This Product will perform in conformity with the description thereof contained in this operating
manual and accompanying labels and/or inserts, when assembled, operated, maintained and
repaired in accordance with the instructions provided. This Product must be checked periodically.
A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted or
contaminated should be replaced immediately. Should such repair or replacement become
necessary, Datex-Ohmeda recommends that a telephonic or written request for service advice be
made to the nearest Datex-Ohmeda Field Service Support Center. This Product or any of its parts
should not be repaired other than in accordance with written instructions provided by DatexOhmeda and by Datex-Ohmeda trained personnel. The Product must not be altered without the
prior written approval of Datex-Ohmeda’s Quality Assurance Department. The user of this Product
shall have the sole responsibility for any malfunction which results from improper use, faulty
maintenance, improper repair, damage, or alteration by anyone other than Datex-Ohmeda.
CAUTIONCAUTION
w
CAUTIONU. S. Federal and Canadian law restrict this device to sale by or on the order of a
CAUTIONCAUTION
licensed medical practitioner. Outside the U. S. A. and Canada, check local laws for
any restrictions that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product
group code, the year of manufacture and a sequential unit number for identification.
AAAA12345
This alpha character indicates the year of product manufacture
and when the serial number was assigned; “Y” = 1995, “Z” = 1996, “A” =
1997, etc. “I” and “O” are not used.
1001-0809-000 11/20/99
Table of Contents
1/Introduction
Introduction to the Excel MRI Compatible Anesthesia System1-1
How to use this manual1-2
Symbols used in the manual or on the equipment 1-3
2/General Information
Identifying Excel MRI components and controls2-1
Identifying 5125 O
3/Set Up
Before starting to setup the system3-1
Mounting gas cylinders3-1
Mounting the 5125 O
Circuit and monitoring connections3-4
4/Preoperative Checkout
Before starting the checkout4-1
MRI compatibility check4-1
Initial Checks4-2
Checking vaporizer mounting4-3
Checking cylinder and pipeline supplies4-3
Checking vaporizer back pressure4-4
Leak checking the low pressure circuitry4-4
Checking the gas flow controls4-6
Monitor components and controls2-6
2
Monitor3-2
2
5/Maintenance
i
Breathing system checks4-7
5125 O
Maintenance schedule5-1
Cleaning and sterilization5-2
O
Monitor checks4-7
2
Cleaning5-2
Sterilization5-3
Special precautions for rubber articles5-3
sensor maintenance5-3
2
Installing a cartridge or disassembling the O
Cleaning and sterilization5-6
Replacing the 5125 O
Monitor battery5-7
2
1001-0809-00011/20/99
sensor for cleaning5-3
2
i
Table of Contents
6/Troubleshooting
Repair policy6-1
7/Illustrated Parts
Appendix
Warranty
Problems with the 5125 O
Calibration and drift6-1
5125 O
Pneumatics problems6-3
Excel specific parts7-1
MRI Compatible accessories7-1
5125 O
Where to find additional part numbers7-2
Excel PneumaticsA-1
Excel MRI System SpecificationsA-4
Monitor alarms6-2
2
Monitor accessories7-2
2
Monitor6-1
2
ii
1001-0809-00011/20/99
ii
1/Introduction
WARNINGWARNING
w
WARNINGUsing an MRI system in conjunction with general anesthesia on cardiac patients,
WARNINGWARNING
febrile patients, and patients with impaired ability to perspire may present a patient
health hazard.
In this sectionIn this section
In this section
In this sectionIn this section
Introduction to the Excel MRI Compatible Anesthesia System1-1
How to use this manual1-2
Symbols used in the manual or on the equipment1-3
Introduction to the Excel MRI Compatible Anesthesia SystemIntroduction to the Excel MRI Compatible Anesthesia System
Introduction to the Excel MRI Compatible Anesthesia System
Introduction to the Excel MRI Compatible Anesthesia SystemIntroduction to the Excel MRI Compatible Anesthesia System
The Datex-Ohmeda commitment to meeting your anesthesia needs continues with the Excel
Magnetic Resonance Imaging (MRI) Compatible Excel Anesthesia System.
Basic features
The Excel MRI System includes:
• Three gases, oxygen, nitrous oxide, and air. All gases have pipeline and cylinder connections.
• The 5125 MRI Compatible O
battery and hardware self tests.
• A two vaporizer manifold.
MRI Compatibility
The Excel MRI is constructed primarily of non-ferous materials to help prevent attraction to the
cryogenic magnets in MRI systems.
The Excel MRI and attached accessories performed to specifications when tested together as a
system directly next to an unshielded, 1.5 Tesla MRI unit with a magnetic fringe field below 0.23
Tesla (2300 Gauss).
Approved accessories
A wide variety of Datex-Ohmeda accessories have been approved for use in conjunction with the
Excel MRI under test conditions:
WARNINGWARNING
w
WARNINGThe MRI compatibility of these accessories applies to specific accessory models and
WARNINGWARNING
is limited to use as part of the Excel MRI System. None of these accessories have
been tested for stand alone use in an MRI environment or in magnetic fringe fields
above 0.23 Tesla (2300 Gauss).
Monitor with adjustable high and low O2 alarms, and built in
2
1-1
• Tec 4 or 5 vaporizers; keyed or funnel fill models.
• GMS Absorber, PEEP valve, and Bain Circuit adapter
• Waste Gas Scavenging Valve
• 121 Respirometer
• Standard or free flow suction regulator kits
• Auxiliary oxygen flowmeter
• Flip-up shelf
1001-0809-00011/20/99
1-1
1/Introduction
• Storage cabinets
• Dovetail mounted accessories: a 1 x 3.5 inch post with two inch extension; a 12 inch IV Pole; a
large case Tycos gauge; and holders and regulators for E-size O
Because approval applies only to specific accessory models, consult the Illustrated Parts Section
of this manual for Stock Numbers.
Safety features
The Excel MRI includes several important safety features:
• The Datex-Ohmeda Link 25 Proportion Limiting Control System limits the lowest oxygen
concentration that can be delivered on the Excel MRI to a nominal 25% for O
the common gas outlet.
and N2O cylinders.
2
2/N2
O mixtures at
• An audible low O
(30 psig).
•N
O and air flows stop if O2 supply pressure falls below a nominal 138 kPa (20 psig).
2
• An interlock mechanism that helps prevent more than one Datex-Ohmeda Tec 4 or Tec 5
vaporizer from being “On” at once. An isolation system also helps prevent fresh gas from
entering a vaporizer that is not switched “On.”
• Low battery alarms and built in self tests on the 5125 O
• Differentiated pipeline and cylinder gauges.
• Gas specific pipeline and cylinder connections.
• A guarded O
• Color coded flow controls and pressure gauges.
supply alarm alerts you if the O2 pressure falls below a nominal 207 kPa
2
Monitor.
2
flush button.
2
Operator convenience
For your convenience, the Excel MRI also features:
• An O
• Full length dovetail accessory mounting.
• A single action brake/footrest.
• Shelf space for additional monitoring.
power outlet
2
• Large, easy-running casters and a compact frame.
How to use this manualHow to use this manual
How to use this manual
How to use this manualHow to use this manual
This manual covers the Excel MRI System and the MRI Compatible 5125 O2 Monitor. Vaporizers,
the GMS Absorber, the Waste Gas Scavenging Valve and other major accessories also have
individual operation and maintenance manuals. A complete set of operation and maintenance
manuals comes with the Excel MRI System. Datex-Ohmeda recommends that you keep this
manual and all related manuals available for reference.
If you are using the system for the first time, read this manual first. Then read the operation and
maintenance manuals for the system components. Before you go on to Chapter 2, “General
Information,” make sure that you understand the symbols listed at the end of this chapter.
If you have used the Excel MRI before, complete the step-by-step set up procedure in Chapter 3.
Then turn to Chapter 4 “Preoperative Checkout,” which is required before using the Excel MRI.
1-2
1001-0809-00011/20/99
1-2
1/Introduction
Before cleaning the Excel MRI, read Chapter 5, “Maintenance,” carefully. This section also tells
you how to change the O
cartridge replacement, disassembly for cleaning).
Chapter 6, “Troubleshooting,” helps you solve problems that may occur with the Excel MRI or the
5125 O
you can replace or an internal fault that requires a service call? (If the leak follows a particular
vaporizer and the external o-ring is in place, replace the vaporizer.)
Chapter 7, “Illustrated Parts,” tells you how to order replacement parts and accessories. It also
lists part numbers for all operation and maintenance manuals associated with the Excel MRI
System.
Monitor. For example, how can you tell if a low pressure leak is due to a vaporizer that
2
Monitor battery and assemble and service the O2 sensor (sensor
2
The Appendix lists Excel MRI and 5125 O
pneumatic circuitry of the Excel MRI.
Please also take a moment to review the User Responsibility Statement on the inside of the front
cover; it describes what is expected of you to maintain the Excel MRI. Also read the Warranty on
the back cover; it outlines Datex-Ohmeda’s responsibility in case of a functional defect.
Symbols used in the manual or on the equipmentSymbols used in the manual or on the equipment
Symbols used in the manual or on the equipment
Symbols used in the manual or on the equipmentSymbols used in the manual or on the equipment
WarningsWarnings
w
Warnings and w
WarningsWarnings
obey all of the instructions in this manual.
Warnings tell you about a condition that can cause injury to the operator or the patient. Cautions
tell you about a condition that can cause damage to the equipment. Read and obey all warnings
and cautions.
Other symbols replace words on the equipment or in Datex-Ohmeda manuals. No one device or
manual uses all of the symbols. These symbols include:
ø
O
Cautions Cautions
Cautions tell you about dangerous conditions that can occur if you do not
Cautions Cautions
On (power)
Off (power)
Monitor specifications and describes the internal
2
1-3
o
q
p
œ
†
∏
x
Standby
Standby or preparatory state for a part of the equipment
Systems with this mark agree with European Council Directive (93/42/EEC) for
Medical Devices when they are used as specified in their Operation and
Maintenance Manuals. The xxxx is the certification number of the Notified Body
used by Datex-Ohmeda’s Quality Systems.
