Datex-Ohmeda S-5 Light Monitor User manual

Datex-Ohmeda
S/5 Light Monitor
User’s Reference Manual
0537
Conformity according to the Council Directive 93/42/EEC concerning medical devices
CAUTION: U.S. Federal and Canadian law restricts this device to sale by or on the order of a
licensed medical practitioner. Outside the USA and Canada, check local laws for any
Datex-Ohmeda, Inc.
P.O. Box 7550, Madison WI 53707-7550, USA Tel. +1-608-221-1551
Fax +1-608-222-9147
www.us.datex-ohmeda.com
restriction that may apply.
April 2003
Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 39411 Fax +358 9 146 3310
http://www.datex-ohmeda.com/
Instrumentarium Corp. All rights reserved.
Intended purpose
Datex-Ohmeda S/5 Light Monitor is intended for stationary and mobile monitoring of patient’s ECG, SpO2, non-invasive and invasive blood pressure, temperature, respiration and CO2 in hospitals and ambulances. When used outdoors, the monitor must be in the protective case. The monitor is intended for use by qualified medical personnel only. The S/5 Light Monitor is intended for patients with weight from 5 kg (11 lb.) up. Respiration monitoring is intended for patients three years old and up.
Classification according to IEC 60601-1
CLASS I EQUIPMENT and INTERNALLY POWERED EQUIPMENT according to the type of protection against electrical shock.
TYPE BF or CF equipment according to the degree of protection against electric shock is indicated with a symbol beside each patient connector.
Degree of protection against harmful ingress of water as detailed in the IEC 60529: Monitor: IPX1, vertically falling water drops shall have no harmful effects (applicable when the monitor is in upright position, or tilted backwards). In the protective case IPX4, splash proof, when the case is closed properly, the monitor is intact and operates on battery power. Power adapter: IPX0, ordinary equipment. Power Adapter for Transport Vehicles: IPX1.
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC MIXTURE with air or with oxygen or nitrous oxide.
CONTINUOUS OPERATION according to the mode of operation.
Classification according to EU Medical Device Directive
The monitor is classified as IIb.
Responsibility of the manufacturer
Instrumentarium Corp. is responsible for the effects on safety, reliability and performance of the equipment only if:
assembly, operations, extensions, readjustments, modifications, service and repairs
are carried out by authorized personnel.
the electrical installation of the relevant room complies with appropriate
requirements.
the equipment is used in accordance with this manual.
The CO2 measuring option (N-LCM) is manufactured for Instrumentarium Corp. by Pryon Corp., 8500 S.W. Creekside Place, Beaverton, OR 97008-7107 USA.
The Nellcor option (N-LNSAT) is manufactured for Instrumentarium Corp. by Mallinckrodt Inc., 675 McDonnell Boulevard, P.O. Box 5840, St. Louis, MO 63134 USA.
Trademarks
ComWheel, Datex-Ohmeda AS/3, Datex-Ohmeda CS/3, Light, EarSat, FingerSat and Flexlite are trademarks of Instrumentarium Corp.
© Instrumentarium Corp. All rights reserved.
General
Table of contents
1. How to Use THIS MANUAL 1
Related documents...............................................................................................1
Symbols and abbreviations.....................................................................................2
Symbols on the equipment................................................................................2
Symbols used in this manual ..............................................................................4
Chapter symbols..............................................................................................4
Abbreviations..................................................................................................5
Conventions used in this manual.............................................................................6
2. SAFETY 7
Warnings .............................................................................................................7
EXPLOSION HAZARD ..........................................................................................7
ELECTRICAL SHOCK HAZARD...............................................................................7
PATIENT SAFETY................................................................................................8
Cautions..............................................................................................................9
BATTERY CAUTIONS.........................................................................................10
3. Unpacking and Installation 11
Unpacking..........................................................................................................11
Placing..............................................................................................................11
Electrical Power Adapter .......................................................................................12
Charge batteries............................................................................................13
Connecting to network.........................................................................................13
Configuration.....................................................................................................13
Disposal............................................................................................................14
4. Description 15
Parts of the monitor .............................................................................................15
Rear panel .........................................................................................................16
Keys and controls................................................................................................17
ComWheel....................................................................................................17
Display..............................................................................................................18
Measurement connectors .....................................................................................19
5. Operation 20
Getting started....................................................................................................20
During monitoring ................................................................................................20
In-between patients.............................................................................................20
At the end of the day...........................................................................................20
Using menus......................................................................................................21
i
Datex-Ohmeda S/5 Light Monitor
6. Batteries 22
Charging batteries...............................................................................................22
Charging in the monitor...................................................................................22
Removing batteries from the Battery Module...........................................................23
Battery alarms .................................................................................................... 23
Storing and disposing of batteries..........................................................................23
7. Cleaning 25
Dust filter..........................................................................................................26
Monitor casing....................................................................................................26
Pulse oximetry sensors........................................................................................26
ECG cables, temperature probes, NIBP hose...........................................................26
NIBP cuff...........................................................................................................27
Invasive pressure cables ...................................................................................... 27
Airway accessories .............................................................................................. 27
Adapters.......................................................................................................27
Mainstream CO2 sensor .................................................................................. 27
8. ALARMS 29
Alarm categories ................................................................................................. 29
Smart start alarms .......................................................................................... 30
Silencing alarms.................................................................................................30
Silencing alarms for two minutes.......................................................................31
Turning audible alarms on and off.....................................................................31
Alarm messages.................................................................................................32
Alarms setup and adjustments..............................................................................32
Adjusting alarm limits ...................................................................................... 32
Other settings................................................................................................33
Automatic alarm recording .................................................................................... 33
Alarm tables....................................................................................................... 34
Alarm sources................................................................................................ 34
Default limits................................................................................................. 34
Auto alarm limits ............................................................................................ 36
9. MONITOR SETUP 38
Screen setup ..................................................................................................... 38
Selecting waveforms....................................................................................... 38
Selecting numerical display fields ..................................................................... 39
Setting time & date ............................................................................................. 40
Install/Service menu........................................................................................... 41
Default settings .................................................................................................. 41
ii
General
10. TRENDS 45
Viewing graphic trends ......................................................................................... 46
Graphic trend pages........................................................................................ 46
Viewing numerical trend ....................................................................................... 48
Changing trend setup........................................................................................... 49
Erasing trend data............................................................................................... 49
11. Patient Data 50
12. Recording and Printing 52
Recording .......................................................................................................... 52
Selecting the waveforms to be recorded............................................................. 52
Recording trends................................................................................................. 53
Other recorder settings......................................................................................... 53
Recorder paper replacement................................................................................. 54
Printing.............................................................................................................. 54
Selecting the printer ....................................................................................... 54
Selecting the paper size .................................................................................. 55
Printer installation............................................................................................... 55
Printer setup.................................................................................................. 55
13. Network and DataCard 56
Networking......................................................................................................... 56
Patient view and multi view on the Datex-Ohmeda Central screen ........................... 56
Monitor-to-Monitor communication.................................................................... 57
DataCard........................................................................................................... 57
Inserting and removing the card ........................................................................ 57
Collecting data............................................................................................... 58
14. ECG 59
Monitoring ECG...................................................................................................59
Applying electrodes.............................................................................................59
Patient preparation.........................................................................................59
Placing electrodes..........................................................................................60
Electrode coding............................................................................................60
Selecting trunk cable......................................................................................61
What should happen when electrodes are applied ?.............................................61
Selecting ECG lead.........................................................................................61
ECG display........................................................................................................61
Cascaded ECG...............................................................................................62
ECG adjustments................................................................................................62
ECG setup..........................................................................................................63
Setting heart rate alarm.......................................................................................64
iii
Datex-Ohmeda S/5 Light Monitor
Pacemaker patients............................................................................................64
15. Respiration 65
Monitoring respiration..........................................................................................65
Patient connections............................................................................................65
Respiration setup...............................................................................................65
Limitations of measuring......................................................................................66
Measuring principle ............................................................................................. 67
16. Pulse Oximetry 68
Monitoring pulse oximetry.....................................................................................68
Selecting sensor ................................................................................................. 69
Applying sensors.................................................................................................69
What should happen when sensor is applied ? .................................................... 70
Pulse oximetry display.....................................................................................70
Measuring limitations..........................................................................................70
Adjustments.......................................................................................................71
Setting SpO2 alarms ...........................................................................................72
Measuring principle ............................................................................................. 72
Alternative SpO2 measurement.............................................................................73
Nellcor sensors...........................................................................................73
17. TEMPERATURE 74
Monitoring temperature ....................................................................................... 74
Temperature display ....................................................................................... 74
Temperature testing ............................................................................................ 74
Temperature setup.............................................................................................. 75
18. NON-INVASIVE BLOOD PRESSURE (NIBP) 76
Monitoring NIBP.................................................................................................. 76
Display of NIBP measurement .......................................................................... 77
Automatic NIBP double check........................................................................... 78
Using NIBP functions ........................................................................................... 78
Starting/Stopping autocycle ............................................................................ 79
Starting a manual measurement....................................................................... 79
Canceling any measurement............................................................................ 79
Starting a continuous measurement (STAT) ......................................................... 80
Using venous stasis............................................................................................. 80
NIBP adjustments ............................................................................................... 81
Calibration check................................................................................................ 81
NIBP alarms ....................................................................................................... 82
Measuring principle............................................................................................. 82
iv
General
19. INVASIVE BLOOD PRESSURE 83
Monitoring invasive pressures ............................................................................... 83
Preparing pressure line.................................................................................... 83
Zeroing transducer.......................................................................................... 84
Invasive pressure display................................................................................. 85
Adjustments....................................................................................................... 86
Labeling the invasive line................................................................................. 86
Changing the scale ......................................................................................... 87
Other invasive pressure setup adjustments......................................................... 87
Invasive pressure alarms ...................................................................................... 88
Measuring principle ............................................................................................. 88
20. Carbon Dioxide, CO
Monitoring mainstream CO2.................................................................................. 89
Display of CO2 measurement................................................................................ 90
CO2 adjustments ................................................................................................ 91
Calibration check ................................................................................................ 91
Unit conversions ................................................................................................. 92
Measuring principle ........................................................................................ 92
2
89
21. MAINTENANCE AND TROUBLESHOOTING 93
Maintenance......................................................................................................93
Once a month................................................................................................93
Once a year ...................................................................................................93
Checklist...........................................................................................................93
General........................................................................................................93
ECG/Respiration............................................................................................94
Pulse oximetry...............................................................................................94
Temperature..................................................................................................95
Non-invasive blood pressure NIBP .....................................................................95
Invasive blood pressure...................................................................................96
CO2.............................................................................................................96
Battery packs and Module...............................................................................96
Alarm and information messages...........................................................................97
ECG.............................................................................................................97
Impedance respiration....................................................................................98
Invasive blood pressure...................................................................................98
NIBP.............................................................................................................99
Temperature................................................................................................101
Pulse oximetry.............................................................................................101
CO2...........................................................................................................102
DataCard....................................................................................................103
v
Datex-Ohmeda S/5 Light Monitor
Battery ....................................................................................................... 104
Recorder....................................................................................................104
Other messages ........................................................................................... 105
Troubleshooting................................................................................................106
22. GENERAL SPECIFICATIONS 107
Parameter specifications....................................................................................107
ECG ........................................................................................................... 107
Impedance respiration..................................................................................107
Pulse oximetry (SpO2)/Pleth .......................................................................... 108
Temperature................................................................................................109
Non-invasive blood pressure .......................................................................... 109
Invasive blood pressure (F-LMP1 models only)..................................................110
General...........................................................................................................111
Environmental conditions..............................................................................112
Display.......................................................................................................112
Trends........................................................................................................ 113
Alarms........................................................................................................ 113
Light Monitor Battery Module (optional) (w/2 battery slots) ................................. 113
S/5 Light Monitor Backup Battery (optional)..................................................... 113
External Power Adapter ................................................................................. 114
Power Adapter for transport vehicles (optional) ................................................. 114
Option codes.................................................................................................... 114
Nurse call and defibrillator synchronization signals ............................................ 115
23. Accessories 117
24. INDEX i
vi
1. HOW TO USE THIS MANUAL
This manual describes how to use and maintain the Datex-Ohmeda S/5 Light Monitors F-LM1 and F-LMP1.
