Datex-Ohmeda
S/5™ Anesthesia Monitor and Critical Care Monitor
Technical Reference Manual
|
|
|
|
|
|
All specifications are subject to change without notice. |
||||||||||||
|
|
|
|
CAUTION: U.S. Federal law restricts this device |
to |
sale by |
or on |
|
the |
order |
of a |
licensed medical |
practitioner. |
|||||
|
|
|
|
Outside |
the |
USA, |
check |
local |
laws |
for |
any |
restriction that |
may apply. |
|||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
M1021563 |
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
June, 2005 |
||
Datex-Ohmeda |
Inc. |
|
|
|
|
|
|
|
|
GE |
Healthcare |
Finland |
Oy |
|||||
Р.О. Вох 7550 |
|
|
|
|
|
|
|
|
|
|
|
Helsinki, |
Finland |
|||||
Madison, WI 53707-7550, USA |
|
|
|
|
|
|
|
|
|
|
P.0. |
Box |
900 |
|||||
Tel: |
+1 |
608 |
221 |
1551 |
|
|
|
|
|
|
|
|
|
FI-00031 GE, FINLAND |
||||
Fax: |
+1 |
608 |
222 |
9147 |
|
|
|
|
|
|
|
|
|
Tel: |
+358 |
10 |
39411 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Fax: |
+358 |
9 1463310 |
||
|
|
|
|
|
|
|
www |
datex-ohmeda.com |
www.gehealthcare.com |
|||||||||
|
|
|
|
|
|
|
|
O |
2005 Copyright |
General |
Electric Company |
Intended purpose (Indications for use)
The Datex-Ohmeda S/5 Anesthesia Monitor with L-ANEO5 or L-ANEO5A software is intended for multiparameter patient monitoring with optional patient care documentation.
The S/5 Anesthesia Monitor with L-ANEO5 and L-ANEOSA software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/ regional perfusion, Bispectral index (BIS), Entropy (State Entropy and Response Entropy) and neurophysiological status of all hospital patients.
The |
S/5 Anesthesia Monitor with L-ANEO5 and L-ANEOSA |
software |
when |
using |
BIS |
is for |
monitoring the |
state |
of |
||||||||||
the |
brain by data acquisition and processing of electroencephalograph signals and |
may |
be |
used |
as |
an |
aid in |
|
|||||||||||
monitoring |
the |
effects |
of certain |
anesthetic |
agents. |
|
|
|
|
|
|
|
|
|
|
|
|
||
The |
S/5 Anesthesia Monitor with L-ANEO5 and L-ANEOSA |
software |
is |
also indicated for |
documenting |
patient |
|
||||||||||||
care |
related information. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||
The |
S/5 Anesthesia Monitor with L-ANEO5 and L-ANEODA |
software |
is |
indicated |
for |
use |
by |
qualified |
medical |
|
|||||||||
personnel |
only. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
The |
Datex-Ohmeda S/5 Critical |
Care Monitor with L-ICUO5 |
or L-ICUO5A |
software |
is |
intended |
for multiparameter |
||||||||||||
patient |
monitoring. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
The |
S/5 Critical Care Monitor with L-ICUO5 and L-ICUO5A software is indicated for |
monitoring of |
hemodynamic |
||||||||||||||||
(including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, |
|
||||||||||||||||||
Bispectral |
index |
(BIS) |
and neurophysiological status of all |
hospital |
patients. |
|
|
|
|
|
|
|
|
||||||
The |
S/5 |
Critical |
Care |
Monitor with L-ICUO5 and L-ICUO5A |
software |
when |
using BIS |
is for |
monitoring |
the |
state |
of |
|||||||
the |
brain by data acquisition and processing of electroencephalograph signals and |
may |
be |
used |
as |
an |
aid in |
|
|||||||||||
monitoring |
the |
effects |
of certain |
anesthetic |
agents. |
|
|
|
|
|
|
|
|
|
|
|
|
||
The |
S/5 Critical Care Monitor with L-ICUO5 and L-ICUO5A software is indicated for |
use by qualified |
medical |
|
|||||||||||||||
personnel |
only. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Classifications
In accordance with IEC 60601-1
Class | and internally powered equipment - the type of protection against electric shock.
Type BF or CF equipment. The degree of protection against electric shock is indicated by a symbol on each parameter module.
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
Continuous operation according to the mode of operation. In accordance with IEC 60529
With F-CU8 Central Unit: IPXO - the degree of protection against harmful ingress of water. With F-CU5(P) Central Unit: IPX1 - the degree of protection against harmful ingress of water. In accordance with EU Medical Device Directive
The Datex-Ohmeda S/5 Anesthesia Monitor is classified as IIb. The Datex-Ohmeda S/5 Critical Care Monitor is classified as IIb. In accordance with CISPR 11:
Group 1, Class B (F-CU5(P), F-CPU and N-AC: Class A)
s |
Group |
1 contains all ISM (industrial, scientific |
and |
medical) equipment |
in which there is |
intentionally |
|||||
|
generated and/or used conductively coupled radio-frequency energy |
which is necessary |
for the |
internal |
|||||||
|
functioning of the equipment |
itself. |
|
|
|
|
|
|
|
||
e |
Class B |
equipmentis |
suitable |
for use |
in domestic |
establishments and |
in |
establishments |
directly |
connected |
|
|
to a low |
voltage power |
supply |
network |
which |
supplies buildings used |
for |
domestic purposes. |
|
eClass A equipment is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.
