DC Power Supply
GPD-3303 Series
USER MANUAL
GW INSTEK PART NO.
This manual contains proprietary information, which is protected by copyrights. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
ISO-9001 CERTIFIED MANUFACTURER
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan.
TABLE OF CONTENTS |
|
Table of Contents |
|
SAFETY INSTRUCTIONS .................................................. |
5 |
OVERVIEW....................................................................... |
9 |
Introduction........................................... |
9 |
Series Lineup / Main Features ............. |
11 |
Principle of Operation.......................... |
12 |
Front Panel Overview ........................... |
14 |
Rear Panel Overview ............................ |
17 |
CV/CC Crossover Characteristics ......... |
18 |
SETUP ........................................................................... |
19 |
Power Up ............................................. |
19 |
Load Cable Connection ........................ |
20 |
Output On/Off..................................... |
21 |
Beep On/Off ........................................ |
21 |
Front Panel Lock .................................. |
22 |
OPERATION .................................................................. |
23 |
CH1/CH2 Independent Mode .............. |
23 |
CH3 Independent Mode....................... |
25 |
CH1/CH2 Tracking Series Mode........... |
27 |
CH1/CH2 Tracking Parallel Mode ........ |
32 |
SAVE/RECALL SETUP ..................................................... |
34 |
Save Setup ........................................... |
34 |
Recall Setup ......................................... |
35 |
REMOTE CONTROL ....................................................... |
36 |
Remote Control Setup .......................... |
36 |
Remote Connection Step...................... |
37 |
Command Syntax ................................. |
38 |
|
3 |
GPD-3303 Series User Manual |
|
Error Messages .................................... |
38 |
Command List...................................... |
39 |
Command Details ................................ |
40 |
FAQ................................................................................ |
45 |
APPENDIX ...................................................................... |
46 |
Fuse Replacement ................................ |
46 |
Specifications ...................................... |
47 |
Declaration of Conformity .................... |
49 |
INDEX ............................................................................ |
50 |
4
SAFETY INSTRUCTIONS
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow when operating the GPD3303 series and when keeping it in storage. Read the following before any operation to insure your safety and to keep the best condition for the GPD3303 series.
These safety symbols may appear in this manual or on the GPD-3303 series.
Warning: Identifies conditions or practices that WARNING could result in injury or loss of life.
Caution: Identifies conditions or practices that CAUTION could result in damage to the GPD-3303 series or
to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
5
GPD-3303 Series User Manual
General
Guidelines
CAUTION
•Do not place any heavy object on the GPD-3303 series.
•Avoid severe impacts or rough handling that leads to damaging the GPD-3303 series.
•Do not discharge static electricity to the GPD3303 series.
•Do not block or obstruct the cooling fan vent opening.
•Do not perform measurement at circuits directly connected to Mains (see note below).
•Do not disassemble the GPD-3303 series unless you are qualified as service personnel.
(Measurement categories) EN 61010-1:2001 specifies the measurement categories and their requirements as follows. The GPD-3303 series falls under category I.
•Measurement category IV is for measurement performed at the source of low-voltage installation.
•Measurement category III is for measurement performed in the building installation.
•Measurement category II is for measurement performed on the circuits directly connected to the low voltage installation.
•Measurement category I is for measurements performed on circuits not directly connected to Mains.
Power Supply
WARNING
•AC Input voltage: 100V/120V/220V/230V ±10%, 50/60Hz
•Connect the protective grounding conductor of the AC power cord to an earth ground, to avoid electrical shock.
Fuse |
• |
Fuse type: 100V/120V: T6.3A/250V, |
|
|
220V/230V: T3.15A/250V |
WARNING |
• |
Make sure the correct type of fuse is installed |
|
|
before power up. |
6 |
|
|
SAFETY INSTRUCTIONS
•To ensure fire protection, replace the fuse only with the specified type and rating.
•Disconnect the power cord before fuse replacement.
•Make sure the cause of fuse blowout is fixed before fuse replacement.
Cleaning the GPD-3303 series
•Disconnect the power cord before cleaning.
•Use a soft cloth dampened in a solution of mild detergent and water. Do not spray any liquid.
•Do not use chemicals or cleaners containing harsh products such as benzene, toluene, xylene, and acetone.
