Conformity according to the Council Directive 93/42/EEC concerning medical devices
CAUTION: U.S. Federal and Canadian law restricts this device to sale by or on the order of a
licensed medical practitioner. Outside the USA and Canada, check local laws for any
Datex-Ohmeda, Inc.
P.O. Box 7550, Madison
WI 53707-7550, USA
Tel. +1-608-221-1551
Fax +1-608-222-9147
www.us.datex-ohmeda.com
restriction that may apply.
Document No. 8000056-2
April 2003
Datex-Ohmeda Division,
Instrumentarium Corp.
P.O. Box 900, FIN-00031
DATEX-OHMEDA, FINLAND
Tel. +358 10 39411 Fax +358 9 146 3310
http://www.datex-ohmeda.com/
Instrumentarium Corp. All rights reserved.
Intended purpose
Datex-Ohmeda S/5 Light Monitor is intended for stationary and mobile monitoring of
patient’s ECG, SpO2, non-invasive and invasive blood pressure, temperature, respiration
and CO2 in hospitals and ambulances. When used outdoors, the monitor must be in the
protective case.
The monitor is intended for use by qualified medical personnel only.
The S/5 Light Monitor is intended for patients with weight from 5 kg (11 lb.) up.
Respiration monitoring is intended for patients three years old and up.
Classification according to IEC 60601-1
−CLASS I EQUIPMENT and INTERNALLY POWERED EQUIPMENT according to the type
of protection against electrical shock.
−TYPE BF or CF equipment according to the degree of protection against electric shock
is indicated with a symbol beside each patient connector.
−Degree of protection against harmful ingress of water as detailed in the
IEC 60529:
Monitor: IPX1, vertically falling water drops shall have no harmful effects (applicable
when the monitor is in upright position, or tilted backwards). In the protective case
IPX4, splash proof, when the case is closed properly, the monitor is intact and
operates on battery power.
Power adapter: IPX0, ordinary equipment.
Power Adapter for Transport Vehicles: IPX1.
−EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC
MIXTURE with air or with oxygen or nitrous oxide.
−CONTINUOUS OPERATION according to the mode of operation.
Classification according to EU Medical Device Directive
− The monitor is classified as IIb.
Responsibility of the manufacturer
Instrumentarium Corp. is responsible for the effects on safety, reliability and performance of
the equipment only if:
− assembly, operations, extensions, readjustments, modifications, service and repairs
are carried out by authorized personnel.
− the electrical installation of the relevant room complies with appropriate
requirements.
− the equipment is used in accordance with this manual.
The CO2 measuring option (N-LCM) is manufactured for Instrumentarium Corp. by
Pryon Corp., 8500 S.W. Creekside Place, Beaverton, OR 97008-7107 USA.
The Nellcor option (N-LNSAT) is manufactured for Instrumentarium Corp. by
Mallinckrodt Inc., 675 McDonnell Boulevard, P.O. Box 5840, St. Louis, MO 63134 USA.
Trademarks
ComWheel, Datex-Ohmeda AS/3, Datex-Ohmeda CS/3, Light, EarSat, FingerSat and
Flexlite are trademarks of Instrumentarium Corp.
Nurse call and defibrillator synchronization signals ............................................ 115
23. Accessories 117
24.INDEXi
vi
1. HOW TO USE THIS MANUAL
This manual describes how to use and maintain the Datex-Ohmeda S/5 Light
Monitors F-LM1 and F-LMP1.
The General section gives you information about safety aspects and
installation. Get acquainted with them before using the monitor.
The Basic Operation section helps you through the daily basic use of the
monitor.
The Detailed Operation section describes the monitor functions in detail. These
chapters are necessary if you want to master all the features available in the
monitor.
The Troubleshooting/Specifications and Index sections help you find the
information you need in various situations.
NOTE: Before using your monitor, please read this manual. Pay special attention
to WARNING and CAUTION statements.
General
Related documents
S/5 Light Monitor Technical Reference Manual
The Technical Reference Manual provides detailed technical descriptions and
instructions for the service, maintenance procedures and troubleshooting.
Light Monitor Schematic Diagrams
The Schematic Diagrams provide schematic diagrams and part layouts for the
monitor.
Patient Monitor Supplies and Accessories
The Patient Monitor Supplies and Accessories Catalogue provides information
on all supplies and accessories available for the monitor.
1
Datex-Ohmeda S/5 Light Monitor
Symbols and abbreviations
Symbols on the equipment
Attention, consult accompanying documents.
on the rear panel
indicates the warning:
on the rear panel dust
filter indicates the
reminder:
on the patient
connector panel
indicates the
warnings and
cautions:
on the external power
adapter indicates the
cautions:
Electric shock hazard. Do not open monitor frame. Refer servicing to
qualified personnel.
Do not touch any part of the monitor or patient connections during
defibrillation procedure.
Disconnect power supply before servicing.
Check rear panel dust filter regularly, and clean at least once a
month.
Use only the specified cables and accessories.
Ensure proper contact of the return electrode of the electrosurgery
unit to your patient to avoid possible burns at ECG electrode or other
probe sites.
The output signals are not floating and must not be connected
directly to a patient.
For continued protection against fire hazard, replace only with same
type and rating of fuse.
Make sure voltage selector is set according to local voltage.
For indoor use only. For patient transport outdoors, use the monitor
in its protective case and with battery power only.
Do not immerse in liquid or allow any liquid to enter the interior.
on the battery unit
indicates the
warnings and
cautions:
on the CO2 sensor:Do not immerse the sensor in liquids or autoclave it.
Do not put in fire.
Do not short-circuit any terminals.
Do not disassemble or mutilate.
Use only with Battery Module or External Battery Charger.
Discharge and charge batteries fully once a month to restore full
capacity.
Sealed NiCd battery must be recycled or disposed of properly.
Discharge before disposing.
2
Ni-Cd
d
Ni-C
General
SpO2, CO2: Type BF (IEC-60601-1) protection against electrical
shock, not defibrillation proof.
NIBP: Type BF (IEC-60601-1) defibrillator-proof protection against
electrical shock.
Temperature: Type CF (IEC-60601-1) protection against electrical
shock.
ECG, Impedance Respiration, P1, P2: Type CF (IEC-60601-1)
defibrillator-proof protection against electrical shock.
Battery packs contain heavy metal cadmium (chemical symbol Cd)
and in case of disposal, must be separated from other waste
according to local regulations.
Battery packs contain Ni-Cd and they can be recycled.
IPX-class:
IPX0
IPX1
IPX4
Degree of protection against harmful ingress of water as detailed in
the IEC 60529:
- ordinary equipment
- vertically falling water drops shall have no harmful effects
- splash proof
Battery charging LEDs: LEDs are blinking while batteries are being
charged and stay illuminated when batteries are fully charged and
the monitor is connected to mains.
The monitor is connected to the network (optional).
The DataCard (option) is inserted. If symbol is blinking, the card is
faulty or it is not inserted properly.
A blinking heart next to the heart rate or pulse rate value indicates
the beats detected.
A lung next to the respiration rate value indicates that respiration
rate is calculated from the impedance respiration measurement.
Silence alarm indicator. When displayed on the upper left corner of
the screen, indicates that all the alarms are silenced. When in the
menu or digit fields, indicates that the alarm source has been turned
off.
Alternating current
3
Datex-Ohmeda S/5 Light Monitor
F
Input voltage
Output voltage
Ethernet connectors
Printer connector
Sample gas outlet
Do not immerse the sensor in liquids.
Symbols used in this manual
Indicates the procedure for making adjustments (turn and press) in the
menus.
Submenu. Selecting a menu item with this symbol opens a new menu.
Temptemperature
AirWairway temperature
Axilaxillary temperature
Bladbladder temperature
Esoesophageal temperature
Myomyocardial temperature
Nasonasopharyngeal temperature
Rectrectal temperature
Roomroom temperature
Skinskin temperature
Temptemperature
Tymptympanic temperature
Corecentral temperature
Surfsurface temperature
Names of the keys appear in bold typeface: Normal Screen.
Selectable menu items appear in bold and italic typeface: Previous Menu.
Informative messages displayed on the screen are enclosed in single quotes:
‘Please wait’.
6
Safety
2. SAFETY
Warnings
WARNING Indicates situations in which the user or the patient may be in
danger of injury or death.
EXPLOSION HAZARD
• Do not use the monitor in presence of flammable anesthetics.
ELECTRICAL SHOCK HAZARD
• Do not remove cover. Refer servicing to qualified service personnel.
• Use only hospital-grade grounded electrical outlets and power cord.
Grounding reliability can only be achieved when the equipment is
connected to a receptacle marked ‘hospital grade’.
• Make sure that external equipment is hospital grade grounded before you
connect it to the monitor. Do not connect any external equipment to the
monitor, except that specified by Datex-Ohmeda.
• When connecting external equipment to monitor, make sure that the
whole combination complies with safety standard IEC 60601-1-1 for the
medical electrical systems and with the requirements of local authorities.
A non-medical electrical equipment connected to monitor shall be
supplied from an extra separating transformer with a protectively
grounded power outlet.
• Before cleaning, disconnect the monitor from electrical outlet.
• Do not immerse any part of the monitor in liquids or allow liquid to enter
the interior.
• Do not autoclave any part of the monitor with steam or ethylene oxide.
• After cleaning, or if liquid has accidentally entered the interior of the
monitor, make sure that every part of the monitor is dry before
reconnecting it to the power supply.
7
Datex-Ohmeda S/5 Light Monitor
PATIENT SAFETY
• Always make sure that necessary alarm limits are set and operative when
you start monitoring.
• When the alarms are suppressed, observe the patient frequently.
• Connect only one patient to one monitor at a time.
• Constant attention by a qualified professional is needed whenever a
patient is under anesthesia or connected to a ventilator. Some
equipment malfunctions may pass unnoticed in spite of the monitor
alarm.
• Do not use antistatic or electrically conductive breathing tubes. They may
increase the risk of burns when electrosurgery unit is utilized.
• Do not use the monitor during magnetic resonance imaging (MRI).
• ECG, Impedance Respiration, InvBP: Ensure proper contact of the return
electrode of the electrosurgery unit to your patient to avoid possible burns
at sensor sites.
• ECG, Impedance Respiration, InvBP: Make sure that no part of the patient
connections touches any electrically conductive material including earth.
• Impedance Respiration: In obstructive apnea, thoracic movements and
impedance variations may continue.
• NIBP: The monitor sets the inflation pressure automatically according to
the first measurement. Reset the case to reset the inflation limit before
measuring a new patient.
