Anesthesia recordkeeping keyboard for, K-ARKB, Keyboard
Interface Board, B-ARK and ARK Barcode Reader, N-SCAN
1
2
3
4
5
6
7
8
9
10
11
12
8001020 –5
8001021 –4
8005676
8003476-2
8003934-1
8003487-3
8004390
8005571-1
8001005-5
Memory Module, M-MEM
Interface Module, M-INT
Device Interfacing Solution, N-DISxxxx
BIS Module, M-BIS
Remote Controllers, K-REMCO, K-CREMCO
Single-width Airway Module, M-miniC
Entropy Module, M-ENTROPY
PRETSN Module
Airway Modules, G-AiOV, G-AiO, G-AOV, G-AO, Gas Interface Board
B-GAS
Document No. M1025359
13
14
15
16
17
18
19
20
21
i
Datex-Ohmeda S/5 Modules
ii
Document No. M1025359
INTRODUCTION
This Technical Reference Manual provides information for the maintenance and service of the
Datex-Ohmeda S/5™ M-Modules, record keeping keyboard, remote controllers and Device
Interfacing System. These Datex-Ohmeda devices are designed for use with S/5™ Anesthesia
Monitor, S/5™ Critical Care Montor, S/5™ Compact Anesthesia Monitor, and S/5™ Compact
Critical Care Monitor.
Please see also the Technical Reference Manual of the S/5 monitor for system specific information
e.g. related documentation, conventions used, symbols on equipment, safety precautions, system
description, system installation, interfacing, functional check and planned maintenance.
For more detailed information about compatibility with different monitor software types and levels
se the "Introduction" chapter of the slot of the module or other device.
Introduction
S/5™ parameter modules, Device Interfacing Solution, Record keeping keyboard and Remote Controller
Document No. M1025359
1
Datex-Ohmeda S/5 Modules
Notes to the reader
This Technical Reference Manual is intended for service personnel and engineers who will perform
service and maintenance procedures on the Datex-Ohmeda S/5 Anesthesia Monitor,S/5 Critical
Care Monitor, S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor.
This Technical Reference Manual completes the Technical Reference Manual of the S/5 Anesthesia
Monitor and S/5 Critical Care Monitor or S/5 Compact Anesthesia Monitor and S/5 Compact
Critical Care Monitor.
• Document number M1025359 is the order number for the whole printed manual. This
manual includes Technical Reference Manual Slots and every slot has own document
number.
• The Technical Reference Manual, S/5 Modules gives detailed descriptions of paramter
modules and other products that can be used with all S/5 modular monitors. Service check
for each product is included in these slots.
For monitor or system specific information see:
The Technical Reference Manual of the S/5 Anesthesia Monitor and S/5 Critical Care Monitor or
The Technical Reference Manual of S/5 Compact Anesthesia Monitor and S/5 Compact Critical
Care Monitor.
The manufacturer reserves the right to change product specifications without prior notice. Although
the information in this manual is believed to be accurate and reliable, the manufacturer assumes
no responsibility for its use.
GE Healthcare Finland Oy (GE) assumes no responsibility for the use or reliability of its software in
equipment that is not furnished by GE.
2
Document No. M1025359
Introduction
Conventions used
Throughout this manual, the following conventions are used to distinguish procedures or elements
of text:
Sign the check form after performing the procedure.
"
Hard KeysHard key names on the Command Board, the Remote Controller, and modules are written in bold
D-O Sans (12 pt) typeface, e.g.
Menu ItemsMenu items are written in bold italic, D-O Sans (11 pt) typeface, e.g. ECG Setup.
‘Messages’ Messages displayed on the screen are enclosed in single quotes, e.g. ‘Please wait’.
ChaptersWhen referring to different chapters in the same manual, the chapter name is written in italic
typeface and is enclosed in double quotes, e.g. chapter “Cleaning and Care.”
Other documents
When referring to different documents, the document name is written in italic typeface, e.g. refer to
User’s Reference Manual.
ECG.
Hypertext linksHypertext links on PDF versions are written in blue color.
WARNING Warnings are written in bold typeface (13 pt), for example:
WARNING Use only hospital-grade electrical outlets and power cord.
CAUTION Cautions are written in the following way (13 pt):
CAUTION The circuit boards contain sensitive integrated circuits that can be damaged by an
electrostatic discharge. Careful handling of the boards is therefore essential.