The primary regulator is set to pressure less than 345 kPa
The primary regulator is set to pressure less than 414 kPa
European Union Representative
Low O
alarm limit switch
2
High O
alarm limit switch
2
Battery test
Circuit test
Test switch
1-6
1001-0809-00011/20/99
1-6
2/General Information
In this sectionIn this section
In this section
In this sectionIn this section
Identifying Excel MRI components and controls2-1
Identifying 5125 O
WARNINGWARNING
w
WARNINGThe Datex-Ohmeda Excel System must only be used with non-flammable anesthetic
WARNINGWARNING
Monitor components and controls2-6
2
agents.
Identifying Excel MRI components and controlsIdentifying Excel MRI components and controls
Identifying Excel MRI components and controls
Identifying Excel MRI components and controlsIdentifying Excel MRI components and controls
The figures in this chapter, show control locations on Excel MRI System components. For more
information on GMS Absorber, or vaporizer controls, refer to the individual operation and maintenance manual.
9
8
7
6
1.System Master Switch
2.Flush Button
3.Common Gas Outlet
4.Drawer Cabinet (Optional Accessory)
5.Brake Bar and Footrest
6.Absorber Arm Button
7.O2 Power Outlet
8.Flow Controls
9.5125 O2 Monitor
1
2
3
4
AA.19.039
5
Figure 2-1
Front view of Excel MRI
2-11001-0809-00011/20/99
2-1
2/General Information
Flow controls
With the system master switch set to “On,” turning a flow control counterclockwise increases flow,
turning a flow control clockwise decreases flow. A linkage between the N
controls limits the lowest oxygen concentration that can be set on the Excel MRI to a nominal 25%
for O
up to two additional turns of the knob); N
knob ); air, 0 to 15 l/min (plus up to one additional turn of the knob).
System master switch
Setting the system master switch to “On” allows gas to flow through the Excel MRI.
Flush button
Pressing the O2 Flush button delivers 45-70 l/min of O2 through the common gas outlet.
Common gas outlet
The common gas outlet delivers anesthetic gases to the patient circuit.
Brake
Push down to keep the Excel MRI from rolling. Lift up to release.
O mixtures at the common gas outlet. Flow ranges are: O2, 200 ml/min to 10 l/min (plus
2/N2
O and the O2 flow
2
O, 0 to 10 l/min (plus up to one additional turn of the
2
Absorber arm button
Press this button to adjust the height of the absorber arm.
O2 power outlet
Provides drive gas for MRI compatible pneumatic equipment.
2-21001-0809-00011/20/99
2-2
2/General Information
1
2
1.Cylinder Connection
2.Pipeline Connection
AA.19.040
Figure 2-2
Excel MRI (rear)
2-31001-0809-00011/20/99
2-3
2/General Information
Vaporizer concentration control
Press in the release and turn the concentration knob to the desired agent concentration. The
locking lever must be fully clockwise (locked position) or the control will not turn.
Concentration control release
Pressing the release allows you to turn the concentration knob on the vaporizer.
Vaporizer locking lever
When the locking lever is completely counterclockwise, and the vaporizer concentration is set to “Off,”
you can mount or remove the vaporizer. Turn the lever clockwise to lock the vaporizer in position.
Vaporizer fill port valve
Opens the fill port valve so that agent can be added or drained.
Vaporizer key lock lever
Holds the fill adapter from the agent bottle in place. To secure the adapter, lower the lever, insert
the adapter, and raise the lever.
Vaporizer fill port (spout)
Unscrew and remove the plug to fill or drain vaporizers with funnel fill ports.
Drain Plug
Loosen (do not remove) with hex head portion of spout to drain agent. Tighten drain plug and
install spout before use.
8
1
7
AA.13.004
6
45
1.Vaporizer Concentration Control
2.Vaporizer Fill Port (Spout)
3.Drain Plug and Drain Port
4.Vaporizer Fill Port Valve
5.Vaporizer Fill Port (Keyed)
6.Vaporizer Key Lock Lever
7.Concentration Control Release
8.Vaporizer Locking Lever
AA.13.003
2
3
Figure 2-3
Tec 5 vaporizers
2-41001-0809-00011/20/99
2-4
2/General Information
Absorber drain plug
Unscrew and remove to drain condensate. Replace and tighten before use.
APL valve
This valve limits patient circuit pressure during manual ventilation (bagging).
Absorber check valves
Provide unidirectional flow through the absorber.
Airway pressure gauge
Absorber gauge to display inspiratory pressure.
Absorber release lever
This lever should always be in the lock position during use. The release position lowers the
absorber base so that you can remove the canisters.
Absorber mounting knob
Turn completely clockwise to attach the absorber to the mounting pin. Turn counterclockwise to
remove the absorber.
Bag/Ventilator switch
This switch on the absorber selects either the bag arm (Bag position) or the ventilator bellows
(Vent position).
1.Bag/Ventilator Switch
2.Airway Pressure Gauge
3.APL Valve
4.Absorber Check Valves
5.Fresh Gas Input
6.Absorber Release Lever
7.Absorber Mounting Knob
8.Absorber Drain Plug
2
3
41
AA.15.153
87
5
AA.15.152
6
Figure 2-4
GMS Absorber
2-51001-0809-00011/20/99
2-5
2/General Information
Identifying 5125 OIdentifying 5125 O
Identifying 5125 O
Identifying 5125 OIdentifying 5125 O
Figure 2-5 identifies 5125 O2 Monitor controls and major components.
1
Monitor components and controls Monitor components and controls
Monitor components and controls
Monitor components and controls Monitor components and controls
22
2
22
2
34
5125
5
6
7
8
MD.12.013
MD.12.014
14
10 mV
per
o
%
2
1. Battery
2. Calibration Control
Min
R
L
20M
1213
MD.12.019
3. Test Switch
4. O2 Alarm Limits
5. Display
6. Alarm LEDs
7. Battery Power Switch
8. Alarm Silence
9. Sensor Housing
10. O-Ring
11. Sensor Cartridge
12. O2 Sensor Connection
13. Remote Connection (Refer to Appendix for Signal Specifications)
14. Front Panel
MD.02.032
910119
Figure 2-5
5125 O
Monitor controls
2
2-61001-0809-00011/20/99
2-6
2/General Information
Test switch
Pressing down the top half of the switch (Batt Test) checks the battery condition. Holding down the
bottom half of the switch (Circuit Test) checks the 5125 O
information about using the switch, refer to the section “5125 O
Chapter 4.
Low O2 alarm
Sets the lowest acceptable O2 concentration. O2 concentrations below 18% trigger a low O2 alarm
regardless of the alarm limit. Setting the limit below 18% also triggers an alarm.
High O2 alarm
Sets the highest allowed O2 concentration (0 to 99%). Setting the limit to 00% disables the alarm.
Display
Shows the measured O2 concentration and any alarm messages.
Battery Power switch
Switches the 5125 O2 Monitor “On” and “Off.”
Alarm Silence
Monitor’s electronic circuitry. For
2
Monitor checks” at the end of
2
Pressing the alarm silence button silences all audible alarms, except for low O2, until the next
occurrence. Low O
LEDs stay on continuously.
alarms are silenced for 30 seconds. While the alarm is silenced, any flashing
2
O2 sensor housing and cartridge
The 5125 O2 Monitor uses the standard Datex-Ohmeda O2 sensor. The sensor cartridge is not part
of the O
section “O
sensor assembly and must be installed before use. For additional information refer to the
2
sensor maintenance” in Chapter 5.
2
Battery
The 5125 O2 Monitor uses a “C” size, non-magnetic, 3.9 Vdc lithium battery. During operation, the
“BATT FAIL” alarm continuously monitors battery strength.
Calibration control
Use the calibration control (CAL) to calibrate the 5125 O2 Monitor for 21% and 100% O2. Refer to
“5125 O
Monitor checks” at the end of Chapter 4.
2
2-71001-0809-00011/20/99
2-7
2/General Information
Notes
2-81001-0809-00011/20/99
2-8
3/Setup
In this sectionIn this section
In this section
In this sectionIn this section
Before starting to setup the system3-1
Mounting gas cylinders3-1
Mounting the 5125 O
Circuit and monitoring connections3-4
Before starting to setup the systemBefore starting to setup the system
Before starting to setup the system
Before starting to setup the systemBefore starting to setup the system
WARNINGWARNING
w
WARNINGRemove the Excel MRI from the MRI room before starting the setup procedure. Do not
WARNINGWARNING
add or remove any anesthesia system components while the Excel MRI System is in
the MRI room.
1.Remove the Excel MRI System from the MRI room before starting the setup procedures.
2.Complete the setup procedure outside the MRI environment.
3.Do not move the Excel MRI System into the MRI room until you complete the first three
sections of the checkout procedure (“MRI compatibility check,” “Initial checks,” and “Checking vaporizer mounting”).
Mounting gas cylindersMounting gas cylinders
Mounting gas cylinders
Mounting gas cylindersMounting gas cylinders
WARNINGWARNING
w
WARNINGUse only non-magnetic gas cylinders on the Excel MRI System.
WARNINGWARNING
monitor 3-2
2
1.Set the system master switch to “Off.”
2.Swing out the yoke clamp.
3.Unscrew the tee handle until the screw is flush with the inside of the clamp.
4.Make sure that the cylinder valve is not covered by a dust cap. Remove dust cap if present.
5.Remove the old cylinder gasket if present and install a new gasket.
Note: Make sure that the old gasket does not stick to the cylinder.
CAUTIONCAUTION
w
CAUTIONUse one cylinder gasket per yoke. Extra gaskets may cause a leak. Not using a
CAUTIONCAUTION
cylinder gasket will also cause a leak.
3-11001-0809-00011/20/99
3-1
3/Setup
1.Gasket
2.Tee Handle
3.Yoke Clamp
4.Pipeline Inlet
1 234
AA.11.194
Figure 3-1
Mounting gas cylinders
6.Line the yoke index pins up with the cylinder post and swing the clamp closed.
7.Hand tighten the tee handle to secure the cylinder.
