Conformity according to the Council Directive 93/42/EEC concerning Medical Devices amended by 2007/47/EEC.
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Outside the USA, check local laws for any restriction that may apply.
All specifications subject to change without notice.
Order code for the Datex-Ohmeda E-Modules Technical Reference Manual, paper: M1065282
Order code for the S/5 Technical Reference Manuals CD: M1220141
th
4
edition
November 19, 2012
GE Healthcare Finland Oy
Kuortaneenkatu 2,
FI-00510 Helsinki
Finland
Tel: +358 10 39411
Fax: +358 9 1463310
www.gehealthcare.com
2012, 2011 General Electric Company. All rights reserved.
Page 2
Responsibility of the manufacturer
GE Healthcare Finland Oy (GE) is responsible for the effects on safety, reliability and performance of the
equipment only if:
• assembly, extensions, readjustments, modifications, servicing and repairs are carried out by
personnel authorized by GE.
• the electrical installation of the monitor room complies with appropriate requirements.
• the equipment is used in accordance with the "User's Guide" and serviced and maintained in
accordance with the “Technical Reference Manual”.
The manufacturer reserves the right to change product specifications without prior notice. Although the
information in this manual is believed to be accurate and reliable, the manufacturer assumes no
responsibility for its use.
Trademarks
S/5, D-lite, D-lite+, Pedi-lite, Pedi-lite+, Mini D-fend, D-fend, D-fend+, MemCard, ComBar, ComWheel,
EarSat, FingerSat, FlexSat, PatientO2, Entropy and Patient Spirometry are trademarks of GE Healthcare
Finland Oy.
Datex, Ohmeda, and OxyTip+ are trademarks of GE Healthcare Finland Oy and Datex-Ohmeda, Inc.
A portion of the Entropy software is derived from the RSA Data Security, Inc. MD5 Message-Digest
Algorithm.
Masimo SET
Masimo SET is a licensed trademark of Masimo Corporation.
All other product and company names are property of their respective owners.
Product availability
Some of the products mentioned in this manual may not be available in all countries.
Please, consult your local representative for the availability.
Page 3
Description
Master Table of Contents
Datex-Ohmeda E-Modules
M1065282
Respiratory Modules, E-sCAiOV, E-sCAiO, E-sCOV and E-sCO1
Compact Airway Modules, E-CAiOVX, E-CAiOV,
E-CAiO, E-COVX, E-COV and E-CO
PRESTN Modules, E-PRESTN, E-RESTN, E-PRETN3
Patient Side Modules, E-PSM, E-PSMP rev. 014
Cardiac Output Modules E-COP rev. 01 and E-COPSv rev. 015
EEG Module, E-EEG rev.00 and EEG Headbox, N-EEG rev. 01
BIS Module, E-BIS rev. 01
NeuroMuscular Transmission Module, E-NMT
Device Interfacing Solution, N-DISxxx rev. 01
13
14
15
16
Interface Module, E-INT17
2
Recorder Module, E-REC18
Memory Module, E-MEM19
Remote Controllers, K-REMCO rev. 01, K-CREMCO20
Anesthesia record keeping keyboard, K-ARKB, Keyboard Interface Board,
B-ARK and ARK Barcode Reader, N-SCAN
E-Modules, Spare Parts22
Document no. M1181405B
21
i
Page 4
Datex-Ohmeda E-Modules
ii
Document no. M1181405B
Page 5
About this manual
Notes to the reader
This Technical Reference Manual is intended for service personnel and engineers who will
service and maintain the Datex-Ohmeda E-Modules as well as the anesthesia record keeping
keyboard, K-ARKB, remote controllers, K-REMCO and K-CREMCO, Device Interfacing Solution,
N-DISxxx, keyboard interface board, B-ARK, and ARK barcode reader, N-SCAN.
This Technical Reference Manual completes the S/5 Anesthesia Monitor and S/5 Critical Care
Monitor Technical Reference Manual and the S/5 Compact Anesthesia Monitor and
S/5 Compact Critical Care Monitor Technical Reference Manual. Later in this manual, the
monitors may be referred to as AM, CCM, CAM and CCCM.
The order code for the Datex-Ohmeda E-Modules Technical Reference Manual is M1065282.
The order code for the S/5 Technical Reference Manuals CD is M1220141. S/5 AM, CCM
Technical Reference Manual, S/5 CAM, CCCM Technical Reference Manual and Datex-Ohmeda
E-Modules Technical Reference Manual are included on the CD.
Each manual on the CD has an individual document number and is available for downloading
from GE Common Document Library in Adobe Acrobat PDF format.
•This Technical Reference Manual contains the information needed to maintain, service
and troubleshoot these products. Instructions for visual and functional inspection,
disassembly and reassembly as well as calibration of the modules are included. A service
check form for each product is included in the slots.
•In addition, this Technical Reference Manual contains detailed module specifications and
descriptions on the technical performance and functioning of the modules.
•Read the manual through and make sure that you understand the procedures described
before servicing the modules. To avoid risks concerning safety or health, strictly observe
the warning indications. If you need any assistance concerning the service, please do not
hesitate to contact your authorized distributor.
For information on safety precautions and symbols on equipment, installation, planned
maintenance and interfacing, refer to the AM and CCM Technical Reference Manual or the CAM
and CCCM Technical Reference Manual.
The manufacturer reserves the right to change product specifications without prior notice.
Although the information in this manual is believed to be accurate and reliable, the
manufacturer assumes no responsibility for its use.
GE Healthcare assumes no responsibility for the use or reliability of its software in equipment
that is not furnished by GE Healthcare.
For more specific information about the clinical aspects refer to:
S/5 monitor’s User’s Guide
S/5 monitor’s User’s Reference Manual
1
Document no. M1181405B
Page 6
Datex-Ohmeda E-Modules
Conventions used
Throughout this manual, the following conventions are used to distinguish procedures or
elements of text:
Sign the check form after performing the procedure.
Hard KeysHard key names on the Command Board, the Remote Controller and modules are written in the
following way:
Menu ItemsMenu items are written in bold italic: ECG Setup.
‘Messages’Messages displayed on the screen are written inside single quotes: ‘Please wait’.
“Sections”When referring to different sections in the same manual, the section name is enclosed in
double quotes: section “Cleaning and Service.”
“Other documents”
When referring to different documents, the document name is enclosed in double quotes: refer
to “User’s Reference Manual”.
Hypertext links Hypertext links on PDF versions are written in blue color.
WARNINGWarnings are written in the following way:
WARNINGMake sure that the electrodes, sensor and connectors do not touch any
electrically conductive material, including earth.
CAUTIONCautions are written in the following way:
ECG.
CAUTIONThe module electronics can only be repaired and calibrated at the factory.
NOTENotes are written in following way:
NOTE: Handle all PC boards by their edges.
In this manual, the word “select” means choosing and confirming.
Revision history
RevisionDateComment
1st edition10 May 2011Initial
2nd edition22 Sep 2011Order code for paper manual added.
3rd edition19 June 2012E-PSM rev. 01, E-PSMP rev. 01, E-COP rev.01 and E-COPSv rev. 01
update.
4th edition19 Nov 2012Respiratory Modules E-sCAiOV, E-sCAiO, E-sCOV and E-sCO added .
Conformity according to the Council Directive 93/42/EEC concerning Medical Devices amended by 2007/47/EEC.
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Outside the USA, check local laws for any restriction that may apply.
All specifications subject to change without notice.
Document number M1214853C
June 21, 2012
GE Healthcare Finland Oy
Kuortaneenkatu 2,
FI-00510 Helsinki
Finland
Tel: +358 10 39411
Fax: +358 9 1463310
www.gehealthcare.com
2012 General Electric Company. All rights reserved.
Size (H x W x D) 112 x 37 x 205 mm ( 4.4 x 1.5 x 8.7 in)
Weight 0.75 kg (1.5 lb)
Power consumption3.9 W
1.2Operating characteristics
Warm-up time
- CO2, O2 and N2O measurements: 1 minute
-Anesthetic agent measurement
and identification:5 minutes
Gas sampling rate:120 ±20 ml/min
Automatic compensation for ambient pressure.
Operating conditions
Ambient temperature:+10°C to +40°C
Ambient pressure:660 mbar to 1060 mbar
Ambient humidity:10%RH to 98%RH, non-condensing
1.3Airway gases
1.3.1General characteristics
Specifications are valid at the following normal operating conditions:
Ambient temperature:+18°C to +28°C, within ±5°C of calibration
Ambient pressure:660 mbar to 1060 mbar, ±67 mbar of calibration
Ambient humidity:20%RH to 80%RH, non-condensing, ±20%RH of calibration
Sampling line length:2, 3 and 6 meters
Respiration rate:4 to 70 breaths/minute
NOTE: The displayed ranges of parameter values depend on the host device. For more information, refer
to the host device’s user documentation.
1.3.2Respiration rate
Breath detection:1 vol% change in CO2 level
Measurement range:4 to 100 breaths/min
Accuracy
at 4 to 20 breaths/min:±1 breath/min
at 20 to 100 breaths/min:±5%
RR value is updated breath-by-breath.
(Halothane 4 to 50 breaths/minute)
3
Document no. M1214853C
Page 14
Datex-Ohmeda E-Modules
1.3.3Carbon dioxide
Measurement range:0 vol% to 15 vol%, 0 kPa to 15 kPa,
Accuracy:±(0.2 vol% +2% of reading)
Total system response time:< 3.0 s
Rise time:< 260 ms
drift:< 0.1 vol%
CO
2
EtCO2 and FiCO2 values are updated breath-by-breath.
Description of test method, data rate, and method of ET-calculation
The module uses gas concentration waveforms with data rate of 25Hz to calculate end-tidal (ET) gas
readings.
The module finds the time instant of the highest CO2 concentration in each breath. Concentration at that
instant is the ET CO2 reading. Because nitrous oxide and anesthetic agents are measured by the same
sensor as CO2, the ET-readings of those gases are obtained directly at the time instant of ET CO2. For
calculating ET-readings of oxygen, the module synchronizes the O2-waveform with the CO2 waveform.
The ET-reading of O2 is then determined as O2-concentration at the time instant of ET CO2. If no breaths
are detected for a given time (20s, for example), an apnea situation is triggered. During apnea, the ET
values are updated every two seconds to the current concentration of each gas.
The rated respiration rate range and the corresponding end-tidal gas reading accuracy were tested with
reference gases of known concentrations. The test gases were fed to the gas sampling system of the
module through an electrically actuated valve with very low internal volume. Depending on its actuation
status, the valve directed either room air or a test gas to the gas sampling line. The desired respiration
rates were set by the electrical actuating times of the valve.
The measurement accuracy of the end-tidal gas readings was tested using gas sampling lines of 3 meter
length, connected to the gas sample port on the D-fend Pro water trap. The gas sampled to the sampling
line was switched from room air to the test gases using an electrically actuated valve with low internal
dead space to generate step changes in the gas concentrations. The electric actuating signal of the valve
was generated using a higly accurate signal generator to accurately control the simulated respiration
rate.
The electronic sampling rate of the gas sensor signals is 25Hz, equaling a new data point on the gas
waveform traces every 40 milliseconds.
0 mmHg to 113 mmHg
1.3.4Oxygen
Measurement range:0 vol% to 100 vol%
Accuracy:±(1 vol% +2% of reading)
Total system response time:< 3.0 s
Rise time:< 260 ms
drift:< 0.3 vol%
O
2
and FiO2 values are updated breath-by-breath.
EtO
2
1.3.5Nitrous oxide
Measurement range:0 vol% to 100 vol%
Accuracy:
N2O < 85 vol%)±(2 vol% +2% of reading)
at (0 <
Total system response time:< 3.0 s
Rise time:< 320 ms
O drift:< 0.3 vol%
N
2
O and FiN2O values are updated breath-by-breath.
EtN
2
4
Document no. M1214853C
Page 15
1.3.6Anesthetic agents
Measurement range:
Sevoflurane:0 vol% to 8 vol%
Desflurane:0 vol% to 20 vol%
Isoflurane, enflurane, halothane:0 vol% to 6 vol%
Accuracy:±(0.15 vol% +5% of reading)
Total system response time:< 3.1 s (< 3.5 s for Halothane)
Rise time:< 420 ms (< 800 ms for Halothane)
Hal drift:< 0.1 vol%
Enf drift:< 0.1 vol%
Iso drift:< 0.1 vol%
Sev drift:< 0.1 vol%
Des drift:< 0.3 vol%
EtAA and FiAA values are updated breath-by-breath.
The module automatically identifies the anesthetic agent present in the sampled gas and measures the
concentration of the identified agent.
Identification threshold:0.15 vol%
Identification time:< 20 s
The module automatically identifies mixtures of two anesthetic agents present in the sampled gas and
measures the concentrations of the two identified agents.
agent:0.2 vol% +10% of the concentration of the 1st agent
1.3.7Non-disturbing gases
A gas is considered non-disturbing if its effects to the measured gas are as follows:
CO
:< 0.2 vol%
2
O2, N2O:< 2 vol%
Anesthetic agents:< 0.15 vol%
The following gases are non-disturbing when tested according to ISO21647(2004B):
ethanol, acetone, isopropanol, methane, nitrogen, carbon monoxide, nitric oxide, freon R134A (for CO
O), water vapor.
and N
2
The effects caused by N2O to the measurement of CO2, O2 and anesthetic agents are automatically
compensated for.
