Datex-Ohmeda S-5 M-Modules Technical reference manual

Page 1
Datex-Ohmeda
S/5™ M-Modules
Technical Reference Manual
Datex-Ohmeda, Inc.
P.O. Box 7550, Madison WI 53707-7550, USA Tel. 1-608-221-1551 Fax 1-608-222-9147
All specifications are subject to change without notice.
Document No. M1025359
GE Healthcare Finland Oy
Helsinki, Finland
P.O. Box 900
FIN-00031 GE, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com www.gehealthcare.com
© 2005 Copyright General Electric Company
Page 2

Responsibility of the manufacturer

GE Healthcare Finland Oy(GE) is responsible for the safety, reliability and performance of the equipment only if:
modifications, service and repairs are carried out by personnel authorized by GE.
the electrical installation complies with appropriate requirements.
the equipment is used in accordance with the User’s Guide and serviced and maintained in
accordance with the Technical Reference Manual.

Trademarks

Datex, Ohmeda, S/5, D-lite, D-lite+, Pedi-lite, Pedi-lite+, Mini D-fend, D-fend, D-fend+, OxyTip+, MemCard, ComWheel, EarSat, FingerSat, FlexSat, PatientO
and Tonometrics are trademarks of GE Healthcare Finland Oy. All other product and company names are
property of their respective owners.

Copyright

© 2005 Copyright General Electric Company
All specifications subject to change without notice.
, Entropy, Patient Spirometry
2
Page 3
Table of contents

Master Table of Contents

Technical Reference Manual, S/5™ M-Modules
M1025359
For S/5™ Anesthesia Monitor and Critical Care Monitor and S/5™ Compact Anesthesia Monitor and Compact Critical Care Monitor
Document No. Updated Updated Description
8001008 -6
8001009 –5
8001010 –4
8001011 –4
8001012 –4
8001013 –4
8001014 –4
8001015 –5
8001016 –3
8005426 –1
8001018 –4
8003840 –1
Hemodynamic Modules, M-NE12STPR, M-NE12STR, M-NE12TPR, M-NESTPR, M-NESTR, M-NETPR, M-ESTPR, M-ESTR, M-ETPR
Compact Airway Modules, M-CAiOVX, M-CAiOV,M-CAiO, M-COVX, M-COV, M-CO, M-C
Tonometry Module, M-TONO
EEG Module, M-EEG and EEG Headbox, N-EEG
Cardiac Output Modules, M-COP and M-COPSv
Pressure Module, M-P, Pressure Temp Module, M-PT
Dual Pressure Module, M-PP
NIBP Module, M-NIBP
Recorder Module, M-REC
Oxygen Saturation Modules, M-NSAT, M-OSAT
NeuroMuscular Transmission Module, M-NMT
Anesthesia recordkeeping keyboard for, K-ARKB, Keyboard Interface Board, B-ARK and ARK Barcode Reader, N-SCAN
1
2
3
4
5
6
7
8
9
10
11
12
8001020 –5
8001021 –4
8005676
8003476-2
8003934-1
8003487-3
8004390
8005571-1
8001005-5
Memory Module, M-MEM
Interface Module, M-INT
Device Interfacing Solution, N-DISxxxx
BIS Module, M-BIS
Remote Controllers, K-REMCO, K-CREMCO
Single-width Airway Module, M-miniC
Entropy Module, M-ENTROPY
PRETSN Module
Airway Modules, G-AiOV, G-AiO, G-AOV, G-AO, Gas Interface Board B-GAS
Document No. M1025359
13
14
15
16
17
18
19
20
21
i
Page 4
Datex-Ohmeda S/5 Modules
ii
Document No. M1025359
Page 5

INTRODUCTION

This Technical Reference Manual provides information for the maintenance and service of the Datex-Ohmeda S/5™ M-Modules, record keeping keyboard, remote controllers and Device Interfacing System. These Datex-Ohmeda devices are designed for use with S/5™ Anesthesia Monitor, S/5™ Critical Care Montor, S/5™ Compact Anesthesia Monitor, and S/5™ Compact Critical Care Monitor.
Please see also the Technical Reference Manual of the S/5 monitor for system specific information e.g. related documentation, conventions used, symbols on equipment, safety precautions, system description, system installation, interfacing, functional check and planned maintenance.
For more detailed information about compatibility with different monitor software types and levels se the "Introduction" chapter of the slot of the module or other device.
Introduction
S/5™ parameter modules, Device Interfacing Solution, Record keeping keyboard and Remote Controller
Document No. M1025359
1
Page 6
Datex-Ohmeda S/5 Modules

Notes to the reader

This Technical Reference Manual is intended for service personnel and engineers who will perform service and maintenance procedures on the Datex-Ohmeda S/5 Anesthesia Monitor,S/5 Critical Care Monitor, S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor.
This Technical Reference Manual completes the Technical Reference Manual of the S/5 Anesthesia Monitor and S/5 Critical Care Monitor or S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor.
Document number M1025359 is the order number for the whole printed manual. This
manual includes Technical Reference Manual Slots and every slot has own document number.
The Technical Reference Manual, S/5 Modules gives detailed descriptions of paramter
modules and other products that can be used with all S/5 modular monitors. Service check for each product is included in these slots.
For monitor or system specific information see: The Technical Reference Manual of the S/5 Anesthesia Monitor and S/5 Critical Care Monitor or The Technical Reference Manual of S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor.
The manufacturer reserves the right to change product specifications without prior notice. Although the information in this manual is believed to be accurate and reliable, the manufacturer assumes no responsibility for its use.
GE Healthcare Finland Oy (GE) assumes no responsibility for the use or reliability of its software in equipment that is not furnished by GE.
2
Document No. M1025359
Page 7
Introduction
Conventions used
Throughout this manual, the following conventions are used to distinguish procedures or elements of text:
Sign the check form after performing the procedure.
"
Hard Keys Hard key names on the Command Board, the Remote Controller, and modules are written in bold
D-O Sans (12 pt) typeface, e.g.
Menu Items Menu items are written in bold italic, D-O Sans (11 pt) typeface, e.g. ECG Setup.
‘Messages’ Messages displayed on the screen are enclosed in single quotes, e.g. ‘Please wait’.
Chapters When referring to different chapters in the same manual, the chapter name is written in italic
typeface and is enclosed in double quotes, e.g. chapter “Cleaning and Care.”
Other documents
When referring to different documents, the document name is written in italic typeface, e.g. refer to User’s Reference Manual.
ECG.
Hypertext links Hypertext links on PDF versions are written in blue color.
WARNING Warnings are written in bold typeface (13 pt), for example:
WARNING Use only hospital-grade electrical outlets and power cord.
CAUTION Cautions are written in the following way (13 pt):
CAUTION The circuit boards contain sensitive integrated circuits that can be damaged by an
electrostatic discharge. Careful handling of the boards is therefore essential.
Document No. M1025359
3
Page 8
Datex-Ohmeda S/5 Modules
4
Document No. M1025359
Page 9
S/5
S/5 S/5
Datex-Ohmeda Hemodynamic modules
TM
NE12STPR Module, M-NE12STPR (rev. 02)
TM
NE12STR Module, M-NE12STR (rev. 02)
TM
NE12TPR Module, M-NE12TPR (rev. 02)
TM
S/5
NESTPR Module, M-NESTPR (rev. 01)
TM
S/5 S/5
S/5
NESTR Module, M-NESTR (rev. 01)
TM
NETPR Module, M-NETPR (rev. 01)
TM
ESTPR Module, M-ESTPR (rev. 04)
TM
S/5 S/5
ESTR Module, M-ESTR (rev. 04)
TM
ETPR Module, M-ETPR (rev. 04)
Technical Reference Manual Slot
Datex-Ohmeda, Inc.
P.O. Box 7550, Madison WI 53707-7550, USA Tel. 1-608-221-1551 Fax 1-608-222-9147
www.us.datex-ohmeda.com
mailto:product.support.ussub@us.datex-ohmeda.com
All specifications are subject to change without notice.
Document No. 800 1008-6
October 2003
Datex-Ohmeda Division, Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 394 11 Fax +358 9 146 3310
www.datex-ohmeda.com
Instrumentarium Corp. All rights reserved.
Page 10
Page 11
Table of contents
TABLE OF CONTENTS
HEMODYNAMIC MODULES
TABLE OF CONTENTS i
Table of figures iii
Introduction 1
1 Specifications 3
1.1 General specifications ..............................................................................................................................3
1.2 Typical performance .................................................................................................................................3
1.2.1 NIBP................................................................................................................................................3
1.2.2 ECG.................................................................................................................................................4
1.2.3 Pulse oximetry..................................................................................................................................4
1.2.4 Temperature.....................................................................................................................................5
1.2.5 Invasive blood pressure ....................................................................................................................5
1.2.6 Respiration ......................................................................................................................................5
1.3 Technical specifications............................................................................................................................6
1.3.1 NIBP................................................................................................................................................6
1.3.2 ECG.................................................................................................................................................6
1.3.3 Pulse oximetry..................................................................................................................................7
1.3.4 Temperature.....................................................................................................................................7
1.3.5 Invasive blood pressure ....................................................................................................................7
1.3.6 Respiration ......................................................................................................................................7
2 Functional Description 8
2.1 Measurement principle .............................................................................................................................8
2.1.1 NIBP................................................................................................................................................8
2.1.2 ECG.................................................................................................................................................8
2.1.3 Pulse oximetry..................................................................................................................................8
2.1.4 Temperature...................................................................................................................................10
2.1.5 Invasive blood pressure ..................................................................................................................11
2.1.6 Respiration ....................................................................................................................................11
2.2 Main components...................................................................................................................................11
M-ESTPR/-ETPR/-ESTR modules...............................................................................................................11
M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules .................................................12
2.2.3 NIBP board ....................................................................................................................................13
2.2.4 ECG board in 3-and 5-lead measurement........................................................................................15
2.2.5 ECG board in 12-lead measurement ...............................................................................................17
2.2.6 ECG filtering...................................................................................................................................19
2.2.7 STP board ......................................................................................................................................20
2.3 Connectors and signals...........................................................................................................................25
2.3.1 Module bus connector....................................................................................................................25
2.3.2 Front panel connectors...................................................................................................................26
2.3.3 Test points on boards .....................................................................................................................27
3 Service Procedures 29
3.1 General service information.....................................................................................................................29
3.2 Service check .........................................................................................................................................29
Document No. 8001008-6
i
Page 12
Datex-Ohmeda S/5 monitors
3.2.1 Recommended tools......................................................................................................................29
3.2.2 Recommended parts......................................................................................................................29
3.3 Disassembly and reassembly..................................................................................................................40
3.3.1 M-ESTPR, M-ESTR, and M-ETPR modules........................................................................................40
3.3.2 M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules.......................................40
3.4 Adjustments and calibrations..................................................................................................................41
3.4.1 Pressure safety level detection “OFFSET”.........................................................................................41
3.4.2 NIBP calibrations ...........................................................................................................................41
3.4.3 Temperature calibration .................................................................................................................43
3.4.4 Invasive pressure calibration ..........................................................................................................43
4 Troubleshooting 45
4.1 Troubleshooting charts ...........................................................................................................................45
4.1.1 NIBP..............................................................................................................................................45
4.1.2 NIBP error code explanation ...........................................................................................................48
4.1.3 ECG...............................................................................................................................................49
4.1.4 Pulse oximetry (SpO
4.1.5 Temperature ..................................................................................................................................50
4.1.6 Invasive blood pressure..................................................................................................................51
4.1.7 Impedance respiration ...................................................................................................................52
4.2 Troubleshooting flowcharts .....................................................................................................................53
4.2.1 M-NE12STPR and M-NESTPR module troubleshooting.....................................................................53
4.2.2 M-ESTPR, M-ESTR, and M-ETPR module troubleshooting .................................................................54
).....................................................................................................................49
2
5Service Menu 55
5.1 NIBP service menu .................................................................................................................................56
5.1.1 NIBP demo menu...........................................................................................................................57
5.1.2 NIBP calibration menu....................................................................................................................58
5.1.3 NIBP safety valve menu..................................................................................................................59
5.1.4 NIBP pulse valve menu...................................................................................................................60
5.1.5 NIBP buttons/leds menu................................................................................................................61
5.1.6 NIBP pneumatics menu..................................................................................................................62
5.1.7 NIBP watchdog menu.....................................................................................................................63
5.2 ECG service menu ..................................................................................................................................64
5.2.1 ECG setup menu ............................................................................................................................66
5.3 STP service menu ...................................................................................................................................67
5.3.1 STP calibration menu .....................................................................................................................69
6 Spare Parts 70
6.1 Spare parts list.......................................................................................................................................70
6.1.1 M-ESTP rev. 01, M-ETP rev. 00, M-EST rev. 00 .................................................................................70
6.1.2 M-NESTPR rev. 00, M-NETPR rev. 00, M-NESTR rev. 00....................................................................74
6.1.3 M-NE12STPR rev. 00, M-NE12STR rev. 00, M-NE12TPR rev. 00.......................................................78
7 Earlier Revisions 81
APPENDIX A 83
Service Check Form 1
ii
Document No. 8001008-6
Page 13
Table of contents
TABLE OF FIGURES
Figure 1 S/5 NE12STPR Module, M-NE12STPR ................................................................................1
Figure 2 Absorption of infrared light in the finger probe parts layout and schematic diagram..............10
Figure 3 Front panel of M-ESTPR.....................................................................................................11
Figure 4 Front panel of M-NESTPR ..................................................................................................12
Figure 5 NIBP board functional block diagram.................................................................................13
Figure 6 3- and 5- lead ECG board block diagram ...........................................................................15
Figure 7 12-lead ECG measurement block diagram.........................................................................17
Figure 8 STP board block diagram...................................................................................................20
Figure 9 Temperature measurement principle .................................................................................21
Figure 10 Pressure measurement principle .......................................................................................21
Figure 11 Pulse oximetry measurement block diagram ......................................................................22
Figure 12 Serial communication and opto isolation of M-NESTPR/-NE12STPR ...................................23
Figure 13 Serial communication and opto isolation of M-ESTPR.........................................................24
Figure 14 Module bus connector (X1) pin layout ................................................................................25
Figure 15 M-NE12STPR and M-NESTPR module troubleshooting flowchart .........................................53
Figure 16 M-ESTPR Module Troubleshooting Flowchart ......................................................................54
Figure 17 Exploded view of M-ESTP Module ......................................................................................70
Figure 18 Exploded view of M-NESTPR Module..................................................................................74
Figure 19 Exploded view of M-NE12STPR Module..............................................................................78
Document No. 8001008-6
iii
Page 14
Datex-Ohmeda S/5 monitors
This page intentionally left blank.
iv
Document No. 8001008-6
Page 15
INTRODUCTION
This Technical Reference Manual Slot provides information for the maintenance and service of the hemodynamic modules. The S/5 M-ESTPR/-ESTR/-ETPR and S/5 M-NE12STPR/-NE12STR/­NE12TPR/-NESTPR/-NESTR/-NETPR are double width modules designed for use with S/5 monitors. Later in this manual modules may be referred to w/o the system name S/5 for simplicity.
Please refer to the Technical Reference Manual of the S/5 monitor for information related to system e.g. related documentation, conventions used, symbols on equipment, safety precautions, system description, system installation, interfacing, functional check and planned maintenance.
The M-ESTPR/-ESTR/-ETPR and M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules provide general hemodynamic parameters
S/5 Hemodynamic modules
NOTE: Do not use identical modules in the same monitor simultaneously. The following modules are considered identical: M-ESTP/-EST/-ETP, M-ESTPR/-ESTR/-ETPR, M-NESTPR/-NESTR/-NETPR, M-NE12STPR/-NE12STR/-NE12TPR
Figure 1 S/5 NE12STPR Module, M-NE12STPR
Table 1 Options of S/5 hemodynamic modules
Parameter NE12STPR NESTPR NE(12)
STR
12
12-lead ECG (•)(•)
N
NIBP •••
E
ECG • •••••
S
Pulse oximetry •••
T
Two temperatures • •••••
P
Two invasive blood pressures ••
R
Impedance respiration • •••••
NE(12) TPR
ESTPR ESTR ETPR
Document No. 8001008-6
1
Page 16
Datex-Ohmeda S/5 monitors
NOTE: 12-lead ECG measurement requires Display Controller, B-DISP.
NOTE: M-ESTP rev. 01, M-EST rev. 00 and M-ETP rev. 00 work only with S-STD93, S-STD94, S-ARK94, S-STD95, S­ARK95, S-STD96 and S-ARK96 software.
2
Document No. 8001008-6
Page 17
1 SPECIFICATIONS
1.1 General specifications
Module size 75 × 180 × 112 mm W × D × H3.0 × 7.1 × 4.4 in Operation temperature 10 to 40 °C / 50 to 104 °F
@ M-ESTPR/-ETPR/-ESTR
Module weight 0.6 kg / 1.3 lbs Power consumption 6 W
@ M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR
Module weight 1 kg Power consumption about 9 W
S/5 Hemodynamic modules
1.2 Typical performance
1.2.1 NIBP
NOTE:Non-invasive blood pressure measurement is intended for patients weighing over 5 kg (11 lb.)
Oscillometric measurement principle.
Measurement range adult 25 to 260 mmHg
Pulse rate range accepted 30 to 250 bpm
Measurement interval 1, 2.5, 3, 5, 10, 15, 30, 60 min (=1h), 2h, 4h
Typical measuring time adult 23 s
Initial inflation pressure adult 185 ±10 mmHg
Venous stasis adult 80 ±10 mmHg / 2 min.
child 25 to 195 mmHg infant 15 to 145 mmHg
infant 20 s
child 150 ±10 mmHg infant 120 ±10 mmHg
child 60 ±10 mmHg / 2 min. infant 40 ±10 mmHg / 1 min.
