Datex-Ohmeda S-5 Compact Technical reference manual (2008)

GE Healthcare

Datex-Ohmeda S/5TM Compact Anesthesia Monitor Datex-Ohmeda S/5
TM
Compact Critical Care Monitor

Technical Reference Manual

Conformity according to the Council Directive 93/42/EEC concerning Medical Devices
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the USA, check local laws for any restriction that may apply.
All specifications subject to change without notice.
Order code M1065281
st
edition
1
June 2, 2008
GE Healthcare Finland Oy Helsinki, Finland P.O. Box 900 FI-00031 GE, FINLAND Tel: +358 10 39411 Fax: +358 9 1463310
www.gehealthcare.com
Copyright © 2008 General Electric Company. All rights reserved.
Madison, WI 53707-7550, USA
Datex-Ohmeda Inc.
Tel: +1 608 221 1551
Fax: +1 608 222 9147
P.O. Box 7550
Intended purpose (Indications for use)
The Datex-Ohmeda S/5 Compact Anesthesia Monitor with L-CANE06 or L-CANE06A software is intended for multiparameter patient monitoring with optional patient care documentation.
The S/5 Compact Anesthesia Monitor with L-CANE06 or L-CANE06A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS), Entropy (State Entropy and Response Entropy) and neurophysiological status of all hospital patients.
The S/5 Compact Anesthesia Monitor with L-CANE06 and L-CANE06A software when using BIS is for monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as an aid in monitoring the effects of certain anesthetic agents.
The S/5 Compact Anesthesia Monitor with L-CANE06 or L-CANE06A software is also indicated for documenting patient care related information.
The S/5 Compact Anesthesia Monitor with L-CANE06 or L-CANE06A software is indicated for use by qualified medical personnel only.
The Datex-Ohmeda S/5 Compact Critical Care Monitor with L-CICU06 or L-CICU06A software is intended for multiparameter patient monitoring.
The S/5 Compact Critical Care Monitor with L-CICU06 and L-CICU06A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS) and neurophysiological status of all hospital patients.
The S/5 Compact Critical Care Monitor with L-CICU06 and L-CICU06A software when using BIS is for monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as an aid in monitoring the effects of certain anesthetic agents.
The S/5 Compact Critical Care Monitor with L-CICU06 and L-CICU06A software is indicated for use by qualified medical personnel only.
Classifications
In accordance with IEC 60601-1
Class I and internally powered equipment – the type of protection against electric shock. Type BF or CF equipment. The degree of protection against electric shock is indicated by a symbol on
each parameter module. Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen
or nitrous oxide. Continuous operation according to the mode of operation.
In accordance with IEC 60529 In accordance with EU Medical Device Directive
The Datex-Ohmeda S/5 Compact Anesthesia Monitor is classified as IIb. The Datex-Ohmeda S/5 Compact Critical Care Monitor is classified as IIb. In accordance with CISPR 11: Group 1, Class A
• Group 1 contains all ISM (industrial, scientific and medical) equipment in which there is intentionally generated and/or used conductively coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself.
Responsibility of the manufacturer
GE Healthcare Finland Oy (GE) is responsible for the effects on safety, reliability and performance of the equipment only if:
• assembly, extensions, readjustments, modifications, servicing and repairs are carried out by personnel authorized by GE.
• the electrical installation of the monitor room complies with appropriate requirements.
• the equipment is used in accordance with the "User's Guide" and serviced and maintained in accordance with the “Technical Reference Manual”.
Trademarks
Datex, Ohmeda, S/5, D-lite, D-lite+, Pedi-lite, Pedi-lite+, D-fend, D-fend+, Mini D-fend, OxyTip+, MemCard, ComWheel, ComBar, EarSat, FingerSat, FlexSat, PatientO
are trademarks of GE Healthcare Finland Oy. All other product and company names are property of their respective owners.
A portion of the Entropy software is derived from the RSA Data Security, Inc. MD5 Message-Digest Algorithm.
, Patient Spirometry, Entropy and Tonometrics
2
Product availability
Some of the products mentioned in this manual may not be available in all countries. Please, consult your local representative for the availability.
Technical Reference Manual, Order code: M1065281
Document No. Updated Description
Master Table of Contents
Datex-Ohmeda
S/5TM Compact Anesthesia Monitor
S/5TM Compact Critical Care Monitor
1st edition
Part I, General Service Guide
M1144951 -004 Introduction, System description, Installation,
Interfacing, Functional check, General
troubleshooting
M1144953 -001 Planned Maintenance Instructions 2
Part II, Product Service Guide
Document No. Updated Description
M1144955 -001 CAM, CCCM Service Menu 1
M1144956 -003 Display Unit for F-CM1, F-CMC1 rev. 00...05
F-CMREC1, F-CMCREC1 rev .00...01
Software Licenses L-CANE06(A), L-CICU06(A)
M1144958 -001 Frame Unit for F-CM1, F-CMC1 rev. 00...05
F-CMREC1, F-CMCREC1 rev. 00...01
AC/DC Power Supply Unit
M1144959 -001 Spare parts 4
1
2
3
Document no. M1144951-004
1
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For your notes:
2
Document no. M1144951-004
Table of contents
Table of contents
Table of contents i
Table of figures iv
About this manual 1
1Introduction 3
1.1 Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.1 Symbols on transport packaging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.2 Symbols on equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1.3 Equipment safety symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.1.4 Other symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.2.1 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.2.2 ESD precautionary procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
1.2.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2 System description 15
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2 Bus structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.3 Distributed processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.4 Module communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.5 Software loading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.6 Parameter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3 System installation 19
3.1 Unpacking instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2 Choosing location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3 S/5 Frame for Compact Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3.1 Compact Monitor connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.3.2 Connecting to mains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3.3 Connecting to Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.3.4 Connecting to Wireless Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.3.5 Inserting the parameter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.3.6 E-PSM(P) Mounting Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.3.7 Downloading Monitor Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.3.8 Performing Factory Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.3.9 Installing the Datex-Ohmeda Network and MemCard Upgrade, U-CMMEM . . . . . . . 26
3.3.10 Installing the Datex-Ohmeda Network Upgrade, U-CMNET . . . . . . . . . . . . . . . . . . . . . . 27
3.3.11 Installing the S/5 Wireless LAN Upgrade, U-CMW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.3.12 Installing the S/5 Recorder Upgrade, U-CMREC1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.4 S/5 Remote Controller, K-CREMCO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.5 Anesthesia Record Keeping Keyboard, K-ARKB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6 ARK Barcode Reader, N-SCAN (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.1 Connection to Compact Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Document no. M1144951-004
i
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
3.7 S/5 Compact Airway Modules, E-xxxx / M-xxxxx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7.1 Sample gas exhaust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.7.2 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4Interfacing 35
4.1 Interfacing external monitors via Interface Module, E-INT / M-INT . . . . . . . . . . . . . . . . . . . . . . 35
4.1.1 Connection to external Datex-Ohmeda monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.1.2 Connection to Critikon Dinamap 1846SX, Abbott Oximetrix 3 and Baxter Explorer 37
4.1.3 Connection to Baxter Vigilance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.1.4 Connection to Nellcor N-100 and N-1000. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.1.5 Connection to Nellcor N-200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.2 Interfacing external bedside devices via Device Interfacing Solutions, N-DISxxx. . . . . . . . 38
4.2.1 Device Interfacing Solution components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.2.2 Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.2.3 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.2.4 Selecting the external device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.2.5 Functional check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.2.6 Selecting the parameter data source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.3 Interfacing Datex-Ohmeda Anesthesia Delivery Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.3.1 Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.3.2 Setting interfacing parameters on the S/5 Anesthesia Delivery Unit. . . . . . . . . . . . . . 43
4.3.3 Setting interfacing parameters on the S/5 Compact Anesthesia Monitor . . . . . . . . . 43
4.4 Interfacing Dräger Cicero, Cato, Julian and Narkomed 2C (by NAD) . . . . . . . . . . . . . . . . . . . . 44
4.4.1 Interconnection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.4.2 Setting communication parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.4.3 Setting interfacing parameters on the S/5 Compact Anesthesia Monitor or
S/5 Compact Critical Care Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.5 Interfacing printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4.5.1 Setting S/5 Compact Monitor interface for printers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4.5.2 Connection to serial printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4.5.3 Connection to parallel printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4.5.4 Installing the Serial-to-Parallel Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
4.5.5 Connection to printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.6 Interfacing computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.7 Output signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
4.7.1 Digital outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
4.7.2 Analog outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
4.7.3 S/5 Pressure Temp Module, E-PT or M-PT, output signals . . . . . . . . . . . . . . . . . . . . . . . . 51
5 Functional check 53
5.1 Recommended tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
5.1.1 Hemodynamic patient simulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.2 Visual inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5.3 Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
5.3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
5.3.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.3.3 Keyboard(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.3.4 Frame unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.3.5 Display unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
ii
Document no. M1144951-004
Table of contents
5.3.6 Compact Airway Module, E-CXXXXX/ M-CXXXXX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.7 Single width Airway Module, E-miniC/ M-miniC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.8 Tonometry Module, E-TONO/ M-TONO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.9 Multiparameter Hemodynamic Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5.3.10 Pressure/Pressure Temp Modules, E-P, E-PT, M-P, M-PT . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.3.11 Dual pressure Module, E-PP/ M-PP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.3.12 Cardiac Output Modules, E-COP/ M-COP, E-COPSv/ M-COPSv . . . . . . . . . . . . . . . . . . . . 61
5.3.13 NIBP module, M-NIBP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
5.3.14 Nellcor Compatible Saturation module, E-NSATX/ E-NSAT/ M-NSAT . . . . . . . . . . . . . . 61
5.3.15 Datex-Ohmeda Oxygen Saturation module, M-OSAT . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.3.16 BIS Module, E-BIS/ M-BIS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.3.17 Entropy Module, E-ENTROPY/ M-ENTROPY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.3.18 Data Card and Menu Card function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.3.19 Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.3.20 Network connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.3.21 Wireless Network Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.3.22 Interface Module, E-INT/ M-INT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.3.23 Interface module for PSM, E-INTPSM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.3.24 Device Interfacing Solution, N-DISxxx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
5.3.25 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6 General troubleshooting 65
6.1 Software troubleshooting chart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Appendix A: Functional check form, Datex-Ohmeda S/5 CAM, CCCM A-1
Appendix B: ElectroMagnetic Compatibility B-1
Appendix C: Channel Mask Selections C-1
Document no. M1144951-004
iii
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Table of figures
Figure 1 S/5 Compact Anesthesia Monitor system........................................................................................................................ 3
Figure 2 General bus structure of S/5 system................................................................................................................................15
Figure 3 Principle of UPI section operation ......................................................................................................................................16
Figure 4 General structure of parameter modules with patient isolation.........................................................................17
Figure 5 External connections of Compact Monitor frame F-CM(C)1 rev.05 ....................................................................20
Figure 6 External connections of Compact Monitor frame F-CM(C)1 rev.00...03............................................................21
Figure 7 E-PSM(P) mounting accessories ..........................................................................................................................................24
Figure 8 Installation of the U-CMREC1 ...............................................................................................................................................29
Figure 9 ARK Barcode Reader, N-SCAN connected to Compact Monitor..........................................................................30
Figure 10 Compact Airway Module, E-XXXX.......................................................................................................................................30
Figure 11 Scavenging through ventilator reservoir ........................................................................................................................31
Figure 12 Connecting the gas module to the scavenging connector of S/5 Avance .....................................................32
Figure 13 Sample gas returned to patient circuit in ADU.............................................................................................................33
Figure 14 Connection cables and LED indicators ............................................................................................................................40
Figure 15 An example of interfacing external devices with Device Interfacing Solution..............................................41
Figure 16 Connecting S/5 Compact Monitor to printer, converter model PI130-R2 .......................................................48
Figure 17 Connecting S/5 Compact Monitor to printer, converter model PI 1115A........................................................49
Figure 18 S/5 Compact Monitor general troubleshooting flowchart......................................................................................65
Figure 19 Software troubleshooting ......................................................................................................................................................66
iv
Document no. M1144951-004

About this manual

Intended audience
This Technical reference manual is meant for service representatives and technical personnel who install, configure, maintain, administer, troubleshoot or repair Datex-Ohmeda S/5 Compact
Anesthesia Monitors running the software license L-CANE06 or L-CANE06A and S/5 Compact Critical Care Monitors running the software license L-CICU06 or L-CICU06A.
Notes to the reader
As the monitor setup may vary, some functions described may not be available in the monitor you are using.
The order code for the entire printed manual is M1065281. The manual includes
Technical Reference Manual Slots and every slot has an individual document number.
Part I gives the reader an overview of the monitor. It contains the information needed to
install, interface and troubleshoot the monitors. Instructions for functional check and planned maintenance are also included. Read the manual through and make sure that you understand the procedures described before the installation of the monitor. To avoid risks concerning safety or health, strictly observe the warning indications. If you need any assistance concerning the installation, please do not hesitate to contact your authorized distributor.
Part II contains detailed descriptions of each component of the S/5 CAM, CCCM Monitor,
such as frame unit, parameter modules Remote Controller and Device Interfacing Solution. Service check for each product, service menus and all the spare parts information for the Monitor is included.
For information of parameter modules, Remote Controller and Device Interfacing Solution refer to the “S/5 E-Modules, Technical Reference Manual”. Service check for each of these products is included in these slots.
The manufacturer reserves the right to change product specifications without prior notice. Although the information in this manual is believed to be accurate and reliable, the manufacturer assumes no responsibility for its use.
Installation and service are allowed by authorized service personnel only. GE Healthcare Finland Oy (GE) assumes no responsibility for the use or reliability of its software
in equipment that is not furnished by GE.
Related documentation
S/5 Modules
S/5 E-Modules, Technical Reference Manual S/5 M-Modules, Technical Reference Manual
S/5 Compact Anesthesia Monitor
For instructions for daily use including cleaning and daily maintenance, clinical aspects and basic methods of measurement see:
S/5 Compact Anesthesia Monitor, User’s Guide S/5 Compact Anesthesia Monitor, User’s Reference Manual
Document no. M1144951-004
1
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
S/5 Compact Critical Care Monitor
For instructions for daily use including cleaning and daily maintenance, clinical aspects and basic methods of measurement: S/5 Compact Critical Care Monitors, User’s Guide S/5 Compact Critical Care Monitors, User’s Reference Manual
For more information about the iCentral, S/5 Arrhythmia Workstation and anesthesia record keeping solution, see the “Technical Reference Manuals” and “User’s Reference Manuals” for these products.”
Software options and default settings are described in the “Default Configuration Worksheet” delivered with each monitor.
Available accessories are described in the “Supplies and Accessories” catalog delivered with each monitor.
Conventions used
To help you find and interpret information easily, the manual uses consistent text formats:
Sign the check form after performing the procedure.
"
Hard Keys Names of the hard keys on the Remote Controller, Command Board, side panel and modules
are written in the following way:
Menu Items Software terms that identify window parts or menu items are written in bold italic: ECG Setup.
Menu access is described from top to bottom. For example, the selection of the
Others.
Monitor
Setup hard key, the Screen 1 Setup menu item and the Waveform Fields menu item would be
shown as
‘Messages’ Messages (alarm messages, informative messages) displayed on the screen are written inside
single quotes: ‘Please wait’.
“Sections” When referring to different sections in this manual or to other manuals, manual names and
section names are enclosed in double quotes: See section "Cleaning and care."
Please refer to "User's Reference Manual: Alarms."
Hypertext links Hypertext links on PDF versions are written in blue color.
WARNING Warnings are written in the following way:
Monitor Setup - Screen 1 Setup - Waveform Fields.
WARNING This is a WARNING.
CAUTION Cautions are written in the following way:
CAUTION This is a CAUTION.
NOTE Notes are written in the following way:
NOTE: This is a NOTE.
In this manual, the word “select” means choosing and confirming.
Illustrations and names
All illustrations in this manual are only examples, and may not necessarily reflect your system
settings or data displayed in your system. If a particular selection is not available in your
system, the selection is shown grayed.
2
Document no. M1144951-004