8.Install cylinder plugs and gaskets in all unused yoke positions.
9.Verify that the cylinder wrench is present.
WARNINGWARNING
w
WARNINGAlways close the cylinder valve when you are using a pipeline supply. If a cylinder
WARNINGWARNING
valve is open and the pipeline and regulated cylinder pressures are equal, both
supplies may be used simultaneously, leaving an insufficient reserve in case of
pipeline failure.
Mounting the 5125 OMounting the 5125 O
Mounting the 5125 O
Mounting the 5125 OMounting the 5125 O
1.Tighten the thumb screw on the mounting post. This pushes a spring loaded plate in, reducing
the width of the block.
2.Fit the mounting block into the mounting track on the left hand side of the Excel MRI. Loosen
the thumb screw. This wedges the plate against the sides of the mounting track.
monitor monitor
monitor
monitor monitor
22
2
22
3-21001-0809-00011/20/99
3-2
3/Setup
1
8
7
6
5
MD.12.008
2
3
4
1.O2 Monitor
2.O2 Sensor Connection
3.Nut
4.Clamp
5.Thumb Screw
6.Plate
7.Utility Post
8.Mounting Track
Figure 3-2
Mounting the 5125 O2 Monitor
3.Make sure that the utility post is secure and that both sides of the block are wedged into the
rail.
4.Screw the clamp into the threaded hole in the back of the 5125 O
Monitor.
2
5.Attach the clamp to the utility post.
6.To adjust the angle of the 5125 O
Monitor, loosen the nut behind the monitor, reposition the
2
monitor and tighten the nut.
7.Connect the O
in Chapter 5 for O
sensor to the 5125 O2 Monitor. Refer to the section “O2 sensor maintenance”
2
sensor assembly instructions.
2
3-31001-0809-00011/20/99
3-3
3/Setup
Circuit and monitoring connectionsCircuit and monitoring connections
Circuit and monitoring connections
Circuit and monitoring connectionsCircuit and monitoring connections
Use this section as a quick reference if you are already familiar with system connections. For
detailed information and accessory part numbers, refer to the individual operation and maintenance manual.
WARNINGWARNING
w
WARNINGUse only Datex-Ohmeda specified cables, hoses, and tubing for external connec-
WARNINGWARNING
tions. Alternative cables, hoses, or tubing could cause false sensor readings or
damage to the system.
1.Set the system master switch to “Off.”
2.Make sure that the absorber and the 5125 O
3.Make the monitoring connections:
WARNINGWARNING
w
WARNINGFigure 3-3 and 3-4 show the only approved use of the 121 Respirometer with the
WARNINGWARNING
Monitor are securely mounted.
2
Excel MRI.
• Refer to Figure 3-3 if you have a GMS absorber
• Refer to Figure 3-4 if you do not have a GMS absorber
5
4
Output Sensor
1
3
1.O2 Sensor
2.GMS Absorber
3.Absorber Expiratory Port
4.121 Respirometer
5.5125 O2 Monitor
Figure 3-3
sensor connection to a GMS Absorber
O
2
3-41001-0809-00011/20/99
2
AA.19.036
3-4
3/Setup
WARNINGWARNING
w
WARNINGAny absorber used in an MRI environment must be MRI compatible.
WARNINGWARNING
CAUTIONCAUTION
w
CAUTIONThe cable on the O2 sensor must point up to help keep the front (sensing portion) of
CAUTIONCAUTION
the cartridge free of condensate.
12 3
6
1.O2 Sensor
2.Adapter (22 mm Inner Diameter)
3.Absorber Inspiratory Port
4.Absorber Expiratory Port
5.121 Respirometer
6.5125 O2 Monitor
Output Sensor
5
4
AA.19.037
Figure 3-4
O2 sensor connection to a MRI compatible absorber other than the GMS
4. Determine if the O
a. The O
sensor is ready for immediate use if it has not been disconnected since the last
2
sensor is ready for immediate use or if it must be allowed to stabilize:
2
time you used the Excel System.
b. If you just installed a new O
ing, it must be connected and allowed to stabilize for five minutes before it can be used.
3-51001-0809-00011/20/99
sensor cartridge directly from its sealed protective packag-
2
3-5
3/Setup
c. If the O2 sensor was disconnected or a new O2 sensor was previously removed from its
protective packaging, it must be connected and allowed to stabilize. Allow the O
to stabilize for as many hours as it was disconnected or removed from its package (up to a
maximum of 24 hours).
WARNINGWARNING
w
WARNINGAny ventilator used with the Excel must have a high pressure alarm and relief system
WARNINGWARNING
for the patient circuit.
5. Make the fresh gas and gas scavenging connections:
WARNINGWARNING
w
WARNINGThe fresh gas hose must be securely connected to the common gas outlet and the absorber.
WARNINGWARNING
• Fresh gas input from the Excel MRI common gas output to the absorber input.
WARNINGWARNING
w
WARNINGDo not connect the exhaust directly to a vacuum source. The vacuum may draw necessary
WARNINGWARNING
gases from the patient circuit. Refer to the operation and maintenance manual for the waste
gas scavenging valve for additional precautions and information.
• From the excess gas outlet on the absorber to a gas scavenging system. Use 19 mm corrugated
tubing.
sensor
2
1.Common Gas Outlet
2.Absorber
3.Exhaust (Connect to Waste Gas Scavenging)
Figure 3-5
Excel MRI fresh gas connection
1
2
AA.11.226
3
6. Verify that an O2 cylinder is installed.
7. Set the Ventilator/Bag selector switch to the “Bag” position and install a breathing bag on the
bag arm.
8. Turn all flow controls completely clockwise (minimum flow).
9. Continue with Chapter 4, “Preoperative Checkout.”
3-61001-0809-00011/20/99
3-6
4/Preoperative Checkout
In this sectionIn this section
In this section
In this sectionIn this section
Before starting the checkout4-1
MRI compatibility check4-1
Initial Checks4-2
Checking vaporizer mounting4-3
Checking cylinder and pipeline supplies4-3
Checking vaporizer back pressure4-4
Leak checking the low pressure circuitry4-4
Checking the gas flow controls4-6
Breathing system checks4-7
5125 O
Before starting the checkoutBefore starting the checkout
Before starting the checkout
Before starting the checkoutBefore starting the checkout
This section—“Preoperative Checkout” —describes the minimum checks that should be made
before the Datex-Ohmeda Excel MRI System is used on a patient. Do not use the system if it does
not function correctly, as described in the preoperative checkout procedures; instead call a
qualified service representative.
WARNINGSWARNINGS
w
WARNINGSAlways complete the preoperative checkout procedures in this section before using
WARNINGSWARNINGS
the Excel MRI System on a patient.
wMake sure that you understand the correct connection, use and necessary precau-
tions for all system components before using the Excel MRI System. For additional
precautions and information, refer to the operation and maintenance manuals for
each component.
MRI compatibility checkMRI compatibility check
MRI compatibility check
MRI compatibility checkMRI compatibility check
WARNINGSWARNINGS
w
WARNINGSDo not move the Excel MRI into the MRI room until you have completed the first three
WARNINGSWARNINGS
sections of this chapter “MRI compatibility check,” “Initial checks” and “Checking
vaporizer mounting.” Complete the rest of the checkout using the actual room,
pipeline and electrical supplies and gas cylinders that will be used during the case.
Monitor checks4-7
2
wDatex-Ohmeda strongly recommends using only monitor sensors that are compatible
with MRI applications. Use of monitoring device sensors (ECG monitor leads, oximeter probes, etc.) in an MRI environment can cause injury. Check sensor locations on
the patient’s body for signs of discomfort, heating or warming.
1. Check that the gas cylinders are non-magnetic.
2. Open any drawers and check that they contain only MRI compatible items. Test any suspect
items. Remove any items that are not MRI compatible.
3. If a ventilator is connected that is not MRI compatible (e.g. a Datex-Ohmeda 7000 or a 7800
Ventilator), remove it from the Excel MRI. Any ventilator used must be MRI compatible.
WARNINGWARNING
w
WARNINGUse only MRI compatible accessories with the Excel MRI. This includes monitors.
WARNINGWARNING
4. Remove any other accessories that are not MRI compatible.
4-11001-0809-00011/20/99
4-1
4/Preoperative Checkout
Initial checksInitial checks
Initial checks
Initial checksInitial checks
CAUTIONCAUTION
w
CAUTIONDo not exceed the following shelf weight limits: top shelf, 23 kg (50 lbs); middle
CAUTIONCAUTION
shelf, 11 kg (24 lbs). Verify that all equipment on the top shelf is securely strapped in
place.
1. Check the overall integrity of the machine. Make sure that:
• The casters are securely attached
• All accessories are properly mounted
• Setting the brake helps prevent the front casters from turning
2. Make sure the breathing circuit is complete, undamaged, and, if appropriate, contains
adequate CO
WARNINGWARNING
w
WARNINGEnsure that all hoses, tubing, and other circuit connections are made properly before
WARNINGWARNING
using this anesthesia system. Failure to do so may result in patient injury. Refer to the
operation and maintenance manuals for these devices.
3. Make sure that the following are not damaged:
absorbent.
2
• Cylinder yokes
• Pipeline inlets
• Flowmeters and flow control valves
• Pressure gauges
• Vaporizers
• Monitors and cables
• All hoses and tubing
4. Make sure the breathing circuit is closed and connected to a gas scavenging system.
5. Check that the cylinders are properly installed.
6. Check that the vaporizers are properly installed.
7. Check that the cylinder wrench is available.
8. Check that the brake is set.
9. Make sure that required emergency equipment is available and in good working order.
You can use both Tec 4 and Tec 5 vaporizers with the Excel MRI System. If a Tec 4 and a Tec 5
vaporizer are used together, the interlock system still helps prevent more than one vaporizer from
being switched “On” at the same time.
WARNINGWARNING
w
WARNINGDo not attempt to modify the system to accept Datex-Ohmeda Tec 3 vaporizers. Tec 3
WARNINGWARNING
vaporizers are not MRI compatible.
1. Make sure that the tops of the vaporizers are parallel to the top edge of the manifold. Re-
mount any vaporizers that are out of line.