Cuff widths please see User’s Guide
3
Document No. 8001008-6
Page 18
Datex-Ohmeda S/5 monitors
1.2.2 ECG
Lead selection, 12-lead ECG I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 Lead selection, oher modules I, II, III, aVR, aVL, aVF, V Sweep speeds 12.5, 25, 50 mm/sec
DISPLAY FILTER
Diagnostic, 12-lead ECG 0.05 to 150 Hz Diagnostic, other modules 0.05 to 100 Hz
Monitoring 0.5 to 30 Hz (-3 dB, with 50 Hz reject filter)
ST filter 0.05 to 30 Hz (-3 dB, with 50 Hz reject filter)
HEART RATE FROM ECG
Range 30 to 250 bpm Accuracy ±5 bpm or ±5 %, whichever is greater Resolution 1 bpm Update interval 5 s Averaging time 10 s
0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)
0.05 to 40 Hz (-3 dB, with 60 Hz reject filter)
ST LEVELS (in main software)
ST level range -9 to +9 mm (-0.9 to +0.9 mV) Resolution 0.1 mm (0.01 mV) Averaging calculated from 8 QRS
SYNCHRONIZATION
Direct ECG analog output of ECG, 1 V/1 mV Pacer 5 V and 0.5 to 2.5 ms pulse, < 30 ms after pacer peak Defibrillator 5 V and 10 ms pulse, < 35 ms after R-point synchronization
1.2.3 Pulse oximetry
Measurement range 40 to 100 % Accuracy 100 to 80 %, ±2 digits
(% SpO
Display resolution 1 digit = 1 % of SpO Display averaging time 20, 10 sec, beat-to-beat Pulse beep pitch varies with SpO
The monitor is calibrated over the measurement range against functional saturation SpO
PULSE RATE FROM PLETH
Measurement range 30 to 250 bpm Accuracy 30 to 100, ±5 bpm,
Resolution 1 bpm Display averaging 10 s
2 ±1 SD)
1
80 to 50 %, ±3 digits 50 to 40 %, unspecified
2
2 level
func.
2
100 to 250, ±5 %
1
1 SD (standard deviation) = 68 % of all readings in the specified range in stable conditions.
4
Document No. 8001008-6
Page 19
Adjustable pulse beep volume.
PLETH WAVEFORM
Scales 2, 5, 10, 20, 50 mod%, Auto
Start up scale is 20 mod% if AUTO is not selected to be the default setting.
1.2.4 Temperature
Measurement range 10 to 45 °C (50 to 113 °F) (In rev. ESTP 03/ EST 02/ETP 02 or earlier: 15 to 45 °C (59 to 113 °F))
Measurement accuracy ±0.1 °C (25 to 45.0 °C)
Display resolution 0.1 °C (0.1 °F) Temperature test automatic (every 10 min) Probe type compatible with YSI 400 series
1.2.5 Invasive blood pressure
Measurement range -40 to 320 mmHg Measurement accuracy ±2 mmHg or ±5 % Zero adjustment range ±150 mmHg Calibration range ±20 % Scales upper limit is adjustable between 10 and 300 mmHg in steps of
S/5 Hemodynamic modules
±0.2 °C (10 to 24.9 °C)
10. Lower limit is 10 % of selected upper limit below zero.
Sweep speed 12.5, 25, 50 mm/s
DIGITAL DISPLAY
Range -40 to 320 mmHg Resolution ±1 mmHg
WAVEFORM DISPLAY
Range -30 to 300 mmHg
PULSE RATE FROM ARTERIAL PRESSURE
Measurement range 30 to 250 bpm Resolution 1 bpm Accuracy ±5 bpm or ±5 % whichever is greater
1.2.6 Respiration
NOTE:The respiration measurement is intended for patients over three years old
Measurement range 4 to 120 bpm Accuracy ±5 bpm or ±5 % Resolution 1 bpm Averaging time 30 s Update interval 10 s
RESPIRATION WAVEFORM
Sweep Speeds 6.25 mm/s and 0.625 mm/s
Document No. 8001008-6
5
Page 20
Datex-Ohmeda S/5 monitors
1.3 Technical specifications
1.3.1 NIBP
Deflation rate, PR dep. 5 to 13 mmHg/s Inflation time 20 to 185 mmHg, 1 to 5 s
Automatic software control, max. inflation pressure
Over pressure limit, stops measurement after 2 seconds
The safety valve limits the maximum cuff pressure to 320 mmHg in adult/child mode or 165 mmHg in infant mode. Independent timing circuit limits pressurizing (>15 mmHg) time to 2 minutes 10 seconds maximum in adult/child mode, and 1 minute 5 seconds in infant mode.
adult 280 ±10 mmHg child 200 ±10 mmHg infant 150 ±10 mmHg
adult 320 mmHg child 220 mmHg infant 165 mmHg
1.3.2 ECG
Zeroing to ambient pressure is done automatically.
Inflation pressure is adjusted according to the previous systolic pressure, typically 40 mmHg above. If the systolic pressure is not found, inflation pressure is increased typically 50 mmHg.
Max. measurement time adult 2 min
child 2 min infant 1 min
Pressure transducer accuracy is better than ±3 mmHg or ±2 % whichever is greater. Max. error ±4 mmHg.
Protection against electrical shock Type BF defibrillation proof
Defibrillation protection 5000 V, 360 J Recovery time 2 s Input impedance >2.5 M (10 Hz) CMRR >100 dB (ST) System noise <40 µV (p-p, RTI) Allowable offset ±300 mVDC Gain range 0.2 to 5.0 cm/mV Pacemaker pulse detection 2 to 500 mV, 0.5 to 2 ms pulses
6
Document No. 8001008-6
Protection against electrical shock Type CF defibrillator proof
Page 21
1.3.3 Pulse oximetry
Protection against electrical shock Type BF defibrillation proof
1.3.4 Temperature
Measurement accuracy ±0.1 °C (25.0 to 45.0 °C)
Protection against electrical shock Type CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on transducer/probe used. Please refer to the transducer/probe specification.
1.3.5 Invasive blood pressure
DIGITAL DISPLAY AVERAGING
Digital displays Art and P1 are averaged over 5 seconds and updated at 5 seconds intervals. All other pressures have respiration artifact rejection.
S/5 Hemodynamic modules
±0.2 °C (10.0 to 24.9 °C)
Accuracy ±5 % or ±2 mmHg, whichever is greater Transducer and input sensitivity
Input voltage 5VDC max current 20 mA Filter 0 to 4 - 22 Hz adjustable Zero set accuracy ±1 mmHg Calibration resolution ±1 mmHg Zero time less than 15 s Protection against electrical shock Type CF defibrillation proof
NOTE: The accuracy of the measurement may be different from the specified, depending on transducer/probe used. Please refer to the transducer/probe specification.
1.3.6 Respiration
Excitation frequency, 12-lead ECG 62.5 kHz
Excitation frequency, other modules 31.25 kHz
Breath detection automatic, range 0.3 to 6 manually adjustable minimum
Input dynamic range 0.2 to 6 Input impedance range 100 to 5000 Respiration Rate min. 4 bpm
Lead off detection >3 M
5 µV/V/mmHg
detection: 0.2, 0.4, 0.6, 0.8, 1.0
max. 120 bpm
Document No. 8001008-6
7
Page 22
Datex-Ohmeda S/5 monitors
2 FUNCTIONAL DESCRIPTION
2.1 Measurement principle
2.1.1 NIBP
NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure.
The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is inflated with a pressure slightly higher than the presumed systolic pressure, and deflated at a speed based on the patient’s pulse, collecting data from the oscillations caused by the pulsating artery. Based on these oscillations, values for systolic, mean, and diastolic pressures are calculated.
The following parts are necessary for the NIBP measurement:
M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR (or M-NIBP) module
twin hose (adult or infant model)
blood pressure cuffs (various sizes)
2.1.2 ECG
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical potential produced with electrodes placed on the surface of the body.
ECG reflects:
electrical activity of the heart
normal/abnormal function of the heart
effects of anesthesia on heart function
effects of surgery on heart function
See the User's Guide or theUser's Reference Manual for electrodes positions and other information.
2.1.3 Pulse oximetry
A pulse oximeter measures the light absorption of blood at two wavelengths, one in the near infrared (about 900 nm) and the other in the red region (about 660 nm) of the light spectrum. These wavelengths are emitted by LEDs in the SpO peripheral tissue and is finally detected by a PIN-diode opposite the LEDs in the probe. The pulse oximeter derives the oxygen saturation (SpO between the relative absorption at the two wavelengths and the arterial oxygen saturation SaO
probe, the light is transmitted through
2
) using an empirically determined relationship
2
.
2
8
Document No. 8001008-6
In order to measure the arterial saturation accurately, pulse oximeters use the component of light absorption giving variations synchronous with heart beat as primary information on the arterial saturation.
Page 23
S/5 Hemodynamic modules
A general limitation of pulse oximetry is that due to the use of only two wavelengths only two hemoglobin species can be discriminated by the measurement.
The modern pulse oximeters are empirically calibrated either against fractional saturation
frac;
SaO
2
2
=
2
fracSaO
HbO
2
++
binDyshemogloHbHbO
Formula 1
or against functional saturation SaO
HbO
=
2
funcSaO
Functional saturation is more insensitive to changes of carboxyhemoglobin and methemoglobin concentrations in blood.
The oxygen saturation percentage SpO against functional saturation SaO measurement relative to SaO2func can be maintained even at rather high concentrations of carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeters are not able to correctly measure oxygen content of the arterial blood at elevated carboxyhemoglobin or methemoglobin levels.
Plethysmographic pulse wave
The plethysmographic waveform is derived from the IR signal and reflects the blood pulsation at the measuring site. Thus the amplitude of the waveform represents the perfusion.
Pulse rate
The pulse rate calculation is done by peak detection of the plethysmographic pulse wave. The signals are filtered to reduce noise and checked to separate artifacts.
func;
2
2
HbHbO
2
+
measured by the Datex-Ohmeda module is calibrated
2
func. The advantage of this method is that the accuracy of SpO
2
Formula 2
2
Probe
The standard probe is a finger clamp probe which contains the light source LEDs in one half and the photodiode detector in the other half. Different kinds of probes are available from Datex­Ohmeda.
9
Document No. 8001008-6
Page 24
Datex-Ohmeda S/5 monitors
Intensity of transmitted light
I
max (DC-component)
I
I
min
max
AC-component
Transmitted light
Incident light
Emitter
RED
Detector
No pulsation
IRED
Pulsatile blood
SpO sensor cable
2
Variable absorption due to pulse added volume of arterial blood
Arterial blood
Venous blood
Tissue
Time
SpO sensor connector
2
6
GND
7
I
LED
4
5
GND
8
V
B
R
C
1
I
S
9
GND
Figure 2 Absorption of infrared light in the finger probe parts layout and schematic
2.1.4 Temperature
The temperature is measured by a probe whose resistance varies when the temperature changes, called NTC (Negative Temperature Coefficient) resistor.
The resistance can be measured by two complementary methods:
10
Document No. 8001008-6
diagram
Applying a constant voltage across the resistor and measuring the current that flows
through it
Applying a constant current through the resistor and measuring the voltage that is
Page 25
generated across it.
In Datex-Ohmeda modules the two methods are combined in the form of a voltage divider. The NTC-resistor is connected in series with a normal resistor and a constant voltage is applied across them. The temperature dependent voltage can be detected at the junction of the resistors, thus producing the temperature signal from the patient. The signal is amplified by analog amplifiers and further processed by digital electronics.
2.1.5 Invasive blood pressure
To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive pressure setup, consisting of connecting tubing, pressure transducer, an intravenous bag of normal saline all connected together by stopcocks, is attached to the catheter. The transducer is placed at the same level with the heart, and is electrically zeroed.
The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor interprets the voltage signal so that pressure data and pressure waveforms can be displayed.
2.1.6 Respiration
Impedance respiration is measured across the thorax between ECG electrodes. The respiration signal is made by supplying current between the electrodes and by measuring the differential current from the electrodes. The signal measured is the impedance change caused by breathing. From these impedance changes, respiration rate is calculated, and the respiration waveform is displayed on the screen.
S/5 Hemodynamic modules
2.2 Main components
2.2.1 M-ESTPR/-ETPR/-ESTR modules
T1
T2
ECG+ Resp
Figure 3 Front panel of M-ESTPR
The M-ESTPR, M-ETPR, and M-ESTR modules contain two main PC boards, the STP board and the ECG board. They work independently. Both of them have their own processor and software EPROM. Some components on the boards are not used in ETPR and ESTR modules.
SpO
P1
P2
2
ECG
Lead
Start
Wedge
Zero
P1
Zero
P2
In the M-ESTPR module, additionally, there are two small boards, the SP input and the ECG input boards, attached to the front panel of the module. The front panel has six connectors and four keys.
11
Document No. 8001008-6
Page 26
Datex-Ohmeda S/5 monitors
The connectors are two for temperature measurement, two for invasive blood pressure measurement, one for ECG, and one for SpO
measurement. The keys are for ECG lead, Start
2
Wedge, P1 zero, and P2 zero.
In the M-ETPR module, there are two small boards, the ECG input board and the 2P input board attached to the front panel of the module. The front panel has five connectors and four keys. The connectors are two for temperature measurement, two for invasive blood pressure measurement, and one for ECG
measurement. The keys are for ECG lead, Start Wedge, P1 zero, and P2 zero.
In the M-ESTR module, there are two small boards: the S input board and the ECG input board, attached to the front panel of the module. The front panel has four connectors and one key. The connectors are two for temperature measurement, one for ECG, and one for SpO The key is for ECG lead select.
NOTE: M-ESTP rev. 03, M-ETP rev. 02 and M-EST rev. 02 and all earlier revisions have separate T and SP input boards.
2.2.2 M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules
measurement.
2
Auto
On/Off
NIBP
T1
ECG+ Resp
P1
Start
Cancel
P2
T2
Zero
SpO
2
P1
Zero
P2
Figure 4 Front panel of M-NESTPR
The M-NESTPR, M-NETPR, and M-NESTR modules contain three main PC boards, the STP board, the ECG board, and the NIBP board. They work independently. Each of these has their own processor and software EPROM.
The M-NE12STPR, M-NE12TPR, and M-NE12STR contain three main PC boards, The STP board, the ECG board and the NIBP board. They work independently. Each of them has their own processor. The STP board and NIBP board have software EPROM. In the ECG board the software is in flash memory. The STP and NIBP boards are the same as in M-NESTPR module but the ECG board and ECG input board are different.
12
Document No. 8001008-6
In the M-NESTPR module, there are two small boards, the SP input and the ECG input board attached to the front panel of the module. The front panel has seven connectors and four keys. The connectors are two for temperature measurement, two for invasive blood pressure measurement, one for ECG, one for NIBP, and one for SpO
measurement. The keys are for NIBP Auto On/Off,
2
NIBP Start/Cancel, P1 zero, and P2 zero. The structure of M-NE12STPR is similar except the ECG board and ECG input board are different.
In the M-NETPR module, there are two small boards, the 2P input board and the ECG input board, attached to the front panel of the module. The front panel has six connectors and four keys. The
Page 27
connectors are two for temperature measurement, two for invasive blood pressure measurement, one for ECG, and one for NIBP. The keys are for Auto On/Off, Start/Cancel, P1 zero, and P2 zero. The structure of M-NE12TPR is similar except the ECG board and ECG input board are different.
In the M-NESTR module, there are two small boards, the ECG input board and the S input board, attached to the front panel of the module. The front panel has five connectors and two keys. The connectors are two for temperature measurement, and one for SpO and one for NIBP. The keys are for Auto On/Off, Start/Cancel. The structure of M-NE12STR is similar except the ECG board and ECG input board are different.
2.2.3 NIBP board
Cuff
Zero
valve
Exhaust
valve 2
Joining chamber
Exhaust valve 1
Check valve
Bleed valve
Pump
S/5 Hemodynamic modules
measurement, one for ECG,
2
Pump
and
valve
M
driver
B1 B2
to/from module bus
X
Pressure transducers
converter
Power-up
reset
RS485
1
EEPROM
(calibration
Figure 5 NIBP board functional block diagram
Pressure transducers
The NIBP board contains two pressure transducers. They are of piezoresistive type. One is used for measuring the pressure of the blood pressure cuff and the pressure fluctuations caused by arterial wall movement (B1). The other is used for detection of cuff hose type, cuff loose and cuff occlusion situations etc. (B2). The transducers are internally temperature compensated. They are supplied by a constant voltage and their output voltage changes up to 40 mV max. (50 kPa, 375 mmHg).
AD-
interf.
data)
Write protection switch
Overpressure control
CPU
80C51FA
Internal
watchdog
Address decoder
Address
latch
Address bus
Watchdog
timer
RAM EPROM
Databus
+15 VD
Front panel
keys
Software
control
NESTPR_NIBP_board_blck_dgrm .vsd
13
Document No. 8001008-6
Page 28
Datex-Ohmeda S/5 monitors
Signal processing
Two signals from the pressure transducers are amplified and sent to the A/D converter. After the converter, digitized signals are sent to the microprocessor for data processing. Before the converter, one of the signals is used to adjust the offset to the pressure safety level.
The NIBP board is controlled with a 80C51FA microprocessor at 16 MHz oscillator frequency.
Memory
NIBP program memory (EPROM) size is 128k × 8. RAM size is 32k × 8 bit and it stores variable values of the NIBP measurement. EEPROM is size 64 × 16 bit and is used to store the calibration values for the pressure transducers, the pulse valve constants gained during measurements, the PC board identification, and module serial number.
Software control
Software controls valves and pump. In addition to the individual on/off signals for each component there is a common power switch for the valves and the pump that can be used at pump/valve failures.
In addition to external RS485 reset line the microprocessor system is equipped with its own power­up reset. See the section in the ECG board’s description: “RS485 communication”
Watchdog timer
The NIBP board is equipped with a software independent safety circuit to disconnect supply voltages from the pump and the valves if the cuff has been pressurized longer than the preset time. As soon as the cuff pressure rises over a specified pressure limit, timer starts counting. The timer is adjusted to stop the pump and open the valves after 2 minutes 10 seconds in adult/child mode and after 1 minute 5 seconds in infant mode.
Valves
Exhaust valves are used for emptying the cuff and the joining chamber after the measurement. Exhaust valve 1 is also used as safety valve in infant mode. The valve opens at 165 mmHg. Exhaust valve 2 is also used as safety valve in adult mode and opens at 320 mmHg. The bleed valve is used for emptying the cuff during measurement. The zero valve is used for connecting the pressure transducer B1 to open air.