The effects caused by anesthetic agents to the measurement of CO2 and N2O are automatically
compensated for.
1.3.8Gas cross effects
Helium (50 vol%):Decreases CO2 readings by less than 0.5 vol% at 5 vol% of CO
Xenon (80 vol%):Decreases CO2 readings by less than 0.5 vol% at 5 vol% of CO
Decreases O2 readings by less than 2 vol% at 50 vol% of O
Decreases O2 readings by less than 1.5 vol% at 14 vol% of O
, O2
2
2
2
2
2
5
Document no. M1214853C
Page 16
Datex-Ohmeda E-Modules
1.4Patient Spirometry
1.4.1General characteristics
These specifications are valid in the following operating conditions:
The module has been operating continuously for 10 minutes
- pediatric patients:4 to 70 breaths/min
I:E ratio:1:4.5 to 2:1
Airway humidity:10 %RH to 100 %RH
Ambient temperature:+10°C to +40°C
Ambient pressure:660 mbar to 1060 mbar
Ambient humidity:10 %RH to 98 %RH (non-condensing)
NOTE: The displayed ranges of parameter values depend on the host device. For more information, refer
to the host device’s user documentation.
1.4.2Airway pressure
Measurement range:-20 cmH2O to +100 cmH2O
Accuracy:±1 cmH
Time resolution:10 ms
Values calculated from the measured airway pressure data:
Peak pressure (Ppeak)
Plateau pressure (Pplat)
Mean pressure (Pmean)
Positive end expiratory pressure (PEEPtot, or PEEPi and PEEPe)
Static positive end expiratory pressure (static PEEPe and static PEEPi)
1.4.3Airway gas flow
Measurement range
- adults:-100 l/min to +100 l/min
- pediatic patients:-25 l/min to +25 l/min
Time resolution:10 ms
Flow measurement has automatic compensation for airway pressure and effects caused by variation in
the concentrations of the gas components measured by the module.
1.4.4Tidal volume
The module calculates the volume by integrating the measured gas flow over time. Tidal volumes (TVinsp
and TVexp) are obtained as the change of volume during inspiration and expiration.
O
2
Measurement range
- with D-lite:150 ml to 2000 ml
- with Pedi-lite:5 ml to 300 ml
Accuracy
- with D-lite:±6% or 30 ml (whichever is greater)
- with Pedi-lite:±6% or 4 ml (whichever is greater)
6
Document no. M1214853C
Page 17
1.4.5Minute volume
The module calculates the inspired and expired minute volumes as the sum of inspired (MVinsp) and
expired (MVexp) gas volume during one minute.
Measurement range
- with D-lite:2 l to 20 l
- with Pedi-lite:0.5 l to 5 l
1.4.6Compliance
The module calculates both the compliance (Compl) and static compliance (static Compl). Compliance is
calculated by dividing the expired gas volume (TVexp) by the change in the airway pressure (Pplat PEEPtot). Static compliance is calculated by dividing TVexp by the difference of static Pplat and static
PEEPtot.
Measurement range
- adults:4 ml/cmH2O to 100 ml/cmH2O
- pediatric patients:1 ml/cmH2O to 100 ml/cmH2O
1.4.7Airway resistance
The module calculates the airway resistance (Raw) by solving the lung model equation P(t) = Raw * F(t) +
V(t) / Compl + PEEPtot
where: P(t),F(t) and V(t) are the time dependent waveforms of pressure, flow, and volume, respectively.
Measurement range:0 cmH
The module measures ratio of the inspiratory and expiratory time (I:E).
The inspiratory time is the time from the start of inspiration to the start of expiration. The end inspiratory
pause, if one exists, is included in the inspiration. Accordingly, expiratory time is the time from the start of
expiration to the start of the next inspiration.
7
Document no. M1214853C
Page 18
Datex-Ohmeda E-Modules
2Functional description
2.1Measurement principle
2.1.1CO2, N2O, and agent measurement
MiniTPX is a side stream gas analyzer, measuring real time concentrations of CO2, N2O, and
anesthetic agents (Halothane, Enflurane, Isoflurane, Desflurane, and Sevoflurane).
Figure 2MiniTPX sensor principle
Anesthetic agents or mixtures of two anesthetic agents are automatically identified, and
concentrations of the identified agents are measured. MiniTPX also detects mixtures of more
than two agents and issues an alarm.
MiniTPX is a non-dispersive infrared analyzer, measuring absorption of the gas sample at
seven infrared wavelengths, which are selected using optical narrow band filters.
The infrared radiation detectors are thermopiles.
Concentrations of CO
and N2O are calculated from absorption measured at 3-5 m.
Identification of anesthetic agents and calculation of their concentrations is performed by
measuring absorptions at five wavelengths in the 8-9 m band and solving the concentrations
from a set of equations.
Figure 4Infrared absorbance of AAs
The measuring accuracy is achieved utilizing numerous software compensations. The
compensation parameters are determined individually for each MiniTPX during the factory
calibration.
9
Document no. M1214853C
Page 20
Datex-Ohmeda E-Modules
2.1.2O2 measurement
The differential oxygen measuring unit uses the paramagnetic principle in a pneumatic bridge
configuration. The signal picked up with a differential pressure transducer unit is generated in
a measuring cell with a strong magnetic field that is switched on and off at a main frequency of
164 Hz. The output signal is a DC voltage proportional to the O
between the gas to be measured and the air reference.
concentration difference
2
Figure 5O2 measurement principle
2.1.3Patient spirometry
In mechanical ventilation, breaths are delivered to the patient by a ventilator with a proper tidal
volume (TV), respiration rate (RR), and inspiration / expiration ratio in time (I:E) determined by
the settings of the ventilator.
The Patient Spirometry monitors patient ventilation.
The following volume parameters are displayed:
Expiratory and inspiratory tidal volume (TV) in ml
Expiratory and inspiratory minute volume (MV) in l/min
Expiratory spontaneous minute volume in l/min
Inspiration/expiration ratio (I:E)
The following airway pressure parameters are displayed:
Peak pressure (P
Mean airway pressure (P
monitors
End inspiratory pressure (P
PEEPi, PEEPe; available only in S/5 Critical Care and Compact Critical Care monitors
Total positive end expiratory pressure (PEEP
Compact Anesthesia monitors
Real time airway pressure waveform (P
Static Positive end expiratory pressures (Static PEEPi and Static PEEPe); available only in
S/5 Critical Care and Compact Critical Care monitors
Static Plateau pressure (Static Pplat); available only in S/5 Critical Care and Compact
Critical Care monitors
Static Compliance (Static Compl); available only in S/5 Critical Care and Compact Critical
Care monitors
peak
)
); available only in S/5 Critical Care and Compact Critical Care
are measured by a pressure transducer on the MiniPVX board.
plat
Ambient pressure is used as a reference in measurement. The pressure measurement is made
from the airway part that is closest to the patient between the patient circuit and intubation
tube.
PEEP
=intrinsic PEEP, PEEP
i
-PEEP
tot
e
Static pressure measurement maneuvers are automatically identified based on an increased
zero flow period at the end of the inspiration or expiration.
Static Compliance is calculated, if Static PEEP and Static P
measurements were made within
plat
a 2 minute period.
The following airway flow parameters are displayed:
The measurement is based on measuring the kinetic gas pressure and is performed using the
Pitot effect. A pressure transducer is used to measure the Pitot pressure. The pressure signal
obtained is linearized and corrected according to the density of the gas. Speed of flow is
calculated from these pressure values and the TV value is then integrated. The MV value is
calculated and averaged using TV and RR (respiratory rate) values.
D-lite
Patient Spirometry uses specific sensors called D-lite+/D-lite and Pedi-lite+/Pedi-lite flow
sensors. Different types of sensors are available: adult sensor for measuring adults and
pediatric sensor for children. Both are available as reusable and disposable versions.
D-lite and Pedi-lite adapters are designed to measure kinetic pressure by a two-sided Pitot
tube. Velocity is calculated from pressure difference according to Bernoulli's equation. Flow is
then determined using the calculated velocity.
(from Bernoulli's equation)Formula 1
where:
2
V’ = flow (l/min), v = velocity (m/s), A = cross area (m
= density (kg/m3)
), dP = pressure difference (cmH2O),
Finally, the volume information is obtained by integrating the flow signal.
Compliance and airway resistance
Compliance is calculated for each breath from the equation
Formula 2
Compliance describes how large a pressure difference is needed to deliver a certain amount of
gas to the patient.
The airway resistance, Raw, is calculated using an equation that describes the kinetics of the
gas flow between the lungs and the D-lite. The equation states that the pressure at the D-lite
can at any moment of the breath be approximated using the equation
11
Document no. M1214853C
Page 22
Datex-Ohmeda E-Modules
where P(t), V’(t) and V(t) are the pressure, flow and volume measured at the D-lite at a time t,
Raw is the airway resistance, Compl is the compliance and PEEP
expiratory pressure (PEEP
2.2Main components
The respiratory modules consist of:
•Gas sampling system
•MiniTPX measuring unit
•MiniOM measuring unit
•MiniPVX measuring unit
•CPU board
2.2.1Controls and connectors
Formula 3
+PEEPi is the total positive end
e
).
tot
Figure 6Front of CARESCAPE Respiratory Module, E-sCAiOV, and the back of
the module
(1) D-fend Pro water trap
12
Document no. M1214853C
(2) Gas sample, sampling line connector on the water trap
(3) Water trap container
(4) Connectors for Patient Spirometry
(5) Gas exhaust, connector for the gas exhaust line (sampling gas out
Save LoopE-sCOV, E-sCAiOVSave Loop saves a reference
Change LoopE-sCOV, E-sCAiOVChange Loop changes a
ConnectorModuleDescription
D25 connectorall modulesModule bus connector
2.2.2Gas sampling system
The gas sampling system draws a 120ml/min sample from the patient's airway to the
module. The sampling system also takes about 30ml/min flow of room air to the oxygen
sensor. When the gas sensors are zeroed, room air is taken through the CO
the gas sensors instead of the sampled gas from the patient's breathing.
The gas sampling line is connected between the patient circuit and the Gas Sample port
on the water trap. The water trap protects the sampling system and gas sensors from
liquids and dust.
The diagram of the gas sampling system is shown in the figure below:
pressure/volume loop to a
flow/volume loop or vice
versa.
-absorber to
2
Figure 7Gas sampling system
13
Document no. M1214853C
Page 24
Datex-Ohmeda E-Modules
The sampling system has a self diagnostics that detects disturbances in the gas flow, reveals
the most common reasons for disturbances, such as occluded sampling line or blocked gas
exhaust line, and communicates relevant status messages to the patient monitor.
The system is designed so that gas the sampled gas will not flow from the sampling line back to
the patient circuit. The parts and connections of the sampling system are streamlined for
minimal dead spaces and turbulences in gas flows.
All gas inputs of the module have dust filters protecting the sampling system and gas sensors.
The water trap acts as a dust filter for the sampled gas and the module should always have the
water trap connected.
NOTE: It is very important to prevent dust from entering the open gas connections during
service operations.
D-fend Pro(+) water trap
The gas sampling line is connected to the input of the water trap where a special membrane
passes gases and vapors but stops liquids. The gas flowing through the membrane continues
via the main flow connector of the water trap to the module. The main flow is about 90% of the
sample flow.
Liquids stopped below the membrane are moved to the water container by a side flow that
goes through the water container and the water separation membrane before entering the
side flow connector of the water trap. Thus, the side flow also is free of liquids when it gets into
the module. In the module, the side flow is connected directly to the pump input and it does not
enter the gas sensors.
NOTE: The water trap acts as a dust filter for the sampling system and gas sensors. Thus, the
module should always have the water trap connected.
Zero valve and CO2 absorber
The zero valve is activated during gas sensor zeroing. Room air is drawn through the
CO
-absorber and the zero valve to the gas sensors, and the main flow of sample gas is
2
stopped. The zero gas comes to the sensors through the CO
. The side flow of the water trap flows in the gas sampling line even during zeroing.
CO
2
During normal monitoring, the zero valve is not activated and the sampled gas gets through
the zero valve to the gas sensors.
Figure 8Absorber
Nafion tubes
The Nafion tube between the water trap and the zero valve equalizes the humidity of the
sampled gas to ambient level. This will prevent calibration errors caused by the difference in
humidities in the sampled breathing gas and the totally dry calibration gas.
Another Nafion tube is used between the CO
condensation of water generated in the CO
1 Nafion is a registered trademark of Perma Pure Inc.
After the zero valve, the gas flows trough the MiniTPX sensor that measures the concentrations
of all gases but oxygen.
The oxygen concentration is measured in the MiniOM sensor that has two inputs. One input
draws in a part of the main flow and the other draws in room air as reference gas for the O
measurement.
2
Sample flow differential pressure transducer
The module measures total flow at the input of the gas pump and reference flow at the OM
reference line. The sample flow is the difference of these two flows.
Working pressure transducer
The working pressure transducer measures absolute working pressure near the MiniTPX unit
and MiniOM unit. It is used for messages: ‘Sample line blocked’, ‘Check D-fend’, ‘Replace D-fend’
and ‘Check sample gas out’.
Pneumatics unit
The pneumatics unit contains the zero valve, the occlusion valve and the pneumatics block
with tubing connections.
