Rotor-Gene
Software Manual
(C)Copyright 2002, Corbett Research
Rotor-Gene I
Table of Contents
Foreword 0
Part I
Introduction
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .................................................................. 5 1 Important: Read Before Running The Rotor-Gene
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...................................................................................... 5 Rotor-Gene Keyboard (Rotor-Gene 2000 Only)
. . . . . . . . . . . . . . . . . . . . . . . . ................................................................................................................................. 5 Rotor-Gene Startup
. . . . . . . . . . . . . . . . . . . . ..................................................................................................................................... 6 Software Version
. . . . . . . . . . . . . . . . . . . ...................................................................................................................................... 7 Welcome Screen
5
Part II Rotor-Gene Description
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..................................................................................... 9 1 32-Well and Dual-Channel Rotor-Gene
. . . . . . . . . . . . . . . . . . . . . . . . . . . ........................................................................................................ 10 2 Multi-Channel Rotor-Gene
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ................................................................................... 11 3 Locking Ring (For 36 Well Rotor Only)
Part III
Two Different Rotor Systems
(Tubes)
. . . . . . . . . . . . . . . . . . . ................................................................................................................ 12 1 Rotor 36 Well System
. . . . . . . . . . . . . . . . . . . ................................................................................................................ 12 2 Rotor 72 Well System
Part IV Installation and Maintenance
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ................................................................................................. 12 1 Installation (Rotor-Gene 2000)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ................................................................................................. 13 2 Installation (Rotor-Gene 3000)
. . . . ............................................................................................................................... 13 3 Maintenance
Part V
Functional Overview
. . ................................................................................................................................. 13 1 Workspace
. . . . . . . . . . . . . . . . ................................................................................................................... 13 2 Toolbar Workspace
. . . . . . . . . . . . . . . . ................................................................................................................... 14 3 View Raw Channels
. . . . . . . . . . ......................................................................................................................... 14 4 Toggle Samples
................................................................................................................................... 15 5 File Menu
. . . . . . . . . . . . . . . . . . . . ..................................................................................................................................... 15 New Experiment
. . . . . . . . . . . . . . . . . . . . . . . . . . . .............................................................................................................................. 16 Opening and Saving
. . . . ..................................................................................................................................................... 16 Export
. . . ...................................................................................................................................................... 17 Setup
. . . . . . . . ........................................................................................................................... 18 6 Analysis Menu
. . . . . . . . . . . . . . . . . . . . . .................................................................................................................................... 18 Analysis Toolbar
. . . . . . . . . . . . . . ........................................................................................................................................... 18 Quantitation
. . . . . . ............................................................................................................................................ 19 Reports
. . . . . . . . . . . . . . . . . ................................................................................................................................. 19 Standard Curve
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ............................................................................................................... 20 Standard Curve Calculation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . ...................................................................................................................... 21 Import Standard Curve
. . . . . . . . . . . . . . . . . . ................................................................................................................................ 21 Invert Raw Data
9
12
12
13
© 2002 Corbett Research
. . . . . . . . . . . . . . . . . . . . .............................................................................................................................. 22 Calculation of CT
. . . . . ............................................................................................................................................. 23 Results
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....................................................................................................... 24 Dynamic Tube Normalisation
. . . . . . . . . . . . . . . . . . . . . . . . . . . ................................................................................................................ 24 Noise Slope Correction
. . . . . . . . . . ................................................................................................................................. 25 Ignore First
. . . . . . . . . . . . . . . . ........................................................................................................................... 25 Quant. Settings
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...................................................................................... 25 Slope, Amplification, Reaction Efficiency
. .......................................................................................................................................... 26 Offset
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................................................................ 27 Why Are Rotor-Gene Standard Curves Different?
. . . . . . . . . . . . . . .................................................................................................................................... 29 Main Window
. . . . . . . . . . . . . . . . . . . . . . . . . . ............................................................................................................................... 30 Melt Curve Analysis
. . . . . ............................................................................................................................................. 30 Sidebar
. . . . . . ............................................................................................................................................ 31 Reports
. . . . . ............................................................................................................................................. 31 Results
. . . . . . . . . . . ....................................................................................................................................... 31 Genotyping
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . ............................................................................................................................ 32 Allelic Discrimination
. . . . . . ............................................................................................................................................ 32 Reports
. . . . . ............................................................................................................................................. 32 Results
. . . . . . . . . . . . . . . . . . . . . . . . . . . ....................................................................................................................... 32 Normalisation Options
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ................................................................................................................... 33 Discrimination Threshold
. . . . . . . . . . ........................................................................................................................................ 33 Genotypes
. . . . . . . . . . . . . . . . . . . . ..................................................................................................................................... 33 Scatter Analysis
. . . . . . ............................................................................................................................................ 34 Reports
. . . . . ............................................................................................................................................. 34 Results
. . . . . . . . . . . . . . . . . . . . . . . . . . . ....................................................................................................................... 34 Normalisation Options
. . . . . . . . . . ........................................................................................................................................ 35 Genotypes
. . . . . . . . . . . . . . . ................................................................................................................................... 35 Scatter Graph
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...................................................................................................................... 35 Comparitive Quantitation
. . . . . . . . . . . . . . . . . . . . . . . . ................................................................................................................................. 37 EndPoint Analysis
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...................................................................................................... 38 Terms Used In EndPoint Analysis
. . . . . . . . . . . . . . . . . . . . . . . . .......................................................................................................................... 39 Profile Configuration
. . . . . . ............................................................................................................................................ 40 Analysis
. . . . . . . . . . . . . . . . . ................................................................................................................................. 40 Define Controls
. . . . . . . . . . . . . . .................................................................................................................................... 41 Normalisation
. . . . . . . . . ......................................................................................................................................... 43 Threshold
. . . . . . . . . . ........................................................................................................................................ 44 Genotypes
................................................................................................................................... 44 7 Run Menu
. . . . . . . . . ................................................................................................................................................ 44 Start Run
. . . . . . . . . . . .............................................................................................................................................. 44 Pause Run
. . . . . . . . . ................................................................................................................................................ 44 Stop Run
. .................................................................................................................................. 45 8 View Menu
. . . . . . . . . . . . . . . . . . . . . . . . . . . .............................................................................................................................. 45 Experiment Settings
. . . . . . ............................................................................................................................................ 45 General
. . . . . . . . . . . . . . . . . . . ............................................................................................................................... 46 Machine Options
. . . . . . . . . ......................................................................................................................................... 46 Messages
. . . . . . . . .......................................................................................................................................... 47 Channels
. . . . . . . . . . . . ...................................................................................................................................... 48 Tube Layout
. . . . . . ............................................................................................................................................ 49 Security
. . . . . . . . . . . . . . . . . . . . . . . . . . ............................................................................................................................... 50 Temperature Graph
. . . . . . . . . . . . . . . .......................................................................................................................................... 50 Profile Editor
.................................................................................................................................................. 51 Hold
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ................................................................................................................... 51 Optical Denature Cycling
II Contents
© 2002 Corbett Research
Rotor-Gene III
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .............................................................................................. 51 What is Optical Denature Cycling?
. . . . . . . . . . . . ............................................................................................................................... 53 Configuration
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....................................................................................... 53 Adding a New Optical Denature Cycling
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..................................................................... 55 Changing An Existing Step to Use Optical Denature
. . . . . ............................................................................................................................................. 56 Cycling
. . . . . . . . . . ........................................................................................................................................ 56 Acquisition
.................................................................................................................................................. 57 Melt
. . . . . . . . . . . . . . . . . . . . ..................................................................................................................................... 57 Profile Progress
. . . . . . . . . . . . . . . .......................................................................................................................................... 58 Edit Samples
. . . . . . . . . . . . . . . . . . . . ..................................................................................................................................... 60 Gain Calibration
. . . . . . . . . . . . . . . . . . ................................................................................................................................ 60 Auto-Calibration
. . . . . . . . . . . . . . . . . . . . . . ............................................................................................................................ 62 Manual Calibration
. . . . . . . . . . . . . . . . . . . . ..................................................................................................................................... 63 Display Options
. .................................................................................................................................. 64 9 Gain Menu
. . . . . . . ............................................................................................................................ 64 10 Window Menu
................................................................................................................................... 64 11 Help Menu
. . . . . . . . . . . . . . . . . . . . . . . . . . . .............................................................................................................................. 65 Send Support E-Mail
Part VI
General Functions Used Over
Several Windows
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ................................................................................................ 65 1 Opening A Second Experiment
. . . . . .............................................................................................................................. 66 2 Spanner Icon
. . . . . . . . . . ......................................................................................................................... 67 3 Scaling Options
. . . . . . ............................................................................................................................. 67 4 Abbreviations
. .................................................................................................................................. 67 5 Auto-Scale
Part VII Quick Start Using The New
Experiment Wizard
. . . . . . . . . . . ........................................................................................................................ 68 1 Welcome Screen
................................................................................................................................... 69 2 Page 1
................................................................................................................................... 70 3 Page 2
................................................................................................................................... 71 4 Page 3
................................................................................................................................... 72 5 Page 4
Part VIII
Rotor-Gene Hardware Information
. . . . . . . . . . . . . . . . ................................................................................................................... 72 1 Filter Specifications
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..................................................................................................... 72 32 Well and Dual-Channel Machines
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ........................................................................................................................ 73 Multi-Channel Machines
65
68
72
Part IX Troubleshooting
. . . . . .............................................................................................................................. 74 1 Log Archives
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .......................................................................................... 74 2 Troubleshooting Experiment Files
. . . . . . . . . . . . . . . . . . . . . . . . . . . .............................................................................................................................. 75 Initial Denature Step
. . . . . . . . . . . . . . . . . . ....................................................................................................................................... 75 Cycling Profile
. . . . . . . . . . . . . . . . . . . . . . . . . . ............................................................................................................................... 75 Melt Curve Analysis
74
© 2002 Corbett Research
. . . . . . . . . . . . . . . . ......................................................................................................................................... 76 Gain Settings
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..................................................................................................................... 76 General Setup Conditions
. . . . . . . . . . . . . . . . . . . . . . . . . . . .............................................................................................................................. 76 Experiment Settings
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ............................................................................................ 77 3 Regional Settings In Windows 98
IV Contents
Part X
Rotor-Gene 3000 Setup
Instructions
Part XI Rotor-Gene 2000 Setup
Instructions
Part XII
Rotor-Gene 2000 IMPORTANT
SETTINGS
Index
77
78
80
0
© 2002 Corbett Research
5 Rotor-Gene
Before running the Rotor-Gene you should pay attention to the following:
1) Always run the unit with 36 (32) or 72 tubes in the rotor. If you do not have this many samples load
empty tubes into the rotor. This ensures the thermal load in the chamber is identical for every run. If
0.1ml tubes are used it is sufficient to place empty tubes without lids into the rotor.
Do not use caps on the blank tubes. If caps are loaded on blank tubes and run multiple times
there is a risk that caps may come off the tube causing damage to the rotor.
To ensure that tubes sit firmly in the 72 well rotor, a 72 well locking ring is
available. After loading the tubes, the rotor can simply be put over the rotor.
2) Ensure the software is not operating in the Virtual Machine Mode. Go to the
Check that the Virtual Machine box is not ticked.
cannot be selected then it has been disabled and your distributor should have preset the
3) Before running a program ensure the Rotor-Gene shows
shown the machine will not set the Gains for the channels and no data will be collected.
The keyboard can be used to set the unit to a temperature and hold. In addition, the Rotor-Gene
keyboard display shows the lid status, either Open or Closed.
dvances the window forward to the next display and shows Lid Status then
keys to change the set temperature, once set press
The system then will go through a Start sequence that turns on the rotor, then the chamber fan and
The unit will heat or cool and hold the set temperature. This is useful if incubation at a particular
temperature is required before thermal cycling samples.
The keyboard can be used to set a manual temperature but is not used in regular operation.