Power supply section
All connections are established via 25-pin connector (D-type, female). The module needs +5 V, ±15 V, and +15 VD (dirty) power supply to operate. The pump and the valves use separate +15 VD power line. The supply voltages are generated in the power supply section of the S/5 monitor. The reference voltages ±5 V
and +10 V
ref
are generated on the NIBP board.
ref
14
Document No. 8001008-6
Page 29
2.2.4 ECG board in 3-and 5-lead measurement
Patient signals are connected to overload protection circuits (resistors and gas-filled surge arresters) and analog switches to instrumentation amplifiers. Then the signals are amplified by 480 and limited by slew rate. Then they are A/D-converted, analyzed and transferred to module bus in digital form.
PATIENT
Overload protection
RL F CN
Defibrillation
detector
uP
uP
uP
Slew-rate
detector
HP
filter
Anti­alias
Analog switches
Slew-rate
detector
uP
HP
filter
uP
Anti­alias
uP
uP
Slew-rate
detector
HP
filter
Anti­alias
uP
uP
Slew-rate
detector
HP
filter
Anti­alias
S/5 Hemodynamic modules
Lead selection logic
controlled by ECG lead selection
signals from uP
Resp
amplifier
Sync. Rectifier
31 kHz
from uP
Figure 6 3- and 5- lead ECG board block diagram
Analog ECG section
The ECG cable is connected to connector pins E1 to E6 on the input board which contains an overload protection circuit. Leads are connected to amplifiers via analog switches. States of the switches depend on the cable type. Lead-off, noise and pacemaker are detected by a slew rate
ISOLATION
Supply
voltages
Power
source
Power
source
Opto-
coupler
Opto-
coupler
to STP Board
Amplifier
A/D
Micro-
processor
(uP)
Direct
ECG
Direct
ECG
NESTPR_ECG_brd_blck_dgrm.vsd
15
Document No. 8001008-6
Page 30
Datex-Ohmeda S/5 monitors
detector. Lower frequency is determined by high pass (HP) filter 0.5 Hz (monitor bandwidth) or
0.05 Hz (diagnostic or ST- bandwidth).
Respiration section
3-lead cable The analog switches control the current supply source of the impedance respiration measurement,
and the lead selection for the 3-lead cable can be seen from the following table:
Table 2 Lead selection and coding for the 3-lead cable
Selected lead Current source between Signal measured from
IR - L N
II R - N L
III L - N R
Position on
IEC standard coding AAMI standard coding
body surface
right arm R = red RA = white
left arm L = yellow LA = black
left leg F = green LL = red
5-lead cable When the 5-lead cable is used, the current source is between L-F and the signal is measured from
the N, independently on the lead selection.
The respiration amplifier consist of the operational amplifiers, and the components around them. There is an analog switch for controlling the gain of the first stage of the preamplifier. Synchronous rectifier consists of the analog switches, which are used for detecting the respiration signal from 31 kHz amplitude modulated raw signal. The amplifier stage consists of the differential amplifier and the last amplifier. The differential amplifier consists of the operational amplifiers and the components around them. This stage is AC-coupled on both sides for minimising the offset voltages. The last amplifier is used for amplifying the signal derived from differential amplifier stage. The respiration signal is zeroed at the beginning of the measurement. Zeroing is also used for fast recovering the measurement after the motion artefact. This is done in amplifier section.
NOTE: The respiration measurement is switched OFF for 20 seconds when defibrillation is detected at the defibrillation detector.
Microprocessor section
Microprocessor contains RAM and EPROM memories. The processor uses external EEPROM memory. The microprocessor’s internal 8-channel A/D-converter converts the ECG-signals to digital form. See the section in ECG board’s description: “RS485 communication
Serial communication
Communication with the module bus is made through RXD and TXD pins. See the section in STP board’s description: “Serial communication”.
16
Document No. 8001008-6
Page 31
S/5 Hemodynamic modules
Isolated section
The patient isolation of ECG is 5 kV.
NOTE: The isolation has been changed from the earlier revisions.
WARNING Do not touch battery-operated monitor during defibrillation procedure.
See the “Isolated section” in STP board description.
Power supply section
See the “Power supply section” in STP board description. There is a test connector (X20) on the board for voltages +5 VREF, +5 V, +12 V, GND and -12 V.
2.2.5 ECG board in 12-lead measurement
The 12-lead ECG measurement consists of the functions, which are shown in the figure 7. All functions are located in the ECG board except the front panel connector and the ECG input board.
Front panel connector and ECG input board
Input protection and filtering
Respiration impedance amplifiers
RS 485 communication
Respiration impedance supply
ECG CPU
Power supply
ECG preamplifiers
ECG amplifiers
and
Baseline
restoration
NV memory
Pacer
detection
Isolation
to STP board
Isolation
Figure 7 12-lead ECG measurement block diagram
17
Document No. 8001008-6
Page 32
Datex-Ohmeda S/5 monitors
Front panel connector and ECG input board
The connector for the 12-lead ECG cable is a green 12 pin Nicolay type connector. 3- or 5-lead cables with blue connectors cannot be connected to this connector. The ECG input board contains high voltage resistors and a connector for ECG board.
Input protection and filtering
The input protection is implemented with protection diodes, which are connected to analog power supply voltage and ground. The input filtering for ECG measurement is done with discrete components. The measured signal is AC-coupled for respiration measurement. The signal from the respiration supply is AC coupled. There are also the overload protection diodes for respiration measurement supply.
ECG preamplifiers
The buffer amplifiers are used for each lead except N/RL. The leads off detection is implemented by measuring the level of the input buffer amplifiers with A/D converter of CPU. The ECG signals are measured using differential amplifiers.
ECG amplifiers and baseline restoration
The function of the ECG amplifiers and baseline restoration is to amplify the signal and to restore the baseline of the signal in the middle of the display after the change of the signal level e.g. after the change of the DC offset voltage.
Pacer detection
Pacer detection has been made by using two slew rate detector circuits. The pacer detection amplifiers have been realized at the front of the slew rate detectors independently from the ECG measuring channels.
Respiration impedance supply
The 62.5 kHz sine wave generator is used as the respiration measurement signal supply. Analog switches are used for connecting the sine wave to the ECG leads to be measured.
Respiration impedance amplifiers
Buffer amplifiers are used in respiration measurement. Analog switches are used for selecting the measurement leads. There are also additional amplifiers for increasing the respiration signal gain. Respiration is always measured between R and F, independently on the ECG lead selection.
ECG CPU
18
Document No. 8001008-6
The CPU is a 16 bit H8/3048 single-chip microcomputer. It contains 128 kbytes of flash memory and 4 kbytes of RAM. The clock frequency is 16 MHz.
Page 33
RS485 communication
The communication to the CPU board of the monitor uses RS485 protocol. The RS485 driver circuits are optically isolated from the processor of the module. PWM signal is used for direct ECG signal. Direct ECG signal is available from the X2 connector of the UPI board or from the PT module.
Power supply
The ECG board has a driver controlled half bridge switching power supply with 5 kV isolation. The supply voltages have been regulated with linear regulators.
2.2.6 ECG filtering
The S/5 monitors have three ECG filtering modes:
MONITORING 0.5 to 30 Hz (with 50 Hz reject filter)
DIAGNOSTIC 12-lead ECG 0.05 to 150 Hz DIAGNOSTIC other modules 0.05 to 100 Hz
S/5 Hemodynamic modules
0.5 to 40 Hz (with 60 Hz reject filter)
ST FILTER 0.05 to 30 Hz (with 50 Hz reject filter)
0.05 to 40 Hz (with 60 Hz reject filter)
The purpose of filtering is to reduce high frequency noise and low frequency (e.g. respiratory) movement artifacts.
Monitor filter is used in normal monitoring. Diagnostic filter is used if more accurate diagnostic information is needed. ST filter gives more accurate information of ST segment, but reduces high frequency noise.
The high-pass filters 0.5 Hz and 0.05 Hz are done with hardware. The monitor sends a command to the hemodynamic module determining which of the corner frequencies 0.5 Hz or 0.05 Hz is to be used.
The 50 Hz and 60 Hz reject filters are both low-pass filters with zero at 50 Hz or 60 Hz correspondingly and they are done with software. They are for the mains supply filtering. When these filters are used, 3 dB value for low-pass filter is 30 Hz or 40 Hz.
In diagnostic mode the upper frequency is limited by hardware and the -3 dB frequency is 100 Hz for 3 or 5 lead ECG measurement. For 12 lead ECG the upper frequency is 150 Hz and it is limited by software.
19
Document No. 8001008-6
Page 34
Datex-Ohmeda S/5 monitors
2.2.7 STP board
Patient connectors
Front panel
keys
Power
isolation
section
Isolation
transformer
measuring
Temp AD
TEMP
unit
Press AD
AD-converter
- 8 chn
- 12 bit
µprocessor unit
RAM internal 2K
external 16K
EPROM 48K
RS communication
Patient isolation
INV PRESS measuring
unit
Pox AD
Serial device communication
POX pre­amplifier
POX gain
control
Red
measuring
Module bus data
Reset
ed IR
Opto
isolation
g
rin
su
ea m
Intensities
Non volatile
memory
ver
i
Ds dr
ox LE P
Figure 8 STP board block diagram
Microprocessor unit
An Intel 80C196KC-16 processor is used. There are external memories, an 8-bit data bus, a 16 MHz oscillator, an open collector reset, and a watchdog timer. Three A/D-converters within the processor are used. The processor’s internal UART communicates with the CPU board.
High speed I/O is used to obtain pulse control sequence necessary for pulse oximetry measurement. It gets its timing clock from the oscillator.
20
Document No. 8001008-6
Power
non-isolation
section
Power for
module
Power
reset
Power for
Communication
Module bus connector or
connector to NIBP Board (NESTPR)
Module reset Module data
RS485 Driver
for module
reset
RS485 Driver
for data
NESTPR_STP_brd_blck_dgrm.vsd
Page 35
Temperature measurement unit
The value of the NTC-resistor in the probe depends on the patient’s temperature. It is measured with the following principle described below.
The temperature signal(s) is produced by voltage dividers, part of which is the patient probe (YSI 400-series thermistor). The output is amplified by the calibrated amplifier(s) whose offset voltage makes its output spread on both sides of zero. A wider output range (measurement range) means better resolution.
0°C
=> 7K357 15 °C => 3K541 25 °C => 2K253 38 °C => 1K301 45 °C => 984R1
NTC
+5V
S/5 Hemodynamic modules
reference
A1
to AD converter
Figure 9 Temperature measurement principle
Invasive blood pressure measurement unit
An isolated +5 V voltage is supplied to the pressure transducer. From the bridge connection a differential voltage, which depends on pressure and supplied voltage, is calculated (see the formula below).
= Uin × pressure × 5 V, where Uin is 5 V
U
out
= 25 V × pressure [mmHg]
Þ U
out
Pressure amplification is realized in the instrumentation amplifier. The gain of the amplifier is set so that the level of the signal transferred to A/D converter stays within the measurement range even when there are circumstantial offsets or offsets caused by the transducer. There is a filter before the amplifier to attenuate high frequency disturbances.
Vin
Pressure transducer
offset
NESTPR_temp_meas_princ.vsd
Instrumentation amplifier
Vout
Input Filter
Figure 10 Pressure measurement principle
G
to AD converter
pressure_meas_principle.vsd
21
Document No. 8001008-6
Page 36
Datex-Ohmeda S/5 monitors
Pulse oximetry measurement section
I=5-350mA
Probe
Preamplifier Current - to - Voltage converter
LED driving circuit
Level of LED current measurement and feedback circuit
G =1/4096-1
Digitally controlled attenuator
Amplifier
IRed LED intensity adjustment
Red LED intensity adjustment
Level of LED current indication (to CPU)
IR DC level
G = 275
Two step AC amplifier
G =16 or 63
G =16 or 63
IRed AC signal
for NESTPR G =18 or 125
Red AC signal
Red DC level
Figure 11 Pulse oximetry measurement block diagram
LED control signals
The processor sends pulse width modulated signals, IRED intensity and RED intensity, which are converted to DC voltages and filtered. Using switches either RED or IRED intensity is sent to the amplifier in LED driving circuit.
LED driving circuit
The voltage difference which corresponds to LED current, is measured by the differential amplifier circuit and its output is sent back to the processor in 0 to 5 V level. There are feedback circuits from LED current measurement and LED intensity control.
Background light is measured by picking up a sample from the signal. The sample is modified to 0 to 5 V level and sent to the processor.
Measured signal preamplification
The preamplifier is current-to-voltage converter with gain selection. A higher gain is used for measuring thin tissue.
Digitally controlled amplifier
22
Document No. 8001008-6
The D/A converter is a digitally controlled amplifier after which there is another constant amplifier.
Page 37
Red and infrared channel separation
Red and infrared channels are separated from each other by switches. An operational amplifier functions as a buffer and after this the infrared DC signal is sent to the processor. A capacitor separates theC signal from the DC and the AC signal is sent to the processor after amplification. There is a switch to choose the amplification constant.
Serial communication
Serial communication between the module and the frame is done by an RS485 type bus whose buffers get their supply voltage (+5 VDC) from the frame and in the isolation section get the supply voltage (+5 V) from the isolated power supply.
The serial communication buffers are controlled by a Reset signal so that when the Reset is active, the buffer does not transfer data.
Reset is also RS485 type and additionally, there is an auxiliary logic power reset, which keeps the reset active for about 500 ms despite the state of reset in the module bus. Time constant determines the power-up reset time. There are components to prevent the module from sending data during reset. Data transmission rate is 500 kbps.
S/5 Hemodynamic modules
NIBP Board
or s
send/receive
dule proses o
m
STP Board
r
osesso r
send/receive
le p
du o
m
ECG Board
ssor
se
send/ recei ve
Receive data
Send d ata
Res et
Rece ive dat a Send data
Res et
Rec ei ve da ta
Send data
Res et
Opto isolation
PatientIso
Opto isolation
n tio
la
Receive data
Send data
Rese t
Recei ve d ata
Send data
connector for ECG board
Rec ei ve d at a
Send dat a
send/recei ve
Res e t
Res et
RS485 Dri ver
send/receive
RS485 Dri ver
RS4 85 Dri v er
send/r eceiv e
RS4 85 Dri v er
Dat a
NDat a
Rese t in
NRes et in
Dat a
NDat a
Rese t i n
NRes et i n
connector for STP boardconnector for NIBP board
module bus connector
module pro
t n e ti
Isolation
Pa
connector for STP board
Figure 12 Serial communication and opto isolation of M-NESTPR/-NE12STPR
Document No. 8001008-6
23
Page 38
Datex-Ohmeda S/5 monitors
STP Board
send/receive
module processor
ECG Board
send/receive
module processor
Receive data Send data
Res et
Receive data
Send data
Res et
Opto isolation
Patient
Opto isolation
Patient
isolation
isolation
Receive data
Send data
connector for ECG board
Recei ve dat
Send data
send/receive
Reset
Rese t
RS485 Driver
send/receive
RS485 Driver
Connector for STP board
Data
NDat a
Reset in
NRese t in
Connector for ECG board
module bus connector
Figure 13 Serial communication and opto isolation of M-ESTPR
Isolated section
There are two opto isolators. The signal is processed on a logical high-low level even though the outputs of the opto isolators are analog signals in the isolated section.
The reset line is an open collector type with a pull-up resistor. Thus the processor is able to use its internal watch-dog function.
Power supply section
The isolated supply voltage of the module is developed from +15 Vdirty voltage from the Central Unit. The power supply is a switched-mode circuit, where the FET transistor switch is controlled by an oscillator using bipolar timer. The frequency of the oscillator is about 30 kHz and pulse ratio 50 %. Controlling the FET switch is slowed to suppress spurious interference.
A special pulse transformer is used in the circuit. In the secondary circuit normal linear regulators are used except for +5 V (low drop type linear regulator).
24
Document No. 8001008-6
Page 39
2.3 Connectors and signals
2.3.1 Module bus connector
S/5 Hemodynamic modules
13
25
1
14
Figure 14 Module bus connector (X1) pin layout
Table 3 Module bus connector description
Pin No I/O Signal Note
1 I RESET_RS485
2 I -15 VDC ∗∗
3 I +15 VDIRTY
4 I +15 VDC ∗∗
5 I/O -DATA_RS485
6 I/O DATA_RS485
7 Ground & Shield
8 I -RESET_RS485
9 I CTSB
10 O RTSB
11 I RXDB
12 O TXDB
13 Ground & Shield
14 I +32 VDIRTY
15 I GroundDIRTY
16 I CTSC
17 O RTSC
18 I RXDC
19 O TXDC
20 ON/STANDBY
21 O PWM_ECG
22 RXDD_RS232
23 TXDD_RS232
24 I +5 VDC
25 I +5 VDC
= Not used ∗∗ = Used only by M-ESTPR, M-ETPR, M-ESTR and M-NIBP modules
25
Document No. 8001008-6
Page 40
Datex-Ohmeda S/5 monitors
2.3.2 Front panel connectors
Table 4 Front panel connectors
12-lead ECG connector Temp connector (T1, T2)
8
1
2
3
Pin No Signal Pin No Signal
1
Right arm electrode (R)
2
Left arm electrode (L)
3
N
4
Left leg electrode (F)
5
Chest electrode (C1)
6
Chest electrode (C2)
7
Chest electrode (C3)
8
Chest electrode (C4)
9
Chest electrode (C5)
10
11
12
Chest electrode (C6)
Cable type
Ground
9
101112
45
7
6
1
Temperature probe
2
Temperature probe
ECG connector (ECG) SpO2 connector (SpO2) Invasive blood pressure
connectors (P1, P2)
6
3
1
2
9
4
7
0
5
8
6
3
1
2
9
4
7
0
5
8
3
1
4
2
5
Pin No Signal Pin No Signal Pin No Signal
10
1
Right arm electrode (R)
2
Left arm electrode (L)
3
Right leg electrode (RL)
4
Left leg electrode (F)
5
Chest electrode (C)
6
Cable shield
7
Not connected
8
3/5 lead identification
9
Lead connection check
Ground
10
1
Feedback resistor
2
Ground
3
Not Connected
4
Cable shield + probe identification ground
5
Probe identification
6
LED drive ground
7
LED drive current
8
Input signal current
9
Ground
Ground
10
1
Pressure +
2
Pressure -
3
Polarisation - (ground)
4
Polarisation +
5
Not connected
6
Not connected
7
Not connected
8
Not connected
9
Ground
Cable detection
6
9
7
0
8
26
Document No. 8001008-6
Page 41
2.3.3 Test points on boards
12-lead ECG board
X
2
0
1
n
o
n
f
l
o
d
n
g
1
5
+
1
6 8
5
+
1
v
S/5 Hemodynamic modules
d
a
o
r
b
G
C
d
E
l
a
2
e
1
­t
s
i
d
c
o
m
p
g
a
t
i
n
Y
T
v
I
D
R
o
e
n
e
n
NIBP board
2
X
3
1
f
l
o
a
t
i
n
g
1
G
N
D
l
o
6
5
+
G
0
1
g
a
n
V
a
D
N
NESTPR_12-lead_ECG_board.vsd
X
0
1
X
f 1 2
1
9
6
l
o
a
t
i
n
g
-
5
V
2
.
