Please read this operating manual completely before using the device for the first time. Also
observe the instructions for use enclosed with the accessories.
Please view this operating manual as part of the product and keep it somewhere easily
accessible.
When passing the device on to third parties, be sure to include this operating manual.
If this manual is lost, please request another one. The current version can be found on our
website www.eppendorf.com
1.2 Danger symbols and danger levels
1.2.1Danger symbols
Toxic substancesElectric shock
Hazard pointMaterial damage
Cuts
1.2.2Danger levels
The danger level is a part of a safety note and distinguishes the possible results of
non-observance from each other.
1 Grid head for holding the capillary 2 Front knurled screw
3 Piezo element4 Connection for pressure tube
5 Connection to PiezoXpert
6
Page 7
Eppendorf PiezoXpert™ — Operating manual
Abb. 3:Foot control
A
1
Fig. 3: Foot control
1 Left foot control: parameter set A2 Right foot control: parameter set B
B
2
2.2 Delivery package
QuantityDescription
1PiezoXpert
1Actuator
1Foot control
1Power cable
1Spacer plate (for installation on TransferMan NK 2 and PatchMan NP 2)
1PiezoXpert operating manual
Store the shipping box and packing material in order to store the device or transport it at
Hint!
a later time.
Check the delivery for completeness.
Check all parts for damage in transit.
EN
Operating manual
2.3 Features
The PiezoXpert is used in micromanipulation. Penetration into cells is supported by piezo
impulses which are directly transmitted to the capillaries loss-free. Intuitive operation and an
additional adjustment range enable reproducible work.
• Optimal transmission of piezo impulses: The piezo element is located at the front of the
capillary holder.
• Loss-free transmission of piezo impulses: The actuator (capillary holder with piezo element)
is rigidly mounted to the micromanipulator.
• Reproducible work: Intensity, speed (frequency) and quantity of piezo impulses can be
simultaneously set in two application-specific sets of parameters. Three memory locations
are available for storing the settings.
• Cleaning function: Parameters for removing contaminations from the capillary can be set and
stored.
• Simple operation: The device can be operated intuitively. piezo impulses can be triggered on
the device or by using the foot control.
7
Page 8
Eppendorf PiezoXpert™ — Operating manual
EN
3 Safety
3Safety
3.1 Intended use
The PiezoXpert supports micromanipulation and microinjection in research. The PiezoXpert is
intended exclusively for indoor use.
3.2 User profile
This device may only be operated by trained and skilled personnel.
Before using the device, read the operating manual carefully and familiarize yourself with the
device's mode of operation.
Operating manual
3.3 Information on product liability
In the following cases, the designated protection may compromise the safety of the device.
Liability for material damage and personal injury is transferred to the operator:
• The device is not used in accordance with the operating manual.
• The device is used outside of its intended use.
• The owner has made unauthorized modifications to the device.
3.4 Warnings for intended use
WARNING!
WARNING!
WARNING! Damage to health from toxic, radioactive or aggressive chemicals as
well as infectious liquids and pathogenic germs.
Observe the national regulations for handling these substances, the biological
security level of your laboratory, the material safety data sheets and the
manufacturer's application notes.
Wear personal protective equipment (PPE).
WARNING! Lethal voltages inside the device.
Ensure that the housing is always closed and undamaged so that no parts inside
the device can be contacted by accident.
Do not remove the housing of the device.
Do not allow any liquids to penetrate the inside of the housing.
Do not allow the device to be opened by anyone except service personnel who have
been specifically authorized by Eppendorf.
WARNING! Electric shock due to damage to device or mains cable.
Only switch on the device if the device and mains cable are undamaged.
WARNING!
Only use devices that have been properly installed or repaired.
In case of danger, disconnect the device from the mains supply.
WARNING! Risk from incorrect supply voltage
Only connect the device to power supplies which correspond with the electrical
WARNING!
requirements on the nameplate.
