CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Datex-Ohmeda Inc.
P.O. Box 7550
Madison, WI 53707-7550, USA
Tel: +1 608 221 1551
Fax: +1 608 222 9147
All specifications are subject to change without notice.
Outside the USA, check local laws for any restriction that may apply.
Order Code M1021567
March, 2006
GE Healthcare Finland Oy
Helsinki, Finland
P.O. Box 900
FI-00031 GE, FINLAND
Tel: +358 10 39411
Fax: +358 9 1463310
www.gehealthcare.com
Copyright 2006 General Electric Company
Page 2
Intended purpose (Indications for use)
The Datex-Ohmeda S/5 Compact Anesthesia Monitor with L-CANE05 or L-CANE05A software is intended for
multiparameter patient monitoring with optional patient care documentation.
The S/5 Compact Anesthesia Monitor with L-CANE05 or L-CANE05A software is indicated for monitoring of
hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory,
gastrointestinal/regional perfusion, Bispectral index (BIS), Entropy (State Entropy and Response Entropy) and
neurophysiological status of all hospital patients.
The S/5 Compact Anesthesia Monitor with L-CANE05 and L-CANE05A software when using BIS is for
monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may
be used as an aid in monitoring the effects of certain anesthetic agents.
The S/5 Compact Anesthesia Monitor with L-CANE05 or L-CANE05A software is also indicated for documenting
patient care related information.
The S/5 Compact Anesthesia Monitor with L-CANE05 or L-CANE05A software is indicated for use by qualified
medical personnel only.
The Datex-Ohmeda S/5 Compact Critical Care Monitor with L-CICU05 or L-CICU05A software is intended for
multiparameter patient monitoring.
The S/5 Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated for monitoring of
hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory,
gastrointestinal/regional perfusion, Bispectral index (BIS) and neurophysiological status of all hospital
patients.
The S/5 Compact Critical Care Monitor with L-CICU05 and L-CICU05A software when using BIS is for monitoring
the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as
an aid in monitoring the effects of certain anesthetic agents.
The S/5 Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated for use by qualified
medical personnel only.
Classifications
In accordance with IEC 60601-1
Class I and internally powered equipment – the type of protection against electric shock.
Type BF or CF equipment. The degree of protection against electric shock is indicated by a symbol on each
parameter module.
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Continuous operation according to the mode of operation.
In accordance with IEC 60529
IPX1- degree of protection against harmful ingress of water.
In accordance with EU Medical Device Directive
The Datex-Ohmeda S/5 Compact Anesthesia Monitor is classified as IIb.
In accordance with CISPR 11:
Group 1, Class B
• Group 1 contains all ISM (industrial, scientific and medical) equipment in which there is intentionally
generated and/or used conductively coupled radio-frequency energy which is necessary for the internal
functioning of the equipment itself.
• Class B equipment is suitable for use in domestic establishments and in establishments directly connected
to a low voltage power supply network which supplies buildings used for domestic purposes.
Page 3
Responsibility of the manufacturer
GE Healthcare Finland Oy (GE) is responsible for the safety, reliability and performance of the product only if:
• modifications, service and repairs are carried out by personnel authorized by GE.
• the electrical installation of the room complies with appropriate requirements.
• the equipment is used in accordance with the User's Guide and serviced and maintained in accordance
with the Technical Reference Manual.
Trademarks
Datex, Ohmeda, S/5, D-lite, D-lite+, Pedi-lite, Pedi-lite+, Mini D-fend, D-fend, D-fend+, OxyTip+, MemCard,
ComWheel, EarSat, FingerSat, FlexSat, PatientO2, Entropy, Patient Spirometry and Tonometrics are trademarks
of GE Healthcare Finland Oy. All other product and company names are property of their respective owners.
A portion of the Entropy software is derived from the RSA Data Security, Inc. MD5 Message-Digest Algorithm.
Page 4
Page 5
Technical Reference Manual, Order code: M1021567
Document No.UpdatedDescription
S/5
S/5
Master table of contents
Datex-Ohmeda
TM
Compact Anesthesia Monitor
TM
Compact Critical Care Monitor
Part I, General Service Guide
M1031517-02Introduction, System description, Installation,
Interfacing, Functional check, General troubleshooting
M1031518-02Planned Maintenance Instructions2
Part II, Product Service Guide
Document No.UpdatedDescription
M1031519-01CAM, CCCM Service Menu1
M1031520-02 Display Unit for F-CM1, F-CMC1 rev. 00...04
F-CMREC1, F-CMCREC1 rev .00...02
Software Licenses L-CANE05(A), L-CICU05(A)
M1044470-01Frame Unit for F-CM1, F-CMC1 rev. 00...04
Figure 10Scavenging through ventilator reservoir............................................................................................... 27
Figure 11Connecting the gas module to the scavenging connector of S/5 Avance .............................................. 27
Figure 12Sample gas returned to patient circuit in ADU ..................................................................................... 28
Figure 13Connection cables and LED indicators................................................................................................ 36
Figure 14An example of interfacing external devices with Device Interfacing Solution.......................................... 37
Figure 15Connecting S/5 Compact Monitor to printer, converter model PI130-R2 .............................................. 43
Figure 16Connecting S/5 Compact Monitor to printer, converter model PI 1115A............................................... 44
Figure 17S/5 Compact Monitor general troubleshooting flowchart..................................................................... 59
iv
Document no. M1031517-02
Page 11
Table of figures
Document no. M1031517-02
v
Page 12
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
List of tables
Table 1Transference of parameters, Datex-Ohmeda monitors ............................................................ 32
Table 2Transference of parameters, external monitors .......................................................................32
Table 3DIS modules and interfaced devices...................................................................................... 34
Table 4Parameters transferred from S/5 Compact Anesthesia Monitor to S/5 Anesthesia
Delivery Unit ......................................................................................................................... 39
Table 5Events transferred from S/5 Anesthesia Delivery Unit to the S/5 Compact Anesthesia Monitor 40
Table 6Parameters transferred from Dräger Cicero monitor to S/5 Compact Anesthesia Monitor or
S/5 Compact Critical Care Monitor........................................................................................ 41
Table 7Parameters transferred from Dräger Cato, Julian and Narkomed 2C (NAD) monitor to the S/5
Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor....................................... 42
Table 8Coding element connector, X4............................................................................................... 45
Table 11Patient simulators’ compatibility with each hemodynamic module.......................................... 49
Table 12Adapter cables for hemodynamic patient simulators.............................................................. 50
Table 13Guidance and manufacturer’s declaration – electromagnetic emissions ................................B-1
Table 14Guidance and manufacturer’s declaration – electromagnetic immunity .................................B-2
Table 15Guidance and manufacturer’s declaration – electromagnetic immunity .................................B-3
Table 16Recommended separation distances between portable and mobile RF communications
equipment and the S/5™ CAM/ CCCM.................................................................................B-5
vi
Document no. M1031517-02
Page 13
About this manual
Notes to the reader
This Technical Reference Manual is intended for service personnel and engineers who will install,
service and maintain Datex-Ohmeda S/5 Compact Anesthesia Monitors running the software
license L-CANE05 or L-CANE05A and S/5 Compact Critical Care Monitors running the software
license L-CICU05 or L-CICU05A. As the monitor setup may vary, some functions described may not be
available in the monitor you are using.
•The order code for the entire printed manual is M1021567. The manual includes Technical
Reference Manual Slots and every slot has an individual document number.
•Part I gives the reader an overview of the monitor. It contains the information needed to install,
interface and troubleshoot the monitors. Instructions for functional check and planned
maintenance are also included. Read the manual through and make sure that you understand
the procedures described before the installation of the monitor. To avoid risks concerning safety
or health, strictly observe the warning indications. If you need any assistance concerning the
installation, please do not hesitate to contact your authorized distributor.
•Part II contains detailed descriptions of each component of the S/5 CAM, CCCM Monitor,
For information of parameter modules, Remote Controller and Device Interfacing Solution refer to the
“S/5 E-Modules, Technical Reference Manual”. Service check for each of these products is included
in these slots.
The manufacturer reserves the right to change product specifications without prior notice. Although
the information in this manual is believed to be accurate and reliable, the manufacturer assumes no
responsibility for its use.
Installation and service are allowed by authorized service personnel only.
GE Healthcare Finland Oy (GE) assumes no responsibility for the use or reliability of its software in
equipment that is not furnished by GE.
Related documentation
S/5 Modules
S/5 E-Modules, Technical Reference Manual
S/5 M-Modules, Technical Reference Manual
S/5 Compact Anesthesia Monitor
For instructions for daily use including cleaning and daily maintenance, clinical aspects and basic
methods of measurement see:
S/5 Compact Anesthesia Monitor, User’s Guide
S/5 Compact Anesthesia Monitor, User’s Reference Manual
S/5 Compact Critical Care Monitor
For instructions for daily use including cleaning and daily maintenance, clinical aspects and basic
methods of measurement:
S/5 Compact Critical Care Monitors, User’s Guide
S/5 Compact Critical Care Monitors, User’s Reference Manual
For more information about the iCentral, S/5 Arrhythmia Workstation and anesthesia record keeping
solution, see the “Technical Reference Manuals” and “User’s Reference Manuals” for these products.
1
Document no. M1031517-02
Page 14
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Software options and default settings are described in the “Default Configuration Worksheet”
delivered with each monitor.
Available accessories are described in the “Supplies and Accessories” catalog delivered with each
monitor.
Conventions used
To help you find and interpret information easily, the manual uses consistent text formats:
"
Sign the check form after performing the procedure.
Hard KeysNames of the hard keys on the Remote Controller, Command Board, side panel and modules are
written in the following way:
Menu ItemsSoftware terms that identify window parts or menu items are written in bold italic: ECG Setup.
Menu access is described from top to bottom. For example, the selection of the
key, the Screen 1 Setup menu item and the Waveform Fields menu item would be shown as
Monitor Setup - Screen 1 Setup - Waveform Fields.
‘Messages’Messages (alarm messages, informative messages) displayed on the screen are written inside single
quotes: ‘Please wait’.
“Sections”When referring to different sections in this manual or to other manuals, manual names and section
names are enclosed in double quotes:
See section "Cleaning and care."
Please refer to "iCentral User's Reference Manual: Alarms."
Hypertext linksHypertext links on PDF versions are written in blue color.
WARNINGWarnings are written in the following way:
WARNINGThis is a WARNING.
CAUTIONCautions are written in the following way:
Others.
Monitor Setup hard
CAUTIONThis is a CAUTION.
NOTENotes are written in the following way:
NOTE: This is a NOTE.
In this manual, the word “select” means choosing and confirming.
Illustrations and names
All illustrations in this manual are only examples, and may not necessarily reflect your system
settings or data displayed in your system. If a particular selection is not available in your system, the
selection is shown grayed.
2
Document no. M1031517-02
Page 15
1Introduction
The modular design makes the system flexible and easy to upgrade. In addition to patient parameter
modularity and easy upgrades, the monitor can be upgraded to anesthesia record keeping, wired
and wireless networking and memory card operation. Additionally, external devices can be interfaced
to the monitor with interface modules.
NOTE: Your system may not include all these components. Consult your local representative for the
available components.
Introduction
1
5
Figure 1S/5 Compact Anesthesia Monitor system
(1) S/5 Compact Anesthesia Monitor with modules inside
(2) Remote Controller, K-CREMCO
(3) Printer
(4) Wireless Network (WLAN) Access Point, N-WAP
(5) Network (LAN)
2
3
4
Optional components
Optional components are:
•Built-in recorder, N-CMREC1
•Remote controller
•Keyboard for anesthesia record keeping solution, barcode reader
•Wireless network option, N-CMW
Possible software options are:
•MemCard option, N-CMMEM
NOTE: Insert the WLAN PC card in the first slot from the front.
NOTE: Memory Module, M-MEM/ E-MEM, is not compatible with the S/5 Compact Monitors.
3
Document no. M1031517-02
Page 16
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Communication between monitors
You can use the Compact Monitor as a stand-alone monitor or for
−viewing and receiving data (alarms, vital signs) from other patient monitors
−gathering and storing data during transportation.
To view other patient monitors, the monitor needs to be connected to the network (LAN or WLAN). To
gather, store, and transfer data between different Datex-Ohmeda monitors, use memory cards or
connect the monitor to the network.
NOTE: With WLAN, use only Data Card.
4
Document no. M1031517-02
Page 17
1.1Symbols
1.1.1Symbols on transport packaging
The contents of the transport package are fragile and must be handled with care.
Indicates the correct upright position of the transport package.
The transport package must be kept in a dry environment.
Indicates the temperature limitations within which the transport package should be
stored.
Introduction
1.1.2Symbols on equipment
This battery contains Ni-MH and can be recycled.
Ni-MH
Pb
Dangerous voltage.
When using the ARK Barcode Reader, N-SCAN, do not stare into beam. The N-SCAN
Barcode Reader is a Class 2 laser product.
Document no. M1031517-02
5
Page 18
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
1.1.3Equipment safety symbols
-Attention, consult accompanying documents.
-When displayed next to the O2 value, indicates that the FiO2 low alarm limit is set
below 21%.