• PATIENTS WITH PACEMAKERS OR ARRHYTHMIAS: Monitor may count the
pacemaker pulses as heart beats during cardiac arrest, some
arrhythmias, and with certain types of pacemakers particularly in ON R
mode. Do not rely entirely upon rate meter alarms. Keep patients with
pacemakers and arrhythmias under close surveillance.
• PACEMAKER PATIENTS: The impedance respiration measurement may
cause rate changes in Minute Ventilation Rate Responsive Pacemakers.
Set the pacemaker rate responsive mode off or turn off the impedance
respiration measurement on the monitor.
• SpO
: A damaged sensor or a sensor soaked in liquid may cause burns
2
during electrosurgery.
8
Safety
• SpO
: Change measuring site frequently. Change the sensor site and
2
check skin and circulatory status every 2-4 hours with adults and every
hour with small children.
• The output signals are not floating and must not be connected directly to
a patient.
• Do not use delayed analog signals for defibrillator or intra-aortic balloon
pump synchronization.
• The monitor is intended for use by qualified medical personnel only.
• InvBP: All invasive procedures involve risks to the patient. Use aseptic
• InvBP: Use only defibrillator proof cables and transducers.
Cautions
CAUTION Indicates situations in which the unit or devices connected to it may
be damaged.
• Use licensed patient cables and accessories only, see Patient Monitor
supplies and accessories catalogue. Other cables and accessories may
damage the system or interfere with measurement.
• Before connecting the power cord, ensure that the input voltage selector
is set correctly in the external power adapter.
• Use the monitor outdoors with the battery power only.
• Do not store the monitor beyond the specified temperature range.
• Do not use hypochlorite, ammonia-, phenol- or acetone-based cleaners.
They may damage the monitor.
• Clean rear panel fan dust filter once a month or whenever needed.
• Leave space around the monitor for ventilation to prevent the monitor
from overheating.
• Do not apply pressurized air to any outlet or tubing connected to the
monitor. Pressure may destroy sensitive elements.
• When used in moving vehicle, mount the monitor properly.
• Dispose of the device and parts thereof in accordance with local
regulations.
9
Datex-Ohmeda S/5 Light Monitor
• Vibrations during transport may disturb SpO
• CO
• DataCards: Do not subject cards to excessive heat, bending, or magnetic
• InvBP: Mechanical shock to invasive blood pressure transducer may
• Electromagnetic interference from radio-frequency transmitting devices in
BATTERY CAUTIONS
• Do not put in fire.
• Do not short-circuit any terminals.
• Do not use damaged or leaking batteries.
• Do not disassemble or break the batteries.
• Use only with Battery Module or External Battery Charger.
, ECG, impedance
2
respiration and NIBP measurements.
, SpO2: Do not apply force to sensors or sensor cables.
2
fields.
cause severe shifts in zero balance and calibration and produce
erroneous readings.
ambulances may affect the performance of this medical device.
• Charge and discharge batteries fully once a month to restore full
capacity.
• Sealed NiCd battery must be recycled or disposed of properly. Discharge
before disposing.
10
3. UNPACKING AND INSTALLATION
Unpacking
The monitor is delivered in one box containing typically (as minimum):
− Datex-Ohmeda S/5 Light Monitor
− Datex-Ohmeda S/5 Light Monitor Power Adapter with power cord or
optional Power Adapter for Transport Vehicles
− User’s Reference Manual
Check that the contents correspond to your order and that all additional options
you have ordered are included. If any of the items appears damaged or is
missing, contact your sales person immediately.
Placing
The monitor can be placed on a flat surface or hung with the handle from bed or
wall rails. Make sure that the surface or rail holds up to at least 10 kg/22 lb.
Installation
Figure 3-1Installation
A protective case is also available. If using the case, make sure that the monitor
is always in a standing position to ensure proper ventilation. The monitor is
splash-proof (IPX4) when it is in the protective case and all the openings of the
case are closed properly. Make sure the monitor is intact and operates on
battery power. When used outdoors, the monitor must be in the protective case.
NOTE: The pocket in the protective case is not water-proof. Do not place items in
the pocket that may be damaged by moisture.
Universal mounting solutions are sold separately. Contact your local distributor.
11
Datex-Ohmeda S/5 Light Monitor
Electrical Power Adapter
Use only the power cord and power adapter provided with the monitor.
• Set the voltage selector on the adapter to the right position (115V/230V)
e.g. with a pen.
• Connect the adapter cord to the connector on the rear panel.
• Connect the power cord to the adapter and then plug onto the wall
electrical outlet.
NOTE: To use the power adapter, it must be lying on a flat surface or hanging
from the holder.
Do not let any liquid enter the adapter.
When external power supply is in use, a green led is lit.
Figure 3-2Placing Power Adapter
WARNING ELECTRICAL SHOCK HAZARD: Use only hospital grade grounded
electrical outlets.
CAUTION Before connecting the power cord, ensure that the
input voltage selector is set correctly in the power
adapter.
In vehicles with 12 - 32 V power source, use the Power Adapter for Transport
Vehicles, N-LPOWT. For information on how to install and use this adapter, refer
to the instruction sheet provided with it, or contact your local distributor.
12
Installation
CAUTION Electromagnetic interference from radio-frequency transmitting
devices in ambulances may affect the performance of this medical
device.
Charge batteries
Keep the monitor connected to the mains outlet. The batteries are fully charged
when the battery capacity indicators on the screen show FULL and green LEDs
in the Battery Module stop flashing.
CAUTION When stored for longer periods, or used mostly with mains power,
charge and discharge the batteries fully once a month.
Connecting to network
The monitor can be connected to a monitor network for central monitoring and
printing services.
• Connect the Ethernet and location ID plug to the corresponding
connectors on the monitor’s rear panel.
• Configure the location IDs in the Datex-Ohmeda Central.
When the monitor is connected to the network and turned on, the network
symbol is displayed on the upper right corner of the screen:
Configuration
When the monitor is delivered, it is configured with the factory default settings
and is ready to use.
If you want to change the settings and make your own permanent configuration,
refer to chapters “Alarms” and “Monitor setup” of this manual.
Disposal
Dispose of the monitor, or parts of it, in accordance with local environmental
and waste disposal regulations.
13
Datex-Ohmeda S/5 Light Monitor
14
4. DESCRIPTION
Parts of the monitor
3
2
Operation
4
5
6
7
8
9
1
12
l
ll
11
Figure 4-1S/5 Light Monitor
(1) Battery (2 pieces)
(2) Recorder (option)
(3) DataCard slot (option)
(4) Handle with rail hooks
(5) Alarm indicators
(6) Function keys
(7) ComWheel
(8) Power On/Standby
(9) External power indicator
(10) External power adapter, or Power Adapter for Transport Vehicles
(11) Battery Module (option)
(12) Battery charge status LEDs
10
15
Datex-Ohmeda S/5 Light Monitor
Rear panel
4
3
5
2
1
Figure 4-2Rear panel
(1) Dust filter
(2) Feature connector, X3
(3) Connector for power adapter, X1
(4) Printer connector (serial), X2
(5) Mounting attachment
(6) External VGA connector, X7
(7) Connector for network location ID plug, X6
(8) Network connector, X5
7
8
6
16
Keys and controls
Operation
On/Standby.
Silence
Alarms
Start/Cancel
Zero ALL
Freeze
Recorder
Start/Stop
ComWheel
Normal
Screen
Menu
NIBP
Returns back to normal monitoring display.
Silences the alarm sound for two minutes.
Calls up the main menu which enables changing the monitor
settings.
Opens NIBP mini menu for autocycling or a single non-invasive
blood pressure measurement. Cancels any NIBP measurement.
Zeroes all invasive pressure lines (except ICP which needs to be
zeroed separately, see chapter “Invasive pressure”).
Stops the waveform sweep for 60 seconds. Pressing it again
unfreezes the sweep.
Starts/stops recording of selected waveforms with the built-in
recorder.
Menu functions are controlled by turning and pressing the
ComWheel.
17
Datex-Ohmeda S/5 Light Monitor
Display
1
5
2
3
4
Figure 4-3Parts of the display
(1) Alarm message field (for five messages)
(2) Battery status field
(3) Actual time
(4) Up to four waveform fields with numeric information
(5) Up to four digit fields
The waveform and digit fields are configurable, see chapter “Monitor setup.”
18
Measurement connectors
Connectors for measurement sensors are in the right side panel. Connectors are
marked with the measurement abbreviations and symbols specifying the type of
protection against electrical shock.
Operation
P1
NIBP
3
P2
2
SpO
2
CO
2
ECG +
1
Resp
4
5
T
Signal
Out
6
7
Figure 4-4Connector panel
(1) Mainstream CO2 connector
(2) Connectors for invasive pressure lines
(3) NIBP hose connector
(4) Pulse oximetry sensor connector
(5) ECG trunk cable connector, also used for impedance respiration
measurement
(6) Temperature probe connector
(7) Signal out connector
19
Datex-Ohmeda S/5 Light Monitor
5. OPERATION
Getting started
• Turn on the power.
After 10 seconds a start-up display will appear, the CO2 measurement needs to
warm up about 80 seconds in room temperature. If the monitor has been stored
in the temperature below freezing point, it takes approximately one hour to
warm up.
• Make patient connections, check the accessories.
• Zero invasive lines.
• Check that desired parameters are displayed in digit and waveform fields.
If necessary, make appropriate adjustments.
During monitoring
• Ensure that ECG electrodes have a good skin contact.
• Change pulse oximetry measuring site frequently.
• Observe the cuffed limb frequently during NIBP measurement.
• If the monitor is connected to the network, the patient information and
alarms can be followed on the screens of the Datex-Ohmeda Central and
the S/5 monitors connected to the system.
In-between patients
• Print trends: see chapter “Recorder.”
• Reset previous case: Press Menu and select Patient Data - Reset ALL.
• Change single use accessories; clean reusable ones.
• Check the monitor and accessories before use.
• Plug in the power cord to charge batteries.
At the end of the day
• Turn the monitor to standby.
• Remove the power cord.
• Clean the monitor and reusable accessories. Let dry completely.
• Plug in the power cord to charge batteries.
20
Using menus
Operation
• Press the Menu key.
• Select the item by turning
and pressing the ComWheel.
• Change settings by turning
and pressing the ComWheel.
Need help?
Short instructions appear in
every menu.
• PressNormal Screen.
21
Normal
Screen
Datex-Ohmeda S/5 Light Monitor
6. BATTERIES
The monitor can be equipped with either the Battery Module (F-LBAT) with two
removable batteries, or the Backup Battery (N-LBB) with a built-in battery pack
in the monitor basement.
Battery capacity is displayed on the battery status field. The arrow shows the
discharging (left) and recharging (right) level of the battery.
