Druck DPI 610LP, DPI 615LP Operating Manual

GE
Sensing
& Inspection Technologies
Druck DPI 610/615
Portable Pressure Calibrator Series
User manual - K0415
English
//www.gesensing.com/toolsupport/manuals.htm
Français
Pour choisir le manuel dans une langue disponible, accédez à :
//www.gesensing.com/toolsupport/manuals.htm
Deutsch
Um das Handbuch in einer vorhandenen Sprache auszuwählen, gehen Sie zu:
//www.gesensing.com/toolsupport/manuals.htm
Italiano
Per scaricare il manuale in una delle lingue disponibili consultare la pagina:
//www.gesensing.com/toolsupport/manuals.htm
Español
Para seleccionar el manual en uno de los idiomas disponibles vaya a:
//www.gesensing.com/toolsupport/manuals.htm
Português
Para selecionar o manual em uma língua disponível vá:
//www.gesensing.com/toolsupport/manuals.htm
Polski
Aby wybrać podr ęcznik w dostępnym j ęzyku, przejdź do st ro ny:
//www.gesensing.com/toolsupport/manuals.htm
Язык
Для выбора руководства на имеющемся языке зайдите на web-сайт
//www.gesensing.com/toolsupport/manuals.htm
语言
选择手册的语言,请访问:
//www.gesensing.com/toolsupport/manuals.htm
言語
利用可能な言語のマニュアルを選択するには次のサイトへアクセスしてください :
//www.gesensing.com/toolsupport/manuals.htm
i K0415 Issue No. 2
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. Do not use this equipment for any other purpose than that stated. This publication contains operating and safety instructions that must be followed to ensure safe operation and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage. Use suitably qualified * technicians and good engineering practice for all procedures in this publication.
Pressure
Do not apply pressures greater than the safe working pressure to this equipment.
Maintenance
The equipment must be maintained using the procedures in this publication. Further manufacturer’s procedures should be carried out by authorized service agents or the manufacturer’s service departments.
www.gesensing.com
For technical advice contact the manufacturer.
* A qualified technician must have the necessary technical knowledge,
documentation, special test equipment and tools to carry out the required work on this equipment.
Symbols
This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.
This symbol, on the instrument, indicates that the user should refer to the user manual. This symbol, in this manual, indicates a hazardous operation.
This symbol, on the instrument, indicates do not throw-away in domestic bin, hazardous material, dispose correctly in accordance with local regulations.
K0415 Issue No. 2 ii
Specification
Safe working pressure
20 bar range (300 psi) 1.75 x full-scale 350 bar range (5000 psi) 1.2 x full-scale 400 bar range (6000 psi) 1.5 x full-scale All other ranges 2 x full-scale
Accuracy
Combined non-linearity, hysteresis and repeatability
±70 mbar range (2 inHg) 0.05% F.S. up to ±150 mbar (4.4 inHg) 0.05% span 200 mbar to 20 bar (3 psi to 300 psi) [Calibrator]): 0.025% F.S. 35 bar to 700 bar (500 psi to 10000 psi) [Indicator] 0.025% F.S. 70 bar to 400 bar (1000 psi to 6000 psi) [Hydraulic] 0.025% F.S.
Pressure Ranges
Refer to the pressure range matrix in the data sheet.
Temperature Effects
±0.004% of reading/°C (averaged over -10° to +40°C w.r.t. 20°C) ±0.002% of reading/°F (averaged over +14° to 104°F w.r.t. 68°F)
Power supply
Batteries 6 x 1.5 V C cells, alkaline (up to 60 hours nominal use at 20°C) Rechargeable NiCad battery pack (20 hours nominal use) supplied with charger/ adaptor, supplies power to instrument while charging batteries.
Voltage Inputs
Range: ±50V Accuracy ±0.05% rdg, ±0.004% F.S. Resolution 100µV max
Voltage Outputs
Range: ±10V Accuracy ±0.1% Load 10mA
Range: ±24V Accuracy ±5% Load 26mA