1 Introduction

1
2
3
5
4
The modular design makes the system flexible and easy to upgrade. In addition to patient
parameter modularity and easy upgrades, the monitor can be upgraded to anesthesia record
keeping, wired and wireless networking and memory card operation. Additionally, external
devices can be interfaced to the monitor with interface modules.
NOTE: Your system may not include all these components. Consult your local representative
for the available components.
Introduction
Figure 1 S/5 Compact Anesthesia Monitor system
(1) S/5 Compact Anesthesia Monitor with modules inside
(2) Remote Controller, K-CREMCO
(3) Printer
(4) Wireless Network (WLAN) Access Point, N-WAP
(5) Network (LAN)
Optional components
Optional components are:
Built-in recorder, N-CMREC1
Remote controller
Keyboard for anesthesia record keeping solution, barcode reader
Wireless network option, N-CMW
Possible software options are:
MemCard option, N-CMMEM
NOTE: Insert the WLAN card in the first slot from the front.
NOTE: Memory Module, M-MEM/ E-MEM, is not compatible with the S/5 Compact Monitors.
Document no. M1144951-004
3
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Communication between monitors
You can use the Compact Monitor as a stand-alone monitor or for:
viewing and receiving data (alarms, vital signs) from other patient monitors
gathering and storing data during transportation.
To view other patient monitors, the monitor needs to be connected to the network (LAN or
WLAN). To gather, store, and transfer data between different Datex-Ohmeda monitors, use
memory cards or connect the monitor to the network.
NOTE: With WLAN, use only Data Card.
4
Document no. M1144951-004

1.1 Symbols

Pb
Ni-MH
1.1.1 Symbols on transport packaging
The contents of the transport package are fragile and must be handled with care.
Indicates the correct upright position of the transport package.
The transport package must be kept in a dry environment.
Indicates the temperature limitations within which the transport package should be stored.
Introduction
This package can be recycled.
1.1.2 Symbols on equipment
This battery contains Ni-MH and can be recycled.
Dangerous voltage.
When using the ARK Barcode Reader, N-SCAN, do not stare into beam. The N-SCAN Barcode Reader is a Class 2 laser product.
5
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
1.1.3 Equipment safety symbols
- Attention, consult accompanying documents.
- When displayed next to the O is set below 21%.
- When displayed next to the HR value, indicates that the pacer is set on R.
- On the modules or frames indicates that modules with identical measurements should not be used in the same monitor. If such modules have been inserted, remove the module that has been most recently connected. You can also remove both modules and re-connect the new module after five seconds.
- On the Interface Module E-INT, M-INT, indicates that it is for connecting external devices. Do not connect patient cables to the module.
- On the E-TONO, M-TONO module indicates that the module should only be used with Tonometrics catheters.
- On the E-miniC module indicates that airway gases should be calibrated every six months in normal use and every two months in continuous use.
- BIS: On the Aspect DSC indicates that the converter must not be opened for any reason or autoclaved.
value, indicates that the FiO2 low alarm limit
2
- On the E-PRESTN, E-PRETN, E-RESTN, E-PSM, E-PSMP, E-P, E-PP, E-PT, E-COP and E-COPSv module indicates that protection against cardiac defibrillator discharge is due in part to the accessories for pulse oximetry (SpO2), temperature (T) and invasive pressure (P) measurement.
- On the E-NMT, M-NMT module indicates the following warnings:
- Do not place the NMT stimulating electrodes on the patient’s chest.
- Always stop the NMT measurement before handling the stimulating
electrodes.
- Never subject a patient with an implanted electronic device to electrical
stimulation without consulting a medical specialist first.
- On the rear or bottom panel this symbol indicates the following warnings and cautions:
- Electric shock hazard. Do not open the cover or the back. Refer servicing
to qualified service personnel.
- For continued protection against fire hazard, replace the fuse only with
one of the same type and rating.
- Do not touch a battery-operated monitor during defibrillation procedure.
- Disconnect from the power supply before servicing.
- Do not use the monitor without manufacturer approved mounting
attached.
- Lithium battery on the CPU board: follow the regional regulations for
disposal.
6
Document no. M1144951-004
Introduction
Type BF (IEC 60601-1) protection against electric shock.
Type BF (IEC 60601-1) defibrillator-proof protection against electric shock.
Type CF (IEC 60601-1) protection against electric shock.
Type CF (IEC 60601-1) defibrillator-proof protection against electric shock.
When displayed in the upper left corner of the screen, indicates that the alarms are silenced. When displayed in the menu or digit fields, indicates that the alarm source has been turned off or alarm does not meet the alarm-specific activation criteria.
ESD warning symbol for electrostatic sensitive devices. Pins of connectors identified with the ESD warning symbol should not be touched. Connections should not be made to these connectors unless ESD precautionary procedures are used. For details, see section “1.2.2. ESD precautionary procedures”.
1.1.4 Other symbols
SN, S/N
Symbol for non-ionizing electromagnetic radiation. Interference may occur in the vicinity of equipment marked with this symbol.
Equipotentiality. Monitor can be connected to potential equalization conductor.
Alternating current
Fuse. Replace the fuse only with one of the same type and rating.
Serial Number
Battery operation and remaining capacity (green bar)
Battery charging (white bar)
Submenu. Selecting an alternative marked with this symbol in a menu opens a new menu.
Document no. M1144951-004
7
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
The monitor is connected to the Datex-Ohmeda Network (LAN).
The monitor is connected to the Datex-Ohmeda Network (WLAN).
Data Card (green) and/or Menu Card (white) is inserted.
WLAN signal strength. The number of segments corresponds to the signal strength: four segments indicate strong signal, one segment weak signal. When connection to access point is being searched, the segments scroll from zero to four and back.
Ethernet connector
A blinking heart next to the heart rate or pulse rate value indicates the beats detected.
A lung next to the respiration rate value indicates that respiration rate is calculated from the impedance respiration measurement.
Gas inlet
Gas outlet
Do not reuse.
Use by. Indicates the last use day.
Date of manufacturer
Does not contain Latex.
8
Document no. M1144951-004
Do not immerse the sensor in liquids.
Introduction
20
IPX class:
IPX0 IPX1 IPX2 IPX3 IPX4 IPX7 IPX8
Degree of protection against harmful ingress of water as detailed in the IEC 60529:
- Ordinary equipment
- Protection against vertically falling water drops.
- Protection against vertically falling water drops when enclosure tilted
up to 15 °.
- Protected against spraying water.
- Protected against splashing water.
- Protected against the effects of temporary immersion in water.
- Protected against the effects of continuous immersion in water.
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.
This symbol indicates the product contains hazardous materials in excess of the limits established by the Chinese standard SJ/T11364-2006 Requirements for Concentration Limits for Certain Hazardous Substances in Electronic Information Products. The number in the symbols is the Environment-friendly Use Period (EFUP), which indicates the period during which the toxic or hazardous substances or elements contained in electronic information products will not leak or mutate under normal conditions so that the use of such electronic information products will not result in any severe environmental pollution, any bodily injury or damage to any assets. The unit of the period is “Year”.
In order to maintain the declared EFUP, the product shall be operated normally according to the instructions and environmental conditions as defined in the product manual, and the periodic maintenance schedules specified in Product Maintenance Procedures shall be followed strictly.
Consumables or certain parts may have their own label with an EFUP value less than the product. Periodic replacement of those consumables or parts to maintain the declared EFUP shall be done in accordance with the Product Maintenance Procedures.
This product must not be disposed of as unsorted municipal waste, and must be collected separately and handled properly after decommissioning.
9
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors

1.2 Safety

The following list contains general warnings and cautions you should know before installing, maintaining or servicing the system. Warnings and cautions specific to the use of the system can be found in the User’s Guide and User’s Reference Manual.
1.2.1 Safety precautions
Warnings
WARNING A WARNING indicates a situation in which the user or the patient may be in
danger of injury or death.
The device is not able to withstand unpacked drops from a height of 1 m without
damaging the module latches. If the device is dropped, please service the device before taking it back into use.
Power connection
Always check that the power cord and plug are intact and undamaged.
Use only hospital-grade grounded power outlets and power cord. Do not remove the
grounding pin from the power plug.
Use only an intact power cord. Replace the power cord if it is cracked, frayed, broken or
otherwise damaged.
Do not apply tension to the power cord otherwise the cord may get damaged.
Do not use an additional multiple socket outlet, extension cord or adapters of any kind.
Before starting to use the system, ensure that the whole combination complies with the
international standard IEC 60601-1-1 and with the requirements of the local authorities. Do not connect any external devices to the system other than those specified.
If the integrity of the external protective earth conductor arrangement is in doubt, use the
monitor with battery operation.
To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth.
Installation
Keep the monitor horizontal when the Compact Airway Module is used. Tilting the monitor
may cause erroneous results in the Compact Airway Module’s readings and damage the module.
The monitor or its components should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the monitor and its components should be observed to verify normal operation in the configuration in which it will be used.
Pins of connectors identified with the ESD warning symbol should not be touched.
Connections should not be made to these connectors unless ESD precautionary procedures are used. For details, see section “1.2.2. ESD precautionary procedures
After transferring or reinstalling the monitor, always check that it is properly connected
and all parts are securely attached. Pay special attention to this in case of stacked mounting.
Do not use the monitor in high electromagnetic fields (for example, during MRI.)
Never install the monitor so that it is above the patient.
10
Document no. M1144951-004
Introduction
A printer or computer must be supplied from an additional transformer providing at least
basic isolation (isolating or separating transformer).
If you accidentally drop the monitor, modules or frames, have them checked by
authorized service personnel prior to clinical use.
To avoid explosion hazard, do not use the monitor in presence of flammable anesthetics.
The monitor measures only non-flammable anesthetics.
Do not touch the patient, table, instruments, modules or the monitor during defibrillation.
Laser radiation
When using the ARK Barcode Reader, N-SCAN, do not stare into the beam. The N-SCAN is
a Class 2 laser product.
External connection
Do not connect any external devices to the monitor other than those specified.
Fuse replacement
Replace a fuse only with one of the same type and rating.
Explosion hazard
To avoid explosion hazard do not use the monitor in the presence of flammable
anesthetics.
Patient safety
Do not perform any testing or maintenance on the monitor while it is being used on a
patient.
PACEMAKER PATIENTS: The impedance respiration measurement may cause rate
changes in Minute Ventilation Rate Responsive Pacemakers. In this case, set the pacemaker rate responsive mode off or turn the monitor impedance respiration measurement off.
Never install the monitor so that it is above the patient.
The monitor must not be used without manufacturer approved mounting attached.
Operation of the monitor outside the specified values may cause inaccurate results.
Cleaning and service
Only trained personnel with proper tools and test equipment should perform the tests
and repairs described in this manual. Unauthorized service may void the monitor warranty.
Always unplug the monitor before cleaning or service. After cleaning or service ensure
that every part of the monitor is dry before reconnecting it to the power supply.
Do not touch any exposed wire or conductive surface while any cover is removed and the
monitor is energized. The voltages present can cause injury or death.
Pins of connectors identified with the ESD warning symbol should not be touched.
Connections should not be made to these connectors unless ESD precautionary procedures are used. For details, see section “1.2.2. ESD precautionary procedures”.
Always perform an electrical safety check and a leakage current test on the monitor after
service.
11
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Handle the water trap and its contents as you would any body fluid. Infectous hazard
may be present.
Accessories
Use only accessories, including mounts and batteries, and defibrillator-proof cables and
invasive pressure transducers approved by GE Healthcare. For a list of approved supplies and accessories, see the “Supplies and Accessories” catalog delivered with the monitor. Other cables, batteries, transducers and accessories may cause a safety hazard, damage the equipment or the system, result in increased emissions or decreased immunity of the equipment or system or interfere with the measurement. Protection against cardiac defibrillator discharge is due in part to the accessories for pulse oximetry (SpO2), temperature (T) and invasive pressure (P) measurement.
Single use accessories are not designed to be reused. Reuse may cause a risk of
contamination and/or affect the measurement accuracy.
Cautions
CAUTION A CAUTION indicates a condition that may lead to equipment damage or
malfunction.
Installation
Leave space for air circulation to prevent the monitor from overheating.
Before connecting the power cord to the power supply, check that the local voltage and
frequency correspond with the rating stated on the device plate.
Before use
Allow two minutes for warm-up and note any error messages or deviations from normal
operation.
Do not connect a sampling line to the female Patient Spirometry connector while the
other end of the sampling line is connected to the D-fend water trap. The pressure in the gas sampling system may cause damage to the PVX unit pressure transducers.
Autoclaving and sterilizing
Do not autoclave any part of the monitor.
Do not gas sterilize the modules.
Cleaning and service
Do not use hypoclorite, ammonia-, phenol-, or acetone based cleaners. These cleaners
may damage the monitor surface.
Do not immerse any part of the device in any liquid, or allow liquid to enter the monitor or
modules.
Do not apply pressurized air to any outlet or tubing connected to the monitor.
Electrostatic discharge through the PC boards may damage the components. Before
handling PC boards, wear a static control wrist strap. Handle all PC boards by their non-conductive edges and use anti-static containers when transporting them.
Do not break or bypass the patient isolation barrier when testing PC boards.
If liquid has accidentally entered the system or its parts, disconnect the power cord from
the power supply and have the equipment serviced by authorized service personnel.
12
Document no. M1144951-004
Introduction
Do not clean the spirometry tubes with high pressure air or O
spirometry tubes are connected to Patient Spirometry connector. High differential pressure may damage PVX unit pressure transducers.
flushing while the
2
Special components
Special components are used in these monitors that are vital to assure reliability and
safety. GE Healthcare assumes no responsibility for damage, if replacement components not approved by GE Healthcare are used.
A lithium battery on the CPU Board. Dispose of the faulty IC containing the battery
according to local regulations.
Batteries
Refresh the batteries completely every six months. To replace the batteries safely, please refer to the service instructions in this manual.
Do not short-circuit the battery terminals, this may produce a very high current, which
will damage the battery.
Do not dispose of the battery into open flame, nor put the battery near fire, as it may
explode.
Do not dismantle the battery.
See also section “Symbols”.
Storage and transport
Do not store or transport the monitor outside the specified temperature, pressure and humidity ranges:
Temperature -10...+50 °C/14...122 °F Atmospheric pressure 660...1060 hPa/500...800 mmHg/660...1060 mbar Relative humidity 10...90% noncondensing For display specific environmental requirements see specifications in the “Display” slot.
1.2.2 ESD precautionary procedures
To avoid electrostatic charges building up, it is recommended to store, maintain and use
the equipment at a relative humidity of 30% or greater. Floors should be covered by ESD dissipative carpets or similar. Non-synthetic clothing should be used when working with the component.
To prevent applying a possible electrostatic discharge to the ESD sensitive parts of the
equipment, one should touch the metallic frame of the component or a large metal object located close to the equipment. When working with the equipment and specifically when the ESD sensitive parts of the equipment may be touched, a grounded wrist strap intended for use with ESD sensitive equipment should be worn. Refer to the documentation provided with the wrist straps for details of proper use.
ESD precautionary procedure training
It is recommended that all potential users receive an explanation of the ESD warning symbol and training in ESD precautionary procedures.
The minimum contents of an ESD precautionary procedure training should include an introduction to the physics of electrostatic charge, the voltage levels that can occur in normal practice and the damage that can be done to electronic components if they are touched by an operator who is electrostatically charged. Further, an explanation should be given of methods
13
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
to prevent build-up of electrostatic charge and how and why to discharge one’s body to earth or to the frame of the equipment or bond oneself by means of a wrist strap to the equipment or the earth prior to making a connection.
1.2.3 Disposal
Dispose of the whole device, parts of it and its packing material and manuals in accordance with local environmental and waste disposal regulations.
14
Document no. M1144951-004

2 System description

486
Processor
PCMCIA Serial
Ethernet Sound UPI
Chipset
Display
Controller
486 LOCAL BUS
ISA-BUS
DRAM
Reset
Logic
Non-
Volatile
Memory
Boot
Memory
Code
Memory
Patient Module
Patient
Module
Patient
Modules
MODULE-BUS
cam_p1_gnrl_bus_strctr_cm.vsd

2.1 Introduction

Datex-Ohmeda monitors build up a freely configurable modular system. The architecture is designed to enable different module combinations so that the user is able to get the desirable parameter and feature set. This modular approach makes it possible to add new features when they are needed.

