This Product will perform in conformity with the description thereof contained in this
operating manual and accompanying labels and/or inserts, when assembled, operated,
maintained and repaired in accordance with the instructions provided. This Product
must be checked periodically. A defective Product should not be used. Parts that are
broken, missing, plainly worn, distorted or contaminated should be replaced
immediately. Should such repair or replacement become necessary, Datex-Ohmeda
recommends that a telephonic or written request for service advice be made to the
nearest Datex-Ohmeda Field Service Support Center. This Product or any of its parts
should not be repaired other than in accordance with written instructions provided by
Datex-Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be
altered without the prior written approval of Datex-Ohmeda’s Quality Assurance
Department. The user of this Product shall have the sole responsibility for any
malfunction which results from improper use, faulty maintenance, improper repair,
damage, or alteration by anyone other than Datex-Ohmeda.
CAUTIONw U. S. Federal and Canadian law restrict this device to sale by or on the order of
a licensed medical practitioner. Outside the U. S. A. and Canada, check local
laws for any restrictions that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a
product group code, the year of manufacture and a sequential unit number for
identification.
AAA A 12345
This alpha character indicates the year of product manufacture
and when the serial number was assigned; “Y” = 1995, “Z” = 1996,
“A” = 1997, etc. “I” and “O” are not used.
Fig.17 Tec 5 Vaporizer Gas Flow.....................................................................20
Contents 2Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 6
1.0Introduction
1.1Operation And Maintenance Manual
This Operation And Maintenance (O&M) Manual contains the information required in order to install,
operate and maintain the Tec 5 Continuous Flow Vaporizer.
Requests for servicing facilities, advice or assistance must be addressed either to a local Datex-Ohmeda
Field Operations Unit or to an Datex-Ohmeda Authorised Distributor.
Additional copies of this manual, quoting Tec 5 Continuous Flow Vaporizer O & M Manual Part No.
1105-0100-000, can be requested from a local Datex-Ohmeda Field Operations Unit or from an DatexOhmeda Authorised Distributor.
It is recommended that all relevant documentation, including the O & M Manual and accompanying
labels and/or inserts, is immediately available to all prospective operators.
1.2Precautions
A number of Warnings and Cautions are used throughout this manual to draw attention to the possible
hazards and/or adverse conditions which may occur if the information and instructions provided are not
strictly observed.
Cautions and Warnings are preceded by a w symbol and are used to draw attention to a condition
which can endanger either the patient or the operator and can result in damage to the equipment.
Special attention must be paid to each Warning and Caution as it appears in the manual.
1.3Symbols
Systems with this mark agree with the European Council Directive (93/42/EEC) for
Medical Devices when they are used as specified in their operation and maintenance
manuals. The xxxx is the certification number of the Notified Body used by
Datex-Ohmeda's Quality Systems.
European Union Representative
Tec 5 VaporizerPage 1
O & M Manual Part No. 1105-0100-000August 1999
Page 7
2.0Servicing Policy
During initial installation of a Tec 5 vaporizer, the Next Service Date label located on the vaporizer dial
must be completed by the Technician/Engineer/Customer responsible for the installation.
Datex-Ohmeda recommends that all Tec 5 Vaporizers are serviced every three years, irrespective of
conditions of use.
Servicing procedures for this Product must be performed by Datex-Ohmeda trained personnel in
accordance with written instructions provided by Datex-Ohmeda.
wWarning: Only Technicians/Engineers trained and certificated by Datex-Ohmeda to repair and/or
service the Tec 5 Vaporizer should attempt to repair and/or service it and it must be repaired and/or
serviced at an Datex-Ohmeda Authorised Service Centre in accordance with written instructions
provided by Datex-Ohmeda. Detailed information for more extensive repairs is included in the Tec 5
Vaporizer Service Centre Manual which is available only to Authorised Service Centres.
Warranty repair and service procedures must be performed at an Datex-Ohmeda Authorised Service
Centre. A Datex-Ohmeda Service Representative can be contacted at the nearest Datex-Ohmeda Field
Operations Unit or Datex-Ohmeda Authorised Distributor.
Do not use malfunctioning equipment. If any assistance is required, contact the nearest Datex-Ohmeda
Field Operations Unit or Datex-Ohmeda Authorised Distributor.
If the equipment is to be transported to the nearest Datex-Ohmeda Field Operations Unit, drain the
vaporizer, package it securely for protection in its original packaging and ship it prepaid. Enclose the
following items as applicable:
1. A letter describing in detail any difficulties experienced with the equipment.
2. Warranty information, such as a copy of the invoice or other applicable documentation.
3. Purchase order number to cover repair of equipment not under warranty.
4. Ship to and bill to information.
5. The name and telephone number of the person to contact.
Page 2 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 8
3.0Description
3.1General
wWarning: This manual and all its associated documentation must be studied thoroughly before any
attempt is made to install, operate or maintain any part of the Tec 5 Vaporizer.
The Tec 5 Vaporizer is designed for out of circuit use in continuous flow techniques of inhalation
anaesthesia.
The vaporizer is temperature, flow and pressure compensated so that its output remains relatively
constant despite cooling due to evaporation, variations in inlet flow and fluctuating pressures as
described in Section 7.2 Effect of Variables.
Each vaporizer is agent specific and is clearly labelled with the name of the anaesthetic agent for which
it is designed.
The vaporizer is designed to be used on Selectatec Series Mounted Manifolds. The vaporizer can be
installed on other Selectatec Manifolds but the interlock system is designed to function on Selectatec
Series Mounted Manifolds only.
Mounting a Tec 5 vaporizer on a Selectatec 7 Compatibility Block is not recommended.
"It is a requirement of European Standard prEN 740 - Anaesthetic Workstations And Their Modules, that
an appropriate gas monitor is used to monitor the concentration of anaesthetic agent vapour in the
inspiratory gas when the vaporizer is in operation, in order to provide protection against hazardous
output in the event of a device malfunction".
wWarning: If the vaporizer has been inverted, connect it to a gas scavenging system, set the dial to
5% and purge the vaporizer with the carrier gas at 5 litres/minute for 5 minutes.
wCaution: Although the vaporizer incorporates an integral non-spill system, the vaporizer is intended
to be operated in its upright position.
wCaution: Turn the Vaporizer OFF when it is not in use.
3.2Control Dial
A single control dial with a concentration scale calibrated in percentage of anaesthetic agent vapour per
total volume (% v/v) is employed to set the desired concentration of the anaesthetic agent.
A dial release is incorporated in the dial assembly to prevent accidental displacement of the control dial
from the OFF position. To select an ON setting it is necessary to pull in the dial release and
simultaneously rotate the dial counter-clockwise.
The dial and dial release are designed to enable an ON setting to be selected using only one hand.
3.3Safety Interlocks
The vaporizer incorporates an interlock mechanism. This mechanism also interfaces with the Selectatec
Series Mounted Manifold to comprise an interlock system which is designed to help ensure that the
following criteria are satisfied:
1. The vaporizer must be locked on to the manifold before it can be turned ON.
2. Only one vaporizer at a time can be turned ON when two or more Tec 4, Tec 5 or Tec 6 vaporizers
are fitted on a Selectatec Series Mounted Manifold.
