This Product will perform in conformity with the description thereof contained in
this operating manual and accompanying labels and/or inserts, when
assembled, operated, maintained, and repaired in accordance with the
instructions provided. This Product must be checked periodically. A defective
Product should not be used. Parts that are broken, missing plainly worn,
distorted, or contaminated should be replaced immediately. Should repair or
replacement become necessary, Datex-Ohmeda recommends that a telephonic
or written request for service advice be made to the nearest Datex-Ohmeda
Field Service Support center. This Product or any of its parts should not be
repaired other than in accordance with written instructions provided by DatexOhmeda and by Datex-Ohmeda trained personnel. The Product must not be
altered without the prior written approval of Datex-Ohmeda’s Quality Assurance
Department. The user of this Product shall have the sole responsibility for any
malfunction which results from improper use, faulty maintenance, improper
repair, damage, or alteration by anyone other than Datex-Ohmeda.
wWARNINGU.S. Federal and Canadian law restrict this device to sale by or on the or-
der of a licensed medical practitioner. Outside the U.S.A. and Canada,
check local laws for any restriction that may apply.
Inhaled Nitric Oxide mixtures must be handled and stored in compliance
with federal, state and local regulations.
Datex-Ohmeda products have unit serial numbers with coded logic which
indicates a product group code, the year of manufacture and a sequential unit
number for identification.
A12345
AAA
This alpha character indicates the year of product
manufacture and when the serial number was
assigned; “B” = 1998, “C” = 1999, “D” = 2000, etc.
“I” and “O” are not used.
Symbols used in this manual or on the system . . . . . . . . . . . . 1-4
.
2
1605-0014-00002/08/00
1-1
Page 10
INOvent delivery system
Important
Before using the INOvent delivery system, read through this manual
and the related manuals provided. Read through the manuals for the
ventilator, humidifier and any other accessory items used. Follow the
manual instructions and obey Warnings and Cautions.
Know the proper setup and operation of the ventilator and the
humidifier you are using.
Know the proper information provided in this Operation and
Maintenance manual before operating the INOvent delivery system.
Keep this manual readily available to answer questions.
Read through each step or procedure to understand it before doing it.
All specifications in this manual are nominal.
Read the User Responsibility statement on the inside front cover of
this manual; it describes what the user must do to maintain this
product.
w
WARNING The INOvent delivery system must only be used in accordance
with the indications, usage, contraindications, warnings and
precautions described in the nitric oxide drug package inserts
and labeling. Drug package inserts and labeling are packaged
with each drug cylinder. Refer to this material prior to use.
The use of devices which radiate high-intensity electrical fields
may affect the operation of the delivery system. Do not drape the
leads of these devices across the delivery system. Constant
surveillance of all monitoring and life support equipment is
mandatory whenever interfering devices are in operation on or
near a patient.
1-2
02/08/00 1605-0014-000
Page 11
Definitions and abbreviations
% v/v:% volume/volume
Button:A pushbutton control used to select a function.
Control wheel:Rotary control used to change and confirm
Display:The electronic information panel on the front of
HFOV:High frequency oscillating ventilator.
Menu:A list of available choices for an operation.
Introduction
settings or actions.
the delivery system.
Menu area:Area in the display where choice selections are
shown.
Message area:Area in the display where information is shown.
N
:Nitrogen.
2
NO:Nitric oxide.
NO
:Nitrogen dioxide.
2
NO/N
:Nitric oxide (NO) and nitrogen (N
2
) gas mixture.
2
Numeric area:Area in the display where values and settings
are shown.
:Oxygen.
O
2
Pop-up box:Menu or message which appears on the display.
1605-0014-00002/08/00
Resolved alarm:An alarm whose cause has been corrected.
Set NO:The concentration of NO delivered by the
system as set by the user.
1-3
Page 12
INOvent delivery system
Symbols used in the manual or on the system
WARNINGS and CAUTIONS tell you about dangerous conditions that
can occur if you do not obey all of the instructions in this manual.
Warnings tell you about a condition that can cause injury to the
operator or the patient.
Cautions tell you about a condition that can cause damage to the
equipment. Read and obey all warnings and cautions.
Other symbols replace words on the equipment or in this manual.
These symbols include:
Product “Use by” information (indicating
warranty expiration date as YYYY-MM).
Replace fuse only as marked.
Indicates operating-on-battery.
European Union Representative
A system with this mark agrees with the
European Council Directive (93/42/EEC)
XXXX
for Medical Devices when it is used as
specified in the Operation and
Maintenance Manual. The xxxx is the
certification number of the Notified Body
used by Datex-Ohmeda’s Quality
Systems.
• The INOvent delivery system delivers nitric oxide therapy gas into
the inspiratory limb of the patient breathing circuit in a way that
provides a constant concentration of NO, as set by the user, to the
patient throughout the inspired breath.
• The INOvent delivery system uses a specially designed Injector
Module which enables the tracking of ventilator waveforms and the
delivery of a synchronized and proportional dose of NO.
• The INOvent delivery system may be used with most ventilators.
• The INOvent delivery system provides continuous on-line integrated
monitoring of delivered O
, NO
, and NO, and a comprehensive
2
2
alarm system.
• The interactive control panel and display provide a simple-tounderstand interface for operation of the INOvent delivery system.
• The battery system provides up to 30 minutes of uninterrupted nitric
oxide delivery in the absence of an external power source.
• The INOvent delivery system provides an integrated Manual NO
delivery system for administration of a fixed concentration of NO/O
therapy with a resuscitator bag.
2
2-2
02/08/00 1605-0014-000
Page 17
Configuration Options
The INOvent delivery system is available in two configuration options:
• Shelf or table option and
• Mounted on a transport cart which holds two NO therapy gas
cylinders.
The options let you choose the setup which meets the needs of the
work environment.
2/General Information
1605-0014-00002/08/002-3
Page 18
INOvent delivery system
Main component views
8
7
6
5
1
2
1. Display and operating panel
2. Manual NO Delivery System Activity Indicator
3. NO/N2 Injector Tube connector
4. Injector Module Electrical Cable connector
5. Fluid Trap Bottle
6. Inspiratory Gas Sample Line connector
7. Inspiratory Gas Sample Line
8. Fluid Trap Filter Cartridge
Figure 2-1 • INOvent delivery system, front view
34
AA.69.034
2-402/08/00 1605-0014-000
Page 19
2/General Information
1
2
3
4
15
14
13
12
11
10
1. NO Sensor
2. NO2 Sensor
3. Nurse Call connector
4. Serial Port connector
5. Power Entry Module
6. Power Cord connector
7. External Ground connection point
8. ON/STANDBY switch
9. Cooling Fan and Filter
10. O2 Sensor
11. Sample gas outlet
12. Regulator Low Pressure Hose with Quick Connector (2)
13. NO/N2 Input Hose Connector (2)
14. Resuscitator Bag O2 Tube connector
15. Connector to the O2 Flowmeter
5
6
7
8
9
AA.69.147
Figure 2-2 • INOvent delivery system, rear view
1605-0014-00002/08/002-5
Page 20
INOvent delivery system
4
1
23
1. NO/N2 injector tube connection to the INOvent delivery system
2. Inspiratory limb breathing circuit connection; to humidifier
3. Inspiratory limb breathing circuit connection; from ventilator
4. Electrical cable connection to the INOvent delivery system
Figure 2-3 • Injector Module, side view
AA.69.266
2-602/08/00 1605-0014-000
Page 21
Display and user controls
The INOvent delivery system display and control panel has a control
wheel, buttons, and an easy-to-view electroluminescent display. All
the INOvent delivery system controls and monitoring are in one place
which reduces the need for extensive visual scanning.
The buttons and control wheel on the INOvent delivery system
control panel perform an assortment of functions and access various
features.
The steps for performing any operation, or accessing any function, on
the INOvent delivery system require the same interaction with the
control panel, namely:
1. Select (press) a button associated with the desired function or
operation,
2. Rotate the control wheel to select a menu item or value,
3. Confirm the selection by pressing either the control wheel or the
function button.
2/General Information
The control panel display, buttons and control wheel are used to:
• Set the concentration of delivered NO,
• Adjust alarms,
• Pause NO delivery,
• Silence alarms,
• Review alarm history,
• Define setup options,
• Calibrate the monitors.
A single tone indicates:
• Your choice of an action with a button or the control wheel,
• An invalid action with a button or the control wheel or
• A message box appearance on the display.
In normal operation, a typical display presents these major sources of
information:
• The Set NO value,
• The measured values of three inspired gases O
• Alarm information.
, NO2 and NO,
2
Menus guide you through the steps needed to deliver NO. Display
and user controls are also discussed in the sections of this manual
where system operations are explained.
1605-0014-00002/08/002-7
Page 22
INOvent delivery system
23
1
45
6
7
14
13
12
11
10
9
8
16
1. Alarm Silence button
2. Alarm Silenced Countdown Timer
3. Alarm message area
4. Alarm History indication are
5. Alarm History button
6. NO Bargraph
7. Control Wheel
8. Set NO button
9. Pause Flow button
10. Inspired NO Measured Value
11. Calibration button
12. Inspired NO2 Measured Value
13. Setup button
14. Inspired O2 Measured Value
15. Alarms button
16. Mains Power light
Figure 2-4 • User controls and indicators on the front
15
AA.69.034
Following are brief descriptions of display and user controls shown in
figure 2-4. See the other areas in the manual for more information.
1. Alarm Silence
button
2. Alarm Silenced
Countdown Timer
2-802/08/00 1605-0014-000
Pushing the Alarm Silence button silences an audible alarm and
starts the alarm silence countdown timer in the upper left location of
figure 2-4 (item 2). The counter shows the remaining silence time. A
new alarm will sound even if the Alarm Silence function has
previously been activated.
The Alarm Silenced Countdown Timer shows a silenced bell and the
remaining silence time when the Alarm Silence button is pushed. The
timer counts down from 120 seconds to zero for high and medium
priority alarms.
Page 23
2/General Information
3. Alarm message areaThe highest priority most recent alarm message appears in the top
part of this alarm message area first. On high priority alarms, the
elapsed-time counter counts up. Alarm messages are also
accompanied by audio signals. Refer to section 7/Alarms for
additional information on alarms.
4. Alarm History
indication area
This normally blank area displays “Alarm History” if there are resolved
alarms. See section 7/Alarms for information on resolved alarms.
5. Alarm History buttonPushing the Alarm History button shows the Alarm History menu
listing the resolved alarms. There is an option to clear the displayed
resolved alarms.
6. NO BargraphThe NO Bargraph shows the Set NO value in ppm.
7. Control WheelThe Control Wheel’s function is defined by the information on the
display. When a menu is displayed, turning the wheel in either
direction selects a choice or adjusts device settings on the menu.
Pressing the wheel then selects the current menu item or confirms the
adjusted setting.
8. Set NO buttonPushing the Set NO button allows adjustment of the Set NO bargraph.
9. Pause Flow buttonPushing the Pause Flow button displays the “Pause Flow” message
box but does not pause the NO flow. To pause NO flow, you must
push the Pause Flow button a second time.