Current Inputs
Range: ±55mA Accuracy ±0.05% rdg, ±0.004% F.S. Resolution 1µA max
Current Output
Range: 24mA Accuracy ±0.05% rdg, ±0.01% F.S. Resolution 1µA max
iii K0415 Issue No. 2
Display
Size: 60 x 60 mm (2.36” x 2.36”) LCD Graphics Reading ±99999, update rate 2 readings/sec
Environment
Operating Temperature: -10°C to 50°C (+14°F to 122°F) Calibrated Temperature: -10°C to 40°C (+14°F to 104°F) Storage Temperature: -20°C to 60°C (-4°F to 140°F) Calibration Temperature: 21°C ±2°C (70°F ±4°F)
Sealing
Physical
Size: 300 x 170 x 140 mm (11.8” x 6.7” x 5.5”) Weight: 3 kg (6.6lb)
Specification (contd.)
Sealed to IP54 (NEMA 4)
K0415 Issue No. 2 iv
Introduction
General
Description of Procedures 1
Summary of Functions
Using the Guide 2
Getting Started
Fitting Batteries 7 Switching On 7 Change Pressure Units 8 Voltage and Current Measurement 8 Typical Calibration Set-up (Pressure to Voltage) 9 Zero Display Reading 9
OPERATOR CONTROLS 3 DISPLAY 3 HARD KEY FUNCTIONS 4 SOFT KEYS 5 CURSOR KEYS 5 ELECTRICAL CONNECTIONS 6
Task Selection
Task Key 10 Using Task Functions 10 Set Units 10 Set 24 Volts 10 Cal Mode (DPI 615 versions only) 11 Basic Mode (Task BASIC) 11
Taking Measurements
Pressure Transmitter (P-I) Task 12 Voltage Output Pressure Transmitter (P-V) Task 12 Pressure Converter (P-P) Task 13 Current to Pressure Converter (I-P) Task 14 Pressure Switch Test (P-Switch) Task 14 Pressure Switch Testing with Contact Resistance Measurement 15 Pressure to Display (P-Display) Task 16 Leak Test (Leak Test) Task 17 Transmitter Simulator (TX SIM) Task 18 Relief Valve Test (REL VALVE) Task 19
v K0415 Issue No. 2
Advanced Task
Electrical Outputs (Loop Power) 25
General 20 Select Input 20 Ambient Temperature Measurement 20 Process Functions 21 Tare Process Function 22 Min/Max Process Function 23 Filter Process Function 23 Flow Function 24 % Span 24 Select Output 25
mA Step 26 mA Ramp 27 mA Value 28 24 Volt 28 Define New Task 29 Clear Task 29
Memory Operations
Saving Display or Data Log 30 Store Operations (Screen Snapshots) 30 Recalling Stored Data (Screen Snapshots) 30 Datalog Operations 31
Using Set-up
General 37 Store Mode 37 Contrast 37 Settings - Select Set-up Option 38
Auto Log (Timer) 31
Manual Logging 31 Recall Data Log Files 32 Uploading Data Log Files 33 Delete Data Log and Procedure Files 33 Downloading Procedure Files (DPI 615 versions only) 34 Running Procedure Files (DPI 615 versions only) 35 Recalling Data Files 36
Units 38
Define Special Units 38
Language 39
RS232 39
Powerdown 40
K0415 Issue No. 2 vi
Date and Time (Real Time Clock) 41
Backlight 42
Calibration
General 43 Calibration Check 43 Calibration Adjustment 43 General Procedures 44 Using The Calibration Menu 45 Change PIN 45 Pressure 46 Voltage Input Range (5 Volts) 47 Voltage Input Range (50 Volts) 49 Current Input Range (55 mA) 51 Current Output Range (24 mA) 53 Temperature 56 External Pressure Range 57 Add External Sensor 58
Hydraulic Calibrator Versions
Calibration 40
Date Format 41
Set Date 41
Set Time 41
Introduction 61 Safety Instructions 62 Preparation for Use 62 Bleeding the System 63 Draining the Hydraulic Fluid 64 Flushing, Replenishing or Changing the Hydraulic Fluid 65
Low Pressure Calibrator Versions
Introduction 69 Preparation for Use Low Volume Systems 70 Method of Testing with Low Volume Systems 70 Preparation for Use Larger Volume Systems 72 Method of Testing with Larger Volume Systems 74 Calibration 75 Calibrate Internal Pressure Range 75
Appendix 1 - Datalog File Example
Typical Uploaded Datalog File (DPI 610) 79 Typical Uploaded Procedure Data File (DPI 615) 80
vii K0415 Issue No. 2
K0415 Issue No. 2 viii