The Battery Module works for two hours (one hour / battery) and Backup Battery
for 30 minutes with fully charged batteries.
NOTE: Make sure that there is enough battery capacity for a patient transport.
Charging batteries
• Charge the batteries in +10...40 °C (50...104 °F) only.
Charging in the monitor
• Keep the monitor connected to power outlet. The batteries are charged
when the monitor is either ON or STANDBY.
• If monitor is turned ON, an arrow to the right is displayed on the Battery
Status field during charging.
• The green LEDs on the battery module keep flashing until the battery is
fully charged.
Charging empty batteries to full capacity for Battery Module takes
approximately 3.5 hours/battery, and for Backup Battery 1.5 hours. The
protective case may increase the operating temperature so check the charging
status when using it.
22
Removing batteries from the Battery Module
• Push the release latch down.
• The battery pops out from the frame.
Figure 6-1Removing battery packs from the Battery Module
Replace the batteries with new ones if the total operation time is less than one
hour. Part number is 896895.
Battery alarms
Operation
If something out of normal routine happens, a corresponding message appears.
These battery alarm and note messages are explained in chapter “Maintenance
and troubleshooting.”
Storing and disposing of batteries
Keep the batteries in the monitor or external charger. If they are stored for longer
periods, store them fully charged in a dry place. Before use, charge them fully.
CAUTION Do not put in fire.
CAUTION Do not short-circuit any terminals.
CAUTION Do not disassemble or mutilate.
CAUTION Use only with Battery Module or External Battery Charger.
CAUTION Do not insert any other than specified batteries into Battery Module
or External Charger.
23
Datex-Ohmeda S/5 Light Monitor
d
When stored for longer periods or used mostly with mains power; charge and
discharge the batteries fully once a month.
This battery contains heavy metal cadmium (Cd), and when disposed, must be
separated from other waste according to local regulations.
Ni-Cd
Battery packs can be recycled. Discharge before disposing.
Ni-C
24
Operation
7. CLEANING
WARNING ELECTRICAL SHOCK HAZARD:
Before cleaning, disconnect the monitor from electrical outlet.
Do not immerse any part of the monitor in liquids or allow liquid
to enter the interior.
Do not autoclave any part of the monitor with steam or ethylene
oxide.
After cleaning, or if liquid has accidentally entered the interior
of the monitor, make sure that every part of the monitor is dry
before reconnecting it to the power supply.
CAUTION Do not use hypochlorite, acetone-, phenol-, or ammonia-based
cleaners.
Cleaning
Why ?Which items ?How ?
To remove
dirt and most
microbes.
Disinfecting
To kill and
remove
pathogenic
microbes.
Sterilization
To kill the
microbes and
bacterial
spores.
All parts daily.
The items to be
disinfected or
sterilized.
Accessories and
supplies in contact
with the patient’s
skin or airways.
Accessories and
supplies
penetrating the
patient’s tissue.
NOTE: Always detach accessories from the monitor and the patient before
cleaning.
• Wipe away visible organic material using a moist cloth and mild
detergent, e.g. Datex-Ohmeda Cleaning Fluid.
• Rinse as recommended by the detergent manufacturer.
• Allow to dry completely.
With heat:
• Wash in a washing machine (at least 1 min at 85°C/185°F).
With chemicals:
1. Immerse in 2% glutaraldehyde for 10 minutes (pH=6.5). After
tuberculosis, HIV or hepatitis patient, the treatment time should be 20
min.
2. Rinse carefully and let dry.
or
• Wipe with alcohol and let dry.
• Steam autoclave at maximum 121°C (250 °F) for 15 min.
• Use ethylene oxide mixture at 50 to 60°C (120 to 140°F). Follow the
sterilizer manufacturer’s recommendations for required aeration times.
• Immerse in 2% glutaraldehyde for three hours, rinse carefully and let dry.
25
Datex-Ohmeda S/5 Light Monitor
Dust filter
Clean the dust filter at least once a month:
1. Pull out the filter frame on the rear panel. Do not remove any screws!
2. Shake the filter and blow out the dust.
If the filter is damaged, replace it with a new one (p/n 886841).
Monitor casing
1. Wipe gently using a moist cloth and mild detergent.
2. Let dry completely before reconnecting the power cord.
Pulse oximetry sensors
1. Wipe the sensor with mild detergent solution.
2. Allow it to dry completely before use.
The Sat Sensors may be sterilized using ethylene oxide.
The Sat Sensors are latex free. Take possible patient allergies into account also
when selecting the cleaning agent.
With the option using Nellcor sensors, see instructions in the sensor package.
WARNING PATIENT SAFETY: A damaged sensor, or a sensor soaked in
liquid, may cause burns during electrosurgery.
ECG cables, temperature probes, NIBP hose
Wipe the cables with mild detergent solution. When necessary, use disinfectant.
Always follow the manufacturer’s instructions. Disinfect only the items that
tolerate the process.
NIBP cuff
Clean only when necessary:
1. Remove the bladder from the cuff.
26
2. Wash the bladder and the cuff in mild detergent solution. Do NOT use
alcohol.
Invasive pressure cables
1. Wipe the cables with sterile alcohol-based detergent.
2. After cleaning wipe the surfaces with a cloth dampened with sterile water.
Dry with a dry cloth.
Airway accessories
Adapters
Replace the single patient use adapter after each patient.
If the adapter is occluded, it can be rinsed with water and air dried. It may then
be replaced into the ventilator circuit of the same patient.
Multi-use adapters can be rinsed with water and steam autoclaved. The adapter
is designed to withstand 20 sterilization cycles. Before using, check that the
adapter windows are dry and residue-free and that the adapter has not been
damaged.
Operation
Make sure that all traces of alcohol or detergent are rinsed away or dried before
connecting to the patient.
Mainstream CO2 sensor
WARNING ELECTRICAL SHOCK HAZARD: Do not immerse the sensor in
liquids or autoclave it.
Wipe the sensor with a slightly damp cotton swab. When necessary, use
disinfectant.
27
Datex-Ohmeda S/5 Light Monitor
28
8. ALARMS
Alarms
In alarm:
• Messages (max. five messages) appear on the alarm message field at the
top of the display in the priority order from left to right.
• Audible alarm is heard.
• External Nurse Call signal is active, and if the monitor is connected to the
network, the alarm is passed into it.
• Yellow or red alarm indicator (LED) next to the Silence Alarms key is
illuminated according to the alarm category.
• The measured value, which activates the alarm, flashes and an
explanation may appear on the waveform, the digit field, or the alarm
message field.
Alarms are active also when the parameter is not selected on the screen unless
the alarm source is not turned off. See default settings in “Alarm tables/alarmsources” later in this chapter.
WARNING PATIENT SAFETY: Always make sure that alarm limits are set and
alarms are on when monitoring a patient.
Alarm categories
Alarms are classified into two types, red and yellow alarms. The alarm type
depends primarily upon the physiological significance and the duration of the
alarm; e.g., ‘Asystole’ advances rapidly to the red category.
The alarm messages are displayed in the order of priority from left to right.
Audio alarms follow either General or ISO pattern. This selection can be made in
Install/Service menu and is stored in permanent memory. Select the same
pattern for all the monitors in a hospital.
29
Datex-Ohmeda S/5 Light Monitor
Table 8-1Alarm priority categories
CategoryVisualAudio patternMeaning
Red alarmRed LED
If the alarm is
connected to
measured values,
this number field
is flashed.
Yellow
alarm
NoteMessage is
Yellow LED
If the alarm is
connected to
measured values,
this number field
is flashed.
displayed.
Smart start alarms
Alarms are automatically active once the patient connection cables are
attached. Some alarms have their own specific requirements before they
become active:
ISO:
3+2 beeps every
6 seconds.
--- -- --- --
Triple beep every
20 seconds.
--- ---
Single beep.
-
General:
Continuous string
of beeps.
-------------
Double beep every
5 seconds.
-- -- -- --
Single beep.
-
Life
threatening
situation.
Serious
problem.
Advisory note
or equipment
related alarm.
• Invasive pressures must be within alarm limits for 20 seconds after
zeroing.
• Apnea alarm requires at least three detected breaths.
• When apnea alarm is activated, the high and low alarms of CO
respiration will not be activated.
Silencing alarms
WARNING PATIENT SAFETY: Do not turn off audible alarms without
assuming continuous, direct observation of the patient.
Silencing alarms also disables the external Nurse Call signal.
30
and
2
Silencing alarms for two minutes
• Press Silence Alarms .
Alarms
The symbol
with a countdown timer appears on the screen, and the
message is cleared from the display. Messages of new alarms also appear
during silencing period. After two minutes, all silenced alarms and Nurse Call
function will return to normal alarm operation.
To reactivate alarms:
• Press the Silence Alarms key during the silencing period.
The alarm sounds of new alarms and the Nurse Call signal are activated.
Turning audible alarms on and off
It is possible to silence only apnea, ECG and apnea alarms, or all the alarms:
If an active alarm is suppressed, the monitor will give a reminder beep every two
minutes. (Reminder beep volume can be adjusted by selecting ReminderVolume).
If asystole alarm is silenced, and ECG is used as the heart rate source, the HR
high and low limits are also silenced.
Alarm messages
All messages are listed in chapter “Maintenance and troubleshooting.”
Alarms setup and adjustments
You can adjust alarm limits and volume, silence alarms permanently, and view
alarm status (a 10-minute trend) in Alarms Setup menu.
• Press Menu.
• Select Alarms Setup.
Adjusting alarm limits
Alarm limits can be adjusted individually (Adjust Limits) or set near to real
measured patient values (Auto Limits).
To adjust alarm limits:
• Select Adjust Limits.
32
Other settings
Alarms
• Select the desired high or low alarm limit by turning and pressing the
ComWheel.
• Adjust the alarm limit up or down by turning the wheel.
• Press the wheel to confirm the adjusted alarm limit.
• Turn the ComWheel to move the highlight either to the next alarm limit or
to the Exit box.
• Press the ComWheel.
• Press Normal Screen to return to normal monitoring.
To set auto limits
Auto limits are calculated from the measured patient values at the time when
auto limits are selected. These limits enable close patient control with individual
alarms.
• Select Auto Limits.
• Press Normal Screen.
Default Limits Sets the alarms to the default alarm limits.
Cancel Changes Returns all the values to those that were valid when the Alarms Setup menu
was entered.
Alarm SourcesNIBP and invasive pressure alarms have alternative sources, (systolic, diastolic,
mean pressure, or OFF).
Alarm VolumeThe volume of the audio alarms can be adjusted from 1 (soft) to 10 (loud).
Audio ON/OFF All alarms or asystole alone can be turned on and off.