The Rotor-Gene is setup using the following procedure:
1. Turn the desktop PC on and wait for Windows to boot up.
2. Turn on the Rotor-Gene on the left-hand side of the unit wait for the ready message to be displayed
1 Introduction
1.1 Important: Read Before Running The Rotor-Gene
1.1.1 Rotor-Gene Keyboard (Rotor-Gene 2000 Only)
1.1.2 Rotor-Gene Startup
© 2002 Corbett Research
3. Double click on the Rotor-Gene icon.
4. The Rotor-Gene is now ready for use.
Note for Rotor-Gene 2000 users: If the Rotor-Gene is turned ON without having Rotor-Gene
software loaded and running on the PC, the instrument will emit a buzzing sound. By running
the Rotor-Gene software the buzzing will stop.
1.1.3 Software Version
Software development for the Rotor-Gene system is ongoing. To check on your version number click
The latest software version is available for download at our web
This screen displays general information about the software, and specifically includes the version of
the software, serial number of the machine, as well as the date it was last updated.
6 Introduction
© 2002 Corbett Research
7 Rotor-Gene
This screen appears when a new version of the Rotor-Gene software has been installed and the
Rotor-Gene icon is double clicked for the first time after the installation.
Type in the serial number (six digits), which is located at the back of the
Type in the Offset coefficient, which can be found next to the serial number of the
machine. If no number can be found, the value 0 should be entered.
Chose your type of machine.
1.1.4 Welcome Screen
© 2002 Corbett Research
8 Introduction
Choose type of 36 well Rotor:
(for demonstration): Ticking the box allows installing the Rotor-Gene software
on a computer without Rotor-Gene attached. The software is fully functional and can even simulate
NOTE: If this box is ticked with a Rotor-Gene connected to your computer, a message will
appear before you start your run: "you about to run in virtual mode". To be able to perform a
"real" run, the setup (see setup) has to be changed.
When all parameters are set, press Begin. A window will come up initializing machine. Wait
until the machine is initialized, which might take a few seconds. If virtual mode was chosen the
following screen appears:
If the box stays unticked the machine is initialized and opens up the Rotor-Gene software
This topic explains the new features in this version of Rotor-Gene and answers
commonly asked questions about the location of existing features that have been moved. These
features incorporate design changes and requests by users.
© 2002 Corbett Research
9 Rotor-Gene
The Rotor-Gene is a real-time thermal cycling system used for DNA amplification and hybridization.
The unit holds 36 x 0.2ml standard micro-fuge tubes (32 x 0.2), which are loaded into a 36-position
rotor (32). With a Dual-Channel machine, a 72-well rotor is also supplied that can run special strip
tubes that are 0.1ml in strips of four. During the run the rotor spins at approximately 500 rpm as the
tubes are thermally cycled in a low-mass air oven.
NOTE: Always run a full rotor of tubes, for both the 72-well and 36-well rotors. If the number
of samples is small fill the remaining positions with empty tubes. This ensures the thermal
load of plastic tubes is the same from run to run.
There are two detection modules in the Rotor-Gene, Channel (1) is setup to detect 510nm and
Channel (2) is setup to detect 555nm. This allows for multiplexed samples to be run.
As the sample tubes spin past the detector modules, a high powered LED strobes the sample and a
photomultiplier (PMP) collects the fluorescent energy. See the figure below.
This data is sent to a Desktop PC that averages the energy of each sample over a number of
revolutions. This data is then displayed in real-time on the screen as fluorescence versus cycle
number or temperature plot.
2 Rotor-Gene Description
2.1 32-Well and Dual-Channel Rotor-Gene
© 2002 Corbett Research
2.2 Multi-Channel Rotor-Gene
The principal of the Multi-Channel Rotor-Gene is very similar to the 32-well and the Dual-Channel
Rotor-Gene. However, the Multi-Channel has one LED with four excitation filters.
From the side view of the heating cooling chamber we see the LED source irradiating the tube from the
side wall and the photomultiplier detecting the energy from the base of the chamber. The detection
filter wheel has six different filters, four are band-pass that are used in four Channel multiplex runs, the
other two are high pass filters used with other non-standard fluorophores.
10 Rotor-Gene Description
© 2002 Corbett Research
11 Rotor-Gene
Opening and closing 0.2 tubes several times before running might loosen the closing mechanism of
tubes. The 0.2 tubes run on the 36-well rotor on the Rotor-Gene should only be closed once before
running, otherwise it might result in popping of tubes.
Also make sure that the 0.2ml tubes are
closed properly by firmly pressing down the lid of the tube.
If popping is experienced or wants to
be avoided, a locking ring can be used on the Rotor-Gene.
The Locking Ring is designed to use
for flat top 0.2 ml tubes only.
The below graph shows schematically how the locking ring has to be used:
1) Shows the schematic representation of the 36-well rotor.
2) Screw down the screw with threads on both sides. Make sure the nipple and the red dot faces the
3) Simply place the locking ring over the top of the tubes.
4) Screw down the locking ring holder (4) over the screw with threads on both ends (2) until the
locking ring (3) is fixed.
2.3 Locking Ring (For 36 Well Rotor Only)
© 2002 Corbett Research
3 Two Different Rotor Systems (Tubes)
It is recommend that you use flat cap, 0.2ml tubes from Axygen
, supplied by Corbett Research.
If using your own tubes, be aware of the following:
1. The tube cap must fit down tightly otherwise there is a risk that the cap may 'pop' open during
thermal cycling. This has been observed with some domed capped 0.2ml tubes. If necessary the
locking ring could be used to secure the tubes.
2. Be sure that the tubes fit into the rotor firmly. Tubes that are excessively loose may vibrate during a
run and interfere with the data acquisition and give poor results.
3. In addition, ensure that the tubes are not too large. If the diameter of the tube is too large it will not fit
down far enough into the rotor. Samples will not be optically aligned over the detection system and may
cause a reduction in fluorescent signal acquired and sensitivity.
4. Do not use tubes from different manufacturers within the same run. This may result in
5. Locking Ring can only be used with
Corbett Research manufactures the 72-well tubes and caps in strips of four especially for the Rotor-
It is not advised to autoclave the 72-well tubes and strip caps as they may distort.
The tubes can become visibly deformed and are no longer straight but bent slightly. When running
tubes that have been autoclaved it has been noted that the caps may detach from the tube.
This has the effect of the loose caps being circulated around the chamber causing damage to the
insulation material. It is also advisable to exchange the empty 72-well tubes after at least ten runs.
Caps might start popping.
Do not use caps on the blank tubes. If caps are loaded on blank tubes and run multiple times
there is a risk that caps may come off the tube causing damage to the rotor.
The Rotor-Gene 2000 and the Desktop PC should be connected with the special "Y" mains cable
supplied. This special cable minimizes the connection length between the two devices.
NOTE FOR Rotor-Gene 2000 Users: If you do not set up the mains connection as described
you may see spikes in the collected data due to mains noise.
3.1 Rotor 36 Well System
12 Two Different Rotor Systems (Tubes)
3.2 Rotor 72 Well System
4 Installation and Maintenance
4.1 Installation (Rotor-Gene 2000)
© 2002 Corbett Research
13 Rotor-Gene
No special installation steps are required for the Rotor-Gene 3000. The Rotor-Gene 3000 requires
only a standard serial cable to be connected to a communications port on the back of the Laptop or
The only maintenance the user may have to perform is keeping the lenses relatively clean and free
from dust that may build up over time.
It is recommended that when the Rotor-Gene is not in use, the lid is in the closed position to minimize
dust falling into the heater chamber.
Looking down into the heating chamber you will notice one (36/72-Well, Multi-Channel) or two (36/72-
Well, Dual-Channel) round lenses, which are the detector objectives for the different Channels. Should
those lenses be covered with dust or any other debris you should clean them using a cotton bud and
On the side-wall of the heater chamber there are also one (36/72-Well, Multi-Channel) or two (36/72-
Well, Dual-Channel) round lenses for the high power diode emitter. These lenses are not likely to build
up dust, however, if you place tubes into the rotor that are dirty or wet you are likely to spin those
particles around the inner wall of the heater chamber and possibly produce a build up on these lenses.
If these lenses need to be cleaned, the rotor can be removed by unscrewing the black lock-down in the
center and removing the rotor. These lenses can be cleaned with a cotton bud and ethyl alcohol.
The following chapter will help to familiarize you with elements in the Rotor-Gene user interface.
The Rotor-Gene workspace is the backdrop of the main window. This is the area in which you can
open up graphs of raw data, temperature and analysis results. If you have several windows currently
open, you can organize them by clicking the
on the toolbar. There are several
additional options available, which you can access by clicking on the
These buttons are shortcuts to frequently used operations. To see what their function is, hover your
mouse over them and a brief explanation will appear. These commands can also be accessed via their
corresponding menu items with the same name.
4.2 Installation (Rotor-Gene 3000)
4.3 Maintenance
5 Functional Overview
5.1 Workspace
5.2 Toolbar Workspace
© 2002 Corbett Research
5.3 View Raw Channels
Click on these buttons to view different channels in the experiment.
Use this control to hide or show different samples. Semi-greyed samples are deselected. The
is used to display the next group of samples. You can toggle samples individually by clicking on
them, or you can hide/show all currently visible (on the current 'bank') by clicking on
. To select a range of samples, click on the first sample and drag your mouse to the end sample.
When you release the mouse button, the selected samples will either be toggled on or off. Clicking
will only show those samples you have given a name to; a quick way to show only relevant
will show all/ none 36 or 72 samples. Pressing the
button opens the samples window.
Toggle samples ID display:
If a 72-well rotor is used the samples are shown from A1 to A8, B1 to
B8, etc. Using the toggle samples ID display button lets the user switch to a numerical order of
5.4 Toggle Samples
14 Functional Overview
© 2002 Corbett Research
15 Rotor-Gene
This screen presents you with a selection of available experiment templates. In a laboratory
environment, you can set up standard templates by placing them in the Rotor-Gene template directory
(usually C:\Program Files\Rotor-Gene\Templates). These templates will then appear in this window.
Double-clicking this icon will create a new experiment based on the last experiment
run. To always begin with a blank experiment, choose File, Save As and save a template in the Rotor-
Gene template directory as "Blank Template.ret". Then, double-click this icon instead of New
Experiment when starting a new run.
Opens the template selection window to create a new experiment.
Pressing Browse... opens a window in which you can select the template you would like to
use for the next run. This option is also available under Other Templates.
This icon has the same effect as clicking the Browse... button, and displays an
open dialogue to load custom experiment templates.
Use Experiment Wizard to Configure Experiment:
Upon selecting a template, the New Experiment
Wizard will be displayed. Unticking the Use Experiment Wizard option can disable the wizard.
Show this screen when Rotor-Gene opens:
Untick the box if you don't want to see this window
coming up before using the Wizard.
5.5 File Menu
5.5.1 New Experiment
© 2002 Corbett Research
5.5.2 Opening and Saving
Opens a previously saved Rotor-Gene Experiment (REX file) or Rotor-Gene Experiment
Displays the last four files you have opened or saved.
Saves the experiment, prompting you for a filename if this is the first time the experiment has
been run. You have the option of saving the experiment as a template, meaning that you can base
future experiments on it. You can also save in the Rotor-Gene Experiment Archive (REA) format that
compresses the experiment. This is ideal for emailing, as it does not need to be compressed before
Saves the experiment, always prompting you for a filename. It also lets you choose what
type of file you want to save the document as.
A file can be saved as an experiment, a template, an experiment archive or an experiment in a previous
RG software version (4.3).
Opens the Export window where you can export channel and sample data to a format readable by
other software. You can export Rotor-Gene data to third party software such as spreadsheet and
statistical applications. This screen allows you to export channel information and the sample list.
The experiment's sample list will be saved in a format, which can be opened by
popular spreadsheet applications. Once you have saved the samples, drag the export file into your
You can export both raw data obtained from the machine and the data used to
perform analysis after normalization and other operations have been performed. Select the channel to
export, and the type of data you wish to export, then click Export. The resultant file can be opened by
dragging it into the target application.