2.2 Bus structure

System description
Figure 2 General bus structure of S/5 system
The CPU bus is a communication channel used only for internal data transfer. It is based on the ISA bus used in IBM PC computers. Data is transferred on this 16 bit wide bus using the CPU clock frequency.
The module bus is for the parameter modules. The bus is based on the industry standard RS-485, which uses a differential serial method to transfer data. This type of bus is robust and it allows parameter modules to be inserted or removed while the power is on. The module bus uses a 500 kbps data transfer rate.
The RS-485 type of serial communication supports so-called multidrop or party line connections. This means that all parameter modules connected to the module bus use exactly the same lines for communication. The advantage of this is that all bus connectors are identical and the modules can be connected in any order and position.
Document no. M1144951-004
15
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
UPI
Processor
Shared
SRAM
Control
Logic
ISA-BUS
Voltage
Meas
Temp Meas
Keypad Interface
ComWheel I nterface
Module Bus
Analog & Digit al outputs
cam_p1_prncpl_UPI_operation.vsd

2.3 Distributed processing

A system assembled from Datex-Ohmeda products is a multiprocessor system. All parameter modules have their own microprocessor, which performs functions such as module key control, waveform filtering, parameter related computing and pneumatic control, etc. At the same time the main CPU performs higher level tasks such as trending and alarm control. While the parameter modules and CPU are performing their tasks, the UPI (Universal Peripheral Interface) microprocessor handles all functions needed to transfer data between the parameter modules and the CPU.
This kind of parallel processing gives one major advantage to centralized processing. When new parameter modules are added to the system, the processing power is increased. As a result, the system does not slow down when new features are added.

2.4 Module communication

The communication master controlling data transfers between the CPU and parameter modules is called UPI processor. It sends data to each connected module 100 times a second. Modules respond to each data request immediately by sending a data package, whose length depends on the type of the module. This communication protocol ensures that each module receives and sends data every 10 ms. If a module does not respond to data requests, the UPI processor presumes that the module is disconnected.
Parameter modules may hold a static (fixed) or dynamic address, which the UPI processor uses when sending out data. Two parameter modules of the same type must not be fitted onto the same monitor since they might reply to a data request simultaneously, thus causing communication errors.
Figure 3 Principle of UPI section operation
The UPI processor collects and stores all data that is received from the parameter modules into a shared SRAM, which is mapped directly to the address space of the main CPU. The main CPU reads data from the memory while the UPI processor guarantees that the data is up to date. This operation also works in the other direction. In this the main CPU fills the shared SRAM with data and the UPI processor distributes it to the parameter modules.
16
Document no. M1144951-004

2.5 Software loading

Module Bus
Isolation
transformer
RS485 drivers
Peripheral
drivers
A/D
converter
CPU
Analog
electronics
Opto
isolation
RAM
EEPROM
Patient isolation
PATIENT
Module
keys
Data
+13...16 V
VMOD
+5 V
+12 V
The program memory on the CPU board is loaded with monitor software at the factory. The software is used for running all the functions that are integrated into the PC board. For service and upgrade procedures, the CPU board is fitted with a PCMCIA card drive through which new software can be loaded.

2.6 Parameter modules

System description
Figure 4 General structure of parameter modules with patient isolation
The detailed structure of a parameter module depends on the specific needs for each individual parameter. However, some common parts are used in the parameter modules. The electronics inside the module is usually divided into isolated (floating) and non-isolated sections. Typically, the non-isolated section consists of buffers to interface the parameter module to the module bus while the rest of the electronics is located in the isolated section. The isolated section includes the microcontroller together with memory components, the front-end analog electronics (amplifiers, etc.) and sensor drivers.
17
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For your notes:
18
Document no. M1144951-004

3 System installation

3.1 Unpacking instructions

1. Confirm that the packing box is undamaged. If the box is damaged, contact the shipper.
2. Open the top of the box and carefully unpack all components.
3. Confirm that all components are undamaged. If any of the components are damaged, contact the shipper.
4. Confirm that all components are included. If any of the components are missing, contact your GE Healthcare distributor.

3.2 Choosing location

Consider the following aspects:
lighting
space
connections
System installation
electromagnetic and radio frequency interference, see Appendix B. ElectroMagnetic
Compatibility
environment
WARNING The monitor or its components should not be used adjacent to or stacked
with other equipment. If adjacent or stacked use is necessary, the monitor and its components should be observed to verify normal operation in the configuration in which it will be used.
WARNING Never install the monitor so that it is above the patient.
CAUTION The monitor display is fragile. Ensure that it is not placed near a heat source or
exposed to mechanical shocks, pressure, moisture or direct sunlight.

3.3 S/5 Frame for Compact Monitor

Mounting of S/5 Compact Monitor to the Bed Mount, Wall Mount or Roll Stand is described in a separate instruction sheet delivered with each mount.
WARNING After transferring or reinstalling the monitor, always check that it is
properly connected and all parts are securely attached. Pay special attention to this in case of stacked mounting.
19
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
1
7
8
9
2
4
3
5
6
14
12
13
11
10
3.3.1 Compact Monitor connections
Figure 5 External connections of Compact Monitor frame F-CM(C)1 rev.05
(1) Air filter (2) NET connector (3) NET ID connector (not used with N-CMW) (4) Recorder (optional) (5) SYNC connector (6) Connector for external keyboard (anesthesia record keeping solution keyboard),
K-CREMCO or Barcode Reader
(7) Potential equalization (8) Serial port (9) Receptacle for power cord (10) WLAN card LED (11) WLAN card (12) Slot for Memory Cards (PCMCIA) and WLAN card (13) Connector for Device Interfacing Solution (14) Modules
NOTE: Insert the WLAN card in the first PC card slot from the front.
NOTE: WLAN CF card with PCMCIA adapter is compatible only with Compact Monitor frame F-CM(C)1..05 or later.
20
Document no. M1144951-004
System installation
4
5
2
1
3
11
9
7
6
12
11
13
14
10
8
Figure 6 External connections of Compact Monitor frame F-CM(C)1 rev.00...03
(1) Dust filters (2) Recorder unit (optional) (3) Connector for factory use only (4) Synchronization connector (5) Connector for external keyboard (anesthesia record keeping solution keyboard),
K-CREMCO or Barcode Reader
(6) Potential equalization connector (7) Serial port connector (8) Receptacle for power cord (9) NET ID connector (10) NET connector (11) WLAN antenna cover (optional) (12) Connector for Device Interface Solution (13) WLAN antenna card LED WLAN antenna card (optional)
CAUTION Switch the power to standby before making any connections.
3.3.2 Connecting to mains
Connect the power cord to the mains power inlet on the side of the power supply unit and to the wall socket.
NOTE: Before taking the monitor into use for the first time, the batteries should be fully charged. Keep the monitor connected to the mains until the Battery charging symbol
21
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
disappears, or in STBY mode the Stby LED stops blinking (may take up to 10 hours if the batteries are fully discharged).
WARNING The power cord may only be connected to a three-wire, grounded, hospital
grade receptacle.
3.3.3 Connecting to Network
Make sure that the Datex-Ohmeda Network Option, N-CMNET or S/5 MemCard and Network Option, N-CMMEM is installed.
NOTE: With L-CANE03(A) and L-CICU03(A) or newer software, the N-CMNET option is automatically installed.
Use the Monitor-Network cable to connect the monitor to the network.
1. Make sure that the power is switched off.
2. Connect the Identification Plug and one RJ-45 connector to the connectors X4 and X5.
3. Connect the other RJ-45 connector to the corresponding connector on the wallbox.
4. Switch on the power. Confirm that the network symbol and 'Connected to Network' message are displayed on the upper part of the screen.
3.3.4 Connecting to Wireless Network
Make sure that the S/5 Wireless LAN Option, N-CMW and the antenna cover are installed. Use the wireless LAN antenna card to connect the monitor to the network.
NOTE: The N-CMW option requires also that either N-CMNET or N-CMMEM option is installed.
1. Make sure that the power is switched off.
2. Connect the wireless LAN antenna card into the foremost card drive slot.
3. Switch on the power. Confirm that the wireless LAN network symbol and the wireless LAN signal strength symbol are displayed on the upper part of the screen.
22
Document no. M1144951-004
3.3.5 Inserting the parameter modules
Insert each plug-in parameter module into a module slot. Firmly press the module in position.
NOTE: Ensure that the module is properly oriented (i.e. the module release latch is facing downward). Do not use two or more modules with identical functions in the monitor.
Modules with identical functions are:
Hemodynamic multiparameter modules, E-PRESTN, E-PRETN, E-RESTN, E-PSM, E-PSMP,
M-PRESTN, M-PRETN, M-RESTN, M-ESTPR, M-ETPR, M-ESTR, M-NESTPR, M-NETPR, M-NESTR, M-NE12STPR, M-NE12TPR, M-NE12STR
Dual Pressure Modules, E-PP/M-PP
Pressure Modules, E-P/ M-P, Pressure Temp Module, E-PT/ M-PT
Cardiac Output Modules, E-COP, E-COPSv, M-COP and M-COPSv
NIBP Modules, M-NIBP and hemodynamic multiparameter modules w/ N measurement
Compact Airway Modules, E-CO, E-COV, E-COVX, E-CAiOVX, E-CAiOV, E-CAiO, M-C, M-CO,
M-COV, M-CAiO, M-CAiOV, M-CAiOVX, Single-width Airway Modules M-miniC, E-miniC
The Recorder Module, E-REC/M-REC and the built-in recorders in F-CMREC1 or
F-CMCREC1 and U-CMREC1 or N-CMREC1
Interface Modules, E-INT / M-INT
System installation
Oxygen Saturation Modules, E-NSATX/ E-NSAT/ M-NSAT and M-OSAT
Tonometry Modules, E-TONO/ M-TONO
NeuroMuscular Transmission Modules, E-NMT/ M-NMT
EEG Modules, E-EEG/ M-EEG
BIS Modules, E-BISX/ E-BIS/ M-BIS
3.3.6 E-PSM(P) Mounting Accessories
Intended use
The Frame Mount for PSM and the Pole Mount for PSM, short and long, are intended to be used with the stationary docking station of the E-PSM(P) module. The Interface Module for PSM, E-INTPSM, is intended for connecting the Frame Mount and the Pole Mount to the Compact Monitor frame.
The Frame Mount for PSM can be attached directly to the Compact Monitor frame. The Pole Mount for PSM, short or long, can be attached to an IV pole or to an anesthesia machine rail with a diameter of 10mm* 25mm.
With Interface Module for PSM, E-INTPSM, the Frame Mount or the Pole Mount can be connected to the Compact Monitor frame equipped with software license 04 or later. The E-PSM(P) module can be attached to the Frame Mount or Pole Mount and used like a Datex-Ohmeda S/5 modular module.
23
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
1
2
3
4
Figure 7 E-PSM(P) mounting accessories
1. M1054424 Interface Module for PSM, E-INTPSM
2. M1051021 Frame Mount for PSM
3. M1049197 Pole Mount for PSM, short
4. M1051023 Pole Mount for PSM, long
Frame Mount for PSM – Instructions connecting to the Compact Monitor frame
1. See the picture above for used parts and assembly.
2. Remove the 4 screws on the back of the Compact Monitor frame.
3. Install the enclosed plate and tighten the screws. Install the Frame Mount for PSM, screws M5*10mm,
4. Insert the E-INTPSM module and connect the cable.
5. Attach the E-PSM(P) module and check the module communication.
6. Assemble the PSM frame, screws and the plate on the opposite side of the frame wall and tighten the screws.
24
Document no. M1144951-004
System installation
WARNING Make sure that the Pole Mount for PSM is always used in vertical position to
prevent water from entering the E-PSM(P) module.
Pole Mount for PSM – Instructions for connecting to an IV pole, vertical position
Fasten the Pole Mount for PSM with the fastening screw of the clamp and tighten properly to an IV pole.
Pole Mount for PSM - Instructions for installing in horizontal position.
Remove the 2 screws from the clamp, turn the clamp and insert and tighten the screws back. Fasten the Pole Mount for PSM with the fastening screw of the clamp and tighten properly to a horizontal tube or rail with a diameter of 10 mm*25 mm.
Pole mount for PSM – Instructions for connecting to monitor
1. Attach the E-PSM(P) module to the Pole Mount.
2. Connect the cable of the Pole Mount for PSM to the
3. Check the module communication of the E-PSM(P)
3.3.7 Downloading Monitor Software
The following instructions apply to downloading of new monitor software in case of upgrade or service. Detailed instructions for downloading software are supplied with software PCMCIA cards.
NOTE: All user settings will be lost after downloading of new monitor software.
NOTE: During the downloading of software, the serial number of the monitor is written on the software card. The software can then be downloaded again onto the same monitor, but not onto any other monitor.
1. Make sure that the monitor is switched to standby.
2. Press and hold the service reset button on the rear panel for at least five seconds or until the service reset indicator LED turns off.
3. Open the cover for card drive slots, if installed. Insert the software card into one of the card drive slots and press the software card firmly in position.
4. Make sure that at least 30 seconds have passed since the service reset, then switch the monitor on.
Compact Monitor with the E-INTPSM module (M1054424).
module.
25
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
5. Wait for approximately 80 seconds. After the start-up screen appears, enter the Service View and make sure that the information regarding monitor software has been updated. Memorize the serial number of new software.
6. Remove the software card.
7. Replace the original device plate for monitor software by the new one supplied with the software card.
8. Perform Factory Reset, see “Performing Factory Reset” . Make sure that the monitor functions normally after the restart.
9. Set the time and date.
10. Set the monitor's network communication according to the network software in use, if necessary. Enter the service menu: Monitor Setup -Install/Service (16-4-34) - Service (26-23-8)- Frame - Network- Config and set
Network software S-CNET99 -> DRI Level = 1999
Network software S-CNET01 -> DRI Level = 2001
Network software S-CNET02 -> DRI Level = 2001 or 2002 (WLAN)
Network software L-NET03 -> DRI Level = 2003
Network software L-NET05 -> DRI Level = 2005
NOTE: The DRI level 1999 is not selectable in all monitor software versions.
NOTE: If the DRI level is changed, the monitor will restart automatically.
11. Fill out all necessary documentation regarding the new monitor software.
NOTE: The license agreement of monitor software needs to be in accordance with the monitor software serial number. Make sure you archive the license agreement in a secure location.
NOTE: The first start-up after software loading takes considerably longer.
3.3.8 Performing Factory Reset
NOTE: The factory reset is necessary after replacing monitor software.
NOTE: The factory reset will restore all customized defaults, including language selection, to factory defaults.
1. Press the
2. Select Install/Service and password (16-4-34).
3. Select Service and password (26-23-8).
4. Select Set/ Test and Factory Reset.
5. The monitor will perform an automatic restart. After the restart is completed, restart the monitor also manually.
Monitor Setup key.
3.3.9 Installing the Datex-Ohmeda Network and MemCard Upgrade, U-CMMEM
The following instructions apply to upgrading the monitor with both the MemCard and the Network options. Detailed upgrade instructions are supplied with the corresponding upgrade product.
NOTE: You can download the option software only onto one monitor. During the downloading of option software, the serial number of the monitor is written on the software card, and if the downloading for some reason would fail, the software can be downloaded again onto the same monitor, but not onto any other monitor.
1. Make sure that the monitor is switched to standby.
26
Document no. M1144951-004
2. Detach the cover plate on the network connectors at the rear of the frame unit (above the power supply unit). Leave the collar on the connectors attached.
3. Open the cover for card drive slots (on the left side of the monitor). Insert the option software card into one of the card drive slots and press the software card firmly in position.
4. Switch the monitor on.
5. Wait until the normal screen appears and check that a message 'Field upgrade OK' is displayed as a note of successful upgrade.
6. Remove the option software card.
7. Enter the service menu and make sure that both options were loaded (Monitor Setup
-Install/Service - Service - Frame - Options - "Net+DC").
8. Install the device plate for option software on inside the cover for card drive slots. Close the cover.
9. Test the monitor with a MemCard and by connecting it to Datex-Ohmeda Network.
10. Fill out all necessary documentation regarding the upgrade.
3.3.10 Installing the Datex-Ohmeda Network Upgrade, U-CMNET
The following instructions apply to upgrading the monitor with the Network option. Detailed upgrade instructions are supplied with the corresponding upgrade product.
NOTE: You can download the option software only onto one monitor. During the downloading of option software, the serial number of the monitor is written on the software card, and if the downloading for some reason would fail, the software can be downloaded again onto the same monitor, but not onto any other monitor.
System installation
1. Make sure that the monitor is switched to standby.
2. Detach the cover plate on the network connectors at the rear of the frame unit (above the power supply unit). Leave the collar on the connectors attached.
3. Open the cover for card drive slots (on the left side of the monitor). Insert the option software card into one of the card drive slots and press the software card firmly in position.
4. Switch the monitor on.
5. Wait until the normal screen appears and check that a message 'Field Upgrade OK' is displayed as a note of successful upgrade.
6. Remove the option software card.
7. Enter the service menu and make sure that the option was loaded (Monitor Setup
-Install/Service - Service - Frame - Options - "Net").
8. Install the device plate for option software on inside the cover for card drive slots. Close the cover.
9. Test the monitor by connecting it to Datex-Ohmeda Network.
10. Fill out all necessary documentation regarding the upgrade.
3.3.11 Installing the S/5 Wireless LAN Upgrade, U-CMW
NOTE: WLAN CF card with PCMCIA adapter is compatible only with Compact Monitor frame F-CM(C)1..05 or later.
The following instructions apply to upgrading the monitor with the Wireless LAN option. Detailed upgrade instructions are supplied with the corresponding upgrade product. You can download the option software only onto one monitor. During the downloading of
27
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
option software, the serial number of the monitor is written on the software card, and if the downloading for some reason would fail, the software can be downloaded again onto the same monitor, but not onto any other monitor.
1. Make sure that the monitor is switched to standby.
2. Open the cover for card drive slots (on the left side of the monitor). Insert the option software card into one of the card drive slots and press the software card firmly in position.
3. Switch the monitor on.
4. Wait until the normal screen appears and check that a message 'Field Upgrade OK' is displayed as a note of successful upgrade.
5. Remove the option software card.
6. Switch the monitor to standby.
7. Detach the cover for card drive slots and install the antenna cover.
8. Insert the wireless LAN antenna card with adapter into the foremost card drive slot.
9. Switch the monitor on.
10. Enter the service menu and make sure that the option was downloaded successfully (Monitor Setup - Install/Service - Service - Frame - Options - "WLAN" or "All").
11. Enter the WLAN service menu (Monitor Setup - Install/Service (16-4-34)- Service (26-23-8)- Frame - Network - WLAN - WLAN Config).
Enter appropriate Network ID, WEP Algorithm, Key ID and Encryption Key using the corresponding menu selections.
For selecting allowed communication channels, enter the appropriate Channel Mask using the corresponding menu selection. See Appendix C. Channel Mask Selections.
Save the monitor’s wireless LAN configuration by selecting Save Configuration - Save.
NOTE: The settings must match up with the ones that are defined in the Access Point configuration. Refer to the related site configuration documentation for correct settings, if necessary.
12. Check all installed options on the monitor software device plate and attach the Virtual ID sticker under the device plate at the monitor rear panel.
13. Test the monitor on Datex-Ohmeda Network.
14. Fill out all necessary documentation regarding the upgrade.
3.3.12 Installing the S/5 Recorder Upgrade, U-CMREC1
NOTE: Only S/5 Compact Anesthesia Monitor rev. 02 or later and S/5 Compact Critical Care Monitor rev. 02 or later can be upgraded with this S/5 Recorder Upgrade kit.
1. Turn the monitor off.
2. Disconnect the power cord.
3. Push the Service Reset button for at least 5 seconds until the service reset indicator LED turns off (the fan should turn off).
4. Remove the plate that covers the recorder installation place in the side panel.
5. Open the recorder (by pressing the button on the right upper corner at the front of the recorder).
6. Insert the recorder from outside to the side panel (opening button upwards).
28
Document no. M1144951-004
7. Attach the recorder to the side panel with two screws at the back of the recorder (the screws can be accessed inside the recorder).
8. Check that the recorder and the side panel are aligned and the recorder tightly attached.
Figure 8 Installation of the U-CMREC1
9. Attach the device plate to the back cover of the monitor, near the other device plates.