Alarm HistoryDisplays a list of the last 10 alarms message with respective times. Press the
ComWheel to get back to Alarms Setup menu.
Remove MenuOnly the alarm limits and 10 minute trends are displayed.
Automatic alarm recording
You can record following alarm situations:
1. Asystole
2. HR high/low and Art low simultaneously (in model with invasive
pressure)
When the alarm reaches red state, the recorder will print the ECG + P1
waveforms. Printing appears 12 seconds after the alarm situation has started.
33
Datex-Ohmeda S/5 Light Monitor
2
To set alarm recording, press Menu and select
Record&Print - Recorder Setup - Start on Alarms YES/NO.
Alarm tables
Alarm sources
If alarm source is OFF, no alarms of this parameter occur and the parameter box
with the 10 minute trend is not displayed in the Alarm Setup menu.
Table 8-2Alarm source alternatives and defaults
AlarmSource alternatives
P1
P2
Art
CVP
PA
ICP
NIBP
[Default selection]
[Sys] Dia Mean OFF
Sys Dia Mean [OFF][Sys] Dia Mean OFF
Sys Dia Mean [OFF]
Sys Dia Mean [OFF]
Sys Dia Mean [OFF][Sys] Dia Mean OFF
Default limits
The monitor is delivered with a certain set of alarm limits, so called factory
default limits. These are listed for each parameter in the following tables.
It is possible to save alarm adjustments as defaults, see chapter “Monitorsetup”, submenu “Install/Service menu.”
Do not change default settings without common consent of all users. If you
change the default alarm limit, write the new limits in the following table.
Table 8-3Adjustment ranges and default settings of alarm limits
ParameterLimitLimit range
min-max
HRHigh
SpO
NIBP SysHigh
Low
High
Low
Low
35-250 bpm
30-245 bpm
51-100 %,OFF
50-100 %
15-265, OFF
OFF, 15-265
Factory defaultUser
default
160
40
OFF
90
180 (120 infant)
80(70 infant)
34
Alarms
2
2
MeanHigh
DiaHigh
Inv-P1 SysHigh
MeanHigh
DiaHigh
Inv-P2 SysHigh
MeanHigh
DiaHigh
Art SysHigh
MeanHigh
DiaHigh
CVP SysHigh
MeanHigh
DiaHigh
PA SysHigh
MeanHigh
DiaHigh
ICP SysHigh
MeanHigh
DiaHigh
EtCO
FiCO
RespHigh
TempHigh
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
High
Low
High
Low
Low
Low
15-265, OFF
OFF, 15-265
15-265, OFF
OFF, 15-265
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
-40 -300, OFF
OFF, -40-300
0.1 - 13%, OFF
OFF, 0 - 12.9%
0 - 3.0 %, OFF
OFF, 0 - 2.9%
5 - 60, OFF
OFF,4 - 59
10 - 45 °C, OFF
(50 - 113 °F, OFF)
OFF, 10 - 45 °C
(OFF, 50 - 113 °F)
140(90 infant)
60(50 infant)
100(70 infant)
40(40 infant)
180
80
140
60
100
40
20
5
15
0
10
-5
180
80
140
60
100
40
20
5
15
0
10
0
40
10
30
5
20
5
20
0
15
0
10
0
8%(60 mmHg, 8.1 kPa)
3%(23 mmHg, 3 kPa)
3%(23 mmHg, 3 kPa)
OFF
60
4
38 °C (100 °F)
15 °C (59 °F)
35
Datex-Ohmeda S/5 Light Monitor
Auto alarm limits
Auto Limits are based on the measured patient values according to the
following tables.
Table 8-4Auto alarm limits
AlarmLimitAuto limit calculation
HRHigh
Low
Pulse OximetryLowSpO2 - 5%
NIBP SysHigh
Low
MeanHigh
Low
DiaHigh
Low
Inv-P1 SysHigh
Low
MeanHigh
Low
DiaHigh
Low
Inv-P2 SysHigh
Low
MeanHigh
Low
DiaHigh
Low
Art SysHigh
Low
MeanHigh
Low
DiaHigh
Low
CVP SysHigh
Low
MeanHigh
Low
DiaHigh
Low
HR x 1.25
HR x 0.75
NIBP x 1.25 + 10
NIBP x 0.75 - 10
NIBP x 1.25 + 10
NIBP x 0.75 - 10
NIBP 1.25 +10
NIBP x 0.75 - 10
P1x 1.25 + 10
P1 x 0.75 -10
P1 x 1.25 + 10
P1 x 0.75 -10
P1 x 1.25 + 10
P1 x 0.75 -10
P2 x 1.25 + 5
P2 x 0.75 - 5
P2 x 1.25 + 5
P2 x 0.75 - 5
P2 x 1.25 + 5
P2 x 0.75 - 5
P1 x 1.25 + 10
P1 x 0.75 -10
P1 x 1.25 + 10
P1 x 0.75 -10
P1x 1.25 + 10
P1 x 0.75 -10
CVP x 1.25 + 5
CVP x 0.75 - 5
CVP x 1.25 + 5
CVP x 0.75 - 5
CVP x 1.25 + 5
CVP x 0.75 - 5
36
Alarms
PA SysHigh
MeanHigh
DiaHigh
ICP SysHigh
MeanHigh
DiaHigh
EtCO
2
FiCO
2
RespHigh
TempHigh
Low
Low
Low
Low
Low
Low
High
Low
HighOFF
Low
Low
PA x 1.25 + 10
PA x 0.75 -10
PA x 1.25 + 10
PA x 0.75 -10
PA x 1.25 + 10
PA x 0.75 -10
ICP x 1.25 + 5
ICP x 0.75 - 5
ICP x 1.25 + 5
ICP x 0.75 - 5
ICP x 1.25 + 5
ICP x 0.75 - 5
EtCO2 + 1%
EtCO2 - 1%
RR *1.25 + 2
RR*.75 -2
Temp +1 °C (1.8 °F)
Temp +1 °C (1.8 °F)
37
Datex-Ohmeda S/5 Light Monitor
9. MONITOR SETUP
The Monitor Setup menu enables you to select the information displayed on
the screen, to set the time and date and to change the default settings via the
Install/ Service menu.
Getting to Monitor Setup:
• Press Menu.
• Select Monitor Setup.
Screen setup
Select the parameters displayed in the waveform and lower digit fields.
A parameter can be displayed in one place only at the time.
The monitor turns to the default settings every time it is turned off. If you want to
save a certain screen setup as default, see Install/Service adjustments later in
this chapter.
Selecting waveforms
Four waveforms can be displayed. If one field is set to ‘OFF’, the remaining
waveforms are enlarged to fill the area.
38
• Select Waveform Fields.
• Select the field you want to rearrange.
Monitor setup
• Select the parameter from the list.
• Press Normal Screen.
Selecting Combine Pressures YES displays invasive pressures in the same field
(only in models with invasive pressure).
Selecting Thick Waves YES draws the displayed waveforms with thicker trace.
Selecting numerical display fields
There are four fields in the lower part of the screen for numeral information,
fields are numbered from left to right. If a digit field is turned OFF the digit field
to its left is automatically enlarged to fill the space.
39
Datex-Ohmeda S/5 Light Monitor
• Select Digit Fields.
• Select the field you want to rearrange.
• Select the parameter from the displayed list.
• Press Normal Screen.
Setting time & date
• Press Menu. Select Monitor Setup -Set Time and Date.
The time is shown in the upper right corner of the screen. Turning the monitor off
does not affect the clock.
You can adjust hours, minutes, seconds, day, month, and year, or select the
clock format of either 12 or 24 hours.
CAUTION Do not change the monitor’s time settings during patient
monitoring. Trend data may be lost.
NOTE: If the monitor is connected to network, the time is set by the network, and
cannot be adjusted in the monitor.
40
Monitor setup
Install/Service menu
To enter Install/Service menu, a password 16-4-34 must be given.
Service ViewService View is for a qualified technician only. It is entered with a password.
Alarm TonesSelect ISO or General pattern, see chapter “Alarms.”
Save As Defaults
Saves permanently your new settings and configurations.
The choices are Save ALL, Screen Setup, Trend Setup, Alarm Settings,
Recorder Setup and Param. Defaults.
Select Save to save settings, Cancel to cancel changes, or Factory to return to
the factory settings.
Analog outputs Select the parameter for four analog outputs, as default, they all are off.
ChannelPinSelection range
1#3 in Signal Out connectorECG or test signals
2#4 in Signal Out connectorP1 or test signals
3#5 in feature connectorfreely selectable
4#10 in feature connectorfreely selectable
Previous MenuReturn to Monitor Setup menu.
Default settings
Default settings are always on when the monitor is turned on. They are marked
in the following tables with brackets.
If you change the default settings using Save as Defaults, please write them
down in the empty column. Do not change default settings without common
consent of all users.
Table 9-1Monitor setup selections and factory defaults
Monitor settings
Parameter
Waveform Fields
Field 1OFF [ECG] P1 P2 PLETH CO2 Resp
Field 2[OFF] ECG P1 P2 PLETH CO2 Resp (*P1, **Pleth, ***P1)
Field 3OFF ECG P1 P2 [PLETH] CO2 Resp (**CO2)
Field 4[OFF] ECG P1 P2 PLETH CO2 Resp (***CO2)
Combine PressuresYES [NO]
Thick Waves[YES] NO
Digit Fields
Lower Field 1OFF HR P1 P2 [NIBP] SpO2 CO2 Resp Temp
Alternatives [Default selection]
User default
41
Datex-Ohmeda S/5 Light Monitor
Lower Field 2[OFF] HR P1 P2 NIBP SpO2 CO2 Resp Temp(*P2, ***P2)
Lower Field 3OFF HR P1 P2 NIBP SpO2 CO2 Resp [Temp]
Lower Field 4OFF HR P1 P2 NIBP SpO2 CO2 [Resp] Temp (**OFF)
Set Time and Date
Clock Format12 [24]
Parameter settings
ECG
ECG LeadI [II] III
Size[1.0] 0.2-5.0
Beat Sound Volume[0] 0-10
ECG Setup
Sweep Speed12.5 [25] 50
HR Source[AUTO] ECG Pleth
GridON [OFF]
PacemakerShow [Hide] On R
Start on AlarmsYES [NO]
Waveform Delay[12 s] OFF
Trend Resolution1 [5] 10 30
Paper Speed6.25 12.5 [25]
Recording Length[30 s] Cont
Alternatives [Default selection]
Monitor setup
User default
Display Trends
Parameter
Trend Setup / Trend Page 1
Field 1Real-time ECG in all models
Field 2OFF ECG [HR] P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp
Field 3OFF ECG HR P1 P2 [NIBP] SpO2 ****Pleth CO2 Resp Temp
Field 4OFF ECG HR P1 P2 NIBP [SpO2] ****Pleth CO2 Resp Temp
Field 5OFF ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 [Resp] Temp (**CO2,
Trend Setup / Trend Page 2
Field 1Real-time ECG in all models
Field 2OFF ECG [HR] P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp
Field 3OFF ECG HR P1 P2 NIBP [SpO2] ****Pleth CO2 Resp Temp (*P1,
Field 4OFF ECG HR P1 P2 NIBP SpO2 [****Pleth] CO2 Resp Temp (*P2,
Field 5OFF ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp [Temp]
Alternatives [Default selection]
***CO2)
(**SpO2)
**Pleth, ***P1)
**CO2, ***P2)
(**Resp)
43
User default
Datex-Ohmeda S/5 Light Monitor
Trend Setup / Trend Page 3
Field 1Real-time ECG in all models
Field 2[OFF] ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp (*HR,
Field 3[OFF] ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp (*SpO2,
Field 4[OFF] ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp (*Pleth,
Field 5[OFF] ECG HR P1 P2 NIBP SpO2 ****Pleth CO2 Resp Temp
**HR, ***SpO2)
**SpO2, ***Pleth)
**Temp, ***CO2)
(*Temp,***Resp)
*) default in models with invasive pressure,
**) default in models with CO2,
***) default in models with invasive pressure and CO
2
****) not available with the Nellcor option, N-LNSAT
44
10. TRENDS
The monitor accumulates and displays trend data for user selected periods
between 2 and 24 hours.