Document No. M1025359
3
Datex-Ohmeda S/5 Modules
4
Document No. M1025359
S/5
S/5
S/5
Datex-Ohmeda Hemodynamic modules
TM
NE12STPR Module, M-NE12STPR (rev. 02)
TM
NE12STR Module, M-NE12STR (rev. 02)
TM
NE12TPR Module, M-NE12TPR (rev. 02)
TM
S/5
NESTPR Module, M-NESTPR (rev. 01)
TM
S/5
S/5
S/5
NESTR Module, M-NESTR (rev. 01)
TM
NETPR Module, M-NETPR (rev. 01)
TM
ESTPR Module, M-ESTPR (rev. 04)
TM
S/5
S/5
ESTR Module, M-ESTR (rev. 04)
TM
ETPR Module, M-ETPR (rev. 04)
Technical Reference Manual Slot
Datex-Ohmeda, Inc.
P.O. Box 7550, Madison
WI 53707-7550, USA
Tel. 1-608-221-1551 Fax 1-608-222-9147
www.us.datex-ohmeda.com
mailto:product.support.ussub@us.datex-ohmeda.com
All specifications are subject to change without notice.
Document No. 800 1008-6
October 2003
Datex-Ohmeda Division, Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.
Table of contents
TABLE OF CONTENTS
HEMODYNAMIC MODULES
TABLE OF CONTENTSi
Table of figuresiii
Introduction1
1Specifications3
1.1 General specifications ..............................................................................................................................3
2.2 Main components...................................................................................................................................11
2.3 Connectors and signals...........................................................................................................................25
2.3.1 Module bus connector....................................................................................................................25
2.3.2 Front panel connectors...................................................................................................................26
2.3.3 Test points on boards .....................................................................................................................27
3Service Procedures29
3.1 General service information.....................................................................................................................29
3.2 Service check .........................................................................................................................................29
5.1 NIBP service menu .................................................................................................................................56
5.2 ECG service menu ..................................................................................................................................64
5.2.1 ECG setup menu ............................................................................................................................66
5.3 STP service menu ...................................................................................................................................67
5.3.1 STP calibration menu .....................................................................................................................69
6Spare Parts70
6.1 Spare parts list.......................................................................................................................................70
Figure 17Exploded view of M-ESTP Module ......................................................................................70
Figure 18Exploded view of M-NESTPR Module..................................................................................74
Figure 19Exploded view of M-NE12STPR Module..............................................................................78
Document No. 8001008-6
iii
Datex-Ohmeda S/5 monitors
This page intentionally left blank.
iv
Document No. 8001008-6
INTRODUCTION
This Technical Reference Manual Slot provides information for the maintenance and service of the
hemodynamic modules. The S/5 M-ESTPR/-ESTR/-ETPR and S/5 M-NE12STPR/-NE12STR/NE12TPR/-NESTPR/-NESTR/-NETPR are double width modules designed for use with S/5
monitors. Later in this manual modules may be referred to w/o the system name S/5 for simplicity.
Please refer to the Technical Reference Manual of the S/5 monitor for information related to
system e.g. related documentation, conventions used, symbols on equipment, safety precautions,
system description, system installation, interfacing, functional check and planned maintenance.
The M-ESTPR/-ESTR/-ETPR and M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR
modules provide general hemodynamic parameters
S/5 Hemodynamic modules
NOTE: Do not use identical modules in
the same monitor simultaneously.
The following modules are considered
identical:
M-ESTP/-EST/-ETP,
M-ESTPR/-ESTR/-ETPR,
M-NESTPR/-NESTR/-NETPR,
M-NE12STPR/-NE12STR/-NE12TPR
child60 ±10 mmHg / 2 min.
infant40 ±10 mmHg / 1 min.
