Datex-Ohmeda S-5 Anesthesia Monitor, S-5 Critical Care Monitor User manual

0 (0)

Datex-Ohmeda

S/5™ Anesthesia Monitor and Critical Care Monitor

Technical Reference Manual

 

 

 

 

 

 

All specifications are subject to change without notice.

 

 

 

 

CAUTION: U.S. Federal law restricts this device

to

sale by

or on

 

the

order

of a

licensed medical

practitioner.

 

 

 

 

Outside

the

USA,

check

local

laws

for

any

restriction that

may apply.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M1021563

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

June, 2005

Datex-Ohmeda

Inc.

 

 

 

 

 

 

 

 

GE

Healthcare

Finland

Oy

Р.О. Вох 7550

 

 

 

 

 

 

 

 

 

 

 

Helsinki,

Finland

Madison, WI 53707-7550, USA

 

 

 

 

 

 

 

 

 

 

P.0.

Box

900

Tel:

+1

608

221

1551

 

 

 

 

 

 

 

 

 

FI-00031 GE, FINLAND

Fax:

+1

608

222

9147

 

 

 

 

 

 

 

 

 

Tel:

+358

10

39411

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fax:

+358

9 1463310

 

 

 

 

 

 

 

www

datex-ohmeda.com

www.gehealthcare.com

 

 

 

 

 

 

 

 

O

2005 Copyright

General

Electric Company

Intended purpose (Indications for use)

The Datex-Ohmeda S/5 Anesthesia Monitor with L-ANEO5 or L-ANEO5A software is intended for multiparameter patient monitoring with optional patient care documentation.

The S/5 Anesthesia Monitor with L-ANEO5 and L-ANEOSA software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/ regional perfusion, Bispectral index (BIS), Entropy (State Entropy and Response Entropy) and neurophysiological status of all hospital patients.

The

S/5 Anesthesia Monitor with L-ANEO5 and L-ANEOSA

software

when

using

BIS

is for

monitoring the

state

of

the

brain by data acquisition and processing of electroencephalograph signals and

may

be

used

as

an

aid in

 

monitoring

the

effects

of certain

anesthetic

agents.

 

 

 

 

 

 

 

 

 

 

 

 

The

S/5 Anesthesia Monitor with L-ANEO5 and L-ANEOSA

software

is

also indicated for

documenting

patient

 

care

related information.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The

S/5 Anesthesia Monitor with L-ANEO5 and L-ANEODA

software

is

indicated

for

use

by

qualified

medical

 

personnel

only.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The

Datex-Ohmeda S/5 Critical

Care Monitor with L-ICUO5

or L-ICUO5A

software

is

intended

for multiparameter

patient

monitoring.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The

S/5 Critical Care Monitor with L-ICUO5 and L-ICUO5A software is indicated for

monitoring of

hemodynamic

(including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion,

 

Bispectral

index

(BIS)

and neurophysiological status of all

hospital

patients.

 

 

 

 

 

 

 

 

The

S/5

Critical

Care

Monitor with L-ICUO5 and L-ICUO5A

software

when

using BIS

is for

monitoring

the

state

of

the

brain by data acquisition and processing of electroencephalograph signals and

may

be

used

as

an

aid in

 

monitoring

the

effects

of certain

anesthetic

agents.

 

 

 

 

 

 

 

 

 

 

 

 

The

S/5 Critical Care Monitor with L-ICUO5 and L-ICUO5A software is indicated for

use by qualified

medical

 

personnel

only.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Classifications

In accordance with IEC 60601-1

Class | and internally powered equipment - the type of protection against electric shock.

Type BF or CF equipment. The degree of protection against electric shock is indicated by a symbol on each parameter module.

Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.

Continuous operation according to the mode of operation. In accordance with IEC 60529

With F-CU8 Central Unit: IPXO - the degree of protection against harmful ingress of water. With F-CU5(P) Central Unit: IPX1 - the degree of protection against harmful ingress of water. In accordance with EU Medical Device Directive

The Datex-Ohmeda S/5 Anesthesia Monitor is classified as IIb. The Datex-Ohmeda S/5 Critical Care Monitor is classified as IIb. In accordance with CISPR 11:

Group 1, Class B (F-CU5(P), F-CPU and N-AC: Class A)

s

Group

1 contains all ISM (industrial, scientific

and

medical) equipment

in which there is

intentionally

 

generated and/or used conductively coupled radio-frequency energy

which is necessary

for the

internal

 

functioning of the equipment

itself.

 

 

 

 

 

 

 

e

Class B

equipmentis

suitable

for use

in domestic

establishments and

in

establishments

directly

connected

 

to a low

voltage power

supply

network

which

supplies buildings used

for

domestic purposes.

 

eClass A equipment is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.

Responsibility of the manufacturer

GE Healthcare Finland Oy (GE) is responsible for the safety, reliability and performance of the product only if: modifications, service and repairs are carried out by personnel authorized by GE.

the

electrical installation of the room complies with appropriate reguirements.

the

equipment is

used in accordance with the User's Guide and serviced and maintained in accordance

with

the Technical

Reference Manual.

Master table of contents

Datex-Ohmeda S/5™ Anesthesia and Critical Care Monitors

Technical Reference Manual, Order code: M1021563

Part I, General Service Guide

Document No. | Updated

Description

 

 

M1027807

Introduction, System

description,

Installation,

 

Interfacing, Functional check, General troubleshooting

M1027808

Planned

Maintenance

Instructions

Part II, Product Service Guide

Document No.