Only use sockets with a protective earth (PE) conductor and a suitable mains cable.
8
Page 9
Eppendorf PiezoXpert™ — Operating manual
WARNING! Risk of injury due to flying capillaries and glass splinters.
If exposed to high pressures, capillaries can detach themselves from the grip heads and
WARNING!
become projectiles.
Capillaries can crack as a result of incorrect handling.
Wear protective goggles.
Never aim capillaries at people.
Use capillaries with an outer diameter that matches the grip head specifications.
Always mount capillaries depressurized.
Mount the capillary correctly in the grip head.
Do not touch the capillary with the Petri dish or other objects.
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf
CAUTION!
may impair the safety, function and precision of the device. Eppendorf cannot be held
liable or accept any liability for damage resulting from the use of incorrect or
non-recommended accessories and spare parts or from the improper use of such
equipment.
Only use accessories and original spare parts recommended by Eppendorf.
CAUTION! Risk of cuts when unpacking the capillaries
Capillaries can break as a result of incorrect unpacking.
CAUTION!
Do not reach into the capillary transport protection.
EN
Operating manual
CAUTION! Risk of cuts from broken capillaries.
Capillaries are made of glass and are very fragile.
CAUTION!
Wear personal protective equipment (PPE).
Always mount capillaries depressurized.
Never aim capillaries at people.
Handle the capillaries very carefully.
3.5 Safety instructions on the device
DepictionMeaningLocation
Follow the instructions in the
operating manual.
Rear side of the device
9
Page 10
Eppendorf PiezoXpert™ — Operating manual
EN
4 Installation
4Installation
4.1 Selecting the location
Please select the location for the PiezoXpert according to the following criteria:
• Mains connection (230 V/120 V/100 V) according to the name plate. This is located on the
rear side of the device.
• At least 10 cm away from adjacent devices and walls.
• Solid bench with stable, horizontal and even work surface.
4.2 Installing the actuator on the micromanipulator
Operating manual
The actuator is mounted on the micromanipulator in the same manner as the capillary holder.
NOTICE!
NOTICE!
Hint!
NOTICE! Damage to the actuator.
Impacts or vibrations can cause the piezo element to malfunction or fail.
Do not allow the actuator to fall.
Do not expose the actuator to strong vibrations.
If you suspect the actuator has been damaged, have it inspected by Eppendorf
Service.
NOTICE! Damage to the piezo element.
The piezo element may become damaged if is twisted.
It may only be rotated on the knurled screws.
Do not use the piezo element as a lever.
When the actuator is mounted for the first time, the micromanipulator must be converted
or realigned as necessary. Follow the instructions in the operating manual of your
micromanipulator.
10
Page 11
Eppendorf PiezoXpert™ — Operating manual
4.2.1Installing the actuator on the InjectMan® NI 2 or micromanipulators from other
1. Loosen the knurled screw on the micromanipulator tool holder.
2. Insert the actuator into the support.
3. Tighten the knurled screw.
4. Set an angle between 0° and 25°. The flatter the angle, the more direct the effect of the
manufacturers
piezo impulses.
The PiezoXpert will only optimally transmit the piezo impulses to the capillaries if it is
Hint!
firmly mounted.
• Ensure that the actuator is secured and cannot be moved in the support.
EN
Operating manual
11
Page 12
Eppendorf PiezoXpert™ — Operating manual
EN
4.2.2Installing the actuator on the TransferMan® NK 2 or the PatchMan NP 2
In the following description, the TransferMan NK 2 or PatchMan NP 2 is mounted on the right
side of the microscope. If necessary, the X head must be converted in order to mount the
actuator in the same manner as the capillary holder.
A conversion is not necessary if, e.g., the TransferMan NK 2 is operated using a 10°
plate and an A head.
