary head (F) electrically connected to the signal
converter by signal and field current cables.
ALTOFLUX 3080 K Compact flowmeter (K), IFC 090 K signal converter mounted direct
on the primary head.
Versions for hazardous locations
M900 and 3080 K are approved as electrical equipment to the harmonized European Standards
and to Factory Mutual (FM).
Test certificate, certificate of conformity and wiring instructions for these devices are attached to
the ”Ex” installation instructions, provided only with hazardous-duty equipment.
6 Installation and operating instruction ALTOFLUX M900 and 3080 K