3. The gas flow enters only the vaporizer which is turned ON.
4. Any unwanted anaesthetic trace vapour is minimised after a vaporizer is turned OFF.
Tec 5 VaporizerPage 3
O & M Manual Part No. 1105-0100-000August 1999
Page 9
wWarning: Earlier versions of the Selectatec Series Mounted Manifold which provide mounting
positions for three vaporizers require that if only two vaporizers are fitted then the centre position
must be occupied. If the centre position is not occupied the interlock which helps to ensure that only
one vaporizer at a time can be turned ON is ineffective.
Later versions of the Selectatec Series Mounted Manifold which provide mounting positions for three
vaporizers incorporate an additional interlock which helps to ensure that only one vaporizer at a time
can be turned ON even if the centre position is not occupied.
3.4Vaporizer Identification Label
A vaporizer identification label is affixed
to the back panel of the vaporizer as
illustrated on Fig. 1.
AA.13.051
An anaesthesia system fitted with a
vaporizer identification unit uses this label
to identify the vaporizer type.
Do not affix any additional labels or
markings to the back panel, they may
adversely affect the operation of the
vaporizer identification unit.
4.0Operating Instructions
4.1General
The vaporizer is designed to be used on Selectatec Series Mounted Manifolds. The vaporizer can be
installed on other Selectatec Manifolds but the interlock system is designed to function on Selectatec
Series Mounted Manifolds only.
T
E
C
5
S
E
V
O
F
L
U
R
A
N
E
Fig. 1 Vaporizer Identification Label
VAPORIZER
NAME TEC 5
IDENTIFICATION
CODE
AGENT NAME
VAPORIZER
IDENTIFICATION
LABEL
Mounting a Tec 5 vaporizer on a Selectatec 7 Compatibility Block is not recommended.
wWarning: It is a requirement of European Standard prEN 740 - Anaesthetic Workstations And Their
Modules, that the gas monitoring device referred to at 3.1 of this manual is in operating condition,
by being enabled and functioning , prior to use of the vaporizer.
wWarning: Handle the vaporizer with care at all times. Do not lift or support the vaporizer by holding
the control dial.
wWarning: Before mounting a vaporizer on to the manifold ensure that only one 'O' ring is fitted to
each manifold port valve. Ensure that each 'O' ring is intact and that there is no foreign matter
around the mating surfaces. A damaged 'O' ring and/or foreign matter around the mating surfaces
can cause leaks.
wWarning: Do not use a vaporizer if the liquid level decreases to below the minimum level.
wWarning: Before using a vaporizer allow it to attain the ambient temperature of the location in
which it is to be used.
Page 4 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 10
4.2Vaporizer Mounting Procedure
1. Remove the red coloured dust caps fitted to the vaporizer interlock block ports, illustrated on Fig. 2.
2. Check that each port valve 'O' ring is intact and not damaged. Renew each 'O' ring if necessary.
3. Make sure that the vaporizer control dial is in the OFF position.
4. Make sure that the vaporizer locking lever is at the unlocked position, that is, in the up position.
CHECK DIAL IS AT OFF
CHECK INTERLOCK BLOCK
PORTS ENGAGE CORRECTLY
WITH MANIFOLD PORT VALVES
CHECK LOCKING LEVER
IS UNLOCKED (UP)
REMOVE DUST CAPS
FROM INTERLOCK BLOCK
PORTS
CHECK EACH MANIFOLD
PORT VALVE 'O' RING. FIT
NEW 'O' RING(S) IF
NECESSARY
AA.13.052
Fig. 2 Vaporizer Mounting Procedure
5. Carefully lower the vaporizer on to the manifold so that the vaporizer interlock block covers the two
manifold port valves with the interlock block ports correctly engaged with the manifold port valves.
wCaution: Push the locking lever down to the full extent of its travel before attempting to turn it. The
mechanism may be damaged if an attempt is made to turn the lever before it is at the full extent of
its vertical travel.
6. Press down the locking lever 1 to the full extent of its travel and then turn it clockwise 2 to the
locked position 3 to lock the vaporizer on to the manifold as illustrated on Fig. 3.
PUSH DOWN
TO FULL EXTENT
LOCKING
LEVER
DIAL
RELEASE
Tec 5 VaporizerPage 5
O & M Manual Part No. 1105-0100-000August 1999
1
TURN
2
LOCKING
LEVER
DIAL
RELEASE
3
AA.13.053
LOCKED
Fig. 3 Locking The Vaporizer On To A Manifold
Page 11
4.3Checking For Correct Mounting Of The Vaporizer
wWarning: To help to ensure correct operation, do not use a vaporizer which is visibly out of line on the
manifold or which can be lifted off the manifold when the locking lever is in the locked position.
1. If more than one vaporizer is fitted, visually check to make sure that the tops of vaporizers are square
to the manifold. If a vaporizer is visibly out of line, remove it from the manifold and attempt to mount it
correctly. Do not use a vaporizer which cannot be mounted correctly.
2. When the vaporizers appear to be level and their locking levers are in the locked position, attempt to lift
each vaporizer from the manifold.
3. If a vaporizer can be lifted OFF the manifold, remove it from the manifold and attempt to mount it
correctly. Do not use a vaporizer which cannot be mounted correctly.
4. Make sure only one vaporizer at a time can be turned ON, thus checking that the interlock rods are in
alignment.
5. Check the anaesthesia system for leaks in accordance with the relevant O&M Manual with the vaporizer
dial turned to 0% and then repeat the check with the vaporizer dial turned OFF.
4.4Turning ON The Vaporizer
Note: Do not attempt to turn the dial if the vaporizer is
not properly locked on to the manifold.
1.Pull in the dial release illustrated on Fig. 4 and
rotate the dial counter-clockwise. Note that it is
not possible to turn ON the vaporizer if an
adjacent Tec 4, Tec 5 or Tec 6 vaporizer is turned
ON.
2. The vaporizer should not be used between OFF
and the first graduation mark. To avoid inadvertent
delivery of small concentrations the control dial
should be turned OFF when the vaporizer is not in
use.
4.5Vaporizer Removal
1. Rotate the dial clockwise to turn OFF the vaporizer. If it is not completely turned to the OFF position it
cannot be released from the manifold.
2. Referring to Fig. 5, unlock the vaporizer from the manifold by turning the locking lever
clockwise to the unlocked position. Release the locking lever 2, check that the locking lever springs
up to the unlocked condition 3 and then carefully lift the vaporizer straight up and off the manifold.
LOCKING
LEVER
1
TURN
2
RELEASE
AA.13.054
Fig. 4 Turning ON The Vaporizer
UNLOCKED
LOCKING
LEVER
3
counter-
1
AA.13.053
Fig. 5 Unlocking The Vaporizer From A Manifold
Page 6 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 12
4.6Filling And Draining
4.6.1General
wWarning: Do not fill the vaporizer with any agent other than the agent specified on the front label.
The vaporizer is designed for that agent only. Filling the vaporizer with any agent other than that
specified can prove to be dangerous to a patient.
wWarning: Do not fill the vaporizer unless the control dial is in the OFF position.
wWarning: Do not turn the dial ON during filling or attempt to fill beyond the ¥ mark.
wWarning: Do not drain the agent into any container other than a properly marked drug container.
wWarning: Ensure that the filler cap is tightened prior to use.
Periodically check the agent level. The vaporizer should be refilled at appropriate intervals. The
vaporizer is designed to function according to specification as long as there is agent visible above
the mark.