16 Functional Overview
5.5.3 Export
© 2002 Corbett Research
17 Rotor-Gene
Choosing File, Setup... opens the Rotor-Gene setup window.
Select the machine you are operating from the option buttons available.
This option enables you to operate the software even when a machine is not
attached for demonstration purposes.
Allow access to this screen:
Untick this box, once the machine has been correctly selected to
prevent users from altering this setting. If you need to have access to this screen again, contact your
5.5.4 Setup
© 2002 Corbett Research
5.6 Analysis Menu
This toolbar enables you to create new analyses and displays existing ones. The tabs at the top of the
bar enable you to select a method of analysis to perform. Once you have done this, a list of the
channels which can be analyzed using this method will be listed below. Those that have already been
analyzed will have a green tick next to them, such as the first two channels shown in the screen shot.
This means that threshold and normalization settings have been remembered for this analysis. To
analyze a channel, or view an existing channel, simply double-click on the channel to view. The
specific analysis window will then appear.
Ticking the box shrinks the window when it is not used. Moving the cursor over
the window enlarges the window again.
Organizing Your Workspace:
Each time you double-click on a new analysis, its windows will be
arranged to fit in with those already on the screen. With many windows, this can be difficult to use.
Simply close the windows you do not require, then click
on the toolbar. The windows will
automatically be rearranged according to the
method. You can select another method of
arrangement by clicking the
next to the toolbar button.
Clicking the right mouse button over the analysis window also allows to
Double click quantitation or press
to open the channel of interest. Three windows will be opened
automatically, the main screen, the standard curve and the results. Going from left to right, the
following buttons are displayed in the main window:
5.6.1 Analysis Toolbar
18 Functional Overview
5.6.2 Quantitation
© 2002 Corbett Research
19 Rotor-Gene
Opens the Quantitation Report selection window where you can choose a report to preview
of the currently selected Quantitation analysis. There are three different options, Standard Report, Full
Report and Concise Report.
Using the buttons on the top, the reports can be
This button opens the standard curve graph. By default, this window is opened when an
analysis is opened. If you close the window, it can be re-opened by using this command.
On the standard curve the values are recalculated dynamically as the threshold level is varied, by
on the curve show the samples that have been defined as standards and
the unknown sample data points.
NOTE: If redefining standards to recalculate the standard curve, toggling the standard colors
ON or OFF will remove it from standard curve calculation.
The Reaction Efficiency of the experiment. This value is discussed in more detail in
Amplification, Reaction Efficiency
(correlation coefficient):
value (as displayed with the superscript),
is the percentage of the data which is consistent with the statistical hypothesis. In the Quantitation
context, this is the percentage of the data which matches the hypothesis that the given standards form
a standard curve. If the R
value is low, then the given standards cannot be easily fit onto a line of best
fit. This means that the results obtained (ie. the calculated concentrations) may not be reliable. A good
NB: It is still possible to achieve a high R^2 value with a poor standard curve, if not enough
Reports
Standard Curve
© 2002 Corbett Research
20 Functional Overview
standards have been run. The R^2 value will improve as the number of standards decreases.
(square root of correlation coefficient):
The R-value of the calculation is the square root
of the R^2 value. Unless you have a specific statistical application, the R2 value is more useful in
) of a standard curve are
automatically calculated and shown in the top right corner of the standard curve window.
With the pointer on the standard curve, click on the right button to show the option
This function is used when no replicates on a standard have been run or
defined. With the function enabled the standard curve shows a line between the Given concentration
and the Calculated concentration for the standard series. The threshold is then optimized, by moving
the level to minimize the distance between each Given and Calculated standard concentration.
When multiple quantitation runs have been performed in the same experiment, it is possible
to overlay the standard curves in the same window. This is useful for graphically viewing the difference
between different thresholds on the statistical results. Below is a screenshot of this feature:
represents the equation used to relate CT values and concentrations. Type can be
either floating or fixed. If floating, an optimal standard curve equation is calculated as you move the
threshold. If Fixed, the equation does not change because it has been imported from another
Note: The Rotor-Gene standard curve formula is displayed in a different format to
competitors' standard curves. Competitors express their curve in terms of concentrations
instead of CT values, which does not affect results, but does affect the display of the M
(gradient) and B (offset) values. Please read the section
Why Are Rotor-Gene Standard
Standard Curve Calculation
© 2002 Corbett Research
21 Rotor-Gene
Importing a standard curve allows you to perform estimates of concentrations when a standard curve
is not available in an experiment and you are certain that the reaction efficiency has not varied between
the two runs. You can import curves from another channel, or from another experiment by clicking on
the standard curve, or not to adjust. Adjusting means modifying the offset
(b) in the standard curve to be consistent with the current experiment. Whether you should adjust the
standard curve or not depends on the chemistry application.
A reference should be defined in the target experiment to adjust the curve to match the target
experiment. You can define a reference by setting a sample's type to Standard, and entering a
concentration value in the Sample Editor. Multiple copies of the same reference can be entered to
improve accuracy of the method. Note that you cannot define more than 1 reference concentration. For
example, it is possible to have 3 replicate references of 1000 copies, but not to have one reference of
1000 copies, and another with 100 copies in the same experiment.
Once the curve has been imported, the standard curve type will be changed to Fixed. Click
set the curve type back to Floating.
Below is a screenshot of the Import Standard Curve screen:
Using this screen, you can import the standard curve from another channel you have analyzed in the
current experiment, or you can load a standard curve from another experiment.
When this option button is selected, quantitation analyses on other channels
will be listed with their corresponding standard curves.
Selecting this option button will bring up an Open Dialog in which you can
select an experiment to open. If any quantitation analysis has been performed for the experiment, you
will see standard curves listed for each channel analyzed.
Lists the analyzed channels and their respective standard curve formulas.
Some chemistries produce a fluorescent readout that decreases exponentially instead of increasing. It
is still possible to analyse these using Quantitation, but the "Invert Raw Data" checkbox should be
For all other quantitation analysis, this option must remain unticked.
Import Standard Curve
Invert Raw Data
© 2002 Corbett Research
NB: Techniques such as Quenched FRET
signals to quantitate may have less accurate results than those that increase. Such
techniques have not been widely verified yet in the scientific community.
Calculation of CT
The Ct value is the value where the threshold line crosses the amplification curve. By
setting a threshold line and calculating the intersection with each of the sample curves, the Ct values
for each sample are established.
(Manual Set): To set the threshold click on the icon (grid with red arrow) then click and
hold on the graph and drag a threshold line to the desired level, or enter a log value. Alternatively the
Auto-Find Threshold function can be used to automatically determine the best level. When setting a
threshold manually, it should be set in the exponential phase of the experiment, significantly above the
background level to avoid noise.
The automatic threshold function will scan the darkened region of the graph to
find a threshold setting which delivers optimal estimates of given concentrations. You can change the
region to be modified by entering new upper and lower bounds in the text boxes. For most quantitation
analyses, the default region is suitable. Based on the standards that have been defined the function
then scans the range of threshold levels to obtain the best fit of the standard curve through the
samples that have been defined as standards, (i.e. maximizes the R value to approach 1.0).
To set click on the icon (grid with red arrow) then click and hold on the
graph and drag the threshold line to the right. This eliminates the threshold line for low cycle numbers.
Note: This is useful when there is noise on the signal during the initial cycles.
22 Functional Overview
© 2002 Corbett Research
23 Rotor-Gene
Opens the quantitation results grid. By default, this window is opened when you open an analysis. If
you close it, it can be re-opened by using this command.
The results obtained from the experiment are summarized in a table. Clicking the right mouse button,
will export the table to Excel. There is no need to open the Excel
program, as this will be done automatically.If you would like to copy the data into an existing
option instead, then open your spreadsheet, then select paste.
To make calculations easier, a feature called
is introduced which automatically calculates
the Average, Standard deviation, Minimum and Maximum values of samples of interest. Simply move
the mouse over the area of interest and the values are given in a small table displayed below the
sample list on the right-hand side of the screen.
Results
© 2002 Corbett Research
24 Functional Overview
In this screenshot, samples 1 to 5 are analysed:
This option is ticked by default and is used to determine the average background of each individual
sample just before amplification commences. Standard Normalization simply takes the first five cycles
and uses this as an indicator for the 'background' level of each sample. All data points for the sample
are then divided by this value to normalize the data. This process is then repeated for all samples. This
can be inaccurate as for some samples the background level over the first five cycles may not be
indicative of the background level just prior to amplification. Dynamic Tube Normalization uses the
second derivative of each sample trace to determine a starting point for each sample. The background
level is then averaged from cycle 1 up to this starting cycle number for each sample.
gives the most precise quantitation results.
Alternatively with some data sets it may be necessary
to disable the dynamic tube normalization. If this is the case the average background for each of the
samples is only calculated over the first 5 cycles. If the background is not constant over the cycles
before amplification it will result in less precise data.
The background fluorescence (Fl) of a sample should ideally remain constant before amplification.
However, sometimes the Fl-level can show an increase or decrease due to the effect of the chemistry
being run and produce a skewed noise level. The Noise Slope Correction option uses a line-of-best-fit
to determine the noise level instead of an average, and normalizes to that instead. Turning on this
option can tighten replicates if your sample baselines are noticeably sloped.
This function improves the data when raw data backgrounds are seen to slope upward or downward
before the amplification take off point (Ct). It is very helpful for experiments when for example the FAM
background is seen to creep upwards over the initial cycles.
Dynamic Tube Normalisation
Noise Slope Correction
© 2002 Corbett Research
25 Rotor-Gene
The first couple of cycles in a quantitation run are not usually representative of the rest of the
experiment. For this reason, you may get better results if you select to ignore the first few cycles. If the
first cycles look similar to cycles after them, you will gain better results by disabling this function, as
the normalization algorithm will have more data to work with. You can ignore up to ten cycles.
The quantitation setting in the quantitation screen allows excluding samples or NTCs, which have a
slight drift upwards, due to probe degradation or other effects. All samples with a change below the
NTC threshold will not be reported in the quantitation screen. The percentage is relative to the largest
maximum change found in any tube. For example, if you have one sample which began at a
background of 2Fl and went to 47Fl, then 45Fl represents 100%. An NTC threshold of 10% would
consider as noise any sample with a reaction less than 4.5Fl.
The slope (m) of a reaction (shown in the standard curve window), once slightly transformed, can be
used to determine the exponential amplification and efficiency of a reaction.
Calculate a gradient g = 1 / m. This gradient g is used in the following equations, and not the m value.
This is because the slope used in the Rotor-Gene software and in other applications is reported in a
different but equivalent manner. The reason for this is explained in the topic "Why are Rotor-Gene
Standard Curves different?".
The following calculations give some important results:
exponential amplification = 10
orreaction efficiency = [10
Optimal values for g, amplification and reaction efficiencies are
–3.322, 2 or 1, respectively. The efficiency of a reaction is reported in the quantitation report.
The slope is calculated of being the change in Ct divided by the change in log input (for example copy
number). A 100% efficient amplification means a doubling of amplification product in each cycle
resulting in a g value of -3.322, an amplification factor of 2 and a reaction efficiency of 1.
Given a g value of –3.322, the calculations are as follows:
Here are two examples for two different slope values.
A g value of 3.8 means that the reaction has an amplification value of ~1.83 and a reaction efficiency
There could be several reasons for this value. If the value needs to be improved, optimization steps like
Ignore First
Quant. Settings
Slope, Amplification, Reaction Efficiency
© 2002 Corbett Research
primer or probe concentrations, MgCl
- or Sybr-Green I concentrations could be improved.
A g value of 3 means that the reaction is more than 100% efficient. A reason for this could be an
unproportional digestion of probe compared to the amplicon produced. In addition, if the R-value is low,
then statistical error can cause an unexpected reaction efficiency.
