Getinge 56M, 56A Technical Data Sheet

Page 1
Technical Data Sheet
6002466802 Getinge Aquadis 56 56M, 56A
2021-12-14 EN REV.A
Page 2
The manual, or parts thereof, are a part of the product. This manual must not be modified, reproduced or copied without Getinge’s wrien consent. The content may not be reused by third parties or used for unauthorized purposes. Violators will be prosecuted. Additional copies of this manual can be obtained from Getinge for a fee.
© Copyright 2013 Getinge. Reprinting prohibited.
Getinge Disinfection AB Ljungadalsgatan 11 352 46 Växjö SWEDEN
Internet: www.getinge.com
Page 3

Contents

1 DIMENSIONS AND VOLUMES.................................................................................... 5
2 UTILITY REQUIREMENTS........................................................................................... 6
3 OPERATING AND ENVIRONMENTAL REQUIREMENTS........................................... 8
4 TECHNICAL DATA COMPONENTS........................................................................... 10
5 AIR BARRIER SEPARATION (OPTION).................................................................... 12
6 ELECTRICAL DATA.................................................................................................... 13
7 IT NETWORK.............................................................................................................. 15
Contents
1.1 External dimensions............................................................................................. 5
1.2 Chamber dimensions and volumes...................................................................... 5
1.3 Weight and floor loading...................................................................................... 5
1.4 Weight and size of package................................................................................. 5
2.1 Water, steam, drain, condensate connections..................................................... 6
2.2 Treated water limit values.................................................................................... 7
2.3 Detergent values.................................................................................................. 7
2.4 Material type and plate thickness......................................................................... 7
3.1 Operating requirements....................................................................................... 8
3.2 Requirements for storage during transport.......................................................... 8
3.3 Ventilation requirements....................................................................................... 8
3.4 Consumption........................................................................................................ 8
3.5 Heat emissions..................................................................................................... 9
5.1 Air barrier separation for 56M (option)............................................................... 12
5.2 Air barrier separation for 56A (option)................................................................ 12
6.1 Electrical coupling alternative - electrical heating.............................................. 13
6.2 Electrical coupling alternative - electrical heating turbo..................................... 13
6.3 Electrical coupling alternative - steam heating................................................... 14
6.4 Electrical coupling alternative - combined electrical and steam heating............ 14
7.1 Firewall and router settings................................................................................ 15
7.2 Intended information flow................................................................................... 16
7.3 Potential hazardous situation............................................................................. 17
Alphabetical index............................................................................................................ 18
3 (20)Technical Data Sheet Getinge Aquadis 56
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
Page 4
Contents
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
4 (20) Technical Data Sheet Getinge Aquadis 56
Page 5

1 DIMENSIONS AND VOLUMES

1 DIMENSIONS AND VOLUMES

1.1 External dimensions

Measurement 56M 56A
External width 650 mm (25 9/16 in)
External depth 690 mm (27 3/16 in) 760 mm (29 15/16 in)
Unit height
Total height
Height with CCU 2270 mm (89 3/8 in)
1
Height of front of machine recessed in a wall.
2
Total height of machine at the highest point.
1
2

1.2 Chamber dimensions and volumes

Measurement 56M 56A
Chamber width 555 mm (21 27/22 in)
Chamber depth 618 mm (24 21/64 in)
Chamber height 667 mm (26 17/64 in)
Effective chamber volume 229 l (61 gal)
Gross chamber volume 298 I (79 gal)
1950 mm (76 49/64 in)
2000 mm (78 47/64 in)

1.3 Weight and floor loading

Measurement 56M 56A
Total machine weight 232 kg (511 lb) 270 kg (595 lb)
Total weight incl. water and load 332 kg (732 lb) 370 kg (816 lb)
Maximum weight of wash cart and load 75 kg (165 lb)
Load per machine foot (4 pieces) 58 kg (128 lb) 67.5 kg (149 lb)
Specific surface load
Floor load per machine foot
7.27 kN/m2 (152 lbf/ft2) 7.35 kN/m2 (154 lbf/ft2)
470.8 kN/m
(9832.9 lbf/ft2)
2
524.7 kN/m
(10958.6 lbf/ft2)

1.4 Weight and size of package

Measurement 56M 56A
Transport package weight 269 kg (593 lb) 307 kg (676 lb)
Transport package width 817 mm (32 11/64 in)
Transport package depth 967 mm (38 5/64in)
Transport package height 2190 mm (86 7/32 in)
2
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
5 (20)Technical Data Sheet Getinge Aquadis 56
Page 6