The General section gives you information about safety aspects and installation. Get acquainted with them before using the monitor.
The Basic Operation section helps you through the daily basic use of the monitor.
The Detailed Operation section describes the monitor functions in detail. These chapters are necessary if you want to master all the features available in the monitor.
The Troubleshooting/Specifications and Index sections help you find the information you need in various situations.
NOTE: Before using your monitor, please read this manual. Pay special attention to WARNING and CAUTION statements.
General
Related documents
S/5 Light Monitor Technical Reference Manual
The Technical Reference Manual provides detailed technical descriptions and instructions for the service, maintenance procedures and troubleshooting.
Light Monitor Schematic Diagrams
The Schematic Diagrams provide schematic diagrams and part layouts for the monitor.
Patient Monitor Supplies and Accessories
The Patient Monitor Supplies and Accessories Catalogue provides information on all supplies and accessories available for the monitor.
1
Datex-Ohmeda S/5 Light Monitor
Symbols and abbreviations
Symbols on the equipment
Attention, consult accompanying documents.
on the rear panel indicates the warning:
on the rear panel dust filter indicates the reminder:
on the patient connector panel indicates the warnings and cautions:
on the external power adapter indicates the cautions:
Electric shock hazard. Do not open monitor frame. Refer servicing to qualified personnel. Do not touch any part of the monitor or patient connections during defibrillation procedure. Disconnect power supply before servicing.
Check rear panel dust filter regularly, and clean at least once a month.
Use only the specified cables and accessories. Ensure proper contact of the return electrode of the electrosurgery
unit to your patient to avoid possible burns at ECG electrode or other probe sites.
The output signals are not floating and must not be connected directly to a patient.
For continued protection against fire hazard, replace only with same type and rating of fuse.
Make sure voltage selector is set according to local voltage. For indoor use only. For patient transport outdoors, use the monitor
in its protective case and with battery power only. Do not immerse in liquid or allow any liquid to enter the interior.
on the battery unit indicates the warnings and cautions:
on the CO2 sensor: Do not immerse the sensor in liquids or autoclave it.
Do not put in fire. Do not short-circuit any terminals. Do not disassemble or mutilate. Use only with Battery Module or External Battery Charger. Discharge and charge batteries fully once a month to restore full capacity.
Sealed NiCd battery must be recycled or disposed of properly. Discharge before disposing.
2
Ni-Cd
d
Ni-C
General
SpO2, CO2: Type BF (IEC-60601-1) protection against electrical shock, not defibrillation proof.
NIBP: Type BF (IEC-60601-1) defibrillator-proof protection against electrical shock.
Temperature: Type CF (IEC-60601-1) protection against electrical shock.
ECG, Impedance Respiration, P1, P2: Type CF (IEC-60601-1) defibrillator-proof protection against electrical shock.
Battery packs contain heavy metal cadmium (chemical symbol Cd) and in case of disposal, must be separated from other waste according to local regulations.
Battery packs contain Ni-Cd and they can be recycled.
IPX-class:
IPX0 IPX1 IPX4
Degree of protection against harmful ingress of water as detailed in the IEC 60529:
- ordinary equipment
- vertically falling water drops shall have no harmful effects
- splash proof Battery charging LEDs: LEDs are blinking while batteries are being
charged and stay illuminated when batteries are fully charged and the monitor is connected to mains.
The monitor is connected to the network (optional). The DataCard (option) is inserted. If symbol is blinking, the card is
faulty or it is not inserted properly. A blinking heart next to the heart rate or pulse rate value indicates
the beats detected. A lung next to the respiration rate value indicates that respiration
rate is calculated from the impedance respiration measurement. Silence alarm indicator. When displayed on the upper left corner of
the screen, indicates that all the alarms are silenced. When in the menu or digit fields, indicates that the alarm source has been turned off.
Alternating current
3
Datex-Ohmeda S/5 Light Monitor
F
Input voltage
Output voltage
Ethernet connectors
Printer connector
Sample gas outlet
Do not immerse the sensor in liquids.
Symbols used in this manual
Indicates the procedure for making adjustments (turn and press) in the menus.
Submenu. Selecting a menu item with this symbol opens a new menu.
Chapter symbols
General chapters Measuring parameter chapters
General information ECG
Operation NIBP
Alarms Pulse oximetry
Trends
4
°C
Temperature
°
General
Recorder Invasive pressure
Abbreviations
Monitor setup CO
2
Networking and DataCard Impedance respiration
Patient data
Blood pressure:
Art arterial pressure CVP central venous pressure ICP intracranial pressure Inv. invasive Mean mean of diastolic and systolic pressures NIBP non-invasive blood pressure P blood pressure PA pulmonary arterial pressure S/D systolic/diastolic
Temperature:
Temp temperature AirW airway temperature Axil axillary temperature Blad bladder temperature Eso esophageal temperature Myo myocardial temperature Naso nasopharyngeal temperature Rect rectal temperature Room room temperature Skin skin temperature Temp temperature Tymp tympanic temperature Core central temperature Surf surface temperature
5
Datex-Ohmeda S/5 Light Monitor
Miscellaneous:
Casc cascaded ECG waveform ECG electrocardiography EL electroluminescence (display) HR heart rate LED light emitting diode Pleth plethysmograph waveform P/N part number PR pulse rate SpO
2
MS mainstream CO2 measurement SN, S/N serial number
oxygen saturation measured by pulse oximeter
Conventions used in this manual
Names of the keys appear in bold typeface: Normal Screen. Selectable menu items appear in bold and italic typeface: Previous Menu. Informative messages displayed on the screen are enclosed in single quotes:
‘Please wait’.
6
Safety
2. SAFETY
Warnings
WARNING Indicates situations in which the user or the patient may be in
danger of injury or death.
EXPLOSION HAZARD
Do not use the monitor in presence of flammable anesthetics.
ELECTRICAL SHOCK HAZARD
Do not remove cover. Refer servicing to qualified service personnel.
Use only hospital-grade grounded electrical outlets and power cord.
Grounding reliability can only be achieved when the equipment is connected to a receptacle marked ‘hospital grade’.
Make sure that external equipment is hospital grade grounded before you
connect it to the monitor. Do not connect any external equipment to the monitor, except that specified by Datex-Ohmeda.
When connecting external equipment to monitor, make sure that the
whole combination complies with safety standard IEC 60601-1-1 for the medical electrical systems and with the requirements of local authorities. A non-medical electrical equipment connected to monitor shall be supplied from an extra separating transformer with a protectively grounded power outlet.
Before cleaning, disconnect the monitor from electrical outlet.
Do not immerse any part of the monitor in liquids or allow liquid to enter
the interior.
Do not autoclave any part of the monitor with steam or ethylene oxide.
After cleaning, or if liquid has accidentally entered the interior of the
monitor, make sure that every part of the monitor is dry before reconnecting it to the power supply.
7
Datex-Ohmeda S/5 Light Monitor
PATIENT SAFETY
Always make sure that necessary alarm limits are set and operative when
you start monitoring.
When the alarms are suppressed, observe the patient frequently.
Connect only one patient to one monitor at a time.
Constant attention by a qualified professional is needed whenever a
patient is under anesthesia or connected to a ventilator. Some equipment malfunctions may pass unnoticed in spite of the monitor alarm.
Do not use antistatic or electrically conductive breathing tubes. They may
increase the risk of burns when electrosurgery unit is utilized.
Do not use the monitor during magnetic resonance imaging (MRI).
ECG, Impedance Respiration, InvBP: Ensure proper contact of the return
electrode of the electrosurgery unit to your patient to avoid possible burns at sensor sites.
ECG, Impedance Respiration, InvBP: Make sure that no part of the patient
connections touches any electrically conductive material including earth.
Impedance Respiration: In obstructive apnea, thoracic movements and
impedance variations may continue.
NIBP: The monitor sets the inflation pressure automatically according to
the first measurement. Reset the case to reset the inflation limit before measuring a new patient.
PATIENTS WITH PACEMAKERS OR ARRHYTHMIAS: Monitor may count the
pacemaker pulses as heart beats during cardiac arrest, some arrhythmias, and with certain types of pacemakers particularly in ON R mode. Do not rely entirely upon rate meter alarms. Keep patients with pacemakers and arrhythmias under close surveillance.
PACEMAKER PATIENTS: The impedance respiration measurement may
cause rate changes in Minute Ventilation Rate Responsive Pacemakers. Set the pacemaker rate responsive mode off or turn off the impedance respiration measurement on the monitor.
SpO
: A damaged sensor or a sensor soaked in liquid may cause burns
2
during electrosurgery.
8
Safety
SpO
: Change measuring site frequently. Change the sensor site and
2
check skin and circulatory status every 2-4 hours with adults and every hour with small children.
The output signals are not floating and must not be connected directly to
a patient.
Do not use delayed analog signals for defibrillator or intra-aortic balloon
pump synchronization.
The monitor is intended for use by qualified medical personnel only.
InvBP: All invasive procedures involve risks to the patient. Use aseptic
technique. Follow catheter manufacturer’s instructions.
InvBP: Use only defibrillator proof cables and transducers.
Cautions
CAUTION Indicates situations in which the unit or devices connected to it may
be damaged.
Use licensed patient cables and accessories only, see Patient Monitor
supplies and accessories catalogue. Other cables and accessories may damage the system or interfere with measurement.
Before connecting the power cord, ensure that the input voltage selector
is set correctly in the external power adapter.
Use the monitor outdoors with the battery power only.
Do not store the monitor beyond the specified temperature range.
Do not use hypochlorite, ammonia-, phenol- or acetone-based cleaners.
They may damage the monitor.
Clean rear panel fan dust filter once a month or whenever needed.
Leave space around the monitor for ventilation to prevent the monitor
from overheating.
Do not apply pressurized air to any outlet or tubing connected to the
monitor. Pressure may destroy sensitive elements.
When used in moving vehicle, mount the monitor properly.
Dispose of the device and parts thereof in accordance with local
regulations.
9
Datex-Ohmeda S/5 Light Monitor
Vibrations during transport may disturb SpO
CO
DataCards: Do not subject cards to excessive heat, bending, or magnetic
InvBP: Mechanical shock to invasive blood pressure transducer may
Electromagnetic interference from radio-frequency transmitting devices in
BATTERY CAUTIONS
Do not put in fire.
Do not short-circuit any terminals.
Do not use damaged or leaking batteries.
Do not disassemble or break the batteries.
Use only with Battery Module or External Battery Charger.
, ECG, impedance
2
respiration and NIBP measurements.
, SpO2: Do not apply force to sensors or sensor cables.
2
fields.
cause severe shifts in zero balance and calibration and produce erroneous readings.
ambulances may affect the performance of this medical device.
Charge and discharge batteries fully once a month to restore full
capacity.
Sealed NiCd battery must be recycled or disposed of properly. Discharge
before disposing.
10
3. UNPACKING AND INSTALLATION
Unpacking
The monitor is delivered in one box containing typically (as minimum):
Datex-Ohmeda S/5 Light Monitor
Datex-Ohmeda S/5 Light Monitor Power Adapter with power cord or
optional Power Adapter for Transport Vehicles
User’s Reference Manual
Check that the contents correspond to your order and that all additional options you have ordered are included. If any of the items appears damaged or is missing, contact your sales person immediately.
Placing
The monitor can be placed on a flat surface or hung with the handle from bed or wall rails. Make sure that the surface or rail holds up to at least 10 kg/22 lb.
Installation
Figure 3-1 Installation
A protective case is also available. If using the case, make sure that the monitor is always in a standing position to ensure proper ventilation. The monitor is splash-proof (IPX4) when it is in the protective case and all the openings of the case are closed properly. Make sure the monitor is intact and operates on battery power. When used outdoors, the monitor must be in the protective case.
NOTE: The pocket in the protective case is not water-proof. Do not place items in the pocket that may be damaged by moisture.
Universal mounting solutions are sold separately. Contact your local distributor.