10. Inspired NO
Measured Value
The measured value of inspired NO (in ppm) is displayed in large
numbers. Low and high alarm settings for inspired NO appear below
these numbers.
11. Calibration buttonPushing the Calibration button displays the menu for calibrating the
high and low range of the sensors. Refer to section 4/Calibration for
information on calibration.
12. Inspired NO
Measured Value
The measured value of inspired NO2 in ppm is displayed in large
2
numbers. The high alarm setting for inspired NO2 appears below
these numbers.
13. Setup buttonPushing the Setup button displays the Setup menu for setting the
alarm volume and display brightness or returning to the normal
display. See section 3/Setup and Calibration for additional
information.
1605-0014-00002/08/002-9
Page 24
INOvent delivery system
14. Inspired O2 Measured
Value
The inspired O2 measured value in % is displayed in large numbers.
The low and high alarm settings for inspired O2 appear below these
numbers.
15. Alarms buttonPushing the Alarms button displays the menu for setting the user-
adjustable alarm limits. High limits can be set for O2, NO2 and NO.
Low limits can be set for O2 and NO. See section 7/Alarms for
additional information on alarms.
16. Mains Power lightWhen the green Mains Power light is on, the system is connected to a
functioning mains power outlet.
2-1002/08/00 1605-0014-000
Page 25
2
Theory of Operation
I
S
The INOvent delivery system delivers nitric oxide therapy gas into the
inspiratory limb of the patient breathing circuit in a way that provides a
constant concentration of NO to the patient throughout the inspired
breath.
The INOvent delivery system uses a specially designed Injector
Module which enables the tracking of ventilator waveforms and the
delivery of a synchronized and proportional dose of NO.
2/General Information
NO Flow from the
Cylinder to the
Patient Breathing
Circuit
nspired Gas
ample Port
Cells
Sample
O2
Pump
NO
2
NO
Monitoring
and
Alarm Board
Controls and Display
Cylinder
Controller
Board
High Pressure
Regulator
Pressure
Switch
Quick Connectors
Low Pressure
Regulator
Shut Off Valve
Flow
Controllers
AA.69.121
Injector
Module
Nitric Oxide
Sample Gas
Electronic
Figure 2-5 • NO Flow from the cylinder to the patient breathing circuit
See figure 2-5.
Nitric oxide therapy gas is stored as a gas mixture (Nitric Oxide [NO]
and Nitrogen [N2] balance gas) in an aluminum cylinder at a maximum
pressure of 2200 psi. When the cylinder valve is opened, nitric oxide
therapy gas is regulated down by a high pressure NO regulator.
1605-0014-00002/08/002-11
Page 26
INOvent delivery system
The regulator has a safety relief valve which opens if the regulator
malfunctions. The regulator also has a supply pressure gauge which
indicates the pressure in the cylinder.
Regulator and gauge location varies with the INOvent delivery system
configuration option chosen. Each INOvent delivery system has two
sets of regulators and gauges.
The nitric oxide therapy gas enters the INOvent delivery system
through the NO/N
input hose connectors (quick-connects.) When the
2
gas enters the INOvent delivery system, a pressure switch monitors if
adequate pressure is available.
A low pressure regulator reduces the pressure to a nominal 26 psi.
The output of this regulator supplies a shut-off valve where it waits for
a delivery demand from the user.
When the shut-off valve is opened, NO flows through one of two flow
controllers: a high flow or a low flow controller. The high and low flow
controllers ensure accurate delivery of nitric oxide therapy gas over a
wide range of ventilator flow rates and desired NO concentrations.
Nitric oxide therapy gas is introduced into the inspiratory limb of the
breathing circuit through the INOvent Injector Module. The Injector
Module must be between the ventilator and the humidifier chamber.
Nitrogen dioxide (NO
), an unwanted by-product of mixing NO and O2,
2
is reduced by injecting the nitric oxide therapy gas downstream of the
ventilator.
The Injector Module uses an integrated respiratory flow sensor to
measure ventilator flow rate in the breathing circuit inspiratory limb.
This lets the INOvent delivery system track ventilator waveforms and
inject a synchronized and proportional flow rate of NO. This supplies a
constant concentration to the patient.
The INOvent delivery system “dual-channel” design ensures safe
delivery of nitric oxide therapy gas. One channel has the delivery
components: controller board, flow controllers, and injector module.
The other channel has the monitoring components: monitor and alarm
board, monitor sensors (NO, NO
, and O2 cells) and inspired-gas
2
sample system.
2-1202/08/00 1605-0014-000
Page 27
2/General Information
2
The dual-channel approach to delivery and monitoring permits
delivering NO independent of monitoring if the monitor malfunctions.
The INOvent delivery system also has the safety features of an
integrated system such as automatic shutdown if the NO monitored
value becomes greater than 100 ppm.
When nitric oxide therapy gas delivery is required, the user selects
the desired NO dose level in parts per million (ppm.)
The INOvent delivery system processes both the desired level for NO
delivery as well as the instantaneous patient circuit flow from the
INOvent Injector Module. Using this information, the INOvent delivery
system injects the proportional amount of NO necessary to deliver a
constant concentration throughout the breath.
The INOvent delivery system tracks ventilator waveforms and injects
a synchronized, proportional, flow of NO into the inspiratory limb of
the breathing circuit. This delivers a constant concentration of NO to
the patient, independent of ventilator modes and flow rates as shown
in figure 2-6.
Figure 2-6 • INOvent delivery system injection method: a constant NO
concentration
1605-0014-00002/08/002-13
AA.69.120
Page 28
INOvent delivery system
Effect of inserting the
INOvent into a
breathing circuit
There are two main effects of connecting and using the INOvent
delivery system in a ventilator breathing circuit.
First, the INOvent delivery system adds gas to the breathing circuit in
proportion to the NO setting. For example, at an 80 ppm NO setting
(the maximum NO setting with a 800 ppm NO cylinder concentration),
the INOvent delivery system adds 10% more gas to that delivered by
the ventilator, 5% more for a 40 ppm setting, etc.
Second, the INOvent delivery system subtracts gas from the
breathing circuit via the gas sampling system at a nominal flowrate of
230 mL/min.
These two effects of adding and subtracting gas from the ventilator
breathing circuit have the following effects.
Oxygen DilutionThe INOvent delivery system adds gas to the breathing circuit in
proportion to the NO setting as described above. The gas being
added by the INOvent delivery system is a mixture of NO with the
Volume Control
Ventilation
balance gas being Nitrogen (N
the breathing circuit gas is reduced because of the NO/N2 injection.
This reduction is proportional to the NO setting. At an NO setting of 80
ppm with a cylinder concentration of 800 ppm NO, the added gas is
10%. Thus, the O
value. For example, if the original O2 concentration was 50% v/v, then
the value after injection, at the 80 ppm setting, is 45% v/v.
When using volume control ventilation with the INOvent delivery
system, the measured tidal volume delivered to the patient shows
small changes depending on the NO setting being used. Some minor
ventilator adjustments may be required to the volume setting to
compensate for this addition and subtraction of gases.
For example, assume a respiratory rate of 20 bpm and a tidal volume
of 500 mL. The reduction in tidal volume due to the sampling system
is 230 / 20 bpm. This is approximately 10 mL per breath. The increase
in tidal volume due to an NO setting of 80 ppm with a cylinder
concentration of 800 ppm is 10% of 500 mL. This is 50 mL per breath.
Thus, the total effect on the delivered tidal volume is between -10 mL
to +40 mL for an NO set range between 0 to 80 ppm.
concentration is reduced by 10% of its original
2
.) Thus, the oxygen concentration in
2
2-1402/08/00 1605-0014-000
Page 29
2/General Information
2
Pressure Control
Ventilation
In pressure control ventilation there are no significant effects of the
INOvent delivery system on the pressures in the breathing circuit and
hence to the tidal volume being delivered to the patient. This is
because the ventilator actively controls the pressure in the breathing
circuit and makes adjustments to compensate for the effect of the
INOvent delivery system.
However, if a ventilator has inspired tidal volume measurement which
is monitored inside the ventilator, this measurement reflects the
change in the ventilator gas flow necessary because of the effect of
the INOvent delivery system.
For example, assuming a respiratory rate of 20 bpm and a tidal
volume of 500 mL, the net effect on the inspired tidal volume is -40 mL
to +10 mL for an NO set range of 0 to 80 ppm.
Trigger SensitivityThe addition and subtraction of gases by the INOvent delivery system
may effect the trigger sensitivity of the ventilator when using
synchronize modes of ventilation.
In ventilators which have flow trigger modes where the trigger flow is
set to less than 1 L/min, this may cause the ventilator to auto-trigger.
The trigger sensitivity of the ventilator should be checked after
connecting the INOvent delivery system.
High Frequency
Oscillatory Ventilator
Circle Anesthesia
Ventilator Systems
The effect of the INOvent delivery system on a High Frequency
Oscillatory Ventilator (HFOV) is minor changes to the bias flow
because of the gas sampling system and the addition of gas flow in
proportion to the NO setting.
For example, with a bias flow of 10 L/min, the reduction in bias flow
due to the sampling system is 0.23 L/min. The increase in bias flow
due to an NO setting of 80 ppm with a cylinder concentration of 800
ppm is 10% of 10 L/min which is 1 L/min. Thus, the bias flow will vary
between 9.77 and 10.77 for an NO set range between 0 to 80 ppm.
The effect of the INOvent delivery system on circle anesthesia
ventilator systems which use volume control ventilation is to cause
small changes in the tidal volume being delivered to the patient
depending on the NO setting being used. Minor changes to the
volume setting on the ventilator may be necessary to compensate for
this addition and subtraction of gases.
For example, assume a respiratory rate of 10 bpm and a tidal volume
of 500 mL.
1605-0014-00002/08/002-15
Page 30
INOvent delivery system
The reduction in tidal volume caused by the sampling system is 230 /
10 bpm. This is approximately 20 mL per breath. The increase in tidal
volume caused by an NO setting of 40 ppm with a cylinder
concentration of 800 ppm is 10% of 500 mL. This is 50 mL per breath.
Thus, the total effect on the delivered tidal volume is between -20 mL
to +30 mL for an NO set range between 0 to 80 ppm.
Recirculation of gases in circle breathing systems should be avoided
to ensure that the accumulated gases in the ventilator bellows are not
introduced into the inspiratory limb of the breathing circuit. The gas in
the ventilator bellows may contain undesirable levels of NO
and other
2
reaction products that may not be removed by the absorbent when it
contains appreciable moisture as might occur in typical use.
Recirculation of gases can be avoided by using fresh gas flowrates
equal to a patient’s minute volume. This can be explained as follows.
During the expiratory phase of the breathing cycle, the fresh gas from
the common gas outlet enters the absorber and flows in the reverse
direction displacing some of the expired gas that is present.
During the next inspiratory phase, the accumulated gas in the
ventilator bellows displaces the fresh gas in the absorber, pushing it
into the inspiratory limb where it mixes with the fresh gas from the
common gas outlet and flows to the patient.
As long as the fresh gas flow is equal to or greater than the patient’s
minute volume, there is sufficient fresh gas in the absorber such that
no accumulated gas from the ventilator bellows reaches the
inspiratory limb of the breathing circuit and hence to the patient.
