1 Product description

The Eppendorf Microplate VIS, 96/F-PS is suitable for the absorbance measurement in a plate reader with wavelengths higher than 340 nm.

2 Notes for use

WARNING! Contamination risk from damaged plates during centrifugation.
The plates withstand heavy loads during centrifugation. However, they can be destroyed and release the substances they contain if used improperly.
Always observe the maximum permitted centrifugation forces.Read the operating manual of the centrifuge used.Centrifuge stacked plates at lower speed only.Note that organic solvents may reduce the mechanical load capability
of the plates. If in doubt please contact the Eppendorf Application Support (support@eppendorf.com).
WARNING! Damages to health due to infectious liquids and pathogenic germs.
When handling infectious liquids and pathogenic germs, observe the
national regulations, the biological security level of your laboratory, the material safety data sheets, and the manufacturer's application notes.
Wear personal protective equipment.Consult the "Laboratory Biosafety Manual" (Source: World Health
Organization, Laboratory Biosafety Manual, as amended) for comprehensive regulations on the handling of risk group II germs or biological materials.
Instructions for use
2
Microplate VIS, 96/F-PS English (EN)
Only use optically perfect and undamaged products.Reagents, consumables contaminated by reagents and materials used for
cleaning and disinfecting must be disposed of in accordance with laboratory regulations.
When using aggressive chemicals like (e. g. acetone) please contact the
Eppendorf Application Support (support@eppendorf.com).

3 Technical data

Material: Polystyrene, colorless
Dimensions: Conforms to ANSI/SLAS 1-2004 through
ANSI/SLAS 4-2004
Operating temperature: -86 °C to +70 °C
Centrifuging: Can be centrifuged up to 3 000 x g**
When arranged in a stack (5 plates) can be centrifuged up to 300 x g*
Sealing: You can close the plates with Eppendorf Storage
Film or Eppendorf Storage Foil or with the Eppendorf Lid
Purity: PCR clean
*The centrifugation stability of each plate type generally depends on the centrifuge and its accessories, the ambient conditions and the liquid used.

4 Ordering Information

Instructions for use
Microplate VIS, 96/F-PS
English (EN)
3
Order no. (International)
Order no. (North America)
0030 730.020 0030730020
6141 000.002
6141000010
0030 127.870 0030127870
0030 127.889 0030127889
0030 131.517 0030 131.525
0030131517 0030131525
Description
Microplate VIS, 96/F-PS
PCR clean
PlateReader AF2200
230 V / 50 – 60 Hz 120 V / 50 – 60 Hz (US)
Storage Film
PCR clean, adhesive, 100 pieces
Storage Foil
PCR clean, adhesive, 100 pieces
Eppendorf Plate Lid
PCR clean, 80 pieces sterile, 80 pieces
Instructions for use
Microplate VIS, 96/F-PS
English (EN)
Your local distributor: www.eppendorf.com/contact
Eppendorf AG · 22331 Hamburg · Germany eppendorf@eppendorf.com · www.eppendorf.com
and the Eppendorf logo are registered trademarks of Eppendorf AG.
®
Eppendorf