Offset
In a formula describing the relation between two variables, the intercept is expressed with the letter "b"
(y = mx + b). The intercept is also sometimes referred to as the
The intercept used in the standard curve of the Rotor-Gene software is the maximum theoretical
number of copies that could have been detected at cycle number 1.
The below shown standard curve has the following values:
A dilution series has been performed starting with a copy number of 10
copies. Four dilutions have
been performed down to 10
The intercept in this example is B = 10.393. The doted blue line below represents a CT-value of 1 and
the red line is an extrapolation of the standard curve. The crossing point of those two lines gives the
maximal theoretical amount of copies which could have been detected. The calculation is as follows:
26 Functional Overview
© 2002 Corbett Research
27 Rotor-Gene
The short answer is that the Rotor-Gene standard curves and the competitor standard curves display
exactly the same information, but in a different manner. If you are not satisfied with this explanation,
then this section is of interest to you.
Many competitors provide a standard curve as a linear function from concentrations to CT values. This
means, the standard curve formula they produce allows to you to find out the CT value for a sample if
you know its concentration. The Rotor-Gene contains a formula written the other way around. That is,
given a CT value, tell me the concentration. Of course, this is exactly what the Quantitation process is
all about -- finding out concentrations of samples given their CT values. This is the reason we chose
this method of representing the information. The different approaches of the Rotor-Gene and
competitors has, however, been the source of some confusion.
This is the relationship between the values:
"competitor m value" = 1 / "Rotor-Gene m value"
"competitor b value" = -"Rotor-Gene b value" * "competitor m value"
Here is the proof of the correlation between the Rotor-Gene and Competitor standard curves. It is a
basic rearrangement of two functions.
Let rm, rb be the Rotor-Gene gradient and offset.
Let cm, cb be the Competitor gradient and offset.
Let Concentration be the calculated concentration for a sample.
Let CT be the CT value for a sample.
Log(Concentration) = rm * CT + rb
(Rotor-Gene has a function from CTs to Concentrations)
CT = cm * Log(Concentration) + cb
(Competitors have a function from concentrations to CTs)
Log(Concentration) = rm * CT + rb
(Log(concentration) - rb) / rm = CT
(Log(concentration) - rb) * 1 / rm = CT
(Log(concentration) - rb) * q = ct
(Log(concentration) - rb) * q = ct
Log(concentration) * q - rb * q = ct
CT = Log(Concentration) * f + g
Which is exactly the same as the competitor equation, by substituting
Why Are Rotor-Gene Standard Curves Different?
© 2002 Corbett Research
28 Functional Overview
cm = 1 / rm and cb = -rb * cm
© 2002 Corbett Research
29 Rotor-Gene
This screen shows that taking the average background energy for each of the samples normalizes the
data which is then converted to a Log scale.
From this Log graph a threshold can be determined to calculate the Ct value for each of the samples.
The Ct value denotes the starting cycle for the amplification. This Ct value can be related directly to the
starting copy number of the sample.
on the bottom of the screen takes you directly from the
The difference between the Linear and the Log Scale is only how the data is presented. The lower
values in the Log-Scale are "blown up" to make data readable easier. The Log Scale emphasizes the
normalized data approaching zero. The linear scale is the normalized RAW data, we see all samples
with the same starting background.
The log-scale is the LOG of the normalized data, we see the small increase at the take off point
expanded so we can accurately determine the Ct value. This could easily be checked by turning on the
pinpointer in both graphs and compare the numbers obtained for the fluorescence. It can be seen that
there is no difference between the actual data.
Note: This process can also be performed while the Rotor-Gene is running. This real-time
monitoring of quantitation data provides the user with the possibility to gain results as soon
as the curves show an exponential growth. Preliminary conclusions could be drawn already
and decisions made for the next run.
Main Window
© 2002 Corbett Research
5.6.3 Melt Curve Analysis
analyses the derivative of the raw data, after smoothing. You can use it to
perform genotyping, to find take-off points in reactions and numerous other applications. Peaks in the
curve are grouped into bin groups, and all peaks below the threshold are discarded. You then map
bins to genotypes through the Genotype, Edit Genotypes menu.
Typically after a cycling run has been finished a melt step can be added to visualize the melt
characteristics of the products. The sample temperature is increased at a linear rate and the
fluorescence of each sample is recorded.
Below is a typical melt curve analysis shown for a Sybr Green I amplification.
Before defining peaks ensure the dF/dT sign is correct for the data set to give
In the Melt Curve Analysis screen peaks can be defined and reported using different
methods. One is to automatically call all the peaks for each sample. The other is to assign peaks to
bins, which is useful for genotyping samples.
Any peak that is within a range of +/-2°C of the bin center will be assigned to the bin. If there are two
peak bins close together then the peak will be assigned to the closest bin.
Bins are used to define the general area where you expect peaks to occur. The melt analysis software
clusters peaks into bin groups, based upon actual peak values in the curve.
30 Functional Overview
Sidebar
© 2002 Corbett Research
31 Rotor-Gene
Note: The peak bins should not be used for estimating peak positions, use the Automatic
Peak Calling table instead for this purpose.
To define a peak bin click on the
button, then click and hold on the graph to
define the center of the peak bin. To add another bin repeat the process or use the
To set the threshold, click on the icon (grid with red arrow) then click and hold on
the graph and drag a threshold line to the desired level.
To set click on the icon (grid with red arrow) then click and hold on the
graph and drag the threshold line to the right. This eliminates the threshold line for the lower
Note: This is useful when there is noise on the signal at low temperatures.
Opens the Melt Report selection window where you can choose a report to preview. You can generate
a report based upon the currently selected channel, or you can generate a Multi-Channel genotyping
Displays the result grid, showing peaks in samples.
Click on the genotyping toolbar and select the genotype definitions.
This screen lets you assign genotypes to the incidence of peaks in bins. The default genotype
configuration is shown in the screenshot, with heterozygous samples having two peaks, homozygous a
peak in the first and wild type samples in the second bin. Next to the name of each genotype is a field
for typing in an abbreviation. This is used when printing multi-channel genotyping reports so that all
results from multiple channels can be read easily across the screen.
For multiplex analysis, genotypes have to be setup in each channel. If for example a dual
channel FRET analysis is run, where a wild type, heterozygous and homozygous is expected
in each channel, the above setup has to be performed for each channel. The results will than
be given in a multiplex report.
Reports
Results
Genotyping
© 2002 Corbett Research
5.6.4 Allelic Discrimination
For Allelic Discrimination, it is not sufficient to double-click on the channel you would like to analyze, as
this analysis is performed using multiple channels simultaneously. To perform this analysis, either hold
down SHIFT and click to highlight each channel you wish to analyze, or drag your mouse over these
channels. Once the desired channels have been highlighted, click
. The list will update to show
all the channels on one line, with a tick next to them. This indicates that they are all being used in one
analysis. You can "break apart" these channels by right clicking on the analysis and selecting
You will then be able to include those channels in another Allelic Discrimination analysis.
Please note that a channel can only be used once in each type of analysis.
Allelic Discrimination Analysis allows you to perform genotyping using dual labeled probes. The
genotype of samples is displayed in the result window.
Opens the Allelic Discrimination Report for preview.
Displays the genotyping results grid. This grid is opened when the analysis is first displayed by
A variety of options are available to optimise the way in which the data is normalised:
Dynamic Tube (Dynamic Tube Normalization)
Slope Correct (Noise Slope Correction)
Ignore First x (First x Cycles Noise Correction)
These terms are explained in the chapter
32 Functional Overview
Reports
Results
Normalisation Options
© 2002 Corbett Research
33 Rotor-Gene
Discrimination Threshold:
Enter values in these text boxes to position the discrimination threshold.
All curves surpassing this line will be considered as having amplified for the purposes of genotyping.
Click on the button to the right of each text box then drag the threshold on the graph to set these
Opens the genotype window to define which genotype is detected in which channel.
This screen lets you assign genotypes to reacting channels for an Allelic Discrimination analysis. In
the above example, a sample is heterozygous if readings in channels Cycling A.CH1 and Cycling
A.CH2 cross the threshold.
For Scatter Analysis, it is not sufficient to double-click on the channel you would like to analyze, as this
analysis is performed using multiple channels simultaneously. To perform this analysis, either hold
down SHIFT and click to highlight each channel you wish to analyze, or drag your mouse over these
channels. Once the desired channels have been highlighted, click
. The list will update to show
all the channels on one line, with a tick next to them. This indicates that they are all being used in one
analysis. You can "break apart" these channels by right clicking on the analysis and selecting
You will then be able to include those channels in another Scatter Analysis analysis.
Please note that a channel can only be used once in each type of analysis.
Scatter Analysis allows you to observe how different samples will react using different channels,
reactions are graphed on a scatter plot allowing a visual comparison across both channels.
Discrimination Threshold
Genotypes
5.6.5 Scatter Analysis
© 2002 Corbett Research
34 Functional Overview
Opens the Scatter Analysis Report for preview.
Displays the genotyping results grid. The genotype for each sample is determined by the custom
regions which the user may defined using the scatter graph.
A variety of options are available to optimise the way in which the data is normalised:
Dynamic Tube (Dynamic Tube Normalization)
Slope Correct (Noise Slope Correction)
Ignore First x (First x Cycles Noise Correction)
These terms are explained in the chapter
Reports
Results
Normalisation Options
© 2002 Corbett Research
35 Rotor-Gene
Opens the genotype window to define which genotype is detected in which channel.
This screen lets you assign genotypes to reacting channels for a scatter analysis. Defining a genotype
here, gives you the option to define a user-defined genotype on the
user-defined genotypes from the scatter graph that will determine the results, default genotypes
defined here are for display purposes only and will appear in the appropriate corners of the scatter
The scatter graph is a plot of both channels on their respective axes, giving a visual comparison so we
can easily see how the sample react on different channels.
The user is given the option to define custom genotypes by clicking on the graph and dragging to
define a region. The regions that are available can be modified through the
" is used to compare the relative expressions of samples to a
control sample in a run when a standard curve is not available. Users testing results from Microarray
analysis frequently use this feature.
To perform the analysis, go to
channel to analysis. Chose a control sample by using the pull down menu on the right-hand side of the
Genotypes
Scatter Graph
5.6.6 Comparitive Quantitation
© 2002 Corbett Research
36 Functional Overview
screen below the sample toggler. The table below the graph will automatically calculate the results.
The first column of the table shows the names of the samples. The second column is called "Take off"
and gives the take off point of the samples. To calculate the take off point, the second derivative of the
raw data is taken. Then, a peak is found where the reaction is increasingly most rapidly. This is the
peak of the exponential reaction and occurs shortly after the take-off of the reaction. The take-off point
cannot be determined exactly, but is estimated by finding the first point to be 80% below the peak level.
This graph shows a second derivative of a quantitation reaction, showing the peak, and the 20% level
used to indicate the level:
The third column gives the efficiency of the particular sample. A 100% efficient reaction would result in
an amplification value of 2 for every sample, which means that a doubling of an amplicon takes place
in every cycle. In terms of the raw data, the signal should increase by a doubling amount in the
exponential phase. So, if the signal was 50 at cycle 12, then went to 51 at cycle 13, it should go to 53
fluorescence units at cycle 14. All of the amplification values for each sample are averaged to produce
the amplification value that is shown on the right-hand side of the screen. The more variation there is
between the estimated amplification values of each sample, then the larger the confidence interval will
be (indicated by the value after the ± sign). The confidence interval gives a 68.3% probability that the
true amplification of the samples lies within this range (1 standard deviation). By doubling the ±
interval, one achieves a 95.4% confidence interval.
The average amplification is needed to calculate how much more or less a sample is expressed. If for
example the amplification value was lower, a certain absolute copy number of an amplicon is obtained
later than if the amplification value was higher. The last column finally gives the comparative
quantitation value. Based on the take off point and the reaction efficiency it calculates the relative
concentration of each sample compared to the control sample that was chosen by the user. The
© 2002 Corbett Research
37 Rotor-Gene
number given is expressed in scientific notation.