11
Datex-Ohmeda S/5 Light Monitor
Electrical Power Adapter
Use only the power cord and power adapter provided with the monitor.
Set the voltage selector on the adapter to the right position (115V/230V)
e.g. with a pen.
Connect the adapter cord to the connector on the rear panel.
Connect the power cord to the adapter and then plug onto the wall
electrical outlet.
NOTE: To use the power adapter, it must be lying on a flat surface or hanging from the holder. Do not let any liquid enter the adapter.
When external power supply is in use, a green led is lit.
Figure 3-2 Placing Power Adapter
WARNING ELECTRICAL SHOCK HAZARD: Use only hospital grade grounded
electrical outlets.
CAUTION Before connecting the power cord, ensure that the
input voltage selector is set correctly in the power adapter.
In vehicles with 12 - 32 V power source, use the Power Adapter for Transport Vehicles, N-LPOWT. For information on how to install and use this adapter, refer to the instruction sheet provided with it, or contact your local distributor.
12
Installation
CAUTION Electromagnetic interference from radio-frequency transmitting
devices in ambulances may affect the performance of this medical device.
Charge batteries
Keep the monitor connected to the mains outlet. The batteries are fully charged when the battery capacity indicators on the screen show FULL and green LEDs in the Battery Module stop flashing.
CAUTION When stored for longer periods, or used mostly with mains power,
charge and discharge the batteries fully once a month.
Connecting to network
The monitor can be connected to a monitor network for central monitoring and printing services.
Connect the Ethernet and location ID plug to the corresponding
connectors on the monitor’s rear panel.
Configure the location IDs in the Datex-Ohmeda Central.
When the monitor is connected to the network and turned on, the network symbol is displayed on the upper right corner of the screen:
Configuration
When the monitor is delivered, it is configured with the factory default settings and is ready to use.
If you want to change the settings and make your own permanent configuration, refer to chapters “Alarms” and “Monitor setup” of this manual.
Disposal
Dispose of the monitor, or parts of it, in accordance with local environmental and waste disposal regulations.
13
Datex-Ohmeda S/5 Light Monitor
14
4. DESCRIPTION
Parts of the monitor
3
2
Operation
4
5
6
7
8
9
1
12
l
ll
11
Figure 4-1 S/5 Light Monitor
(1) Battery (2 pieces) (2) Recorder (option) (3) DataCard slot (option) (4) Handle with rail hooks (5) Alarm indicators (6) Function keys (7) ComWheel (8) Power On/Standby (9) External power indicator (10) External power adapter, or Power Adapter for Transport Vehicles (11) Battery Module (option) (12) Battery charge status LEDs
10
15
Datex-Ohmeda S/5 Light Monitor
Rear panel
4
3
5
2
1
Figure 4-2 Rear panel
(1) Dust filter (2) Feature connector, X3 (3) Connector for power adapter, X1 (4) Printer connector (serial), X2 (5) Mounting attachment (6) External VGA connector, X7 (7) Connector for network location ID plug, X6 (8) Network connector, X5
7
8
6
16
Keys and controls
Operation
On/Standby.
Silence Alarms
Start/Cancel
Zero ALL
Freeze
Recorder
Start/Stop
ComWheel
Normal Screen
Menu
NIBP
Returns back to normal monitoring display.
Silences the alarm sound for two minutes.
Calls up the main menu which enables changing the monitor settings.
Opens NIBP mini menu for autocycling or a single non-invasive blood pressure measurement. Cancels any NIBP measurement.
Zeroes all invasive pressure lines (except ICP which needs to be zeroed separately, see chapter “Invasive pressure”).
Stops the waveform sweep for 60 seconds. Pressing it again unfreezes the sweep.
Starts/stops recording of selected waveforms with the built-in recorder.
Menu functions are controlled by turning and pressing the ComWheel.
17
Datex-Ohmeda S/5 Light Monitor
Display
1
5
2
3
4
Figure 4-3 Parts of the display
(1) Alarm message field (for five messages) (2) Battery status field (3) Actual time (4) Up to four waveform fields with numeric information (5) Up to four digit fields
The waveform and digit fields are configurable, see chapter “Monitor setup.”
18
Measurement connectors
Connectors for measurement sensors are in the right side panel. Connectors are marked with the measurement abbreviations and symbols specifying the type of protection against electrical shock.
Operation
P1
NIBP
3
P2
2
SpO
2
CO
2
ECG +
1
Resp
4
5
T
Signal Out
6
7
Figure 4-4 Connector panel
(1) Mainstream CO2 connector (2) Connectors for invasive pressure lines (3) NIBP hose connector (4) Pulse oximetry sensor connector (5) ECG trunk cable connector, also used for impedance respiration
measurement
(6) Temperature probe connector
(7) Signal out connector
19
Datex-Ohmeda S/5 Light Monitor
5. OPERATION
Getting started
Turn on the power.
After 10 seconds a start-up display will appear, the CO2 measurement needs to warm up about 80 seconds in room temperature. If the monitor has been stored in the temperature below freezing point, it takes approximately one hour to warm up.
Make patient connections, check the accessories.
Zero invasive lines.
Check that desired parameters are displayed in digit and waveform fields.
If necessary, make appropriate adjustments.
During monitoring
Ensure that ECG electrodes have a good skin contact.
Change pulse oximetry measuring site frequently.
Observe the cuffed limb frequently during NIBP measurement.
If the monitor is connected to the network, the patient information and
alarms can be followed on the screens of the Datex-Ohmeda Central and the S/5 monitors connected to the system.
In-between patients
Print trends: see chapter “Recorder.”
Reset previous case: Press Menu and select Patient Data - Reset ALL.
Change single use accessories; clean reusable ones.
Check the monitor and accessories before use.
Plug in the power cord to charge batteries.
At the end of the day
Turn the monitor to standby.
Remove the power cord.
Clean the monitor and reusable accessories. Let dry completely.
Plug in the power cord to charge batteries.
20
Using menus
Operation
Press the Menu key.
Select the item by turning
and pressing the ComWheel.
Change settings by turning and pressing the ComWheel.
Need help? Short instructions appear in every menu.
Press Normal Screen.
21
Normal Screen
Datex-Ohmeda S/5 Light Monitor
6. BATTERIES
The monitor can be equipped with either the Battery Module (F-LBAT) with two removable batteries, or the Backup Battery (N-LBB) with a built-in battery pack in the monitor basement.
Battery capacity is displayed on the battery status field. The arrow shows the discharging (left) and recharging (right) level of the battery.
The Battery Module works for two hours (one hour / battery) and Backup Battery for 30 minutes with fully charged batteries.
NOTE: Make sure that there is enough battery capacity for a patient transport.
Charging batteries
Charge the batteries in +10...40 °C (50...104 °F) only.
Charging in the monitor
Keep the monitor connected to power outlet. The batteries are charged
when the monitor is either ON or STANDBY.
If monitor is turned ON, an arrow to the right is displayed on the Battery
Status field during charging.
The green LEDs on the battery module keep flashing until the battery is
fully charged.
Charging empty batteries to full capacity for Battery Module takes approximately 3.5 hours/battery, and for Backup Battery 1.5 hours. The protective case may increase the operating temperature so check the charging status when using it.
22
Removing batteries from the Battery Module
Push the release latch down.
The battery pops out from the frame.
Figure 6-1 Removing battery packs from the Battery Module
Replace the batteries with new ones if the total operation time is less than one hour. Part number is 896895.
Battery alarms
Operation
If something out of normal routine happens, a corresponding message appears. These battery alarm and note messages are explained in chapter “Maintenance
and troubleshooting.”
Storing and disposing of batteries
Keep the batteries in the monitor or external charger. If they are stored for longer periods, store them fully charged in a dry place. Before use, charge them fully.
CAUTION Do not put in fire. CAUTION Do not short-circuit any terminals. CAUTION Do not disassemble or mutilate. CAUTION Use only with Battery Module or External Battery Charger. CAUTION Do not insert any other than specified batteries into Battery Module
or External Charger.
23
Datex-Ohmeda S/5 Light Monitor
d
When stored for longer periods or used mostly with mains power; charge and discharge the batteries fully once a month.
This battery contains heavy metal cadmium (Cd), and when disposed, must be separated from other waste according to local regulations.
Ni-Cd
Battery packs can be recycled. Discharge before disposing.
Ni-C
24
Operation
7. CLEANING
WARNING ELECTRICAL SHOCK HAZARD:
Before cleaning, disconnect the monitor from electrical outlet. Do not immerse any part of the monitor in liquids or allow liquid to enter the interior. Do not autoclave any part of the monitor with steam or ethylene oxide. After cleaning, or if liquid has accidentally entered the interior of the monitor, make sure that every part of the monitor is dry before reconnecting it to the power supply.
CAUTION Do not use hypochlorite, acetone-, phenol-, or ammonia-based
cleaners.
Cleaning
Why ? Which items ? How ?
To remove dirt and most microbes.
Disinfecting
To kill and remove pathogenic microbes.
Sterilization
To kill the microbes and bacterial spores.
All parts daily. The items to be
disinfected or sterilized.
Accessories and supplies in contact with the patient’s skin or airways.
Accessories and supplies penetrating the patient’s tissue.
NOTE: Always detach accessories from the monitor and the patient before cleaning.
Wipe away visible organic material using a moist cloth and mild detergent, e.g. Datex-Ohmeda Cleaning Fluid.
Rinse as recommended by the detergent manufacturer.
Allow to dry completely.
With heat:
Wash in a washing machine (at least 1 min at 85°C/185°F).
With chemicals:
1. Immerse in 2% glutaraldehyde for 10 minutes (pH=6.5). After tuberculosis, HIV or hepatitis patient, the treatment time should be 20 min.
2. Rinse carefully and let dry.
or
Wipe with alcohol and let dry.
Steam autoclave at maximum 121°C (250 °F) for 15 min.
Use ethylene oxide mixture at 50 to 60°C (120 to 140°F). Follow the
sterilizer manufacturer’s recommendations for required aeration times.
Immerse in 2% glutaraldehyde for three hours, rinse carefully and let dry.
25
Datex-Ohmeda S/5 Light Monitor
Dust filter
Clean the dust filter at least once a month:
1. Pull out the filter frame on the rear panel. Do not remove any screws!
2. Shake the filter and blow out the dust. If the filter is damaged, replace it with a new one (p/n 886841).
Monitor casing
1. Wipe gently using a moist cloth and mild detergent.
2. Let dry completely before reconnecting the power cord.
Pulse oximetry sensors
1. Wipe the sensor with mild detergent solution.
2. Allow it to dry completely before use. The Sat Sensors may be sterilized using ethylene oxide. The Sat Sensors are latex free. Take possible patient allergies into account also
when selecting the cleaning agent. With the option using Nellcor sensors, see instructions in the sensor package.
WARNING PATIENT SAFETY: A damaged sensor, or a sensor soaked in
liquid, may cause burns during electrosurgery.
ECG cables, temperature probes, NIBP hose
Wipe the cables with mild detergent solution. When necessary, use disinfectant. Always follow the manufacturer’s instructions. Disinfect only the items that
tolerate the process.
NIBP cuff
Clean only when necessary:
1. Remove the bladder from the cuff.
26
2. Wash the bladder and the cuff in mild detergent solution. Do NOT use alcohol.
Invasive pressure cables
1. Wipe the cables with sterile alcohol-based detergent.
2. After cleaning wipe the surfaces with a cloth dampened with sterile water. Dry with a dry cloth.
Airway accessories
Adapters
Replace the single patient use adapter after each patient. If the adapter is occluded, it can be rinsed with water and air dried. It may then
be replaced into the ventilator circuit of the same patient. Multi-use adapters can be rinsed with water and steam autoclaved. The adapter
is designed to withstand 20 sterilization cycles. Before using, check that the adapter windows are dry and residue-free and that the adapter has not been damaged.
Operation
Make sure that all traces of alcohol or detergent are rinsed away or dried before connecting to the patient.
Mainstream CO2 sensor WARNING ELECTRICAL SHOCK HAZARD: Do not immerse the sensor in
liquids or autoclave it.
Wipe the sensor with a slightly damp cotton swab. When necessary, use disinfectant.
