Datex-Ohmeda D-Tec, D-Tec Vaporizer Operation And Maintenance Manual

D-Tec Vaporizer Operation and Maintenance Manual
User Responsibility
This Product will perform in conformity with the description thereof contained in this operating manual and accompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted or contaminated should be replaced immediately. Should such repair or replacement become necessary, Datex-Ohmeda recommends that a telephonic or written request for service advice be made to the nearest Datex-Ohmeda Field Service Support Center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Datex-Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be altered without the prior written approval of Datex-Ohmeda’s Quality Assurance Department. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Datex-Ohmeda.
CAUTION w U. S. Federal and Canadian law restrict this device to sale by or on the order of
a licensed medical practitioner. Outside the U. S. A. and Canada, check local laws for any restrictions that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group code, the year of manufacture and a sequential unit number for identification.
AAA A 12345
This alpha character indicates the year of product manufacture and when the serial number was assigned; “Y” = 1995, “Z” = 1996, “A” = 1997, etc. “I” and “O” are not used.
D-Tec Vaporizer
Contents
1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.1 Operation And Maintenance Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.0 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2.2 Component Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2.1 Power Supply Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2.2 Filler Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2.3 Drain Plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2.4 Power Supply Connector And Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2.5 Equi-Potential Stud . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.6 Sump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.7 Heater Elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.8 Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.9 Agent Level Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.10 Tilt Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.0 Controls And Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.1 Dial Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.2 Auditory And Visual Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3.2 Operational . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3.3 Agent Level Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3 Status Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3.1 Warm-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3.4 Low Agent Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3.5 No Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3.6 Alarm Battery Low . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4.0 Vaporizer Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1 Fitting The Battery And The Mains Lead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.2 Vaporizer Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5.0 Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
5.1 Alarm And Display Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.2 Preoperative Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
5.3 Turning The Dial To The Required Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.4 Filling The Vaporizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5.4.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5.4.2 Filling Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
6.0 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.1.1 External Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.1.2 Internal Contamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.3 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.2 Changing The Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
D-Tec Vaporizer Contents 1 O & M Manual Part No. 1107-0622-000 August 2001
7.0 Principle Of Operation (Fig. 19) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7.1 Alarm And Indicator Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
7.2 Delivery Of Gas/Agent Vapor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
8.0 Fault Diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
9.0 Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.1 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.2 Electrical Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.3 Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.4 Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
9.5 Weight And Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
9.6 Flow Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
9.7 Effects Of Output At Varied Altitudes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
9.8 Effects Of Ambient Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
9.9 Effects Of Back Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
9.10 Effects Of Carrier Gas Composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
10.0 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
11.0 Servicing Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Illustrations
Fig. 1 Filler Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Fig. 2 Vaporizer Base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Fig. 3 Dial Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Fig. 4 Front Display Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Fig. 5 Warm-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Fig. 6 Operational . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fig. 7 Low Agent Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Fig. 8 No Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Fig. 9 Alarm Battery Low . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Fig. 10 Fitting The Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Fig. 11 Fitting The Mains Lead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Fig. 12 Mounting The Vaporizer Onto A Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Fig. 13 Locking The Vaporizer Onto A Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Fig. 14 Routing The Mains Lead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Fig. 15 Unlocking The Vaporizer From A Manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Fig. 16 Alarm And Display Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Fig. 17 Amber WARM-UP Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Fig. 18 Green OPERATIONAL Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Fig. 19 Turning The Vaporizer Dial . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Fig. 20 Inserting The Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Fig. 21 Filling The Vaporizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Fig. 22 Changing The Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Fig. 23 Vaporizer Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Page 2 D-Tec Vaporizer August 2001 O & M Manual Part No. 1107-0622-000
1.0 Introduction
1.1 Operation And Maintenance Manual
This Operation And Maintenance (O & M) Manual contains the information required in order to install, operate and maintain the D-Tec Vaporizer.
Maintenance procedures are restricted to those detailed in Section 6.0 of this manual. Requests for servicing facilities, advice or assistance must be addressed either to a local Datex-Ohmeda
Office or to an Datex-Ohmeda Authorized Distributor. Additional copies of this manual, quoting Datex-Ohmeda D-Tec Vaporizer O & M Manual Part No.
