External depth690 mm (27 3/16 in)760 mm (29 15/16 in)
Unit height
Total height
Height with CCU2270 mm (89 3/8 in)
1
Height of front of machine recessed in a wall.
2
Total height of machine at the highest point.
1
2
1.2 Chamber dimensions and volumes
Measurement56M56A
Chamber width555 mm (21 27/22 in)
Chamber depth618 mm (24 21/64 in)
Chamber height667 mm (26 17/64 in)
Effective chamber volume229 l (61 gal)
Gross chamber volume298 I (79 gal)
1950 mm (76 49/64 in)
2000 mm (78 47/64 in)
1.3 Weight and floor loading
Measurement56M56A
Total machine weight232 kg (511 lb)270 kg (595 lb)
Total weight incl. water and load332 kg (732 lb)370 kg (816 lb)
Maximum weight of wash cart and load75 kg (165 lb)
Load per machine foot (4 pieces)58 kg (128 lb)67.5 kg (149 lb)
Specific surface load
Floor load per machine foot
7.27 kN/m2 (152 lbf/ft2)7.35 kN/m2 (154 lbf/ft2)
470.8 kN/m
(9832.9 lbf/ft2)
2
524.7 kN/m
(10958.6 lbf/ft2)
1.4 Weight and size of package
Measurement56M56A
Transport package weight269 kg (593 lb)307 kg (676 lb)
Transport package width817 mm (32 11/64 in)
Transport package depth967 mm (38 5/64in)
Transport package height2190 mm (86 7/32 in)
2
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
5 (20)Technical Data Sheet Getinge Aquadis 56
Page 6
2 UTILITY REQUIREMENTS
2
UTILITY REQUIREMENTS
2.1 Water, steam, drain, condensate connections
MeasurementConnectionPressureFlow rateMaximum
temperature
Cold water
(CW)
Hot water (HW)Top:
Treated water
(option)
DrainAdaptable:
ø 60 mm (2 23/64”)
Steam (option)
4
ISO G- 1/2 “ Male
Condensate
(option)
Machine
ISO G- 1/2 “ Male
1xø76 mm (1xø3”)-
exhaust
Building
exhaust
3
Maximum back pressure from building drain.
4
To ensure proper operation, incoming steam must be pressure regulated, saturated and filtered to 5
The machine must be supplied with sufficient ventilation to keep the temperature
at<40 °C (104 °F) in the space above the machine.
microns.
5
Pressure difference between supply and return lines must be >100 kPa.
6
Steam pressure affects the process time. Low pressure extends program time.
7
Maximum return pressure.
Top:
ISO G- 3/4 “ Male
Floor:
ISO G- 1/2
“ Female
ISO G- 3/4 “ Male
Floor:
ISO G- 1/2
“ Female
Top:
ISO G- 3/4 “ Male
Floor:
ISO G- 1/2
“ Female
ø 50 mm (2”)
Floor:
Floor:
200-600 kPa
(29-87 Psi)
200-600 kPa
(29-87 Psi)
200-600 kPa
(29-87 Psi)
800 kPa
(116 Psi)
300-500 kPa
5
(44-73 Psi)
30 kPa
5
(4.3 Psi)
Minimum:
30 L/min
20 °C
(68 °F)
(8 gal./min)
Minimum:
30 L/min
45-60 °C
(113-140 °F)
(8 gal./min)
Minimum:
30 L/min
60 °C
(140 °C)
(8 gal./min)
3
37.5 L/min
(9.9 gal./min)
90 °C (194 °F)
60 °C (140°F)
With drain cooling:
61.2 L/min
(16.2 gal./min)
6
0.9-1.0 kg/min
160 °C (320°F)
(2-4.4 lb./min)
7
--
160 m3/h
90 °C (194 °F)
during drying phase
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
6 (20)Technical Data Sheet Getinge Aquadis 56
Page 7
2.2 Treated water limit values
2 UTILITY REQUIREMENTS
MeasurementUnitDrinking waterSoftened
DiW/RO
8
pH valueN/A6.5 - 8.56.5 - 8.55 - 7.5
Hardnessmg/l< 120< 10N/A
Conductivity (at 25 °C)μS/cm< 700< 700≤ 15
Chlorine (Cl)mg/l< 60< 60≤ 0.2
Ironmg/l< 0.3< 0.3≤ 0.1
Coppermg/l< 0.1< 0.1≤ 0.1
Silicon dioxide (SiO2)mg/l≤ 2≤ 2≤ 1
Free chlorinemg/l< 0.5< 0.5≤ 0.5
Phosphatemg/l< 10< 10≤ 0.5
8
DiW/RO = Treated water/Reverse osmosis.