INTRODUCTION Summary of Functions

General
The versions of the DPI 610 and DPI 615 instruments are: pneumatic indicator, pneumatic calibrator, hydraulic calibrator and low pressure pneumatic calibrator. All instruments measure and display pneumatic and hydraulic pressure applied to the test port or to an externally connected pressure sensor. Pressure measurement can be absolute, gauge and sealed gauge and in ranges from 2.5 mbar to 700 bar (1.0 inH
The calibrator versions of this instrument contain pneumatic or hydraulic pressure generation components to produce pneumatic pressure ranges between -1 to 20 bar (-14.5 psi to 300 psi) and hydraulic pressure ranges up to 400 bar (6000 psi). The electrical connections, on the front of the instrument, enable the instrument to measure ±50 volts d.c. and ±55 mA and produce 10 volts d.c. or 24 volts d.c. and a maximum of 24 mA. An integral sensor provides measurement of ambient temperature. Additional sensors (option B1) connect to an external connector and extend the pressure measurement range and include differential pressure measurement. The instrument has an RS232 connector to enable uploading of test data to a compatible documenting system. The DPI 615 has the ability to download, from a PC, pre-defined calibration and test routines. Six alkaline C size batteries or (option A) rechargeable batteries with a charger/adaptor, power the instrument.
Important Notice
Zinc-carbon and zinc-chloride cells should NOT be used in this instrument.
Use only the battery types as shown in the table on page 7.
O to 10000 psi).
2
Description of Procedures
In the procedures outlined in this user guide, hard (fixed function) and soft (variable function) key operations are shown in bold type (e.g.) TASK and F1. These statements mean press the TASK key and press the F1 key. Soft key operations can be allocated to both the F1 and F2 keys. Where a specific soft function is referred to it is written in bold italics (e.g.) PROCESS. This instrument has a number of operating modes that are described in a simplified form in the following sections. Diagrams accompanying the procedures give typical selection sequences and shaded controls indicate that this control key should be pressed in the appropriate sequence. Diagrams should be read from left to right, top to bottom where appropriate. A shaded display soft box indicates that the function key immediately below that soft box should be pressed (either F1 for the left hand soft box or F2 for the right).
In the above diagram the following key sequence is indicated. (a) Press the F2 key (the key immediately below the
(b) Use the
shaded, use all these keys to select or enter data).
(c) Press the ENTER key.
Up
and
Down
cursor keys (only) to select the required option. (If all keys
PRESSURE UNITS
1 K0415 Issue No. 2
soft box).

INTRODUCTION Summary of Functions

Using this guide
The following key symbols are used in the procedure diagrams that follow:
SELECT VALUE
Shaded cursor keys indicate that a combination of these four keys, Up, Down, Left and Right should be used to (e.g.) enter an alpha numeric value or to select a function.
Indicates the ENTER key. Used to confirm an operation or a selection. Shading indicates key operation.
Exit key, used to clear current menu selection and return to next menu level above current level. Used as an escape key from current operation. Shading indicates key operation.
Hardkey (total 7). Legend beside key symbol indicates function. Shading indicates key operation.
Maximum Instrument Ratings
The following table shows the maximum measurement input ratings of the instrument that should not be exceeded.
PRESSURE 120% FULL SCALE
VOLTAGE 50 V d.c.
CURRENT 55 mA d.c.
Note 1: The display flashes if the input pressure, voltage or current overrange.
Note 2: Max applied voltage for external loop supply = 30V dc (see page 8).
K0415 Issue No. 2 2
PRESSURE INT
bar
VOLTAGE
V
F2F1
6
8
7
1
CURRENT
PRESSURE
UNITS
TASK: BASIC
DPI 615
2
3
4
5
PRESSURE INT bar
VOLTAGE
V
CURRENT
UNITS
PRESSURE
TASK: BASIC
+
-
Status display
Input display
Output display
Soft boxes