If you use very long capillaries, the TransferMan NK 2 motor module can be moved
further away from the operating point. A more precise description can be found in the
TransferMan NK 2 operating manual in the chapter "Changing the position of the motor
module on the X-axis".
234
1
1. Loosen the M3x12 cheese head screw and remove the X head.
2. Rotate the X head 180°. Place the fitting pin
(on the rear side of the X head) in the center
bore of the tool holder.
Tighten the X head using the M3x12 cheese
head screw.
3. Loosen the knurled screw. Remove the knurled screw and pressure plate from the X head.
12
Page 13
1
Eppendorf PiezoXpert™ — Operating manual
4. Place the provided spacer plate on the bore
of the X head.
Secure the spacer plate using the knurled
screw and pressure plate.
5. Place the actuator in the upper (1) or lower
(2) groove of the spacer plate.
6. Tighten the knurled screw to secure the
actuator between the spacer plate and
pressure plate.
7. Set an angle between 0° and 25°. The flatter the angle, the more direct the effect of the
piezo impulses.
EN
Operating manual
2
The PiezoXpert will only optimally transmit the piezo impulses to the capillaries if it is
Hint!
firmly mounted.
• Ensure that the actuator is secured and cannot be moved in the support.
• Ensure that the piezo element is not pushed against a surface while it is warped.
13
Page 14
Eppendorf PiezoXpert™ — Operating manual
EN
4.3 Mounting the capillary
WARNING!
Operating manual
CAUTION!
NOTICE!
Hint!
WARNING! Risk of injury due to flying capillaries and glass splinters.
If exposed to high pressures, capillaries can detach themselves from the grip heads and
become projectiles.
Capillaries can crack as a result of incorrect handling.
Wear protective goggles.
Never aim capillaries at people.
Use capillaries with an outer diameter that matches the grip head specifications.
Always mount capillaries depressurized.
Mount the capillary correctly in the grip head.
Do not touch the capillary with the Petri dish or other objects.
CAUTION! Risk of cuts from broken capillaries.
Capillaries are made of glass and are very fragile.
Wear personal protective equipment (PPE).
Always mount capillaries depressurized.
Never aim capillaries at people.
Handle the capillaries very carefully.
NOTICE! Damage to the piezo element.
The piezo element may become damaged if is twisted.
It may only be rotated on the knurled screws.
Do not use the piezo element as a lever.
Standard capillary: Only use the 0 gear head with capillaries with an outer diameter of
1.0 mm to 1.1 mm. If you would like to use other capillaries, order the matching gear
head.
For the piezo impulses to be optimally transmitted to the capillaries, the capillary must
Hint!
come into contact with the metal in the inside of the piezo element. Make sure that the
capillaries are pushed into the actuator up to the internal stop.
1. Loosely screw the grip head into the front
knurled screw of the actuator.
14
Page 15
Eppendorf PiezoXpert™ — Operating manual
2. Push the capillary into the grip head as far
as it will go.
3. Tighten grip head into the front knurled screw.
4. Bring the capillary into the microscope focus range.
5. Rotate the front knurled screw to align the
angled capillary.
Use the straight or angled capillaries with an angle up to 25°. If you would like to use
Hint!
polished capillaries with spikes, check their aptitude using pretests.
EN
Operating manual
15
Page 16
Eppendorf PiezoXpert™ — Operating manual
EN
4.4 Connecting PiezoXpert and the actuator
WARNING!
NOTICE!
Operating manual
1 Connection (currently non-functioning) 2 Power switch
3 Fuse holder4 Power connection socket
5 Connection for foot control6 USB interface (only for Technical
7 Actuator connection
WARNING! Risk from incorrect supply voltage
Only connect the device to power supplies which correspond with the electrical
requirements on the nameplate.
NOTICE! Material damage from incorrect connections.
Only electrical connections may be made to devices described in the operating
manual.
Other connections are permitted only following consultation and agreement with
Eppendorf AG.