3.4 S/5 Remote Controller, K-CREMCO

System installation
To connect Remote Controller, K-CREMCO, to a Compact Monitor:
1. Make sure that the monitor is switched to standby.
2. Connect the K-CREMCO to the 5-pin DIN connector X9 that is located on the right side of the Compact Monitor display unit.

3.5 Anesthesia Record Keeping Keyboard, K-ARKB

To connect Anesthesia Record Keeping Keyboard, K-ARKB, to a Compact Monitor:
1. Make sure that the monitor is switched to standby.
2. Connect the K-ARKB with an appropriate cable to the 5-pin DIN connector X9 that is located on the right side of the Compact Monitor display unit.

3.6 ARK Barcode Reader, N-SCAN (optional)

3.6.1 Connection to Compact Monitor
1. Make sure that the monitor is switched to standby.
2. Connect the N-SCAN with an appropriate cable to the 5-pin DIN connector X9 that is located on the right side of the Compact Monitor display unit.
29
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Figure 9 ARK Barcode Reader, N-SCAN connected to Compact Monitor

3.7 S/5 Compact Airway Modules, E-xxxx / M-xxxxx

This chapter provides information for installing Compact Airway Modules E-xxxxx.
Figure 10 Compact Airway Module, E-XXXX
Connection to frame
1. Ensure that the module is properly orientated (i.e. module release latch facing downward).
30
Document no. M1144951-004
System installation
SAMPLE GAS OUT
2. Align the module insertion guide slot with the insertion guide.
3. Push the module into the frame until it clicks.
3.7.1 Sample gas exhaust
Preventing operation room pollution
When N2O or volatile anesthetics are used, pollution of the operation room by these gases should be prevented. Connect the sample gas outlet of the monitor to the scavenging system or return it to the patient circuit.
NOTE: The sample gas from the E-miniC/ M-miniC can only be connected to the scavenging, not returned to the patient circuit.
Connect the sample gas outlet of the monitor to the scavenging system either
through the ventilator, or
directly to the vacuum scavenging system.
CAUTION Strong scavenging suction may change the operating pressure of the Airway
Module and cause inaccurate readings or internal damage. To prevent this, there must for example be an opening to room air.
Connect the sample gas outlet only to an open scavenging system where gas is removed in room pressure.
Scavenging through the ventilator reservoir
Figure 11 Scavenging through ventilator reservoir
1. Connect an exhaust line to the sample gas outlet on the module’s front panel.
2. Attach the other end of the line to the ventilator reservoir. Make sure that the reservoir tube diameter is at least 2 - 3 times larger than the exhaust line.
Scavenging through the anesthesia gas scavenging system
Anesthesia machines are equipped with an anesthesia gas scavenging system (AGSS), and in some machines the sample gas outlet can be connected directly to that.
For example, connect the sample gas outlet to the S/5 Avance:
31
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Figure 12 Connecting the gas module to the scavenging connector of S/5
Avance
NOTE: Refer to the anesthesia machine’s documentation to find out where and how the sample gas can be connected.
Connecting directly to the vacuum scavenging system
1. Connect the exhaust line to the monitor’s sample gas outlet.
2. Connect the exhaust line only to an open scavenging system where gas is removed at room pressure.
NOTE: Do not connect the monitor directly to a vacuum scavenging system.
Returning sampling gas to the patient circuit
The sampling gas can also be returned to the patient circuit. If you use the S/5 Anesthesia Delivery Unit (ADU), you need an optional adapter connected to the patient breathing tubes.
Take special care when returning sample gas to the patient circuit. For further information, please contact your GE Healthcare distributor.
NOTE: If E-miniC is being used, do not return the sample gas to the patient circuit.
32
Document no. M1144951-004
Figure 13 Sample gas returned to patient circuit in ADU
SAMPLE GAS OUT
SCAVENGING
System installation
3.7.2 Troubleshooting
If a problem occurs during a functional examination, check the components of the monitor according to the following troubleshooting chart. If the problem persists, please refer to Part II of this Technical Reference Manual.
Trouble Treatment
Nothing functions. Unplug and re-plug Remote Controller Cable. Also confirm that the cable is intact.
Unplug and re-plug the Power Cord. Also confirm that the cable is intact. Confirm that the fuses are intact.
A plug-in module does not function.
Compact Airway Module does not function.
Remove and replace the module. Confirm that the desired parameters are configured to be displayed.
Confirm that ‘Occlusion’ or ‘Calibrating Gas Sensor’ messages are not displayed. Confirm that a D-fend water trap and a sample tube are attached. Confirm that the desired parameters are configured to be displayed. Remove and replace the module.
33
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For your notes:
34
Document no. M1144951-004

4 Interfacing

External devices can be interfaced with the S/5 Compact Anesthesia and S/5 Compact Critical Care Monitors via the Compact Monitor’s serial port, via the Interface Module, E-INT / M-INT, and via the Device Interfacing Solution, N-DISxxx.
The serial port can be used for interface with:
S/5 Anesthesia Delivery Unit
Printers
Computers
An E-INT / M-INT can be used for interface with:
Datex-Ohmeda monitors
Various other manufacturers’ monitors
Various anesthesia machines
Printers and computers can be interfaced via the monitor’s serial port.
Device specific N-DISxxx modules can be used with:
Ventilators/ anesthesia machines
Interfacing
Heart-lung machines
Monitors
Blood-gas analyzers

4.1 Interfacing external monitors via Interface Module, E-INT / M-INT

It is possible to interface Datex-Ohmeda monitors, Critikon Dinamap 1846SX, Abbott Oximetrix 3, Baxter Explorer and Vigilance, Nellcor N-100, N-200 and N-1000 to the S/5 Compact Anesthesia Monitor or to the S/5 Compact Critical Care Monitor via Interface Module, E-INT / M-INT. The parameters that are transferred from external monitors are summarized in tables:
Table 4 Transference of parameters, Datex-Ohmeda monitors and Table 5 Transference of parameters, external monitors.
The Interface Module, E-INT / M-INT, has two serial/analog connectors (X7 and X8).
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the requirements of local authorities.
35
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Table 4 Transference of parameters, Datex-Ohmeda monitors
Device Waveforms
Numerics Alarms
(analog)
Cardiocap E-INT / M-INT -> CO2Et&Fi Airway gases,
Respiration rate, SpO2, Pulse rate
,
Capnomac Capnomac II
E-INT / M-INT -> CO Pleth
Capnomac Ultima E-INT / M-INT -> CO
Normocap 200
E-INT / M-INT -> CO
Normocap 200 OXY
Oscar Oscar II
E-INT / M-INT -> CO Pleth
Oscar OXY
Satlite
E-INT / M-INT -> CO Satlite II Satlite Plus
Satlite Trans None SpO
Et&Fi Airway gases,
2
Respiration rate
Et&Fi Airway gases,
2
Respiration rate, Spirometry, SpO
Pulse rate, Pleth amplitude
Et&Fi Airway gases,
2
Respiration rate
,
Et&Fi Airway gases,
2
Respiration rate, SpO2, Pulse rate,
Pleth amplitude
SpO2, Pulse rate,
2
Pleth amplitude
, Pulse rate,
2
Pleth amplitude
2
None
None
CO agent, Respiration
,
rate, Apnea, Occlusion, SpO
Pulse rate
None
None
None
None
, O2, Anesthesia
2
,
2
Table 5 Transference of parameters, external monitors
Device Waveforms Numerics Alarms
Critikon Dinamap 1846SX
Abbott Oximetrix 3
Baxter Explorer
Baxter Vigilance
Nellcor N-100 N-200 N-1000
None NIBP None
None SvO
None C.O., SvO
/SaO2, CO None
2
REF,
2,
Tblood
None C.O., SvO
C.C.O.,
2,
Tblood
Pleth (analog) SpO
, Pulse rate None
2
4.1.1 Connection to external Datex-Ohmeda monitors
Use the INT-External Device Cable.
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT. Tighten the finger screws.
None
None
36
Document no. M1144951-004
Interfacing
3. Connect the 25 pin D-connector to the corresponding connector on the other monitor. Tighten the finger screws.
4.1.2 Connection to Critikon Dinamap 1846SX, Abbott Oximetrix 3 and Baxter Explorer
Use the INT-External Device Cable.
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT. Tighten the finger screws.
3. Connect the 25 pin D-connector to the connector on the external monitor. Tighten the finger screws.
4.1.3 Connection to Baxter Vigilance
Use the INT-Baxter Vigilance Cable.
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT. Tighten the finger screws.
3. Connect the other 9 pin D-connector to a corresponding connector on the external monitor. Tighten the finger screws.
Communication parameters are: baud rate 19200, no parity (none), data bits 8, stop bits 1. Mode of communication port: IFM out.
4.1.4 Connection to Nellcor N-100 and N-1000
Use the Monitor-Nellcor Cable.
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT. Tighten the finger screws.
3. Connect the other connector to the corresponding connector on the external monitor. Tighten the finger screws. Connect also the mono connectors on the external monitor.
4.1.5 Connection to Nellcor N-200
Use the Monitor-Nellcor Cable.
1. Make sure that the power to both monitors is turned off.
2. Connect one of the connectors to the corresponding connector on Interface Module, E-INT / M-INT. Tighten the finger screws.
3. Connect the other connector to the corresponding connector on the external monitor. Tighten the finger screws. Connect also the mono connectors on the external monitor. Connect also the mono connectors on the external monitor.
37
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors

4.2 Interfacing external bedside devices via Device Interfacing Solutions, N-DISxxx

The Device Interfacing Solution, N-DISxxx provides means for transferring physiological, waveform and event data from various bedside patient care devices to the Datex-Ohmeda monitoring system. The real-time and trended data can be displayed on the monitor screen and used for record keeping purposes. The interfacing module reads the data coming from the external device, converts it to a suitable format and sends it to the monitor.
See the following table of DIS modules and devices that you can interface with the Device Interfacing Solution.
NOTE: The Device Interfacing Solution (DIS) is only compatible with the S/5 Compact Anesthesia and S/5 Compact Critical Care Monitor when the monitor has S/5 Monitor software of rev. 02 or later installed.
Table 6 DIS modules and interfaced devices
Device Ventilators
N-DISEV4
N-DISPRIM
N-DIS7200
N-DIS840
N-DISS300
N-DIS7900
N-DISAEST
N-DISVENT
5
1
Evita 4
1
Primus
NOTE: Not available in the US
7200 Series Ventilator System
840 Ventilator System
Servo Ventilator 300
3
7900 SmartVent Ventilator
4
Aestiva/5
2
2
4
S/5 Aespire4
Aestiva/5
Aisys Carestation
S/5 Avance
Centiva/5
Engström Carestation
4
4
4
4
4
1 Trademark of Dräger Medical AG & Co 2 Trademark of Nellcor Puritan Bennet Inc 3 Trademark of Maquet Critical Care AB par t of the Getinge Group (previously trademark of Siemens) 4 Trademark of GE Healthcare Finland Oy 5 Replaced by N-DISVENT
N-DISOXIM3
N-DISQVUE
N-DISA2000
N-DISVIGIL
N-DISPICCO
N-DISRGM
38
Document no. M1144951-004
Device Monitors
Oximetrix 3
QVue /Q2
A-2000 Bispectral Index Monitoring System
Baxter-Vigilance
PiCCO-Technology4 NOTE: Not available in the US
RGM Monitor
1
1
2
3
5
Interfacing
N-DISTONO
N-DISWHITE
5
Tonocap
Capnomac, Capnomac II
Multicap, Normocap, CD2-O2
Capnomac Ultima
Normocap CD-200
5
5
5
5
Oscar Oxy, Cardiocap 1GS, Cardiocap 2GS
Satlite, Satlite Trans, Satlite Plus
5
5
1 Trademark of Hospira Inc. (previously trademark of Abbott Laboratories) 2 Trademark of Aspect Medical Systems 3 Trademark of Edwards Lifesciences Corporation 4 Trademark of Pulsion Medical Systems 5 Trademark of GE Healthcare Finland Oy
Device Blood gas analyzers
N-DISOPT
AVL Opti CCA
1
Device Heart-lung machines
N-DISHL20
Jostra HL-20
1 Trademark of Diamond Diagnostics Inc 2 Trademark of MAQUET GmbH & Co. KG part of the Getinge Group
For specific information on parameters transferred from the interfaced device to the Datex-Ohmeda monitor and the applicable software versions of the device refer to the Installation guide accompanying each DIS module.
2
4.2.1 Device Interfacing Solution components
The Device Interfacing Solution consists of:
a device specific interfacing module
a device specific cable
a bus cable
a connector for another bus cable
label specifying the external device
4.2.2 Connections
Connect the device specific cable to the external device and the bus cable to the monitor's DIS connector or to another interfacing module.
39
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
(1) label specifying the external device (2) LED indicators (3) black bus cable from another
interfacing module, if needed
(4) grey device specific cable to the
communication port of the external device
(5) black bus cable to the monitor’s DIS
connector (or to another interfacing module)
Figure 14 Connection cables and LED indicators
WARNING The Compact Monitor, interfacing modules and interfaced devices must be
situated in the same patient environment (as defined in IEC 60601-1-1).
WARNING Connecting electrical equipment together or using the same extension cord
for more than one device may cause their leakage currents to exceed the limits specified in relevant safety standards. Always make sure that the combination complies with the international safety standard IEC 60601-1-1 for medical electrical systems and with the requirements of local authorities.
WARNING The manufacturer guarantees a reliable functioning of the devices with
tested software versions only. Always refer to the Installation guide accompanying the DIS module and verify the compatibility before use.
4.2.3 Mounting
The DIS module can be mounted on the side panel of the external device. Also IV pole placement is possible.
NOTE: As the Device Interfacing Solution works only with the device specified in the label of the interfacing module, it is recommended that the interfacing module always travels along with the external device.
For mounting accessories, please refer to the “Supplies and Accessories” catalog. See the following figure for an example of a device interfacing.
40
Document no. M1144951-004
1
4
Interfacing
Figure 15 An example of interfacing external devices with Device Interfacing
Solution
(1) Datex-Ohmeda S/5 Compact Anesthesia Monitor (with software L-CANE02(A) or later) (2) Aestiva/5 anesthesia machine (3) RGM monitor (4) Interfacing module
NOTE: You can connect up to ten (10) interfacing modules to one system simultaneously. Check the maximum number of modules: one meter cable = max. four ten modules, three meter cable = max. three modules, six meter cable = one module.
CAUTION Make sure that the interfacing module is always used in vertical position to
prevent water from entering the module.
CAUTION Make sure that you are connecting the interfacing module to the device
specified in the label.
4.2.4 Selecting the external device
1. Turn off the monitor.
2. Turn off the external device.
3. Connect the interfacing module to the monitor’s connector for N-DIS or to another interfacing module.
4. Connect the device specific cable to the external device and turn the external device on.
5. Turn the monitor on. The monitor identifies the connected device automatically.
41
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
4.2.5 Functional check
There are two ways to check the function of the Device Interfacing Solution:
1. Press the
2. Select Interfacing and open the Status Page menu. The status page shows you the current communication status of the interfacing module (1 - 10).
NOTE: The status message ‘Connected appears on the Status Page after you have connected the external device to the interfacing module and turned it on. Note also that the monitor and the interfacing module must be operational.
Monitor Setup key.
Check the LED indicators on the interfacing module (the green LED indicates physical
connections, the yellow LED software selections)
:
GREEN YELLOW INDICATION
lit dark
dark
lit lit
dark dark
lit
Physical connections between the monitor, interfacing module and external device are in order and the device has been selected in the menu.
There is something wrong with the physical connections between the monitor, interfacing module and external device. The external device has not been selected in the menu.
Physical connections between the monitor, interfacing module and external device are in order but the external device has not been selected in the menu.
The interfacing module is not connected to the monitor.
4.2.6 Selecting the parameter data source
Select the external device via Monitor Setup - Interfacing menu:
Select the desired parameter, for example Ga s e s .
Select the desired source by name, for example Aestiva.
NOTE: The name of the device is visible on the list only if the device is correctly connected.
Detailed information about interfacing module related mountings, connections and settings is included in the installation guides that are delivered with the interfacing modules.
42
Document no. M1144951-004
Interfacing

4.3 Interfacing Datex-Ohmeda Anesthesia Delivery Unit

It is possible to interface the Datex-Ohmeda Anesthesia Delivery Unit to the S/5 Compact Anesthesia Monitor via the serial interface port. The data link is bi-directional. The parameters transferred to the Anesthesia Delivery Unit are summarized in Table 7 and the events transferred to the S/5 Compact Anesthesia Monitor are summarized in Table 8. The events are transferred continuously once a minute or during the system check procedure and appear in the trend pages or in the anesthesia record. The transfer of events requires record-keeping software.
4.3.1 Interconnection
Use the ADU-AS/3 Monitor Cable.
4.3.2 Setting interfacing parameters on the S/5 Anesthesia Delivery Unit
1. Press the Setup key.
2. Select Install password 10.
3. Select Interfacing and choose S/5 AM.
4. Press the
Normal Screen key.
4.3.3 Setting interfacing parameters on the S/5 Compact Anesthesia Monitor
No settings are required on the S/5 Compact Anesthesia Monitor.
Table 7 Parameters transferred from S/5 Compact Anesthesia Monitor to S/5
Anesthesia Delivery Unit
Waveforms Numerics
CO
2
Et & Fi CO2 % Et & Fi O O
2
(I-E) Et & Fi N
Et & Fi AA% Vol Exp MV Vol Exp TV
%
2
% difference
O%
2
43
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Table 8 Events transferred from S/5 Anesthesia Delivery Unit to the
S/5 Compact Anesthesia Monitor
Events transferred once a minute
Ventilator settings Ventilation (mode)
Tidal Volume Minute Volume Resp. Rate I:E Times: inp I:E Times: exp InspPause
Ventilator measurements Ppeak
Pplat Peep Pmin
Fresh gas settings Agent name
Agent % in fresh gas Total flow O
flow
2
O flow
N
2
Air flow
Events transferred during system check
System check test results Gas Delivery: Agent
Gas Delivery: N Leak Tests: AUTO
AUTO Leak (ml/min) Leak Test: MAN MAN Leak (ml/min) Number of confirmed checklist items Bypass Check
O
2

4.4 Interfacing Dräger Cicero, Cato, Julian and Narkomed 2C (by NAD)

It is possible to interface Dräger Cicero and Cato to the S/5 Compact Anesthesia Monitor or to the S/5 Compact Critical Care Monitor via the Interface Module, E-INT / M-INT.
The parameters transferred to the S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor are summarized in Table 9 and Table 10.
NOTE: Some of the summarized parameters (waveforms or numerics) may not be available in all device versions.
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the requirements of local authorities.
4.4.1 Interconnection
Use the INT-External Device Cable, except for Julian and Cicero EM, in which case use the INT-Julian/Cicero EM cable.
44
Document no. M1144951-004
Interfacing
1. Make sure that the power to both monitors is turned off.
2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT. Tighten the finger screws.
3. Connect the 25 pin D-connector to a corresponding connector on the anesthesia machine. The connector is labeled 'RS-232-C'. Tighten the finger screws.
4.4.2 Setting communication parameters
Set the communication parameters for Cicero and Cato to RS-232: 9600, e, 8, 1 (MEDIBUS 3.00). In all cases please refer to the documentation provided by Dräger.
4.4.3 Setting interfacing parameters on the S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor
1. Press the Monitor Setup key.
2. Select Install/Service (password 16-4-34).
3. Select Service (password 26-23-8) - Parameters - More...
4. Select Interface and combine a required parameter with the external monitor. Two letters, al, denote alarm integration. The selection will be automatically stored in permanent memory.
5. Press the
Table 9 Parameters transferred from Dräger Cicero monitor to S/5 Compact
Normal Screen key.
Anesthesia Monitor or S/5 Compact Critical Care Monitor
Selection Waveforms Numerics States
Gases / SSS CO
(kPa, mmHg, %) CO2 Fi & Et (kPa, %)
2
Fi & Et (%)
O
2
O Fi & Et (%)
N
2
RR (CO
) (1/min)
2
None
AA Fi & Et (%) Tidal Volume insp. (1) Minute Volume exp. (1) Ppeak (mbar) Pplat (mbar), PEEP (mbar)
SpO
2
Pleth (% full scale) SpO2 (%)
Probe Off
Pulse rate (1/min)
45
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Table 10 Parameters transferred from Dräger Cato, Julian and Narkomed 2C
(NAD) monitor to the S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor
Selection Waveforms Numerics States
Gases / SSS CO2 (kPa, mmHg, %) CO2 Fi & Et (kPa, %)
O
Fi & Et(%)
2
N2O Fi & Et (%) RR (CO2) (1/min) AA Fi & Et (%)
Tidal Volume insp. (1) Minute Volume exp. (1) Ppeak (mbar) Pplat (mbar); not in Narkomed 2C PEEP (mbar)
SpO
2
Pleth (% full scale) SpO2 (%)
Pulse rate (1/min)
NOTE: The transference of parameters illustrated in Table 7 and Table 8 represents the maximum. Some of these parameters may not be available in all device versions.
NOTE: The transference of waveforms requires the MEDIBUS real-time extension and, in Cato monitors, waveforms are available only on port 1.
None
Probe Off
46
Document no. M1144951-004
Interfacing

4.5 Interfacing printer

It is possible to interface a laser printer (serial or parallel) with the S/5 Compact Monitors via the monitors’ serial port. The printer should be PCL5 compatible and should contain at least 2 MB of memory. Parallel printers require the use of Serial-to-Parallel Converter, order code 78030, model PI 130-R2 or PI 1115A.
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the requirements of local authorities.
WARNING Connecting the power supply cord of the printer to the wall power outlet
may cause the printer leakage current to exceed the limit specified for medical equipment. A printer must be supplied from an additional transformer providing at least basic isolation (isolating or separating transformer).
4.5.1 Setting S/5 Compact Monitor interface for printers
1. Press the Print/Record key.
2. Select Printer Connection.
3. Select Serial.
4. Press the
Normal Screen key.
4.5.2 Connection to serial printers
A serial printer is connected to the serial port connector X11 on the frame unit. Contact your authorized GE Healthcare distributor for advice on suitable serial printers.
4.5.3 Connection to parallel printers
A parallel printer is connected to the serial interface port connector X11 on the frame unit via the Serial-to-Parallel Converter, order code 78030, model PI130-R2 or PI1115A.
Contact your authorized GE Healthcare distributor for advice on suitable parallel printers.
4.5.4 Installing the Serial-to-Parallel Converter
Order code 78030, model PI130-R2
A common serial modem cable, gender changer and parallel printer cable are required when connecting the serial to parallel converter between the monitor and a parallel printer. The converter gets power from the connected devices.
1. Make sure that power is switched off on both devices.
2. Connect a standard PC-to-parallel-printer cable (order code 713701) to the parallel printer.
47
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Position
Flow Control: hardware
LED: Enabled
Parity/D, Bits/S, Bit: None/8/1
D.Rate (kbps): 115,2
1
OFF
2
ON
3
OFF4OFF
5
OFF
6
OFF7ON8OFF
DB9F plugs into CM1
Gender changer plugs into converter
DB25M end plugs into gender changer
Centronics
end plugs
into printer
Standard Async Modem Cable
Serial to Parallel
Converter
DB25F/DB25F
Gender Changer
Standard Parallel Printer
Adapter Cable
Parallel printer
1234
3. Check the DIP-switch settings on the converter (order code 78030, model PI130-R2):
4. Connect a DB25 female-to-female gender changer (order code 78032) between the printer cable and converter.
Figure 16 Connecting S/5 Compact Monitor to printer, converter model
PI130-R2
5. Connect a common RS-232 modem cable’s (order code 78031) DB9 connector to the serial port connector X4 on the monitor, and the DB25 connector to the converter.
Order code 78030, model PI1115A
A common serial modem cable, power adapter and parallel printer cable are required when connecting the serial to parallel converter between the monitor and a parallel printer. The converter gets power from an external power supply and requires an external power adapter.
NOTE: The order code 78030 includes the new serial to parallel converter PI1115A and the European power adapter PS1002-EU (230 VAC/50 Hz). Power adapters for US, UK or Asian markets must be locally sourced from Black Box Corporation through its international distributor network, see http://www.blackbox.com - Contact us - Contacts by Country.
To connect the S/5 Compact Monitor to printer:
1. Make sure that power is switched off on both devices.
2. Connect a standard PC-to-parallel-printer cable (order code 713701) to the parallel printer.
3. Check the DIP-switch settings on the converter (order code 78030, model PI1115A).
Setting Position
12345678
115.2K Baud ON ON ON
XON/XOFF OFF
8 Data bits OFF
48
Document no. M1144951-004
Interfacing
DB9F plugs into CM1
Centronics
end plugs
into printer
Standard Async Modem Cable
Serial to Parallel
Converter
Standard Parallel Printer
Adapter Cable
Parallel printer
cam p1 inst serial parallel conv p1115A a
PS1002-EU, Power Adapter, 230 VAC Europe, for PI1115A (included in p/n 78030), plugs to power supply
p/n 78030 (PI1115A) p/n 713701
p/n 78031
No parity OFF
Even parity OFF
Serial to Par. OFF
4. Connect a common RS-232 modem cable’s (order code 78031) DB9 connector to the serial port connector X4 on the monitor, and the DB25 connector to the converter.
5. Connect the converter to external power supply using the power adapter PS1002-xx.
Figure 17 Connecting S/5 Compact Monitor to printer, converter model PI
1115A
4.5.5 Connection to printer Setting the printing parameters
Refer to the documentation provided with the printer.

4.6 Interfacing computer

A computer is connected to the serial port connector X11 on the frame unit. Contact your authorized GE Healthcare distributor for further advice on computer interface.
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the requirements of local authorities.
WARNING Connecting the power supply cord of the computer to the wall power outlet
may cause the computer leakage current to exceed the limit specified for medical equipment. A computer must be supplied from an additional transformer providing at least basic isolation (isolating or separating transformer).

