The Eppendorf InjectMan
for use in microinjection.
Its optimum functionality and user-friendliness make the InjectMan® NI 2 the ideal solution for
easy injection. This is illustrated particularly by the large number of injected cells.
The Eppendorf InjectMan® NI 2 micromanipulator combines the classic benefits of a
mechanical system (direct feeling for tool movements) with the advantages of a precise
motorised system.
The direct and sturdy adapter between the module unit and the microscope tripod (optionally
positioned at the left or right) and the separate control board prevent oscillation and vibration.
Connection to the FemtoJet®, FemtoJet® express or Transjector 5246 also permits
semi-automated injections. The injection motion and pressure are activated simultaneously at
the press of a button.
The highly resource-conserving way in which the microcapillaries are injected axially at any
desired angle into the cell, and also extracted axially, considerably increases the effectiveness
and broadens the range of application of microinjection.
No manual wobble of the capillary can damage the cell. The microelectronics and the software
help to conserve the cells to the maximum extent.
The clearly laid out keypad with the additional display also help to avoid operator errors and so
protect the preparation and the tool.
The unit can be attached to any commonly used inverted microscopes by way of an appropriate
adapter. The capillary can be aligned at any angle and in any direction.
Defined working planes and limits largely eliminate the possibility of damage to the capillaries.
An integral CLEAN/HOME function greatly eases the handling and changing of capillaries.
Simple, self-explanatory user control also permits inexperienced users to begin performing
injections into adherent cells quickly and successfully. Additional functions are available for the
more experienced user.
®
NI 2 micromanipulator is specially designed as a micromanipulator
58
2 Intended use and safety precautions
The InjectMan NI 2 enables an electronically controlled axial injection movement. It is designed
for controlled, gentle, semi-automatic microinjection in (adherent) cells and small organisms. By
coupling with an external pressure source/external pressure sensor (e.g., FemtoJet (express) or
Transjector 5246), the injection movement and injection pressure are monitored, and simultaneously activated at the push of a button. Microelectronics and software control the cell-preserving execution of the microinjection process. The cell wall is minimally perforated, thus
avoiding cell lysis (via the potential introduction of excess volumes of liquid). In the optimal scenario, inspecting the work area and injection limits reduces damage to the capillaries.
The InjectMan NI 2 is only intended for use in research labs. It may only be operated by qualified
lab personnel.
It is essential to observe the following safety precautions!
•
The device must be commissioned and subsequently used in compliance with the instructions set out in this Manual, in order to protect the user and prevent damage to the device!
•
The device may only be used for the purpose for which it is intended!
•
The device is to be used solely for micromanipulation in accordance with the appropriate
legislative requirements!
•
Do not operate the device if there is visible damage to the device itself and/or to its
packaging.
•
Before starting up the device, please compare your power supply with the voltage
requirements on the identification plate!
•
Take care when transporting the control board!
Do
not
transport the device by carry the joystick!
•
When installing and dismantling, make sure the weight of the motor module (1.4 kg) can be
safely carried! Otherwise injury may be caused, or the device may be damaged!
•
A weight of max. 100 g may be applied to the motor module.
•
The motor modules must not be moved against a mechanical obstacle, as this may damage
the drive units.
•
Before transporting the micromanipulator, dismantle the module unit and, if possible,
transport it in its original box. The delivered transport guards are to be mounted according
to the red supplement sheet.
•
•
•
Important!
The mains plug must not be plugged into the socket outlet until all connections
between the control console, power supply unit and motor module have been
made.
Caution – risk of trapping/crushing!
Do not reach into areas of the motor modules in which your hand may become
trapped (the gap between the guide and the module or between the module
and any obstacles), and do not insert any other objects into such gaps!
Caution when mounting tools!
Mounting must not be performed with pressure. Glass capillaries can be
damaged during mounting if they are not handled carefully. Wear eye
protection to prevent the risk of injury from glass splinters!
2
Intended use and safety precautions
59
2 Intended use and safety precautions
•
•
2
•
Electrical connections to devices not mentioned in this Manual are permitted only with the
•
consent of the manufacturer!
•
Repairs must be carried out only by authorised Service personnel!
The Manual, or any part of it, may be reproduced only with the written consent of
•
Eppendorf AG.
•
Liability for the functioning of the device shall in all cases be transferred to the owners or
operators if the device is serviced or repaired by persons not authorised by the Eppendorf
approved Service or if any action is performed which is not in conformance with the
designated use of the device. Eppendorf accepts no liability for damage arising from failure
to comply with the instructions set out in the Manual. Guarantee and conditions of liability of
Intended use and safety precautions
the terms of sale and delivery of Eppendorf AG are not expanded by the above notes.
Eppendorf AG reserves the right to make technical modifications to this device!
•
Transfer
•
If the device is passed on to someone else, please include the instruction manual.
•
Disposal
In case the product is to be disposed of, the relevant legal regulations are to be observed.
Information on the disposal of electrical and electronic devices in the European
•
Community
The disposal of electrical devices is regulated within the European Community by national
regulations based on EU Directive 2002/96/EC on waste electrical and electronic equipment
(WEEE).
According to these regulations, any devices supplied after 13.08.05 in the business-tobusiness sphere, to which this product is assigned, may no longer be disposed of in
municipal or domestic waste. They are marked with the following symbol to indicate this.
Caution when mounting tools!
When the tool is mounted, the motors are swung out forwards and the
capillaries face the front. Handle with care!
Caution!
If the work is not carried out properly, for example if glass capillaries contact
a surface, they may smash and flying splinters may endanger eyes. You should
therefore always protect your eyes, such as by wearing goggles!
Never point capillaries at persons!
Capillaries can suddenly be loosened from their capillary grips under high
pressure and become a missile.
As disposal regulations within the EU may vary from country to country, please
contact your supplier if necessary.
60
2 Intended use and safety precautions
2.1 Hazards when unpacking and other information and instructions
Before checking that the device has been delivered complete and installing it, please read the
following section thoroughly.
•
The control board must not be carried or dragged by the joystick!
The joystick is a highly sensitive control element which transmits movement commands in
the 0.1 µm range! Always carry the unit by the housing only!
•
Never place the control board upside down on the joystick
•
Treat the motor modules with care!
•
Never transport the motor modules while they are still assembled!
•
Whenever possible, transport the motor modules in their original packaging!
•
Do not tilt the assembled motor modules!
•
After the warm-up phase (at least half an hour) the device is virtually drift-free.
•
After switching on, wait for the initialisation process to complete before operating the joystick. Completion of initialisation is indicated by an acoustic signal or by the LED returning
to its standby condition.
•
Enter an injection time >0.0 for the Eppendorf injection system being used, otherwise an
error message will be delivered
2
Intended use and safety precautions
61
The Eppendorf InjectMan
–Control board
3
–Module unit
–Mains power supply unit
Description
3 Description
3.1 Components
10
5
5
10
®
NI 2 micromanipulator is made up of three components:
Coarse
Fine
Pos 1
Menu
I
n
j
e
c
t
M
a
n
N
I
2
Limit
Home
Axial
Pos 2
Standby
Fig. 1: InjectMan® NI 2 micromanipulator, without mains power supply unit
Module unit for right side mounting (without tool mount)
3.2 Module unit
The module unit can be mounted at the left or right on any commonly used microscopes by way
of an adapter.
The module unit comprises the motor unit for horizontal movement (X, Y axis,
X/Y module) and the axial motor unit (A axis). The injection angle is adjusted manually using the
angle guide. By means of the angle setting and appropriate mounting, injection angles between
60° and 30° are possible.