Only devices which comply with safety requirements according to IEC 60950-1 may
be connected.
123
4567
Service)
Requirement
• PiezoXpert is switched off.
• The power plug is disconnected.
1. Connect the actuator cable to the "Actuator connection" 7.
2. Connect the foot control to the "Connection for foot control" 5.
3. Optional: Connect the pressure tube (from CellTram or FemtoJet) to the actuator (pressure
tube connection).
4. Plug the power cable into the "Power connection socket" 4.
5. Plug the power plug into the socket.
6. Switch on the PiezoXpert at the power switch.
16
Page 17
Eppendorf PiezoXpert™ — Operating manual
5Operation
5Operation
5.1 Overview of operating controls
Become acquainted with the operating controls and the display before using the PiezoXpert for
the first time.
Abb. 4:Operating controls and display
prog
Fig. 4: Operating controls and display
1 Program keys2 prog clean program key
3Display4clean key
5 B rotary knob6 pulse program key
7 speed program key8 int program key
9A rotary knob
12
1
prog2prog
int
8
speed
3
prog clean
pulse
7
6
3
clean
59
4
EN
Operating manual
1Cursor
2 Parameter set A
Marks the active and modifiable
parameters in parameter set A.
3 Parameter set B4 Cursor
Marks the active and modifiable
parameters in parameter set B.
4321
17
Page 18
int
speed
Eppendorf PiezoXpert™ — Operating manual
EN
5.2 Setting the piezo impulses
The parameters for parameter sets A and B can be set independent of each other using the A
and B rotary knobs.The prog 1, prog 2 and prog 3 program keys can be used to save both
parameter sets at the same time.
5.2.1Setting parameters for piezo impulses
ParameterFunction
Operating manual
pulse
1. Press on one of the program keys: int, speed or pulse.
The control LED above the key lights up. The and cursors mark the parameter in the
display.
2. Rotate the A rotary knob to change the parameter in parameter set A. Rotate the B rotary
knob to change the parameter in parameter set B.
The set parameters are activated immediately.
Intensity of piezo impulses
Select a value between 1 and 50 to define the strength of the piezo
impulses.
Speed (frequency) of the piezo impulses
Select a value between 1 and 10 to define how many piezo impulses
should be emitted per second.
Quantity of piezo impulses
Select a value between 1 and 10 or ∞ (infinity) to define the total number
of piezo impulses to be emitted.
∞ (infinity): piezo impulses will be emitted as long as the rotary knob or
foot control is depressed. After the impulse series, the quantity of emitted
piezo impulses will be displayed next to the ∞ symbol.
18
If the int, speed or pulse parameter keys are pressed, the corresponding parameter
Hint!
will stay marked for 8 seconds. The control LED above the parameter key lights up. The
control LED goes out when the parameter key is pressed again or a program key or the
clean key is pressed.
If a parameter is marked and triggered using the rotary knob or foot control, the marking
is extended by 8 seconds. This allows you to set, test and, if necessary, immediately
change a parameter.
Page 19
Eppendorf PiezoXpert™ — Operating manual
5.2.2Saving parameters
1. Set the parameters for parameter sets A and B.
2. Press a program key for longer than 1 second.
prog 1
The control LED above the program key lights up. The parameters are saved.
If you select settings for parameter sets A and B that are already saved, the LED above
Hint!
the program key will light up.
5.2.3Accessing saved parameters
prog 2
prog 3
Briefly press the prog 1, prog 2 or prog 3 program key to access the saved parameters.
The saved parameters will be shown in the display.
5.3 Triggering piezo impulses
1. Set the required parameters using the A or B rotary knobs or access the saved parameters
using the program keys.
2. Press the left foot control or A rotary knob to trigger the parameter set A. Press the right foot
control or B rotary knob to trigger the parameter set B.
As long as piezo impulses are emitted (rotary knob or foot control pressed), the rotary knob is
underlaid with a blue pool of light.