The zero valve is activated during the zero level calibrations of gas sensors. The occlusion and
zero valves are activated when the sampling line or water trap is occluded. With the activated
valves, the gas pump generates maximal suction trough the “side flow” connector of the water
trap, thus maximizing the transfer of liquids from the wet side of the water trap to the
container.
The pneumatics block contains a network of constrictions to divide the sampled gas in correct
proportions to different parts in the module. The first branching takes place in the water trap
where incoming flow is divided to the “main flow” and “side flow”. The second branching takes
place before the MiniOM sensor.
The pneumatics block also contains a pneumatic low pass filter between gas sensors and gas
pump. The filter consists of constrictions (resistors) and volumes (capacitors) and it attenuates
the pressure pulsation generated in the gas pump so that they do not disturb the operation of
the gas sensors.
Gas pump unit
The gas pump is a membrane pump run by a brushless DC-motor. The pump is adjusted so
that the sample gas flow is kept close to its nominal value even when the flow resistances in
the sampling line of water trap change.
15
Document no. M1214853C
Page 26
Datex-Ohmeda E-Modules
)LOWHU
&2
$EVRUEHU
0LQL73;
8QLW
0LQL20
8QLW
)DVW)HQG
=HUR
9DOYH
6DPSOH
OLQH
2FFOXVLRQ
9DOYH
9ROXPH
9ROXPH
3XPS
8QLW
1DILRQ
1DILRQ
5RRP
$LU
*DV
5HWXUQ
3QHXPDWLF
%ORFN
PLQL73;
:RUNLQJ
3UHVVXUH
%
PLQL20
:RUNLQJ
3UHVVXUH
%
7RWDO
JDVIORZ
3UHVVXUH
VHQVRU
%
$PELHQW
3UHVVXUH
%
The pump is in a plastic enclosure to minimize the operating noise and mechanical vibration of
the pump unit. A pneumatic damping chamber is integrated to enclosure to attenuate the
pressure pulsation and noise conducted to the gas exhaust port.
Pressure measurements
The four pressure sensors on the CPU board are used to measure ambient pressure, working
pressure of the MiniTPX and MiniOM sensors and pressure of the reference gas flow to the
MiniOM sensor.
Sample flow control
The gas flow in the sampling line is monitored by measuring the gas flow at the input of the gas
pump and the reference flow to the oxygen sensor is estimated by measuring the pressure in
the reference gas flow branch. The sample flow is calculated by subtracting the reference flow
from the total gas flow. A control loop adjusts the rotation speed of the pump motor so that the
gas flow is kept close to 120ml/min.
Gas sampling self-diagnostics
The sample flow and the vacuum in the sampling system are used for continuous monitoring
of the gas sampling system. The vacuum is calculated in real time as difference of the
measured ambient and working pressures.
The self-diagnostics of the gas sampling system sends the following status data to the patient
monitor when specific triggering conditions are met: ‘Check water trap’, ‘Check sample gas
out’, ‘Replace water trap’, ‘Sample line blocked’ and ‘Continuous blockage’.
The gas pump is stopped when the 'Sample line blocked' has lasted for more than 1 minute.
The module automatically restarts the pump to check whether the abnormal situation has
been resolved so that normal gas sampling operation is possible.
The gas pump repeats 1 minute full pump, 30 seconds pump off when the ‘Continuous
blockage’ message is shown.
Figure 9Gas tubing layout
2.2.3MiniTPX measuring unit
The MiniTPX unit is a non dispersive infrared analyzer, measuring the absorption of the gas
sample at seven infrared wavelengths, which are selected using optical narrow band filters.
The IR source is a micro-machined heating element with an integrated collimator. From the
output of the source, the radiation is passed to a flow optimized measuring chamber.
From the sample chamber, radiation goes via a specially designed beam splitter to two
detector units, each with four thermopile detectors and integrated optical filters. The miniTPX
measuring unit has two detector units for redundancy purposes. A more detailed description of
the measuring principle can be found in section “2.1.1. CO2, N2O, and agent measurement”.
Each detector unit also measures the unit's temperature. The module CPU uses it for further
processing and temperature compensation of the measured raw signals.
The miniTPX unit includes an amplifier board with the following functions:
•On-board 5V regulator and 2.5V reference source.
•Preamplifiers for the eight thermopile detectors and for the two temperature sensors.
A 16 channel buffered multiplexer is used to transfer the signals to the CPU board.
•PWM controlled power for the IR source.
•An EEPROM memory for storing factory calibration coefficients of the sensor.
The input to the amplifier board comprises a 7V DC feed and CPU control signals for the PWM,
the multiplexer and the EEPROM. When the module starts up, the calibration coefficients are
read to the module CPU and then used for calculating the gas concentrations from the raw
data received from the sensor multiplexer.
Figure 10MiniTPX measuring unit
2.2.4MiniOM Oxygen sensor
The miniOM sensor measures the concentration of Oxygen in the gas sample.
The measurement is based on the magnetic properties of oxygen. The sensor measures the
sound pressure generated in the air gap of the magnet at the 164Hz operating frequency. Two
microphones are used for detection and the Oxygen concentration is calculated from the RMS
value of the difference of the microphone outputs. The measurement principle is described in
more detail in section ”2.1.2. O2 measurement”.
The sensor consists of the following functional parts
•Pneumatic system
•Amplifier board
•MiniOM board
•Magnet
17
Document no. M1214853C
Page 28
Datex-Ohmeda E-Modules
The sensor is shown in the picture below.
Figure 11MiniOM oxygen sensor
NOTE: The sensor is assembled in the module using flexible suspension to prevent the
mechanical vibrations of the gas pump and cooling fan from disturbing the Oxygen
measurement. All gas lines to the sensor must also be carefully assembled so that they do not
pick up mechanical vibrations of the module mechanics.
Pneumatic System
The pneumatic system, together with the gas sampling system of the module creates the gas
flows and pressures needed for the oxygen measurement and protection of the microphones
from excessive pressure. About 30 ml/min flow of sampled gas comes to the In connector on
the MiniOM sensor. Room air is drawn to the Ref input of MiniOM also at 30 ml/min rate. About
75% of these flows are conducted to a pressure equalization chamber so that only about a
8 ml/min flow of the two gas streams continue into the air gap of the magnet. All the internal
gas flows finally get to a volume enclosed by the sensor board and the sensor body, and then
flow out through the Out connection of the sensor. Some of the gas channels and flow
restrictors are integrated into the preamplifier electronics board utilizing the multi-layer
structure of the LTCC (Low Temperature Co-fired Ceramics) circuit board technology.
NOTE: It is very important to prevent dust or liquids from getting into the pneumatic circuit of
MiniOM and thus, the gas connections should always be closed with a protecting cap when the
sensor is not connected to the module pneumatics.
Amplifier Board
The amplifier board located in the sensor has two electric microphones for the differential
detection of pressure pulses generated in the magnet's air gap. The microphone signals are fed
to two identical signal conditioning channels with a band-pass filter and a digitally controlled
amplifier. The voltage gains of the amplifiers are set during factory calibration so that the
responses of the microphone channels match in spite of differences in microphone's
sensitivities. The amplifier board also has an amplifier for the thermistor measuring the
temperature of the magnet.
MiniOM Board
The MiniOM board has five functions
•Drive the magnet coil.
•Convert the microphone and temperature signals into digital format.
•Filter digitally the microphone signals and perform the RMS-conversion.
•Communicate digitally with the module CPU.
18
Document no. M1214853C
Page 29
•Store factory calibration data in permanent memory and communicate them to the
module CPU.
The module CPU provides the coil drive and communication enabling signals and also clock
signal for MiniOM board. The FPGA takes care of the coil drive and has also back-up clock in
case of CPU clock does not work. The FPGA takes care of the A/Dconversions which are
performed with a serial controlled SAR A/D-converter.
The digital band pass filtering and RMS conversion of the microphone signals is made with
FPGA circuit controlled by the VHDL code stored in the circuit. In order to filter out the
disturbances caused by acoustic noise, mechanical vibration and amplifier noise, the band
pass filters are designed to have as narrow a pass band as possible without slowing down the
filter's response to changes in the amplitude of the 164 Hz signal.
The FPGA circuit takes care of the digital communication between the miniOM sensor and the
module CPU.
The factory calibration coefficients of the sensor are stored in an EEPROM memory on the
miniOM board. When the module starts up, the calibration coefficients are read to the module
CPU and then used for calculating the O2 concentration from the Oxygen raw data received
from the sensor.
NOTE: Never apply the overpressure or negative pressure of more than 300 cmH2O to the flow
and volume tubing. Differential pressure max 25 cmH
when connecting tubes.
When Patient Spirometry is used, a special sensor, D-lite, replaces the normal airway adapter
in the patient circuit. A double lumen tubing is attached to the two connectors on the adapter
and on the module front panel.
The Patient Spirometry provides patient respiration monitoring capabilities using the D-lite and
Pedi-lite flow sensors.
O is allowed on one port at a time e.g.
2
Figure 12MiniPVX measuring unit
The measurement is based on measuring the kinetic gas pressure and is performed using the
Pitot effect. A pressure transducer is used to measuring the Pitot pressure. The signal is then
linearized and corrected according to the density of the gas. Speed of the flow is calculated
from the pressure and TV is integrated from it.
19
Document no. M1214853C
Page 30
Datex-Ohmeda E-Modules
'&'&
3RZHU
6XSSO\
&38
$WPHO
$750
56
WUDQVFHLYHU
&U\VWDO
0+]
&U\VWDO
N+]
0RG%XVB'DWD
0RGXOH
%XV
&RQQHFWRU
'
$QDORJ
WR
'LJLWDO
&RQYHUVLRQ
=HUR9DOYH
&RQQHFWRU
2FFOXVLRQ
9DOYH
&RQQHFWRU
)DQ
&RQQHFWRU
3UHVVXUH
VHQVRUV
P73;
&RQQHFWRU
P20
&RQQHFWRU
P39;
&RQQHFWRU
3XPS
&RQQHFWRU
)DQ
&RQWURO
9DOYH
&RQWURO
+XPLGLW\
7HPSHUDWXUH
VHQVRU
(7&9DOYH
&RQQHFWRU
3XPS
&RQWURO
0HPRULHV
125)/$6+
0E\WH
6'5$0
0E\WH
9
56
9
9
9
9
9&&B20
8VHU%XWWRQ
&RQQHFWRU
Patient Spirometry consists of airway connections, two pressure transducers, valves and
preamplifiers. The preamplifiers are connected to the A/D-converter on the module main CPU.
The patient’s breathing flow passing through the D-lite adapter creates a pressure difference.
This pressure difference is measured by a pressure transducer, B1. Overpressure and negative
pressure in airways are measured by another pressure transducer, B2.
2.2.6CPU board
The CPU board contains the processor, memories and an A/D-converter that is common to the
whole module.
The CPU board also contains sensors for pressure, temperature and humidity as well as drivers
for valves, the fan and the pump. The module is connected to the module bus through an
RS-485 serial channel.
Figure 13Signal processing on CPU board
2.2.7MiniOM board
20
Document no. M1214853C
The miniOmM board contains electronics specific to the MiniOM sensor: FPGA circuit, coil drive,
A/D-converter etc. It also contains EEPROM memory that stores calibration data of the oxygen
measurement.
The MiniPVX board contains pressure sensors for airway pressure and flow measurement and
preamplifiers for those. It also contains EEPROM memory that stores calibration data of the
spirometry measurement.
2.2.9Main Component Interactions
The figure below describes the functionality of the module and the division of tasks between
different components.
To help ensure the equipment remains in proper operational and functional order, adhere to a
good maintenance schedule.
WARNINGOnly perform maintenance procedures specifically described in the manual.
WARNINGPlanned maintenance should be carried out annually. Failure to implement
the recommended maintenance schedule may cause equipment failure and
possible health hazards.
CAUTIONDo not apply pressurized air to any outlet or tubing connected to the module.
NOTE: The manufacturer does not, in any manner, assume the responsibility for performing the
recommended maintenance schedule, unless an Equipment Maintenance Agreement exists.
The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device.
Corrective maintenance
Service personnel shall perform the following checkout procedure after any corrective
maintenance, before taking the module back into clinical use:
Required checkout procedure
Performed service activity
Front panel replacementAll stepsCheck “Module Keys” only
OM Reference gas filter
assembly
Module case opened either for
troubleshooting purpose or for
replacing any of the internal
parts.
Visual inspections
(section 3.2)
All stepsCheck “Sample Flow Rate Check”
All stepsAll steps
Functional check
(section 3.3)
Planned maintenance
Service personnel shall perform the following checkout procedure completely every 12 months
after installation:
Replace the following parts that wear in use at the recommended interval.
DescriptionPiecesReplacement interval
Nafion Tube, 230 mm (mainflow)1Once a year
OM Reference gas filter assembly including
O-ring
PM sticker1Once a year
Nafion tube, 85 mm (zero line)1Once every 4 years
CO
absorber1Once every 4 years
2
It is also recommended to replace the D-fend Pro water trap, the gas sampling line and the
spirometry tube as part of the planned maintenance procedure.
NOTE: See
compatible accessories.
the supplies and accessories document delivered with the manual for
3.1.2Planned Maintenance Kits
The required planned maintenance parts are included in a PM kit.
Part numberDescription
M1206554Planned Maintenance Kit for CARESCAPE E-sCxxx Respiratory modules.
The PM kit includes the required Nafion tubes, the OM reference gas filter
assembly with an O-ring and a PM sticker.