Cuff widthsplease see User’s Guide
3
Document No. 8001008-6
Datex-Ohmeda S/5 monitors
1.2.2 ECG
Lead selection, 12-lead ECGI, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Lead selection, oher modulesI, II, III, aVR, aVL, aVF, V
Sweep speeds 12.5, 25, 50 mm/sec
DISPLAY FILTER
Diagnostic, 12-lead ECG0.05 to 150 Hz
Diagnostic, other modules0.05 to 100 Hz
Monitoring0.5 to 30 Hz (-3 dB, with 50 Hz reject filter)
ST filter0.05 to 30 Hz (-3 dB, with 50 Hz reject filter)
HEART RATE FROM ECG
Range30 to 250 bpm
Accuracy±5 bpm or ±5 %, whichever is greater
Resolution1 bpm
Update interval5 s
Averaging time10 s
0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)
0.05 to 40 Hz (-3 dB, with 60 Hz reject filter)
ST LEVELS (in main software)
ST level range-9 to +9 mm (-0.9 to +0.9 mV)
Resolution0.1 mm (0.01 mV)
Averagingcalculated from 8 QRS
SYNCHRONIZATION
Direct ECGanalog output of ECG, 1 V/1 mV
Pacer5 V and 0.5 to 2.5 ms pulse, < 30 ms after pacer peak
Defibrillator5 V and 10 ms pulse, < 35 ms after R-point synchronization
1.2.3 Pulse oximetry
Measurement range40 to 100 %
Accuracy100 to 80 %, ±2 digits
(% SpO
Display resolution1 digit = 1 % of SpO
Display averaging time20, 10 sec, beat-to-beat
Pulse beep pitchvaries with SpO
The monitor is calibrated over the measurement range against functional saturation SpO
PULSE RATE FROM PLETH
Measurement range 30 to 250 bpm
Accuracy30 to 100, ±5 bpm,
Resolution1 bpm
Display averaging10 s
2±1 SD)
1
80 to 50 %, ±3 digits
50 to 40 %, unspecified
2
2 level
func.
2
100 to 250, ±5 %
1
1 SD (standard deviation) = 68 % of all readings in the specified range in stable conditions.
4
Document No. 8001008-6
Adjustable pulse beep volume.
PLETH WAVEFORM
Scales2, 5, 10, 20, 50 mod%, Auto
Start up scale is 20 mod% if AUTO is not selected to be the default setting.
1.2.4 Temperature
Measurement range10 to 45 °C (50 to 113 °F)
(In rev. ESTP 03/ EST 02/ETP 02 or earlier: 15 to 45 °C (59 to 113 °F))
Measurement accuracy±0.1 °C (25 to 45.0 °C)
Display resolution0.1 °C (0.1 °F)
Temperature testautomatic (every 10 min)
Probe typecompatible with YSI 400 series
1.2.5 Invasive blood pressure
Measurement range-40 to 320 mmHg
Measurement accuracy±2 mmHg or ±5 %
Zero adjustment range±150 mmHg
Calibration range±20 %
Scalesupper limit is adjustable between 10 and 300 mmHg in steps of
S/5 Hemodynamic modules
±0.2 °C (10 to 24.9 °C)
10. Lower limit is 10 % of selected upper limit below zero.
Sweep speed12.5, 25, 50 mm/s
DIGITAL DISPLAY
Range-40 to 320 mmHg
Resolution±1 mmHg
WAVEFORM DISPLAY
Range-30 to 300 mmHg
PULSE RATE FROM ARTERIAL PRESSURE
Measurement range30 to 250 bpm
Resolution1 bpm
Accuracy±5 bpm or ±5 % whichever is greater
1.2.6 Respiration
NOTE:The respiration measurement is intended for patients over three years old
Measurement range4 to 120 bpm
Accuracy±5 bpm or ±5 %
Resolution1 bpm
Averaging time30 s
Update interval10 s
RESPIRATION WAVEFORM
Sweep Speeds6.25 mm/s and 0.625 mm/s
Document No. 8001008-6
5
Datex-Ohmeda S/5 monitors
1.3 Technical specifications
1.3.1 NIBP
Deflation rate, PR dep.5 to 13 mmHg/s
Inflation time20 to 185 mmHg, 1 to 5 s
Over pressure limit, stops measurement after 2 seconds
The safety valve limits the maximum cuff pressure to 320 mmHg in adult/child mode or 165 mmHg
in infant mode. Independent timing circuit limits pressurizing (>15 mmHg) time to 2 minutes 10
seconds maximum in adult/child mode, and 1 minute 5 seconds in infant mode.
Zeroing to ambient pressure is done automatically.
Inflation pressure is adjusted according to the previous systolic pressure, typically 40 mmHg
above. If the systolic pressure is not found, inflation pressure is increased typically 50 mmHg.
Max. measurement timeadult2 min
child2 min
infant1 min
Pressure transducer accuracy is better than ±3 mmHg or ±2 % whichever is greater.
Max. error ±4 mmHg.
