Ivoclar Vivadent SR Ivocap Operating Instructions Manual

Page 1
Operating Instructions
SR Ivocap
®
Equipment System
Page 2
2
Page 3
3
Index
Page
1. Introduction / Signs and Symbols 5
1.1 Introduction
1.2 Signs and symbols
1.3 Notes regarding the Operating Instructions
2. Safety First 6
2.1 Indications
3. Product Description 7
3.1 Components
3.2 Hazardous areas and safety equipment
3.3 Functional description
3.4 Accessories
4. Installation 8
4.1 Unpacking and checking the contents
4.2 Selecting the location
5. Operation 9
5.1 Operating elements at the pressure apparatus
5.2 Operating elements at the clamping frame
6. Procedure 10
6.1 Practical use
7. Maintenance, Cleaning, and Diagnosis 11
7.1 Monitoring and maintenance
7.2 Cleaning
7.3 Changing the Compression spring
8. What if…? 12
8.1 Error messages
8.2 Repairs
9. Product Specifications 13
9.1 Delivery form
9.2 Technical data
9.3 Acceptable operating conditions
9.4 Acceptable transportation and storage conditions
Page 4
4
Views of the Equipment System, List of Parts
A) Pressure apparatus
"ID3"
A1 = Locking valve A2 = Manometer A3 = Extrusion piston A4 = Vise shank A5 = Safety loop
B) Clamping frame “JS3”
B2 = Thrust collar with lever B3 = Pressing plate
C) Flask
C1 = Flask half M C2 = Flask half W C3 = Flask lid C4 = Injection funnel
D) Capsule plunger
E) Investment aid
F) Funnel
G) Timer
D
E
F
G
C
C2
C4
C1
C3
A
A2
A4
A5
B
B3
B1
B2
A1
A3
Page 5
5
1.1 Introduction
The SR Ivocap System is an injection technique which has been especially developed to compensate curing shrinkage. Heat/ pressure curing allows shrinkage of the acrylate during polymerization to be compensated by inflowing material.
The SR Ivocap injection technique, therefore, permits the fabrication of high quality products with excellent physical properties.
Indication
• Complete dentures
• Partial dentures
• Relines
• Orthodontic appliances
• Occlusal splints
Contraindication
Keep unpolymerized
material away from intraoral areas. Do not use the material if the patient is known to be allergic to any of the ingredients listed.
The SR Ivocap System
may only be
operated by trained
personnel.
1.2 Signs and symbols
The signs and symbols in these Operating Instructions facilitate the finding of important points and have the following meaning:
Risks and dangers
Important information
Contraindication
1.3 Notes regarding the Operating Instructions
These Operating Instructions must be read prior to working with the SR Ivocap System.
Equipment concerned: SR Ivocap System Target group: Dental technologists
The Operating Instructions facilitate the correct and economic use of the SR Ivocap System.
These Operating Instructions are divided in several, clearly structured chapters. They should enable you to locate specific topics quickly and easily.
The SR Ivocap material is not described in these instruc­tions. Please refer to the corresponding Instructions for Use of the Ivocap material.
The inform your about risks/dangers, important information, and contraindi­cations, these Instructions contain corresponding signs/ symbols (pictographs) to mark important paragraphs.
We recommend keeping the Instructions in a safe place near the equipment to have immediate access to the information if necessary.
Should you lose the Operating Instructions, extra copies can be ordered at a nominal fee from your local Ivoclar Vivadent Service Center.
1. Introduction / Signs and Symbols
Page 6
6
2. Safety First
This chapter is especially important for personnel who work with the SR Ivocap System or who have to carry out maintenance or repair work. This chapter must be read and the corresponding instructions followed.
2.1 Indications
The SR Ivocap System must only be used for injecting the SR Ivocap material and it should be used for this purpose only. Other uses than the ones stipulated, i.e. injecting other materials, etc. are contraindicated. The manufacturer does not assume any liability for damage resulting from misuse. The user is solely responsible for any risk resulting from failure to observe these Instructions.
Further instructions to assure proper use of the system: – The instructions,
regulations, and notes in these Operating Instructions must be observed.
– The system must be
operated under the indicated environmental and operating conditions (Chapter 9).
– The SR Ivocap System
must be properly maintained (Chapter 7).
2.1.1
Risk of crushing
Do not reach between pressing plunger and safety loop with your fingers. There is a risk of crushing when air is compressed.
Use protective thermo-gloves to touch hot metal parts.
Do not use the water bath without the plastic floaters. Splash water presents a burn hazard.
Wear protective clothing to protect yourself from splash water.
2.1.2
Burn hazard
2.1.3
Burn hazard
2.1.4
Burn hazard
Page 7
3.2 Hazardous areas and safety equipment
Description of the risk areas of the SR Ivocap System:
See also Chapter 2.