Responsibility of the manufacturer
GE Healthcare Finland Oy (GE) is responsible for the safety, reliability and performance of the product only if: modifications, service and repairs are carried out by personnel authorized by GE.
the |
electrical installation of the room complies with appropriate reguirements. |
|
the |
equipment is |
used in accordance with the User's Guide and serviced and maintained in accordance |
with |
the Technical |
Reference Manual. |
Master table of contents
Datex-Ohmeda S/5™ Anesthesia and Critical Care Monitors
Technical Reference Manual, Order code: M1021563
Part I, General Service Guide
Document No. | Updated |
Description |
|
|
M1027807 |
Introduction, System |
description, |
Installation, |
|
Interfacing, Functional check, General troubleshooting |
||
M1027808 |
Planned |
Maintenance |
Instructions |
Part II, Product Service Guide
Document No. |
Updated |
Description |
|
|
|
M1027809 |
|
AM, CCM |
Service |
Menu |
|
M1027810 |
|
8-Module |
Frame, |
F-CU8 |
|
M1027811 |
|
5-Module Frame, F-CU5(P) |
|||
M1027812 |
|
CPU Board, B-CPU5 |
|||
M1027813 |
|
UPINET Board, B-UPI4NET |
|||
M1023412 |
|
Displays and Display Controller Boards |
|||
M1027814 |
|
Command |
Boards and |
Bars |
|
M1027815 |
|
Interface Board, B-INT |
|||
M1027816 |
|
Extension Frame, F-EXT, Extension |
Module, |
M-EXT |
|
M10 39372 |
|
AM, |
CCM Spare |
Parts |
F
NR
AK oon©
ona
pa©
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
About this manual
Notes to the reader
This Technical Reference Manual is intended for service personnel and engineers who will service and maintain Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
e |
The order |
code for |
the |
entire printed |
manual |
is |
M1021563. |
The manual |
includes Technical |
||||||
|
Reference |
Manual |
Slots |
and |
every |
slot |
has an |
individual document number. |
M102 7807 is the |
||||||
|
document |
number |
of |
this |
first |
slot. |
|
|
|
|
|
|
|
|
|
e |
Parti gives |
the |
reader |
an |
overview |
of |
the |
S/5 |
Anesthesia Monitor and S/5 Critical Care Monitor. |
||||||
|
It contains |
the |
information needed |
to |
install, |
interface and |
troubleshoot |
the |
monitors. |
Instructions for functional check and planned maintenance are also included. Read the manual through and make sure that you understand the procedures described before the installation of the monitor. To avoid risks concerning safety or health, strictly observe the warning indications. İf you need any assistance concerning the installation, please do not hesitate to contact your authorized distributor.
e |
Partll contains detailed descriptions |
of each component |
of |
the |
S/5 AM, CCM, such as frame |
||
|
unit and displays. Service check for |
each |
product is |
included |
in |
these slots. Service Menu slot |
|
|
contains all the service menus and Spare |
Parts slot |
all the |
spare |
parts information for this |
||
|
monitor. |
|
|
|
|
|
|
For |
information |
of |
parameter modules, |
Remote |
Controller |
and |
Device |
Interfacing Solution refer to |
the |
||||||
S/5 E-Modules Technical Reference Manual. Service check for each |
of these |
products |
is |
included |
in |
||||||||||
these slots. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
The |
manufacturer |
reserves the |
right to |
change |
product specifications |
without |
prior notice. |
Although |
|||||||
the |
information |
in |
this manual |
is believed |
to |
be |
accurate |
and |
reliable, |
the manufacturer |
assumes |
no |
|||
responsibility for |
its use. |
|
|
|
|
|
|
|
|
|
|
|
|
||
GE Healthcare Finland Oy (GE) |
assumes |
no |
responsibility |
for |
the use |
or reliability of its software in |
|
||||||||
eguipment that |
is |
not furnished |
by GE. |
|
|
|
|
|
|
|
|
|
|
|
Related documentation
S/5 Modules
S/5 E-Modules Technical Reference Manual
S/5 M-Modules Technical Reference Manual
Other devices closely related to the S/5 Anesthesia Monitor: iCentral, User's Reference Manual
3/5 Arrhythmia Workstation User’s Reference Manual
S/5 |
Anesthesia Monitor |
|
||
Instructions for daily use, clinical |
aspects and basic methods of measurement: |
|||
S/5 |
Anesthesia |
Monitor, |
User's |
Guide |
S/5 |
Anesthesia |
Monitor, |
User's |
Reference Manual |
S/5 |
Critical Care Monitor |
|
|||
Instructions for daily use, clinical |
aspects and basic methods of measurement: |
||||
S/5 |
Critical |
Care |
Monitor, |
User's |
Guide |
S/5 |
Critical |
Care |
Monitor, |
User's |
Reference Manual |
1
Document no. M1027807
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
Conventions used
To help you find and interpret information easily, the manual uses consistent text formats:
BS |
|
Sign |
the |
check form |
after |
performing the procedure. |
|
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
Hard |
Keys |
Names |
of |
the |
hard keys on |
the Remote |
Controller, |
Command Bar and |
modules are |
written in the |
|
|
|
following |
way: |
Others. |
|
|
|
|
|
||
Hard |
Keys |
Menu |
ItemsSoftware |
terms |
that identify |
window |
parts or menu items |
are written in |
bold italic: ECG |
||
|
|
Setup. |
|
|
|
|
|
|
|
|
|
Menu access is described from top to bottom. For example, the selection of the Monitor Setup hard
key, the Screen 1 Setup menu item and the Waveform Fields menu item would be shown as
Monitor Setup - Screen 1 Setup - Waveform Fields.
‘Messages’ |
Messages |
(alarm messages, informative messages) displayed on the screen are |
written |
inside single |
||||||||||||||||||
|
quotes: |
‘Please |
wait’. |
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||
“Sections” |
When |
referring |
to |
different |
sections |
in this |
manual |
or to |
other manuals, |
manual |
names |
and section |
||||||||||
|
names are |
enclosed |
in |
|
double |
quotes: |
|
|
|
|
|
|
|
|||||||||
|
See section |
"Cleaning |
|
and |
care." |
|
|
|
|
|
|
|
|
|
||||||||
|
Please refer |
to |
"iCentral |
User's |
Reference |
Manual: |
Alarms. " |
|
|
|
||||||||||||
Hypertext links |
Hypertext |
links |
on |
versions |
are |
written |
in blue |
color. |
|
|
|
|||||||||||
WARNING |
Warnings |
are |
written in |
|
the |
following |
way: |
|
|
|
|
|
|
|||||||||
WARNING |
This |
is |
a |
WARNING. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|||||
CAUTION |
Cautions |
are |
written |
in |
the |
following |
way: |
|
|
|
|
|
|
|||||||||
CAUTION |
This |
is |
a |
CAUTION. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|||||
NOTE |
Notes |
are |
written |
in |
following way: |
|
|
|
|
|
|
|
|
|
||||||||
|
NOTE: |
This |
is |
a |
NOTE. |
|
|
|
|
|
|
|
|
|
|
|
|
|
||||
|
In |
this |
manual, |
the |
word |
“select” |
means |
choosing |
and |
confirming. |
|
|
|
|||||||||
Illustrations |
and |
names |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
|
All |
illustrations |
in |
this |
manual |
are |
only |
examples, |
and may not necessarily reflect your system |
|||||||||||||
|
settings |
or |
data |
displayed |
in your |
system. |
If a particular |
selection is not |
available in your system, the |
|||||||||||||
|
selection |
is shown |
grayed. |
|
|
|
|
|
|
|
|
|
|
|
2
Document no. M1027807
Introduction
1 Introduction
The Datex-Ohmeda S/5 Anesthesia Monitor is a modular multiparameter patient monitor used during anesthesia in operating rooms.