2. Make sure the vaporizers are locked in place. Check the vaporizers’ locking knobs to ensure
that the vaporizers are locked in place.
3. Attempt to lift each vaporizer off of the manifold. Remount any vaporizer that is not securely
locked in place.
WARNINGWARNING
w
WARNINGDo not use a vaporizer that continuously lifts off the manifold when its locking lever is
WARNINGWARNING
in the lock position.
4. Check that the interlock rods are aligned. Attempt to switch “On” more than one vaporizer at a
time. Try every combination that is possible with your system. If more than one vaporizer can
be switched “On” at a time, remount the vaporizers and repeat this step.
WARNINGWARNING
w
WARNINGDo not use the system if the vaporizer interlock system allows more than one vapor-
WARNINGWARNING
izer to be switched “On” at a time.
5. Fill the vaporizers as described in the vaporizer operation and maintenance manual.
Checking cylinder and pipeline suppliesChecking cylinder and pipeline supplies
Checking cylinder and pipeline supplies
Checking cylinder and pipeline suppliesChecking cylinder and pipeline supplies
Complete this section and the rest of Chapter 4 in the MRI room. The MRI System should be
switched “On” during this portion of the checkout procedure.
1. With the cylinders closed, make sure the cylinder gauges read zero.
If the cylinder gauges do not read zero, the cylinders may have been previously opened.
Briefly set the system master switch to “On” and open the flow control valves to bleed
pressure.
CAUTIONSCAUTIONS
w
CAUTIONSOpen cylinder valves slowly to help avoid damaging the regulators.
CAUTIONSCAUTIONS
wForcing the flow controls can cause serious damage to the Excel MRI.
2. Close all flow control valves and open the cylinders. Check that the cylinder gauges show
adequate pressure.
3. Close all of the cylinder valves and note the value on each cylinder pressure gauge. The
gauges must show less than a 690 kPa (100 psig) pressure drop in a five minute period. If
the pressure drop exceeds this limit, the high-pressure circuit has an unacceptable leak.
4-31001-0809-00011/20/99
4-3
4/Preoperative Checkout
If there is an unacceptable leak:
a. Check for a defective cylinder gasket, an extra cylinder gasket, or a loose tee handle.
Replace the gasket(s) with a single, undamaged gasket or tighten the tee handle.
b. Repeat the leak check. If the circuit still leaks, do not use the system for clinical applica-
tions. Call a qualified service representative for repairs.
4. Connect pipeline supplies and check that pipeline pressure gauges read approximately
310-345 kPa (45-50 psig ).
Checking vaporizer back pressureChecking vaporizer back pressure
Checking vaporizer back pressure
Checking vaporizer back pressureChecking vaporizer back pressure
WARNINGWARNING
w
WARNINGFollow the appropriate agent evacuation/collection procedures. Use the hospital gas
WARNINGWARNING
evacuation system.
CAUTIONCAUTION
w
CAUTIONSudden surges can damage the flow control assembly. Make sure that all flow
CAUTIONCAUTION
controls are fully clockwise (minimum flow or off) before setting the system master
switch to “On.”
1. Set the system master switch to “On” and adjust the O2 flow to 6 l/min. Verify that the O2 float
does not oscillate or stick.
Note: The pneumatic alarm sounds briefly when the Excel MRI is switched “On.”
2. Adjust vaporizer concentration from 0 to 1% click by click and observe the O
3. Return the vaporizer to “Off.” If O
qualified service personnel.
Note: If replacing the vaporizer allows the Excel MRI System to pass the vaporizer back
pressure test, the fault is in the original vaporizer. Remove the original vaporizer from use and
contact qualified service personnel to repair it.
4. Repeat steps 2 and 3 for each mounted vaporizer.
Leak checking the low pressure circuitryLeak checking the low pressure circuitry
Leak checking the low pressure circuitry
Leak checking the low pressure circuitryLeak checking the low pressure circuitry
1. Check the low pressure leak test device:
• Block the inlet and squeeze the bulb until it is fully collapsed.
• If the bulb inflates in less than 60 seconds, replace the leak test device.
2. Set the system master switch to “Off:”
Note: The pneumatic alarm sounds when the Excel MRI is switched “Off.”
2
flow.
2
flow dropped by more than 1 l/min during step 2, contact
3. Verify that all vaporizers are “Off.” This changes gas routing.
4. Leak check the low pressure circuitry:
4-41001-0809-00011/20/99
4-4
4/Preoperative Checkout
WARNINGWARNING
w
WARNINGA low pressure leak means that metabolic gases and anesthetic agent are leaking
WARNINGWARNING
into the atmosphere instead of going into the patient circuit. Low pressure leaks must
be repaired before the Excel MRI is used.
Note: Gas supplies can be left “On” without affecting test results. With the system master switch
set to “Off” there should be no flow through the system.
a. Turn the flow controls one and a half turns counter-clockwise.
b. Connect the low pressure leak check device to the common gas outlet.
c. Repeatedly squeeze and release the bulb until it collapses.
Note: The vacuum from the leak check device may cause the floats to rise slightly. This is
normal.
d. If the bulb inflates again in 30 seconds or less, there is an unacceptable leak in the low
pressure circuitry.
e. Disconnect the low pressure leak test device.
f. Set one of the mounted vaporizers to 1% and repeat steps b through d.
g. Remove the test device to relieve the vacuum. Then, switch the vaporizer “Off.”
h. Set the second mounted vaporizers to 1% and repeat steps b through d.
i.Remove the test device to relieve the vacuum. Then, switch the vaporizer “Off.”
j.Store the test device in the Excel MRI drawer.
k. Turn all flow controls completely clockwise (minimum flow). Do not over tighten.
CAUTIONCAUTION
w
CAUTIONVerify that all vaporizers are “Off” before continuing.
CAUTIONCAUTION
WARNINGWARNING
w
WARNINGAfter performing the low pressure leak test, do not use the anesthesia system until
WARNINGWARNING
the system has been purged with oxygen. Using a system that has not been purged
with oxygen may result in incorrect gas mixtures and injury to the patient.
5. Reconnect the common gas outlet to the breathing circuit and the scavenging system, set the
system master switch to “On,” adjust the O
for one minute.
6. Return the system master switch to “Off.”
flow to 1 l/min and purge the Excel MRI with O
2
2
4-51001-0809-00011/20/99
4-5
4/Preoperative Checkout
Checking the gas flow controlsChecking the gas flow controls
Checking the gas flow controls
Checking the gas flow controlsChecking the gas flow controls
WARNINGWARNING
w
WARNINGThe Link 25 Proportion Limiting Control System sets a minimum O2 concentration in
WARNINGWARNING
the fresh gas stream when only O2 and N2O are used. Recirculating through an
absorber or using a third gas may still cause a hypoxic mixture to be delivered,
especially at low O2 flow rates.
1. Either connect the pipeline supplies or slowly open the cylinder valves.
2. Set the system master switch to “On” and turn all flow control valves completely clockwise
(minimum flow).
3. The oxygen flowmeter should show about 200 ml/min. The other flowmeters should show no
gas flow.
4. Use the Link 25 Proportion Limiting Control System to check the N
O flow control. Observe the
2
following precautions:
• Adjust only the N
• Start with the N
• Increase the N
O flow control.
2
O and O2 flow controls at the minimum setting.
2
O flow as specified in the table and make sure that the O2 flow is in the
2
allowed range.
• If you overshoot a setting, turn the O
flow control clockwise until the N2O flow decreases
2
to the previous level.
Set the NSet the N
Set the N
Set the NSet the N
to (l/min):to (l/min):
to (l/min):
to (l/min):to (l/min):
O flow controlO flow control
O flow control
O flow controlO flow control
22
2
22
The OThe O
flow must be (l/min) flow must be (l/min)
The O
flow must be (l/min)
The OThe O
flow must be (l/min) flow must be (l/min)
22
2
22
MinimumMinimum
Minimum
MinimumMinimum
MaximumMaximum
Maximum
MaximumMaximum
0.90.240.36
1.50.400.61
3.00.791.22
6.01.582.44
9.02.373.66
WARNINGWARNING
w
WARNINGDuring operation, always use the O2 Flow control to increase total gas flow and the
WARNINGWARNING
N2O flow control to decrease total gas flow.
5. Test link system tracking. Observe the following precautions:
• Start with the N
• Start with an initial O
O flow control set to 9.0 l/min. Adjust only the O2 flow control.
2
flow between 2.40 and 3.66 l/min and go from higher to lower flow
2
rates.
• If you overshoot a setting, turn the N
rises to the previous level.
Set the OSet the O
Set the O
Set the OSet the O
to (l/min):to (l/min):
to (l/min):
to (l/min):to (l/min):
flow control flow control
flow control
flow control flow control
22
2
22
3.07.3611.41
1.02.463.80
0.51.231.90
0.30.741.14
4-61001-0809-00011/20/99
O flow control counterclockwise until the O2 flow
2
The NThe N
O flow must be (l/min)O flow must be (l/min)
The N
O flow must be (l/min)
The NThe N
O flow must be (l/min)O flow must be (l/min)
22
2
22
MinimumMinimum
Minimum
MinimumMinimum
MaximumMaximum
Maximum
MaximumMaximum
4-6
4/Preoperative Checkout
WARNINGWARNING
w
WARNINGDo not use the anesthesia system if the Datex-Ohmeda Link 25 Proportion Limiting
WARNINGWARNING
Control System does not operate within permitted ranges. Using an incorrectly
operating control system may result in incorrect gas mixtures, and injury to the
patient.
6. Adjust all of the gas flows to mid scale. While you are turning the flowmeter knobs, the
flowmeter floats must move smoothly.
7. Shut “Off” the oxygen supply either by closing the oxygen cylinder valve, or by disconnecting
the oxygen pipeline supply. As pressure bleeds off:
• The oxygen-supply failure alarm must sound.
• All gas flow must fall to zero, with oxygen being the last gas to stop flowing.
8. Turn all of the flow control valve knobs completely clockwise to the minimum flow or closed
position. Do not over tighten the valves.