Note: If a large value is displayed after the +- in the Average Amplification, then the results
are likely to be unreliable. The interval that the average amplification represents the range for
which the results are valid.
Steps taken to calculate relative concentrations:
1. The take-off points of each sample are calculated by looking at the second derivative peaks.
2. The average increase in raw data 4 points following the Take-Off is calculated and becomes the
3. The amplications are averaged to become an experiment "Average Amplification".
4. The relative concentration for a sample is calculated as Amplification ^ (ControlTakeOff -
5. The result is displayed in scientific notation.
is a technique which allows the determination of relative levels of concentration for
runs performed in a non-RT thermal cycler.
Below is a screenshot of the EndPoint Analysis software module:
is similar to Allelic Discrimination, in that the results are of the form "reaction/no
5.6.7 EndPoint Analysis
© 2002 Corbett Research
reaction", and that names can be assigned to certain permutations of reactions over different
is that only a single reading is available instead of a
cycle-by-cycle reading for each sample. This means that the user must supply additional information to
help facilitate the analysis, namely, the identification of
To facilitate the reading of this information, a Normalisation process using known positives and
negatives for each channel presents the samples in terms of their expression relative to this. The user
then selects a percentage
as the reaction threshold.
Terms Used In EndPoint Analysis
Below are explanations of the terms and concepts used in EndPoint Analysis:
A sample which is known to have reacted, and which
represents the 100% expression level for samples on the same
A sample which is known not to have reacted. This represents
the 0% expression level for samples.
A percentage level of expression above which a sample is said
to have reacted. This setting must be adjusted by the user for
A percentage value which represents the fluorescence as the
difference between the highest positive control and the lowest
An interpretation of different permutations of reactions on
different channels. For example, one could assign a genotype
of "Heterozygous" to samples which reacted in both channels
"FAM/Sybr" and "JOE". The genotype can also be used for
reporting results of reactions with internal controls. For
example, you may wish to report results such as "Inhibited",
"Positive", "Negative" on the basis of whether a reaction was
seen in certain channels.
38 Functional Overview
© 2002 Corbett Research
39 Rotor-Gene
To perform an EndPoint analysis, you should cycle the samples for the appropriate time on your other
thermal cycling instrument.
Once this has finished, transfer the tubes to the Rotor-Gene, and perform a profile with a hold at 60
degrees for several minutes, then a cycling step with 1 step, on 60 degrees for 20 seconds, acquiring
on the required channel. Set the number of repeats to 5, as shown below:
Please note that these times are a guide only, and may vary for your particular application.
The more repeats in the profile, the more information will be available to perform the analysis. The
analysis module will automatically average all the readings taken into a single value for each sample.
There is no set number of repeats required, and unless you are performing an experiment where a
very fine level of accuracy is required, 5 repeats should be largely sufficient.
Profile Configuration
© 2002 Corbett Research
Analysis
You can perform EndPoint analysis on an number of channels simultaneously. To create a new
analysis, click on the EndPoint tab, select the channels by dragging over them with your mouse, and
. The EndPoint analysis window will be displayed.
When you open an EndPoint analysis for the first time, you will likely not have set up your Positive and
Negative controls. The following message will be displayed :
40 Functional Overview
Define Controls
© 2002 Corbett Research
41 Rotor-Gene
After clicking OK, the sample editor appears, allowing you to define Positive and Negative controls
set a sample to be a Positive or Negative control, click on the cell containing the sample's type, then
select the relevant control type from the drop-down box which appears next. The samples to be
positive or negative controls must be turned on before you enter this screen.
This screen functions in an identical manner to the
You can also access this screen
by clicking Define Controls in the
Normalisation of the EndPoint data brings all samples on all channels analysed, within the range of 0-
100. You must have selected at least one
will need more than this if you are analysing multiple channels and your standards are not multi-plexed.
You should also use more than one control of each type if there is the risk that your positive control
For each channel, all the Positive Controls are sampled and one with the highest fluorescence is
set to be 100%. This allows for a Positive Control to fail without affecting your experiment, if you
are running duplicate controls.
All the Negative Controls are then sampled, and the one with the lowest fluorescence level is set
The raw fluorescence values of the other samples are then scaled relative to the highest Positive
Control and the lowest Negative Control.
Normalisation
© 2002 Corbett Research
42 Functional Overview
This experiment was a success, as the two positive and negative controls are both close together, and
fluorescence values of the samples 5 and 6.
Here are the normalised values:
As Sample 1 was the Positive Control with the highest fluorescence, it was set to 100% expression,
with the other positive control slightly lower. Sample 4, the lowest Negative Control was set to 0%
expression. It is now obvious that Sample 5 is more likely to have reacted, whereas Sample 6 is likely
If Positive and Negative Controls are not selected carefully, it is possible to achieve expression
levels of greater than 100%, and lower than 0%. This can be interpreted as "the sample expressed
itself more than the positive controls", or "it is less likely that the sample reacted than that the negative
controls reacted". Such situations are, however, unlikely to cause errors.
© 2002 Corbett Research
43 Rotor-Gene
If you receive a message stating that the Negative Controls are higher than the Positive Controls, then
you have incorrectly set up your samples. For example, your Negative Controls had a level of 10Fl, but
the Positive Controls had a level of 5Fl. Since there is no way to interpret this in a logical manner, you
will receive the following warning:
Normalisation on Multiple Channels
If a sample is a positive control for at least one channel, mark its type as Positive Control.
Otherwise leave it as type Sample. The same applies for Negative Controls.
If you are analysing data on multiple channels, then you still only define one page of negative and
positive controls. So, if a sample is a Positive Control on FAM/Sybr, but not on JOE, you should still
define it as a Positive Control. So long as you have another Positive Control on JOE which will react,
the control for FAM/Sybr will be ignored. The only limitation of this method of describing the samples is
that a single sample cannot be a Positive Control on FAM/Sybr, and a Negative Control on JOE, for
is used to determine the percentage level of expression required for a reaction on
have been defined, all channels will be
normalised to the same scale, 0-100%. This is the reason why only one threshold is needed, even
when analysing multiple channels.
Click and drag the threshold line to an area between 0-100%. Your threshold should not be too near
samples on either side of the line -- this indicates that the experiment was not conclusive. If the
difference between a sample reacting or not reacting is a matter of a couple of percent, then the same
reaction run again could appear on the other side of the line.
Threshold
© 2002 Corbett Research
Genotypes
As in Allelic Discrimination, this option opens the genotype window to define which
genotype is detected in which channel.
This screen lets you assign genotypes to reacting channels. In the above example, a sample is
heterozygous if readings in channels Cycling A.CH1 and Cycling A.CH2 cross the threshold.
Starts the defined Temperature Profile with the current Gain settings. Before the Rotor-Gene starts the
Profile Run Confirmation screen is shown. A graphical representation of the temperature profile to be
run is displayed along with the Gain settings for each channel.
You can use this screen to pause and resume an experiment. As performing this operation can
seriously affect the results of an experiment, a marker is drawn across all channels acquired to by the
currently running cycle. A message is also placed in the message screen.
When stopping a run you will be prompted to confirm the end of the run.
44 Functional Overview
5.7 Run Menu
5.7.1 Start Run
5.7.2 Pause Run
5.7.3 Stop Run
© 2002 Corbett Research
45 Rotor-Gene
This window allows the setup of experiment information, experiment name, analysis date, operator and
The experiment information window contains all information except for the profile required to configure
an experiment. The following information is displayed after a run has finished: records the type of
machine, Gain settings, number of channels and time of start/finish.
If you were a previous Rotor-Gene user, this screen now contains elements, which were found in
separate windows via the View menu.
5.8 View Menu
5.8.1 Experiment Settings
General
© 2002 Corbett Research
Machine Options
This tab contains settings for the configuration of the Rotor-Gene machine.
Spike Suppression Options:
Spike suppression is used to remove spikes caused by electrical
disturbance. It is strongly recommended to leave this option on.
The High-Speed Rotor option will spin the rotor more quickly when the
machine is cooling. This option is useful as the high speed firms the tubes in their places and prevents
them from moving around. This option should generally be left on.
The machine carousel size should be set to the size of the carousel currently
installed in your Rotor-Gene. If opening an existing experiment, this setting will reflect the rotor that
was installed in the machine at that time. The options available depend on your machine and the
This tab displays messages where the operator has changed functions such as pausing the machine
or skipping cycles during a run. It also displays warnings received during the experiment. You should
check this tab if you receive unusual results.
46 Functional Overview
Messages
© 2002 Corbett Research
47 Rotor-Gene
If configuring a new experiment, the channels tab displays the current configuration of the channels
available in the machine. If you are viewing an existing experiment, then the information displayed
represents the configuration of the channels when the experiment was run. You can modify the Gain
and create and modify channels if you are using a Multi-Channel system. To restore default channels
if an experiment corrupts your machine's channel settings, click Reset Defaults.
Describes the name of the channels.
Describes the type of machine chosen (Dual-Channel/Multi Channel). This should be set to
Multi-Channel, as on a Dual-Channel filters cannot be changed.
Specifies the wavelength of the source LED.
Specifies the wavelength of the detection filter.
Specifies the Gain for that particular channel.
This feature allows creating new channels. Pressing this button opens a window, which
asks for a new name, source and detection filter. The filters can be chosen by using the pull down
menu next to each window.
Examples for valid Channel Configurations are:
Channels
© 2002 Corbett Research
48 Functional Overview
are the standard configuration for 4 channel multiplex detection. Quencher
molecules need to be used on the 3' end of the dual labeled probes so as not to occupy spectral
bandwidth. The best combination is BHQ (black hole quenchers from Biosearch Technologies)
FAM/BHQ1, JOE/BHQ1, ROX/BHQ2, Cy5/BHQ3.
can be used for single channel Sybr Green I detection and gives more sensitivity than
Channel 1 due to the fact that a 585hp (high pass filter is used) and so more energy is captured.
is used to detect eg. Cy5 or LC640 in a FRET reaction where an increase in energy is
detected. This could be used for quantitation or mutation detection.
For mutation detection using FRET probes, the information is contained in the differential of
the melt curve and it makes no difference whether the increase in Cy5 or the decrease in FAM is
monitored. This is helpful when looking at multiplexing a mutation detection assay as a FAM/BHQ1
and JOE/BHQ1 probe sets can be used in conjunction with Channel 1 and 2 configuration.
If you are using a 72-well system, you may wish to reorganize samples to more closely
match the labeling that you are using on a 9x8 block. The tube layout tab allows you to label samples
sequentially (the default), so that samples are listed in the order they are placed in the machine, or by
every 9th sample (1A, 1B, 1C, etc). This is useful when loading samples with a multi-channel pipette.
If modifying this setting, you should perform a test run to confirm that the organization option you have
selected works in the manner that you expect.
Tube Layout
© 2002 Corbett Research
49 Rotor-Gene
The security tab displays information about the Experiment Signature. The experiment signature is a
non-reversible key which is regenerated every time the file is changed. If any section of the REX file is
modified outside of the software, the signature and the file will no longer match. Using the signature,
you can ensure that a scientist does not modify the raw data outside the application, that the profile
has not been tampered with and that the temperature graph is valid. The signature also protects
against non-malicious corruption, such as file-system errors.
The experiment signature only exists in recent versions of the software. The absence of a
signature in old files does not mean that a file has been modified. However, any new experiments you
create should contain a signature and so can be validated. You should treat with suspicion any file
where the signature is invalid.
You can ensure security for your files by making a laboratory policy that all experiment files must be
run using 5.0.8 or later. Experiments run in this version will display a warning if they have not been
signed, whereas older experiments will not. However, all reports will include the version number of the
software, so you can also treat as suspicious newly run files on old versions of the software.