Updated

Description

 

 

 

M1027809

 

AM, CCM

Service

Menu

M1027810

 

8-Module

Frame,

F-CU8

M1027811

 

5-Module Frame, F-CU5(P)

M1027812

 

CPU Board, B-CPU5

M1027813

 

UPINET Board, B-UPI4NET

M1023412

 

Displays and Display Controller Boards

M1027814

 

Command

Boards and

Bars

M1027815

 

Interface Board, B-INT

M1027816

 

Extension Frame, F-EXT, Extension

Module,

M-EXT

M10 39372

 

AM,

CCM Spare

Parts

F

NR

AK oon©

ona

pa©

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

About this manual

Notes to the reader

This Technical Reference Manual is intended for service personnel and engineers who will service and maintain Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

e

The order

code for

the

entire printed

manual

is

M1021563.

The manual

includes Technical

 

Reference

Manual

Slots

and

every

slot

has an

individual document number.

M102 7807 is the

 

document

number

of

this

first

slot.

 

 

 

 

 

 

 

 

e

Parti gives

the

reader

an

overview

of

the

S/5

Anesthesia Monitor and S/5 Critical Care Monitor.

 

It contains

the

information needed

to

install,

interface and

troubleshoot

the

monitors.

Instructions for functional check and planned maintenance are also included. Read the manual through and make sure that you understand the procedures described before the installation of the monitor. To avoid risks concerning safety or health, strictly observe the warning indications. İf you need any assistance concerning the installation, please do not hesitate to contact your authorized distributor.

e

Partll contains detailed descriptions

of each component

of

the

S/5 AM, CCM, such as frame

 

unit and displays. Service check for

each

product is

included

in

these slots. Service Menu slot

 

contains all the service menus and Spare

Parts slot

all the

spare

parts information for this

 

monitor.

 

 

 

 

 

 

For

information

of

parameter modules,

Remote

Controller

and

Device

Interfacing Solution refer to

the

S/5 E-Modules Technical Reference Manual. Service check for each

of these

products

is

included

in

these slots.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The

manufacturer

reserves the

right to

change

product specifications

without

prior notice.

Although

the

information

in

this manual

is believed

to

be

accurate

and

reliable,

the manufacturer

assumes

no

responsibility for

its use.

 

 

 

 

 

 

 

 

 

 

 

 

GE Healthcare Finland Oy (GE)

assumes

no

responsibility

for

the use

or reliability of its software in

 

eguipment that

is

not furnished

by GE.

 

 

 

 

 

 

 

 

 

 

 

Related documentation

S/5 Modules

S/5 E-Modules Technical Reference Manual

S/5 M-Modules Technical Reference Manual

Other devices closely related to the S/5 Anesthesia Monitor: iCentral, User's Reference Manual

3/5 Arrhythmia Workstation User’s Reference Manual

S/5

Anesthesia Monitor

 

Instructions for daily use, clinical

aspects and basic methods of measurement:

S/5

Anesthesia

Monitor,

User's

Guide

S/5

Anesthesia

Monitor,

User's

Reference Manual

S/5

Critical Care Monitor

 

Instructions for daily use, clinical

aspects and basic methods of measurement:

S/5

Critical

Care

Monitor,

User's

Guide

S/5

Critical

Care

Monitor,

User's

Reference Manual

1

Document no. M1027807

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

Conventions used

To help you find and interpret information easily, the manual uses consistent text formats:

BS

 

Sign

the

check form

after

performing the procedure.

 

 

 

 

 

 

 

 

 

 

 

 

 

Hard

Keys

Names

of

the

hard keys on

the Remote

Controller,

Command Bar and

modules are

written in the

 

 

following

way:

Others.

 

 

 

 

 

Hard

Keys

Menu

ItemsSoftware

terms

that identify

window

parts or menu items

are written in

bold italic: ECG

 

 

Setup.

 

 

 

 

 

 

 

 

 

Menu access is described from top to bottom. For example, the selection of the Monitor Setup hard

key, the Screen 1 Setup menu item and the Waveform Fields menu item would be shown as

Monitor Setup - Screen 1 Setup - Waveform Fields.

‘Messages’

Messages

(alarm messages, informative messages) displayed on the screen are

written

inside single

 

quotes:

‘Please

wait’.

 

 

 

 

 

 

 

 

 

 

 

 

 

“Sections”

When

referring

to

different

sections

in this

manual

or to

other manuals,

manual

names

and section

 

names are

enclosed

in

 

double

quotes:

 

 

 

 

 

 

 

 

See section

"Cleaning

 

and

care."

 

 

 

 

 

 

 

 

 

 

Please refer

to

"iCentral

User's

Reference

Manual:

Alarms. "

 

 

 

Hypertext links

Hypertext

links

on

PDF

versions

are

written

in blue

color.

 

 

 

WARNING

Warnings

are

written in

 

the

following

way:

 

 

 

 

 

 

WARNING

This

is

a

WARNING.

 

 

 

 

 

 

 

 

 

 

 

 

 

CAUTION

Cautions

are

written

in

the

following

way:

 

 

 

 

 

 

CAUTION

This

is

a

CAUTION.

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE

Notes

are

written

in

following way:

 

 

 

 

 

 

 

 

 

 

NOTE:

This

is

a

NOTE.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In

this

manual,

the

word

“select”

means

choosing

and

confirming.