4.7 Output signals

Analog/ digital output signals on the connectors X4 and X8 can be used for interfacing with other devices. The pin assignments are illustrated in tables/pictures below.
49
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
5
9
1
6
Table 11 Coding element connector, X4
9 pin female D-connector Pin Signal
Table 12 Synchronization connector, X8
4 pin female connector Pin Signal
4.7.1 Digital outputs
1 2 3 4 5 6 7 8 9
1 2 3 4
IDCS1 (chip select) IDCL (clock) IDDI (data in) IDDO (data out) IDPE (protect enable) +5Vdc Direct ECG Nurse call GND
GND Def_Sync Direct ECG Pressure Out
The digital output signals are as follows:
Defibrillation Sync (X8 pin 2)
The defibrillation sync signal is generated by the ECG. When activated, the signal is set to a high level and then set back to a low level after 10 ms. The signal is regenerated only after returning to the low level. The high level ranges from 2.8 to 5 V, while the low level ranges from 0 to 0.8 V. The delay from the R wave peak to the start of the signal is 35 ms maximum.
Nurse Call (X4 pin 8)
The Nurse Call signal is generated by red and yellow alarms. When activated, the signal is set to the high state and remains at the high state until the alarm situation is over or the
ALARM
key is pressed. The high level range is from 2.8 to 5 V, while low state range is from 0 to
0.8 V.
4.7.2 Analog outputs
Direct ECG (X4 pin 7 and X8 pin 3)
The signal requires input impedance of 100 kΩ.
NOTE: The ECG signal is based on the ECG measurement of the Hemodynamic Modules, M-ESTPR (all variations), M-NES(12)TPR (all variations), E-PRESTN/ M-PRESTN (all variations) and
SILENCE
Delay (max.): 15 ms Gain ECG (out)/ECG (in): 1 V/1 mV Pacer: 5 V and 2 ms pulse
50
Document no. M1144951-004
Interfacing
E-PSM / E-PSMP. The ECG signal from the modules is channel 1 (ECG1). The channels 2 and 3 (ECG2, ECG3) are not transmitted. Make sure that the signal in channel 1 (ECG1) is good enough (extensive QRS for IABP synchronization). For further information, please contact your authorized GE Healthcare distributor.
Pressure out (X8 pin 4)
P1 from Hemodynamic Module
NOTE: With V/100 mmHg, originally ranging from 0 to 300 mmHg and with a delay of approximately 25ms. The signal requires an input impedance of 100k.
monitor software version 03 or earlier, the Invasive pressure output signal is 1
4.7.3 S/5 Pressure Temp Module, E-PT or M-PT, output signals
The signal output connector on the Pressure Temp Module, E-PT or M-PT, can be used to interface some models of IABPs to the S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor. The pin assignments are illustrated in Table 13. Please contact your local distributor for more information.
WARNING Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the requirements of local authorities.
Table 13 Signal output connector pin assignments
Pin Signal
1 2 3 4
ECG out, ECG1 Pressure out, P3 ECG out, ECG1 Ground
Output signals
ECG out, ECG1 1V / 1mV Pressure out, P3 1V / 100mmHg ECG out, ECG1 1V / 1mV
51
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For your notes:
52
Document no. M1144951-004

5 Functional check

These instructions include procedures for a functional check for Datex-Ohmeda S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor. The functional check is recommended to be performed after monitor installation.
These instructions include a “Functional check form, Datex-Ohmeda S/5 CAM, CCCM” to be
Functional check
filled in when performing the procedures. The symbol
performed procedure should be signed in the check form. The procedures should be performed in ascending order, bypassing those that are not applicable for a particular monitor.
All menu selections related to Datex-Ohmeda products are written in following typeface: e.g. Parameters - Gas Unit.
As you enter the service menus, you need the following passwords:
" in the instructions means that the
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
In case you evaluate the measurement accuracy with a patient simulator, add simulator’s accuracy specification to the one of the monitor.
An electrical safety check and a leakage current test are recommended to be performed prior to the monitor installation.

5.1 Recommended tools

NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the specified calibrations and adjustments to ensure accuracy.
For product(s) Tool Order No.
Airway modules
Compact Airway Module, E-CAiO(VX)/ M-CAiO(VX)
Calibration gas and regulator 755583/755534*
Compact Airway Module, E-COVX/ M-COVX
Compact Airway Module, E-CO(V), M-C, M-CO(V)
E-miniC/ M-miniC Calibration gas and regulator 755580/755534*
Compact Airway Module, E-COVX, E-CAiOVX, M-COVX, M-CAiOVX
All Airway modules w/ (V) Spirometry tube, 2 m 890031
Hemodynamic modules
Hemodynamic modules w/ (P), M-NIBP
Hemodynamic modules w/ (P)E-PSMP
Calibration gas and regulator 755587/755534*
Calibration gas and regulator 755581/755534*
Sampling line, 2 m 73318
D-lite 733950
NIBP cuff and hose 2753E (cuff)
877235 (hose)
Pressure manometer
InvBP transducer 70077-001
53
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For product(s) Tool Order No.
Multi-Link ECG accessories, IEC
E-PSM(P)/ E-PRESTN/ M-PRESTN
- Multi-link 3-leadwire set 412682-003
- Multi-link 5-leadwire set 412681-003
- Multi-link 5-leadwire set, C2-C6 416467-004
E-PRESTN/ M-PRESTN w/ (E12) - Multi-link 12-lead ECG trunk
416035-002
cable
or Multi-Link ECG accessories, AHA
E-PSM(P)/ E-PRESTN/ M-PRESTN - Multi-link 3-leadwire set 412682-001
- Multi-link 5-leadwire set 416681-001
- Multi-link 5-leadwire set, V2-V6 416467-003
E-PRESTN/ M-PRESTN w/ (E12) - Multi-link 12-lead ECG trunk
416035-001
cable
M-NESTPR 3-lead ECG trunk cable 545302
5-lead ECG cable 545303
M-NESTPR w/ (E12) 10-leadwire ECG cable 545324
E-PSM(P)/ E-PRESTN SpO
M-PRESTN SpO
finger probe OXY-F-UN
2
Interconnect Cable OXY-ES3
SpO
2
finger probe OXY-F4-N or SAS-F4
2
E-NSATX 40xxxxx-001 Nellcor OxiMax Spo2
interconnect cable
E-NSAT/ M-NSAT Nellcor SpO
finger probe with
2
DOC-10 cable
E-OSAT OxyTip® + Integrated Finger Sensor OXY-F4-N
Hemodynamic modules w/ (T) Temperature test set 884515
Hemodynamic E-Modules w/ (NIBP) Adult NIBP cuff hose with cuff ID 2021285-001
NIBP cuff 2753E
Pressure manometer
Infant cuff hose without cuff ID 414874-001
Hemodynamic M-Modules w/ (NIBP)
Adult NIBP cuff 572435
Adult NIBP cuff hose 877235
E-COP/ E-COPSv/ M-COP/ M-COPSv C.O. Catheter cable 16590
E-COPSv/ M-COPSv SvO
simulator 890121
2
54
Document no. M1144951-004
For product(s) Tool Order No.
ECG
Functional check
Tonometry Module, E-TONO/ M-TONO
Calibration gas and regulator 755580/755534*
Sampling line 733251
Tonometrics
TM
catheter
TONO-14F or another suitable TONO-_F catheter
Luer plug
Pressure manometer
BIS Module, E-BIS/ M-BIS BIS simulator or
BIS Sensor simulator
900509 900508
E-ENTROPY/ M-ENTROPY Entropy simulator N-ES
Entropy sensor cable 8002964
MemCard option MemCard – Data or Menu 893860 (Menu, English)
887045 (Data, English)
NOTE: * Ensure that the calibration gas and regulator are functioning properly before calibration. Perform annual maintenance on the regulator as required. For more information see section “Adjustments and calibrations” in Compact Airway modules slot in E-Modules Technical Reference Manual.
For details on recommended accessories see “Supplies and Accessories” catalog.
5.1.1 Hemodynamic patient simulators
The following tables present the patient simulators’ compatibility with each hemodynamic module, and the accessories needed:
Table 14 Patient simulators’ compatibility with each hemodynamic module
Patient simulator
Module Parameter
E-PSM(P), E-PRESTN ECG Ok Multilink ECG acc. Not compatible
T 402015-004 402015-004 and
InvBP Ok M1010858
M1010831 MedSim 874027
M1010832
55
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
NIBP
ECG
SpO
T1
T2
P1
P2
2
Auto
On/Off
Start
Cancel
Zero
P1
Zero P2
T2
P1
P2
T1
NIBP
ECG+ Resp
SpO
2
Patient simulator
Module Parameter
M1010831 MedSim 874027
M-PRESTN ECG Ok Multilink ECG acc. Not compatible
TOk M1010832 M1010832
InvBP Not compatible Ok Ok
M-NESTPR ECG Not compatible 300 series acc. Ok
12 lead ECG Not compatible 300 series acc. Not compatible
TOk M1010832 M1010832
InvBP Not compatible Ok Ok
Table 15 Adapter cables for hemodynamic patient simulators
Patient simulator Adapter cables for simulators
Hemodynamic patient simulator - Dual temperature adapter cable 402015-004
Hemodynamic patient simulator - Dual Inv.BP adapter cable 2005772-001
Medsim - Temperature adapter cable M1010832
Medsim - Inv.BP adapter cable M1010858
Lionheart & MPS450 - Temperature adapter cable M1010846
Lionheart & MPS450 - Inv.BP adapter cable M1010862

5.2 Visual inspection

Make sure that the monitor is switched to standby. Disconnect the mains power cord from the monitor. If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable from the monitor. Remove any memory cards, if installed.
1. Check all units visually Check that all parts are intact and that the cables and screws are connected and
tightened properly. Especially check the following parts:
the display unit locking screw is in a locking position.
sampling line is connected to the Airway Module, if installed.
Check that modules go in smoothly and lock up properly in all module slots.
56
Document no. M1144951-004
Functional check
CAUTION Ensure that the module is properly orientated (i.e. module release latch facing
downward) before insertion.
"

5.3 Functional inspection

WARNING Handle the water trap and its contents as you would any body fluid.
Infectous hazard may be present.
5.3.1 General
1. Connect the mains power cord. Check that the LED above the main switch is lit continuously (full charged batteries) or periodically (charging batteries).
2. Switch the monitor on. Check that the monitor starts up properly, i.e. a normal start-up sound is heard from the loudspeaker, the alarm LEDs turn on and off, and the monitoring screen appears. No error messages should appear on the screen.
3. Configure the screen for the parameters that are connected.
4. Enter the Service Menu. When applicable, check from the corresponding Parameters submenu that the Timeouts,
Bad checksums and Bad c-s by mod values of inserted modules are not increasing faster than by 5 per second. Check also that the module memories have passed the internal memory test, i.e. RAM, ROM and EEPROM all state OK. If installed, the recorder should record two lines of start-up information.
"
Preset the measurement settings for those parameters that are connected, for example:
Recorder - Record Waveforms - Waveform 1 - ECG1
- Waveform 2 - P1
- Waveform 3 - P2
Invasive Pressures - P1 ‘ART’ Setup -- Label – ART
- P2 ‘CVP’ Setup -- Label – CVP
- P3 Setup -Label –PA
- P4 Setup -Label -P4
- P5 Setup -Label -P5
- P6 Setup -Label -P6
Pulse Oximetry -Pleth Scale –AUTO
or
Others -SPO2 Setup - Pleth Scale –AUTO Airway Gas -Spirometry View -Scaling -Indep.
-Paw Scale –20
-Flow Scale –15
or
Ventil.-Spirometry Setup -Scaling -Indep.
-Paw Scale –20
-Flow Scale –15
57
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Others -Resp Setup -Size -1.0
- Resp Rate Source – AUTO
- Measurement – ON
- Detection Limit – AUTO
5.3.2 Display
1. Check that the picture on the screen is correct.
"
5.3.3 Keyboard(s)
1. Tests with all the connected keyboards:
Press the
the cursor in the menu moves correspondingly. Select Normal Screen and check that the menu disappears from the screen.
Tests with the Command Board:
Press the
in the menu moves correspondingly. Select Normal Screen and check that the menu disappears from the screen. Check the rest of the menu keys by pressing them one by one.
Tests with the ARK Keyboard:
Enter the Keyboard service menu
Check functioning of the ComWheel.
Press all keys, except
from the loudspeaker, or the Message count value in the service menu increases.
Press the
on the monitor screen.
Tests with the Remote Controller:
Enter the Keyboard service menu.
Check the function of the ComWheel.
Press all keys. Check that each key produces a sound from the loudspeaker, or the
Message count value in the service menu increases.
Monitor Setup key. Turn the ComWheel in both directions and check that
Help key. Turn the ComWheel in both directions and check that the cursor
Modify and Print. Check that each key produces a sound
Modify and Print keys and check that the corresponding menus appear
"
5.3.4 Frame unit
1. Check that the frame fan is running.
"
5.3.5 Display unit
1. Check that the clock on the screen shows correct time. Readjust the time and date, if necessary.
"
58
Document no. M1144951-004
5.3.6 Compact Airway Module, E-CXXXXX/ M-CXXXXX
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1. Check that the fan is running.
2. If the module contains membrane keys on the front panel, press each of the keys for at least one second and check that they are identified.
"
Compact Airway Modules with the Patient Spirometry option
3. Connect a clean spirometry tube and D-lite to the module. Connect the sampling line. Breathe through the wider side of the D-lite. Check that the flow waveform moves downwards when you breathe in and upwards when you breathe out.
"
For all Compact Airway Modules
4. Block the tip of the sampling line with your finger and check that the message ‘Sample line blocked’ appears on the monitor screen within 30 seconds.
5. Detach the D-fend and check that the message ‘Check D-fend’ appears on the monitor screen within 30 seconds.
6. Breathe to the sampling line briefly. Check that the CO2 information is updated on the screen.
Functional check
"
5.3.7 Single width Airway Module, E-miniC/ M-miniC
Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
1. Block the tip of the sampling line with your finger and check that the message ‘Sample line blocked’ appears on the monitor screen within 30 seconds.
2. Detach the Mini D-fend and check that the message ‘Check D-fend’ appears on the monitor screen within 30 seconds.
Breathe to the sampling line briefly. Check that the CO2 information is updated on the screen.
"
5.3.8 Tonometry Module, E-TONO/ M-TONO
1. Press each of the front panel membrane keys for at least one second and check that they are identified.
"
5.3.9 Multiparameter Hemodynamic Modules
ECG and RESP measurements
1. Check the function of the front panel membrane key ECG, Lead, if available.
Document no. M1144951-004
59
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2. Connect an ECG cable to the module. Connect the cable leads to a patient simulator. Check that all ECG and impedance respiration information is shown on the monitor screen as configured on the simulator.
Turn the simulator off. Check that the ‘Asystole’ and ‘Apnea’ messages are displayed.
"
Temperature measurement
3. Check the temperature channels with a patient simulator. Check that temperature measurement information is shown on the monitor screen as
configured on the simulator.
"
Invasive blood pressure measurement
4. Check the function of the front panel membrane keys.
5. Check the InvBP channels with a patient simulator.
6. Zero the InvBP channels and check that the values and waveforms correspond to the simulator settings.
"
SpO2 measurement
7. Connect an SpO2 finger probe to the module. Check that the message ‘Probe off’ is shown when the probe is not connected to a finger.
8. Attach the SpO2 probe to your finger. Check that a reading of 95-99 and a pleth waveform appear on the screen
"
Non invasive blood pressure measurement
9. Check the function of the front panelmembrane keys.
10. Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check that the module identifies the cuff, i.e. the text ‘Adult’ appears in the NIBP digit field for a short time.
Check that the module gives a reasonable measured result.
"
5.3.10 Pressure/Pressure Temp Modules, E-P, E-PT, M-P, M-PT
Invasive blood pressure measurement
1. Check the function of the front panel Zero P3 key.
2. Check the InvBP channel with a patient simulator. Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
"
60
Document no. M1144951-004
Temperature measurement
3. Check the temperature channels with a patient simulator. Check that the temperature measurement information is shown on the monitor screen as
configured on the simulator.
"
5.3.11 Dual pressure Module, E-PP/ M-PP
1. Check the function of the front panel Zero P5 and Zero P6 keys.
2. Check the InvBP channels with a patient simulator. Zero the InvBP channels. Then check that the values and waveforms correspond to the
simulator settings.
"
5.3.12 Cardiac Output Modules, E-COP/ M-COP, E-COPSv/ M-COPSv
Invasive blood pressure measurement
1. Check the function of the front panel Zero P4 key.
2. Check the InvBP channel with a patient simulator. Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
Functional check
"
Cardiac Output measurement
3. Check the function of the front panel Start C.O. key.
"
5.3.13 NIBP module, M-NIBP
1. Check the function of the front panel membrane keys.
2. Attach an adult NIBP cuff onto your arm and perform one NIBP measurement. Check that the module identifies the cuff, i.e. the text ‘Adult ’ appears on the NIBP digit
field for a short time. Check that the module gives a reasonable measured result.
"
5.3.14 Nellcor Compatible Saturation module, E-NSATX/ E-NSAT/ M-NSAT
1. Connect a Nellcor SpO2 finger probe to the module. Check that the message ‘Pulse search ’ is shown. Check also that the shown message changes to ‘Check probe ’ within 30 seconds.
61
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2. Attach the SpO2 probe on your finger. Check that a reading of 95-100 and a proper SpO2 waveform appear.
"
5.3.15 Datex-Ohmeda Oxygen Saturation module, M-OSAT
1. Connect an OxyTip ® + Reusable Finger Sensor such as Integrated Finger Sensor OXY-F4-N to the module. Check that the message ‘No Probe’ in the screen changes to ‘Check probe’.
2. Attach the SpO2 probe on your finger. Check that a reading of 95-100 and a proper SpO2 waveform appear.
"
5.3.16 BIS Module, E-BIS/ M-BIS
1. Connect the BIS module to the monitor frame. Check that ‘Cable off’ is displayed in the BIS waveform field.
2. Connect the DSC to the module. Check that the ‘No sensor’ message appears in the waveform field.
3. Check the function of the front panel membrane keys. Open the BIS Setup menu:
BIS - BIS Setup and check that Test DSC shows PASS.
NOTE: If the sensor is not connected, the
Check Sensor key is inoperative.
"
5.3.17 Entropy Module, E-ENTROPY/ M-ENTROPY
1. Check the function of the front panel membrane keys.
2. Connect the Entropy sensor cable and Entropy simulator to the module. Check that ‘Checking sensor’ text and an image appear in the waveform numeric field. Wait for a while and check that all sensors show PASS.
3. Check that the EntrEEG waveform and RE and SE values appear on the monitor screen.
"
5.3.18 Data Card and Menu Card function
4. Insert a Data card or a Menu card to the slot. Check that the corresponding symbol appears on the monitor screen.
"
5.3.19 Recorder
1. Press the Record Wave module key and check that the module starts recording the selected waveforms. Press the
62
Document no. M1144951-004
Stop module key to stop recording.
2. Check that the quality of the recordings is acceptable.
"
5.3.20 Network connection
1. Check that the Mon-Net cable connector and the Identification plug are clean and intact, then connect them to the frame unit.
Check that the monitor connects to the network, i.e. the network symbol appears on the upper right-hand corner of the screen. Also a message regarding the connected Central should appear in the message field on the screen.
"
5.3.21 Wireless Network Option
1. Check that the green LED on the wireless antenna card blinks and the WLAN signal strenght symbol scrolls between zero and full or stays fixed on the monitor screen.
2. Check that the wireless LAN network symbol appears on the upper right-hand corner of the screen when the monitor connects to the Datex-Ohmeda Network.
Functional check
NOTE: If the monitor does not connect to the Datex-Ohmeda Network, check the WLAN configuration on the monitor and on the network.
"
5.3.22 Interface Module, E-INT/ M-INT
1. Make sure that the monitor receives all necessary parameter data from the connected devices. Check the screen configuration and the related interfacing settings, if necessary.
Monitor Setup – Interfacing
"
5.3.23 Interface module for PSM, E-INTPSM
1. Make sure that the monitor receives all necessary parameter data from the connected devices. Check the screen configuration and the related interfacing settings, if necessary.
Monitor Setup – Interfacing
"
5.3.24 Device Interfacing Solution, N-DISxxx
1. Make sure that the monitor receives all necessary parameter data from the connected devices. Check the screen configuration and the related interfacing settings, if necessary.
Monitor Setup - Interfacing - Status Page
"
Document no. M1144951-004
63
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
5.3.25 General
Switch the monitor to standby
Perform final cleaning
Fill in all necessary documents
"
64
Document no. M1144951-004