The Datex-Ohmeda S/5 Critical Care Monitor provides full patient profile throughout the care period.
The modular design makes the system flexible and |
easy to upgrade. In |
addition |
to parameter |
||
changes, the modularity |
includes |
an easy upgrade |
to anesthesia record |
keeping, |
monitor networking |
and interfacing with other |
external |
devices. |
|
|
|
Re eso
Ara,
РМval|
3
Document no. M1027807
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
1.1Symbols
1.1.1Symbols on transport packaging
5 |
The |
contents |
of the transport package |
are fragile and |
must be handled with care. |
||
| | |
Indicates the correct upright position |
of |
the |
transport |
package. |
||
|
The |
transport package must be kept |
in |
a dry |
environment. |
||
Y |
Indicates the |
temperature limitations |
within |
which the |
transport package should be |
||
|
stored. |
|
|
|
|
|
1.1.2Symbols on equipment
NA |
The battery |
contains |
lead |
acid, |
and |
in |
the event of disposal must be separated from other |
waste according to local |
regulations. |
|
|||||
Pb |
|
|
|
|
|
|
|
GÖ |
This battery |
contains |
lead |
and |
can |
be |
recycled. |
Pb |
|
|
|
|
|
|
|
|
Dangerous |
voltage. |
|
|
|
|
|
|
When using the |
ARK |
Barcode |
Reader, N-SCAN, do not stare into beam. The N-SCAN |
AR |
Barcode Reader |
is a |
Class 2 |
laser product. |
4
Document no. M1027807
Introduction
1.1.3Equipment safety symbols
Attention, |
consult accompanying |
documents. |
|
|
|
|
|
|
|
|
|
||||||||||||||
When |
displayed next |
|
to |
the |
02 |
|
value, |
indicates |
that |
the Fi02 |
low alarm |
limit is |
set |
||||||||||||
below |
21 |
%. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
When |
displayed |
next |
to |
the |
HR |
value, |
indicates |
that |
there |
is |
a |
risk that |
the |
|
|||||||||||
monitor |
counts |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||||
* |
pacemaker |
spikes |
|
because |
the |
pacer is |
set |
on |
R. |
|
|
|
|
|
|
||||||||||
* |
T-waves |
because |
a |
wide |
QRS |
is |
selected. |
|
|
|
|
|
|
|
|
|
|
||||||||
On |
|
the |
15” |
display, |
D-VMC15, |
|
indicates that |
the |
display should be supplied from |
||||||||||||||||
the |
mains |
outlet |
(or |
from |
the |
Central |
Unit, |
F-CU8 |
Rev.03 |
- 09). |
|
|
|
||||||||||||
On |
the |
15” |
display, |
D-VNC15, |
|
indicates that the display must be supplied from |
|||||||||||||||||||
the |
mains |
outlet via an appropriate additional separating |
transformer (or from |
|
|||||||||||||||||||||
the |
Central |
Unit, F-CU8 Rev.03 |
- |
09). |
|
|
|
|
|
|
|
|
|
|
|
||||||||||
On |
the |
15" |
LCD display, D-LCC15, indicates that |
the display must only be used |
|||||||||||||||||||||
together |
with |
the |
original type |
|
of |
D-LCC15 |
power |
adapter. |
The |
display |
should |
be |
supplied from the mains outlet (or from the Central Unit, F-CU8 Rev.03 - 09).
On |
the |
17” |
display, |
D-LCC17 |
indicates that the display |
must be used only |
|
together |
with the original D-LCC17 power |
adapter. |
|
||||
On |
the |
17” |
display, |
D-VHC17 |
rev.00-01, |
indicates that |
the display must only be |
supplied from the mains outlet, not from the Central Unit, F-CU8.
On the 17” display, D-VHC17 rev. 02 or higher, indicates that the display should
be supplied from the mains outlet via an appropriate additional separating |
|
|||||||||||||||||||
transformer (or |
from |
the |
Central Unit, |
F-CU8 |
Rev.03 |
- 09). |
|
|
|
|||||||||||
On |
the |
19” |
display, |
D-LCC19, |
indicates |
that |
the |
display |
must |
only |
be supplied |
|||||||||
from |
the |
mains |
outlet |
via |
an |
appropriate |
additional |
separating |
transformer |
and |
||||||||||
the original D-LCC19 |
|
power |
adapter, |
not |
from the Central |
Unit, |
F-CU8. |
|
||||||||||||
On |
the |
21” |
display, |
D-VSC21, |
indicates |
that |
the |
display |
must |
only |
be supplied |
|||||||||
from |
the |
mains |
outlet |
via |
an |
appropriate |
additional |
separating |
transformer, |
not |
||||||||||
from |
the |
Central |
Unit, |
F-CU8. |
|
|
|
|
|
|
|
|
|
|
|
|||||
On |
the |
Interface |
Module |
E-INT, |
M-INT, |
indicates |
that |
it is |
for connecting external |
On the E-TONO, M-TONO module indicates that the module should only be used with Tonometrics catheters.
On the E-miniC module indicates that airway gases should be calibrated every six months in normal use and every two months in continuous use.
BIS: On |
the Aspect DSC indicates that the converter must not be opened for any |
reason |
or autoclaved. |
On the E-PRESTN, |
E-PRETN, E-RESTN, E-PSM, E-PSMP, E-P, |
E-PP, E-PT, E-COP and |
||||||
E-COPSv |
module |
indicates that |
protection |
against |
cardiac |
defibrillator |
discharge |
|
is due in |
part to the |
accessories |
for pulse |
oximetry |
(SpO2), |
temperature |
(T) and |
|
invasive |
pressure |
(P) |
measurement. |
|
|
|
|
5
Document no. M1027807
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
- © On the E-NMT, M-NMT module indicates the following warnings:
* |
Do not |
place |
the |
NMT |
stimulating electrodes on the patient’s chest. |
* |
Always |
stop |
the |
NMT |
measurement before handling the stimulating electrodes. |
* Never subject a patient with an implanted electronic device to electrical Stimulation without consulting a medical specialist first.