 

 

 

Illustrations

and

names

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

All

illustrations

in

this

manual

are

only

examples,

and may not necessarily reflect your system

 

settings

or

data

displayed

in your

system.

If a particular

selection is not

available in your system, the

 

selection

is shown

grayed.

 

 

 

 

 

 

 

 

 

 

 

2

Document no. M1027807

Introduction

1 Introduction

The Datex-Ohmeda S/5 Anesthesia Monitor is a modular multiparameter patient monitor used during anesthesia in operating rooms.

The Datex-Ohmeda S/5 Critical Care Monitor provides full patient profile throughout the care period.

The modular design makes the system flexible and

easy to upgrade. In

addition

to parameter

changes, the modularity

includes

an easy upgrade

to anesthesia record

keeping,

monitor networking

and interfacing with other

external

devices.

 

 

 

Re eso

Ara,

РМval|

3

Document no. M1027807

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

1.1Symbols

1.1.1Symbols on transport packaging

5

The

contents

of the transport package

are fragile and

must be handled with care.

| |

Indicates the correct upright position

of

the

transport

package.

 

The

transport package must be kept

in

a dry

environment.

Y

Indicates the

temperature limitations

within

which the

transport package should be

 

stored.

 

 

 

 

 

1.1.2Symbols on equipment

NA

The battery

contains

lead

acid,

and

in

the event of disposal must be separated from other

waste according to local

regulations.

 

Pb

 

 

 

 

 

 

 

This battery

contains

lead

and

can

be

recycled.

Pb

 

 

 

 

 

 

 

 

Dangerous

voltage.

 

 

 

 

 

 

When using the

ARK

Barcode

Reader, N-SCAN, do not stare into beam. The N-SCAN

AR

Barcode Reader

is a

Class 2

laser product.

4

Document no. M1027807

Introduction

1.1.3Equipment safety symbols

Attention,

consult accompanying

documents.

 

 

 

 

 

 

 

 

 

When

displayed next

 

to

the

02

 

value,

indicates

that

the Fi02

low alarm

limit is

set

below

21

%.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

When

displayed

next

to

the

HR

value,

indicates

that

there

is

a

risk that

the

 

monitor

counts

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

*

pacemaker

spikes

 

because

the

pacer is

set

on

R.

 

 

 

 

 

 

*

T-waves

because

a

wide

QRS

is

selected.

 

 

 

 

 

 

 

 

 

 

On

 

the

15”

display,

D-VMC15,

 

indicates that

the

display should be supplied from

the

mains

outlet

(or

from

the

Central

Unit,

F-CU8

Rev.03

- 09).

 

 

 

On

the

15”

display,

D-VNC15,

 

indicates that the display must be supplied from

the

mains

outlet via an appropriate additional separating

transformer (or from

 

the

Central

Unit, F-CU8 Rev.03

-

09).

 

 

 

 

 

 

 

 

 

 

 

On

the

15"

LCD display, D-LCC15, indicates that

the display must only be used

together

with

the

original type

 

of

D-LCC15

power

adapter.

The

display

should

be

supplied from the mains outlet (or from the Central Unit, F-CU8 Rev.03 - 09).

On

the

17”

display,

D-LCC17

indicates that the display

must be used only

together

with the original D-LCC17 power

adapter.

 

On

the

17”

display,

D-VHC17

rev.00-01,

indicates that

the display must only be

supplied from the mains outlet, not from the Central Unit, F-CU8.

On the 17” display, D-VHC17 rev. 02 or higher, indicates that the display should

be supplied from the mains outlet via an appropriate additional separating

 

transformer (or

from

the

Central Unit,

F-CU8

Rev.03

- 09).

 

 

 

On

the

19”

display,

D-LCC19,

indicates

that

the

display

must

only

be supplied

from

the

mains

outlet

via

an

appropriate

additional

separating

transformer

and

the original D-LCC19

 

power

adapter,

not

from the Central

Unit,

F-CU8.

 

On

the

21”

display,

D-VSC21,

indicates

that

the

display

must

only

be supplied

from

the

mains

outlet

via

an

appropriate

additional

separating

transformer,

not

from

the

Central

Unit,

F-CU8.

 

 

 

 

 

 

 

 

 

 

 

On

the

Interface

Module

E-INT,

M-INT,

indicates

that

it is

for connecting external

On the E-TONO, M-TONO module indicates that the module should only be used with Tonometrics catheters.

On the E-miniC module indicates that airway gases should be calibrated every six months in normal use and every two months in continuous use.

BIS: On

the Aspect DSC indicates that the converter must not be opened for any

reason

or autoclaved.

On the E-PRESTN,

E-PRETN, E-RESTN, E-PSM, E-PSMP, E-P,

E-PP, E-PT, E-COP and

E-COPSv

module

indicates that

protection

against

cardiac

defibrillator

discharge

is due in

part to the

accessories

for pulse

oximetry

(SpO2),

temperature

(T) and

invasive

pressure

(P)

measurement.

 

 

 

 

5

Document no. M1027807

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

- © On the E-NMT, M-NMT module indicates the following warnings:

*

Do not

place

the

NMT

stimulating electrodes on the patient’s chest.

*

Always

stop

the

NMT

measurement before handling the stimulating electrodes.

* Never subject a patient with an implanted electronic device to electrical Stimulation without consulting a medical specialist first.