Maximum NO DeliveryThe INOvent delivery system is limited to a maximum NO flow of 6.2
L/min. This means the maximum deliverable NO concentration will
vary based on the ventilator flow rate.
Figure 2-7 illustrates the maximum deliverable NO concentration
versus constant inspiratory flow rate (for an 800 ppm cylinder
concentration). The graph depicts the decrease in maximum
deliverable concentration by the INOvent delivery system as a
function of constant inspiratory flow rate.
2-1602/08/00 1605-0014-000
Page 31
2/General Information
2
The INOvent delivery system will deliver a maximum concentration of
80 ppm (using 800 ppm cylinder concentration) up to an inspiratory
flow rate of 55 L/min. As inspiratory ventilator flows increase, the
maximum deliverable NO concentration decreases gradually until the
inspiratory flow rate reaches 120 L/min. The maximum deliverable NO
concentration at 120 L/min (using 800 ppm cylinder concentration) is
40 ppm. The NO flow rate for a constant inspiratory flow rate can be
calculated from the following formula:
(NO ) x (Fresh gas flow rate)
set
(Cylinder concentration) - (NO )
set
When intermittent inspiratory flow rates are used, peak ventilator
flows may be attained which exceed the INOvent flow rate
specification of 120 L/min. Peak inspiratory flow rates are transient
and extremely short in duration. As a result, the portion of the breath
which is not matched by the INOvent delivery system is extremely
small and virtually negligible with respect to the entire range of the
breath.
80
1
40
200406080100120
2
1. Maximum deliverable NO concentration (ppm)
2. Constant inspiratory flowrate (L/min)
AA.69.193
Figure 2-7 • Maximum deliverable NO concentration vs. constant
Inspiratory flow rate
1605-0014-00002/08/002-17
Page 32
INOvent delivery system
Notes:
2-1802/08/00 1605-0014-000
Page 33
3
3/Setup
In this sectionThis section provides connection information for the INOvent delivery system.
To set the alarm loudness . . . . . . . . . . . . . . . . . . . . . . 3-10
To set the display brightness . . . . . . . . . . . . . . . . . . . . 3-11
1605-0014-00002/08/003-1
Page 34
INOvent delivery system
INOvent delivery system connections
See figures 2-1 or 2-2 if you need to identify additional parts.
Mounting
Refer to the following mounting instructions:
• when first using the system or
• when installing a replacement system.
Cart mountingIf the optional transport cart is used, the INOvent delivery system
must be properly mounted and secured to the cart before making
system connections. See figure 3-1.
1. Place the INOvent delivery system on the cart top shelf with the four
mounting feet inserted into the four mounting slots.
• Make sure the INOvent delivery system is pushed to the back and
is against the shelf backstop.
2. Tighten the mounting screw at the back of the cart top shelf to
attach the INOvent delivery system.
1
3
2
1. System mounting feet (4)
2. Cart top shelf
3. Mounting slots on cart top shelf (4)
4. Shelf backstop
5. Mounting screw
Figure 3-1 • Cart mounting
3-202/08/00 1605-0014-000
4
5
AA.69.034
AA.69.037
Page 35
3/Setup
Shelf or table mountingPlace the INOvent delivery system on the shelf or table where it will
be used.
• Allow safe and convenient routing for nitric oxide and therapy gas
hoses.
System connections
ImportantTo remove this type of hose connector, pull the widest part of the
connection away from the back of the system until the hose connection
releases. You can then pull the hose connector out.
1. Remove all protective caps from the connectors and ports on the
INOvent delivery system.
See figure 3-2.
2. Connect the low pressure NO/N
hose from the regulator to the
2
back of the INOvent delivery system by inserting until the red sleeve
pops into place.
1
AA.69.147
3
2
4
5
6
1. Regulated low pressure NO/N
hose connectors (2)
2
2. Cylinder gauge boot (2)
3. Regulator low pressure NO/N
hoses (2)
2
4. Regulator high pressure hoses (2)
5. Gas cylinder (2)
6. Dovetail rail (2)
Figure 3-2 • System connections
1605-0014-00002/08/003-3
AA.69.037
Page 36
INOvent delivery system
wwwwCAUTIONMake sure the cart is stable after the accessories are attached.
3. Check the therapy gas cylinders for the correct product identity
labels and NO concentration. The NO concentration must match
that shown in the Setup menu for Cylinder Concentration. For
additional cylinder information see section 13/Appendix B Cylinder Information.
4. If necessary, attach the cylinder gauge boots to the gauges on the
front of the cart by pressing the boots into the gauge openings.
Gauge boots are shown in figure 3-2.
5. If a cart is used, lock accessory supports into the dovetail rail before
attaching accessories. Place heavier accessories near the base of
the cart.
• Make sure that any load is evenly distributed to help maximize
stability when the system is moved.
6. Place the cart where it will be used and push down on the wheel
locks with your foot to lock the casters. To unlock a caster, push
down on the opposite end of the wheel lock.
wwwwCAUTIONDo not lean on, or lean anything against, the cart.
7. Connect and tighten the fittings on the high pressure hose from the
regulator to the NO therapy gas supplies. Refer to figure 3-3.
a. Make sure the hose tip or seal is in place and not damaged
when connecting to the cylinders.
b. Note: over-tightening a fitting can reduce the useful life of the tip.
• Section 13/Appendix B - Cylinder Information has information on
cylinders and leak checks.
1
2
1. High pressure hose fitting (CGA type)
2. High pressure hose fitting tip
3. High pressure hose from the regulator
4. Therapy gas cylinder
3
1
4
AA.69.132, AA.69.141
Figure 3-3 • Cylinder connections
3-402/08/00 1605-0014-000
Page 37
3/Setup
8. Make sure two NO gas cylinders, with more than 200 psi in each
cylinder, are available.
9. Empty and clean the fluid trap bottle (figure 3-4) before each patient
and whenever the trap is more than half full. See “Emptying the
Fluid Trap” in section 6/Maintenance.
10.Connect the inspiratory gas sample line (figure 3-4) to the
inspiratory gas sample line connector.
2
1
AA.69.034
1. Fluid trap bottle
2. Inspiratory gas sample line connector
Figure 3-4 • Fluid trap bottle and inspiratory gas sample line connector
11.Check that the three sensors (NO, NO
and O2) are installed on the
2
back of the INOvent delivery system. The O2 sensor has a short
cable which must be plugged into the sensor and into the back of
the system. See section 9/Maintenance for sensor installation
information.
2
1
3
4
1. NO sensor
sensor
2. NO
2
sensor
3. O
2
4. Short cable for O2 sensor
AA.69.147
Figure 3-5 • Sensor items
1605-0014-00002/08/003-5
Page 38
INOvent delivery system
12.Connect either end of the Injector Module electrical cable to the
Injector Module as shown in figure 3-6.
• Align the connector before pushing the connector firmly into place.
• A keyway is identified by a red dot on both the connector and the
Injector Module.
13.Connect the NO/N
Injector Tube to the Injector Module. See figure
2
3-6.
ImportantTo remove this type of electrical connector, the knurled sleeve on the
connector must be pulled outward before removing the connector from
the Injector Module or the front panel.
21
1. Electrical cable to the delivery system
2. NO/N
Injector Tube to the delivery system
2
Figure 3-6 • Injector Module connections
14.Connect the other end of the Injector Module electrical cable to the
front of the INOvent delivery system.
• A keyway is identified by a red dot on the connector and on the
front panel.
15. Connect the other end of the NO/N
Injector Tube to the front of the
2
INOvent delivery system.
AA.69.024
16.Tie the NO/N
Injector Tube to the Injector Module electrical cable
2
using hose clips as shown in figure 3-7.
2
1
3
1. Injector Module Electrical Cable
2. Hose Clip
3. NO/N
Injector Tube
2
Figure 3-7 • Hose clip connection
3-602/08/00 1605-0014-000
AA.69.034
Page 39
3/Setup
Note: The Injector Module is inserted later into the inspiratory limb to
measure the ventilator flow rate and inject the therapy gas.
Figure 3-8 shows a typical system connection diagram.
54321
6
1514131211109
1. Patient wye
2. Ventilator
3. Ventilator Expiratory port
4. Ventilator Inspiratory port
5. Patient gas sample line input connection
6. INOvent delivery system
7. NO/N
Injector Tube front panel connection
2
8. Injector Module Cable front panel connection
9. Injector Module Electrical Cable connection
10.NO/N2 Injector Tube connection
11.Humidifier inlet
12.Humidifier
13.Humidifier outlet
14.Patient Gas sample line connection
15.Sample Tee
7
8
AA.69.017
Figure 3-8 • Typical system connection diagram (ICU ventilator)
1605-0014-00002/08/003-7
Page 40
INOvent delivery system
Electrical connections
1. If you are using a nurse call system, refer to your nurse call system
operation manual for the proper connection procedure. Connection
is to the Nurse Call connector on the back of the INOvent delivery
system.
• See section 12/Appendix A - Specifications for Nurse Call
connector information.
wwwwCAUTIONThe correct AC line voltage for your local area must appear in the Power
Entry Module window on the back.
2. Check that the correct AC line voltage appears in the Power Entry
Module window on the back: 100V, 120V, 220V or 240V, whichever
is appropriate for your local voltage.
• If the voltage shown is not correct, see “Fuse Replacement and
Line Voltage Selection” in section 10/Maintenance to set the
voltage selection to the correct local voltage.
1
1. Power Entry Module window
Figure 3-9 • Power Entry Module window
3. Connect the power cord to the INOvent delivery system (see figure
3-10) and tighten the power cord clamp screws using a #1 Phillips
screwdriver.
1
1
2
AA.69.267
AA.69.267
1. Power cord clamp screws
2. Power cord
Figure 3-10 • Power cord and clamp
3-802/08/00 1605-0014-000
Page 41
3/Setup
4. Connect the INOvent delivery system power cord to an emergencypower-backed hospital-grade outlet.
• The front panel green Mains Power light must light to indicate that
line voltage is present.
• The power cord must always be connected to an electrical outlet
to maintain a full battery charge.
5. Switch the ON/STANDBY control to ON. (The system may take a
minute before operation.)
• The Power-ON display (figure 3-11) appears during self-tests
followed by a typical display as shown in figure 3-12.
Figure 3-11 • Power-ON display
Figure 3-12 • Typical display
SCR_1.tif
510k_2.tif
1605-0014-00002/08/003-9
Page 42
INOvent delivery system
Setup menu
The Setup menu lets you change alarm volume and display
brightness. You can view, but not change, the software revision, NO
cylinder concentration configuration and hourmeter.
Push the Setup button to see the Setup menu (shown in figure 3-13).
Cylinder
concentration
Hourmeter display
To set the alarm
loudness
The INOvent delivery system is factory-set for an 800 ppm cylinder
concentration. This set concentration is indicated in the Setup menu.
Alternate cylinder concentrations may be available by a special
request to Datex-Ohmeda. See 17/Appendix F.
When connecting an NO cylinder to the INOvent delivery system,
always check to see it is the same concentration for which the system
is configured as appears in the setup menu.