-Qn the rear or bottom panel this symbol indicates the following warnings and Cautions:
* |
Electric |
shock |
hazard. |
Do |
not open the cover or the |
back. |
Refer servicingto |
||||||
qualified |
service |
personnel. |
|
|
|
|
|
|
|||||
* |
For |
continued |
protection |
against |
fire hazard, |
replace |
the |
fuse only |
with one of |
||||
the same |
type |
and |
rating. |
|
|
|
|
|
|
||||
* |
Disconnect |
from |
the power |
supply |
before servicing. |
|
|
|
|||||
* |
Do |
not |
use |
the |
monitor |
without manufacturer |
approved |
mounting |
attached. |
-On the rear panel of the power unit (N-AC) this symbol indicates the following
warnings and cautions:
*Electric shock hazard. Do not open the cover. Refer servicing to qualified service personnel.
*Does not contain field replaceable fuses.
Type BF (IEC 60601-1) protection against electrical shock.
Type BF (IEC 6060 1-1) defibrillator-proof protection against electric shock.
Type CF (IEC 60601-1) protection against electric shock.
Type CF (IEC 60601-1) defibrillator-proof protection against electric shock.
When displayed in the upper left corner of the screen, indicates that the alarms are silenced. When displayed in the menu or digit fields, indicates that the alarm source has been turned off or alarm does not meet the alarm-specific activation criteria.
A
ÂLA
ESD |
warning |
symbol |
for electrostatic |
sensitive devices. Pins of connectors |
identified |
with |
the ESD |
warning |
symbol should |
not be touched. Connections should |
not be |
made to these connectors unless ESD precautionary procedures are used. For details, see “ESD precautionary procedures” in the User's Reference Manual.
(4) |
Symbol |
for non-ionizing electromagnetic radiation. Interference may occur in the |
|
vicinity |
of equipment marked with this symbol. |
Document no. M1027807
1.1.4Other symbols
<
о SN, S/N
B
剛 のeo@@
|
+)
©)
==
LE }{ }
Introduction
Equipotentiality. Monitor can be connected to potential equalization conductor.
Alternating current
Fuse. |
Replace the fuse only with one of the same type and rating. |
Serial |
Number |
Connector for color display
Signal/power output
Signal/power input
Signal/power input/output
Connector for defibrillator synchronization
Connectorforthe 5/5 Device Interfacing Solution, DIS
Power |
input |
Signal |
input |
Power |
input |
Submenu. Selecting an alternative marked with this symbol in a menu opens a new
menu.
The monitor is connected to the iNetwork (LAN).
7
Document no. M1027807
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
品
IPX class:
Data Card (green) and/or Menu Card (white) is inserted.
A blinking heart next to the heart rate or pulse rate value indicates the beats detected.
A lung next to the respiration rate |
value indicates that respiration rate is calculated |
from the impedance respiration |
measurement. |
Gas inlet
Gas outlet
Do not reuse.
Use by. Indicates the last use day.
Do not immerse the sensor in liquids.
Degree of protection against harmful ingress of water as detailed in the IEC 60529:
IPXO
Al
Z
-Ordinary equipment
Date |
of manufacture |
|
|
This |
symbol |
indicates that the waste |
of electrical and electronic equipment must not |
be disposed as unsorted municipal |
waste and must be collected separately. Please |
||
contact an authorized representative of the manufacturerfor information |
|||
concerning |
the decommissioning of |
your equipment. |
Ethernet connector
Document no. M1027807
Introduction
1.2Safety
The following list contains general warnings and cautions you should know before installing, maintaining or servicing the system. Warnings and cautions specific to the use of the system can be found in the User's Guide and User's Reference Manual.
1.2.1 Safety precautions
Warnings |
|
|
|
|
|
|
|
|
|
|
|
WARNING |
A |
WARNING |
indicates a situation in which the user or the patient may be |
in |
|||||||
|
danger of injury or death. |
|
|
|
|
|
|||||
|
Power connection |
|
|
|
|
|
|
||||
|
e |
Use |
only |
hospital-grade |
grounded power outlets |
and power |
cord. Do |
not remove the |
grounding |
||
|
|
pin |
from |
the |
power |
plug. |
|
|
|
|
|
|
e |
Use |
only |
an |
intact |
power |
cord. Replace the power |
cord if it is |
cracked, |
frayed, broken |
or |
|
otherwise |
damaged. |
s |
Do not apply tension to the power cord otherwise the cord may get damaged. |
|
«e |
Donotuse |
extension cords or adapters of any type. |
eBefore starting to use the system, ensure that the whole combination complies with the international standard IEC 60601-1-1 and with the requirements of the local authorities.
Installation |
|
|
|
|
||
s |
Keep |
the monitor |
horizontal |
when the Compact |
Airway Module is |
used. Tilting the monitor may |
|
Cause |
erroneous |
results in |
the Compact Airway |
Module’s readings |
and damage the module. |
e The monitor or its components should not be used adjacentto or stacked with other equipment.
|
If adjacent |
or stacked |
use |
is |
necessary, the monitor and its |
components should |
be observed |
to |
|||||||||||
|
verify |
normal |
operation |
in |
the |
configuration |
in which |
it will |
be used. |
|
|
|
|
||||||
s |
Pins |
of |
connectors |
|
identified |
with the ESD warning symbol |
should not be touched. Connections |
||||||||||||
|
should not be made to these connectors unless ESD precautionary |
procedures |
are used. For |
||||||||||||||||
|
details, |
see |
section |
"ESD |
precautionary |
procedures.” |
|
|
|
|
|
|
|||||||
e |
After |
transferring |
or |
reinstallingthe monitor, |
always |
check |
that it |
is |
properly |
connected and |
all |
||||||||
|
parts |
are |
securely |
attached. |
Pay special |
attention to |
this in |
case |
of |
stacked |
mounting. |
|
Laser radiation
e When using the ARK Barcode Reader, N-SCAN, do not stare into the beam. The N-SCAN is a Class 2 laser product.
External connection
© Do notconnect any external devices to the monitor other than those specified.
Fuse replacement
e Replace a fuse only with one of the same type and rating.
Explosion hazard
e Do not use the monitor in the presence of flammable anesthetics.