There are two types of trend screens: graphic trend displays measured data in
bars or dots, and numerical trends give the information in tabular form.
In addition a 10-minute trend is displayed in the Alarm Setup menu for every
parameter with alarm limits.
During trend display, the monitor is active and the numerical information is
continuously updated and displayed. Trend data is stored in memory for 15
minutes after the monitor is turned off.
Trends
45
Datex-Ohmeda S/5 Light Monitor
Viewing graphic trends
• Press Menu.
• Select Trends.
• Select Graph 2 h, 4 h, 12 h or 24 h.
1
2
Figure 10-1Graphic trend
3
4
5
(1) List of selections
(2) Labels
(3) Real time ECG above the trend information
(4) Parameter trend fields
(5) Movable trend cursor
Graphic trend pages
There are three pages of graphic trends. Pages are turned by selecting Next
Graph. The page that is viewed last is stored in memory. It appears first when
*) not available with option N-LNSAT
Parameters displayed on each page can be selected in Trend Setup.
Resolution
Resolution of trend display depends on the trend length selected.
Table 10-2Trend lengths and resolutions
Trend length (hours)Resolution (minutes)
21
42
126
2412
47
Datex-Ohmeda S/5 Light Monitor
Viewing numerical trend
You may view trend data also in numerical format. The trend information is
updated every five minutes and always after NIBP measurement. The latest
information is displayed at the bottom of the page.
• Select Numerical.
Figure 10-2Numerical trend page
(1) List of selections
(2) Real time ECG above the trend information
(3) Numeric trend information
(4) Movable cursor
Parameters displayed are:
HR; SpO
; NIBP sys/dia; P1/Art sys/dia; P1/CVP mean; CO2 Et/Fi; RR
2
48
Changing trend setup
• Select Trend Setup in Trends menu.
• Select the graphic trend page you want to change.
• Select the parameter for each field, or select OFF.
Trends
If you want to make your trend setup permanent you can store it, see chapter
“Monitor setup/Default settings.”
Erasing trend data
• Press Menu.
• Go to Patient Data.
• Select Reset Trends.
• Select YES to confirm resetting trends.
49
Datex-Ohmeda S/5 Light Monitor
11. PATIENT DATA
Patient Data menu is used for identifying the patient case and erasing trends
and settings.
To stop a monitoring case, reset trends between patients. Alternatively you can
reset all data and settings. If the trends are just being printed, they cannot be
erased. A message ‘Resetting not possible during printing. Please wait until
printing is completed.’ appears.
The Patient Data menu will open automatically when the monitor is started less
than 15 minutes after previous case, and the trends have not been erased. This
is to confirm that the collected data is connected to the right patient.
• Press Menu.
• Select Patient Data.
Last NameGive patients last name, max. 14 characters, with ComWheel.
First NameGive patients first name, max. 14 characters, with ComWheel.
Patient IDGive patients ID, max. 14 characters, with ComWheel. This starts the case, the
message ‘Case started’ is displayed.
Reset TrendsErase trend memory between patients. Select YES to confirm the erasing, the
message ‘Case ended’ is displayed.
Reset ALLReturn all the settings to their defaults and erase trend and alarm memory.
Select YES to confirm the erasing, the message ‘Case ended’ is displayed.
Main MenuReturn to Main Menu.
50
Patient data
The Last Name, First Name and Patient ID can be entered also from the DatexOhmeda Central via network. The name is displayed as long as there are less
than five alarm messages displayed simultaneously.
51
Datex-Ohmeda S/5 Light Monitor
12. RECORDING AND PRINTING
The optional built-in recorder provides single, dual and triple real time waveform
recording, and single and dual recordings of numerical information. In addition
it can record up to 24 hours of graphic and numerical trend data.
NOTE: Printings on thermal paper may be destroyed when exposed to light,
heat, alcohol etc. Take a photocopy for archives.
Graphical trend data can also be printed on the laser printer either from the
local port, or, if the monitor is connected to network, using a central printer.
Recording
• Press Recorder Start/Stop to start and stop recording.
NOTE: Scaling of recorded waveforms follows the displayed parameter scaling,
when applicable.
Selecting the waveforms to be recorded
• Select up to three waveforms you wish to record in
Menu - Record/Print - Wave Setup menu.
Waveform xSelect the recorded waveforms.
52
Recording and printing
Recording trends
• Press Menu and select Record/Print.
• Start recording by selecting Record Graph. Trend, or Record Num.
Trend, or Record Tab. Trend.
• Stop printing by selecting Stop Graph./Num./Tab. Trend.
To select the graphical trend you want to record, go to Menu - Record/Print -
Graph - Trend Setup.
Other recorder settings
To make the following selections, go to Menu - Record/Print - Recorder
Setup:
Start on Alarms Select if an alarm activates the recorder or not (YES / NO).
Recording time is 30 seconds: 12 seconds recording from the recorder
memory, 18 seconds real time recording.
Alarm source annotated.
Waveform Delay Select the amount of waveform data (in seconds) that will be stored in the
recorder memory and printed before real-time waveform. The choices are: OFF
or 12 seconds.
Paper SpeedSelect the paper speed 6.25, 12.5, or 25 mm/seconds.
Recording Length
Select the printing time, either 30 seconds or continuous.
Previous Menu Return to Record&Print menu.
53
Datex-Ohmeda S/5 Light Monitor
Recorder paper replacement
Figure 12-1Placing paper roll
Printing
Graphical trend pages P1, P2, and P3 can be printed either separately or all
together. Print with a local laser printer, or if the monitor is connected to the
network, with a central laser printer.
To print:
• Press Menu.
• Select Record/Print.
• Select Print Graphical.
• Select Print Graphs.
• Press Menu.
• Select Trends.
• Select Print Page.
Selecting the printer
• Press Menu.
• Select Record/Print.
• Select Print Graphical.
or
54
• Select Printer Location.
• Select Local for local printing or an appropriate network alternative.
Network printers are named in your Datex-Ohmeda Central.
Selecting the paper size
• Press Menu.
• Select Record/Print.
• Select Print Graphical.
• Select Printer Paper Size and the right one of A4 or Letter size.
Printer installation
A laser printer can be connected to the monitor’s serial connector and used as
the local printer. The printer must be able to use PCL5, and a special converter
cable (p/n 894193) is required when using the parallel connection. The printers
with serial connection can be connected directly to the monitor with a standard
cable.
Recording and printing
Purchase the printer locally. The printer memory should be at least 2 MB.
WARNING ELECTRIC SHOCK HAZARD: The printer must be supplied from an
appropriate extra separating transformer.
WARNING ELECTRIC SHOCK HAZARD: Before starting to use the system,
ensure that the entire combination complies with the
international standard IEC 60601-1-1 and with the
requirements of the local authorities.
Printer setup
For instructions refer to the printer’s documentation. Make sure that the printer
settings are set to serial interface.
55
Datex-Ohmeda S/5 Light Monitor
13. NETWORK AND DATACARD
The monitor can be equipped with following options to enhance data continuum
capabilities in the Datex-Ohmeda S/5 information system:
N-LNETNetwork
N-LDATADataCard
N-LDNETDataCard and network.
Networking
If the monitor is connected to the network, the symbol is displayed:
When the monitor is connected to network
• Time and date are set by the network and cannot be adjusted in the
monitor.
• The patients name can be entered in the Patient Data menu to be
displayed on the Datex-Ohmeda Central’s and the monitor’s screen and
printed on the laser printouts.
• Measured and entered patient data can be viewed on the Datex-Ohmeda
Central.
• Measured patient data and alarms can be viewed on other
Datex-Ohmeda S/5 monitoring systems, but not vice versa.
Patient view and multi view on the Datex-Ohmeda Central screen
The Patient View on the Datex-Ohmeda Central screen looks like monitor’s own
screen.
In the Multi View, the measured values from each of the monitor shown on the
screen are:
8-Patient mode16-Patient mode
ECG waveformECG waveform
HR, P1, SpO2, RR or
EtCO2numbers
Alarm fieldAlarm field
HR numbers
56
Monitor-to-Monitor communication
When the monitor is viewed on another S/5 monitor, the following parameters
are displayed:
Network and DataCard
WaveformsEGC, Pleth, Art, CO
NumbersHR, SpO2, Art, CO
1-hour numerical trendHR, SpO2, NIBP, P1, P2, CO
You cannot view other monitors on the S/5 Light Monitor.
DataCard
The DataCard is a rewritable flash memory card. With the DataCard, the patient
trend data collected with the S/5 Light Monitor can be transferred to other S/5
monitors with record keeping or intensive care softwares, as well as to S/5
monitors with ARK software and S/5 monitors.
Data from any card cannot be loaded to the S/5 Light Monitor, nor loaded from
the network to the S/5 Light Monitor or vice versa.
Inserting and removing the card
2
2
2
l
l
l
• Push the card into the slot on the left side panel of the monitor.
When the DataCard is inserted, this symbol is displayed. If the symbol is
blinking, the card is not inserted properly.
If any other card or a faulty card is inserted, a message ‘Faulty DataCard change card’ is displayed.