-Qn the rear or bottom panel this symbol indicates the following warnings and Cautions:

*

Electric

shock

hazard.

Do

not open the cover or the

back.

Refer servicingto

qualified

service

personnel.

 

 

 

 

 

 

*

For

continued

protection

against

fire hazard,

replace

the

fuse only

with one of

the same

type

and

rating.

 

 

 

 

 

 

*

Disconnect

from

the power

supply

before servicing.

 

 

 

*

Do

not

use

the

monitor

without manufacturer

approved

mounting

attached.

-On the rear panel of the power unit (N-AC) this symbol indicates the following

warnings and cautions:

*Electric shock hazard. Do not open the cover. Refer servicing to qualified service personnel.

*Does not contain field replaceable fuses.

Type BF (IEC 60601-1) protection against electrical shock.

Type BF (IEC 6060 1-1) defibrillator-proof protection against electric shock.

Type CF (IEC 60601-1) protection against electric shock.

Type CF (IEC 60601-1) defibrillator-proof protection against electric shock.

When displayed in the upper left corner of the screen, indicates that the alarms are silenced. When displayed in the menu or digit fields, indicates that the alarm source has been turned off or alarm does not meet the alarm-specific activation criteria.

A

ÂLA

ESD

warning

symbol

for electrostatic

sensitive devices. Pins of connectors

identified

with

the ESD

warning

symbol should

not be touched. Connections should

not be

made to these connectors unless ESD precautionary procedures are used. For details, see “ESD precautionary procedures” in the User's Reference Manual.

(4)

Symbol

for non-ionizing electromagnetic radiation. Interference may occur in the

 

vicinity

of equipment marked with this symbol.

Document no. M1027807

1.1.4Other symbols

<

о SN, S/N

B

eo@@

|

+)

©)

==

LE }{ }

Introduction

Equipotentiality. Monitor can be connected to potential equalization conductor.

Alternating current

Fuse.

Replace the fuse only with one of the same type and rating.

Serial

Number

Connector for color display

Signal/power output

Signal/power input

Signal/power input/output

Connector for defibrillator synchronization

Connectorforthe 5/5 Device Interfacing Solution, DIS

Power

input

Signal

input

Power

input

Submenu. Selecting an alternative marked with this symbol in a menu opens a new

menu.

The monitor is connected to the iNetwork (LAN).

7

Document no. M1027807

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

IPX class:

Data Card (green) and/or Menu Card (white) is inserted.

A blinking heart next to the heart rate or pulse rate value indicates the beats detected.

A lung next to the respiration rate

value indicates that respiration rate is calculated

from the impedance respiration

measurement.

Gas inlet

Gas outlet

Do not reuse.

Use by. Indicates the last use day.

Do not immerse the sensor in liquids.

Degree of protection against harmful ingress of water as detailed in the IEC 60529:

IPXO

Al

Z

-Ordinary equipment

Date

of manufacture

 

This

symbol

indicates that the waste

of electrical and electronic equipment must not

be disposed as unsorted municipal

waste and must be collected separately. Please

contact an authorized representative of the manufacturerfor information

concerning

the decommissioning of

your equipment.

Ethernet connector

Document no. M1027807

Introduction

1.2Safety

The following list contains general warnings and cautions you should know before installing, maintaining or servicing the system. Warnings and cautions specific to the use of the system can be found in the User's Guide and User's Reference Manual.

1.2.1 Safety precautions

Warnings

 

 

 

 

 

 

 

 

 

 

 

WARNING

A

WARNING

indicates a situation in which the user or the patient may be

in

 

danger of injury or death.

 

 

 

 

 

 

Power connection

 

 

 

 

 

 

 

e

Use

only

hospital-grade

grounded power outlets

and power

cord. Do

not remove the

grounding

 

 

pin

from

the

power

plug.

 

 

 

 

 

 

e

Use

only

an

intact

power

cord. Replace the power

cord if it is

cracked,

frayed, broken

or

 

otherwise

damaged.

s

Do not apply tension to the power cord otherwise the cord may get damaged.

«e

Donotuse

extension cords or adapters of any type.

eBefore starting to use the system, ensure that the whole combination complies with the international standard IEC 60601-1-1 and with the requirements of the local authorities.

Installation

 

 

 

 

s

Keep

the monitor

horizontal

when the Compact

Airway Module is

used. Tilting the monitor may

 

Cause

erroneous

results in

the Compact Airway

Module’s readings

and damage the module.

e The monitor or its components should not be used adjacentto or stacked with other equipment.

 

If adjacent

or stacked

use

is

necessary, the monitor and its

components should

be observed

to

 

verify

normal

operation

in

the

configuration

in which

it will

be used.

 

 

 

 

s

Pins

of

connectors

 

identified

with the ESD warning symbol

should not be touched. Connections

 

should not be made to these connectors unless ESD precautionary

procedures

are used. For

 

details,

see

section

"ESD

precautionary

procedures.”

 

 

 

 

 

 

e

After

transferring

or

reinstallingthe monitor,

always

check

that it

is

properly

connected and

all

 

parts

are

securely

attached.

Pay special

attention to

this in

case

of

stacked

mounting.

 

Laser radiation

e When using the ARK Barcode Reader, N-SCAN, do not stare into the beam. The N-SCAN is a Class 2 laser product.

External connection

© Do notconnect any external devices to the monitor other than those specified.