NOTE: The PM kit does not include the CO
1Once a year
absorber. Order it separately.
2
3.1.3Replacement procedures
Replace the specified planned maintenance parts according to the chapter “3.4. Disassembly
and reassembly”.
1.Replace the CO
2.Replace the special tubes (Nafion) and check the condition of the internal tubing.
absorber every 4 years.
2
•Check that the tubing inside the module is not contaminated. Any contamination
inside the tubing may indicate that the valves or sensors are contaminated, too. This
can increase a risk of faulty operation in valves or sensors. The valves or gas sensors
are not possible to clean in the field. Therefore, if you noticed any contamination in
the module tubing, send the module to GE Healthcare for factory service.
NOTE: The nafion tubes do not include the silicon fittings they connect to. Use the original
silicon fittings unless they are damaged or leaking.
3. Replace the OM reference gas filter assembly.
4.Check that the fan and ventilation hole are not covered in dust.
3.2Visual inspections
Detach the module from the module slot and check that:
•the front cover is intact
•all connectors are intact and are attached properly
•the module box and latch are intact
•the D-fend Pro and its connectors are clean and intact
•the module and the applied parts are clean
The cleaning precautions, cleaning requirements, cleaning procedures, and recommended
cleaning solutions for the monitor are described in the S/5 monitor user’s manual. For details
about cleaning, disinfecting and sterilizing the accessories, see the instructions for use in the
accessory package.
24
Document no. M1214853C
Page 35
3.3Functional check
Turn the monitor on. Wait until the normal monitoring screen appears.
3.3.1Test setup
Required tools
A barometer
A mass flowmeter for measuring air flow, minimum measurement range from 0 to
200ml/min, accuracy 5% or better in the 0 to 200 ml/min range.
P/N: 755534-HEL Calibration Gas Regulator
P/N: 755583-HEL Calibration gas, CO2, O2, N2O, DESF, package of 1 can (with E-sCAiO,
E-sCAiOV modules)
P/N: 755581-HEL QUICK CAL calibration gas, CO2, O2, N2O, package of 4 cans (with
E-sCO, E-sCOV modules)
P/N: M1006864, Calibration Gas Regulator, US only
P/N: 755571-HEL, Calibration Gas, 5% CO2, 54.5% O2, 36.0% N2O, 2.0% DESFLURANE,
D-fend Pro water trap
3 m / 10 ft anesthesia gas sampling line
Spirometry tube, 3 m/10 ft (with E-sCOV and E-sCAiOV modules)
Adult D-Lite sensor
A pressure manometer with either an integrated or a separate pressure pump
Tubing for spirometry leak tests
Forceps
NOTE: See the supplies and accessories document delivered with the manual for compatible
accessories.
Connections
•
Monitor configuration
1.Configure the CO2, O2, AA, and Flow waveform fields to the monitor screen.
2.Configure the Spiro 1 split screen to the monitor screen.
3.Select the Spirometry Setup in the Airway Gas menu and configure:
3.3.2Procedure
Mark each task as complete on the checkout form.
1.Gas Sampling System Leak Test
NOTE: The gas module shall be disconnected from the monitor during the leak test.
Disconnect the module from the monitor, if connected.
Scaling: Auto
Sensor Type: Adult
TV or MV: TV
25
Document no. M1214853C
Page 36
Datex-Ohmeda E-Modules
Check the gas sampling system for possible leakages.
a.Disconnect the module from the monitor.
b.Detach the module front cover and casing, see chapters“Detaching the Front Cover”
c. Block the OM reference tube with the forceps. Correct positioning of the forceps is
on page 33 and “Detaching the Module Casing” on page 34.
indicated by the figure below.
NOTE: Be careful when attaching the forceps to the tube and avoid stretching the
tube. Short pieces of silicone tubing on the forcep jaws can be used to protect the
tube from breaks that may appear when the tube is compressed between the jaws.
d.Connect a new D-fend Pro water trap to the module.
e.Connect a new gas sampling line to the sampling line connector in the water trap.
f.Connect the other end of the gas sampling line to a pressure manometer and a
pressure pump.
g.Block the sample gas out (gas exhaust) connector.
h.Carefully pump 80 mmHg 20 mmHg pressure to the gas sampling system. Let the
pressure stabilize for 10 - 20 seconds.
i.Check that the pressure reading does not drop more than 2 mmHg during 25
seconds.
j.Release the forceps, and reassemble the module casing. Make sure that the tubing
fits nicely into the module casing.
2.Spirometry System Leak Test
NOTE: Perform this test only for E-sCOV and E-sCAiOV modules.
NOTE: The gas module shall be disconnected from the monitor during the leak test.
NOTE: The spirometry pressure transducers are very sensitive for differential overpressure. A
momentary differential pressure between the two spirometry connectors exceeding 25 cmH
(18 mmHg) may damage the pressure sensors. To ensure that both pressure channels are
equally pressurized, make sure that the tubing between the manometer and the two
spirometry connectors is connected tightly, the tubes are equally long and thick and not
kinked.
NOTE: Do not overpressure the spirometry sampling system. A static pressure exceeding
300 cmH
O (220 mmHg) may damage the pressure sensor.
2
Check the spirometry sampling system for possible leakages.
a.Ensure the module is disconnected from the monitor.
b.Connect a pressure manometer to the spirometry connectors.
c.Pump ~68 cmH
O (50 mmHg ±10 mmHg) pressure to the Spirometry sampling
2
system. Let the pressure stabilize for approximately 10 seconds.
d.Verify that the pressure reading does not drop more than 4 cmH
O (3 mmHg) during
2
one minute.
3.Sample Flow Rate Check
Check the sample flow rate.
Connect the module to the monitor.
NOTE: Anesthetic gas measurement is not available during the first 1 to 5 minutes after the
module is connected due to warming up. A message 'Calibrating Gas Sensor' is shown in the
waveform field. Wait until warm-up is completed before proceeding with the next steps.
NOTE: The ambient temperature and air pressure influence the flow rate measured by the flow
meter. A flow meter, which has been calibrated at 21.11 °C (70 °F) and 760 mmHg
(1033 cmH2O), measures the flow rate correctly under the same conditions, i.e. in room
temperature at sea level. A flow rate correction as instructed by the manufacturer of the flow
meter needs to be performed when measuring flow rate under other conditions, for example in
high altitude.
a.Connect the gas sampling line(3 m / 10 ft with E-sCO, E-sCOV, E-sCAiO and
E-sCAiOV) to the Sampling line connector.
b.Connect the other end of the gas sampling line to a flowmeter.
c.Check the sample flow rate reading from the flowmeter. The flow rate shall be within
the specification limit 120 ± 20 ml/min.
NOTE: Readjustment is needed, if the measured value is not within the specification limit. Adjust
the sample gas flow rate according to the instructions in section “3.5.1. Sample Flow Rate
Adjustment”.
27
Document no. M1214853C
Page 38
Datex-Ohmeda E-Modules
4.Reference Gas Flow Rate Check
Check the flow rate in reference gas inlet:
a.Connect the module to the monitor.
b.Leave the other end of the gas sampling line open to room air.
c.Connect the flowmeter to the OM reference gas inlet on the side of the module with
d.Check that the Reference Flow is within the following range:
e.Detach the water trap.
f.Attach the front cover.
5.Fan
a.Check that the gas module's fan is running behind the D-fend Pro water trap.
b.Attach the water trap.
6.Module Keys
NOTE: Perform this test only for E-sCOV and E-sCAiOV modules.
a piece of tubing.
10 - 50 ml/min with E-sCO, E-sCOV, E-sCAiO and E-sCAiOV modules
a.Press the
b.Check that the spirometry loop is changed from Flow / Vol loop to Paw/Vol loop, or
vice versa.
c.Leave the Flow / Vol loop on the screen.
7.Zero Valve Operation
Test the zero valve functionality:
a.Connect the gas regulator to the calibration gas container.
b.Connect the end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
c.Select
d.Start feeding the specified calibration gas. Wait until the gas values shown in the
Gas calibration menu rise approximately to the level indicated in the labelling of the
calibration gas container.
NOTE: The gas values in the Gas Calibrations menu is in percentages (%).
e.Open the zero valve to room air by selecting Zero valve Ctrl > Zero (zero position).
f.Check that the CO2, N2O and anesthesia agent values drop back near 0% and the
O2 reading near 21% (room air).
g.Stop feeding the calibration gas.
h.Turn the zero valve back to the normal measurement position by selecting Zero
valve Crtl > Meas (measurement position).
Change Loop module key.
Monitor Setup> Install/Service >Service > Parameters > Gas out > Gases
8.Gas Calibration
Perform gas calibration according to the instructions in section “3.5.2. Gas Calibration”.
9.Agent Identification
NOTE: Perform this test only for E-sCAiO and E-sCAiOV modules.
a.Feed the specified calibration gas for at least 30 seconds.
b.Check that the anesthesia agent is identified as Desflurane and the ID unrel value
(=agent ID unreliability) shown in the
Parameters > Gas Unit > Gases menu is lower than 75.
If the value is higher, repeat the gas calibration and check the value again.
10. Ambient Pressure
Use a barometer to check the operation of the absolute pressure sensor.
Monitor Setup > Install / Service > Service >
•Check that the ambient pressure value shown in the Gas Calibrations menu does not
differ more than ± 10 mmHg from the value shown by the barometer.
NOTE: The ambient pressure value in the Gas Calibrations menu is in mmHg.
11. Occlusion detection
a.Block the tip of the sampling line by your finger.
b.Check that a 'Sample line blocked' and a ‘Low sample flow’ message appear on the
screen within 30 seconds.
12. Air Leak detection
a.Detach the D-fend Pro water trap.
b.Check that the message 'Check Water Trap' appears on the screen within 30
seconds.
c.Attach the water trap.
13. Gas exhaust blockage
a.Block the gas exhaust connector with your finger.
b.Check that the message 'Sample gas out' appears on the screen within 30
seconds.
14. Airway Gases
a.Breathe a minimum of 5 times to the tip of the sampling line.
b.Check that a normal CO2 waveform appears and the EtCO2 and FiCO2 values are
updated on the screen .
15. Apnea detection
a.Stop breathing to the gas sampling line.
b.Check that an 'Apnea' alarm appears on the screen within 30 seconds.
16. Flow waveform
NOTE: Perform this test only for E-sCOV and E-sCAiOV modules.
a.Connect a clean spirometry tube and D-lite to the module.
b.Breathe through the wider side of the D-lite.
c.Check that the flow waveform responds when you breathe in and out. The setting of
the inspiratory flow may be positive or negative.
29
Document no. M1214853C
Page 40
Datex-Ohmeda E-Modules
3.3.3Test completion
Select Discharge patient or Reset case to discard any changes made to the monitor
•
configuration during checkout.
•Disconnect and reconnect the module before starting a new case.
•Complete the “Appendix A. Service check form”.
3.4Disassembly and reassembly
3.4.1Disassembly guidelines
WARNINGDisconnect the module from any monitoring system before performing any
repair.
WARNINGAlways perform gas sampling system leak test after the module cover is
reassembled.
WARNINGAlways perform gas calibration after any planned or corrective
maintenance.
Field service of the module is limited to replacing the serviceable parts listed below (see also
chapter ”5. Spare parts”). Attempting a field repair on any other parts could jeopardize the safe
and effective operation of the module, and void the warranty.
NOTE: Only a qualified service technician should perform field replacement procedures.
NOTE: Perform the checkout procedure described in chapter ”3. Service Procedures” after you
have disassembled and reassembled the module.
Serviceable parts
CO2 Absorber
•
•D-fend Pro
•Nafion tubes
•Front chassis unit
•MiniPVX Unit
•Pump
•OM reference filter
•Latch and spring
•Mechanical parts listed in chapter ”5. Spare parts”
Service limitations
The following parts are not serviceable:
•MiniOM Measuring unit
•MiniTPX measuring unit
NOTE: Due to the complicated and sensitive mechanical construction of the oxygen measuring
unit, no repairs should be attempted inside the unit. Instead, if the fault has been found in the
measuring unit itself, the entire module should be replaced and the faulty module be sent to
GE Healthcare for repair.
NOTE: The MiniTPX measuring unit can only be repaired and calibrated at the factory. In case of
failure, the entire module should be replaced and the faulty module be sent to GE Healthcare
for repair.
ESD precautions
WARNINGProtect module from electrostatic discharge.
All external connectors of the module are designed with protection from ESD damage.
However, if the module requires service, exposed components and assemblies inside are
susceptible to ESD damage. This includes human hands, non-ESD protected work stations or
improperly grounded test equipment. The following guidelines may not guarantee a 100%
static-free workstation, but can greatly reduce the potential for failure of any electronic
assemblies being serviced:
•Discharge any static charge you may have built up before handling semiconductors or
assemblies containing semiconductors.
•A grounded, antistatic wristband or heel strap should be worn at all times while handling
or repairing assemblies containing semiconductors.
•Use properly grounded test equipment.
•Use a static-free work surface while handling or working on assemblies containing
semiconductors.
•Do not remove semiconductors or assemblies containing semiconductors from antistatic
containers until absolutely necessary.
•Do not slide semiconductors or electrical/electronic assemblies across any surface.
•Do not touch semiconductor leads unless absolutely necessary.
•Semiconductors and electronic assemblies should be stored only in antistatic bags or
boxes.