The Power-ON display (figure 3-11) and the Setup menu (figure 3-13)
provide an hourmeter indication which can be used for maintenance
and service purposes. The hourmeter counts the hours when the ON/
STANDBY switch is in the ON position.
1. Push the Setup button to see the Setup menu.
2. Turn the control wheel to highlight the Alarm Loudness value
(shown in figure 3-13.)
Figure 3-13 • Setup menu - Alarm Loudness level highlighted
3. Push the control wheel to confirm the Alarm Loudness level
selection. Figure 3-14 shows the same level in a selection box after
pushing the control wheel.
3-1002/08/00 1605-0014-000
510k_9.tif
Page 43
3/Setup
3
510k_10.tif
To set the display
brightness
Figure 3-14 • Select menu - Alarm Loudness level selected for
adjustment
4. Turn the control wheel to indicate the loudness level you want.
Choices range from 1 (softest) to 5 (loudest.)
5. Push the control wheel to confirm your selection.
6. If you are finished with the Setup menu, turn the control wheel to
select Exit to Normal Display.
• If you want to set the display brightness, continue with step 2 of
“To set the display brightness.”
7. Push the control wheel to exit.
1. Push the Setup button.
2. Turn the control wheel to highlight the Display Brightness level on
the Setup menu.
3. Push the control wheel to confirm the display brightness level for
change.
4. Turn the control wheel to show the level you want. Choices range
from 1 (darkest) to 10 (brightest.)
5. Push the control wheel to confirm your selection.
6. If you are finished with the Setup menu, turn the control wheel to
select Exit to Normal Display. Push the control wheel to exit.
1605-0014-00002/08/003-11
Page 44
INOvent delivery system
Notes:
3-1202/08/00 1605-0014-000
Page 45
4
4/Calibration
In this sectionThis section describes how to calibrate the NO, NO
calibration is needed. For example:
• after replacing a sensor,
• during Pre-Use Procedures,
• Monthly System Checkout or
• when messages ask you to do so (i.e., “Calibration Due”.)
wwwwCAUTIONCalibrate only with the correct calibration gas.
Make sure that the calibration gas expiration date is still
valid.
Calibrate the sensors in the environment in which they will
be used: for example, in the ICU.
NO therapy can be administered during the sensor calibration process
but inspired gases are not monitored and gas monitoring alarms are
disabled.
The ventilator settings must not be changed during sensor calibration
or an alarm condition may occur.
You calibrate the low ranges of the three sensors at the same time
using room air.
The high ranges must be calibrated using their respective NO and
NO
calibration gases and O2 using a medical grade 100% O2 from a
2
wall or cylinder source.
The calibration process can be stopped at anytime by pushing the
control wheel to cancel an operation or to exit to a normal display.
Also, calibration can be stopped by selecting another control button
(for example, Pause Flow.)
If calibrations fail, see the 9/Troubleshooting section in this manual.
4-202/08/00 1605-0014-000
Page 47
4/Calibration
NO, NO2 and O2 Sensor low range calibration (daily,
room air)
Procedure
Use the Calibration menus to calibrate the low range of the three gas
sensors (NO, NO2 and O2) at the same time using room air; the
calibration may take up to five minutes.
NO, NO2 and O2 room air levels are displayed during low range
calibration and trend toward 0 ppm, 0 ppm and 21% respectively.
1. From a normal display with the INOvent delivery system operating,
push the Calibration button to reach the first Calibration menu. It
shows the Low Range (room air) “NO, NO
highlighted (figure 4-1).
and O2” option
2
Figure 4-1 • Calibration menu - low range sensor calibration option
2. Push the control wheel to start the low range calibration. All three
sensors calibrate automatically at the same time. The display in
figure 4-2 appears during the calibration. Calibration may take a
few minutes.
• The sample line does not have to be disconnected to do this step.
The system draws in room air from within the system enclosure
through an NO/NO
highlighted
2
scrubber.
510k_11scr_12
Figure 4-2 • Calibration menu - low range sensor calibration in
1605-0014-00002/08/004-3
progress
Page 48
INOvent delivery system
3. When the calibration is successful, the display in figure 4-3
appears.
• If calibration is unsuccessful for any of the sensors, an appropriate
message as in figure 4-4 is displayed. In this case follow the
instructions on the display.
• You may turn the control wheel to “Exit to Normal Display” or
continue on to “High range calibration (calibration gas.)”
510k_13510k_14
Figure 4-3 • Calibration menu - low range calibration completed
Figure 4-4 • Calibration menu - NO low range calibration failed
4. To continue on to “High range calibration (calibration gas),” turn the
control wheel to select “Go to Calibration of Sensors.”
5. Push the control wheel to accept “Go to Calibration of Sensors” and
continue with the following steps.
4-402/08/00 1605-0014-000
Page 49
NO, NO2 and O2 Sensor high range calibration
wwwwCAUTIONNever connect the sample line directly to a high pressure source. This
could damage the sampling system.
4/Calibration
Procedure
Use the Calibration menus to calibrate the high range of the NO, NO
and O2 sensors.
Procedures for high range calibration of the three sensors are
essentially the same as for the low range, but a different calibration
gas is used for each sensor. Calibration must be done individually:
• NO and NO
each with their respective NO and NO2 calibration
2
gases,
•O
with a medical grade 100% O2 from a wall or cylinder source.
2
Note: Contact the drug supplier for NO and NO
calibration gases.
2
1. To calibrate the NO or NO2 sensor high range: connect the
calibration gas for the sensor you are calibrating as shown in figure
4-5. When calibrating NO, connect NO calibration gas; when
calibrating NO
, connect NO2 calibration gas. Then, go to step 3.
2
2
3
1
2
4
6
5
1. Calibration gas cylinder pressure gauge
2. Sample Line
3. One-way valve
4. Vent to atmosphere or scavenge - Do not occlude
5. Cylinder ON/OFF control
6. NO or NO
Figure 4-5 • Connection diagram for NO and NO
1605-0014-00002/08/004-5
calibration gas cylinder
2
calibration
high range
2
AA.69.110
Page 50
INOvent delivery system
2. To calibrate the O2 sensor high range, connect the O2 calibration
gas as shown in figure 4-6. Then, go to step 3.
O2 to the
4
123
100% O
2
FlowmeterSample Tee
Sample System
5
To the
6
Atmosphere
AA.69.260
1. 100% O
2
2. Flowmeter
3. Sample Tee
4. O2 to the Sample System
5. INOvent delivery system Sample Line
6. To the atmosphere
Figure 4-6 • Connection diagram for O
high range calibration
2
3. On the Calibration menu) turn the control wheel to highlight the NO,
or O2 selection of “High Range (cal. gas).” Example shown in
NO
2
figure 4-7.
Figure 4-7 • Calibration menu - high range NO sensor calibration
option highlighted
4. Make sure that the calibration gas cylinder ON/OFF control is
turned OFF.
5. Attach the system sample line to the calibration gas source.
6. Turn the calibration gas source ON.
4-602/08/00 1605-0014-000
scr_11b
Page 51
4/Calibration
7. Push the control wheel to start the high range calibration of the
selected sensor. A display similar to figure 4-8 appears.
8. Follow the instructions given on the display. These instructions are
shown here:
• 1. Flow NO [NO
or 100% O2] calibration gas into the sample line.
2
• 2. Wait for “Measured NO, NO2 or 100% O2” to stabilize.
(Note: The “Measured” value is the monitored value
of the calibration gas during sampling. This value
updates continuously during the calibration process.
After the “Measured” value has stabilized, push the
control wheel.)
• 3. Set “Calibration XX” to cal. gas value.
(Select “Calibration XX.” Then, turn the control
wheel to set the value to that marked on the
calibration gas cylinder label [or 100% for
4-8 shows the “Calibration NO” value set to match a
calibration gas cylinder label value of 45 ppm.) Then,
push the control wheel to enter. “In Progress...” will
be shown on the Calibration NO line as in figure 4-9.
O
.] Figure
2
510k_15510k_16
Figure 4-8 • High range calibration menu example - Calibration NO
adjusted to 45 ppm
Figure 4-9 • High range calibration menu example - Calibration NO “In
Progress” highlighted
1605-0014-00002/08/004-7
Page 52
INOvent delivery system
9. If the high range calibration fails, follow the instructions given on
the display (shown in figure 4-10).
510k_17
Figure 4-10 • High range calibration menu example - NO high range
calibration failed
10.When the current sensor calibration is finished, select “Go to
Calibration of Sensors” and push the control wheel.
11.When the “Calibration of Sensors” display appears, select the next
sensor for calibration.
If you are finished with all three sensor calibrations:
12.Select “Exit to Normal Display” and push the control wheel.
13.Turn the calibration gas cylinder ON/OFF control to OFF.
14.Reconnect the sample line to the breathing circuit.
4-802/08/00 1605-0014-000
Page 53
5
5/Pre-Use Procedures
ImportantThe Pre-Use Procedures consist of three tests
which must be done before delivering NO
therapy to a patient:
1. Initial connections and leak test.
2. System purge and performance test.
3. Manual NO delivery system purge and performance test.
These Pre-Use Procedures are for 800 ppm concentration
configurations only.
In this section1. Initial connections and leak test . . . . . . . . . . . . . . . . . . . . . 5-2
2. System purge and performance test . . . . . . . . . . . . . . . . . . 5-3
3. Manual NO delivery system purge and performance test . . 5-5
1605-0014-00002/08/005-1
Page 54
INOvent delivery system
1. Initial connections and leak test
1. If necessary, connect the INOvent delivery system as described in
section 3/Setup.
2. Check the cables and hoses for signs of wear or damage and
replace as needed.
3. Turn the INOvent delivery system ON and confirm that the buzzer
and the speaker sound. Wait for the start-up routine to finish. (The
Power-ON display appears during self-tests followed by a normal
display.)
4. Confirm from your records that a Monthly System Checkout has
been done in the last month. If it hasn’t, do a Monthly System
Checkout. See section 10/Maintenance.
5. Do a system high pressure leak test.
a. With the INOvent delivery system ON, turn each cylinder valve
ON then OFF. Check for adequate cylinder pressures. Wait for
30 seconds and check for pressure decrease.
b. If there is any decrease, make sure that the auxiliary oxygen
flowmeter to the Manual NO System is not turned on and check
for leaks around the hose connections using soapy water.
c. Check for leaks at the NO cylinder valve outlet connection using
soapy water.
Note: The pressure drop leak test will not detect
leaks at the cylinder valve outlet connection because
of the check valve in the regulator hose. See 13/
Appendix B - Cylinder Information if you suspect a
leak.
5-202/08/00 1605-0014-000
Page 55
2. System purge and performance test
5/Pre-Use Procedures
wwwwWARNINGMake sure a purge has been completed within five minutes before the
start of NO therapy. If not, repeat the performance test as described
here, making sure the NO
is less than the value listed for the cylinder
2
concentration being used.
1. Connect the Injector Module to an O
auxiliary flowmeter using O2
2
tubing, a 4.5 mm to 15 mm adapter and a breathing circuit adapter
to condition the flow before the Injector Module. See figure 5-1.
2. Connect a sample tee to the outlet of the injector module with 300
mm of 22 mm hose to ensure gas mixing prior to gas sampling (see
figure 5-1.) Attach a sampling line from the sample tee to the
sample connector on the front panel.