Pulse ∞ (infinity): Piezo impulses will be emitted as long as the rotary knob or foot
Hint!
control are pressed. The quantity of emitted piezo impulses is displayed after the
impulse series.
5.4 Optimizing parameters for capillaries
Clarify the following questions to ensure that the piezo impulses is transmitted immediately and
loss-free:
• Which capillary is best suited for the planned application?
• Which angle will the capillary be used in?
• Does the capillary need to be filled? With what?
Set the piezo impulses according to the application, the capillary used and the filling
(see Optimizing parameters on p. 20).
5.4.1Preparing capillaries with larger opening diameters
The following section contains a description of how to prepare capillaries with larger opening
diameters, e.g., for ES cell transfer to morula or blastocysts.
EN
Operating manual
WARNING! Damage to health due to toxic, radioactive or aggressive chemicals.
Observe the national regulations for handling these substances as well as the
WARNING!
material safety data sheets and manufacturer's application notes.
Wear personal protective equipment (PPE).
19
Page 20
Eppendorf PiezoXpert™ — Operating manual
EN
1. Mount the capillary in the actuator grip head (see Mounting the capillary on p. 14).
2. Using the CellTram vario, fill the mounted capillary with oil from the rear.
3. Absorb the medium via the front capillary opening.
4. Initially, select a higher setting (e.g., 20) for the Int. (intensity) parameter.
Alternative capillary filling
• Weigh down the capillary by filling it with liquids with higher specific densities.
• Fill the capillary with a liquid with a high specific density, free of air bubbles, and then with oil.
In this case, reduce all parameters, especially Int. (intensity).
5.4.2Preparing capillaries with smaller opening diameters
Operating manual
The following section contains a description of how to prepare capillaries with smaller opening
diameters, e.g., for mouse ICSI.
WARNING!
1. Fill the capillary from the rear using a microloader, e.g., Fluorinert™ FC-77.
2. Mount the capillary in the actuator grip head (see Mounting the capillary on p. 14).
3. Absorb the medium via the front capillary opening.
4. Initially, select a lower setting (e.g., 10) for the Int. (intensity) parameter.
5.4.3Optimizing parameters
The PiezoXpert offers you a wide range of setting options. To determine the optimal parameters
for your application, proceed as follows:
1. Set the parameters for the speed and quantity of the piezo impulses to 1.
Speed: 1
Pulse: 1
2. Set the parameter for Int. (intensity) to a low value (e.g., 10).
Int.: 10 (example)
3. Incrementally increase the value for Int. until the piezo impulses is strong enough to penetrate
the cell wall.
4. Now, fine tune the Speed and Pulse parameters.
Or, you can start with a high value for the intensity (e.g., Int.: 30) and optimize the strength of the
pulse as the values decrease.
If you do not have success with the optimized parameters, you have to take into consideration
that the PiezoXpert cannot deliver the required contribution on its own.
WARNING! Damage to health due to toxic, radioactive or aggressive chemicals.
Observe the national regulations for handling these substances as well as the
material safety data sheets and manufacturer's application notes.
Wear personal protective equipment (PPE).
20
Page 21
Eppendorf PiezoXpert™ — Operating manual
5.5 Clean function
Contamination on the outside of the capillaries, for example, can be removed using the Clean
function.
5.5.1Accessing the Clean function
To access the Clean function, press the
prog clean program key.
• The display shows the parameters set for
Clean.
• The following parameters are factory preset
for Clean:
Int.: 20
Speed: 20
To exit the Clean function, briefly press the prog 1, prog 2 or prog 3 program keys.
You can also exit the Clean function by pressing the A or B rotary knob or the foot
Hint!
control. The display shows the most recently used set of A and B parameters. Piezo
impulses are transmitted to the capillary.
5.5.2Setting the parameters for the Clean function
1. Press the clean key or the prog clean program key.
2. Press the int or speed parameter key.
The control LED above the parameter key
lights up.The cursor marks Int. or Speed in
the display.