The vaporizer must be filled and used in an upright position. Small deviations from the upright position
do not affect either the output or the safety of the vaporizer, but because the agent depth is shallow in
relation to the diameter of the vaporizing chamber, more frequent checks of the agent level should be
performed to avoid a misleading impression of the amount of agent in the vaporizer when small
deviations from the upright position occur.
At intervals, ideally not exceeding two weeks, the vaporizer should be drained into the correct drug
bottle when the agent is low. This helps to preserve the drug purity by removing oxidized impurities,
accumulated contaminants and stabilisers. Discard the agent in a manner consistent with local policies
and guidelines. Less frequent intervals may be used when the anaesthetic agent does not contain
additives or stabilising agents.
4.6.2Filling A Vaporizer Which Incorporates A Screw Cap Filler
wWarning: Before filling a vaporizer fitted with a screw cap filler, turn the cap slowly to allow any
pressure to vent slowly.
1. Turn the vaporizer dial clockwise to the OFF position.
FILLER CAP
HEXAGON
AA.13.024
SIGHT GLASS
DRAIN PLUG
Fig. 6 Filling A Vaporizer Which Incorporates A Screw Cap Filler
Tec 5 VaporizerPage 7
O & M Manual Part No. 1105-0100-000August 1999
Page 13
2. Remove the filler cap by turning it counter-clockwise. Ensure that the drain plug is closed by
tightening it with the hexagonal end of the filler cap.
3. Verify that the anaesthetic agent is the same as that specified on the vaporizer front label. Observe
the agent level through the sight glass indicator on the side of the filler body and pour the agent
slowly into the filling port, as illustrated on Fig. 6, until the level reaches the ¥ mark. The level may
decrease slightly as the wicks absorb the agent.
4. Replace the filler cap and tighten it to minimise any possibility of leaks.
4.6.3Draining A Vaporizer Which Incorporates A Screw Cap Filler
wCaution: Do not allow a bottle to become completely full during draining procedures.
The vaporizer must only be drained into a properly marked container, as follows:
1. Remove the filler cap and insert the hexagonal end of the cap into the drain plug situated below the
filling port on the filler body as illustrated on Fig. 7.
2. Position a properly marked container under the drain spout.
3. Unscrew but do not remove the drain plug to allow the vaporizer contents to pour from the drain
spout into the container.
4. After draining is complete, tighten the drain plug to minimise any possibility of leaks.
5. Replace the filler cap and tighten it to minimise any possibility of leaks.
FILLING PORT
AA.13.028
DRAIN PLUG
HEXAGONAL
END
FILLER CAP
DRAIN SPOUT
CONTAINER
Fig. 7 Draining A Vaporizer Which Incorporates A Screw Cap Filler
Page 8 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 14
4.6.4Filling A Vaporizer Which Incorporates A Keyed Filler
wCaution: Always hold the bottle below the level of the end of the bottle adaptor and also below the
level of the filler port until the adaptor is clamped into position.
wCaution: If it is suspected that there is liquid in the adaptor air tube, remove the bottle adaptor from
the bottle and carefully shake the adaptor two or three times to clear the tube.
The filling system consists of three elements, the bottle collar, the bottle adaptor and the filling and
draining unit fitted to the vaporizer. The vaporizer must only be filled using the correct agent specific
filling system, as follows:
1. Screw the agent specific bottle adaptor firmly on to the bottle.
2. Ensuring that the bottle (1) is held below the level of the end of the bottle adaptor and below the
level of the filler port, insert the end of the bottle adaptor into the filler port (2) as illustrated on
Fig. 8.
3. Tighten the locking clamp by pulling the locking clamp lever forward and downward (3) until the
bottle adaptor is clamped.
wCaution: When it is set correctly, the clamp secures the bottle adaptor without using the full travel
of the locking lever. The lever must not be forced to its lowest limit of travel.
4. Raise the bottle above the level of the filler port (4) and then open the filler port valve by pulling the
port valve lever forward (5) to its full extent.
AA.13.045
AA.13.046
BOTTLE
3
4
5
2
AGENT SPECIFIC
BOTTLE ADAPTOR
6
LOCKING
CLAMP
LEVER
LOCKING
CLAMP
1
LEVER
PORT
VALVE
LEVER
AGENT SPECIFIC
BOTTLE COLLAR
021A T5 OM
AGENT SPECIFIC
BOTTLE COLLAR
Fig. 8 Filling A Vaporizer Which Incorporates A Keyed Filler
Tec 5 VaporizerPage 9
O & M Manual Part No. 1105-0100-000August 1999
Page 15
Note: The vaporizer normally fills in less than one minute. If the vaporizer is dry the level decreases
slightly as the wick absorbs the agent.
5. When the vaporizer is full, return the port valve lever (5) to the closed condition to close the valve.
6. Lower the bottle to below the level of the filler port (6) and allow any agent in the tube to drain
back into the bottle.
7. Release the locking clamp lever, return it to the up position and then remove the bottle adaptor
from the filler port.
4.6.5Draining Vaporizer Which Incorporates A Keyed Filler
wCaution: Do not allow a bottle to become completely full during draining procedures.
The vaporizer must only be drained into a properly marked container, as follows:
1. Position the bottle below the level of the filler port and insert the end of the bottle adaptor into the
filler port (1) as illustrated on Fig. 9.
2. Tighten the locking clamp by pulling the lever forward and downward (2) until the adaptor is
clamped.
3. Open the filler port valve by pulling the lever forward to its full extent (3).
4. Hold the bottle below the level of the filler port to allow the vaporizer contents to flow into the
bottle (4).
5. When the vaporizer is drained, close the port valve.
6. Release the locking clamp and return it to the up position.
7. Remove the bottle adaptor from the port.
AA.13.026
AA.13.027
2
3
1
AGENT SPECIFIC
BOTTLE ADAPTOR
LOCKING
CLAMP
LEVER
4
FILLER
PORT
VALVE
LEVER
LOCKING
CLAMP
LEVER
022A T5 OM
AGENT SPECIFIC
BOTTLE COLLAR
Fig. 9 Draining A Vaporizer Which Incorporates A Keyed Filler
Page 10 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 16
4.6.6Filling A Vaporizer Which Incorporates A Quik-Fil™ Filler
wWarning: Ensure that the drain plug screw (1), located on the lower front of the vaporizer, is
correctly tightened to prevent loss of liquid agent.
1. Remove the protective cap from the anaesthetic agent bottle filler, checking that the bottle and filler
mechanism are not damaged.
2. Remove the filler cap (2) and insert the bottle nozzle into the filler block. Rotate the bottle to align
the bottle filler nozzle keys (3) with the index slots (4) in the filler block as illustrated on Fig. 10.
3. Press the agent bottle fully into the vaporizer filler block. Allow the liquid to flow into the vaporizer
until the maximum level mark ¥ is reached, paying particular attention to the level in the sight
glass and the air return bubbles flowing into the bottle.
4. Release the bottle when the vaporizer is full and the continuous stream of bubbles ceases.
5. Withdraw the bottle from the vaporizer filler and replace the filler cap (2) and the cap on the agent
bottle. Ensure that the filler cap is tightened to minimise any possibility of leaks.