Breathing system checksBreathing system checks
Breathing system checks
Breathing system checksBreathing system checks
Note: Refer to the GMS Absorber Operation and Maintenance Manual for additional information. A
condensed version of the checkout is found on the back of the GMS Absorber.
5125 O5125 O
5125 O
5125 O5125 O
1. Complete the GMS absorber checkout procedure.
2. Reconnect the unit for manual ventilation. Remove all occlusions from the circuit. Remove the
test plug from the ‘Y’ piece and connect a test lung. Press the Flush button to fill the breathing
bag. Set the Bag/Ventilator switch to the “Bag” position and adjust the APL valve to relieve at
the desired pressure.
3. If the Excel MRI will not be used immediately, set the system master switch to “Off” and close
all gas cylinder valves.
Monitor checks Monitor checks
Monitor checks
Monitor checks Monitor checks
22
2
22
1. Switch the O2 Monitor “On.”
Note: To avoid a long stabilization period, install a new cartridge (five minute stabilization time if
just removed from package).
2. Set the alarm switch to 20% and the alarm switch to “00.”
3. Press down and release switch. Check that the “BATT OK” message appears
for at least five seconds.
4. Press down and hold the switch. Check that:
O2%
O2%
• The alarm speaker beeps once.
• All four messages appear (Figure 4-1).
• An oxygen concentration between 88 and 102% is displayed.
• Both alarm LEDs illuminate.
• When you release the switch “BATT OK” remains for approximately five seconds.
4-71001-0809-00011/20/99
4-7
4/Preoperative Checkout
Figure 4-1
Circuit test display
MD.07.020
5. Calibrate the O
• Expose the O
Monitor at 21% O2:
2
sensor to room air and allow the reading to stabilize for two minutes as
2
room air fills the sensor housing.
• If necessary, remove the two corner screws, open the lower front panel, and adjust the
calibration control (CAL) until the display reads 20%. Then, adjust the CAL control until
the display just indicates 21%.
• If you cannot calibrate the O
Monitor, replace the sensor cartridge. Refer to “O2 sensor
2
maintenance,” in Chapter 5.
6. Set the alarm switch to 22% (1% over the display reading) and check that:
O2%
• The red LED flashes.
LOW O2LOW O2
•
LOW O2 appears in the display.
LOW O2LOW O2
• A two-pitch alarm sounds.
• Pressing the
• Adjusting the alarm switch to 20% (1% below the display reading) stops the alarm.
O2%
button stops the audible alarm for 30 seconds.
å
O2%
7. Set the alarm switch to 20% (1% below the display reading) and check that:
• The yellow LED flashes.
• HIGH O2 appears in the display.
• An intermittent alarm sounds.
O2%
• Adjusting the alarm switch to 22% (1% over the display reading) stops the alarm.
8. At least once a month and following sensor cartridge replacement, calibrate the O
for 100% O
:
2
O2%
• Adjust the alarm switch to 00% (alarm disabled).
• Expose the O
sensor to pure oxygen and allow the display to stabilize for two minutes as
2
oxygen fills the patient circuit.
• Remove the two corner screws, open the lower front cover, and adjust the calibration control
(CAL) until the display reads 99%.
• Expose the O
sensor to room air and allow the display to stabilize for two minutes as
2
room air fills the sensor housing. If the final reading is outside the allowed range 21 ± 3%
(18 to 24%), the sensor cartridge is no longer linear and must be replaced. Refer to “O
sensor maintenance,” in Chapter 5.
9. If the system will not be used immediately, switch the 5125 O
4-81001-0809-00011/20/99
Monitor “Off.”
2
Monitor
2
2
4-8
5/Maintenance
In this sectionIn this section
In this section
In this sectionIn this section
Maintenance schedule5-1
Cleaning and sterilization5-2
Cleaning5-2
Sterilization5-3
Special precautions for rubber articles5-3
O
sensor maintenance5-3
2
Installing a cartridge or disassembling the O
Cleaning and sterilization5-6
sensor for cleaning5-3
2
Replacing the 5125 O
WARNINGSWARNINGS
w
WARNINGSThis manual specifies Krytox® as an oxygen service lubricant. Do not use any lubri-
WARNINGSWARNINGS
Monitor battery5-7
2
cant on the Excel MRI that is not specifically approved for use on anesthesia or
oxygen equipment. Oil and grease based lubricants burn violently and may explode in
the presence of oxygen.
wStatic electricity is a fire hazard. Use only anti-static materials to cover the Excel MRI
System and its components.
Maintenance scheduleMaintenance schedule
Maintenance schedule
Maintenance scheduleMaintenance schedule
This schedule lists the minimum maintenance required, based on normal use and typical environmental conditions. Heavier use or unusual environments may require more frequent maintenance. Before
any cleaning or sterilization procedure check the section labeled “Cleaning and sterilization.”
Before each useBefore each use
Before each use
Before each useBefore each use
Perform preoperative checkout procedure (includes 21% O2 calibrations).
DailyDaily
Daily
DailyDaily
Clean the external surfaces.
Every two weeksEvery two weeks
Every two weeks
Every two weeksEvery two weeks
5-1
Drain and discard anesthetic agent from vaporizers. Less frequent changes may be required if the
agent does not contain additives or stabilizing agents.
MonthlyMonthly
Monthly
MonthlyMonthly
Calibrate the O
Apply Krytox to the threads on the yoke tee handle. Do not lubricate the absorber post assembly.
Every three monthsEvery three months
Every three months
Every three monthsEvery three months
Contact trained service personnel to perform a full checkout and scheduled service maintenance
on the Excel MRI System (Excel MRI, absorber, and O
AnnuallyAnnually
Annually
AnnuallyAnnually
Replace the external vaporizer port o-rings.
Service Datex-Ohmeda Tec 4 vaporizers at an authorized Datex-Ohmeda service center.
Replace the O
(77°F). Different operating condition (higher O2 concentration, high temperature, etc.) can
shorten cartridge life expectancy. Freezing may damage the sensor cartridge.
®
Krytox is a registered trademark of Dupont de Nemours E.I. & Company Inc.
Monitor with 100% O2.
2
monitor).
2
sensor cartridge. Cartridge life expectancy is one year at 50% O2 and 25°C
2
1001-0809-00011/20/99
5-1
5/Maintenance
Every three yearsEvery three years
Every three years
Every three yearsEvery three years
Service Datex-Ohmeda Tec 5 vaporizers at an authorized Datex-Ohmeda service center.
As requiredAs required
As required
As requiredAs required
Install new cylinder gaskets.
Replace the absorbant in the GMS.
Replace the oxygen monitor battery.
Cleaning and sterilizationCleaning and sterilization
Cleaning and sterilization
Cleaning and sterilizationCleaning and sterilization
CAUTIONSCAUTIONS
w
CAUTIONSDo not sterilize the Excel MRI or the 5125 O2 Monitor.
CAUTIONSCAUTIONS
wFollowing ethylene oxide sterilization, quarantine the items in a well ventilated area
to allow dissipation of absorbed ethylene oxide gas. Follow the sterilizer
manufacturer’s recommendations for specific sterilization periods.
Cleaning
Use this section as a quick reference once you are familiar with the cleaning procedures found in
the individual operation and maintenance manuals.
External surfacesExternal surfaces
External surfaces
External surfacesExternal surfaces
To clean external surfaces use a damp cloth and a mild detergent.
Use the minimum amount of liquid necessary since excess liquid may leak into monitoring
connections or other electrical components.
Sensor cleaning precautionsSensor cleaning precautions
Wipe the O2 sensor cable and housing with a damp cloth. Never immerse the O2 sensor assembly
in liquid. Exposure to liquids can damage the electrical contacts.
sensor must be disassembled and the sensor cartridge removed for separate cleaning. The
The O
2
housing can be cleaned with a damp cloth (water, detergent solution or isopropyl alcohol).
Corrosion from leaked sensor electrolyte can be removed with white vinegar under a fume hood.
sensor cartridge contains an electrolyte (caustic). It can be wiped clean with a damp cloth
The O
2
(water, liquid disinfectant, or white vinegar; do not use alcohols). Special precautions are required. Refer to the “O
Precautions for painted, metal, or plastic surfacesPrecautions for painted, metal, or plastic surfaces
Precautions for painted, metal, or plastic surfaces
Precautions for painted, metal, or plastic surfacesPrecautions for painted, metal, or plastic surfaces
Do not use abrasive cleaners. They can mar the finish.
sensor maintenance” section in this chapter for detailed instructions.
2
5-2
Do not use anesthetic agent or glass cleaners on plastic or painted surfaces. They can mar the
surface finish.
Always check cleaning product information to make sure that it is safe for aluminum, painted, or
plastic surfaces.
1001-0809-00011/20/99
5-2
5/Maintenance
Sterilization
Use this section as a quick reference once you are familiar with the sterilization procedures found
in the individual operation and maintenance manuals.
Items that cannot be sterilized by any methodItems that cannot be sterilized by any method
Items that cannot be sterilized by any method
Items that cannot be sterilized by any methodItems that cannot be sterilized by any method
Excel MRI; 5125 O
Items that can be sterilized with ethylene oxideItems that can be sterilized with ethylene oxide
Items that can be sterilized with ethylene oxide
Items that can be sterilized with ethylene oxideItems that can be sterilized with ethylene oxide
These items can be sterilized with ethylene oxide: the front half of the oxygen sensor housing and
rubber and plastic articles. Refer to individual operation and maintenance manuals for disassembly instructions and additional information.
Items that can be sterilized with liquid agentsItems that can be sterilized with liquid agents
Items that can be sterilized with liquid agents
Items that can be sterilized with liquid agentsItems that can be sterilized with liquid agents
The rubber and plastic articles can be sterilized with a cold germicidal solutions.
Special precautions for rubber articles
Rubber goods deteriorate over time and are considered expendable. The presence of oxygen,
ozone, ether, mineral or vegetable oils, phenols, cresols, terpenes, hydrocarbon solvents,
chlorinated hydrocarbons, esters, or oxidizing agents will hasten deterioration.
