Eppendorf In situ Adapter User Manual

In situ Adapter
Bedienungsanleitung
Operating Manual
Mode d'emploi
Bedienungsanleitung. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Operating Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Mode d'emploi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Contents
1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3 Function and operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.1 Delivery package and installation . . . . . . . . . . . . . . . . . . . . . . . . 12
3.2 Technical description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3 Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4 The
4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
in situ
reaction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
9
1 General information
The
in situ
Mastercycler® gradient enables a specially developed carrier system. This allows the Mastercycler® and the Mastercycler® gradient to be used for the conv entional PCR technique in different tube systems as well as for the
The can accommodate up to four glass slides. To ensure optimum temperature control of the samples, the entire carrier system also has a detachable covering hood. This covering hood is transparent and double-walled, which allows access to the sample at all times and also prevents heat loss.
The adapter is extremely easy to handle: The carrier system is positioned onto the thermoblock of the Mastercycler samples are placed on up to four object carriers, fixed into place and sealed hermetically prior to the PCR. Eppendorf ideal sealing system for the (volumes) (see "Ordering information"). The transparent covering hood is then closed and the PCR may be started.
Thanks to the wide selection of different tube types, the software is specially tailored to suit the quality heat transition necessary for optimal results.
Adapter for the Eppendorf Mastercycler® and the Eppendorf
in situ
Adapter is a high-temperature-conductive metal carrier, which
The
in situ
Adapter must be set up and used in accordance with the operating manual provided in order to protect both the user and the device. The instructions in the operating manual of the Mastercycler® or Mastercycler® gradient must also be observed.
When operating the device, please observe the instructions contained in the corresponding applicational regulations!
in situ
PCR reactions to be carried out on
in situ
technique.
®
or Mastercycler® gradient. The
in situ
in situ
PCR and are availab le in diff erent siz es
in situ
Adapter, which guarantees the high-
Frames form the
10
2 Safety precautions
Voltage
Before startup, please ensure that your power supply corresponds to the voltage specifications on the identification plate! Mastercycler or Mastercycler gradient must be plugged into a grounded socket. Before carrying out maintenance or cleaning work or replacing the fuses, switch off the Mastercycler, Mastercycler gradient or from the mains supply.
Explosive substances
Do not use explosive or highly reactive substances in the Mastercycler® or Mastercycler® gradient or in combination with the
in situ
Adapter! Do not operate the Mastercycler®, Mastercycler® gradient or the explosive environment!
Liquids
Do not allow any liquid to enter into the device!
Danger of burns
The thermoblock, the micro test tubes and microtiter plates attain a temperature in excess of 50 °C extremely rapidly.
Close the covering hood of the after starting the PCR. Ensure that the covering hood remains closed until the temperature is 30 °C or lower.
Repairs may be carried out by trained specialists only. Only use original Eppendorf spare parts. Disconnect the mains supply before opening the Mastercycler® or Mastercycler® gradient.
Eppendorf AG reserves the right to make technical modifications to this product.
This operating manual may be reproduced in part or in full only with the prior written permission of Eppendorf AG.
in situ
adapter and disconnect them
in situ
Adapter in a potentially
in situ
Adapter, inside of the heated lid and
in situ
Adapter immediately
11
3 Function and operation
3.1 Delivery package and installation
Delivery package
– Metal carrier –Transparent covering hood (double-walled) –Transparent plate attachment – Operating manual
Together with a transparent plate attachment, the covering hood offers protection during transport. The plate attachment and the covering hood should also be used for storing the adapter.
When an inserted into the 0.5 ml positions of the thermoblock. There is no privileged direction (left/right).
Caution:
Mastercycler® gradient!
in situ
PCR is carried out, the "pegs" on the underside are
Only a
cold
adapter may be attached to the Mastercycler® and
Mastercycler gradient
12
in situ
Adapter
3.2 Technical description
The glass slides are placed onto the four contact surfaces of the metal carrier:
Up to four glass slides can be attached to the carrier
3.3 Programming
The Mastercycler / Mastercycler gradient is programmed in accordance with the operating manual for these devices. The by the Mastercycler and Mastercycler gradient like all other tube types
in situ
PCR,
"in situ"
(i.e. for the
is selected instead of the tube types "0.2 ml", "0.5 ml" and "plate"). Specific settings for the adapter are thus selected.