7
3.1 Components
The SR Ivocap basic assortment consists of the following components:
• Pressure apparatus
• Flask
• Clamping frame
• Capsule plunger
• Investment aid
• Funnel
• Compression spring
• Allen key
Furthermore, the following accessories are required (not contained in the assortment):
• Cap Vibrator (Ivoclar Vivadent)
• Polymerization bath (various manufacturers)
• Hydraulic press (various manufacturers)
• Plastic floaters (Ivoclar Vivadent)
3. Product Description
3.3 Functional description
The SR Ivocap System is an injection technique which has been especially developed to compensate curing shrinkage. Controlled polymerization allows shrinkage of the material during polymerization to be compensated by inflowing material. The SR Ivocap injection technique, therefore, permits the fabrication of high quality products with outstanding physical properties.
3.4 Accessories
• Cap Vibrator (Ivoclar Vivadent)
• Wall bracket for 4 pressure apparatus (e.g. from BDT)
• Polymerization bath (e.g. from BDT)
• Hydraulic press (various manufacturers)
• Various accessories (BDT)
Address of BDT: BDT GmbH, Industriestrasse 27, D-77656 Offenburg (Germany)
There is a burn hazard if no protective clothing and gloves are worn.
There is a burn hazard if the equipment is removed from the boiling water without protective gloves.
Type of riskHazardous area
Flask and clamping frame
Splash water from the polymerization bath
Page 8
8
4. Installation
4.1 Unpacking and checking the contents
Remove system components from their packaging and place them on a suitable table.
Check the delivery for completeness (see delivery form in Chapter 9) and transportation damage. If certain parts are damaged or missing, contact your local Ivoclar Vivadent Customer Service. We recommend keeping the original packaging for future transportation purposes.
4.2 Selecting the location
The diagram below depicts one possible installation.
Important information
For safety reasons, the installation
of the pressure, power, and water connections may only be carried out by qualified technicians (electricians, plumbers).
Furthermore, the national installation requirements of the corresponding country must be observed.
Quick-fitting pipe union
Air pressure 6 bar
Flexible pressure tubing
6 bar pressure Wall bracket for pressure apparatus
Timer
Water outlet
Water level
Polymerization bath
Cap-Vibrator
Hot water supply
Cold water
Hydraulic press, 3 t or 80 bar
Page 9
9
5. Operation
5.1 Operating elements on the pressure apparatus
Compressed air locking valve (A1)
This locking valve is used to close or open the pressure cylinder for the compressed air.
Operation:
Locking valve open: perpendicular position
Locking valve closed: horizontal position
5.2 Operating elements on the clamping frame
Thrust collar with lever (B2)
Operation:
1) Thrust collar with lever closed
(push to the right until it stops)
2) Thrust collar with lever open
(push to the left)
Pressure onto the clamping frame: 3 tons
Safety loop (A5)
The safety loop is used to secure the vise shanks.
Operation:
Safety loop open: Towards the pressure cylinder
Safety loop closed: Towards the clamping frame
open
closed
1
2
B1
B2
Page 10
10
6.1 Practical use
6. Procedure
The most important steps:
Injection - 5 minutes
Polymerization – 35 minutes in boiling water (as of the beginning of boiling)
Cooling in a cold-water bath – at least 30 minutes; the injection pressure must be maintained for at least 20 minutes thereof
3
2
1
Compressed air 6 bar
Water level
Pressure apparatus
Capsule
Flask
Mark for water level
Flowing cold water
For information on how to process the material please refer to the material Instructions for Use of the SR Ivocap System.
Page 11
11
This chapter describes the maintenance and cleaning procedures which can be carried out by dental professionals. All other tasks must be performed by qualified service personnel at a certified Ivoclar Vivadent Service Center.
7. Maintenance, Cleaning, and Diagnosis
7.2 Cleaning
Metal parts can be cleaned with warm water and a cleaning brush. After cleaning, the flask may be treated with Vaseline to facilitate the removal of plaster residue.
7.3 Changing the Compression spring
Please take note of the informational sheet added to the 'spring washers' set.
The Compression spring are to be changed by a certified Ivoclar Vivadent Service Center.
Disconnect power and compressed air connection before maintenance and cleaning.
7.1 Monitoring and maintenance
The time for these maintenance procedures depends on the frequency of use and work­ing habits of the users. For that reason, the recommended times are only approximates.