2 UTILITY REQUIREMENTS

2
UTILITY REQUIREMENTS

2.1 Water, steam, drain, condensate connections

Measurement Connection Pressure Flow rate Maximum
temperature
Cold water (CW)
Hot water (HW) Top:
Treated water (option)
Drain Adaptable:
ø 60 mm (2 23/64”)
Steam (option)
4
ISO G- 1/2 “ Male
Condensate (option)
Machine
ISO G- 1/2 “ Male
1xø76 mm (1xø3”) -
exhaust
Building exhaust
3
Maximum back pressure from building drain.
4
To ensure proper operation, incoming steam must be pressure regulated, saturated and filtered to 5
The machine must be supplied with sufficient ventilation to keep the temperature at<40 °C (104 °F) in the space above the machine.
microns.
5
Pressure difference between supply and return lines must be >100 kPa.
6
Steam pressure affects the process time. Low pressure extends program time.
7
Maximum return pressure.
Top:
ISO G- 3/4 “ Male
Floor:
ISO G- 1/2
“ Female
ISO G- 3/4 “ Male
Floor:
ISO G- 1/2
“ Female
Top:
ISO G- 3/4 “ Male
Floor:
ISO G- 1/2
“ Female
ø 50 mm (2”)
Floor:
Floor:
200-600 kPa
(29-87 Psi)
200-600 kPa
(29-87 Psi)
200-600 kPa
(29-87 Psi)
800 kPa (116 Psi)
300-500 kPa
5
(44-73 Psi)
30 kPa
5
(4.3 Psi)
Minimum:
30 L/min
20 °C
(68 °F)
(8 gal./min)
Minimum:
30 L/min
45-60 °C
(113-140 °F)
(8 gal./min)
Minimum:
30 L/min
60 °C
(140 °C)
(8 gal./min)
3
37.5 L/min
(9.9 gal./min)
90 °C (194 °F)
60 °C (140°F)
With drain cooling:
61.2 L/min
(16.2 gal./min)
6
0.9-1.0 kg/min
160 °C (320°F)
(2-4.4 lb./min)
7
- -
160 m3/h
90 °C (194 °F)
during drying phase
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
6 (20) Technical Data Sheet Getinge Aquadis 56
Page 7

2.2 Treated water limit values

2 UTILITY REQUIREMENTS
Measurement Unit Drinking water Softened
DiW/RO
8
pH value N/A 6.5 - 8.5 6.5 - 8.5 5 - 7.5
Hardness mg/l < 120 < 10 N/A
Conductivity (at 25 °C) μS/cm < 700 < 700 ≤ 15
Chlorine (Cl) mg/l < 60 < 60 ≤ 0.2
Iron mg/l < 0.3 < 0.3 ≤ 0.1
Copper mg/l < 0.1 < 0.1 ≤ 0.1
Silicon dioxide (SiO2) mg/l ≤ 2 ≤ 2 ≤ 1
Free chlorine mg/l < 0.5 < 0.5 ≤ 0.5
Phosphate mg/l < 10 < 10 ≤ 0.5
8
DiW/RO = Treated water/Reverse osmosis.

2.3 Detergent values

Measurement All models
Dosage volume
Viscosity
9
PH value
9
Recommended values to achieve accuracy of ± 5 % of nominal dosed volume in detergent system.
10
The machine is designed to be compatible with the most common detergents used for washer-
9
10
disinfectors but was type-tested with Getinge Clean products.
10 ml - 750 ml (concentration 0.5 ml/l - 30 ml/l)
1 cSt - 10 cSt at 20 °C
1 - 13.5

2.4 Material type and plate thickness

Measurement Chamber Exterior panels Frame work
Material type Stainless steel
EN 1.4404/ASTM 316L
Thickness 1.25 mm (3/64 in) 1.0 mm (3/64 in) -
Surface Bright annealed Brushed Galvanized
Stainless steel
EN 1.4301/ASTM 304
Steel
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
7 (20)Technical Data Sheet Getinge Aquadis 56
Page 8

3 OPERATING AND ENVIRONMENTAL REQUIREMENTS

3
OPERATING AND ENVIRONMENTAL REQUIREMENTS

3.1 Operating requirements

Measurement All models
Air humidity during operation Maximum 80 % at 31 °C (88 °F)
decreasing linearly to 50 % at 40 °C (104 °F)
Surrounding temperature during operation 5 °C - 40 °C (41 °F - 104 °F)
Maximum surface temperature 50 °C (123 °F)
Maximum installation height
Pollution degree 2
Overvoltage category CAT II
Protection class
Connection to mains supply Permanent
Mains supply voltage +/-10 %
Sound levels
11
When installing at higher altitudes >500 m (1640ft) above sea level, modification must be made
due to low air pressure.
12
The IP rate classifies the degree of protection against the intrusion of objects and water in electrical enclosures. IP21 means the device is protected from contact with fingers and objects larger than 12 mm, from dripping water.
13
The sound level values apply for sensitivity programs.
12
13
11
3000 m (9842.5 ft) above sea level
IP21
66 dB, LwA (washing)
70 dB, LwA (drying)
67 dB, LwA (average during program)