Check rubber articles regularly. Replace them when any of the following signs of deterioration
appear, swelling, tackiness, or cracking.
Conductive rubber goods loose their electrical conductivity with age.
To extend the useful life of rubber articles:
• Remove metal connectors immediately after use.
Monitor
2
• Store rubber articles in the dark away from ozone sources (florescent light fixtures, electric
motors, and diathermy machines).
WARNINGWARNING
w
WARNINGDo not use talc, zinc stearate, calcium carbonate, corn starch or similar substances
WARNINGWARNING
to prevent tackiness on rubber articles. Any substances used could contaminate or
irritate the patient’s respiratory tract.
OO
sensor maintenance sensor maintenance
O
sensor maintenance
OO
sensor maintenance sensor maintenance
22
2
22
Installing a cartridge or disassembling the O2 sensor for cleaning
WARNINGWARNING
w
WARNINGUse protective gloves and eyewear when you open the O2 sensor in case the cartridge
WARNINGWARNING
is leaking. The sensor cartridge contains an electrolyte (caustic).
After servicing the O2 sensor, complete the checkout procedure “5125 O2 Monitor checks” at the
end of chapter 4.
5-3
1001-0809-00011/20/99
5-3
5/Maintenance
Note: The oxygen sensor cartridge is located inside the sensor housing. Handle the sensor
cartridge with care to avoid damage.
Figure 5-1
Hold the sensor housing, and depress the locking latch using a tool.
MD.12.015
Figure 5-2
Twist to open, and unscrew the top half of the sensor housing. Lift the old sensor out.
MD.12.016
MD.02.032
5-4
Figure 5-3
Install the new sensor cartridge with the circular contacts toward the rear of the sensor housing.
Verify that the top half of the housing is unobstructed to ensure proper flow.
1001-0809-00011/20/99
5-4
5/Maintenance
Figure 5-4
Twist to close, and tighten the housing to form a gas-tight seal.
MD.12.017
10 mV
per
o
%
2
Min
R
L
20M
MD12.019
Figure 5-5
If necessary, immediately connect the O2 sensor to the 5125 O2 Monitor.
WARNINGWARNING
w
WARNINGPerform the 5125 O2 Monitor Checks in chapter 4 after replacing the sensor cartridge
WARNINGWARNING
to ensure the monitor is working properly. Allow at least 5 minutes for the sensor to
stabilize.
5-5
1001-0809-00011/20/99
5-5
5/Maintenance
Cleaning and sterilization
WARNINGSWARNINGS
w
WARNINGSUse protective gloves and eyewear when you open the O2 sensor in case the cartridge
WARNINGSWARNINGS
is leaking. The sensor cartridge contains potassium acetate (caustic).
wDo not inhale any fumes generated by the oxygen sensor cleaning procedure. Such
fumes can cause respiratory system or skin damage. The sensor cartridge contains
potassium acetate (caustic).
A
B
MD.12.018
MD.07.009
5125
AC
Figure 5-6
Cleaning and sterilization methods
Use the following methods to clean and sterilize the parts as labeled in Figure 5-6.
Method A: Wipe with damp cloth (liquid disinfectant, mild detergent solution, isopropyl alcohol);
do not immerse, autoclave or gas sterilize; remove leaked electrolyte under a fume hood with
vinegar.
Method B: Gas sterilize with ethylene oxide or clean with liquid disinfectant, mild detergent
solution, isopropyl alcohol.
CAUTIONCAUTION
w
CAUTIONFollowing ethylene oxide sterilization, quarantine the items in a well ventilated area
CAUTIONCAUTION
MD.02.032
B
to allow dissipation of absorbed ethylene oxide gas. Follow the sterilizer
manufacturer’s recommendations for specific sterilization periods.
5-6
Method C: Wipe with damp cloth (water, white vinegar, liquid disinfectant).
1001-0809-00011/20/99
5-6
5/Maintenance
Replacing the 5125 O2 Monitor battery
WARNINGWARNING
w
WARNINGUse only the specified Datex-Ohmeda, non-magnetic batteries. Standard batteries
WARNINGWARNING
are magnetic and can cause injuries if used in an MRI environment. Refer to the
“Illustrated Parts” section for Stock Numbers.
The 5125 O2 Monitor uses a non-magnetic, 3.9 Vdc lithium battery. The O2 monitor batteries are
not rechargeable.
The battery compartment is located behind the front cover. Remove the two corner screws and
open the cover to gain access to the battery.
After replacing the battery, complete the checkout procedure “5125 O
chapter 4.
1
Monitor checks” at the end of
2
5-7
1.Battery
Figure 5-7
Replacing the battery
1001-0809-00011/20/99
MD.12.014
5-7
Notes
5/Maintenance
5-8
1001-0809-00011/20/99
5-8
6/Troubleshooting
In this sectionIn this section
In this section
In this sectionIn this section
Repair policy6-1
Repair policyRepair policy
Repair policy
Repair policyRepair policy
CAUTIONCAUTION
w
CAUTIONNo repair should ever be attempted by anyone not having experience in the repair of
CAUTIONCAUTION
Problems with the 5125 O
Calibration and drift6-1
5125 O
Pneumatics problems6-3
Do not use malfunctioning equipment. Make all necessary repairs, or have the equipment
serviced by an Authorized Datex-Ohmeda Service Representative. After repair, test the equipment
to ensure that it is functioning properly, in accordance with the manufacturer’s published specifications.
To ensure full reliability, have all repairs and service done by an Authorized Datex-Ohmeda Service
Representative. If this cannot be done, replacement and maintenance of those parts listed in this
manual may be undertaken by a competent, trained individual having experience in the repair of
devices of this nature.
devices of this nature.
Monitor alarms6-2
2
Monitor6-1
2
Replace damaged parts with components manufactured or sold by Datex-Ohmeda. Then test the
unit to ascertain that it complies with the manufacturer’s published specifications.
Contact the nearest Datex-Ohmeda Service Center for service assistance. In all cases, other than
where Datex-Ohmeda’s warranty is applicable, repairs will be made at Datex-Ohmeda’s current
list price for the replacement part(s) plus a reasonable labor charge.
Problems with the 5125 OProblems with the 5125 O
Problems with the 5125 O
Problems with the 5125 OProblems with the 5125 O
Calibration and drift
Cannot calibrate monitorCannot calibrate monitor
Cannot calibrate monitor
Cannot calibrate monitorCannot calibrate monitor
Replace the O
Drift in ODrift in O
Drift in O
Drift in ODrift in O
Monitor O
sensor cartridge. Refer to “O2 sensor maintenance,” in Chapter 5.
2
readings readings
readings
readings readings
22
2
22
readings. If O2 readings fail to stabilize, replace the sensor cartridge.
2
Monitor Monitor
Monitor
Monitor Monitor
22
2
22
6-11001-0809-00011/20/99
6-1
6/Troubleshooting
5125 O2 Monitor alarms
Note: When one message covers multiple conditions, the distinguishing feature is highlighted.
HIGH O2HIGH O2
HIGH O2
HIGH O2HIGH O2
O2 Display Shows: O2 concentration
Alarm LED: Flashing Yellow
Alarm Silence: Until next occurrence
Measured O
concentration is above the alarm limit. Set the alarm limit to “00” to disable high O
Low pressure leak test fails when vaporizer “On”Low pressure leak test fails when vaporizer “On”
Low pressure leak test fails when vaporizer “On”
Low pressure leak test fails when vaporizer “On”Low pressure leak test fails when vaporizer “On”
Check for correct vaporizer mounting. Tighten the vaporizer fill caps.
Make sure that there are external o-rings on all vaporizer manifold ports. Replace missing or
damaged o-rings.
If the leak continues, move the vaporizer to another port. If the leak follows the vaporizer, use a
different vaporizer. Do not use the leaking vaporizer until it has been serviced. If the leak stays with
the port, i.e. any vaporizer put on that port leaks and the external o-rings are installed, contact
qualified service personnel to repair the vaporizer manifold.
High pressure leak or inadequate ventilator gas supplyHigh pressure leak or inadequate ventilator gas supply
High pressure leak or inadequate ventilator gas supply
High pressure leak or inadequate ventilator gas supplyHigh pressure leak or inadequate ventilator gas supply
Make sure that the gas cylinder is correctly aligned and the tee handle is tight. Check that there is
only one cylinder gasket and it is in good condition. Check external hose connections to the O
power outlet.
High pressure leak at gas blockHigh pressure leak at gas block
High pressure leak at gas block
High pressure leak at gas blockHigh pressure leak at gas block
Make sure that the cylinder is correctly aligned and the tee handle is tight. Install yoke plugs as
required. Check that there is only one cylinder gasket and it is in good condition.
2
alarm (pneumatic) alarm (pneumatic)
Low OLow O
alarm (pneumatic)
Low O
alarm (pneumatic) alarm (pneumatic)
Low OLow O
22
2
22
This alarm sounds briefly when the Excel is first switched “On.”
If it occurs during operation, check the gas supply pressure (cylinder or pipeline). If necessary,
switch to a reserve cylinder.
6-31001-0809-00011/20/99
6-3
6/Troubleshooting
Notes
6-41001-0809-00011/20/99
6-4
7/Illustrated Parts
In this sectionIn this section
In this section
In this sectionIn this section
Excel specific parts7-1
MRI Compatible accessories7-1
5125 O
Where to find additional part numbers7-2
This manual lists only customer replaceable parts used directly by the Excel or the 5125 O
Monitor. To locate additional part information find the part description in the section titled “Where
to find additional part numbers.”
Excel specific partsExcel specific parts
Excel specific parts
Excel specific partsExcel specific parts
1102-3016-000Vaporizer port o-rings, external (six)
0236-0042-801Fresh gas hose
0219-3372-600Yoke tee handle
0206-7129-525Yoke plug
0210-5022-300Cylinder gasket
0219-7210-300Test lung
1001-3854-000Krytox
0309-1318-800Low pressure leak test device
Monitor accessories7-2
2
2
2900-0001-000Test plugs, bag port
0216-6764-870GMS Mounting Pin (older style)
Included with AbsorberGMS Mounting Pin (New style)
WARNINGThe MRI compatibility of these accessories applies to specific accessory models and
WARNINGWARNING
is limited to use as part of the Excel MRI System. None of these accessories have
been tested for stand alone use in an MRI environment or in magnetic fringe fields
above 0.23 Tesla (2300 Gauss).