27
Datex-Ohmeda S/5 Light Monitor
28
8. ALARMS
Alarms
In alarm:
Messages (max. five messages) appear on the alarm message field at the
top of the display in the priority order from left to right.
Audible alarm is heard.
External Nurse Call signal is active, and if the monitor is connected to the
network, the alarm is passed into it.
Yellow or red alarm indicator (LED) next to the Silence Alarms key is
illuminated according to the alarm category.
The measured value, which activates the alarm, flashes and an
explanation may appear on the waveform, the digit field, or the alarm message field.
Alarms are active also when the parameter is not selected on the screen unless the alarm source is not turned off. See default settings in “Alarm tables/alarm sources” later in this chapter.
WARNING PATIENT SAFETY: Always make sure that alarm limits are set and
alarms are on when monitoring a patient.
Alarm categories
Alarms are classified into two types, red and yellow alarms. The alarm type depends primarily upon the physiological significance and the duration of the alarm; e.g., ‘Asystole’ advances rapidly to the red category.
The alarm messages are displayed in the order of priority from left to right. Audio alarms follow either General or ISO pattern. This selection can be made in
Install/Service menu and is stored in permanent memory. Select the same pattern for all the monitors in a hospital.
29
Datex-Ohmeda S/5 Light Monitor
Table 8-1 Alarm priority categories
Category Visual Audio pattern Meaning
Red alarm Red LED
If the alarm is connected to measured values, this number field is flashed.
Yellow alarm
Note Message is
Yellow LED If the alarm is connected to measured values, this number field is flashed.
displayed.
Smart start alarms
Alarms are automatically active once the patient connection cables are attached. Some alarms have their own specific requirements before they become active:
ISO: 3+2 beeps every 6 seconds.
--- -- --- --
Triple beep every 20 seconds.
--- ---
Single beep.
-
General: Continuous string of beeps.
-------------
Double beep every 5 seconds.
-- -- -- --
Single beep.
-
Life threatening situation.
Serious problem.
Advisory note or equipment related alarm.
Invasive pressures must be within alarm limits for 20 seconds after
zeroing.
Apnea alarm requires at least three detected breaths.
When apnea alarm is activated, the high and low alarms of CO
respiration will not be activated.
Silencing alarms
WARNING PATIENT SAFETY: Do not turn off audible alarms without
assuming continuous, direct observation of the patient.
Silencing alarms also disables the external Nurse Call signal.
30
and
2
Silencing alarms for two minutes
Press Silence Alarms .
Alarms
The symbol
with a countdown timer appears on the screen, and the
message is cleared from the display. Messages of new alarms also appear during silencing period. After two minutes, all silenced alarms and Nurse Call function will return to normal alarm operation.
To reactivate alarms:
Press the Silence Alarms key during the silencing period.
The alarm sounds of new alarms and the Nurse Call signal are activated.
Turning audible alarms on and off
It is possible to silence only apnea, ECG and apnea alarms, or all the alarms:
Press Menu.
Select Alarms Setup.
Select Audio ON/OFF.
Select Silence ECG, Silence Apnea, Silence Apnea + ECG or
Silence ALL.
31
Datex-Ohmeda S/5 Light Monitor
If an active alarm is suppressed, the monitor will give a reminder beep every two minutes. (Reminder beep volume can be adjusted by selecting Reminder Volume).
If asystole alarm is silenced, and ECG is used as the heart rate source, the HR high and low limits are also silenced.
Alarm messages
All messages are listed in chapter “Maintenance and troubleshooting.”
Alarms setup and adjustments
You can adjust alarm limits and volume, silence alarms permanently, and view alarm status (a 10-minute trend) in Alarms Setup menu.
Press Menu.
Select Alarms Setup.
Adjusting alarm limits
Alarm limits can be adjusted individually (Adjust Limits) or set near to real measured patient values (Auto Limits).
To adjust alarm limits:
Select Adjust Limits.
32
Other settings
Alarms
Select the desired high or low alarm limit by turning and pressing the
ComWheel.
Adjust the alarm limit up or down by turning the wheel.
Press the wheel to confirm the adjusted alarm limit.
Turn the ComWheel to move the highlight either to the next alarm limit or
to the Exit box.
Press the ComWheel.
Press Normal Screen to return to normal monitoring.
To set auto limits
Auto limits are calculated from the measured patient values at the time when auto limits are selected. These limits enable close patient control with individual alarms.
Select Auto Limits.
Press Normal Screen.
Default Limits Sets the alarms to the default alarm limits. Cancel Changes Returns all the values to those that were valid when the Alarms Setup menu
was entered.
Alarm Sources NIBP and invasive pressure alarms have alternative sources, (systolic, diastolic,
mean pressure, or OFF).
Alarm Volume The volume of the audio alarms can be adjusted from 1 (soft) to 10 (loud). Audio ON/OFF All alarms or asystole alone can be turned on and off. Alarm History Displays a list of the last 10 alarms message with respective times. Press the
ComWheel to get back to Alarms Setup menu.
Remove Menu Only the alarm limits and 10 minute trends are displayed.
Automatic alarm recording
You can record following alarm situations:
1. Asystole
2. HR high/low and Art low simultaneously (in model with invasive
pressure)
When the alarm reaches red state, the recorder will print the ECG + P1 waveforms. Printing appears 12 seconds after the alarm situation has started.
33
Datex-Ohmeda S/5 Light Monitor
2
To set alarm recording, press Menu and select Record&Print - Recorder Setup - Start on Alarms YES/NO.
Alarm tables
Alarm sources
If alarm source is OFF, no alarms of this parameter occur and the parameter box with the 10 minute trend is not displayed in the Alarm Setup menu.
Table 8-2 Alarm source alternatives and defaults
Alarm Source alternatives
P1 P2 Art
CVP
PA
ICP
NIBP
[Default selection]
[Sys] Dia Mean OFF Sys Dia Mean [OFF] [Sys] Dia Mean OFF Sys Dia Mean [OFF] Sys Dia Mean [OFF] Sys Dia Mean [OFF] [Sys] Dia Mean OFF
Default limits
The monitor is delivered with a certain set of alarm limits, so called factory default limits. These are listed for each parameter in the following tables.
It is possible to save alarm adjustments as defaults, see chapter “Monitor setup”, submenu “Install/Service menu.”
Do not change default settings without common consent of all users. If you change the default alarm limit, write the new limits in the following table.
Table 8-3 Adjustment ranges and default settings of alarm limits
Parameter Limit Limit range
min-max
HR High
SpO
NIBP Sys High
Low
High
Low
Low
35-250 bpm 30-245 bpm
51-100 %,OFF
50-100 %
15-265, OFF OFF, 15-265
Factory default User
default
160
40
OFF
90
180 (120 infant)
80(70 infant)
34
Alarms
2
2
Mean High
Dia High
Inv-P1 Sys High
Mean High
Dia High
Inv-P2 Sys High
Mean High
Dia High
Art Sys High
Mean High
Dia High
CVP Sys High
Mean High
Dia High
PA Sys High
Mean High
Dia High
ICP Sys High
Mean High
Dia High
EtCO
FiCO
Resp High
Temp High
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
High
Low
High
Low
Low
Low
15-265, OFF OFF, 15-265
15-265, OFF OFF, 15-265
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
-40 -300, OFF OFF, -40-300
0.1 - 13%, OFF OFF, 0 - 12.9%
0 - 3.0 %, OFF
OFF, 0 - 2.9%
5 - 60, OFF
OFF,4 - 59
10 - 45 °C, OFF
(50 - 113 °F, OFF)
OFF, 10 - 45 °C
(OFF, 50 - 113 °F)
140(90 infant)
60(50 infant)
100(70 infant)
40(40 infant)
180
80
140
60
100
40 20
5
15
0
10
-5
180
80
140
60
100
40 20
5
15
0
10
0
40 10
30
5
20
5
20
0
15
0
10
0
8%(60 mmHg, 8.1 kPa)
3%(23 mmHg, 3 kPa) 3%(23 mmHg, 3 kPa)
OFF
60
4
38 °C (100 °F)
15 °C (59 °F)
35
Datex-Ohmeda S/5 Light Monitor
Auto alarm limits
Auto Limits are based on the measured patient values according to the following tables.
Table 8-4 Auto alarm limits
Alarm Limit Auto limit calculation
HR High
Low Pulse Oximetry Low SpO2 - 5% NIBP Sys High
Low
Mean High
Low
Dia High
Low Inv-P1 Sys High
Low
Mean High
Low
Dia High
Low Inv-P2 Sys High
Low
Mean High
Low
Dia High
Low Art Sys High
Low
Mean High
Low
Dia High
Low CVP Sys High
Low
Mean High
Low
Dia High
Low
HR x 1.25 HR x 0.75
NIBP x 1.25 + 10
NIBP x 0.75 - 10
NIBP x 1.25 + 10
NIBP x 0.75 - 10
NIBP 1.25 +10
NIBP x 0.75 - 10
P1x 1.25 + 10
P1 x 0.75 -10
P1 x 1.25 + 10
P1 x 0.75 -10
P1 x 1.25 + 10
P1 x 0.75 -10 P2 x 1.25 + 5
P2 x 0.75 - 5
P2 x 1.25 + 5
P2 x 0.75 - 5
P2 x 1.25 + 5
P2 x 0.75 - 5
P1 x 1.25 + 10
P1 x 0.75 -10
P1 x 1.25 + 10
P1 x 0.75 -10
P1x 1.25 + 10
P1 x 0.75 -10
CVP x 1.25 + 5
CVP x 0.75 - 5
CVP x 1.25 + 5
CVP x 0.75 - 5
CVP x 1.25 + 5
CVP x 0.75 - 5
36
Alarms
PA Sys High
Mean High
Dia High
ICP Sys High
Mean High
Dia High
EtCO
2
FiCO
2
Resp High
Temp High
Low
Low
Low
Low
Low
Low High
Low High OFF
Low
Low
PA x 1.25 + 10
PA x 0.75 -10
PA x 1.25 + 10
PA x 0.75 -10
PA x 1.25 + 10
PA x 0.75 -10 ICP x 1.25 + 5
ICP x 0.75 - 5 ICP x 1.25 + 5
ICP x 0.75 - 5 ICP x 1.25 + 5
ICP x 0.75 - 5
EtCO2 + 1%
EtCO2 - 1%
RR *1.25 + 2
RR*.75 -2
Temp +1 °C (1.8 °F) Temp +1 °C (1.8 °F)
37
Datex-Ohmeda S/5 Light Monitor
9. MONITOR SETUP
The Monitor Setup menu enables you to select the information displayed on the screen, to set the time and date and to change the default settings via the Install/ Service menu.
Getting to Monitor Setup:
Press Menu.
Select Monitor Setup.
Screen setup
Select the parameters displayed in the waveform and lower digit fields. A parameter can be displayed in one place only at the time.
The monitor turns to the default settings every time it is turned off. If you want to save a certain screen setup as default, see Install/Service adjustments later in this chapter.
Selecting waveforms
Four waveforms can be displayed. If one field is set to ‘OFF’, the remaining waveforms are enlarged to fill the area.
38
Select Waveform Fields.
Select the field you want to rearrange.
Monitor setup
Select the parameter from the list.
Press Normal Screen.
Selecting Combine Pressures YES displays invasive pressures in the same field (only in models with invasive pressure).
Selecting Thick Waves YES draws the displayed waveforms with thicker trace.
Selecting numerical display fields
There are four fields in the lower part of the screen for numeral information, fields are numbered from left to right. If a digit field is turned OFF the digit field to its left is automatically enlarged to fill the space.
39
Datex-Ohmeda S/5 Light Monitor
Select Digit Fields.
Select the field you want to rearrange.
Select the parameter from the displayed list.
Press Normal Screen.
Setting time & date
Press Menu. Select Monitor Setup - Set Time and Date.
The time is shown in the upper right corner of the screen. Turning the monitor off does not affect the clock.
You can adjust hours, minutes, seconds, day, month, and year, or select the clock format of either 12 or 24 hours.
CAUTION Do not change the monitor’s time settings during patient
monitoring. Trend data may be lost.