Fuse replacement

e Replace a fuse only with one of the same type and rating.

Explosion hazard

e Do not use the monitor in the presence of flammable anesthetics.

9

Document no. M1027807

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

Patient safety

Do not perform any testing or maintenance on the monitor while it is being used on a patient.

PACEMAKER PATIENTS: The impedance respiration

measurement

may

Cause

rate

changes in

Minute Ventilation Rate Responsive Pacemakers. In

this case,

set

the

pacemaker

rate

responsive

mode

off or turn

the

monitor

impedance

respiration

measurement

off.

 

Never install

the

monitor or

the

displays

so that they

are above

the

patient.

 

 

The monitor must not be used without manufacturer approved mounting attached.

Cleaning

and service

 

 

Only trained personnel with proper tools and test equipment should

perform

the tests and

repairs

described in this manual. Unauthorized service may void the

monitor

warranty.

Turn the power off and unplug the power cord before cleaning or service. Completely remove any moisture before reconnecting the power cord to the mains outlet.

Do

not

touch any exposed wire or conductive

surface

while

any

cover

is removed

and the

monitor

is energized. The voltages present

can

 

cause

injury

or

death.

 

 

 

Pins

of

connectors identified with the ESD

warning symbol should not be touched.

Connections

should

not be

made

to these

connectors

unless

ESD

precautionary

procedures

are

used. For

details,

see 1.2.2. ESD precautionary procedures.

 

 

 

 

 

 

Always

perform

an

electrical

safety check

and

a

leakage current test

on the monitor

after

service.

 

 

 

 

 

 

 

 

 

 

 

 

 

Accessories

Use only accessories, including mounts and batteries, and defibrillator-proof cables and

invasive pressure transducers approved by GE Healthcare. For a list of approved supplies and accessories, see the “Supplies and Accessories” catalog delivered with the monitor. Other

cables, batteries, transducers

and accessories may cause a safety hazard, damage the

equipment or

the

system,

result

in increased

emissions

or decreased

immunity of

the

equipment or

system

or

interfere

with the

measurement.

Protection against cardiac

defibrillator

discharge

is

due

in part

to

the

accessories

for

pulse oximetry (SpO2),

temperature

(T) and

invasive

pressure

(P)

measurement.

 

 

 

 

 

 

Single

use

accessories are not designed to

be reused.

Reuse may

cause a risk

of

contamination

 

ane

affect

tha

Pasas rarn ant

Aair o,

 

 

 

 

 

 

απ

ATPULL

UNO

MILUOJUTUTIITUTTTUT

 

 

 

 

 

 

Cautions

 

 

 

 

 

 

 

 

 

 

 

CAUTION

A CAUTION

indicates a condition that may

lead to

equipment

damage

or

malfunction.

installation

Leave space for air circulation to prevent the monitor from overheating.

Ensure that the module is properly orientated (i.e. module release latch facing downward) before insertion.

Before connecting the power cord to the mains outlet, check that the local voltage and

frequency correspond with the rating stated on the device plate on the rear panel of the monitor. See instructions for different displays in section “Displays”.

Turn off the power before making any rear panel connections.

10

Document no. M1027807

Introduction

Before use

eAllow two minutes for warm-up and note any error messages or deviations from normal operation.

e

Cleanthe rear

panel

fan dust

filters once

a

month or

whenever necessary.

s

Donotconnect

a

sampling

line to

the

female Patient

Spirometry connector while the other end

 

of the

sampling

line

is connected

to the

D-fend water

trap. The pressure in the gas sampling

 

system

may cause

damage

to the

PVX

unit

pressure

transducers.

Autoclaving and sterilizing

e Do not autoclave any part of the monitor.

e Do not gas sterilize the modules.

Cleaning and service

e Donotusehypoclorite, ammonia-, phenol-, or acetone based cleaners. These cleaners may damage the monitor surface.

eDonotimmerseanypartof the device in any liquid, or allow liquid to enter the monitor or modules.

©

Donotapplypressurizedairto

any outlet or

tubing

connected to

monitor.

 

 

 

e

Electrostatic

discharge

through

the PC boards may damage the

components.

Before handling

 

PC boards, wear a static

control

wrist

strap.

Handle

all

PC boards

by their

non-conductive

edges

 

and use anti-static containers

when

transporting them.

 

 

 

 

e

Do not break or bypass the patient isolation barrier when testing PC boards.

 

 

®

Donotclean

the spirometry tubes with high

pressure

air 02 flushing while

the

spirometry

tubes

are connected to Patient Spirometry connector. High differential pressure may damage PVX unit pressure transducers.

Special components

e Special components are used in these monitors that are vital to assure reliability and safety. GE Healthcare assumes no responsibility for damage, if replacement components not approved by GE Healthcare are used.

 

A lithi

С

an

atala

ar

AA

»

Ρζο

 

 

 

 

 

 

 

 

O

 

 

 

 

 

 

 

local

regulations.

 

 

 

 

 

 

 

 

Batteries

 

 

 

 

 

 

 

 

 

 

The

battery packages

in

the

central

unit, F-CPU

and in the power supply

unit of F-CU8

contain lead

acid

(Pb)

which

is hazardous

to

the

environment

and therefore needs to

be disposed

of carefully

according to local regulations.

To replace the batteries safely, please refer to the service instructions in this manual.

e

Do

notshort-circuit the battery terminals, this

may

produce

a very

high

current,

which

will

 

damage

the battery.