+
5
V
1
4
f
l
o
a
t
i
n
g 1 2
.
5
+
V
d
i
g
There are test pad blocks on solder side. X8 and X6 pads and voltages are:
X8
10 o o o o o 2
9 o o o o o 1
X6
X5
X7
X8
X6
1 2 o o o o o o o o 8 7
nibp_text_conn_in_nestpr.vsd
X8 X6
Pin No Signal Pin No Signal
1 GND 1 GND
2 WD out 2 A1 output
3 reset 3 - 5 V
4 +5 V 4 +5 V ref
5 +15 V dirty 5 B1 out - (A1 input)
6 +15 V 6 B1 out +
7 -15 V 7 B2 out +
8- 8B2 out -
9-
10 GND
27
Document No. 8001008-6
Page 42
Datex-Ohmeda S/5 monitors
ECG and STP board
There are identical test pin blocks both on STP and ECG boards. Pins and voltages are as follows:
ESTPR
NESTPR
X11 pin 1 +5 Vref
pin 2 +5 V
pin 3 +12 V
pin 4 Gnd
pin 5 -12 V
X12 pin 1 -5 V (STP board only)
X11 pin 1 +5 Vref
pin 2 +5 V
pin 3 +7 V
pin 4 Gnd
pin 5 -7 V
X12 pin 1 -5 V (STP board only)
X11
X11
X12
X12
28
Document No. 8001008-6
Page 43
S/5 Hemodynamic modules
3 SERVICE PROCEDURES
3.1 General service information
Field service of the hemodynamic modules is limited to replacing faulty printed circuit boards or mechanical parts. Faulty printed circuit boards should be returned to Datex-Ohmeda for repair.
Datex-Ohmeda is always available for service advice. Please provide the unit serial number, full type designation, and a detailed description of the fault.
CAUTION Only trained personnel with appropriate equipment should perform the tests and repairs outlined in
this section. Unauthorized service may void warranty of the unit.
3.2 Service check
These instructions include complete procedures for a service check. The service check should be performed after any service repair, however, the service check procedures can also be used for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (Appendix A) which should be filled in when performing the procedures.
The mark
check form.
? in the instructions means that the performed procedure should be signed in the
3.2.1 Recommended tools
Tool Order No. Notes
Patient simulator -
Pressure manometer -
Temperature test set 884515
3-lead ECG trunk cable
5-lead ECG cable
10-leadwire ECG cable
SpO2 finger probe OXY-F4-N or SAS-F4
InvBP transducer
Adult NIBP cuff & hose
Infant NIBP cuff & hose
Screwdriver
3.2.2 Recommended parts
Part Order No. Notes
NIBP pump filter 57142
29
Document No. 8001008-6
Page 44
Datex-Ohmeda S/5 monitors
All modules
Detach the module box by removing the two screws from the back of the module. Be careful with loose latch and spring pin for locking.
1. Check internal parts:
screws are tightened properly
cables are connected properly
all IC’s that are on sockets are attached properly
EMC covers are attached properly
there are no loose objects inside the module
?
2. Check external parts:
the front cover and the front panel sticker are intact
all connectors are intact and are attached properly
the module box, the latch and the spring pin are intact
?
3. Replace the NIBP pump filter in NE12STPR/NE12TPR/NE12STR/NESTPR/NETPR/NESTR
modules, if necessary.
?
Reattach the module box and check that the latch is moving properly.
Switch the monitor on and wait until the monitoring screen appears. Configure the monitor
screen so that all the needed parameters are shown, for example as follows:
Monitor Setup - Waveform Fields - Field 1 - ECG1
Field 2 - ECG2 Field 3 - P1 Field 4 - P2 Field 5 - Pleth Field 6 - Resp
Digit Fields - Lower Field 2 - NIBP
Lower Field 3 - T1+T2
30
Document No. 8001008-6
Page 45
S/5 Hemodynamic modules
4. Plug in the module. Check that it goes in smoothly and locks up properly
?
5. Check that the module is recognized, i.e. all needed parameter information, except invasive
blood pressure, starts to show on the screen.
?
Preset ECG, Respiration, InvBP and SpO2 measurement settings:
ECG - ECG Setup - Hr Source - Auto
Pacemaker - Show
Others - Resp Setup - Size - 1.0
Resp Rate Source - Auto Measurement - On Detection Limit - Auto
Invasive Pressures - P1 ‘Art’ Setup - Label - Art
P2 ‘Cvp’ Setup - Label - Cvp
ECG measurement
6. Enter the service menu:
7. Enter the ESTP : ECG service menu:
8. Check the front panel membrane key ECG Lead (not available in NE12STPR/NESTPR type
Pulse Oximetry - Pleth Scale - Auto
Monitor Setup - Install/Service (password 16-4-34) -
Service (password 26-23-8) - Parameters
Take down the information regarding module software by selecting Scroll Vers and turning the ComWheel.
?
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing faster than by 5 per second. Check also that the ECG/RESP board memories have passed the internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ state all OK.
?
modules). Press the key at least for two seconds. Check that the selected ECG lead is changing on the screen and the state for ‘Button’ in the service menu.
?
31
Document No. 8001008-6
Page 46
Datex-Ohmeda S/5 monitors
9. Check that the power frequency value has been set according to the current mains power
frequency. Change the setting by selecting Power Freq, if necessary.
?
10. @ M-ESTPR, M-ETPR , M-ESTR, M-NESTPR, M-NETPR and M-NESTR modules: connect a 5-
lead ECG cable to the module. Check that the ‘Cable type’ shows 5 lead. If it shows 3 lead, make sure the used 5-lead ECG cable contains the necessary wiring for cable recognition (pins 0, 8 and 9 connected together).
@ M-NE12STPR, M-NE12TPR and M-NE12STR modules: connect a 10-leadwire ECG cable to the module. Connect limb lead electrodes and one electrode from the chest lead set to the same potential. Check that the ‘Cable type’ shows 10 lead.
?
11. Connect a 3-lead ECG trunk cable without a lead set to the module. Check that the
message "Leads off" is displayed on the screen.
?
12. Check that all the electrodes show OFF in the service menu and the message ‘Leads Off’ is
shown on the screen.
Connect all the leads together, for example to a suitable screwdriver. Check that all the electrodes show ON and the message ‘Asystole’ appears.
Disconnect one of the leads and check that the corresponding electrode in the service menu shows OFF within 10 seconds from the disconnection, then reconnect the lead. Check the rest of the leads using the same method.
NOTE: When the ground lead (black) is disconnected all the electrodes should show OFF.
NOTE: The ‘Asystole’ and ‘Different leads off’ messages are shown using certain priority, so even one of the leads is disconnected, the lead related ‘Leads off’ message may not appear onto the screen.
NOTE: When RA, LA or LL electrode is disconnected, all six V electrodes show OFF.
NOTE: With NESTPR/ESTPR type modules and 5 lead cable the state of V2, V3, V4, V5 and V6 electrodes follow the state of the V electrode.
?
32
Document No. 8001008-6
13. Connect the leads to a patient simulator.
The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
ECG - BASE - BPM - 160 PACE - WAVE - NSR
Check that normal ECG waveform is shown, the HR -value is 160 (±5) and the ‘Pacer count’
-value is not increasing in the service menu. Check the lead selections by pressing
Page 47
ECG Lead key on the module (not available in NE12STPR/NESTPR type modules).
the
ECG - PACE - WAVE - ASNC
Check that pacemaker spikes are shown on the ECG waveform, the ‘HR’ -value changes to 75 (±5) and the ‘Pacer count’ -value is increasing according to shown pacemaker spikes.
Set the pacemaker option off:
ECG - PACE - WAVE - NSR
?
Respiration measurement
13. Check that the ‘Resp Available’ and ‘RESP Measurement’ show both ON in the ESTP: ECG
service menu.
?
14. Check the respiration measurement with a patient simulator.
S/5 Hemodynamic modules
The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
BASELINE IMPEDANCE -switch - 500 LEAD SELECT-switch - II/RL-LL
RESP - WAVE - NORM
RATE - 20 OHMS - 1.0 RATIO - 1/1 APNEA - OFF SHIFT - OFF
Check that the RESP waveform is shown and the ‘RR’ -value is 20 (±5). Change the position of the BASELINE IMPEDANCE -switch and check that appropriate RESP waveform and ‘RR’ ­value are shown again within 30 seconds.
RESP - APNEA - 32 S
Check that the monitor gives the APNEA -alarm.
NOTE: Make sure that only the ECG leads are connected to the simulator during the apnea ­test. If other cables are connected at the same time, the respiration signal from the simulator may be disturbed, and therefore, the APNEA -alarm may not be activated.
NOTE: When you have ECG service menu open, spikes will appear on the respiration waveform. These spikes represent the threshold level for detecting inspiration and expiration.
?
Document No. 8001008-6
33
Page 48
Datex-Ohmeda S/5 monitors
Temperature measurement
15. Enter the ESTP : STP service menu:
Parameters - ESTP : STP
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing faster than by 5 per second. Check also that the STP board memories have passed the internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ show all OK.
?
16. Check that the ‘Cable’ and ‘Probe’ show OFF for both channels, T1 and T2, when no probes
are connected.
Connect a temperature test plug into the connector T1. Check that the ‘Cable’ and ‘Probe’ for T1 show ON and the corresponding temperature value appears onto the monitor screen. Perform the same check also for the channel T2.
?
17. Check the temperature calibrations using temperature test plugs.
If the deviation on a temperature reading on the screen is more than 0.1 °C, calibrate the temperature channels according to the instructions in the chapter 3.4.3 Temperature
calibration.
?
18. Activate the temperature test by selecting Temp Test from the menu and pressing the
ComWheel twice. When the message ‘Performing temp test’ disappears from the digit field, check that no error messages appear and ‘Temp error’ shows OFF for both channels in the service menu.
?
19. Check that the module configuration has been set correctly. The configuration in use is
shown beside the text ‘Configuration’ in the service menu and it can be either STP, ST or TP. Change the configuration in the Calibrations menu, if necessary.
?
Invasive blood pressure measurement
20. Check the front panel membrane keys that are related to the InvBP or temperature
measurement. Press each of the keys at least for one second. Check that the pressed key is identified, i.e. one of the texts for ‘Buttons’ changes from OFF to ON in the service menu.
34
Document No. 8001008-6
?
21. Check that the ‘Cable’ and ‘Probe’ for P1 show OFF. Plug a cable with an invasive blood
pressure transducer into the front panel connector P1 and check that the ‘Cable’ and
Page 49
S/5 Hemodynamic modules
‘Probe’ show ON and the corresponding pressure waveform appears onto the screen.
Perform the same check also for the InvBP channel P2.
?
22. Calibrate the InvBP channels P1 and P2 according to the instructions in the chapter 3.4.4.
Invasive pressure calibration
?
23. Check the InvBP channels with a patient simulator.
The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
SENSITIVITY -switch - 5 µV/V/mmHg
ECG - BASE - BPM - 60 - BP - 1 - WAVE - ATM
2 - WAVE - ATM
SpO2 measurement
24. Check that the message ‘No probe’ is shown when no SpO2 sensor is connected to the
Restore the normal monitoring screen by pressing the key Normal Screen.
Connect cables from the channels BP1 and BP2 to the module connectors P1 and P2. Zero the InvBP channels by pressing the keys ZERO P1 and ZERO P2 on the module front panel.
BP - 1 - WAVE - ART 2 - WAVE - CVP
Check that appropriate InvBP waveforms are shown and the InvBP values are approximately 120/80 (±3 mmHg) for the channel P1 and 15/10 (±2 mmHg) for the channel P2.
Check that HR- value is calculated from P1 when ECG is not measured (ECG cable disconnected).
?
module. Connect a SpO shown when the probe is not connected to a finger.
finger probe to the module. Check that the message ‘Probe off’ is
2
?
25. Connect the SpO2 probe onto your finger. Check that the reading of
95-99 and SpO and InvBP (P1) are not measured.
waveform appear. Check that HR- value is calculated from SpO2 when ECG
2
?
35
Document No. 8001008-6
Page 50
Datex-Ohmeda S/5 monitors
Non Invasive Blood Pressure measurement
26. Enter the NIBP module service menu:
Parameters - NIBP
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing faster than by 5 per second. Check also that the NIBP board memories have passed the internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ show all OK.
?
27. Check the front panel membrane keys.
Select Buttons/Leds.
Press each of the two NIBP related membrane keys at least for one second. Check that the pressed key is identified, i.e. the corresponding text changes from OFF to ON in the menu.
?
28. Check the pump and valves.
Highlight Pneumatics from the NIBP menu. Connect a pressure manometer to the NIBP module cuff connector.
Select Start Pump and press the ComWheel. Check that the pump turns on and the pressure inside the tubing system starts to increase. Stop the pump by pressing the ComWheel again when the pressure reaches 280 mmHg.
Highlight Open Exh1. Press the ComWheel and check that the pressure inside the tubing system starts to drop then press the ComWheel again. Check the other exhaust valve by the same way by selecting Open Exh2 from the menu. If necessary, turn the pump on again for a moment to increase the pressure inside the tubing system.
Highlight Set Valve. Press the ComWheel and set the value under the text ‘Pulse Valve’ to number 150 by turning the ComWheel. Press the ComWheel again and check that the pressure inside the tubing system starts to drop. Finish the test by selecting Previous Menu.
?
36
Document No. 8001008-6
Page 51
S/5 Hemodynamic modules
29. Check the NIBP tubing system for leakages.
Select Calibrations from the NIBP service menu.
Connect the pressure manometer to the NIBP module cuff connector. Start the active leak test from the menu by pressing the ComWheel. The module pumps a pressure of about 265 mmHg and then the pump stops. Wait for 15 seconds for the pressure to stabilize then check that the pressure does not drop more than 6 mmHg per one minute. Release the pressure by pressing the ComWheel once more.
?
30. Calibration check.
Recalibrate the NIBP measurement according to the instructions in the chapter 3.4 Adjustment and Calibration. Calibration,Remember to set the calibration protection back on after the calibration.
Disconnect the pressure manometer. Select Calibrations and then highlight Calibration Check. Press the ComWheel and take down the zero offset values for both pressure transducers, B1 and B2. The values should be within ±10 mmHg.
Connect the pressure manometer to the cuff connector and check the calibration with pressures 100 mmHg, 200 mmHg and 260 mmHg. The zero offset value must be added to the displayed pressure value in order to determine the real pressure.
?
31. Check the watchdog timer activation pressure.
Select Pneumatics from the NIBP service menu.
Keep the pressure manometer connected to the cuff connector. Pump up the pressure very slowly and note the value on the manometer when your hear a signal from the loudspeaker. The pressure at where the watchdog timer should activate with an audible signal is 13 mmHg (11 to 15 mmHg). Adjust the limit with the trimmer on the NIBP board, if necessary.
?
37
Document No. 8001008-6
Page 52
Datex-Ohmeda S/5 monitors
32. Check the watchdog timer.
Select Watchdog from the NIBP service menu.
Check the watchdog timer in the adult mode. Activate the timer by highlighting Test ADULT and then pressing the ComWheel. Check that the time beside the text ‘Watchdog Interval’ starts to run. Wait until you hear a signal from the loudspeaker and then check the time again. The time from the adult test should fall within 120 to 140 seconds.
Check the watchdog timer also in the infant mode by first selecting Test INFANT from the menu. The time from the infant test should fall within 60 to 70 seconds.
?
33. Check the safety valve.
Select Safety Valve from the NIBP service menu. Keep the pressure manometer connected to the cuff connector.
NOTE: Make sure your pressure manometer can be used to measure pressures over 300 mmHg. If such a pressure manometer is not available, perform the check with an adult cuff that is connected around some round object, for example a calibration gas bottle.
Highlight Start Test. Start the adult safety valve test by pressing the ComWheel. Wait until the pump stops and the pressure is deflated. Check the pressure values ‘Max press’ and ‘2 s after stop’ for both transducers. All the values should be within 290 - 330 mmHg.
Highlight ADULT. Press the ComWheel and check that the text changes now to INFANT. Select Start Test and wait until the pump stops and the pressure values on the screen have been updated. Check that the values ‘Max press’ and ‘2 s after stop’ are all now within 154 to 165 mmHg.
Return to the normal monitoring mode by pressing
Normal Screen.
?
34. Connect an adult NIBP cuff to the cuff connector and disconnect one of its hoses.
Start NIBP measurement by pressing the key the message ‘Cuff loose’ appears on the screen within 30 seconds. Reconnect the hose and then bend it with your fingers. Restart the measurement and check that the message ‘Cuff occlusion’ appears on the screen within 30 seconds.
Start/Cancel on the module and check that
?