What: Part: When:
Check all compressed air connections for correct fit
Compressed air connections
Before use
Check the components for mechanical damage
Pressure apparatus
Flask
Clamping frame
Before use
Check the plastic parts of the flask for correct fit
Clean contaminated clamping frame (wax), e.g. by boiling out
clamping frame in case of
contamination
Plastic parts of the flask
Before use
The insulation shoulder of the flask must only be replaced by a certified Service Center, since remounting requires a special mounting assembly and a special material. Only the correct mounting procedure ensures sound function of the flask.
Page 12
12
8. What if...?
8.1 Error messages
List of possible error messages and their meaning:
This chapter will help you to recognize malfunctions and take appropriate measures or, if possible, to perform some repairs.
8.2 Repairs
Repairs may only be carried out by a certified Ivoclar Vivadent Service Center. Please refer to the addresses in Chapter 10.
If repairs during the warranty period are not carried out by a certified Ivoclar Vivadent Service Center, the warranty will expire immediately.
Please also read the safety information in Chapter 2.
Error
1
2
3
4
5
6
7
8
9
10
11
Flasks do not close. The flask covers do not sit on the flask margins.
The model to be invested does not fit in the flask.
The pressure is difficult to be transferred to the flask and/or clamping frame.
Pressure in the clamping frame fluctuates.
Injection plunger enters the capsule and deforms the bottom of the capsule.The material is extruded at the wrong end of the capsule.
The rod of the injection plunger cannot be moved, even if the connection to the compressed air source is interrupted. It seems to be stuck.
Injection plunger does not move.The SR Ivocap material is not injected.
The resin flows out of the flask during polymerization.
The locking mechanism is blocked.
The locking mechanism is contaminated with wax.
Capsule reservoir has polymerized after 35 minutes.
Description Notes for users
Flask margins are not clean; plaster residue on the injection funnel.
The model should be trimmed prior to investment.
Clamping frame is not correctly positioned, i.e. in the centre of the press.
Compression spring in the clamping frame are old or dirty and cannot transfer the pressure.
Replace the compression spring or clean the
clamping frame (see Error 12).
The injection plunger is not correctly placed on the capsule.There is an empty space between the injection plunger and capsule bottom. Avoid empty space and place the injection plunger evenly on the capsule bottom.
– No compressed air present – Capsule cover was not removed – There is plaster residue or the funnel is blocked, e.g. by plaster
residue – The injection channels are too narrow. – The flow properties of the resin are poor because of incorrect
storage
– Leaks as a result of incorrect investment. – Flask defective. – Inadequate clamping pressure on the frame. – A foreign object is stuck between the two flask halves.
The mechanical components are damaged as a result of improper handling. In case of considerable damage, the mechanical components must be replaced.
Wax residue in the water bath has contaminated the clamping frame. Boil out clamping frame with clean, hot water.
– Water level surpasses the marking on the clamping frame. – An unsuitable polymerization bath was used. – The system was not placed in cold water immediately after
polymerization. – Reduce polymerization time by 1–2 minutes if necessary
Negative pressure in the pressure cylinder may prevent the plunger from moving, open valve
Page 13
13
9. Product Specifications
This chapter contains all relevant product specifications.
9.1 Delivery form
1 Pressure apparatus 1 Clamping frame 2 Flasks 1 Timer 1 Compression spring 1 Capsule plunger 2 Investment aids 1 Allen key
9.2 Technical data
Weight:
Flask 1.3 kg Clamping frame ID3 2.3 kg Pressure apparatus 2.0 kg Total 5.6 kg
Max. operating pressure:
Pressure apparatus 6.5 bar Clamping frame 3 tons
Noise emissions:
< 70 dB (A)
9.3 Acceptable operating conditions
Acceptable humidity range
Maximum relative humidity at 31 °C (87 °F) gradually decreasing to 50% at 40 °C (104 °F), condensation excluded.
Acceptable ambient pressure
The unit is tested for altitudes of up to 2000 m above sea level.
9.4 Acceptable transporation and storage conditions
Acceptable temperature range
–20 °C to 50 °C (–4 °F to 122 °F)
Acceptable humidity range
Maximum relative humidity at 31 °C (87 °F) gradually decreasing to 50% at 40 °C (104 °F), condensation excluded.
Acceptable ambient pressure
500 mbar to 1060 mbar
Use only original packaging together with the respective foam material for shipping purposes.
Page 14
14
Page 15
15
Page 16
Ivoclar Vivadent – worldwide
This apparatus has been developed solely for use in dentistry. Start-up and operation should be carried out strictly according to the Instructions for Use. Liability cannot be accepted for damage resulting from misuse or failure to observe the Instructions. The user is solely responsible for testing the apparatus for its suitability for any purpose not explicitly stated in the Instructions. Descriptions and data constitute nor warranty of attributes and are not binding.
