GW Instek GFG-813 User Manual

4 (1)
GW Instek GFG-813 User Manual

PE-C4-410

PAGE 1

GOODWILL INSTRUMENT ( MALAYSIA ) SDN. BHD.

CALIBRATION STANDARD

PROCEDURES

FUNCTION GENERATOR

MODEL : GFG-813

APPLICABLE ODM MODEL :

FG-813; BK-3040

GW GOODWILL INSTRUMENT ( M ) SDN BHD.

APPROVED BY CHECKED BY PREPARED BY

TOTAL PAGE : 15

( INCLUDING COVER PAGE )

 

 

 

 

 

 

 

 

 

 

 

 

 

FOR INTERNAL USE ONLY

 

 

MODEL: GFG-813

 

 

PAGE

:

2

 

 

 

CALIBRATION PROCESS

 

 

 

 

STATION : PRE-ADJUST

 

 

 

 

 

 

 

 

 

 

 

NO

Adjustment

Location

 

Particular

 

 

Specification

 

Remarks

1

Initial Setting.

 

 

POWER Switch = ON;

 

 

 

 

 

Instrument Used :

 

 

 

 

EXT COUNTER : release all buttons;

 

 

 

 

 

20Mhz O'scope

 

 

 

 

MAIN GENERATOR : AMPL=Max, ATT= Release,

 

 

 

 

 

 

 

 

 

 

 

OFFSET & SYM= Push, TRIGGER PHASE=Free run;

 

 

 

 

 

 

 

 

 

 

 

MODULATION : AM, FM & SWP=Release,

 

 

 

 

 

 

 

 

 

 

 

MODULATION Output=Max, SYM=CAL.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Puncture [ HI POT] Test.

 

 

Connect AC cord's GND, Life & Neutral pins respective.

 

 

 

 

GPT-515A :

 

 

 

 

 

 

 

 

 

 

 

 

O/P voltage

= 1.5 KV

 

 

 

 

 

 

 

 

 

 

 

Cutoff Current = 1 mA

 

 

 

 

 

 

 

 

 

 

 

Timer

= 60 Sec

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

Functioning Check.

 

 

Check the function of all VRs & button switches

 

 

 

 

 

20Mhz O'scope

 

 

 

 

corresponing to the output waveform.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4

Burn - in.

 

 

Deliver the sets to the burn-in chamber.

 

 

50 / 4 Hours

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GOODWILL INSTRUMENT

 

APPROVED BY

CHECKED BY

PREPARED BY

 

 

 

APPLICABLE ODM MODEL :

(MALAYSIA) SDN. BHD.

 

 

 

 

 

 

 

 

 

FG-813

ENGINEERING DEPARTMENT

 

 

 

 

 

 

 

 

 

BK-3040

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FOR INTERNAL USE ONLY

 

MODEL: GFG-813

 

 

PAGE

:

3

 

 

 

CALIBRATION PROCESS

 

 

 

STATION :

1

of

5

 

 

 

 

 

 

 

 

 

 

 

NO

Adjustment

Location

 

Particular

 

 

Specification

 

 

Remarks

1

Initial Setting

 

POWER Switch = ON;

 

 

 

 

 

Instrument Used :

 

 

 

EXT COUNTER : release all buttons;

 

 

 

 

 

1. Dmm [GDM-8045G]

 

 

 

MAIN GENERATOR : AMPL=Max, ATT= Release,

 

 

 

 

2. 20Mhz O'scope

 

 

 

OFFSET & SYM= Push, TRIGGER PHASE=Free run;

 

 

 

 

3. 10 Mhz Standard Timebase

 

 

 

MODULATION : AM, FM & SWP=Release,

 

 

 

 

 

4. Leveled Sinewave

 

 

 

MODULATION Output=Max, SYM=CAL.

 

 

 

 

 

Generator [SG-503]

 

 

 

 

 

 

 

 

2

+ 17V Adjust.

VR101

Measure the output voltage of R101 lead (right) relative

 

+ 17 +/- 0.01V

 

DMM : 20 VDC

 

 

{GFG-427}

to GND, adjust VR101 to get +17.0V show on DMM.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

- 17V Adjust.

VR102

Measure the output voltage of R102 lead (left) relative

 

- 17 +/- 0.01V

 

DMM : 20 VDC

 

 

{GFG-427}

to GND, adjust VR102 to get -17.0V show on DMM.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4

Offset Null Adjust.