Check that automatic inflation limits are in use:
38
Document No. 8001008-6
NIBP - NIBP Setup - Inflation Limits - Auto - Previous Menu
Page 53
S/5 Hemodynamic modules
35. Connect the cuff onto your arm, highlight Start Ven.Stasis in the NIBP menu and press the
ComWheel. Check the module identifies the cuff, i.e. the text ‘Adult’ appears into the NIBP digit field for a short moment.
Keep the pressure inside the cuff for about half a minute in order to find out that the cuff is not leaking, then press the ComWheel again. Select
Normal Screen.
?
36. Keep the cuff on your arm and perform one NIBP measurement. Check that the module
gives a reasonable measuring result.
?
37. Connect an infant cuff to cuff connector and wrap it around your fingers.
Start NIBP measurement and check that the module identifies the cuff, i.e. the text ‘Infant’ appears into the NIBP digit field. Cancel the measurement after the cuff identification.
?
All modules
38. Perform electrical safety check and leakage current test.
?
39. Check that the module functions normally after the performed electrical safety check.
?
40. Clean the module with suitable detergent.
?
Fill in all necessary documents.
39
Document No. 8001008-6
Page 54
Datex-Ohmeda S/5 monitors
3.3 Disassembly and reassembly
3.3.1 M-ESTPR, M-ESTR, and M-ETPR modules
Disassemble the M-ESTPR/-ESTR/-ETPR module in the following way. See the exploded view of the module in 6.1.1.
1. Remove the two screws from the back of the module.
2. Pull the module box slowly rearwards and detach it from the main body. Be careful with
loose latch and spring pin for locking.
3. To detach the ECG board, remove four screws, disconnect the ribbon cable from the STP
board, and the ribbon cable from the front panel. Slide the board rearward to disconnect the fixed 10-pin connector from the ECG input board.
4. To detach the STP board, remove two screws and disconnect the two connectors from the
SP input board. The T-input connector cables must be disconnected as well.
CAUTION When reassembling the module, make sure that all cables are reconnected
properly.
3.3.2 M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR modules
Disassemble the M-NE12STPR/-NE12STR/-NE12TPR/-NESTPR/-NESTR/-NETPR module in the following way. See the exploded view of the module in 6.1.3.
1. Remove the two screws from the back of the module.
2. Pull the module box slowly rearwards and detach it from the main body. Be careful with
loose latch and spring pin for locking.
3. To detach the ECG board, detach four screws, disconnect ribbon cable from the STP board
(supply voltage), and ribbon cable from the ECG input board.
4. When the ECG board is removed, the STP board can be detached by removing four screws,
disconnecting the cable from the membrane keypad, the cable from the temperature connectors, and cables from the SP input board. Also disconnect the NIBP hoses and the ribbon cable from the NIBP board.
5. When the ECG board and the STP board are removed, the NIBP board can be detached by
removing four screws. The joining chamber can be detached by removing three screws and disconnecting the hoses from the pressure transducers and the pump. The pump can be detached by removing two screws. If the filter for the air inlet of the pump is removed, it must be replaced.
40
Document No. 8001008-6
Page 55
3.4 Adjustments and calibrations
3.4.1 Pressure safety level detection “OFFSET”
Remove two screws at the rear of the module. Remove the module box. Connect first the service cable (e.g. a long Gas Interface Cable) to the module connector inside the monitor frame and then to the rear connector of the module. Turn the monitor on. Enter to the NIBP service menu and select Pneumatics. Pump reference pressure 13 mmHg into the module. Adjust the trimmer until AD5 signal sign changes from negative to positive. Re-check the adjustement, then lock the trimmer with for example nail polish.
3.4.2 NIBP calibrations
The electronics of the NIBP pressure measurement is calibrated at the factory. Zeroing pressure is automatically maintained by the processor. If the zero point of the pressure transducer drifts more than specified, an error message is given and the NIBP board should be recalibrated or replaced.
Check the NIBP measurement, and recalibrate if necessary , once a year. The checking and recalibrating can be done in the NIBP service menu.
S/5 Hemodynamic modules
The calibration of the primary pressure channel can also be checked from the NIBP setup menu (NIBP - NIBP Setup - Calibration Check). In this case the auto zeroing is performed at start ­remove hose before entering to ensure atmospheric pressure to the pressure transducers - the primary pressure is displayed. The zero-offset value should then be zero.
Check the intake air filter as part of the calibration check.. Change the filter if it is visibly dirty.
Calibration check
1. Enter Calibration menu.
2. Select Calibration Check and press the ComWheel.
3. Connect an external precision manometer to the module.
41
Document No. 8001008-6
Page 56
Datex-Ohmeda S/5 monitors
4. Pump the following pressures to manometer and check the difference between the
manometer and monitor pressure display:
Table 5 NIBP calibration check pressures
Pressure Max. error Example
0 mmHg ±9 mmHg (=zero offset) -2
100 mmHg 100 + zero offset ±2 mmHg 98 ±2
200 mmHg 200 + zero offset ±3 mmHg 198 ±2
If the error of pressure channel B1 is larger than specified above, the module should be recalibrated. The error of B2 is allowed to be even twice as large because it has no effect on blood pressure measurement accuracy. However, we recommend recalibrating the module when the error of B2 is larger than specified above to ensure best possible operation.
Calibration
1. Enter Calibration menu.
2. Remove hoses from front panel connector to enable proper zeroing.
3. Select Calibration. If it is not available, perform the steps A, B, and C.
NOTE: Do not pull out the NIBP module from the monitor frame. The module must be in the frame during the whole procedure.
A. Turn the toggle switch at the bottom of the NIBP module to enable the calibration. Turn the
switch to the right by, for example, a sharp pencil. This enables menu selection Protection. The message ‘Calibration switch ON!’ appears.
B. Select Protection OFF in the Calibration menu and press the ComWheel.
C. Return the toggle switch to the left. Menu selection Calibration is now enabled, and
Protection is disabled. When the calibration is enabled, a message ‘Calibration not protected’ appears.
Start Calibration by pressing the ComWheel. Messages ‘Zeroing’ and ‘Zeroed’ will appear in
the NIBP message field. After this a pressure bar will appear.
Connect an external mercury manometer with pump to module through the both tubes of
the hose - both transducers B1 and B2 must be calibrated simultaneously. Pump up to a pressure about 200 mmHg according to the manometer. Calibration is possible in the range 150 to 300 mmHg.
42
Document No. 8001008-6
Verify that both pressure values in the prompt field match the manometer reading. If not,
adjust by turning the ComWheel. When the values of the pressure bar and the manometer are equal, press the ComWheel to confirm the calibration. The message ‘Calibrating’ will appear onto the NIBP digit field. After a few seconds it is followed by ‘Calibrated’, which means that the calibration has succeeded, and the new calibration data has been saved into EEPROM.
Page 57
NOTE! When calibrating NIBP, always change the displayed pressure value slightly with the
ComWheel, even in cases where the value would be correct (i.e. change the value for example one step higher and then back one step lower). “Calibrated” text should appear in the display. This ensures that the calibration procedure is correctly registered and stored by the module.
To set the protection on:
Turn the toggle switch to the right. Select Protection ON and push the ComWheel. Then turn the toggle switch back to the left.
Remove the module from the frame and plug it back again. Then perform Calibration check
(see the preceding page) to verify the new calibration.
3.4.3 Temperature calibration
NOTE: For the temperature calibration, separate, accurate test plugs (25 °C and 45 °C) are needed. A test set of two plugs is available from Datex-Ohmeda, order code 884515.
Calibrate temperature when measured test values deviate more than ±0.1 °C, and always after STP board replacement.
S/5 Hemodynamic modules
1. Enter ESTPR: STP service menu.
2. Enter Calibrations menu.
3. Press the protect button at the bottom of the module and choose OFF in protect mode.
Release the button.
4. Select Calibrate T1/Calibrate T2.
5. Insert calibration plug (25 °C) into T1/T2 connector.
6. Press the ComWheel.
7. Insert calibration plug (45 °C) into T1/T2 connector.
8. Press the ComWheel.
9. Press in the protect button at the bottom of the module and choose ON in protect mode.
Release the button.
3.4.4 Invasive pressure calibration
Calibrate invasive pressure when the pressure transducer (probe) is replaced with a different type of transducer, and when STP board is replaced.
1. Enter ESTPR: the STP service menu.
Monitor Setup, Install/Service (password 16-4-34), Service (password 26-23-8),
( Parameters).
2. Enter Calibrations menu.
3. Connect a pressure transducer with a pressure manometer to the P1/P2 connector. Choose
Calibrate P1 or Calibrate P2. Leave the transducer to room air pressure.
4. Press the ComWheel to start zeroing.
43
Document No. 8001008-6
Page 58
Datex-Ohmeda S/5 monitors
5. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
6. Set the pressure on the display to match the pressure reading on the manometer and press
the ComWheel. A tolerance of ±1 mmHg is allowed.
7. The message ‘Calibrated’ will appear on the display.
44
Document No. 8001008-6
Page 59
S/5 Hemodynamic modules
4 TROUBLESHOOTING
4.1 Troubleshooting charts
See also the User’s Reference Manual for more troubleshooting procedures.
4.1.1 NIBP
TROUBLE CAUSE TREATMENT
No NIBP value displayed NIBP not selected on screen. Check monitor setup.
NIBP menu fading No M-NE(12)STPR module, module
not properly connected, or NIBP and NE(12)STPR module connected at the same time.
Artifacts-message Unsuccessful measurement due to
patient movement or shivering.
Weak pulsation-message Weak or unstable oscillation pulses
due to:
artifacts (accurate diastolic
pressure difficult to measure)
marked arrhythmia
marked drop in diastolic
pressure
diastolic pressure difficult to
measure
improper cuff position or
attachment
too few pulses detected
Plug in the module.
Check patient condition and retry.
Check any leaks and retry.
Use proper size of cuff. Check attachment.
Call service Error X-message
weak or unusual blood
circulation
may give systolic value
NIBP hardware error. X = error number.
See the description of the error message code in 4.1.2, the causes and the solutions listed in the next chapter.
45
Document No. 8001008-6
Page 60
Datex-Ohmeda S/5 monitors
TROUBLE CAUSE TREATMENT
Cuff loose-message 1. Hose and/or cuff not connected. 1. Connect the hose and the cuff.
2. Hose and cuff connected.
2.
Reasons:
cuff loosely wrapped tighten the cuff
leakage in cuff or hose replace cuff/hose
leakage inside module check internal tubing and air
chamber, and fix if necessary
pump does not work check pump connector; if OK, replace
pump
no pulses during the last three
check cuff positioning
measurements
Air leakage-message 1. Hose or cuff leaking. Reasons: 1. Replace cuff
cuff damaged replace cuff
cuff connector damaged replace cuff connector (if the fault is
in hose connector,)
O-ring damaged or missing replace O-ring
hose double connector
replace hose
damaged
Unable to measure Sys­message
2. Hose and cuff OK. Reasons: 2. Connect or replace tube
leakage inside the module replace the whole tubing
tube disconnected or damaged fix connections
air chamber leaking
tubes or valve(s) damaged replace tubes/valve(s)
Systolic blood pressure probably higher than the maximum inflation
Automatic retrial with increased pressure.
pressure.
46
Document No. 8001008-6
Page 61
TROUBLE CAUSE TREATMENT
S/5 Hemodynamic modules
Cuff occlusion-message 1. Cuff and/or hose occluded.
Reason:
cuff tube kinked straighten tube
tube inside module kinked straighten tube
occlusion inside/outside
module
2. Cuff, hose, and tubes OK. Reason:
fault in pressure transducer replace the NIBP board
fault in A/D converter replace the NIBP board
faulty calibration check calibration
missing voltages recalibrate
Calibration switch on ­message
EEPROM protection switch at the bottom of the module is turned to right.
Calibration not protected ­message.
Calibration protection is set to OFF.
1.
remove occlusion
2.
Enables setting the protection OFF in the Calibration menu. Turn switch to left if you are not going to calibrate.
Set the protection ON in the NIBP Calibration menu.
47
Document No. 8001008-6
Page 62
Datex-Ohmeda S/5 monitors
4.1.2 NIBP error code explanation
Code Explanation Treatment
0 RAM failure; memory failure Change NIBP board.
1 ROM checksum error; memory failure Change NIBP board.
2 +15 V failure Check short circuits. Change NIBP
3 -15 V failure Check short circuits. Change NIBP
board.
board.
4 EEPROM protection switch error (only
with S-STD93)
5 Calibration not protected. (only with S-
STD93)
6 ADC error ADC circuit failure. Change NIBP board.
7 Watchdog time too short Change NIBP board.
8 Watchdog time too long Change NIBP board.
9 Watchdog activated Change NIBP board.
10 EEPROM checksum error; memory failure Change NIBP board.
11 Auto zero range exceeded Calibrate NIBP.
12 Communication break; temporal break
down of communication from monitor detected
13 - -
14 Too early Auto Start (needs 25 seconds
without pressure)
Turn the toggle switch to the left at the bottom of the module.
Protect calibration by selecting Protection ON in the NIBP calibration menu.
Automatic recovery
48
Document No. 8001008-6
Page 63
4.1.3 ECG
TROUBLE CAUSE TREATMENT
S/5 Hemodynamic modules
HR numerical display shows ‘---’
Unacceptable ECG waveform Poor electrode or poor electrode
No ECG trace Waveform not selected on screen.
No heart rate available. If no ECG waveform, check LEADS OFF
skin contact.
Poor electrode condition. Electrodes are dried out.
Improper site of electrodes. Check that electrodes are not placed
Improper skin preparation. Remove body hair. Clean attachment
Improper bandwidth filter. Check filter.
message and connect the leads.
If ECG waveform exists, check heart rate source e.g. in the ECG Setup menu behind ECG key.
Electrodes from different manufacturers are used. /Too much/little gel is used.
over bones, active muscles, or layers of fat.
site carefully with alcohol.
Press the make adjustments.
Monitor Setup key and
Module not plugged in correctly. Plug in.
Noise-message High frequency or 50/60 Hz noise. Isolate noise source.
4.1.4 Pulse oximetry (SpO2)
TROUBLE CAUSE TREATMENT
Message ‘NO PROBE’ No probe connected to the
monitor.
Probe faulty. Change the probe.
Message ‘PROBE OFF’ though probe properly attached to the patient
Finger probe falls off 1. Probe is slippery. 1. Wipe with 70 % isopropyl alcohol
Unsuitable site. Try another site.
Probe faulty. Try another probe.
Probe connection cable not connected to probe.
Check probe connections.
Connect the cable to probe.
and allow to dry.
49
Document No. 8001008-6
Page 64
Datex-Ohmeda S/5 monitors
TROUBLE CAUSE TREATMENT
2. Finger is too thin or thick. 2. Try other fingers, or other probe types.
Weak signal artifacts Poor perfusion. Try another place.
Movement artifacts.
Shivering.
Message ‘NO PULSE’ Pulse search > 20 sec. and low
or low pulse rate.
SpO
2
Message ‘ARTIFACT’ Pulse modulation exceeds the
Try other fingers.
Try another place or another probe.
present scale.
Message ‘CHECK PROBE’ DC value not in balance. Try another probe.
Message ‘POOR SIGNAL’ Modulation (Red or Ired) < 0.25 % Patient may be cold.
Message ‘FAULTY PROBE’ Probe is faulty. Change the probe.
No SpO
2
No waveform selected on screen. Check selected SpO2 waveforms by
pressing
Monitor Setup key and
selecting Modify waveforms.
Wrong configuration setting. Check the configuration settings from
the ESTPR:STP/Calibrations menu
Monitor Setup - Install/Service -
( Service - Parameters)
4.1.5 Temperature
TROUBLE CAUSE TREATMENT
Message ‘TEMPERATURE ERROR’ Faulty calibration. Perform calibration. If it does not
help, check that front panel connector is properly connected to STP board.
No temperature displayed Wrong type of probe. Use correct probe.
Temperature out of measurable
The range is between 10 and 45 °C.
range.
Temperature calibration not protected.
Set the protection ON in the Service Menu.
50
Document No. 8001008-6
Page 65
S/5 Hemodynamic modules
4.1.6 Invasive blood pressure
TROUBLE CAUSE TREATMENT
Abnormally low pressure Transducer wrongly positioned. Check mid-heart level and reposition
transducer.
No pressure Defective transducer. Check transducer.
No pressure module plugged in. Check the module.
No waveform selected on screen. Check selected pressure waveforms
by pressing Monitor Setup key and selecting modify waveforms.
Check that pressure transducer is open to patient.
Wrong configuration setting Check the configuration setting from
the ESTP:STP/Calibrations menu
Monitor Setup - Install/Service -
( Service - Parameters)
Not zeroed -message Measurement on, channel not
zeroed.
Zeroing failed -message Unsuccessful zeroing of P1 /P2
(number field).
Calibration failed -message Unsuccessful calibrating of
P1/P2 (number field), possibly due to pulsating waveform
Out of range < 40 mmHg Measurement pressure is beyond
measurement range.
Out of range > 320 mmHg Measurement pressure is beyond
measurement range.
Zero the channel.
Possibly due to pulsating pressure waveform. Open the transducer to air and zero the channel.
Offset is > 150 mmHg. Open the transducer to air and zero the channel.
Defective transducer. Replace it and zero the channel.
Turn the transducer to sphygmomanometer and try again (zeroing takes place first).
Gain is beyond the limits (± 20 % of the default gain). Replace the transducer.
Check transducer level. Zero the channel.
Check transducer level. Zero the channel. The patient may also have high pressure.
Zero adj. > 100 mmHg Offset when zeroing is > 100
mmHg (but < 150 mmHg) from the absolute zero of the module (with default gain).
Check transducer. The waveform may hit the top and the numeric display not shown.
51
Document No. 8001008-6
Page 66
Datex-Ohmeda S/5 monitors
TROUBLE CAUSE TREATMENT
Out of range Measured pressure is beyond the
internal measurement range of the module.
The waveform hits the top and the numeric display not shown. Check transducer and its level. Zero the channel.
4.1.7 Impedance respiration
TROUBLE CAUSE TREATMENT
No resp trace Waveform not selected on the
screen
Module not plugged in correctly Plug in
Unacceptable resp waveform Poor electrode or poor electrode
skin contact
Poor electrode condition Electrodes are dried out.