3.2 Requirements for storage during transport

Measurement All models
Maximum air humidity during transport and storage 90 % at 30 °C (86 °F)
Temperature during transport and storage -20 °C - 60 °C (-4 °F - 140 °F)

3.3 Ventilation requirements

Measurement All models
Maximum air quantity for ventilation
Maximum temperature for ventilation 90 °C (194 °F)
Air humidity for ventilation 60-100 %
160 m3/h

3.4 Consumption

Measurement All models
Water consumption 16 - 28 l/phase (3.9 - 7.4 gal./phase)
Energy consumption 3.5 - 7 kWh
Steam consumption 8.6 -10.7 kg/process
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
8 (20) Technical Data Sheet Getinge Aquadis 56
Page 9
3 OPERATING AND ENVIRONMENTAL REQUIREMENTS

3.5 Heat emissions

Measurement All models
Maximum heat emission
14, 15
Maximum temperature of cover plates 50 °C (123 °F)
Heat emission from the front and doors (when a program is running)
15
Heat emission from open door (after program
15
end)
Heat emission from wash cart (after program end)
14
Applies at ambient temperature of 20 ± 2°C (68 ± 36°F).
15
Measured during continuous operation of the machine, other operating conditions and programs
15
affect heat emissions. Conditions that also affect heat emissions are machine frequency used, type of program used, additional equipment used, type of items to be processed, and amount of loaded items.
16
Estimated value that decreases with time.
17
Measured using a wash cart with a total weight of 75 kg (165 lb). The wash cart emits heat to the
environment for one hour during cooling.
1 550 W
460 W (clean side)
1100 W (soiled side)
1000 W
350 W
16
17
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
9 (20)Technical Data Sheet Getinge Aquadis 56
Page 10

4 TECHNICAL DATA COMPONENTS

4
TECHNICAL DATA COMPONENTS
WATER CIRCULATION SYSTEM
Measurement All models
Design pressure 200 kPa (20.0 Psi)
Maximum operating pressure 100 kPa (14.5 Psi)
Design temperature 100 °C (212 °F)
Operating temperature 93 °C (200 °F)
CIRCULATION MAIN PUMP (WASH CART)
Measurement All models
Maximum flow 600 L/min ( 158 gpm)
Motor 1.0 kW (1.34 HP)
Material of construction SPS, Silicon carbide, Carbon, EPDM, AISI 304
CIRCULATION SECONDARY PUMP (CHAMBER)
Measurement All models
Maximum flow 80 L/min ( 21 gpm)
Motor 0.25 kW (0.34 HP)
Material of construction SPS, Silicon carbide, Carbon, EPDM, AISI 304
DRAIN PUMP
Measurement All models
Maximum flow 60 L/min (16 gpm)
Motor 0.19 kW (0.25 HP)
Material of construction SPS, Silicon carbide, Carbon, EPDM, AISI 304
DOSAGE PUMP
Measurement All models
Maximum flow
HEATER STEAM
Measurement Depend on steam pressure
Heating velocity 13.7 °C/min at 5 Bar
HEATER ELECTRICAL
Measurement Depends on voltage
Heating velocity 4.1 - 9.3 °C/min
Installed power 7.5 kW - 16 kW
0.3 l/min (0.305 in3/s)
15 kW - 32 kW (option)
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
10 (20) Technical Data Sheet Getinge Aquadis 56
Page 11
4 TECHNICAL DATA COMPONENTS
HEATER COMBINED
Measurement Depends on steam pressure and voltage
Heating velocity 17.8 °C/min at 5 Bar
Installed power 7.5 kW - 16 kW
DRYER
Measurement Depends on voltage
Installed power, heaters 3 - 5.8 kW
Installed power, motors 1.1 kW
CONTROL SYSTEM
Measurement All models
Memory capacity 99 program places
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
11 (20)Technical Data Sheet Getinge Aquadis 56
Page 12

5 AIR BARRIER SEPARATION (OPTION)

007278
10 20 30 40 50
0
10
20
30
40
50
60
70
80
90
Pressure (Pa)
Standard
Option
Standard + Open door Clean side
Option + Open door Clean side
Flow (l/s)
Standard
Option
Standard + Open door Clean side
Option + Open door Clean side
007277
Flow (l/s)
140
120
100
80
60
40
20
10 20 30 40 50
0
Pressure (Pa)
5
AIR BARRIER SEPARATION (OPTION)
A tolerance of + 10 % applies to cart values. The barrier conversion in relation to the area, or the total front area in the 56-series, is 1.26 m2 (13.56 ft2).