1001-8968-000Add on cylinder, O2 DISS
1001-8970-000Add on cylinder, N
1603-3000-000Respirometer 121 style
1010-8015-000Suction regulator kit
1010-8016-000Free flow suction regulator kit
1010-8021-000Bracket for suction regulator
1010-8017-000Auxiliary O
flowmeter
2
O DISS
2
7-11001-0809-00011/20/99
7-1
7/Illustrated Parts
1001-8930-000Flip-up shelf
0216-6819-8001” x 3.5” Post dovetail mount
0216-6814-80012” IV Pole Dovetail mount
1001-8957-000Large Tycos Case Dovetail mount
0211-1100-300BP adult inflation system Tycos
0207-8022-801Bag to ventilator switch valve
5125 O5125 O
5125 O
5125 O5125 O
Where to find additional part numbersWhere to find additional part numbers
Where to find additional part numbers
Where to find additional part numbersWhere to find additional part numbers
Monitor accessories Monitor accessories
Monitor accessories
Monitor accessories Monitor accessories
22
2
22
1001-3741-000Non-magnetic lithium 3.9 Vdc C size battery
0237-2034-700O
6050-0005-406O
6050-0005-438O
0210-0499-300O-ring small
0212-0763-100Sensor adapter
System operation and maintenance manualsSystem operation and maintenance manuals
System operation and maintenance manuals
System operation and maintenance manualsSystem operation and maintenance manuals
All major system components have individual operation and maintenance manuals:
• GMS Absorber0178-1742-000
• GMS Bain Circuit Adapter0178-1752-000
2
2
2
sensor cartridge
sensor (without cartridge)
sensor front housing w/ o-ring
• GMS PEEP Valve0178-1753-000
• Tec 4 Vaporizer0205-7106-300
• Tec 5 Vaporizer1105-0100-000
• Waste Gas Scavenging Valve0178-1728-000
Parts listed in the GMS Absorber Operation and Maintenance ManualParts listed in the GMS Absorber Operation and Maintenance Manual
Parts listed in the GMS Absorber Operation and Maintenance Manual
Parts listed in the GMS Absorber Operation and Maintenance ManualParts listed in the GMS Absorber Operation and Maintenance Manual
Look the following parts up in the GMS Absorber Operation and Maintenance Manual:
• Unidirectional (inspiratory and expiratory) check valve components
• Pressure gauge and inspiratory pressure sensing tee
• Drain plug
• O-rings
• Canister gaskets, screens, hose connections and seals
7-21001-0809-00011/20/99
7-2
Appendix
In this sectionIn this section
In this section
In this sectionIn this section
Excel PneumaticsA-1
Excel MRI System SpecificationsA-4
Excel MRI PneumaticsExcel MRI Pneumatics
Excel MRI Pneumatics
Excel MRI PneumaticsExcel MRI Pneumatics
The Excel Gas Machine consists of the gas supply modules, the flow tubes and controls responsible
for gas concentration, the vaporizer assemblies, and specialized alarm and “On” /“Off” functions
provided by the oxygen supply. Most of the pneumatic circuitry is located underneath the tabletop.
The Excel MRI comes with gas supply modules for nitrous oxide, oxygen, and air.
Gas supply modules contain two parallel branches, pipeline and cylinder. The pipeline supply
branch consists of a filtered pipeline connection and a pipeline pressure gauge. The cylinder branch
consists of a filtered cylinder connection, a cylinder gauge, and a high pressure gas regulator that
reduces cylinder pressure to a nominal 311 kPa (45 psig). When only one source is connected or
switched “On,” check valves keep gas from leaking out through the unused branch. When both
pipeline and cylinder supplies are connected, make sure that the cylinder valve is closed. Otherwise, cylinder gas could be depleted since the pipeline pressure is almost the same as the regulated cylinder pressure. A pressure relief valve in each gas supply module limits maximum pressure
to 518 kPa (75 psig).
Shut-off valves control the flow of the nitrous oxide and the air. When the system master switch is
set to “Off,” these valves are closed. When the system master switch is set to “On,” the pilot oxygen
pressure holds them open. As long as the pilot pressure remains above the critical level, approximately 138 kPa (20 psig), the valves remain open. If the oxygen pilot pressure drops below this
level, the shut-off valves close, stopping the flow of nitrous oxide and the air. You should notice the
pilot oxygen flow through the oxygen flow tube whenever the system master switch is “On,” even if
the oxygen flow control is turned completely clockwise (minimum flow).
Secondary regulators reduce the oxygen and nitrous oxide pressures to the levels required by the
flow controls. The air goes directly from the primary regulator to the flow control. A gear linkage on
the nitrous oxide and oxygen flow control knobs helps limit the lowest O
nominal 25% of the N
flows.
The mixed gas output of the flowmeter assembly goes through the selected vaporizer and into the
patient circuit. A pressure relief valve on the common gas outlet limits pressure to approximately
27.6 to 38 kPa (4 to 5.5 psig) at 200 ml/min.
Pressing the “Flush” button connects pipeline or regulated cylinder oxygen directly to the gas
machine outlet regardless of the position of the system master switch.
The same pilot pressure that opens the shut-off valves pressurizes the alarm canister. A regulator
keeps the oxygen in the canister as long as supply pressure exceeds approximately 207 kPa (30
psig); range: 186 to 228 kPa (27 to 33 psig). If the oxygen supply pressure falls below this level,
gas exits the canister through a reed alarm.
O flow control setting. A system of mechanical stops sets the maximum
2
flow control setting to a
2
A-1
1001-0809-00011/20/99
A-1
Appendix
Air
11
10
FM
N2O
11
1010
O
2
11
Ohmeda
Link 25
9
2d
2d
VM
5b
12
9
GSM N2O
GSM Air
PLCYL
1111
333
4a
GSM N2OGSM O
PLCYL
2a
4a4a
5b
5b
PLCYL
5b
12
2
2a2a
13
CGO
88
PO
O
2
4b
7
PA
2c2b
6
5a
A-2
Figure A-1
Typical Excel pneumatic circuitry
1001-0809-00011/20/99
AA.11.213
A-2
Appendix
Key to Subassembly AbbreviationsKey to Subassembly Abbreviations
Key to Subassembly Abbreviations
Key to Subassembly AbbreviationsKey to Subassembly Abbreviations
CGOCommon Gas Outlet
CYLCylindery Yoke
GSMGas Supply Module (O2, N2O, Air)
FMFlowmeter
PAPneumatic Alarm (triggers betwen 186 and 228 kPa [27 and 33 psig])
PLPipeline Connection (nominal 345 kPa [50 psig])
POPower Outlet (O2 for ventilator Drive gas)
VMVaporizer Manifold
Key to Numbered ComponentsKey to Numbered Components
Key to Numbered Components
Key to Numbered ComponentsKey to Numbered Components
1. Pipeline Pressure Gauge
2. Regulators
a. Primary Regulator
b. Alarm Regulator (set threshold)
c. Flow Regulator (alarm duration)
d. Secondary Regulators (262 ± 3.5 kPa for N
psig for N
3. Cylinder Pressure Gauge
4. Pressure Relief Valves
a. 517 kPa (75 psig)
b. 27.6 to 38 kPa (4 to 5.5 psig) at 200 ml/min
5.Switches
a. System Master Switch
b. Vaporizer Selector Switch
6. Alarm Reservoir
7. Whistle
8. Shut-Off Valve
9. Secondary Regulator Test Points
10. Flow Control Needle Valves
(maximum flows: 10-14 1/min O
(minimum flows: 180 to 220 mil O
11. Flowtube Assembly
12 Anesthesia Vaporizer
13. Flush Button/Valve
O; 20.75 ± 3.75 psig for O2 at 2 l/min])
2
O; 143 ± 26 kPa for O2 at 2 l/min [38 ± 0.5
2
; N2O 10-12 l/min; Air 15-17 1/min.)
2
; 0 ml all other gases)
2
Key to Unnumbered SymbolsKey to Unnumbered Symbols
Key to Unnumbered Symbols
Key to Unnumbered SymbolsKey to Unnumbered Symbols
Filter
Check Valve
Gas Circuit Fitting
Gas Circuit Connection
AA.19.041
A-3
Direction of Gas Flow
1001-0809-00011/20/99
A-3
Appendix
Excel MRI System SpecificationsExcel MRI System Specifications
Excel MRI System Specifications
Excel MRI System SpecificationsExcel MRI System Specifications
Note: All specifications are nominal and subject to change without notice.
Dimensions
Weight:Weight:
Weight:
Weight:Weight:
Excel 210: 93.5 kg (206 lbs)
Options: Vaporizer, 17.6 kg (17 lbs); Absorber; 9.6 kg (21 lbs)
Dimensions:Dimensions:
Dimensions:
Dimensions:Dimensions:
Height: 168 cm (66 in) floor to upper most part of top shelf.
Depth: 76 cm (30 in) with casters and brake bar.
Width: 73.6 cm (29 in)
Shelves and drawers:Shelves and drawers:
Shelves and drawers:
Shelves and drawers:Shelves and drawers:
Middle Shelf:
Size: 29 x 31 cm (11.4 x 12.4 in)
Height: 19.8 cm (7.8 in)
Maximum Load: 11 kg (24 lb)
Top Shelf:
Size: 59.7 x 36 cm (23.5 x 14.2 in)
Maximum Load: 23 kg (50 lb)
Table-Top:
Height (above floor): 84 cm (33.5 in)
Size: 55.0 x 29.0 cm (21.7 x 11.7 in)
Drawer Cabinet (option):
Contains two 39.4 x 26.7 x 10.2 cm (15.5 x 10.5 x 4 in) drawers.