INTRODUCTION Summary of Functions

OPERATOR CONTROLS (Figure 1 and 2)
These divide into two groups, the operator/display controls (Figure 1) and the pressure/ vacuum generation components (Figure 2). The operator controls and a typical display, common to all instrument versions, is shown below.
1 Display 2 Electrical Input Sockets 3 Electrical Output Sockets 4 Cursor 5 Enter Key 6 On/Off Key 7 Function (soft) Keys 8 Hard Keys
Figure 1 - DPI 610/615 Key-pad
Display
The display section of the instrument basically divides into four distinct sections. The two main sections of the display are used to display a input and an output. The remaining sections show the status display area and define the soft key functions. A typical display is shown below:
3 K0415 Issue No. 2

INTRODUCTION Summary of Functions

HARD KEY FUNCTIONS (Figure 1)
Key Function
I/O
SETUP*
ZERO
INPUT*
TASK
OUTPUT*
This key is used to turn the instrument ON and OFF. 7
The
SETUP
key provides access to the instrument’s general configuration parameters that are set up to certain default parameters on delivery.
The
ZERO
key can be used to zero either the selected input or output display, if the display reading is within 5% of zero. Attempts to zero a larger offset result in an error message,
The
INPUT
key is used to select the input parameter to be displayed. 20, 21
The
TAS K
key is used as a means of rapidly configuring the instrument for a number of different types of external device calibration. There are twenty task configurations available, eleven of which are pre­programmed and nine are user definable
The
OUTPUT
key is used to select output parameter to be displayed. 25-28
Zero too large.
Page
reference
37
9
10
STORE*
RECALL*
ENTER
EXIT
Depending upon how the instrument’s is used either to store up to 20 display screens (in or to manually log a screen in
This key is used to recall a previously stored screen to the display. Depending on the either the snapshot of a previously stored screen or datalog file. In
STORE
mode, selection displays the last screen stored. By using the cursor keys, the operator can scroll either forward or back through memory locations.
The
ENTER
conjunction with the soft keys, to accept a given selection.
The
EXIT
keys to exit from the current screen or menu level, to the level immediately preceding it. To quit completely from any menu level, press
key is used either to enter data (accept entered data), or, in
key operates in conjunction with all the other hard and soft
EXIT
until the
STORE
MEASURE/SOURCE
DATALOG
mode set-up, operation of this key recalls
STORE
mode.
screen is displayed.
* These key functions are not available in BASIC mode
mode is setup, this key
SNAPSHOT
mode),
30-36
30, 32-36
2
2
K0415 Issue No. 2 4
7
8
9
5
6
4
3
11
10
12
1
2
5

INTRODUCTION Summary of Functions

SOFT KEYS (Figure 1)
Three soft keys, designated F1, EXIT and F2, are situated immediately below the display as shown below. These keys have their function allocated by the instrument software which is indicated in the bottom of the display (Voltage for F1 and Units for F2 in this example). They are used to select menu (program) options and are fully described under the appropriate section headings.
CURSOR KEYS (Figure 1)
The cursor keys consist of a block of four keys, designated up , down , left and right . In programs where options need to be selected from a list, (e.g.) the TASK
selection program, the up and down cursor keys are used to highlight one of the options, from which it can be selected by the ENTER key. In TASK mode, where more
than one page of options are provided, the left and right cursor keys will switch between pages.
1 Test port, connect to unit under test 2 Hard keys 3 Cover (external interfaces) 4 Electrical inputs 5 Electrical outputs 6 Cursor keys 7 Function (soft) keys 8 Release valve (releases pressure through 9) 9 Vent port 10 Select positive or negative pressure 11 Pump 12 Fine pressure adjuster
Figure 2 - DPI 610/615 Calibrator Controls
5 K0415 Issue No. 2
1
2
3
4
5
PRESSURE INT
bar
VOLTAGE
V
F2F1
2
CURRENT
UNITS
5
3
4
1
PRESSURE
TASK: BASIC
DPI 615
INPUTS
OUTPUTS