9
Document no. M1027807
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
Patient safety
Do not perform any testing or maintenance on the monitor while it is being used on a patient.
PACEMAKER PATIENTS: The impedance respiration |
measurement |
may |
Cause |
rate |
changes in |
||||||
Minute Ventilation Rate Responsive Pacemakers. In |
this case, |
set |
the |
pacemaker |
rate |
||||||
responsive |
mode |
off or turn |
the |
monitor |
impedance |
respiration |
measurement |
off. |
|
||
Never install |
the |
monitor or |
the |
displays |
so that they |
are above |
the |
patient. |
|
|
The monitor must not be used without manufacturer approved mounting attached.
Cleaning |
and service |
|
|
Only trained personnel with proper tools and test equipment should |
perform |
the tests and |
|
repairs |
described in this manual. Unauthorized service may void the |
monitor |
warranty. |
Turn the power off and unplug the power cord before cleaning or service. Completely remove any moisture before reconnecting the power cord to the mains outlet.
Do |
not |
touch any exposed wire or conductive |
surface |
while |
any |
cover |
is removed |
and the |
||||||
monitor |
is energized. The voltages present |
can |
|
cause |
injury |
or |
death. |
|
|
|
||||
Pins |
of |
connectors identified with the ESD |
warning symbol should not be touched. |
Connections |
||||||||||
should |
not be |
made |
to these |
connectors |
unless |
ESD |
precautionary |
procedures |
are |
used. For |
||||
details, |
see 1.2.2. ESD precautionary procedures. |
|
|
|
|
|
|
|||||||
Always |
perform |
an |
electrical |
safety check |
and |
a |
leakage current test |
on the monitor |
after |
|||||
service. |
|
|
|
|
|
|
|
|
|
|
|
|
|
Accessories
Use only accessories, including mounts and batteries, and defibrillator-proof cables and
invasive pressure transducers approved by GE Healthcare. For a list of approved supplies and accessories, see the “Supplies and Accessories” catalog delivered with the monitor. Other
cables, batteries, transducers |
and accessories may cause a safety hazard, damage the |
||||||||||||
equipment or |
the |
system, |
result |
in increased |
emissions |
or decreased |
immunity of |
the |
|||||
equipment or |
system |
or |
interfere |
with the |
measurement. |
Protection against cardiac |
defibrillator |
||||||
discharge |
is |
due |
in part |
to |
the |
accessories |
for |
pulse oximetry (SpO2), |
temperature |
(T) and |
|||
invasive |
pressure |
(P) |
measurement. |
|
|
|
|
|
|
Single |
use |
accessories are not designed to |
be reused. |
Reuse may |
cause a risk |
of |
contamination |
|||
|
ane |
affect |
tha |
Pasas rarn ant |
Aair o, |
|
|
|
|
|
|
|
απ |
ATPULL |
UNO |
MILUOJUTUTIITUTTTUT |
|
|
|
|
|
|
|
Cautions |
|
|
|
|
|
|
|
|
|
|
|
CAUTION |
A CAUTION |
indicates a condition that may |
lead to |
equipment |
damage |
or |
malfunction. |
installation
Leave space for air circulation to prevent the monitor from overheating.
Ensure that the module is properly orientated (i.e. module release latch facing downward) before insertion.
Before connecting the power cord to the mains outlet, check that the local voltage and
frequency correspond with the rating stated on the device plate on the rear panel of the monitor. See instructions for different displays in section “Displays”.
Turn off the power before making any rear panel connections.
10
Document no. M1027807
Introduction
Before use
eAllow two minutes for warm-up and note any error messages or deviations from normal operation.
e |
Cleanthe rear |
panel |
fan dust |
filters once |
a |
month or |
whenever necessary. |
||||
s |
Donotconnect |
a |
sampling |
line to |
the |
female Patient |
Spirometry connector while the other end |
||||
|
of the |
sampling |
line |
is connected |
to the |
D-fend water |
trap. The pressure in the gas sampling |
||||
|
system |
may cause |
damage |
to the |
PVX |
unit |
pressure |
transducers. |
Autoclaving and sterilizing
e Do not autoclave any part of the monitor.
e Do not gas sterilize the modules.
Cleaning and service
e Donotusehypoclorite, ammonia-, phenol-, or acetone based cleaners. These cleaners may damage the monitor surface.
eDonotimmerseanypartof the device in any liquid, or allow liquid to enter the monitor or modules.
© |
Donotapplypressurizedairto |
any outlet or |
tubing |
connected to |
monitor. |
|
|
|
||||
e |
Electrostatic |
discharge |
through |
the PC boards may damage the |
components. |
Before handling |
||||||
|
PC boards, wear a static |
control |
wrist |
strap. |
Handle |
all |
PC boards |
by their |
non-conductive |
edges |
||
|
and use anti-static containers |
when |
transporting them. |
|
|
|
|
|||||
e |
Do not break or bypass the patient isolation barrier when testing PC boards. |
|
|
|||||||||
® |
Donotclean |
the spirometry tubes with high |
pressure |
air 02 flushing while |
the |
spirometry |
tubes |
are connected to Patient Spirometry connector. High differential pressure may damage PVX unit pressure transducers.
Special components
e Special components are used in these monitors that are vital to assure reliability and safety. GE Healthcare assumes no responsibility for damage, if replacement components not approved by GE Healthcare are used.
|
A lithi |
С |
an |
atala |
ar |
AA |
» |
Ρζο |
|
|
|
|
|
|
|
|
O |
|
|
|
|
|
|
|
|
local |
regulations. |
|
|
|
|
|
|
|
|
|||
Batteries |
|
|
|
|
|
|
|
|
|
|
||
The |
battery packages |
in |
the |
central |
unit, F-CPU |
and in the power supply |
unit of F-CU8 |
contain lead |
||||
acid |
(Pb) |
which |
is hazardous |
to |
the |
environment |
and therefore needs to |
be disposed |
of carefully |
according to local regulations.