Absorber post mounting assembly:Absorber post mounting assembly:
Absorber post mounting assembly:
Absorber post mounting assembly:Absorber post mounting assembly:
Absorber Arm Length (varies slightly with arm style): 35.5 cm (14 in)
Vertical Adj (above floor): 25.4 to 66 cm (10 to 26 in)
Horizontal Adj: 16.5 to 27.9 cm (6.5 to 11 in)
Casters:Casters:
Casters:
Casters:Casters:
12.7 cm (5 in) diameter; front casters have a foot-operated brake-bar lock.
Common gas outlet:Common gas outlet:
Common gas outlet:
Common gas outlet:Common gas outlet:
Equipped with a latching, positive engagement, bayonet type connector. The common gas outlet
connector will also accept standard 22 mm OD or 15 mm ID conical friction fit connectors.
Note: Flowtube assemblies are calibrated at 20°C (68°F) and 101.3 kPa (760 mm Hg). Different
breathing circuit pressures, barometric pressures or temperatures will affect flowtube accuracy. In
some cases, deviations due to different conditions may exceed the specified tolerances.
Common gas outlet relief valve:Common gas outlet relief valve:
Common gas outlet relief valve:
Common gas outlet relief valve:Common gas outlet relief valve:
27.6 to 38 kPa (4.0 to 5.5 psig) at minimum flow.
pressure alarm: pressure alarm:
Low OLow O
pressure alarm:
Low O
pressure alarm: pressure alarm:
Low OLow O
22
2
22
An alarm sounds for at least seven seconds if the O
supply falls below approximately 207 kPa
2
(30 psig); range: 186 to 228 kPa (27 to 33 psig).
A-5
Oxygen flush button:Oxygen flush button:
Oxygen flush button:
Oxygen flush button:Oxygen flush button:
Recessed, self-closing push button provides a flow of 45-70 l/min when fully depressed.
1001-0809-00011/20/99
A-5
Appendix
5125 O2 Monitor specifications
Range:Range:
Range:
Range:Range:
0-100% O
Display resolution:Display resolution:
Display resolution:
Display resolution:Display resolution:
2
1%
Display update:Display update:
Display update:
Display update:Display update:
Three times per second
Alarms:Alarms:
Alarms:
Alarms:Alarms:
Audible and visual alarms for: high and low O
, low battery, sensor malfunction or disconnection,
1Analog ground
2Shield
3 & 4Normally open switch connection for remote alarm silence.
High to low transition silences alarm.
5% O
; 10 mv = 1% O2 at 10 KΩ load impedance
2
6Alarm status:
high (>4.5 Vdc) = normal, no alarm;
10 msec pulse = alarm in progress; number of pulses in
320 msec specifies the highest priority alarm (1 = low O
2 = high O
4 = dead battery; 5 = low O
; 3 = sensor malfunction/disconnection;
2
limit <18%; 6 = weak battery)
2
2
alarm line stays low for remainder of 320 msec.
;
A-6
1001-0809-00011/20/99
A-6
Warranty
Appendix
This Product is sold by Datex-Ohmeda under the warranties set forth in the following paragraphs.
Such warranties are extended only with respect to the purchase of this Product directly from
Datex-Ohmeda or Datex-Ohmeda’s Authorized Dealers as new merchandise and are extended to
the Buyer thereof, other than for the purpose of resale.
For a period of twelve (12) months from the date of original delivery to Buyer or to Buyer’s order,
but in no event for a period of more than two years from the date of original delivery by DatexOhmeda to a Datex-Ohmeda Authorized Dealer, this Product, other than its expendable parts, is
warranted against functional defects in materials and workmanship and to conform to the
description of the Product contained in this operation manual and accompanying labels and/or
inserts, provided that the same is properly operated under the conditions of normal use, that
regular periodic maintenance and service is performed and that replacements and repairs are
made in accordance with the instructions provided. This same warranty is made for a period of
thirty (30) days with respect to expendable parts. The foregoing warranties shall not apply if the
Product has been repaired other than by Datex-Ohmeda or in accordance with written instructions
provided by Datex-Ohmeda, or altered by anyone other than Datex-Ohmeda, or if the Product has
been subject to abuse, misuse, negligence, or accident.
Datex-Ohmeda’s sole and exclusive obligation and Buyer’s sole and exclusive remedy under the
above warranties is limited to repairing or replacing, free of charge, at Datex-Ohmeda’s option, a
Product, which is telephonically reported to the nearest Datex-Ohmeda Field Service Support
Center and which, if so advised by Datex-Ohmeda, is thereafter returned with a statement of the
observed deficiency, not later than seven (7) days after the expiration date of the applicable
warranty, to the Datex-Ohmeda Service and Distribution Center during normal business hours,
transportation charges prepaid, and which, upon Datex-Ohmeda’s examination, is found not to
conform with above warranties. Datex-Ohmeda
including but not limited to incidental damages, consequential damages, or special damages.
shall not be otherwise liable for any damages
There are no express or implied warranties which extend beyond the warranties hereinabove set
Datex-Ohmeda
forth.
respect to the product or parts thereof.
makes no warranty of merchantability or fitness for a particular purpose with
A-7
1001-0809-00011/20/99
W-1
A-7
Corporate Office
Datex-Ohmeda Division
Instrumentarium Corp.
PO Box 900
FIN-00031 Helsinki
Finland
Tel 358 10 394 11
Fax 358 9 146 3310
North America
United States
Customer Service, Technical Support
and Distribution Center
Datex-Ohmeda, Inc.
PO Box 7550
Madison, WI 53707-7550, USA
Tel 1 800 345 2700
Fax 1 608 221 4384
Equipment Service Center
Datex-Ohmeda, Inc.
1315 West Century Drive
Louisville, CO 80027-9560, USA
Tel 1 800 345 2755
Canada
Datex-Ohmeda (Canada) Inc.
1093 Meyerside Drive, Unit 2
Mississauga, Ontario
L5T 1J6
Canada
Tel 1 800 268 1472
Tel 1 905 565 8572
Fax 1 905 565 8592
Asia/Pacific
China
Datex-Ohmeda Pte. Ltd.
Room B416, COFCO Plaza
8 Jianguomennei Avenue
Beijing 100005, PR China
Tel 86 10 6526 9773
Fax 86 10 6526 0653
Datex-Ohmeda Pte. Ltd.
Room 1708, Yunlong Mansion
No. 122 Luoguo Street
Chengdu 610017, PR China
Tel 86 28 661 4424
Fax 86 28 676 2703
Datex-Ohmeda Pte. Ltd.
403 Huan Shi Dong Road
Room 1602, GIE Tower
Guangzhou, 510095, P R China
Tel 86 20 8732 2521
Fax 86 20 8732 2518
Datex-Ohmeda Pte. Ltd.
Room 2509 Lippo Plaza
No. 222 Huaihai Road (M)
Shanghai 200021, P.R. China
Tel 8621 5382 5657
Fax 8621 5382 1691
Datex-Ohmeda Pte. Ltd.
Room 809, Truroll Plaza
Wusheng Road
Wuhan 430033, P R China
Tel 86 27 8571 2536
Fax 86 27 8571 2655
India
Datex-Ohmeda (India) Pvt. Ltd.
Block EP & GP, Sector V
Plot XI-16, Salt Lake City
Calcutta 700091
India
Tel 91 33 3574002
Fax 91 33 3574001
Indonesia
Datex-Ohmeda Pte. Ltd.
Wisma Danamon Aetna Life 19th Floor
Jln. Jend Sudirman Kav. 45-46
Jakarta 12930, Indonesia
Tel 62 21 575 0864
Fax 62 21 575 0865
Japan
Datex-Ohmeda K. K.
TRC Annex 9F
6-1-1 Heiwajima
Ohta-ku, Tokyo 143-0006
Japan
Tel 81 3 5763 6801
Fax 81 3 5763 6838
Datex-Ohmeda K. K.
Technical Center
TRC A Bldg. AE 4-8
6-1-1 Heiwajima
Ohta-ku, Tokyo 143-0006
Japan
Tel 81 3 5763 6850
Fax 81 3 5763 6852
Korea
Datex-Ohmeda Pte. Ltd.
10th Floor, Sam Sung Building
36 - 1, Yoido-Dong, Youngdeungpo-Ku
Seoul, Korea
Tel 82 2 786 7421
Fax 82 2 786 7420
Malaysia
Datex-Ohmeda Pte. Ltd.
Level 2 Bangunan O'Connor
13 Jalan 223
46100 Petaling Jaya
Selangor, West Malaysia
Tel 60 3 754 7872
Fax 60 3 757 6948
Singapore
Datex-Ohmeda Pte. Ltd.
152 Beach Road
#12-05/07 Gateway East
Singapore 189721
Tel 65 391 8618
Fax 65 291 6618
Thailand
Datex-Ohmeda Pte. Ltd.
12th Floor (Unit F) Grand Amarin Tower
1550 New Petchburi Road, Makasan,
Rajathevi,
Bangkok 10320, Thailand
Tel 66 2 2071012/13
Fax 66 2 207 1014
Taiwan and Philippines
Datex-Ohmeda Pte. Ltd.
2nd Floor, No. 85, Chien-Kuo North
Road, Sec. 2
Taipei, Taiwan
Republic of China
Tel 886-2 2515-0457
Fax 886-2 2501-9136
Vietnam
Datex-Ohmeda Pte. Ltd.
522G Nguyen Tri Phuong St.
Ho Chi Minh City, Dist. 10 Vietnam
Tel 848 865 5875
Fax 848 862 5501
Australia
Datex-Ohmeda Pty. Ltd.
Units 1 & 2
149 Arthur Street
P O Box 356
Homebush
NSW 2140
Australia
Tel 61 132 229
Fax 61 297 461796
Europe
France
Datex-Ohmeda S.A.S.
Parc de Pissaloup, BP 10
8 Avenue Jean d’Alembert
F-78191 Trappes-Cédex
France
Tel 33 1 30 68 60 00
Fax 33 1 30 68 60 01
Datex-Ohmeda S.A.S.
17 rue Jean-Elysée Dupuy
F-69410 Champagne Au Mont d'Or
France
Tel 33 1 30 68 60 00
Fax 33 4 78 43 26 58