57
Datex-Ohmeda S/5 Light Monitor
To remove the card:
• Push the lever on the side of the slot.
Collecting data
When the card is inserted, the continuous patient trend data is automatically
stored on it. When the card is filled, the oldest data is erased, but only if it is
older than two days. You can retrieve information of approximately the last 24
hours. The memory capacity depends on the complexity of the saved data.
58
14. ECG
ECG monitoring is used to: measure the heart rate, detect arrhythmias,
monitor pacemaker function, and myocardial ischaemia.
ECG reflects the electrical activity generated by the heart muscle.
You can use 5- and 3-lead sets and trunk cables, but only leads I, II, or III can
be measured.
Monitoring ECG
1.Connect the patient cable to the blue ECG
2.Place the electrodes on the patient.
ECG
connector on the monitor’s side panel.
Connect the leadwires to the electrodes.
NOTE: Always attach all the electrodes to the
patient, also when using a 5-lead cable.
3.The label of the lead is displayed in the ECG field
Applying electrodes
WARNING PATIENT SAFETY:
Ensure proper contact of the electrosurgery return electrode to
avoid possible burns on the patient via ECG electrodes and
probes.
Make sure that no part of the patient connections touches any
electrically conductive material, including earth.
Patient preparation
• Ensure good skin contact with the electrodes. Remove excessive body
hair, skin oil etc.
• Check, that the pre-gelled electrodes are moist and have not dried out
during storage.
and above the HR numerics.
59
Datex-Ohmeda S/5 Light Monitor
Placing electrodes
Avoid bones close to the skin, obvious layers of fat and major muscles.
Figure 14-1Lead selections in typical electrode positioning
Electrode coding
Table 14-1IEC and AAMI codes for ECG electrodes
3-LEAD IEC STANDARD5-LEAD IEC STANDARD
L = yellow (left arm)
F = green (foot)
R = red (right arm)
RED (IEC )
WHITE ( AAMI)
RED (IEC)
WHITE (AAMI)
LEAD I
GREEN (IEC)
RED ( AAMI)
YELLOW (IEC)
BLACK (AAMI)
LEAD I
LEAD II
YELLOW (IEC)
BLACK (AAMI)
LEAD III
GREEN (IEC)
RED (AAMI)
Left: Standard 3-Lead Positioning
Middle: Modified CB5 3-Lead Positioning
Right: Standard 5-Lead Positioning, place the fifth electrode in
one of the six places indicated.
R = red (right arm)
N = black (neutral)
L = yellow (left arm)
F = green (foot)
C = white (chest)
3-LEAD AAMI STANDARD5-LEAD AAMI STANDARD
LA = black
LL = red
RA = white
60
RA = white
RL = green
LA = black
LL = red
V = brown
Selecting trunk cable
Use a 5-lead set with a 5-lead trunk cable, and accordingly a 3-lead set with
3-lead trunk cable. (If you have a 3-lead set connected to a 5-lead trunk cable,
you must set in the ECG Setup menu the selection 5 Lead Cable to 3elect).
What should happen when electrodes are applied ?
• ECG display appears.
• A beep is heard when a QRS complex is detected.
Selecting ECG lead
• Press the Menu key and select ECG & Resp.
• Select ECG Lead.
• Select one of I, II, or III.
NOTE: Only 3 leads are measured, though a 5-lead set and cable are used.
NOTE: To minimize the interference from the electrosurgery unit (ESU), place the
electrodes and select the lead so that the current from the surgical burning area
to the return electrode does not go through the measured ECG lead.
ECG
ECG display
2
1
4
3
Figure 14-2Display of ECG and HR
(1) Selected lead label
(2) 1 mV reference bar
(3) Heart rate (HR) label
(4) Heart rate calculation source and message field for HR messages
61
Datex-Ohmeda S/5 Light Monitor
Cascaded ECG
If two fields are configured (See “Monitor setup/Waveform fields”) for ECG
display the waveform continues from one field to another. Thus more QRS
complexes are displayed at the same time.
ECG adjustments
• Press the Menu key.
• Select ECG & Resp.
ECG LeadSelect the displayed lead: I, II, or III.
SizeAdjust from 0.2 to 5 in steps of 0.1.
The bar on the left side of the ECG display is 1 mV.
Beat Sound Volume
A beat (QRS complex, or a plethysmographic or pressure pulse) produces a
beep, and its volume is adjustable from 0 to 10 in steps of 1. Adjusting also
changes the beat sound volume of SpO2.
(When SpO2 is monitored, the tone of the pulse beep rises with increasing
oxygen saturation and falls as saturation decreases.)
ECG SetupOpens a submenu for ECG settings, see below.
HR AlarmOpens a submenu for adjusting heart rate alarm, see chapter “Alarms.”
Resp SetupOpens a submenu for Respiration settings, see chapter “Respiration.”
Resp Rate Alarm Opens a submenu for adjusting respiration rate alarm, see chapter “Alarms.”
62
ECG
ECG setup
Sweep Speed Set the speed of the waveform sweep. Speeds are: 12.5, 25, and 50 mm/s on
a 9” screen.
The 12.5 mm/sec speed is suggested for monitoring patients with a slow HR to
show more QRS complexes per sweep.
The 50 mm/sec speed is recommended when monitoring patients with a high
HR or when using cascaded ECG.
NOTE: The sweep speed is same for ECG, plethysmographic, and invasive
pressure waveforms.
HR Source Select the heart rate source. If ECG signal is too noisy for reliable heart rate
calculation, heart rate can be calculated from pressure (Art) or pulse oximetry
(Pleth). The heart rate source appears above the numerical display of the heart
rate.
The AUTO selects the first available of ECG, Art, and Pleth.
5 Lead CableIf you use a 5-lead trunk cable with a 3-lead set, select 3elect on, otherwise the
‘Leads Off’ message will be displayed.
GridSelect ON to display ECG waveform with 0.5 mV grid lines.
PacemakerSelects how to display the pacing pulse of a cardiac pacemaker, Show, Hide or
ON R.
Hide: The pacing pulse is filtered from the waveform.
Show: As above, but the pulse is marked with a constant height marker.
63
Datex-Ohmeda S/5 Light Monitor
ON R: Pacing pulses are not filtered from the waveform. This improves ECG
monitoring with A-V pacemaker patients, as the QRS complexes are counted
even if the pacing pulse hits the complex. On the other hand, during asystole
the monitor may count pacing pulses as heart rate.
Setting heart rate alarm
Heart rate alarms can be adjusted in the Alarms Setup menu or accessed
through the ECG & Resp. menu. See more detailed instructions in the “Alarms”
chapter.
Alarm messages are explained in chapter “Maintenance and troubleshooting;
Alarm and information messages.”
Pacemaker patients
The monitor detects and rejects pacemaker pulses, see selection Pacemaker in
the ECG Setup menu. Sometimes this may lead to unnecessary asystole
alarms. The pacemaker may well change the shape of QRS so much that QRS
detection may be affected.
WARNING PATIENT SAFETY: Monitor may count pacemaker pulses as heart
rate during cardiac arrest, some arrhythmias, or with certain
types of pacemakers, particularly in ON R mode. Do not rely
entirely upon alarms. Keep pacemaker patients under close
surveillance.
64
Respiration
15. RESPIRATION
Respiration measurement uses ECG electrodes to measure the impedance
changes caused by breathing.
The measurement is intended for patients three years old and up.
WARNING PACEMAKER PATIENTS: The impedance respiration
measurement may cause rate changes in Minute Ventilation
Rate Responsive Pacemakers. Set the pacemaker rate
responsive mode off or turn off the impedance respiration
measurement on the monitor.
Monitoring respiration
Respiration measurement is activated when a digit or waveform field is selected
for it:
• Press the Menu key.
• Select Monitor Setup.
• Select Waveform Fields or Digit Fields.
• Select Resp. to be displayed on an appropriate field.
Figure 15-1Respiration display
Patient connections
Respiration uses the same electrodes as ECG, refer to the “ECG” section of this
manual.
Respiration setup
• Press the Menu key.
• Select ECG & Resp.
65
Datex-Ohmeda S/5 Light Monitor
• Select Resp Setup.
SizeAdjust the scale height 1...50 mm/Ohm. The bar on the left displays always 1
Ohm.
Sweep Speed Set the speed of the waveform sweep. Speeds are: Fast 6.25 and Slow 0.625
mm/s on a 9” screen.
Resp. Rate Source
Select the respiration rate source of AUTO, CO2 and Imped. The AUTO sets the
CO2 first, and if it is not available, impedance respiration is used. Default
selection is AUTO.
Measurement Select ON to measure impedance respiration and OFF to set the measurement
off. Default selection is ON.
Detection Limit Adjusts measurement sensitivity. Selections are AUTO, 20%, 40%, 60%, or
100%. The percentage is the ratio to the 1 Ohm reference bar which is 100%.
Limitations of measuring
Movement artifacts
Changing the patient position, moving the head, moving the arms or shaking
will result in movement artifacts. Also the heart may cause noticeable
movement, and sometimes this may interfere with the respiration
measurement.
66
Respiration
Electrical interference
Electrical devices that emit electromagnetic disturbance will result in artifacts or
disable the respiration measurement completely. Such devices are
electrosurgery units and infrared heaters, among others.
Intermittent mechanical ventilation
During spontaneous breathing the ventilator may at times support the patient’s
ventilation with an extra inspiration. If these ventilator inspirations are
substantially larger than the spontaneous breaths, the respiration calculation
may mistakenly count only the ventilator produced inspirations and expirations.
You can correct this by manually adjusting the detection limits.
1. Press the Menu key.
2. Select ECG & Resp.
3. Select Resp Setup.
4. Select Detection limit.
Measuring principle
ECG electrodes are used to measure impedance change across the thorax.
Impedance change is caused by movements and changing air volume in the
lungs.
Lead selection
using 3-lead
cables
IL and RF*
IIF* and RL
IIIF* and LR
Using 5-lead cablesL and FN
*) may be neutral (N) in some older 3-lead sets.
Signal fed
between
electrodes
Signal measured
67
Datex-Ohmeda S/5 Light Monitor
16. PULSE OXIMETRY
Pulse oximetry is used for monitoring oxygen saturation (SpO2), pulse rate,
and plethysmographic waveform.
The plethysmographic pulse waveform reflects the changes in blood
pulsation at the measuring site. The amplitude of the waveform reflects the
strength of perfusion, except when using the Nellcor option, N-LNSAT.
Pulse oximetry is measured with a sensor from finger, toe, ear, or with little
children, from palm or feet.