1107-0622-000, can be requested from a local Datex-Ohmeda Field Operations Unit or from an Datex-Ohmeda Authorized Distributor.
It is recommended that all relevant documentation, including the O & M Manual and accompanying labels, is immediately available to all prospective operators.
1.2 Precautions
A number of Warnings and Cautions are used throughout this manual to draw attention to the possible hazards and/or adverse conditions which may occur if the information and instructions provided are not strictly observed.
Cautions and Warnings are preceded by a w symbol and are used to draw attention to a condition which can endanger either the patient or the operator and can result in damage to the equipment.
Special attention must be paid to each Warning and Caution as it appears in the manual.
w Warning: Do not fill the vaporizer with any substance other than Suprane™ (desflurane) as specified on
the front label. If any substance other than Suprane™ (desflurane) is used, patient injury could occur.
w Warning: U.S. Federal and Canadian law restrict this device to sale by or on the order of a licensed
medical practitioner.
w Warning: Before using the D-Tec Vaporizer, the Suprane™ (desflurane) package insert must be
studied. Failure to conform to the recommendations on the insert may result in patient injury.
Important:
The vaporizer manifold O-rings required for your NAD / Dräger anesthesia machine are dependent on the place of manufacture of the anesthesia machine vaporizer manifold. Datex-Ohmeda cannot determine the specific requirements of your NAD / Dräger anesthesia machine. Therefore, per a written request from Dräger Medical, Inc., this D-Tec series vaporizer does not include replacement O-rings for the anesthesia machine vaporizer manifold. Please contact your Dräger Medical representative to obtain the correct O-rings for your system.
D-Tec Vaporizer Page 1 O & M Manual Part No. 1107-0622-000 August 2001
1.3 Symbols
Symbol Location Facility/Rating
Control Dial Standby setting Interlock Block Setting mark - dial marking alignment point for required setting Interlock Block Equi-potential stud - connection for minimizing electrical potential difference Front Panel Auditory alarm mute button Display Panel 240 ml refill mark - indicates that an additional 240 ml of agent can be added Display Panel Symbolizes nominal agent level display when 390 ml of agent in sump (max.) Display Panel Symbolizes nominal agent level display when only 60 ml of agent in sump (min.) Rear Label Mains electrical supply must be single phase sinusoidal alternating current Top Plate Label Direction of flow
w
Various Warning
2.0 Description
2.1 General
The D-Tec Vaporizer is designed to add Suprane™ (desflurane) inhalant anesthetic vapor to the medical gases supplied to a patient. The vaporizer must only be mounted on a Dräger compatible anesthesia machine which is equipped with an interlocking type manifold. The vaporizer must be supplied with dry medical gas and it must be connected to a suitable electrical mains supply. However the D-Tec vaporizer cannot be used on the Narkomed Mobile anesthesia system.
The D-Tec Vaporizer is designed to meet UL 2601-1, IEC 601-1 and IEC 601-2-13 recommendations. When the vaporizer is connected to the electrical supply, the dial is turned to the q (Stand-by) symbol and the green
OPERATIONAL light located on the front display panel is illuminated, the vaporizer is in the electrical stand-by condition and is ready for use, as described in Section 5.0 Operating Instructions.
Page 2 D-Tec Vaporizer August 2001 O & M Manual Part No. 1107-0622-000
Mechanical and electrical interlocks are incorporated into the vaporizer to help ensure that the following criteria are satisfied:
1. The vaporizer dial can only be turned from q when the vaporizer is locked onto the manifold and the green OPERATIONAL light is illuminated to indicate that the vaporizer has attained the correct operating temperature.