910
1
8
2
34567
1O
Flowmeter
2
AA.69.032
2O2 Tubing
315M x 4.5 mm Adapter
422M / 15F x 22M / 15F Adapter
5Injector Module
6300 mm of 22 mm hose
7Sample Tee
8Inspiratory Gas Sample Line
9NO/N
Injector Tube
2
10 Injector Module Electrical Cable
Figure 5-1 • Circuit connections for purge and performance test
1605-0014-00002/08/005-3
Page 56
INOvent delivery system
3. Do a low range calibration of the NO, NO2 and O2 monitors. (See
section 4/Calibration.)
4. Make sure both NO cylinders are turned off after being turned on.
5. Set the oxygen flow to 15 L/min and the Set NO concentration to its
maximum setting.
6. Make sure both cylinder high pressure gauges go to zero to purge
both high pressure circuits.
7. Check that the NO
NO
monitor is functional and
2
value is higher than 0.2 ppm to verify that the
2
8. Make sure these two alarms occur:
• Low NO/N2 Pressure”
• Delivery Failure”
(This may take a few minutes depending on the
cylinder pressures at the start of the test.)
9. When the alarms have occurred, turn on the NO cylinder with which
you intend to start therapy. Leave the other cylinder turned off.
10.Set the O2 flow on the auxiliary flowmeter to 15 L/min and the Set
NO to 40 ppm.
a. Wait for 3 minutes or until the monitor readings are stable.
b. Make sure that the O
, NO2 and NO readings are within the
2
acceptable ranges given below.
Note: These acceptable readings include the errors
of the NO delivery system and the NO therapy gas.
O2 %v/v
(±3 %v/v)
NO2 ppm
(max)
NO ppm
(min/max)
Set NO = 40 ppm:951.532/48
11.Do the monitor high range calibrations if required.
5-402/08/00 1605-0014-000
Page 57
5/Pre-Use Procedures
3. Manual NO delivery system purge and performance
test
wwwwWARNINGThe Manual NO delivery system must be set up and available at all times.
The Manual NO delivery system must be configured as shown in figure
8-2.
Do not connect an O
the rear of the INOvent delivery system.
wwwwCAUTIONSupplemental O
delivery system is in use.
1. Connect the Manual NO delivery system as described in section 3/
Setup.
2. Disconnect the resuscitator bag. Connect the 22 mm / 15 mm
sample tee (with the sample line connected) to the oxygen tubing
using a 15M x 4.5 mm adapter. See figure 5-2.
3
2
1
supply, wall or cylinder, directly to the O2 inlet on
2
should only be turned on and used when the manual
2
4
5
}
6
987
1 Inspiratory Gas Sample Line
2 Oxygen tubing (user supplied)
3 NO/O
4O
5O
6 Pressure Compensated O
7 NO/N
8 15M x 4.5 mm Adapter
9 Sample Tee
outlet
2
inlet
2
supply hose from the O2 flowmeter
2
flowmeter (user supplied), shown set to OFF
(no float activity)
input hoses from the regulators
2
2
Figure 5-2 • Sample tee connection to the Manual NO delivery system
1605-0014-00002/08/005-5
AA.69.160
Page 58
INOvent delivery system
3. Flow 15 L/min of O2 from the auxiliary O2 flowmeter into the Manual
delivery system and make sure the float moves to the middle of the
NO Flow Indicator Window.
4. Wait for the gas to flow through the oxygen tubing. Make sure that
the NO reading is 20 ± 8 ppm and the NO2 reading is less than 1.0
ppm. If the NO2 reading is greater than 1.0, continue flowing gas
until the limit is reached.
Note: These acceptable readings include the errors
of the NO delivery system and the NO therapy gas.
5. Reduce the O2 flow to 1 L/min and make sure that the float drops to
the bottom of the NO Flow Indicator.
6. Set the O
4.5 mm adapter from the oxygen tubing.
7. Reconnect the resuscitator bag to the oxygen tubing.
flow to zero and disconnect the sample tee and 15M x
2
- End of the Pre-Use Procedures -
5-602/08/00 1605-0014-000
Page 59
6
6/Operation
wwwwWARNINGUse only “Latex-Free” breathing circuits and ventilators when using the
INOvent delivery system.
The use of devices which radiate high-intensity electrical fields may
affect the operation of the delivery system. Do not drape the leads of
these devices across the delivery system. Constant surveillance of all
monitoring and life support equipment is mandatory whenever
interfering devices are in operation on or near a patient.
Make sure a purge has been completed within five minutes before the
start of NO therapy. If not, repeat the performance test as described in
section 5/Pre-Use Procedures making sure the NO
listed for the cylinder concentration being used.
Set the INOvent delivery system alarm thresholds for the current patient
conditions to monitor any inadvertent changes in treatment. For alarm
information, see section 7/Alarms.
Be certain all cables and hoses are positioned to help prevent damaging
or occluding them.
Patient disconnect and high pressure alarms are required for the
ventilator.
is less than the value
2
wwwwCAUTIONUse distilled water in the humidifier to prevent the formation of bases or
acids.
In this sectionThis section 6 provides information for connecting and operating the INOvent
1. Before operating the INOvent delivery system in the ventilator
patient breathing circuit, make sure the Pre-Use Procedures of
section 5 have been completed. The procedures include:
• Initial connections and leak test.
• System purge and performance test.
• Manual NO delivery purge and performance test.
2. Make sure the NO therapy gas cylinder pressure gauge shows an
adequate cylinder supply pressure.The cylinder must be replaced
when its pressure is less than 200 psig.
• To change the cylinder, see “Changing the NO therapy cylinder
and purging the regulator assembly” at the end of this section.
3. Setup the ventilator. See the ventilator manual for its setup and
operation.
4. Do a breathing circuit test as described in the ventilator’s operation
manual.
5. Empty and clean the fluid trap. See section 10/Maintenance for the
procedure.
AA.69.121
AA.69.034
6-2
Connect the INOvent delivery system to the patient circuit. Use one of
the following the connection diagrams for the patient breathing circuit
you are using and follow the steps shown.
02/08/00 1605-0014-000
Page 61
6/Operation
Connection diagrams in this section include the following:
Connection to a high frequency oscillatory ventilator circuit:
Figure 6-5System connection diagram, high frequency oscillatory
ventilator circuit
Connection to a circle anesthesia system:
Figure 6-6System connection diagram, anesthesia system with
ventilator
Connection to a transport ventilator circuit:
Figure 6-7System connection diagram for a transport ventilator
circuit
6-3
1605-0014-00002/08/00
Connection to bagging systems:
Figure 6-8• System connection diagram for a hyper-inflation
bag
Figure 6-9• System connection diagram for a self-filling
resuscitator bag
Figure 6-10• System connection diagram for a spontaneous
breathing circuit
6-3
Page 62
6-4
INOvent delivery system
Connection to an ICU
ventilator circuit
ww
ww
WARNING
ww
ww
CAUTION
The INOvent delivery system subtracts gas from the breathing circuit via
Note the airflow direction arrow on the Injector Module: flow out of the
the gas sampling system; this can effect the sensitivity of a flow
triggered synchronized breath mode of some ventilators. The trigger
sensitivity of the ventilator should be checked after connecting the
INOvent delivery system to the breathing circuit
The humidifier chamber volume should not be more than 480 mL to help
prevent elevated NO
ventilator must be through the Injector Module in the direction of the
arrow on the module.
values.
2
The Injector Module should only be used in the dry part of the breathing
circuit to make sure flow measurement is correct.
To condition ventilator flow and make sure flow measurements are
accurate, connect the Injector Module to the humidifier chamber; then,
connect to the ventilator inspiratory port using breathing circuit tubing.
This can also be done by placing a breathing circuit filter between the
Injector Module and the ventilator.
To avoid medications interfering with the gas monitoring system,
administer surfactants or other medications only on the patient side of
the sample tee.
Procedure:
• When connecting breathing tubing, make sure it is long enough for
any required equipment movement.
• Make sure that the system is not wheeled over tubing connected to
a patient or to the equipment.
Refer to the connection diagrams in figures 6-1 through 6-4 for the
following steps.
6-4
1. Connect a breathing hose from the ventilator inspiratory port to the
Injector Module
2. Connect a breathing hose or adapter from the output end of the
Injector Module to the humidifier inlet port.
3. Connect a breathing hose from the humidifier outlet port to a
sample tee.
input end using the appropriate adapters.
02/08/00 1605-0014-000
Page 63
6/Operation
54321
6
1514131211109
Note: The above connection diagram
is also printed on the left side of the
INOvent delivery system.
a
b
7
8
AA.69.017
AA.69.266
c
6-5
a. Injector Module
b. Input end
c. Output end
1. Patient wye
2. Ventilator
3. Ventilator Expiratory port
4. Ventilator Inspiratory port
5. Patient Gas sample line input connection
6. INOvent delivery system
7. NO/N
Injector Tube front panel connection
2
8. Injector Module Electrical Cable front panel connection
9. Injector Module Electrical Cable connection
10.Injector Module NO/N
Injector Tube connection
2
11.Humidifier inlet
12.Humidifier
13.Humidifier outlet
14.Patient Gas sample line connection
15.Sample Tee
Figure 6-1 •
System connection diagram, ICU ventilator
1605-0014-00002/08/00
6-5
Page 64
INOvent delivery system
4. Connect a breathing hose from the sample tee to the patient wye.
Important This breathing hose between the sample tee and the
patient wye should be between 6 and 12 inches (150
to 300 mm) long: greater than 6 inches to minimize
the sampling of mixed inspired/expired
concentrations, less than 12 inches to help ensure
correct patient NO
measurement.
2
6-6
5. Connect the gas sample line from the “Sample Inlet” port on the
front of the fluid trap to the sample tee. Make sure the sample port
at the T-piece is pointing upward to help avoid fluid accumulation in
the sample line.
6. Connect a breathing hose from the ventilator expiratory port to the
expiratory limb of the patient wye.
7. The trigger sensitivity of the ventilator should be checked after
connecting the INOvent delivery system to the breathing circuit.
8. Proceed to “Operation” - “Setting the delivered NO concentration.”
AA.69.281
6-6
1. To INOvent delivery system
Figure 6-2 •
Adult circuit connection diagram
1
02/08/00 1605-0014-000
Page 65
6/Operation
AA.69.280
1
6-7
1. To INOvent delivery system
Figure 6-3 •
Neonatal circuit connection diagram
AA.69.282
1
1605-0014-00002/08/00
1. To INOvent delivery system
Figure 6-4 •
Pediatric circuit connection diagram
6-7
Page 66
6-8
INOvent delivery system
Connection to a
high frequency
oscillatory
ventilator circuit
ww
ww
WARNING
ww
ww
CAUTION
Omission of the one-way valve may result in high NO delivery.
Note the airflow direction arrow on the one-way valve. The flow
out of the Injector Module must be through the one-way valve in
the direction of the arrow on the valve.
Note the airflow direction arrow on the Injector Module; the flow
out of the ventilator must be through the Injector Module in the
direction of the arrow on the module.