To replace the batteries safely, please refer to the service instructions in this manual.
e |
Do |
notshort-circuit the battery terminals, this |
may |
produce |
a very |
high |
current, |
which |
will |
|
|
damage |
the battery. |
|
|
|
|
|
|
|
|
© |
Do |
not |
dispose of the battery into open flame, |
nor |
put the |
battery |
near |
fire, as |
it may |
explode. |
11
Document no. M1027807
Datex-Ohmeda S/5 Anesthesia |
and Critical |
Care |
Monitors |
|
|
|
|
|
|
|
|||||
s |
Donotdismantle |
the |
battery. |
It contains |
electrolyte, which may damage clothing |
or |
cause |
injury |
|||||||
|
to skin |
or |
eyes. If |
exposed |
to |
electrolyte, |
wash the injured area with |
plenty of |
clean |
water |
and |
||||
|
contact |
a |
doctor. |
|
|
|
|
|
|
|
|
|
|
|
|
See |
also |
section “Symbols”. |
|
|
|
|
|
|
|
|
|
||||
Storage |
and |
transport |
|
|
|
|
|
|
|
|
|
||||
Do |
not store or |
transport the monitor outside |
the |
specified temperature, |
pressure |
and |
humidity |
|
|||||||
ranges: |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Temperature |
|
|
|
|
|
-10...+50 °C/14...122 °F |
|
|
|
|
|
||||
Atmospheric |
pressure |
|
|
|
660...1060 hPa/500...800 mmHg/660...1060 |
mbar |
|
||||||||
Relative |
humidity |
|
|
|
10...90% |
non-condensing |
|
|
|
|
|
For display specific environmental requirements see specifications in the “Display” slot.
1.2.2 ESD precautionary procedures
e To avoid electrostatic charges to build up, it is recommended to store, maintain and use the
|
equipment |
at |
a relative humidity |
of 30% or greater. |
Floors |
should |
be |
covered |
by |
ESD |
dissipative |
||||||
|
carpets |
or |
similar. Non-synthetic clothing should be used |
when |
working |
with |
the |
component. |
|||||||||
e |
To prevent |
applying a possible electrostatic discharge |
to the ESD sensitive parts |
of |
the |
|
|||||||||||
|
equipment, |
one should touch the metallic frame of the |
component or to |
a large |
metal |
object |
|||||||||||
|
located |
close |
to the |
equipment. |
When |
working with |
the |
equipment |
and |
specifically |
when the |
||||||
|
ESD sensitive |
parts |
of the equipment |
may be touched, |
a |
grounded |
wrist |
strap |
intended |
for use |
with ESD sensitive equipment should be worn. Refer to the documentation provided with the wrist straps for details of proper use.
ESD precautionary |
procedure |
training |
|
|
|
|
|
|
|
|
|
|
|
|
|
||
It is |
recommended |
that all potential users receive an |
explanation |
of |
the |
ESD warning symbol |
and |
|
|||||||||
training in |
ESD |
precautionary |
procedures. |
|
|
|
|
|
|
|
|
|
|
|
|
||
The |
minimum |
contents of an |
ESD precautionary |
procedure |
training |
should |
include an |
introduction |
to |
||||||||
the physics of electrostatic charge, the voltage levels |
that can |
occur |
in |
normal |
practice |
and |
the |
|
|||||||||
damage that can be done to electronic components |
if they |
are |
touched |
by |
an |
operator |
who |
is |
|
||||||||
electrostatically charged. Further, an explanation should be |
given |
of |
methodsto |
prevent |
build-up of |
|
|||||||||||
electrostatic charge and how and why to discharge one’s body to |
earth or to |
the frame |
of the |
|
|
||||||||||||
equipment |
or |
bond |
oneself by means of a wrist |
strap |
to the |
equipment or the |
earth priorto making |
a |
|||||||||
connection. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
1.2.3Disposal
Dispose of the whole device and parts of it according to local environmental and waste disposal regulations.
12
Document no. M1027807
System description
2System description
2.1Introduction
Datex-Ohmeda monitors build up a freely configurable modular system. The architecture is designed to enable different module combinations so that the user is able to get the desirable parameter and feature set. This modular approach makes it possible to add new features when they are needed.
2.2Bus structure
The operation |
of |
Datex-Ohmeda monitors |
is based |
on |
two communication |
channels, |
the |
CPU |
bus and |
||||||||||
module |
bus. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
In the 5-Module |
Central |
Unit, |
PC |
boards |
receive |
power from the |
F-CPU |
power |
supply |
and |
the |
|
|||||||
parameter modules |
receive |
power |
from a |
separate power supply in the 5-Module |
Frame |
unit. These |
|||||||||||||
power supplies |
|
are |
both |
fed |
by the N-AC |
Power |
Unit. |
In the 8-Module |
Central |
Unit, |
F-CU8, all |
PC |
|||||||
boards connected to the CPU |
bus, as well as the parameter modules attached to |
the |
module |
bus |
|||||||||||||||
receive |
power |
from |
the |
same |
power supply, which |
is |
an integral |
part of the |
Central |
Unit, |
F-CU8. |
||||||||
|
UP! |
|
|
|
|
CPU |
|
Display |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Controller |
|
|
|
|
|
|
|
|
|
|
Module |
Bus |
|
Parameter |
Parameter |
|
Parameter |
Module |
Module |
000 |
Module |
Figure1-—Generalstructure-ofbusS/5 system
The |
CPU |
bus |
is a |
communication |
channel |
used |
only |
for |
internal data |
transfer. |
Itis based on |
the |
|
ISA |
|||||||||||
bus |
used |
in |
IBM |
PC |
computers. |
Data |
is transferred |
on this 16 |
bit |
wide |
bus using |
the CPU |
clock |
|
|
||||||||||
frequency. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
The |
module |
bus |
is used to connect the parameter modules to |
the Central Unit. The bus is based |
on |
||||||||||||||||||||
the |
industry |
standard RS-485, which |
usesa |
differential |
serial |
method |
to |
transfer |
data. This |
type |
of |
||||||||||||||
bus |
is |
robust |
and |
it allows |
parameter |
modules |
to be inserted |
or |
removed |
while the power |
is |
on. |
|
The |
|||||||||||
module |
|
bus |
uses |
a |
500 kbps data transfer rate |
and |
can |
be used |
for longer distances |
than |
the |
CPU |
|||||||||||||
bus, |
e.g. |
for |
external |
frame |
connections. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||
The |
RS-485 |
type |
of |
serial communication |
supports |
so-called |
multidrop or party line connections. This |
||||||||||||||||||
means |
that |
all parameter modules connected to the module |
bus |
use |
exactly |
the |
same |
lines |
for |
|
|
||||||||||||||
communication. |
The |
advantage |
of this is that |
all bus |
connectors |
are identical |
and |
the modules |
can be |
||||||||||||||||
connected in any |
order and |
position. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
13 |
Document no. M1027807
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
2.3Distributed processing