 

 

 

 

 

 

 

©

Do

not

dispose of the battery into open flame,

nor

put the

battery

near

fire, as

it may

explode.

11

Document no. M1027807

Datex-Ohmeda S/5 Anesthesia

and Critical

Care

Monitors

 

 

 

 

 

 

 

s

Donotdismantle

the

battery.

It contains

electrolyte, which may damage clothing

or

cause

injury

 

to skin

or

eyes. If

exposed

to

electrolyte,

wash the injured area with

plenty of

clean

water

and

 

contact

a

doctor.

 

 

 

 

 

 

 

 

 

 

 

See

also

section “Symbols”.

 

 

 

 

 

 

 

 

 

Storage

and

transport

 

 

 

 

 

 

 

 

 

Do

not store or

transport the monitor outside

the

specified temperature,

pressure

and

humidity

 

ranges:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temperature

 

 

 

 

 

-10...+50 °C/14...122 °F

 

 

 

 

 

Atmospheric

pressure

 

 

 

660...1060 hPa/500...800 mmHg/660...1060

mbar

 

Relative

humidity

 

 

 

10...90%

non-condensing

 

 

 

 

 

For display specific environmental requirements see specifications in the “Display” slot.

1.2.2 ESD precautionary procedures

e To avoid electrostatic charges to build up, it is recommended to store, maintain and use the

 

equipment

at

a relative humidity

of 30% or greater.

Floors

should

be

covered

by

ESD

dissipative

 

carpets

or

similar. Non-synthetic clothing should be used

when

working

with

the

component.

e

To prevent

applying a possible electrostatic discharge

to the ESD sensitive parts

of

the

 

 

equipment,

one should touch the metallic frame of the

component or to

a large

metal

object

 

located

close

to the

equipment.

When

working with

the

equipment

and

specifically

when the

 

ESD sensitive

parts

of the equipment

may be touched,

a

grounded

wrist

strap

intended

for use

with ESD sensitive equipment should be worn. Refer to the documentation provided with the wrist straps for details of proper use.

ESD precautionary

procedure

training

 

 

 

 

 

 

 

 

 

 

 

 

 

It is

recommended

that all potential users receive an

explanation

of

the

ESD warning symbol

and

 

training in

ESD

precautionary

procedures.

 

 

 

 

 

 

 

 

 

 

 

 

The

minimum

contents of an

ESD precautionary

procedure

training

should

include an

introduction

to

the physics of electrostatic charge, the voltage levels

that can

occur

in

normal

practice

and

the

 

damage that can be done to electronic components

if they

are

touched

by

an

operator

who

is

 

electrostatically charged. Further, an explanation should be

given

of

methodsto

prevent

build-up of

 

electrostatic charge and how and why to discharge one’s body to

earth or to

the frame

of the

 

 

equipment

or

bond

oneself by means of a wrist

strap

to the

equipment or the

earth priorto making

a

connection.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.2.3Disposal

Dispose of the whole device and parts of it according to local environmental and waste disposal regulations.

12

Document no. M1027807

System description

2System description

2.1Introduction

Datex-Ohmeda monitors build up a freely configurable modular system. The architecture is designed to enable different module combinations so that the user is able to get the desirable parameter and feature set. This modular approach makes it possible to add new features when they are needed.

2.2Bus structure

The operation

of

Datex-Ohmeda monitors

is based

on

two communication

channels,

the

CPU

bus and

module

bus.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the 5-Module

Central

Unit,

PC

boards

receive

power from the

F-CPU

power

supply

and

the

 

parameter modules

receive

power

from a

separate power supply in the 5-Module

Frame

unit. These

power supplies

 

are

both

fed

by the N-AC

Power

Unit.

In the 8-Module

Central

Unit,

F-CU8, all

PC

boards connected to the CPU

bus, as well as the parameter modules attached to

the

module

bus

receive

power

from

the

same

power supply, which

is

an integral

part of the

Central

Unit,

F-CU8.

 

UP!

 

 

 

 

CPU

 

Display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Controller

 

 

 

 

 

 

 

 

 

 

Module

Bus

 

Parameter

Parameter

 

Parameter

Module

Module

000

Module

Figure1-—Generalstructure-ofbusS/5 system

The

CPU

bus

is a

communication

channel

used

only

for

internal data

transfer.

Itis based on

the

 

ISA

bus

used

in

IBM

PC

computers.

Data

is transferred

on this 16

bit

wide

bus using

the CPU

clock

 

 

frequency.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The

module

bus

is used to connect the parameter modules to

the Central Unit. The bus is based

on

the

industry

standard RS-485, which

usesa

differential

serial

method

to

transfer

data. This

type

of

bus

is

robust

and

it allows

parameter

modules

to be inserted

or

removed

while the power

is

on.

 

The

module

 

bus

uses

a

500 kbps data transfer rate

and

can

be used

for longer distances

than

the

CPU

bus,

e.g.

for

external

frame

connections.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The

RS-485

type

of

serial communication

supports

so-called

multidrop or party line connections. This

means

that

all parameter modules connected to the module

bus

use

exactly

the

same

lines

for

 

 

communication.

The

advantage

of this is that

all bus

connectors

are identical

and

the modules

can be

connected in any

order and

position.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

13

Document no. M1027807

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

2.3Distributed processing

Asystem

assembled from

S/5

products

is a multiprocessor

system. All

parameter modules

have their

Own microprocessor, which performs functions such as module key control, waveform filtering,

parameter

related computing

and

pneumatic

control,

etc.