VR207

Dial scale=Max, Main Generator RANGE=100;

 

 

 

 

 

DMM : 200 mVDC

 

 

{GFG-428}

Measure the voltage drop on R237 or R238 relative to

 

 

 

 

 

 

 

 

 

GND by using DMM. Adjust VR207 to get a reading of

 

 

 

 

 

 

 

 

 

110mV.

 

 

 

 

 

 

 

 

5

EXT Counter

VR001

COUNTER=Ext, GATE Time=0.1sec, 1/10 @ 1/1

 

 

<= 15 mVrms

 

SG 503 : 15 mVrms

 

Sensitivity Adjust.

{GFG-429}

button=Release; Apply a 10 Mhz 15 mVrms Sinewave

 

 

 

 

Freq : 10.0 Mhz &

 

 

 

signal to EXT COUNTER Terminal, adjust VR001 until

 

 

 

 

30.0Mhz

 

 

 

the display is stable. Check the display steadiness of

 

 

 

 

 

 

 

 

 

multiple level X1, X.1 & X.01[SG 503] for EXT ATT=1/1

 

 

 

 

 

 

 

 

 

and X1 & X.1 for EXT ATT=1/10. Push 30M/10M button

 

 

 

 

 

 

 

 

 

to select 30M; Repeat the above checking procedures

 

 

 

 

 

 

 

 

 

for the range of 30 Mhz by apply 30Mhz 15 mVrms

 

 

 

 

 

 

 

 

 

sinewave signal to EXT COUNTER Terminal.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6

EXT Counter 10Mhz

VC001

GATE Time=1sec,10M/30M=Release;

 

 

10 Mhz +/- 100 hz

 

 

 

 

Timebase Adjust.

{GFG-429}

Apply 10 Mhz standard Time base to EXT COUNTER

 

 

 

 

 

 

 

 

 

Terminal, adjust VC001 until the counter reads 10Mhz.

 

 

 

 

 

 

GOODWILL INSTRUMENT

APPROVED BY

CHECKED BY

PREPARED BY

 

 

 

APPLICABLE ODM MODEL :

(MALAYSIA) SDN. BHD.

 

 

 

 

 

 

 

 

 

FG-813

ENGINEERING DEPARTMENT

 

 

 

 

 

 

 

 

 

BK-3040

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FOR INTERNAL USE ONLY

 

 

MODEL: GFG-813

 

PAGE

:

4

 

 

 

CALIBRATION PROCESS

 

 

 

STATION :

1

of

5

 

 

 

 

 

 

 

 

 

 

NO

Adjustment

Location

 

Particular

 

Specification

 

 

Remarks

7

Main Generator Output

 

 

Connect MAIN O/P terminal to the o'scope without 50

Without 50 Ohm Load :

 

O'scope Setting :

 

Level Check.

 

 

Ohm load, check output level of sinewave, trianglewave

20 ~ 25 Vpp

 

T/D = 1 mS

 

 

 

 

& squarewave. Repeat the checking procedures by

 

With 50 Ohm Load :

 

V/D = 5 V [Without Load]

 

 

 

 

connect the O'scope to MAIN O/P terminal through a

10 ~ 13 Vpp

 

V/D = 2 V [With Load]

 

 

 

 

50 Ohm load.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8

SYNC Output Level

 

 

Set GFG-813 as step 6, Connect SYNC OUTPUT to

Without 50 Ohm Load :

 

O'scope Setting :

 

Check.

 

 

O'scope, check output level of sinewave, trianglewave

> 1 Vpp

 

 

T/D = 1 mS

 

 

 

 

& squarewave. Repeat the checking procedures by

 

With 50 Ohm Load :

 

V/D=0.2V [Without Load]

 

 

 

 

connect the O'scope to MAIN O/P terminal through a

> 0.5 Vpp

 

 

V/D=0.2V [With Load]

 

 

 

 

50 Ohm load.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

9

Modulation Generator

 

 

MODULATION : RANGE=10K, Freq=Max;

 

 

 

 

 

 

 

Output Level Check.

 

 

Connect MOD. OUTPUT to O'scope, check sinewave,

 

 

 

 

 

 

 

 

 

trianglewave & squarewave output level.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GOODWILL INSTRUMENT

 

APPROVED BY

CHECKED BY

PREPARED BY

 

 

APPLICABLE ODM MODEL :

(MALAYSIA) SDN. BHD.