-When displayed next to the HR value, indicates that the pacer is set on R.
-On the modules or frames indicates that modules with identical measurements
should not be used in the same monitor. If such modules have been inserted,
remove the module that has been most recently connected. You can also remove
both modules and re-connect the new module after five seconds.
-On the Interface Module E-INT, M-INT, indicates that it is for connecting external
devices. Do not connect patient cables to the module.
-On the E-TONO, M-TONO module indicates that the module should only be used
with Tonometrics catheters.
-On the E-miniC module indicates that airway gases should be calibrated every six
months in normal use and every two months in continuous use.
-BIS: On the Aspect DSC indicates that the converter must not be opened for any
reason or autoclaved.
-On the E-PRESTN, E-PRETN, E-RESTN, E-PSM, E-PSMP, E-P, E-PP, E-PT, E-COP and
E-COPSv module indicates that protection against cardiac defibrillator discharge
is due in part to the accessories for pulse oximetry (SpO2), temperature (T) and
invasive pressure (P) measurement.
-On the E-NMT, M-NMT module indicates the following warnings:
* Do not place the NMT stimulating electrodes on the patient’s chest.
* Always stop the NMT measurement before handling the stimulating electrodes.
* Never subject a patient with an implanted electronic device to electrical
stimulation without consulting a medical specialist first.
-On the rear or bottom panel this symbol indicates the following warnings and
cautions:
* Electric shock hazard. Do not open the cover or the back. Refer servicing to
qualified service personnel.
* For continued protection against fire hazard, replace the fuse only with one of
the same type and rating.
* Do not touch a battery-operated monitor during defibrillation procedure.
* Disconnect from the power supply before servicing.
* Do not use the monitor without manufacturer approved mounting attached.
6
Document no. M1031517-02
Page 19
Introduction
Type BF (IEC 60601-1) protection against electric shock.
Type BF (IEC 60601-1) defibrillator-proof protection against electric shock.
Type CF (IEC 60601-1) protection against electric shock.
Type CF (IEC 60601-1) defibrillator-proof protection against electric shock.
When displayed in the upper left corner of the screen, indicates that the alarms are
silenced. When displayed in the menu or digit fields, indicates that the alarm source
has been turned off or alarm does not meet the alarm-specific activation criteria.
ESD warning symbol for electrostatic sensitive devices. Pins of connectors identified
with the ESD warning symbol should not be touched. Connections should not be
made to these connectors unless ESD precautionary procedures are used. For
details, see section “1.2.2. ESD precautionary procedures”.
1.1.4Other symbols
SN, S/N
Symbol for non-ionizing electromagnetic radiation. Interference may occur in the
vicinity of equipment marked with this symbol.
Equipotentiality. Monitor can be connected to potential equalization conductor.
Alternating current
Fuse. Replace the fuse only with one of the same type and rating.
Serial Number
Battery operation and remaining capacity (green bar)
Battery charging (white bar)
Submenu. Selecting an alternative marked with this symbol in a menu opens a new
menu.
Document no. M1031517-02
7
Page 20
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
The monitor is connected to the network (LAN).
The monitor is connected to the wireless network (WLAN).
Data Card (green) and/or Menu Card (white) is inserted.
WLAN signal strength. The number of segments corresponds to the signal strength:
four segments indicate strong signal, one segment weak signal. When connection to
access point is being searched, the segments scroll from zero to four and back.
A blinking heart next to the heart rate or pulse rate value indicates the beats
detected.
A lung next to the respiration rate value indicates that respiration rate is calculated
from the impedance respiration measurement.
Gas inlet
Gas outlet
Do not reuse.
Use by. Indicates the last use day.
Date of manufacture
Do not immerse the sensor in liquids.
8
Document no. M1031517-02
Page 21
Introduction
IPX class:
IPX0
IPX1
IPX2
IPX3
IPX4
IPX7
IPX8
Degree of protection against harmful ingress of water as detailed in the
IEC 60529:
- Ordinary equipment
- Protection against vertically falling water drops.
- Protection against vertically falling water drops when enclosure tilted up to 15 °.
- Protected against spraying water.
- Protected against splashing water.
- Protected against the effects of temporary immersion in water.
- Protected against the effects of continuous immersion in water.
This symbol indicates that the waste of electrical and electronic equipment must not
be disposed as unsorted municipal waste and must be collected separately. Please
contact an authorized representative of the manufacturer for information concerning
the decommissioning of your equipment.
Ethernet connector
Document no. M1031517-02
9
Page 22
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
1.2Safety
The following list contains general warnings and cautions you should know before installing,
maintaining or servicing the system. Warnings and cautions specific to the use of the system can be
found in the User’s Guide and User’s Reference Manual.
1.2.1Safety precautions
Warnings
WARNINGA WARNING indicates a situation in which the user or the patient may be in
danger of injury or death.
Power connection
Use only hospital-grade grounded power outlets and power cord. Do not remove the grounding
•
pin from the power plug.
•Use only an intact power cord. Replace the power cord if it is cracked, frayed, broken or
otherwise damaged.
•Do not apply tension to the power cord otherwise the cord may get damaged.
•Do not use extension cords or adapters of any type.
•Before starting to use the system, ensure that the whole combination complies with the
international standard IEC 60601-1-1 and with the requirements of the local authorities. Do
not connect any external devices to the system other than those specified.
•If the integrity of the external protective earth conductor arrangement is in doubt, use the
monitor with battery operation.
•To avoid the risk of electric shock, this equipment must only be connected to a supply mains
with protective earth.
•Do not use an additional multiple socket outlet or extension cord.
Installation
Keep the monitor horizontal when the Compact Airway Module is used. Tilting the monitor may
•
cause erroneous results in the Compact Airway Module’s readings and damage the module.
•The monitor or its components should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the monitor and its components should be observed to
verify normal operation in the configuration in which it will be used.
•Pins of connectors identified with the ESD warning symbol should not be touched. Connections
should not be made to these connectors unless ESD precautionary procedures are used. For
details, see section “1.2.2. ESD precautionary procedures”
•After transferring or reinstalling the monitor, always check that it is properly connected and all
parts are securely attached. Pay special attention to this in case of stacked mounting.
•Never install the monitor so that it is above the patient.
•A printer or computer must be supplied from an additional transformer providing at least basic
isolation (isolating or separating transformer).
•If you accidentally drop the monitor, modules or frames, have them checked by authorized
service personnel prior to clinical use.
10
Document no. M1031517-02
Page 23
Introduction
Laser radiation
When using the ARK Barcode Reader, N-SCAN, do not stare into the beam. The N-SCAN is a
•
Class 2 laser product.
External connection
Do not connect any external devices to the monitor other than those specified.
•
Fuse replacement
Replace a fuse only with one of the same type and rating.
•
Explosion hazard
Do not use the monitor in the presence of flammable anesthetics.
•
Patient safety
Do not perform any testing or maintenance on the monitor while it is being used on a patient.
•
•PACEMAKER PATIENTS: The impedance respiration measurement may cause rate changes in
Minute Ventilation Rate Responsive Pacemakers. In this case, set the pacemaker rate
responsive mode off or turn the monitor impedance respiration measurement off.
•Never install the monitor so that it is above the patient.
•The monitor must not be used without manufacturer approved mounting attached.
•Operation of the monitor outside the specified values may cause inaccurate results.
Cleaning and service
•
Only trained personnel with proper tools and test equipment should perform the tests and
repairs described in this manual. Unauthorized service may void the monitor warranty.
•Turn the power off and unplug the power cord before cleaning or service. Completely remove
any moisture before reconnecting the power cord to the mains outlet.
•Do not touch any exposed wire or conductive surface while any cover is removed and the
monitor is energized. The voltages present can cause injury or death.
•Pins of connectors identified with the ESD warning symbol should not be touched. Connections
should not be made to these connectors unless ESD precautionary procedures are used. For
details, see section “1.2.2. ESD precautionary procedures”.
•Always perform an electrical safety check and a leakage current test on the monitor after
service.
Accessories
Use only accessories, including mounts and batteries, and defibrillator-proof cables and
•
invasive pressure transducers approved by GE Healthcare. For a list of approved supplies and
accessories, see the “Supplies and Accessories” catalog delivered with the monitor. Other
cables, batteries, transducers and accessories may cause a safety hazard, damage the
equipment or the system, result in increased emissions or decreased immunity of the
equipment or system or interfere with the measurement. Protection against cardiac defibrillator
discharge is due in part to the accessories for pulse oximetry (SpO2), temperature (T) and
invasive pressure (P) measurement.
11
Document no. M1031517-02
Page 24
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
•Single use accessories are not designed to be reused. Reuse may cause a risk of contamination
and/or affect the measurement accuracy.
Cautions
CAUTIONA CAUTION indicates a condition that may lead to equipment damage or malfunction.
Installation
Leave space for air circulation to prevent the monitor from overheating.
•
•Before connecting the power cord to the power supply, check that the local voltage and
frequency correspond with the rating stated on the device plate.
Before use
Allow two minutes for warm-up and note any error messages or deviations from normal
•
operation.
•Do not connect a sampling line to the female Patient Spirometry connector while the other end
of the sampling line is connected to the D-fend water trap. The pressure in the gas sampling
system may cause damage to the PVX unit pressure transducers.
Autoclaving and sterilizing
Do not autoclave any part of the monitor.
•
•Do not gas sterilize the modules.
Cleaning and service
•
Do not use hypoclorite, ammonia-, phenol-, or acetone based cleaners. These cleaners may
damage the monitor surface.
•Do not immerse any part of the device in any liquid, or allow liquid to enter the monitor or
modules.
•Do not apply pressurized air to any outlet or tubing connected to the monitor.
•Electrostatic discharge through the PC boards may damage the components. Before handling
PC boards, wear a static control wrist strap. Handle all PC boards by their non-conductive edges
and use anti-static containers when transporting them.
•Do not break or bypass the patient isolation barrier when testing PC boards.
•Do not clean the spirometry tubes with high pressure air or O2 flushing while the spirometry
tubes are connected to Patient Spirometry connector. High differential pressure may damage
PVX unit pressure transducers.
Special components
Special components are used in these monitors that are vital to assure reliability and safety. GE
•
Healthcare assumes no responsibility for damage, if replacement components not approved by
GE Healthcare are used.
• A lithium battery on the CPU Board. Dispose of the faulty IC containing the battery according to
local regulations.
Batteries
To replace the batteries safely, please refer to the service instructions in this manual.
12
Document no. M1031517-02
Page 25
•Do not short-circuit the battery terminals, this may produce a very high current, which will
damage the battery.
•Do not dispose of the battery into open flame, nor put the battery near fire, as it may explode.
•Do not dismantle the battery.
See also section “Symbols”.
Storage and transport
Do not store or transport the monitor outside the specified temperature, pressure and humidity
ranges:
For display specific environmental requirements see specifications in the “Display” slot.
1.2.2ESD precautionary procedures
To avoid electrostatic charges building up, it is recommended to store, maintain and use the
•
equipment at a relative humidity of 30% or greater. Floors should be covered by ESD dissipative
carpets or similar. Non-synthetic clothing should be used when working with the component.
•To prevent applying a possible electrostatic discharge to the ESD sensitive parts of the
equipment, one should touch the metallic frame of the component or a large metal object
located close to the equipment. When working with the equipment and specifically when the
ESD sensitive parts of the equipment may be touched, a grounded wrist strap intended for use
with ESD sensitive equipment should be worn. Refer to the documentation provided with the
wrist straps for details of proper use.
Introduction
ESD precautionary procedure training
It is recommended that all potential users receive an explanation of the ESD warning symbol and
training in ESD precautionary procedures.
The minimum contents of an ESD precautionary procedure training should include an introduction to
the physics of electrostatic charge, the voltage levels that can occur in normal practice and the
damage that can be done to electronic components if they are touched by an operator who is
electrostatically charged. Further, an explanation should be giv en o f m e th o ds to p re v en t b u il d -u p o f
electrostatic charge and how and why to discharge one’s body to earth or to the frame of the
equipment or bond oneself by means of a wrist strap to the equipment or the earth prior to making a
connection.
1.2.3Disposal
Dispose of the whole device, parts of it and its packing material in accordance with local
environmental and waste disposal regulations.
13
Document no. M1031517-02
Page 26
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2System description
2.1Introduction
Datex-Ohmeda monitors build up a freely configurable modular system. The architecture is designed
to enable different module combinations so that the user is able to get the desirable parameter and
feature set. This modular approach makes it possible to add new features when they are needed.
2.2Bus structure
Patient
Patient
Patient
Modules
Module
Module
MODULE-BUS
PCMCIASerial
Non-
Volatile
Memory
Reset
Logic
Boot
Memory
EthernetSoundUPI
ISA-BUS
Code
Memory
Display
Controller
Chipset
486 LOCAL BUS
DRAM
486
Processor
cam_p1_gnrl_bus_strctr_cm.vsd
Figure 2General bus structure of S/5 system
The CPU bus is a communication channel used only for internal data transfer. It is based on the ISA
bus used in IBM PC computers. Data is transferred on this 16 bit wide bus using the CPU clock
frequency.