Improper site of electrodes Check that electrodes are not placed
Press the Monitor Setup key and make adjustments
Electrodes from different manufacturers are used. Too much/little gel is used.
over bones, active muscles, or layers of fat.
Improper skin preparation Remove body hair. Clean
attachment site carefully with alcohol.
Message: ‘SMALL RESP CURVE’ Respiration signal is very small With 3-lead cable in ESTPR/NESTPR
try another lead connection I, II, III or try 5-lead cable.
Message: ‘APNEA ALARM’, and respiration waveform normal
Respiration source is CO
2
Check respiration source and change it to correct one.
52
Document No. 8001008-6
Page 67
4.2 Troubleshooting flowcharts
4.2.1 M-NE12STPR and M-NESTPR module troubleshooting
Possible faulty
START
module with NIBP
parameter.
S/5 Hemodynamic modules
Insert NXXX module
and turn power on.
Does fault
still appear?
YES
Module ID
on screen?
YES
Error
message on
screen?
YES
See error code explanation in Service Manual and
fix it.
NO
NO
NO
Remove all
modules from
Monitor Frame.
Fault not in
NXXX module.
NO
Does
another module
work in same
slot?
Key check
in Service Menu
OK?
YES
YES
NO
Check and
replace
NIBP board.
Check keyboard
connector. If OK,
change front panel.
Replace
NXXX module.
NO
OK?
Check and
replace
NIBP board.
NO
OK?
Do
they work?
NO
Check tubes and
connectors. Find leak and
fix it.
Figure 15 M-NE12STPR and M-NESTPR module troubleshooting flowchart
Pump check
in Service Menu
OK?
YES
Leak test
in Service Menu
OK?
NO
NO
YES
Check pump
connector. If OK,
change pump.
NESTPR_module_troubles_flowch.vsd
53
Document No. 8001008-6
Page 68
Datex-Ohmeda S/5 monitors
4.2.2 M-ESTPR, M-ESTR, and M-ETPR module troubleshooting
Possibly fault in module with
ESTPR parameters.
Check module configuration in Serv ice menu (ESTPR/STP/ Cal ibration s)
Is it
correct?
YES Remove all modul es from
the frame
Insert ESTPR module and t urn power on
Does faul t
still appear?
YES
Enter the
Service menu
STP module ID on screen?
YES
NO
NO
NO
Select the correct configuration
Fault not in ESTPR module
NO
Does
another
mod ule wo rk
in same
slot?
YES
Remove
ECG board
Replace
STP board
NO
STP module ID on screen?
YES
ECG module ID on screen?
YES Check front panel key functions in Service menu
Do
they wo rk?
NO
Check front p anel unit, ECG board, and STP board to find culprit. Replace if necessary.
Figure 16 M-ESTPR Module Troubleshooting Flowchart
54
Document No. 8001008-6
NO
YES
Has ECG
measurement
failed?
YES
Replace ECG board
NO
board failed?
Has STP
YES
Replace STP board
NO
Replace ECG board
Has invBP
measurement
failed?
YES
Replace STP board
NO
Has Temp
measurement
failed?
YES
Replace STP bo ard
NO
measurement
Has SpO
failed?
YES
Replace STP board
NESTPR_MESTPR_module_trou
bles_flowch.vsd
2
Page 69
5 SERVICE MENU
NOTE: The Service Menu pictures are for reference only. Details on the menu page can vary depending on the software version.
S/5 Hemodynamic modules
1. Press the Monitor Setup key.
2. Select Install/Service (password 16-4-34).
3. Select Service (password 26-23-8).
4. Select Parameters - NIBP.
NOTE: Parameter values in Service Data fields are for reference only on this chapter.
55
Document No. 8001008-6
Page 70
Datex-Ohmeda S/5 monitors
5.1 NIBP service menu
Service Data
Pressure shows measured pressure multiplied by 10. Zero shows pressure at auto zeroing multiplied by 10 and changes between +20 and -20 mmHg. Absolute pressure is the sum of Pressure and Zero.
Protect handle indicates hardware protection for EEPROM memory. It should be ON all the time in
normal operation. If it is OFF data can not be read from or written to EEPROM, only the calibration protection can be set or reset by software. It can be turned to OFF by turning the toggle switch to the right at the bottom of the module, which also enables Protection ON/OFF menu selection in the calibration menu.
Calibr. prot. shows software calibration protection and it should be OFF to enable calibration.
+15 V power indicates the condition of the supply voltage +15 Vdirty for the pump and valves. It
exists (ON) or not (OFF) depending on service menu function. The supply voltage can be turned on by selecting the previous Menu and then the desired menu again.
AD0 to AD7 show the values of each eight channels of A/D converter.
Timeouts is a cumulative number that indicates how many times the module has not responded to
the monitor’s inquiry. Bad checksums is a cumulative number that indicates how many times communication from the module to monitor broke down. Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected. The monitor starts counting these items at power up and resets to zero at power off. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) indicates either serial communication failure, or module not in place. Also other modules can cause communication errors that cause these numbers rise.
56
Document No. 8001008-6
RAM indicates the state of the RAM memory. ROM indicates whether the checksum in the EPROM is in accordance with the one the software has
calculated.
Page 71
EEPROM indicates if the values stored in the permanent memory are valid. The state is either OK, Fail or ? (module not in place or a communication error).
5.1.1 NIBP demo menu
S/5 Hemodynamic modules
A service menu for demonstrating the oscillometric method of NIBP measurement. The menu shows the realtime pressure signals that are measured from the NIBP cuff. The measurement result is shown in the adjoining digit field.
Wave Recording Wave Recording is for selecting the recording option. If ON is selected, the pressure signals are
recorded in realtime onto the M-REC paper.
Remove menu Remove menu widens the displayed waveform area.
Previous Menu The menu can be closed by selecting the Previous Menu or just by pressing the ComWheel if the
Remove menu was selected.
57
Document No. 8001008-6
Page 72
Datex-Ohmeda S/5 monitors
5.1.2 NIBP calibration menu
Active Leak Test Wrap an adult cuff around a pipe and connect the cuff to the module. Select the active leak test
(ON). The module automatically pumps a pressure of 260 mmHg into the cuff. Wait for several seconds until the pressure stabilizes. Then check that the pressure reading does not drop more than 6 mmHg per minute. If it does, leaking point(s) should be detected and fixed. Cancel the test by selecting Active leak test OFF.
Calibration Check
After the calibration check is selected (ON), manually pump pressure into the module and make sure that the same pressure values are shown both on the display and on manometer. Pressure of both pressure channels B1 and B2 are shown. Note that if the display shows +2 mmHg at zero pressure and if you pumped +200 mmHg into the module, the display should show +202 mmHg.
Protection Software calibration protection (ON/OFF). Select OFF when calibrating. Protection can be set to ON
or OFF only when the toggle switch at the bottom of the module is set to the right.
Calibration Calibration selection is available only when protection is OFF.
NIBP calibration can be performed in the NIBP Service menu as follows:
NOTE: Both channels B1 and B2 must be calibrated simultaneously.
1. If Protection is ON change it to OFF by first turning the toggle switch to the right at the
bottom of the module, which enables the Protection selection. Then turn the toggle switch to the left to enable Calibration.
NOTE : Do not disconnect the module from the frame when turning the switch. The module must be in the frame during the whole procedure.
58
Document No. 8001008-6
NOTE: When the switch is at the right, the NIBP field shows an error message ‘Calibration switch on!’.
NOTE: When calibration is enabled, a message ‘Calibration not protected’ appears.
2. For proper zeroing to take place, remove the hose from the front panel connector. Select
Calibration and push the ComWheel. Messages ‘Zeroing’ and ‘Zeroed’ will appear in the NIBP message field. After this a pressure bar will appear beside the menu.
Page 73
3. Connect an external mercury manometer with pump to module through the both tubes of the
hose. Pump up to about 200 mmHg pressure (range of 150 to 300 mmHg allowed) according to the manometer. Verify that both pressure values in the prompt field match the manometer reading. If not, adjust by turning the ComWheel.
4. When the values are equal, push the ComWheel to confirm the calibration. First the message
‘Calibrating’ will appear in the digit fields for NIBP followed after a few seconds ‘Calibrated’, which means that the calibration data has now been saved. NOTE! When calibrating NIBP, always change the displayed pressure value slightly with the ComWheel, even in cases where the value would be correct (i.e. change the value for example one step higher and then back one step lower). “Calibrated” text should appear in the display. This ensures that the calibration procedure is correctly registered and stored by the module.
5. Use the bottom switch to enable Protection setting and set it ON, and finally disable
Protection setting.
5.1.3 NIBP safety valve menu
S/5 Hemodynamic modules
Start Test Start test is for starting and Stop test is for stopping the Safety Valve test.
Safety Valve Data
See NIBP Service menu in chapter 5.1 for information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc.
Max. press and 2 s after stop show the measured values at Safety Valve test.
Safety Valve Test Adult/Infant
Wrap an adult cuff around a pipe and connect the cuff to the module. Highlight Start test and give the ComWheel a push. The test ends automatically or when Stop test (appears in place of Start
test) is pushed.
Max. press indicates the pressure at which the safety valve opens and is normally 310 ±15 mmHg
for adult and 150 mmHg ±15 mmHg for infant. 2 s after stop indicates the pressure at 2 seconds after the pump has stopped and is normally > 280 mmHg for adult and > 120 mmHg for infant. If the value is less, check leakage by the active leak test.
59
Document No. 8001008-6
Page 74
Datex-Ohmeda S/5 monitors
5.1.4 NIBP pulse valve menu
Start Test Start test is for starting and Stop test is for stopping the test.
Set Valve Set Valve lets you adjust the opening of the pulse valve.
Pulse Valve Data
See NIBP Service menu in chapter 5.1 for information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc.
Pulse Valve Checking
Wrap an adult cuff around a pipe and connect the cuff to the module. Select the Start test and push the ComWheel. The pressure rises beyond 240 mmHg and stops. The pulse valve opens. The module counts the time it takes for the pressure to go down from 240 mmHg to 50 mmHg and displays it on the screen. The test can be manually stopped by selecting Stop test.
The valve can be adjusted between 0 and 255 (0 for fully closed and 255 for fully open). First select Set Valve and push the ComWheel. See the pulse valve value and adjust it by turning the ComWheel. Then push the ComWheel to confirm the value.
The ‘Interval 240 mmHg -> 50 mmHg’ time should be less than 60 seconds when the valve is ‘150’ and less than 10 when fully opened (255). When fully closed (0), the system should be airtight and the pressure does not drop. Depending on an individual, the pulse valve may remain closed up to approx. value 100.
60
Document No. 8001008-6
If the measured time deviates much from those above, then the pulse valve or its tubes are faulty.
Page 75
5.1.5 NIBP buttons/leds menu
S/5 Hemodynamic modules
The selections Auto ON/OFF, Manual ON/OFF, STAT ON/OFF, and Measur. ON/OFF have no effect on the module.
Buttons/Leds Data
See NIBP Service menu in chapter 5.1 for information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc.
Buttons Checking
The front panel keys function is confirmed by pressing the key and observing OFF turns to ON at
Auto On/Off, and Start Cancel.
61
Document No. 8001008-6
Page 76
Datex-Ohmeda S/5 monitors
5.1.6 NIBP pneumatics menu
Start Pump/Stop Pump
A manual control for the pump. The selection changes to Stop Pump when the pump turns on.
Open Exh1/Close Exh1
A manual control for the exhaust valve 1. The selection changes to Close Exh1 when the valve is opened.
Open Exh2/Close Exh2
A manual control for the exhaust valve 2. The selection changes to Close Exh2 when the valve is opened.
Set Valve With Set Valve, the opening of the pulse valve is adjusted between 0 and 255 (0 for fully closed
and 255 for fully open). First push the ComWheel, then turn it to adjust the value on screen and finally push to set the value.
Reset Clock Reset Clock will zero the time on the display.
Pneumatics Data field
See NIBP service menu in chapter 5.1 for information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc.
Pump, Exh1 Valve, and Exh2 Valve show their states.
Pulse Valve shows how much the valve is opened (0 to 255) during Valve Setting.
62
Document No. 8001008-6
Interval 20 mmHg -> 185 mmHg Checking
Select the Start pump at different combinations of the valves open/closed and push the ComWheel. The module counts the time it takes for the pressure to go up from 20 mmHg to 185 mmHg and displays it. When all the valves are closed, the pump should be able to pump the pressure in about 1 to 4 seconds into an adult cuff wrapped around a pipe. The pump does not stop without selecting the Stop Pump by pushing the ComWheel.
Watchdog BEEP
Connect manometer to the front panel and pump pressure into the module. When the AD5 value
Page 77
changes from negative to positive value (at about 13 mmHg) a beep is heard. This is the watchdog threshold pressure. Beyond this pressure the watchdog is active and cut pressures at about 2 min. (adult).
5.1.7 NIBP watchdog menu
S/5 Hemodynamic modules
Test ADULT Test ADULT is to test watchdog timer in adult mode (120 to 140 seconds).
Test INFANT Test INFANT is to test watchdog timer in infant mode (about 60 to 70 seconds).
Stop Test Stop Test is for stopping the test.
Watchdog Data field
See NIBP Service menu in chapter 5.1 for information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc.
Watchdog Interval shows the time the +15 Vdirty stays on during the test.
Adult watchdog time testing
Select Test ADULT and push the ComWheel. Watchdog interval starts counting up seconds and keeps on counting as long as the +15 Vdirty is on. The time should be 120 to 140 seconds.
Infant watchdog time testing
Select Test INFANT and push the ComWheel. Watchdog interval starts counting up seconds and keeps on counting as long as the +15 Vdirty is on. The time should be 60 to 70 seconds.
63
Document No. 8001008-6
Page 78
Datex-Ohmeda S/5 monitors
5.2 ECG service menu
Power freq Set power frequency; 50 Hz/60 Hz.
Filter low Set filter low frequency; 0.05 Hz/0.5 Hz.
Filter high Set filter high frequency; 30 Hz (40 Hz if power freq is 60 Hz) /100 Hz or 150 Hz @ NE12STPR.
Service Data field
Power freq, and Cable type show the values chosen or detected, Filter low and high defines the
selected filter (Monitor/Diagnostic/ST).
Quick zero @ NESTPR and ESTPR modules is ON when the signal in any of the three internal amplifier goes beyond scale, and therefore, a capacitor connected to the related channel discharges overvoltage. At least one of Quick zero values is OFF when 3-lead cable is used. All three values are OFF when 5-lead cable is used. Quick zero also takes place when lead is changed in 3-lead measurement. @ NE12STPR Quick zero is on if any of the ECG amplifiers goes beyond the scale.
Cable shows ON when an ECG cable is connected.
Electrode shows ON when each of these electrodes are connected. Pacer count is a running number for pacemaker users.
The front panel ECG key function is confirmed by pressing the key and observing OFF turns to ON at
Button.
64
Document No. 8001008-6
NOTE: M-NE12STPR and M-NESTPR Module does not consist ECG key.
Resp Available indicates that ECG hardware is capable of measuring impedance respiration.
Measurement shows ON when the respiration measurement is on.
Amp zero shows ON when zeroing of the respiration amplifier takes place.
Waveform VALUE will be updated in one second interval.
Page 79
S/5 Hemodynamic modules
Timeouts is a cumulative number that indicates how many times the module has not responded to the monitor’s inquiry. Bad checksums is a cumulative number that indicates how many times communication from the module to monitor broke down. Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected. The monitor starts counting these items at power up and resets to zero at power off. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) indicates either serial communication failure, or module not in place. Also other modules can cause communication errors that cause these numbers rise.
RAM indicates the state of the RAM memory. ROM indicates whether the checksum at the EPROM is in accordance with the one the software
has calculated. EEPROM indicates if the values stored in the permanent memory are valid. The state is either OK, Fail or ? (module not in place or a communication error).
65
Document No. 8001008-6
Page 80
Datex-Ohmeda S/5 monitors
5.2.1 ECG setup menu
Filter Filters the ECG signal high frequency noise and slow respiratory artefacts.
Monit (monitor) filter is used in routine monitoring. It effectively filters the artefacts caused by the
electrosurgery unit and respiration.
Diagn (diagnostic) filter is used if more accurate information of the waveform is needed (e.g., of P- wave or AV block). The diagnostic filter is more susceptible to both high frequencies and baseline wander than monitor filter.
STfilt (ST filter) permits more accurate information of ST segment. It filters the high frequency artefacts caused by electrosurgery unit but catches the slow changes in ST segment. The ST filter is more susceptible to baseline wander than the monitor filter.
Pacemaker Selects how to display the pacing pulse of cardiac pacemaker. The selections are Show, Hide, ON
R and Sensit.
Hide, the pacing pulse is filtered away from ECG data.
Show, the pacer pulse is filtered away from ECG data but the pulse is displayed as a constant
height marker.
ON R, pacing pulses are not filtered away from ECG data. This improves ECG monitoring with A-V pacemaker patients, as QRS complexes are counted even if the pacing pulse hits the QRS complex. However, during asystole the monitor may count pacing pulses as heart beats.
Sensit selection uses a more sensitive pacemaker detection. Pacemaker spike is displayed on ECG.
66
Document No. 8001008-6
Page 81
S/5 Hemodynamic modules
5.3 STP service menu
Record Data Record Data prints out the shown service data and board information (id, serial number and sw id)
onto the recorder module, M-REC.
Temp Test Temp Test activates the automatic temperature test for the temperature channels T1 and T2. The
result from the test is shown in the service data field.
NOTE: The Temp Test needs to be selected twice before the test starts.
Service Data field
Gain is a coefficient to compensate gain error. Usually the values for P1 and P2 are between
17000 and 25000 and for T1 and T2 between 13000 and 14300. Zero indicates offset compensation value of each parameter in A/D converter. Typically the values for P1 and P2 are within ±1000 and for T1 and T2 between -150 and +300. Calibrate if zero and/or gain value is outside the ranges.
Cable shows ON when a corresponding cable is connected to the front panel and Probe shows ON when a corresponding probe is connected to the cable.
Under Value the measured numeric values are displayed simultaneously. Pressure values are real time values and shown in mmHg. Temperature values are shown in degrees Celsius.