Monitoring pulse oximetry
1.Attach the sensor to the gray connector on the
2.Attach the sensor to the patient. The SpO2 and
monitor’s side panel.
pulse rate values, and the plethysmographic
pulsewave appear on the screen.
3.Change the sensor site frequently.
4.Remove the sensor from the patient and clean it.
68
Selecting sensor
Pulse oximetry
Reusable Sensors
OXY-F4-N
OXY-F-UN
OXY-F-DB
OxyTip+ Adhesive Sensors
OXY-AP
OXY-AF
Figure 16-1Sat Sensors for different applications
NOTE: These sensors cannot be used when using the Nellcor option, N-LNSAT.
Please refer to chapter “Alternative SpO
measurement.”
2
Applying sensors
• Use dry and clean sensors only.
• Confirm that red light lights up in the sensor when it is connected to the
monitor.
• Clean the application site.
Remove nail polish, artificial fingernail, earrings etc.
Clip long finger nails.
• Attach the sensor cable to the wrist or bed clothes to prevent cable and
sensor from moving.
WARNING PATIENT SAFETY: Change measuring site frequently.
69
Datex-Ohmeda S/5 Light Monitor
• Change sensor site and check skin and circulatory status every 2-4 hours
with adults, and every hour with little children.
What should happen when sensor is applied ?
• The message ‘Pulse Search’ appears.
• When ‘Pulse Search’ is completed, the pulse rate and SpO
appear.
• The plethysmographic waveform will appear. The waveform scale is
established according to the measured amplitude.
• Pulse beep is heard. The tone of the pulse beep will rise with increasing
oxygen saturation and fall as saturation decreases.
Pulse oximetry display
value will
2
1
Figure 16-2Plethysmographic waveform with oxygen saturation number
(1) Bar reflecting the signal amplitude, not available with the Nellcor option,
The waveform is a good indicator of measurement validity. If the wave is noisy,
amplitude is low, or the normal waveform cannot be seen, attach the sensor
again.
Dyshemoglobins
2
3
The pulse oximeter cannot distinguish between oxyhemoglobin and
dyshemoglobins. Thus, the SpO2 values may be erroneously high in smokers or
in patients who have burns or carbon monoxide (CO) intoxication.
70
Adjustments
Pulse oximetry
Electrosurgery
Under some circumstances electrosurgery may cause noise on the display.
Therefore, be careful in interpreting the results, especially the plethysmographic
pulse waveform, during electrosurgery.
External interference
Ambient light, intravascular dyes and vasoconstrictive drugs may affect the
accuracy of the measurement. Improper sensor attachment (measuring site too
thick or thin, sensor ends not aligned), or patient movement may also interfere
with the measurement.
• Press the Menu key.
• Select SpO2 & Temp.
• Select and adjust by turning and pressing the ComWheel.
Beat Sound Volume
Adjust beat volume from 0 to 10.
ScaleSelect Scale from 50, 20, 10, 5, 2 or use AUTO which automatically adjusts
the scale according to the measured waveform.
NOTE: The selection is not available with the Nellcor option, N-LNSAT.
Sweep SpeedSet the speed of the display sweep for the waveform display. The speeds are
12.5, 25, or 50 mm/sec on a 9" screen.
NOTE: The sweep speed is same for ECG, plethysmographic, and invasive
pressure waveforms.
71
Datex-Ohmeda S/5 Light Monitor
HR Source Select the heart rate source. If ECG signal is too noisy for reliable heart rate
calculation, heart rate can be calculated from pressure (Art) or pulse oximetry
(Pleth). The heart rate source appears above the numerical display of the heart
rate.
The AUTO (default) selects the source in the following order: ECG, pressure, and
plethysmographic pulse waveform.
SpO2 Response Select how fast the SpO2 value follows the measurement. The selections are:
Beat-to-beat, Normal (10 s), or Slow (20 s, default setting). The selections with
the Nellcor option, N-LNSAT, are Fast, Normal, and Slow.
SpO2 AlarmsOpens a submenu for adjusting alarms, see the “Alarms” chapter.
Temp SetupOpens a submenu for adjusting temperature settings, see the “Temperature”
chapter.
Main MenuReturn to Main Menu.
Setting SpO2 alarms
The alarms can be adjusted in the Alarms Setup menu, accessed through the
SpO2 & Temp menu. See more detailed instructions in the chapter “Alarms.”
Alarm messages are explained in the chapter “Maintenance and
troubleshooting; Alarm and information messages.”
Measuring principle
Oxygen saturation, SpO2, expressed as a percent, defines the amount of oxygen
carried compared to total capacity (also called functional or in vivo oxygen
saturation). It is measured by a two wavelength pulse oximeter.
The SpO2 value is measured by light absorption technique. Red and infrared
light (660 nm and 910 nm) is emitted from the light source. The amount of
transmitted light is detected in the detector half. When the pulsative part of the
light signal is examined, the amount of light absorbed by arterial hemoglobins is
resulted, and the saturation level can be calculated.
The saturation percentage is displayed numerically, and the pulse beep sound
varies according to the saturation level.
Plethysmographic pulse wave
The plethysmographic pulse wave is derived from variations of the intensity of
the transmitted light and reflects the blood pulsation at the measuring site.
Pulse rate
Pulse rate is calculated from the plethysmographic pulse wave.
72
Alternative SpO2 measurement
As an alternative to the Datex-Ohmeda pulse oximetry, the monitor can be
equipped with the Nellcor compatible saturation option, N-LNSAT. With this
option you can use the sensors listed below.
NOTE: When the monitor has the N-LNSAT option, the monitor is equipped with
the Nellcor measurement only and cannot use Datex-Ohmeda Sat Sensors.
Plethysmographic scale
You cannot adjust the scaling when using the N-LNSAT option; it is
automatically set by the monitor.
Nellcor sensors
Use one of the following Nellcor sensors with the N-LNSAT module:
One temperature value can be measured from various sites using YSI-400
probes, or equivalents.
Monitoring temperature
• Attach the probe to the patient.
• Connect the temperature probe to the connector on the right side of the
monitor.
• The measured value appears on the digit field.
Temperature display
Example:
2
1
Figure 17-1Temperature display
(1) Label
(2) Temperature value
Temperature testing
The monitor automatically tests the temperature measuring function at the
start-up and after that every 10 minutes. If the test fails contact your local
service representative.
74
F
Temperature setup
• Press Menu.
• Select SpO2 & Temp.
• Select Temp Setup&Alarm.
Temperature
°C
°
UnitSelects unit, °C or °F.
Temp Label Select the label:
AirWairway temperature
Axilaxillary temperature
Bladbladder temperature
Esoesophageal temperature
Myomyocardial temperature
Nasonasopharyngeal temperature
Rectrectal temperature
Roomroom temperature
Skinskin temperature
Temptemperature
Tymptympanic temperature
Corecentral temperature
Surfsurface temperature
Previous Menu Return to SpO2 & Temp menu.
75
Datex-Ohmeda S/5 Light Monitor
18. NON-INVASIVE BLOOD PRESSURE (NIBP)
The monitor can measure NIBP automatically after set intervals, or take single
measurements, or measure continuously for five minutes in STAT mode.
WARNING PATIENT SAFETY: The monitor sets the inflation pressure
according to the first measurement. Reset the monitor between
patients to reset the inflation limit.
CAUTION Vibrations during transport may disturb the NIBP measurement.
Monitoring NIBP
• Select the appropriate cuff and hose.
Table 18-1NIBP cuff sizes
CuffColorWidth
cminch
Large adult
Standard adult
Small adult
Child
Infant
Disposable infant cuff #5
NOTE: Always use the white hose when measuring little children. Make also sure
the Inflation Limits selection is set to 150 mmHg or Auto.
• Attach the cuff to the patient and the hose to the NIBP connector located
on the right side panel of the monitor.
• Make sure that NIBP tubes are not bent, pressed or stretched.
Measurement may be impaired.
• Press the NIBP key to open the NIBP mini menu.
• Select Start Cycling or Start Manual and press the ComWheel.
• You can also adjust the cycling time by selecting Set Cycle Time. Adjust
the time by turning and pressing the ComWheel.
• Observe the cuffed limb frequently. The measurement may impair blood
circulation.
Display of NIBP measurement
E
N
I
L
X
E
D
N
I
Use smaller cuff
Correct size
Use large cuff
1
2
3
4
(1) Systolic pressure
(2) Diastolic pressure
(3) Field shows the selected measurement (Manual, STAT), or the time since
last autocycle measurement
77
Datex-Ohmeda S/5 Light Monitor
(4) Mean pressure value of non-invasive blood pressure
At the beginning of the measurement the cuff pressure is displayed in the mean
pressure value field.
If motion artifacts are detected, the monitor automatically retains deflation until
the motion stops (maximum of 30 seconds). If the artifacts interrupt proper
measurement, a new measurement is automatically activated.
When the measurement is ready, a short beep is heard and the result starts
flashing.
Automatic NIBP double check
If the NIBP value exceeds the alarm limits, a new measurement is taken
automatically (immediately, with MANUAL measurements, and after 30
seconds, with AUTO measurements). If an alarm situation persists, an alarm
signal is given out.
Using NIBP functions
To use NIBP functions:
• Press the NIBP key to open the NIBP mini menu
or
• Press Menu and select NIBP & Inv. Press to open the
NIBP & Inv. Pressure menu.
78
Non-invasive blood pressure
Starting/Stopping Autocycle, Starting a Manual Measurement and Canceling
Any Measurement are available through both menus. STAT mode and Venous
Stasis are available through the NIBP & Inv. Pressure menu only.
Starting/Stopping autocycle
• Select Start Cycling/Stop Cycling to start/stop automatic NIBP
measurement at selected intervals.
The bar at the bottom of the NIBP display shows the time remaining to the next
measurement.
Autocycling is synchronized to actual time so that if the first measurement
happens at 12.02, the next measurement is done at 12.05 and again at 12.10
(5 min intervals).
Starting a manual measurement
• Select Start/Stop Manual.
Canceling any measurement
• Press the NIBP key,
or
• select the corresponding menu item.
79
Datex-Ohmeda S/5 Light Monitor
Starting a continuous measurement (STAT)
You can start the continuous measurement for five minutes (STAT mode)
through the NIBP & Inv. Pressure menu only.
In STAT mode the early systolic value is measured and displayed until the final
result is available.
After five minutes the monitor automatically returns to the previously selected
cycling interval or to manual mode.
To start the measurement:
• Press the Menu key and select NIBP & Inv. Press.
• Select Start STAT.
Using venous stasis
You can hold a constant pressure in the cuff to help venous cannulation
(Venous Stasis). This function is only available through the NIBP & Inv.Pressure menu. The pressure is maintained in the cuff for a certain time
according to the inflation limits.