Operation |
• |
Location: Indoor, no direct sunlight, dust free, |
Environment |
|
almost non-conductive pollution (note below) |
|
• Relative Humidity: < 80% |
|
|
• |
Altitude: < 2000m |
• Temperature: 0°C to 40°C
(Pollution Degree) EN 61010-1:2001 specifies the pollution degrees and their requirements as follows. The GPD-3303 series falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
• Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence.
• Pollution degree 2: Normally only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
• Pollution degree 3: Conductive pollution occurs, or dry, nonconductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
7
|
|
|
GPD-3303 Series User Manual |
|
|
|
|
|
|
|
|
|
|
|
|
Storage |
• |
Location: Indoor |
|
environment |
• |
Relative Humidity: < 70% |
|
|
|
• Temperature: −10°C to 70°C
When using the GPD-3303 series in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the following code:
Green/ Yellow: |
Earth |
Blue: |
Neutral |
Brown: |
Live (Phase) |
As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth
terminal marked with the letter E or by the earth symbol or coloured Green or Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring must be carried out in accordance with the information detailed on this label.
8
OVERVIEW
OVERVIEW
This chapter describes the GPD-3303 series in a nutshell, including its main features and front / rear panel introduction. After going through the overview, follow the Setup chapter (page19) to properly power up and set operation environment.
Overview |
GPD-3303, the regulated DC power supply series, |
|
are light weight, adjustable, multifunctional work |
|
stations. They have three independent outputs: |
|
two with adjustable voltage level and one with |
|
fixed level selectable from 2.5V, 3.3V and 5V. The |
|
GPD-3303 series can be used for logic circuits |
|
where various output voltage or current are |
|
needed, and for tracking mode definition systems |
|
where plus and minus voltages with insignificant |
|
error are required. |
|
|
Independent / |
The three output modes of GPD-3303 series, |
Tracking Series / |
independent, tracking series, and tracking parallel, |
Tracking Parallel |
can be selected through pressing the TRACKING |
|
key on the front panel. In the independent mode, |
|
the output voltage and current of each channel are |
|
controlled separately. The isolation degree, from |
|
output terminal to chassis or from output terminal |
|
to output terminal, is 300V. In the tracking modes, |
|
both the CH1 and CH2 outputs are automatically |
|
connected in series or parallel; no need to connect |
|
output leads. In the series mode, the output |
|
voltage is doubled; in the parallel mode, the |
|
output current is doubled. |
|
9 |
GPD-3303 Series User Manual
Constant Voltage/ Except for CH3, each output channel is completely Constant Current transistorized and well-regulated, and works in
constant voltage (CV) or constant current (CC) mode. Even at the maximum output current, a fully rated, continuously adjustable output voltage is provided. For a big load, the power supply can be used as a CV source; while for a small load, a CC source. When in the CV mode (independent or tracking mode), output current (overload or short circuit) can be controlled via the front panel. When in the CC mode (independent mode only), the maximum (ceiling) output voltage can be controlled via the front panel. The power supply will automatically cross over from CV to CC operation when the output current reaches the target value. The power supply will automatically cross over from CC to CV when the output voltage reaches the target value. For more details about CV/CC mode operation, see page18.
Automatic |
The front panel display (CH1, CH2) shows the |
tracking mode |
output voltage or current. When operating in the |
|
tracking mode, the power supply will |
|
automatically connect to the autotracking mode. |
|
|
Dynamic load |
When used in audio production lines, the power |
|
supply will provide a continuous or dynamic load |
|
connector. When the connectors are connected to |
|
the position “ON”, a stable DC current power will |
|
be provided for audio power amplifiers. |
10
OVERVIEW
Model |
V Meter |
A Meter |
USB |
Tracking Error |
|
|
|
|
|
GPD-3303D |
3 digit |
3 digit |
Yes |
≤ 0.5% + 50mV of Master |
|
|
|
|
|
GPD-3303S |
5 digit |
4 digit |
Yes |
≤ 0.5% + 10mV of Master |
Performance |
• |
Low noise: Cooling fan controlled by Heatsink |
|
|
temperature |
|
• Compact size, light weight |
|
|
|
|
Operation |
• |
Constant Voltage / Constant Current operation |
|
• Tracking Series / Tracking parallel operation |
|
|
• |
Output On/Off control |
|
• 3 outputs: 30V/3A x 2, 2.5V/3.3V/5V/3A x 1 |
|
|
• |
Digital panel control |
|
• 4 sets of panel setup save/recall |
|
|
• Coarse and fine Voltage/Current control |
|
|
• |
Software calibration |
|
• |
Buzzer output |
|
• |
Key lock function |
|
|
|
Protection |
• |
Overload protection |
|
• |
Reverse polarity protection |
|
|
|
Interface |
• |
USB for remote control |
11
GPD-3303 Series User Manual
Overview |
The power supply consists of the following. |
|
|
• |
AC input circuit |
|
• |
Transformer |
|
• |
Bias power supply including rectifier, filter, |
pre-regulator and reference voltage source
• Main regulator circuit including the main rectifier and filter, series regulator, current comparator, voltage comparator, reference voltage amplifier, remote device and relay control circuit
The block diagram below shows the circuit arrangement. The single phase input power is connected to the transformer through the input circuit. Details of each part are described in the next page.