The front panel STP keys functions are confirmed by pressing each key and observing OFF turns to ON at Button.
shows measured beat-to-beat SpO2 value. Modpr is a modulation % that indicates AC/DC
SpO
2
ratio in the measured signal. Hr is a pulse rate calculated from every beat.
Cable and Probe can be either OFF or ON, and these indicate the state PROBE OFF.
Under them there is a message field for SpO
. It can be OK, PULSE SEARCH, NO PROBE, PROBE
2
OFF, NO PULSE, ARTEFACT, POOR SIGNAL, or CHECK PROBE.
Balance between leds is adjusted by changing the intensity of red/infrared. Intensity of infrared (Ired int.) is in the range of 40 to 255 and red intensity (red int.)is in the range of 40 to 255.
Document No. 8001008-6
67
Page 82
Datex-Ohmeda S/5 monitors
DC gain shows the gain of DC signal adjusted by the module.
IDC is the value of infrared signal.
RDC is the dc value of red signal.
AC gain is the gain of infrared and red ac signals. AC gain values can be 1 or 0. Value 1 means high
ac gain and 0 means low gain.
Pre gain is a preamplifier gain for infrared and red signals. Pre gain values can be 1 or 0. Value 1 means normal operation. Value 0 means that signal levels are very low and extra gain is taken into use.
Temp error shows the status of the temperature test. No errors found shows the status (OFF) and errors found (ON).
Protect key shows normally OFF but turns to ON when the button at the bottom of the module is pressed.
Protect mode is normally ON. It turns to OFF when Protect is switched to OFF for the temperature calibration in Calibration Menu.
Configuration shows the chosen module configuration: TP, ST, or STP.
Timeouts is a cumulative number that indicates how many times the module has not responded to
the monitor’s inquiry. Bad checksums is a cumulative number that indicates how many times communication from the module to monitor broke down. Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected. The monitor starts counting these items at power up and resets to zero at power off. The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) indicates either serial communication failure, or module not in place. Also other modules can cause communication errors that cause these numbers rise.
RAM indicates the state of the RAM memory. ROM indicates whether the checksum at the EPROM is in accordance with the one the software
has calculated. EEPROM indicates if the values stored in the permanent memory are valid. The state is either OK, Fail or ? (module not in place or a communication error).
68
Document No. 8001008-6
Page 83
S/5 Hemodynamic modules
5.3.1 STP calibration menu
Protection Protection for the configuration and temperature calibrations can be set ON and OFF only when
protect button at the bottom of the module is pressed.
Set Config The module configuration should be set according to the module type. The setting is possible only
when the protection is set OFF. The available selections are TP, ST or STP. The configuration setting should be checked if the STP board is replaced.
Calibrate T1 / Calibrate T2
The functions are for calibrating the temperature channels T1 and T2. The calibrations are possible only when the protection is set OFF. The temperature calibration requires accurate test plugs of value 25 °C and 45 °C.
Calibration:
1. Select Calibrate T1/Calibrate T2
2. Insert the test plug 25 °C into the T1/T2 connector
3. Press the ComWheel
4. Insert the test plug 45 °C into the T1/T2 connector
5. Press the ComWheel
Calibrate P1/Calibrate P2
The functions are for calibrating the invasive blood pressure channels P1 and P2. The calibrations require a pressure transducer (with an appropriate cable) and a pressure manometer.
1. Connect the pressure transducer with the pressure manometer to the P1/P2 connector.
Select Calibrate P1/Calibrate P2. Leave the transducer to room air pressure.
2. Press the ComWheel to start zeroing.
3. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
4. Set the pressure on the display to match the pressure reading on the manometer and press
the ComWheel.
69
Document No. 8001008-6
Page 84
Datex-Ohmeda S/5 monitors
6 SPARE PARTS
6.1 Spare parts list
6.1.1 M-ESTP rev. 01, M-ETP rev. 00, M-EST rev. 00
Figure 17 Exploded view of M-ESTP Module
Item Description Order No. Replaced by
1 Module box (wide) 886168
2 Latch for module box 879181
3 Spring pin 879182
4 Cross recess screw M3x8 black 616215
5 Front mask unit, M-EST (rev.00) 880946
5 Front mask unit, M-EST (rev.01-02) 882129
5 Front mask unit, M-EST (rev.03-04) 887155
5 Front mask unit, M-ESTP (rev. 01) 880337
5 Front mask unit, M-ESTP (rev.02-03) 882127
5 Front mask unit, M-ESTP (rev.04-05) 887153
5 Front mask unit, M-ESTPR (rev.03-04) 894161
5 Front mask unit, M-ESTR (rev.03-04) 894162
5 Front mask unit, M-ETP (rev.00) 880941
5 Front mask unit, M-ETP (rev.01-02) 882128
5 Front mask unit, M-ETP (rev.03-04) 887154
5 Front mask unit, M-ETPR (rev.03-04) 894163
6 PP board, M-PP, M-ESTPR 891442 8000801
6 STP board, M-ESTPR 8000801 8002575
6 STP board, M-ESTP (rev.01), M-P (rev.00-01) 880339
70
Document No. 8001008-6
Page 85
S/5 Hemodynamic modules
Item Description Order No. Replaced by
6 STP board, M-ESTP (rev.03-05), M-P(rev.02) 882627
6 STP board, M-ESTPR 8002575
6 STP board (rev.02) 882130 882627
7 ECG board 884609 888956
7 ECG board, M-ESTP 888957
7 ECG board, M-ESTP (rev.03) 883119 888957
7 ECG board, M-ESTP (rev.04) 886748 888957
7 ECG board, M-ESTP(rev.01) 880338
7 ECG board,M-ESTP (rev.02) 882025 888957
7 ECG/RESP board, M-ESTPR (Rev.02) 888956
8 Membrane keypad 879374
10 Front Panel sticker, DA ; M-ESTPR (rev.01-03) 892207 898504
10 Front Panel sticker, DA ; M-ESTPR (rev.04) ; S/5 898504
10 Front Panel sticker, DA ; M-ESTR (rev.01-03) 892208 898565
10 Front Panel sticker, DA ; M-ESTR (rev.04) ; S/5 898565
10 Front Panel sticker, DA ; M-ETPR (rev.01-03) 892209 898578
10 Front Panel sticker, DA ; M-ETPR (rev.04) ; S/5 898578
10 Front Panel sticker, DE ; M-EST (rev.00) 880453
10 Front Panel sticker, DE ; M-EST (rev.01--) 881961
10 Front Panel sticker, DE ; M-ESTP (rev.01) 880552
10 Front Panel sticker, DE ; M-ESTP (rev.02--) 881955
10 Front Panel sticker, DE ; M-ESTPR (rev.04) ; S/5 898495
10 Front Panel sticker, DE ; M-ESTPR (rev.01-03) 886966 898495
10 Front Panel sticker, DE ; M-ESTR (rev.01-03) 886968 898556
10 Front Panel sticker, DE ; M-ESTR (rev.04) ; S/5 898556
10 Front Panel sticker, DE ; M-ETP (rev.00) 880560
10 Front Panel sticker, DE ; M-ETP (rev.01--) 881958
10 Front Panel sticker, DE ; M-ETPR (rev.01-03) 886967 898569
10 Front Panel sticker, DE ; M-ETPR (rev.04) ; S/5 898569
10 Front Panel sticker, EN ; M-EST (rev.00) 880138
10 Front Panel sticker, EN ; M-EST (rev.01--) 881959
10 Front Panel sticker, EN ; M-ESTP (rev.01) 879481
10 Front Panel sticker, EN ; M-ESTP (rev.02--) 881953
10 Front Panel sticker, EN ; M-ESTPR (rev.01-03) 886603 898494
10 Front Panel sticker, EN ; M-ESTPR (rev.04) ; S/5 898494
10 Front Panel sticker, EN ; M-ESTR (rev.01-03) 886605 898555
10 Front Panel sticker, EN ; M-ESTR (rev.04) ; S/5 898555
10 Front Panel sticker, EN ; M-ETP (rev.00) 880428
10 Front Panel sticker, EN ; M-ETP (rev.01--) 881956
10 Front Panel sticker, EN ; M-ETPR (rev.01-03) 886604 898568
10 Front Panel sticker, EN ; M-ETPR (rev.04) ; S/5 898568
10 Front Panel sticker, ES ; M-EST (rev.01--) 885044
10 Front Panel sticker, ES ; M-ESTP (rev.02--) 884200
10 Front Panel sticker, ES ; M-ESTPR (rev.01-03) 886943 898498
10 Front Panel sticker, ES ; M-ESTPR (rev.04) ; S/5 898498
10 Front Panel sticker, ES ; M-ESTR (rev.01-03) 886945 898559
10 Front Panel sticker, ES ; M-ESTR (rev.04) ; S/5 898559
10 Front Panel sticker, ES ; M-ETP (rev.01--) 885043
71
Document No. 8001008-6
Page 86
Datex-Ohmeda S/5 monitors
Item Description Order No. Replaced by
10 Front Panel sticker, ES ; M-ETPR (rev.01-03) 886944 898572
10 Front Panel sticker, ES ; M-ETPR (rev.04) ; S/5 898572
10 Front Panel sticker, FI ; M-ESTPR (rev.01-03) 888868 898501
10 Front Panel sticker, FI ; M-ESTPR (rev.04) ; S/5 898501
10 Front Panel sticker, FI ; M-ESTR (rev.01-03) 888870 898562
10 Front Panel sticker, FI ; M-ESTR (rev.04) ; S/5 898562
10 Front Panel sticker, FI ; M-ETPR (rev.01-03) 888869 898575
10 Front Panel sticker, FI ; M-ETPR (rev.04) ; S/5 898575
10 Front Panel sticker, FR ; M-EST (rev.00) 880140
10 Front Panel sticker, FR ; M-EST (rev.01--) 881960
10 Front Panel sticker, FR ; M-ESTP (rev.01) 880158
10 Front Panel sticker, FR : M-ESTP (rev.02--) 881954
10 Front Panel sticker, FR ; M-ESTPR (rev.01-03) 886963 898496
10 Front Panel sticker, FR ; M-ESTPR (rev.04) ; S/5 898496
10 Front Panel sticker, FR ; M-ESTR (rev.01-03) 886964 898557
10 Front Panel sticker, FR ; M-ESTR (rev.04) ; S/5 898557
10 Front Panel sticker, FR ; M-ETP (rev.00) 880429
10 Front Panel sticker, FR ; M-ETP (rev.01--) 881957
10 Front Panel sticker, FR ; M-ETPR (rev.01-03) 886965 898570
10 Front Panel sticker, FR ; M-ETPR (rev.04) ; S/5 898570
10 Front Panel sticker, IT ; M-EST (rev.01--) 886755
10 Front Panel sticker, IT ; M-ESTP (rev.02--) 886753
10 Front Panel sticker, IT ; M-ESTPR (rev.01-03) 886925 898499
10 Front Panel sticker, IT ; M-ESTPR (rev.04) ; S/5 898499
10 Front Panel sticker, IT ; M-ESTR (rev.01-03) 886927 898560
10 Front Panel sticker, IT ; M-ESTR (rev.04) ; S/5 898560
10 Front Panel sticker, IT ; M-ETP (rev.01--) 886754
10 Front Panel sticker, IT ; M-ETPR (rev.01-03) 886926 898573
10 Front Panel sticker, IT ; M-ETPR (rev.04) ; S/5 898573
10 Front Panel sticker, JA ; M-ESTPR (rev.01-03) 888303 898505
10 Front Panel sticker, JA ; M-ESTPR (rev.04) ; S/5 898505
10 Front Panel sticker, JA ; M-ESTR (rev.01-03) 888304 898566
10 Front Panel sticker, JA ; M-ESTR (rev.04) ; S/5 898566
10 Front Panel sticker, JA ; M-ETPR (rev.01-03) 888305 898579
10 Front Panel sticker, JA ; M-ETPR (rev.04) ; S/5 898579
10 Front Panel sticker, NL ; M-EST (rev.01--) 886045
10 Front Panel sticker, NL ; M-ESTP (rev.02--) 886044
10 Front Panel sticker, NL ; M-ESTPR (rev.01-03) 886937 898497
10 Front Panel sticker, NL ; M-ESTPR (rev.04) ; S/5 898497
10 Front Panel sticker, NL ; M-ESTR (rev.01-03) 886939 898558
10 Front Panel sticker, NL ; M-ESTR (rev.04) ; S/5 898558
10 Front Panel sticker, NL ; M-ETP (rev.01--) 886046
10 Front Panel sticker, NL ; M-ETPR (rev.01-03) 886938 898571
10 Front Panel sticker, NL ; M-ETPR (rev.04) ; S/5 898571
10 Front Panel sticker, NO ; M-ESTPR (rev.01-03) 893554 898503
10 Front Panel sticker, NO ; M-ESTPR (rev.04) ; S/5 898503
10 Front Panel sticker, NO ; M-ESTR (rev.01-03) 893556 898564
10 Front Panel sticker, NO ; M-ESTR (rev.04) ; S/5 898564
72
Document No. 8001008-6
Page 87
S/5 Hemodynamic modules
Item Description Order No. Replaced by
10 Front Panel sticker, NO ; M-ETPR (rev.01-03) 893555 898577
10 Front Panel sticker, NO ; M-ETPR (rev.04) ; S/5 898577
10 Front Panel sticker, PT ; M-ESTPR (rev.01-03) 895249 898500
10 Front Panel sticker, PT ; M-ESTPR (rev.04) ; S/5 898500
10 Front Panel sticker, PT ; M-ESTR (rev.01-03) 895250 898561
10 Front Panel sticker, PT ; M-ESTR (rev.04) ; S/5 898561
10 Front Panel sticker, PT ; M-ETPR (rev.01-03) 895251 898574
10 Front Panel sticker, PT ; M-ETPR (rev.04) ; S/5 898574
10 Front Panel sticker, SV ; M-EST (rev.01--) 885859
10 Front Panel sticker, SV ; M-ESTP (rev.02--) 885857
10 Front Panel sticker, SV ; M-ESTPR (rev.01-03) 886928 898502
10 Front Panel sticker, SV ; M-ESTPR (rev.04) ; S/5 898502
10 Front Panel sticker, SV ; M-ESTR (rev.01-03) 886930 898563
10 Front Panel sticker, SV ; M-ESTR (rev.04) ; S/5 898563
10 Front Panel sticker, SV ; M-ETP (rev.01--) 885856
10 Front Panel sticker, SV ; M-ETPR (rev.01-03) 886929 898576
10 Front Panel sticker, SV ; M-ETPR (rev.04) ; S/5 898576
11 Metal frame 879183
11 Metal frame 879184
13 Cross cylinder-head screw M3x6 61721
14 Cross cylinder-head scerw M3x12 628700
15 T-Input Board, M-ESTP (rev.02-03) 882090
15 T-Input connectors 887152
16 EMC cover 884099
17 EMC cover 886818
A front panel unit includes all the connectors and input boards.
NOTE: The STP board 882627 can be used as a replacement only with STP software 884342.
NOTE: The ECG board 888957 can be used as a replacement only with ECG software 883806.
NOTE: The STP board 891442 can be used as a replacement only with STP software 891401.
NOTE: The STP board 8000801 includes the STP software 8000717.
NOTE: The STP board 8002575 includes the STP software 8002573.