5.1 Air barrier separation for 56M (option)

5.2 Air barrier separation for 56A (option)

<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
12 (20) Technical Data Sheet Getinge Aquadis 56
Page 13

6 ELECTRICAL DATA

6 ELECTRICAL DATA

6.1 Electrical coupling alternative - electrical heating

Voltage (V)
480 3+PE 60 22.0 C30
415 3N+PE 50 16.3 C20
400 3N+PE 60 16.9 C20
400 3N+PE 50 16.9 C20
380 3N+PE 60 17.2 C20
380 3N+PE 50 17.2 C20
240 3+PE 60 22.1 C30
230 3+PE 50 22.9 C32
220 3+PE 60 23.8 C32
208 3+PE 60 26.6 C30
200 3+PE 60 27.5 C32
200 3+PE 50 27.5 C32
Connection Frequency
(Hz)
Maximum operating current (A)
Maximum fuse (A)

6.2 Electrical coupling alternative - electrical heating turbo

Voltage (V)
480 3+PE 60 41.3 C50
415 3N+PE 50 30.0 C40
400 3N+PE 60 31.1 C40
400 3N+PE 50 31.1 C40
380 3N+PE 60 32.1 C40
380 3N+PE 50 32.1 C40
240 3+PE 60 40.1 C60
230 3+PE 50 41.7 C60
220 3+PE 60 43.5 C50
208 3+PE 60 49.1 C60
200 3+PE 60 50.9 C60
200 3+PE 50 50.9 C60
Connection Frequency
(Hz)
Maximum operating current (A)
Maximum fuse (A)
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
13 (20)Technical Data Sheet Getinge Aquadis 56
Page 14
6 ELECTRICAL DATA

6.3 Electrical coupling alternative - steam heating

Voltage (V)
415 3N+PE 50 18.9 C20
400 3N+PE 60 18.1 C20
400 3N+PE 50 18.1 C20
380 3N+PE 60 17.3 C20
380 3N+PE 50 17.3 C20
240 3+PE 60 18.9 C25
230 3+PE 50 18.1 C25
220 3+PE 60 17.3 C30
208 3+PE 60 16.5 C30
200 3+PE 60 15.9 C32
200 3+PE 50 15.9 C32
Connection Frequency
(Hz)
Maximum operating current (A)
Maximum fuse (A)

6.4 Electrical coupling alternative - combined electrical and steam heating

Voltage (V)
480 3+PE 60 22.0 C30
415 3N+PE 50 16.3 C20
400 3N+PE 60 16.9 C20
400 3N+PE 50 16.9 C20
380 3N+PE 60 17.2 C20
380 3N+PE 50 17.2 C20
240 3+PE 60 22.1 C30
230 3+PE 50 22.9 C32
220 3+PE 60 23.8 C32
208 3+PE 60 26.6 C30
200 3+PE 60 27.5 C32
200 3+PE 50 27.5 C32
Connection Frequency
(Hz)
Maximum operating current (A)
Maximum fuse (A)
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
14 (20) Technical Data Sheet Getinge Aquadis 56
Page 15

7 IT NETWORK

7 IT NETWORK
Ethernet communications up to 100 Mbit/s, half-duplex.
Standard DHCP and IP addressing technologies.
Routing and firewalling: blocking of specific ports and traffic.
If T-DOC is used: either fixed IP addresses or a DHCP server with reserved IP addresses, to provide the same IP address to the G1 controller, as the IP address that is used by T-DOC in order to find the G1 controller on the network.
If Getinge Online is used: DHCP with access to a DNS server and a route to the Internet.
The IT network must be protected from unauthorized access. The responsible organization must take necessary steps to ensure that the G1 medical device on the network is not subjected to probing, attacks, or other non-essential traffic.