Protection against electrical
shockType BF defibrillation proof
Defibrillation protection5000 V, 360 J
Recovery time2 s
Input impedance>2.5 MΩ (10 Hz)
CMRR>100 dB (ST)
System noise<40 µV (p-p, RTI)
Allowable offset±300 mVDC
Gain range0.2 to 5.0 cm/mV
Pacemaker pulse detection2 to 500 mV, 0.5 to 2 ms pulses
6
Document No. 8001008-6
Protection against electrical
shockType CF defibrillator proof
1.3.3 Pulse oximetry
Protection against electrical
shockType BF defibrillation proof
1.3.4 Temperature
Measurement accuracy±0.1 °C (25.0 to 45.0 °C)
Protection against electrical
shockType CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on
transducer/probe used. Please refer to the transducer/probe specification.
1.3.5 Invasive blood pressure
DIGITAL DISPLAY AVERAGING
Digital displays Art and P1 are averaged over 5 seconds and updated at 5 seconds intervals. All
other pressures have respiration artifact rejection.
S/5 Hemodynamic modules
±0.2 °C (10.0 to 24.9 °C)
Accuracy±5 % or ±2 mmHg, whichever is greater
Transducer and input sensitivity
Input voltage 5VDC
max current 20 mA
Filter0 to 4 - 22 Hz adjustable
Zero set accuracy±1 mmHg
Calibration resolution±1 mmHg
Zero timeless than 15 s
Protection against electrical
shockType CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on
transducer/probe used. Please refer to the transducer/probe specification.
1.3.6 Respiration
Excitation frequency, 12-lead ECG62.5 kHz
Excitation frequency, other modules31.25 kHz
Breath detectionautomatic, range 0.3 to 6 Ω manually adjustable minimum
Input dynamic range0.2 to 6 Ω
Input impedance range100 to 5000 Ω
Respiration Ratemin. 4 bpm
Lead off detection>3 MΩ
5 µV/V/mmHg
detection: 0.2, 0.4, 0.6, 0.8, 1.0
max. 120 bpm
Document No. 8001008-6
7
Datex-Ohmeda S/5 monitors
2 FUNCTIONAL DESCRIPTION
2.1 Measurement principle
2.1.1 NIBP
NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure.
The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is
inflated with a pressure slightly higher than the presumed systolic pressure, and deflated at a
speed based on the patient’s pulse, collecting data from the oscillations caused by the pulsating
artery. Based on these oscillations, values for systolic, mean, and diastolic pressures are
calculated.
The following parts are necessary for the NIBP measurement:
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical
potential produced with electrodes placed on the surface of the body.
ECG reflects:
• electrical activity of the heart
• normal/abnormal function of the heart
• effects of anesthesia on heart function
• effects of surgery on heart function
See the User's Guide or theUser's Reference Manual for electrodes positions and other
information.
2.1.3 Pulse oximetry
A pulse oximeter measures the light absorption of blood at two wavelengths, one in the near
infrared (about 900 nm) and the other in the red region (about 660 nm) of the light spectrum.
These wavelengths are emitted by LEDs in the SpO
peripheral tissue and is finally detected by a PIN-diode opposite the LEDs in the probe. The pulse
oximeter derives the oxygen saturation (SpO
between the relative absorption at the two wavelengths and the arterial oxygen saturation SaO
probe, the light is transmitted through
2
) using an empirically determined relationship
2
.
2
8
Document No. 8001008-6
In order to measure the arterial saturation accurately, pulse oximeters use the component of light
absorption giving variations synchronous with heart beat as primary information on the arterial
saturation.
S/5 Hemodynamic modules
A general limitation of pulse oximetry is that due to the use of only two wavelengths only two
hemoglobin species can be discriminated by the measurement.
The modern pulse oximeters are empirically calibrated either against fractional saturation
frac;
SaO
2
2
=
2
fracSaO
HbO
2
++
binDyshemogloHbHbO
Formula 1
or against functional saturation SaO
HbO
=
2
funcSaO
Functional saturation is more insensitive to changes of carboxyhemoglobin and methemoglobin
concentrations in blood.
The oxygen saturation percentage SpO
against functional saturation SaO
measurement relative to SaO2func can be maintained even at rather high concentrations of
carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeters are not able
to correctly measure oxygen content of the arterial blood at elevated carboxyhemoglobin or
methemoglobin levels.