At the same

time the main CPU

performs

higher level tasks such as

trending

and

alarm

control.

While the parameter modules and CPU are

performing

their tasks, the UPI (Universal Peripheral Interface) microprocessor handles all functions

needed to

transfer data between the parameter modules and the CPU. At the same time the Display

controller

microprocessor

performs

pixel

calculations

for graphics.

 

 

Module

[ Dual-port

|

Er

RAM

Il

A

Display

Controller Board

Display

Display

memo

 

a

=

Qs

 

 

 

Command

Board/Bar

Figure 2

Distributed processing in S/5 system

 

 

This kind of parallel processing gives one major advantage

to

centralized processing. When new

parameter modules or PC boards are added

to the system,

the

processing power is increased. As a

result, the

system does not slow down when

new features

are

added.

2.4Module communication

The communication master controlling data transfers between the CPU and parameter modules is

called

UPI processor.

It sends data

to each

connected

module

100 times

a second.

Modules

respond

to

each

data request

immediately

by sending a data package,

whose

length

depends

on

the

type

of

the

module.

This communication

protocol

ensures that each module receives and

sends

data

every

10

ms.

If a

module does not respond to data requests,

the UPI

processor

presumes that

the

module

is

disconnected.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Parameter modules may hold a static (fixed) or dynamic address, which

the

UPI processor

uses when

sending out data. Two parameter

modules of the same type must not

be

fitted onto

the

same

monitor

since they

might reply

to a data request simultaneously, thus

causing

communication

errors.

 

 

14

Document no. M1027807

System description

UPI Board

Dual-port

RAM

E prt

a Е

Microcontroller

[=

BES

Figure 3

Principle of

UPI

section

operation

 

 

 

 

 

 

The UPI processor collects and

stores

all

data that is received from

the

parameter

modules

into

a dual

port RAM,

which is mapped

directly to

the address space of the main CPU. The main

CPU

reads

data

from the memory while the UPI

processor guarantees that the data

is up to date. This operation

also

works in the other direction. In

this the

main CPU fills the dual port

RAM

with data

and

the

UPI

 

processor

distributes it to the

parameter

modules.

 

 

 

 

 

 

2.5Software loading

The program memory on CPU board is loaded with monitor software at the factory. The software is

used for running all the functions that are procedures the CPU board is fitted with a loaded.

integrated into the PC board. For service and upgrade PCMCIA card drive through which new software can be

iService!

 

 

| Software

 

 

 

 

 

 

 

 

I

 

 

 

JPCMCIA!

 

UPI

CPU

~

>

Display

Board

Board

 

 

Controller

 

 

 

 

Board

 

 

Flash

 

 

Program

|

Program

 

Program

Memory

 

Memory

 

Memory

CPU Bus

Figure 4 Software loading

15

Document no. M1027807

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

2.6Parameter modules

Module E::.

keys

Patient

electronics

E:

eee

000000],

 

Peripheral

È

 

 

 

drivers

Е

 

 

 

 

 

 

Bus

Module

Figure 5

General structure of parameter modules with patient isolation

 

 

 

 

The

detailed

structure

of a

parameter

module

depends

on the

specific needs for

each

individual

 

parameter.

However,

some

common

parts

are

used

in the parameter modules. The electronics

inside

the

module

is usually

divided

into isolated

(floating)

and non-isolated sections. Typically, the

 

non-isolated

section

consists of buffers to interface

the parameter module to the

module

bus while

the rest of the electronics

is

located

in the

isolated

section. The

isolated

section

includes

the

 

microcontroller together with memory components,

the

front-end analog

electronics

(amplifiers,

etc.)

and

sensor

drives.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16

Document no. M1027807

System installation

3 System installation

3.1 Unpacking instructions

1.

Confirm that the packing box

is

undamaged.

If the box is damaged,

contact

the

shipper.

2.

Open the

top

of

the box and

carefully unpack

all components.

 

 

 

 

3.

Confirm

that

all

components

are

undamaged.

If any

of the components

are

damaged, contact

 

the shipper.

 

 

 

 

 

 

 

 

 

 

4.

Confirm

that

all

components

are

included. If any of

the components

are

missing,

contact your

 

GE Healthcare

distributor.

 

 

 

 

 

 

 

 

3.2Choosing location

Consider the following aspects:

e

lighting

s

space

e

connections

selectromagnetic and radio frequency interference, see Appendix 8. ElectroMagnetic Compatibility

 

e

environment

 

 

 

 

 

 

 

WARNING

The

monitor or

its

components

should not be

used adjacent to

or

stacked

with

 

other equipment.

If adjacent

or stacked use

is necessary, the

monitor and

its

 

components should be observed to verify normal operation in

the

configuration in

 

which it will be

used.

 

 

 

 

 

3.3Central Unit; S/5 8-Module Frame, F-CU8

Figure 6 Central Unit: S/5 8-Module Frame, F-CU8

The front of the Central

Unit, F-CU8, houses

plug-in

parameter

modules which interface the

system

with external devices.

The back of the

Central

Unit

houses the

system

circuit

boards,

together

with

expansion

slots which

accommodate

PC boards which interface with

external

devices

and

the

Datex

Ohmeda

Network.

