Ivoclar Vivadent SR Ivocap Equipment System User Manual

Page 1
SR Ivocap
®
Equipment System
Operating Instructions
Page 2
2
Page 3
Index
Views of the Furnace, List of Parts 4
1. Introduction / Signs and Symbols 5
1.3 Notes regarding the Operating Instructions
2. Safety First 6
2.1 Indications
3. Product Description 7
3.1 Components
3.2 Hazardous areas and safety equipment
3.3 Functional description
3.4 Accessories
4. Installation 8
4.1 Unpacking and checking the contents
4.2 Selecting the location
Page
.2 Signs and symbols
5. Operation 9
5.1 Operating elements at the pressure apparatus
5.2 Operating elements at the clamping frame
6. Procedure 10
6.1 Practical use
7. Maintenance, Cleaning, and Diagnosis 11
7.1 Monitoring and maintenance
7.2 Cleaning
7.3 Changing the Compression spring
8. What if…? 12
8.1 Error messages
8.2 Repairs
9. Product Specifications 13
9.1 Delivery form
9.2 Technical data
9.3 Acceptable operating conditions
9.4 Acceptable transportation and storage conditions
3
Page 4
Views of the Equipment System, List of Parts
A
A) Pressure apparatus
D) Capsule plunger
"ID3"
E) Investment aid
) Funnel
F
G) Timer
A2
A1
A1 = Locking valve A2 = Manometer A3 = Extrusion piston A4 = Vise shank
5 = Safety loop
A
B) Clamping frame “JS4”
B2 = Thrust collar with lever B3 = Pressing plate
C) Flask
C1 = Flask half M C2 = Flask half W C3 = Flask lid C4 = Injection funnel
A5
A3
B
A4
C
C2
C4
B1
B2
B3
C3
D
E
G
C1
F
4
Page 5
1. Introduction / Signs and Symbols
1.1 Introduction
The SR Ivocap System is an injection technique which has been especially
eveloped to compensate
d curing shrinkage. Heat/ pressure curing allows shrinkage of the acrylate during polymerization to be compensated by inflowing material.
The SR Ivocap injection technique, therefore, permits the fabrication of high quality products with excellent physical properties.
Indication
• Complete dentures
• Partial dentures
• Relines
• Orthodontic appliances
• Occlusal splints
Contraindication
Keep unpolymerized
material away from intraoral areas. Do not use the material if the patient is known to be allergic to any of the ingredients listed.
The SR Ivocap System
may only be
operated by trained
personnel.
1.2 Signs and symbols
The signs and symbols in these Operating Instructions facilitate the finding of
mportant points and have
i the following meaning:
Risks and dangers
Important information
Contraindication
1.3 Notes regarding the Operating Instructions
These Operating Instructions must be read prior to working with the SR Ivocap System.
Equipment concerned: SR Ivocap System Target
roup: Dental technologists
g
The Operating Instructions facilitate the correct and economic use of the SR Ivocap System.
These Operating Instructions are divided in several, clearly structured chapters. They should enable you to locate specific topics quickly and easily.
The SR Ivocap material is not described in these instruc­tions. Please refer to the corresponding Instructions for Use of the Ivocap material.
The inform your about risks/dangers, important information, and contraindi­cations, these Instructions contain corresponding signs/ symbols (pictographs) to mark important paragraphs.
We recommend keeping the Instructions in a safe place near the equipment to have immediate access to the information if necessary.
Should you lose the Operating Instructions, extra copies can be ordered at a nominal fee from your local Ivoclar Vivadent Service Center.
5
Page 6
2. Safety First
This chapter is especially important for personnel who work with the SR Ivocap System or who have to carry out maintenance or repair work. This chapter must be read and the corresponding
nstructions followed.
i
2.1 Indications
The SR Ivocap System must only be used for injecting the SR Ivocap material and it should be used for this purpose only. Other uses
han the ones stipulated, i.e. injecting other
t materials, etc. are contraindicated. The manufacturer does not assume any liability for damage resulting from misuse. The user is solely responsible for any risk resulting from failure to observe these Instructions.