Plethysmographic pulse wave
The plethysmographic waveform is derived from the IR signal and reflects the blood pulsation at
the measuring site. Thus the amplitude of the waveform represents the perfusion.
Pulse rate
The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The
signals are filtered to reduce noise and checked to separate artifacts.
func;
2
2
HbHbO
2
+
measured by the Datex-Ohmeda module is calibrated
2
func. The advantage of this method is that the accuracy of SpO
2
Formula 2
2
Probe
The standard probe is a finger clamp probe which contains the light source LEDs in one half and
the photodiode detector in the other half. Different kinds of probes are available from DatexOhmeda.
9
Document No. 8001008-6
Datex-Ohmeda S/5 monitors
Intensity of
transmitted
light
I
max (DC-component)
I
I
min
max
AC-component
Transmitted
light
Incident light
Emitter
RED
Detector
No pulsation
IRED
Pulsatile blood
SpO sensor cable
2
Variable absorption
due to pulse added
volume of arterial
blood
Arterial blood
Venous blood
Tissue
Time
SpO sensor connector
2
6
GND
7
I
LED
4
5
GND
8
V
B
R
C
1
I
S
9
GND
Figure 2Absorption of infrared light in the finger probe parts layout and schematic
2.1.4 Temperature
The temperature is measured by a probe whose resistance varies when the temperature changes,
called NTC (Negative Temperature Coefficient) resistor.
The resistance can be measured by two complementary methods:
10
Document No. 8001008-6
diagram
•Applying a constant voltage across the resistor and measuring the current that flows
through it
•Applying a constant current through the resistor and measuring the voltage that is
generated across it.
In Datex-Ohmeda modules the two methods are combined in the form of a voltage divider. The
NTC-resistor is connected in series with a normal resistor and a constant voltage is applied across
them. The temperature dependent voltage can be detected at the junction of the resistors, thus
producing the temperature signal from the patient. The signal is amplified by analog amplifiers and
further processed by digital electronics.
2.1.5 Invasive blood pressure
To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive
pressure setup, consisting of connecting tubing, pressure transducer, an intravenous bag of normal
saline all connected together by stopcocks, is attached to the catheter. The transducer is placed at
the same level with the heart, and is electrically zeroed.
The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor
interprets the voltage signal so that pressure data and pressure waveforms can be displayed.
2.1.6 Respiration
Impedance respiration is measured across the thorax between ECG electrodes. The respiration
signal is made by supplying current between the electrodes and by measuring the differential
current from the electrodes. The signal measured is the impedance change caused by breathing.
From these impedance changes, respiration rate is calculated, and the respiration waveform is
displayed on the screen.
S/5 Hemodynamic modules
2.2 Main components
2.2.1 M-ESTPR/-ETPR/-ESTR modules
T1
T2
ECG+
Resp
Figure 3Front panel of M-ESTPR
The M-ESTPR, M-ETPR, and M-ESTRmodules contain two main PC boards, the STP board and the
ECG board. They work independently. Both of them have their own processor and software EPROM.
Some components on the boards are not used in ETPR and ESTR modules.
SpO
P1
P2
2
ECG
Lead
Start
Wedge
Zero
P1
Zero
P2
In the M-ESTPR module, additionally, there are two small boards, the SP input and the ECG input
boards, attached to the front panel of the module. The front panel has six connectors and four keys.
11
Document No. 8001008-6
Datex-Ohmeda S/5 monitors
The connectors are two for temperature measurement, two for invasive blood pressure
measurement, one for ECG, and one for SpO
measurement. The keys are for ECG lead, Start
2
Wedge, P1 zero, and P2 zero.
In the M-ETPR module, there are two small boards, the ECG input board and the 2P input board
attached to the front panel of the module. The front panel has five connectors and four keys. The
connectors are two for temperature measurement, two for invasive blood pressure measurement,
and one for ECG
measurement. The keys are for ECG lead, Start Wedge, P1 zero, and P2 zero.
In the M-ESTR module, there are two small boards: the S input board and the ECG input board,
attached to the front panel of the module. The front panel has four connectors and one key. The
connectors are two for temperature measurement, one for ECG, and one for SpO
The key is for ECG lead select.
NOTE: M-ESTP rev. 03, M-ETP rev. 02 and M-EST rev. 02 and all earlier revisions have separate T
and SP input boards.