•Handle all PCB assemblies by their edges.
•Do not flex or twist a circuit board.
Protection from dust
WARNINGModule must be handled to prevent dust from entering the gas sampling
system.
The gas sampling system must be protected from dust entering the tubes, valves and other
components. In order to achieve this goal, the following measures must be taken:
•Have the D-fend Pro water trap always connected to the module.
•Clean and dust free working environment during all service procedures.
•Minimize the times with any open connections in the gas sampling system.
•Always close the open tube connections of the sampling system when not working on the
module.
•Remove the protective caps on the gas pump only immediately before assembling it to
the module
31
Document no. M1214853C
Page 42
Datex-Ohmeda E-Modules
•Take the CO
the module.
-absorber out from the plastic bag only immediately before assembling it to
2
•The clothing of the service person must be such that the dust risk is taken into account.
Before disassembly
Note the positions of any sampling tubes, wires or cables. Mark them if necessary to
•
ensure that they are reassembled correctly.
•Save and set aside all hardware for reassembly.
Required tools
- Torx T8 and T10 screwdrivers
- flat blade screwdriver
-forceps
- antistatic wristband
3.4.2Disassembly and reassembly procedure
Disassembling the module (see the exploded view of the module in chapter ”5. Spare parts”:
Reassembling the module: reverse the order of the disassembly steps.
Check that:
•all screws are tightened properly
•all cables are connected properly
•tubes are not pinched and there are no sharp bends on them
•all tubes are connected properly
NOTE: Make sure that the Nafion tubes are routed in such a way that they don’t come near the
fan, and there is no risk of the fan being obstructed by the tubes. An obstructed fan will result in
degraded ventilation inside the module, and a ‘Sensor inop’ message being displayed.
Use this workflow diagram to find the simplest way to disassemble the required parts of the
module. Follow the arrows from the top down to the required part and disassemble the module
by following the steps in between.
Detaching the Front Cover
1. Remove the D-fend Pro.
2. Release the two snaps on both sides of the
module by using a flat blade screwdriver.
3. Detach the front cover.
33
Document no. M1214853C
Page 44
Datex-Ohmeda E-Modules
Detaching the Module Casing
1. Remove the two T8 screws mounting the D25
connector shield.
2. Detach the connector shield.
3. Remove the two T8 screws to detach the module
casing.
4. Push the latch and pull the module casing.
NOTE: When reassembling ensure that the module
casing does not damage the conductive sealings on
the front chassis unit.
1. Carefully remove the main flow nafion tube and
every 4
th
year the shorter zero line nafion tube.
NOTE: Remember the route of the tubes and
reassemble correctly.
NOTE: Make sure that the Nafion tubes are routed in
such a way that they don’t come near the fan, and
there is no risk of the fan being obstructed by the
tubes. An obstructed fan will result in degraded
ventilation inside the module, and sensor inop
message being displayed.
NOTE: The nafion tubes do not include the silicon
fittings they are connected to. Use the original
silicon fittings unless they are not damaged or
leaking.
2. Pull out the OM reference filter assembly with
forceps.
3. Push the new filter assembly until it is on the
same level with the front chassis.
Replacement of CO
Absorber
2
1. Lift the CO
absorber from the slot.
2
2. Detach the tubes from the absorber.
3. Connect the tubes to a new CO
absorber and
2
place it to the slot.
35
Document no. M1214853C
Page 46
Datex-Ohmeda E-Modules
Detaching the Latch
1. Pull the latch from the front chassis.
NOTE: Remember to detach the front cover first.
NOTE: When reassembling insert the pump silicone
tube in the front chassis connector.
Detaching the Main Flow Connector
Original Main Flow Connector is required to maintain proper gas flow restriction in the module.
When the Front Chassis Unit is replaced move the original connector to the new unit.
1. Carefully detach the lock pin holding the main
flow connector.
2. Carefully attach the main flow connector to the
new front chassis unit.
1. Carefully detach the three tubes from the OM
unit.
2. Carefully disconnect the OM flex cable from the
CPU board.
3. Carefully pass the flex cable through metal
frame.
4. Lift the OM unit with the holder from the metal
brackets.
5. Detach the holder.
41
Document no. M1214853C
Page 52
Datex-Ohmeda E-Modules
3.5Adjustments and calibration
3.5.1Sample Flow Rate Adjustment
Sample flow rate shall be adjusted:
•if the sample flow rate check in section “3.3.2. Procedure” failed.
Calibration setup
Required tools
A mass flowmeter for measuring air flow, minimum measurement range from 0 to
200ml/min, accuracy 5% or better in the 0 to 200 ml/min range.
3 m / 10 ft anesthesia gas sampling line.
NOTE: See the supplies and accessories document delivered with the manual for compatible
accessories.
NOTE: Use only accurate, properly maintained, calibrated and traceable calibration tools for
the parameter calibration to ensure measurement accuracy.
NOTE: If the flow meter unit is not ml/min, it shall be converted to ml/min according to the
instructions of the flow meter manufacturer.
NOTE: Gas module sample flow rate is calibrated in the factory to ambient air conditions
corresponding the flow at the end of 3 m sampling line. Make sure that your meter is also
showing the flow at ambient conditions (= ATP).
NOTE: Refer to the flowmeter documentation for user instructions.
Connections
1.Ensure that the module is connected to the monitor.
2.Ensure that you have a new D-fend Pro water trap in use.
3.Connect a new gas sampling line to the sampling line connector in the water trap.
4.Connect the other end of the gas sampling line to the flow meter.
NOTE: Before checking or adjusting the sample flow, make sure there is no leakage in the
sampling system.
Sample Flow Rate Adjustment
1.Select Monitor Setup>Install / Service > Service > Parameters > Gas Unit > Gases
2.Select Sample gain adj.
3.Adjust the sample flow to the nominal value 120 ml/min by increasing or decreasing the
Sample Flow Gain:
NOTE: On S/5 monitor screen the 120 ml/min flow rate is shown as 200 ml/min to be
compatible with the old S/5 monitors.
•To decrease the sample flow rate measured by the flow meter by approximately
7,5 ml / min, add gain value by 0.05.
•To increase the sample flow rate measured by the flow meter by approximately
4.Press Confirm to check the effect of the gain adjustment. Wait until the sample flow value
shown in the Calibration menu returns near to the nominal value 120 ml/min and then
check the actual measured flow rate from the flow meter.
5.Repeat steps 3 and 4 until the flow meter shows a 120 ± 20 ml /min flow rate.
NOTE: Adjust the flow rate according to the reading in the flow meter. The flow rate reading in
the Calibration menu is measured by the internal electronics and settles always back to the
nominal 120 ml /min independent on the real flow rate.
3.5.2Gas Calibration
WARNINGFailure in zeroing or calibrating gases might cause inaccurate readings.
WARNINGSince calibration gas contains anesthetic agents, always ensure sufficient
ventilation of the room during calibration.
Gas calibration shall be performed:
•each time planned maintenance is performed.
•each time corrective maintenance is performed.
NOTE: Gas calibration is a normal user action. Refer to the monitor user’s manual for the
recommendation for gas calibration interval in clinical use.
Calibration setup
Required tools
P/N: 755534-HEL Calibration Gas Regulator
P/N: 755583-HEL Calibration gas, CO2, O2, N2O, DESF, package of 1 can (with E-sCAiO,
E-sCAiOV modules)
P/N: 755581-HEL QUICK CAL calibration gas, CO2, O2, N2O, package of 4 cans (with
E-sCO, E-sCOV modules)
P/N: M1006864, Calibration Gas Regulator, US only
P/N: 755571-HEL, Calibration Gas, 5% CO2, 54.5% O2, 36.0% N2O, 2.0% DESFLURANE,
NOTE: Use only the specified GE Healthcare calibration gas for the gas calibration to ensure
measurement accuracy. Do not use any other calibration gases. Check the calibration gas
container's labelling to ensure that the calibration gas has not expired.
NOTE: Ensure that the gas regulator is functioning properly before gas calibration. Refer to the
gas regulator's "Instructions for Use" letter for the annual maintenance instructions.
Connections
1.Ensure that the module is connected to the monitor.
2.Ensure that you have a new D-fend Pro water trap in use.
3.Connect the gas regulator to the calibration gas container.
4.Connect a new gas sampling line to the sampling line connector in the water trap.
5.Connect the other end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
43
Document no. M1214853C
Page 54
Datex-Ohmeda E-Modules
Figure 15Connecting a gas regulator to the calibration gas container and
Procedure
NOTE: Gas calibration is not available during the first 5 minutes after the module is connected.
A message 'Gas calibration is not available during first 5 minutes' is shown in the lower left
corner of the Calibration menu. For maximum accuracy, let the monitor to warm up for 30
minutes before starting calibration.
NOTE: Gas calibration is not available during a 'Sample line blocked', 'Check Dfend' and 'Check
sample gas out’ alarm condition. A message 'Gas calibration is not available during gas
sampling warning' is shown in the lower left corner of the calibration menu. Resolve the alarm
condition before starting calibration.
connecting a sampling line to the gas regulator.
1.Select Airway Gas > Gas Calibration
2.The monitor will start automatic zeroing of the gas sensors. Wait until the message
'Zeroing' is replaced by a message 'Zero Ok' for all measured gases.
3.Open the regulator after a message 'Feed gas' is shown for all measured gases. The
measured gas concentrations are shown in real-time in the gas calibration menu.
Continue feeding the calibration gas until the measured gas concentrations are stabilized
and a message 'Adjust' is shown for all measured gases. Close the regulator.
4.Adjust the gas readings shown in the Calibration menu to match with the gas readings in
the labelling of the calibration gas container. Select Accept to accept the adjusted values
when the gas readings match each other.
5.Wait until a message 'Ok' is shown for all measured gases.
NOTE: A message 'Zero Error' is shown in case the zeroing fails.
NOTE: A message 'Calibration Error' is shown, if you do not start feeding gas within 1 minute
after the automatic zeroing is completed, or if the calibration fails due to too large gain
adjustment.
NOTE: If zeroing or calibration failed, select the
procedure from the beginning.
Recalibrate button to restart the calibration
44
Document no. M1214853C
Page 55
3.5.3Spirometry Calibration
Patient spirometry does not require regular service calibration during planned maintenance, or
after the MiniPVX unit has been replaced. Calibration is only needed if there is a permanent
difference between the measured inspiratory and expiratory volumes.
The MiniPVX measuring unit is calibrated at the factory and due to the unit's design, spirometry
calibration is not regularly needed in the field. The calibration data is saved into the board's
EEPROM.
If calibration is desired, it is recommended to perform the calibration both with adult values
using the D-lite, and with pediatric values using the Pedi-lite.
•Refer to the "Instructions for Use" -letter of the spirometry tester to see the setup.
3HGLDWULF
Monitor configuration
$GXOW
•Configure the Flow waveform field to the monitor screen with adequate priority.
•Select the Spirometry Setup in the Airway Gas menu and configure:
Scaling: Auto
Sensor Type: Adult
TV or MV: TV
45
Document no. M1214853C
Page 56
Datex-Ohmeda E-Modules
Ventilator configuration
•Configure the ventilator to use air as fresh gas.
•Set the Tidal Volume (TV) to 500 ml/min when doing calibration check and calibration with
adult sensor and 100 ml/min with pediatric sensor.
•Set the RR =15, I/E =1/2 and PEEP 0cmH2O.
Calibration check
1.Perform the calibration check according to the steps 1 through 12a in the "Instructions for
Use" -letter of the spirometry tester.
NOTE: Let the gas module to warm up at least for 10 minutes before performing the calibration
check or flow calibration.
2.The measured flow values are shown in real-time in the TV Insp and TV Exp fields in the
Flow parameter window. Compare these measured values to the TV value reading
(highest water level) in the spirometry tester.
Acceptance criteria:
•If the TV Insp and TV Exp values differ less than ± 6% of the value read from the
spirometry tester, flow calibration is not needed.
•If the TV Insp and TV Exp values differ more than ± 6% of the value read from the
spirometry tester, perform flow calibration according to section “Flow calibration”.
Flow calibration
1.Select Monitor Setup > Install /Service > Service > Parameters > Gas Unit >
Spirometry
2.Ensure that the Sensor Type is correct and that Spirometry Zeroing is Enabled.
3.Wait until the MiniPVX sensor performs an automatic zeroing. It will show a message
‘zeroing’ in the Flow parameter window when zeroing takes place.
4.Adjust the Exp Flow Gain and/or Insp Flow Gain separately to calibrate the measured TV Exp ml and TV Insp ml values:
To increase the TV Exp ml flow value, increase the Exp Flow Gain.
To decrease the TV Exp ml flow value, lower the Exp Flow Gain.
To increase the TV Insp ml flow value, increase the TV Flow Gain.
To decrease the TV Insp ml flow value, lower the TV Flow Gain.
5.Press Confirm to check the effect of the gain adjustment to the flow readings.
6.Repeat steps 3 and 4 until the flow values are within the specification.
46
Document no. M1214853C
Page 57
4Troubleshooting
The problems and solutions in this chapter represent only a few of the faults that you may
encounter and are not intended to cover every possible problem that may occur.
This chapter focuses on troubleshooting technical problems. Refer also to the troubleshooting
hints on the S/5 monitor user’s manual for troubleshooting monitoring problems, performance
issues and clinical configuration issues.