3. Rotate the B rotary knob to change the parameter.
For Int., select a value between 1 and 86.
For Speed, select a value between 1 and 40.
The set parameters are activated immediately.
5.5.3 Saving the parameters for the Clean function
1. Set the parameters Int. and Speed for Clean.
2. Press the prog clean program key longer than 1 second.
The control LED above prog clean lights up. The parameters are saved.
EN
Operating manual
If you select settings for Clean that are already saved, the LED above the prog clean
Hint!
program key lights up.
5.5.4Accessing parameters saved for the Clean function
Briefly press the prog clean program key to access the saved parameters.
The saved parameters will be shown in the display.
21
Page 22
Eppendorf PiezoXpert™ — Operating manual
EN
5.5.5Activating the Clean function
Hold the clean key pressed down.
Piezo impulses will be transmitted to the capillary as long as the clean key is pressed. The
display shows the parameters set for Clean.
If the clean key is released, the display shows the most recently used parameter sets, A and B,
after approx. 5 seconds.
5.6 Setting the display contrast
1. Press the int and speed keys simultaneously.
Operating manual
The display shows the LCD-Contrast menu.
2. Rotate the B rotary knob to set the contrast
of the display.
Highest contrast: 0
Lowest contrast: 100
3. Press the pulse key to exit the menu.
6 Troubleshooting
6Troubleshooting
6.1 General errors
Control functions in the device ensure that the PiezoXpert can continue to be used even if an
error code appears in the display. Errors are saved in a list that can be accessed.
Symptom/messageRemedy
An error code appears in the display.
An error code reappears.
Error messages can be acknowledged
using any key.
Switch the device off and back on again.
6.1.1Accessing error codes
This function can be used to display the error codes of the last 10 errors.
1. Press the int and speed keys simultaneously.
The display shows the LCD-Contrast menu.
2. Rotate the A rotary knob to display the
stored error codes.
The display shows the error codes of the
last 10 errors under LAST ERRORS.
22
Page 23
A B
Please connect
Actuator!
Eppendorf PiezoXpert™ — Operating manual
6.2 Error messages
If the recommended troubleshooting measures are not successful after several
Hint!
attempts, please contact your local Eppendorf partner. The addresses can be found on
the website: www.eppendorf.com. The addresses of Eppendorf subsidiaries can be
found on the second to last page of this operating manual.
Symptom/
message
Display remains
dark.
The display shows:
An error message
appears in the
display.
CauseRemedy
• The device was not connected
to the power supply or the
power switch was not switched
on.
Check the power connection
and power cable.
Switch on the device.
• The actuator is not connected. Error messages can be
acknowledged using any key.
Connect the actuator.
Check the actuator connection
to the PiezoXpert.
• The actuator is defective. Error messages can be
acknowledged using any key.
Replace the actuator.
Have the defective actuator
inspected by Eppendorf.
• A control function has detected
an error.
Error messages can be
acknowledged using any key.
Switch the device off and back
on again.
Contact Technical Service if
error messages continue to
appear.
EN
Operating manual
7 Maintenance
7Maintenance
7.1 Replacing fuses
DANGER!
The fuse holder is located between the power connection socket and the power switch. The fuses
may only be replaced with the same type of fuses.
1. Disconnect the power plug.
DANGER! Electric shock.
Switch off the device and disconnect the power plug before starting maintenance or
cleaning work.
23
Page 24
1
2
Eppendorf PiezoXpert™ — Operating manual
EN
2. Press the left and right plastic springs 1
together. Completely remove fuse holder 2.
3. Replace the defective fuses. Reinsert the
Operating manual
fuse holder.
7.2 Cleaning
DANGER!
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect it from the power supply before starting
cleaning or disinfecting work.
Do not allow any liquids to penetrate the inside of the housing.