Vertical movements (Z movements) are executed by the interaction between the A and X axes.
62
3 Description
On the A module is the tool mount for the probe holders. Tools with a diameter, or width, of
4 mm and weighing max. 100 g can be inserted, e.g. capillary holders for microinjection.
10
5
1
2
3
Description
5
10
3
Fig. 2: Module unit (here right side mounted)
1 Motor unit for horizontal movements (X/Y module)
2 Axial motor unit (A module, without tool mount)
3 A module arc guide
3.3 Control board
The control board accommodates the keypad, the display and the joystick with integral button,
as well as the turn wheel on the side. The movements performed on the joystick are transmitted
to the module unit. The movement modes, directions and speeds, as well as movement
sequences, are pre-set and triggered by way of the keypad, the turn wheel and the joystick.
Underside
The bay underneath the control board
accommodates the connector sockets for
the module unit and the mains power supply
as well as the serial port and the connection
for the FemtoJet
®
and an optional foot
switch.
Note
: Do not place the control board down
resting on the joystick! When connecting
cables (section 4.4), place the control board
on its side on an even surface and secure it
against toppling over!
63
3 Description
3.3.1 Tool movement
Moving the dynamic joysticks moves the capillaries in one plane.
The speed of the motion is determined by the speed of the
deflection movement on the joystick. Releasing the joystick stops
3
the movement, the joystick springs back to its neutral position.
Description
3.3.2 Keypad
The keypad can be used to make settings or initiate movements directly. Frequently used
settings can be entered by means of key combinations. The LEDs indicate the active condition
of a function. The display shows the current settings and assists user control with visual aids.
Keys
To understand the key functions, users must have read the detailed Manual.
Some of the key functions are triggered when the keys are released.
Coarse/Fine
Speed
selected mode is indicated by the LED associated with the key and by the
selected speed on the on-screen display.
Home
Move out of working radius and return to the precise same spot.
Rotating the joystick causes the tool to move up and down.
Clockwise rotation: down
Counterclockwise rotation: up
Overlaid movements in all three coordinate directions are
possible.
selector key: Toggle between
FINE
and
COARSE
manipulation. The
64
3 Description
Axial
Activate or disable axial movement.
Pos 1 / Pos 2
Approach a stored position.
Store current position by pressing and holding down the key.
Clear approached position by pressing and holding down the key.
Limit
Set/cancel limits for injection movement.
Arrow keys
Adjusting the limit.
3
Description
Menu
Toggle between menu and back. Parameters can be selected or changed in
the menu using the joystick.
Standby
Switches the device off or on.
The function is used as a reset function.
Joystick button
Initiate the Inject function with a set limit.
65
3
3.3.3 Display
Description
3 Description
fineX: –1200µm
Y:300µm
Z: –21323µm
Turn wheel
Set the speed in the current speed mode.
Display of selected speed on-screen.
In normal operation the display shows the coordinates
of the three axes (X, Y, Z; Cartesian) and the pre-set
mode with the selected speed at maximum joystick
deflection. Additional functions are displayed in
theactive condition.
Parameter settings can be made and changed in the onscreen menu using the joystick.
coarseX: –1200µm
pos 1 pos2
StepInj user 3
–21323 z-limit+search
Defining the operating parameters and functions, and how to program and change them, is
described in section 5.
Y:300µm
Z: –21323µm
axial
▲
▲
Display with all display options.
66
4 Installation
4.1 Unpacking
–The packaging and the transport guards should be kept for use if the device is transported
again later!
–The control board must not be carried or dragged by the joystick!
–The joystick is a highly sensitive control element which transmits movement commands in
the 0.1 µm range. Always carry the unit by the housing only!
–Never place the control board upside down on the joystick!
–Treat the motor modules with care!
–Only remove the motor modules from their packaging together! Never tug on the module
connecting cable!
–Never transport the motor modules while they are still assembled!
–Whenever possible, transport the motor modules with the transport guards in their original
packaging!
–Do not tilt the assembled motor modules!
The packaging comprises two levels. The first level contains the accessories. The bottom level
contains the module unit and the control board.
Check that the standard accessories are present and complete.
The standard accessories are:
1 Control board
1 Motor module
1 Mains power supply unit
1 Mains power cable
1 Hinged bracket
1 Set of Allen keys (metric)
2 Screws M5x16
1 FemtoJet
2 Positioning aids for universal capillary holders
1 Operating manual
1 Set of spare screws
(2 x M5x16, 2 x M4x10, 2 x M3x16, 2 x M3x12)
®
interface cable (5247)
4
Installation
Carefully lift the control board and the motor modules out of the packaging using both hands.
1. Loosen the
screws and
remove the
transport guard
(yellow).
2. Remove the colored
transport guard
only after comple-tely mounting the
device on the
microscope!
67
10
10
5
5
4 Installation
4.2 Assembling the module unit
4.2.1 Components of the module unit
4
10
5
Installation
5
10
Fig. 3: Module unit, assembled (here right side mounted)
1 Motor unit for horizontal movements (X/Y module)
2 Axial motor unit (A module, without tool mount)
3 A module arc guide
4.2.2 Angle guide
The A module is attached by bolting on the angle guide.
Depending on the mounting point of the angle guide, injection angles between 60° and 30° can
be set.
For the sake of clarity, the A module is not shown.
10
Example:
The angle guide is attached to the X/Y module in the
centre hole, corresponding to 45°. The A module can
be adjusted to between 55° and 35° (45°±10°) by
turning the spindle.
10
1
2
3
68
5
10
10
5
5
4 Installation
The hole spacing is 5°. By moving the mounting point
10
10
5
5
10
5
10
4.2.3 Mounting the axial module (A module)
10
10
of the angle guide 5° upwards, the adjustable angle
changes each time to 50° to 30°.
By moving the mounting point downwards, angles of
up to 60° can be set. This requires sufficient space to
be available on the microscope.
The angle guide is already bolted to the A module.
The spindle may need to be turned until the desired
mounting point (arrow, e.g. here centre hole) is
accessible.
Note
face of the A module must be slackened prior to
adjusting the angle and must then be re-tightened in
order to ensure reliable operation.
4
Installation
: The orange fixing screw on the narrow front
Set the angle guide with the
free hole (2) at the mount (1) of
the X/Y unit. The fitting pins
must be fitted into the holes
1
10
5
5
10
provided. Fix the angle guide
to the X/Y module by the
M4x10 cheese head screw.
2
69
4.2.4 Angle adjuster
4
Installation
4 Installation
10
5
5
10
45°
40°
35°
Set the desired angle.
Mounting the angle guide at the centre hole and
adjusting the A module to 0° corresponds to an angle
of 45°.
To perform the adjustment, the lines on the angle
guide are moved to match the outer edge of the A
module.
Example – 35°: The angle guide is attached by the
centre hole, the bottom edge of the A module is set to
the lower 10° marking.
Note
: The mechanically set angle must be entered
subsequently in the Install menu! Otherwise Z axis
movements cannot be executed correctly! See
section 5.4.2.6.
Note
: The orange fixing screw on the narrow face of
the A module must be slackened prior to adjusting the
angle and must then be re-tightened in order to
ensure reliable operation.
4.2.5 Assembling the hinged bracket
70
Set up the hinged bracket according to the side on
1
which the module unit is to be mounted.
Example of left side mounting:
The hinge (2) is positioned at the left outer, aligned
upwards and to the rear.