AA.13.056
3
4
2
1
Fig. 10 Filling A Vaporizer Which Incorporates A Quik-Fil™ Filler
Tec 5 VaporizerPage 11
O & M Manual Part No. 1105-0100-000August 1999
Page 17
4.6.7Draining A Vaporizer Which Incorporates A Quik-Fil™ Filler
wCaution: Do not allow a bottle to become completely full during draining procedures.
The vaporizer must only be drained into a properly marked container, as follows:
2
1. Remove the cap from the vaporizer
filler.
2. Remove the protective cap from an
empty bottle. Insert the bottle nozzle
into the drain funnel. Rotate the
bottle to align the index slots in the
drain funnel (1) with the bottle filler
nozzle keys (2) and screw the drain
funnel onto the empty bottle as
illustrated on Fig. 11.
3. Fully insert the drain funnel into the
keyed drain slot (3) in the bottom of
the vaporizer filler block.
4. Unscrew the drain plug (4) with the
key (5). Continue to drain the
vaporizer until empty.
3
1
AA.13.057
5. After draining is complete, tighten
the drain plug to minimise the
possibility of leaks and withdraw
the drain funnel.
6. Replace the filler cap and tighten it
to minimise the possibility of leaks.
7. Unscrew the drain funnel from the
bottle and refit the protective cap.
Note: If further drain funnels are required
they should be obtained from the local
Abbott Laboratories representative.
Fig. 11 Draining A Vaporizer Which Incorporates A Quik-Fil™ Filler
AA.13.058
4
5
AA.13.059
Page 12 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 18
5.0Maintenance
wWarning: Do not modify, tamper with or disassemble the vaporizer because of the dangers of
damaging the unit and altering the accuracy of graduation.
Observation of the instructions previously provided, regular servicing and normal professional vigilance
is normally all that is required to maintain the vaporizer in a safe working condition.
5.1Schedule
5.1.1Every Two Weeks
When the agent level is low, drain the contents of the vaporizer into an appropriately marked container
and discard the agent. Less frequent intervals may be used when the anaesthetic agent does not
contain additives or stabilising agents, but the procedure must be performed at least once every year.
5.1.2Every Three Years
Datex-Ohmeda recommends that all Tec 5 Vaporizers are serviced every three years, irrespective of
conditions of use.
The vaporizer should be serviced at an Datex-Ohmeda authorised Service Centre.
This service includes the following:
1. Complete disassembly of components.
2. Thorough cleaning.
3. Inspection for damage and wear.
4. Renewal of wicks, seals and damaged or worn components.
5. Where appropriate, replacement of discontinued parts with more up-to-date parts.
6. Lubrication where necessary.
7. Checks of delivered vapour concentrations under closely defined conditions at different
temperatures and flows with regraduation or adjustment where necessary.
5.2Cleaning
wWarning: Do not put water or any other solvent into a vaporizer. A vaporizer should be filled with the
specified anaesthetic agent only.
wWarning: Do not immerse the vaporizer in any liquid, including water.
wWarning: Do not autoclave the vaporizer.
Clean the exterior of the vaporizer with a damp cloth.
Never allow cleaning agents to accumulate either in the filler, the gas inlet and outlet ports or around
the control dial.
Tec 5 VaporizerPage 13
O & M Manual Part No. 1105-0100-000August 1999
Page 19
5.3Contamination
If the vaporizer is either filled or partly filled with an incorrect agent or any other contaminant, water for
example, proceed as follows:
1. Drain the vaporizer and discard all liquid.
2. Return the vaporizer to a Service Centre stating, if possible, the type of contaminant used.
5.4Repairs
Repairs must only be performed by the Manufacturer's Service Representatives.
Page 14 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 20
6.0Principle Of Operation
6.1Interlock Mechanism
The vaporizer locking lever, illustrated on Fig. 12, is interlocked with the vaporizer percentage control dial in
such a way that the control dial release, located at the rear of the dial, cannot be actuated until the vaporizer
locking lever is in the locked position.
With the vaporizer locking lever in the locked position the dial release can be pressed in towards the dial to
operate the interlock mechanism which simultaneously actuates the following:
1. The interlock extension rods, which extend to minimise the possibility of adjacent vaporizers being
turned ON.
2. The two port valve actuating spindles, which actuate the Selectatec Manifold port valves to allow fresh
gas to flow through to the vaporizer.
3. The vaporizer percentage control dial release, which allows the dial to be turned to select the required
percentage.
Turning the control dial to OFF automatically reverses the operating sequence, which allows the dial release
to move out to lock the dial in the OFF position, closes the manifold port valves and vents the vaporizer gas
connecting ports and retracts the extension rods to allow an adjacent vaporizer to be turned ON.
Turning the locking lever to the unlocked position releases the vaporizer allowing it to be removed from the
manifold.
PERCENTAGE
CONTROL DIAL
RELEASE
VAPORIZER
LOCKING LEVER
PORT VALVE
ACTUATING
SPINDLE
PORT
VALVES
CONTROL DIAL
EXTENSION
ROD
PORT
VALVES
AA.13.020
SELECTATEC
SERIES
MOUNTED
MANIFOLD
Fig. 12 Vaporizer Interlock Mechanism
Tec 5 VaporizerPage 15
O & M Manual Part No. 1105-0100-000August 1999
Page 21
When two vaporizers are locked on to a manifold and both are turned OFF as illustrated on Fig. 13, each set
of interlock extension rods is retracted allowing either vaporizer to be turned ON.
VAPORIZER B
012A T5 OM
INTERLOCK
EXTENSION
ROD RETRACTED
VAPORIZER A
AA.13.019
INTERLOCK
EXTENSION
ROD RETRACTED
Fig. 13 Two Vaporizers Locked On To The Manifold - Both Turned OFF
Operating the dial release on vaporizer A, illustrated on Fig. 14, extends the interlock extension rods and
turning the dial to a % setting locks the extension rods in the extended condition to prevent the possibility
of Vaporizer B being turned ON.
The actuating spindles are also operated to open the port valves so that carrier gas can flow through
vaporizer A.
VAPORIZER B
INTERLOCK
EXTENSION
ROD RETRACTED
012A T5 OM
INTERLOCK EXTENSION ROD
EXTENDED TO PREVENT THE
POSSIBILITY OF TURNING
ON VAPORIZER B
ACTUATING
SPINDLE
VAPORIZER A
INTERLOCK
EXTENSION
ROD EXTENDED
AA.13.018
Fig. 14 Two Vaporizers Locked On To The Manifold - Vaporizer A Turned ON
Page 16 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 22
6.2Selectatec Series Mounted Manifold Gas Circuit
Fresh gas is introduced from the flowmeters into the Selectatec Series Mounted Manifold which incorporate
pairs of series-connected, two-way port valves.
When a vaporizer is locked on to a manifold and turned ON, both associated port valves are opened. Fresh
gas from the manifold flows into the vaporizer via the inlet port valve and the gas/agent mixture flows from
the vaporizer into the manifold via the outlet port valve.
When the vaporizer is turned OFF, or if no vaporizer is fitted, each port valve is closed to provide a seal to
atmosphere and also to allow the fresh gas to by-pass the vaporizer via the manifold by-pass circuit.
Vaporizer B illustrated on Fig. 15 is turned OFF, therefore both port valves to which vaporizer B is connected
are closed so that the carrier gas flows through the manifold by-pass circuit to vaporizer A.