The Injector Module should only be used in the dry part of the
breathing circuit to make sure flow measurement is correct.
To avoid medications interfering with the gas monitoring
system, administer surfactants or other medications only on the
patient side of the sample tee.
• When connecting breathing tubing, make sure it is long enough for
any required equipment movement.
• Make sure that the system is not wheeled over tubing connected to
a patient or to the equipment.
Procedure:
Refer to the connection diagram in figure 6-5 for the following steps.
1. Connect tubing from the ventilator outlet to the INOvent delivery
system Injector Module input end using the appropriate adapters
2. Connect the output end of the Injector Module to the input end of a
one-way valve.
Important: Note the airflow direction arrow on the one-way
valve. The flow out of the Injector Module must be
through the one-way valve in the direction of the
arrow on the valve.
6-8
3. Connect the output end of the one-way valve to the humidifier inlet
port.
4. Connect tubing from the humidifier outlet port to the bias flow tube
on the high frequency oscillating ventilator.
02/08/00 1605-0014-000
Page 67
c
b
a
6/Operation
5. Connect the gas sample line from the “Sample Inlet” port on the
front of the INOvent delivery system fluid trap to a sampling tee on
the Flexible Breathing Circuit of the high frequency oscillating
ventilator.
1234
Tubing
Sensor Medics
3100A Ventilator
11
6-9
5
6
One way
Tubing
valve
10
Humidifier
9
1. High Frequency Oscillatory Ventilator
2. Ventilator outlet
3. Injector Module
4. INOvent delivery system
5. NO/N
Injector Tube connection
2
6. Injector Module Electrical Cable connection
7. One-way valve
8. Humidifier inlet
9. Patient gas sample line connection
10. Humidifier outlet
11. Bias Flow Tube
7
8
AA.69.126
AA.69.266
a. Injector Module
b. Input end
c. Output end
Figure 6-5 •
System connection diagram, high frequency oscillatory
ventilator
1605-0014-00002/08/00
6-9
Page 68
6-10
INOvent delivery system
Notes:
6-10
02/08/00 1605-0014-000
Page 69
Connection to a circle
anesthesia system
ww
ww
WARNING
Avoid recirculation of gases. Undesired recirculation of gases will occur
if fresh gas flows are less than the patient minute volume and may result
in:
Use only Sodasorb in combination with Isoflurane or Desflurane. Other
anesthetic agents and CO
reactive by-products.
6/Operation
• Higher NO
absorbent to remove NO
• Higher NO concentrations than those set due to NO recirculated through the absorber.
• Reduction in O
gas for nitric oxide and will be present in the re-circulated
gases.
levels due to the limited ability of the carbon dioxide
2
.
2
concentration because nitrogen is the balance
2
absorbents have not been evaluated for
2
6-11
ww
ww
CAUTION
Note the airflow direction arrow on the Injector Module: the flow out of
the absorber must be through the Injector Module in the direction of the
arrow on the module.
O will also affect the Set NO versus the measured NO value. For a 50%
N
2
O, 50% O
N
2
7% indicated less than the same Set NO value at 100% O
composition, the measured NO value will be approximately
2
. For example,
2
at a Set NO value of 20 ppm, measured NO will be approximately 18 ppm.
Similarly, the effect of 2%v/v Isofluorane will result in a high measured
NO value of approximately 3% indicated for the same Set NO value at
100% O
.
2
Sudden changes in anesthetic agent concentration may cause brief
transient changes in the measured NO and NO
values.
2
Procedure:
• When connecting breathing tubing, make sure it is long enough for
any required equipment movement.
• Make sure that the system is not wheeled over tubing connected to
a patient or to the equipment.
1605-0014-00002/08/00
6-11
Page 70
INOvent delivery system
Refer to the System connection diagram in figure 6-6 for the following
steps.
6-12
1. Connect the Injector Module’s
connects to the inspiratory limb of the breathing circuit.
input end to the absorber port which
Note: For OR ventilator systems with inspiratory flow
measurement at the inspiratory port of the absorber, place
the Injector Module upstream of the inspiratory flow
sensor.
2. Connect an inspiratory breathing hose from the output end of the
Injector Module to a sample tee.
3. Connect an inspiratory breathing hose from the output end of the
Injector Module to a sample tee.
4. Connect an inspiratory limb breathing hose from the sample tee to
the patient wye:
ImportantThis breathing hose between the sample
tee and the patient Y should be between 6
and 12 inches (150 to 300 mm) long:
greater than 6 inches to minimize the
sampling of mixed inspired/expired
concentrations and less than 12 inches to
help ensure correct patient NO
2
measurement.
6-12
5. Connect the gas sample line from the “Sample Inlet” port on the front
of the fluid trap to the sample tee. Make sure the sample port at the
T-piece is pointing upward to help avoid fluid accumulation in the
sample line.
6. Connect a breathing hose from the patient wye the absorber port
which connects to the expiratory limb of the breathing circuit.
Note: With a circle anesthesia breathing circuit, the
INOvent delivery system will perform as specified in
the technical specifications with fresh gas flow rates
equal to or more than the patient minute ventilation.
7. Proceed to “Operation” - “Setting the delivered NO concentration.”
02/08/00 1605-0014-000
Page 71
6/Operation
c
b
a
1234567
6-13
11a
11
11b121314
10
89
a. Injector Module
b. Input end
c. Output end
1 Patient wye
2 Patient Gas sample line
3 Patient Gas sample line connection (“Sample Inlet” port)
4 INOvent delivery system
5 Bellows assembly
6 Ventilator drive gas
7 Ventilator (rear view)
8 Absorber
9 Absorber expiratory port
10 Absorber Inspiratory port
11 Injector Module
a. Injector Module input end
b. Injector Module output end
12 Hose
13 22M / 15F x 22M / 15F Adapter
14 Sample Tee
AA.69.191
AA.69.266
Figure 6-6 •
System connection diagram, anesthesia system with
ventilator
1605-0014-00002/08/00
6-13
Page 72
INOvent delivery system
Connection to a
transport ventilator
circuit
ww
ww
WARNINGIf the INOvent delivery system is to be mounted in a transportation
vehicle, it must be equipped with four (4) vibration isolation devices
positioned near the feet of the device. See Appendix A for validated
vibration isolators.
6-14
Connection to a BIO-MED MVP-10 or an Infrasonics Infant Star 100
ventilator is shown in figure 6-7.
When operating on the battery, keeping the display brightness and the
audio alarm loudness to the minimum needed will help prolong
battery charge life.
Display brightness and alarm loudness can be changed if desired by
pushing the Setup key to display the Setup menu.
Note: The display brightness automatically dims to
the lowest setting (1) when running on battery.
If a humidifier is not being used in the breathing circuit, a “Dry Sample
Gas” alarm may occur. This alarm occurs if the sample gas has less
than 15% relative humidity.
The design of the gas sampling system allows its use without
humidification under certain conditions.
• When the ambient humidity is greater than approximately 45%
RH, the sample gas is humidified as it flows through the INOvent
delivery system and probably will not display the alarm message.
• Short term use of the INOvent delivery system with the “Dry
Sample Gas” alarm present should not affect the O
sensors.
, NO and NO2
2
6-14
Long-term use with the alarm present can reduce the life of all the
sensors and may result in slightly lower NO readings.
After connection as shown in the figure for the configuration you are
using, proceed to “Setting the delivered NO concentration.”
02/08/00 1605-0014-000
Page 73
6/Operation
5
4
3
2
6
1
7
8
6-15
AA.69.167
AA.69.266
13
12
11
910
a
b
c
a. Injector Module
b. Input end
c. Output end
1 Patient wye
2 Expiratory Breathing Circuit Hose
3 Patient Gas sample line
4 Ventilator Expiratory Valve
5 Ventilator
6 INOvent delivery system
7 Ventilator Inspiratory port
8 22M / 15F x 22M / 15F Adapter
9 Injector Module Electrical Cable
10 NO/N
Figure 6-7 • System connection diagram for a transport ventilator circuit
1605-0014-00002/08/006-15
Page 74
INOvent delivery system
Connection to
bagging systems
circuit
The INOvent delivery system can be used with self-inflating and
hyper-inflation manual resuscitator bagging systems and with
spontaneous breathing circuits.
6-16
wwwwWARNINGThe hyperinflation bag will, under some conditions, contain NO2 in
excess of 1 ppm. Use of large tidal volume breaths may expose the
patients to the NO
present in the bag, for part of the breath. In general,
2
if the inspiratory flow rate induced by the manual ventilation does not
exceed the fresh gas flow rate, the patient should not be exposed to the
concentrations of NO
Adult and infant hyperinflation bags generate more NO
present in the hyperinflation bag.
2
when used at
2
lower minute ventilation. If use of the bag is interrupted (for example to
adjust the tracheal tube), before resuming ventilation of the patient the
user should squeeze the bag several time to empty residual gas from the
bag.
Because of the potential for inhalation of excessive concentrations of
, and the difficulty in monitoring the peak inhaled NO2
NO
2
concentrations, ventilation with a hyperinflation bag or self-inflating bag
is intended only for short term use.
The Datex-Ohmeda monitors within the INOvent will not detect
generation of NO
devices, and the alarms for excessive NO
within the hyperinflation bag or self-inflating bag
2
cannot warn of NO2 produced
2
within the manual bag system.
To minimize the delivered concentration of NO
, the following steps
2
should be taken for use with the manual resuscitator bags:
a. Concentrations of greater than 20 ppm NO should not be
used due to excessive NO
generation.
2
b. Use the smallest bag adequate to deliver the desired tidal
volume.
c. Inspiratory tubing lengths greater than 72 inches should
not be used.
d. Use the highest fresh gas flow rated (up to 15 L/min) that is
practical.
e. Use the lowest practical inspired oxygen concentration.
f. After starting fresh gas flow, squeeze the bag several times
to empty residual gas in the bag prior to using the system
to ventilate a patient.
Testing has only been conducted using specific hyperinflation and self-
inflating bag systems. See Appendix A for validated manual resuscitator
breathing systems.
6-1602/08/00 1605-0014-000
Page 75
Procedure:
6/Operation
wwwwCAUTIONNew O
tubing must be used each time for optimal fit on 4.5 mm adapter.
2
If a humidifier is not being used in the breathing circuit, a “Dry Sample
Gas” alarm may occur.
This alarm occurs if the sample gas has less than 15% relative
humidity. The design of the gas sampling system allows its use
without humidification under certain conditions.
• When the ambient humidity is greater than approximately 45%
RH, the sample gas is humidified as it flows through the INOvent
delivery system and probably will not display the alarm message.
• Short term use of the INOvent delivery system with the “Dry
Sample Gas” alarm present should not affect the O
, NO and NO2
2
sensors. Long-term use with the alarm present can reduce the life
of all the sensors and may result in slightly lower NO readings.
Note the values shown in Tables 6-1 and 6-2.
Bagging system connections are shown in:
6-17
• Figure 6-8 for a hyper-inflation bag,
• Figure 6-9 for a self-filling resuscitator bag,
• Figure 6-10 for a spontaneous breathing circuit.