2
4 Installation
Opening the hinged bracket
0° – 45°
1
0° – 45°
For right side mounting, the hinge must be at the right
outer (not visible here, on the rear).
The module unit is mounted from above onto the
hinged bracket (1).
4
Installation
To open the hinged bracket beyond its normal
opening angle between 0° and 45° for mounting
purposes, push the pin on the hinge side up. The
slider is purposely hard to move. The unit is usually
only mounted once. The pin can be moved with the 4
mm Allen key.
45° – 90°
The hinged bracket can then be opened to 90°. The
mounting holes on the module are then accessible.
90°
71
4
Installation
4 Installation
The hinged bracket is opened up and joined to
the motor module by its inner inclined surface by
means of the M5x16 screw. Close the hinged
bracket.
Note
: The A module is not shown.
The module unit for left side mounting is mounted
on the hinged bracket.
72
Right side mounting
The hinge is not visible, the mounting point of the
A module is facing the middle of the microscope.
4 Installation
4.3 Mounting the module on the microscope
4.3.1 Right and left side mounting
The module unit mounting is dependent on the
chosen side of the microscope.
The module unit can be mounted on the most
commonly used microscopes (usually inverted
microscopes). The unit is mounted on the microscope
tripod by way of a separately ordered microscopespecific adapter. Mounting of the adapter is described
in the relevant mounting instructions. If a different
type of mounting is required, special adapters will be
needed.
The hinged bracket is mounted on the adapter by one
of the two shorter rollers as shown in the illustration
(here right side mounting; for the sake of clarity the
mounted A module is not shown).
The M5x16 screw is supplied as part of the
InjectMan® NI 2 accessory set; the roller is a
component of the microscope-specific adapter.
The rollers are adapter-specific.
The module unit (here on the right) is joined
to the adapter by way of the roller with two
M5x16 screws. The InjectMan® NI 2 is
mounted at the front position.
The screws are component parts of the
microscope-specific adapter.
Note
: In the "Install" menu (section 5.4,
subsection 5.4.2.6) the parameters
Side = RIGHT
Direction = FLAT
Angle = e.g. 45°
are set in the given sequence according to the
mode of mounting.
4
Installation
73
4
Installation
4.3.2 Assembling the capillary holder
4 Installation
According to the mounting mode on the left
side of the adapter.
Note
: In the "Install" menu (section 5.4,
subsection 5.4.2.6) the parameters
Side = LEFT
Direction = FLAT
Angle = e.g. 45°
are set in the given sequence according to the
mode of mounting.
On the A module is the A head (1) for tool
mounting.
It is attached to the A module by two M3x16
5
10
screws as shown in the illustration.
The A head can accommodate tools with a
1
diameter of 4 mm, such as the universal
capillary holder (accessory FemtoJet
®
or
CellTram).
74
10
5
The universal capillary holder is inserted into
the A head after the knurled screw (arrow) has
been slackened, and then fixed in place by retightening the knurled screw.
5
10
4 Installation
To avoid having to reposition every time a capillary is changed, the
positioning aid can be mounted on the universal capillary holder so it fits
in the centre hole in the A head.
4.4 Connecting cables – Startup
Note
: The connections between the control board, the mains power supply unit and the motor
module must only be plugged and unplugged with the device switched off.
Note
: Do not place the control board down resting on the joystick! When connecting cables,
place the control board on its side on an even surface and secure it against toppling over!
–The connection to the motor module/injector (e.g. FemtoJet
corresponding connectors, as shown in Figure 4.
–Check the voltage and frequency against the rating plate before connecting the mains
power!
–Use only the mains power supply unit supplied by Eppendorf!
–The cable is plugged into the mains power supply unit before being connected to the mains.
Important
–Check that the connectors are firmly attached.
: The main power connection is the last one to be connected.
®
) is made via the
4
Installation
Fig. 4: Underside of control board
1 Connection for optional foot switch (SER2; SubD9, male)
2 Connection for injector (FemtoJet
3 Connection for mains power supply unit
(
Important
4 Connection for motor module (SubD25, female)
5 Serial port (SER0; SubD9, male) for computer (RS 232)
: Connect only after the other components)
®
, Transjector 5246) (SER1; SubD9, male)
1
2
3
4
5
75
4.4.1 Transmission cable 5181/5246
For optimum use of the INJECT function (see section 5.1), the InjectMan® NI 2
micromanipulator and the microinjector (e.g. Eppendorf FemtoJet® or Transjector 5246) must
be interconnected.
4
Making the cable connection
–Switch off the device.
–Plug the FemtoJet® interface cable (accessory) or interface cable 5246/5242 (optional) into
–Switch on the device.
Installation
Note
micromanipulator with the Eppendorf microinjector 5242.
4 Installation
socket 2.
The display shows "Injector ready" if the connected injector is ready for operation.
: Interface cable 5246 can be used to combine the Eppendorf InjectMan® NI 2
76
5 User control and mode of operation
Section 5.1 provides a brief summary of how the commands can be used for the injection
procedure. Section 5.2 provides additional information on the device's mode of operation.
Sections 5.3 and 5.4 set out detailed descriptions of all the commands.
5.1 First injection
Precondition
Prior to performing the first injection, it is assumed that there is a filled capillary in the correctly
positioned capillary holder (see sections 5.2.1 and 4.3.2).
Other requirements: The device is properly initialised (pre-set angle, mounting side defined,
etc.) and an Eppendorf injector (e.g. FemtoJet
The InjectMan® NI 2 micromanipulator is switched on
and in the following "start mode": The capillary is at the
top position and to the side of the sample. The LED of
the "Speed" key is lit in FINE mode. The display shows
only the speed setting for FINE, the coordinates for the
X, Y and Z axes and "Injector ready".
On the microscope table is a focussed sample, e.g.
adherent cells in a petri dish, coated with medium.
Movement towards sample
Toggle "Speed" key to COARSE mode.
Coarse
Fine
The capillary is moved to the surface of the medium
by pushing/rotating the joystick in the direction of the
sample.
Visual monitoring of the movement.
The speed can be
increased by turning the
wheel on the right hand
side of the control board.
The change of speed is
indicated on the display.
®
) is connected.
fineX:0µm
Injector ready
coarseX: –1200µm
Injector ready
Y:0µm
Z:0µm
Y:300µm
Z: –21323µm
COARSE
7500
µm/s
5
User control and mode of operation
77
5 User control and mode of operation
The new speed is shown on the display as a modified
block diagram.
5
Coarse
Fine
right brings the tip of the capillary into the focal plane (microscopic control). Adjust the speed as
necessary by way of the turn wheel.
The focal plane is adjusted carefully to the cells. The capillary tip is brought into the focal plane.
The set limit is indicated on the bottom line of the display
directly above the "Limit" key. The number before the
User control and mode of operation
limit indicator is the level of the limit. The value is stored.
The capillary cannot be moved any lower.
Move the capillary up slightly by rotating the joystick to the left. The height to which the capillary
moves is termed the "search plane".
By moving the joystick the capillary tip is moved over a cell in the search plane.
injection time has elapsed the capillary is returned to the search plane.
This procedure automatically injects the cell with the pre-set parameters.
If the capillary touches the medium surface for example (visually monitored),
switch to FINE mode.
The capillary is moved into the beam path of the microscope by changing the
plane of the joystick. Carefully lowering the capillary by rotating the joystick to the
Press the "Limit" key.
The Inject function cannot be executed
without a defined limit.
Press the joystick button initiates the
automatic injection movement. The
capillary is moved to the side and then
axially to the limit plane.