The module bus is for the parameter modules. The bus is based on the industry standard RS-485,
which uses a differential serial method to transfer data. This type of bus is robust and it allows
parameter modules to be inserted or removed while the power is on. The module bus uses a 500 kbps
data transfer rate.
The RS-485 type of serial communication supports so-called multidrop or party line connections. This
means that all parameter modules connected to the module bus use exactly the same lines for
communication. The advantage of this is that all bus connectors are identical and the modules can be
connected in any order and position.
14
Document no. M1031517-02
Page 27
2.3Distributed processing
A system assembled from S/5 products is a multiprocessor system. All parameter modules have their
own microprocessor, which performs functions such as module key control, waveform filtering,
parameter related computing and pneumatic control, etc. At the same time the main CPU performs
higher level tasks such as trending and alarm control. While the parameter modules and CPU are
performing their tasks, the UPI (Universal Peripheral Interface) microprocessor handles all functions
needed to transfer data between the parameter modules and the CPU.
This kind of parallel processing gives one major advantage to centralized processing. When new
parameter modules are added to the system, the processing power is increased. As a result, the
system does not slow down when new features are added.
2.4Module communication
The communication master controlling data transfers between the CPU and parameter modules is
called UPI processor. It sends data to each connected module 100 times a second. Modules respond
to each data request immediately by sending a data package, whose length depends on the type of
the module. This communication protocol ensures that each module receives and sends data every
10 ms. If a module does not respond to data requests, the UPI processor presumes that the module is
disconnected.
Parameter modules may hold a static (fixed) or dynamic address, which the UPI processor uses when
sending out data. Two parameter modules of the same type must not be fitted onto the same monitor
since they might reply to a data request simultaneously, thus causing communication errors.
System description
Voltage
Temp
Meas
Meas
Shared
SRAM
Control
ISA-BUS
Logic
Processor
UPI
Analog & Digit al outputs
Module Bus
ComWheel I nterface
Keypad Interface
cam_p1_prncpl_UPI_operation.vsd
Figure 3Principle of UPI section operation
The UPI processor collects and stores all data that is received from the parameter modules into a
shared SRAM, which is mapped directly to the address space of the main CPU. The main CPU reads
data from the memory while the UPI processor guarantees that the data is up to date. This operation
also works in the other direction. In this the main CPU fills the shared SRAM with data and the UPI
processor distributes it to the parameter modules.
15
Document no. M1031517-02
Page 28
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2.5Software loading
The program memory on the CPU board is loaded with monitor software at the factory. The software is
used for running all the functions that are integrated into the PC board. For service and upgrade
procedures, the CPU board is fitted with a PCMCIA card drive through which new software can be
loaded.
2.6Parameter modules
PATIENT
Module
keys
Analog
electronics
Peripheral
drivers
A/D
converter
RAM
EEPROM
+5 V
+12 V
CPU
Isolation
transformer
Patient isolation
Opto
isolation
+13...16 V
VMOD
Data
RS485
drivers
Module Bus
Figure 4General structure of parameter modules with patient isolation
The detailed structure of a parameter module depends on the specific needs for each individual
parameter. However, some common parts are used in the parameter modules. The electronics inside
the module is usually divided into isolated (floating) and non-isolated sections. Typically, the
non-isolated section consists of buffers to interface the parameter module to the module bus while
the rest of the electronics is located in the isolated section. The isolated section includes the
microcontroller together with memory components, the front-end analog electronics (amplifiers, etc.)
and sensor drivers.
eneral_struct_of_parm_mod.vsd
16
Document no. M1031517-02
Page 29
3System installation
3.1Unpacking instructions
1.Confirm that the packing box is undamaged. If the box is damaged, contact the shipper.
2.Open the top of the box and carefully unpack all components.
3.Confirm that all components are undamaged. If any of the components are damaged, contact
the shipper.
4.Confirm that all components are included. If any of the components are missing, contact your
GE Healthcare distributor.
3.2Choosing location
Consider the following aspects:
•lighting
•space
•connections
•electromagnetic and radio frequency interference, see Appendix B. ElectroMagnetic
Compatibility
System installation
•environment
WARNINGThe monitor or its components should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the monitor and its
components should be observed to verify normal operation in the configuration in
which it will be used.
WARNINGNever install the monitor so that it is above the patient.
CAUTIONThe monitor display is fragile. Ensure that it is not placed near a heat source or
exposed to mechanical shocks, pressure, moisture or direct sunlight.
3.3S/5 Frame for Compact Monitor
Mounting of S/5 Compact Monitor to the Bed Mount, Wall Mount or Roll Stand is described in a
separate instruction sheet delivered with each mount.
WARNINGAfter transferring or reinstalling the monitor, always check that it is properly
connected and all parts are securely attached. Pay special attention to this in
case of stacked mounting.
17
Document no. M1031517-02
Page 30
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
3.3.1Compact Monitor connections
1
9
10
11
12
2
3
4
14
5
8
13
6
7
Figure 5External connections of Compact Monitor frame F-CM(C)1 rev.04
(1) Air filter
(2) NET connector
(3) NET ID connector (not used with N-CMW)
(4) Recorder (optional)
(5) SYNC connector
(6) Connector for external keyboard (anesthesia record keeping solution keyboard), K-CREMCO or
Barcode Reader
(7) Potential equalization
(8) Serial port
(9) Receptacle for power cord
(10) WLAN PC card LED
(11) WLAN PC card
(12) Slot for Memory Cards (PCMCIA) and WLAN PC card
(13) Connector for Device Interfacing Solution
(14) Modules
NOTE: Insert the WLAN PC card in the first PC card slot from the front.
18
Document no. M1031517-02
Page 31
11
System installation
11
1
2
3
4
5
10
9
13
8
7
6
14
12
Figure 6External connections of Compact Monitor frame F-CM(C)1 rev.00...03
(1) Dust filters
(2) Recorder unit (optional)
(3) Connector for factory use only
(4) Synchronization connector
(5) Connector for external keyboard (anesthesia record keeping solution keyboard), K-CREMCO or
Barcode Reader
(6) Potential equalization connector
(7) Serial port connector
(8) Receptacle for power cord
(9) NET ID connector
(10) NET connector
(11) WLAN antenna cover (optional)
(12) Connector for Device Interface Solution
(13) WLAN antenna card LED
WLAN antenna card (optional)
CAUTIONSwitch the power to standby before making any connections.
Document no. M1031517-02
19
Page 32
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
3.3.2Connecting to mains
Connect the power cord to the mains power inlet on the side of the power supply unit and to the wall
socket.
NOTE: Before taking the monitor into use for the first time, the batteries should be fully charged. Keep
the monitor connected to the mains until the Battery charging symbol disappears, or in STBY mode
the Stby LED stops blinking (may take up to 10 hours if the batteries are fully discharged).
WARNINGThe power cord may only be connected to a three-wire, grounded, hospital grade
receptacle.
3.3.3Connecting to Network
Make sure that the Datex-Ohmeda Network Option, N-CMNET or S/5 MemCard and Network Option,
N-CMMEM is installed.
NOTE: With L-CANE03(A) and L-CICU03(A) or newer software, the N-CMNET option is automatically
installed.
Use the Monitor-Network cable to connect the monitor to the network.
1.Make sure that the power is switched off.
2.Connect the Identification Plug and one RJ-45 connector to the connectors X4 and X5.
3.Connect the other RJ-45 connector to the corresponding connector on the wallbox.
4.Switch on the power. Confirm that the network symbol and 'Connected to Network' message
are displayed on the upper part of the screen.
3.3.4Connecting to Wireless Network
Make sure that the S/5 Wireless LAN Option, N-CMW and the antenna cover are installed. Use the
wireless LAN antenna card to connect the monitor to the network.
NOTE: The N-CMW option requires also that either N-CMNET or N-CMMEM option is installed.
1.Make sure that the power is switched off.
2.Connect the wireless LAN antenna card into the foremost card drive slot.
3.Switch on the power. Confirm that the wireless LAN network symbol and the wireless LAN signal
strength symbol are displayed on the upper part of the screen.
20
Document no. M1031517-02
Page 33
3.3.5Inserting the parameter modules
Insert each plug-in parameter module into a module slot. Firmly press the module in position.
NOTE: Ensure that the module is properly oriented (i.e. the module release latch is facing downward).
Do not use two or more modules with identical functions in the monitor.
•The Recorder Module, E-REC/M-REC and the built-in recorders in F-CMREC1 or F-CMCREC1
and U-CMREC1 or N-CMREC1
System installation
3.3.6Downloading Monitor Software
The following instructions apply to downloading of new monitor software in case of upgrade or
service. Detailed instructions for downloading software are supplied with software PCMCIA cards.
NOTE: All user settings will be lost after downloading of new monitor software.
NOTE: During the downloading of software, the serial number of the monitor is written on the software
card. The software can then be downloaded again onto the same monitor, but not onto any other
monitor.
1.Make sure that the monitor is switched to standby.
2.Press and hold the service reset button on the rear panel for at least five seconds or until the
service reset indicator LED turns off.
3.Open the cover for card drive slots, if installed. Insert the software card into one of the card drive
slots and press the software card firmly in position.
4.Make sure that at least 30 seconds have passed since the service reset, then switch the monitor
on.
5.Wait for approximately 80 seconds. After the start-up screen appears, enter the Service View
and make sure that the information regarding monitor software has been updated. Memorize
the serial number of new software.
6.Remove the software card.
7.Replace the original device plate for monitor software by the new one supplied with the software
card.
8.Perform Factory Reset, see “Performing Factory Reset” . Make sure that the monitor functions
normally after the restart.
9.Set the time and date.
21
Document no. M1031517-02
Page 34
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
10. Set the monitor's network communication according to the network software in use, if
necessary. Enter the service menu: Monitor Setup -Install/Service (16-4-34) - Service (26-23-8)- Frame - Network- Config and set
NOTE: The DRI level 1999 is not selectable in all monitor software versions.
NOTE: If the DRI level is changed, the monitor will restart automatically.
11. Fill out all necessary documentation regarding the new monitor software.
NOTE: The license agreement of monitor software needs to be in accordance with the monitor
software serial number. Make sure you archive the license agreement in a secure location.
NOTE: The first start-up after software loading takes considerably longer.
3.3.7Performing Factory Reset
NOTE: The factory reset is necessary after replacing monitor software.
NOTE: The factory reset will restore all customized defaults, including language selection, to factory
defaults.
1.Press the
2.Select Install/Service and password (16-4-34).
3.Select Service and password (26-23-8).
4.Select Set/ Test and Factory Reset.
5.The monitor will perform an automatic restart. After the restart is completed, restart the monitor
also manually.
Monitor Setup key.
3.3.8Installing the Datex-Ohmeda Network and MemCard Upgrade, U-CMMEM
The following instructions apply to upgrading the monitor with both the MemCard and the Network
options. Detailed upgrade instructions are supplied with the corresponding upgrade product.
NOTE: You can download the option software only onto one monitor. During the downloading of
option software, the serial number of the monitor is written on the software card, and if the
downloading for some reason would fail, the software can be downloaded again onto the same
monitor, but not onto any other monitor.
1.Make sure that the monitor is switched to standby.
2.Detach the cover plate on the network connectors at the rear of the frame unit (above the power
supply unit). Leave the collar on the connectors attached.
3.Open the cover for card drive slots (on the left side of the monitor). Insert the option software
card into one of the card drive slots and press the software card firmly in position.
4.Switch the monitor on.
22
Document no. M1031517-02
Page 35
5.Wait until the normal screen appears and check that a message 'Field upgrade OK' is displayed
as a note of successful upgrade.
6.Remove the option software card.
7.Enter the service menu and make sure that both options were loaded (Monitor Setup
-Install/Service - Service - Frame - Options - "Net+DC").
8.Install the device plate for option software on inside the cover for card drive slots. Close the
cover.
9.Test the monitor with a MemCard and by connecting it to Datex-Ohmeda Network.
10. Fill out all necessary documentation regarding the upgrade.
3.3.9Installing the Datex-Ohmeda Network Upgrade, U-CMNET
The following instructions apply to upgrading the monitor with the Network option. Detailed upgrade
instructions are supplied with the corresponding upgrade product.
NOTE: You can download the option software only onto one monitor. During the downloading of
option software, the serial number of the monitor is written on the software card, and if the
downloading for some reason would fail, the software can be downloaded again onto the same
monitor, but not onto any other monitor.
System installation
1.Make sure that the monitor is switched to standby.
2.Detach the cover plate on the network connectors at the rear of the frame unit (above the power
supply unit). Leave the collar on the connectors attached.
3.Open the cover for card drive slots (on the left side of the monitor). Insert the option software
card into one of the card drive slots and press the software card firmly in position.
4.Switch the monitor on.
5.Wait until the normal screen appears and check that a message 'Field Upgrade OK' is
displayed as a note of successful upgrade.
6.Remove the option software card.
7.Enter the service menu and make sure that the option was loaded (Monitor Setup
-Install/Service - Service - Frame - Options - "Net").
8.Install the device plate for option software on inside the cover for card drive slots. Close the
cover.