6 General troubleshooting

Monitor not functioning
Power cord
connected?
Connect power
cord
Monitor can be
started now?
Batteries were
empty
On/Standby LED
is ON or blinks?
PMB failure.
Replace PMB.
Monitor can be
started now?
Display Unit
failure.
No
Yes
Yes
Yes
No
Possible AC/DC
unit failure.
Replace AC/DC
unit.
No
No
Frame Unit fan is
running?
Yes
No
AC/DC Unit
failure.
Yes
CM1_trbl_Flowchart.vsd
General troubleshooting
Figure 18 S/5 Compact Monitor general troubleshooting flowchart
Document no. M1144951-004
65
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
SERVICE PROCEDURE:
Downloading of Service software or
card onto the CPU board.
Start-up display
appears?
Has the information
regarding monitor software
been updated on the
Service View?
Remove the software card. Perform Factory Reset.
Yes
No
No
Yes
Insert the software card into
the card slot.
Turn the power ON.
Check the software card's connection
in the card slot
and try again.
Wait for 80
seconds.
Try with another
software card.
Please read the
note below.
Replace the CPU board and try again.
OK ?
No
Yes
OK
OK
No
The software card may be defective, or all available software has already been downloaded from the
OK.
A
number of new software.
Upgrade software from a software

6.1 Software troubleshooting chart

ttach the device plate with the serial
Figure 19 Software troubleshooting
NOTE:
software card.
66
Document no. M1144951-004
Appendix A, Functional check form, Datex-Ohmeda S/5 CAM, CCCM

APPENDIX A: Functional check form, Datex-Ohmeda S/5 CAM, CCCM

Customer
Service
Service engineer
Monitor Installation
Frame: F-CM________ System options: E/M-
Software: N- E/M-
K- N- E/M-
K- N- E/M-
Measuring equipment / test gases used:
Equipment / tool / gas: Manufacturer: Model/Type/Part
Number:
Date
Serial Number / ID:
Calibration Date:
OK = Test OK N.A. = Test not applicable Fail = Test failed
Visual Inspection OK N.A. Fail
1. Check all units visually
Functional Inspection OK N.A. Fail
5.3.1. General
5.3.2. Display
5.3.3. Keyboard(s) S/N
5. Functional check S/N
5.3.5. Display unit
A-1(4)
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Functional Inspection OK N.A. Fail
5.3.6. Compact Airway Module, E-CXXXXX/ M-CXXXXX S/N
. Compact Airway Modules with the Patient Spirometry option
. For all Compact Airway Modules
5.3.7. Single width Airway Module, E-miniC/ M-miniC S/N
Notes
5.3.8. Tonometry Module, E-TONO/ M-TONO S/N
5.3.9. Multiparameter Hemodynamic Modules S/N
. ECG and RESP measurements
. Temperature measurement
. Invasive blood pressure measurement
. SpO2 measurement
. Non invasive blood pressure measurement
Notes
5.3.10. Pressure/Pressure Temp Modules, E-P, E-PT, M-P, M-PT
. Invasive blood pressure measurement
. Temperature measurement
5.3.11. Dual pressure Module, E-PP/ M-PP S/N
Notes
5.3.12. Cardiac Output Modules, E-COP/ M-COP, E-COPSv/ M-COPSv
. Invasive blood pressure measurement
. Cardiac Output measurement
S/N
S/N
Notes
A-2(4)
Document no. M1144951-004
Appendix A, Functional check form, Datex-Ohmeda S/5 CAM, CCCM
Functional Inspection OK N.A. Fail
5.3.13. NIBP module, M-NIBP S/N
5.3.14. Nellcor Compatible Saturation module, E-NSATX/ E-NSAT/ M-NSAT
5.3.15. Datex-Ohmeda Oxygen Saturation module, M-OSAT
5.3.16. BIS Module, E-BIS/ M-BIS S/N
5.3.17. Entropy Module, E-ENTROPY/ M-ENTROPY S/N
5.3.18. Data Card and Menu Card function S/N
5.3.19. Recorder S/N
5.3.20. Network connection S/N
5.3.21. Wireless Network Option S/N
5.3.22. Interface Module, E-INT/ M-INT S/N
5.3.23. Interface module for PSM, E-INTPSM S/N
5.3.24. Device Interfacing Solution, N-DISxxx S/N
S/N
S/N
Notes
5.3.25. General
Perform final cleaning
Fill in all necessary documents
Notes
Signature
A-3(4)
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For your notes:
A-4(4)
Document no. M1144951-004
Appendix B, ElectroMagnetic Compatibility

APPENDIX B: ElectroMagnetic Compatibility

Table 1 Guidance and manufacturer’s declaration – electromagnetic emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
(1S/5™ CAM/ CCCM equipped with N-CMW option
Group 1 The S/5™ CAM/ CCCM uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Group 2
Class B The S/5™ CAM/ CCCM is suitable for use in all establishments,
Class A
Complies
(1
The S/5™ CAM/ CCCM must emit electromagnetic energy in order to perform its intended function. Nearby electronic equipment may be affected.
including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
B-1(6)
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Table 2 Guidance and manufacturer’s declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment -
guidance
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transients/bursts IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply lines IEC 61000-4-11
±6 kV contact
±8 kV air
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode ±2 kV common mode
<5 % U
T
(>95 % dip in UT) for 0.5 cycle
40 % U
T
(60 % dip in UT) for 5 cycles
70 % U
T
(30 % dip in UT) for 25 cycles
±6 kV contact
±8 kV air
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode ±2 kV common mode
<5 % U
T
(>95 % dip in UT) for 0.5 cycle
40 % U
T
(60 % dip in UT) for 5 cycles
70 % U
T
(30 % dip in UT) for 25 cycles
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If user of the S/5™ CAM/ CCCM requires continued operation during power mains interruptions, it is recommended that the S/5™ CAM/ CCCM be powered from an uninterruptible power supply or a battery.
<5 % U (>95 % dip in UT) for 5 sec
Power frequency
3 A/m 3 A/m Power frequency magnetic field should (50/60 Hz) magnetic field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
B-2(6)
Document no. M1144951-004
T
<5 % U
T
(>95 % dip in UT) for 5 sec
be at levels characteristic of a typical location in a typical commercial or hospital environment.
Appendix B, ElectroMagnetic Compatibility
d 1.2 P=
d 3.5 P=
d 1.2 P=
80 MHz to 800 MHz
d 2.3 P=
800 MHz to 2.5 GHz
d 3.5 P=
80 MHz to 800 MHz
d 7 P=
800 MHz to 2.5 GHz
Table 3 Guidance and manufacturer’s declaration – electromagnetic immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The S/5™ CAM/ CCCM is intended for use in the electromagnetic environment specified below. The customer or the user of the S/5™ CAM/ CCCM should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms 150 kHz to 80 MHz
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to 2.5 GHz
3 V/m 80 MHz to 2.5 GHz
3 Vrms
1 Vrms
3 V/m
1 V/m
(1
(2
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the S/5™ CAM/ CCCM, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
b
a
B-3(6)
Document no. M1144951-004
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the S/5™ CAM/ CCCM is used exceeds the applicable RF compliance level above, the S/5™ CAM/ CCCM should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the S/5™ CAM/ CCCM.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
(1
For impedance RESP measurement . For invasive pressure measurement in M-series
modules except M-PRE(S)TN. The invasive pressure measurement compliance level is dependent on the transducer used, and only the lower level is guaranteed for all transducers.
(2
For BIS measurement in E-BIS/M-BIS, RESP measurement in M-(P)RE(S)TN and invasive
pressure measurement in M-COPSv/E-COPSv and M-COP/E-COP modules. The invasive pressure measurement compliance level is dependent on the transducer used, and only the lower level is guaranteed for all transducers.
B-4(6)
Document no. M1144951-004
Appendix B, ElectroMagnetic Compatibility
d 1.2 P=
d 3.5 P=
d 1.2 P=
d 3.5 P=
d 2.3 P=
d 7 P=
Table 4 Recommended separation distances between portable and mobile RF
communications equipment and the S/5™ CAM/ CCCM
Recommended separation distances between portable and mobile RF communications equipment and the S/5™ CAM/ CCCM.
The S/5™ CAM/ CCCM is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the S/5™ CAM/ CCCM can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the S/5™ CAM/ CCCM as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
0.01 0.12
0.1 0.38
11.2
10 3.8
100 12
150 kHz to 80 MHz
Separation distance according to frequency of transmitter
m
80 MHz to 800 MHz
(1
(2
80 MHz to 800 MHz
0.12
0.35
(1
0.35
(2
0.70
0.38
1.1
(1
1.1
(2
1.2
(1
3.5
3.5 (2
7.0
3.8
11
(1
11
(2
12
35
(1
35
(2
0.23
0.73
2.2
2.3
7.3
22
23
70
(2
(2
(2
(2
(2
(2
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
(1
For impedance RESP measurement. For invasive pressure measurement in M-series
modules except M-PRE(S)TN. The invasive pressure measurement compliance level is dependent on the transducer used, and only the lower level is guaranteed for all transducers.
(2
For BIS measurement in E-BIS/M-BIS, RESP measurement in M-(P)RE(S)TN and invasive pressure measurement in M-COPSv/E-COPSv and M-COP/E-COP modules. The invasive pressure measurement compliance level is dependent on the transducer used, and only the lower level is guaranteed for all transducers.
Document no. M1144951-004
B-5(6)
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For your notes:
B-6(6)
Document no. M1144951-004
Appendix C, Channel Mask Selections