2.1.1
Risk of crushing
Further instructions to assure proper use of the system: – The instructions,
regulations, and notes in these Operating
nstructions must be
I observed.
– The system must be
operated under the indicated environmental and operating conditions (Chapter 9).
– The SR Ivocap System
must be properly maintained (Chapter 7).
Do not reach between pressing plunger and safety loop with your fingers. There is a risk of crushing when air is compressed.
2.1.2
Burn hazard
2.1.3
Burn hazard
2.1.4
Use protective thermo-gloves to touch hot metal parts.
Do not use the water bath without the plastic floaters. Splash water presents a burn hazard.
Wear protective clothing to protect yourself from splash
Burn hazard
6
water.
Page 7
3. Product Description
3.1 Components
The SR Ivocap basic assortment consists of the following components:
Pressure apparatus
• Flask
• Clamping frame
• Capsule plunger
• Investment aid
• Funnel
• Compression spring
• Allen key
urthermore, the following accessories
F are required (not contained in the assortment):
• Cap Vibrator (Ivoclar Vivadent)
• Polymerization bath (various manufacturers)
• Hydraulic press (various manufacturers)
• Plastic floaters (Ivoclar Vivadent)
3.2 Hazardous areas and safety equipment
Description of the risk areas of the
R Ivocap System:
S
Hazardous area
Flask and clamping frame
Splash water from the polymerization bath
See also Chapter 2.
3.3 Functional description
The SR Ivocap System is an injection technique which has been especially developed to compensate curing shrinkage. Controlled polymerization allows shrinkage of the material during polymerization to be compensated by inflowing material. The SR Ivocap injection technique, therefore, permits the fabrication of high quality products with outstanding physical properties.
Type of risk
There is a burn hazard if the equipment is removed from the boiling water without protective gloves.
There is a burn hazard if no protective clothing and gloves are worn.
3.4 Accessories
• Cap Vibrator (Ivoclar Vivadent)
• Wall bracket for 4 pressure apparatus (e.g. from BDT)
• Polymerization bath (e.g. from BDT)
• Hydraulic press (various manufacturers)
• Various accessories (BDT)
Address of BDT: BDT GmbH, Industriestrasse 27, D-77656 Offenburg (Germany)
7
Page 8
4. Installation
4.1 Unpacking and checking the contents
Remove system components from their
ackaging and place them on a suitable
p table.
Check the delivery for completeness (see delivery form in Chapter 9) and transportation damage. If certain parts are damaged or missing, contact your local Ivoclar Vivadent Customer Service. We recommend keeping the original packaging for future transportation purposes.
Wall bracket for pressure apparatus
6 bar pressure
4.2 Selecting the location
The diagram below depicts one possible installation.
mportant information
I
For safety reasons, the installation
of the pressure, power, and water connections may only be carried out by qualified technicians (electricians, plumbers).
Furthermore, the national installation requirements of the corresponding country must be observed.
Flexible pressure tubing
Timer
Cold water
Water outlet
Water level
Hot water supply
Quick-fitting pipe union
Air pressure 6 bar
Polymerization bath
Cap-Vibrator
Hydraulic press, 3 t or 80 bar
8
Page 9
5. Operation
5.1 Operating elements on the pressure apparatus
Compressed air locking valve (A1)
This locking valve is used to close or open the pressure cylinder for the compressed air.
closed
Operation:
Locking valve open: perpendicular position
Locking valve closed: horizontal position
open
Safety loop (A5)
The safety loop is used to secure the vise shanks.
Operation:
Safety loop open: Towards the pressure cylinder
Safety loop closed: Towards the clamping frame
5.2 Operating elements on the clamping frame
hrust collar with lever (B2)
T
B2
2
Operation:
1) Thrust collar with lever closed
(push to the right until it stops)
2) Thrust collar with lever open
(push to the left)
Pressure onto the clamping frame: 3 tons
B1
1
9
Page 10
6. Procedure
6.1 Practical use
Compressed air 6 bar
Water level
Pressure apparatus
Capsule
Flask
Mark for water level
The most important steps:
Injection - 5 minutes
1
Polymerization – 35 minutes in
2
boiling water (as of the beginning
f boiling)
o
Cooling in a cold-water bath –
3
at least 30 minutes; the injection pressure must be maintained for at least 20 minutes thereof
Flowing cold water
For information on how to process the material please refer to the material Instructions for Use of the SR Ivocap System.