NOTE: If the monitor is connected to network, the time is set by the network, and cannot be adjusted in the monitor.
40
Monitor setup
Install/Service menu
To enter Install/Service menu, a password 16-4-34 must be given.
Service View Service View is for a qualified technician only. It is entered with a password. Alarm Tones Select ISO or General pattern, see chapter “Alarms.” Save As Defaults
Saves permanently your new settings and configurations. The choices are Save ALL, Screen Setup, Trend Setup, Alarm Settings,
Recorder Setup and Param. Defaults. Select Save to save settings, Cancel to cancel changes, or Factory to return to
the factory settings.
Analog outputs Select the parameter for four analog outputs, as default, they all are off.
Channel Pin Selection range
1 #3 in Signal Out connector ECG or test signals 2 #4 in Signal Out connector P1 or test signals 3 #5 in feature connector freely selectable 4 #10 in feature connector freely selectable
Previous Menu Return to Monitor Setup menu.
Default settings
Default settings are always on when the monitor is turned on. They are marked in the following tables with brackets.
If you change the default settings using Save as Defaults, please write them down in the empty column. Do not change default settings without common consent of all users.
Table 9-1 Monitor setup selections and factory defaults
Monitor settings
Parameter Waveform Fields
Field 1 OFF [ECG] P1 P2 PLETH CO2 Resp Field 2 [OFF] ECG P1 P2 PLETH CO2 Resp (*P1, **Pleth, ***P1) Field 3 OFF ECG P1 P2 [PLETH] CO2 Resp (**CO2) Field 4 [OFF] ECG P1 P2 PLETH CO2 Resp (***CO2) Combine Pressures YES [NO] Thick Waves [YES] NO
Digit Fields
Lower Field 1 OFF HR P1 P2 [NIBP] SpO2 CO2 Resp Temp
Alternatives [Default selection]
User default
41
Datex-Ohmeda S/5 Light Monitor
Lower Field 2 [OFF] HR P1 P2 NIBP SpO2 CO2 Resp Temp(*P2, ***P2) Lower Field 3 OFF HR P1 P2 NIBP SpO2 CO2 Resp [Temp] Lower Field 4 OFF HR P1 P2 NIBP SpO2 CO2 [Resp] Temp (**OFF)
Set Time and Date
Clock Format 12 [24]
Parameter settings
ECG
ECG Lead I [II] III Size [1.0] 0.2-5.0 Beat Sound Volume [0] 0-10
ECG Setup
Sweep Speed 12.5 [25] 50 HR Source [AUTO] ECG Pleth Grid ON [OFF] Pacemaker Show [Hide] On R
Pulse Oximetry
Beat Sound Volume [0] 0-10 Pleth Scale **** AUTO 50 [20] 10 5 2
SpO2/Pleth Setup
Sweep Speed 12.5 [25] 50 HR Source [AUTO] ECG Pleth SpO2 Response B-to-B Normal [Slow]
Temperature
Temp Setup
Unit [°C] °F Label [Temp] Eso Naso Tymp Rect Blad Axil Skin AirW Room Myo Core Surf
Invasive Pressures
P1 Setup
Label P1 [Art] CVP ICP PA Scale [200] 10-300 Digit Format [S/D] Mean Sweep Speed 12.5 [25] 50 HR Source [AUTO] ECG Pleth
P2 Setup
Label P2 Art [CVP] ICP PA Scale [20] 10-300 Digit Format S/D [Mean] Sweep Speed 12.5 [25] 50 HR Source [AUTO] ECG Pleth
with N-LNSAT selections are: Fast Normal [Slow]
42
NIBP
Set Cycle Time 1 2.5 3 5 10 [15] 30 60
NIBP Setup
Ready Prompt [5] 0-10 Inflation Limits
150 200 [280] mmHg
CO2
Scale [6] 10 15 (%, kPa) (50 80 100 mmHg) Unit [%] kPa mmHg Sweep speed [Fast] Slow MS or OFF [MS] OFF
Recording and Printing
Parameter
Waveform 1 OFF [ECG] P1 P2 PLETH Waveform 2 [OFF] ECG P1 P2 PLETH (*P1, **Pleth, ***P1, ) Waveform 3 OFF ECG P1 P2 [PLETH] (**CO2, ***CO2) Graphic Trend 1 [HR] P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp Graphic Trend 2 HR P1 P2 NIBP [SpO2] ****Pleth CO2 Resp Temp
Recorder Setup
Start on Alarms YES [NO] Waveform Delay [12 s] OFF Trend Resolution 1 [5] 10 30 Paper Speed 6.25 12.5 [25] Recording Length [30 s] Cont
Alternatives [Default selection]
Monitor setup
User default
Display Trends
Parameter Trend Setup / Trend Page 1
Field 1 Real-time ECG in all models Field 2 OFF ECG [HR] P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp Field 3 OFF ECG HR P1 P2 [NIBP] SpO2 ****Pleth CO2 Resp Temp Field 4 OFF ECG HR P1 P2 NIBP [SpO2] ****Pleth CO2 Resp Temp Field 5 OFF ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 [Resp] Temp (**CO2,
Trend Setup / Trend Page 2
Field 1 Real-time ECG in all models Field 2 OFF ECG [HR] P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp
Field 3 OFF ECG HR P1 P2 NIBP [SpO2] ****Pleth CO2 Resp Temp (*P1,
Field 4 OFF ECG HR P1 P2 NIBP SpO2 [****Pleth] CO2 Resp Temp (*P2,
Field 5 OFF ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp [Temp]
Alternatives [Default selection]
***CO2)
(**SpO2)
**Pleth, ***P1)
**CO2, ***P2)
(**Resp)
43
User default
Datex-Ohmeda S/5 Light Monitor
Trend Setup / Trend Page 3
Field 1 Real-time ECG in all models Field 2 [OFF] ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp (*HR,
Field 3 [OFF] ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp (*SpO2,
Field 4 [OFF] ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp (*Pleth,
Field 5 [OFF] ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp
**HR, ***SpO2)
**SpO2, ***Pleth)
**Temp, ***CO2)
(*Temp,***Resp)
*) default in models with invasive pressure, **) default in models with CO2, ***) default in models with invasive pressure and CO
2
****) not available with the Nellcor option, N-LNSAT
44
10. TRENDS
The monitor accumulates and displays trend data for user selected periods between 2 and 24 hours.
There are two types of trend screens: graphic trend displays measured data in bars or dots, and numerical trends give the information in tabular form.
In addition a 10-minute trend is displayed in the Alarm Setup menu for every parameter with alarm limits.
During trend display, the monitor is active and the numerical information is continuously updated and displayed. Trend data is stored in memory for 15 minutes after the monitor is turned off.
Trends
45
Datex-Ohmeda S/5 Light Monitor
Viewing graphic trends
Press Menu.
Select Trends.
Select Graph 2 h, 4 h, 12 h or 24 h.
1
2
Figure 10-1 Graphic trend
3
4
5
(1) List of selections (2) Labels (3) Real time ECG above the trend information (4) Parameter trend fields (5) Movable trend cursor
Graphic trend pages
There are three pages of graphic trends. Pages are turned by selecting Next Graph. The page that is viewed last is stored in memory. It appears first when
trends are viewed again.
46
Table 10-1 The parameters trended in three pages
2
2
2
2
2
2CO2
2
2
2
2
2
2
2
2
2
1st Page
Basic model Model with invasive
pressure
ECG ECG ECG ECG HR HR HR HR NIBP NIBP NIBP NIBP SpO Resp Resp CO
SpO
Model with CO
SpO
Model with CO2 and invasive pressure
SpO
2nd Page
Basic model Model with invasive
pressure
ECG ECG ECG ECG HR HR SpO SpO Pleth* P2 CO Temp Temp Resp Temp
P1 Pleth* P1
Model with CO
Model with CO2 and invasive pressure
HR
P2
3rd Page
Basic model Model with invasive
pressure
ECG ECG ECG ECG Off HR HR SpO Off SpO Off Pleth* Temp CO Off Temp Off Resp
Model with CO
SpO
Model with CO2 and invasive pressure
Pleth*
Trends
*) not available with option N-LNSAT Parameters displayed on each page can be selected in Trend Setup.
Resolution
Resolution of trend display depends on the trend length selected.
Table 10-2 Trend lengths and resolutions
Trend length (hours) Resolution (minutes)
2 1
4 2 12 6 24 12
47
Datex-Ohmeda S/5 Light Monitor
Viewing numerical trend
You may view trend data also in numerical format. The trend information is updated every five minutes and always after NIBP measurement. The latest information is displayed at the bottom of the page.
Select Numerical.
Figure 10-2 Numerical trend page
(1) List of selections
(2) Real time ECG above the trend information
(3) Numeric trend information
(4) Movable cursor
Parameters displayed are: HR; SpO
; NIBP sys/dia; P1/Art sys/dia; P1/CVP mean; CO2 Et/Fi; RR
2
48
Changing trend setup
Select Trend Setup in Trends menu.
Select the graphic trend page you want to change.
Select the parameter for each field, or select OFF.
Trends
If you want to make your trend setup permanent you can store it, see chapter “Monitor setup/Default settings.”
Erasing trend data
Press Menu.
Go to Patient Data.
Select Reset Trends.
Select YES to confirm resetting trends.
49
Datex-Ohmeda S/5 Light Monitor
11. PATIENT DATA
Patient Data menu is used for identifying the patient case and erasing trends and settings.
To stop a monitoring case, reset trends between patients. Alternatively you can reset all data and settings. If the trends are just being printed, they cannot be erased. A message ‘Resetting not possible during printing. Please wait until printing is completed.’ appears.
The Patient Data menu will open automatically when the monitor is started less than 15 minutes after previous case, and the trends have not been erased. This is to confirm that the collected data is connected to the right patient.
Press Menu.
Select Patient Data.
Last Name Give patients last name, max. 14 characters, with ComWheel. First Name Give patients first name, max. 14 characters, with ComWheel. Patient ID Give patients ID, max. 14 characters, with ComWheel. This starts the case, the
message ‘Case started’ is displayed.
Reset Trends Erase trend memory between patients. Select YES to confirm the erasing, the
message ‘Case ended’ is displayed.
Reset ALL Return all the settings to their defaults and erase trend and alarm memory.
Select YES to confirm the erasing, the message ‘Case ended’ is displayed.
Main Menu Return to Main Menu.
50
Patient data
The Last Name, First Name and Patient ID can be entered also from the Datex­Ohmeda Central via network. The name is displayed as long as there are less than five alarm messages displayed simultaneously.
51
Datex-Ohmeda S/5 Light Monitor
12. RECORDING AND PRINTING
The optional built-in recorder provides single, dual and triple real time waveform recording, and single and dual recordings of numerical information. In addition it can record up to 24 hours of graphic and numerical trend data.
NOTE: Printings on thermal paper may be destroyed when exposed to light, heat, alcohol etc. Take a photocopy for archives.
Graphical trend data can also be printed on the laser printer either from the local port, or, if the monitor is connected to network, using a central printer.
Recording
Press Recorder Start/Stop to start and stop recording.
NOTE: Scaling of recorded waveforms follows the displayed parameter scaling, when applicable.
Selecting the waveforms to be recorded
Select up to three waveforms you wish to record in
Menu - Record/Print - Wave Setup menu.
Waveform x Select the recorded waveforms.
52
Recording and printing
Recording trends
Press Menu and select Record/Print.
Start recording by selecting Record Graph. Trend, or Record Num.
Trend, or Record Tab. Trend.
Stop printing by selecting Stop Graph./Num./Tab. Trend.
To select the graphical trend you want to record, go to Menu - Record/Print - Graph - Trend Setup.
Other recorder settings
To make the following selections, go to Menu - Record/Print - Recorder Setup:
Start on Alarms Select if an alarm activates the recorder or not (YES / NO).
Recording time is 30 seconds: 12 seconds recording from the recorder memory, 18 seconds real time recording. Alarm source annotated.
Waveform Delay Select the amount of waveform data (in seconds) that will be stored in the
recorder memory and printed before real-time waveform. The choices are: OFF or 12 seconds.