Asystem |
assembled from |
S/5 |
products |
is a multiprocessor |
system. All |
parameter modules |
have their |
|||
Own microprocessor, which performs functions such as module key control, waveform filtering, |
||||||||||
parameter |
related computing |
and |
pneumatic |
control, |
etc. |
At the same |
time the main CPU |
performs |
||
higher level tasks such as |
trending |
and |
alarm |
control. |
While the parameter modules and CPU are |
|||||
performing |
their tasks, the UPI (Universal Peripheral Interface) microprocessor handles all functions |
|||||||||
needed to |
transfer data between the parameter modules and the CPU. At the same time the Display |
|||||||||
controller |
microprocessor |
performs |
pixel |
calculations |
for graphics. |
|
|
Module
[ Dual-port |
| |
Er |
RAM |
Il |
A |
Display
Controller Board
Display
Display
memo
|
a |
= |
Qs |
|
|
|
|
Command
Board/Bar
Figure 2 |
Distributed processing in S/5 system |
|
|
|
This kind of parallel processing gives one major advantage |
to |
centralized processing. When new |
||
parameter modules or PC boards are added |
to the system, |
the |
processing power is increased. As a |
|
result, the |
system does not slow down when |
new features |
are |
added. |
2.4Module communication
The communication master controlling data transfers between the CPU and parameter modules is
called |
UPI processor. |
It sends data |
to each |
connected |
module |
100 times |
a second. |
Modules |
respond |
||||||||
to |
each |
data request |
immediately |
by sending a data package, |
whose |
length |
depends |
on |
the |
type |
of |
||||||
the |
module. |
This communication |
protocol |
ensures that each module receives and |
sends |
data |
every |
||||||||||
10 |
ms. |
If a |
module does not respond to data requests, |
the UPI |
processor |
presumes that |
the |
module |
is |
||||||||
disconnected. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|||
Parameter modules may hold a static (fixed) or dynamic address, which |
the |
UPI processor |
uses when |
||||||||||||||
sending out data. Two parameter |
modules of the same type must not |
be |
fitted onto |
the |
same |
monitor |
|||||||||||
since they |
might reply |
to a data request simultaneously, thus |
causing |
communication |
errors. |
|
|
14
Document no. M1027807
System description
UPI Board
Dual-port
RAM
E prt
a Е
Microcontroller |
[= |
BES |
Figure 3 |
Principle of |
UPI |
section |
operation |
|
|
|
|
|
|
|
The UPI processor collects and |
stores |
all |
data that is received from |
the |
parameter |
modules |
into |
a dual |
|||
port RAM, |
which is mapped |
directly to |
the address space of the main CPU. The main |
CPU |
reads |
data |
|||||
from the memory while the UPI |
processor guarantees that the data |
is up to date. This operation |
also |
||||||||
works in the other direction. In |
this the |
main CPU fills the dual port |
RAM |
with data |
and |
the |
UPI |
|
|||
processor |
distributes it to the |
parameter |
modules. |
|
|
|
|
|
|
2.5Software loading
The program memory on CPU board is loaded with monitor software at the factory. The software is
used for running all the functions that are procedures the CPU board is fitted with a loaded.
integrated into the PC board. For service and upgrade PCMCIA card drive through which new software can be
iService!
|
|
| Software |
|
|
|
|
— |
|
|
|
|
I |
|
|
|
JPCMCIA! |
|
||
UPI |
CPU |
~ |
> |
Display |
Board |
Board |
|
|
Controller |
|
|
|
|
Board |
|
|
Flash |
|
|
Program |
| |
Program |
|
Program |
Memory |
|
Memory |
|
Memory |
CPU Bus
Figure 4 Software loading
15
Document no. M1027807
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
2.6Parameter modules
Module E::.
keys
Patient |
electronics |
E: |
eee |
000000], |
|
Peripheral |
È |
|
|
|
drivers |
Е |
|
|
|
|
|
|
Bus
Module
Figure 5 |
General structure of parameter modules with patient isolation |
|
|
|
|
|||||||||||
The |
detailed |
structure |
of a |
parameter |
module |
depends |
on the |
specific needs for |
each |
individual |
|
|||||
parameter. |
However, |
some |
common |
parts |
are |
used |
in the parameter modules. The electronics |
inside |
||||||||
the |
module |
is usually |
divided |
into isolated |
(floating) |
and non-isolated sections. Typically, the |
|
|||||||||
non-isolated |
section |
consists of buffers to interface |
the parameter module to the |
module |
bus while |
|||||||||||
the rest of the electronics |
is |
located |
in the |
isolated |
section. The |
isolated |
section |
includes |
the |
|
||||||
microcontroller together with memory components, |
the |
front-end analog |
electronics |
(amplifiers, |
etc.) |
|||||||||||
and |
sensor |
drives. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
16
Document no. M1027807
System installation
3 System installation
3.1 Unpacking instructions
1. |
Confirm that the packing box |
is |
undamaged. |
If the box is damaged, |
contact |
the |
shipper. |
|||||
2. |
Open the |
top |
of |
the box and |
carefully unpack |
all components. |
|
|
|
|
||
3. |
Confirm |
that |
all |
components |
are |
undamaged. |
If any |
of the components |
are |
damaged, contact |
||
|
the shipper. |
|
|
|
|
|
|
|
|
|
|
|
4. |
Confirm |
that |
all |
components |
are |
included. If any of |
the components |
are |
missing, |
contact your |
||
|
GE Healthcare |
distributor. |
|
|
|
|
|
|
|
|
3.2Choosing location
Consider the following aspects:
e |
lighting |
s |
space |
e |
connections |
selectromagnetic and radio frequency interference, see Appendix 8. ElectroMagnetic Compatibility
|
e |
environment |
|
|
|
|
|
|
|
WARNING |
The |
monitor or |
its |
components |
should not be |
used adjacent to |
or |
stacked |
with |
|
other equipment. |
If adjacent |
or stacked use |
is necessary, the |
monitor and |
its |
|||
|
components should be observed to verify normal operation in |
the |
configuration in |
||||||
|
which it will be |
used. |
|
|
|
|
|
3.3Central Unit; S/5 8-Module Frame, F-CU8
Figure 6 Central Unit: S/5 8-Module Frame, F-CU8
The front of the Central |
Unit, F-CU8, houses |
plug-in |
parameter |
modules which interface the |
system |
||||||
with external devices. |
The back of the |
Central |
Unit |
houses the |
system |
circuit |
boards, |
together |
with |
||
expansion |
slots which |
accommodate |
PC boards which interface with |
external |
devices |
and |
the |
Datex |
|||
Ohmeda |
Network. |
|
|
|
|
|
|
|
|
|
|
17
M1027807
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
3.3.1Connecting to mains
Connect the power cord to the mains power inlet on the rear of the Central Unit and to the wall socket.