2.3 Detergent values
MeasurementAll models
Dosage volume
Viscosity
9
PH value
9
Recommended values to achieve accuracy of ± 5 % of nominal dosed volume in detergent system.
10
The machine is designed to be compatible with the most common detergents used for washer-
9
10
disinfectors but was type-tested with Getinge Clean products.
10 ml - 750 ml (concentration 0.5 ml/l - 30 ml/l)
1 cSt - 10 cSt at 20 °C
1 - 13.5
2.4 Material type and plate thickness
MeasurementChamberExterior panelsFrame work
Material typeStainless steel
EN 1.4404/ASTM 316L
Thickness1.25 mm (3/64 in)1.0 mm (3/64 in)-
SurfaceBright annealedBrushedGalvanized
Stainless steel
EN 1.4301/ASTM 304
Steel
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
7 (20)Technical Data Sheet Getinge Aquadis 56
Page 8
3 OPERATING AND ENVIRONMENTAL REQUIREMENTS
3
OPERATING AND ENVIRONMENTAL REQUIREMENTS
3.1 Operating requirements
MeasurementAll models
Air humidity during operationMaximum 80 % at 31 °C (88 °F)
decreasing linearly to 50 % at 40 °C (104 °F)
Surrounding temperature during operation5 °C - 40 °C (41 °F - 104 °F)
Maximum surface temperature50 °C (123 °F)
Maximum installation height
Pollution degree2
Overvoltage categoryCAT II
Protection class
Connection to mains supplyPermanent
Mains supply voltage+/-10 %
Sound levels
11
When installing at higher altitudes >500 m (1640ft) above sea level, modification must be made
due to low air pressure.
12
The IP rate classifies the degree of protection against the intrusion of objects and water in
electrical enclosures. IP21 means the device is protected from contact with fingers and objects larger
than 12 mm, from dripping water.
13
The sound level values apply for sensitivity programs.
12
13
11
3000 m (9842.5 ft) above sea level
IP21
66 dB, LwA (washing)
70 dB, LwA (drying)
67 dB, LwA (average during program)
3.2 Requirements for storage during transport
MeasurementAll models
Maximum air humidity during transport and storage90 % at 30 °C (86 °F)
Temperature during transport and storage-20 °C - 60 °C (-4 °F - 140 °F)
3.3 Ventilation requirements
MeasurementAll models
Maximum air quantity for ventilation
Maximum temperature for ventilation90 °C (194 °F)
Air humidity for ventilation60-100 %
160 m3/h
3.4 Consumption
MeasurementAll models
Water consumption16 - 28 l/phase (3.9 - 7.4 gal./phase)
Energy consumption3.5 - 7 kWh
Steam consumption8.6 -10.7 kg/process
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
8 (20)Technical Data Sheet Getinge Aquadis 56
Page 9
3 OPERATING AND ENVIRONMENTAL REQUIREMENTS
3.5 Heat emissions
MeasurementAll models
Maximum heat emission
14, 15
Maximum temperature of cover plates50 °C (123 °F)
Heat emission from the front and doors (when a
program is running)
15
Heat emission from open door (after program
15
end)
Heat emission from wash cart (after program end)
14
Applies at ambient temperature of 20 ± 2°C (68 ± 36°F).
15
Measured during continuous operation of the machine, other operating conditions and programs
15
affect heat emissions. Conditions that also affect heat emissions are machine frequency used, type
of program used, additional equipment used, type of items to be processed, and amount of loaded
items.
16
Estimated value that decreases with time.
17
Measured using a wash cart with a total weight of 75 kg (165 lb). The wash cart emits heat to the
environment for one hour during cooling.