VAPORIZER A
VAPORIZER B
(ISOLATED)
FROM
FLOWMETERS
AA.13.021
MANIFOLD
BY-PASS
TO COMMON
GAS OUTLET
Fig. 15 Manifold By-pass Circuit
Vaporizer A is turned ON therefore both port valves to which vaporizer A is connected are open so that
carrier gas flows through the vaporizer to the common gas outlet.
The Selectatec Series Mounted Manifold and the vaporizer connection system previously described forms
an interlock system that is designed to allow only one vaporizer at a time to be turned ON. This manifold
system accepts Tec 4, Tec 5 and Tec 6 vaporizers.
wWarning: Connecting any vaporizer other than a Datex-Ohmeda Tec 4, Tec 5 or Tec 6 vaporizer to the
Selectatec Series Mounted Manifold defeats the interlock system. This action results in the ability to
administer anaesthetic agent mixtures, possibly resulting in patient injury.
Tec 5 VaporizerPage 17
O & M Manual Part No. 1105-0100-000August 1999
Page 23
6.3Vaporizer Valve And Sump Assembly
The vaporizer incorporates a vaporizing chamber and cover assembled between a cover base and an
interlock block, and is enclosed by an outer wrap.
A duct system lies within the cover and a rotary valve is located above the cover. The lower face of the
rotary valve incorporates ducts and a curved vapour control channel.
The internal baffle system is designed to limit any liquid anaesthetic agent from reaching the valve
mechanism if the vaporizer is tilted or even inverted. This system minimises the risk of drug
contaminating the valve mechanism and entering the patient circuit.
The spindle of the rotary valve passes through the interlock block and is connected to the control dial.
In the 0% position the rotary valve closes all gas passages except a link between the inlet and outlet
within the vaporizer. When the dial is turned past 0% to ON, the carrier gas is split into two streams
respectively designated by-pass flow and vaporizing chamber flow.
The vaporizing chamber consists of a woven helical wick which transmits gas flow. Surrounding this
helical wick is a woven outer wick which extends below the helical wick into the liquid agent.
The amount of agent picked up in the vaporizing stream can vary, due either to variation in room
temperature or to the cooling which takes place when the agent is vaporized. Variation of either
temperature or cooling rate can cause changes in the vapour pressures of the anaesthetic agents and
unless some form of compensating device is used the output of the vaporizer for a given flow and dial
setting can vary with time and temperature.
The vaporizer incorporates a temperature compensating device (thermostat), which utilises a bimetallic
strip that deflects according to its temperature to control the resistance offered to the flow of gas
through it. This deflection varies the relative proportions of the vaporizing chamber flow and the bypass flow.
If the temperature of the vaporizer decreases, the thermostat closes, the by-pass flow through the
thermostat decreases and therefore the vaporizing chamber flow increases.
Conversely, if the temperature of the vaporizer increases, the thermostat opens, the by-pass flow
through the thermostat increases and therefore the vaporizing chamber flow decreases.
Thus the output of the vaporizer remains relatively constant under conditions of changing temperature.
Page 18 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 24
6.4Gas Flow Through The Vaporizer
6.4.1General
The output concentration of the Tec 5 vaporizer is regulated by the 'variable flow-split' method
described in the following text and illustrated on Fig. 16 and 17:
A total flow of fresh gas from upstream flowmeters enters the vaporizer at 1 where it is immediately
split into two streams. One stream flows into the fresh gas by-pass circuit and the other stream flows
through the vaporizing chamber where it is enriched with the vapour of the liquid anaesthetic agent.
6.4.2By-pass Circuit
The by-pass circuit includes the gas transfer manifold and also a thermostat assembly which is located
at the base of the vaporizer.
The fresh gas flows through the by-pass circuit vertically downwards from a across the sump baseb , through the thermostat to c and back up the gas transfer manifold via d to e as illustrated on
Fig. 16.
The thermostat deflects according to its temperature to control the resistance offered to the flow of gas
through it. This deflection varies the relative proportions of gas flowing through the by-pass and
vaporizing chamber circuits, as described in Section 6.3.
SHUT-OFF
OPEN
FROM
FLOWMETER
1
a
bcd
Fig. 16 Vaporizer Flow - Schematic Diagram
6.4.3Vaporizing Chamber Circuit
The fresh gas flow through the vaporizing chamber, illustrated on Fig. 17, flows from
sump cover 2 where it is diverted via 3 through the central cavity of the rotary valve and back
through the Intermittent Positive Pressure Ventilation (IPPV) compensating assembly via
Gas now flows from the IPPV assembly via 7 down through the tubular wick assembly where it picks
up anaesthetic vapour and then flows across the base of the vaporizing chamber above the liquid agent
to 8.
ROTARY
VALVE
43
VAPORIZING
CHAMBER
THERMOSTAT
FLOW CONTROL
7
(VAPOUR CHANNEL)
014A T5 OM
10
AA.13.062
TO COMMON
GAS OUTLET
11
e
across the
1
4, 5
and 6.
From the base of the vaporizing chamber the gas/agent mixture flows via 9 through the sump cover to
the proportional radial drug control groove of the rotary valve and then back into the sump cover
bk
where it combines bl with the fresh gas from the by-pass circuit.
The combined total flow then flows out from the vaporizer and via the Selectatec circuitry to the anaesthesia
system common gas outlet.
Tec 5 VaporizerPage 19
O & M Manual Part No. 1105-0100-000August 1999
Page 25
With the vaporizer correctly fitted and turned OFF the vaporizer is isolated from the gas flow.
With the vaporizer correctly fitted and turned ON, the gas flow through the vaporizing chamber circuit is
via 1 through to bl and the gas flow through the by-pass circuit is via a through to e .