9.Test the monitor by connecting it to Datex-Ohmeda Network.
10. Fill out all necessary documentation regarding the upgrade.
3.3.10 Installing the S/5 Wireless LAN Upgrade, U-CMW
The following instructions apply to upgrading the monitor with the Wireless LAN option. Detailed
upgrade instructions are supplied with the corresponding upgrade product.
You can download the option software only onto one monitor. During the downloading of option
software, the serial number of the monitor is written on the software card, and if the downloading for
some reason would fail, the software can be downloaded again onto the same monitor, but not onto
any other monitor.
1.Make sure that the monitor is switched to standby.
2.Open the cover for card drive slots (on the left side of the monitor). Insert the option software
card into one of the card drive slots and press the software card firmly in position.
3.Switch the monitor on.
23
Document no. M1031517-02
Page 36
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
4.Wait until the normal screen appears and check that a message 'Field Upgrade OK' is
displayed as a note of successful upgrade.
5.Remove the option software card.
6.Switch the monitor to standby.
7.Detach the cover for card drive slots and install the antenna cover.
8.Insert the wireless LAN antenna card into the foremost card drive slot.
9.Switch the monitor on.
10. Enter the service menu and make sure that the option was downloaded successfully (Monitor Setup - Install/Service - Service - Frame - Options - "WLAN" or "All").
11. Enter the WLAN service menu and enter appropriate Network ID, WEP Algorithm, Key ID and
Encryption Key information (Monitor Setup - Install/Service - Service - Frame - Network - WLAN - WLAN Config).
12. Check all installed options on the monitor software device plate and attach the Virtual ID sticker
under the device plate at the monitor rear panel.
13. Test the monitor on Datex-Ohmeda Network.
14. Fill out all necessary documentation regarding the upgrade.
3.3.11 Installing the S/5 Recorder Upgrade, U-CMREC1
NOTE: Only S/5 Compact Anesthesia Monitor rev. 02 or later and S/5 Compact Critical Care Monitor
rev. 02 or later can be upgraded with this S/5 Recorder Upgrade kit.
1.Turn the monitor off.
2.Disconnect the power cord.
3.Push the Service Reset button for at least 5 seconds until the service reset indicator LED turns
off (the fan should turn off).
4.Remove the plate that covers the recorder installation place in the side panel.
5.Open the recorder (by pressing the button on the right upper corner at the front of the recorder).
6.Insert the recorder from outside to the side panel (opening button upwards).
7.Attach the recorder to the side panel with two screws at the back of the recorder (the screws can
be accessed inside the recorder).
8.Check that the recorder and the side panel are aligned and the recorder tightly attached.
Figure 7Installation of the U-CMREC1
9.Attach the device plate to the back cover of the monitor, near the other device plates.
24
Document no. M1031517-02
Page 37
3.4S/5 Remote Controller, K-CREMCO
To connect S/5 Remote Controller, K-CREMCO, to an S/5 Compact Monitor:
1.Make sure that the monitor is switched to standby.
2.Connect the K-CREMCO to the 5-pin DIN connector X9 that is located on the right side of the
Compact Monitor display unit.
3.5Anesthesia Record Keeping Keyboard, K-ARKB
To connect Anesthesia Record Keeping Keyboard, K-ARKB, to an S/5 Compact Monitor:
1.Make sure that the monitor is switched to standby.
2.Connect the K-ARKB with an appropriate cable to the 5-pin DIN connector X9 that is located on
the right side of the Compact Monitor display unit.
3.6ARK Barcode Reader, N-SCAN (optional)
3.6.1Connection to Compact Monitor
1.Make sure that the monitor is switched to standby.
2.Connect the N-SCAN with an appropriate cable to the 5-pin DIN connector X9 that is located on
the right side of the Compact Monitor display unit.
System installation
Figure 8ARK Barcode Reader, N-SCAN connected to Compact Monitor
Document no. M1031517-02
25
Page 38
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
3.7S/5 Compact Airway Modules, E-xxxx / M-xxxxx
This chapter provides information for installing Compact Airway Modules E-xxxxx.
Figure 9Compact Airway Module, E-XXXX
Connection to frame
1.Ensure that the module is properly orientated (i.e. module release latch facing downward).
2.Align the module insertion guide slot with the insertion guide.
3.Push the module into the frame until it clicks.
3.7.1Sample gas exhaust
Preventing operation room pollution
When N2O or volatile anesthetics are used, pollution of the operation room by these gases should be
prevented. Connect the sample gas outlet of the monitor to the scavenging system or return it to the
patient circuit.
NOTE: The sample gas from the E-miniC/ M-miniC can only be connected to the scavenging, not
returned to the patient circuit.
Connect the sample gas outlet of the monitor to the scavenging system either
•through the ventilator, or
•directly to the vacuum scavenging system.
CAUTIONStrong scavenging suction may change the operating pressure of the Airway Module
and cause inaccurate readings or internal damage. To prevent this, there must for
example be an opening to room air.
Connect the sample gas outlet only to an open scavenging system where gas is removed in room
pressure.
26
Document no. M1031517-02
Page 39
Scavenging through the ventilator reservoir
Figure 10Scavenging through ventilator reservoir
1.Connect an exhaust line to the sample gas outlet on the module’s front panel.
2.Attach the other end of the line to the ventilator reservoir. Make sure that the reservoir tube
diameter is at least 2 - 3 times larger than the exhaust line.
System installation
SAMPLE GAS OUT
Scavenging through the anesthesia gas scavenging system
Anesthesia machines are equipped with an anesthesia gas scavenging system (AGSS), and in some
machines the sample gas outlet can be connected directly to that.
For example, connect the sample gas outlet to the S/5 Avance:
Figure 11Connecting the gas module to the scavenging connector of S/5 Avance
NOTE: Refer to the anesthesia machine’s documentation to find out where and how the sample gas
can be connected.
Document no. M1031517-02
27
Page 40
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Connecting directly to the vacuum scavenging system
1.Connect the exhaust line to the monitor’s sample gas outlet.
2.Connect the exhaust line only to an open scavenging system where gas is removed at room
pressure. Do not connect the monitor directly to a vacuum scavenging system.
Returning sampling gas to the patient circuit
The sampling gas can also be returned to the patient circuit. If you use the S/5 Anesthesia Delivery
Unit (ADU), you need an optional adapter connected to the patient breathing tubes.
Take special care when returning sample gas to the patient circuit. For further information, please
contact your GE Healthcare distributor.
NOTE: If E-miniC is being used, do not return the sample gas to the patient circuit.
SCAVENGING
SAMPLE GAS OUT
Figure 12Sample gas returned to patient circuit in ADU
28
Document no. M1031517-02
Page 41
System installation
3.7.2Troubleshooting
If a problem occurs during a functional examination, check the components of the monitor according
to the following troubleshooting chart. If the problem persists, please refer to Part II of this Technical
Reference Manual.
TroubleTreatment
Nothing functions.Unplug and re-plug Remote Controller Cable. Also confirm that the cable is intact.
Unplug and re-plug the Power Cord. Also confirm that the cable is intact.
Confirm that the fuses are intact.
A plug-in module does not
function.
Compact Airway Module
does not function.
Remove and replace the module.
Confirm that the desired parameters are configured to be displayed.
Confirm that ‘Occlusion’ or ‘Calibrating Gas Sensor’ messages are not displayed.
Confirm that a D-fend water trap and a sample tube are attached.
Confirm that the desired parameters are configured to be displayed.
Remove and replace the module.
29
Document no. M1031517-02
Page 42
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For your notes:
30
Document no. M1031517-02
Page 43
4Interfacing
External devices can be interfaced with the S/5 Compact Anesthesia and S/5 Compact Critical Care
Monitors via the Compact Monitor’s serial port, via the Interface Module, E-INT / M-INT, and via the
Device Interfacing Solution, N-DISxxx.
The serial port can be used for interface with:
•S/5 Anesthesia Delivery Unit
•Printers
•Computers
An E-INT / M-INT can be used for interface with:
•Datex-Ohmeda monitors
•Various other manufacturers’ monitors
•Various anesthesia machines
Printers and computers can be interfaced via the monitor’s serial port.
Device specific N-DISxxx modules can be used with:
Interfacing
•Ventilators/ anesthesia machines
•Heart-lung machines
•Monitors
•Blood-gas analyzers
4.1Interfacing external monitors via Interface Module, E-INT / M-INT
It is possible to interface Datex-Ohmeda monitors, Critikon Dinamap 1846SX, Abbott Oximetrix 3,
Baxter Explorer and Vigilance, Nellcor N-100, N-200 and N-1000 to the S/5 Compact Anesthesia
Monitor or to the S/5 Compact Critical Care Monitor via Interface Module, E-INT / M-INT. The
parameters that are transferred from external monitors are summarized in tables: Table 1
Transference of parameters, Datex-Ohmeda monitors and Table 2 Transference of parameters,
external monitors.
The Interface Module, E-INT / M-INT, has two serial/analog connectors (X7 and X8).
WARNINGAlways make sure that the combination complies with the international safety
standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
31
Document no. M1031517-02
Page 44
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Table 1Transference of parameters, Datex-Ohmeda monitors
DeviceWaveforms (analog) NumericsAlarms
CardiocapE-INT / M-INT -> CO
Et&Fi Airway gases,
2
Respiration rate, SpO
Pulse rate
Capnomac
Capnomac II
Capnomac UltimaE-INT / M-INT -> CO
E-INT / M-INT -> CO
Pleth
,
Et&Fi Airway gases,
2
Respiration rate
Et&Fi Airway gases,
2
Respiration rate,
Spirometry, SpO
rate, Pleth amplitude
Normocap 200
Normocap 200 OXY
Oscar
Oscar II
Oscar OXY
E-INT / M-INT -> CO
E-INT / M-INT -> CO
Pleth
Et&Fi Airway gases,
2
Respiration rate
,
Et&Fi Airway gases,
2
Respiration rate, SpO2,
Pulse rate, Pleth
amplitude
Satlite
Satlite II
E-INT / M-INT -> CO
SpO2, Pulse rate, Pleth
2
amplitude
Satlite Plus
Satlite TransNoneSpO
, Pulse rate, Pleth
2
amplitude
Table 2Transference of parameters, external monitors
, Pulse
2
None
,
2
None
CO
, O2, Anesthesia
2
agent, Respiration rate,
Apnea, Occlusion, SpO2,
Pulse rate
None
None
None
None
DeviceWaveformsNumericsAlarms
Critikon
Dinamap 1846SX
Abbott
Oximetrix 3
Baxter
Explorer
Baxter
Vigilance
Nellcor
N-100
N-200
N-1000
NoneNIBPNone
NoneSvO
NoneC.O., SvO
NoneC.O., SvO
Pleth (analog)SpO
4.1.1Connection to external Datex-Ohmeda monitors
Use the INT-External Device Cable.
1.Make sure that the power to both monitors is turned off.
2.Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT.
Tighten the finger screws.
3.Connect the 25 pin D-connector to the corresponding connector on the other monitor. Tighten
the finger screws.
/SaO2, CONone
2
REF, TbloodNone
2,
C.C.O., TbloodNone
2,
, Pulse rateNone
2
32
Document no. M1031517-02
Page 45
Interfacing
4.1.2Connection to Critikon Dinamap 1846SX, Abbott Oximetrix 3 and Baxter Explorer
Use the INT-External Device Cable.
1.Make sure that the power to both monitors is turned off.
2.Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT.
Tighten the finger screws.
3.Connect the 25 pin D-connector to the connector on the external monitor. Tighten the finger
screws.
4.1.3Connection to Baxter Vigilance
Use the INT-Baxter Vigilance Cable.
1.Make sure that the power to both monitors is turned off.
2.Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT.
Tighten the finger screws.
3.Connect the other 9 pin D-connector to a corresponding connector on the external monitor.
Tighten the finger screws.
Communication parameters are: baud rate 19200, no parity (none), data bits 8, stop bits 1. Mode of
communication port: IFM out.
4.1.4Connection to Nellcor N-100 and N-1000
Use the Monitor-Nellcor Cable.
1.Make sure that the power to both monitors is turned off.
2.Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT.
Tighten the finger screws.
3.Connect the other connector to the corresponding connector on the external monitor.
Tighten the finger screws. Connect also the mono connectors on the external monitor.
4.1.5Connection to Nellcor N-200
Use the Monitor-Nellcor Cable.
1.Make sure that the power to both monitors is turned off.
2.Connect one of the connectors to the corresponding connector on Interface Module, E-INT /
M-INT. Tighten the finger screws.
3.Connect the other connector to the corresponding connector on the external monitor. Tighten
the finger screws. Connect also the mono connectors on the external monitor.
Connect also the mono connectors on the external monitor.
33
Document no. M1031517-02
Page 46
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
4.2Interfacing external bedside devices via S/5 Device Interfacing
Solutions, N-DISxxx
The Device Interfacing Solution, N-DISxxx provides means for transferring physiological, waveform
and event data from various bedside patient care devices to the Datex-Ohmeda monitoring system.
The real-time and trended data can be displayed on the monitor screen and used for record keeping
purposes. The interfacing module reads the data coming from the external device, converts it to a
suitable format and sends it to the monitor.