2. The gas flow only enters the vaporizer which is in operation.
3. Unwanted anesthetic trace vapor is minimized after a vaporizer is turned to q.
Note: It is a requirement of the ASTM standard F1850-98 and European Standard EN 740 - Anaesthetic Workstations And Their Modules, that an appropriate gas monitor is used to monitor the concentration of anaesthetic agent vapor in the inspiratory gas when the vaporizer is in operation, in order to provide protection against hazardous output in the event of a device malfunction.
w Warning: This manual and all its associated documentation must be studied before any attempt is
made to install, operate or clean any part of the D-Tec Vaporizer.
w Warning: The performance of the anesthetic machine and vaporizer may be degraded if the machine
and vaporizer are mismatched.
w Warning: Do not install a D-Tec vaporizer in the right hand position of a Narkomed 6000 as this will
result in an incorrect fit and may cause leaks in the fresh gas delivery system.
w Warning: To avoid explosive hazards, flammable anesthetic agents such as Ether and Cyclopropane
must not be used in this vaporizer. Only anesthetic agents which comply with the requirements for non­flammable anesthetic agents in the IEC 601-2-13 Standard, particular requirements for the safety of anesthesia machines, are suitable for use in the presence of this vaporizer.
w Warning: As this vaporizer is not suitable for use with flammable anesthetic agents such as Ether or
Cyclopropane, the use of antistatic breathing tubes and face masks is not necessary. The use of antistatic or electrically conductive breathing tubes when utilizing high frequency electric surgery equipment may cause burns and is therefore not recommended in any application of this vaporizer.
w Warning: The functioning of this vaporizer may be adversely effected by the operation of equipment
such as high frequency surgical apparatus or short-wave therapy equipment in the vicinity.
w Warning: Do not operate the vaporizer other than with dry medical gases. w Warning: The D-Tec Vaporizer must not be used in a Magnetic Resonance Imaging environment. w Warning: The functioning of this D-Tec vaporizer may be adversely affected by electromagnetic
interference exceeding the levels specified in EN 60601-1-2.
w Warning: This product contains a small quantity of Mercury in the Tilt Switch mechanism and should be
disposed of according to local regulations.
w Caution: If the vaporizer is not likely to be used for a period of 12 months, the battery must be
removed. Failure to do so may result in damage to the vaporizer.
w Caution: 'No Output' alarm could be caused by downstream high frequency ventilation at greater than
60 breaths/min with an airway pressure of 100 cm H2O or pulsing the O2 flush faster than 1 pulse/sec over a time period of greater than 10 secs.
w Caution: Refer to the User’s Manual for the Dräger Julian Anaesthesia Workstation for recommended
ventilator parameter limitations when using the D-Tec with a Julian.
D-Tec Vaporizer Page 3 O & M Manual Part No. 1107-0622-000 August 2001
2.2 Component Information
2.2.1 Power Supply Unit
The D-Tec Vaporizer uses a 100 to 120V a.c. 50/60 Hz electrical supply model. A label on the rear panel of the vaporizer states the voltage requirement for each model.
The integral power supply unit consists of a transformer and an a.c. to d.c. converter which provides the 12V d.c. and 5V d.c. supplies for the electrical system.
2.2.2 Filler Assembly
The filler assembly, illustrated on Fig. 1, incorporates an agent specific fitting which helps to ensure that only a Suprane (desflurane) specific bottle equipped with a "Saf-T-Fill" valve can be inserted in order to fill the vaporizer.
FILLER ASSEMBLY
AA.43.037
FILLER PORT
FILLER LABEL
Use Only
Desflurane
Fig. 1 Filler Assembly
2.2.3 Drain Plug
w Warning: Draining the vaporizer may result in rapid loss of pressure and/or agent which could lead to
injury to the operator.
A drain plug is located at the base of the vaporizer as illustrated on Fig. 2. The drain plug must not be removed except at an Datex-Ohmeda Authorized Service Center.
2.2.4 Power Supply Connector And Battery
The electrical power supply connector is located in the base of the vaporizer as illustrated on Fig. 2.
If the mains power supply fails, a Duracell 1604 or VARTA Energy 2000 battery incorporated in the base of the vaporizer provides power for the auditory and visual alarms only. The battery must be replaced annually.
POWER SUPPLY CONNECTOR
Page 4 D-Tec Vaporizer August 2001 O & M Manual Part No. 1107-0622-000
AA.43.055
BATTERY COVER
Fig. 2 Vaporizer Base
2.2.5 Equi-Potential Stud
The equi-potential stud Y is located on the top of the vaporizer. It provides a means of connecting a vaporizer to the anesthesia system to minimize any electrical potential difference between the vaporizer casing and the anesthesia system.