ROTARY
VALVE
ON
IPPV
COMPENSATING
ASSEMBLY
VAPOUR
CONTROL
CHANNEL
ENRICHED
FRESH
GAS OUT
11
3
a
1
2
e
COMBINED
FRESH GAS
AND ENRICHED
GAS OUT
10
FRESH GAS
4
BY-PASS
d
FRESH
9
GAS OUT
7
5
6
WICK
ASSEMBLY
VAPORIZING
CHAMBER
b
8
c
AA.13.063
015A T5 OM
THERMOSTAT
Fig. 17 Tec 5 Vaporizer Gas Flow
Page 20 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 26
7.0Performance
7.1Performance Curves
7.1.1Isotec 5%
Effect of Flowrate at 21 ± 2° C with oxygen flowing
5
4
3
% ISOFLURANE
2
1
0
1
0.210
5
Flowrate (litres/minute oxygen)
002A T5 OM
5
4
3
DIAL SETTING
2
1
0.6
0.2
15
5
4
3
% ISOFLURANE
2
1
0
15
Effect of Temperature at 5 litres/minute oxygen flow
20
Temperature ° C
003A T5 OM
5
4
3
DIAL SETTING
2
1
0.6
0.2
30
3525
Tec 5 VaporizerPage 21
O & M Manual Part No. 1105-0100-000August 1999
Page 27
7.1.2Fluotec 5%
5
Effect of Flowrate at 21 ± 2° C with oxygen flowing
004A T5 OM
5
4
3
% HALOTHANE
2
1
0
0.2 1
5
1015
Flowrate (litres/minute oxygen)
Effect of Temperature at 5 litres/minute oxygen flow
4
3
DIAL SETTING
2
1
0.6
0.2
005A T5 OM
5
4
3
% HALOTHANE
2
1
0
15
5
4
3
DIAL SETTING
2
1
0.6
0.2
20
25
3035
Temperature ° C
Page 22 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 28
7.1.3Enfluratec 5%
5
Effect of Flowrate at 21 ± 2° C with oxygen flowing
006A T5 OM
5
4
3
% ENFLURANE
2
1
0
0.2
4
3
DIAL SETTING
2
1
0.6
0.2
1
5
10
15
Flowrate (litres/minute oxygen)
Effect of Temperature at 5 litres/minute oxygen flow
007A T5 OM
5
5
4
4
3
% ENFLURANE
2
1
0
1525
20
30
3
2
1
0.6
0.2
35
Temperature ° C
Tec 5 VaporizerPage 23
O & M Manual Part No. 1105-0100-000August 1999
DIAL SETTING
Page 29
7.1.4Sevotec 5%
5
4
Effect of Flowrate at 21 ± 2° C with oxygen flowing
008A T5 OM
5
4
% SEVOFLURANE
5
3
2
1
0
0.2
1
5
10
3
DIAL SETTING
2
1
0.6
0.2
15
Flowrate (litres/minute oxygen)
Effect of Temperature at 5 litres/minute oxygen flow
009A T5 OM
5
4
3
% SEVOFLURANE
2
1
0
15
20
25
30
4
3
2
1
0.6
0.2
35
Temperature ° C
Page 24 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
DIAL SETTING
Page 30
7.1.5Sevotec 8%
8
7
6
Effect of Flowrate at 21 ± 2° C with oxygen flowing
024A T5 OM
5
4
% SEVOFLURANE
3
2
1
0
0.2
8
8
7
6
5
4
3
2
DIAL SETTING
1
0.6
0.2
1
510
15
Flowrate (litres/minute oxygen)
Effect of Temperature at 5 litres/minute oxygen flow
023A T5 OM
8
7
7
6
5
4
% SEVOFLURANE
3
2
1
0
15
20
2535
30
6
5
4
2
1
3
0.6
0.2
DIAL SETTING
Temperature ° C
Tec 5 VaporizerPage 25
O & M Manual Part No. 1105-0100-000August 1999
Page 31
7.2Effects Of Variables
7.2.1Anaesthetic Agent Consumption
7.2.1.1Isoflurane, Enflurane And Halothane
The rate of consumption of anaesthetic agent depends primarily on flowrate and vapour output
concentration. As an approximate working figure, 1 millilitre of liquid anaesthetic is required to provide
200 millilitres of vapour.
The approximate hourly consumption of anaesthetic agents can be expressed as 3 x % x F, where %
represents the setting of the vaporizer output percentage and F represents the input flowrate in litres/
minute.
Example:If a vaporizer is set to deliver 2% at 6 litres/minute total gas input flowrate then the
approximate rate of consumption = 3 x 2 x 6 = 36 ml/hour.
The figures are approximate and are intended for general guidance only.
7.2.1.2Sevoflurane
The rate of consumption of anaesthetic agent depends primarily on flowrate and vapour output
concentration. As an approximate working figure, 1 millilitre of liquid anaesthetic is required to provide
200 millilitres of vapour.
The approximate hourly consumption of anaesthetic agents can be expressed as 3.3 x % x F, where %
represents the setting of the vaporizer output percentage and F represents the input flowrate in litres/
minute.
Example:If a vaporizer is set to deliver 2% at 6 litres/minute total gas input flowrate then the
approximate rate of consumption = 3.3 x 2 x 6 = 39.6 ml/hour.
The figures are approximate and are intended for general guidance only.
7.2.2Temperature
The effects of variation in temperature are normally negligible at commonly used combinations of dial
setting and ambient temperature.
If the vaporizer temperature is above the range shown on the performance curves, the vaporizer output
may be unpredictably high, particularly if the temperature of the agent approaches the agent boiling
point specified by the agent Manufacturer.
If the vaporizer temperature is below the range shown on the performance curves, the vaporizer output
may be lower than expected.
To help to avoid inaccuracies due to extreme temperatures, before using the vaporizer it must be
allowed to attain a temperature within the range shown on the performance curves.
7.2.3Barometric Pressure
Vaporizers are graduated in % v/v at 760 mm Hg. If the ambient pressure changes the % v/v changes,
so that at an ambient pressure P mm Hg the delivered percentage (D% v/v) is calculated as follows:
Equation 1: D =where % is the nominal setting of the vaporizer.
To obtain a consistent depth of anaesthesia when gross changes of barometric pressure occur, the
% v/v must be changed in inverse proportion to the barometric pressure.
% x 760
P
The vaporizer automatically makes this % v/v change and for practical clinical purposes the effects of
barometric pressure can be ignored.
Page 26 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 32
7.2.4Back Pressure (Steady)
7.2.4.1Low And Moderate Pressures
The vaporizer cannot distinguish between pressures at the outlet due to barometric pressure and
pressures in excess of barometric due to steady back pressures applied by downstream components.
Equation 1 therefore applies with the term P now being the absolute pressure at the outlet, that is,
barometric pressure plus back pressure. Steady back pressure reduces the % v/v.
Currently it is unlikely that the steady back pressure imposed by commonly used downstream
components, other than some ventilators, exceeds 30 mm Hg at commonly used flowrates. Back
pressures as high as 30 mm Hg would reduce the delivered % v/v, at 760 mm Hg barometric pressure,
to the following:
Equation 2:= 0.96 of what would otherwise be expected.
Under normal clinical circumstances effects of this magnitude can be ignored.
7.2.4.2High Pressures
760
790
wWarning: Pressures in excess of 400 mm Hg must not be imposed on the vaporizer since pressures
of this magnitude may overcome the internal pressure balance and cause a variation in output.
7.2.5Back Pressure (Fluctuating)
Fluctuating back pressures may be imposed on the vaporizer by downstream components and/or
assisted or controlled ventilation to the patient. These fluctuating back pressures can affect the
vaporizer and increase the concentration by intermittently altering the pressures, and hence the flow
distribution, within the vaporizer.
The greatest effects are observed at combinations of very low flowrates and low dial setting with large
and rapid pressure fluctuations. The effects become progressively less important as the dial setting and
flowrate increase and the magnitude and rate of cycling of the pressure fluctuations decrease.
7.2.6Carrier Gas Composition
Small effects on the output can occur when the carrier gas composition is changed. These effects may
either be transient, arising from the varying solubility of gases in anaesthetic agent, or steady, due to
the variations in gas density and viscosity affecting the flow split within the vaporizer.
When either air or nitrous oxide at low flows is employed as the carrier gas, the output is depressed
compared to the output when oxygen is the carrier gas. This effect is greatest when nitrous oxide is
employed, but using nitrous oxide reduces the required inspired concentrations of volatile agent
thereby mitigating the small depression in output from the vaporizer.
At high flows and low dial settings a small increase in output may be observed.
If the proportion of nitrous oxide passing into the vaporizer is increased, a temporary decrease in
vapour output is obtained. Conversely, if the proportion of nitrous oxide is decreased, a temporary
increase in output is obtained.
wWarning: Do not operate the vaporizer other than with dry medical gases.