In parallel with the injection movement,
the connected microinjector is
activated and the injection pressure is
built up. When the microinjector's
coarseX: –3200µm
Injector ready
fineX: –3200µm
Injector ready
–21455 z-limit
INJECT
INJECT-Position
Y:300µm
Z: –21323µm
Y:300µm
Z: –21455µm
–3200
300
–21455
Moving to
78
5 User control and mode of operation
Changing the limit
As a result of moving the petri dish, the limit may need to be adjusted. This is done using the
arrow keys to the left and right of the "Limit" key.
Each key press changes the value of the limit upwards or downwards. The new value of the limit
is checked by a test injection. Alternatively, the stored limit can be cleared by pressing the
"Limit" key again and then be redefined.
Home
To change capillaries, the Home function
can be used. Move the capillary up slightly
by rotating the joystick to the left. Press
and hold down the "Home" key. Swivel out
the motor module and change the capillary. Swivel the
motor module back into place and press the "Home"
key. The capillary is returned to its home position.
If necessary, the capillary can also be moved manually to
the sample by moving the joystick in the desired direction.
Standby
When the operation is complete, the InjectMan® NI 2
micromanipulator can be switched off by the “Standby“
key. The display illumination is switched off and Standby
is displayed.
HOME
To move back use HOME
or joystick
STANDBY
–3200
300
–21455
5
User control and mode of operation
5.2 More information on the mode of operation
5.2.1 Setting up the capillary
–Mount the capillary holder (without capillary) at as high a position as possible on the
A module. Use the positioning aid if necessary.
–From the menu select the CentrMot function (see sections 5.4 and 5.4.7). Make sure there
are no obstacles blocking the field of movement!
–Run the CentrMot function. The motor unit moves towards the microscope lens.
–Swivel out module (see section 5.2.4)
–Attach the capillary to the universal capillary holder without dismantling the holder.
–Swivel motor back into place.
79
5 User control and mode of operation
–Carefully slacken knurled screw of tool mount and positioning aid a little way. Carefully bring
the capillary holder with the capillary into the focal plane by hand. If necessary, bring the
capillary into the field of view by means of horizontal movements of the joystick (X and Y
axis).
5
–When the capillary is in the field of view, bring it carefully into focus (e.g. at a 4 times zoom
factor).
–Tighten knurled screw and positioning aid.
This procedure adjusts the capillary so that approximately 80 % of the field of movement is then
above the current position. This setting can be used for all capillaries of the same length. By
means of the positioning aid (section 4.2.3) the universal capillary holder can be dismantled and
refitted in the tool mount as necessary. This does not affect the alignment.
Other setups can also be used specific to application, without any problems.
5.2.2 Definition of Z limit
To protect the tool and the preparation against damage, and for the automatic Inject function, a
limit must be defined. This is relevant mainly to the application of microinjection into adherent
cells.
Once the limit is defined, the capillary is moved upwards manually above the cells. This plane,
from which the various cells are injected, is termed the search plane.
User control and mode of operation
Microcapillary
Cell
Search plane
Z limit
Glass slide
Fig. 5: Capillary above cell on search plane, Z limit set.
To define the limit, the capillary is set to the desired
position by lightly pressing in the cell.
–Press the "Limit" key.
80
Microcapillary
Cell
Focal plane
Glass slide
5 User control and mode of operation
Or the capillary is brought into the focal plane like the
cell.
–Press the "Limit" key.
Cell
Focal plane
Microcapillary
5
Glass slide
The defined limit is indicated on the display.
To clear the limit, press the "Limit" key again.
To set a limit at a lower position, the limit must first be
cleared. Only then can the capillary be moved lower and
a new limit be defined.
To adapt to uneven surfaces (primarily
when working with plastic petri dishes) or
to differing cell morphology, the limit can
be adjusted upwards or downwards using
the arrow keys. Each key press changes
the limit by approximately 0.2 µm.
Press and hold down an arrow key to change the limit
continuously. Each 1 µm change is signalled
acoustically.
To check the success of the change, an injection can be triggered directly.
In the special z-limit+search application,
the limit changes made also affect the
search plane. The capillary changes its
Z position by the same distance as the
limit is changed.
fineX: –3200µm
Injector ready
–21444 z-limit
fineX: –3200µm
Injector ready
–21442 z-limit
coarseX: –1200µm
Injector ready
–21323 z-limit+search
Y:300µm
Z: –21455µm
Y:300µm
Z: –21455µm
Y:300µm
Z: –21346µm
User control and mode of operation
81
5 User control and mode of operation
5.2.3 Injection
Pressing the joystick button activates the Inject function.
The capillary moves automatically from its current position, the search plane, to the side and
5
then axially to the limit plane. When the injection is complete the capillary returns by the same
route to its starting position. The injection process is shown on the display.
Injection with pre-set Synchronisation Immediate (factory default) is shown schematically in
Fig. 6.
12
microcapillary
cell
Z-Limit
slide
34
injection
User control and mode of operation
search level
Fig. 6: Inject function
1 Starting position, capillary in search plane
2 The manipulator performs a backswing movement
3 The capillary injects in axial direction into the cell as far as the Z limit
4 When the injection is complete the capillary returns to the search plane
Injection options with a connected microinjector (e.g. FemtoJet®; Automatic injection mode,
ready; Readout on display of InjectMan® NI 2: Injector ready):
–With Synchronisation Immediate the connected injector in Fig. 6.1 begins to build up
pressure. The states as shown in Fig. 6.3 is maintained until the injector's injection time has
elapsed.
–Synchronisation Limit: The pressure build-up only begins when the limit is reached
(Fig. 6.3). When the injection time elapses the capillary is moved back (Fig. 6.4).
–Synchronisation Pressure: There is no injection movement, the pressure build-up of the
connected microinjector is started immediately. The injection is made at the location of the
capillary. A manual injection movement can be performed beforehand.
82
5 User control and mode of operation
–Synchronisation Move: No pressure build-up occurs. If the joystick button is held down the
capillary stops as shown in Fig. 6.3 and a separate injection can be triggered with a
(manual) injector.
If the injection is triggered via the connected injector, the InjectMan
warning. With this synchronisation the movement can be triggered only by way of the
InjectMan® NI 2.
®
NI 2 delivers a
5
5.2.4 Capillary exchange
Before swivelling it out, make sure the capillary cannot be impeded by any obstacles!
Wear goggles to protect against possible glass splinters! Swivel out module unit on
X/Y module.
Swivel module unit for capillary exchange out of working radius. The capillary swivels out
forwards and upwards with the module unit.
–Before changing the capillary, bring the capillary to a position over the petri dish either
manually or by way of the Home function.
–Grip module unit at rear motor unit (X/Y module) and swivel out.
–Do not use the A module!
–Exchange capillaries and swivel module unit back into place.
–Bring the new capillary back into the focal plane.
10
5
5
10
Fig. 7: Module unit in working position and swivelled out (here without capillary holder)
User control and mode of operation
83
Z
5 User control and mode of operation
5.2.5 Field of movement
To utilise the benefits of axial injection to the full, only one motor (A module) is used. This
permits less harmful injection, with minimal damage to the cell.
5
As a result, vertical movements are set with the A module and the X axis together. To perform
a downward movement, the X axis is moved outwards and the A module is moved forwards,
simultaneously. This results in the downward Z movement.
This has effects on user control of the micromanipulator in the boundary zones of the
X axis/Z axis.
In the case of a three-dimensional manipulator with a motor standing for each coordinate, the
field of movement forms a cube with equal sides.