NOTE: Perform the checkout procedure described in chapter “3. Service Procedures” each time
after you have opened the module casing.
4.1Visual inspection
Before beginning any detailed troubleshooting, complete thorough visual inspection to be sure
that:
•the front cover is intact
•the water trap connection and disconnection functions properly
•all connectors are intact, clean and are attached properly
•the metal D-fend Pro connectors are clean and intact
If in doubt of having any loose parts or cable connections inside the module, detach the
module box by removing the four screws from the back of the module and check that:
•all screws are tightened properly
•all cables are connected properly
•tubes are not pinched and there are no sharp bends on them
•all tubes are connected properly
•there are no loose objects inside the module
4.2Troubleshooting checklist
The following simple troubleshooting hints may help you to localize and isolate a functional
problem to the correct unit. Ensure that the monitor is turned on and all modules are
connected:
•Check if there are any messages shown in the message field. Find the possible cause and
solution from the “Messages” section later in the chapter.
•Check that the module in doubt is compatible with the monitor. Compatibility information
can be found from the Carescape Respiratory Modules User’s Manual.
•Check that there are no duplicate modules connected to the monitor. List of identical
modules can be found from the Carescape Respiratory Modules User’s Manual.
•Connect the accessories to the module in doubt. Check that the parameters measured by
the module are configured to the display (
Monitor Setup > Screen 1 Setup).
•Do a visual check to the accessories used with the module. If in doubt, replace the
accessories with known good ones.
47
Document no. M1214853C
Page 58
Datex-Ohmeda E-Modules
After troubleshooting if the problem remains, contact service. Make sure you have all
necessary information of the product at hand. Describe the problem and the troubleshooting
done so far. Provide Webmin Device Information and Service logs, if requested.
4.2.1Gas sampling system troubleshooting
Faults which can occur in the sampling system are: leaks or blockages in the tubing,
•
failure of the sampling pump or the magnetic valves, or diminishing of the flow rates
because of dirt or other matter accumulating in the internal tubing or failure of pressure
sensors.
•Whenever suspecting the sampling system and always after having done any work on
the sampling system, check the sampling system for leakages and check the flow rate.
•The D-fend Pro water trap should be replaced, when the 'REPLACE D-FEND' message
appears.
•If any liquid has entered the MiniTPX measuring unit due to water trap filter failure,
contact GE Healthcare service.
•Check that the tubing inside the module is not contaminated.
Any contamination inside the tubing may indicate that the valves or sensors are
contaminated, too. This can increase a risk of faulty operation in valves or sensors. The
valves or gas sensors are not possible to clean in the field. Therefore, if you noticed any
contamination in the module tubing, send the module to GE Healthcare for factory
service.
NOTE: All internal tubes are mechanically fragile. Sharp bends may cause leaks and occlusions.
4.2.2MiniOM Measuring unit troubleshooting
Due to the complicated and sensitive mechanical construction of the oxygen measuring
•
unit, no repairs should be attempted inside the unit. Instead, if the fault has been found in
the measuring unit itself, the entire module should be replaced and the faulty module be
sent to GE Healthcare for repair.
•In cases of no response to O
blockages, and leaks.
or strong drift, check the tubing for loose connections,
2
•Check also the OM reference gas filter assembly, and replace if needed.
•If the O
has tension.
NOTE: Never apply overpressure to the O
permanently damaged.
signal is noisy, check the measurement unit suspension and if the MiniOM tubing
2
measuring unit, as the pressure transducer may be
2
4.2.3MiniTPX Measuring unit troubleshooting
The MiniTPX measuring unit can only be repaired at the factory. In case of failure, the
•
entire module should be replaced and the faulty module be sent to GE Healthcare for
repair.
4.2.4MiniPVX Measuring unit troubleshooting
In case of failure, the MiniPVX unit can be replaced.
•
•Perform spirometry system leak test to check if there is any leakages in the internal or
external spirometry tubing.
48
Document no. M1214853C
Page 59
4.2.5CPU board troubleshooting
Due to the complexity of the large scale integrated circuitry, there are few faults in the
•
CPU digital electronics that can be located without special equipment.
•Check that all connectors and screws are properly installed.
•In case of failure, the entire module should be replaced and the faulty module be sent to
GE Healthcare for repair.
4.3Service Interface
The monitor has a service menu, which is a useful tool to examine monitor functions and
troubleshoot it in case a fault occurs.
To enter to the service menu see instructions in chapter Service Menu in the S/5 monitor’s
technical reference manual.
Failure in Agent IDAgent ID has failed.Perform gas calibration. Check
agent ID unreliability (see
functional check). If it does not
help, send the module to GE
Healthcare for factory repair.
ZeroingZeroing is in progress.Wait until zeroing is completed
successfully.
Zero errorAutozeroing during the measurement or in
the beginning of the gas calibration failed.
Calibrating errorFeeding the calibration gas was not started
within 1 minute after the automatic zeroing
was completed.
Calibration was failed due to too large gain
adjustment.
Wrong calibration gas is used.
Over rangeMeasured FiO2>103%Perform gas calibration.
Apnea deactivatedApnea alarm start-up conditions are not
reached.
Sensor INOPIR Lamp failure.
Ambient pressure is too high or low.
CPU failure.
No response from the gas module, high
temperature inside the module, or EEPROM
checksum failure.
Check the zero valve operation.
Replace the zero absorber and
Nafion tube in zero line. If it does
not help, send the module to GE
Healthcare for factory repair.
Recalibrate.
Recalibrate.
Use the specified calibration
gas.
Apnea alarm detection is
activated after the 3 breaths are
detected.
Check miniTPX flex cable
connection.
Check the ambient pressure
from the Gas Calibrations menu.
Replace CPU.
Return the module to GE
Healthcare for service.
51
Document no. M1214853C
Page 62
Datex-Ohmeda E-Modules
MessagePossible causesPossible solutions
Sensor INOPMiniOM unit failure
- temperature
- internal supply voltages
- other internal failure
MiniTPX unit failure
- temperature
- internal supply voltages
- IR lamp failure
MiniPVX unit failure
- pressure sensor failure
- internal supply voltages
CPU failure
- internal supply voltages
- pressure sensor failure
- a/d-converter system failure
Other failure
-Fan failure
- Pump failure
- Valve (Zero, Occlusion) failure
-Zeroing fails too many times
-CO
Calibrating gas sensorO
reference signal differs too much
2
from CO
, CO2 and N2O measurements are not
2
signal
2
available during the first minute after the
module is connected due to warm-up.
Anaesthesia agent measurement is not
available during the first 5 minutes after the
module is connected due to warm-up.
Check flex cable connection.
Check that fan can rotate freely.
Check fan, pump or valve wire
connection.
Wait until the warm-up is
completed.
Over ScaleIncorrect waveform scale for the parameter.
The waveform clipped because measured
gas concentration exceeds the upper limit of
the current scale.
Low gas sample flowSample flow deviates to less than 80% of the
module specific nominal flow value.
Gas sampling line, gas output, water trap, or
internal tubing is blocked.
Pump failure.
Change to the appropriate
waveform scale.
For detailed instructions refer to
the S/5 monitor user’s manual.
Check sample flow rate. Adjust,
if needed.
Check or replace the gas
sampling line, water trap, or
internal tubing.
Replace the pump unit.
Incompatible gas moduleIncompatible gas module detected by the
monitor.
Check the compatibility of the
gas module.
Gas measurements removed The module is disconnected.Reconnect the module.
Identical gas modulesThe monitor detects gas measurement from
ProblemPossible clinical causePossible technical causeWhat to do
exp TV> insp TV- spirometry tube leak
- water inside D-lite or
tubings
Loop overscale
Monitored volumes <
set volumes
Strongly vibrating
loop
Too large or too
small volumes
Fluctuating Raw-mucus in airways or
- mucus in ET tube- water or secretions in
tubings
- breathing effort
against the ventilator
- patient triggered
breathes
- wrong scale selected
- wrong TV base selected
- leak between ventilator
and D-lite
hoses or D-lite
- wrong mode vs. sensor
selection
- incorrect sensor type
selection
- ventilator exp. valve
causes fluctuations
during exp. flow
- check leakages — perform
leak test
- change tubings and D-lite
- do not use active
humidification
- change scaling
- Select correct TV base
(ATPD / BTPS / NTPD/ STPD)
- check ventilator
connections
- check the patient status
- change dry D-lite and/or
empty the water from
hoses
- check mode and sensor
- D-lite for adult
- Pedi-lite for pediatric
- check the tubings and
D-fend Pro
- check the patient status
Too high Raw
Raw value invalid
Too high Ppeak- bronchospasm
Compl value invalid- spontaneous breaths- compliance cannot be
- kink in tubing
- mucus
- asthmatic patient
- bronchospasm
- spontaneous breaths
- breathing efforts
against the ventilator
- patient triggered
breaths
- patient is coughing
- patient breaths
against the ventilator
- obstruction in airways
-HME obstructed
- check the tubing
- check the patient status
- check the patient status
- check the patient circuit
status
calculated
55
Document no. M1214853C
Page 66
Datex-Ohmeda E-Modules
5Spare parts
5.1Ordering parts
To order parts, Contact GE Healthcare. Contact information is available at
www.gehealthcare.com. Make sure you have all necessary information at hand.
NOTE: Perform the checkout procedure described in chapter “3. Service Procedures” after you
have disassembled and reassembled the module.
5.1.1Planned Maintenance Kits
The required planned maintenance parts are included in a PM kit.
Part numberDescription
M1206554
Planned Maintenance Kit for CARESCAPE E-sCxxx Respiratory modules. The
PM kit includes the required Nafion tubes, OM reference filter assembly with
O-ring and PM Sticker.
Conformity according to the Council Directive 93/42/EEC concerning Medical Devices amended by 2007/47/EEC.
CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Outside the USA, check local laws for any restriction that may apply.
All specifications subject to change without notice.
Document number M1027822-02
May 2, 2011
GE Healthcare Finland Oy
Kuortaneenkatu 2,
FI-00510 Helsinki
Finland
Tel: +358 10 39411
Fax: +358 9 1463310
Figure 2Absorbance of N2O and CO2 ................................................................................................................................................. 7
Figure 3Infrared absorbance of AAs .................................................................................................................................................... 8
Figure 7Gas sampling system layout.................................................................................................................................................14
Figure 9TPX measuring unit ...................................................................................................................................................................15
Figure 14Calibration data stored in EEPROM....................................................................................................................................18
Figure 15Pneumatic unit and reference gas connection block ..............................................................................................25
Document no. M1027822-02
iii
Page 80
Datex-Ohmeda E-Modules
iv
Document no. M1027822-02
Page 81
Introduction
This Technical Reference Manual slot provides information for the maintenance and service of
the Datex-Ohmeda S/5 Compact Airway modules. The Compact Airway modules are double
width plug-in modules. E-CO, E-COV, E-COVX, E-CAiO, E-CAiOV, E-CAiOVX and
E-CAiOVX/SERVICE/SERVICE are designed for use with the S/5 Monitors. Later in this manual
modules may be referred to without S/5 for simplicity.
The service menu is described in a separate “Service Menu“ slot and the spare part lists in the
“E-Modules Spare Parts” slot.
The Compact Airway modules provide airway and respiratory measurements.
Letters in the module name stand for:
C = CO
agents, and i = agent identification
About E-CAiOVX/SERVICE module
The E-CAiOVX/SERVICE module is meant for service purposes only. It can be used as a loan
module, if the module in the hospital should be sent to the factory for repair. The specifications
that apply to the E-CAiOVX apply also to the E-CAiOVX/SERVICE module. Module differences:
the color of the front mask is green, the front panel has a “SERVICE” text and there are no front
panel keys equipped.
Table 1Options for Compact Airway modules
and N2O, O = patient O2, V = patient spirometry, X = gas exchange, A = anesthetic
2
Compact Airway Modules
ModulesParameters / measurements
CO
E-CAiOVX
E-CAiOV
E-CAiO
E-COV
E-COVX
E-CO
E-CAiOVX/SERVICE
N2OO2Anesthetic
2
agents
Agent ID SpirometryGas
exchange
XXX XXXX
XXX XXX
XXX XX
XXXX
XXXXX
XXX
XXX XXXX
NOTE: Do not use identical modules in the same monitor simultaneously.
The E-miniC, E-CO, E-COV, E-COVX, E-CAiO, E-CAiOV, E-CAiOVX and E-CAiOVX/SERVICE and the
M-miniC, M-C, M-CO, M-COV, M-COVX, M-CAiO, M-CAiOV, M-CAiOVX and M-CAiOVX/SERVICE are
considered identical modules.
NOTE: The Compact Airway Module or Single-Width Airway Module and Airway Module,
G-XXXX, cannot be used simultaneously in the same monitor.
NOTE: The Compact Airway modules cannot be used in the Extension Frame, F-EXT4.
NOTE: Anesthetic agents and N
but when present in the module they are calculated for compensation of CO
O values are not displayed with Critical Care main software,
2
and O2.
2
Document no. M1027822-02
1
Page 82
Datex-Ohmeda E-Modules
1Specifications
1.1General specifications
Module size, W x D x H75 x 228 x 112 mm / 3.0 x 9.0 x 4.4 in
Module weight1.6 kg / 3.5 lb.