During such programming, please observe that the temperature of the sample (TUBE) is regulated. The lid heating (corresponds to LID=0°), irrespective of the entry in the LID field.
in situ
Adapter is treated
switches off automatically
and the temperature control of the lid switches off
FILES... Edit UNNAMED Edit CNTRL TUBE Load LID=0° Standard NOWAIT AUTO New 1 T=94.0° 00:02:00 Delete 2 T=94.0° 00:00:15 UNNAMED 3 T=44.0° 00:00:15 – –/22.0° 10:22:06
Temperature regulation is set to TUBE
13
For Mastercycler gradient only: The Gradient function switches off automatically, independent of the entry in the Gradient field:
The gradient is switched off
FILES... Edit UNNAMED Edit 3 T=44.0° 0:00:15 Load 4 T=72.0° 0:00:30 Standard +0.0° +0:01 New R= 3.0°/s +0.1°/s Delete G=0.0° UNNAMED 5 GOTO2 REP24 – –/22.0° 10:22:06
3.4 The in situ reaction
– Place the – Prepare the glass slides (sample, reagents and sealing method) and
place them onto the metal carriers.
– After preparing the glass slides, the transparent covering hood is closed
over this arrangement of thermoblock and metal carrier. The lid of the Mastercycler
– The Mastercycler
manner. The desired program must contain the CNTRL-TUBE command, the tube type
"in situ"
the program should be started by pressing :
in situ
Adapter onto the thermoblock.
or Mastercycler gradient remains open!
or Mastercycler gradient is started up in the usual
in situ
program is then selected. As an
in situ
should be selected in the following menu using the key and
Select the
Main-Menu Run a Program: Start CNTRL TUBE FILES... Tube: OPTIONS... Lid Incubate IN SITU – –/22.0° 10:22:06
14
in situ
in situ
Adapter
4 Maintenance
The
in situ
Adapter does not require regular maintenance.
Contamination can be removed using water or a mild laboratory cleaning agent. The covering hood should be wiped using a moist cloth only in order to prevent water or other liquids from entering the lateral ventilation slits of the double chamber.
Do not allow any parts of the organic solvents or aggressive chemical. Do not clean the transparent plastic parts using alcohol or acetone.
in situ
Adapter to come into contact with
5 Technical data
Type of instrument:
Sample capacity: Max. 4 glass slides of 76 x 26 mm
Temperature range: 4 to 99 °C Temperature distribution 40 °C: ± 0,6 °C
across the adapter: 72 °C: ± 0,6 °C
Temperature-control speed: Heating rate 1,0 °C/s,
Dimensions of the adapter: Metal carrier: Width: 112 mm
Weight: Metal carrier: approx. 60 g Conditions of use: 0 to 105 °C,
Storage conditions: –20 to +70 °C,
in situ
Adapter for Mastercycler
or Mastercycler gradient
(3 x 1 inch)
95 °C: ± 1,0 °C
measured on the adapter Cooling rate 0,7 °C/s, measured on the adapter
Depth: 79 mm Height: 9 mm
Covering hood: Width: 148 mm
25 to 100 % relative humidity
25 to 85 % relative humidity
Depth: 105 mm Height: 31 mm
15
6 Ordering information
Order number Article
5331 220.002 5331 221.904 Operating manual
0030 127.501 0030 127.510 0030 127.528 0030 127.536
in situ
Adapter
German, English, French
For sample preparation