The M-NESTPR, M-NETPR, and M-NESTRmodules contain three main PC boards, the STP board,
the ECG board, and the NIBP board. They work independently. Each of these has their own
processor and software EPROM.
The M-NE12STPR, M-NE12TPR, and M-NE12STR contain three main PC boards, The STP board,
the ECG board and the NIBP board. They work independently. Each of them has their own
processor. The STP board and NIBP board have software EPROM. In the ECG board the software is
in flash memory. The STP and NIBP boards are the same as in M-NESTPR module but the ECG
board and ECG input board are different.
12
Document No. 8001008-6
In the M-NESTPR module, there are two small boards, the SP input and the ECG input board
attached to the front panel of the module. The front panel has seven connectors and four keys. The
connectors are two for temperature measurement, two for invasive blood pressure measurement,
one for ECG, one for NIBP, and one for SpO
measurement. The keys are for NIBP Auto On/Off,
2
NIBP Start/Cancel, P1 zero, and P2 zero. The structure of M-NE12STPR is similar except the ECG
board and ECG input board are different.
In the M-NETPR module, there are two small boards, the 2P input board and the ECG input board,
attached to the front panel of the module. The front panel has six connectors and four keys. The
connectors are two for temperature measurement, two for invasive blood pressure measurement,
one for ECG, and one for NIBP. The keys are for Auto On/Off, Start/Cancel, P1 zero, and P2 zero.
The structure of M-NE12TPR is similar except the ECG board and ECG input board are different.
In the M-NESTR module, there are two small boards, the ECG input board and the S input board,
attached to the front panel of the module. The front panel has five connectors and two keys. The
connectors are two for temperature measurement, and one for SpO
and one for NIBP. The keys are for Auto On/Off, Start/Cancel. The structure of M-NE12STR is
similar except the ECG board and ECG input board are different.
2.2.3 NIBP board
Cuff
Zero
valve
Exhaust
valve 2
Joining chamber
Exhaust valve 1
Check valve
Bleed valve
Pump
S/5 Hemodynamic modules
measurement, one for ECG,
2
Pump
and
valve
M
driver
B1B2
to/from module bus
X
Pressure
transducers
converter
Power-up
reset
RS485
1
EEPROM
(calibration
Figure 5NIBP board functional block diagram
Pressure transducers
The NIBP board contains two pressure transducers. They are of piezoresistive type. One is used for
measuring the pressure of the blood pressure cuff and the pressure fluctuations caused by arterial
wall movement (B1). The other is used for detection of cuff hose type, cuff loose and cuff occlusion
situations etc. (B2). The transducers are internally temperature compensated. They are supplied by
a constant voltage and their output voltage changes up to 40 mV max. (50 kPa, 375 mmHg).
AD-
interf.
data)
Write protection switch
Overpressure control
CPU
80C51FA
Internal
watchdog
Address
decoder
Address
latch
Address bus
Watchdog
timer
RAMEPROM
Databus
+15 VD
Front
panel
keys
Software
control
NESTPR_NIBP_board_blck_dgrm .vsd
13
Document No. 8001008-6
Datex-Ohmeda S/5 monitors
Signal processing
Two signals from the pressure transducers are amplified and sent to the A/D converter. After the
converter, digitized signals are sent to the microprocessor for data processing. Before the
converter, one of the signals is used to adjust the offset to the pressure safety level.
The NIBP board is controlled with a 80C51FA microprocessor at 16 MHz oscillator frequency.
Memory
NIBP program memory (EPROM) size is 128k × 8. RAM size is 32k × 8 bit and it stores variable
values of the NIBP measurement. EEPROM is size 64 × 16 bit and is used to store the calibration
values for the pressure transducers, the pulse valve constants gained during measurements, the
PC board identification, and module serial number.
Software control
Software controls valves and pump. In addition to the individual on/off signals for each component
there is a common power switch for the valves and the pump that can be used at pump/valve
failures.
In addition to external RS485 reset line the microprocessor system is equipped with its own powerup reset. See the section in the ECG board’s description: “RS485 communication”
Watchdog timer
The NIBP board is equipped with a software independent safety circuit to disconnect supply
voltages from the pump and the valves if the cuff has been pressurized longer than the preset time.
As soon as the cuff pressure rises over a specified pressure limit, timer starts counting. The timer is
adjusted to stop the pump and open the valves after 2 minutes 10 seconds in adult/child mode
and after 1 minute 5 seconds in infant mode.