10
Page 11
7. Maintenance, Cleaning, and Diagnosis
This chapter describes the maintenance and cleaning procedures which can be carried out by dental professionals. All other tasks must be performed by qualified service personnel at a certified Ivoclar Vivadent Service Center.
Disconnect power and compressed
ir connection before maintenance
a and cleaning.
The insulation shoulder of the flask must only be replaced by a certified Service Center, since remounting requires a special mounting assembly and a special material. Only the correct mounting procedure ensures sound function of the flask.
7.1 Monitoring and maintenance
he time for these maintenance procedures
T depends on the frequency of use and work­ing habits of the users. For that reason, the recommended times are only approximates.
What: Part: When:
Check all compressed air connections for correct fit
Check the components for mechanical damage
Check the plastic parts of the flask for correct fit
Clean contaminated clamping frame (wax), e.g. by boiling out
Compressed air connections
Pressure apparatus
Before use
Before use
Flask
Clamping frame
Plastic parts of the flask
clamping frame in case of
Before use
contamination
7.2 Cleaning
Metal parts can be cleaned with warm water and a cleaning brush. After cleaning, the flask may be treated with Vaseline to facilitate the removal of plaster residue.
7.3 Changing the Compression spring
Please take note of the informational sheet added to the 'spring washers' set.
The Compression spring are to be changed by a certified Ivoclar Vivadent Service Center.
11
Page 12
8. What if...?
This chapter will help you to recognize malfunctions and take appropriate measures or, if possible, to perform some repairs.
Error
Description Notes for users
Flasks do not close.
1
The flask covers do not sit on the flask margins.
The model to be invested does not fit in the flask.
2
The pressure is difficult to be transferred to the flask and/or clamping frame.
3
Pressure in the clamping frame fluctuates.
4
Injection plunger enters the capsule and deforms the bottom of the capsule.The material is
5
extruded at the wrong end of the capsule.
The rod of the injection plunger cannot be moved, even if the connection to the compressed
6
air source is interrupted. It seems to be stuck.
Injection plunger does not move.The SR Ivocap material is not injected.
7
The resin flows out of the flask during polymerization.
8
The locking mechanism is blocked.
9
8.1 Error messages
List of possible error messages and their meaning:
Flask margins are not clean;plaster residue on the injection funnel.
The model should be trimmed prior to investment.
Clamping frame is not correctly positioned, i.e. in the centre of the press.
Compression spring in the clamping frame are old or dirty and cannot transfer the pressure.
clamping frame (see Error 12).
The injection plunger is not correctly placed on the capsule. There is an empty space between the injection plunger and capsule bottom. Avoid empty space and place the injection plunger evenly on the capsule bottom.
Negative pressure in the pressure cylinder may prevent the plunger from moving, open valve
– No compressed air present – Capsule cover was not removed – There is plaster residue or the funnel is blocked, e.g. by plaster
residue – The injection channels are too narrow. – The flow properties of the resin are poor because of incorrect
storage
– Leaks as a result of incorrect investment. – Flask defective. – Inadequate clamping pressure on the frame. – A foreign object is stuck between the two flask halves.
The mechanical components are damaged as a result of improper handling. In case of considerable damage, the mechanical components must be replaced.
R
eplace the compression spring or clean the
The locking mechanism is contaminated with wax.
10
Capsule reservoir has polymerized after 35 minutes.
11
8.2 Repairs
Repairs may only be carried out by a certified Ivoclar Vivadent Service Center. Please refer to the addresses in Chapter 10.
If repairs during the warranty period are not carried out by a certified Ivoclar Vivadent Service Center, the warranty will expire immediately.