Operating temperature+10 to +40 °C
Storage temperature-25 to +70 °C
Atmospheric pressure666 to 1060 hPa /
Relative humidity10 to 95% non-condensing (in airway 0 to 100%, condensing)
Power consumption12.6 W P
Protection against electrical shockType BF
1.2Typical performance
(67 to 106 kPa)
(500 to 800 mmHg)
(666 to 1060 mbar)
, 14.6 W momentary
rms
CO
2
Measurement range0 to 15 vol% (0 to 15 kPa, 0 to 113 mmHg)
Measurement rise time< 400 ms typical
Accuracy(0.2 vol% +2% of the reading)
Gas cross effects< 0.2 vol% (O
O
2
Measurement range0 to 100 vol%
Measurement rise time< 400 ms typical
Accuracy± (1 vol% + 2% of the reading)
Gas cross effects< 1 vol%; anesthetic agents
Fi-Et differenceresolution 0.1 vol%
O
2
N2O
Measurement range0 to 100%
Measurement rise time< 400 ms typical
Accuracy± (2 vol% + 2% of the reading) (0%<N2O<85%)
Gas cross effects< 2 vol%; anesthetic agents
Respiration Rate (RR)
Measurement range4 to 60 breaths/min
Detection criteria1% variation in CO
Measuring range
Hal, Enf, Iso0 to 6 vol%
Sev0 to 8 vol%
Des0 to 20 vol%
Measurement rise time<600 ms typical (<1000 ms for Hal, typical)
Gas cross effects< 0.15 vol% N
Resolution is two digits, when the AA concentration is below 1.0 vol%.
If AA concentration is below 0.1 vol%, 0.0% is displayed.
Agent identification
Identification threshold0.15 vol% typical
Identification time< 20 s (for pure agents)
Mixture identification threshold for 2. agent:
MAC
Range0...9.9 MAC
Equation:
O
2
0.2 vol% +10% of total conc.
Compact Airway Modules
where x(AA): Hal=0.75%, Enf=1.7%, Iso=1..15%, Sev=2.05%, Des=6.0%
1.3Gas specifications
Airway humidity0...100%, condensing
Sampling rate200 ±20 ml/min. (sampling line 2-3 m, normal conditions)
Sampling delay2.5 seconds typical with a 3 m sampling line
Total system response time2.9 seconds typical with a 3 m sampling line, including
Display update ratebreath-by-breath
Warm up time2 min. for operation with CO
Gas values are measured in ATPD conditions (ambient temperature and pressure, dry). When
CO2 is displayed as a partial pressure (kPa, mmHg), the value can be alternatively shown as
wet (BTPS, body temperature and pressure saturated).
Automatic compensation for barometric pressure , CO
effect.
Auto zeroing intervalImmediately after calibrating the gas sensor and 2, 5, 10, 15, 30,
Formula 1
sampling delay and rise time
, O2, and N2O
2
5 min. for operation of anesthetic agents
30 min. for full specifications
O and CO2-O2 collision broadening
2-N2
45, 60 minutes after start-up, then every 60 minutes
1.3.1Normal conditions
Accuracy specifications apply in normal conditions (after 30 minutes warm-up period):
Ambient temperature18 to 28 °C, within ±5 °C of calibration
Ambient pressure500 to 800 mmHg, ±50 mmHg of cal.
Ambient humidity20 to 80% RH, ±20% RH of cal.
3
Document no. M1027822-02
Page 84
Datex-Ohmeda E-Modules
Non-disturbing gases
Ethanol C
Acetone< 0.3%
Methane CH
Freon 21< 1%
Freon R134A<1%
Nitrogen N
Carbon monoxide CO
Nitric Oxide NO < 200 ppm
water vapor
Maximum effect on readings
CO
2
O
, N2O < 2 vol%
2
anesthetic agents < 0.15 vol%
Effect of Heliumdecreases CO
Effect of Xenondecreases CO
OH< 0.3%
2H5
< 0.3%
4
2
1.3.2Conditions exceeding normal
< 0.2 vol%
decreases O
readings < 0.6 vol% typically
2
readings < 3 vol% typically
2
readings < 0.4 vol% typically
2
Accuracy specifications under the following conditions; nopq:
n Ambient temperature10 to 40 °C, within ±5 °C of calibration
Ambient pressure500 to 800 mmHg, ±50 mmHg of calibration
Ambient humidity10 to 98% RH, ±20% RH of calibration
o During warm-up 2 to 10 minutes (anesthetic agents 5-10 minutes), under normal conditions
p During warm-up 10 to 30 minutes, under normal conditions
q N
O > 85%, under normal conditions
2
Accuracy under different conditions (see above)
Condition n and p Condition o Condition q
CO
2
O
2
O±(3 vol% + 3% of reading)±(3 vol% + 5% of reading)± (2vol% + 8% of reading)
Measurement range0 to 40 cmH2O/ l/s
Resolution1 cmH
O/ l/s
2
Other parameters
Specifications apply in conditions listed in patient spirometry specifications.
Dead space of the sensor
9.5 ml (adult), 2.5 ml (pediatric)
Resistance of the sensor
0.5 cmH2O at 30 l/min (adult), 1.0 cmH2O at 10 l/min (pediatric)
1.4.2Conditions exceeding normal
Accuracy specifications under the following condition (during warm-up 2 to 10 minutes):
Airway Pressure (Paw)
Accuracy±2 cmH2O
Tidal volume
Accuracy±10% or 100 ml (adult), ±10% or 10 ml (pediatric)
1.5Gas exchange specifications
Mathematical integration of airway flow and gas concentration for intubated, mechanically
ventilated and/or partly spontaneously breathing patients.
NOTE: These specifications apply only, when a 2 meter gas sampling line is used, and a Y-piece
with a physical dead space less than 8 ml.
5
Document no. M1027822-02
Page 86
Datex-Ohmeda E-Modules
NOTE: These specifications only apply, if the FiO2 level delivered to the patient is varying by less
NOTE: than 0.2% at the measurement point during the inspiratory cycle.
VO2 and VCO
Measurement range 50 to 1000 ml/min
Resolution 10 ml/min
Accuracy ±10% or 10 ml/min; when FiO
2
±15% or 15 ml/min; when 65% < FiO
RQ
Measurement range 0.6...1.2
Resolution0.05
< 65%
2
< 85%
2
6
Document no. M1027822-02
Page 87
2Functional description
73;BVHQVRUBSULQFSOHYVG
7KHUPRSLOH
GHWHFWRUV
7HPSVHQVRU
6DPSOH
JDVLQ
6DPSOH
JDVRXW
6DPSOH
FKDPEHU
absorbance_N2O_CO2.vsd
2.1Measurement principle
2.1.1CO2, N2O, and agent measurement
TPX is a side stream gas analyzer, measuring real time concentrations of CO2, N2O and
anesthetic agents (Halothane, Enflurane, Isoflurane, Desflurane, and Sevoflurane).
Compact Airway Modules
Figure 1TPX sensor principle
Anesthetic agents or mixtures of two anesthetic agents are automatically identified and
concentrations of the identified agents are measured. TPX also detects mixtures of more than
two agents and issues an alarm.
TPX is a non dispersive infrared analyzer, measuring absorption of the gas sample at seven
infrared wavelengths, which are selected using optical narrow band filters.
The infrared radiation detectors are thermopiles.
Concentrations of CO
and N2O are calculated from absorption measured at 3-5 μm.
2
Figure 2Absorbance of N2O and CO
2
7
Document no. M1027822-02
Page 88
Datex-Ohmeda E-Modules
absorbance_AA.vsd
6ZLWFKHG
PDJQHWLF
ILHOG
0L[WXUH
RXW
(OHFWURPDJQHW
0LFURSKRQH
6DPSOHLQ
5HIHUHQFHLQ
2BPHDVBSULQFLSOHYVG
Identification of anesthetic agents and calculation of their concentrations is performed by
measuring absorptions at five wavelengths in the 8-9 μm band and solving the concentrations
from a set of five equations.
Figure 3Infrared absorbance of AAs
The measuring accuracy is achieved utilizing numerous software compensations. The
compensation parameters are determined individually for each TPX during the factory
calibration.
2.1.2O2 measurement
The differential oxygen measuring unit uses the paramagnetic principle in a pneumatic bridge
configuration. The signal picked up with a differential pressure transducer is generated in a
measuring cell with a strong magnetic field that is switched on and off at a frequency of 165
Hz. The output signal is a DC voltage proportional to the O
the two gases to be measured.
concentration difference between
2
Figure 4O2 measurement principle
8
Document no. M1027822-02
Page 89
2.1.3Patient spirometry
In mechanical ventilation, breaths are delivered to the patient by a ventilator with a proper tidal
volume (TV), respiration rate (RR), and inspiration / expiration ratio in time (I:E) determined by
the settings of the ventilator.
The Patient Spirometry monitors patient ventilation. The following parameters are displayed:
−Expiratory and inspiratory tidal volume (TV) in ml
−Expiratory and inspiratory minute volume (MV) in l/min
−Expiratory spontaneous minute volume in l/min
−Inspiration/expiration ratio (I:E)
Airway pressure
−Peak pressure (P
−Mean airway pressure (P
monitors
−End inspiratory pressure (P
−PEEPi, PEEPe; available only in S/5 Critical Care and Compact Critical Care monitors
−Total positive end expiratory pressure (PEEP
Compact Anesthesia monitors
−Real time airway pressure waveform (P
−Static Positive end expiratory pressures (Static PEEP
S/5 Critical Care and Compact Critical Care monitors
−Static Plateau pressure (Static Pplat); available only in S/5 Critical Care and Compact
Critical Care monitors
−Static Compliance (Static Compl); available only in S/5 Critical Care and Compact Critical
Care monitors
PEEP, P
Atmospheric pressure is used as a reference in measurement. The pressure measurement is
made from the airway part that is closest to the patient between the patient circuit and
intubation tube.
PEEP
Static pressure measurement maneuvers are automatically identified based on an increased
zero flow period at the end of the inspiration or expiration.
Static Compliance is calculated, if Static PEEP and Static P
a 2 minute period.
, P
peak
mean
=intrinsic PEEP, PEEP
i
, and P
Compact Airway Modules
)
peak
are measured by a pressure transducer on the PVX board.
plat
tot
); available only in S/5 Critical Care and Compact Critical Care
mean
)
plat
); available only in S/5 Anesthesia and
tot
)
-PEEP
aw
e
and Static PEEPe); available only in
i
measurements were made within
plat
Airway flow
−Real time flow waveform (V')
−Compliance (Compl)
−Airway resistance (Raw)
−Pressure volume loop
−Flow volume loop
The measurement is based on measuring the kinetic gas pressure and is performed using the
Pitot effect. A pressure transducer is used to measure the Pitot pressure. The pressure signal
obtained is linearized and corrected according to the density of the gas. Speed of flow is
calculated from these pressure values and the TV value is then integrated. The MV value is
calculated and averaged using TV and RR (respiratory rate) values.
Compliance and airway resistance
Compliance is calculated for each breath from the equation
Formula 2
9
Document no. M1027822-02
Page 90
Datex-Ohmeda E-Modules
Compliance describes how large a pressure difference is needed to deliver a certain amount of
gas to the patient.
The airway resistance, Raw, is calculated using an equation that describes the kinetics of the
gas flow between the lungs and the D-lite. The equation states that the pressure at the D-lite
can at any moment of the breath be approximated using the equation
where P(t), V’(t) and V(t) are the pressure, flow and volume measured at the D-lite at a time t,
Raw is the airway resistance, Compl is the compliance and PEEP
expiratory pressure (PEEP
D-lite
Patient Spirometry uses specific sensors called D-lite+/D-lite and Pedi-lite+/Pedi-lite flow
sensors. Different types of sensors are available: adult sensor for measuring adults and
pediatric sensor for children. Both are available as reusable and disposable versions.
D-lite and Pedi-lite adapters are designed to measure kinetic pressure by a two-sided Pitot
tube. Velocity is calculated from pressure difference according to Bernoulli's equation. Flow is
then determined using the calculated velocity.
Formula 3
+PEEPi is the total positive end
e
).
tot
(from Bernoulli's equation)Formula 4
where:
F = flow (l/min), v = velocity (m/s), A = cross area (m
ρ
= density (kg/m3)
Finally the volume information is obtained by integrating the flow signal.
2.1.4Gas exchange measurement
The gas exchange measurement uses the D-lite flow sensor and the gas sampler.
The basic data which is needed to obtain O
concentrations.
Concentrations have been corrected for delay and deformation during the transport of the gas
sample in a sidestream gas measurement sensor.
To obtain the amount of O
from the amount that is inhaled.
To obtain the amount of CO
from the amount that is exhaled.
These amounts can be obtained by multiplying each measured volume piece (dv) by the
corresponding gas concentration:
consumed in ml/min, the amount which is exhaled is subtracted
2
produced in ml/min, the amount which is inhaled is subtracted
2
2
), dP = pressure difference (cmH2O),
consumption and CO2 production are volumes and
2
and
10
Document no. M1027822-02
Formula 5
Page 91
Compact Airway Modules
Formula 6
Using inspiratory and expiratory minute volumes MVi and MVe and volume-weighted
inspiratory concentrations fi and fe, these equations can be rewritten as:
Formula 7
Formula 8
To obtain results which are less sensitive to errors in volume measurements, the so-called
Haldane transformation is used. This means taking advantage of the fact that the patient is not
consuming nor producing nitrogen: the amount of nitrogen inhaled is equal to the amount
exhaled fi
and VCO2 can then be written as:
VO
2
× MVi=feN2×MVe.
