Druck MCX-II Data sheet

MCX II
MCX II
Portable documenting calibrator
n Reduces instrumentation maintenance costs
n Over 90 Input/Output ranges: Pressure,
n Interchangeable pressure modules -14.7 to 5000 psi
n HART® module for Smart transmitters
n Memory card for procedure and data transfer
cd
n Eliminates field calibration errors
USMCX II -PDS-A132 09/00
1
MCX II
Portable documenting calibrator
A CALIBRATION WORKSHOP IN A SINGLE INSTRUMENT
The Druck MCX II portable documenting calibrator is the most comprehensive field calibration tool available. It is the culmination of many years combined field experience with the Druck and Unomat series of portable pressure, temperature and electrical calibrators.
Designed for field use, this rugged, self-contained, battery powered package simulates and reads RTD’s, thermocouples and resistance, as well as sourcing and reading milliamps, millivolts, volts and frequency. With interchangeable single and dual sensor pressure modules over 90 input and output ranges are available.
The MCX II saves time and money with the calibration, maintenance and commissioning of instrumentation for process plants, production lines, utility processing and distribution by:
n Reducing the burden imposed by quality systems such as ISO 9000. n Reducing calibration, maintenance and commissioning time. n Reducing documentation time and errors. n Replacing several standard test instruments. n Reducing test instrument calibration costs. n Minimising down time and maximising efficient field usage.
For example, a typical thermocouple transmitter calibration can take one hour using a mV source, look-up tables and a milliammeter. In just five minutes the MCX II can make an automatic calibration and document the results while virtually eliminating human errors.
A PCMCIA memory card provides data storage and gives total flexibility to suit different working practices. By simply exchanging PCMCIA cards, the MCX II can remain permanently in the field and when compared to serial data transfer methods this can save one to two hours a day. With a single item to calibrate the cost of re-calibration is reduced and the inconvenience of down time is minimized.
HART® communicator for SMART transmitters
The HART® communicator allows digital field adjustment of smart transmitters. Typical adjustments to sensor and analogue trims can take up to 40 minutes using conventional test equipment and a hand held communicator. With a single MCX II this time can be reduced to less than 10 minutes, including a fully documented calibration.
High precision and multi-functional
Typical accuracy: 0.01% Rdg +/-0.003% FS for mA measurement
Input: mA, mV, volts, T/C’s, RTD’s, pressure, ohms, frequency and switch state. Output: mA, mV, volts, T/C’s, RTD’s, ohms, and frequency. Pressure modules: Interchangeable single and dual ranges from -14.7 to
HART® communicator: HART® digital communicator for SMART transmitters. C/J compensation: Internal, external and manual. Loop Power: Dual 24 Vdc. Temperature probe: 1/5 DIN accuracy P100 probe. Data storage: 1 Mbyte PCMCIA card. Data transfer: PCMCIA card or RS 232 interface. PC software: Linkpak-W and Intecal-W.
Easy to use
The multi-lingual user interface is an easy to use Input/Output menu with dual parameter readout. The Input and Output connectors are standard 4mm gold plated sockets which are separated and clearly labelled. The rugged impact resistant enclosure is surrounded by a durable protective carry case which allows access to all the instrument features and provides convenient pockets for storing test leads and accessories. Whether the MCX II is horizontal or vertical the rotatable display provides the optimal viewing angle. On the bench, test leads and pressure modules connect to the front face. In the field, with the MCX II held vertically by the wide neck strap, the connections are made to the rear face. With safety a major design concern, these features reduce the possibility of dropping equipment as the operators hands are kept free.
0.05% Rdg for pressure measurement,
5000 psi including gauge, absolute and differential.