1605-0014-00002/08/006-17
Page 76
INOvent delivery system
6-18
wwwwWARNINGThe values of NO
tracheal tube under the specific test conditions. Use of lower fresh gas
flows than shown in the tabulated information will increase the
concentration of NO
Table 6-1 • Neonatal
Hyperinflation Manual
Resuscitator Breathing Circuit
Table
Set
NO(ppm)
100.40.2150208
100.10.11504015
100.10.150808
100.10.1504015
200.40.2150208
200.10.41504015
NO
2
tracheal tube (ppm)
in the following table are the values expected in the
2
in the inhaled gas.
2
The following table shows the maximum measured NO
values in the
2
inspiratory limb and at the rear of the breathing bag after five minutes
of bagging. Carrier gas = 100% oxygen, I:E = 1:2.
measured at
NO2 measured
in bag (ppm)
Vt(mL)
Rate
(bpm)
Flow Rate
(L/min)
200.10.450208
200.10.3504015
ww
ww
WARNING Concentrations greater than 20 ppm NO should not be used
because of excessive NO
generation.
2
400.31.350208
400.21.4504015
400.60.3150208
400.21.51504015
800.85.250208
800.84.1504015
801.03.6150208
800.85.41504015
6-1802/08/00 1605-0014-000
Page 77
6/Operation
wwwwWARNINGThe values of NO
tracheal tube under the specific test conditions. Use of lower fresh gas
flows than shown in the tabulated information will increase the
concentration of NO
Table 6-2 • Adult
Hyperinflation Manual
Resuscitator Breathing Circuit
Table
Set
NO(ppm)
100.20.2700205
100.00.15004015
100.00.1300205
100.00.12004015
200.50.6700205
200.10.45004015
NO
2
tracheal tube (ppm)
in the following table are the values expected in the
2
in the inhaled gas.
2
The following table shows the maximum measured NO
values in the
2
inspiratory limb and at the rear of the breathing bag after five minutes
of bagging. Carrier gas = 100% oxygen, I:E = 1:2.
measured at
NO2 measured
in bag (ppm)
Vt(mL)
Rate
(bpm)
Flow Rate
(L/min)
6-19
200.10.4300205
200.10.42004015
ww
ww
WARNING Concentrations greater than 20 ppm NO should not be used
because of excessive NO
generation.
2
400.32.1300205
400.42.32003015
400.91.9700205
400.42.45003015
801.29.0300205
801.49.02003015
802.88.8700205
801.59.15003015
1605-0014-00002/08/006-19
Page 78
6-20
INOvent delivery system
123456
1O
2
78910111213
Flowmeter
AA.69.166
2 Injector Module Electrical Cable
3 NO/N2 Injector Tube
4O2 Tubing
5O2 Tubing Sample Tee
6 Patient Gas sample line
7 Hyper-Inflation Bag
8 Pressure Gauge
9 15M x 4.5 mm Adapter for the output end of the Injector Module
10 Injector Module
11 22M / 15F x 22M / 15F Adapter for the input end of the Injector
Module
12 15M x 4.5 mm Adapter
Tubing
13 O
2
Figure 6-8 • System connection diagram for a hyper-inflation bag
6-2002/08/00 1605-0014-000
Page 79
6/Operation
10
1
6-21
9
8
2
4
3
56
7
AA.69.165
1O2 Flowmeter
2O2 Tubing
3 15M x 4.5 mm Adapter
4 22M / 15F x 22M / 15F Adapter for the input end of the Injector
Module
5 Injector Module
6 15M x 4.5 mm Adapter for the output end of the Injector Module
7O
Tubing
2
8O2 Tubing Sample Tee
9 Patient Gas sample line
10 Resuscitator bag with O2 reservoir
11 NO/N2 Injector Tube
12 Injector Module Electrical Cable
Figure 6-9 • System connection diagram for a self-filling resuscitator
bag
1605-0014-00002/08/006-21
Page 80
6-22
INOvent delivery system
13141516
AA.69.164
12345678 91011 12
1O
Tubing
2
2 15M x 4.5 mm Adapter
3 22M / 15F x 22M / 15F Adapter
4 Breathing Circuit Tee
5 Breathing Circuit Bag
6 Injector Module
7 Breathing Circuit Hose
8 Sample Tee
9 22M / 15F x 22M / 15F Adapter
10 One-way valve
11 Sealed Face Mask
12 One-way valve
13 Patient Gas Sample Line
14 NO/N
Figure 6-10 • System connection diagram for a spontaneous breathing
circuit
6-2202/08/00 1605-0014-000
Page 81
Operation
6/Operation
Setting or changing
the delivered NO
concentration
After connecting the INOvent delivery system to the patient circuit,
turn the INOvent delivery system ON.
• Turning the system on delivers nitric oxide concentration when
there is flow in the patient circuit.
• Figure 6-11 shows a normal state display with the Set NO
bargraph at off after the startup tests are completed.
6-23
510k_3.tif510k_4.tif
Figure 6-11 • Set NO display - Set NO is 0.0 ppm (OFF)
To set or change the NO concentration delivered:
1. Push the Set NO button.
• The bargraph border now flashes
• The bargraph top now becomes a moveable double line indicating
the current concentration setting.
2. Turn the control wheel clockwise to increase the NO concentration
or counterclockwise to decrease the concentration. Figure 6-12
shows the concentration line set at 20 ppm.
Figure 6-12 • Set NO display - NO setpoint is a double line
1605-0014-00002/08/006-23
Page 82
6-24
INOvent delivery system
3. Push either the Set NO button or the control wheel to confirm the
newly-entered concentration.
• If you don’t push the button or the wheel within five seconds, a
reminder message, shown in figure 6-13, appears to prompt you
to do so. A highlight box around the display flashes.
• If you don’t respond to the message within 15 seconds, the
system defaults to its previous NO setting.
510k_5.tif510k_6.tif
Figure 6-13 • Set NO display - confirm new NO setpoint reminder
message
• Figure 6-14 shows the NO concentration set to 20 ppm after the
Set NO button or the control wheel is pressed. The system is now
set to deliver the new concentration.
Figure 6-14 • Set NO display - normal state, NO bargraph showing 20
ppm setting
4. After the monitored values have stabilized, you can set or change
the user-adjustable alarms. (See section 7/Alarms for changing the
alarm settings.)
Note: NO delivery dilutes the therapy gas O
concentration up to
2
10% depending upon the Set NO concentration.
The system is now ready for use. The new NO concentration is
delivered when there is flow in the patient circuit.
6-2402/08/00 1605-0014-000
Page 83
During operation
Pausing NO flow
wwwwWARNINGDo not pause flow more than two times consecutively to avoid elevated
NO
values when NO delivery is resumed.
2
The Low NO alarm is disabled while NO flow is paused.
Nitric oxide flow can be paused for two-minute periods.
To pause nitric oxide flow during operation:
1. Press the Pause Flow button to display the Pause Flow
confirmation message box shown in figure 6-15.
6/Operation
6-25
Figure 6-15 • Pause Flow confirmation message box
2. Then:
• Press the Pause Flow button a second time to pause the nitric
oxide flow. The FLOW PAUSED message box appears with a
two-minute countdown timer as shown in figure 6-16.
or:
• Press the Control Wheel to remove the Pause Flow confirmation
message box. The Pause Flow confirmation message box
disappears after 15 seconds if no action is taken.
1605-0014-00002/08/006-25
510k_18.tif
Page 84
INOvent delivery system
510k_19.tif
6-26
Changing the NO
therapy cylinder and
purging the regulator
assembly
wwwwWARNINGYou MUST purge the regulator assembly IMMEDIATELY BEFORE using a
new NO cylinder to make sure the patient continues to receive the
correct NO concentration and does not receive greater than 3.0 ppm of
NO
ImportantReplace an NO therapy gas cylinder when its pressure is less than 200 psig.
Figure 6-16 • FLOW PAUSED message box
Once nitric oxide flow is paused, it will resume:
• if the Resume Flow button is pressed or
• after two minutes have passed.
When the nitric oxide flow is resumed, the FLOW PAUSED message
is removed.
.
2
The regulator assembly must be purged immediately before using a
new NO therapy gas cylinder. A small purge manifold allows you to
purge the regulator assembly. The manifold is located either between
the regulators on the transport cart or attached to the wall mount
regulator inlet fitting.
To purge the regulator assembly immediately before using a new NO
therapy gas cylinder do the following.
1. Determine which regulator assembly is not being used and requires
purging. The cylinder valve should be closed on this regulator
assembly that is not being used.
6-2602/08/00 1605-0014-000
Page 85
6/Operation
2. On this regulator assembly that requires purging, disconnect the
low pressure hose quick-connect fitting from the NO/N
rear of the INOvent delivery system (shown in figure 6-17.)
Figure 6-18 • Inserting the low pressure hose quick-connect fitting into
the purge manifold
7. Firmly push and hold the quick-connect fitting in place while the
pressure drops to zero on the regulator gauge.
8. After the pressure drops to zero (approximately 15 seconds),
reconnect the low pressure hose quick-connect fitting to the NO/N2
input on the rear of the INOvent delivery system.
AA.69.152
1605-0014-00002/08/006-27
Page 86
6-28
INOvent delivery system
To use the new cylinder:
1. Open the cylinder valve on the new cylinder.
2. Close the cylinder valve on the empty cylinder and label
appropriately.
3. Replace the empty NO therapy gas cylinder (see 13/Appendix B.)
Leave the replacement cylinder valve turned off until needed. The
purge procedure will be performed on this cylinder immediately
before its use.
Electronic delivery
shutdown
If the electronic NO delivery fails:
• A loud buzzer operates and
• The system stops NO delivery to the patient and
• Displays a message similar to the one in figure 6-19.
Figure 6-19 • System shutdown display
1. Record the service information as requested on the display.
2. Turn the system to STANDBY.
3. Use the Manual NO delivery system which can still be used. If you
need to know how to use the manual system, see “Manual NO
Delivery System operation” in section 8/Manual NO delivery
system.”
6-2802/08/00 1605-0014-000
SD1
Page 87
Monitoring the environment
You can use the system monitor to measure the environmental levels
of NO and NO2. You will need a Luer fitting cap for the patient line
sample tee.
1. Disconnect the sample line connector from the sample tee.
Figure 6-20 • Sample line connector at the sample tee
6/Operation
6-29
AA.69.017
2. Cap the Luer fitting on the sample tee to prevent gas loss from the
patient circuit.
wwwwWARNINGAfter removing the sample line connector, cap the Luer fitting on the
sample tee to prevent gas loss from the patient circuit.
3. With the sample system turned on, sample the room air with the
sample line and read the NO and NO
press the Alarm Silence button to turn it off.
4. After environmental monitoring, remove the Luer fitting cap on the
sample tee and reconnect the sample line.
Note: Refer to the your ventilator manual for information on the
Apnea, High Pressure and High Pressure Relief alarms.
An Alarm message table is provided at the end of this section.
7-2
High, medium and
low priority alarms
All alarms have tones and visible messages. You’ll see and hear an
alarm when:
• the measured value of a gas exceeds the high or low limits shown in
the measured value display area,
• a device alarm condition occurs, such as an expended monitor cell.