Please also read the safety information in Chapter 2.
Wax residue in the water bath has contaminated the clamping frame. Boil out clamping frame with clean, hot water.
– Water level surpasses the marking on the clamping frame. – An unsuitable polymerization bath was used. – The system was not placed in cold water immediately after
polymerization. – Reduce polymerization time by 1–2 minutes if necessary
12
Page 13
9. Product Specifications
This chapter contains all relevant product specifications.
9.1 Delivery form
1 Pressure apparatus 1 Clamping frame 2 Flasks 1 Timer 1 Compression spring 1 Capsule plunger 2 Investment aids
Allen key
1
9.2 Technical data
Weight:
Flask 1.3 kg Clamping frame JS4 2.3 kg
ressure apparatus ID3 2.0 kg
P Total 5.6 kg
Max. operating pressure:
Pressure apparatus 6.5 bar Clamping frame 3 tons
Noise emissions:
< 70 dB (A)
9.3 Acceptable operating conditions
Acceptable humidity range
aximum relative humidity at 31 °C
M (87 °F) gradually decreasing to 50% at 40 °C (104 °F), condensation excluded.
Acceptable ambient pressure
The unit is tested for altitudes of up to 2000 m above sea level.
9.4 Acceptable transporation and storage conditions
Acceptable temperature range
–20 °C to 50 °C (–4 °F to 122 °F)
Acceptable humidity range
Maximum relative humidity at 31 °C (87 °F) gradually decreasing to 50% at 40 °C (104 °F), condensation excluded.
Acceptable ambient pressure
500 mbar to 1060 mbar
Use only original packaging together with the respective foam material for shipping purposes.
13
Page 14
14
Page 15
15
Page 16
Ivoclar Vivadent – worldwide
Ivoclar Vivadent AG
Bendererstrasse 2
L-9494 Schaan
F
iechtenstein
L Tel. +423 235 35 35 Fax +423 235 33 60
ww.ivoclarvivadent.com
w
Ivoclar Vivadent Pty. Ltd.
1 – 5 Overseas Drive
.O. Box 367
P
oble Park, Vic. 3174
N Australia Tel. +61 3 979 595 99
ax +61 3 979 596 45
F
ww.ivoclarvivadent.com.au
w
Ivoclar Vivadent GmbH
remschlstr. 16
B
ostfach 223
P A-6706 Bürs Austria
el. +43 5552 624 49
T
ax +43 5552 675 15
F www.ivoclarvivadent.com
Ivoclar Vivadent Ltda.
Rua Geraldo Flausino Gomes, 78 – 6.º andar Cjs. 61/62
airro: Brooklin Novo
B CEP: 04575-060 São Paulo – SP
razil
B
el. +55 11 3466 0800
T Fax +55 11 3466 0840 www.ivoclarvivadent.com.br
Ivoclar Vivadent Inc.
2785 Skymark Avenue, Unit 1 Mississauga Ontario L4W 4Y3 Canada Tel. +1 905 238 5700 Fax +1 905 238 5711 www.ivoclarvivadent.us
Ivoclar Vivadent Marketing Ltd.
Rm 603 Kuen Yang International Business Plaza No. 798 Zhao Jia Bang Road Shanghai 200030 China Tel. +86 21 5456 0776 Fax +86 21 6445 1561 www.ivoclarvivadent.com
Ivoclar Vivadent Marketing Ltd.