APPENDIX C: Channel Mask Selections

Country Selection
Argentina 07FF 2.400 - 2.4835 01 - 11
Australia 1FFF 2.400 - 2.4835 01 - 13
Austria 1FFF 2.400 - 2.4835 01 - 13
Belgium 1FFF 2.400 - 2.4835 01 - 13
Canada 1FFF 2.400 - 2.4835 01 - 13
Chile 07FF 2.400 - 2.4835 01 - 11
Colombia 07FF 2.400 - 2.4835 01 - 11
Czech Republic 1FFF 2.400 - 2.4835 01 - 13
Denmark 1FFF 2.400 - 2.4835 01 - 13
Ecuador 07FF 2.400 - 2.4835 01 - 11
Finland 1FFF 2.400 - 2.4835 01 - 13
Allowed Band DS Channels
France 1C00 2.4465 - 2.4835 11 - 13
Germany 1FFF 2.400 - 2.4835 01 - 13
Greece 1FFF 2.400 - 2.4835 01 - 13
Hong Kong 1FFF 2.400 - 2.4835 01 - 13
Hungary 1FFF 2.400 - 2.4835 01 - 13
Iceland 1FFF 2.400 - 2.4835 01 - 13
Ireland 1FFF 2.400 - 2.4835 01 - 13
Israel 00F0 2.418 - 2.457 05 - 08
Italy 1FFF 2.400 - 2.4835 01 - 13
Japan 1FFF 2.400 - 2.4835 01 - 13
Liechtenstein 1FFF 2.400 - 2.4835 01 - 13
Luxembourg 1FFF 2.400 - 2.4835 01 - 13
Netherlands 1FFF 2.400 - 2.4835 01 - 13
Norway 1FFF 2.400 - 2.4835 01 - 13
Philippines 1FFF 2.400 - 2.4835 01 - 13
Poland 1FFF 2.400 - 2.4835 01 - 13
Portugal 1FFF 2.400 - 2.4835 01 - 13
Singapore 1FFF 2.400 - 2.4835 01 - 13
South Africa 1FFF 2.400 - 2.4835 01 - 13
South Korea 1FFF 2.400 - 2.4835 01 - 13
Spain 1FFF 2.400 - 2.4835 01 - 13
Document no. M1144951-004
C-1(2)
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
Country Selection
Sweden 1FFF 2.400 - 2.4835 01 - 13
Switzerland 1FFF 2.400 - 2.4835 01 - 13
Taiwan 1FFF 2.400 - 2.4835 01 - 13
Thailand 1FFF 2.400 - 2.4835 01 - 13
Turkey 1FFF 2.400 - 2.4835 01 - 13
United Kingdom 1FFF 2.400 - 2.4835 01 - 13
United States 07FF 2.400 - 2.4835 01 - 11
Venezuela 07FF 2.400 - 2.4835 01 - 11
Vietnam 1FFF 2.400 - 2.4835 01 - 13
Allowed Band DS Channels
C-2(2)
Document no. M1144951-004
Datex-Ohmeda
S/5
Compact Anesthesia Monitor
TM
TM
S/5
Compact Critical Care Monitor

Planned Maintenance Instructions

Conformity according to the Council Directive 93/42/EEC concerning Medical Devices
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the USA, check local laws for any restriction that may apply.
All specifications subject to change without notice.
Document number M1144953-001
April 9, 2008
GE Healthcare Finland Oy Helsinki, Finland P.O. Box 900 FI-00031 GE, FINLAND Tel: +358 10 39411 Fax: +358 9 1463310
www.gehealthcare.com
Copyright © 2008 General Electric Company. All rights reserved.
Madison, WI 53707-7550, USA
Datex-Ohmeda Inc.
Tel: +1 608 221 1551
Fax: +1 608 222 9147
P.O. Box 7550
Table of contents
Table of contents
Table of contents i
1 Planned maintenance instructions 1
1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Recommended tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.1 Hemodynamic patient simulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Recommended parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Planned maintenance parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4.1 PM parts for Compact Airway Modules, E-CXX/ M-CXX without CO2 absorber
- Anesthesia 5
1.4.2 PM parts for Compact Airway Modules, E-CXX/ M-CXX without CO2 absorber
- Critical Care 6
1.4.3 PM parts for Compact Airway Modules, E-CXX/ M-CXX with CO2 absorber
- Anesthesia. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.4.4 PM parts for Compact Airway Modules, E-CXX/ M-CXX with CO2 absorber
- Critical Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4.5 PM parts for Airway Module, E-miniC/ M-miniC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Planned maintenance check list 9
2.1 Visual inspection/preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1.2 Compact Airway Module, E-CXX/ M-CXX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1.3 Single-width Airway Module, E-miniC/ M-miniC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.1.4 Tonometry Module, E-TONO/ M-TONO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.1.5 Recorder Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1.6 Modules with NIBP measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2 Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2.1 General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.2.3 Keyboard(s). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.2.4 Display Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.2.5 Frame Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.2.6 Compact Airway Module, E-CXXXXX/ M-CXXXXX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.2.7 Single-width Airway Module, E-miniC/ M-miniC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2.8 Tonometry Module, E-TONO/ M-TONO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2.9 Multiparameter Hemodynamic Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.2.10EEG Module, E-EEG/ M-EEG and EEG Headbox, N-EEG. . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.2.11BIS Module, E-BIS/ M-BIS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.2.12Entropy Module, E-ENTROPY/ M-ENTROPY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.2.13Pressure/Pressure Temp Modules, E-P, E-PT, M-P, M-PT . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.2.14Dual pressure Module, E-PP, M-PP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.2.15Cardiac Output Modules, E-COP, E-COPSv, M-COP, M-COPSv . . . . . . . . . . . . . . . . . . . . . . 20
2.2.16NIBP module, M-NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2.2.17NeuroMuscular Module, E-NMT, M-NMT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Document no. M1144953-001
i
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
2.2.18Nellcor Compatible Saturation module, E-NSATX, E-NSAT, M-NSAT. . . . . . . . . . . . . . . . 22
2.2.19Datex-Ohmeda Oxygen Saturation module, M-OSAT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.2.20Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.2.21Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.2.22Network Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.23Wireless Network Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.24Interface Module, E-INT, M-INT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.25Device Interfacing Solution, N-DISxxx . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.26General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Appendix A: Planned maintenance check form, S/5 Compact Monitors A-1
ii
Document no. M1144953-001

1 Planned maintenance instructions

1.1 Introduction
These instructions include procedures for planned maintenance (PM) for the Datex-Ohmeda S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor. The Planned maintenance should be performed once a year.
These instructions include “Planned maintenance check form, S/5 Compact Monitors” to be filled in when performing the corresponding procedures.
The symbol
the check form. The procedures should be performed in ascending order, bypassing those that are not
applicable for a particular monitor. If you need further information on how to perform a certain Planned maintenance procedure,
please refer to the corresponding slot in the Technical Reference Manual. All menu selections related to the Datex-Ohmeda monitors are written in the following
typeface: e.g. Parameters - Gas Unit As you enter the service menus, you need following passwords:
Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
In case you evaluate the measurement accuracy with a patient simulator, add the simulator’s accuracy specification to the one for the monitor.
" in the instructions means that the procedure performed should be signed in
WARNING Handle the water trap and its contents as you would any body fluid.
Infectous hazard may be present.
CAUTION Failure on the part of all responsible individuals, hospitals or institutions,
employing the use of this device, to implement the recommended maintenance schedule may cause equipment failure. The manufacturer does not, in any manner, assume the responsibility for performing the recommended maintenance schedule, unless an equipment maintenance agreement exists. The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device.
CAUTION Only trained personnel with appropriate equipment should perform the tests
and repairs outlined in this section. Unauthorized service may void warranty of the unit.
CAUTION Wear a static control wrist strap when handling PC boards. Electrostatic
discharge may damage components on the board.
Document no. M1144953-001
1
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
1.2 Recommended tools
NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the specified calibrations and adjustments to ensure accuracy.
For product(s) Tool Order No.
All Airway modules Flowmeter
Compact Airway Module, E-CAiO(VX)/ M-CAiO(VX)
Compact Airway Module, E-COVX/ M-COVX Calibration gas and regulator 755587/755534*
Compact Airway Module, E-CO(V), M-C, M-CO(V)
E-miniC/ M-miniC Calibration gas and regulator 755580/755534*
Compact Airway Module, E-COVX, E-CAiOVX, M-COVX, M-CAiOVX
All Airway modules w/ (V) Spirometry tube, 2 m 890031
All Airway modules w/ (V) D-lite 733950
All Airway modules Sampling line, 6 m/19.7 ft 73306
Hemodynamic modules
Hemodynamic modules w/ (P), M-NIBP E-PSM(P)
Hemodynamic modules w/ (P)
Calibration gas and regulator 755583/755534*
Calibration gas and regulator 755581/755534*
Sampling line, 2 m 73318
NIBP cuff and adult NIBP cuff hose with cuff ID
Pressure manometer
InvBP transducer 70077-001
Multi-Link ECG accessories, IEC
2753E (cuff) 2021285-001 (hose)
E-PSM(P)/ E-PRESTN/ M-PRESTN - Multi-link 3-leadwire set 412682-003
- Multi-link 5-leadwire set 412681-003
- Multi-link 5-leadwire set, C2-C6 416467-004
E-PRESTN/ M-PRESTN w/ (E12) - Multi-link 12-lead ECG trunk
cable
or Multi-Link ECG accessories, AHA
E-PSM(P) / E-PRESTN/ M-PRESTN - Multi-link 3-leadwire set 412682-001
- Multi-link 5-leadwire set 4162681-001
- Multi-link 5-leadwire set, V2-V6 416467-003
E-PRESTN/ M-PRESTN w/ (E12) - Multi-link 12-lead ECG trunk
cable
M-NESTPR 3-lead ECG trunk cable 545302
M-NESTPR 5-lead ECG cable 545303
M-NE12STPR 10-leadwire ECG cable 545324
2
Document no. M1144953-001
416035-002
416035-001
For product(s) Tool Order No.
E-PSM(P) / E-PRESTN SpO
M-PRESTN SpO
finger probe OXY-F-UN
2
Interconnect Cable OXY-ES3
SpO
2
finger probe OXY-F4-N or SAS-F4
2
E-NSATX 40xxxxx-001 Nellcor OxiMax Spo2
interconnect cable
E-NSAT/ M-NSAT Nellcor SpO
finger probe with
2
DOC-10 cable
M-OSAT OxyTip® + Integrated Finger Sensor OXY-F4-N
Hemodynamic modules w/ (T) Temperature test set 884515
Hemodynamic M-Modules w/ (NIBP) Adult NIBP cuff 572435
Adult NIBP cuff hose 877235
E-COP, E-COPSv, M-COP, M-COPSv C.O. Catheter cable 16590
E-COPSv/ M-COPSv SvO
simulator 890121
2
Tonometry Module, E-TONO/ M-TONO Calibration gas and regulator 755580/755534
Sampling line 733251
Luer plug
Pressure manometer
Tonometrics
TM
catheter
TONO-14F or another suitable TONO-_F catheter
EEG Module, E-EEG/ M-EEG Earphones
EEG simulator 90502
BIS Module, E-BIS/ M-BIS BIS simulator or
BIS Sensor simulator
900509 900508
E-ENTROPY/ M-ENTROPY Simulator for M-ENTROPY N-ES
Entropy sensor cable 8002964
E-NMT/ M-NMT NMT simulator 871251
NMT ElectroSensor 888416
NMT sensor cable 888415
MemCard – Menu 893860 (English)
MemCard – Data 887045 (English)
All M-Modules Screwdrivers
All E-Modules Torx screwdrivers; T8, T10
NOTE: * Ensure that the calibration gas and regulator are functioning properly before calibration. Perform annual maintenance on the regulator as required. For more information see section “Adjustments and calibrations” in Compact Airway modules slot in E-Modules Technical Reference Manual.
Document no. M1144953-001
3
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
ECG
NIBP
ECG
SpO
T1
T2
P1
P2
2
Auto
On/Off
Start
Cancel
Zero
P1
Zero
P2
T2
P1
P2
T1
NIBP
ECG+ Resp
SpO
2
1.2.1 Hemodynamic patient simulators
The following tables present the patient simulators’ compatibility with each hemodynamic module, and the accessories needed:
Table 1 Patient simulators’ compatibility with each hemodynamic module
Patient simulator
Module Parameter
E-PSM(P),E-PRESTN ECG Cable included Multilink ECG acc. Not compatible
T 402015-004 402015-004 and
InvBP Cable included M1010858
M1010831 MedSim 874027
M1010832
M-PRESTN ECG Ok Multilink ECG acc. Not compatible
TOk M1010832 M1010832
InvBP Not compatible Ok Ok
M-NESTPR ECG Not compatible 300 series acc. Ok
12 lead ECG Not compatible 300 series acc. Not compatible
TOk M1010832 M1010832
InvBP Not compatible Ok Ok
Table 2 Adapter cables for hemodynamic patient simulators
Patient simulator
Hemodynamic patient simulator Dual temperature adapter cable 402015-004
Hemodynamic patient simulator Dual Inv.BP adapter cable 2005772-001
Medsim Temperature adapter cable M1010832
4
Document no. M1144953-001
Patient simulator
Medsim Inv.BP adapter cable M1010858
Lionheart & MPS450 Temperature adapter cable M1010846
Lionheart & MPS450 Inv.BP adapter cable M1010862
1.3 Recommended parts
For product(s) Part Order No.
CM(C) Frame Unit Fan filter 896085
CM(C) Display Unit Fan filter 874594
Central Processing Board Battery for SRAM/Timekeeper 197230
Recorder paper 74205
Tonometry Module, E-TONO/ M-TONO 733382
1.4 Planned maintenance parts
1.4.1 PM parts for Compact Airway Modules, E-CXX/ M-CXX without CO2 absorber
- Anesthesia
Part Order No. For product(s)
Special tube (Nafion) 733382 All Airway modules
Ref. gas sticker 893110 Compact Airway Modules
Filter (2 pcs) 886136 Compact Airway Modules
Filter assembly 896025 Compact Airway Modules
Fan filter 886236 Compact Airway Modules
D-fend O-ring (2 pcs) 653125 All Airway modules
D-fend (black) 876445 All Airway modules
Sampling line 2.0 m 73318 Compact Airway Modules
PM sticker 893108 All Airway modules
NOTE: Corresponding PM kit is available with Order No. 8001758.
Document no. M1144953-001
5
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
1.4.2 PM parts for Compact Airway Modules, E-CXX/ M-CXX without CO2 absorber
- Critical Care
Part Order No. For product(s)
Special tube (Nafion) 733382 All Airway modules
Ref. gas sticker 893110 Compact Airway Modules
Filter (2 pcs) 886136 Compact Airway Modules
Filter assembly 896025 Compact Airway Modules
Fan filter 886236 Compact Airway Modules
D-fend O-ring (2 pcs) 653125 All Airway modules
D-fend+ (green) 893200 All Airway modules
Sampling line 2.0 m 733162 Compact Airway Module, M-COVX
PM sticker 893108 All Airway modules
NOTE: Corresponding PM kit is available with Order No. 8001759.
1.4.3 PM parts for Compact Airway Modules, E-CXX/ M-CXX with CO2 absorber
- Anesthesia
Part Order No. For product(s)
Special tube (Nafion 2 pcs) 733382 All Airway modules
Ref. gas sticker 893110 Compact Airway Modules
Filter (2 pcs) 886136 Compact Airway Modules
Filter assembly 896025 Compact Airway Modules
Fan filter 886236 Compact Airway Modules
D-fend O-ring (2 pcs) 653125 All Airway modules
D-fend (black) 876445 All Airway modules
Sampling line 2.0 m 73318 Compact Airway Modules
PM sticker 893108 All Airway modules
NOTE: Corresponding PM kit is available with Order No. 8001760.
6
Document no. M1144953-001
1.4.4 PM parts for Compact Airway Modules, E-CXX/ M-CXX with CO2 absorber
- Critical Care
Part Order No. For product(s)
Special tube (Nafion 2 pcs) 733382 All Airway modules
Ref. gas sticker 893110 Compact Airway Modules
Filter (2 pcs) 886136 Compact Airway Modules
Filter assembly 896025 Compact Airway Modules
Fan filter 886236 Compact Airway Modules
D-fend O-ring (2 pcs) 653125 All Airway modules
D-fend+ (green) 893200 All Airway modules
Sampling line 2.0 m 733162 Compact Airway Module, M-COVX
PM sticker 893108 All Airway modules
NOTE: Corresponding PM kit is available with Order No. 8001761.
1.4.5 PM parts for Airway Module, E-miniC/ M-miniC
Part Order No. For product(s)
Special tube (Nafion 2 pcs) 733382 All Airway modules
Mini D-fend O-ring (2 pcs) 656565 E-miniC/ M-miniC
Mini D-fend 8002174 (pkg of 10 pcs) E-miniC/ M-miniC
Zero valve air filter M1011471 E-miniC, every 3 years
Zero valve air filter 571421 M-miniC
CO
Sampling line 3.0 m 733163 E-miniC/ M-miniC
2
PM sticker 893108 All Airway modules
For details on recommended accessories see the “Supplies and Accessories” catalog.
Document no. M1144953-001
7
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
For your notes:
8
Document no. M1144953-001
Loading...