7.1 Firewall and router settings

Port and protocol Direction Usage Comment
3332/tcp Inbound to this port and
Outbound from this port to sender port.
68/udp Inbound to this port and
Outbound from this port to sender port.
21/tcp FTP Outbound to this port
on a Network Printer or FTP Server, and Inbound from this port to the medical device.
ICMP echo request, ICMP echo reply
445/tcp 137/udp 139/tcp
443/tcp Outbound to this port to
Configured port/tcp Outbound to the
Outbound echo request and Inbound echo reply to and from the IP address of the network printer.
Outbound to these ports on a Network Storage server (SMB/ CIFS), and Inbound response from these ports to the medical device.
a server on the Internet. Inbound from this server to the medical device.
configured port to a local proxy server.
T-DOC connection Enable if using T-DOC
DHCP Enable if using
Reserved IP address from a DHCP server.
Network Printing Network Storage (FTP Server)
Network Printing Network Storage (FTP Server)
Network Storage (CIFS Server)
Network Storage (CIFS Server)
Getinge Online Enable when using the
Getinge Online Enable when using
Enable if using the network printing option or if using the Network Storage with FTP Server option.
Enable if using the network printing option or if using the Network Storage with FTP Server or CIFS Server option.
Enable if using Network Storage with CIFS server option.
Getinge Online feature.
Getinge Online with a local proxy server.
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
15 (20)Technical Data Sheet Getinge Aquadis 56
Page 16
7 IT NETWORK
Port and protocol Direction Usage Comment
Inbound from this port
to the medical device.
17002/tcp Outbound to this port
on a Getinge Online Global Access Server.
DNS Outbound to local DNS
server as configured by the DHCP assignment. Inbound from this server.
Configuration. HTTP
CONNECT default port I 8080, and SOCKS default port is 1080, but are user-configurable.
Getinge Online Enable when using the
Getinge Online feature.
Getinge Online Enable when using the
Getinge Online feature.

7.2 Intended information flow

Feature Connection initiatedbyInformation flow Contents
T-DOC T-DOC program on
Windows PC on IT network.
Getinge Online G1 Controller G1 Controller to
Network Printing G1 Controller G1 Controller to printer
Network Storage G1 Controller G1 Controller to
G1 Controller to T-DOC program.
Getinge Online servers on the Internet.
located on the IT network.
network server providing storage services on the IT network.
Operating status of the G1 Controller, continuous process data, and final process reports. The T-DOC client will request: Status updates every second (1.5 Kbytes) and process reports at the end of each process cycle (1-2 Mbytes).
Operating status of the G1 Controller, continuous process data, and final process reports.
PDF files containing the final process reports from each completed process cycle and other types of user-prompted reports.
PDF files containing the final process reports from each completed process cycle and other types of user-prompted reports.
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
16 (20) Technical Data Sheet Getinge Aquadis 56
Page 17
7 IT NETWORK

7.3 Potential hazardous situation

Failure mode Potential hazard Comment
Failure to prevent flooding attacks, for example flood of packets to open ports on the device.
Loss of availability of the network interface and potential loss of availability of the washer-disinfector.
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
17 (20)Technical Data Sheet Getinge Aquadis 56
Page 18
Alphabetical index
Alphabetical index
AIR BARRIER SEPARATION (OPTION) ........................................................................ 12
Air barrier separation for 56A (option) ............................................................................. 12
Air barrier separation for 56M (option) ............................................................................ 12
Chamber dimensions and volumes ................................................................................... 5
Consumption ..................................................................................................................... 8
Detergent values ............................................................................................................... 7
DIMENSIONS AND VOLUMES ........................................................................................ 5
Electrical coupling alternative - combined electrical and steam heating ......................... 14
Electrical coupling alternative - electrical heating ........................................................... 13
Electrical coupling alternative - electrical heating turbo .................................................. 13
Electrical coupling alternative - steam heating ................................................................ 14
ELECTRICAL DATA ........................................................................................................ 13
External dimensions .......................................................................................................... 5
Firewall and router settings ............................................................................................. 15
Heat emissions .................................................................................................................. 9
Intended information flow ................................................................................................ 16
IT NETWORK .................................................................................................................. 15
Material type and plate thickness ...................................................................................... 7
OPERATING AND ENVIRONMENTAL REQUIREMENTS ............................................... 8
Operating requirements ..................................................................................................... 8
Potential hazardous situation .......................................................................................... 17
Requirements for storage during transport ....................................................................... 8
TECHNICAL DATA COMPONENTS ............................................................................... 10
Treated water limit values ................................................................................................. 7
UTILITY REQUIREMENTS ............................................................................................... 6
Water, steam, drain, condensate connections .................................................................. 6
Weight and floor loading .................................................................................................... 5
Weight and size of package .............................................................................................. 5
Ventilation requirements .................................................................................................... 8
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
18 (20)
Technical Data Sheet Getinge Aquadis 56
Page 19
Page 20
Manufacturer: Getinge Disinfection AB Ljungadalsgatan 11 352 46 Växjö SWEDEN
www.getinge.com
2021-12-14 EN REV.A