Block diagram
12
|
|
OVERVIEW |
|
|
|
|
|
|
Rectifier |
The auxiliary rectifiers D1011~ D1014 provide bias |
|
|
|
voltage filtered by the capacitors C102 and C103, |
|
|
for the pre-regulators U101 and U102. They |
|
|
provide a regulated voltage for other modules. |
|
|
|
Main Rectifier |
The main rectifier is a full wave bridge rectifier. It |
|
|
|
provides the power after the rectifier is filtered by |
|
|
the capacitor C101, and then regulated via a series- |
|
|
wound regulator, which is finally delivered to the |
|
|
output terminal. |
|
|
|
Current Limiter |
U104 acts as a current limiter. When the current is |
|
|
|
over predetermined rating, U104 is activated and |
|
|
decreases the current. U208 provides a reference |
|
|
voltage. U206 is the inverter amplifier. U103 is a |
|
|
comparator amplifier which compares reference |
|
|
voltage and feedback voltage, and then delivers to |
|
|
Q103, Q104, which then calibrates the output |
|
|
voltage. |
|
|
|
Overload |
When the unit is overloaded, Q107 activates to |
|
|
|
control the current magnitude of Q104, to limit the |
|
|
output current. The relay control circuit controls |
|
|
the power dissipation in the series-wound |
|
|
regulated circuit. |
GPD-3303 Series User Manual
Display |
VoltMeter Displays CH1 or CH2 output voltage.
GPD-3303S (5 digits) |
GPD-3303D (3 digits) |
||||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
AmpMeter Displays CH1 or CH2 output current.
GPD-3303S (4 digits) |
GPD-3303D (3 digits) |
||||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
13 |
14 |
|
OVERVIEW |
Control Panel |
|
Memory Keys |
Saves or recalls panel settings. Four |
|
settings, 1 ~ 4, are available. For |
|
save/recall details, see page34. |
CH1/CH2/Beep |
Selects the output channel for level |
Keys |
adjustment. For level setting details, |
|
see page23. Pressing and holding |
|
CH2 key enables beep sound. For |
|
details, see page21. |
Parallel/Series |
Activates Tracking Parallel operation |
Keys |
or Tracking Series operation, For |
|
details, see page27. |
Lock Key |
Locks or unlocks the front panel |
|
settings. For details, see page22. |
Output Key |
Turns the output on or off. |
Voltage Knobs |
Adjusts the output voltage level for |
|
CH1 or CH2. Pressing the knob |
|
switches coarse and fine level setting. |
Current Knobs |
Adjusts the output current level for |
|
CH1 or CH2. Pressing the knob |
|
switches coarse and fine level setting. |
Power Switch |
Turns On or Off the main power. |
|
For power up sequence, see page19. |
|
15 |
GPD-3303 Series User Manual
Terminals
Default Terminals |
European Terminals |
GND Terminal |
|
|
Accepts a grounding wire. |
|
|
||
|
|
|
|
|
|
|
|
CH1 Output |
|
|
|
Outputs CH1 voltage and current. |
|
|
|
|
|
|
|
|
|
|
CH1 CV/CC |
Indicates CH1 Constant Voltage or |
Indicator |
Constant Current state. |
CH2 Output |
Outputs CH2 voltage and current. |
CH2 |
Indicates CH2 Constant Voltage, |
CV/CC/PAR |
Constant Current, or Tracking |
Indicator |
Parallel operation mode. |
CH3 Output |
Outputs CH3 voltage and current. |
CH3 Overload |
Indicates when CH3 output current is |
Indicator |
overloaded. |
CH3 Voltage |
Selects CH3 output voltage: 2.5V, |
Selector |
3.3V, or 5V. |
16 |
|