7.2.7Time Out Of Service
If the anaesthetic machine on which the vaporizer is fitted is left for a period of time with no gases
flowing, small concentrations may be detected at the machine outlet immediately after the gas flow is
turned ON. This is a normal machine characteristic and is caused by residual vapour left in the machine
from previous use.
Tec 5 VaporizerPage 27
O & M Manual Part No. 1105-0100-000August 1999
Page 33
When the vaporizer is turned from OFF after a period out of use, a brief high concentration may occur
which rapidly stabilises to the set concentration within approximately 10 seconds at 5 litres/minute.
The preceding phenomena are normal characteristics of vaporizers. In use the volume of vapour
involved is small compared to the volume of the breathing circuit.
7.2.8Other Variables
Ambient temperature, input flowrate and duration of use can affect delivered concentration, particularly
when the vaporizers are used at extremes of the usual clinical range.
Note: Use of the vaporizer at high gas flows and high dial concentrations may affect the accuracy of
delivered concentrations. Refer to Performance Curves in Section 7.1 for full information.
The valve design and temperature compensation system of Tec 5 Vaporizers reduces the effects to
levels such that, under most clinical conditions, their effect on vaporizer performance is not clinically
significant. The nominal performance characteristics should be consulted for further details.
8.0Checking The Calibration
The performance of most vaporizers which are in clinical use can be confirmed by observing patient
signs and consumption of anaesthetic agents. Some users may, however, wish to employ analysers
either as a routine procedure or as part of an investigation to determine whether any abnormalities of
performance have developed.
In order to help to achieve the reliability and consistency of performance of Tec 5 Vaporizers, DatexOhmeda use closely specified test conditions, test methods and detailed protocol in conjunction with
training, experience and quality auditing systems. For these reasons the full programme necessary to
help to ensure that a vaporizer complies with Datex-Ohmeda specifications cannot practicably be
carried out in a field situation.
The following points must be considered when any measurements are being carried out on vaporizers
to assist in determining whether any abnormalities of performance have developed.
1. In order to predict the concentration that the vaporizer can be expected to deliver, the detailed
nominal performance data and the preceding comments must be taken into account.
2. The method of test used must not be such that it bears little relation to normal conditions of
clinical use.
3. Any sampling techniques used must be such as to ensure the following:
a) The sample is fully representative of the vaporizer output, which may not be a homogeneous
mixture at the vaporizer outlet.
b) The absorption of agent by any connecting tubing is negligible.
4. If a number of vaporizers are being examined at the same time the probability of all of them being
consistently in error is so remote as to be negligible and the cause of any apparent error probably
lies in the test method employed.
5. Consistent and reproducible analytical techniques must be used.
6. If unexpected results are obtained it is a wise precaution to repeat the observation because the
vaporizer may be more reliable than the techniques used to observe its performance.
7. If unexpected results occur it is also worthwhile checking for sources of error such as the
flowmeter, leaks or absorption by adjacent components.
8. Full account must be taken of any extraneous effects on the analyser which may arise from
changes in the carrier gas composition.
Page 28 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 34
9. If the anaesthetic machine on which the vaporizer is fitted is left for a period of time with no gases
flowing, sensitive analysers may detect small concentrations of agent for a short time at the
machine outlet after the gas flow is turned ON with the vaporizer turned OFF. This is a normal
machine characteristic caused by residual vapour left in the machine from previous use.
10. When the vaporizer is turned from OFF to 0% or above after a period out of use, an increased
concentration may occur which rapidly stabilises to the set concentration within approximately
10 seconds at 5 litres/minute flow.
11. At the 0 % setting it is not abnormal for small steady concentrations to be observable on sensitive
analysers.
9.0Analytical Techniques
For field checking of the state of calibration, many techniques and analysers are available. DatexOhmeda would not recommend any one technique or analyser in preference to another, but the
calibration and reliability of analysers must be realistically considered and account must be taken of
errors of use.
The following method of checking can be used where special equipment is not available or where a
secondary check of analysers is desired. The characteristics of the vaporizer are such that if the
vaporizer is satisfactory at one dial setting it should be satisfactory at all other graduations.
1. Ensure that the vaporizer is at least half full and has been at an ambient temperature of 21 ± 2° C
for at least three hours.
2. With the vaporizer securely mounted, drain the vaporizer as detailed in Section 4.6 and after
draining ensure that either the drain plug and the filler cap are both securely tightened or the port
valve is fully closed and the locking clamp is in the up position, as appropriate.
3. Check that the control dial is turned OFF and then carefully and quickly pour a measured 50
millilitres of agent into the vaporizer without spilling.
4. Leave the vaporizer at a nominal temperature of 21 ± 2° C for one hour to help to ensure that the
temperature has stabilised.
5. Set the flowrate to 5 litres/minute oxygen.
6. Turn the control dial to 2%, note the time and check that the flowrate is still 5 litres/minute.
Readjust the flowrate as necessary.
7. Leave the vaporizer at this setting for 30 minutes. Periodically check and adjust the flowrate as
necessary. Turn the vaporizer OFF and turn the oxygen OFF.
8. Drain the vaporizer as detailed previously in Instruction 2 and measure the amount of liquid
drained off. The amount of liquid consumed should be as follows:
Appropriate action must be taken to handle the exhaust gases and spillage.
It must be appreciated that the preceding method is designed to be a quick and easy check of vaporizer
operation and that the method is somewhat imprecise. Nevertheless, it would be unusual for the
measured liquid consumption to vary by more than 25% of the preceding values.
Tec 5 VaporizerPage 29
O & M Manual Part No. 1105-0100-000August 1999
Page 35
10.0Specification
10.1Calibration
Vaporizers are calibrated at 21° C using an oxygen carrier gas at a flow of 5 litres/minute and they are
temperature, flow and pressure compensated within the specified operating range.
10.2Resistance To Gas Flow With Vaporizer Out Of Circuit
ResistanceVaporizer SettingFlow rateTemperature
10 -15 cm wgON + 0%5 litres/minute, Oxygen21 ± 2° C
10.3Liquid Capacity
The amount of liquid to fully charge a vaporizer with dry wicks = 300 millilitres (nominal).
The amount retained by the wick system = 75 millilitres (nominal).
5% Vaporizer
The amount of liquid to fill the vaporizer from the minimum to the maximum mark = 170 milliliters
(nominal).
8% Vaporizer (Sevotec)
The amount of liquid to fill the vaporizer from the minimum to the maximum mark = 137 milliliters
(nominal).
10.4Weight And Dimensions
WeightHeightDepthWidth
7.0 Kg250 mm164 mm110 mm
10.5Environmental Conditions
Operating temperature Range = 18°C to 35°C
Storage temperature range = –40°C to +65°C
Note: The vaporizer packaging should be protected from condensation.
10.6Performance
Accuracy at 5 litres/minute O2 @ 21°C ± 2°C:
Vaporizer 5%±0.25% of delivered agent or
Vaporizer 8%±0.4% of delivered agent or
±15% of dial setting (whichever is greater).
±15% of dial setting (whichever is greater).
Page 30 Tec 5 Vaporizer
August 1999O & M Manual Part No. 1105-0100-000
Page 36
11.0Warranty
Such warranties are extended only with respect to the purchase of this Product directly from DatexOhmeda or Datex-Ohmeda’s Authorized Dealers as new merchandise and are extended to the first
Buyer thereof, other than for purpose of resale.