73
Document No. 8001008-6
Page 88
Datex-Ohmeda S/5 monitors
6.1.2 M-NESTPR rev. 00, M-NETPR rev. 00, M-NESTR rev. 00
Figure 18 Exploded view of M-NESTPR Module
Item Description Order No. Replaced by
1 Module box (wide) 886168
2 Spring pin 879182
3 Latch for module box 879181
4 ECG/RESP board, M-NESTPR (rev.00) 889988
5 STP board M-NESTPR (rev.00) 890007 8000800
5 STP board M-NESTPR (rev.00) 890007 8002574
5 STP-board 8000800 8002574
5 STP-board, M-NE12STPR 8002574
6 EMC cover 884099
7 NIPB board, M-NESTPR (rev.00) 887520
NIBP board software 888237
8 EMC cover 892305
9 NIPB pump, M-NESTPR, M-NIBP (rev.04) 889993
10 EMC cover 892307
11 Air filter (NIBP) 57142
12 Check valve 58542
13 Damping chamber, M-NESTPR, M-NIBP (rev.04) 888240
14 Bleed valve, M-NESTPR, M-NIBP (rev.04) 58566
15 Magnetic valve 58562
16 NIBP Cuff connector 64654
17 T-Input connectors 887152
74
Document No. 8001008-6
Page 89
S/5 Hemodynamic modules
Item Description Order No. Replaced by
18 ECG Input Board, M-NESTPR (rev.00) 889985
19 P Input Board, M-NESTPR (rev.00) 890834
19 SP Input Board, M-NESTPR (rev.00) 890006
19 SPO2 Input Board, M-NESTR (rev.00), M-MRIP (rev.00) 890833
20 Fitting plate 879510
23 Membrane keypad 888242
24 Front panel unit, M-NESTPR 888273
24 Front panel unit, M-NESTR 891748
24 Front panel unit, M-NETPR 891772
25 Metal frame 888230
26 Cross recess screw M3x8 black 616215
27 Cross cylinder-head screw M3x6 61721
27 Cross cylinder-head screw M3x6 61721
28 Cross cylinder-head scerw M3x12 628700
28 Cross cylinder-head scerw M3x12 628700
29 Front Panel sticker, DA ; M-NESTPR (rev.00) 892204 898429
29 Front Panel sticker, DA ; M-NESTPR (rev.01) ; S/5 898429
29 Front Panel sticker, DA ; M-NESTR (rev.00) 892205 898458
29 Front Panel sticker, DA ; M-NESTR (rev.01) ; S/5 898458
29 Front Panel sticker, DA ; M-NETPR (rev.00) 892206 898492
29 Front Panel sticker, DA ; M-NETPR (rev.01) ; S/5 898492
29 Front Panel sticker, DE ; M-NESTPR (rev.00) 891166 898420
29 Front Panel sticker, DE ; M-NESTPR (rev.01) ; S/5 898420
29 Front Panel sticker, DE ; M-NESTR (rev.00) 891671 898449
29 Front Panel sticker, DE ; M-NESTR (rev.01) ; S/5 898449
29 Front Panel sticker, DE ; M-NETPR (rev.00) 891678 898483
29 Front Panel sticker, DE ; M-NETPR (rev.01) ; S/5 898483
29 Front Panel sticker, EN ; M-NESTPR (rev.00) 889264 898419
29 Front Panel sticker, EN ; M-NESTPR (rev.01) ; S/5 898419
29 Front Panel sticker, EN ; M-NESTR (rev.00) 891456 898448
29 Front Panel sticker, EN ; M-NESTR (rev.01) ; S/5 898448
29 Front Panel sticker, EN ; M-NETPR (rev.00) 891457 898482
29 Front Panel sticker, EN ; M-NETPR (rev.01) ; S/5 898482
29 Front Panel sticker, ES ; M-NESTPR (rev.00) 891171 898423
29 Front Panel sticker, ES ; M-NESTPR (rev.01) ; S/5 898423
29 Front Panel sticker, ES ; M-NESTR (rev.00) 891674 898452
29 Front Panel sticker, ES ; M-NESTR (rev.01) ; S/5 898452
29 Front Panel sticker, ES ; M-NETPR (rev.00) 891681 898486
29 Front Panel sticker, ES ; M-NETPR (rev.01) ; S/5 898486
29 Front Panel sticker, FI ; M-NESTPR (rev.00) 891170 898426
29 Front Panel sticker, FI ; M-NESTPR (rev.01) ; S/5 898426
29 Front Panel sticker, FI ; M-NESTR (rev.00) 891676 898455
29 Front Panel sticker, FI ; M-NESTR (rev.01) ; S/5 898455
29 Front Panel sticker, FI ; M-NETPR (rev.00) 891683 898489
29 Front Panel sticker, FI ; M-NETPR (rev.01) ; S/5 898489
29 Front Panel sticker, FR ; M-NESTPR (rev.00) 891167 898421
29 Front Panel sticker, FR ; M-NESTPR (rev.01) ; S/5 898421
29 Front Panel sticker, FR ; M-NESTR (rev.00) 891672 898450
75
Document No. 8001008-6
Page 90
Datex-Ohmeda S/5 monitors
Item Description Order No. Replaced by
29 Front Panel sticker, FR ; M-NESTR (rev.01) ; S/5 898450
29 Front Panel sticker, FR ; M-NETPR (rev.00) 891679 898484
29 Front Panel sticker, FR ; M-NETPR (rev.01) ; S/5 898484
29 Front Panel sticker, IT ; M-NESTPR (rev.00) 891172 898424
29 Front Panel sticker, IT ; M-NESTPR (rev.01) ; S/5 898424
29 Front Panel sticker, IT ; M-NESTR (rev.00) 891675 898453
29 Front Panel sticker, IT ; M-NESTR (rev.01) ; S/5 898453
29 Front Panel sticker, IT ; M-NETPR (rev.00) 891682 898487
29 Front Panel sticker, IT ; M-NETPR (rev.01) ; S/5 898487
29 Front Panel sticker, JA ; M-NESTPR (rev.00) 894963 898430
29 Front Panel sticker, JA ; M-NESTPR (rev.01) ; S/5 898430
29 Front Panel sticker, JA ; M-NESTR (rev.00) 894964 898459
29 Front Panel sticker, JA ; M-NESTR (rev.01) ; S/5 898459
29 Front Panel sticker, JA ; M-NETPR (rev.00) 894965 8000380
29 Front Panel Sticker, JA ; M-NETPR (rev.01) ; S/5 8000380
29 Front Panel sticker, NL ; M-NESTPR (rev.00) 891168 898422
29 Front Panel sticker, NL ; M-NESTPR (rev.01) ; S/5 898422
29 Front Panel sticker, NL ; M-NESTR (rev.00) 891673 898451
29 Front Panel sticker, NL ; M-NESTR (rev.01) ; S/5 898451
29 Front Panel sticker, NL ; M-NETPR (rev.00) 891680 898485
29 Front Panel sticker, NL ; M-NETPR (rev.01) ; S/5 898485
29 Front Panel sticker, NO ; M-NESTPR (rev.00) 893566 898428
29 Front Panel sticker, NO ; M-NESTPR (rev.01) ; S/5 898428
29 Front Panel sticker, NO ; M-NESTR (rev.00) 893568 898457
29 Front Panel sticker, NO ; M-NESTR (rev.01) ; S/5 898457
29 Front Panel sticker, NO ; M-NETPR (rev.00) 893567 898491
29 Front Panel sticker, NO ; M-NETPR (rev.01) ; S/5 898491
29 Front Panel sticker, PT ; M-NESTPR (rev.00) 895234 898425
29 Front Panel sticker, PT ; M-NESTPR (rev.01) ; S/5 898425
29 Front Panel sticker, PT ; M-NESTR (rev.00) 895235 898454
29 Front Panel sticker, PT ; M-NESTR (rev.01) ; S/5 898454
29 Front Panel sticker, PT ; M-NETPR (rev.00) 895236 898488
29 Front Panel sticker, PT ; M-NETPR (rev.01) ; S/5 898488
29 Front Panel sticker, SV ; M-NESTPR (rev.00) 891169 898427
29 Front Panel sticker, SV ; M-NESTPR (rev.01) ; S/5 898427
29 Front Panel sticker, SV ; M-NESTR (rev.00) 891677 898456
29 Front Panel sticker, SV ; M-NESTR (rev.01) ; S/5 898456
29 Front Panel sticker, SV ; M-NETPR (rev.00) 891684 898427
29 Front Panel sticker, SV ; M-NETPR (rev.01) ; S/5 898490
30 Flat cable, STP 890874
30 Flat cable, STP 890874
31 Flat cable, ECG 890876
32 Flat cable, M-PP 891573
76
Document No. 8001008-6
A front panel unit includes all the connectors and input boards.
NOTE: The STP board 890007 can be used as a replacement only with STP software 891401.
NOTE: The STP board 8000800 includes the STP software 8000717.
Page 91
NOTE: The STP board 8002574 includes the STP software 8002573.
S/5 Hemodynamic modules
77
Document No. 8001008-6
Page 92
Datex-Ohmeda S/5 monitors
6.1.3 M-NE12STPR rev. 00, M-NE12STR rev. 00, M-NE12TPR rev. 00
Figure 19 Exploded view of M-NE12STPR Module
Item Description Order No. Replaced by
1 Module box (wide) 886168
2 Spring pin 879182
3 Latch for module box 879181
4 12-LEAD ECG Board, M-NE12STPR 894284 8000995
4 12-LEAD ECG BOARD II, N-NE12STPR 8000995
5 STP board M-NESTPR (rev.00) 890007 8000800
5 STP board M-NESTPR (rev.00) 890007 8002574
5 STP-board 8000800 8002574
5 STP-board, M-NE12STPR 8002574
6 EMC cover 884099
7 NIPB board, M-NESTPR (rev.00) 887520
NIBP board software 888237
8 EMC cover 892305
9 NIPB pump, M-NESTPR, M-NIBP (rev.04) 889993
10 EMC cover 892307
11 Air filter (NIBP) 57142
13 Damping chamber, M-NESTPR, M-NIBP (rev.04) 888240
14 Bleed valve, M-NESTPR, M-NIBP (rev.04) 58566
15 Magnetic valve 58562
16 NIBP Cuff connector 64654
78
Document No. 8001008-6
Page 93
S/5 Hemodynamic modules
Item Description Order No. Replaced by
17 T-Input connectors 887152
18 12-lead ECG Input Unit 896913
19 P Input Board, M-NESTPR (rev.00) 890834
19 SP Input Board, M-NESTPR (rev.00) 890006
19 SPO2 Input Board, M-NESTR (rev.00), M-MRIP (rev.00) 890833
20 Fitting plate 879510
23 Membrane keypad 888242
24 Front Mask Unit, M-NE12STPR 896910
24 Front Mask Unit, M-NE12STR 896911
24 Front Mask Unit, M-NE12TPR 896912
25 Metal frame 888230
26 Cross recess screw M3x8 black 616215
27 Cross cylinder-head screw M3x6 61721
27 Cross cylinder-head screw M3x6 61721
28 Cross cylinder-head scerw M3x12 628700
28 Cross cylinder-head scerw M3x12 628700
29 Front Panel sticker, DA ; M-NE12STPR (rev.00) 897718 898442
29 Front Panel sticker, DA ; M-NE12STPR (rev.01) ; S/5 898442
29 Front Panel sticker, DA ; M-NE12STR (rev.00) 897622 898470
29 Front Panel sticker, DA ; M-NE12STR (rev.01) ; S/5 898470
29 Front Panel sticker, DA ; M-NE12TPR (rev.00) 897655 898481
29 Front Panel sticker, DA ; M-NE12TPR (rev.01) ; S/5 898481
29 Front Panel sticker, DE ; M-NE12STPR (rev.00) 897709 898433
29 Front Panel sticker, DE ; M-NE12STPR (rev.01) ; S/5 898433
29 Front Panel sticker, DE ; M-NE12STR (rev.00) 897613 898461
29 Front Panel sticker, DE ; M-NE12STR (rev.01) ; S/5 898461
29 Front panel sticker, DE ; M-NE12TPR (rev.00) 897646 898472
29 Front Panel sticker, DE ; M-NE12TPR (rev.01) ; S/5 898472
29 Front Panel sticker, EN ; M-NE12STPR (rev.00) 896985 898432
29 Front Panel sticker, EN ; M-NE12STPR (rev.01) ; S/5 898432
29 Front Panel sticker, EN ; M-NE12STR (rev.00) 896986 898460
29 Front Panel sticker, EN ; M-NE12STR (rev.01) ; S/5 898460
29 Front Panel sticker, EN ; M-NE12TPR (rev.00) 896987 898471
29 Front Panel sticker, EN ; M-NE12TPR (rev.01) ; S/5 898471
29 Front Panel sticker, ES ; M-NE12STPR (rev.00) 897712 898436
29 Front Panel sticker, ES ; M-NE12STPR (rev.01) ; S/5 898436
29 Front Panel sticker, ES ; M-NE12STR (rev.00) 897616 898464
29 Front Panel sticker, ES ; M-NE12STR (rev.01) ; S/5 898464
29 Front Panel sticker, ES ; M-NE12TPR (rev.00) 897649 898475
29 Front Panel sticker, ES ; M-NE12TPR (rev.01) ; S/5 898475
29 Front Panel sticker, FI ; M-NE12STPR (rev.00) 897715 898439
29 Front Panel sticker, FI ; M-NE12STPR (rev.01) ; S/5 898439
29 Front Panel sticker, FI ; M-NE12STR (rev.00) 897619 898467
29 Front Panel sticker, FI ; M-NE12STR (rev.01) ; S/5 898467
29 Front Panel sticker, FI ; M-NE12TPR (rev.00) 897652 898478
29 Front Panel sticker, FI ; M-NE12TPR (rev.01) ; S/5 898478
29 Front Panel sticker, FR ; M-NE12STPR (rev.00) 897710 898434
29 Front Panel sticker, FR ; M-NE12STPR (rev.01) ; S/5 898434
79
Document No. 8001008-6
Page 94
Datex-Ohmeda S/5 monitors
Item Description Order No. Replaced by
29 Front Panel sticker, FR ; M-NE12STR (rev.00) 897614 898462
29 Front Panel sticker, FR ; M-NE12STR (rev.01) ; S/5 898462
29 Front Panel sticker, FR ; M-NE12TPR (rev.00) 897647 898473
29 Front Panel sticker, FR ; M-NE12TPR (rev.01) ; S/5 898473
29 Front Panel sticker, IT ; M-NE12STPR (rev.00) 897713 898437
29 Front Panel sticker, IT ; M-NE12STPR (rev.01) ; S/5 898437
29 Front Panel sticker, IT ; M-NE12STR (rev.00) 897617 898465
29 Front Panel sticker, IT ; M-NE12STR (rev.01) ; S/5 898465
29 Front Panel sticker, IT ; M-NE12TPR (rev.00) 897650 898476
29 Front Panel sticker, IT ; M-NE12TPR (rev.01) ; S/5 898476
29 Front Panel sticker, JA ; M-NE12STPR (rev.00) 897719 8000377
29 Front Panel sticker, JA ; M-NE12STPR (rev.01) ; S/5 8000377
29 Front Panel sticker, JA ; M-NE12STR (rev.00) 897623 8000378
29 Front Panel sticker, JA ; M-NE12STR (rev.01) ; S/5 8000378
29 Front Panel sticker, JA ; M-NE12TPR (rev.00) 897656 8000379
29 Front Panel sticker, JA ; M-NE12TPR (rev.01) ; S/5 8000379
29 Front Panel sticker, NL ; M-NE12STPR (rev.00) 897711 898435
29 Front Panel sticker, NL ; M-NE12STPR (rev.01) ; S/5 898435
29 Front Panel sticker, NL ; M-NE12STR (rev.00) 897615 898463
29 Front Panel sticker, NL ; M-NE12STR (rev.01) ; S/5 898463
29 Front Panel sticker, NL ; M-NE12TPR (rev.00) 897648 898474
29 Front Panel sticker, NL ; M-NE12TPR (rev.01) ; S/5 898474
29 Front Panel sticker, NO ; M-NE12STPR (rev.00) 897717 898441
29 Front Panel sticker, NO ; M-NE12STPR (rev.01) ; S/5 898441
29 Front Panel sticker, NO ; M-NE12STR (rev.00) 897621 898469
29 Front Panel sticker, NO ; M-NE12STR (rev.01) ; S/5 898469
29 Front Panel sticker, NO ; M-NE12TPR (rev.00) 897654 898480
29 Front Panel sticker, NO ; M-NE12TPR (rev.01) ; S/5 898480
29 Front Panel sticker, PT ; M-NE12STPR (rev.00) 897714 898438
29 Front Panel sticker, PT ; M-NE12STPR (rev.01) ; S/5 898438
29 Front Panel sticker, PT ; M-NE12STR (rev.00) 897618 898466
29 Front Panel sticker, PT ; M-NE12STR (rev.01) ; S/5 898466
29 Front Panel sticker, PT ; M-NE12TPR (rev.00) 897651 898477
29 Front Panel sticker, PT ; M-NE12TPR (rev.01) ; S/5 898477
29 Front Panel sticker, SV ; M-NE12STPR (rev.00) 897716 898440
29 Front Panel sticker, SV ; M-NE12STPR (rev.01) ; S/5 898440
29 Front Panel sticker, SV ; M-NE12STR (rev.00) 897620 898468
29 Front Panel sticker, SV ; M-NE12STR (rev.01) ; S/5 898468
29 Front Panel sticker, SV ; M-NE12TPR (rev.00) 897653 898479
29 Front Panel sticker, SV ; M-NE12TPR (rev.01) ; S/5 898479
30 Flat cable, STP 890874
30 Flat cable, STP 890874
32 Flat cable, M-PP 891573
33 Cross cylinder head screw M3x4 61719
80
Document No. 8001008-6
Page 95
S/5 Hemodynamic modules
7 EARLIER REVISIONS
For service information on the earlier revisions, please refer to:
Module revision Manual Note!
ESTP Module revision 01 Service Manual p/n 880850
ETP Module revision 00 Service Manual p/n 880850
ESTP Module revision 02 Service Manual p/n 882580
ETP and EST Modules revision 01 Service Manual p/n 882580
ESTP Module revision 03-04, and ETP and EST Modules revision 02-03
NE12STPR Module, M-NE12STPR (rev. 00)
NE12STR Module, M-NE12STR (rev. 00)
NE12TPR Module, M-NE12TPR (rev. 00)
NESTPR Module, M-NESTPR (rev. 00)
NESTR Module, M-NESTR (rev. 00)
NETPR Module, M-NETPR (rev. 00)
ESTPR Module, M-ESTPR (rev. 03)
ESTR Module, M-ESTR (rev. 03)
ETPR Module, M-ETPR (rev. 03)
Technical Reference Manual Slot 896620-1. Main manual is 896 624.
81
Document No. 8001008-6
Page 96
APPENDIX A, Service check form, S/5 NE12STPR, NESTPR, ESTPR modules
Document No. 88001008-6
82
Page 97
APPENDIX A, Service Check Form, S/5 NE12STPR, NESTPR, ESTPR modules
APPENDIX A
83
Document No. 8001008-6
Page 98
Datex-Ohmeda S/5 monitors
This page intentionally left blank.
84
Document No. 88001008-6
Page 99
APPENDIX A, Service check form, S/5 NE12STPR, NESTPR, ESTPR modules
SERVICE CHECK FORM
DATEX-OHMEDA HEMODYNAMIC MODULES
Customer
Service
Service engineer Date
OK = Test OK N.A. = Test not applicable Fail = Test Failed
All modules
1. Internal parts 2. External parts
3. NIBP pump filter 4. Installation
5. Recognition
Notes
ECG measurement
OK N.A. Fail OK N.A. Fail
Module type S/N
S/N
6. Module software (serial numbers)
ECG/RESP
STP
NIBP
OK N.A. Fail OK N.A. Fail
7. Communication and
memories
9. Power frequency 10. Cable recognition
11. Lead detection 12. Test with patient
Notes
8. Membrane key
simulator
A-1(4)
Document No. 8001008-6
Page 100
Datex-Ohmeda S/5 monitors
RESP measurement
OK N.A. Fail OK N.A. Fail
13. RESP measurement
recognition
14. Test with patient
simulator
Notes
TEMP measurement
OK N.A. Fail OK N.A. Fail
15. Communication and
memories
16. Temperature probe
detection
17. Calibration check 18. Temp test -function
19. Configuration
STP/ST/TP
Notes
InvBP measurement
S/N
S/N
S/N
OK N.A. Fail OK N.A. Fail
20. Membrane keys 21. Cable and transducer
detection
22. Calibration 23. Test with patient
simulator
Notes
SpO2 measurement
OK N.A. Fail OK N.A. Fail
24. SpO2 probe detection 25. Test measurement
Notes
S/N
A 2 (4)
Document No. 8001008-6
Loading...