INTRODUCTION Summary of Functions

ELECTRICAL CONNECTIONS
1 Cover, closed when
not using connectors 2 External transducer 3 RS232 connector 4Temperature sensor 5 DC power input
Figure 3 - Electrical System Connections
Measurement inputs and Source outputs are made via the control panel sockets as shown below.
1 Status display 2 Input display 3 Electrical input sockets 4 Electrical output sockets 5 Output display
Figure 4 - Electrical Measurement Inputs/Outputs
K0415 Issue No. 2 6
1
2
3

Getting Started

Fitting Batteries
1Cover fixing screws.
2 Rechargeable NiCad pack Part no.
191-126
3 Six alkaline C-cells Type no. LR14.
Caution: Old batteries can leak and cause corrosion. Never leave
discharged batteries in the instrument.
Note: After fitting a rechargeable NiCad battery pack the display may show the battery
low indication caused by the battery pack not being fully charged. Use the power adaptor/charger to fully charge the pack.
Switching On
Press the I/O switch on the front panel and proceed as follows.
The first time the instrument powers up, it will be in BASIC mode with the main screen displaying voltage in the input display area and pressure in the output display area. To switch to CURRENT as input, press F1 as shown. Similarly, F1 to return to VOLTAGE.
Note: No other keys are active in this mode and the instrument can only be reconfigured
by pressing the TASK key and selecting another mode.
7 K0415 Issue No. 2
55mA
Vin
mA in
mA Sink
mA out 10V out
Xmtr
24V out
max 50V
CAT II
max
55mA
Vin
mA in
mA Sink
mA out 10V out
Xmtr
24V out
max 50V
CAT II
max
+
-
V
55mA
Vin
mA in
mA Sink
mA out 10V out
Xmtr
24V out
max 50V
CAT II
max
+
-
LOAD

Getting Started

Change Pressure Units
To change the pressure units proceed as follows. If the four units displayed are not the units required, press TAS K and select any task, other than BASIC, press SETUP and proceed as detailed on page 37. To return to BASIC mode, press TASK and select BASIC.
In BASIC mode, the instrument is configured to carry out basic Pressure to Voltage (P to
V) or Pressure to Current (P to I) tests, a typical test procedure follows.
Voltage and Current Measurement
Connect the electrical input sockets as follows for voltage and current measurements. Use the test leads provided and DO NOT push bare wires into the sockets.
Note: Maximum applied voltage = 50V dc,. Maximum input current = 55mA dc
Note: Maximum applied voltage for external loop supply = 30V dc
K0415 Issue No. 2 8
F2F1
P
V
-+
+
-
+
-
bar
V
CURRENT
PRESSURE
UNITS
TASK: BASIC
DPI 615
A
B
C
ZERO
VOLTAGE
PRESSURE INT bar
V
YES
ZERO VOLTAGE DC ?
NO
VOLTAGE
PRESSURE INT bar
V
ZERO
INPUT
ZERO
OUTPUT
TASK: BASIC TASK: BASIC
ZERO
VOLTAGE
PRESSURE INT bar
V
YES
ZERO INTERNAL SENSOR ?
NO
VOLTAGE
PRESSURE INT bar
V
ZERO
INPUT
ZERO
OUTPUT
TASK: BASIC TASK: BASIC