Valves
Exhaust valves are used for emptying the cuff and the joining chamber after the measurement.
Exhaust valve 1 is also used as safety valve in infant mode. The valve opens at 165 mmHg. Exhaust
valve 2 is also used as safety valve in adult mode and opens at 320 mmHg.
The bleed valve is used for emptying the cuff during measurement. The zero valve is used for
connecting the pressure transducer B1 to open air.
Power supply section
All connections are established via 25-pin connector (D-type, female). The module needs +5 V,
±15 V, and +15 VD (dirty) power supply to operate. The pump and the valves use separate +15 VD
power line. The supply voltages are generated in the power supply section of the S/5 monitor. The
reference voltages ±5 V
and +10 V
ref
are generated on the NIBP board.
ref
14
Document No. 8001008-6
2.2.4 ECG board in 3-and 5-lead measurement
Patient signals are connected to overload protection circuits (resistors and gas-filled surge
arresters) and analog switches to instrumentation amplifiers. Then the signals are amplified by 480
and limited by slew rate. Then they are A/D-converted, analyzed and transferred to module bus in
digital form.
PATIENT
Overload protection
RL F CN
Defibrillation
detector
uP
uP
uP
Slew-rate
detector
HP
filter
Antialias
Analog switches
Slew-rate
detector
uP
HP
filter
uP
Antialias
uP
uP
Slew-rate
detector
HP
filter
Antialias
uP
uP
Slew-rate
detector
HP
filter
Antialias
S/5 Hemodynamic modules
Lead selection logic
controlled by ECG lead selection
signals from uP
Resp
amplifier
Sync. Rectifier
31 kHz
from uP
Figure 63- and 5- lead ECG board block diagram
Analog ECG section
The ECG cable is connected to connector pins E1 to E6 on the input board which contains an
overload protection circuit. Leads are connected to amplifiers via analog switches. States of the
switches depend on the cable type. Lead-off, noise and pacemaker are detected by a slew rate
ISOLATION
Supply
voltages
Power
source
Power
source
Opto-
coupler
Opto-
coupler
to STP Board
Amplifier
A/D
Micro-
processor
(uP)
Direct
ECG
Direct
ECG
NESTPR_ECG_brd_blck_dgrm.vsd
15
Document No. 8001008-6
Datex-Ohmeda S/5 monitors
detector. Lower frequency is determined by high pass (HP) filter 0.5 Hz (monitor bandwidth) or
0.05 Hz (diagnostic or ST- bandwidth).
Respiration section
3-lead cableThe analog switches control the current supply source of the impedance respiration measurement,
and the lead selection for the 3-lead cable can be seen from the following table:
Table 2Lead selection and coding for the 3-lead cable
Selected leadCurrent source betweenSignal measured from
IR - LN
IIR - NL
IIIL - NR
Position on
IEC standard codingAAMI standard coding
body surface
right armR = redRA = white
left armL = yellowLA = black
left legF = greenLL = red
5-lead cableWhen the 5-lead cable is used, the current source is between L-F and the signal is measured from
the N, independently on the lead selection.
The respiration amplifier consist of the operational amplifiers, and the components around them.
There is an analog switch for controlling the gain of the first stage of the preamplifier. Synchronous
rectifier consists of the analog switches, which are used for detecting the respiration signal from 31
kHz amplitude modulated raw signal. The amplifier stage consists of the differential amplifier and
the last amplifier. The differential amplifier consists of the operational amplifiers and the
components around them. This stage is AC-coupled on both sides for minimising the offset
voltages. The last amplifier is used for amplifying the signal derived from differential amplifier
stage. The respiration signal is zeroed at the beginning of the measurement. Zeroing is also used
for fast recovering the measurement after the motion artefact. This is done in amplifier section.
NOTE: The respiration measurement is switched OFF for 20 seconds when defibrillation is detected
at the defibrillation detector.
Microprocessor section
Microprocessor contains RAM and EPROM memories. The processor uses external EEPROM
memory. The microprocessor’s internal 8-channel A/D-converter converts the ECG-signals to
digital form. See the section in ECG board’s description: “RS485 communication
Serial communication
Communication with the module bus is made through RXD and TXD pins. See the section in STP
board’s description: “Serial communication”.
16
Document No. 8001008-6
Loading...
+ 876 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.