Do not spray clean/spray disinfect the housing.
Only connect the device to the power supply if it is fully dried inside and out.
NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as
NOTICE!
strong and weak bases, strong acids, acetone, formaldehyde, halogenated
hydrocarbons or phenol.
If the device becomes contaminated with aggressive chemicals, clean it
immediately with a mild cleaning agent.
NOTICE! Damage to the actuator due to penetrating liquid.
Penetrating liquid can damage the piezo element.
NOTICE!
Do not clean the actuator under running water.
Moisten a towel with mild cleaning fluid and dimineralized water and remove contamination
from the outside of the device and actuator.
24
Page 25
Eppendorf PiezoXpert™ — Operating manual
8 Technical data
8Technical data
8.1 Power supply
Power connection100 V to 240 V ±10 %, 50 Hz to 60 Hz
Automatically adjusted to the voltage.
Fuses for 100 V to 240 VT3.15A, 250 V (2 pcs.)
Power consumption18 W
Protection classI
Overvoltage categoryII
8.2 Ambient conditions
AmbienceOnly for use in indoors.
Ambient temperature15 °C to 35 °C (59 °F to 95 °F)
Relative humidity30 % to 65 %, dewing not permitted
Atmospheric pressure80 kPa to 106 kPa
Suitable for use up to a height of 2000 m above sea level
Degree of pollution2
8.3 Weight/dimensions
EN
Operating manual
Weight2,8 kg (6.17 lb)
DimensionsWidth: 17 cm (6.69 in)
Height: 11.5 cm (4.53 in)
Depth: 23 cm (9.06 in)
25
Page 26
Eppendorf PiezoXpert™ — Operating manual
EN
8.4 Interfaces
USBOnly for Technical Service.
RS232Currently non-functioning.
NOTICE!
Operating manual
8.5 Piezo impulse parameters
Intensity (Int.)1 to 50
Speed (Speed)1 to 10
Quantity (Pulse)1 to 10, ∞
Clean function
Intensity (Int.)
Clean function
Speed (Speed)
NOTICE! Material damage from incorrect connections.
Only electrical connections may be made to devices described in the operating
manual.
Other connections are permitted only following consultation and agreement with
Eppendorf AG.
Only devices which comply with safety requirements according to IEC 60950-1 may
be connected.
• 1–22 can be set in increments of 1
• 22–50 can be set in increments of 4
• 1–10 can be set in increments of 1, 1 Hz per step
• 1–10 can be set in increments of 1
1 to 86
• 1–22 can be set in increments of 1
• 22–86 can be set in increments of 4
1 to 40
• 1–40 can be set in increments of 1
• 1–10: 1 Hz per step, 10–20: 2 Hz per step, 20–34: 5 Hz
per step, 34–40: 25 Hz per step
Technical specifications subject to change.
26
Page 27
Eppendorf PiezoXpert™ — Operating manual
9 Ordering Information
9Ordering Information
9.1 PiezoXpert
Order No.
(International)
5194 000.016––
5194 075.105 5194075105
5194 075.202 5194075202
5194 075.300
Order No.
(North America)
5194000024
5194075300
9.2 Accessories
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf
CAUTION!
may impair the safety, function and precision of the device. Eppendorf cannot be held
liable or accept any liability for damage resulting from the use of incorrect or
non-recommended accessories and spare parts or from the improper use of such
equipment.
Only use accessories and original spare parts recommended by Eppendorf.
EN
Operating manual
Description
Eppendorf PiezoXpert
for piezo-assisted micromanipulation
with mains/power plug EU
with mains/power plug USA/Japan
Actuator
for PiezoXpert
Foot control
for PiezoXpert
Spacer Plate
for mounting the PiezoXpert on the TransferMan NK 2, or
PatchMan NP 2
9.2.1Capillaries
Order No.