Getting Started

Typical Calibration Set-up (Pressure to Voltage)
Connect a device under test to the instrument as shown below:
A - External pressure source (indicator instruments only) B - Pressure regulator C - Excitation 10V
General procedure
Use the hand-pump to pressurize the system to the required level as indicated on the
display. Allow the display to settle and screw the volume adjuster in or out as a fine
adjustment to the required pressure. Record the input (e.g.) Voltage, reading at each
applied pressure.
Zero Display Reading
Both the input and output readings can be set to zero using the ZERO key and if the displayed reading is within 5% of zero. To zero either the INPUT or OUTPUT displays, proceed as follows:
9 K0415 Issue No. 2
BASIC
USER TASKS
P - SWITCH
ADVANCED
REL VALVE
LEAK TEST
TX SIM
TASKS TASKS TASKS
P - DISPLAY
P-I
I-P
P-P P-V
Current (Page 2) (Page 3)
OUTPUT
SELECT
FUNCTION
UNITS
ON
inHg
hPa Pa
bar
SELECT FUNCTION
SELECT UNITS OF PRESSURE
SELECT UNITS
24 VOLT
OFF
OUTPUT
SELECT ON/OFF
SELECT FUNCTION
SELECT OUTPUT
24 VOLT
OFF
UNITS
24 VOLT
ON
UNITS
SELECT OUTPUT
ON

Task Selection

TASK key
The TASK key is used to set-up the instrument for a number of specific types of test. There are two modes BASIC and ADVANCED and nine other specific types of test that automatically configure the instrument on selection from the TAS K menu. The tasks available under the TAS K menu are held on three pages shown below. To select a task from the menu, press the TASK key, position the cursor over the required task and press the ENTER key as shown below. Use the right/left cursor keys to switch between pages.
Using Task Functions
Specific tasks are selected as shown above. The following diagrams show how to connect the unit under test (UUT) for each task selectable under the TA SK menu.
Input and output units, where applicable, can be selected by pressing either the INPUT or OUTPUT keys as shown below. The output key also provides the facility of turning the 24V output supply ON and OFF. When not in use, the 24V supply output should be turned OFF to conserve battery power.
Set Units
Note: If the four units displayed are not the units required, press SETUP, select SETTINGS
and refer to page 38.
Set 24 Volts
K0415 Issue No. 2 10
0.00
2000.00
OUTPUT
MAX. ERROR
ERROR TYPE
CHANGE
VALUE
psi
=
=
=
=
=
mA
4.00
20.00
LINEAR
0.05%
%span
PI
SELECT VALUE
psi
mA
TASK : P-I DATALOG
24V OFF
..........
. %span FAIL
TURN OFF CAL MODE
TURN ON
CAL MODE
mA
TASK : P-I DATALOG
24V OFF
..........
psi

Task Selection

Cal mode (DPI 615 versions only)
Cal mode, which is available in tasks P-I, P-P, P-V, P-P, P-DISPLAY and P-SWITCH, provides a method of setting-up test parameters manually. Downloaded test procedures can also automatically set-up and turn on the Cal Mode function. The method of turning on and setting-up Cal Mode is shown below for a P-I task. The method is similar for all the other tasks available in Cal Mode.
Pressing the F1 key (TURN ON CAL MODE), provides the set-up screen for the CAL mode. Initially, the cursor is placed in the UUT SPAN field to allow the required span range to be entered. The corresponding values for the UUT output parameter (current) are then set, followed by the maximum error value and error type (% rdg or % span). When all test parameters have been set-up, the screen changes to display the input and output and the test results. The test result can only be displayed to within a range of ±9.99%. If the test result is outside this range, either the left pointing (-ve error) or right pointing (+ve error) chevrons are displayed. Within this error band, the actual tolerance value is displayed. Test results can either be stored as snapshots or logged as datalog files, depending upon how the instrument has been set-up.
Basic Mode (Task BASIC)
This instrument powers up in this mode the first time that it is used. To select BASIC from any other task, press the TASK key and select BASIC and press the ENTER key. BASIC mode is fully described in the Getting Started, section (see page 7).
11 K0415 Issue No. 2
P
I
+
-
F2F1
bar
mA
TASK : P-I SNAPSHOT MODE
24V OFF
DPI 615
A
B
C
F2F1
P
V
-+
+
-
+
-
bar
V
CURRENT
PRESSURE
UNITS
TASK: BASIC
DPI 615
A
B
C