1 550 W
460 W (clean side)
1100 W (soiled side)
1000 W
350 W
16
17
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
9 (20)Technical Data Sheet Getinge Aquadis 56
Page 10
4 TECHNICAL DATA COMPONENTS
4
TECHNICAL DATA COMPONENTS
WATER CIRCULATION SYSTEM
MeasurementAll models
Design pressure200 kPa (20.0 Psi)
Maximum operating pressure100 kPa (14.5 Psi)
Design temperature100 °C (212 °F)
Operating temperature93 °C (200 °F)
CIRCULATION MAIN PUMP (WASH CART)
MeasurementAll models
Maximum flow600 L/min ( 158 gpm)
Motor1.0 kW (1.34 HP)
Material of constructionSPS, Silicon carbide, Carbon, EPDM, AISI 304
CIRCULATION SECONDARY PUMP (CHAMBER)
MeasurementAll models
Maximum flow80 L/min ( 21 gpm)
Motor0.25 kW (0.34 HP)
Material of constructionSPS, Silicon carbide, Carbon, EPDM, AISI 304
DRAIN PUMP
MeasurementAll models
Maximum flow60 L/min (16 gpm)
Motor0.19 kW (0.25 HP)
Material of constructionSPS, Silicon carbide, Carbon, EPDM, AISI 304
DOSAGE PUMP
MeasurementAll models
Maximum flow
HEATER STEAM
MeasurementDepend on steam pressure
Heating velocity13.7 °C/min at 5 Bar
HEATER ELECTRICAL
MeasurementDepends on voltage
Heating velocity4.1 - 9.3 °C/min
Installed power7.5 kW - 16 kW
0.3 l/min (0.305 in3/s)
15 kW - 32 kW (option)
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
10 (20)Technical Data Sheet Getinge Aquadis 56
Page 11
4 TECHNICAL DATA COMPONENTS
HEATER COMBINED
MeasurementDepends on steam pressure and voltage
Heating velocity17.8 °C/min at 5 Bar
Installed power7.5 kW - 16 kW
DRYER
MeasurementDepends on voltage
Installed power, heaters3 - 5.8 kW
Installed power, motors1.1 kW
CONTROL SYSTEM
MeasurementAll models
Memory capacity99 program places
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
11 (20)Technical Data Sheet Getinge Aquadis 56
Page 12
5 AIR BARRIER SEPARATION (OPTION)
007278
1020304050
0
10
20
30
40
50
60
70
80
90
Pressure (Pa)
Standard
Option
Standard + Open
door Clean side
Option + Open
door Clean side
Flow (l/s)
Standard
Option
Standard + Open
door Clean side
Option + Open
door Clean side
007277
Flow (l/s)
140
120
100
80
60
40
20
1020304050
0
Pressure (Pa)
5
AIR BARRIER SEPARATION (OPTION)
A tolerance of + 10 % applies to cart values. The barrier conversion in relation to the area, or the total
front area in the 56-series, is 1.26 m2 (13.56 ft2).
5.1 Air barrier separation for 56M (option)
5.2 Air barrier separation for 56A (option)
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
12 (20)Technical Data Sheet Getinge Aquadis 56
Page 13
6 ELECTRICAL DATA
6 ELECTRICAL DATA
6.1 Electrical coupling alternative - electrical heating
Voltage
(V)
4803+PE6022.0C30
4153N+PE5016.3C20
4003N+PE6016.9C20
4003N+PE5016.9C20
3803N+PE6017.2C20
3803N+PE5017.2C20
2403+PE6022.1C30
2303+PE5022.9C32
2203+PE6023.8C32
2083+PE6026.6C30
2003+PE6027.5C32
2003+PE5027.5C32
ConnectionFrequency
(Hz)
Maximum operating current
(A)
Maximum fuse
(A)
6.2 Electrical coupling alternative - electrical heating turbo
Voltage
(V)
4803+PE6041.3C50
4153N+PE5030.0C40
4003N+PE6031.1C40
4003N+PE5031.1C40
3803N+PE6032.1C40
3803N+PE5032.1C40
2403+PE6040.1C60
2303+PE5041.7C60
2203+PE6043.5C50
2083+PE6049.1C60
2003+PE6050.9C60
2003+PE5050.9C60
ConnectionFrequency
(Hz)
Maximum operating current
(A)
Maximum fuse
(A)
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
13 (20)Technical Data Sheet Getinge Aquadis 56
Page 14
6 ELECTRICAL DATA
6.3 Electrical coupling alternative - steam heating
Voltage
(V)
4153N+PE5018.9C20
4003N+PE6018.1C20
4003N+PE5018.1C20
3803N+PE6017.3C20
3803N+PE5017.3C20
2403+PE6018.9C25
2303+PE5018.1C25
2203+PE6017.3C30
2083+PE6016.5C30
2003+PE6015.9C32
2003+PE5015.9C32
ConnectionFrequency
(Hz)
Maximum operating current
(A)
Maximum fuse
(A)
6.4 Electrical coupling alternative - combined electrical and steam
heating
Voltage
(V)
4803+PE6022.0C30
4153N+PE5016.3C20
4003N+PE6016.9C20
4003N+PE5016.9C20
3803N+PE6017.2C20
3803N+PE5017.2C20
2403+PE6022.1C30
2303+PE5022.9C32
2203+PE6023.8C32
2083+PE6026.6C30
2003+PE6027.5C32
2003+PE5027.5C32
ConnectionFrequency
(Hz)
Maximum operating current
(A)
Maximum fuse
(A)
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
14 (20)Technical Data Sheet Getinge Aquadis 56
Page 15
7 IT NETWORK
7 IT NETWORK
•Ethernet communications up to 100 Mbit/s, half-duplex.