(International)
5176 210.003920007414
5176 212.006920007708
5176 214.009920007716
5176 207.002920007406
5176 196.000920005870
Order No.
(North America)
Description
Grip head 0
For microcapillaries with an outer diameter of 1.0 mm to
1.1 mm, set of 2
Grip head 1
For microcapillaries with an outer diameter of 1.2 mm to
1.3 mm, set of 2
Grip head 2
For microcapillaries with an outer diameter of 1.4 mm to
1.5 mm, set of 2
Grip head 3
For microcapillaries with an outer diameter of 0.7 mm to
0.9 mm, set of 2
O-ring set
For grip heads 0-3 with removal tool
27
Page 28
Eppendorf PiezoXpert™ — Operating manual
EN
Order No.
(International)
5175 220.005930001091
5175 250.001930001104
Operating manual
5175 108.000930001015
5175 240.0065175240006
VacuTip and Vacutip FCH are registered as medical devices in Europe (Medical Devices
Directive 93/42/EEC).
9.2.2Microinjection and micromanipulator
Order No.
(International)
5176 000.025920002030CellTram Oil
5176 000.033920002111CellTram vario
for mouse ICSI
25° angle, 6 mm rigid flange, sterile, set of 25
Piezo Drill Tip (ES)
for transfer of ES cells into blastocysts
25° angle, 6 mm rigid flange, 15 μm inner diameter,
sterile, set of 25
VacuTip
25 pieces, sterile
VacuTips FCH
25 pieces
Description
TransferMan NK 2
Proportional micromanipulator for microinjection into
suspension cells
with mains/with mains/power plug EU
with mains/power plug USA/Canada
with mains/power plug UK/Hong Kong
with mains/power plug Japan
with mains/power plug Australia
InjectMan NI 2
Micromanipulator for semi-automatic injection in
adherent cells
with mains/with mains/power plug EU
with mains/power plug USA/Canada
with mains/power plug UK/Hong Kong
with mains/power plug Japan
with mains/power plug Australia
28
Page 29
Eppendorf PiezoXpert™ — Operating manual
10 Transport, storage and disposal
10Transport, storage and disposal
10.1 Storage
Air temperatureRel. humidityAtmospheric
In transport packaging –25 °C to 55 °C10 % to 95 %70 kPa to 106 kPa
Without transport
packaging
–5 °C to 45 °C10 % to 95 %70 kPa to 106 kPa
pressure
10.2 Decontamination before shipment
If you are shipping the device to the authorized Technical Service for repairs or to your authorized
dealer for disposal please note the following:
WARNING! Risk to health from contaminated device
1. Follow the instructions in the decontamination certificate. It is available in PDF
WARNING!
format on our homepage (www.eppendorf.com/decontamination)
2. Decontaminate all the parts you want to dispatch.
3. Enclose the fully-completed decontamination certificate for returned goods
(including the serial number of the device) with the dispatch.
.
10.3 Transport
The device may only be transported in its original packaging.
Air temperatureRel. humidityAtmospheric
General transport–25 °C to 60 °C10 % to 95 %30 kPa to 106 kPa
Air freight–40 °C to 55 °C10 % to 95 %30 kPa to 106 kPa
pressure
EN
Operating manual
29
Page 30
Eppendorf PiezoXpert™ — Operating manual
EN
10.4 Disposal
In case the product is to be disposed of, the relevant legal regulations are to be observed.
Information on the disposal of electrical and electronic devices in the European
Community:
Within the European Community, the disposal of electrical devices is regulated by national
regulations based on EU Directive 2002/96/EC pertaining to waste electrical and electronic
equipment (WEEE).
According to these regulations, any devices supplied after August 13, 2005, in the
business-to-business sphere, to which this product is assigned, may no longer be disposed of in
municipal or domestic waste. To document this, they have been marked with the following
identification:
Operating manual
Because disposal regulations may differ from one country to another within the EU, please
contact your supplier if necessary.