Taking Measurements

Pressure Transmitter (P-I) Task
Select the P-I task from the task menu and connect the Unit Under Test (UUT) to the calibrator as shown below.
A - External pressure source (indicator instruments only) B - Pressure regulator C - Pressure to current 24V device
If required, select the output units as described on page 10.
If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
Voltage Output Pressure Transmitter (P-V) Task
Select the P-V task from the task menu and connect the UUT to the calibrator as shown below. Voltage output transducers with a 10 V supply and outputs of ±10 V can be calibrated using the 10V output sockets (supply) and the transducer output connected to a voltmeter (refer to Method on page 13 for a test procedure).
A - External pressure source (indicator instruments only) B - Pressure regulator C - Excitation 10V
If required, select the output units as described on page 10.
If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
K0415 Issue No. 2 12
A
B
C
D
E

Taking Measurements

Pressure Converter (Pressure to Pressure) Task
Select the P-P task from the task menu and connect the UUT to the calibrator as shown below. Testing a converter requires one pressure to be applied to the unit under test (UUT) and another (converter output) to be measured. The additional measurement is provided by the external transducer option.
Method
Connect the UUT to the calibrator and plug the external transducer into the
calibrator as shown below.
A - External pressure source (indicator instruments only) B - Pressure regulator C - External pressure supply D - Pressure to pressure device E - External transducer
Press the TASK key and select the P-P task. Providing the external transducer has
been calibrated and its parameters stored in the calibrator, the display shows
External pressure in the input window and calibrator Output pressure in the output
window. If an error message “NO SENSOR OR CAL INVALID” is displayed, this
indicates that the external transducer has not been entered and/or calibrated with
the calibrator. Refer to page 45 for details of adding an external transducer. If an
external transducer change is made, switch the calibrator off and then on to load
new transducer data.
If required, select the input and output units as described on page 10.
If applicable, turn on Cal Mode and set-up test parameters as detailed on page 11.
Note: Match pressure ranges to give required accuracy and avoid overpressure. If
external pressure is required as the output and internal pressure the input, use ADVANCED mode for this set-up.
13 K0415 Issue No. 2
P
I
+
-
F2F1
bar
OUTPUT
NEW VALUE
mA
OK
TASK : I-P SNAPSHOT MODE
24V OFF
DPI 615
TURN ON
CAL MODE
A
B
F2F1
bar
RUN
CONTACTSTATE
TASK : P-SWITCH SNAPSHOT MODE
24V OFF
DPI 615
A
B
C

Taking Measurements

Current to Pressure Converter (I-P) Task
A - External pressure supply B - Pressure to current device (24V)
•Use the up and down cursor keys to adjust the loop current to the required
value. Alternatively, press ENTER and use cursor keys to enter a finite value. Cursor
keys can then be used to nudge the output either up or down. If required, change
pressure units with INPUT key.
Pressure Switch Test (P-SWITCH) Task
A - External pressure source (indicator instruments only) B - Pressure regulator C - Pressure switch under test
Contact state will be shown on display. When contacts close, buzzer sounds.
To run switch test, close vent valve and press the RUN (F1) key.
Using the hand-pump, increase the applied pressure to just below the switch
operating point. Screw the volume adjuster in until the switch operates (the display
then shows the operating pressure of the switch).
Reduce pressure until the switch releases (indicated by the switch symbol). The
display then shows the release pressure and the hysteresis value.
K0415 Issue No. 2 14
F2F1
bar
RUN
CONTACTSTATE
TASK : P-SWITCH SNAPSHOT MODE
24V OFF
DPI 615
A
B
C
INPUT
SELECT
RESISTANCE
SELECT
ZERO
SELECT
FUNCTION
RESIST
OFF
ZERO
SELECT
FUNCTION
RESIST
ON
ZERO
Connect Vin +, SW + and mA in + together.
Connect Vin +, SW + and mA in
+ together.
Connect switch test leads across switch symbol.
Connect switch
test leads
across switch
symbol.
Open circuit test leadsOpen circuit test leads
Press ENTER to continue.Press ENTER
to continue.
Switch Resistance input now ready for use.
Switch Resistance input now
ready for use.
Press ENTER to continue.Press ENTER
to continue.
Open circuit test leads.Open
circuit test leads.
Now short test leads.Now short test leads.
Press ENTER to continue.Press ENTER
to continue.
bar
RUN
CONTACT STATE
TASK : P-SWITCH SNAPSHOT MODE
24V OFF
24V OFF
RUN
bar
CONTACT STATE
CLOSED AT 15.060 bar OPENED AT 14.059 bar HYSTERESIS 1.001 bar RESISTANCE 3 m
W
TASK: P-SWITCH SNAPSHOT MODE