A secondary alarm circuit activates when the main alarm circuit fails,
providing a continuous, non-silenceable buzzing tone. In this case the
system must be turned off.
Alarms have priorities of high, medium and low. Higher priority alarm
messages are displayed in the upper left alarm message area first.
See figure 7-1. The most recent high priority alarm message (for
example “High NO”) is shown in the top left part of the alarm message
area. High priority alarm messages are shown on a
light background.
7-2
Figure 7-1 •
The most recent medium priority alarm message (for example “High
O
” in figure 7-2) appears in normal video in the top right or the top left
2
part of the alarm message area.
High priority alarm message shown in most recent alarm
location (upper left)
02/08/00 1605-0014-000
510k_20
Page 91
7/Alarms
510k_21‘510K_22
7-3
Figure 7-2 •
Medium priority alarm message shown in most recent
alarm location (upper left)
Alarm silencing
The most recent low priority alarm message (for example “Sample
Line/Filter Block” in figure 7-3) appears in
right part of the alarm message area.
Figure 7-3 •
Push the Alarm Silence button to silence high and medium priority
alarms for 120 seconds. Most low priority alarms are permanently
silenced.
Low priority alarm message shown in low priority alarm
location
normal video in the bottom
1605-0014-00002/08/00
Alarm messages, however, remain visible during the alarm silence
period as long as the alarm cause is active.
Note: Tone loudness is set through the Setup button menu.
The following table provides alarm tone information.
7-3
Page 92
INOvent delivery system
7-4
Alarm
Priority
High:
Medium:
Low:
Figure 7-4 shows that a High NO alarm message has been silenced.
Note the alarm silenced symbol in the top left location indicating a
silenced alarm. An alarm silenced countdown timer has counted down
from 120 seconds and now indicates 114 seconds remaining of the
silence period.
2
1
Tone
Frequency
400 Hz5-pulse groupRepeats after 10
400 Hz3-pulse groupRepeats after 25
400 Hz1 pulseRepeats after 40
3
Tone
Description
Comment
seconds if not silenced.
seconds if not silenced.
seconds if not silenced.
510k_23
Changing the alarm
settings
7-4
1. Countdown timer
2. Alarm silenced symbol
3. High nitric oxide alarm message
Figure 7-4 •
Each alarm setting is displayed below its respective measured gas
value. See figure 7-5.
1. Measured gas value
2. Low alarm limit
3. High alarm limit
Figure 7-5 •
Silenced alarm and countdown timer example (upper left
of the display)
1
2
Alarm setting displays
3
02/08/00 1605-0014-000
510k_23
Page 93
7/Alarms
When an alarm occurs for a measured gas concentration, the
measured value and the alarm limit setting are displayed on a light
background. (See figure 7-6.)
Alarm adjustment
ranges
AlarmsAdjustmentStep valueDefault Detection range Priority
High NO (ppm)1 to 100 *1 90 *NO > high limitHigh
Low NO (ppm)
(Can be OFF)
High NO
High O
(Can be OFF)
(ppm)0 to 50.13NO
2
(% v/v)
2
1
1. Measured gas alarming
2. Alarm limit exceeded
Figure 7-6 •
OFF, 0.1 to
Alarm active display
0.1OffNO < low limitHigh
9.9*
10 to 100*1.0
21 to 1001100 %O
2
> high limitHigh
2
> high limitMedium
2
510k_21
7-5
Low O
Alarm adjustment
procedure
1605-0014-00002/08/00
(% v/v)18 to 1001 %21 %O
2
* Values shown are for maximum Set NO range of 80 ppm. For other
maximum Set NO ranges, the high adjustment range and the default
value is 1.5 X maximum NO Set.
A user-adjustable low limit alarm cannot be set above the high limit
setting.
1. With the monitor ON and after the monitored values have
stabilized, push the Alarms button to show the Alarms menu.
< low limitHigh
2
7-5
Page 94
INOvent delivery system
This menu lets you:
7-6
• Set a high limit for measured values of the three gases (O
, NO
2
and NO.)
• Set a low limit for the measured values of NO and O
.
2
• Exit to the normal display.
2. Turn the control wheel to highlight the alarm setpoint to be
changed. In figure 7-7, the Low NO alarm setting is highlighted.
Figure 7-7 •
Alarms menu - low NO alarm setting highlighted
3. Push the control wheel to select the highlighted alarm setpoint for
change; in figure 7-8 the Low NO setpoint is shown in a selection
box.
2
510k_7.tif510k_8.tif
7-6
Figure 7-8 •
Alarms menu - low NO alarm setting selected for
adjustment
4. Turn the control wheel until the value you wanted is displayed in the
selection box.
5. Push the control wheel to confirm your selection. The alarm limit is
displayed below its respective measured gas value when the
normal display is shown. See figure 7-9.
02/08/00 1605-0014-000
Page 95
7/Alarms
510K_22
Clearing
resolved alarm
messages
Figure 7-9 •
6. If you are finished with the Alarms menu, turn the control wheel to
select Exit to Normal Display. To adjust another value, continue
using the above procedure. Push the control wheel to exit when
you are finished.
An “Alarm History” indication in the top right location of
the display indicates there are resolved alarms that can
be reviewed and cleared.
A resolved alarm is an alarm that occurred but is no longer true: that
is, the alarm message came and went. A history of up to the last ten
resolved alarm messages is stored for analysis or troubleshooting.
Push the Alarm History button to see the Alarm History menu. The
messages are displayed in chronological order, starting with the most
recent resolved alarm.
New nitric alarm limit set
7-7
1605-0014-00002/08/00
Note: The alarm history list is cleared if the INOvent delivery system is
turned OFF and then ON.
To clear alarm messages on the resolved alarm history list:
1. Push the Alarm History button to see the Alarm History menu.
• Figure 7-10 shows page 1 of the menu when there are more than
five alarm messages.
• Figure 7-11 shows page 2 of the menu used when there are more
than five alarm messages.
7-7
Page 96
INOvent delivery system
510k_24.tif
7-8
Figure 7-10 •
Resolved Alarm History menu - six or more resolved
alarms, page 1
Figure 7-11 •
Resolved Alarm History menu - 6 or more resolved
alarms, page 2
2. Turn the control wheel to highlight the message or action wanted.
3. Push the control wheel to complete the action you want or go to the
next page. Figure 7-12 shows the menu with all alarm messages
removed.
510k_25.tif510k_26.tif
7-8
4. When finished, turn the control wheel to “Exit to Normal Display.”
Push the wheel to exit.
Figure 7-12 •
Resolved Alarm History menu - resolved alarms cleared
A list of alarm messages starts on the next page.
02/08/00 1605-0014-000
Page 97
Alarm message table
ww
ww
WARNING
The INOvent delivery system should not be connected to the patient
during troubleshooting or service activities.
When an alarm occurs, check the patient condition and then check the
equipment.
Use the table below for information on conditions.The last column
references numbers in the Troubleshooting Guide table of section 9/
Troubleshooting.
7/Alarms
7-9
See Troubleshooting
AlarmPriorityConditions
Battery
Inoperable
LowDisplayed when running on mains power and
battery not functioning.
Low BatteryHighDisplayed when running on battery and charge is
Guide number:
13
14
low.
Low Battery
Charge
Running on
LowDisplayed only when running on mains power and
15
battery charge is low.
LowDisplayed only when running on battery.16
Battery
Set NO ErrorLowSystem detects a Set NO communication error.7
Water Bottle FullLowSystem detects full water bottle.18
Sample Line/
Filter Block
LowSystem failed to clear blockage after 3 purge
routines.
17
Dry Sample GasLowSample gas < 15% relative humidity.11
Wet Sample
LowSample gas > 95% relative humidity.35
Gas
Calibration DueLowLoss of calibration data.25
Failed NO CellLowSystem detects absence or failure of an NO cell.26
Failed NO
Failed O
CellLowSystem detects absence or failure of an NO
2
CellLowSystem detects absence or failure of an O
2
cell.27
2
cell.28
2
Weak NO CellLowWeak NO cell detected during calibration.29
Weak NO
Weak O
CellLowWeak NO
2
CellLowWeak O
2
cell detected during calibration.30
2
cell detected during calibration.31
2
High NOHighMeasured NO value > User set limit; default: 1.5 x
Bargraphmax*, but not greater than 90 ppm.
1605-0014-00002/08/00
2
7-9
Page 98
INOvent delivery system
7-10
High NO FlowHighDuring calibration, NO flow changed and NO
concentration may exceed 100 ppm or O
dilution
2
> 20%.
High NO
2
HighMeasured NO
value > User set limit; default: 3.0
2
ppm.
High O
2
MediumMeasured O
value > User set limit; default: 100
2
%.
Low NOHighMeasured NO value < User set limit; default: Off.1, 3
Low NO/N
2
HighNO therapy gas supply pressure < 32 psig.4
Pressure
Low O
2
Decreased NO
Flow
Increased NO
Flow
Measured NO <
Set NO
Measured NO >
Set NO
Injector Module
HighMeasured O
value < User set limit; default: 21 %.21
2
LowNO flow < 60% of NO flow when calibration was
initiated.
LowNO flow > 160% of NO flow when calibration was
initiated.
LowNOmeas < 50% of NOset and NOmeas < NOset -
NOoffset.**
LowNOmeas > 150% of NOset and NOmeas > NOset
+ NOoffset.**
HighSystem cannot detect a working Injector Module.34
Fail
23
12
22
8
9
1, 5
1, 6
Delivery FailureHighCalculated NO flow is outside acceptable range or
leak rate was unacceptable at Power-ON.
Monitoring
Failure
LowMonitor fails to communicate correctly or reports a
fault.
System FailureShutdown NOmeas > 2 x Bargraphmax or 100 ppm or other
system failure. *
Service
LowSystem detects a condition which needs service.33
Advisory
* Bargraphmax =Maximum value settable on the Set NO
Bargraph.
** NOoffset =5 ppm when Set NO range = 0 - 20
7 ppm when Set NO range = 0 - 40
10 ppm when Set NO range = 0 - 80.
- End of Alarm message table -
10
17
24
7-10
02/08/00 1605-0014-000
Page 99
8
8/Manual NO delivery system
In this sectionManual NO delivery system description . . . . . . . . . . . . . . . . . 8-2
The Manual NO delivery system permits continued NO delivery if the
ventilator or the main INOvent delivery system fails. The manual
system is completely pneumatic and is not linked to the primary
delivery system.
The Manual NO delivery system is used with a pressure compensated
(PC) O
(user supplied) to deliver a 20 ppm concentration of NO to the patient
with an O2 flowrate of 15 L/min and an 800 ppm NO cylinder.
flowmeter (user supplied) and a manual resuscitator bag
2
The pressure compensated O
function and is used to set the O
flowmeter provides the ON/OFF
2
flow at 15 L/min. A shutoff valve in
2
the delivery system prevents NO flow if there is insufficient O2 flow.
When the manual NO delivery system is operating correctly, the float
in the NO Flow Indicator Window will be in the position shown in figure
8-1.
1
NO/O
2
2
1. Float
2. NO Flow Indicator Window
69.023
Figure 8-1 • Manual NO delivery system: NO Flow Indicator Window
and Float
8-202/08/00605-0014-000
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.