Security
© 2002 Corbett Research
5.8.2 Temperature Graph
The temperature graph shows the temperature of the samples during a run. As the run proceeds the
Set, Actual and Hold time is shown for each step of the program. When loading an existing experiment
file the temperature graph shows the temperature history during the experiment. The vertical scale is
20-100°C and the horizontal scale is shown in minutes. Use the scrollbar to scroll backwards and
forwards throughout the temperature history.
This screen allows you to preview and modify the profile the machine will follow to heat, cool and
acquire data from the tubes in the Rotor-Gene. A profile is made up of cycles, which are listed in order
of running in the list box.
Add five cycles (when visible):
When cycling has already begun, it cannot be modified. However, it
can be lengthened by clicking on this button, enabling you to prolong an experiment. The add 5 cycles
button is only visible during cycling. Five cycles could also added in the Profile Progress Screen.
Displays a pop-up menu where you can select the type of cycle to insert into the profile.
The new cycle will be inserted after the currently selected cycle.
Identical to the Insert After button, except that the new cycle will be inserted just before
the currently selected cycle.
50 Functional Overview
5.8.3 Profile Editor
© 2002 Corbett Research
51 Rotor-Gene
Removes the currently selected cycle from the profile.
Depending on which cycle is currently selected, the lower half of the screen will change to provide
options specific to that cycle type. Each cycle type and its related options are discussed below.
This cycle has a dual purpose. It can either instruct the Rotor-Gene to remain at a temperature for a
given amount of time, or it can perform an optical denature calibration.
Click directly on the value you would like to change, such as 60 degrees to bring up a seperate edit
window. This window will allow you to modify the value using either a slider control, or by entering it
directly into the textbox.
To perform an optical denature calibration, the hold temperature must be 90 degrees or
above. After changing this, the Denature option becomes available. If you are using
you should also tick "Calibration Step" to allow the software to locate the melt point of the samples
The calibration settings seen below are the same as those described in the section
is not selected, ticking the Denature box has no effect other than changing the
is an exciting new technique which performs real-time melt analysis to
determine the melt peak of a reference sample. This indicates with greater precision when your
product has denatured than by setting a particular denature temperature for a hold time. To perform
this technique, simply place a tube of pre-amplified product in tube position 1 in your rotor. The
reference tube must also contain a detection chemistry that enables determination of the dissociation
Hold
Optical Denature Cycling
What is Optical Denature Cycling?
© 2002 Corbett Research
52 Functional Overview
of the strands of the product. We suggest using Sybr Green I.
When first heating the machine to the initial denature temperature, a melt on FAM/Sybr is performed
starting from around 80 degrees until 95 degrees (although these parameters may vary). From this
data, a melt curve is generated and automatically analysed:
is referenced back to the raw data to obtain a
Optical Denature step, the machine is heated as quickly as possible and data acquired continuously.
has hit this fluorescence level, the machine is immediately cooled and
move onto the anneal step. While cycling, a peak is not calculated, rather, it is the fluorescence level
that is referenced to the melt peak and designates the denature threshold.
In this graph, the raw fluorescence readings and the first derivative have been overlayed. It shows the
1-to-1 correspondence between the
What do I need to perform an Optical Denature run?
To perform an Optical Denature run, you will need:
A pre-amplified sample which you should place in position 1 in the rotor. This sample should contain
the same product as samples of interest and a detection chemistry, such as Sybr Green, that allows
for the monitoring of product dissociation.
© 2002 Corbett Research
53 Rotor-Gene
An optical denature profile. You can find out how to set up an Optical Denature Run in the topic
Adding a New Optical Denature Cycling
What will I see when performing an Optical Denature run?
An Optical Denature run, from the user perspective, appears almost identical. The most striking
differences are the melt step automatically inserted at the beginning of the profile, and the sharp profile
of the denature step during cycling. The denature step does not require defined hold times as the
dissociation of the product is monitored at each cycle.
To perform this technique, the software must know several things about your experiment:
The Initial Denaturation temperature. This is the same temperature as your denature step in a
standard cycling profile.
The tube position of a sample which has already been amplified, and which will produce a melt
curve on the FAM/Sybr Channel.
The tube position of a sample which has already been amplified,
and which will produce a melt curve on the FAM/Sybr Channel. For example, if you are performing a
Sybr-Green I run or a run with dual labeled probes, put a sample of pre-amplified Sybr product in
tube position 1, and your other samples in the remaining positions.
The time of denaturation required.
The time depends on the used Taq-Polymerase and can vary
between 1 and 15 minutes.
You must also define an Optical Denature cycling profile. This is enabled by default in this version of
the software. However, your existing templates from older versions of the software will not use this
To add an optical denature step, open the Profile Editor. Then click New. A default profile containing a
Denature step and an Optical Denature Cycling step will appear: The profile is displayed as:
The blue shaded region at the beginning of the run represents the optical denature calibration process.
The Green dots represent the acquisitions taken each cycle during heating. The lighter blue dots
represent the acquisition at the end of the anneal step at 60 degrees. Note that while the profile shows
each step going to the same denature temperature, this may not be the case. If the sample requires
slightly longer to melt towards the end of the run, the optical denature process will wait for the melt in
the fluorescent data, and not off the times. For this reason, the temperature trace for optical denature
By clicking on the Cycling step, the information about the 2-step profile appears in the lower half of the
Profile Editor window. Click on the second half of the graph with the Optical Denature symbol
Configuration
Adding a New Optical Denature Cycling
© 2002 Corbett Research
54 Functional Overview
The Calibration settings information appears on the left-hand side of the screen:
Most of the time, this information will be correct. You can, however, modify it by clicking
Calibration Settings window will then be displayed:
The tube indicated in the Tube Position contains a sample of pre-amplified product that will show a
melt peak on the FAM/Sybr channel.
The Final Ramp temperature will not burn the sample, yet will be high enough to allow it to melt.
The hold time is sufficient to denature the samples.
© 2002 Corbett Research
55 Rotor-Gene
You can also define a Denature step yourself, and configure the calibration settings via its screen:
The calibration settings are sychronised with the denature settings, so a change to the hold time in the
denature step will automatically update the calibration's hold time. This is because the calibration
process and denaturation are equivalent in
To change an existing denature step in a Cycling, select the cycle in the Profile Editor's list. Then,
select the denature hold by clicking on it in the preview graph at the bottom of the screen:
Click on the word "Timed Step" and select "Optical Denature". The Temperature and Hold Time will be
removed and the Optical Denature icon
will be displayed in their place.
If you are using Optical Denature Cycling, you
make sure that your denature step is also a
calibration step. Otherwise, the Cycling step will automatically include one as well. The profile will
appear with a dip in the denature, as shown below:
To correct this problem, tick the
box in the Denature step.
Changing An Existing Step to Use Optical Denature
© 2002 Corbett Research
Cycling
As its name implies, Cycling loops through the given cycle points a given number of times, acquiring at
certain points, even gradually cooling or lengthening certain cycle points if desired as it repeats.
The white graph represents a single repeat in a cycling profile. More cycle points can be added to the
end, or the tail cycle point removed by using the [+] and [-] buttons. This gives you the flexibility to
define just one cycle point (in which case you can acquire at a constant temperature), or with up to five
To modify the temperature, hold time or acquisition properties of a cycle point, first click on that portion
of the graph. The graph will become greyed-in, such as the 94 degree step shown here. Then, click on
the value you are interested in in the left-hand bar. For example, to modify the temperature, click on 72
degrees to display a pop-up window where you can modify this value. The operation is identical for
hold time. You can also drag the graph line up and down to modify temperature, and from left to right
Touchdown can be enabled to decrease the temperature during the initial cycles. Set
the number of cycles and the program decreases the Anneal Temperature by 1°C every cycle for the
number of cycles specified. This is reflected in the graphical cycle representation.
Long range can be enabled to increment the hold time of the selected step with each
new cycle. This has the effect of increasing the extension time as the cycling proceeds.
For acquisition, a separate window will appear in which you can modify which channels will be
Use the [<], [>] and [<<] buttons to move channels between the two lists displayed. All channels you
56 Functional Overview
Acquisition
© 2002 Corbett Research
57 Rotor-Gene
put in the right-hand side will be acquired to. To not acquire on this cycle point, click on
. Any changes you make will be reflected back on the Profile Edit window.
For a melt cycle, specify a start temperature, an end temperature, the time you want to wait at before
the first acquiring point, the amount of time to remain at each point and a ramp will be generated going
between the two temperatures. The
clicking on the button. The same screen as for Cycling will appear and the channels to acquire to can
When running a standard melt the temperature is increased by increments of 1ºC.
This screen shows a graphical representation of the thermal profile associated with the experiment.
When running an experiment the shaded portion of the window indicates the number of cycles that
have been completed. There is also an estimate of how many minutes the program will take to finish
There are two further buttons on the buttons,
. While skip allows skipping any
steps of the profile, the "add 5 cycles" button allows to add five cycles to the cycling profile.
Melt
5.8.4 Profile Progress
© 2002 Corbett Research
5.8.5 Edit Samples
This window has identical functionality to the sample edit grid in the New Experiment Wizard, except
that the toolbar functions are also available in the File and Edit menus. This menu can also be
accessed by right clicking over the sample list on the right side of the main screen.
Three menus are given on the top of the screen,
. The file menu is used to create a
new (blank) sample sheet, open an existing sample template or save sample names as a template for
future use. The extension of those files is *.smp
Given concentration format:
This drop down menu is used to choose a concentration suitable
format. Due to the fact that different countries show formats in different ways, various formats are
This drop down menu allows choosing an appropriate unit for your assay.
Pressing this button opens the color scheme. Selected samples (up to 36 or 72) will be changed
according to the color chosen. Click OK and all the selected samples will be assigned this color.
Click this to reset all selected color cells back to their default color values.
The gradient function allows choosing a gradient form the first to the last selected color. A
gradient has been used in the above sample list. Several gradients can be chosen in a sample setup.
To export data into an Excel spreadsheet simply use the right mouse button. Excel
will automatically be opened and exports the samples. Using the
main window will also create a file that can be can exported into a spreadsheet application. Importing
samples from Excel into the Rotor-Gene software is also possible by simply using the copy and paste
58 Functional Overview
© 2002 Corbett Research
59 Rotor-Gene
The Tab key can be used to navigate around the sample editor.
Clears the sample grid in preparation for data entry.
Brings up a dialog box in which you can select a Rotor-Gene Sample file to import. Note:
A 32-well sample setup cannot be imported into a 72-well setup and vice versa. The number of
samples in the open sheet and the file being imported must match.
Brings up a dialog box in which you can enter the name of a file to save the current
Copies the selected cells.
Pastes cells which had been selected with the copy command onto the currently selected
Pressing the Excel icon prompts you for a file name. Excel is than opened automatically.
There are several sample types listed.
No sample in that position.
Unknown sample to be quantitated or melted.
Values are used to construct a standard curve to calculate
unknown sample concentrations.
This new function allows the user to choose various standard curves in the same run.
Use the arrow buttons to move from one standard set the next. The software only allows you to move
further if standards in the previous standard set have been defined.
This feature is very useful when different standard curves need to be run, either in one channel (for
example Sybr-Green I or any other fluorphore) or in different channels (with up to four different
Shows the concentration for each of the standards defined. The units can be defined
as a decimal or log number. If a dilution series has to be typed in it is only necessary to type the first
two standards. By pressing ENTER, the program automatically adds the 10 fold concentration, if there
are further Standards defined below.
You can modify the style of the line to improve readability of graphs on black and white
printers. You can also add emphasis to certain lines by modifying their style. To access this feature,
click on the right arrow button next to the Edit button:
© 2002 Corbett Research
60 Functional Overview
The toolbar will then change to show the default style "Solid". You can change this to "Dashed",
"Dotted" or a number of other possibilities. When you have finished, click the left arrow button to return
to the Edit, Reset Default and Gradient view.