Paper Speed Select the paper speed 6.25, 12.5, or 25 mm/seconds. Recording Length
Select the printing time, either 30 seconds or continuous.
Previous Menu Return to Record&Print menu.
53
Datex-Ohmeda S/5 Light Monitor
Recorder paper replacement
Figure 12-1 Placing paper roll
Printing
Graphical trend pages P1, P2, and P3 can be printed either separately or all together. Print with a local laser printer, or if the monitor is connected to the network, with a central laser printer.
To print:
Press Menu.
Select Record/Print.
Select Print Graphical.
Select Print Graphs.
Press Menu.
Select Trends.
Select Print Page.
Selecting the printer
Press Menu.
Select Record/Print.
Select Print Graphical.
or
54
Select Printer Location.
Select Local for local printing or an appropriate network alternative.
Network printers are named in your Datex-Ohmeda Central.
Selecting the paper size
Press Menu.
Select Record/Print.
Select Print Graphical.
Select Printer Paper Size and the right one of A4 or Letter size.
Printer installation
A laser printer can be connected to the monitor’s serial connector and used as the local printer. The printer must be able to use PCL5, and a special converter cable (p/n 894193) is required when using the parallel connection. The printers with serial connection can be connected directly to the monitor with a standard cable.
Recording and printing
Purchase the printer locally. The printer memory should be at least 2 MB.
WARNING ELECTRIC SHOCK HAZARD: The printer must be supplied from an
appropriate extra separating transformer.
WARNING ELECTRIC SHOCK HAZARD: Before starting to use the system,
ensure that the entire combination complies with the international standard IEC 60601-1-1 and with the requirements of the local authorities.
Printer setup
For instructions refer to the printer’s documentation. Make sure that the printer settings are set to serial interface.
55
Datex-Ohmeda S/5 Light Monitor
13. NETWORK AND DATACARD
The monitor can be equipped with following options to enhance data continuum capabilities in the Datex-Ohmeda S/5 information system: N-LNET Network N-LDATA DataCard N-LDNET DataCard and network.
Networking
If the monitor is connected to the network, the symbol is displayed:
When the monitor is connected to network
Time and date are set by the network and cannot be adjusted in the
monitor.
The patients name can be entered in the Patient Data menu to be
displayed on the Datex-Ohmeda Central’s and the monitor’s screen and printed on the laser printouts.
Measured and entered patient data can be viewed on the Datex-Ohmeda
Central.
Measured patient data and alarms can be viewed on other
Datex-Ohmeda S/5 monitoring systems, but not vice versa.
Patient view and multi view on the Datex-Ohmeda Central screen
The Patient View on the Datex-Ohmeda Central screen looks like monitor’s own screen.
In the Multi View, the measured values from each of the monitor shown on the screen are:
8-Patient mode 16-Patient mode
ECG waveform ECG waveform HR, P1, SpO2, RR or
EtCO2 numbers Alarm field Alarm field
HR numbers
56
Monitor-to-Monitor communication
When the monitor is viewed on another S/5 monitor, the following parameters are displayed:
Network and DataCard
Waveforms EGC, Pleth, Art, CO Numbers HR, SpO2, Art, CO 1-hour numerical trend HR, SpO2, NIBP, P1, P2, CO
You cannot view other monitors on the S/5 Light Monitor.
DataCard
The DataCard is a rewritable flash memory card. With the DataCard, the patient trend data collected with the S/5 Light Monitor can be transferred to other S/5 monitors with record keeping or intensive care softwares, as well as to S/5 monitors with ARK software and S/5 monitors.
Data from any card cannot be loaded to the S/5 Light Monitor, nor loaded from the network to the S/5 Light Monitor or vice versa.
Inserting and removing the card
2
2
2
l l
l
Push the card into the slot on the left side panel of the monitor.
When the DataCard is inserted, this symbol is displayed. If the symbol is blinking, the card is not inserted properly.
If any other card or a faulty card is inserted, a message ‘Faulty DataCard ­change card’ is displayed.
57
Datex-Ohmeda S/5 Light Monitor
To remove the card:
Push the lever on the side of the slot.
Collecting data
When the card is inserted, the continuous patient trend data is automatically stored on it. When the card is filled, the oldest data is erased, but only if it is older than two days. You can retrieve information of approximately the last 24 hours. The memory capacity depends on the complexity of the saved data.
58
14. ECG
ECG monitoring is used to: measure the heart rate, detect arrhythmias, monitor pacemaker function, and myocardial ischaemia.
ECG reflects the electrical activity generated by the heart muscle. You can use 5- and 3-lead sets and trunk cables, but only leads I, II, or III can
be measured.
Monitoring ECG
1. Connect the patient cable to the blue ECG
2. Place the electrodes on the patient.
ECG
connector on the monitor’s side panel.
Connect the leadwires to the electrodes. NOTE: Always attach all the electrodes to the
patient, also when using a 5-lead cable.
3. The label of the lead is displayed in the ECG field
Applying electrodes
WARNING PATIENT SAFETY:
Ensure proper contact of the electrosurgery return electrode to avoid possible burns on the patient via ECG electrodes and probes. Make sure that no part of the patient connections touches any electrically conductive material, including earth.
Patient preparation
Ensure good skin contact with the electrodes. Remove excessive body
hair, skin oil etc.
Check, that the pre-gelled electrodes are moist and have not dried out
during storage.
and above the HR numerics.
59
Datex-Ohmeda S/5 Light Monitor
Placing electrodes
Avoid bones close to the skin, obvious layers of fat and major muscles.
Figure 14-1 Lead selections in typical electrode positioning
Electrode coding
Table 14-1 IEC and AAMI codes for ECG electrodes
3-LEAD IEC STANDARD 5-LEAD IEC STANDARD
L = yellow (left arm) F = green (foot) R = red (right arm)
RED (IEC ) WHITE ( AAMI)
RED (IEC) WHITE (AAMI)
LEAD I
GREEN (IEC) RED ( AAMI)
YELLOW (IEC) BLACK (AAMI)
LEAD I
LEAD II
YELLOW (IEC) BLACK (AAMI)
LEAD III
GREEN (IEC) RED (AAMI)
Left: Standard 3-Lead Positioning Middle: Modified CB5 3-Lead Positioning Right: Standard 5-Lead Positioning, place the fifth electrode in one of the six places indicated.
R = red (right arm) N = black (neutral) L = yellow (left arm) F = green (foot) C = white (chest)
3-LEAD AAMI STANDARD 5-LEAD AAMI STANDARD
LA = black LL = red RA = white
60
RA = white RL = green LA = black LL = red V = brown
Selecting trunk cable
Use a 5-lead set with a 5-lead trunk cable, and accordingly a 3-lead set with 3-lead trunk cable. (If you have a 3-lead set connected to a 5-lead trunk cable, you must set in the ECG Setup menu the selection 5 Lead Cable to 3elect).
What should happen when electrodes are applied ?
ECG display appears.
A beep is heard when a QRS complex is detected.
Selecting ECG lead
Press the Menu key and select ECG & Resp.
Select ECG Lead.
Select one of I, II, or III.
NOTE: Only 3 leads are measured, though a 5-lead set and cable are used. NOTE: To minimize the interference from the electrosurgery unit (ESU), place the
electrodes and select the lead so that the current from the surgical burning area to the return electrode does not go through the measured ECG lead.
ECG
ECG display
2
1
4
3
Figure 14-2 Display of ECG and HR
(1) Selected lead label (2) 1 mV reference bar (3) Heart rate (HR) label (4) Heart rate calculation source and message field for HR messages
61
Datex-Ohmeda S/5 Light Monitor
Cascaded ECG
If two fields are configured (See “Monitor setup/Waveform fields) for ECG display the waveform continues from one field to another. Thus more QRS complexes are displayed at the same time.
ECG adjustments
Press the Menu key.
Select ECG & Resp.
ECG Lead Select the displayed lead: I, II, or III. Size Adjust from 0.2 to 5 in steps of 0.1.
The bar on the left side of the ECG display is 1 mV.
Beat Sound Volume
A beat (QRS complex, or a plethysmographic or pressure pulse) produces a beep, and its volume is adjustable from 0 to 10 in steps of 1. Adjusting also changes the beat sound volume of SpO2.
(When SpO2 is monitored, the tone of the pulse beep rises with increasing oxygen saturation and falls as saturation decreases.)
ECG Setup Opens a submenu for ECG settings, see below. HR Alarm Opens a submenu for adjusting heart rate alarm, see chapter “Alarms.” Resp Setup Opens a submenu for Respiration settings, see chapter “Respiration.” Resp Rate Alarm Opens a submenu for adjusting respiration rate alarm, see chapter “Alarms.”
62
ECG
ECG setup
Sweep Speed Set the speed of the waveform sweep. Speeds are: 12.5, 25, and 50 mm/s on
a 9” screen. The 12.5 mm/sec speed is suggested for monitoring patients with a slow HR to
show more QRS complexes per sweep. The 50 mm/sec speed is recommended when monitoring patients with a high
HR or when using cascaded ECG. NOTE: The sweep speed is same for ECG, plethysmographic, and invasive
pressure waveforms.
HR Source Select the heart rate source. If ECG signal is too noisy for reliable heart rate
calculation, heart rate can be calculated from pressure (Art) or pulse oximetry (Pleth). The heart rate source appears above the numerical display of the heart rate.
The AUTO selects the first available of ECG, Art, and Pleth.
5 Lead Cable If you use a 5-lead trunk cable with a 3-lead set, select 3elect on, otherwise the
‘Leads Off’ message will be displayed.
Grid Select ON to display ECG waveform with 0.5 mV grid lines. Pacemaker Selects how to display the pacing pulse of a cardiac pacemaker, Show, Hide or
ON R. Hide: The pacing pulse is filtered from the waveform. Show: As above, but the pulse is marked with a constant height marker.
63
Datex-Ohmeda S/5 Light Monitor
ON R: Pacing pulses are not filtered from the waveform. This improves ECG monitoring with A-V pacemaker patients, as the QRS complexes are counted even if the pacing pulse hits the complex. On the other hand, during asystole the monitor may count pacing pulses as heart rate.
Setting heart rate alarm
Heart rate alarms can be adjusted in the Alarms Setup menu or accessed through the ECG & Resp. menu. See more detailed instructions in the “Alarms” chapter.
Alarm messages are explained in chapter “Maintenance and troubleshooting;
Alarm and information messages.”
Pacemaker patients
The monitor detects and rejects pacemaker pulses, see selection Pacemaker in the ECG Setup menu. Sometimes this may lead to unnecessary asystole alarms. The pacemaker may well change the shape of QRS so much that QRS detection may be affected.
WARNING PATIENT SAFETY: Monitor may count pacemaker pulses as heart
rate during cardiac arrest, some arrhythmias, or with certain types of pacemakers, particularly in ON R mode. Do not rely entirely upon alarms. Keep pacemaker patients under close surveillance.
64
Respiration
15. RESPIRATION
Respiration measurement uses ECG electrodes to measure the impedance changes caused by breathing.
The measurement is intended for patients three years old and up.
WARNING PACEMAKER PATIENTS: The impedance respiration
measurement may cause rate changes in Minute Ventilation Rate Responsive Pacemakers. Set the pacemaker rate responsive mode off or turn off the impedance respiration measurement on the monitor.
Monitoring respiration
Respiration measurement is activated when a digit or waveform field is selected for it:
Press the Menu key.
Select Monitor Setup.
Select Waveform Fields or Digit Fields.
Select Resp. to be displayed on an appropriate field.
Figure 15-1 Respiration display
Patient connections
Respiration uses the same electrodes as ECG, refer to the “ECG” section of this manual.
Respiration setup
Press the Menu key.
Select ECG & Resp.
65
Datex-Ohmeda S/5 Light Monitor
Select Resp Setup.
Size Adjust the scale height 1...50 mm/Ohm. The bar on the left displays always 1
Ohm.
Sweep Speed Set the speed of the waveform sweep. Speeds are: Fast 6.25 and Slow 0.625
mm/s on a 9” screen.
Resp. Rate Source
Select the respiration rate source of AUTO, CO2 and Imped. The AUTO sets the CO2 first, and if it is not available, impedance respiration is used. Default selection is AUTO.