As a result of the arrangement of the motor axes of the InjectMan® NI 2, this cube is tilted in
X direction by the mounting angle of the A module. Consequently, it is not possible to execute
purely Z movements at the sharp angled corners of the cube. Here axial movements are
necessary, executed by simultaneously moving the joystick back and forth and rotating it.
The actual field of movement (provided the capillary is mounted properly) relevant to the
experiment is much smaller than the tilted cube. Within this working area, conventional working
in the three coordinates is possible.
User control and mode of operation
Y
1
2
3
X
Fig. 8: a) Cartesian field of movementb) Field of movement with axial component
1 Capillary
2 Field of movement in experiment
3 Height of capillary
4 Position of capillary after start of
CentrMot function (section 5.4.2.7)
84
4
5 User control and mode of operation
5.3 Use of the keypad
The keypad can be used to make settings or initiate movements directly. Frequently used
settings can be entered by means of key combinations. The LEDs indicate the active condition
of a function. The display shows the current settings and assists user control with visual aids.
Some of the key functions are triggered when the keys are released.
5.3.1 Coarse/Fine
Selector key
Toggle between
manipulation.
The selected mode is indicated by the LED
associated with the key and by the
selected speed on the on-screen display
().
Speed
FINE
selector key:
and
COARSE
fineX: –1200µm
coarseX: –1200µm
Y:300µm
Z: –21323µm
Y:300µm
Z: –21323µm
5
User control and mode of operation
85
5 User control and mode of operation
5.3.2 Setting the speed
The speed can be set in three ways:
Directly using the turn wheel, by pressing and holding
down the key, or by way of the menu.
5
Turn wheel
Press and hold down the Speed key
Pressing and holding down the "Speed" key in the
selected speed mode (Coarse/Fine) activates the setting,
which is then also indicated on the display. The speed is
User control and mode of operation
reduced in stages.
Each stage is also indicated acoustically. If you switch
directly from the minimum to the maximum radius a long
acoustic signal sounds. There are five different levels,
each roughly halving the speed.
FINE
1000 µm/s
The turn wheel can be used to make speed settings directly.
Turn wheel backwards: increase speed to maximum possible
Turn wheel forwards:reduce speed to minimum possible
The setting is made in stages. Finer settings can be made by
way of the menu.
The turn wheel can also be used to navigate round the menu.
FINE
250 µm/s
86
5 User control and mode of operation
5.3.3 Home
The tools moves axially out of the working radius.
The function is intended for changing capillaries, or for quick removal of capillaries
out of the working radius.
HOME: Press (and hold down as necessary) to move out
as far as motors' end position.
The red LED flashes during the movement, and is then lit
steadily when the end position is reached.
Press the key again to move the tool back to its starting
point. To return the tool manually the joystick can be
moved directly. This terminates the Home function.
Clean moves the capillary out of the medium. In the
process, cell particles adhering to the medium boundary
layer are removed from the capillary.
When Clean is active:
CLEAN: Tap to move out a pre-set distance.
The red LED flashes during the movement and at the end
point.
Press the key again to move the tool back precisely to its
starting position, or move the tool manually.
If the parameter BackMan in the "Home" menu is set to
"off", it is then not possible to terminate the Home or
Clean function by moving the joystick
(see figure under notice text).
Interruption of the function is possible only by a
Reset ("Standby" button).
–3200
300
HOME
To move back use HOME
or joystick
CLEAN
To move back use HOME
or joystick
Home
To move back use HOME
–21455
–3200
300
–21455
–3200
300
–21455
5
User control and mode of operation
5.3.4 Axial
Activate or disable axial movement
Rotating the joystick causes the tool to
move axially up or down. The active
function is shown by the lit LED and is
indicated on the display.
Clockwise rotation: Movement
diagonally downwards
Counterclockwise rotation: Movement
diagonally upwards
When the Z limit is reached the only movement is in X direction.
coarseX: –1200µm
Y:300µm
Z: –21323µm
axial
87
5 User control and mode of operation
5.3.5 Pos 1 / Pos 2
Approach
Store
5
The stored position is shown by the LED and indicated on the display. Reaching of the position
is also indicated by the LED and on the display (here position 2, for example).
The current position of the tool is stored in the three spatial coordinates. Storing new positions
overwrites existing ones. Positions can also be cleared by performing a reset ("Standby" key) or
via the "Position" menu.
5.3.6 Limit
User control and mode of operation
Long key press: see section 5.3.11
holding down the key (acoustic
acknowledgement).
Clear
holding down the key again (acoustic
acknowledgement).
Set/cancel limits for injection movement.
An indicator on the bottom line of the
display above the Limit key indicates
whether the limit is set. To the left of it is
the stored value for the limit.
a stored position.
current position by pressing and
current position by pressing and
coarseX: –1200µm
pos 1pos 2
coarseX: –1200µm
–21323 z-limit
Y:300µm
Z: –21323µm
▲
Y:300µm
Z: –21323µm
▲
5.3.7 Arrow keys
Adjusting the limit.
By moving the petri dish, the position of
the sample may shift relative to the limit.
The value of the limit can be adjusted. To
do so, press the relevant arrow key. The
stored limit value changes by
approximately 0.2 µm per key press. Press
and hold down the key to change the limit
value continuously. In the event of changes by 1 µm
(cf. value on display) an additional acoustic signal is delivered.
88
coarseX: –1200µm
–21324 z-limit
Y:300µm
Z: –21323µm
5 User control and mode of operation
5.3.8 Menu
Toggles between menu and back.
Parameters can be selected or changed in
the menu using the joystick. For a detailed
description see section 5.4.
The left hand column of the display lists the main menu
options. The current selected option is indicated by
arrows.
The right hand column, under "Submenu", lists the parameters associated with the selected
main menu option.
5.3.9 Standby
Switches the device off or on. When the
device is off, the unlit display shows
"Standby".
To switch on, press the "Standby" key
again. The device runs through the power-
up routine.
This functionality is also used as a reset function.
5.3.10 Joystick button, triggering of injection
Triggering the Inject function.
Pressing the joystick button triggers the
automatic injection movement. The
capillary is moved to the side and then
axially to the limit level.
In parallel with the injection movement, the connected
microinjector is activated and the injection pressure is
built up. When the microinjector's injection time has elapsed the capillary is returned to the
search plane.
This procedure automatically injects the cell with the pre-set parameters.
The injection speed can also be fine-adjusted by way of
the "Inject" menu option.
5.3.12 Turn wheel
5.4 Parameter settings via menu
User control and mode of operation
The right hand column lists the submenu parameters
associated with the option. When a new main menu
option is selected the left hand column changes
accordingly.
the display. The injection speed is reduced
When the "Menu" key is pressed the main
menu options are listed on the display at
the left. The selected option is framed by
two arrows (cursor).
The main menu options are selected by
moving the joystick back and forth or
rotating it.
Alternatively, the turn wheel can be used.
INJECT
600
Setting the speed in the current speed mode.
Display of selected speed on-screen.
The turn wheel can be used to navigate round the menu and
select options.
Adjustable from 7,500 to 5 µm/s
in 5-µm/s increments.
5
Synchron: Synchronisation of connected
injector (e.g. FemtoJet®).
Setting options:
Immediate: The injection pressure is
built upwhen the Inject function
is activated.
Limit: The injection pressure is built up when the (Z) limit is reached.
Move:No injection pressure is built up;
only the injection movement is executed.
Pressure: The injection pressure is activated;
no injection movement occurs.
The synchronisation also applies to the special StepInj application.