Calle 134 No. 7-B-83, Of. 520
ogotá
B
olombia
C Tel. +57 1 627 33 99 Fax +57 1 633 16 63
ww.ivoclarvivadent.com
w
Ivoclar Vivadent SAS
B.P. 118
-74410 Saint-Jorioz
F
rance
F Tel. +33 450 88 64 00 Fax +33 450 68 91 52
ww.ivoclarvivadent.fr
w
voclar Vivadent GmbH
I
Dr.Adolf-Schneider-Str. 2
-73479 Ellwangen, Jagst
D
ermany
G Tel. +49 (0) 79 61 / 8 89-0 Fax +49 (0) 79 61 / 63 26
ww.ivoclarvivadent.de
w
Ivoclar Vivadent Marketing Ltd. (Liaison Office)
503/504 Raheja Plaza 15 B Shah Industrial Estate Veera Desai Road, Andheri( West)
umbai, 400 053
M India
el. +91 (22) 2673 0302
T
ax +91 (22) 2673 0301
F www.ivoclarvivadent.com
Ivoclar Vivadent s.r.l. & C. s.a.s
Via Gustav Flora, 32 39025 Naturno (BZ) Italy Tel. +39 0473 67 01 11 Fax +39 0473 66 77 80 www.ivoclarvivadent.it
Ivoclar Vivadent K.K.
1-28-24-4F Hongo Bunkyo-ku Tokyo 113-0033 Japan Tel. +81 3 6903 3535 Fax +81 3 5844 3657 www.ivoclarvivadent.jp
Ivoclar Vivadent S.A. de C.V.
Av. Mazatlán No. 61, Piso 2
ol. Condesa
C
6170 México, D.F.
0 Mexico Tel. +52 (55) 5062-1000
ax +52 (55) 5062-1029
F
ww.ivoclarvivadent.com.mx
w
Ivoclar Vivadent Ltd.
2 Omega St, Albany
1
O Box 5243 Wellesley St
P Auckland, New Zealand Tel. +64 9 914 9999
ax +64 9 814 9990
F
ww.ivoclarvivadent.co.nz
w
Ivoclar Vivadent
olska Sp. z.o.o.
P
l. Jana Pawla II 78
u PL-00175 Warszawa Poland
el. +48 22 635 54 96
T
ax +48 22 635 54 69
F www.ivoclarvivadent.pl
Ivoclar Vivadent Marketing Ltd.
Derbenevskaja Nabereshnaya 11, Geb. W 115114 Moscow
ussia
R Tel. +7 495 913 66 19
ax +7 495 913 66 15
F
ww.ivoclarvivadent.ru
w
Ivoclar Vivadent Marketing Ltd.
171 Chin Swee Road #02-01 San Centre Singapore 169877 Tel. +65 6535 6775 Fax +65 6535 4991 www.ivoclarvivadent.com
Ivoclar Vivadent S.L.U.
c/ Emilio Muñoz Nº 15 Entrada c/ Albarracin E-28037 Madrid Spain Tel. + 34 91 375 78 20 Fax + 34 91 375 78 38 www.ivoclarvivadent.es
Ivoclar Vivadent AB
Dalvägen 14
-169 56 Solna
S
weden
S Tel. +46 (0) 8 514 93 930 Fax +46 (0) 8 514 93 940
ww.ivoclarvivadent.se
w
Ivoclar Vivadent Liaison Office
Ahi Evran Caddesi No 1
olaris Is Merkezi Kat: 7
P
0670 Maslak
8 Istanbul Turkey
el. +90 212 346 04 04
T
ax +90 212 346 04 24
F
ww.ivoclarvivadent.com
w
voclar Vivadent Limited
I
round Floor Compass Building
G Feldspar Close Warrens Business Park
nderby
E
eicester LE19 4SE
L United Kingdom Tel. +44 116 284 78 80 Fax +44 116 284 78 81 www.ivoclarvivadent.com
voclar Vivadent, Inc.
I
175 Pineview Drive
mherst, N.Y. 14228
A
SA
U Tel. +1 800 533 6825 Fax +1 716 691 2285 www.ivoclarvivadent.us
Version: 5 Issued: 04/2009
This apparatus has been developed solely for use in dentistry. Start-up and operation should be carried out strictly according to the Instructions for Use. Liability cannot be accepted for damage resulting from misuse or failure to observe the Instructions. The user is solely responsible for testing the apparatus for its suitability for any purpose not explicitly stated in the Instructions. Descriptions and data constitute nor warranty of attributes and are not binding.
Printed in Austria © Ivoclar Vivadent AG, Schaan/Liechtenstein 551548/0409/e