GW Instek GPD-3303S, GPD-3303D Instruction Manual

0 (0)

DC Power Supply

GPD-3303 Series

USER MANUAL

GW INSTEK PART NO.

This manual contains proprietary information, which is protected by copyrights. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.

The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.

ISO-9001 CERTIFIED MANUFACTURER

Good Will Instrument Co., Ltd.

No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan.

TABLE OF CONTENTS

Table of Contents

 

SAFETY INSTRUCTIONS ..................................................

5

OVERVIEW.......................................................................

9

Introduction...........................................

9

Series Lineup / Main Features .............

11

Principle of Operation..........................

12

Front Panel Overview ...........................

14

Rear Panel Overview ............................

17

CV/CC Crossover Characteristics .........

18

SETUP ...........................................................................

19

Power Up .............................................

19

Load Cable Connection ........................

20

Output On/Off.....................................

21

Beep On/Off ........................................

21

Front Panel Lock ..................................

22

OPERATION ..................................................................

23

CH1/CH2 Independent Mode ..............

23

CH3 Independent Mode.......................

25

CH1/CH2 Tracking Series Mode...........

27

CH1/CH2 Tracking Parallel Mode ........

32

SAVE/RECALL SETUP .....................................................

34

Save Setup ...........................................

34

Recall Setup .........................................

35

REMOTE CONTROL .......................................................

36

Remote Control Setup ..........................

36

Remote Connection Step......................

37

Command Syntax .................................

38

 

3

GPD-3303 Series User Manual

Error Messages ....................................

38

Command List......................................

39

Command Details ................................

40

FAQ................................................................................

45

APPENDIX ......................................................................

46

Fuse Replacement ................................

46

Specifications ......................................

47

Declaration of Conformity ....................

49

INDEX ............................................................................

50

4

SAFETY INSTRUCTIONS

SAFETY INSTRUCTIONS

This chapter contains important safety instructions that you must follow when operating the GPD3303 series and when keeping it in storage. Read the following before any operation to insure your safety and to keep the best condition for the GPD3303 series.

Safety Symbols

These safety symbols may appear in this manual or on the GPD-3303 series.

Warning: Identifies conditions or practices that WARNING could result in injury or loss of life.

Caution: Identifies conditions or practices that CAUTION could result in damage to the GPD-3303 series or

to other properties.

DANGER High Voltage

Attention Refer to the Manual

Protective Conductor Terminal

Earth (ground) Terminal

5

GPD-3303 Series User Manual

Safety Guidelines

General

Guidelines

CAUTION

Do not place any heavy object on the GPD-3303 series.

Avoid severe impacts or rough handling that leads to damaging the GPD-3303 series.

Do not discharge static electricity to the GPD3303 series.

Do not block or obstruct the cooling fan vent opening.

Do not perform measurement at circuits directly connected to Mains (see note below).

Do not disassemble the GPD-3303 series unless you are qualified as service personnel.

(Measurement categories) EN 61010-1:2001 specifies the measurement categories and their requirements as follows. The GPD-3303 series falls under category I.

Measurement category IV is for measurement performed at the source of low-voltage installation.

Measurement category III is for measurement performed in the building installation.

Measurement category II is for measurement performed on the circuits directly connected to the low voltage installation.

Measurement category I is for measurements performed on circuits not directly connected to Mains.

Power Supply

WARNING

AC Input voltage: 100V/120V/220V/230V ±10%, 50/60Hz

Connect the protective grounding conductor of the AC power cord to an earth ground, to avoid electrical shock.

Fuse

Fuse type: 100V/120V: T6.3A/250V,

 

 

220V/230V: T3.15A/250V

WARNING

Make sure the correct type of fuse is installed

 

 

before power up.

6

 

 

SAFETY INSTRUCTIONS

To ensure fire protection, replace the fuse only with the specified type and rating.

Disconnect the power cord before fuse replacement.

Make sure the cause of fuse blowout is fixed before fuse replacement.

Cleaning the GPD-3303 series

Disconnect the power cord before cleaning.

Use a soft cloth dampened in a solution of mild detergent and water. Do not spray any liquid.

Do not use chemicals or cleaners containing harsh products such as benzene, toluene, xylene, and acetone.

Operation

Location: Indoor, no direct sunlight, dust free,

Environment

 

almost non-conductive pollution (note below)

 

Relative Humidity: < 80%

 

Altitude: < 2000m

Temperature: 0°C to 40°C

(Pollution Degree) EN 61010-1:2001 specifies the pollution degrees and their requirements as follows. The GPD-3303 series falls under degree 2.

Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.

• Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence.

• Pollution degree 2: Normally only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.

• Pollution degree 3: Conductive pollution occurs, or dry, nonconductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.

7

 

 

 

GPD-3303 Series User Manual

 

 

 

 

 

 

 

 

 

 

 

 

Storage

Location: Indoor

environment

Relative Humidity: < 70%

 

 

Temperature: 10°C to 70°C

Power cord for the United Kingdom

When using the GPD-3303 series in the United Kingdom, make sure the power cord meets the following safety instructions.

NOTE: This lead/appliance must only be wired by competent persons

WARNING: THIS APPLIANCE MUST BE EARTHED

IMPORTANT: The wires in this lead are coloured in accordance with the following code:

Green/ Yellow:

Earth

Blue:

Neutral

Brown:

Live (Phase)

As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows:

The wire which is coloured Green & Yellow must be connected to the Earth

terminal marked with the letter E or by the earth symbol or coloured Green or Green & Yellow.

The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.

The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.

If in doubt, consult the instructions provided with the equipment or contact the supplier.

This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.

Any moulded mains connector that requires removal /replacement must be destroyed by removal of any fuse & fuse carrier and disposed of immediately, as a plug with bared wires is hazardous if a engaged in live socket. Any re-wiring must be carried out in accordance with the information detailed on this label.

8

OVERVIEW

OVERVIEW

This chapter describes the GPD-3303 series in a nutshell, including its main features and front / rear panel introduction. After going through the overview, follow the Setup chapter (page19) to properly power up and set operation environment.

Introduction

Overview

GPD-3303, the regulated DC power supply series,

 

are light weight, adjustable, multifunctional work

 

stations. They have three independent outputs:

 

two with adjustable voltage level and one with

 

fixed level selectable from 2.5V, 3.3V and 5V. The

 

GPD-3303 series can be used for logic circuits

 

where various output voltage or current are

 

needed, and for tracking mode definition systems

 

where plus and minus voltages with insignificant

 

error are required.

 

 

Independent /

The three output modes of GPD-3303 series,

Tracking Series /

independent, tracking series, and tracking parallel,

Tracking Parallel

can be selected through pressing the TRACKING

 

key on the front panel. In the independent mode,

 

the output voltage and current of each channel are

 

controlled separately. The isolation degree, from

 

output terminal to chassis or from output terminal

 

to output terminal, is 300V. In the tracking modes,

 

both the CH1 and CH2 outputs are automatically

 

connected in series or parallel; no need to connect

 

output leads. In the series mode, the output

 

voltage is doubled; in the parallel mode, the

 

output current is doubled.

 

9

GPD-3303 Series User Manual

Constant Voltage/ Except for CH3, each output channel is completely Constant Current transistorized and well-regulated, and works in

constant voltage (CV) or constant current (CC) mode. Even at the maximum output current, a fully rated, continuously adjustable output voltage is provided. For a big load, the power supply can be used as a CV source; while for a small load, a CC source. When in the CV mode (independent or tracking mode), output current (overload or short circuit) can be controlled via the front panel. When in the CC mode (independent mode only), the maximum (ceiling) output voltage can be controlled via the front panel. The power supply will automatically cross over from CV to CC operation when the output current reaches the target value. The power supply will automatically cross over from CC to CV when the output voltage reaches the target value. For more details about CV/CC mode operation, see page18.

Automatic

The front panel display (CH1, CH2) shows the

tracking mode

output voltage or current. When operating in the

 

tracking mode, the power supply will

 

automatically connect to the autotracking mode.

 

 

Dynamic load

When used in audio production lines, the power

 

supply will provide a continuous or dynamic load

 

connector. When the connectors are connected to

 

the position “ON”, a stable DC current power will

 

be provided for audio power amplifiers.