When entering standards, if you place them in consecutive tubes, begin entry on
the first tube, then press Enter. Enter the second concentration, then press Enter. The software has
automatically predicted the next concentration in the set. Keep pressing Enter to fill in the remaining
If you need to enter the same information for several rows at once, select all the
rows, then begin to type. The information will be entered into each row. This works for selecting
sample types, choosing colors or entering concentrations.
To quickly select a sample type, just enter the first letter of its name. So, to set
5 samples to be Non Template Controls, select them in the sample type column, then press N for
NTC. All samples will be converted to NTC's. NOTE: Since Sample and Standard have the same
beginning letter, press S until the one you want is selected.
A complete sample description can be saved as a sample file (*.smp) and loaded into future
experiments with the same sample configuration.
When setting up a new experiment with reactions that have not previously been run on the Rotor-Gene
it is helpful to use the Gain calibration function. This screen allows you to set the Gains of each of the
channels and the set temperature.
Note: The gain calibration will never be 100% correct. This could be due to changes in
fluorescence after the first denature step. Nevertheless, the result of the Gain calibration will
give you a good indication on what fluorescence level the run will be started.
The Gain for each channel is 1 to 10, 1 = least sensitive and 10 = most sensitive.
When running reactions for the first time it is advisable to prepare a test sample containing all reaction
components. This is then placed in the machine and Gain Calibration is run to determine the best Gain
This window lets you calibrate your machine by automatically adjusting your Gain settings until the
readings for all selected channels fall below a certain threshold. You can select to calibrate all
channels, or just those that you will be using in the current experiment.
Note: This is only available
on Rotor-Gene 3000, or the 4-channel, Multi-Filter Rotor-Gene 2000.
5.8.6 Gain Calibration
Auto-Calibration
© 2002 Corbett Research
61 Rotor-Gene
Before reading, the machine will be heated or cooled to match the given
Calibrate All / Calibrate Acquiring:
"Calibrate All" will attempt to calibrate for all channels known by
the software. Selecting "Calibrate Acquiring" will instead only calibrate those that you have used in the
thermal profile defined in the experiment (cycling and melt).
This is a pull down menu allowing you to add additional channels to the gain
calibration window. Choose the channel of interest and
Opens a window where the fluorescence range of the sample can be determined. The auto-
calibration process begins on Gain 5, reading from each channel. It chooses the first Gain, which has
a fluorescence reading equal to or below the level you set in this box. In the example below, tube
position 1 was chosen with a target sample range between 5 and 10 Fl.
Removes the highlighted channel or all channels.
© 2002 Corbett Research
62 Functional Overview
Begins the calibration process. A Gain will be chosen which is within the given range. If no
value is found within the range, the closest match will be chosen.
Opens the Manual Calibration screen (see below).
Closes this window. You cannot close the calibration window while the machine is in operation.
Perform Calibration Before 1st Acquisition:
This tick box performs the Gain calibration at the first
cycle where data acquisition occurs. This is useful as with some probes after the initial denaturation
the background fluorescence of the probes could change substantially.
Perform Calibration At [x] Degrees At Beginning of Run:
This tick box performs the Gain
calibration just before starting the experiment. The machine is heated to the given temperature, the
gain calibration is performed, and then cycling begins on the first step, usually a Denature. This can
be useful when a calibration during the run may impact too much on the time spent on the initial step.
Perform Calibration Before 1st Acquisition
should be the ideal option as
calibration is performed as close as possible to experiment conditions.
Changing Gain During a Run:
If the Gain at the beginning of the run was accidentally chosen to
high or to low it can be changed within the first ten cycles. A vertical line in the main screen will appear
where the Gain has been changed. Due to the drop or increase in fluorescence the cycles before the
change will be excluded from the analysis.
This window lets you view in real time the fluorescent readings at any given temperature. It is used
when the background of a sample is unknown and therefore the Gain has to be determined to ensure
the sample signal is sufficiently on scale.
By default, the first eight samples are toggled on. If more than the first samples would like to be
monitored, turn on the samples of interest. Using manual gain calibration means that data are always
It takes approximately 4 seconds to acquire data per channel. During this time the user interface is
inactivated and so it is best to 1) Start the test 2) wait for the temperature to stabilize 3) note the end
point fluorescence (Fl) reading 4) stop the unit 5) make the appropriate Gain change 6) and restart the
Manual Calibration
© 2002 Corbett Research
63 Rotor-Gene
Change this value to set the temperature of the machine. Note: The temperature will
only be sent when the start button is pressed. Adjust the temperature on the Gain calibration screen to
reflect the required acquisition temperature for the run.
Opens the Edit Gains window. The Gains will be sent the next time the machine is started
Begins the run, setting the machine temperature to that displayed on the screen. The
temperature and channel graphs will start to display data. Ensure the Rotor-Gene displays READY
before starting another calibration run.
Stops the experiment. If the experiment is still acquiring data when you click the button, then the
machine will first finish acquiring, and then stop the machine. This process can take up to 5 seconds
for each channel data is being acquired.
NOTE: The aim of the Gain setting is to have all data on the screen. The Gain doesn't
influence your data. If the Gain was chosen too high, curves might go off scale and data
could be lost. It is therefore a good idea to start a Sybr-Green I or dual-labeled probe run with
a low fluorescence (an increase is expected) and a quenched FRET run with a higher
fluorescence (a decrease is expected).
Show at most two analysis windows:
This option ticked shows a maximum of two analysis
windows at once. By performing a quantitation analysis, three windows are opened by default. If more
than six or seven windows are opened at once, the overall view might not be clear. This option ticked
will close the first analysis window and replace it with the last opened analysis window. If the option is
unticked more than three analysis windows could be displayed. This could be useful when comparing
quantitation screens of four different channels.
To improve readability, the software removes unused windows when new
windows are opened. This option is enabled by default, as it keeps the Rotor-Gene work area clear.
You may, however, need to see more than 6 windows at once, in which case you should uncheck this
5.8.7 Display Options
© 2002 Corbett Research
Reset all "Don't Show This Message again" Dialogs:
This window forces the software to re-
display all dialog boxes again. These include messages about suspicious settings which you may have
previously set to not display again. This might be useful if a new user is now using the machine who is
unfamiliar with the Rotor-Gene or the Rotor-Gene software.
5.9 Gain Menu
for the current experiment and modify them if the experiment
has not yet been run. Use the
arrows next to each text field to modify the fields without
needing to enter in a value with the keyboard. Click OK when you have finished.
Sets the Gain of the specified channel before running a program. Gain settings will be retained from
Change gain during the first ten cycles:
This new software version also offers the possibility to
change the gain during the initial cycles. A red line will be drawn in the appropriate channel showing
where the gain has been changed. The cycles before the gain has been changed will be excluded of
the calculation of the quantitation data.
Note: This is only possible for 4-Channel Rotor-Gene, not for dual channel or 32-well Rotor-
This menu allows you to tile the window either vertically, horizontally or arrange windows in a cascade.
A further option of tiling windows is found under Arrange in the tool menu.
contains all the help files sorted by names. Simply go to contents and choose the area
gives a brief overview of new features added to the previous software
will open the help file according to the activated window in the Rotor-
will automatically open our Web page, whereas
the forum can be accessed by pressing
several details for example the serial number of the Rotor-Gene or the software version you are using.
64 Functional Overview
5.10 Window Menu
5.11 Help Menu
© 2002 Corbett Research
65 Rotor-Gene
option in the Help menu allows you to send a support email to Corbett
Research which contains all relevant information to an experiment. The
the information to a file which you can copy onto a floppy disk or onto a network if you do not have
access to email on the Rotor-Gene machine.
While running an experiment, a second already finished experiment can be opened and analyzed. Note
that several functions like
are not activated in this second window. Even closing the
first experiment will not bring up those functions. A new run can only be started from the first
experiment, once the experiment has finished. If a run has to be started from the second experiment, it
needs to be closed and reopened to access the appropriate functions.
5.11.1 Send Support E-Mail
6 General Functions Used Over Several Windows
6.1 Opening A Second Experiment
© 2002 Corbett Research
6.2 Spanner Icon
By clicking the spanner icon, you have access to a number of tools, which simplifies manual analysis
of data. Those tools could also be accessed using the right mouse button in the appropriate screens.
Saves the graph in the JPEG image format and
This function allows for the current graphic screen to be printed.
Modifies the currently selected digital filter on the graph. The digital filter smoothes data
using a sliding window of points. To open the window, click the right mouse button over a window
Opens a window with exact coordinates of the current cursor position.
Note: The following options are no longer available via the Spanner icon. You should first
drag a selection area on the graph, and then, upon releasing the mouse, a context menu will
appear. Choose one of the following options to select samples or zoom the graph.
Shows only the samples that you swipe in a rectangular region.
With the zoom function you can zoom in certain areas of the screen. It can be horizontally and
66 General Functions Used Over Several Windows
© 2002 Corbett Research
67 Rotor-Gene
will bring up a window in which you can manually enter a scale, or interactively select
one. To chose this option simply press the right mouse button over the appropriate screen.
Rotor-Gene Experiment File
attempts to fit the scale to the maximum and minimum readings in the data.
will reset the scale to show from 0-100 fluorescence units.
6.3 Scaling Options
6.4 Abbreviations
6.5 Auto-Scale
© 2002 Corbett Research
7 Quick Start Using The New Experiment Wizard
To perform a new experiment, click on the New icon .
By default the setup Wizard is active and you should see the wizard screen appear. If the Wizard does
not appear tick the appropriate box in New Experiment.
The wizard comes up with a welcome screen. After reading the introduction, the box
introductory screen in the future
can be ticked. The welcome screen will then automatically be
7.1 Welcome Screen
68 Quick Start Using The New Experiment Wizard
© 2002 Corbett Research
69 Rotor-Gene
The setup wizard guides you through all the necessary steps to setting up an experiment.
7.2 Page 1
© 2002 Corbett Research
7.3 Page 2
can be modified. Simply press the
appropriate buttons and modify as outlined in previous chapters.
70 Quick Start Using The New Experiment Wizard
© 2002 Corbett Research
71 Rotor-Gene
The last screen gives a summary of what you are about to run. Always check the parameters. If you
are satisfied with the parameters click
and you will be prompted for a file name. Before
starting the run make sure that the Rotor-Gene is in
NOTE: If the unit does not display READY on the LCD screen, the detectors will not be
activated and no data will be collected.
7.4 Page 3
© 2002 Corbett Research
7.5 Page 4
Once the experiment has begun, you can enter in sample types and descriptions while you wait for it to
complete. The functionality of this screen is identical to the
can be detected on Channel 1 and Channel 2, when labeled with FAM or JOE
for the above fluorophores could be BHQ1 or TAMRA.
FRET style quenched probes
could also be used labeled with FAM and JOE (VIC).
72 Quick Start Using The New Experiment Wizard
8 Rotor-Gene Hardware Information
8.1 Filter Specifications
8.1.1 32 Well and Dual-Channel Machines
© 2002 Corbett Research
73 Rotor-Gene
NOTE: The above quenched dyes can be multiplexed in the same tube with no cross-talk in
Channel 1 and less than 1% cross-talk in Channel 2.
is detected on Channel 1 (FAM Channel) or Sybr Channel.
For multiplex reactions we recommend using FAM, JOE, ROX and Cy5, since those fluorophores after
optimization give a cross talk less then 1%.
to be used for the above dyes:
TAMRA for FAM or JOE (VIC)
8.1.2 Multi-Channel Machines
© 2002 Corbett Research
FAM and JOE for example have to be detected on the above mentioned channels (quenched FRET).