 

 

 

 

 

 

 

 

 

 

17

M1027807

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

3.3.1Connecting to mains

Connect the power cord to the mains power inlet on the rear of the Central Unit and to the wall socket.

WARNING

The

power cord

may

only

be

connected to a

three-wire,

grounded, hospital grade

 

 

receptacle.

 

 

 

 

 

 

 

 

 

 

3.3.2

Connecting to

Datex

Ohmeda

Network

 

 

 

 

 

 

 

 

To connect

the monitorto

the

Datex Ohmeda

Network,

make sure

a

Network Board

B-UPI4NET

is

 

 

installed.

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

the Monitor-Network

cable to connect the monitor the network as follows:

 

 

 

 

1.

Make

sure that the

power to

the Central

Unit is turned off.

 

 

 

 

 

 

2.

Connect the RJ-45

connector

and the Identification Plug to

the

corresponding

connectors

on

 

 

 

the B-UPI4NET.

 

 

 

 

 

 

 

 

 

Connect the other RJ-45 connector the corresponding connector on the wall box.

Turn on the power to the Central Unit. Enter the Network service menu:

Monitor Setup - Install/Service (password 16-4-34) -

Service (password 26-23-8) -

Frame - Network

Make sure that the

monitor's

network communication has

been

set according to the used

network software:

 

 

 

 

 

 

 

Network software S-CNETO1 -> DRI

Level

=

2001

 

 

Network software

S-CNETO2 -> DRI

Level

=

2001

 

 

Network software L-NETO3 -> DRI Level = 2003

 

 

Network software

LNETO5 ->

DRI

Level

= 2005

 

 

If necessary, change the monitor's

network

communication

by

selecting DRI Level and turning

the ComWheel.

 

 

 

 

 

 

 

NOTE: If the DRI level is changed, the monitor will restart automatically.

Confirm that the network symbol and ‘Connected to Network’ message are displayed on the upper part of the screen.

3.3.3 Inserting the parameter modules

1. Ensure that the module is properly orientated (i.e. module release latch facing downward). 2. Align the module insertion guide slot with the insertion guide.

3. Push the module into the frame until it clicks.

18

System installation

Figure 7

Module insert

NOTE: The Compact Airway Modules cannot be placed into the third and fourth slot from the right- hand side of the F-CU8 Central Unit.

NOTE:

Use

only one Extension

Frame F-EXT4 in one F-CU8 Central Unit.

NOTE:

Do

not use two or more

parameter modules with identical functions in the monitor system.

Take special care not to do this when using the Extension Frame, F-EXT4.

Parameter modules or boards with identical functions are:

e

Hemodynamic

multiparameter

modules,

E-PRESTN, E-PRETN, E-RESTN, M-PRESTN, M-PRETN,

 

M-RESTN, M-ESTPR, M-ETPR, M-ESTR, M-NESTPR, M-NETPR, M-NESTR, M-NE12STPR, M-

 

NE12TPR, M-NE12STR

 

 

 

e

Dual Pressure Modules, E-PP/M-PP

 

 

e

Pressure Modules, E-P/ M-P, Pressure Temp Module, E-PT/ M-PT

e

Cardiac Output

Modules, E-COP, E-COPSv, M-COP

and M-COPSv

s

NIBP Modules, M-NIBP and hemodynamic modules w/ N measurement

s —

Airway Modules, E-CO, E-COV, E-COVX,

E-CAIOVX,

E-CAiOV, E-CAIiO, E-miniC, M-C, M-CO, M-

 

COV, M-CAiO, M-CAiOV, M-CAiOVX, G-O, G-OV, G-AO, G-Ai0, G-AOV, G-AiOV, M-MiniC

e

Interface Module, E-INT / M-INT and Interface Board, B-INT

e

Oxygen Saturation Modules, E-NSAT/ M-NSAT and M-OSAT

e

Tonometry Modules, E-TONO/ M-TONO

 

 

e

NeuroMuscular

Transmission

Modules,

E-NMT/ M-NMT

e

EEG

Modules,

E-EEG/ M-EEG

 

 

 

e

BIS

Modules,

E-BIS/ M-BIS

 

 

 

19

M1027807

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

3.3.4 Positioning of PC boards

©

©

o EEBo

“ER

-|

+

0

| j ceo

(7)

~

im

Figure 8 Rear view and positioning, F-CU8 rev. 10

(1)Mains power inlet

(2) Battery 24 Vde input connector (F-CU8 rev. 10)/ Display power outlet (F-CU8 rev. 03 to 09)

(3)UPINET Board, B-UPI4NET

(4)

CPU Board, B-CPU5

 

( 5)

Display Controller Board, B-DISPX

(primary/screen 1)

( 6) )

Display Controller Board, B-DISPX (secondary/screen 2),

 

Keyboard Interface Board, B-ARK

 

(7)

Interface Board, B-INT,

 

 

Gas Interface Board, B-GAS,

 

 

Display Controller Board, B-DISPX

(3 screen)

(8)Service reset button

(9)Equipotential connector

(10)Back plate lid for CPU Board, B-CPU5

NOTE: Authorized personnel only may open the cover of 24Vdc input connector (use aflat screwdriver).

3.3.5Replacing PC Boards

For service procedures refer to Part Il.

1. Make sure that the power is turned off the monitor and unplug the power cord. Press and hold the service reset button on the rear panel for at least five seconds or until an audible tone is heard (see Figure 9).

20

Loading...
+ 494 hidden pages