10

OVERVIEW

Series Lineup / Main Features

Series Lineup

Model

V Meter

A Meter

USB

Tracking Error

 

 

 

 

 

GPD-3303D

3 digit

3 digit

Yes

≤ 0.5% + 50mV of Master

 

 

 

 

 

GPD-3303S

5 digit

4 digit

Yes

≤ 0.5% + 10mV of Master

Main Features

Performance

Low noise: Cooling fan controlled by Heatsink

 

 

temperature

 

Compact size, light weight

 

 

 

Operation

Constant Voltage / Constant Current operation

 

Tracking Series / Tracking parallel operation

 

Output On/Off control

 

3 outputs: 30V/3A x 2, 2.5V/3.3V/5V/3A x 1

 

Digital panel control

 

4 sets of panel setup save/recall

 

Coarse and fine Voltage/Current control

 

Software calibration

 

Buzzer output

 

Key lock function

 

 

 

Protection

Overload protection

 

Reverse polarity protection

 

 

 

Interface

USB for remote control

11

GPD-3303 Series User Manual

Principle of Operation

Overview

The power supply consists of the following.

 

AC input circuit

 

Transformer

 

Bias power supply including rectifier, filter,

pre-regulator and reference voltage source

Main regulator circuit including the main rectifier and filter, series regulator, current comparator, voltage comparator, reference voltage amplifier, remote device and relay control circuit

The block diagram below shows the circuit arrangement. The single phase input power is connected to the transformer through the input circuit. Details of each part are described in the next page.

Block diagram

12

GW Instek GPD-3303S, GPD-3303D Instruction Manual

 

 

OVERVIEW

 

 

 

 

 

 

Rectifier

The auxiliary rectifiers D1011~ D1014 provide bias

 

 

voltage filtered by the capacitors C102 and C103,

 

 

for the pre-regulators U101 and U102. They

 

 

provide a regulated voltage for other modules.

 

 

 

Main Rectifier

The main rectifier is a full wave bridge rectifier. It

 

 

provides the power after the rectifier is filtered by

 

 

the capacitor C101, and then regulated via a series-

 

 

wound regulator, which is finally delivered to the

 

 

output terminal.

 

 

 

Current Limiter

U104 acts as a current limiter. When the current is

 

 

over predetermined rating, U104 is activated and

 

 

decreases the current. U208 provides a reference

 

 

voltage. U206 is the inverter amplifier. U103 is a

 

 

comparator amplifier which compares reference

 

 

voltage and feedback voltage, and then delivers to

 

 

Q103, Q104, which then calibrates the output

 

 

voltage.

 

 

 

Overload

When the unit is overloaded, Q107 activates to

 

 

control the current magnitude of Q104, to limit the

 

 

output current. The relay control circuit controls

 

 

the power dissipation in the series-wound

 

 

regulated circuit.

GPD-3303 Series User Manual

Front Panel Overview

Display

VoltMeter Displays CH1 or CH2 output voltage.

GPD-3303S (5 digits)

GPD-3303D (3 digits)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AmpMeter Displays CH1 or CH2 output current.

GPD-3303S (4 digits)

GPD-3303D (3 digits)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

13

14

 

OVERVIEW

Control Panel

 

Memory Keys

Saves or recalls panel settings. Four

 

settings, 1 ~ 4, are available. For

 

save/recall details, see page34.

CH1/CH2/Beep

Selects the output channel for level

Keys

adjustment. For level setting details,

 

see page23. Pressing and holding

 

CH2 key enables beep sound. For

 

details, see page21.

Parallel/Series

Activates Tracking Parallel operation

Keys

or Tracking Series operation, For

 

details, see page27.

Lock Key

Locks or unlocks the front panel

 

settings. For details, see page22.

Output Key

Turns the output on or off.

Voltage Knobs

Adjusts the output voltage level for

 

CH1 or CH2. Pressing the knob

 

switches coarse and fine level setting.

Current Knobs

Adjusts the output current level for

 

CH1 or CH2. Pressing the knob

 

switches coarse and fine level setting.

Power Switch

Turns On or Off the main power.

 

For power up sequence, see page19.

 

15

GPD-3303 Series User Manual

Terminals

Default Terminals

European Terminals

GND Terminal

 

 

Accepts a grounding wire.

 

 

 

 

 

 

 

 

 

 

CH1 Output

 

 

 

Outputs CH1 voltage and current.

 

 

 

 

 

 

 

 

 

 

CH1 CV/CC

Indicates CH1 Constant Voltage or

Indicator

Constant Current state.

CH2 Output

Outputs CH2 voltage and current.

CH2

Indicates CH2 Constant Voltage,

CV/CC/PAR

Constant Current, or Tracking

Indicator

Parallel operation mode.

CH3 Output

Outputs CH3 voltage and current.

CH3 Overload

Indicates when CH3 output current is

Indicator

overloaded.

CH3 Voltage

Selects CH3 output voltage: 2.5V,

Selector

3.3V, or 5V.

16

 