Search+L: When the parameter is active (ON), changing the limit value additionally changes
the position of the tool – and thus of the search plane – by the same value.
The distance between the limit and the search plane remains constant.
Setting options: ON / OFF
LimAxis: The parameter sets the axis on which the limit is to be used.
User control and mode of operation
InjAxial: The active parameter (ON) causes an axial injection at the pre-set angle.
Setting options: Z-axis / X-axis.
The Z axis setting should be chosen for adherent injection.
The X axis setting is used for special injections (not currently possible).
If the parameter is disabled, the injection is performed vertically downwards
(in Z direction with setting "LimAxis = Z-axis";
with "LimAxis = X-axis" the parameter is irrelevant).
Setting options: ON / OFF
This special parameter enables the tool to be move
forward automatically from its current position by a
specific distance (D) at a specific speed, to perform an
injection and then return to its starting position.
StepInj: Switches the function on or off.
Setting options: OFF / ON
StpSpeed: Speed of
step injection.
Adjustable from 7,500 to 5 µm|s
in 5-µm/s increments.
StpDist:Distance the tool is to move forward.
Adjustable from 5 to 2,000 µm in 5-µm increments.
StpAxial: Activates or disables the axial movement of the step injection.
Setting options: ON / OFF
The synchronisation set in Inject/Synchron parameter also has an effect on the StepInj
command.
5.4.2.4 Home
HomSpeed: Sets the speed for the Home and Clean
movement.
Adjustable from 7,500 to 5 µm/s
in 5-µm/s increments.
Offset: Sets the height of the position to which the
tool is to move on completion of the Home
movement.
Adjustable from 0 to 2,000 µm in 5-µm
increments.
The offset position is on the axial axis of the start point.
Home: Switches the Home function on or off.
Setting options: ON / OFF
Clean: Switches the Clean function on or off.
Setting options: ON / OFF
Distance: Sets the distance to be used for the Clean function.
Only operative when Clean is activated.
Adjustable from 0 to 15,000 µm in 5-µm increments.
BackMan:ON: The home position can be left by
moving the joystick.
OFF: The home position can be left by
pressing the Home key again. All other
operations are not possible, except for Reset/Standby.
PosSpeed: Sets the speed at which the stored position
is approached.
Adjustable from 7,500 to 5 µm/s
5
ClearAll: When the function is run (Execute) all stored
5.4.2.6 Install
The default setting of the InjectMan® NI 2 installation parameters is right, flat and 45°.
Side: Indication of which side of microscope the
Direction: Indicates how the large motor module
User control and mode of operation
Angle: Angle at which the axial module (A module) is mounted to the plane of the
LCDLight: Switches the display illumination on or off.
Beeper: Switches the speaker on or off.
MultUser: Indicates the number of possible users potentially working with differing
Invert Y: The Invert Y parameter can be used to invert the direction of movement of the
in 5-µm/s increments.
positions are cleared.
module unit is mounted on.
Adjustable: Right or Left
(X/Y module) is mounted.
Adjustable: Horizontal (Flat) or vertical (Up).
The setting automatically changes the value
for the Angle parameter. Check both
parameters! For vertical (Up) different components are required, as well as a
different mode of mounting. See ordering information, "Non-standard accessories".
microscope table. Changing the Direction parameter changes the value of the Angle
parameter. Check both parameters!
Setting options: ON / OFF
Setting options: ON / OFF
parameters.
Setting options: 1 – 4
By way of the MultUser function, parameters can be stored separately specific to
the application and retrieved as needed. Installation parameters are stored globally.
Y axis. This is useful for microscopes that show a mirrored image, e.g. upright
microscopes.
OFF: The movement of the tool follows the joystick movement.
ON: The movement of the tool is in the opposite direction to the joystick movement.
ZeroCord: This function sets the coordinates of the
X, Y and Z axes to zero.
CentrMot: This function sets the motors to a mid
position. See section 5.2.5.
Before executing the function,
remove the probe holder and
capillary! Remove objects from the
travel range of the micromanipulator! Avoid
collisions!
If the function is nonetheless executed with a capillary present, make sure it is not
destroyed! Wear goggles as a precaution!
If you do not want to execute the function, press the "Menu" key. The current
function can be aborted by pressing the "Standby" key.
Selftest: This function triggers a self-test.
Before executing the function remove the probe holder and capillary!
Remove objects from the travel range of the micromanipulator!
Avoid collisions!
All motor axes are moved. In case of error an error message is delivered. If you do
not want to execute the function or want to stop it, press the "Menu" key.
MotorOff: This function switches off the motor currents. The motors can no longer be
operated. The tool may move to a preferred position.
UsrDeflt: This function resets the user parameters to defaults (User Default). If you do not
want to execute the function, press the "Menu" key.
In addition to the familiar speed setting (press and hold down "Speed" or "Limit" key) and
position storage key functions, there are also other special functions operated via the keypad.
5.5.1 General-Reset
The general reset resets the device to its delivery defaults. All the parameters, including the
installation parameters and the number of users, are reset to the defaults set at the factory.
Execution:
–Switch device to Standby.
–Press "Standby" to switch device on again and then immediately press and release "Home"
key.
The device runs through a lengthy initialisation phase.
97
5 User control and mode of operation
5.5.2 Software version polling
As an additional software version polling facility (also visible during the power-up routine on the
display) the version number can be indicated acoustically.
The version number is given in three signal sequences, with the digit 0 represented as 10.
5
Execution:
–Switch device to Standby
–Press "Standby" to switch device on again and then immediately press and release "Axial"
key.
The device runs through the initialisation phase and the acoustic output begins.
5.5.3 Installation
As an alternative to the Install menu option, the installation can also be performed by way of the
keypad.
Execution:
–device is on.
–Press "Speed" (Coarse/Fine) and "Axial" keys simultaneously.
–The Installation subfunction is activated and is indicated on the display and by the two
LEDs on the "Speed" key being lit.
–The "Home" key is used to enter right or left side
User control and mode of operation
mounting. In addition, a RIGHT setting is indicated
by a short acoustic signal and a LEFT setting by a
long one.
As shown on the display, pressing
the Home key would activate left side mounting.
(Home =>LEFT)
–The "Pos 1" key is used to enter horizontal or upright
mounting of the X/Y module. In addition, an UP (upright) setting is indicated by a short
acoustic signal and a FLAT (horizontal) setting by a long one.
–Pressing and holding down the "Axial" key enters the mounting angle. In addition to the
display readout, the change of angle per 5° increment is indicated by a short acoustic
signal, a 90° setting by a long acoustic signal and a 45° setting by a double one.
–Pressing and holding down the "Menu" key initiates loading of a software update. To be
undertaken only by Service!
–Press the "Coarse/Fine" key to terminate installation and save the parameters directly.
The display switches to the routine display and only the LED of the selected speed range is
lit.
The device runs through a lengthy initialisation phase.
Coarse
Fine
Coarse
Fine
Axial
Installation
Home=> LEFT
POS1=> UP
AXIAL(long)=> 40°
Menu(long)=> LOAD
Press COARSE to quit
98
6 Cleaning and service
6.1 Cleaning
Before cleaning the InjectMan
Ensure that no fluids enter into the inside of the micromanipulator in order to avoid short-circuits
in the electrical installation as well as corrosion.
Wipe painted parts and aluminum surfaces using a cloth and mild detergent and then with a dry
cloth.
Warning
Clean the InjectMan® NI 2 before disinfecting.
Wipe every part of the InjectMan® NI 2, including accessories and connector cables, with an
alcohol-based disinfectant (Isopropanol, Alcohol).