Printed in Austria © Ivoclar Vivadent AG, Schaan/Liechtenstein 551548/0804/e
Version: 5 Issued: 08/2004
Ivoclar Vivadent AG
Bendererstrasse 2 FL-9494 Schaan Liechtenstein Tel. +423 235 35 35 Fax +423 235 33 60 www.ivoclarvivadent.com
Ivoclar Vivadent Pty. Ltd.
1 – 5 Overseas Drive P.O. Box 367 Noble Park,Vic. 3174 Australia Tel. +61 3 979 595 99 Fax +61 3 979 596 45 www.ivoclarvivadent.com.au
Ivoclar Vivadent GmbH
Bremschlstr. 16 Postfach 223 A-6706 Bürs Austria Tel. +43 5552 624 49 Fax +43 5552 675 15 www.ivoclarvivadent.com
Ivoclar Vivadent Ltda.
Rua Maestro João Gomes de Araújo 50; Salas 92/94 Sao Paulo, CEP 02332-020 Brazil Tel. +55 11 69 59 89 77 Fax +55 11 69 71 17 50 www.ivoclarvivadent.com
Ivoclar Vivadent Inc.
2785 Skymark Avenue, Unit 1 Mississauga Ontario L4W 4Y3 Canada Tel. +1 905 238 57 00 Fax +1 905 238 5711 www.ivoclarvivadent.us.com
Ivoclar Vivadent Marketing Ltd.
Rm 603 Kuen Yang International Business Plaza No. 798 Zhao Jia Bang Road Shanghai 200030 China Tel. +86 21 5456 0776 Fax. +86 21 6445 1561 www.ivoclarvivadent.com
Ivoclar Vivadent Marketing Ltd.
Calle 134 No. 13-83, Of. 520 Bogotá Colombia Tel. +57 1 627 33 99 Fax +57 1 633 16 63 www.ivoclarvivadent.com
Ivoclar Vivadent SAS
B.P. 118 F-74410 Saint-Jorioz France Tel. +33 450 88 64 00 Fax +33 450 68 91 52 www.ivoclarvivadent.fr
Ivoclar Vivadent GmbH
Dr.Adolf-Schneider-Str. 2 D-73479 Ellwangen, Jagst Germany Tel. +49 (0) 79 61 / 8 89-0 Fax +49 (0) 79 61 / 63 26 www.ivoclarvivadent.de
Ivoclar-Vivadent Marketing Ltd
114, Janki Centre Shah Industrial Estate Veera Desai Road, Andheri (West) Mumbai 400 053 India Tel. +91 (22) 673 0302 Fax. +91 (22) 673 0301 www.ivoclarvivadent.firm.in
Ivoclar Vivadent s.r.l.
Via dell’Industria 16 I-39025 Naturno (BZ) Italy Tel. +39 0473 67 01 11 Fax +39 0473 66 77 80 www.ivoclarvivadent.it
Ivoclar Vivadent S.A. de C.V.
Av. Mazatlán No. 61, Piso 2 Col. Condesa 06170 México, D.F. Mexico Tel. +52 (55) 55 53 00 38 Fax +52 (55) 55 53 14 26 www.ivoclarvivadent.com.mx
Ivoclar Vivadent Ltd
12 Omega St, Albany PO Box 5243 Wellesley St Auckland, New Zealand Tel. +64 9 914 9999 Fax +64 9 630 61 48 www.ivoclarvivadent.co.nz
Ivoclar Vivadent Polska Sp. z.o.o.
ul. Jana Pawla II 78 PL-01-501 Warszawa Poland Tel. +48 22 635 54 96 Fax +48 22 635 54 69 www.ivoclarvivadent.pl
Ivoclar Vivadent S.A.
c/Emilio Muñoz, 15 Esquina c/Albarracín E-28037 Madrid Spain Tel. + 34 91 375 78 20 Fax + 34 91 375 78 38 www.ivoclarvivadent.com
Ivoclar Vivadent AB
Dalvägen 14 S-169 56 Solna Sweden Tel. +46 8 514 93 943 Fax +46 8 514 93 940 www.ivoclarvivadent.se
Ivoclar Vivadent UK Limited
Ground Floor Compass Building Feldspar Close Warrens Business Park Enderby Leicester LE19 4SE United Kingdom Tel. +44 116 284 78 80 Fax +44 116 284 78 81 www.ivoclarvivadent.co.uk
Ivoclar Vivadent, Inc.
175 Pineview Drive Amherst, N.Y. 14228 USA Tel. +1 800 533 6825 Fax +1 716 691 2285 www.ivoclarvivadent.us.com
Dentigenix Inc.
551 N. 34th Street Seattle, WA 98103 USA Tel. +1 425 822 0902 Fax +1 425 822 3101 www.dentigenix.com
Loading...