Table 18-2Inflation limits and venous stasis times
Inflation limitVenous stasis pressureVenous stasis time
150 mmHg40 mmHg1 min
200 mmHg60 mmHg2 min
280 mmHg80 mmHg2 min
The pressurization time is displayed in the NIBP field and during the last 15
seconds the word ‘STASIS’ is flashed.
• Press the Menu key.
• Select NIBP & Inv. Press. and Start Ven. Stasis.
To release the pressure before the time is expired:
• Press the NIBP key.
80
Non-invasive blood pressure
NIBP adjustments
Set Cycle TimeSelect desired interval time of 1, 2.5, 3, 5, 10, 15, 30, or 60 minutes.
Ready Prompt Vol.
Adjust the volume of the beep that informs when the NIBP measurement is
ready. The range is from 1 (soft) to 10 (loud), or 0 (OFF).
Inflation Limits Select the inflation pressure limit: 280, 200, or 150 mmHg.
Calibration Check
Start the pressure calibration check with an external manometer. See
instructions below.
NIBP AlarmOpens a submenu for adjusting the NIBP alarms, see the “Alarms” chapter.
Previous Menu Return to the NIBP & Inv. Pressure menu.
Calibration check
• Remove the cuff from the connector before entering the menu.
• Select Calibration Check ON in NIBP Setup menu.
• When the calibration menu appears, attach an external manometer with
pump to the connector.
• Pump ca. 200 mmHg and compare the readings of the manometer and
display. If the difference is greater than 4 mmHg, the calibration must be
checked by an authorized service person.
81
Datex-Ohmeda S/5 Light Monitor
NIBP alarms
NIBP Alarms are set in the Alarm Setup menu, see the “Alarms” chapter. The
alarm messages are explained in chapter “Maintenance and troubleshooting.”
Measuring principle
The non-invasive blood pressure (NIBP) measurement uses the oscillometric
measuring principle. The cuff is inflated with a pressure slightly higher than the
presumed systolic pressure, then slowly deflated at a speed based on the
patient’s heart rate, collecting data from the oscillations produced by the
pulsating artery. Based on this data, the unit calculates values for systolic,
mean and diastolic pressures.
82
Invasive blood pressure
19. INVASIVE BLOOD PRESSURE
Two invasive blood pressures can be measured with the model F-LMP1.
WARNING PATIENT SAFETY:
All invasive procedures involve patient risks. Use the aseptic
technique. Follow catheter manufacturer’s instructions.
WARNING Ensure proper contact of the return electrode of the
electrosurgery unit to your patient to avoid possible burns at
sensor sites.
WARNING Make sure that no part of the patient connections touches any
electrically conductive material including earth.
WARNING Use only defibrillator proof transducers and cables.
Monitoring invasive pressures
Preparing pressure line
• Prepare the transducer kit according to transducer manufacturer’s
instructions. Ensure there is no air in the line.
• Connect the kit to the transducer cable and the cable to the red
connector in the monitor.
• Mount the transducer kit at heart level.
• Connect the patient catheter to the pressure line.
• Zero the transducer.
• Open the line to the patient.
83
Datex-Ohmeda S/5 Light Monitor
1
2
Reusable
Disposable
3
4
Figure 19-1Invasive pressure setup for disposable and reusable
(1) Heparinized fluid bag with pressure infusor
(2) Flushing set
(3) Transducer
(4) Adapter cable for using disposable transducers
Zeroing transducer
Always zero the pressure transducer before monitoring and check zero level
after power interruptions.
To zero the transducer:
• Open the transducer to air.
• Press the Zero ALL key,
Reusable
transducers
or
select Zero Px in the NIBP & Inv. Pressure menu.
The messages ‘Zeroing’ and ‘Zeroed’ are displayed accordingly. After
the channel is zeroed, the zeroing time appears in the menu.
If the zeroing fails, the previous zero is used.
84
Invasive blood pressure
• Open the transducer to the patient.
NOTE: ICP can be zeroed via the NIBP & Inv. Pressure menu only.
CAUTION Mechanical shock to pressure transducer may change zero balance
and calibration.
Invasive pressure display
After the line is zeroed and opened to the patient, the pressure wave appears.
If two pressures are measured they can be combined with individual scales.
(Select separate or combined waveforms in the Monitor Setup/ WaveformFields menu, see chapter “Monitor setup”).
2
3
1
Figure 19-2Combined pressure waveforms
(1) Invasive blood pressure waveforms with zero and reference lines
(2) Selected pressure label
(3) Selected pressure scale
(4) Systolic, diastolic and mean pressures
(5) Ventilation mode selected in the Px Setup&Alarm menu
3
2
2
2
4
5
85
Datex-Ohmeda S/5 Light Monitor
Adjustments
• Press Menu.
• Select NIBP&Inv. Press.
• Select Px Setup&Alarm.
Labeling the invasive line
The label of the pressure channel sets its scale, alarm source and alarm limits.
The label alternatives are:
Table 19-1Waveform scales, alarm sources and limits that are
Changing the scale
Scales can be adjusted between 10 and 300 mmHg in steps of 10.
• Go to setup menu of the desired pressure.
• Select Scale.
• Adjust the scale by turning and pressing the ComWheel.
Invasive blood pressure
assigned when the label is selected
LabelP1, ArtP2, CVPPAICP
Scale (mmHg)200206020
Alarm sourcesysoffoffoff
Digit formatS/DMeanS/DMean
Filter (Hz)22999
Other invasive pressure setup adjustments
Digit FormatSelect the displayed digit, either, systole and diastole (S/D) or Mean.
Sweep Speed Set the display sweep speed for the waveform display. The speeds are: 12.5,
25, or 50 mm/sec.
NOTE: The sweep speed is same for ECG, plethysmographic, and invasive
pressure waveforms.
Ventilation Mode
Select Spont or Contrl for spontaneous or controlled ventilation. Respiration
may cause artifacts in invasive pressure measurement.
HR SourceSelect the heart rate source. If the ECG signal is affected by too much noise for
a reliable heart rate calculation, heart rate can be calculated from pressure (Art)
or Plethysmographic pulse waveform (Pleth). The selected heart rate source is
shown above the numerical display of the heart rate.
AUTO selects the first available of: ECG, Art, and Pleth.
Px AlarmOpens a submenu for adjusting Alarm settings, see the “Alarms” chapter.
Previous Menu Return to the NIBP & Inv. Pressure menu.
87
Datex-Ohmeda S/5 Light Monitor
Invasive pressure alarms
Invasive pressure alarm is active only after pressure channel has been zeroed.
For alarm setup, see the “Alarms” chapter.
Alarm messages are explained in the chapter “Maintenance and
troubleshooting.”
Measuring principle
The transducer converts the pressure variations into electrical signals. The
electrical signals are amplified and displayed as numeric pressure values and
waveforms.
88
Carbon dioxide
20. CARBON DIOXIDE, CO
CO2 can be measured with mainstream method (N-LCM option) using
Mainstream CO2 Sensor and airway adapter.
Respiration rate is the frequency of the peak (end tidal) CO2 value per minute.
Monitoring mainstream CO
2
2
Figure 20-1Connecting mainstream CO2 sensor
1. Connect the special airway adapter between the elbow adapter and
Y-piece with the saddles upright.
2. Attach the Mainstream CO2 Sensor to the adapter between the saddles so
that the sensor cable fits into the slot on the farther saddle. Ensure the
connection between the sensor and the adapter with the Velcro tape
supplied with the sensor.
3. Connect the sensor to the connector on the monitor’s side panel.
Place the adapter as close to the patient airway as possible. Make sure that the
sensor remains in upright position.
89
Datex Ohmeda S/5 Light Monitor
To remove the adapter:
• Detach the sensor before removing the adapter from the circuit.
CAUTION Do not apply force to the sensor or sensor cable.
Select the airway adapter according to the patient’s status. Refer to the
instructions for use provided in the adapters package.
Using the adult/pediatric mainstream airway adapter (endotracheal diameter ≥
5.5 mm) increases the dead space by 6 ml. With the low dead-space airway
adapter (diameter < 5.0 mm), the increase is 0.6 ml.
NOTE: Airway adapter selection recommendations should be used as a
guideline. It is important for the clinician to assess each patient when selecting
the appropriate airway adapter.
WARNING Do not allow the CO2 sensor to come in contact with the
patient’s skin for a prolonged period of time. Mainstream sensor
is heated to prevent the fogging of the airway windows.
WARNING Do not immerse the sensor in liquids or autoclave it.
WARNING Do not use damaged sensors or adapters.
WARNING Do not clean or sterilize a single use adapter.
Display of CO2 measurement
1
Figure 20-2Display of the CO2 waveform
(1) Continuous CO2 waveform
(2) Values of end tidal (ET) and fraction of inspiratory (FI) CO
(3) Respiration rate
2
2
3
90
Carbon dioxide
CO2 adjustments
MeasurementSet on mainstream (MainSt) or turn the measurement OFF.
ScaleSelects 0 - 6 %, 0 - 10 %, 0 - 15 %, or kPa and 0 - 50 mmHg,
0 - 80 mmHg, 0 - 100 mmHg. Scale 0 - 6% is used in normal situations,
0 - 10% and 0 - 15% scales are used if hypercarbia is expected.
N2O/O2 CompSelects compensation: OFF, N2O or O2. In the breathing circuit when there is
50% or more of N2O, use N2O compensation; when ther is 60% or more of O2,
use O2. In other cases, turn off the compensation.
UnitSelects CO2 unit: %, kPa or mmHg. See “Unit conversions” later in this chapter.
Sweep SpeedSelects Fast (6.25 mm/sec) or Slow speed (0.62 mm/sec). Slow waveforms
have a sweep speed one tenth of normal, for a full screen sweep (five minutes).
Slow waveforms show changes better than fast waveforms.
Resp Rate Source
Change respiration rate source: AUTO, CO2 or impedance respiration.
CO2 AlarmBrings you to the Alarms Setup menu to adjust CO2 alarms.
Main MenuReturns you to the Main Menu.
Calibration check
No routine calibration by user is required. Calibration check is recommended to
be performed once a year by a qualified technical personnel. See the “TechnicalReference Manual” for instructions.
91
Datex Ohmeda S/5 Light Monitor
Unit conversions
Relationship between gas concentration and its partial pressure:
reading in mmHg = (ambient pressure mmHg * gas concentration in %) / 100
reading in kPa = (ambient pressure in mmHg * gas concentration in % ) / 750
Measuring principle
The CO2 measurement is based on the infrared absorption. In the mainstream
method the measurement bench is located at the patient’s airway adapter.
92
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.