Spray disinfection is not advisable, as disinfectants can enter into the micromanipulator.
The disinfection method used must comply with current disinfection regulations and guidelines.
A decontamination certificate must accompany the InjectMan® NI 2.
A printed form ("Decontamination certificate for return of goods" and general instructions for
decontamination are available at our www.eppendorf.com homepage.
The serial number of the device must be entered in the decontamination certificate.
: Do not use any corrosive, solvent or abrasive detergents or polishes.
6.2 Disinfection
6.3 Decontamination prior to shipping
If the device is to be inspected, repaired or calibrated by the service partners of the
Eppendorf AG, it must be free of hazardous substances and clean!
®
NI 2, disconnect the plug.
6
Cleaning and service
6.4 Service
The dealer is responsible for warranty and service.
99
7 Troubleshooting
The following error situations are conceivable:
– Motor axes move in wrong direction or do not
7
Troubleshooting
match joystick movement.
– Parameters for certain user actions are not
accessible.
– Machine not responding to key presses,
Home function still active.
– Tool does not move far enough down.– Cancel limit function
– Tool does not move straight downwards
when joystick is rotated, but also moves to
the side.
– Tool moves straight downwards during
injection
– When connected, machine has no display
and will not switch on
Remedy
– Check installation parameters match
motor module setup, e.g. left side
mounting etc.
– Enter parameters again or perform a
general reset and then reset the
machine.
– Press "Home" key again,
tool moves downwards
or move joystick.
or re-adjust tool
– Disable AXIAL function.
– Adapt InjAxial parameter.
– Check mains cable and plug.
If mains power supply unit defective:
Service
100
7 Troubleshooting
The following warnings or error messages may be displayed:
Remedy
– Warning 001
Check Injector and cable:
The injector is not ready or there is no
connection to the theoretically connected
injector.
Long injection path triggered at injector,
timeout of manipulator
– Warning 002
Sync not immediate, limit or pressure:
An injection was triggered on the connected
injector even though the Move parameter is
entered on the InjectMan
injection synchronisation.
– Warning 004
Connect motor module:
Occurs in self-test. The reason for this
warning is most likely that the motor module
is not connected to the control board.
– Error 010 – 020
Certain hardware components of the control
board are missing.
– Error 021 – 023
Joystick hardware fault.
– Error 024
Wrong internal joystick parameters.
– Error 025
For specific joystick axes wrong values are
defined. Reason may be an incorrect joystick
setting on power-up or in a general reset.
– Error 030 – 033
Internal software errors.
– Error 034 – 039
Internal hardware faults.
– Error 040 – 041
Internal software errors.
– Error 042 – 043
Hardware faults
®
NI 2 as the
– Make the injector ready or check the
connecting cable.
– Trigger "Inject" at manipulator with
long injection paths.
– Do not trigger the injection by the
injector in this case or change the
Synchon parameter.
– Connect motor module.
– Service
– Service
– Reset, general reset.
If error recurs: Service
– General reset with joystick at mid
position.
If error recurs: Service
– Reset, general reset.
If error recurs: Service
– If recurring: Service
– Reset, general reset.
If error recurs: Service
– If recurring: Service
7
Troubleshooting
101
7 Troubleshooting
– Error 044 – 046
In self-test wrong values defined for
7
Troubleshooting
displayed axis.
– Error 060 – 062
False data has been imported via the serial
ports.
There may be errors in the external PC
software.
– Error 098
Unable to write to EEPROM:
No internal data can be written to the
machine's EEPROM.
– Error 099
Internal Error: Internal error
Remedy
– Repeat self-test, reset, general reset.
If error recurs: Service
– Transfer correct data.
– General reset or Service (new
software download or hardware).
– Reset, general reset.
If error recurs: Service
102
8 Technical data
Power supply
Input:120 – 240 V AC; 50 – 60 Hz; < 50 W
Output: 15 – 18 V DC; 25 W
Control board with power supply:2.7 kg
Module unit:1.4 kg, incl. guide
Dimensions
XY module:120 x 100 x 68 mm
A module:120 x 60 x 50 mm
Control board:270 x 205 x 160 mm
Max. travel:> 18 mm (each axis)
Resolution per microstep:approx. 40 nm
Max. tool travel speed:7,500 µm/s
The device carries the mark.
15 – 18 V DC; 25 W
II
II
(IEC 61010-1)
8
Technical data
Technical specifications subject to change!
103
9 Ordering information
InjectMan® NI 2, Europe5181 000.017
InjectMan® NI 2, USA5181 000.025
9
InjectMan
InjectMan® NI 2, Japan5181 000.041
InjectMan
InjectMan® NI 2, China5181 000.084
Delivery package
1 Control board
1 Motor module
1 Mains power supply unit
1 Mains power cable
1 Hinged bracket
Ordering information
1 Set of Allen keys (metric)
2 Screws M5x16
1 Interface cable FemtoJet
2 Positioning aidsfor universal capillary holders5171 130.001
1 Operating manual5181 900.018
1 Set of spare screws 2 x M5x16, 2 x M4x10, 2 x M3x16, 2 x M3x12
Adapters
Adapter for Leica DMIL5181 201.004
Adapter for Leica DMIRE25181 202.000
Adapter for Leica DMI 60005181 203.007
Adapter for Nikon TMD5181 210.003
Adapter for Nikon TS-1005181 212.006
Adapter for Nikon Ti-E/-U/-S5181 214.009
Adapter for Olympus IX 50/70 IX 51/71/815181 220.009
Adapter for Olympus IMT-25181 221.005
Adapter for Olympus CK 30/40 / CKX 31/415181 222.001
Adapter for Zeiss 100/135/100 S5181 230.004
Adapter for Zeiss 10/355181 233.003
Adapter for Zeiss Axiovert 200 / AxioObserver5181 235.006
Adapter for Zeiss 25/25C / 40C/40CFL5181 237.009
Hund Wilovert 305181 234.000
Universal stand5181 250.005
Extension plate5181 236.002
®
NI 2, UK/Hongkong5181 000.033
®
NI 2, Australia5181 000.076
Order no.
®
(5247)5181 070.015
104
9 Ordering information
Accessories
Universal capillary holder for accommodating Femtotips®,
Femtotips® II or grip heads
Grip head 0 for universal capillary holder
for microcapillaries with an outer diameter of 1.0 to 1.1 mm
Grip head 1 for universal capillary holder
for microcapillaries with an outer diameter of 1.2 to 1.3 mm
Grip head 2 for universal capillary holder
for microcapillaries with an outer diameter of 1.4 to 1.5 mm
Grip head 3 for universal capillary holder
for microcapillaries with an outer diameter of 0.7 to 0.9 mm
O-Ring Set
for grip heads 0 – 3 with removal tool
Service kit for grip heads and capillary holder,
consisting of O-Ring Set, adapter for Femtotips
20 Femtotips
20 Femtotips® II5242 957.000
2x 96 Microloader5242 956.003
Interface cable 5246/52425181 150 060
Foot control 51815181 150.051
Arc guide5181 150.019
"A" head (used with arc guide)5181 150.027
PC cable update5181 150.094
Service kit
(Screws M3x8, M3x16, M4x10, M4x14, M5x16 and war. M4, M5, M8)
Straight-line guide to connect X module to Y/Z module5181 150.035
"X" head with angle adjuster5181 150.043
Spacer (41 mm) for Universal Stand 5181 255.007
(non-standard)
®
Order no.