See the following table of DIS modules and devices that you can interface with the Device Interfacing
Solution.
NOTE: The Device Interfacing Solution (DIS) is only compatible with the S/5 Compact Anesthesia and
S/5 Compact Critical Care Monitor when the monitor has S/5 Monitor software of rev. 02 or later
installed.
Table 3DIS modules and interfaced devices
Device Ventilators
N-DISEV4
N-DISPRIM
N-DIS7200
N-DIS840
N-DISS300
N-DIS7900
N-DISAEST
N-DISVENT
e
a
Evita 4
a.
Primus
NOTE: Not available in the US
7200 Series Ventilator System
840 Ventilator System
Servo Ventilator 300
7900 SmartVent Ventilator
Aestiva/5
S/5 Aespire
d.
Aestiva/5d.
Aisys Carestation
S/5 Avance
Centiva/5
Engström Carestation
c
d.
d.
d.
b
b.
d
d.
d.
a. Trademark of Dräger Medical AG & Co
b. Trademark of Nellcor Puritan Bennet Inc
c. Trademark of Maquet Critical Care AB part of the Getinge Group (previously trademark of Siemens)
d. Trademark of GE Healthcare Finland Oy
e. Replaced by N-DISVENT
Device Monitors
N-DISOXIM3
N-DISQVUE
N-DISA2000
N-DISVIGIL
N-DISPICCO
Oximetrix 3
QVue /Q2a.
A-2000 Bispectral Index Monitoring System
Baxter-Vigilance
PiCCO-Technology
a
c
d
NOTE: Not available in the US
34
Document no. M1031517-02
b
Page 47
Interfacing
N-DISRGM
N-DISTONO
N-DISWHITE
a. Trademark of Hospira Inc. (previously trademark of Abbott Laboratories)
b. Trademark of Aspect Medical Systems
c. Trademark of Edwards Lifesciences Corporation
d. Trademark of Pulsion Medical Systems
e. Trademark of GE Healthcare Finland Oy
RGM Monitor
Tonocap
Capnomac, Capnomac II
Multicap, Normocap, CD2-O2
Capnomac Ultima
Normocap CD-200
Oscar Oxy, Cardiocap 1GS, Cardiocap 2GS
Satlite, Satlite Trans, Satlite Plus
e
e.
e.
e.
e.
e.
e.
e.
Device Blood gas analyzers
N-DISOPT
AVL Opti CCA
a
Device Heart-lung machines
N-DISHL20
a. Trademark of Diamond Diagnostics Inc
b. Trademark of MAQUET GmbH & Co. KG part of the Getinge Group
Jostra HL-20
b
For specific information on parameters transferred from the interfaced device to the Datex-Ohmeda
monitor and the applicable software versions of the device refer to the Installation guide
accompanying each DIS module.
4.2.1Device Interfacing Solution components
The Device Interfacing Solution consists of:
−a device specific interfacing module
−a device specific cable
−a bus cable
−a connector for another bus cable
−label specifying the external device
4.2.2Connections
Connect the device specific cable to the external device and the bus cable to the monitor's DIS
connector or to another interfacing module.
35
Document no. M1031517-02
Page 48
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
(1) label specifying the external device
(2) LED indicators
(3) black bus cable from another interfacing
module, if needed
(4) grey device specific cable to the
communication port of the external
device
(5) black bus cable to the monitor’s DIS
connector (or to another interfacing
module)
Figure 13Connection cables and LED indicators
WARNINGThe Compact Monitor, interfacing modules and interfaced devices must be
situated in the same patient environment (as defined in IEC 60601-1-1).
WARNINGConnecting electrical equipment together or using the same extension cord for
more than one device may cause their leakage currents to exceed the limits
specified in relevant safety standards. Always make sure that the combination
complies with the international safety standard IEC 60601-1-1 for medical
electrical systems and with the requirements of local authorities.
WARNINGThe manufacturer guarantees a reliable functioning of the devices with tested
software versions only. Always refer to the Installation guide accompanying the
DIS module and verify the compatibility before use.
4.2.3Mounting
The DIS module can be mounted on the side panel of the external device. Also IV pole placement is
possible.
NOTE: As the Device Interfacing Solution works only with the device specified in the label of the
interfacing module, it is recommended that the interfacing module always travels along with the
external device.
36
Document no. M1031517-02
Page 49
Interfacing
For mounting accessories, please refer to the “Supplies and Accessories” catalog. See the figure
below for an example of a device interfacing.
1
4
Figure 14An example of interfacing external devices with Device Interfacing Solution
NOTE: You can connect up to ten (10) interfacing modules to one system simultaneously.
Check the maximum number of modules: one meter cable = max. four ten modules, three meter
cable = max. three modules, six meter cable = one module.
CAUTIONMake sure that the interfacing module is always used in vertical position to prevent
water from entering the module.
CAUTIONMake sure that you are connecting the interfacing module to the device specified in
the label.
4.2.4Selecting the external device
1.Turn off the external device.
2.Connect the interfacing module to the monitor’s connector for S/5 DIS or to another interfacing
module.
3.Connect the device specific cable to the external device and turn the external device on.
4.The monitor identifies the connected device automatically.
37
Document no. M1031517-02
Page 50
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
4.2.5Functional check
There are two ways to check the function of the S/5 Device Interfacing Solution:
1.Press the
Monitor Setup key.
2.Select Interfacing and open the Status Page menu. The status page shows you the current
communication status of the interfacing module (1 - 10).
NOTE: The status message ‘Connected’ appears on the Status Page after you have connected the
external device to the interfacing module and turned it on. Note also that the monitor and the
interfacing module must be operational.
•Check the LED indicators on the interfacing module (the green LED indicates physical
connections, the yellow LED software selections)
:
GREEN YELLOW INDICATION
lit dark
Physical connections between the monitor, interfacing module
and external device are in order and the device has been
selected in the menu.
dark
lit lit
lit
There is something wrong with the physical connections
between the monitor, interfacing module and external device.
The external device has not been selected in the menu.
Physical connections between the monitor, interfacing module
and external device are in order but the external device has not
been selected in the menu.
dark dark
The interfacing module is not connected to the monitor.
4.2.6Selecting the parameter data source
Select the external device via Monitor Setup - Interfacing menu:
•Select the desired parameter, for example Gases.
•Select the desired source by name, for example Aestiva.
NOTE: The name of the device is visible on the list only if the device is correctly connected.
NOTE: Detailed information about interfacing module related mountings, connections and settings is
included in the installation guides that are delivered with the interfacing modules.
38
Document no. M1031517-02
Page 51
4.3Interfacing Datex-Ohmeda Anesthesia Delivery Unit
It is possible to interface the Datex-Ohmeda Anesthesia Delivery Unit to the S/5 Compact Anesthesia
Monitor via the serial interface port. The data link is bi-directional. The parameters transferred to the
Anesthesia Delivery Unit are summarized in Table 4 and the events transferred to the S/5 Compact
Anesthesia Monitor are summarized in Table 5. The events are transferred continuously once a
minute or during the system check procedure and appear in the trend pages or in the anesthesia
record. The transfer of events requires record-keeping software.
4.3.1Interconnection
Use the ADU-AS/3 Monitor Cable.
4.3.2Setting interfacing parameters on the S/5 Anesthesia Delivery Unit
1.Press the Setup key.
2.Select Install password 10.
3.Select Interfacing and choose S/5 AM.
4.Press the
Normal Screenkey.
Interfacing
4.3.3Setting interfacing parameters on the S/5 Compact Anesthesia Monitor
No settings are required on the S/5 Compact Anesthesia Monitor.
Table 4Parameters transferred from S/5 Compact Anesthesia Monitor to S/5 Anesthesia
Delivery Unit
WaveformsNumerics
CO
2
Et & Fi CO2 %
Et & Fi O
O
2
Et & Fi N
Et & Fi AA%
Vol Exp MV
Vol Exp TV
%
2
% difference (I-E)
O%
2
39
Document no. M1031517-02
Page 52
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Table 5Events transferred from S/5 Anesthesia Delivery Unit to the S/5 Compact
It is possible to interface Dräger Cicero and Cato to the S/5 Compact Anesthesia Monitor or to the
S/5 Compact Critical Care Monitor via the Interface Module, E-INT / M-INT.
The parameters transferred to the S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care
Monitor are summarized in Table 6 and Table 7.
NOTE: Some of the summarized parameters (waveforms or numerics) may not be available in all
device versions.
WARNINGAlways make sure that the combination complies with the international safety
standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
40
Document no. M1031517-02
Page 53
Interfacing
4.4.1Interconnection
Use the INT-External Device Cable, except for Julian and Cicero EM, in which case use the
INT-Julian/Cicero EM cable.
1.Make sure that the power to both monitors is turned off.
2.Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT.
Tighten the finger screws.
3.Connect the 25 pin D-connector to a corresponding connector on the anesthesia machine. The
connector is labeled 'RS-232-C'. Tighten the finger screws.
4.4.2Setting communication parameters
Set the communication parameters for Cicero and Cato to RS-232: 9600, e, 8, 1 (MEDIBUS 3.00). In
all cases please refer to the documentation provided by Dräger.
4.4.3Setting interfacing parameters on the S/5 Compact Anesthesia or Compact Critical
Care Monitor
1.Press the Monitor Setup key.
2.Select Install/Service (password 16-4-34).
3.Select Service (password 26-23-8) - Parameters - More...
4.Select Interface and combine a required parameter with the external monitor. Two letters, al,
denote alarm integration. The selection will be automatically stored in permanent memory.
5.Press the
Normal Screen key.
Table 6Parameters transferred from Dräger Cicero monitor to S/5 Compact Anesthesia
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Table 7Parameters transferred from Dräger Cato, Julian and Narkomed 2C (NAD) monitor
to the S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor
SelectionWaveformsNumericsStates
Gases / SSSCO2 (kPa, mmHg, %)CO2 Fi & Et (kPa, %)
O
Fi & Et(%)
2
N2O Fi & Et (%)
RR (CO2) (1/min)
AA Fi & Et (%)
Tidal Volume insp. (1)
Minute Volume exp. (1)
Ppeak (mbar)
Pplat (mbar); not in Narkomed 2C
PEEP (mbar)
SpO
2
Pleth (% full scale)SpO2 (%)
Pulse rate (1/min)
NOTE: The transference of parameters illustrated in Table 4 and Table 5 represents the maximum.
Some of these parameters may not be available in all device versions.
NOTE: The transference of waveforms requires the MEDIBUS real-time extension and, in Cato
monitors, waveforms are available only on port 1.
4.5Interfacing printer
It is possible to interface a laser printer (serial or parallel) with the S/5 Compact Monitors via the
monitors’ serial port. The printer should be PCL5 compatible and should contain at least 2 MB of
memory. Parallel printers require the use of Serial-to-Parallel Converter, order code 78030, model
PI 130-R2 or PI 1115A.
None
Probe Off
WARNINGAlways make sure that the combination complies with the international safety
standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
WARNINGConnecting the power supply cord of the printer to the wall power outlet may
cause the printer leakage current to exceed the limit specified for medical
equipment. A printer must be supplied from an additional transformer providing
at least basic isolation (isolating or separating transformer).
4.5.1Setting S/5 Compact Monitor interface for printers
1.Press the Print/Record key.
2.Select Printer Connection.
3.Select Serial.
4.Press the
Normal Screen key.
4.5.2Connection to serial printers
A serial printer is connected to the serial port connector X11 on the frame unit.
Contact your authorized GE Healthcare distributor for advice on suitable serial printers.
42
Document no. M1031517-02
Page 55
4.5.3Connection to parallel printers
A parallel printer is connected to the serial interface port connector X11 on the frame unit via the
Serial-to-Parallel Converter, order code 78030, model PI130-R2 or PI1115A.
Contact your authorized GE Healthcare distributor for advice on suitable parallel printers.
4.5.4Installing the Serial-to-Parallel Converter
Order code 78030, model PI130-R2
A common serial modem cable, gender changer and parallel printer cable are required when
connecting the serial to parallel converter between the monitor and a parallel printer. The converter
gets power from the connected devices.
1.Make sure that power is switched off on both devices.
2.Connect a standard PC-to-parallel-printer cable (order code 713701) to the parallel printer.
3.Check the DIP-switch settings on the converter (order code 78030, model PI130-R2):
Position
2
ON
3
OFF4OFF
Flow Control:
hardware
LED: Enabled
Parity/D, Bits/S, Bit:
None/8/1
D.Rate (kbps):
115,2
1
OFF
5
6
OFF
OFF7ON8OFF
Interfacing
4.Connect a DB25 female-to-female gender changer (order code 78032) between the printer
cable and converter.
DB9F plugs
into CM1
1234
Standard Async Modem Cable
Gender changer
plugs into
converter
Serial to Parallel
Converter
DB25M end
plugs into gender
changer
DB25F/DB25F
Gender Changer
Centronics
end plugs
into printer
Standard Parallel Printer
Adapter Cable
Parallel printer
Figure 15Connecting S/5 Compact Monitor to printer, converter model PI130-R2
5.Connect a common RS-232 modem cable’s (order code 78031) DB9 connector to the serial
port connector X4 on the monitor, and the DB25 connector to the converter.