Measurement Select ON to measure impedance respiration and OFF to set the measurement
off. Default selection is ON.
Detection Limit Adjusts measurement sensitivity. Selections are AUTO, 20%, 40%, 60%, or
100%. The percentage is the ratio to the 1 Ohm reference bar which is 100%.
Limitations of measuring
Movement artifacts
Changing the patient position, moving the head, moving the arms or shaking will result in movement artifacts. Also the heart may cause noticeable movement, and sometimes this may interfere with the respiration measurement.
66
Respiration
Electrical interference
Electrical devices that emit electromagnetic disturbance will result in artifacts or disable the respiration measurement completely. Such devices are electrosurgery units and infrared heaters, among others.
Intermittent mechanical ventilation
During spontaneous breathing the ventilator may at times support the patient’s ventilation with an extra inspiration. If these ventilator inspirations are substantially larger than the spontaneous breaths, the respiration calculation may mistakenly count only the ventilator produced inspirations and expirations. You can correct this by manually adjusting the detection limits.
1. Press the Menu key.
2. Select ECG & Resp.
3. Select Resp Setup.
4. Select Detection limit.
Measuring principle
ECG electrodes are used to measure impedance change across the thorax. Impedance change is caused by movements and changing air volume in the lungs.
Lead selection using 3-lead cables
I L and R F* II F* and R L III F* and L R Using 5-lead cables L and F N
*) may be neutral (N) in some older 3-lead sets.
Signal fed between electrodes
Signal measured
67
Datex-Ohmeda S/5 Light Monitor
16. PULSE OXIMETRY
Pulse oximetry is used for monitoring oxygen saturation (SpO2), pulse rate, and plethysmographic waveform.
The plethysmographic pulse waveform reflects the changes in blood pulsation at the measuring site. The amplitude of the waveform reflects the strength of perfusion, except when using the Nellcor option, N-LNSAT.
Pulse oximetry is measured with a sensor from finger, toe, ear, or with little children, from palm or feet.
Monitoring pulse oximetry
1. Attach the sensor to the gray connector on the
2. Attach the sensor to the patient. The SpO2 and
monitor’s side panel.
pulse rate values, and the plethysmographic pulsewave appear on the screen.
3. Change the sensor site frequently.
4. Remove the sensor from the patient and clean it.
68
Selecting sensor
Pulse oximetry
Reusable Sensors
OXY-F4-N
OXY-F-UN
OXY-F-DB
OxyTip+ Adhesive Sensors
OXY-AP
OXY-AF
Figure 16-1 Sat Sensors for different applications
NOTE: These sensors cannot be used when using the Nellcor option, N-LNSAT. Please refer to chapter “Alternative SpO
measurement.”
2
Applying sensors
Use dry and clean sensors only.
Confirm that red light lights up in the sensor when it is connected to the
monitor.
Clean the application site.
Remove nail polish, artificial fingernail, earrings etc. Clip long finger nails.
Attach the sensor cable to the wrist or bed clothes to prevent cable and
sensor from moving.
WARNING PATIENT SAFETY: Change measuring site frequently.
69
Datex-Ohmeda S/5 Light Monitor
Change sensor site and check skin and circulatory status every 2-4 hours
with adults, and every hour with little children.
What should happen when sensor is applied ?
The message ‘Pulse Search’ appears.
When ‘Pulse Search’ is completed, the pulse rate and SpO
appear.
The plethysmographic waveform will appear. The waveform scale is
established according to the measured amplitude.
Pulse beep is heard. The tone of the pulse beep will rise with increasing
oxygen saturation and fall as saturation decreases.
Pulse oximetry display
value will
2
1
Figure 16-2 Plethysmographic waveform with oxygen saturation number
(1) Bar reflecting the signal amplitude, not available with the Nellcor option,
N-LNSAT (2) Plethysmographic waveform (3) Oxygen saturation (SpO2)
Measuring limitations
Plethysmograph
The waveform is a good indicator of measurement validity. If the wave is noisy, amplitude is low, or the normal waveform cannot be seen, attach the sensor again.
Dyshemoglobins
2
3
The pulse oximeter cannot distinguish between oxyhemoglobin and dyshemoglobins. Thus, the SpO2 values may be erroneously high in smokers or in patients who have burns or carbon monoxide (CO) intoxication.
70
Adjustments
Pulse oximetry
Electrosurgery
Under some circumstances electrosurgery may cause noise on the display. Therefore, be careful in interpreting the results, especially the plethysmographic pulse waveform, during electrosurgery.
External interference
Ambient light, intravascular dyes and vasoconstrictive drugs may affect the accuracy of the measurement. Improper sensor attachment (measuring site too thick or thin, sensor ends not aligned), or patient movement may also interfere with the measurement.
Press the Menu key.
Select SpO2 & Temp.
Select and adjust by turning and pressing the ComWheel.
Beat Sound Volume
Adjust beat volume from 0 to 10.
Scale Select Scale from 50, 20, 10, 5, 2 or use AUTO which automatically adjusts
the scale according to the measured waveform. NOTE: The selection is not available with the Nellcor option, N-LNSAT.
Sweep Speed Set the speed of the display sweep for the waveform display. The speeds are
12.5, 25, or 50 mm/sec on a 9" screen.
NOTE: The sweep speed is same for ECG, plethysmographic, and invasive pressure waveforms.
71
Datex-Ohmeda S/5 Light Monitor
HR Source Select the heart rate source. If ECG signal is too noisy for reliable heart rate
calculation, heart rate can be calculated from pressure (Art) or pulse oximetry (Pleth). The heart rate source appears above the numerical display of the heart rate.
The AUTO (default) selects the source in the following order: ECG, pressure, and plethysmographic pulse waveform.
SpO2 Response Select how fast the SpO2 value follows the measurement. The selections are:
Beat-to-beat, Normal (10 s), or Slow (20 s, default setting). The selections with the Nellcor option, N-LNSAT, are Fast, Normal, and Slow.
SpO2 Alarms Opens a submenu for adjusting alarms, see the “Alarms” chapter. Temp Setup Opens a submenu for adjusting temperature settings, see the “Temperature”
chapter.
Main Menu Return to Main Menu.
Setting SpO2 alarms
The alarms can be adjusted in the Alarms Setup menu, accessed through the SpO2 & Temp menu. See more detailed instructions in the chapter “Alarms.”
Alarm messages are explained in the chapter “Maintenance and troubleshooting; Alarm and information messages.”
Measuring principle
Oxygen saturation, SpO2, expressed as a percent, defines the amount of oxygen carried compared to total capacity (also called functional or in vivo oxygen saturation). It is measured by a two wavelength pulse oximeter.
The SpO2 value is measured by light absorption technique. Red and infrared light (660 nm and 910 nm) is emitted from the light source. The amount of transmitted light is detected in the detector half. When the pulsative part of the light signal is examined, the amount of light absorbed by arterial hemoglobins is resulted, and the saturation level can be calculated.
The saturation percentage is displayed numerically, and the pulse beep sound varies according to the saturation level.
Plethysmographic pulse wave
The plethysmographic pulse wave is derived from variations of the intensity of the transmitted light and reflects the blood pulsation at the measuring site.
Pulse rate
Pulse rate is calculated from the plethysmographic pulse wave.
72
Alternative SpO2 measurement
As an alternative to the Datex-Ohmeda pulse oximetry, the monitor can be equipped with the Nellcor compatible saturation option, N-LNSAT. With this option you can use the sensors listed below.
NOTE: When the monitor has the N-LNSAT option, the monitor is equipped with the Nellcor measurement only and cannot use Datex-Ohmeda Sat Sensors.
Plethysmographic scale
You cannot adjust the scaling when using the N-LNSAT option; it is automatically set by the monitor.
Nellcor sensors
Use one of the following Nellcor sensors with the N-LNSAT module:
OXISENSOR II Oxygen transducer
D-25, D-25L Adult D-20 Pediatric I-20 Infant N-25 Neonatal/Adult R-15 Adult Nasal
Pulse oximetry
OXICLIC Oxygen transducer
A Adult P Pediatric
DURASENSOR Oxygen transducer
DS-100A Adult
DURA-Y Oxygen transducer
D-YS Adult/Pediatric/Infant/Neonatal
OXIBAND Oxygen transducer
OXI-A/N Adult/Neonatal OXI-P/I Pediatric/Infant
73
Datex-Ohmeda S/5 Light Monitor
17. TEMPERATURE
One temperature value can be measured from various sites using YSI-400 probes, or equivalents.
Monitoring temperature
Attach the probe to the patient.
Connect the temperature probe to the connector on the right side of the
monitor.
The measured value appears on the digit field.
Temperature display
Example:
2
1
Figure 17-1 Temperature display
(1) Label (2) Temperature value
Temperature testing
The monitor automatically tests the temperature measuring function at the start-up and after that every 10 minutes. If the test fails contact your local service representative.
74
F
Temperature setup
Press Menu.
Select SpO2 & Temp.
Select Temp Setup&Alarm.
Temperature
°C
°
Unit Selects unit, °C or °F. Temp Label Select the label:
AirW airway temperature Axil axillary temperature Blad bladder temperature Eso esophageal temperature Myo myocardial temperature Naso nasopharyngeal temperature Rect rectal temperature Room room temperature Skin skin temperature Temp temperature Tymp tympanic temperature Core central temperature Surf surface temperature
Previous Menu Return to SpO2 & Temp menu.
75
Datex-Ohmeda S/5 Light Monitor
18. NON-INVASIVE BLOOD PRESSURE (NIBP)
The monitor can measure NIBP automatically after set intervals, or take single measurements, or measure continuously for five minutes in STAT mode.
WARNING PATIENT SAFETY: The monitor sets the inflation pressure
according to the first measurement. Reset the monitor between patients to reset the inflation limit.
CAUTION Vibrations during transport may disturb the NIBP measurement.
Monitoring NIBP
Select the appropriate cuff and hose.
Table 18-1 NIBP cuff sizes
Cuff Color Width
cm inch
Large adult Standard adult Small adult Child Infant Disposable infant cuff #5
NOTE: Always use the white hose when measuring little children. Make also sure the Inflation Limits selection is set to 150 mmHg or Auto.
Red Blue Gray Green Tan
15 6 12 4.7 9 3.5 6 2.4
5.1 2
5.5 2.2
Limb circumference cm inch
33 - 47 13 - 18 25 - 35 9 - 14 18 - 26 7 - 10 10 - 19 4 - 7 9 - 14 3 - 5 10 - 14 4 - 6
Hose
Black Black Black Black White White
76
Over brachial artery
Non-invasive blood pressure
Attach the cuff to the patient and the hose to the NIBP connector located
on the right side panel of the monitor.
Make sure that NIBP tubes are not bent, pressed or stretched.
Measurement may be impaired.
Press the NIBP key to open the NIBP mini menu.
Select Start Cycling or Start Manual and press the ComWheel.
You can also adjust the cycling time by selecting Set Cycle Time. Adjust
the time by turning and pressing the ComWheel.
Observe the cuffed limb frequently. The measurement may impair blood
circulation.
Display of NIBP measurement
E
N
I
L
X
E
D
N
I
Use smaller cuff
Correct size
Use large cuff
1
2
3
4
(1) Systolic pressure (2) Diastolic pressure (3) Field shows the selected measurement (Manual, STAT), or the time since
last autocycle measurement
77
Datex-Ohmeda S/5 Light Monitor
(4) Mean pressure value of non-invasive blood pressure At the beginning of the measurement the cuff pressure is displayed in the mean
pressure value field. If motion artifacts are detected, the monitor automatically retains deflation until
the motion stops (maximum of 30 seconds). If the artifacts interrupt proper measurement, a new measurement is automatically activated.
When the measurement is ready, a short beep is heard and the result starts flashing.
Automatic NIBP double check
If the NIBP value exceeds the alarm limits, a new measurement is taken automatically (immediately, with MANUAL measurements, and after 30 seconds, with AUTO measurements). If an alarm situation persists, an alarm signal is given out.