 

 

 

 

 

 

 

 

 

FG-813

ENGINEERING DEPARTMENT

 

 

 

 

 

 

 

 

 

BK-3040

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FOR INTERNAL USE ONLY

 

MODEL: GFG-813

 

 

PAGE

:

5

 

 

 

CALIBRATION PROCESS

 

 

 

STATION :

2

of

5

 

 

 

 

 

 

 

 

 

 

 

NO

Adjustment

Location

 

Particular

 

 

Specification

 

 

Remarks

1

Initial Setting.

 

MAIN GENERATOR : AMPL=Max, SYNC=Push,

 

 

 

 

Instrument Used :

 

 

 

OFFSET=Push, ATT=Release, TRIGGER PHASE =

 

 

 

 

1. Frequency Counter

 

 

 

Free Run;

 

 

 

 

 

 

[GFC-8130G]

 

 

 

MODULATION : AM, FM & SWP = Release.

 

 

 

 

 

2. 100 Mhz O'scope

 

 

 

 

 

 

 

 

 

2

Main Generator Dial

VR201

Dial Scale=Max, RANGE=1K, Counter EXT/INT=Release

 

< 0.5 %

 

 

100Mhz O'scope :

 

Scale MAX Squarewave

VR204

Main FUNCTION=Squarewave;Connect SYNC output

 

 

 

 

V/D=5V , T/D=2uS;

 

Symmetry Adjust.

{GFG-428}

to Freq Counter [GFC] & MAIN output to a 100Mhz

 

 

 

 

GFC-8130G Setting :

 

 

 

O'scope.Adjust VR201 so that the reading show on GFC

 

 

 

 

FUCTION=FREQ A

 

 

 

is 13.4Khz. Adjust SWEEP VAR of O'scope to get a

 

 

 

 

GATE TIME=Push & Min

 

 

 

positive halfcycle of 10 div show on CRT. Select the

 

 

 

 

TRI LEVEL=Push

 

 

 

the SLOPE to -ve,adjust VR204 to get a balanced

 

 

 

 

 

LPF,ATT,COUP=Release

 

 

 

waveform.

 

 

 

 

 

 

 

 

3

Main Generator Dial

VR205

O'scope sweep variable=lock;

 

 

 

 

 

100Mhz O'scope :

 

Scale MIN Squarewave

VR206

Dial Scal=Min, Adjust VR206 to get a 5 Div positive

 

 

 

 

V/D=5V , T/D=1mS;

 

Symmetry Adjust.

{GFG-428}

cycle and adjust VR205 to get 5 Div negative cycle.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4

Main Generator Dial

VR201

Loosen dial scale lock nut, adjuts dial scale to get a

 

 

 

 

 

GFC-8130G Setting :

 

Scale Accuracy Adjust.

{GFG-427}

balance on 1 & 10 scale; tighten the lock nut, turn dial

 

 

 

 

FUCTION=FREQ A,

 

 

 

scale to 10, adjust VR201 to get 10Mhz. Make sure

 

 

 

 

GATE TIME=Push & Min,

 

 

 

the output freq of dial scale Max is greater than 13Mhz.

 

 

 

 

TRI LEVEL=Push,

 

 

 

 

 

 

 

 

5

Main Generator Dial

VR205

Dial Scale=Min; Connect SYNC output to GFC input

 

94.5 to 95.5 Hz

 

100Mhz O'scope :

 

Scale MIN Frequency &

VR206

terminal & MAIN output to O'scope with BNC-BNC.

 

 

 

 

V/D=5V , T/D=0.5mS;

 

Squarewave Symmetry

{GFG-427}

Set O'scope T/D=0.5mS, adjust O'scope SWEEP VAR

 

 

 

 

GFC-8130G Setting :

 

Adjust.

 

to get a 10 Div positive cycle. Set O'scope triggering

 

 

 

 

FUCTION=FREQ A,

 

 

 

to -ve slope; adjust VR205 to get a 10 Div negative cycle.

 

 

 

 

GATE TIME=Push & Min,

 

 

 

Check GFC reading to make sure that the output freq

 

 

 

 

TRI LEVEL=Push,

 

 

 

is within 94.5 to 95.5 Hz. If not, re-adjust VR205 &

 

 

 

 

LPF,ATT,COUP=Release

 

 

 

VR206 with condition that the duty cycle must balance.

 

 

 

 

 

 

GOODWILL INSTRUMENT

APPROVED BY

CHECKED BY

PREPARED BY

 

 

 

APPLICABLE ODM MODEL :

(MALAYSIA) SDN. BHD.

 

 

 

 

 

 

 

 

 

FG-813

ENGINEERING DEPARTMENT

 

 

 

 

 

 

 

 

 

BK-3040

 

 

 

 

 

 

 

 

 

 

 

 