2.2.6 Sump
The sump provides a means of containing the agent and includes the agent filling port, the drain, the heaters and the agent level sensor.
The sump has a total nominal capacity of 390 milliliters (ml) which consists of a nominal 330 ml indicated volume and a nominal 60 ml reserve.
2.2.7 Heater Elements
Two 100W heater elements, located in the base of the sump, heat the agent to a nominal 39°C (102°F) to generate the working pressure.
Two 100W heater elements located in the upper part of the vaporizer help to prevent condensation of the anesthetic agent in the vaporizer.
The current supplied to power the heaters alternates between the two heaters in the sump and the two heaters in the upper part of the vaporizer. This minimizes the current requirement.
The casing of the vaporizer is normally warm to the touch while it is connected to the electrical supply.
2.2.8 Monitors
Electro/mechanical monitors are incorporated to monitor the fresh gas/agent vapor pressure balance and the agent volume.
The monitors control the operational status display and activate the appropriate alarm indication if a monitored parameter malfunction is detected.
2.2.9 Agent Level Sensor
An agent level sensing probe is mounted in the sump to measure the agent level. When the sump contains between 60 and 390 ml of agent, the LCD level display indicates the level of agent in the sump. If the agent level decreases below 60 ml, the amber LOW AGENT light on the display panel flashes and the auditory alarm is activated. If the agent level decreases below approximately 20 ml, the red NO OUTPUT light flashes and the auditory alarm is activated. Vapor output then ceases.
2.2.10 Tilt Switch
If the vaporizer is tilted for 10-15 seconds while it is in operational mode and delivering vapor, vapor delivery will be stopped, as indicated by the flashing No Output light and auditory alarm.
D-Tec Vaporizer Page 5 O & M Manual Part No. 1107-0622-000 August 2001
3.0 Controls And Indicators
3.1 Dial Assembly
The manually operated dial is used to select the level of agent concentration from 1% to 18%. The dial is marked at intervals of 1% from 1% to 10% and at intervals of 2% from 10% to 18%. Selections can be made at any setting between 1% and 18%.
The dial release incorporated in the dial assembly must be operated in order to turn the dial from the q (Standby) setting and it must be operated again to turn the dial from 12% to a higher % setting. All other settings can be made without operating the dial release.
A solenoid interlock mechanism is also incorporated to help to ensure that the dial can be operated only when the green OPERATIONAL light located on the display panel is illuminated.
3.2 Auditory And Visual Indicators
DIAL RELEASE
EQUI-POTENTIAL STUD
DIAL
AB.54.001
Fig. 3 Dial Assembly
Visual indicators and an auditory alarm are used to provide the vaporizer status display and also to indicate an alarm condition, as described in Section 3.3.
Visual indicators are located on the front display panel illustrated in Fig. 4 and the
AUDITORY ALARM MUTE BUTTON
auditory alarm is mounted behind the upper part of the front display panel.
An auditory alarm mute button located above the display panel can be used to mute the alarm temporarily under certain conditions as described in Sections 3.3,
3.3.4 and 3.3.5. The LCD agent level display indicates the
level of agent contained in the vaporizer sump, as described in Section 3.3.3. The range of indication is between a minimum of 60 ml and a maximum of 390 ml. When the indication is below the 240 ml refill mark , illustrated in Fig. 4, it indicates that an additional 240 ml of agent can be added.
GREEN
RED
AMBER
AMBER
AMBER
DESFLURANE
OPERATIONAL
NO OUTPUT
LOW AGENT
WARM-UP
ALARM BATTERY LOW
390 ml NOMINAL
LCD AGENT LEVEL DISPLAY
240 ml REFILL MARK
AA.43.038
Page 6 D-Tec Vaporizer August 2001 O & M Manual Part No. 1107-0622-000
60 ml NOMINAL
Fig. 4 Front Display Panel
Loading...
+ 23 hidden pages