WARNING |
The |
power cord |
may |
only |
be |
connected to a |
three-wire, |
grounded, hospital grade |
||||||
|
|
receptacle. |
|
|
|
|
|
|
|
|
|
|
||
3.3.2 |
Connecting to |
Datex |
Ohmeda |
Network |
|
|
|
|
|
|
||||
|
|
To connect |
the monitorto |
the |
Datex Ohmeda |
Network, |
make sure |
a |
Network Board |
B-UPI4NET |
is |
|||
|
|
installed. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Use |
the Monitor-Network |
cable to connect the monitor the network as follows: |
|
|
||||||||
|
|
1. |
Make |
sure that the |
power to |
the Central |
Unit is turned off. |
|
|
|
|
|||
|
|
2. |
Connect the RJ-45 |
connector |
and the Identification Plug to |
the |
corresponding |
connectors |
on |
|||||
|
|
|
the B-UPI4NET. |
|
|
|
|
|
|
|
|
|
Connect the other RJ-45 connector the corresponding connector on the wall box.
Turn on the power to the Central Unit. Enter the Network service menu:
Monitor Setup - Install/Service (password 16-4-34) -
Service (password 26-23-8) -
Frame - Network
Make sure that the |
monitor's |
network communication has |
been |
set according to the used |
|||
network software: |
|
|
|
|
|
|
|
Network software S-CNETO1 -> DRI |
Level |
= |
2001 |
|
|
||
Network software |
S-CNETO2 -> DRI |
Level |
= |
2001 |
|
|
|
Network software L-NETO3 -> DRI Level = 2003 |
|
|
|||||
Network software |
LNETO5 -> |
DRI |
Level |
= 2005 |
|
|
|
If necessary, change the monitor's |
network |
communication |
by |
selecting DRI Level and turning |
|||
the ComWheel. |
|
|
|
|
|
|
|
NOTE: If the DRI level is changed, the monitor will restart automatically.
Confirm that the network symbol and ‘Connected to Network’ message are displayed on the upper part of the screen.
3.3.3 Inserting the parameter modules
1. Ensure that the module is properly orientated (i.e. module release latch facing downward). 2. Align the module insertion guide slot with the insertion guide.
3. Push the module into the frame until it clicks.
18
System installation
Figure 7 |
Module insert |
NOTE: The Compact Airway Modules cannot be placed into the third and fourth slot from the right- hand side of the F-CU8 Central Unit.
NOTE: |
Use |
only one Extension |
Frame F-EXT4 in one F-CU8 Central Unit. |
NOTE: |
Do |
not use two or more |
parameter modules with identical functions in the monitor system. |
Take special care not to do this when using the Extension Frame, F-EXT4.
Parameter modules or boards with identical functions are:
e |
Hemodynamic |
multiparameter |
modules, |
E-PRESTN, E-PRETN, E-RESTN, M-PRESTN, M-PRETN, |
||
|
M-RESTN, M-ESTPR, M-ETPR, M-ESTR, M-NESTPR, M-NETPR, M-NESTR, M-NE12STPR, M- |
|||||
|
NE12TPR, M-NE12STR |
|
|
|
||
e |
Dual Pressure Modules, E-PP/M-PP |
|
|
|||
e |
Pressure Modules, E-P/ M-P, Pressure Temp Module, E-PT/ M-PT |
|||||
e |
Cardiac Output |
Modules, E-COP, E-COPSv, M-COP |
and M-COPSv |
|||
s |
NIBP Modules, M-NIBP and hemodynamic modules w/ N measurement |
|||||
s — |
Airway Modules, E-CO, E-COV, E-COVX, |
E-CAIOVX, |
E-CAiOV, E-CAIiO, E-miniC, M-C, M-CO, M- |
|||
|
COV, M-CAiO, M-CAiOV, M-CAiOVX, G-O, G-OV, G-AO, G-Ai0, G-AOV, G-AiOV, M-MiniC |
|||||
e |
Interface Module, E-INT / M-INT and Interface Board, B-INT |
|||||
e |
Oxygen Saturation Modules, E-NSAT/ M-NSAT and M-OSAT |
|||||
e |
Tonometry Modules, E-TONO/ M-TONO |
|
|
|||
e |
NeuroMuscular |
Transmission |
Modules, |
E-NMT/ M-NMT |
||
e |
EEG |
Modules, |
E-EEG/ M-EEG |
|
|
|
e |
BIS |
Modules, |
E-BIS/ M-BIS |
|
|
|
19
M1027807
Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
3.3.4 Positioning of PC boards
©
©
o EEBo
“ER
-|
+
0
| j ceo
(7)
~
im
Figure 8 Rear view and positioning, F-CU8 rev. 10
(1)Mains power inlet
(2) Battery 24 Vde input connector (F-CU8 rev. 10)/ Display power outlet (F-CU8 rev. 03 to 09)
(3)UPINET Board, B-UPI4NET
(4) |
CPU Board, B-CPU5 |
|
( 5) |
Display Controller Board, B-DISPX |
(primary/screen 1) |
( 6) ) |
Display Controller Board, B-DISPX (secondary/screen 2), |
|
|
Keyboard Interface Board, B-ARK |
|
(7) |
Interface Board, B-INT, |
|
|
Gas Interface Board, B-GAS, |
|
|
Display Controller Board, B-DISPX |
(3 screen) |
(8)Service reset button
(9)Equipotential connector
(10)Back plate lid for CPU Board, B-CPU5
NOTE: Authorized personnel only may open the cover of 24Vdc input connector (use aflat screwdriver).
3.3.5Replacing PC Boards
For service procedures refer to Part Il.
1. Make sure that the power is turned off the monitor and unplug the power cord. Press and hold the service reset button on the rear panel for at least five seconds or until an audible tone is heard (see Figure 9).
20