Order code 78030, model PI1115A
A common serial modem cable, power adapter and parallel printer cable are required when
connecting the serial to parallel converter between the monitor and a parallel printer. The converter
gets power from an external power supply and requires an external power adapter.
NOTE: The order code 78030 includes the new serial to parallel converter PI1115A and the European
power adapter PS1002-EU (230 VAC/50 Hz). Power adapters for US, UK or Asian markets must be
locally sourced from Black Box Corporation through its international distributor network, see
http://www.blackbox.com - Contact us - Contacts by Country.
43
Document no. M1031517-02
Page 56
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
cam p1 inst serial parallel conv p1115A a
To connect the S/5 Compact Monitor to printer:
1.Make sure that power is switched off on both devices.
2.Connect a standard PC-to-parallel-printer cable (order code 713701) to the parallel printer.
3.Check the DIP-switch settings on the converter (order code 78030, model PI1115A).
SettingPosition
12345678
115.2K BaudONONON
XON/XOFFOFF
8 Data bitsOFF
No parityOFF
Even parityOFF
Serial to Par.OFF
4.Connect a common RS-232 modem cable’s (order code 78031) DB9 connector to the serial
port connector X4 on the monitor, and the DB25 connector to the converter.
5.Connect the converter to external power supply using the power adapter PS1002-xx.
PS1002-EU, Power Adapter, 230 VAC Europe,
for PI1115A (included in p/n 78030),
DB9F plugs
into CM1
plugs to power supply
Centronics
end plugs
into printer
Standard Async Modem Cable
p/n 78031
Serial to Parallel
Converter
p/n 78030 (PI1115A)p/n 713701
Standard Parallel Printer
Adapter Cable
Parallel printer
Figure 16Connecting S/5 Compact Monitor to printer, converter model PI 1115A
4.5.5Connection to printer
Setting the printing parameters
Refer to the documentation provided with the printer.
4.6Interfacing computer
A computer is connected to the serial port connector X11 on the frame unit. Contact your authorized
GE Healthcare distributor for further advice on computer interface.
WARNINGAlways make sure that the combination complies with the international safety
standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
WARNINGConnecting the power supply cord of the computer to the wall power outlet may
cause the computer leakage current to exceed the limit specified for medical
equipment. A computer must be supplied from an additional transformer
providing at least basic isolation (isolating or separating transformer).
44
Document no. M1031517-02
Page 57
4.7Output signals
Analog/ digital output signals on the connectors X4 and X8 can be used for interfacing with other
devices. The pin assignments are illustrated in tables/pictures below.
Table 8Coding element connector, X4
9 pin female D-connectorPinSignal
Interfacing
1
5
9
Table 9Synchronization connector, X8
4 pin female connectorPinSignal
1
6
2
3
4
5
6
7
8
9
1
2
3
4
IDCS1 (chip select)
IDCL (clock)
IDDI (data in)
IDDO (data out)
IDPE (protect enable)
+5Vdc
Direct ECG
Nurse call
GND
GND
Def_Sync
Direct ECG
Pressure Out
4.7.1Digital outputs
The digital output signals are as follows:
Defibrillation Sync (X8 pin 2)
The defibrillation sync signal is generated by the ECG. When activated, the signal is set to a high level
and then set back to a low level after 10 ms. The signal is regenerated only after returning to the low
level. The high level ranges from 2.8 to 5 V, while the low level ranges from 0 to 0.8 V. The delay from
the R wave peak to the start of the signal is 35 ms maximum.
Nurse Call(X4 pin 8)
The Nurse Call signal is generated by red and yellow alarms. When activated, the signal is set to the
high state and remains at the high state until the alarm situation is over or the
pressed. The high state range is from 2.8 to 5 V, while the low state range is from 0 to 0.8 V.
SILENCE ALARM key is
45
Document no. M1031517-02
Page 58
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
4.7.2Analog outputs
Direct ECG (X4 pin 7 and X8 pin 3)
Delay (max.): 15 ms
Gain ECG (out)/ECG (in): 1 V/1 mV
Pacer:5 V and 2 ms pulse
The signal requires input impedance of 100 kΩ.
NOTE: The ECG signal is based on the ECG measurement of the Hemodynamic Modules, M-ESTPR (all
variations), M-NES(12)TPR (all variations), E-PRESTN/ M-PRESTN (all variations). The ECG signal from
the modules is channel 1 (ECG1). The channels 2 and 3 (ECG2, ECG3) are not transmitted. Make sure
that the signal in channel 1 (ECG1) is good enough (extensive QRS for IABP synchronization). For
further information, please contact your authorized GE Healthcare distributor.
Pressure out (X8 pin 4)
−P1 from Hemodynamic Module
NOTE: With monitor software version 03 or earlier, the Invasive pressure output signal is 1 V/100
mmHg, originally ranging from 0 to 300 mmHgand with a delay of approximately 25ms. The signal
requires an input impedance of 100k.
4.7.3S/5 Pressure Temp Module, E-PT or M-PT, output signals
The signal output connector on the Pressure Temp Module, E-PT or M-PT, can be used to interface
some models of IABPs to the S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care
Monitor. The pin assignments are illustrated in Table 10. Please contact your local distributor for more
information.
WARNINGAlways make sure that the combination complies with the international safety
standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
These instructions include procedures for a functional check for Datex-Ohmeda S/5 Compact
Anesthesia Monitor and S/5 Compact Critical Care Monitor. The functional check is recommended to
be performed after monitor installation.
These instructions include a “Functional check form, Datex-Ohmeda S/5 CAM, CCCM” to be filled in
Functional check
when performing the procedures. The symbol
procedure should be signed in the check form. The procedures should be performed in ascending
order, bypassing those that are not applicable for a particular monitor.
All menu selections related to Datex-Ohmeda products are written in following typeface:
e.g. Parameters - Gas Unit.
As you enter the service menus, you need the following passwords:
" in the instructions means that the performed
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
In case you evaluate the measurement accuracy with a patient simulator, add simulator’s accuracy
specification to the one of the monitor.
An electrical safety check and a leakage current test are recommended to be performed prior to the
monitor installation.
5.1Recommended tools
For product(s)ToolOrder No.
Airway modules
Compact Airway Module, E-CAiO(VX)/
M-CAiO(VX)
Compact Airway Module, E-COVX/
M-COVX
Calibration gas and regulator755583/755534*
Calibration gas and regulator755587/755534*
Compact Airway Module, E-CO(V),
M-C, M-CO(V)
E-miniC/ M-miniCCalibration gas and regulator755580/755534*
Tonometry Module, E-TONO/ M-TONO Calibration gas and regulator755580/755534*
Sampling line733251
Tonometrics
TM
catheter
TONO-14F or another suitable
TONO-_F catheter
Luer plug
Pressure manometer
EEG Module, E-EEG/ M-EEGEarphones
EEG simulator90502
48
Document no. M1031517-02
Page 61
For product(s)ToolOrder No.
Functional check
BIS Module, E-BIS/ M-BISBIS simulator or
BIS Sensor simulator
900509
900508
E-ENTROPY/ M-ENTROPYEntropy simulatorN-ES
Entropy sensor cable8002964
E-NMT/ M-NMTNMT simulator871251
M-NMT ElectroSensor888416
M-NMT sensor cable888415
MemCard optionMemCard – Data or Menu893860 (Menu, English)
887045 (Data, English)
NOTE: * Ensure that the calibration gas and regulator are functioning properly before calibration.
Perform annual maintenance on the regulator as required. For more information see section
“Adjustments and calibrations” in Compact Airway modules slot.
For details on recommended accessories see “Supplies and Accessories” catalog.
5.1.1Hemodynamic patient simulators
The following tables present the patient simulators’ compatibility with each hemodynamic module, and the
accessories needed:
Table 11Patient simulators’ compatibility with each hemodynamic module
•Make sure that the monitor is switched to standby.
Disconnect the mains power cord from the monitor.
If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable from
the monitor. Remove any memory cards, if installed.
1.Check all units visually
Check that all parts are intact and that the cables and screws are connected and tightened
properly. Especially check the following parts:
−the display unit locking screw is in a locking position.
−sampling line is connected to the Airway Module, if installed.
Check that modules go in smoothly and lock up properly.
"
50
Document no. M1031517-02
Page 63
5.3Functional inspection
5.3.1General
1.Connect the mains power cord.
Check that the LED above the main switch is lit continuously (full charged batteries) or
periodically (charging batteries).
2.Switch the monitor on.
Check that the monitor starts up properly, i.e. a normal start-up sound is heard from the
loudspeaker, the alarm LEDs turn on and off, and the monitoring screen appears.
No error messages should appear on the screen.
3.Configure the screen for the parameters that are connected.
4.Enter the Service Menu.
When applicable, check from the corresponding Parameters submenu that the Timeouts, Bad
checksums and Bad c-s by mod values of inserted modules are not increasing faster than by 5
per second. Check also that the module memories have passed the internal memory test, i.e.
RAM, ROM and EEPROM all state OK.
If installed, the recorder should record two lines of start-up information.
Functional check
"
Preset the measurement settings for those parameters that are connected, for example:
Recorder - Record Waveforms - Waveform 1 - ECG1
- Waveform 2 - P1
- Waveform 3 - P2
Invasive Pressures - P1 ‘ART’ Setup -- Label – ART
- P2 ‘CVP’ Setup -- Label – CVP
- P3 Setup -Label –PA
- P4 Setup -Label -P4
- P5 Setup -Label -P5
- P6 Setup -Label -P6
Pulse Oximetry -Pleth Scale –AUTO
or
Others -SPO2 Setup - Pleth Scale –AUTO
Airway Gas-Spirometry View -Scaling -Indep.
-Paw Scale –20
-Flow Scale –15
or
Ventil.-Spirometry Setup -Scaling -Indep.
-Paw Scale –20
-Flow Scale –15
Others -Resp Setup -Size -1.0
- Resp Rate Source – AUTO
- Measurement – ON
- Detection Limit – AUTO
51
Document no. M1031517-02
Page 64
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
5.3.2Display
1.Check that the picture on the screen is correct.
"
5.3.3Keyboard(s)
1.Tests with all the connected keyboards:
−Press the
cursor in the menu moves correspondingly. Select Normal Screen and check that the
menu disappears from the screen.
Tests with the Command Board:
−Press the
menu moves correspondingly. Select Normal Screen and check that the menu
disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
Tests with the ARK Keyboard:
−Enter the Keyboard service menu
−Check functioning of the ComWheel.
−Press all keys, except
loudspeaker, or the Message count value in the service menu increases.
−Press the
monitor screen.
Tests with the Remote Controller:
−Enter the Keyboard service menu.
−Check the function of the ComWheel.
−Press all keys. Check that each key produces a sound from the loudspeaker, or the
Message count value in the service menu increases.
Monitor Setupkey. Turn the ComWheel in both directions and check that the
Help key. Turn the ComWheel in both directions and check that the cursor in the
Modify and Print. Check that each key produces a sound from the
Modify and Print keys and check that the corresponding menus appear on the
"
5.3.4Frame unit
1.Check that the frame fan is running.
"
5.3.5Display unit
2.Check that the clock on the screen shows correct time. Readjust the time and date, if necessary.
"
52
Document no. M1031517-02
Page 65
5.3.6Compact Airway Module, E-CXXXXX/ M-CXXXXX
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1.Check that the fan is running.
2.If the module contains membrane keys on the front panel, press each of the keys for at least
one second and check that they are identified.
"
Compact Airway Modules with the Patient Spirometry option
3.Connect a clean spirometry tube and D-lite to the module. Connect the sampling line.
Breathe through the wider side of the D-lite.
Check that the flow waveform moves downwards when you breathe in and upwards when you
breathe out.
"
For all Compact Airway Modules
4.Block the tip of the sampling line with your finger and check that the message ‘Sample line
blocked’ appears on the monitor screen within 30 seconds.
5.Detach the D-fend and check that the message ‘Check D-fend’ appears on the monitor screen
within 30 seconds.
6.Breathe to the sampling line briefly. Check that the CO2 information is updated on the screen.
Functional check
"
5.3.7Single width Airway Module, E-miniC/ M-miniC
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1.Block the tip of the sampling line with your finger and check that the message ‘Sample line
blocked’ appears on the monitor screen within 30 seconds.
2.Detach the Mini D-fend and check that the message ‘Check D-fend’ appears on the monitor
screen within 30 seconds.
Breathe to the sampling line briefly. Check that the CO2 information is updated on the screen.
"
5.3.8Tonometry Module, E-TONO/ M-TONO
1.Press each of the front panel membrane keys for at least one second and check that they are
identified.
"
53
Document no. M1031517-02
Page 66
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
5.3.9Multiparameter Hemodynamic Modules
ECG and RESP measurements
1.Check the function of the front panel membrane key ECG, Lead, if available.
2.Connect an ECG cable to the module. Connect the cable leads to a patient simulator. Check
that all ECG and impedance respiration information is shown on the monitor screen as
configured on the simulator.