Using NIBP functions
To use NIBP functions:
Press the NIBP key to open the NIBP mini menu
or
Press Menu and select NIBP & Inv. Press to open the
NIBP & Inv. Pressure menu.
78
Non-invasive blood pressure
Starting/Stopping Autocycle, Starting a Manual Measurement and Canceling Any Measurement are available through both menus. STAT mode and Venous Stasis are available through the NIBP & Inv. Pressure menu only.
Starting/Stopping autocycle
Select Start Cycling/Stop Cycling to start/stop automatic NIBP
measurement at selected intervals.
The bar at the bottom of the NIBP display shows the time remaining to the next measurement.
Autocycling is synchronized to actual time so that if the first measurement happens at 12.02, the next measurement is done at 12.05 and again at 12.10 (5 min intervals).
Starting a manual measurement
Select Start/Stop Manual.
Canceling any measurement
Press the NIBP key,
or
select the corresponding menu item.
79
Datex-Ohmeda S/5 Light Monitor
Starting a continuous measurement (STAT)
You can start the continuous measurement for five minutes (STAT mode) through the NIBP & Inv. Pressure menu only.
In STAT mode the early systolic value is measured and displayed until the final result is available.
After five minutes the monitor automatically returns to the previously selected cycling interval or to manual mode.
To start the measurement:
Press the Menu key and select NIBP & Inv. Press.
Select Start STAT.
Using venous stasis
You can hold a constant pressure in the cuff to help venous cannulation (Venous Stasis). This function is only available through the NIBP & Inv. Pressure menu. The pressure is maintained in the cuff for a certain time according to the inflation limits.
Table 18-2 Inflation limits and venous stasis times
Inflation limit Venous stasis pressure Venous stasis time
150 mmHg 40 mmHg 1 min 200 mmHg 60 mmHg 2 min 280 mmHg 80 mmHg 2 min
The pressurization time is displayed in the NIBP field and during the last 15 seconds the word ‘STASIS’ is flashed.
Press the Menu key.
Select NIBP & Inv. Press. and Start Ven. Stasis.
To release the pressure before the time is expired:
Press the NIBP key.
80
Non-invasive blood pressure
NIBP adjustments
Set Cycle Time Select desired interval time of 1, 2.5, 3, 5, 10, 15, 30, or 60 minutes. Ready Prompt Vol.
Adjust the volume of the beep that informs when the NIBP measurement is ready. The range is from 1 (soft) to 10 (loud), or 0 (OFF).
Inflation Limits Select the inflation pressure limit: 280, 200, or 150 mmHg. Calibration Check
Start the pressure calibration check with an external manometer. See instructions below.
NIBP Alarm Opens a submenu for adjusting the NIBP alarms, see the “Alarms” chapter. Previous Menu Return to the NIBP & Inv. Pressure menu.
Calibration check
Remove the cuff from the connector before entering the menu.
Select Calibration Check ON in NIBP Setup menu.
When the calibration menu appears, attach an external manometer with
pump to the connector.
Pump ca. 200 mmHg and compare the readings of the manometer and
display. If the difference is greater than 4 mmHg, the calibration must be checked by an authorized service person.
81
Datex-Ohmeda S/5 Light Monitor
NIBP alarms
NIBP Alarms are set in the Alarm Setup menu, see the “Alarms” chapter. The alarm messages are explained in chapter “Maintenance and troubleshooting.”
Measuring principle
The non-invasive blood pressure (NIBP) measurement uses the oscillometric measuring principle. The cuff is inflated with a pressure slightly higher than the presumed systolic pressure, then slowly deflated at a speed based on the patient’s heart rate, collecting data from the oscillations produced by the pulsating artery. Based on this data, the unit calculates values for systolic, mean and diastolic pressures.
82
Invasive blood pressure
19. INVASIVE BLOOD PRESSURE
Two invasive blood pressures can be measured with the model F-LMP1.
WARNING PATIENT SAFETY:
All invasive procedures involve patient risks. Use the aseptic technique. Follow catheter manufacturer’s instructions.
WARNING Ensure proper contact of the return electrode of the
electrosurgery unit to your patient to avoid possible burns at sensor sites.
WARNING Make sure that no part of the patient connections touches any
electrically conductive material including earth.
WARNING Use only defibrillator proof transducers and cables.
Monitoring invasive pressures
Preparing pressure line
Prepare the transducer kit according to transducer manufacturer’s
instructions. Ensure there is no air in the line.
Connect the kit to the transducer cable and the cable to the red
connector in the monitor.
Mount the transducer kit at heart level.
Connect the patient catheter to the pressure line.
Zero the transducer.
Open the line to the patient.
83
Datex-Ohmeda S/5 Light Monitor
1
2
Reusable
Disposable
3
4
Figure 19-1 Invasive pressure setup for disposable and reusable
(1) Heparinized fluid bag with pressure infusor (2) Flushing set (3) Transducer (4) Adapter cable for using disposable transducers
Zeroing transducer
Always zero the pressure transducer before monitoring and check zero level after power interruptions.
To zero the transducer:
Open the transducer to air.
Press the Zero ALL key,
Reusable
transducers
or
select Zero Px in the NIBP & Inv. Pressure menu.
The messages ‘Zeroing’ and ‘Zeroed’ are displayed accordingly. After
the channel is zeroed, the zeroing time appears in the menu. If the zeroing fails, the previous zero is used.
84
Invasive blood pressure
Open the transducer to the patient.
NOTE: ICP can be zeroed via the NIBP & Inv. Pressure menu only.
CAUTION Mechanical shock to pressure transducer may change zero balance
and calibration.
Invasive pressure display
After the line is zeroed and opened to the patient, the pressure wave appears. If two pressures are measured they can be combined with individual scales.
(Select separate or combined waveforms in the Monitor Setup/ Waveform Fields menu, see chapter “Monitor setup”).
2
3
1
Figure 19-2 Combined pressure waveforms
(1) Invasive blood pressure waveforms with zero and reference lines (2) Selected pressure label (3) Selected pressure scale (4) Systolic, diastolic and mean pressures (5) Ventilation mode selected in the Px Setup&Alarm menu
3
2
2
2
4
5
85
Datex-Ohmeda S/5 Light Monitor
Adjustments
Press Menu.
Select NIBP&Inv. Press.
Select Px Setup&Alarm.
Labeling the invasive line
The label of the pressure channel sets its scale, alarm source and alarm limits. The label alternatives are:
P1, P2 Standard labels Art Arterial pressure CVP Central venous pressure PA Pulmonary arterial pressure ICP Intracranial pressure
To change the label
Go to setup menu of the desired pressure.
Select Label.
Select the desired label from the list.
86
Table 19-1 Waveform scales, alarm sources and limits that are
Changing the scale
Scales can be adjusted between 10 and 300 mmHg in steps of 10.
Go to setup menu of the desired pressure.
Select Scale.
Adjust the scale by turning and pressing the ComWheel.
Invasive blood pressure
assigned when the label is selected
Label P1, Art P2, CVP PA ICP
Scale (mmHg) 200 20 60 20
Alarm source sys off off off
Digit format S/D Mean S/D Mean
Filter (Hz) 22 9 9 9
Other invasive pressure setup adjustments
Digit Format Select the displayed digit, either, systole and diastole (S/D) or Mean. Sweep Speed Set the display sweep speed for the waveform display. The speeds are: 12.5,
25, or 50 mm/sec.
NOTE: The sweep speed is same for ECG, plethysmographic, and invasive pressure waveforms.
Ventilation Mode
Select Spont or Contrl for spontaneous or controlled ventilation. Respiration may cause artifacts in invasive pressure measurement.
HR Source Select the heart rate source. If the ECG signal is affected by too much noise for
a reliable heart rate calculation, heart rate can be calculated from pressure (Art) or Plethysmographic pulse waveform (Pleth). The selected heart rate source is shown above the numerical display of the heart rate.
AUTO selects the first available of: ECG, Art, and Pleth.
Px Alarm Opens a submenu for adjusting Alarm settings, see the “Alarms” chapter. Previous Menu Return to the NIBP & Inv. Pressure menu.
87
Datex-Ohmeda S/5 Light Monitor
Invasive pressure alarms
Invasive pressure alarm is active only after pressure channel has been zeroed. For alarm setup, see the “Alarms” chapter. Alarm messages are explained in the chapter “Maintenance and
troubleshooting.”
Measuring principle
The transducer converts the pressure variations into electrical signals. The electrical signals are amplified and displayed as numeric pressure values and waveforms.
88
Carbon dioxide
20. CARBON DIOXIDE, CO
CO2 can be measured with mainstream method (N-LCM option) using Mainstream CO2 Sensor and airway adapter.
Respiration rate is the frequency of the peak (end tidal) CO2 value per minute.
Monitoring mainstream CO
2
2
Figure 20-1 Connecting mainstream CO2 sensor
1. Connect the special airway adapter between the elbow adapter and Y-piece with the saddles upright.
2. Attach the Mainstream CO2 Sensor to the adapter between the saddles so that the sensor cable fits into the slot on the farther saddle. Ensure the connection between the sensor and the adapter with the Velcro tape supplied with the sensor.
3. Connect the sensor to the connector on the monitor’s side panel.
Place the adapter as close to the patient airway as possible. Make sure that the sensor remains in upright position.
89
Datex Ohmeda S/5 Light Monitor
To remove the adapter:
Detach the sensor before removing the adapter from the circuit.
CAUTION Do not apply force to the sensor or sensor cable.
Select the airway adapter according to the patient’s status. Refer to the instructions for use provided in the adapters package.
Using the adult/pediatric mainstream airway adapter (endotracheal diameter
5.5 mm) increases the dead space by 6 ml. With the low dead-space airway adapter (diameter < 5.0 mm), the increase is 0.6 ml.
NOTE: Airway adapter selection recommendations should be used as a guideline. It is important for the clinician to assess each patient when selecting the appropriate airway adapter.
WARNING Do not allow the CO2 sensor to come in contact with the
patient’s skin for a prolonged period of time. Mainstream sensor
is heated to prevent the fogging of the airway windows. WARNING Do not immerse the sensor in liquids or autoclave it. WARNING Do not use damaged sensors or adapters. WARNING Do not clean or sterilize a single use adapter.
Display of CO2 measurement
1
Figure 20-2 Display of the CO2 waveform
(1) Continuous CO2 waveform (2) Values of end tidal (ET) and fraction of inspiratory (FI) CO (3) Respiration rate
2
2
3
90
Carbon dioxide
CO2 adjustments
Measurement Set on mainstream (MainSt) or turn the measurement OFF. Scale Selects 0 - 6 %, 0 - 10 %, 0 - 15 %, or kPa and 0 - 50 mmHg,
0 - 80 mmHg, 0 - 100 mmHg. Scale 0 - 6% is used in normal situations, 0 - 10% and 0 - 15% scales are used if hypercarbia is expected.
N2O/O2 Comp Selects compensation: OFF, N2O or O2. In the breathing circuit when there is
50% or more of N2O, use N2O compensation; when ther is 60% or more of O2, use O2. In other cases, turn off the compensation.
Unit Selects CO2 unit: %, kPa or mmHg. See “Unit conversions” later in this chapter. Sweep Speed Selects Fast (6.25 mm/sec) or Slow speed (0.62 mm/sec). Slow waveforms
have a sweep speed one tenth of normal, for a full screen sweep (five minutes). Slow waveforms show changes better than fast waveforms.
Resp Rate Source
Change respiration rate source: AUTO, CO2 or impedance respiration.
CO2 Alarm Brings you to the Alarms Setup menu to adjust CO2 alarms. Main Menu Returns you to the Main Menu.
Calibration check
No routine calibration by user is required. Calibration check is recommended to be performed once a year by a qualified technical personnel. See the “Technical Reference Manual” for instructions.
91
Datex Ohmeda S/5 Light Monitor
Unit conversions
Relationship between gas concentration and its partial pressure: reading in mmHg = (ambient pressure mmHg * gas concentration in %) / 100 reading in kPa = (ambient pressure in mmHg * gas concentration in % ) / 750
Measuring principle
The CO2 measurement is based on the infrared absorption. In the mainstream method the measurement bench is located at the patient’s airway adapter.
92
Loading...