•Standard DHCP and IP addressing technologies.
•Routing and firewalling: blocking of specific ports and traffic.
•If T-DOC is used: either fixed IP addresses or a DHCP server
with reserved IP addresses, to provide the same IP address to
the G1 controller, as the IP address that is used by T-DOC in
order to find the G1 controller on the network.
•If Getinge Online is used: DHCP with access to a DNS server
and a route to the Internet.
The IT network must be protected from unauthorized access. The
responsible organization must take necessary steps to ensure that
the G1 medical device on the network is not subjected to probing,
attacks, or other non-essential traffic.
7.1 Firewall and router settings
Port and protocolDirectionUsageComment
3332/tcpInbound to this port and
Outbound from this port
to sender port.
68/udpInbound to this port and
Outbound from this port
to sender port.
21/tcp FTPOutbound to this port
on a Network Printer or
FTP Server, and
Inbound from this port
to the medical device.
ICMP echo request,
ICMP echo reply
445/tcp
137/udp
139/tcp
443/tcpOutbound to this port to
Configured port/tcpOutbound to the
Outbound echo request
and Inbound echo reply
to and from the IP
address of the network
printer.
Outbound to these
ports on a Network
Storage server (SMB/
CIFS), and Inbound
response from these
ports to the medical
device.
a server on the Internet.
Inbound from this
server to the medical
device.
configured port to a
local proxy server.
T-DOC connectionEnable if using T-DOC
DHCPEnable if using
Reserved IP address
from a DHCP server.
Network Printing
Network Storage (FTP
Server)
Network Printing
Network Storage (FTP
Server)
Network Storage (CIFS
Server)
Network Storage (CIFS
Server)
Getinge OnlineEnable when using the
Getinge OnlineEnable when using
Enable if using the
network printing option
or if using the Network
Storage with FTP
Server option.
Enable if using the
network printing option
or if using the Network
Storage with FTP
Server or CIFS Server
option.
Enable if using Network
Storage with CIFS
server option.
Getinge Online feature.
Getinge Online with a
local proxy server.
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
15 (20)Technical Data Sheet Getinge Aquadis 56
Page 16
7 IT NETWORK
Port and protocolDirectionUsageComment
Inbound from this port
to the medical device.
17002/tcpOutbound to this port
on a Getinge Online
Global Access Server.
DNSOutbound to local DNS
server as configured by
the DHCP assignment.
Inbound from this
server.
Configuration. HTTP
CONNECT default port
I 8080, and SOCKS
default port is 1080, but
are user-configurable.
Network PrintingG1 ControllerG1 Controller to printer
Network StorageG1 ControllerG1 Controller to
G1 Controller to T-DOC
program.
Getinge Online servers
on the Internet.
located on the IT
network.
network server
providing storage
services on the IT
network.
Operating status of the
G1 Controller,
continuous process
data, and final process
reports. The T-DOC
client will request:
Status updates every
second (1.5 Kbytes)
and process reports at
the end of each process
cycle (1-2 Mbytes).
Operating status of the
G1 Controller,
continuous process
data, and final process
reports.
PDF files containing the
final process reports
from each completed
process cycle and other
types of user-prompted
reports.
PDF files containing the
final process reports
from each completed
process cycle and other
types of user-prompted
reports.
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
16 (20)Technical Data Sheet Getinge Aquadis 56
Page 17
7 IT NETWORK
7.3 Potential hazardous situation
Failure modePotential hazardComment
Failure to prevent flooding
attacks, for example flood of
packets to open ports on the
device.
Loss of availability of the network interface
and potential loss of availability of the
washer-disinfector.
<Doc_INS><Doc_6002466802><Rev.A><Lang_en>
17 (20)Technical Data Sheet Getinge Aquadis 56
Page 18
Alphabetical index
Alphabetical index
AIR BARRIER SEPARATION (OPTION) ........................................................................ 12
Air barrier separation for 56A (option) ............................................................................. 12
Air barrier separation for 56M (option) ............................................................................ 12
Chamber dimensions and volumes ................................................................................... 5