If Cy5 has to be detected (normal FRET) the filter combination must be as follows:
FAM-Cy5, FAM-LC 640, FAM-
9 Troubleshooting
The software keeps an untouched copy of each experiment, along with diagnostic information in its
Log Archive repository. By using the Help, Send Support Email option, you can send an email with all
the diagnostic information the Corbett Research support technicians require:
Previously, customers were required to send the experiment file, the temp.log and rot.log files. This is
To troubleshoot experiments which did not occur as expected, enter the Profile Editor and ask the
9.1 Log Archives
74 Rotor-Gene Hardware Information
9.2 Troubleshooting Experiment Files
© 2002 Corbett Research
75 Rotor-Gene
Is the initial denature temperature and time for the Taq Polymerase used appropriate?
the manufacture's description of the enzyme. The recommended temperature is 95°C and the times
can vary from 2 min to 15 min, depending on the enzyme used. If genomic DNA has been used for the
assay, the initial denature time should be at least 5 min depending on how the DNA has been
Have 50 ml volumes been run on the Rotor-Gene?
The Rotor-Gene is calibrated for 20 – 25
l, in which case we recommend a denaturation time during
cycling of 15 seconds. Some protocols specify 50
l reactions in which case we recommend
increasing the denaturation time during cycling to 30 seconds.
Is the denature temperature appropriate?
Most double stranded DNA has melted at 95°C. In some
cases temperatures from 90°C to 94°C may not denature the DNA properly. Due to partially denatured
DNA, the accessibility of primers and probes could be reduced and could therefore reduce the
reaction efficiency of the assay.
Is the denature time appropriate?
15 to 20 sec is usually enough to denature the amplicon,
however, longer products might need up to 30 sec. A denature time of 60 sec is usually not required.
Is the annealing temperature and time appropriate?
Check the Tm's of the primers and probes.
The annealing time for Sybr-Green I is around 20 to 35 sec. Dual labeled probes are often run as a
two step profile where annealing and extension steps are combined. In this case the
annealing/extension step is between 45 and 60 sec and the temperature is usually 60°C. For FRET
probes the annealing step is between 20 and 30 sec.
Is the extension temperature and time appropriate?
Check the recommendations of the
manufacturer's Taq Polymerase. For a product less than 300 bp the recommended time is between 15
At what step have data been acquired?
For Sybr-Green I data should be acquired at 72°C. This is
where most of the DNA is expected to be double stranded. As mentioned above, dual labeled probes
are often run as two step assays. Data therefore should be acquired at the anneal/extension step. For
FRET assays, data should be acquired at the annealing step. If in doubt about the appropriate data
acquisition point, data could be acquired at several steps for comparison. If no raw data can been seen
on the screen at all, go to edit profile and check if data was acquired at least at one point.
Has a hold step at the same temperature as the beginning of the melt been inserted before
If no hold step has been used before running a melt curve, the results in the
melt curve analysis could be a steep increase in the first few cycles making the actual melt peak hardly
For Sybr-Green I melt curves; was the melt curve run to 99°C?
Some amplicons melt higher than
90°C. The expected melt peak might not occur, due to the fact that DNA has not melted.
9.2.1 Initial Denature Step
9.2.2 Cycling Profile
9.2.3 Melt Curve Analysis
© 2002 Corbett Research
9.2.4 Gain Settings
Have the appropriate gain settings been chosen?
In some cases it has been observed that curves
in the raw data went off scale resulting in a straight line at a fluorescence of 100. Although quantitation
of most data could still be performed, the gain should be reduced to a setting, where the raw data does
In some cases it has been observed that the fluorescence went off scale up from the first cycle. This
is due to the fact that the gain setting was chosen too high. It might look like that no data has been
If a straight line at a fluorescence of 100 at the beginning of a melt curve is observed, the melt curve
should be rerun using a lower gain. There is no need to rerun the amplification, Sybr-Green I or FRET
samples can be reused to a certain extent.
Has the appropriate Quantitation Setting been chosen?
In some cases it has been observed that
the raw data show a perfect amplification but no data could be seen in the quantitation screen. Setting
the Quantitation Setting to zero will show all data. The Quantitation Setting can be used to exclude
small changes of fluorescence, due to probe degradations or other probe related effects, which do not
show true amplification but rather a steadily increasing line.
If the quantitation curves can be excluded with a minimal Quantitation Setting the gain for the next run
should be increased. Note: the gain setting has no influence on the actual data, the expected fold
increase will be similar.
Has the appropriate setup been chosen for the Rotor-Gene?
To check, go to file menu then to
setup and tick the appropriate boxes. If in doubt what each of the boxes means or if setup is not
accessible, please contact the distributor.
Has the appropriate rotor for the run been chosen?
Before running an experiment make sure that
either the 36 or 72 well rotor has been chosen according to the rotor in the Rotor-Gene. Choosing the
wrong rotor might either result in not acquiring every second sample (running 72 well rotor but 36 well
setup was chosen) or acquiring double the amount of data that is actually in the rotor (running 36 rotor
but 72 setup was chosen).
What version of the software is used on the Rotor-Gene?
The software development for the
Rotor-Gene is ongoing and freely downloadable from the Corbett Research web page
. Make sure that a recent version is uploaded and used.
If samples are not detected check also the channel setup. Make sure data has been acquired with the
appropriate excitation and detection filters. This is particularly important when user defined filters have
Depending on the environment in which the Rotor-Gene is setup, spikes could be caused due to the
noise from mains. The Rotor-Gene software has a build in spike suppression and will tell the user if
spikes occurred during a run. If normal or medium spike suppression does not suppress spikes, high
spike suppression should be ticked (version 4.4 or less). More recent versions have only one option to
suppress spikes, by either enabling or disabling spike suppression. Spike suppression (high) is
recommended and can be chosen in the experiment wizard.
What if all the above mentioned conditions were right, but the assay did not look as expected
In the restricted area of our homepage there is a file called "real-time
76 Troubleshooting
9.2.5 General Setup Conditions
9.2.6 Experiment Settings
© 2002 Corbett Research
77 Rotor-Gene
summary". Steps are suggested on how to optimize Sybr-Green I, dual labeled probes and FRET
reactions. In general, the expected fold increase of a Sybr-Green I reaction is between 5 and 50 times
and a dual labeled reaction between 2 and 10 times.
On some computers, the regional settings are configured in a contradictory manner which causes a
conflict within the Rotor-Gene system. The problem occurs when different decimal settings are used
for curencies and numbers. For example, if numbers are displayed in a "123 456,789" format, but
currencies are shown as "$299,192.20", then there is a conflict, since the number "20,123" is
ambiguous. On Windows 98 machines, this prevents the Rotor-Gene software from correctly
interpreting numbers. This problem does not occur on Windows 2000.
To correct this setting, please follow the following steps. The names of these options may be different
in your system's language.
Double-Click on Regional Settings.
Double-Click on Regional Options.
Click on the Numbers tab.
Note down which is the Decimal Symbol, and which is used as the Digit Grouping Symbol. On a
German system, for example, the comma (,) will be used for Decimals, and the period (.) is used as
Click on the Currency tab.
Change the Decimal Symbol and the Digit Grouping Symbol to be the same as on the Numbers tab.
Re-run the Rotor-Gene software.
To run software with the Rotor-Gene 3000 you have to follow these steps:
1. Setting up the Rotor-Gene 3000 Software for the first time:
Insert the Rotor-Gene 3000 CD into your CD drive.
Double Click on My Computer icon.
Double Click on Compact Disc Rotor-Gene 3000.
Double Click on RgSetup folder.
Double Click on executable file Rotor-Gene_X_Y_ZZ.exe (where X is Major version number, Y is Minor
version number, ZZ is Revision version number).
Follow all Installer instructions.
2. Setting up the Rotor-Gene 3000 Software on a PC where it was already installed.
If you install Rotor-Gene 3000 Software on PC where it was installed before, it is necessary to
uninstall any previous version of Rotor-Gene on your system.
In order to uninstall Rotor-Gene Software:
Click on Start button on Taskbar,
choose Settings, choose Control Panel, double click on Add/Remove Programs icon, choose Rotor-
9.3 Regional Settings In Windows 98
10 Rotor-Gene 3000 Setup Instructions
© 2002 Corbett Research
Insert the Rotor-Gene 3000 CD into your CD drive.
Double Click on My Computer icon.
Double Click on Compact Disc Rotor-Gene 3000.
Double Click on RgSetup folder.
Double Click on executable file Rotor-Gene_X_Y_ZZ.exe (where X is Major version number, Y is
Minor version number, ZZ is Revision version number).
11 Rotor-Gene 2000 Setup Instructions
To run software with Rotor-Gene you have to follow through these steps from 1-8 to install the board,
the software drivers and the Rotor-Gene Software.
To run the software with the Rotor-Gene the Virtual Machine must not be ticked.
If you want to install a demonstration version of Rotor-Gene Software you have to do Step 7 only
(Installing the Rotor-Gene Software). When you run software for demonstration purposes leave Virtual
In order to change Virtual Machine box must click <File> <Setup> on the main Rotor-Gene Software
If you Install Rotor-Gene Software on PC running Windows 95/98 you must
disable Interrupt 3 in BIOS.
Boot the computer and Press Delete on keyboard simultaneously.
Choose INTEGRATED PERIPHERALS.
Change On Board Serial Port2: Disabled (Could be different on different PC i.e. Onboard Serial Port2
If you install the Rotor-Gene Software on PC running Windows NT you must disable Interrupt 3 in
Click on the Start Menu, Settings and Control Panel
Double-click on System. Then select the second tab Device Manager.
Click on the [+] sign next to Ports(COM+LPT) then click on COM2 If it still exist there.
Tick Disable in this hardware profile and click OK.
Click Close for Device Manager.
Disable hard disk power down
Click on the Start Menu, Settings and Control Panel
Double Click on Power Management
Choose Never for hard disks
Installing the Data Acquisition board
78 Rotor-Gene 3000 Setup Instructions
© 2002 Corbett Research
79 Rotor-Gene
Set the Interrupt jumper on the board to IRQ3.
Turn on your computer again.
Installing the Computer Boards Software
Insert the Rotor-Gene 2000 CD into your CD drive.
Type in D:\RG2000\UNIVLIB5.04\DISK1\SETUP.EXE then press ENTER. If the drive letter of your
CD Rom is one other than D, replace D with that drive letter.
The installation software will now start up. An screen will ask you to select which files to install. Select
ONLY "Win 95/98/NT 32-bit Universal Library and Instacal", unticking any other options.
When the software asks you if it can make changes to AUTOEXEC.BAT, select "Yes". Do likewise for
The software will then ask you if it can reboot the computer. Select Yes.
Configuring The Computer boards Software
Type c:\cb\inscal32.exe and click OK.
Click on Install, Add Board…
Select the board CIO-DAS802 and click OK.
Double-click on the CIO-DAS802 board.
Set the Interrupt Level to 3.
For each Channel from 0 to 7 in channel configuration, change from DIFF to SE.
Installing the Rotor-Gene Software
If you install Rotor-Gene Software first time go to 7.a .
Before installing any new version of the software, it is necessary to uninstall any previous version of
Rotor-Gene on your system.
Click on Start, Settings, Control Panel.Double click on Add-remove software.
In the list box at the bottom of the window, for each copy of Rotor-Gene in the list, click on Rotor-
Gene, then click on Add/Remove. Click on Close.
You can now run setup.exe, located in D:\ RGSetup\Setup.exe
If the drive letter of your CD Rom is one other than D, replace D with that drive letter.
© 2002 Corbett Research
12 Rotor-Gene 2000 IMPORTANT SETTINGS
IMPORTANT SETTINGS THAT SHOULD NOT
DO NOT Turn off hard disks. Ensure Power-down on hard disks is set
DO NOT Put the system on Standby. Ensure Enter Standby mode is set
DO NOT Install any peripheral devices that are using Interrupt 3.
DO NOT Install computer on network if the network card is using
DO NOT Enable Interrupt 3 in BIOS.
DO NOT install custom screen savers.
NB: These notes only apply to the Rotor-Gene 2000.
80 Rotor-Gene 2000 IMPORTANT SETTINGS
© 2002 Corbett Research