Turn the simulator off. Check that the ‘Asystole’ and ‘Apnea’ messages are displayed.
"
Temperature measurement
3.Check the temperature channels with a patient simulator.
Check that temperature measurement information is shown on the monitor screen as
configured on the simulator.
"
Invasive blood pressure measurement
4.Check the function of the front panel membrane keys.
5.Check the InvBP channels with a patient simulator.
6.Zero the InvBP channels and check that the values and waveforms correspond to the simulator
settings.
"
SpO2 measurement
7.Connect an SpO2 finger probe to the module. Check that the message ‘Probe off’ is shown when
the probe is not connected to a finger.
8.Attach the SpO2 probe to your finger. Check that a reading of 95-99 and a pleth waveform
appear on the screen
"
Non invasive blood pressure measurement
9.Check the function of the front panel membrane keys.
10. Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check that the
module identifies the cuff, i.e. the text ‘Adult’ appears in the NIBP digit field for a short time.
Check that the module gives a reasonable measured result.
1.Connect a Nellcor SpO2 finger probe to the module. Check that the message ‘Pulse search ’ is
shown. Check also that the shown message changes to ‘Check probe ’ within 30 seconds.
2.Attach the SpO2 probe on your finger. Check that a reading of 95-100 and a proper SpO2
waveform appear.
1.Connect an OxyTip ® + Reusable Finger Sensor such as Integrated Finger Sensor OXY-F4-N to
the module. Check that the message ‘No Probe’ in the screen changes to ‘Check probe’.
2.Attach the SpO2 probe on your finger. Check that a reading of 95-100 and a proper SpO2
waveform appear.
"
5.3.16 BIS Module, E-BIS/ M-BIS
1.Connect the BIS module to the monitor frame.
Check that ‘Cable off’ is displayed in the BIS waveform field.
2.Connect the DSC to the module.
Check that the ‘No sensor’ message appears in the waveform field.
3.Check the function of the front panel membrane keys.
Open the BIS Setup menu:
NOTE: If the sensor is not connected, the
BIS - BIS Setup and check that Test DSC shows PASS.
Check Sensor key is inoperative.
"
5.3.17 Entropy Module, E-ENTROPY/ M-ENTROPY
1.Check the function of the front panel membrane keys.
2.Connect the Entropy sensor cable and Entropy simulator to the module. Check that ‘Checking
sensor’ text and an image appear in the waveform numeric field. Wait for a while and check that
all sensors show PASS.
3.Check that the EntrEEG waveform and RE and SE values appear on the monitor screen.
"
5.3.18 Data Card and Menu Card function
1.Insert a Data card or a Menu card to the slot.
Check that the corresponding symbol appears on the monitor screen.
"
56
Document no. M1031517-02
Page 69
5.3.19 Recorder
1.Press the Record Wave module key and check that the module starts recording the selected
waveforms. Press the
2.Check that the quality of the recordings is acceptable.
"
5.3.20 Network connection
1.Check that the Mon-Net cable connector and the Identification plug are clean and intact, then
connect them to the frame unit.
Check that the monitor connects to the network, i.e. the network symbol appears on the upper
right-hand corner of the screen. Also a message regarding the connected Central should appear
in the message field on the screen.
"
5.3.21 Wireless Network Option
Functional check
Stop module key to stop recording.
1.Check that the green LED on the wireless antenna card blinks and the WLAN signal strenght
symbol scrolls between zero and full or stays fixed on the monitor screen.
2.Check that the wireless LAN network symbol appears on the upper right-hand corner of the
screen when the monitor connects to the Datex-Ohmeda Network.
NOTE: If the monitor does not connect to the Datex-Ohmeda Network, check the WLAN configuration
on the monitor and on the network.
"
5.3.22 Interface Module, E-INT/ M-INT
1.Make sure that the monitor receives all necessary parameter data from the connected devices.
Check the screen configuration and the related interfacing settings, if necessary.
Monitor Setup – Interfacing
"
5.3.23 Device Interfacing Solution, N-DISxxx
1.Make sure that the monitor receives all necessary parameter data from the connected devices.
Check the screen configuration and the related interfacing settings, if necessary.
Monitor Setup - Interfacing - Status Page
"
57
Document no. M1031517-02
Page 70
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
5.3.24 General
Switch the monitor to standby
•
•Perform final cleaning
•Fill in all necessary documents
"
58
Document no. M1031517-02
Page 71
6General troubleshooting
Monitor not
functioning
Power cord
connected?
No
Connect power
cord
Yes
General troubleshooting
Monitor can be
started now?
No
On/Standby LED
is ON or blinks?
Yes
Frame Unit fan is
running?
No
PMB failure.
Replace PMB.
Yes
No
Yes
Batteries were
empty
Possible AC/DC
unit failure.
Replace AC/DC
unit.
Display Unit
failure.
Monitor can be
started now?
Yes
AC/DC Unit
failure.
No
CM1_trbl_Flowchart.vsd
Figure 17S/5 Compact Monitor general troubleshooting flowchart
Document no. M1031517-02
59
Page 72
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
60
Document no. M1031517-02
Page 73
Appendix A, Functional check form, Datex-Ohmeda S/5 CAM, CCCM
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For your notes:
A-4
Document no. M1031517-02
Page 77
Appendix B, ElectroMagnetic Compatibility
APPENDIX BElectroMagnetic Compatibility
Table 1Guidance and manufacturer’s declaration – electromagnetic emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or
the user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Group 1The S/5™ CAM/ CCCM uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Group 2
(1
The S/5™ CAM/ CCCM must emit electromagnetic energy in order to
perform its intended function. Nearby electronic equipment may be
affected.
Class BThe S/5™ CAM/ CCCM is suitable for use in all establishments,
including domestic establishments and those directly connected to
Class A
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Complies
S/5™ CAM/ CCCM equipped with N-CMW option
B-1
Document no. M1031517-02
Page 78
Table 2Guidance and manufacturer’s declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the
user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Immunity testIEC 60601 test levelCompliance levelElectromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transients/bursts
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
<5 % U
T
(>95 % dip in UT)
for 0.5 cycle
40 % U
T
(60 % dip in UT)
for 5 cycles
70 % U
T
(30 % dip in UT)
for 25 cycles
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV common mode
<5 % U
T
(>95 % dip in UT)
for 0.5 cycle
40 % U
T
(60 % dip in UT)
for 5 cycles
70 % U
T
(30 % dip in UT)
for 25 cycles
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If user of the S/5™ CAM/
CCCM requires continued operation during
power mains interruptions, it is
recommended that the S/5™ CAM/ CCCM
be powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
B-2
Document no. M1031517-02
<5 % U
T
(>95 % dip in UT)
for 5 sec
<5 % U
T
(>95 % dip in UT)
for 5 sec
3 A/m3 A/mPower frequency magnetic field should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
Page 79
Appendix B, ElectroMagnetic Compatibility
Table 3Guidance and manufacturer’s declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the
user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Immunity testIEC 60601 test levelCompliance
level
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
3 Vrms
1 Vrms
(1
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the S/5™ CAM/
CCCM, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d1.2 P=
d3.5 P=
d1.2 P=
80 MHz to 800 MHz
d2.3 P=
800 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
1 V/m
(2
d3.5 P=
80 MHz to 800 MHz
d7 P=
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
the compliance level in each frequency range.
a
should be less than
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
B-3
Document no. M1031517-02
Page 80
NOTE 1At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the S/5™ CAM/ CCCM is used exceeds the applicable RF compliance level above, the S/5™ CAM/ CCCM
should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the S/5™ CAM/ CCCM.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
(1
For impedance RESP measurement. For invasive pressure measurement in M-series modules
except M-PRE(S)TN. The invasive pressure measurement compliance level is dependent on the
transducer used, and only the lower level is guaranteed for all transducers.
(2
For BIS measurement in E-BIS/M-BIS, RESP measurement in M-(P)RE(S)TN and invasive
pressure measurement in M-COPSv/E-COPSv and M-COP/E-COP modules. The invasive
pressure measurement compliance level is dependent on the transducer used, and only the
lower level is guaranteed for all transducers.
B-4
Document no. M1031517-02
Page 81
Appendix B, ElectroMagnetic Compatibility
Table 4Recommended separation distances between portable and mobile RF
communications equipment and the S/5™ CAM/ CCCM
Recommended separation distances between portable and mobile RF communications
equipment and the S/5™ CAM/ CCCM.
The S/5™ CAM/ CCCM is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the S/5™ CAM/ CCCM can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the S/5™ CAM/ CCCM as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
W
0.010.12
0.10.38
11.2
103.8
10012
150 kHz to 80 MHz
d1.2 P=
d3.5 P=
Separation distance according to frequency of transmitter
m
80 MHz to 800 MHz
80 MHz to 800 MHz
d1.2 P=
(1
d3.5 P=
(2
0.12
0.35
(1
0.35
(2
0.38
1.1
(1
1.1
(2
1.2
(1
3.5
3.5 (2
3.8
11
(1
11
(2
12
35
(1
35
(2
d2.3 P=
d7 P=
0.23
0.70
0.73
2.2
7.0
22
70
2.3
7.3
23
(2
(2
(2
(2
(2
(2
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
(1
For impedance RESP measurement. For invasive pressure measurement in M-series modules
except M-PRE(S)TN. The invasive pressure measurement compliance level is dependent on the
transducer used, and only the lower level is guaranteed for all transducers.
(2
For BIS measurement in E-BIS/M-BIS, RESP measurement in M-(P)RE(S)TN and invasive
pressure measurement in M-COPSv/E-COPSv and M-COP/E-COP modules. The invasive
pressure measurement compliance level is dependent on the transducer used, and only the
lower level is guaranteed for all transducers.
Document no. M1031517-02
B-5
Page 82
For your notes:
B-6
Document no. M1031517-02
Page 83
Datex-Ohmeda
S/5TM Compact Anesthesia Monitor
TM
S/5
Compact Critical Care Monitor
Planned Maintenance Instructions
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Datex-Ohmeda Inc.
P.O. Box 7550
Madison, WI 53707-7550, USA
Tel: +1 608 221 1551
Fax: +1 608 222 9147
All specifications are subject to change without notice.
Outside the USA, check local laws for any restriction that may apply.
These instructions include procedures for planned maintenance (PM) for the Datex-Ohmeda S/5
Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor. The Planned maintenance
should be performed once a year.
These instructions include “Planned maintenance check form, S/5 Compact Monitors” to be filled in
when performing the corresponding procedures.
Planned maintenance instructions
The symbol
check form.
The procedures should be performed in ascending order, bypassing those that are not applicable for
a particular monitor.
If you need further information on how to perform a certain Planned maintenance procedure, please
refer to the corresponding slot in the Technical Reference Manual.
All menu selections related to the Datex-Ohmeda monitors are written in the following typeface:
e.g. Parameters - Gas Unit
As you enter the service menus, you need following passwords:
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
In case you evaluate the measurement accuracy with a patient simulator, add the simulator’s
accuracy specification to the one for the monitor.
" in the instructions means that the procedure performed should be signed in the
1.2Recommended tools
For product(s)ToolOrder No.
All Airway modulesFlowmeter
Compact Airway Module, E-CAiO(VX)/
M-CAiO(VX)
Calibration gas and regulator755583/755534*
Compact Airway Module, E-COVX/ M-COVXCalibration gas and regulator755587/755534*
Compact Airway Module, E-CO(V), M-C,
M-CO(V)
E-miniC/ M-miniCCalibration gas and regulator755580/755534*
Tonometry Module, E-TONO/ M-TONOCalibration gas and regulator755580/755534
Sampling line733251
Luer plug
Pressure manometer
Tonometrics
EEG Module, E-EEG/ M-EEGEarphones
EEG simulator90502
2
Document no. M1031518-02
TM
catheter
TONO-14F or another
suitable TONO-_F catheter
Page 89
Planned maintenance instructions
For product(s)ToolOrder No.
BIS Module, E-BIS/ M-BISBIS simulator or
BIS Sensor simulator
E-ENTROPY/ M-ENTROPYSimulator for M-ENTROPYN-ES
Entropy sensor cable8002964
E-NMT/ M-NMTNMT simulator871251
NMT ElectroSensor888416
NMT sensor cable888415
MemCard – Menu893860 (English)
MemCard – Data887045 (English)
All M-ModulesScrewdrivers
All E-ModulesTorx screwdrivers; T8, T10
NOTE: * Ensure that the calibration gas and regulator are functioning properly before calibration.
Perform annual maintenance on the regulator as required. For more information see section
“Adjustments and calibrations” in Compact Airway modules slot.
900509
900508
1.2.1 Hemodynamic patient simulators
The following tables present the patient simulators’ compatibility with each hemodynamic module, and the
accessories needed:
Table 1Patient simulators’ compatibility with each hemodynamic module
Central Processing BoardBattery for SRAM/Timekeeper 197230
Recorder paper74205
Tonometry Module, E-TONO/ M-TONO733382
4
Document no. M1031518-02
Page 91
Planned maintenance instructions
1.4Planned maintenance parts
1.4.1PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX without CO2 absorber
- Anesthesia
PartOrder No.For product(s)
Special tube (Nafion)733382All Airway modules
Ref. gas sticker893110Compact Airway Modules
Filter (2 pcs)886136Compact Airway Modules
Filter assembly896025Compact Airway Modules
Fan filter886236Compact Airway Modules
D-fend O-ring (2 pcs)653125All Airway modules
D-fend (black)876445All Airway modules
Sampling line 2.0 m73318Compact Airway Modules
PM sticker893108All Airway modules
NOTE: Corresponding PM kit is available with Order No. 8001758.