5176 190.002
5176 210.003
5176 212.006
5176 214.009
5176 207.002
5176 196.000
®
5176 195.004
5242 952.008
5181 150.310
9
Ordering information
Microinjection
FemtoJet®
Programmable microinjector with integrated pressure supply
Version Europe5247 000.013
Version USA and Canada5247 000.056
Version UK5247 000.030
Version Japan 100 V, 50 – 60 Hz5247 000.048
Version Australia5247 000.072
Version China5247 000.080
105
9 Ordering information
FemtoJet® express
Programmable microinjector incl. connection for external pressure supply
9
Version Europe5248 000.017
Version USA and Canada5248 000.050
Version UK/Hongkong5248 000.033
Version Japan 5248 000.041
Version Australia5248 000.076
Version China5248 000.084
Cell transfer
CellTram Oil
manual piston pump for transferring cells (sperms, ES cells) and
as manual injector with universal capillary holder for
Ordering information
capillaries with an outer diameter of 1 mm
CellTram vario
manual piston pump with variable transmission (1:1, 1:10)
to transfer cells (sperms, ES cells) and as manual injector with
universal capillary holder for capillaries with an outer diameter of 1 mm
CellTram Air
manual piston pump for holding cells (e.g. oocytes) with
universal capillary holder for capillaries with an outer diameter of 1 mm
25 TransferTips® (ES)
sterile glass capillaries for transferring ES cells, inner diameter 15 µm,
outer diameter 20 µm, with spike, 20° angle
Headstage holder for preamplifiers HEKA EPC 7 / EPC9
and AXON CV-4 HS-2
Order no.
5176 000.025
5176 000.033
5176 000.017
5175 108.000
5175 107.004
5173 071.002
106
9a Ordering information for North America
Order no.
InjectMan® NI 2, USA920000029
InjectMan® NI 2, international 5181 000.017
InjectMan
InjectMan® NI 2, Japan5181 000.041
InjectMan
InjectMan® NI 2, China5181 000.084
Delivery package
1 Control board
1 Motor module
1 Mains power supply unit
1 Mains power cable
1 Hinged bracket
1 Set of Allen keys (metric)
2 Screws M5x16
1 Interface cable FemtoJet
2 Positioning aidsfor universal capillary holders920005829
1 Operating manual5181 900.018
1 Set of spare screws 2 x M5x16, 2 x M4x10, 2 x M3x16, 2 x M3x12
Adapters
Adapter for Leica DMIL920008500
Adapter for Leica DMIRE2920008526
Adapter for Leica DMI 6000920008496
Adapter for Nikon TMD920008534
Adapter for Nikon TS-100920008551
Adapter for Nikon Ti-E/-U/-S920008658
Adapter for Olympus IX 50/70 IX 51/71/81920008585
Adapter for Olympus IMT-2920008577
Adapter for Olympus CK 30/40 / CKX 31/41920008569
Adapter for Zeiss 100/135/100 S920008607
Adapter for Zeiss 10/35920008593
Adapter for Zeiss Axiovert 200 / AxioObserver920008615
Adapter for Zeiss 25/25C / 40C/40CFL920008666
Hund Wilovert 305181 234.000
Universal stand920008631
Extension plate920008674
®
NI 2, UK/Hongkong5181 000.033
®
NI 2, Australia5181 000.076
®
(5247)920005845
9a
Ordering information for North America
107
9a Ordering information for North America
Accessories
Universal capillary holder for accommodating Femtotips®,
9a
Femtotips® II or grip heads
Grip head 0 for universal capillary holder
for microcapillaries with an outer diameter of 1.0 to 1.1 mm
Grip head 1 for universal capillary holder
for microcapillaries with an outer diameter of 1.2 to 1.3 mm
Grip head 2 for universal capillary holder
for microcapillaries with an outer diameter of 1.4 to 1.5 mm
Grip head 3 for universal capillary holder
for microcapillaries with an outer diameter of 0.7 to 0.9 mm
O-Ring Set
for grip heads 0 – 3 with removal tool
Service kit for grip heads and capillary holder,
consisting of O-Ring Set, adapter for Femtotips
20 Femtotips
20 Femtotips® II930000043
2x 96 Microloader930001007
Interface cable 5246/5242920005853
Foot control 5181920005799
Ordering information for North America
Curved guide920005802
"A" head (used with curved guide)920005748
PC cable update920005837
Service kit
(Screws M3x8, M3x16, M4x10, M4x14, M5x16 and war. M4, M5, M8)
Straight-line guide to connect X module to Y/Z module920005756
"X" head with angle adjuster920005764
Spacer (41 mm) for Universal Stand 920008666
(non-standard)
®
Order no.
920007392
920007414
910012509
910012517
920007406
920005870
®
920005888
930000035
920005781
Microinjection
FemtoJet®
Programmable microinjector with integrated pressure supply
Version USA and Canada920010504
Version Europe5247 000.013
Version UK5247 000.030
Version Japan 100 V, 50 – 60 Hz5247 000.048
Version Australia5247 000.072
Version China5247 000.080
108
9a Ordering information for North America
Order no.
FemtoJet® express
Programmable microinjector incl. connection for external pressure supply
Version USA and Canada5248 000.050
Version Europe5248 000.017
Version UK/Hongkong5248 000.033
Version Japan 5248 000.041
Version Australia5248 000.076
Version China5248 000.084
Cell transfer
CellTram Oil
manual piston pump for transferring cells (sperms, ES cells) and
as manual injector with universal capillary holder for
capillaries with an outer diameter of 1 mm
CellTram vario
manual piston pump with variable transmission (1:1, 1:10)
to transfer cells (sperms, ES cells) and as manual injector with
universal capillary holder for capillaries with an outer diameter of 1 mm
CellTram Air
manual piston pump for holding cells (e.g. oocytes) with
universal capillary holder for capillaries with an outer diameter of 1 mm
25 TransferTips® (ES)
sterile glass capillaries for transferring ES cells, inner diameter 15 µm,
outer diameter 20 µm, with spike, 20° angle
Headstage holder for preamplifiers HEKA EPC 7 / EPC9
and AXON CV-4 HS-2
920002030
920002111
920002021
930001015
930001040
920003168
9a
Ordering information for North America
109
A Appendix – MultiUser
A
MultiUser
From the Install submenu up to 4 Users (users/applications) can be activated. The Users can be
assigned dedicated parameters for speed, injection parameters etc. Non-User-specific parameters such as installation parameters apply globally, and when changed the changes apply to
all Users.
MultUser
Indicates the number of possible users potentially
working with differing parameters.
Setting options: 1 – 4
By way of the MultUser function, parameters can
be stored separately specific to the application and
retrieved as needed. Installation parameters are stored
globally.
When MultUser = 1 is set the part in the display is blank.
Where there are multiple Users, the routine display indicates the relevant User, e.g. User 2.
Users can only be changed by switching to Standby and restarting.
Display after power-up if 2 or more Users are activated:
The User value for the
desired application or
user is set by means of
the turn wheel at the side.
Select User
User:2
Turn speed to
select
Press Home to cont.
After setting the desired User press the "Home" key.
The routine display now additionally indicates the User
you have set:
Note
: The number of Users can only be reset to 1 by way
of User 1.
coarseX: –1200µm
Y:300µm
Z: –21323µm
user 2
The specific parameters of the individual users are
retained even when the MultUser parameter is set to 1.
When an additional user, such as User 2, is reactivated
the parameters are then available once again.
The Multiuser parameters can be deleted and reset by way of a general reset, though this will
also reset all global parameters, such as installation parameters, to their factory defaults.