N2
Formula 9
with
f
Hald
= (1-fi
– fiO2 – fi
CO2
N2O
with Un=Urea Nitrogen Excretion = 13 g/day (for adults only).
2.2Main components
The compact airway modules consist of:
•Gas sampling system
•TPX measuring unit
•OM measuring unit
•PVX measuring unit
•CPU board
•OM board
•PVX board
- fi
Ane1
– fi
Ane2
) / (1-fe
– feO2 – fe
CO2
N2O
- fe
Ane1
– fe
Formula 10
)
Ane2
Formula 11
11
Document no. M1027822-02
Page 92
Datex-Ohmeda E-Modules
14
25
1
13
2.2.1Controls and connectors
Figure 5Front of Compact Airway Module, E-CAiOVX, and the back of the
module
(1) D-fend water trap
(2) Sampling line connector
(3) Water trap latch
(4) Oxygen reference gas inlet
(5) Sample gas outlet
(6) Cooling fan with dust filter
Module keysModuleDescription
Save LoopE-COV, E-COVX, E-CAiOV,
E-CAiOVX
Print LoopE-COV, E-COVX, E-CAiOV,
E-CAiOVX
Change LoopE-COV, E-COVX, E-CAiOV,
E-CAiOVX
ConnectorModuleDescription
D25 connectorall modulesModule bus connector
Save Loop saves a reference
loop.
Print Loop prints the
reference loop.
Change Loop changes a
pressure/volume loop to a
flow/volume loop or vice
versa.
12
Document no. M1027822-02
Page 93
2.2.2Gas sampling system
The sampling system takes care of drawing a gas sample into the analyzers at a fixed rate.
The gas sampling system samples the measured air to the module, and removes water and
impurities from it. A sampling line is connected to the water trap. The pump draws gas through
the sampling line to gas measuring units. After the measurements, the gas is exhausted from
the sample gas out connector.
The E-COVX and E-CAiOVX modules have a different gas sampling system compared to the
other modules. A number of flow restrictors have been changed to create a bigger pressure
difference between ambient pressure and the gas sampling system in the gas sensors. The
sample flow is, however, about the same (200 ml/min).
A larger pressure difference makes the deformations of the gas concentration curves less
sensitive to high variations of the airway pressures, thus also meeting the accuracy
requirements of gas exchange for these applications.
Compact Airway Modules
D-fend
TM
The sample is drawn through a sampling line. Then gas enters the monitor through the water
trap, where it is divided into two flows, a main flow and a side flow. The main flow goes into the
analyzers. This flow is separated from the patient side by a hydrophobic filter. The side flow
creates a slight subatmospheric pressure within the D-fend water trap, which causes fluid
removed by the hydrophobic filter to collect in the bottle.
Zero valve and absorber
Figure 6Absorber
Nafion® tubes
A Nafion tube is used between the water trap and the zero valve to balance the sample gas
humidity with that of ambient air. The tube will prevent errors caused by the effect of water
vapor on gas partial pressure, when humid gases are measured after calibration with dry
gases. Another Nafion tube is used between the absorber and the pneumatic unit to prevent
humidity caused by the absorption of CO
1)
The main flow passes through a magnetic valve before
proceeding to the analyzers. This valve is activated to establish the
zero points for the TPX and OM units. When the valve is activated,
room air is drawn through the absorber into the internal system
and the gas sensors. Paralyme is used as an absorbent.
.
2
Gas analyzers
After the zero valve and Nafion tube, the gas passes through the TPX and OM units. The oxygen
sensor has two inputs. One input accepts the main flow and the other draws in room air for
reference. Both gas flows exit from a single port.
Sample flow differential pressure transducer
The sample flow differential pressure transducer measures pressure drop across an OM inlet
restrictor and calculates sample flow from the pressure difference.
1 Nafion is a registered trademark of Perma Pure Inc.
The working pressure transducer measures absolute working pressure between the TPX unit
and OM unit. It is used for messages: ‘Sample line blocked’, ‘Check D-fend’, ‘Replace D-fend’ and
‘Check sample gas outlet’.
Pneumatic unit
The pneumatic unit contains a zeroing valve, occlusion valve and tubing connections. There is
a series of restrictors and chambers forming a pneumatic filter to prevent pressure oscillations
from the pump to reach the measuring units. The occlusion valve connection to room air
includes a dust filter and the zero valve connection to room air includes an absorber.
Connection block
The connection block contains a sample gas outlet connector and an OM unit reference gas
inlet. The inlet is equipped with a dust filter.
Occlusion valve
The valve is activated, when the sampling line gets occluded. The main flow is then diverted to
the side flow of the D-fend water trap to faster remove the occlusion.
Sampling pump and damping chamber
The gas sampling pump is a membrane pump that is run by a brushless DC-motor. Sample
flow is measured with a differential pressure transducer across a known restriction. The motor
is automatically controlled to maintain a constant flow, even when the D-fend water trap ages
and starts to get occluded. It also enables use of sample tubes with varying lengths and
diameters.
The damping chamber is used to even out the pulsating flow and silence the exhaust flow.
NOTE: In no occasion is the flow reversed towards the patient.
14
Document no. M1027822-02
Figure 7Gas sampling system layout
Page 95
Compact Airway Modules
6DPSOH
OLQH
=HUR
YDOYH
5HIHUHQFHJDV
)URQWSDQHO
3UHVVXUH
WUDQVGXFHU
S
S
73;
2FFOXVLRQ
YDOYH
3XPS
5RRPDLU
)LOWHU
$EVRUEHU
5RRPDLU
3UHVVXUH
WUDQVGXFHU
GS
20
'IHQG
)LOWHU
JDVBWXELQJBOD\RXWYVG
TPX_meas_unit.vsd
Tubing marked with 1) is thinner in E-CAiOVX and E-COVX module.
Figure 8Gas tubing layout
2.2.3TPX measuring unit
The TPX unit is a non dispersive infrared analyzer, measuring absorption of the gas sample at
seven infrared wavelengths, which are selected using optical narrow band filters. The IR lamp is
a 4 W filament, surrounded by thermal isolation. There is a hole in the isolation, passing the
radiation to a conical measuring chamber with 4 mm length.
From the sample chamber, radiation goes into seven tubular light guides with reflective inner
surfaces. At the other end of each light guide, there is a thermopile detector with an optical
filter in front of it.
The Temp sensor measures the TPX units’ temperature and it is used for temperature
compensation.
The TPX unit includes a TPX board located at the end of the unit. Its function is to connect the 7
thermopile signals and the temperature sensor signal to the CPU board.
Figure 9TPX measuring unit
Document no. M1027822-02
15
Page 96
Datex-Ohmeda E-Modules
PVX_meas_unit.vsd
OM measuring unit
The oxygen measurement is based on paramagnetic susceptibility. The gas and the reference
gas, which usually is room air, are conducted into a gap in an electromagnet with a strong
magnetic field switched on and off at a frequency of approximately 165 Hz.
An alternating differential pressure is generated between the sample and reference inputs due
to forces acting to the oxygen molecules in a magnetic field gradient.
The pressure is measured with a sensitive differential transducer, rectified with a synchronous
detector and amplified to produce a DC voltage proportional to the oxygen partial pressure
difference of the two gases.
Figure 10OM measuring unit
2.2.4PVX measuring unit
NOTE: Never apply overpressure or negative pressure of more than 300 cmH2O to the flow and
volume tubing. Differential pressure max 25 cmH
tubes.
When Patient Spirometry is used, a special sensor, D-lite, replaces the normal airway adapter
in the patient circuit. A double lumen tubing is attached to the two connectors on the adapter
and on the module front panel.
The Patient Spirometry provides patient respiration monitoring capabilities using the D-lite and
Pedi-lite flow sensors.
O on one port at a time e.g. when connecting
2
Figure 11PVX measuring unit
The measurement is based on measuring the kinetic gas pressure and is performed using the
Pitot effect. A pressure transducer is used to measure the Pitot pressure. The signal is then
linearized and corrected according to the density of the gas. Speed of the flow is calculated
from the pressure and TV is integrated from it.
16
Document no. M1027822-02
Page 97
Compact Airway Modules
08;
DQG
EXIIHU
&38ERDUG73;ERDUG
39;%RDUG
)ORZ3UHVV
20%RDUG
08;
FK
$'
ELW
&38
&17
$QHV
$JHQWV
1
2
&2
,57KHUPRSLOH6HQVRUV
76HQVRU
0RGXOH%XV
08;
DQG
EXIIHU
[
[
2
56
GULYHU
SUHDPSV
76HQVRU
6SLURPHWU\
NH\ERDUG
VLJQDOBSURFHVVLQJYVG
Patient Spirometry consists of airway connections, two pressure transducers, valves and
preamplifiers. The preamplifiers are connected to the A/D-converter on the module main CPU.
The breathing flow of a patient passing through the D-lite adapter creates a pressure
difference. This pressure difference is measured by a pressure transducer, B1. Overpressure
and negative pressure in airways are measured by another pressure transducer, B2.
Gas exchange
The gas exchange measurement uses the concentrations measured by the TPX measurement
unit and the O
measurement unit, in combination with the flow from the PVX measurement
2
unit. The gas exchange calculation is done by software.
CAUTIONThe gas exchange measurement in the E-CAiOVX and E-COVX modules works
accurately only with 2-meter gas sampling lines.
2.2.5CPU board
The CPU board contains the processor and memories and A/D-converters that are common to
the whole module. The CPU board also contains preamplifiers of TPX-sensor and drivers for
valves, fan, pump and lamp. The module is connected to the module bus through an RS-485
serial channel.
The Oxygen board contains the specific electronics for the oxygen sensor. Sample flow
measurement and sampling system pressure sensors are on this board. It also contains
EEPROMs that store calibration data of both TPX and OM sensors. The spirometry keyboard
connection is on this board.
PVX board
The Spirometry board is connected to the oxygen board. It contains pressure sensors for
airway pressure and flow measurement differential pressure and preamplifiers for those.
Calibration data of spirometry is stored on its own EEPROM.
18
Document no. M1027822-02
Page 99
Compact Airway Modules
3Service procedures
3.1General service information
The field service of the compact airway modules is limited to replacing faulty circuit boards or
mechanical parts. The circuit boards should be returned to GE Healthcare for repair.
GE Healthcare is always available for service advice. Please provide the unit serial number, full
type designation, and a detailed fault description.
WARNINGHandle the water trap and its contents as you would any body fluid.
Infectious hazard may be present.
CAUTIONOnly trained personnel with appropriate equipment should perform the tests
and repairs outlined in this section. Unauthorized service may void the
warranty of the unit.
NOTE: Wear a static control wrist strap when handling PC boards. Electrostatic discharge may
damage components on the board.
3.1.1OM measuring unit
CAUTIONDue to the complicated and sensitive mechanical construction of the O2
measuring unit, no repairs should be attempted inside the unit.
3.1.2TPX measuring unit
CAUTIONThe TPX photometer and its components are repaired/calibrated at the factory.
Attempts to repair/calibrate the unit elsewhere will adversely affect operation
of the unit. The information provided is for reference only.
3.1.3OM, TPX, and PVX measuring unit
CAUTIONThe OM, TPX, and PVX measuring units can be repaired only at the factory.
3.1.4Serviceable parts
Absorber
•
•D-fend
•Nafion tubes
•Fan filter
•Fan
•CPU board
•CPU software
•PVX Unit including PVX board
•Pump
NOTE: After any component replacement, see chapter 3.4. Adjustments and calibrations.
Document no. M1027822-02
19
Page 100
Datex-Ohmeda E-Modules
3.2Service check
These instructions include complete procedures for a service check. The service check is
mandatory after any service repair. However, the service check procedures can also be used
for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (“APPENDIX A:”) which may be used when performing the
procedures.
The symbol
results of the particular procedure.
"in the instructions means that the check form contains space to record the
3.2.1Recommended tools
NOTE: Use only calibrated and traceable measuring equipment.
−A barometer
−A mass flowmeter for measuring air flow, minimum measurement range 100-300ml/min,
accuracy 5% or better in the 100-300ml/min range.
−P/N: 755534-HEL Calibration Gas Regulator
−P/N: 755583-HEL Calibration gas, CO2, O2, N2O, DESF, package of 1 can (with E-CAiO,
E-CAiOV, E-CAiOVX modules)
−P/N: 755581-HEL QUICK CAL calibration gas, CO2, O2, N2O, package of 4 cans (with E-CO,
E-COV and E-COVX modules)
−P/N: M1006864, Calibration Gas Regulator, US only
−3 m / 10 ft anesthesia gas sampling line (with E-CO, E-COV, E-CAiO and E-CAiOV modules)
−2 m / 7 ft anesthesia gas sampling line (with E-COVX and E-CAiOVX modules)
−Spirometry tube, 3 m/10 ft (with E-CO, E-COV, E-CAiO and E-CAiOV modules)
−Spirometry tube, 2 m/7 ft (with E-COVX and E-CAiOVX modules)
−Adult D-Lite sensor
−A pressure manometer with either an integrated or a separate pressure pump
−Tubing for spirometry leak tests
NOTE: Ensure that the calibration gas and the regulator are functioning properly before
calibration. Perform annual maintenance on the regulator as required, see “Calibration gas
regulator flow check” page 33.
20
Document no. M1027822-02
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.