For a period of three (3) years from the date of original delivery to the first Buyer or to Buyer’s order,
this Product, other than its expendable parts, is warranted against functional defects in materials and
workmanship and to conform to the description of the Product contained in the Operation and
Maintenance manual and accompanying labels and/or inserts, provided that the same is properly
operated under conditions of normal use, that regular periodic maintenance is performed and that
replacements and repairs are made in accordance with the instructions provided. This same warranty is
made for a period of thirty (30) days with respect to expendable parts.
The foregoing warranties shall not apply if the Product has been repaired or serviced other than by
Datex-Ohmeda or other than in accordance with written instructions provided by Datex-Ohmeda, or
altered by anyone other than Datex-Ohmeda, or if the Product has been subject to abuse, misuse,
negligence, or accident.
Datex-Ohmeda’s sole and exclusive obligation and the Buyer’s sole and exclusive remedy under the
above warranties is limited to repairing or replacing, free of charge, at Datex-Ohmeda’s option, a
Product, which is confirmed as being defective by Datex-Ohmeda following the Buyer’s notification to
Datex-Ohmeda in accordance with the instructions contained in the Service Policy section of the
Operation and Maintenance Manual, not later than seven (7) days after the expiration date of the
applicable warranty. Datex-Ohmeda shall not be otherwise liable for any damages including but not
limited to incidental damages, consequential damages, or special damages.
There are no express or implied warranties which extend beyond the warranties herein above set
forth. Datex-Ohmeda makes no warranty of merchantability or fitness for a particular purpose with
respect to the Product or parts thereof.
Vaporizers which include an Abbott List No. on the rear of the interlock block were initially distributed
by:
Abbott Laboratories
Abbott Park
Chicago
Illinois 60064 - 3500
Tec 5 VaporizerPage 31
O & M Manual Part No. 1105-0100-000August 1999
Page 37
Corporate Office
Datex-Ohmeda Division
Instrumentarium Corp.
PO Box 900
FIN-00031 Helsinki
Finland
Tel 358 9 39411
Fax 358 9 146 3310
North America
United States
Customer Service and
Distribution Center
Datex-Ohmeda, Inc.
Ohmeda Drive PO Box 7550
Madison, WI 53707-7550, USA
Tel 1 800 345 2700
Fax 1 608 221 4384
Technical Support
Datex-Ohmeda, Inc.
Three Highwood Drive
Tewksbury, MA 01876, USA
Tel 1 800 345 2755
Sales and Service
Datex-Ohmeda, Inc.
Three Highwood Drive
Tewksbury, MA 01876, USA
Tel 1 800 635 6099
Fax 1 978 640 0469
Equipment Service Center
Datex-Ohmeda, Inc.
1315 West Century Drive
Louisville, CO 80027-9560, USA
Tel 1 800 345 2755
Canada
Datex-Ohmeda (Canada) Inc.
1093 Meyerside Drive, Unit 2
Mississauga, Ontario
L5T 1J6
Canada
Tel 1 800 268 1472
Tel 1 905 565 8572
Fax 1 905 565 8592
Asia/Pacific
China
Datex-Ohmeda Pte. Ltd.
Room B416, COFCO Plaza
8 Jianguomennei Avenue
Beijing 100005, PR China
Tel 86 10 6526 9773
Fax 86 10 6526 0653
Datex-Ohmeda Pte. Ltd.
Room 1708, Yunlong Mansion
No. 122 Luoguo Street
Chengdu 610017, PR China
Tel 86 28 661 4424
Fax 86 28 676 2703
Datex-Ohmeda Pte. Ltd.
Suite 1306 Peace World Plaza
362 - 366 Huan Shi Dong Road
Guangzhou, 510060, P R China
Tel 86 20 8384 6914
Fax 86 20 8384 9526
Datex-Ohmeda Pte. Ltd.
Room 2202, Lucky Target Square
No. 500 North Chengdu Road
Shanghai 200003, PR China
Tel 86 21 6361 6383
Fax 86 21 6361 6483
Datex-Ohmeda Pte. Ltd.
Room 809, Truroll Plaza
Wusheng Road
Wuhan 430033, P R China
Tel 86 27 8571 2536
Fax 86 27 8571 2655
India
Datex-Ohmeda (India) Pvt. Ltd.
Block EP & GP, Sector V
Plot XI-16, Salt Lake City
Calcutta 700091
India
Tel 91 33 3574002
Fax 91 33 3574001
Indonesia
Datex-Ohmeda Pte. Ltd.
Wisma Danamon Aetna Life 19th Floor
Jln. Jend Sudirman Kav. 45-46
Jakarta 12930, Indonesia
Tel 62 21 575 0864
Fax 62 21 575 0865
Japan
Datex-Ohmeda K. K.
TRC Annex 9F
6-1-1 Heiwajima
Ohta-ku, Tokyo 143-0006
Japan
Tel 81 3 5763 6801
Fax 81 3 5763 6838
Datex-Ohmeda K. K.
Technical Center
TRC A Bldg. AE 4-8
6-1-1 Heiwajima
Ohta-ku, Tokyo 143-0006
Japan
Tel 81 3 5763 6850
Fax 81 3 5763 6852
Korea
Datex-Ohmeda Pte. Ltd.
10th Floor, Sam Sung Building
36 - 1, Yoido-Dong, Youngdeungpo-Ku
Seoul, Korea
Tel 82 2 786 7421
Fax 82 2 786 7420
Malaysia
Datex-Ohmeda Pte. Ltd.
Level 2 Bangunan O'Connor
13 Jalan 223
46100 Petaling Jaya
Selangor, West Malaysia
Tel 60 3 754 7872
Fax 60 3 757 6948
Singapore
Datex-Ohmeda Pte. Ltd.
152 Beach Road
#12-05/07 Gateway East
Singapore 189721
Tel 65 391 8618
Fax 65 291 6618
Thailand
Datex-Ohmeda Pte. Ltd.
12th Floor (Unit F) Grand Amarin Tower
1550 New Petchburi Road, Makasan,
Rajathevi,
Bangkok 10320, Thailand
Tel 66 2 2071012/13
Fax 66 2 207 1014
Taiwan and Philippines
Datex-Ohmeda Pte. Ltd.
2nd Floor, No. 85, Chien-Kuo North
Road, Sec. 2
Taipei, Taiwan
Republic of China
Tel 886-2 2515-0457
Fax 886-2 2501-9136
Vietnam
Datex-Ohmeda Pte. Ltd.
522G Nguyen Tri Phuong St.
Ho Chi Minh City, Dist. 10 Vietnam
Tel 848 865 5875
Fax 848 862 5501
Australia
Datex-Ohmeda Pty. Ltd.
Units 1 & 2
149 Arthur Street
P O Box 356
Homebush
NSW 2140
Australia
Tel 61 132 229
Fax 61 297 461796
Europe
France
Datex-Ohmeda S.A.S.
Parc de Pissaloup
8 Avenue Jean d’Alembert
78191 Trappes-Cedex
France
Tel 33 1 30 68 60 00
Fax 33 1 30 68 60 01
Datex-Ohmeda S.A.S.
Siege Social Lyon
17 rue Jean-Elysée Dupuy
69410 Champagne Au Mont d'Or
France
Tel 33 04 78 66 62 10
Fax 33 04 78 43 26 58