in situ
Frame, 25 µl
in situ
Frame, 65 µl
in situ
Frame, 125 µl
in situ
Frame, 300 µl
16
Eppendorf Offices
ASEAN
Eppendorf AG Regional Office in Malaysia Tel. +60 3 8023 2769 Fax +60 3 8023 3720 E-Mail: eppendorf@eppendorf.com.my Internet: www.eppendorf.com.my
AUSTRALIA / NEW ZEALAND
Eppendorf South Pacific Pty. Ltd. Tel. +61 2 9889 5000 Fax +61 2 9889 5111 E-mail: Info@eppendorf.com.au Internet: www.eppendorf.com.au
AUSTRIA
Eppendorf AG c/o Schott Austria Tel. +43 1 29017560 Fax +43 1 290175620 E-Mail: gilch.p@eppendorf.de Internet: www.eppendorf.com
BRAZIL
Eppendorf do Brasil Ltda. Tel. +55 11 30 95 93 44 Fax +55 11 30 95 93 40 E-Mail: eppendorf@eppendorf.com.br Internet: www.eppendorf.com.br
CANADA
Brinkmann Instruments, Ltd. Tel. +1 905 826 5525 Fax +1 905 826 5424 E-Mail: canada@brinkmann.com Internet: www.brinkmann.com
CHINA
Eppendorf AG Tel. +86 21 68760880 Fax +86 21 50815371 E-Mail: market.info@eppendorf.cn Internet: www.eppendorf.cn
FRANCE
EPPENDORF FRANCE S.A.R.L. Tel. +33 1 30 15 67 40 Fax +33 1 30 15 67 45 E-Mail: eppendorf@eppendorf.fr Internet: www.eppendorf.fr
GERMANY
Eppendorf Vertrieb Deutschland GmbH Tel. +49 02232 418-0 Fax +49 02232 418-155 E-Mail: vertrieb@eppendorf.de Internet: www.eppendorf.de
INDIA
Eppendorf India Limited Tel. +91 44 52111314 Fax +91 44 52187405 E-Mail: info@eppendorf.co.in Internet: www.eppendorf.co.in
ITALY
Eppendorf s.r.l. Tel. +390 2 55 404 1 Fax +390 2 58 013 438 E-Mail: eppendorf@eppendorf.it Internet: www.eppendorf.it
Eppendorf Offices
JAPAN
Eppendorf Japan Co. Ltd. Tel. +81 3 5825 2363 Fax +81 3 5825 2365 E-Mail: info@eppendorf.jp Internet: www.eppendorf.jp
NORDIC
Eppendorf Nordic Aps Tel. +45 70 22 2970 Fax +45 45 76 7370 E-Mail: nordic@eppendorf.dk Internet: www.eppendorf.dk
SPAIN
Eppendorf Ibérica S.L. Tel. +34 91 651 76 94 Fax +34 91 651 81 44 E-Mail: iberica@eppendorf.de Internet: www.eppendorf.es
SWITZERLAND
Vaudaux-Eppendorf AG Tel. +41 61 482 1414 Fax +41 61 482 1419 E-Mail: vaudaux@vaudaux.ch Internet: www.eppendorf.com
UNITED KINGDOM
Eppendorf UK Limited Tel. +44 1223 200 440 Fax +44 1223 200 441 E-Mail: sales@eppendorf.co.uk Internet: www.eppendorf.co.uk
USA
Eppendorf North America Tel. +1 516 334 7500 Fax +1 516 334 7506 E-Mail: info@eppendorf.com Internet: www.eppendorfna.com
OTHER COUNTRIES
see: www.eppendorf.com/worldwide
is a registered trademark
In touch with life
eppendorf
Your local distributor: www.eppendorf.com/worldwide
Eppendorf AG · 22331 Hamburg · Germany · Tel. +49 40 538 01-0 · Fax +49 40 538 01-556
E-Mail: eppendorf@eppendorf.com
Eppendorf North America, Inc. · One Cantiague Road, P.O. Box 1019
Westbury, N.Y. 11590-0207 USA · Tel. +1 516 334 7500 · Toll free phone +1 800 645 3050
Fax +1 516 334 7506 · E-Mail: info@eppendorf.com
5331 221.904-02/0705 Printed in Germany
Europe, International: Tel. +49 1803 666 789 · E-Mail: support@eppendorf.com
Application Support
North America: Tel. +1 800 645 3050 ext. 2258 · E-Mail: support_NA@eppendorf.com
Asia, Pacific: Tel. +603 8023 2769 · E-Mail: support_Asia@eppendorf.com