Taking Measurements

Pressure Switch Testing with Contact Resistance Measurement
A - External pressure source (indicator instruments only)B - Pressure regulator C - Pressure switch under test
To perform switch test with contact resistance measurement, select P-SWITCH and proceed as follows:
To ensure accurate measurements it is recommended that the zero procedure (that compensates for the resistance of the test leads) is carried out before performing this test.
Note: Allow sufficient time after contact closure for the resistance to stabilize.
The switch test is performed in the same way as the previous section, except that the contact resistance is measured and displayed with the results.
15 K0415 Issue No. 2
F2F1
bar
DISPLAY
CHANGE
VALUE
TASK : P-DISPLAY LOG FILE: NONE
24V OFF
TURN ON
CAL MODE
DPI 615
A
B
C
ENTER VALUE
DISPLAY
CHANGE
VALUE
DISPLAY
CHANGE
VALUE
DISPLAY
LOG NOW
STOP
LOGGING
STORE
+
bar bar
bar
TASK: P-DISPLAY LOG FILE: A1
TASK: P-DISPLAY LOG FILE: A1
TASK: P-DISPLAY LOG FILE: A1
24V OFF 24V OFF 24V OFF
DISPLAY
bar
LOG NOW
STOP
LOGGING
STORE
FINAL POINT
TASK: P-DISPLAY LOG FILE: A1
24V OFF

Taking Measurements

Pressure to Display (P - Display) Task
P-Display is a special application of Datalog. To use this mode, select Datalog from the Store Mode menu as detailed on page 37. Connect the UUT to the instrument as shown below and, if required, turn on and set-up Cal Mode (see page 11).
A - External pressure source (indicator instruments only) B - Pressure regulator C - Dial gauge under test
Press
TASK
and select
P-DISPLAY
. If required, use
OUTPUT
key to change pressure units.
Set-up a data log file as detailed on page 31.
Note: TRIGGER field, automatically set to KEYPRESS, cannot be changed.
Apply a series of test pressures to the device under test. Enter displayed reading at
each pressure and log each point:
After logging final test point, terminate as
follows:
K0415 Issue No. 2 16
A
B
C
PRESSURE INT
bar
F2F1
RUN
CHANGE
VALUE
WAIT DURATION
START PRESS STOP PRESS
PRESS CHANGE LEAK RATE
secs secs
bar bar
bar bar/m
60
60
TASK : LEAK TEST SNAPSHOT MODE
24 V OFF
DPI 615
SET WAIT TIME SET DURATION
CION
WAIT DURATION
START PRESS STOP PRESS
PRESS CHANGE
180
LEAK RATE
secs secs
bar bar
bar bar/m
60
PRESSURE INT bar
CHANGE
VALUE
RUN
TASK : LEAK TEST SNAPSHOT MODE
CION
WAIT DURATION
START PRESS STOP PRESS
PRESS CHANGE
180
LEAK RATE
secs secs
bar bar
bar bar/m
60
PRESSURE INT bar
CHANGE
VALUE
RUN
TASK : LEAK TEST SNAPSHOT MODE
24 V OFF
24 V OFF

Taking Measurements

Leak Test (LEAK TEST) Task
A - External pressure source (indicator instruments only) B - Pressure regulator C - Unit/system under test
If required, use the INPUT key to change pressure units.
Set-up the leak test WAIT and DURATION times to the required values as shown
below. Recommended minimum wait period - 3 minutes.
Close the vent valve and pressurize the unit/system to the required LEAK TEST
pressure.
•Press the RUN (F2) key to start the leak test. When completed, the beeper sounds
and the leak test results are written to the display.
17 K0415 Issue No. 2
Loading...
+ 63 hidden pages