1.4.2PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX without CO2 absorber
- Critical Care
PartOrder No.For product(s)
Special tube (Nafion)733382All Airway modules
Ref. gas sticker893110Compact Airway Modules
Filter (2 pcs)886136Compact Airway Modules
Filter assembly896025Compact Airway Modules
Fan filter886236Compact Airway Modules
D-fend O-ring (2 pcs)653125All Airway modules
D-fend+ (green)893200All Airway modules
Sampling line 2.0 m733162Compact Airway Module, M-COVX
PM sticker893108All Airway modules
NOTE: Corresponding PM kit is available with Order No. 8001759.
Document no. M1031518-02
5
Page 92
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
1.4.3PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX with CO2 absorber
- Anesthesia
PartOrder No.For product(s)
Special tube (Nafion 2 pcs)733382All Airway modules
Ref. gas sticker893110Compact Airway Modules
Filter (2 pcs)886136Compact Airway Modules
Filter assembly896025Compact Airway Modules
Fan filter886236Compact Airway Modules
D-fend O-ring (2 pcs)653125All Airway modules
D-fend (black)876445All Airway modules
Sampling line 2.0 m73318Compact Airway Modules
PM sticker893108All Airway modules
NOTE: Corresponding PM kit is available with Order No. 8001760.
1.4.4PM parts for Compact Airway Modules, E-CXXXXX/ M-CXXXXX with CO2 absorber
- Critical Care
PartOrder No.For product(s)
Special tube (Nafion 2 pcs)733382All Airway modules
Ref. gas sticker893110Compact Airway Modules
Filter (2 pcs)886136Compact Airway Modules
Filter assembly896025Compact Airway Modules
Fan filter886236Compact Airway Modules
D-fend O-ring (2 pcs)653125All Airway modules
D-fend+ (green)893200All Airway modules
Sampling line 2.0 m733162Compact Airway Module, M-COVX
PM sticker893108All Airway modules
NOTE: Corresponding PM kit is available with Order No. 8001761.
6
Document no. M1031518-02
Page 93
Planned maintenance instructions
1.4.5PM parts for Airway Module, E-miniC/ M-miniC
PartOrder No.For product(s)
Special tube (Nafion 2 pcs)733382All Airway modules
Mini D-fend O-ring (2 pcs)656565E-miniC/ M-miniC
Mini D-fend8002174 (pkg of 10 pcs)E-miniC/ M-miniC
Zero valve air filterM1011471E-miniC, every 3 years
Zero valve air filter571421M-miniC
CO2 Sampling line 3.0 m733163E-miniC/ M-miniC
PM sticker893108All Airway modules
For details on recommended accessories see the “Supplies and Accessories” catalog.
Document no. M1031518-02
7
Page 94
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2Planned maintenance check list
2.1Visual inspection/preparation
2.1.1General
NOTE: Wear a static control wrist strap when handling PC boards. Electrostatic discharge may
damage components on the board.
Make sure that the monitor is switched to standby.
Disconnect the mains power cord. If the monitor is connected to the Datex-Ohmeda Network,
disconnect the Mon-Net cable from the monitor. Disconnect the WLAN antenna card, if the
Wireless LAN option is installed.
1.Check all the units visually. Check that all parts are intact and that the cables and screws are
connected and tightened properly.
Especially check the following parts:
−the equipotential tap and all the screws are tightened properly.
−the display unit locking screw is in a locking position.
Check that modules go in smoothly and lock up properly in all module slots.
Insert the WLAN antenna card back into the foremost card slot, if the Wireless LAN option is
installed.
2.Replace the batteries, if necessary.
The manufacturer recommendations are:
−Replace the SRAM/Timekeeper battery on the CPU board every 8 years.
NOTE: The Factory Reset must be performed if the SRAM/Timekeeper battery is replaced.
NOTE: If the monitor is used for frequent short transportation, refresh the batteries every two
weeks and replace the batteries every six months.
3.Clean or replace the fan filters in the frame unit and display unit.
4.Check that the fuses are of the correct rating.
"
2.1.2Compact Airway Module, E-CXXXXX/ M-CXXXXX
1.Detach the module box.
Check that all cables and tubes are connected properly and that there are no loose objects
inside the module.
NOTE: The tubes connected to the Oxygen board pressure transducers should not be pressed
too deep.
NOTE: Make sure that the tubes are not in contact with the sampling pump or the O
its springs.
2.Install the PM Kit:
−Replace the special tubes (Nafion™).
−Replace the Ref. filter assembly.
−Replace the filters in the pneumatic unit (1 or 2 pcs).
8
Document no. M1031518-02
sensor, or
2
Page 95
−Check the D-fend O-rings and replace them, if necessary.
−Replace the D-fend and sampling line.
−Clean or replace the fan filter.
NOTE: Use only approved sampling lines to ensure proper functioning.
Use a 2 m/7 ft. sampling line with Compact Airway Modules E-/ M-COVX and E-/ M-CAiOVX.
3.The manufacturer recommendation is to replace the CO
absorber, if necessary.
Note: All Compact Airway Modules do not contain a CO
"
Connect the Compact Airway Module to a module slot with a long gas interface cable (the
grounding plates of the cable should be removed for the procedure).
2.1.3Single-width Airway Module, E-miniC/ M-miniC
1.Detach the module box.
2.Check that all cables and tubes are connected properly and that there are no loose objects
inside the module.
3.Install the PM Kit:
−Replace the special tube (Nafion™).
−Replace the Zero valve air filter every three years.
−Check the D-fend body connector O-rings and replace them, if necessary.
−Replace the Mini D-fend and the sampling line.
NOTE: Use only approved sampling lines to ensure proper functioning.
4.Connect the module to a module slot with a long gas interface cable (the grounding plates of the
cable should be removed for the procedure).
Planned maintenance check list
absorber every 4 years. Replace the
2
absorber.
2
"
2.1.4Tonometry Module, E-TONO/ M-TONO
1.Detach the module box.
Check that all cables and tubes are connected properly and that there are no loose objects
inside the module.
2.Replace the Special tube (Nafion 2 pcs).
Reattach the module box and check that the latch moves properly. Plug the module back into
the module frame.
"
2.1.5Recorder Unit
1.Clean the recorder.
−Open the paper compartment hatch and remove the paper roll, if installed.
−Remove any paper chaff from the paper compartment.
9
Document no. M1031518-02
Page 96
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
−Clean the thermal printhead and the small glass window in front of the static brush with a
cotton swab dipped in isopropyl alcohol, if necessary.
NOTE: Avoid contact with the rubber paper roller. Be careful to limit the application of alcohol to
the thermal printhead and the window.
−Reinstall the paper roll.
"
2.1.6Modules with NIBP measurement
1.Check the NIBP pump filter. Replace the filter, if necessary.
Note: M-NIBP Rev. 00-03 do not contain a NIBP pump filter.
Plug the module back into the frame.
"
2.2Functional inspection
2.2.1General
1.Connect the mains power cord. Check that the LED above the mains switch is lit continuously
(full charged batteries) or periodically (charging batteries).
2.Switch the monitor on.
Check that the monitor starts up properly, i.e. the alarm LEDs and the display unit fan turn on
shortly, normal start-up sound is heard from the loudspeaker and the monitoring screen
appears.
No error messages should appear on the screen.
3.Configure the screen for the parameters that are connected.
Check that all the connected modules are recognized, i.e. the required parameter information is
shown on the screen.
If installed, the recorder should record two lines of start-up information.
Preset the measurement settings for those parameters that are connected, for example:
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2.2.2Display
1.Check that the picture on the screen is adjusted correctly.
"
2.2.3Keyboard(s)
1.Tests with the Command Board:
−Press the
menu moves correspondingly. Select Normal Screen and check that the menu
disappears from the screen.
Check the rest of the menu keys by pressing them one by one.
Tests with the ARK Keyboard:
−Enter the Keyboard service menu
−Check the function of the ComWheel.
−Press all keys, except
loudspeaker, or the ‘Message count’ value in the service menu increases.
−Press the
monitor screen.
Tests with the Remote Controller:
−Enter the Keyboard service menu.
−Check the function of the ComWheel.
−Press all keys. Check that each key produces a sound from the loudspeaker, or the
Message count value in the service menu increases.
Help key. Turn the ComWheel in both directions and check that the cursor in the
Modify and Print. Check that each key produces a sound from the
Modify and Print keys and check that the corresponding menus appear on the
2.2.4Display Unit
1.Check that the clock on the screen shows correct time.
Readjust the time and date, if necessary.
2.Enter the Service Log service menu.
Check the content of the Service Log for possible problems.
"
2.2.5Frame Unit
1.Check that the Frame unit fan is running.
"
2.2.6Compact Airway Module, E-CXXXXX/ M-CXXXXX
1.Check that the fan is running.
2.If the module contains membrane keys on the front panel, press each of the keys for at least
one second and check that they are identified.
NOTE: The Compact Airway Module membrane keys require S/5 main software.
12
Document no. M1031518-02
Page 99
Planned maintenance check list
3.Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter the
Compact Airway Module General service menu.
Check that the module configuration displayed corresponds with the Compact Airway Module
type being used.
4.Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster
than by 5 per second.
5.Enter the Gases service menu: Check that the ‘Ambient’ value displayed corresponds with the
current ambient pressure (±20 mmHg).
6.Check that the flow measurement offset, i.e. the sample ‘Zero’ value displayed is within ±10
ml/min.
7.Perform a sampling system leak test.
8.Check the flow rates. Adjust the sampling flow, if necessary.
9.Perform a gas calibration:
NOTE: For maximum accuracy, a warm-up time of 30 minutes is recommended.
NOTE: For correct measurement values, modules need different amounts of oxygen in the
calibration. If you do not use the recommended calibrating gases, the calibration does not
succeed.
"
Anesthesia Agent
Compact Airway Modules w/ the option
10. Enter the Gases service menu.
Feed calibrating gas (order code 755583) continuously for at least 30 seconds and check that
the ‘ID’ in the service menu shows ‘DES’ and that the value for ‘ID unrel.’ is lower than 50.
"
Patient Spirometry
Compact Airway Modules w/ the option
11. Enter the Spirometry service menu.
Connect a clean spirometry tube and D-lite to the module.
Perform the spirometry leak test.
12. Connect the sampling line. Breathe through the wider side of the D-lite. Check that the flow
waveform moves downwards when you breathe in, and upwards when you breathe out.
"
General
For all Compact Airway Modules
Switch the monitor to standby and reassemble the module.
•
NOTE: Attach the plastic cover to the CPU Board before fitting the module box. Make sure that
the grounding claws in the front panel frame make good contact with the module box.
13
Document no. M1031518-02
Page 100
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
NOTE: When reassembling the module, make sure that the tubes are not pinched between the
module box and internal parts.
Insert the Compact Airway Module into the frame. Switch the monitor back on and wait until the
message ‘Calibrating gas sensor’ disappears from the screen.
13. Block the tip of the sampling line with your finger and check that the message ‘Sample line
blocked’ appears on the monitor screen within 30 seconds.
14. Detach the D-fend and check that the message ‘Check D-fend’ appears on the monitor screen
within 60 seconds.
"
2.2.7Single-width Airway Module, E-miniC/ M-miniC
1.Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter the Gas
Unit General service menu.
NOTE: The Airway Module M-miniC requires S/5 main software L-CANE02(A) / L-CICU02(A) or
later.
2.Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster
than by 5 per second.
3.Enter the Gases service menu: Check that the ‘Ambient’ value displayed corresponds with the
current ambient pressure (±20 mmHg).
4.Check that the flow measurement offset, i.e. the sample ‘Zero’ value displayed is within ±10
ml/min.
5.Perform a sampling system leak test.
6.Check the flow rates. Adjust the sampling flow, if necessary.
7.Block the tip of the sampling line with your finger and check that the message ‘Sample line
blocked’ appears on the monitor screen within 30 seconds.
Remove the Mini D-fend and check that the message ‘Check D-fend’ appears on the screen
within 30 seconds.
8.Perform a gas calibration:
NOTE: For maximum accuracy, a warm-up time of 30 minutes is recommended.
"
2.2.8Tonometry Module, E-TONO/ M-TONO
1.Enter the TONO service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster
than by 5 per second.
Check that the memories of the module have passed the internal memory test, i.e. RAM, ROM
and EEPROM all state OK.
Check that the ‘general error status’, ‘module pneumatics error status’, ‘module hardware error
status’ and ‘testbit status’ are all zero.
2.Check the function of the front panel
3.Perform the pressure sensor calibration.
4.Perform the system test.
14
Document no. M1031518-02
Start-Stop and Lab Data membrane keys.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.